Abstract
Factors affecting the performance of 1H heteronuclear decoupling sequences for magic‐angle spinning (MAS) NMR spectroscopy of organic solids are explored, as observed by time constants for the decay of nuclear magnetisation under a spin‐echo ( ). By using a common protocol over a wide range of experimental conditions, including very high magnetic fields and very high radio‐frequency (RF) nutation rates, decoupling performance is observed to degrade consistently with increasing magnetic field. Inhomogeneity of the RF field is found to have a significant impact on values, with differences of about 20 % observed between probes with different coil geometries. Increasing RF nutation rates dramatically improve robustness with respect to RF offset, but the performance of phase‐modulated sequences degrades at the very high nutation rates achievable in microcoils as a result of RF transients. The insights gained provide better understanding of the factors limiting decoupling performance under different conditions, and the high values of observed (which generally exceed previous literature values) provide reference points for experiments involving spin magnetisation refocussing, such as 2D correlation spectra and measuring small spin couplings.
Keywords: decoupling, magic-angle spinning, nutation, organic solids, solid-state NMR spectroscopy
1. Introduction
Effective decoupling of 1H nuclear spins is essential for achieving high‐resolution 13C and 15 n solid‐state NMR spectra from typical organic molecules, and is particularly important for correlation experiments that use J (or scalar) couplings to determine molecular connectivity. Such experiments are central to the use of NMR spectroscopy for establishing molecular structure and dynamics. Moreover, measurement of small J couplings, such as those across N−H N hydrogen bonds, provides direct information on molecular assembly.1 As J couplings are small, relatively long periods (tens of ms) of evolution are required to build up the required spin coherences. Imperfect decoupling of the 1H spins leads to significant magnetisation losses during these periods, directly affecting the viability of experiments. For example, the refocussed INADEQUATE experiment2 used to assign the 13C spectra of testosterone solid forms3 required three days, whereas a recent experiment to characterise the organic components of a solid electrolyte interphase4 required a 14 day experimental run, despite 13C labelling. Isotopic enrichment was also used when probing biopolymers in secondary plant cell walls5 and when establishing supramolecular assembly in oxidative polymerisation of aniline6 and in rosette nanotubes (using analogous 15 n experiments),7 whereas specialist dynamic nuclear polarisation techniques have been recently used to obtain correlation spectra of natural abundance samples.8, 9
The viability of experiments exploiting J couplings in organic molecules is directly related to the rate at which 13C magnetisation decays as a result of imperfect decoupling. Although considerable progress has been made in developing approaches to decoupling and understanding how they work,10, 11, 12 there is no comprehensive theory that allows decoupling performance to be quantitatively predicted. Indeed, our earlier work13 has shown that quantitative reproduction of experimental data through simulation is intrinsically difficult owing to the rapid population of high‐order coherences. Existing experimental studies focus on the important goals of improving spectral linewidths, often through new sequences, including studies of how to choose between sequences.14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25, 26, 27 The varied conditions used in these studies, however, make it difficult to establish an overview of the factors determining decoupling performance. Here, we deliberately focus on well‐characterised decoupling sequences under a wide variety of experimental conditions to make more direct and quantitative links between theory and practice. We also concentrate on the regime where the radio‐frequency (RF) nutation rate exceeds the spinning rate. The “low power” regime, where the magic‐angle spinning drives the decoupling, is important for systems that are sensitive to RF heating, such as biological systems. The very different mode of operation,10 however, means it is difficult to compare the regimes, and so we focus on the “high power” regime, which is more typical for chemical applications involving organic solids.
As in our earlier study,13 the decay time constant under spin‐echo, , is chosen as the primary experimental metric because it is unaffected by inhomogeneous contributions to the spectral linewidth, such as shimming, anisotropic bulk magnetic susceptibility or sample inhomogeneity.28, 29 Hence continues increasing as the decoupling efficiency increases even though the spectral resolution has plateaued.30 It has been noted previously when comparing different decoupling sequences,21, 27 or different parameters under the same sequence,31 that the optimal values tend to vary significantly even though the differences in spectral linewidths at the same conditions are small. So while optimising spectral resolution is relevant for many applications, provides more insight when trying to understand the factors determining decoupling performance, and is directly relevant to the challenging experiments discussed above. The values observed here suggest that J couplings as small as a few Hz are measurable and that J‐based 13C correlation experiments should be viable for most systems without the need for isotopic enrichment.
Techniques and Methods
In common with several previous studies, the methylene group of glycine was used as a model system. The strong dipolar coupling network, both homonuclear and heteronuclear, in methylene groups makes them the hardest type of 13C to decouple, ensuring that decoupling that is effective on methylene sites will also be generally effective. Sample data sets acquired by using the methine group of l‐alanine showed similar trends (see Table S1 in the Supporting Information for data sets available), and the conclusions are thus expected to be generally applicable to organic solids.
Experimental Methods
Experimental measurements of were performed on polycrystalline samples of glycine‐2–13C,15 n (99 % 13C, 98 % 15 n) and l‐alanine‐2–13C,15 n (99 % 13C, 98 % 15 n) purchased from CortecNet. The glycine sample was confirmed to be ‐glycine based on the 13C carbonyl peak at 176.5 ppm, which is sensitive to polymorphic changes.32, 33 As expected from the stability range of this form, 5–500 K,34 no transformations were observed during experiments.
