Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2017 Aug 1.
Published in final edited form as: Adv Ophthalmol Optom. 2016 Aug;1(1):231–247. doi: 10.1016/j.yaoo.2016.03.004

Objective Measures of Visual Function in Papilledema

Heather E Moss 1
PMCID: PMC5403134  NIHMSID: NIHMS788136  PMID: 28451649

Synopsis

Visual function is an important parameter to consider when managing patients with papilledema. Though the current standard of care uses standard automated perimetry (SAP) to obtain this information, this test is inherently subjective and prone to patient errors. Objective visual function tests including the visual evoked potential, pattern electroretinogram, photopic negative response of the full field electroretinogram, and pupillary light response have the potential to replace or supplement subjective visual function tests in papilledema management. This article reviews the evidence for use of objective visual function tests to assess visual function in papilledema and discusses future investigations needed to develop them as clinically practical and useful measures for this purpose.

Keywords: papilledema, optic nerve, retinal ganglion cells, visual evoked potential, electroretinogram, photopic negative response, pupillary light response

INTRODUCTION

Papilledema is swelling of optic nerve head that occurs as a consequence of elevated cerebral spinal fluid (CSF) pressure in the optic nerve sheath compressing the retinal ganglion cells (RGCs) that form the optic nerve and the blood vessels that supply them.1,2 Any cause of elevated intracranial pressure (ICP) can cause papilledema, including brain tumors, venous sinus thrombosis, meningitis, and idiopathic intracranial hypertension (IIH). Irreversible blindness due to RGC injury is a devastating consequence of papilledema, occurring in 6–14% of patients with IIH, with 50% suffering some degree of vision loss.3,4 Treatments are aimed at reducing CSF pressure on the optic nerve either directly through optic nerve sheath fenestration, or indirectly by reducing intracranial CSF production through medication or by draining CSF from the cerebral ventricles or spinal canal. These treatments are effective for reversal or stabilization of vision loss from papilledema and sometimes are needed in addition to directed treatment aimed at the underlying cause of ICP elevation.5

Visual function measurement, using standard automated perimetry (SAP), for example, using the Humphrey Visual Field Analyzer (Zeiss Meditech Inc) is a primary basis for papilledema management decisions because it assesses the clinical outcome of interest.6 However, as a psychophysical test, SAP is inherently subjective and prone to patient error. This introduces uncertainty into interpretation of its results with respect to RGC function.7 Objective measures that accurately assess the function of the injured neurons could serve as an alternative or additional clinical marker upon which to base treatment decisions. Identifying and characterizing these measures are of significant importance to the ophthalmology and neurology communities who are striving to improve clinical care of IIH and prevent vision loss from papilledema. A recent example of these efforts is the IIH treatment trial (IIHTT), which was sponsored by the Neuro-Ophthalmology Research Disease Investigator Consortium (NORDIC) and National Eye Institute (NEI).5,8

In addition to overcoming the limitations of SAP, objective measures of visual function have the potential to provide information not currently captured by SAP by representing physiological responses beyond those contributing to conscious visual function. Literature regarding other optic neuropathies suggests that objective visual function measures are more sensitive than SAP because abnormalities on some objective measures of visual function based on electrophysiology may precede vision loss as measured by SAP.9 This could facilitate earlier identification of visual dysfunction or its progression. Furthermore, objective measures of visual function may capture predictive or prognostic information that SAP does not. This article reviews methodologies for objective measurement of retinal ganglion cell function and evidence of their applicability for assessment of visual function in papilledema.

SIGNIFIGANCE

Basis of objective visual function measurements

All measurements of vision aim to determine a person’s ability to perceive a certain light stimulus. Stimuli range from a brief flash of light, as is the case in full field electroretinography (ERG) to formed images of standard sizes, as is the case in Snellen visual acuity testing. For psychophysical measurements, such as visual acuity or perimetry, a patient indicates the presence or absence of perception either verbally or through another conscious action such as pushing a button. In contrast, objective measurements do not ask for a conscious response from the patient, rather they collect physiological measurements from which information regarding function of part or parts of the visual pathway is/are extracted. The most common physiological measurement is an electrical signal, for example from the retina (ERG) or visual cortex (visual evoked potential (VEP)). Other measureable signals include change in pupil size, for example in the pupillary light response, or nystagmus, for example in optokinetic nystagmus (Table 1).

Table 1.

Comparison of subjective and objectives measures of visual function

Subjective Objective
Examples Snellen visual acuity
Perimetry (automated or manual; static or kinetic)
Visual evoked potential Electroretinogram
Pupillometry
Typical stimulus complex lines (e.g. letters)
light stimuli in varying locations
full field light
time varying graphical pattern
Response Subjective by the patient (e.g. verbal or button pressing) Electrical or imaging waveform
What test is measuring Integrated function of visual system Physiologic response of sections of the visual system
Causes of underestimation (false negative) Poor subject cooperation (voluntary* or otherwise)
Incorrect testing protocol
Poor signal recorded
Incorrect testing protocol
*

e.g. malingering, factitious disorder

Electrophysiological and pupil based measures of RGC function offer advantages over SAP because they are objective, and have minimal patient demands. They are appropriate to consider as clinical markers because they are quick and non-invasive. There are multiple tests relating to RGC function that are abnormal when RGCs are compromised, including the visual evoked potential (VEP), pattern electroretinogram (PERG), photopic negative response (PhNR) of the full field electroretinogram (FF-ERG), and pupillary light response (PLR). These vary in their requirements for refraction, dilation, fixation, and maintaining open eyes. They also vary in the anatomical specificity of the physiological signal they measure (Table 2).

