Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2018 Oct 1.
Published in final edited form as: Magn Reson Med. 2016 Oct 31;78(4):1558–1565. doi: 10.1002/mrm.26535

Local SAR near deep brain stimulation (DBS) electrodes at 64 MHz and 127 MHz: A simulation study of the effect of extracranial loops

Laleh Golestanirad a,b, Leonardo M Angelone c, Maria Ida Iacono c, Husam Katnani b,d, Lawrence L Wald a,b, Giorgio Bonmassar a,b
PMCID: PMC5411348  NIHMSID: NIHMS823622  PMID: 27797157

Abstract

Purpose

Magnetic resonance imaging (MRI) may cause brain tissue around deep brain stimulation (DBS) electrodes to become excessively hot, causing lesions. The presence of extracranial loops in the DBS lead trajectory has been shown to affect the specific absorption rate (SAR) of the radiofrequency energy at the electrode tip but experimental studies have reported controversial results. The goal of this study was to perform a systematic numerical study to provide a better understanding of the effects of extracranial loops in DBS leads on the local SAR during MRI at 64 MHz and 127 MHz.

Methods

A total of 160 numerical simulations were performed on patient-derived data, where relevant factors including lead length and trajectory, loop location and topology, as well as frequency of MRI RF transmitter were assessed.

Results

Overall, the presence of extracranial loops reduced the local SAR in the tissue around the DBS tip compared to straight trajectories with the same length. SAR reduction was significantly larger at 127 MHz compared to 64 MHz. SAR reduction was significantly more sensitive to variable loop parameters (e.g., topology and location) at 127 MHz compared to 64 MHz.

Conclusion

There might exist lead management strategies that significantly reduce risks of 3.0 T MRI for DBS patients.

Keywords: atlas-based segmentation, medical implant, RF heating, Finite Element Method (FEM), safety

Introduction

Many patients with deep brain stimulation (DBS) implants will benefit from regular MRI examinations throughout the course of their lives, as MRI is often the standard diagnostic tool of choice for monitoring structural changes in the brain. Moreover, multi-modal studies that utilize functional MRI in combination with DBS have proven to be advantageous in clinical settings to help better understanding the therapeutic effects of electrical stimulation on DBS target structures (1-3). Today however, high-field MRI (>1.5 T) is largely inaccessible to patients with DBS due to safety concerns (4), and postoperative MRI at 1.5 T is practiced in only few centers and under strict safety guidelines (B1+rms fields <2.0 μT or whole-head averaged SAR < 0.1W/kg) that limit image quality (5,6).

As DBS patients become more likely to need MRI over the course of their lives (7), an important safety concern is the so-called “antenna effect” of the DBS leads. Here, the RF fields of the MRI transmitter couple with the long conductive leads and cause the specific absorption rate (SAR) of the radiofrequency (RF) energy to significantly amplify near the implant's tip (8,9). Injuries reported when FDA-approved guidelines were not followed have underscored these safety concerns (10,11).

A major hurdle in devising a systematic methodology that consistently and reliably estimates the implant-induced heating is that the problem has a large parameter space with many interacting factors. The coupling between MR-generated RF fields and DBS leads depends on the frequency, geometry, and electrical implementation of the transmit/receive RF coils, the patient's position inside the magnet (i.e., imaging landmark), the location of implant within the head (12), the lead trajectory (13-16), as well as implant materials (17) and design (18). Accordingly, complex computational models are used by device manufacturers and researchers to evaluate the safety of patients with DBS undergoing MRI (19). Such models may need to include detailed description of RF coils, head model, and implant geometry, therefore requiring considerable computational time and resources (20). One particular difficulty is to extract realistic DBS lead trajectories from post-operative computed tomography (CT) images that are routinely administered for electrode localization (21,22). In many practical cases surgeons choose to keep the DBS extension connector at the level of cranium to avoid positioning at the neck. As a result the lead is looped several times, making segmentation of the trajectory very challenging because of overlapping sections that cannot be easily distinguished on CT images (see Figure 1).

Figure.1.

Figure.1

Top row: Five head models with different DBS lead trajectories. Model A was directly constructed from intra-operative CT images shown in the bottom row. The end of the lead was extended behind the ear, similar to what happens in conventional DBS surgical approach, to model the connection of the lead to the pulse generator extension. Model B is similar to model A in all aspects, except for the loop which was morphed to a concentric shape. Model C is derived from model B by cutting the loop and moving it on top of the head close to the surgical bur hole. The loop's surface and topology were kept unchanged, but its orientation was adjusted to follow the skull's curvature. Model D was constructed by cutting out the loop and leaving the lead with the shorter overall length of ~39 cm. Model E was based on model D, except that the lead was further extended to reach the overall length of ~50 cm same as in models A-C. Bottom row: Steps of DBS lead segmentation. 3D view of intra-operative CT images of a patient with the implant is shown. Image contrast threshold and manual refinements were applied to recover overlapping loop segments.