Table 1 shows the combinations of hardware used. 13C magnetisation was created by using cross‐polarisation (CP), ramped on the 1H nutation frequency35, 36 through the centreband matching condition for , and through the zero‐quantum sideband ( ) for . At , CP matching was done on the double‐quantum sideband ( ), as this required much lower RF powers compared with the zero‐quantum sideband sidebands commonly employed at slower MAS, but gave comparable signal enhancement.37, 38 Cross‐polarisation is expected to excite a smaller sample region than direct excitation of the 13C magnetisation owing to the effects of RF inhomogeneity.39 Tests on a 1.3 mm probe at 25 kHz MAS (hardware configuration 2) showed a larger initial drop in decays when using direct excitation compared with CP, presumably associated with a poorly decoupled sample towards the coil ends, but the long‐term decay and overall fitted values were not significantly different. Although low‐power CP at high MAS rates has been reported to selectively excite the 13C spectrum,40 such effects were avoided by putting the 13C transmitter on the methylene resonance. Relaxation delays were always 4 s, whereas the CP contact times were optimised for each set of experiments, varying in the range 0.8–2.7 ms, as noted in the figure captions. The magnetisation was then measured after a spin‐echo period, – – , during which either CW, two‐pulse phase‐modulated (TPPM),41 XiX42, 43 or SPINAL‐6444 proton decoupling was applied as shown in Figure 1. As originally defined, the different phase angles in SPINAL‐64 were fixed (10°, 15° and 20°), but have subsequently been optimised, either in the fixed ratio 1:1.5:2 with a single optimisation parameter , or additionally optimising the angles and .21, 45 Here, a single phase optimisation was used.
Table 1.
Configuration | [MHz] | Probe | Console |
---|---|---|---|
1 | 300 | Bruker 2.5 mm | Varian InfinityPlus |
2 | 500 | Bruker 1.3 mm | Varian InfinityPlus |
3 | 600 | Bruker 2.5 mm | Bruker Avance II+ |
4 | 850 | Bruker 2.5 mm | Bruker Avance III |
5 | 850 | Bruker 1.3 mm | Bruker Avance III |
6 | 500 | Agilent T3 3.2 mm | Agilent DD2 |
7 | 500 | Agilent BioMAS 3.2 mm | Agilent DD2 |
8 | 400 | 450 μm (i.d.) piggy‐back μMAS30, 47 | Agilent DD2 |
Generally, the same 1H decoupling was used in both spin‐echo and acquisition periods. The only exception was early measurements using the microcoil probe, hardware configuration 8, for which optimised TPPM decoupling at was used during acquisition. As discussed in more detail in Ref. 13 using a fixed decoupling sequence for acquisition gives consistent line‐shapes in the acquired spectra, but significant mismatches between spin‐echo and acquisition decoupling distort fitted values through the orientation dependence of decoupling efficiency.46 The 1H transmitter frequency was positioned on the maximum of the unresolved 1H spectrum at slow MAS. Inversion of 13C magnetisation used a ‐pulse with duration between 5 and 6 .
The 1H decoupling nutation rate, , was measured by using the same sequence with a zero spin‐echo period. The initial 1H pulse width was incrementally increased in at least 100 steps of approximately to acquire a 1H nutation spectrum and the peak position was taken as the nominal . TPPM and SPINAL‐64 pulse widths are expressed as a flip angle, , whereas XiX pulse widths are expressed as a fraction of the rotor period, .
Combinations of MAS and decoupling RF nutation rate were chosen to avoid rotary resonance recoupling conditions at , where is an integer. Although the PISSARRO‐548 heteronuclear decoupling sequence was tested, it was found to give very similar parameter maps to XiX (see the Supporting Information), and so we have focussed on the simpler and more readily analysed sequence. This result is not surprising given that PISSARRO was designed to improve performance close to rotary resonance conditions.
Full decay curves were obtained at selected decoupling conditions by incrementing the evolution time, , linearly in 30–40 steps from zero to approximately twice the maximum expected . The free induction decays were zero‐filled and Fourier transformed (without apodisation) using matNMR.49 The decay of the methylene 13C peak height as a function of was fitted to a decaying exponential to obtain using MATLAB.50 Where detailed parameter maps as a function of the decoupling sequence parameters were acquired, values were inferred from a pair of experiments at and by assuming a mono‐exponential decay of the peak height between these points. As previously discussed in Ref. 13 and reproduced here in the Supporting Information, discrepancies between the values obtained by this quick, but approximate, approach, and those obtained from full decays were corrected by scaling the approximate values to coincide with the more accurate values obtained from full decays. The experiments also provided data for estimating “ ”, the effective time constant describing the linewidth, calculated from , where FWHM is the full width at half‐maximum.
Experiments to assess the dependence of the decoupling performance on magnetic field were performed at 1H Larmor frequencies of , 600 and 850 MHz, at 12 kHz MAS rate and 105 kHz 1H nutation rate, using the same 2.5 mm o.d. rotor and similar probe designs. Additional data was collected at under the same MAS and decoupling nutation rates using hardware configuration 2. For TPPM and SPINAL‐64, full parameter maps were first acquired as a function of both pulse width and phase to locate the positions of optima, then a detailed parameter cross‐section at a fixed pulse phase was recorded through the region of peak decoupling, which was for both sequences under these conditions. The XiX performance was first characterised over a wide range of pulse lengths, and then in more detail over the region of peak decoupling to ensure the narrow optima were well defined.
Decoupling experiments using very high RF decoupling fields, beyond the reach of commercially available probes, were performed using a piggyback μMAS design equipped with a 450 μm inner diameter coil, as described in Ref. 30, 47.
Experiments to assess the influence of RF field inhomogeneities on decoupling were performed using two 3.2 mm MAS probes with different coil geometries: an Agilent T3 MAS probe with standard solenoid coil geometry, and an Agilent BioMAS probe, whose scroll coil geometry exhibits significantly better homogeneity. The same sample rotor, MAS rate, and spectrometer equipment were used in both sets of experiments, and care was taken to adjust the power levels for each probe such that the peak nutation frequencies were the same.
Simulations
Numerical simulations of the decay of 13C magnetisation in the presence of 1H decoupling were performed using pNMRsim,51 as described in Ref. 13 and reproduced here in the Supporting Information. Spin systems are labelled as CHn, with indicating the number of protons coupled to the central carbon atom. The decays were fitted to a single‐exponential function to derive computed dephasing time constants, , describing the loss of the 13C single‐quantum coherence. Note that spin‐echoes are not included in the simulation as there are no inhomogeneous components of the decay to refocus, and computations of are much more sensitive to finite‐sized spin systems.13 The simulations incorporate effects of RF inhomogeneity through nutation spectra acquired using the same equipment and experimental conditions. As described in more detail in the Supporting Information, simulations were performed for a set of 15–20 RF nutation rates chosen to correspond to equal areas of the nutation profile, and the results summed.