Table 2.

Comparison of objective measures of retinal ganglion cell function

Visual evoked potential Full field electroretinogram Pattern electroretinogram Pupillary Light Response
Standard stimulus Pattern reversal: Alternating black/white checkerboard pattern in central visual field Full field brief flashes of light in DA or LA state Pattern reversal: Alternating black/white checkerboard pattern in central visual field Full field brief flashes of light
Specialized stimuli Sinusoidal pattern reversal
Pattern onset
Pattern offset
Full field flash
PhNR: Blue light flash on red background, LA Melanopsin: bright blue light, DA
Cone: Red light, LA
Rod: dim blue light, DA
Recording location Skin over occiput Cornea Cornea Infra-red image of pupil recorded with camera
Subject Requirements Maintain open eye
Maintain Fixation
Refractive correction
Maintain open eye
Dilated pupil
Maintain open eye
Maintain fixation
Refracted correction
Maintain open eye
Structures assessed Entire visual pathway Eye, retina Eye, retina Eye, optic nerve, brainstem
Output electroencephalogram Electroretinogram Electroretinogram pupil size
Measures relevant to RGC function P100 latency & amplitude PhNR amplitude N95:P50 amplitude ratio Magnitude of transient response
Magnitude of sustained response

DA – dark adapted, LA – light adapted, PhNR – photopic negative response, RGC – retinal ganglion cell, P100 – positive peak at approximated 100ms, P50 – positive peak at approximately 50ms, N75 negative peak at approximately 75ms

Visual Evoked Potential (VEP)

The visual evoked potential is a waveform recorded by electrodes placed over the skin overlying the occipital lobe in response to a controlled visual stimulus. The most commonly used stimulus is “pattern reversal” consisting of an alternating black and white checkerboard pattern in the central/mid visual field. Less often used stimuli include pattern onset, pattern offset, and flash, which have more variability between individuals. The measured signal is an electrical potential as a function of time with latency and amplitude of the first positive peak (p100) being the primary outcome measure of the transient response or phase and amplitude in the steady state response. Due to the type of stimulus and the placement of recording electrodes the signal mainly captures the function related to the central area of the visual field. The signal can be altered by pathology anywhere along the visual pathway including retina, optic nerve and hemispheric causes of visual impairment. Stimulus factors that affect the VEP include the spatial frequency of the stimulus (i.e. the size of the checks), the total size of the stimulus visual field, and the temporal frequency of the stimulus (i.e. speed of the stimulus change). Low frequency stimuli capture the transient response, while high frequency stimuli capture the steady start response. Subject factors that affect the VEP include age and refraction. The subject must maintain fixation on the center of the stimulus field for the duration of the test. Therefore refractive correction, subject attention, and subject effort are necessary for high quality measurements.10

The most common VEP abnormality reported in association with papilledema is prolonged p100 latency (transient response) or delayed phase (steady state response)(Table 3).1117 Estimates of the prevalence of prolonged VEP latency in IIH, a disease characterized by papilledema, ranges from 10–55%, with only some studies demonstrating a significant difference between IIH and control groups. One study of children with craniosynostosis, another disease associated with elevated ICP and papilledema, reported normal latency in all 5 patients with papilledema.18 Multiple studies have evaluated VEP longitudinally, uniformly reporting shortening of latency following treatment of ICP.11,13,14,19 One study included a single patient with progressive vision loss due to papilledema;’ this was associated with progressive latency delay.12

Table 3.

Studies reporting visual evoked potential (VEP) results in subjects with papilledema (in chronological order, * indicates statistically significant finding)