The effect of extracranial loops on the MRI-generated heating of the tissue around DBS electrode contacts is not well understood and reports are controversial. Baker et al (16) observed a significant reduction in temperature rise around the tip of an implanted DBS device in a gel phantom during MRI at 1.5 T and 3.0 T, when the lead formed small concentric loops at the level of phantom cranium. In contrary, Shrivastava et al (13,14) predicted a significant increase in the temperature rise at electrode contacts of a DBS lead implanted in a porcine head during MRI at 3.0 T and 7.0 T when extracranial loops were present. Part of the discrepancy is due to the heterogeneity of experimental settings. Indeed, performing experimental studies that control a plurality of factors that affect the heating of implanted leads is substantially cumbersome as underscored in literature (12,23). Simulation studies, on the other hand, provide a great opportunity to study heating mechanism as they allow for a systematic and controlled way of changing each variable to better study its effect differentially, or in conjunction with other variables.

The goal of the present work was to apply a controlled simulation study and assess the presence of extracranial loops on the local SAR amplification at and around DBS electrode contacts during MRI at 64 MHz and 127 MHz. A total of 160 numerical simulations were performed on patient-derived data, where fine features of DBS lead including electrode contacts, insulation, core, and interconnections were represented. To explore the effect of transmit field frequency, lead trajectories with and without extracranial loops were simulated inside 64 MHz and 127 MHz RF coils. Each simulation scenario was repeated at 15 different RF coil-head orientations to account for variations across different coil systems. Both figure-8 and concentric loop topologies were studied at each frequency. Two different loop locations were investigated by positioning the loops at the side and top of the head. Finally, for lead trajectories without loops, additional simulations were performed with extended leads to control for the effect of lead length.

Methods

Secondary use of patient data for modeling and simulations was approved by Massachusetts General Hospital Internal Review Board (IRB). Finite Element Method (FEM) simulations were performed on a homogenous head model which was built by combining different tissue classes of a multimodal imaging-based detailed anatomical (MIDA) model of human head and neck, reported elsewhere (24,25). Post-operative computer tomography (CT) images of a patient operated at Massachusetts General Hospital were used to extract the DBS lead trajectory. The lead trajectory was manually segmented from CT images and co-registered to the MIDA head and neck model. Co-registration of patient's CT data and MIDA's MRI data was performed using a non-linear approach based on free-form deformations (FFDs) and normalized mutual information (NMI). The segmentation of the lead was then projected onto the MIDA model using the deformation vector field resulting from the co-registration.

DBS electrodes composed of four cylindrical contacts (outer diameter=1.27 mm, thickness=150 μm, height=1.5 mm) were constructed at the tip of the lead and were connected through a solid straight central core (diameter=260 μm). Electrode contacts were 0.5 mm apart, similar to those reported in (26). The electrodes and the core were made of 90%:10% (Pt:Ir) platinum-iridium (σ = 4 × 106S/m) and were separated by an insulation made of polyurethane (σ = 10−10S/m, εr = 3.5). Details of electrode contacts and their interconnections are given in Figure 2.

Figure .2.

Figure .2

Top left: FEM setup showing the head model and implant inside the RF coil. Coil rotation angle θ and feed positions are illustrated. Top right: Details of DBS lead model including electrode contacts, core and insulation and a sample of FEM adaptive mesh after convergence. Mesh was inspected visually for all simulations to assure fine details of DBS lead were properly captured. Bottom: Mesh statistics for the DBS lead, insulation and head model in a representative simulation (Model A, θ = 0°). Mesh statistics were similar across simulations.

RF coil models

MRI head coils were modeled as low-pass 16-rung birdcages with 35 cm diameter and 26 cm length. Commercially available software ANSYS Electronics Desktop 16.2 (ANSYS, Canonsburg, PA, USA) was used to perform a combined FEM-circuit analysis as described in previous studies (27,28). In the first step, the multi-port scattering matrix of the RF coil in the presence of the head model and implant was calculated in ANSYS HFSS with tuning and matching circuit components substituted with equivalent lumped ports. The multiport S-parameter matrix was then exported to ANSYS Designer to be tuned and matched at 127 MHz (3.0 T) and 64 MHz (1.5 T). The updated port voltage and current values were then pushed back to ANSYS HFSS to calculate electric and magnetic field distributions. A detailed description of this procedure can be found in (27). This method has shown to give an accurate estimation of magnitude of field values in phantom experiments (29).