2. Results
Determining the optimal decoupling parameters across a range of experimental set‐ups involved acquiring a large number of detailed parameter maps for both and . As these were acquired under a uniform set of conditions, these data should be a useful resource for further study. A summary of the data sets available is given in the Supporting Information.
An initial comparison of the data highlights that ease of optimisation varies significantly between sequences and experimental conditions. These aspects of optimisation have previously been discussed for peak height24 and .21 Optimisation becomes more difficult as the ratio of the RF nutation rate to MAS frequency reduces, for example, going from to 62.5 kHz under kHz decoupling. As illustrated by Figures S9–S11 in the Supporting Information, the parameter maps became fragmented by multiple destructive resonance conditions, requiring finer parameter grid increments. For TPPM and SPINAL‐64, the optimum pulse phase moved away from the commonly prescribed and ,45 respectively, necessitating the optimisation of both pulse width and phase. SPINAL‐64 optima are especially narrow, likely owing to the increased number of resonance conditions due to the longer cycle time of the sequence. The XiX parameter map underwent relatively few changes, making it easier to optimise at high MAS frequencies compared with TPPM or SPINAL‐64.
As has been previously observed, the local minima and maxima across a parameter map qualitatively correspond with those of the map.21 The optima were, however, generally narrower, and optima with similar values tended to have different relative values. In other words, a map cannot be relied upon to provide the best sequence parameters for .
2.1. Decoupling Transmitter Offset
The detrimental effects of off‐resonance irradiation on decoupling efficiency are well‐established for spectral linewidth,11, 19, 22, 25, 52 but to a lesser degree for .53 It was found that the optimal pulse width (and phase) did not change significantly over a range of 1H offsets about the optimum, thus the offset dependence could be determined independently using the same optimal sequence parameters. Figure 2 shows that the characteristic ‘width’ of the offset dependencies is similar between and for TPPM across a range of experimental parameters. CW and SPINAL‐64 (see Figure S5 in the Supporting Information) exhibit the same behaviour. With the exception of XiX decoupling, an increase in the field is accompanied by a consistent increase in linewidth for both and , that is, a vertical offset. XiX has a more complex offset dependence, being more broadband at larger .
The dependence of 13C linewidths on transmitter offset under CW decoupling was shown to be described by a parabola, and was justified theoretically by VanderHart et al.,54 following Mehring.55 It is not clear, however, that more complex sequences share this dependence, and indeed some sequences have been specifically optimised to be robust with respect to offset variations.56 Figure 2 shows that, although the dependence of spectral linewidth on offset for TPPM decoupling is close to parabolic at modest RF nutation rates, it is clearly not for under the same conditions. The parameter shown in Figure 2 measures the steepness of the parabolic curve fitted to the offset dependence of the spectral linewidth. From the expressions given by VanderHart, this parameter might be expected to be a fixed function of the NMR parameters, independent of RF nutation rate, but this is clearly not the case, even for CW decoupling. In particular, the robustness with respect to the offset is significantly improved at increased nutation rates, as measured by the decrease in the parameter. The curves do not fit well to simple parabolas, but show qualitatively similar trends. Although these observations are interesting and worth further investigation, they are not directly relevant to the complex interplay of experimental conditions and spin system dynamics at the heart of the decoupling problem. The following results assume that the 1H transmitter offset is close to the optimal conditions.
2.2. Field Dependence of
Parameter maps of decoupling performance at moderate MAS and 1H nutation rates (12 kHz and 105 kHz respectively) were acquired at several magnetic fields. Figure 3 shows characteristic sections of these maps for the four fields studied. The decoupling optima marked for each decoupling sequence in Figure 3 are collated in Figure 4 as a function of Larmor period, . The trend in Figure 3 a and Figure 4 of shortening values with increasing for CW and TPPM decoupling is consistent with their decoupling performance being dominated by second‐order cross‐terms between the heteronuclear dipolar couplings and 1H chemical shift anisotropy (CSA) tensor, which increase proportionately with .57, 58, 59 The relative complexity of SPINAL‐64 has hindered its theoretical analysis, but its mode of operation is assumed to be essentially the same as TPPM and so might be expected to have a similar dependence. Although the detrimental effects of increasing upon can be inferred from previously published data21 on glycine, and poorer values of at 600 MHz compared with 300 MHz have been noted for XiX and RS‐HEPT decoupling,53 the field dependence has not been explicitly explored. Figures S6 and S7 in the Supporting Information collate results from Ref. 21 together with ours. Note that the apparent extrapolation of the data points for SPINAL‐64 in Figure 4 towards negative values in the limit of infinite magnetic field are likely to be an artefact of increasing off‐resonance effects at higher field; this behaviour is not borne out in the wider collated data of Figure S6. Moreover, unlike TPPM, the SPINAL‐64 parameter map, Figure 3 b, changes shape with , and so it may be more difficult to observe consistent trends for more complex sequences. This may explain, for example, why SPINAL‐64 appeared to perform better at 700 MHz compared with 500 MHz in a previous study,21 whereas all the other results show the opposite trend.
The linewidth under XiX decoupling, and related sequences such as PISSARRO‐5, is dominated by proximity to resonance conditions as well as second‐order cross‐terms between the heteronuclear and homonuclear dipolar couplings,60 and so is not expected to show a strong magnetic field dependence. This is largely confirmed in Figure 3 c, where significant portions of the parameter space have very similar values for all four fields, although the peak performance at high field is measurably poorer. The position of the XiX optimum is very sensitive to under this combination of and , in the same way as for low‐power XiX decoupling.61 This is evident in Figure 3 c for the decoupling optima at , corresponding to ; owing to their proximity to destructive resonances, the XiX peak values are much more sensitive to mis‐adjustments than TPPM or SPINAL‐64. Therefore, the peak XiX values may not be robust, especially at , where the RF inhomogeneity is noticeable poorer (Figure S4 in the Supporting Information). The consensus of these and previous results is that achievable XiX performance does decrease as the magnetic field increases.