Recording/ study stimulus n Papilledema subject details comparison findings Correlation with other measures Follow up
VEP
Kirkham, 198111 Full field flicker (steady state) 2 brain tumor with papilledema (n=2) Lab norms Latency delayed (LA)
Latency normal (DA)
ERG latency prolonged (LA, DA) (see table 4) Latency normalized with ICP improvement
Onofrj, 198121 Pattern reversal (sinusoidal pattern, 2 temporal frequencies) 6 papilledema, w/o
hydrocephalus (n=3)
papilledema w/ hydrocephalus (n=2)
hydrocephalus
w/o papilledema (n=1)
Latency delay associated w/ hydrocephalus, but not papilledema NR In patient with hydrocephalus w/ papilledema VEP normalized s/p VP shunt, prior to papilledema resolution.
Rizzo, 198412 Pattern reversal (check pattern) 27 IIH with papillary stasis (n=7), 2 w/ constricted VF Controls (n=20) Latency delayed in 2/7 IIH patients Latency delayed exclusively in patients with VF constriction (n=2) Latency shortened 24 hours after LP
In patients with VF constriction (n=2)
Sørenson, 198514 (overlap with Krogsaa) Pattern reversal (check pattern) 33 IIH, all with papilledema, 3 w/ normal ICP (n=13) Controls (n=20) Latency delayed in IIH* 6/25 IIH eyes w/latency above control range Latency associated with ICP*
Latency not associated with papilledema degree or TVOs
Latency increase associated with progressive loss of vision (n=1)
Latency decrease associated with resolution of papilledema (n=6)
Krogsaa, 198513 (overlap with Sørenson) Pattern reversal 35 IIH, all with papilledema (n=15), 12 with normal VF Controls (n=20) Latency delayed in IIH* Latency delay preceded vision loss in patient who lost vision Latency improved with papilledema resolution (n=8)
Bobak, 198815 Pattern reversal (sinusoidal pattern, 2 temporal frequencies) 141 IIH (n=11), papilledema in 10, no visual field loss in 8, mild visual field loss in 3 Recovered optic
neuritis (n=8)
thyroid eye
disease (n=7)
controls (n=35)
4/11 IIH w/ abnormal transient latency or steady state phase
no thyroid eye disease or optic neuritis with normal latency and phase
NR NR
Verplanck, 198816 Pattern reversal (check pattern,3 spatial frequencies) 60 IIH (n=15) Controls (n=45) 4/15 IIH w/ abnormal latency (5 eyes) all eyes w/ abn VEP had abn CS 14/18 eyes w/ abn CS had nrl VEP 10/13 eyes w/ abn VF abn had nrl VEP NR
Falsini, 199220 Pattern reversal (vertical sinusoidal pattern, multiple spatial frequencies) 39 IIH, grade 0–2 papilledema, minimal or no VF loss (n=18) Controls (n=21) Absolute amplitude and phase not different between IIH and control subjects
Normalized amplitude decreased at high frequencies in IIH and low frequencies in controls
10/18 IIH patients with at least one abnormality (phase or amplitude at any stimulus frequency)
9/10 IIH w/ VEP abnormality had PERG abnormality NR
Mursch, 199819 Flash VEP (n=48)
Pattern reversal (n=4)
52 none Craniosynostosis (n=52), 1 with optic atrophy, none with papilledema Latency delayed in 12/52
Amplitude suppression in 2/52
NR 4/4 with pathological pre- operative VEP had improvement after ICP lowering surgery
Liasis, 200618 Pattern reversal, ISCEV protocol 8 craniosynostosis w/ papilledema, high ICP (n=5) Craniosynostosis w/o papilledema (n=3) Latency normal in all Amplitude initially normal in all NR Amplitude decreased in all subjects prior to intervention and increased in all following ICP lowering surgery
Kesler, 200917 Pattern reversal (check pattern) 20 Chronic IIH, 14 w/ abn fundus (swelling or atrophy), 6 w/ abn VF (n=20) Lab norms Latency prolonged in 11/20 w/ IIH Latency prolonged in 12/14 eyes w/ abn fundus, 4/10 eyes w/ abn VF, 12/30 eyes w/ nrl VF NR

n – number of subjects, LA – light adapted (photopic), DA – dark adapted (scotopic), ERG – electroretinogram, ICP – intracranial pressure, w/ - with, w/o – without, NR – not reported; VP – ventriculoperitoneal shunt, IIH – idiopathic intracranial hypertension, VF – visual field, LP = lumbar puncture, TVO – transient visual obscurations, abn – abnormal, nrl – normal, PERG – pattern ERG, ISCEV – International Society for Clinical Electrophysiology of Vision

VEP amplitude abnormalities were not found by most investigators between IIH and control subjects. Falsini et al reported the association between VEP amplitude and stimulus spatial frequency to be different in IIH and control subjects with IIH subjects having decrease of normalized amplitudes with high spatial frequency stimuli, while control subjects had decrease of amplitudes with low spatial frequency stimuli.20 Liasis et al reported progressive amplitude decrease prior to ICP lowering surgery in craniosynostosis in 8 patients and recovery following operative intervention to lower ICP in all of them.18

One possibility for the low prevalence of abnormalities is that papilledema pathology may affect the VEP less than other optic nerve diseases. This was suggested by one study that demonstrated VEP abnormalities in IIH to be less prevalent than in other optic neuropathies (e.g. optic neuritis, thyroid eye disease) with similarly mild degrees of vision loss.15 This low sensitivity of VEP abnormalities for papilledema was also demonstrated by Verplanck et al who found only 5/30 eyes in their study to have abnormal VEP, despite 18 eyes and 15 eyes demonstrating abnormal contrast sensitivity or visual fields respectively.16 Similarly Kesler et al reported that only 4/10 eyes with abnormal visual fields had abnormal VEP.17 One study found pattern ERG to be more sensitive for papilledema detection than VEP.20

It is important to keep in mind that VEP interrogates the entire visual system, not just the optic nerve that is structurally affected by papilledema. Some studies present evidence that elevated ICP may affect VEP through effects on intracerebral structures other than the optic nerve. Onofrj et al demonstrated VEP latency delays in patients with hydrocephalus without papilledema, and proposed that there may be an intracranial effect of elevated ICP on the VEP, perhaps through effect on the optic radiations.21 One of their subjects who had both hydrocephalus and papilledema had improvement in VEP latency rapidly following CSF shunting procedure prior to resolution of papilledema. Mursch et al also contribute evidence for central visual pathway contributions to abnormal VEP with their observation of delayed VEP in 12/52 children with craniosynostosis, none of whom had papilledema, and only one of whom had optic atrophy.19

Multifocal VEP is a newer test that measures spatial variation in the VEP signal to create a 2 dimensional VEP map. This is achieved by using a time varying complex signal pattern and complex signal analysis. It has been shown to correlate with visual field defects in other optic neuropathies.22,23 However, a literature review did not locate studies using this technique in papilledema.