DBS lead trajectories

Five different DBS lead trajectories were studied with details given in Figure 1. First, a realistic patient-derived ~50 cm lead was constructed from CT images, consisting of a figure-8 loop on the side of the head (model A). To study the effect of loop topology, the model was morphed to form a concentric loop, with the rest of the trajectory kept unchanged and the overall length preserved (model B). To investigate the effect of loop's position, another model was built where the loop was moved to the top of the head close to the surgical bur hole. The loop's surface and topology were kept unchanged, but its orientation was adjusted to follow the skull's curvature (model C). Finally, two models were constructed in which loops were eliminated. In one, the loop was cut out with the rest of the path kept the same, resulting in a shorter overall length of ~39 cm (model D). In the other, the lead was extended at the neck level to compensate for the shortening of the path (model E). A summary of model characteristics is given below.

  • Model A: figure-8 loop on side of the head, lead ~50 cm

  • Model B: concentric loop on side of the head, lead ~50 cm

  • Model C: concentricl loop on top of the head, lead ~50 cm

  • Model D: no loop, lead ~39 cm

  • Model E: no loop, lead ~50 cm

Each head and lead model was simulated in both 64 MHz and 127 MHz quadrature head coils. Coils were excited at two rungs that were 90° apart in position and signal phase (PI and PQ in Figure. 2). Excitation ports may have arbitrary position from one MRI site to another depending on the installation process. To account for the variability in the relative orientation of the head and RF coils, coils were rotated around each head model with 22.5° increments (a total of 15 coil positions). A reference head model with no lead was also simulated for the purpose of field normalization (not shown in the figure).

Numerical convergence

At the start of each simulation ANSYS HFSS was set to follow an adaptive mesh scheme. The algorithm started with a user-controlled initial tetrahedral mesh which forced a fine resolution on the DBS lead (maximum tetrahedron edge < 0.5 mm). The adaptive algorithm started by refining the mesh by 30% between each two iterative simulations. At each step, the maximum change in the magnitude of S-parameters, ΔS, was defined as ΔS=Maxij|SNij-SN-1ij|, where i and j coveres all matrix entries and N represents the iteration number. The adaptive simulation continued until the threshold of ΔS<0.01 was reached. All simulations converged after N=4 adaptive passes. Once convergence was reached, the mesh was visually inspected to assure that fine details of DBS leads were properly captured. Details of mesh statistics for a typical simulation (model A, feed at θ = 0°) is given in Figure 2.

SAR and B1+ field calculations

The total power absorbed in the head was calculated by computing the volume integral of the loss density in the head. The whole-head-averaged SAR (whSAR) was then calculated by dividing this total absorbed power by the head mass:

whSAR=12∫∫∫HeadVolumeσE2headmass [1]

where |E| is the magnitude of electric field phasor and σ is the conductivity of the tissue. The overall mass of the head used in the calculation was ~ 5.1 kg. 1g-averaged SAR was calculated based on IEEE STD P1528.4 recommendation (30) using the built-in SAR calculator module in ANSYS HFSS 16.2. The maximum 1g-averaged SAR (referred to as MaxSAR1g) was then calculated as the maximum of the 1g-averaged SAR in a 30×30×30 mm3 cubic volume that encompassed all four electrode contacts.

The counter-clockwise rotating component of the RF magnetic field was computed as B1+=0.5(B1x+jB1y) (31). For comparison, the spatial root mean square (rms) of the B1+ field was recorded from an axial reference plane passing through the head, ~7cm from top of the head. The magnitude of the B1+ field phasor was sampled with a 1 mm ×1 mm resolution from this axial plane and samples were used to calculate the B1+rms as:

B1+rms=NB1N+2N [2]

where B1N+ is the magnitude of the Nth sample and the summation is over all samples. MaxSAR1g and whSAR of models A-E are given in Table S1 in the Supporting Material for both 64 MHz and 127 MHz coils and at all 15 feed positions.