Figure 5 collates the optimal values at two magnetic fields (corresponding to and 850 MHz) and two MAS frequencies ( and 62.5 kHz) for which complete data sets were obtained for all the decoupling sequences used. It can be seen that the trends of decreasing with at kHz (solid lines) are similar to those at kHz in Figure 4—that is, XiX is not as dependent on as the other sequences. However, under fast MAS (dashed lines), when the two frequencies, and , become more comparable, the direct impact of on is less clear. This can be attributed to the increased significance of resonance conditions on the parameter space, complicating interpretations of at optima based on a single dominant mechanism by making the optima narrow and very sensitive to small changes in and (Figures S9–S11 in the Supporting Information). Unlike the other sequences considered, the XiX either improves or stays unchanged with increasing MAS rate.
These results complement previous studies of decoupling performance as a function of for slower spinning frequencies,27 where values for the TPPM and SPINAL‐64 decoupling sequences were observed to increase up to 20 kHz MAS before falling off, whereas the performance of XiX and CW systematically increased and decreased, respectively, with fast spinning (Figure S8 in the Supporting Information). Although the qualitative picture that emerges is consistent, it is worth noting the actual values of vary markedly between studies (with the exception of simple CW decoupling). The dependence of values on the optimisation protocol, and potentially, hardware details, illustrates the difficulty of drawing conclusions based on individual studies.
2.3. Field Inhomogeneity
The effects of field inhomogeneities are potentially significant for phase‐modulated decoupling sequences, such as TPPM, where the optimum pulse length is strongly dependent on the 1H nutation frequency.41, 62 Most of the literature focus has been on homonuclear decoupling, where different fields across a sample produce a distribution of scaling factors, dramatically degrading resolution,63 although some heteronuclear decoupling sequences have been expressly designed to be more robust with respect to differences in , such as SDROOPY56 and SWf‐TPPM.52, 64 The influence of inhomo‐ geneity on values has, however, not been explicitly investigated.
Figure 6 shows 1H nutation spectra measured through 13C for two 3.2 mm MAS probes with different coil geometries. The measured nutation spectrum of the T3 probe (conventional solenoid coil) clearly shows a peak at and a long tail of much‐reduced RF, which is known from inhomogeneity imaging experiments (on a Bruker 4 mm probe) to originate from the sample at the ends of the rotor.65 The nutation spectrum of the BioMAS probe (scroll coil geometry) is much narrower, indicating a more homogeneous field across the sample. Owing to the additional effects of CP selectivity, determined by the quality of RF matching between the two channels,39 these nutation spectra are narrower than ones acquired directly on the proton signal (especially for the T3 probe), but are a better representation of RF experienced by the sample visible in 13C CP/MAS spectra. The nutation spectra show additional peaks appearing at frequencies and , where is an integer and . These features were observed with other combinations of probes and experimental conditions used and are likely to result from a time‐dependence of the magnitude and direction of the effective field owing to anisotropic interactions, which are modulated by the MAS frequency.66
The TPPM parameter space was characterised at 12 kHz MAS and in terms of both pulse duration and phase, and a detailed parameter map cross‐section then recorded at the optimum phase (Figure 7 a). The results clearly show a broadening effect of inhomogeneity on the shape of the parameter map as the sample experiences a wider distribution of pulse lengths. There is a corresponding effect on , as seen in Figure 7 b; inhomogeneity leads to a distribution of values across the sample and a significant multi‐exponential character in the decays.
The dashed lines in Figure 7 a show fitted time constants, , of simulations of the decay of 13C magnetisation in a CH7 spin system with and without the effects of inhomogeneity, as described in the Simulations section above and in more detail in Section 4 of the Supporting Information. The RF inhomogeneity profiles were obtained from the nutation spectra in Figure 6. Although the calculated values and experimental values are not directly comparable, the simulations show the same qualitative trends and confirm that inhomogeneity has a significant impact on decoupling performance.
Sequences such as XiX, the timings of which are expressed relative to the MAS period rather than in terms of a nutation angle, might be expected to be relatively robust with respect to homogeneities. simulations using a CH8 spin‐system at , shown in Figure 8 demonstrate, however, that XiX global optima are also quite sensitive to homogeneities, with a reduction of by approximately 50 % for typical probe homogeneity profiles. Also, the strong dependence of XiX optima on calibration under some combinations of and , as evident from the experimental results in Figure 3 c at low and other investigations at high ,61 imply that homogeneity is expected to reduce peak decoupling performance in those cases too.
2.4. Pulse Transients
The deleterious effects of transient variations in the RF experienced by the sample, associated with changes of the driving amplitude or phase, have been appreciated since the early days of pulse NMR.67 Such effects are known to be significant for homonuclear decoupling sequences; Ref. 68 for example, analyses how pulse imperfections influence the effective resolution of windowed PMLG decoupling. Other experiments have been shown to be relatively robust with respect to RF transients,69 but, to the best of our knowledge, there is no published work on the effects of RF transients on heteronuclear decoupling.