Full Field Electroretinography (FF-ERG) and the Photopic Negative Response (PhNR)

The full field electroretinogram is a waveform recorded by electrodes in contact with the cornea (preferred) or skin adjacent to the eye in response to a flash or flicker of light occupying the entire visual field and thus stimulating the entire retina. Pharmacological pupillary dilation is typically used to maximize exposure. Stimulus parameters such as color, luminance, and duration can be manipulated in light or dark adapted conditions to generate or accentuate components of the waveform that localize to specific components of the retinal circuitry.24 The International Society for Clinical Electrophysiology in Vision (ISCEV) has established standards for clinical FF-ERG recordings including technique and minimum stimuli to be administered.25 The measured parameters are the amplitude and timing (implicit time) of peaks in the waveform (a, b) and high frequency components (oscillatory potentials). Due to the diffuse and brief nature or the stimulus, demands on the subject to fixate are minimal. Refractive correction is not needed, though people with high myopia tend to have smaller signals, likely due to increased distance between the electrode and the retina.

Of particular interest with respect to papilledema is the photopic negative response (PhNR), which is the negative wave following the b wave, the main positive wave of the full field ERG. The amplitude of the PhNR is reduced in other optic neuropathies, including glaucoma and optic atrophy. Animal studies and structure function correlations provide additional evidence that it is generated by the RGCs2629. Though discernable on ISCEV standard recordings, studies focusing on the PhNR typically use specialized protocols consisting of a red flash on a blue background, because these accentuate the PhNR.30

In a study of 10 IIH patients with variable papilledema grades, PhNR was reduced compared with controls, and 6/10 IIH patients had abnormal PhNR outside of the range of control measurements (Table 4).31 Log PhNR was linearly associated with a subjective test of visual function, SAP. Log PhNR was highly correlated with ganglion cell atrophy and papilledema severity, both structural measures of ganglion cell changes. In one subject with high-grade papilledema and minimal visual field loss, PhNR was absent and recovered with treatment of ICP and resolution of papilledema.

Table 4.

Studies reporting full field electroretinogram (FF-ERG), pattern electroretinogram (PERG) or pupillary light response (PLR) results in subjects with papilledema (in chronological order, * indicates statistically significant finding)

Recording/ study stimulus n Papilledema subject details comparison findings Correlation with other measures Follow up
FF-ERG, PhNR
Kirkham, 198111 Full field flicker (steady state) 2 brain tumor with papilledema (n=2) Lab norms ERG latency prolonged (LA, DA) VEP latency prolonged LA but not DA (see table 3) Latency normalized with ICP improvement
Moss, 201531 Full field red flash on blue background 25 IIH w/ optic nerve atrophy (n=2), mild papilledema (n=6), severe papilledema (n=2) Controls (n=15) PhNR prolonged in IIH* 6/10 IIH patients with abnormal PhNR PhNR correlated w/ SAP mean deviation* PhNR correlated w/ ganglion cell atrophy and papilledema severity* PhNR normalized w/ resolution of high grade papilledema in a patient w/ mild vision loss (n=1)
PERG
Falsini, 199220 Pattern reversal (vertical sinusoidal pattern, multiple spatial frequencies) 39 IIH, grade 0- 2
papilledema, minimal or no VF loss (n=18)
Controls (n=21) amplitude lower in IIH at intermediate/high spatial frequencies phase in IIH similar to controls
Normalized amplitude decreased w/ increasing stimulus frequency in IIH and peaked at mid frequencies in controls 15/18 IIH w/ at least one abnormality (phase or amplitude at any stimulus frequency)
9/15 patients w/ abn PERG had abn VEP NR
Afonso, 201535 Pattern reversal (check pattern, 2 spatial frequencies, ISCEV standard) 50 IIH (n=21) or VST (n=3),
All resolved papilledema, stable, abn VF for 6 mo
Controls (n=26) P50, N95, P50+N95 amplitude smaller in IIH for large checks*
N95 latency longer in IIH for large checks*
N95, P50+N95 amplitudes smaller in IIH for small checks*
Latencies not affected for small checks
AUC greatest for N95 large checks
Macular and peripapillary OCT thickness parameters correlated w/ N95 and P50+N95 amplitude for large checks*
Macular OCT thickness parameters correlated w/ N95 and P50+N95 amplitudes for small checks*
Visual field loss correlated w/ N95 and P50+N95 amplitudes for both stimulus sizes*
NR
PLR
Park, 201639 Full field flashes (dim blue, bright blue, moderate red) 26 IIH, variable papilledema grade (n=13) Controls (n=13) Receiver operating characteristic for IIH vs. controls: 0.71 rod response (dim blue), 0.77 cone response (mod red), 0.83 melanopsin response (bright blue), 0.90 combined response
Response abn in 6/13 IIH (rod), 9/13 (cone), 6/13 (melanopsin)*
Sustained PLR amplitude (melanopsin response) correlated w/ SAP and log PhNR amplitude* NR