Results

Table S1 in Supporting Material gives the values of MaxSAR1g and whSAR for models A-E at 64 MHz and 127 MHz for different coil orientations. At each field strengths, the reference B1+ rms (B1+ref rms) was calculated as the spatial rms of B1+ field recorded from the reference plane, when the coil operated at its default position (θ = 0°), and generated a whSAR of 3.2 W/kg inside a head model with no implant. In all other simulations, the coil input power was adjusted to generate the same B1+rms on the reference plane as B1+ref rms which equaled to 6.7 μT at 64 MHz and 3.0 μT at 127 MHz.

Effect of extracranial loops on the MaxSAR1g at 127 MHz

Presence of extracranial loops, either concentric (Models B and C) or figure-8 (Model A), significantly reduced the SAR amplification at DBS electrode contacts at 127 MHz (75%-99%). Results were consistent across different coil-head orientations and when compared against both leads with controlled length (Model E) and un-controlled length (Model D).

The extent of SAR reduction however, varied with loop topology and location. Maximum SAR reduction was observed in Model B (2.1±0.4 W/kg), followed by Model A (18.7±4.9 W/kg) and Model C (46.2±16.8 W/kg) with respect to both Models D (207.3±68.7 W/kg) and E (317.0±100.5 W/kg). Interestingly, MaxSAR1g values of Model B were even smaller than SAR values at the periphery of the head far from the electrode (see Figure 5).

Figure. 5.

Figure. 5

Maps of 1g-averaged SAR for Models B and D at 64 MHz and 127 MHz. Coils were operated at their default feed position (θ = 0°) and the input power was normalized to produce B1+rms=B1+refrms.

Effect of extracranial loops on the MaxSAR1g at 64 MHz

When controlled for the total length, the presence of extracranial loops consistently reduced the MaxSAR1g at 64 MHz. Specifically, a 3 fold reduction in MaxSAR1g was observed at all 15 feed orientations for both figure-8 loop (Model A) and concentric loops (Models B and C) with respect to the lead trajectory with the same length but no loop (Model E). However, compared to the lead trajectory with no loop and a shorter length (Model D), concentric loops only introduced a modest reduction in the MaxSAR1g (5% to 15%) and figure-8 loop did not consistently reduced the SAR (−6% to 2%).

Comparative effect of the lead trajectory at 64 and 127 MHz

At both frequencies and for all 15 feed positions, the 50 cm lead with no loop (Model E) generated a higher MaxSAR1g than the 39 cm lead with no loop (Model D). The difference however, was significantly larger at 64 MHz (1586.8±254.2 W/kg vs. 589.4±98.6 W/kg) than at 127 MHz (317.0±100.5 W/kg vs. 207.3±65.7 W/kg). Figures 3 and 4 give means and standard deviations of MaxSAR1g values (calculated over all coil orientations) for 64 MHz and 127 MHz coils, when the input power was adjusted to generate the same B1+ rms as the B1+ref rms or B1+ rms=1μT, respectively. As it can be observed from Figures 3 and 4, presence of loops had a significantly larger SAR reduction effect at 127 MHz compared to 64 MHz. For example, concentric loop on the side (Model B) introduced a ~144 fold SAR reduction factor compared to Model E at 127 MHz, whereas it only reduced the SAR by a factor of 3 at 64 MHz.

Figure.3.

Figure.3

MaxSAR values for models A-E inside 64 MHz and 127 MHz coils. Input power adjusted to produce a B1+rms= B1+refrms. At each transmit frequency, B1+refrms was calculated as the rms of the B1+ field recorded from an axial plane at ~7 cm from top of the head, when the coil was operated at its default feed position (θ = 0°) and generated a whSAR of 3.2 W/kg inside a head model with no implant. B1+refrms was ~7.65 μT at 64 MHz and ~3.0 μT at 127 MHz.

Figure. 4.

Figure. 4

MaxSAR values for models A-E inside 64 MHz and 127 MHz coils. Input power adjusted to produce B1+rms = 1μT.

Interestingly, the differential effect of loop's topology and location was also more significant at 127 MHz than at 64 MHz. Two-tail t-test comparing means of Models A and B (effect of loop's topology) had a p-value =0.04 at 64 MHz and p-value=1.3e-9 at 127 MHz. Two-tail t-test comparing means of Models B and C (effect of loop's location) had a non-significant p-value at 64 MHz (p-value =0.2) but a p-value =3.9e-8 at 127 MHz.