RF transients can be measured experimentally by pick‐up coils placed close to the NMR sample. At least in the case of the dipolar recoupling experiment considered by Carravetta et al.,69 these were shown to fit well both to exact electronic simulations of model tuned coil circuits and to simple mathematical models of the time‐dependence of the RF phase and amplitude. The model described by Equations (10)–(14) of Ref. 69 can be usefully simplified for phase‐modulated sequences (i.e. constant driving RF amplitude). The resulting (in‐phase) and (quadrature) components of the magnetic field, , following a phase change at are [Eq. (1)]:
(1) |
where is the amplitude of the driving RF, and is the phase change relative to the initial phase. is the rate constant for the transient response of the tuned circuit, which is largely determined by the of the probe, and parameterises the amplitude of the quadrature component of the transient response. Although is essentially fixed by the probe, is largely determined by the mismatch between the frequency of the driving RF and the resonant frequency of the tuned circuit. and are readily converted to an overall RF amplitude and an instantaneous phase as a function of time. Illustrative examples of RF transient profiles can be found in recent literature reports.69, 70, 71, 72 Although active compensation of amplitude and phase transients has been implemented for recoupling experiments with promising effects on reproducibility and stability,72 it is not clear whether heteronuclear spin decoupling could be similarly improved. Pulse transients will change the effective field and the Fourier coefficients characterising a decoupling sequence, thereby affecting the residual coupling terms and resonance conditions. It is, therefore, difficult to determine a priori whether compensation of the transient response would improve the performance of a given sequence. Simulations are employed here in lieu of a detailed theoretical description.
The transient response is modelled in the simulations by dividing the evolution into short time steps, , typically of . Calculating propagators for each making up an individual pulse would be extremely time‐consuming, and so the response is only modelled over the first 5 to 6 time constants (i.e. ), and the set phase and amplitude used for the remaining pulse duration. The convergence of the free induction decay (FID) with respect to both parameters was checked on a case‐by‐case basis. The pulses were always sufficiently long compared with so that overlap of the transient responses did not occur. Using a low‐ pick‐up coil, values of and were obtained by fitting the experimental transient response of a 2.5 mm 400 MHz DR Bruker probe to a pulse pair.
The effects of transients are expected to be larger at the high nutation rates routinely available when using a microcoil MAS probe.73 This was investigated for TPPM and SPINAL‐64 decoupling at and a range of nutation frequencies up to kHz. Under conditions of , the TPPM pulse width and phase were easily optimised, as shown by the relative sparsity of the resonance conditions across the parameter maps in Figure 9 compared with those at lower nutation frequencies (Figure S9 in the Supporting Information). In both experiment and simulation, optimal decoupling was found to lie along the line , close to the predicted by Floquet analysis.52, 59
Figure 10 shows optimal experimental values as a function of nutation frequency for TPPM, SPINAL‐64 and CW. Above v 1≃300 kHz, the peak values for the phase‐modulated sequences decrease sharply towards those of simple CW decoupling. Other experimental measurements using this probe have shown a similar drop in peak for both TPPM and SPINAL‐64 above v 1≃250 kHz.30
Simulations were performed with and without transients over a range of RF nutation frequencies. A value of is expected for the microcoil probe based on its 47, 70 (45 at 400 MHz for 1H). The value of depends on the exact probe tuning and, in lieu of precise measurements, the value measured on the Bruker 2.5 mm probe was assumed to be representative. To find the optimum at a given nutation frequency, care was taken to ensure that the phase and pulse width simulation step sizes were small enough to show smooth evolution of dephasing times. Figure 11 demonstrates that the observed decrease in at kHz is due to the combined effects of in‐phase and quadrature transients. The in‐phase transients have little impact on their own, at least for TPPM with . This agrees with our understanding that only a small reduction in dephasing time, owing to a minor reduction in the average RF amplitude, will be observed if the effective nutation axis of the sequence remains in the plane. This is true in the presence of solely amplitude transients. If the nutation axis tilts out of the plane as a result of the presence of both amplitude and quadrature transients, however, then a larger impact on dephasing times is expected, especially at high RF where decoupling optima are narrow.
3. Conclusions
Although heteronuclear decoupling has been widely investigated, the diversity of previous studies has made it difficult to make quantitative comparisons. The difficulty of optimising multi‐parameter decoupling sequences, particularly in regions where the decoupling optima are narrow, means that the results presented in the prior literature are often inconsistent. Measuring decoupling performance by using a well‐defined protocol and a wide variety of experimental conditions has provided large data sets that can be mined to address different questions. Here, we focus on understanding what are the limiting factors determining decoupling performance under different experimental conditions. Table 2 compiles the peak values observed here with previous literature results.21, 27
Table 2.
[ms][a] | [MHz] | MAS rate [kHz] | [kHz] | Sequence | Reference |
---|---|---|---|---|---|
61.1(5) | 500 | 10 | 115 | SWf‐TPPM‐sc | [21] |
133(3) | 500 | 25 | 170 | TPPM | this work |
82(1) | 500 | 62.5 | 170 | TPPM | this work |
38(1) | 850 | 12 | 105 | TPPM | this work |
101(4) | 850 | 25 | 170 | TPPM | this work |
83.5(5) | 850 | 62.5 | 170 | TPPM | this work |
[a] Figures in parentheses are one standard deviation uncertainties from fitting.
The decrease in decoupling performance with increasing magnetic field was expected for decoupling sequences, such as CW and TPPM, which are primarily limited by cross‐terms between the heteronuclear dipolar coupling and the 1H chemical shift anisotropy, but is less expected for sequences such as XiX decoupling (which is primarily limited by purely dipolar terms). This is likely to reflect higher‐order terms involving the 1H CSA and also offsets in 1H NMR frequencies from the decoupling frequency. In practice, the greater magnetisation losses at higher field will be largely offset by the intrinsically greater sensitivity and resolution of spectra obtained at higher field. Note that large 1H CSAs, such as those often observed in amide groups,74 are expected to have an analogous effect on decoupling performance to increasing the magnetic field.
As has previously been observed with 13C linewidths, decoupling performance measured by values is systematically worse at MAS rates above 60 kHz compared with 25 kHz. This reflects the increased number of “resonance” conditions in the area of parameter space where decoupling is typically optimal when the MAS rate is of the order of the 1H nutation frequency. “Low power” decoupling offers distinct advantages in these regimes; indeed, an impressive of 200 ms has been observed on a similar test sample (glycine ethyl ester) using a modified form of XiX decoupling at 90 kHz MAS.75 Resonance conditions are less significant in the low spinning speed regime, and values increase with spinning rates below approximately 22 kHz MAS.21, 27
RF transients associated with phase switches are not found to have a significant impact at typical 1H nutation rates. At nutation rates above , however, simulations and experiments show that phase transients have an increasing impact, particularly from their quadrature components. At nutation rates of 500 kHz or more, achievable in microcoils, the performance of phase‐modulated sequences decreases dramatically towards that of simple CW decoupling. These problems can be addressed by careful tune‐up to minimise quadrature transients and/or development of sequences that are robust with respect to transients.