PhNR – photopic negative response, n – number of subjects, LA – light adapted (photopic), DA – dark adapted (scotopic), VEP – visual evoked potential, ICP – intracranial pressure, IIH – idiopathic intracranial hypertension, w/ - with, SAP – standard automated perimetry, VF – visual field, NR – not reported, ISCEV – International Society for Clinical Electrophysiology of Vision, VST – cerebral venous sinus thrombosis, AUC – area under curve, OCT – optical coherence tomography, abn – abnormal

Waveforms of the full field ERG other than the PhNR have not been widely studied in papilledema, likely because they represent retinal structures of which there is unlikely to be dysfunction associated with papilledema. However, a single study reported prolonged latency of the scotopic and photopic responses to low and high frequency flicker respectively in two patients with papilledema (Table 4). Both patients had normalization of ERG latency and papilledema following surgical intervention.11 The manuscript indicates similar findings in 6 other patients with papilledema. The mechanism of this puzzling observation was not elucidated, though the authors proposed outer retinal dysfunction related to transynaptic signaling or metabolic disturbance related to hemodynamic changes.

Similar to multifocal VEP, multifocal ERG measures spatial variation of the ERG within the central visual field.24 It is much more sensitive to changes in the macula than the full field ERG. Changes have been demonstrated in glaucomatous eyes, but, to our knowledge, it has not been applied to papilledema.32

Pattern Electroretinography (PERG)

The pattern electroretinogram (PERG) is a waveform recorded by electrodes in contact with the cornea in response to a pattern reversal stimulus in the central visual field. Spatial and temporal frequency can be manipulated and there are ISCEV standards for clinical testing.33 Unlike FF-ERG, pupillary dilation is not required. Similar to pattern reversal VEP, refractive correction is required and the subject must maintain visual fixation on the stimulus. The measured parameters are the amplitudes of the P50 and N95 peaks of the PERG waveform and the N95:P50 ratio. Evidence suggests that RGCs contribute partially to P50 and are the predominant contributor to N95. Due to the small size of the signal, responses to repetitive stimuli must be averaged. Subject fixation with minimal blinking is essential to achieve this.34

Studies have measured PERG in mild active papilledema with minimal or no visual field loss20 and in post-papilledema atrophy with visual field loss (Table 4).35 Both found reduced amplitudes to be associated with papilledema. However spatial frequency effects were opposite, with the group that studied mild active papilledema demonstrating more normalized amplitude decrement at higher spatial frequencies (smaller stimulus size), while the group that studied post papilledema atrophy demonstrated a more prominent effect at the lower spatial frequency (larger stimulus size). In one study, PERG detected abnormalities in a higher proportion of subjects than VEP, suggesting that it is more sensitive for detection of papilledema pathology.20 In the other study correlations between PERG and both visual field loss and structural loss of RGCs were demonstrated.35

Pupillary Light Response (PLR)

The pupillary light response is a measure of the change in pupil size in response to light stimulation. 3638 By varying the light color (wavelength), light intensity, and eye adaptation, it is possible to characterize components of the pupillary light response driven by rod, cone, and melanopsin ganglion cell pathways. Rod and cone pathway responses are measured by the amplitude of the transient pupillary constriction, often normalized to baseline size. Melanopsin ganglion cells are in(??) intrinsically photosensitive and do not contribute to conscious visual perception. Their pupil response has a characteristic slow recovery measured as the sustained response. The PLR response has been studied and is abnormal in other early optic neuropathies including glaucomatous, compressive, and ischemic.36

IIH patients demonstrate abnormalities in all components of the PLR, with the sustained melanopsin response demonstrating the largest area under the curve in receiver operating characteristic analysis against control subjects (Table 4).39 The amplitude of the sustained (melanopsin) PLR is correlated with SAP, a subjective measure of visual function and log PhNR, another objective measure of visual function in IIH patients.

CURRENT CLINICAL RELEVANCE

Current clinical use of objective measures of visual function in papilledema mostly consists of VEP, the most widely available electrophysiological test, in cases where there is concern for malingering or somatoform disorder causing false positive subjective visual function testing. VEP has likely not been applied more broadly in papilledema management due to lack of evidence that VEP exceeds subjective visual function testing as a sensitive and specific marker of visual dysfunction associated with papilledema. Subjective visual function testing is also more widely available and less technically challenging. PERG, PhNR, and PLR have not been sufficiently studied in the papilledema and IIH populations to justify widespread clinical adoption. Furthermore, the technical requirements current limit the availability of testing to centers with expertise in electrophysiology.