Discussion and Conclusion

This study provides novel results of a systematic evaluation of the effect of extracranial loops on the SAR amplification around DBS electrode contacts during MRI at 64 MHz and 127 MHz. To date, reports of temperature rise at and around DBS electrodes have drawn controversial conclusions as to the effect of extracranial loops (13,14,16). The discrepancy is partially due to the heterogeneity of the experimental settings and the methodology applied to measure or predict the temperature. For example Baker (16) used fluoroptic probes to directly measure the temperature rise at the distal DBS electrode contact immersed in a phantom gel. Shrivastava (14) on the other hand, applied in-vivo MR thermometry to estimate the temperature rise a few millimeters away from the tip, and used the estimated temperatures to calibrate a numerical thermal equation solver to predict the maximum heating. Apart from obvious differences in the experimental setup (gel phantom vs. animal model) and the methodology (direct measurement vs. numerical interpolation based on MR thermometry) these studies did not control details of the lead trajectory or position and orientation of loops with respect to the RF coil. Such details however, could significantly affect the heating as highlighted in previous reports (12,32).

Performing experimental studies that control a plurality of interacting factors that affect the heating of implants is substantially cumbersome. Simulation studies on the other hand, allow a more affordable systematic evaluation of such factors including lead length and trajectory, loops position and topology, and RF coil's geometry and position. This work was a first attempt to provide better insight into the effect of extracranial loops on the local SAR at the tip of a DBS lead, by carefully crafting models that were identical in all aspects except the one under study.

Overall, we observed that for the specific length of the lead under study (~50 cm) and specific locations and orientations of loops (side and top of the head), the presence of extracranial loops reduced the local SAR at the DBS tip compared to straight trajectories with the same length. SAR reduction however, was significantly larger at 127 MHz compared to 64 MHz. Specifically, at 127 MHz there existed lead configurations (concentric loops on the side of head) that reduced the MaxSAR1g down to background levels observed at the periphery of the head. We also observed that SAR reduction was significantly more sensitive to differential loop parameters (e.g., topology and location) at 127 MHz compared to 64 MHz.

The mechanism by which extracranial loops affect the SAR is not completely understood. Baker (16) suggested two possible scenarios. First, the loops may add a reactive inductance that produces a source of impedance near the end of the electrode, reducing the induced current in the wire. This mechanism is similar to the reduction of RF-induced temperature rise observed for example by Ladd and Quick in catheter by addition of chokes (33) or by coiling the wires (34,35). Another possibility is that the voltage induced in the loops because of coupling with the magnetic field (from Faraday's law) cancels out the voltage generated by coupling of the straight part of the lead with the RF electric field. As the Faraday coupling effect scales up with the rate of change of magnetic flux, we expected that the former mechanism (inductive effect) be more dominant at 64 MHz, while the latter (Faraday coupling) be dominant at 127 MHz. This was in line with our observation of relative insensitivity of SAR reduction to the location and topology of the loop at 64 MHz vs. its high dependency on these parameters at 127 MHz. Specifically at 127 MHz, the loop on the side of the head with a surface exposed to a larger normal magnetic flux reduced the SAR much more effectively than the loop on top of the head where the B1+ field was more tangential to its surface.

Our study also underscores the fact that 3.0 T MRI should not be considered, per se, more dangerous for patients with conductive leads than 1.5 T MRI. Indeed, our results showed that for a B1+ rms=1μT the majority of lead configurations introduced less SAR amplification at DBS electrodes at 127 MHz compared to 64 MHz. It should be noted however, that the sensitivity and variation of the results with frequency, lead length, as well as loop location, orientation, and topology warrant the need for a multivariate analysis of all relevant factors before any definitive conclusion on the safety-enhancing effect of extracranial loops can be drawn. There are specifically two important limitations of the current study. First, in all our simulations the central core of the DBS lead was modelled as a single straight solid wire connecting all four electrode contacts. In reality, DBS electrode contacts are connected through tightly-coiled interleaved spiral wires, each connected to one electrode contact, and isolated from each other (36). Similar technique is applied in manufacturing of cardiac pacemakers and defibrillator leads (37,38). It is conceivable that the presence of these small coils, adding to the overall electric length of the lead, interferes with the hypothesized SAR-reduction mechanism of large extracranial loops. A more detailed model of DBS lead interconnection wires is needed to investigate this issue. Second, we report results of numerical simulation of SAR distribution and have not performed measurements to calculate the associated temperature rises. Although SAR has been routinely used and an indirect quantitative measurement for clinical MRI procedures when conductive implants are present (8,39), local SAR values do not linearly translate to temperature rise as factors such as thermal conduction and perfusion should be taken into account (40). Additionally, even if the SAR near the lead tip is reduced, there may still be high values in other regions of the head. Consequently, temperature calculations/measurements over the entire head should be considered instead of monitoring SAR alone near the tip.