Inhomogeneity of the radio‐frequency field has a significant impact on , essentially by “smoothing off” peak decoupling conditions. Probes with flatter homogeneity profiles produce decay curves that are closer to exponential and with measurably longer values. Such factors contribute to the difficulty of reproducing quantitatively in simulation,13 and introduce a probe‐to‐probe variation that reduces the transferability of optimal decoupling sequences and parameters. This may be particularly problematic at higher NMR frequencies, where approaches the dimensions of the transmitter coil. The option of restricting the sample to regions with a more uniform RF profile is relatively unattractive, as improvements in sensitivity owing to longer will be more than offset by the overall loss of signal. Coil geometries with flatter RF profiles, such as the end‐compensated coils introduced by Yannoni and co‐workers,76 would be a better alternative, provided that overall RF performance can be maintained. There may also be greater scope for optimising probes at higher field; the noticeably poorer RF inhomogeneity of the probe used at 850 MHz (Figure S4 in the Supporting Information) will have had some impact on the values obtained.
The peak values of the time constant for nuclear spin decay, , of C in glycine summarised in Table 2 provide useful “reference” points when setting up experiments involving refocussing. By measuring the signal intensity after the refocussing period as a function of the decoupling parameters, the decoupling can be readily optimised, in a similar fashion to the optimisation of decoupling during acquisition periods. Indeed, can be readily estimated from the reduction of signal intensity relative to a reference experiment without the refocussing period, as used here to acquire parameter maps efficiently. These reference values can be used to judge the scope for further optimisation.
Although the best results were mostly obtained here using straightforward two‐pulse phase‐modulated (TPPM) decoupling, “good enough” decoupling for a given application may be achievable by using more easily optimised sequences. For example, CW20 and variants17, 18, 19 may get close to these values through single‐parameter optimisations, whereas modifications of TPPM, such as SWf‐TPPM64 and its supercycled variant SWf‐TPPM‐sc,77 are more tolerant to parameter mis‐set. Where optimal performance is critical, for example, to measure very small couplings, the detailed parameter maps obtained here, such as Figure S9 (in the Supporting Information), should be invaluable guides to efficient optimisation. The high values obtained have very real experimental significance for experiments involving small couplings; a of 100 ms allows couplings as small as a few Hz to be measured and means that challenging correlation experiments, such as refocussed INADEQUATE, are viable, even at natural isotopic abundance.
Supporting information
Acknowledgements
This work was supported by Engineering and Physical Sciences Research Council (EPSRC) grants EP/H023291/1 and EP/H023321/1. The UK 850 MHz solid‐state NMR Facility used in this research was funded by EPSRC and BBSRC, as well as the University of Warwick including part funding through Birmingham Science City Advanced Materials Projects 1 and 2 supported by Advantage West Midlands (AWM) and the European Regional Development Fund (ERDF). Dr Dinu Iuga is thanked for his practical assistance with experiments carried out at the 850 MHz Facility. John Stringer (Agilent Technologies, now Phoenix NMR) is thanked for helpful discussions. Research data can be found under: https://doi.org/10.15128/r1z316q156h.
I. Frantsuzov, S. K. Vasa, M. Ernst, S. P. Brown, V. Zorin, A. P. M. Kentgens, P. Hodgkinson, ChemPhysChem 2017, 18, 394.
References
- 1. Pham T. N., Griffin J. M., Masiero S., Lena S., Gottarelli G., Hodgkinson P., Filip C., Brown S. P., Phys. Chem. Chem. Phys. 2007, 9, 3416–3423. [DOI] [PubMed] [Google Scholar]
- 2. Lesage A., Bardet M., Emsley L., J. Am. Chem. Soc. 1999, 121, 10987–10993. [Google Scholar]
- 3. Harris R. K., Joyce S., Pickard C. J., Cadars S., Emsley L., Phys. Chem. Chem. Phys. 2006, 8, 137–143. [DOI] [PubMed] [Google Scholar]
- 4. Michan A. L., Divitini G., Pell A. J., Leskes M., Ducati C., Grey C. P., J. Am. Chem. Soc. 2016, 138, 7918–7931. [DOI] [PubMed] [Google Scholar]
- 5. Dupree R., Simmons T. J., Mortimer J. C., Patel D., Iuga D., Brown S. P., Dupree P., Biochemistry 2015, 54, 2335–2345. [DOI] [PubMed] [Google Scholar]
- 6. Zujovic Z., Webber A. L., Travas-Sejdic J., Brown S. P., Macromolecules 2015, 48, 8838–8843. [Google Scholar]
- 7. Fenniri H., Tikhomirov G. A., Brouwer D. H., Bouatra S., El Bakkari M., Yan Z., Cho J.-Y., Yamazaki T., J. Am. Chem. Soc. 2016, 138, 6115–6118. [DOI] [PubMed] [Google Scholar]
- 8. Rossini A. J., Widdifield C. M., Zagdoun A., Lelli M., Schwarzwälder M., Copéret C., Lesage A., Emsley L., J. Am. Chem. Soc. 2014, 136, 2324–2334. [DOI] [PubMed] [Google Scholar]