Future avenues for investigation

The next step to develop PERG, PhNR, and PLR for clinical application in papilledema management is to characterize the responses in terms of associations with other disease markers, variability, and reliability. This will require large-scale studies of populations in whom these measures might be used as clinical markers, for example, patients with IIH and control subjects. Longitudinal studies are necessary to fully understand how the measurements change within individuals over time and how measurements relate to concurrent standard visual markers and future disease course. These studies will also refine how objective visual function markers might most contribute to clinical care of papilledema.

It will next be important to determine if objective visual function testing improves clinical care. This should be considered in a broad sense. Clinical care could be improved by improving visual outcomes or by reducing health care expenditures or by facilitating new patterns of care. For example, if objective visual outcome measures are easy to administer and interpret for a non-expert, they might be of use in the emergency room setting where ophthalmic expertise is not readily available.

Published studies applying PERG, PhNR and PLR have used careful methodology, requiring expertise and equipment that is impractical for wide scale clinical implementation. This will limit broad adoption, even if the data gained from the other proposed experiments supports widespread use. Development of testing platforms that are less operator dependent and less expensive will enhance feasibility of clinical use.

Summary/Discussion

Objective measures of visual function address many of the shortcomings of subjective visual function measures. The visual evoked potential, full field electroretinogram, pattern electroretinogram, and pupillary light response all have demonstrated abnormalities in association with papilledema. Though VEP has been investigated most widely, studies suggest that it is less sensitive to visual function change in papilledema than subjective visual function testing. PERG, PhNR of the FF-ERG, and PLR show more promise as new or additional clinical markers for papilledema due to higher sensitivity and correlation with other markers of papilledema. PERG, PhNR of the FF-ERG, and PLR are all associated with subjective measures of visual function in IIH, suggesting that they are an index of the outcome of interest (i.e., visual function). PhNR and PERG have been shown to be associated with structural measures of ganglion cell abnormality in IIH, supporting its development as a marker of ganglion cell function in IIH. Other potentially interesting measures, namely multifocal VEP and multifocal ERG have not been investigated in the papilledema population.

Objective measures of visual function, in particular PERG, PhNR and PLR have important potential application to the diagnosis and management of IIH either as a replacement for subjective visual function testing or as a supplementary test in addition to subjective measures of visual function. They need not be a perfect test to have clinical utility. For example, if they can be shown to have high specificity for visual pathway compromise associated with papilledema, their results would help to exclude visual dysfunction and avoid overtreatment. Alternatively, if they can be shown to have high sensitivity for visual pathway compromise associated with papilledema, their results would help to screen for visual dysfunction and identify those who might be in need of further testing and/or care. If they provide additional diagnostic or prognostic information beyond what is captured by subjective visual function testing, their results may provide additional information to physicians to guide management of papilledema. While published data suggest that PERG, PhNR, and PLR are promising with respect to these applications, substantial additional investigation is necessary to develop them to the level of widespread clinical implementation in the area of papilledema diagnosis and management.

Key Points.

  1. Though the visual evoked potential has been widely studied in papilledema, its sensitivity for vision loss is less than standard subjective visual function tests.

  2. The photopic negative response of the full field electroretinogram is abnormal and associated with ganglion cell atrophy and papilledema grade in patients with idiopathic intracranial hypertension.

  3. The pattern electroretinogram is abnormal and correlated with ganglion cell loss in post papilledema subjects with stable vision loss.

  4. The sustained pupillary light response is abnormal in patients with idiopathic intracranial hypertension and correlates with subjective and other objective measures of visual function.

  5. Further investigation is needed to characterize changes in objective visual function tests in papilledema to develop them into clinically useful diagnostic tests for this disease.

Acknowledgments

Dr Moss’ research is supported by an unrestricted grant from research to prevent blindness to the UIC Department of Ophthalmology and Visual Sciences (NIH K23 EY-024345)