Finally, this study highlights the prospect of obtaining a significant reduction in the SAR at 127 MHz by administering lead management strategies. More numerical simulations validated with experimental measurements and inclusive of uncertainty analysis are necessary to verify and optimize such strategies, which could provide a useful surgical guideline to the clinical community.

Supplementary Material

Supp Table S1

Acknowledgement

Grant Support:

Preparation of this paper was supported by grants K99EB021320, R01EB006847, P41EB015896 and R21NS074395 from the National Institutes of Health.

Footnotes

Disclaimer: The mention of commercial products, their sources, or their use in connection with material reported herein is not to be construed as either an actual or implied endorsement of such products by the Department of Health and Human Services.

References

  • 1.Arantes PR, Cardoso EF, Barreiros MA, Teixeira MJ, Goncalves MR, Barbosa ER, Sukwinder SS, Leite CC, Amaro E., Jr Performing functional magnetic resonance imaging in patients with Parkinson's disease treated with deep brain stimulation. Mov Disord. 2006;21(8):1154–62. doi: 10.1002/mds.20912. [DOI] [PubMed] [Google Scholar]
  • 2.Jech R, Urgosik D, Tintera J, Nebuzelsky A, Krasenski J, Liscak R, Roth J, Ruzicka E. Functional magnetic resonance imaging during deep brain stimulation: a pilot study in four patients with Parkinson's disease. Movement disorders. 2001;16(6):1126–1132. doi: 10.1002/mds.1217. [DOI] [PubMed] [Google Scholar]
  • 3.Kahan J, Urner M, Moran R, Flandin G, Marreiros A, Mancini L, White M, Thornton J, Yousry T, Zrinzo L. Resting state functional MRI in Parkinson's disease: the impact of deep brain stimulation on “effective” connectivity. Brain. 137(4):1130–1144. doi: 10.1093/brain/awu027. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Medtronic [12/09/2015];MRI Guidelines for Medtronic Deep Brain Stimulation Systems. ( http://manuals.medtronic.com/wcm/groups/mdtcom_sg/@emanuals/@era/@neuro/documents/documents/contrib_228155.pdf). 2015.
  • 5.Larson PS, Richardson RM, Starr PA, Martin AJ. Magnetic resonance imaging of implanted deep brain stimulators: experience in a large series. Stereotactic and functional neurosurgery. 2008;86(2):92–100. doi: 10.1159/000112430. [DOI] [PubMed] [Google Scholar]
  • 6.Tagliati M, Jankovic J, Pagan F, Susatia F, Isaias IU, Okun MS. Safety of MRI in patients with implanted deep brain stimulation devices. Neuroimage. 2009;47:T53–T57. doi: 10.1016/j.neuroimage.2009.04.044. [DOI] [PubMed] [Google Scholar]
  • 7.Falowski S, Safriel Y, Ryan M, L. H. The need for magnetic resonance (MR) imaging in the United States (US) deep brain stimulation (DBS) population. Las Vegas, NV.: 2014. [Google Scholar]
  • 8.Rezai AR, Phillips M, Baker KB, Sharan AD, Nyenhuis J, Tkach J, Henderson J, Shellock FG. Neurostimulation system used for deep brain stimulation (DBS): MR safety issues and implications of failing to follow safety recommendations. Investigative radiology. 2004;39(5):300–303. doi: 10.1097/01.rli.0000124940.02340.ab. [DOI] [PubMed] [Google Scholar]
  • 9.Rezai AR, Finelli D, Nyenhuis JA, Hrdlicka G, Tkach J, Sharan A, Rugieri P, Stypulkowski PH, Shellock FG. Neurostimulation systems for deep brain stimulation: In vitro evaluation of magnetic resonance imaging–related heating at 1.5 tesla. Journal of Magnetic Resonance Imaging. 2002;15(3):241–250. doi: 10.1002/jmri.10069. [DOI] [PubMed] [Google Scholar]
  • 10.Spiegel J, Fuss G, Backens M, Reith W, Magnus T, Becker G, Moringlane JR, Dillmann U. Transient dystonia following magnetic resonance imaging in a patient with deep brain stimulation electrodes for the treatment of Parkinson disease. Case report. J Neurosurg. 2003;99(4):772–4. doi: 10.3171/jns.2003.99.4.0772. [DOI] [PubMed] [Google Scholar]