- 9. Märker K., Pingret M., Mouesca J.-M., Gasparutto D., Hediger S., De Paëpe G., J. Am. Chem. Soc. 2015, 137, 13796–13799. [DOI] [PubMed] [Google Scholar]
- 10. Ernst M., J. Magn. Reson. 2003, 162, 1–34. [DOI] [PubMed] [Google Scholar]
- 11. Hodgkinson P., Prog. Nucl. Magn. Reson. Spectrosc. 2005, 46, 197–222. [DOI] [PubMed] [Google Scholar]
- 12. Madhu P. K., Isr. J. Chem. 2014, 54, 25–38. [Google Scholar]
- 13. Frantsuzov I., Ernst M., Brown S. P., Hodgkinson P., Solid State Nucl. Magn. Reson. 2015, 70, 28–37. [DOI] [PubMed] [Google Scholar]
- 14. Equbal A., Leskes M., Nielsen N. C., Madhu P., Vega S., J. Magn. Reson. 2016, 263, 55–64. [DOI] [PubMed] [Google Scholar]
- 15. Equbal A., Bjerring M., Sharma K., Madhu P., Nielsen N. C., Chem. Phys. Lett. 2016, 644, 243–249. [Google Scholar]
- 16. Equbal A., Bjerring M., Madhu P. K., Nielsen N. C., J. Phys. Chem. 2015, 142, 184201. [DOI] [PubMed] [Google Scholar]
- 17. Equbal A., Bjerring M., Madhu P., Nielsen N. C., Chem. Phys. Lett. 2015, 635, 339–344. [DOI] [PubMed] [Google Scholar]
- 18. Equbal A., Paul S., Mithu V. S., Vinther J. M., Nielsen N. C., Madhu P., J. Magn. Reson. 2014, 244, 68–73. [DOI] [PubMed] [Google Scholar]
- 19. Equbal A., Paul S., Mithu V. S., Madhu P., Nielsen N. C., J. Magn. Reson. 2014, 246, 104–109. [DOI] [PubMed] [Google Scholar]
- 20. Vinther J. M., Nielsen A. B., Bjerring M., Van Eck E. R. H., Kentgens A. P. M., Khaneja N., Nielsen N. C., J. Phys. Chem. 2012, 137, 214202. [DOI] [PubMed] [Google Scholar]
- 21. Mithu V. S., Pratihar S., Paul S., Madhu P. K., J. Magn. Reson. 2012, 220, 8–17. [DOI] [PubMed] [Google Scholar]
- 22. Mithu V. S., Paul S., Kurur N. D., Madhu P. K., J. Magn. Reson. 2011, 209, 359–363. [DOI] [PubMed] [Google Scholar]
- 23. Gerbaud G., Caldarelli S., Ziarelli F., Gastaldi S., J. Magn. Reson. 2011, 210, 75–81. [DOI] [PubMed] [Google Scholar]
- 24. Paul S., Kurur N. D., Madhu P. K., J. Magn. Reson. 2010, 207, 140–148. [DOI] [PubMed] [Google Scholar]
- 25. Thakur R. S., Kurur N. D., Madhu P. K., J. Magn. Reson. 2008, 193, 77–88. [DOI] [PubMed] [Google Scholar]
- 26. Kotecha M., Wickramasinghe N. P., Ishii Y., Magn. Reson. Chem. 2007, 45, S221–S230. [DOI] [PubMed] [Google Scholar]
- 27. De Paëpe G., Lesage A., Emsley L., J. Chem. Phys. 2003, 119, 4833–4841. [Google Scholar]
- 28. De Paëpe G., Giraud N., Lesage A., Hodgkinson P., Böckmann A., Emsley L., J. Am. Chem. Soc. 2003, 125, 13938–13939. [DOI] [PubMed] [Google Scholar]
- 29. Apperley D. C., Harris R. K., Hodgkinson P., Solid-State NMR: Basic Principles & Practice, Momentum Press, New York, 2012, pp. 109–140. [Google Scholar]
- 30. Vasa S. K., Janssen H., Van Eck E. R. H., Kentgens A. P. M., Phys. Chem. Chem. Phys. 2010, 13, 104–106. [DOI] [PubMed] [Google Scholar]
- 31. De Paëpe G., Eléna B., Emsley L., J. Chem. Phys. 2004, 121, 3165. [DOI] [PubMed] [Google Scholar]
- 32. Taylor R. E., Concepts Magn. Reson. Part A 2004, 22, 79–89. [Google Scholar]
- 33. Hughes C. E., Harris K. D. M., Chem. Commun. 2010, 46, 4982–4984. [DOI] [PubMed] [Google Scholar]
- 34. Boldyreva E. V., Drebushchak V. A., Drebushchak T. N., Paukov I. E., Kovalevskaya Y. A., Shutova E. S., J. Therm. Anal. Calorim. 2003, 73, 419–428. [Google Scholar]
- 35. Metz G., Wu X., Smith S., J. Magn. Reson. Ser. A 1994, 110, 219–227. [Google Scholar]
- 36. Hediger S., Meier B. H., Kurur N. D., Bodenhausen G., Ernst R. R., Chem. Phys. Lett. 1994, 223, 283–288. [Google Scholar]
- 37. Meier B. H., Chem. Phys. Lett. 1992, 188, 201–207. [Google Scholar]
- 38. Laage S., Sachleben J. R., Steuernagel S., Pierattelli R., Pintacuda G., Emsley L., J. Magn. Reson. 2009, 196, 133–141. [DOI] [PubMed] [Google Scholar]
- 39. Gupta R., Hou G., Polenova T., Vega A. J., Solid State Nucl. Magn. Reson. 2015, 72, 17–26. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Laage S., Marchetti A., Sein J., Pierattelli R., Sass H. J., Grzesiek S., Lesage A., Pintacuda G., Emsley L., J. Am. Chem. Soc. 2008, 130, 17216–17217. [DOI] [PubMed] [Google Scholar]
- 41. Bennett A. E., Rienstra C. M., Auger M., Lakshmi K. V., Griffin R. G., J. Chem. Phys. 1995, 103, 6951–6958. [Google Scholar]
- 42. Tekely P., Palmas P., Canet D., J. Magn. Reson. Ser. A 1994, 107, 129–133. [Google Scholar]
- 43. Detken A., Hardy E. H., Ernst M., Meier B. H., Chem. Phys. Lett. 2002, 356, 298–304. [Google Scholar]
- 44. Fung B. M., Khitrin A. K., Ermolaev K., J. Magn. Reson. 2000, 142, 97–101. [DOI] [PubMed] [Google Scholar]
- 45. Comellas G., Lopez J. J., Nieuwkoop A. J., Lemkau L. R., Rienstra C. M., J. Magn. Reson. 2011, 209, 131–135. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46. Zorin V. E., Elena B., Lesage A., Emsley L., Hodgkinson P., Magn. Reson. Chem. 2007, 45, S93–S100. [DOI] [PubMed] [Google Scholar]
- 47. Brinkmann A., Vasa S. K., Janssen H., Kentgens A. P. M., Chem. Phys. Lett. 2010, 485, 275–280. [Google Scholar]
- 48. Weingarth M., Tekely P., Bodenhausen G., Chem. Phys. Lett. 2008, 466, 247–251. [Google Scholar]
- 49. van Beek J. D., J. Magn. Reson. 2007, 187, 19–26. [DOI] [PubMed] [Google Scholar]
- 50.MATLAB Release R2013b, The MathWorks Inc., Natick, Massachusetts, United States.