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • 1.Trobe JD. Papilledema: The Vexing Issues. J Neuroophthalmol. 2011;31(2):175. doi: 10.1097/WNO.0b013e31821a8b0b. [DOI] [PubMed] [Google Scholar]
  • 2.Tso MO, Hayreh SS. Optic disc edema in raised intracranial pressure: IV. Axoplasmic transport in experimental papilledema. Arch Ophthalmol. 1977;95(8):1458. doi: 10.1001/archopht.1977.04450080168023. [DOI] [PubMed] [Google Scholar]
  • 3.Corbett JJ, Savino PJ, Thompson HS, et al. Visual loss in pseudotumor cerebri. Follow-up of 57 patients from five to 41 years and a profile of 14 patients with permanent severe visual loss. Arch Neurol. 1982;39(8):461–474. doi: 10.1001/archneur.1982.00510200003001. [DOI] [PubMed] [Google Scholar]
  • 4.Orcutt J, Page N, Sanders M. Factors affecting visual loss in benign intracranial hypertension. Ophthalmology. 1984;91(11):1303. doi: 10.1016/s0161-6420(84)34149-5. [DOI] [PubMed] [Google Scholar]
  • 5.Markey KA, Mollan SP, Jensen RH, Sinclair AJ. Understanding idiopathic intracranial hypertension: mechanisms, management, and future directions. Lancet Neurol. 2016 Jan;15(1):78–91. doi: 10.1016/S1474-4422(15)00298-7. [DOI] [PubMed] [Google Scholar]
  • 6.Rowe FJ, Sarkies NJ. Assessment of visual function in idiopathic intracranial hypertension: a prospective study. Eye. 1998;12(1):111–118. doi: 10.1038/eye.1998.18. [DOI] [PubMed] [Google Scholar]
  • 7.Digre KB, Nakamoto BK, Warner JEA, Langeberg WJ, Baggaley SK, Katz BJ. A comparison of idiopathic intracranial hypertension with and without papilledema. Headache: The Journal of Head and Face Pain. 2009;49(2):185–193. doi: 10.1111/j.1526-4610.2008.01324.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Wall M. Idiopathic Intracranial Hypertension and the Idiopathic Intracranial Hypertension Treatment Trial. J Neuroophthalmol. 2013;33(1):1–3. doi: 10.1097/WNO.0b013e3182819aee. [DOI] [PubMed] [Google Scholar]
  • 9.Calkins DJ. Critical pathogenic events underlying progression of neurodegeneration in glaucoma. Prog retin eye res. 2012 doi: 10.1016/j.preteyeres.2012.07.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Creel D. Visually Evoke Potentials. [Accessed January 5, 2016];Webvision. 2015 http://webvision.med.utah.edu/book/electrophysiology/visually-evoked-potentials/
  • 11.Kirkham T, Coupland S. Abnormal electroretinograms and visual evoked potentials in chronic papilledema using time-difference analysis. Can J Neurol Sci. 1981;8(3):243. doi: 10.1017/s0317167100043274. [DOI] [PubMed] [Google Scholar]
  • 12.Rizzo P, Pierelli F, Pozzessere G, Sancesario G, Boatta M, Santarcangelo G. Pattern visual evoked potentials in pseudotumor cerebri. A longitudinal study. Acta Neurol Belg. 1984 Mar-Apr;84(2):57–63. [PubMed] [Google Scholar]
  • 13.Krogsaa B, Soelberg Sorensen P, Seedorff HH, Trojaborg W, Gjerris F. Ophthalmologic prognosis in benign intracranial hypertension. Acta Ophthalmol Suppl. 1985;173:62–64. doi: 10.1111/j.1755-3768.1985.tb06842.x. [DOI] [PubMed] [Google Scholar]
  • 14.Sorensen PS, Trojaborg W, Gjerris F, Krogsaa B. Visual evoked potentials in pseudotumor cerebri. Arch Neurol. 1985;42(2):150. doi: 10.1001/archneur.1985.04060020064017. [DOI] [PubMed] [Google Scholar]
  • 15.Bobak P, Friedman R, Brigell M, Goodwin J, Anderson R. Visual evoked potentials to multiple temporal frequencies. Use in the differential diagnosis of optic neuropathy. Arch Ophthalmol. 1988 Jul;106(7):936–940. doi: 10.1001/archopht.1988.01060140082029. [DOI] [PubMed] [Google Scholar]
  • 16.Verplanck M, Kaufman DI, Parsons T, Yedavally S, Kokinakis D. Electrophysiology versus psychophysics in the detection of visual loss in pseudotumor cerebri. Neurology. 1988 Nov;38(11):1789–1792. doi: 10.1212/wnl.38.11.1789. [DOI] [PubMed] [Google Scholar]
  • 17.Kesler A, Vakhapova V, Korczyn AD, Drory VE. Visual evoked potentials in idiopathic intracranial hypertension. Clin Neurol Neurosurg. 2009;111(5):433–436. doi: 10.1016/j.clineuro.2008.12.008. [DOI] [PubMed] [Google Scholar]
  • 18.Liasis A, Nischal KK, Walters B, et al. Monitoring visual function in children with syndromic craniosynostosis: a comparison of 3 methods. Arch Ophthalmol. 2006;124(8):1119. doi: 10.1001/archopht.124.8.1119. [DOI] [PubMed] [Google Scholar]
  • 19.Mursch K, Brockmann K, Lang JK, Markakis E, Behnke-Mursch J. Visually evoked potentials in 52 children requiring operative repair of craniosynostosis. Pediatr Neurosurg. 1998 Dec;29(6):320–323. doi: 10.1159/000028746. [DOI] [PubMed] [Google Scholar]
  • 20.Falsini B, Tamburrelli C, Porciatti V, Anile C, Porrello G, Mangiola N. Pattern electroretinograms and visual evoked potentials in idiopathic intracranial hypertension. Ophthalmologica. 1992;205(4):194–203. doi: 10.1159/000310341. [DOI] [PubMed] [Google Scholar]