  • 11.Henderson JM, Tkach J, Phillips M, Baker K, Shellock FG, Rezai AR. Permanent neurological deficit related to magnetic resonance imaging in a patient with implanted deep brain stimulation electrodes for Parkinson's disease: case report. Neurosurgery. 2005;57(5):E1063. doi: 10.1227/01.neu.0000180810.16964.3e. discussion E1063. [DOI] [PubMed] [Google Scholar]
  • 12.Mattei E, Triventi M, Calcagnini G, Censi F, Kainz W, Mendoza G, Bassen HI, Bartolini P. Complexity of MRI induced heating on metallic leads: experimental measurements of 374 configurations. Biomedical engineering online. 2008;7(1):11. doi: 10.1186/1475-925X-7-11. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Shrivastava D, Tian J, Abosch A, Vaughan J. Radio-Frequency Heating at Deep Brain Stimulation Lead Electrodes due to Imaging with Head Coils in 3 T and 7T.
  • 14.Shrivastava D, Abosch A, Hughes J, Goerke U, DelaBarre L, Visaria R, Harel N, Vaughan JT. Heating induced near deep brain stimulation lead electrodes during magnetic resonance imaging with a 3 T transceive volume head coil. Physics in medicine and biology. 2012;57(17):5651. doi: 10.1088/0031-9155/57/17/5651. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Shrivastava D, Abosch A, Hanson T, Tian J, Gupte A, Iaizzo PA, Vaughan JT. Effect of the extracranial deep brain stimulation lead on radiofrequency heating at 9.4 Tesla (400.2 MHz). Journal of Magnetic Resonance Imaging. 2010;32(3):600–607. doi: 10.1002/jmri.22292. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Baker KB, Tkach J, Hall JD, Nyenhuis JA, Shellock FG, Rezai AR. Reduction of magnetic resonance imaging-related heating in deep brain stimulation leads using a lead management device. Neurosurgery. 2005;57(4):392–397. doi: 10.1227/01.neu.0000176877.26994.0c. [DOI] [PubMed] [Google Scholar]
  • 17.Serano P, Angelone LM, Katnani H, Eskandar E, Bonmassar G. A Novel Brain Stimulation Technology Provides Compatibility with MRI. Scientific reports. 2015:5. doi: 10.1038/srep09805. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Cabot E, Lloyd T, Christ A, Kainz W, Douglas M, Stenzel G, Wedan S, Kuster N. Evaluation of the RF heating of a generic deep brain stimulator exposed in 1.5 T magnetic resonance scanners. Bioelectromagnetics. 2013;34(2):104–113. doi: 10.1002/bem.21745. [DOI] [PubMed] [Google Scholar]
  • 19.ISO/TS-10974. Technical Specifications – Assessment of the safety of magnetic resonance imaging for patients with an active implantable medical device. 2012.
  • 20.U.S. Food and Drug Administration [12/15/2015];Reporting of Computational Modeling Studies in Medical Device Submissions - Draft Guidance for Industry and Food and Drug Administration Staff. http://www.fda.gov/RegulatoryInformation/Guidances/ucm371016.htm.
  • 21.Thani NB, Bala A, Swann GB, Lind CR. Accuracy of postoperative computed tomography and magnetic resonance image fusion for assessing deep brain stimulation electrodes. Neurosurgery. 2011;69(1):207–214. doi: 10.1227/NEU.0b013e318218c7ae. [DOI] [PubMed] [Google Scholar]
  • 22.Lunsford L, Latchaw R, Vries J. Stereotactic implantation of deep brain electrodes using computed tomography. Neurosurgery. 1983;13(3):280–286. doi: 10.1227/00006123-198309000-00012. [DOI] [PubMed] [Google Scholar]
  • 23.Langman DA, Goldberg IB, Finn JP, Ennis DB. Pacemaker lead tip heating in abandoned and pacemaker-attached leads at 1.5 tesla MRI. Journal of Magnetic Resonance Imaging. 2011;33(2):426–431. doi: 10.1002/jmri.22463. [DOI] [PubMed] [Google Scholar]
  • 24.Iacono MI, Neufeld E, Akinnagbe E, Bower K, Wolf J, Oikonomidis IV, Sharma D, Lloyd B, Wilm BJ, Wyss M. MIDA: a multimodal imaging-based detailed anatomical model of the human head and neck. PloS one. 2015;10(4):e0124126. doi: 10.1371/journal.pone.0124126. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Iacono MI, Makris N, Mainardi L, Angelone LM, Bonmassar G. MRI-based multiscale model for electromagnetic analysis in the human head with implanted DBS. Computational and mathematical methods in medicine. 2013:2013. doi: 10.1155/2013/694171. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Elwassif MM, Kong Q, Vazquez M, Bikson M. Bio-heat transfer model of deep brain stimulation-induced temperature changes. Journal of neural engineering. 2006;3(4):306. doi: 10.1088/1741-2560/3/4/008. [DOI] [PubMed] [Google Scholar]
  • 27.Kozlov M, Turner R. Fast MRI coil analysis based on 3-D electromagnetic and RF circuit co-simulation. Journal of Magnetic Resonance. 2009;200(1):147–152. doi: 10.1016/j.jmr.2009.06.005. [DOI] [PubMed] [Google Scholar]
  • 28.Lemdiasov RA, Obi AA, Ludwig R. A numerical postprocessing procedure for analyzing radio frequency MRI coils. Concepts in Magnetic Resonance Part A. 2011;38(4):133–147. [Google Scholar]
  • 29.Golestnirad L, Keil B, Angelone LM, Bonmassar G, Mareyam A, Wald LL. Feasibility of using linearly polarized rotating birdcage transmitters and closefitting receive arrays in MRI to reduce SAR in the icinity of deep brain simulation implants. Magnetic resonance in medicine. 2016 doi: 10.1002/mrm.26220. In Press. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.IEEE-P1528.4™/D1.0. Recommended Practice for Determining the Peak Spatial Average Specific Absorption Rate (SAR) in the Human Body from Wireless Communications Devices, 30 MHz - 6 GHz: Requirements for Using the Finite-Element Method for SAR Calculations, specifically involving Vehicle Mounted Antennas and Personal Wireless Devices. 2014.
  • 31.Hoult D. The principle of reciprocity in signal strength calculations—a mathematical guide. Concepts in Magnetic Resonance. 2000;12(4):173–187. [Google Scholar]
  • 32.Calcagnini G, Triventi M, Censi F, Mattei E, Bartolini P, Kainz W, Bassen HI. In vitro investigation of pacemaker lead heating induced by magnetic resonance imaging: role of implant geometry. Journal of Magnetic Resonance Imaging. 2008;28(4):879–886. doi: 10.1002/jmri.21536. [DOI] [PubMed] [Google Scholar]
  • 33.Ladd ME, Quick HH. Reduction of resonant RF heating in intravascular catheters using coaxial chokes. Magnetic resonance in medicine. 2000;43(4):615–619. doi: 10.1002/(sici)1522-2594(200004)43:4<615::aid-mrm19>3.0.co;2-b. [DOI] [PubMed] [Google Scholar]
  • 34.Gray RW, Helfer JL. Magnetic resonance imaging interference immune device. 2008 US Patent 8041433.
  • 35.Bottomley PA, Kumar A, Edelstein WA, Allen JM, Karmarkar PV. Designing passive MRI-safe implantable conducting leads with electrodes. Medical physics. 2010;37(7):3828–3843. doi: 10.1118/1.3439590. [DOI] [PubMed] [Google Scholar]
  • 36.Wahlstrand CD, Hoegh TB, Hrdlicka GA, Cross TE, Jr, Olsen JM. Lead electrode for use in an MRI-safe implantable medical device. 2007 US Patent 20110112615.
  • 37.Perrey C, Clark BA, Foster AJ. Medical device lead including a unifilar coil with improved torque transmission capacity and reduced MRI heating. 2012 US Patent 8332050.
  • 38.Bulkes C, Denker S. MRI compatible implanted electronic medical lead. 2011 US Patent 20080033497.
  • 39.Shellock FG, Hatfield M, Simon BJ, Block S, Wamboldt J, Starewicz PM, Punchard WF. Implantable spinal fusion stimulator: assessment of MR safety and artifacts. Journal of Magnetic Resonance Imaging. 2000;12(2):214–223. doi: 10.1002/1522-2586(200008)12:2<214::aid-jmri2>3.0.co;2-k. [DOI] [PubMed] [Google Scholar]
  • 40.Winter L, Oberacker E, Ozerdem C, Ji Y, Knobelsdorff-Brenkenhoff Fv, Weidemann G, Ittermann B, Seifert F, Niendorf T. On the RF heating of coronary at 7.0 Tesla MRI. Magnetic resonance imaging. 2015;74:999–1010. doi: 10.1002/mrm.25483. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supp Table S1

RESOURCES