- 51.P. Hodgkinson, pNMRsim: a general simulation program for large problems in solid-state NMR, URL: http://www.durham.ac.uk/paul.hodgkinson/pNMRsim.
- 52. Leskes M., Thakur R. S., Madhu P. K., Kurur N. D., Vega S., J. Chem. Phys. 2007, 127, 024501. [DOI] [PubMed] [Google Scholar]
- 53. Griffin J. M., Tripon C., Samoson A., Filip C., Brown S. P., Magn. Reson. Chem. 2007, 45, S198–S208. [DOI] [PubMed] [Google Scholar]
- 54. Vanderhart D. L., Earl W. L., Garroway A. N., J. Magn. Reson. 1981, 44, 361–401. [Google Scholar]
- 55. Mehring M., Principles of High Resolution NMR in Solids , 2nd edn., Springer, Berlin, 1983. [Google Scholar]
- 56. De Paëpe G., Sakellariou D., Hodgkinson P., Hediger S., Emsley L., Chem. Phys. Lett. 2003, 368, 511–522. [Google Scholar]
- 57. Ernst M., Bush S., Kolbert A. C., Pines A., J. Chem. Phys. 1996, 105, 3387. [Google Scholar]
- 58. Ernst M., Samoson A., Meier B. H., J. Chem. Phys. 2005, 123, 064102. [DOI] [PubMed] [Google Scholar]
- 59. Scholz I., Hodgkinson P., Meier B. H., Ernst M., J. Chem. Phys. 2009, 130, 114510. [DOI] [PubMed] [Google Scholar]
- 60. Ernst M., Geen H., Meier B. H., Solid State Nucl. Magn. Reson. 2006, 29, 2–21. [DOI] [PubMed] [Google Scholar]
- 61. Ernst M., Samoson A., Meier B. H., J. Magn. Reson. 2003, 163, 332–339. [DOI] [PubMed] [Google Scholar]
- 62. Purusottam R. N., Bodenhausen G., Tekely P., Chem. Phys. Lett. 2015, 635, 157–162. [Google Scholar]
- 63. Hodgkinson P. in Annual Reports on NMR Spectroscopy (Ed.: G. A. Webb), Academic Press, vol. 72 of Annual Reports on NMR Spectroscopy, 2011, pp. 185–223. [Google Scholar]
- 64. Thakur R. S., Kurur N. D., Madhu P. K., Chem. Phys. Lett. 2006, 426, 459–463. [Google Scholar]
- 65. Odedra S., Wimperis S., J. Magn. Reson. 2013, 231, 95–99. [DOI] [PubMed] [Google Scholar]
- 66. Vega S., Olejniczak E. T., Griffin R. G., J. Phys. Chem. 1984, 80, 4832–4840. [Google Scholar]
- 67. Mehring M., Waugh J. S., Rev. Sci. Instrum. 1972, 43, 649–653. [Google Scholar]
- 68. Bosman L., Madhu P., Vega S., Vinogradov E., J. Magn. Reson. 2004, 169, 39–48. [DOI] [PubMed] [Google Scholar]
- 69. Carravetta M., Edén M., Johannessen O. G., Luthman H., Verdegem P. J. E., Lugtenburg J., Sebald A., Levitt M. H., J. Am. Chem. Soc. 2001, 123, 10628–10638. [DOI] [PubMed] [Google Scholar]
- 70. Vega A. J., J. Magn. Reson. 2004, 170, 22–41. [DOI] [PubMed] [Google Scholar]
- 71. Tabuchi Y., Negoro M., Takeda K., Kitagawa M., J. Magn. Reson. 2010, 204, 327–332. [DOI] [PubMed] [Google Scholar]
- 72. Wittmann J. J., Takeda K., Meier B. H., Ernst M., Angew. Chem. Int. Ed. 2015, 54, 12592–12596; [DOI] [PubMed] [Google Scholar]; Angew. Chem. 2015, 127, 12781–12785. [Google Scholar]
- 73. Kentgens A. P. M., Bart J., van Bentum P. J. M., Brinkmann A., Van Eck E. R. H., Gardeniers J. G. E., Janssen J. W. G., Knijn P., Vasa S., Verkuijlen M. H. W., J. Chem. Phys. 2008, 128, 052202. [DOI] [PubMed] [Google Scholar]
- 74. Tjandra N., Bax A., J. Am. Chem. Soc. 1997, 119, 8076–8082. [Google Scholar]
- 75. Agarwal V., Tuherm T., Reinhold A., Past J., Samoson A., Ernst M., Meier B. H., Chem. Phys. Lett. 2013, 583, 1–7. [Google Scholar]
- 76. Horne D., Kendrick R. D., Yannoni C. S., J. Magn. Reson. 1983, 52, 299–304. [Google Scholar]
- 77. Augustine C., Kurur N. D., Magn. Reson. Chem. 2010, 48, 798–803. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.