  • 21.Onofrj M, Bodis-Wollner I, Mylin L. Visual evoked potential latencies in papilledema and hydrocephalus. Neuroophthalmol. 1981;2(2):85–92. [Google Scholar]
  • 22.Fraser CL, Klistorner A, Graham SL, Garrick R, Billson FA, Grigg JR. Multifocal visual evoked potential analysis of inflammatory or demyelinating optic neuritis. Ophthalmology. 2006 Feb;113(2):323e321–323.e322. doi: 10.1016/j.ophtha.2005.10.017. [DOI] [PubMed] [Google Scholar]
  • 23.Hood DC, Greenstein VC, Odel JG, et al. Visual field defects and multifocal visual evoked potentials: evidence of a linear relationship. Arch Ophthalmol. 2002 Dec;120(12):1672–1681. doi: 10.1001/archopht.120.12.1672. [DOI] [PubMed] [Google Scholar]
  • 24.Creel D. The Electroretinogram and Electro-oculogram: Clinical Applications. [Accessed January 5, 2016];Webvision. 2015 http://webvision.med.utah.edu/book/electrophysiology/the-electroretinogram-clinical-applications/
  • 25.Marmor MF, Fulton AB, Holder GE, Miyake Y, Brigell M, Bach M. ISCEV Standard for full-field clinical electroretinography (2008 update) Doc Ophthalmol. 2009 Feb;118(1):69–77. doi: 10.1007/s10633-008-9155-4. [DOI] [PubMed] [Google Scholar]
  • 26.Machida S. Clinical applications of the photopic negative response to optic nerve and retinal diseases. J Ophthalmol. 2012 doi: 10.1155/2012/397178. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Gotoh Y, Machida S, Tazawa Y. Selective loss of the photopic negative response in patients with optic nerve atrophy. Arch Ophthalmol. 2004;122(3):341. doi: 10.1001/archopht.122.3.341. [DOI] [PubMed] [Google Scholar]
  • 28.Viswanathan S, Frishman LJ, Robson JG, Walters JW. The photopic negative response of the flash electroretinogram in primary open angle glaucoma. Invest Ophthalmol Vis Sci. 2001;42(2):514–522. [PubMed] [Google Scholar]
  • 29.Rangaswamy NV, Frishman LJ, Dorotheo EU, Schiffman JS, Bahrani HM, Tang RA. Photopic ERGs in patients with optic neuropathies: comparison with primate ERGs after pharmacologic blockade of inner retina. Invest Ophthalmol Vis Sci. 2004 Oct;45(10):3827–3837. doi: 10.1167/iovs.04-0458. [DOI] [PubMed] [Google Scholar]
  • 30.Rangaswamy NV, Shirato S, Kaneko M, Digby BI, Robson JG, Frishman LJ. Effects of spectral characteristics of ganzfeld stimuli on the photopic negative response (PhNR) of the ERG. Invest Ophthalmol Vis Sci. 2007;48(10):4818–4828. doi: 10.1167/iovs.07-0218. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Moss HE, Park JC, McAnany JJ. The Photopic Negative Response in Idiopathic Intracranial Hypertension. Invest Ophthalmol Vis Sci. 2015 Jun;56(6):3709–3714. doi: 10.1167/iovs.15-16586. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Kato F, Miura G, Shirato S, Sato E, Yamamoto S. Correlation between N2 amplitude of multifocal ERGs and retinal sensitivity and retinal nerve fiber layer thickness in glaucomatous eyes. Doc Ophthalmol. 2015 Dec;131(3):197–206. doi: 10.1007/s10633-015-9519-5. [DOI] [PubMed] [Google Scholar]
  • 33.Bach M, Brigell MG, Hawlina M, et al. ISCEV standard for clinical pattern electroretinography (PERG): 2012 update. Doc Ophthalmol. 2013 Feb;126(1):1–7. doi: 10.1007/s10633-012-9353-y. [DOI] [PubMed] [Google Scholar]
  • 34.Holder G. The pattern electroretinogram. In: Fishman G, editor. Electrophysiologic testing in disorders of the retina, optic nerve and visual pathway, 2nd edition. San Francisco: Foundation of the American Academy of Ophthalmology; 2001. pp. 197–236. [Google Scholar]
  • 35.Afonso CL, Raza AS, Kreuz AC, et al. Relationship Between Pattern Electroretinogram, Frequency-Domain OCT, and Automated Perimetry in Chronic Papilledema From Pseudotumor Cerebri Syndrome. Invest Ophthalmol Vis Sci. 2015 Jun;56(6):3656–3665. doi: 10.1167/iovs.15-16768. [DOI] [PubMed] [Google Scholar]
  • 36.Kankipati L, Girkin CA, Gamlin PD. The post-illumination pupil response is reduced in glaucoma patients. Invest Ophthalmol Vis Sci. 2011;52(5):2287–2292. doi: 10.1167/iovs.10-6023. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Park JC, Moura AL, Raza AS, Rhee DW, Kardon RH, Hood DC. Toward a clinical protocol for assessing rod, cone, and melanopsin contributions to the human pupil response. Invest Ophthalmol Vis Sci. 2011;52(9):6624–6635. doi: 10.1167/iovs.11-7586. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Kardon R, Anderson SC, Damarjian TG, Grace EM, Stone E, Kawasaki A. Chromatic pupil responses: preferential activation of the melanopsin-mediated versus outer photoreceptor-mediated pupil light reflex. Ophthalmology. 2009;116(8):1564–1573. doi: 10.1016/j.ophtha.2009.02.007. [DOI] [PubMed] [Google Scholar]
  • 39.Park JC, Moss HE, McAnany JJ. The Pupillary Light Reflex in Idiopathic Intracranial Hypertension. Invest Ophthalmol Vis Sci. 2016 Jan 1;57(1):23–29. doi: 10.1167/iovs.15-18181. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES