Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2018 Jun 1.
Published in final edited form as: Anal Chim Acta. 2017 Mar 27;970:23–29. doi: 10.1016/j.aca.2017.03.032

Pyrenyl Carbon Nanostructures for Ultrasensitive Measurements of Formaldehyde in Urine

Gayan Premaratne 1, Sabrina Farias 1, Sadagopan Krishnan 1
PMCID: PMC5443713  NIHMSID: NIHMS863828  PMID: 28433055

Abstract

Measurement of ultra-low (e.g., parts-per-billion) levels of small-molecule markers in body fluids (e.g., serum, urine, saliva) involves a considerable challenge in view of designing assay strategies with sensitivity and selectivity. Herein we report for the first time an amperometric nano-bioelectrode design that uniquely combines 1-pyrenebutyric acid units pi-pi stacked with carboxylated multiwalled carbon nanotubes on the surface of gold screen printed electrodes for covalent attachment of NAD+ dependent formaldehyde dehydrogenase (FDH). The designed enzyme bioelectrode offered 6 ppb formaldehyde detection in 10-times diluted urine with a wide dynamic range of 10 ppb to 10 ppm. Fourier transform infrared, Raman, and electrochemical impedance spectroscopic characterizations confirmed the successful design of the FDH bioelectrode. Flow injection analysis provided lower detection limit and greater affinity for formaldehyde (apparent KM 9.6 ± 1.2 ppm) when compared with stirred solution method (apparent KM 19.9 ± 4.6 ppm). Selectivity assays revealed that the bioelectrode was selective toward formaldehyde with a moderate cross-reactivity for acetaldehyde (~ 25%) and negligible cross-reactivity toward propanaldehyde, acetone, methanol, and ethanol. Formaldehyde is an indoor pollutant, and studies have indicated neurotoxic characteristics and systemic toxic effects of this compound upon chronic and high doses of exposure. Moreover, reported chromatography and mass spectrometry methods identified elevated urine formaldehyde levels in patients with bladder cancer, dementia, and early stages of cognitive impairments compared to healthy people. Results demonstrate that pyrenyl carbon nanostructures-based FDH bioelectrode design represents novelty and simplicity for enzyme-selective electrochemical quantitation of small 30 Da formaldehyde. Broader applicability of the presented approach for other small-molecule markers is feasible that requires only the design of appropriate marker-specific enzyme systems or receptor molecules.

Keywords: Nanotube-pyrene, Urine samples, Small molecules, Stirring vs. flow injection, Noninvasive analysis, Combined carboxylation

Graphical abstract

This is the first report of a pyrenyl carbon nanostructure based enzymatic bioelectrode for quantitation of formaldehyde marker in urine representing a non-invasive small-molecule assay methodology.

graphic file with name nihms863828u1.jpg

1. Introduction

Biomarkers are molecular indicators that play the crucial role in the diagnosis, prognosis, and theranostics of diseases such as cancer.1 As a result, bioanalytical methods that allow sensitive and selective measurements of biomarkers are significant for clinical applications and therapeutics development. In view of diagnostic challenges, the molecular size of a biomarker present in complex body fluid matrices inversely affects the detection sensitivity. This is because small molecules do not yield measurable assay signal changes compared to large biomolecules. Additionally, dilution of samples to minimize clinical matrix effects can further lower the biomarker concentration and its detection. False-positive signals, tedious extraction procedures of analytes from the matrices, simplicity of the assay and detection protocol, and selectivity are other related problems.

Despite the issues described, nanotechnology-based optical and electrochemical methods have allowed detection of large proteins, receptors, antibodies, DNA, and RNA biomarkers at clinically relevant pM to aM concentrations in body fluids.2 However, development of such analytical assays with simplicity, selectivity, sensitivity, and low-cost for detecting small-molecule markers of cancer and other diseases non-invasively remains to be a significant challenge. In particular, bioanalytical methods for detection of ultra-low ppb levels of formaldehyde (HCHO) in a complex clinical matrix (e.g., plasma, serum, urine) suffer by interferences from non-analyte components present in the matrix (non-specific signals), poor detection capability, and issues of enzyme instability and loss of enzyme activity. Furthermore, a non-invasive detection of biomarker levels is advantageous for routine diagnosis of these diseases at an early stage as it eliminates painful invasive procedures (e.g., biopsy, bronchoscopy), which are known to cause tissue and organ damage.3 Moreover, we can eliminate minimally invasive procedures, such as drawing of blood from patients for analysis.4 In this study, we demonstrate for the first time the design of pyrenyl carbon nanostructure modified electrodes for FDH immobilization and electrochemical quantitation of HCHO in a urine matrix with a 6 ppb detection limit.

Aldehydes, including HCHO, have received considerable attention as a key class of volatile organic compound markers that exhibit toxic effects in humans, and are suggested to be relevant for cancer and neural diseases.58 Many techniques have been developed to detect HCHO, including spectrophotometric,9 electrochemical,10,11 optical,12 electronic,13 and colorimetric14 methods. Sensitive detection of urine HCHO by chromatography-mass spectrometry methods is known. However, these methods involve expensive instrumentation, and time consuming laborious sample preparation, measurement, and analysis steps.5 In contrast, electrochemical methods are straightforward and cost effective.15,16 In particular, electrochemical enzymatic amperometric biosensors offer the advantages of direct and quick detection of analytes with simplicity, and by appropriate design of surface strategies, high sensitivity, robustness, and excellent reproducibility can be attained.

FDH (~ 170 kDa), an enzyme of the oxidoreductase family,17,18 is a homotetrameric enzyme containing 398 amino acids per subunit. Ali et al. developed a biosensor made from NAD+ and glutathione-dependent recombinant FDH immobilized on the surface of a Si/SiO2/Si3N4 structure.11 A mediator-based amperometric biosensor was constructed on disks of woven graphite gauze coated with NAD+ and glutathione-independent FDH from Hyphomicrobium zavarzinii strain ZV 58.19 Demkiv et al. modified platinized graphite electrodes with NAD+ and glutathione-dependent FDH isolated from a genetically engineered strain H. polymorpha. In that study, electron transfer between the enzyme and the electrode was established using low molecular weight redox mediators or positively charged cathodic electrodeposition paints modified with Os-bis-N,N-(2,2′-bipyridyl)-chloride ([Os(bpy)2Cl]).20 FDH from Pseudomonas sp. immobilized on mesoporous silica materials21 and multiwalled carbon nanotube-modified screen printed electrodes22 for HCHO detection in buffer solution have been developed.

Herein, we report for the first-time 1-pyrenebutyric acid (PBA) pi-pi stacked with carboxylated multiwalled carbon nanotubes (MWNTs) on the surface of the gold screen printed electrodes (AuSPEs) for covalent FDH immobilization offering highly sensitive and selective ultra-low detection HCHO concentration in a 10-times buffer diluted urine. We combined here the –COOH groups of MWNTs with the non-covalently attached PBA because this combination approach, discovered by us recently, offered 3-fold improved sensitivity for a serum insulin immunosensor compared to the use of only carboxylated MWNTs.23 HCHO was quantitated using a quinone compound (Q: 1, 2-naphthaquinone-4-sulfonic acid sodium salt) as an electron transfer mediator to obtain electrochemical signals in proportion to HCHO concentration. To our knowledge, this is the first report on pyrenyl carbon nanostructure based enzymatic bioelectrode for urine HCHO quantitation.

2. Experimental section

2.1 Materials and chemicals

Gold screen printed electrodes (AuSPEs) with a three-electrode cell configuration integrated on a ceramic substrate (Model: 250 BT, 4 mm diameter Au working electrode, Pt counter electrode, and Ag pseudo reference electrode) were purchased from DropSens Inc. (Spain). Multiwalled carboxylic acid functionalized carbon nanotubes (MWNTs, > 8% carboxylic acid functionalized, avg. diam. 9.5 nm, length 1.5 μm), β-Nicotinamide adenine dinucleotide sodium salt (NAD+), FDH from Pseudomonas sp., PBA, 1,2-naphthaquinone-4-sulfonic acid sodium salt (Q), HCHO solution (36.5 – 38% in water), 1-ethyl-3-[3-dimethylaminopropyl]carbodiimide hydrochloride (EDC), N-hydroxysuccinimide (NHS), and formaldehyde-2,4-dinitrophenylhydrazone analytical standard (HCHO-DNPH) for liquid chromatography were purchased from Sigma and were used as received. Vivaspin 6 ultrafiltration cartridges (molecular weight cut off: 3 kDa, GE Healthcare, Little Chalfont, Buckinghamshire, UK) were used for filtration of urine. All aqueous reagents were prepared in deionized water (DI H2O) using a Milli-DI water purification system (Millipore Ltd., Billerica, Massachusetts, USA). All other chemicals were high purity analytical grade.

Amperometric measurements were performed at room temperature (23°C) using a CHI 6017E electrochemical workstation coupled to a faraday cage and a picoamp booster (Austin, TX, USA). The flow-cell was connected to a syringe pump system (New Era Pump Systems Inc., NY, USA) and a sample injector valve (Rheodyne model 9725i PEEK injector, IDEX Health & Science LLC, CA, USA). The sample loop volume was 200 μL. Spectroscopic characterizations of the bioelectrode fabrication steps were carried out using Fourier transform infrared spectroscopy (FTIR, Thermo Scientific Nicolet iS50) in the attenuated total reflectance (ATR) mode. The bioelectrodes were placed directly on the ATR diamond crystal and 32 scans were taken and averaged to obtain a good signal-to-noise ratio. In addition, Raman spectroscopy (Nicolet NXR FT-Raman module, Nd:YVO4 laser, 1064 nm, 0.2 W, resolution 4 cm−1) and electrochemical impedance spectroscopy (Interface 1000 potentiostat/galvanostat/ZRA, Gamry instruments, Warminster, PA) were used for the characterization of the designed bioelectrode. Liquid chromatography-mass spectrometry (LC-MS) with an electrospray ionization source (Shimadzu LCMS-2010EV) was used for identification of HCHO-DNPH derivative, methanal 2,4-dinitrophenylhydrazone.

2.2 Surface modification of AuSPE

Briefly, on the surface of disposable AuSPEs, 5 μL dispersions of 2 mg mL−1 MWNTs in dimethylformamide (DMF) were added to cover the entire AuSPE surface and allowed to dry at room temperature. The electrodes were washed well with deionized H2O and dried under nitrogen prior to adding 3 μL of 3 mg mL−1 PBA solution in DMF. The electrodes were kept in a cold and moist environment for 1 h to allow the formation of strong pi-pi interactions between MWNTs and PBA. The electrodes then were washed well with deionized H2O and dried under nitrogen.

FDH enzyme was covalently linked via its free surface lysine residues (PDB: 4JLW, Figure S1) to the –COOH terminal groups of both the MWNT and PBA assembly by the established carbodiimide chemistry.24 To carry out this reaction, MWNT/PBA-modified AuSPEs were treated with 5 μL solution of a freshly prepared mixture of 0.35 M EDC and 0.1 M NHS for 15 min to activate the surface carboxylic acid groups of MWNTs and PBA into lysine (Lys) amine reactive N-succinimidyl esters. The electrodes were then washed with deionized H2O to remove unreacted reagents and subsequently introduced to 5 μL of 200 μg/mL FDH solution prepared in phosphate buffered saline (PBS) and incubated for 1 h to obtain covalently linked FDH films. The prepared electrodes were used immediately for measuring HCHO in urine.

2.3 Chronoamperometric detection of formaldehyde in urine samples

Urine samples were collected from a healthy male adult volunteer in a 50 mL sterile polyethylene vial. Samples were then immediately vortexed and divided into 10 mL aliquots and stored at −20°C for up to 30 days to avoid any storage loss until further use. Before use, the frozen samples were thawed at room temperature. Prior to spiking with formaldehyde, 6 mL aliquot of the urine sample was subjected to centrifugal filtration using Vivaspin-6 cartridges at a speed of 4000 rpm for 30 min at room temperature.25 Filtered urine samples were diluted 10-times in PBS. Various concentrations of formaldehyde were spiked in the 10-times diluted urine and were used for analysis.

FDH immobilized AuSPE/MWNT/PBA electrodes were attached to an in-house designed flow cell that was connected to a syringe pump and a sample injector (Figure 1A). The running buffer was made up of PBS containing 5 mM NAD+ and 1 mM Q. Various concentrations of HCHO prepared in 10-times diluted urine in PBS were injected via a manual injector valve onto the bioelectrode surface. Real-time measurements of the oxidation currents of QH2 formed from NADH as a result of FDH catalyzed HCHO oxidation were performed at an applied constant potential of + 0.35 V (Figure 1B).21,26 The current versus HCHO concentration was plotted and the resulting curve was fit by the Michaelis-Menten non-linear regression equation available with the K-graph software used.

Figure 1.

Figure 1

(A) Schematic of the microfluidics system used in this study. (B) Fabrication steps of the AuSPEs with FDH and the reaction sequence for catalyzing HCHO and detection by flow injection or stirred solution amperometry.

3. Results and Discussion

3.1 Characterization of the modified electrodes

Figure 2 shows the baseline corrected FTIR-ATR results for the modified bioelectrode surface that confirmed the covalent immobilization of FDH onto the –COOH groups of AuSPE/MWNT/PBA electrodes. A strong vibrational peak at 1714 cm−1 indicated the presence of carbonyl stretching from the –COOH groups of MWNTs. Pi-pi stacking of PBA with MWNTs red shifted the C=O stretching vibration to 1698 cm−1, which has been attributed to the partial electron transfer from PBA to MWNT resulting in a relatively weaker C-C bond strength in the MWNT/PBA complex than the MWNT alone.27 Additionally, a broad peak at 3136 cm−1 for the O-H stretching was observed.28 Following covalent FDH attachment via coupling of Lys amines with surface carbodiimide activated (EDC/NHS) –COOH groups of PBA and MWNTs forming amide bonds, typical amide-I and amide-II bands from the peptide backbone of FDH were observed at 1683 and 1603 cm−1, respectively.29 The disappearance of the prior O-H stretching and appearance of a new broad band at 3472 cm−1 arising from N-H stretching of FDH indicated amide bond formation and thus the enzyme immobilization. Furthermore, it is possible that a partial non-covalent adsorption of the FDH onto the MWNT/PBA surface cannot be ruled out. Despite this, based on the observed FTIR spectral changes of the surface carboxylic acids the covalent FDH attachment is evident.

Figure 2.

Figure 2

FTIR spectra of AuSPE coated with (a) carboxylated MWNTs, (b) after PBA stacking, and (c) after covalent immobilization of FDH.

Additionally, Raman spectroscopy and Faradaic impedance measurements confirmed successful AuSPE sensor surface fabrication and FDH immobilization (Figures S2 and S3, and associated text of discussion of results, Supporting Information). The enzyme fabricated SPEs were assessed for HCHO oxidation in a 10-times diluted urine matrix. The ability of the enzyme to function well in the urine matrix was very important, thus adequate NAD+ cofactor (5 mM) was supplied. Ten times dilution of urine provided the optimum performance of the FDH bioelectrode with a minimal non-specific signals in the working potential region as discussed below in Section 3.2. FDH-catalyzed detection of HCHO was performed according to the reaction sequence shown in Figure 1B. Q was used as the electron transfer mediator to transfer electrons to the electrode from NADH formed in solution as a result of the FDH-catalyzed HCHO oxidation.21 This in turn regenerates NAD+ to receive subsequent protons upon HCHO oxidation.

3.2 Amperimetric response for the strirred vs flow injection analysis

Upon dilution of the urine matrix, an increase in the amperometric current was observed (Figure S4, tested using 1 ppm of HCHO for method optimization). This could be attributed to the reduction in high salt concentration and other interfering agents for the optimum performance of the sensor and/or the FDH enzyme.30 Diluting the urine samples more than ten times did not significantly increase the amperometric response (at 95% confidence level). Hence, ten times dilution of the urine was selected for further studies.

Figure 3A-a shows the increase in currents of the bioelectrode in stirred solutions containing various concentrations of HCHO from ppb to ppm levels in 10-times diluted urine matrix at an optimum solution stirring rate of 150 rpm. Linearity of the currents vs. concentration of HCHO was observed between 100 ppb and 16 ppm (initial range in Fig. 3A-b), which is a wide dynamic range useful for clinical assay of such markers. A deviation from the linearity was observed at higher concentrations of HCHO (i.e., > 16 ppm) and followed typical electrochemical Michaelis-Menten enzyme kinetics (Fig. 3A-b).31 The estimated apparent Michaelis-Menten constant (KMapp) using KaleidaGraph software (version 4.1) was 19.9 ± 4.6 ppm. The limit of detection (LOD) (the signal at three times the standard deviation of the mean of the HCHO unspiked control urine sample32) of the described stirred solution-based HCHO detection was 73 ppb.

Figure 3.

Figure 3

(A) Stirred solution method: (a) Amperometric responses of the AuSPE/MWNT/PBA-FDH bioelectrode for various concentrations of HCHO in 10-times diluted urine in PBS, pH 7.4, containing 5 mM NAD+ and 1 mM Q at an applied potential of + 0.35 V at 23°C and a constant stirring of solution using a magnetic stirrer at 150 rpm. Inset shows the enlarged view for lower HCHO concentrations. (b) Michaelis-Menten fit of the designed bioelectrode in oxidizing HCHO. (B) Flow injection analysis: (a) Amperometric responses for the AuSPE/MWNT/PBA enzyme bioelectrode upon injection of various concentrations of HCHO at a flow rate of 100 μL min−1. Inset shows the enlarged view for lower HCHO concentrations. (b) The corresponding Michaelis-Menten fit of the experimental data.

To examine improvements on the detection sensitivity of the stirred solution method by facilitating better mass transport of HCHO and other assay reagents, we used flow injection analysis.33 The cofactor and electron transfer mediator containing solution mixture (5 mM NAD+ and 1 mM Q in PBS, pH 7.4) was delivered at an optimum flow rate of 100 μL min−1 onto the FDH bioelectrode surface using a syringe pump. Various concentrations of HCHO spiked in 10-times diluted urine in PBS (pH 7.4) then were injected via a manual injection valve, and the resulting oxidation currents were measured (Figure 3B-a).

The sensitivity of HCHO detection was only moderately enhanced (by two times) in the flow injection method compared to the stirred solution (calculated from the slopes of the initial linear range in Fig. 3A-b and 3B-b). Possible reason could be that the HCHO and mediators used are small molecules. As a result the mass transport by convection (stirring vs fluid flow) does not seem to be greatly different in view of sensitivity (i.e., the slope of response plots covering the linear range of HCHO concentrations). Nevertheless, the flow injection analysis significantly decreased the LOD to 6 ppb, which is 12-fold smaller than the stirred-solution method. This is likely because diffusion of reactants to products (cofactor, urine matrix, and mediator reaching the electrode and the product diffusing away) becomes much more prominent at the lowest concentration corresponding to the LOD. As a result, the flow injection method for the designed bioelectrode seems to be better than the stirred-solution analysis in facilitating effective electronic communication of the low levels of urine HCHO with the surface of the AuSPE/MWNT/PBA enzyme electrode.

Reported studies based on chromatography and mass spectrometry methods identified increased levels of HCHO present in the urine samples of patients with dementia and bladder cancer conditions compared to healthy individuals. Thus, the obtained LOD of the FDH bioelectrode for measuring HCHO in urine meets the reported clinically relevant range [e.g., bladder cancer (> 85 ppb),34 dementia (> 5.8 ppm),8a and early stages of cognitive impairments in older adults (> 330 ppb)8b]. The lower LOD illustrates the advantages of flow injection analysis in improving analyte mass transport and minimizing noise levels, resulting in enhanced signal-to-noise ratio and thus an improved detection limit. The linear dynamic range of our flow analysis for HCHO in 10-times diluted urine was from 10 ppb to 10 ppm (Figure 3B-b). The relationship between current signal and HCHO concentration was fit by the Michaelis-Menten non-linear regression curve (Figure 3B-b) using KaleidaGraph software (version 4.1).

The estimated KMapp of flow injection method was 9.6 ± 1.2 ppm, which is better than the stirred-solution method and likely due to enhanced substrate mass transport facilitating kinetically faster steady state conditions in the flow system. The observed difference between the affinity constants was calculated to be significant at 95% confidence level. Previously reported KM values of HCHO solution bioassays were 5.4 ppm for FDH from genetically engineered H. polymorpha,20 7.5–8.7 ppm for a homologous enzyme from Candida boidinii,35 6.3 ppm for FDH from H. polymorpha,36 and 12.9 ppm for FDH from Pichia pastoris.37 Results indicate that the KMapp values obtained in the present work are comparable with the reported bioassays.

The presence of HCHO in the prepared urine samples was independently verified by 2,4-dinitrophenylhydrazine (DNPH) derivatization of HCHO and identification of the resulting methanal 2,4-dinitrophenylhydrazone derivative (HCHO-DNPH) product using the liquid chromatography-mass spectrometry method (LC-MS, Figure S5 and associated text of discussion of results, Supporting Information).

The presented FDH bioelectrode design offers sensitivity, ultra-low clinically useful LOD, and simplicity, but it is also essential to meet selectivity criteria for clinical diagnosis of small-molecule disease markers. Figure 4 depicts the current signals for HCHO with two immediate aldehyde homologues and with acetone and two types of alcohol. The current responses confirm the high selectivity for HCHO, with ~ 25% cross-reactivity for acetaldehyde (CH3CHO) and negligible cross-reactivity for propanaldehyde, acetone, methanol, and ethanol. The moderate cross-reactivity for acetaldehyde suggests that the substrate binding pocket of the FDH enzyme has slight affinity to this immediate homologous member.

Figure 4.

Figure 4

Selectivity of the designed AuSPE/MWNT/PBA-FDH bioelectrode for HCHO over other similar analytes in stirred solutions. Current signals for 5 ppm of analytes (x-axis) in 10-times diluted urine solutions in PBS are shown.

3.3 Stability and selectivity of the electrode

The fabricated FDH bioelectrodes were tested for film stability in PBS (pH 7.4 at 23°C) for 40 h by non-faradaic impedance spectroscopy at an applied constant frequency of 5 Hz (Figure S6, Supporting Information).38 A stable impedance magnitude of 232 Ω was noted for more than 30 h. It is worth mentioning that AuSPEs are intended to be for disposable use and hence the observed stability upon continuous soaking of the modified, FDH attached SPE surface in a buffer solution is reasonably good.

Excluding the MWNT/PBA surface modification of AuSPEs, the bioelectrode design only required one step of FDH immobilization and subsequent detection and quantitation of HCHO in 10-times diluted urine. The various sample matrices used in the literature for HCHO and the reported linear dynamic range, detection limit, sensitivity, and bioelectrode stability are presented in Table 1. This comparison shows better performance of the pyrenyl-cabon nanostructure bioelectrode than simple buffer solution-based HCHO quantitation methods.

Table 1.

Comparison of the present MWNT/PBA-FDH nano-bioelectrode with relevant reported studies.

Sensor type Construction material Sensing method Sensing element Matrix Linear dynamic range/Sensitivity/Detection limit Ref.
Planar electrochemical transducers Potentiometric sensor – Si3N4-ISFET Conductometric and amperometric sensor – gold interdigitated electrodes vacuum deposited with sintered aluminum oxide Potentiometry, conductometry, and amperomety Alcohol oxidase from Hansenula polymorpha 10 mM Phosphate buffer (pH 7.7) 150 – 9000 ppm 39
Enzyme biosensor Si/SiO2/Si3N4 structure as physical transducers Capacitance NAD+ and glutathione-dependent recombinant formaldehyde dehydrogenase 2.5 mM Borate buffer (pH 8.40) 0.3 – 600 ppm
Sensitivity 31 mV/decade
Detection limit 0.3 ppm
11
Amperometric enzyme sensor Disks of woven graphite gauze Amperometry NAD+ and glutathione-independent formaldehyde dehydrogenase from Hyphomicrobium zavarzinii strain ZV 58 0.1 M KCl, 80mM KH2PO4 (pH 8.0) 0.5 – 15 ppm
Sensitivity 0.39 μA/ppm
19
Amperometric biosensor Platinized graphite electrode modified with low-molecular free-diffusing redox mediators or positively charged cathodic electrodeposition paints modified with Osbis-N,N-(2,2′-bipyridil)-chloride ([Os(bpy)2Cl]) Amperometry NAD+ and glutathione-dependent formaldehyde dehydrogenase isolated from a geneengineered strain of the methylotrophic yeast Hansenula polymorpha 20 mM Phosphate buffer (pH 8.2) 0.3 – 3 ppm
Sensitivity 358 Am−2 M−1
Detection limit 90 ppb
20
Electrochemical biosensor Glassy carbon electrode containing a membrane constructed with mesoporous silica materials Amperometry Formaldehyde dehydrogenase from Pseudomonas sp. Phosphate buffer (pH 7.4) 30 ppb – 30 ppm
Detection limit 36 ppb
21
Electrochemical biosensor Screen-printed carbon electrode modified with MWCNT Amperometry Formaldehyde dehydrogenase from Pseudomonas putida PBS (based on in situ released HCHO from a prodrug treated cancer cells in PBS) 3 ppb – 3 ppm
Detection limit 300 ppb
22
Electrochemical biosensor Screen printed gold electrodes modified with MWNT/PBA-FDH via the –COOH groups of both MWNT and PBA Amperometry Formaldehyde dehydrogenase from Pseudomonas sp. 10-times diluted urine (in PBS, pH 7.4) Stirred solution analysis: 100 ppb – 16 ppm
Sensitivity (based on the initial linear range, Fig. 3) 174 nA/ppm
Detection limit 73 ppb
Flow injection analysis: 10 ppb – 10 ppm
Sensitivity (based on the initial linear range, Fig. 3)
342 nA/ppm
Detection limit 6 ppb
This work

4. Conclusions

Results presented confirm the analytical detection features of pyrenyl carbon nanostructure-modified FDH bioelectrode for sensitive and selective quantitation of urine HCHO to relate to abnormal conditions that are known to result in elevated urine HCHO. The presented approach provides a viable nano-bioelectrode design for non-invasive detection of small-molecule markers of cancer and other diseases at clinically relevant ultra-low levels in complex matrices. Combining with measurements of other biomarkers and assays of significance is expected to allow successful diagnostic outcome of an abnormal condition. By appropriate immobilization of marker specific receptor molecules or enzymes coupled with a detection probe or mechanism, the proposed methodology is expected to allow broader applicability for quantitative measurement of any other small-molecule markers with selectivity.

Supplementary Material

supplement
NIHMS863828-supplement.docx (644.5KB, docx)

Highlights.

  • First pyrenyl carbon nanostructure based bioelectrode for urine formaldehyde quantitation

  • Sensitivity and selectivity with a wide dynamic range and a few parts-per-billion detection limit

  • Comparative analytical features of stirred-solution and flow-injection amperometric analysis

  • Surface design for ultra-low measurements of small molecules present in complex sample matrices

Acknowledgments

Research reported in this publication was supported, in part, by the National Institute of Diabetes and Digestive and Kidney Diseases of the National Institutes of Health (Award Number R15DK103386), and, in part, by the Oklahoma State University (Start-up funds). The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health. We thank Mr. Charuksha Walgama for helpful discussions.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • 1.a) Ludwig JA, Weinstein JN. Biomarkers in cancer staging, prognosis and treatment selection. Nat Rev Cancer. 2005;5:845–856. doi: 10.1038/nrc1739. [DOI] [PubMed] [Google Scholar]; b) de Gramont A, Watson S, Ellis LM, Rodón J, Tabernero J, de Gramont A, Hamilton SR. Pragmatic issues in biomarker evaluation for targeted therapies in cancer. Nat Rev Clin Oncol. 2015;12:197–212. doi: 10.1038/nrclinonc.2014.202. [DOI] [PubMed] [Google Scholar]; c) Kalia M. Biomarkers for personalized oncology: recent advances and future challenges. Metabolism. 2015;64:S16–S21. doi: 10.1016/j.metabol.2014.10.027. [DOI] [PubMed] [Google Scholar]; d) Surinova S, Radová L, Choi M, Srovnal J, Brenner H, Vitek O, Hajdúch M, Aebersold R. Non-invasive prognostic protein biomarker signatures associated with colorectal cancer. EMBO Mol Med. 2015;7:1153–1165. doi: 10.15252/emmm.201404874. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.a) Shipp G. Ultrasensitive measurement of protein and nucleic acid biomarkers for earlier disease detection and more effective therapies. Biotechnol Healthc. 2006;3:35–40. [PMC free article] [PubMed] [Google Scholar]; b) Swierczewska M, Liu G, Lee S, Chen X. High-sensitivity nanosensors for biomarker detection. Chem Soc Rev. 2012;41:2641–2655. doi: 10.1039/c1cs15238f. [DOI] [PMC free article] [PubMed] [Google Scholar]; c) Nimse SB, Sonawane MD, Song KS, Kim T. Biomarker detection technologies and future directions. Analyst. 2016;141:740–755. doi: 10.1039/c5an01790d. [DOI] [PubMed] [Google Scholar]; d) Rusling JF, Kumar CV, Gutkind JS, Patel V. Measurement of biomarker proteins for point-of-care early detection and monitoring of cancer. Analyst. 2010;135:2496–2511. doi: 10.1039/c0an00204f. [DOI] [PMC free article] [PubMed] [Google Scholar]; e) Krishnan S, Mani V, Wasalathanthri D, Kumar CV, Rusling JF. Attomolar detection of a cancer biomarker protein in serum by surface plasmon resonance using superparamagnetic particle labels. Angew Chem Int Ed. 2011;50:1175–1178. doi: 10.1002/anie.201005607. [DOI] [PMC free article] [PubMed] [Google Scholar]; f) Singh V, Krishnan S. An electrochemical mass sensor for diagnosing diabetes in human serum. Analyst. 2014;139:724–728. doi: 10.1039/c3an01542d. [DOI] [PubMed] [Google Scholar]; g) Singh V, Krishnan S. Voltammetric immunosensor assembled on carbon-pyrenyl nanostructures for clinical diagnosis of type of diabetes. Anal Chem. 2015;87:2648–2654. doi: 10.1021/acs.analchem.5b00016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Cheng YSL, Rees T, Wright J. A review of research on salivary biomarkers for oral cancer detection. Clin Transl Med. 2014;3:1. doi: 10.1186/2001-1326-3-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.a) Heneghan HM, Miller N, Lowery AJ, Sweeney KJ, Newell J, Kerin MJ. Circulating microRNAs as novel minimally invasive biomarkers for breast cancer. Ann Surg. 2010;251:499–505. doi: 10.1097/SLA.0b013e3181cc939f. [DOI] [PubMed] [Google Scholar]; b) Wang J, Chen J, Chang P, LeBlanc A, Li D, Abbruzzesse JL, Frazier ML, Killary AM, Sen S. MicroRNAs in plasma of pancreatic ductal adenocarcinoma patients as novel blood-based biomarkers of disease. Cancer Prev Res. 2009;2:807–813. doi: 10.1158/1940-6207.CAPR-09-0094. [DOI] [PMC free article] [PubMed] [Google Scholar]; c) Goossens N, Nakagawa S, Sun X, Hoshida Y. Cancer biomarker discovery and validation. Transl Cancer Res. 2015;4:256. doi: 10.3978/j.issn.2218-676X.2015.06.04. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Zhang J, Sun R, Chen Y, Tan K, Wei H, Yin L, Pu Y. Small Molecule Metabolite Biomarker Candidates in Urine from Mice Exposed to Formaldehyde. Int J Mol Sci. 2014;15:16458–16468. doi: 10.3390/ijms150916458. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Songur A, Ozen OA, Sarsilmaz M. Reviews of environmental contamination and toxicology. Springer; New York: 2010. The toxic effects of formaldehyde on the nervous system; pp. 105–118. [DOI] [PubMed] [Google Scholar]
  • 7.a) Buszewski B, Kęsy M, Ligor T, Amann A. Human exhaled air analytics: biomarkers of diseases. Biomed Chrom. 2007;21:553–566. doi: 10.1002/bmc.835. [DOI] [PubMed] [Google Scholar]; b) Fuchs P, Loeseken C, Schubert JK, Miekisch W. Breath gas aldehydes as biomarkers of lung cancer. Int J Cancer. 2010;126:2663–2670. doi: 10.1002/ijc.24970. [DOI] [PubMed] [Google Scholar]; c) Wehinger A, Schmid A, Mechtcheriakov S, Ledochowski M, Grabmer C, Gastl GA, Amann A. Lung cancer detection by proton transfer reaction mass-spectrometric analysis of human breath gas. Int J Mass Spectrom. 2007;265:49–59. [Google Scholar]
  • 8.a) Tong Z, Zhang J, Luo W, Wang W, Li F, Li H, Luo H, Lu J, Zhou J, Wan Y, He R. Urine formaldehyde level is inversely correlated to mini mental state examination scores in senile dementia. Neurobiol Aging. 2011;32:31–41. doi: 10.1016/j.neurobiolaging.2009.07.013. [DOI] [PubMed] [Google Scholar]; b) Yu J, Su T, Zhou T, He Y, Lu J, Li J, He R. Uric formaldehyde levels are negatively correlated with cognitive abilities in healthy older adults. Neurosci Bull. 2014;30:172–184. doi: 10.1007/s12264-013-1416-x. [DOI] [PMC free article] [PubMed] [Google Scholar]; c) Hao ZH, Li WJ, Li M, He R. Correlation of urine formaldehyde levels and mini mental state examination scores in Alzheimer’s disease. Chin J Gerontol. 2011;31:3442–3444. [Google Scholar]
  • 9.Maruo YY, Nakamura J, Uchiyama M, Higuchi M, Izumi K. Development of formaldehyde sensing element using porous glass impregnated with Schiff’s reagent. Sens Actuators B Chem. 2008;129:544–550. doi: 10.1016/j.talanta.2007.08.017. [DOI] [PubMed] [Google Scholar]
  • 10.a) Vianello F, Stefani A, Di Paolo ML, Lui A, Margesin B, Zen M, Scarpa M, Soncini G. Potentiometric detection of formaldeyde in air by an aldehyde de-hydrogenase FET. Sens Actuators B. 1996;37:49–54. [Google Scholar]; b) Bunde RL, Jarvi EJ, Rosebtreter JJ. A piezoelectric method for monitoring formaldehyde induced cross-link formation between poly-lysine and poly-deoxyguanosine. Talanta. 2000;51:159–171. doi: 10.1016/s0039-9140(99)00265-9. [DOI] [PubMed] [Google Scholar]; c) Kida T, Morinaga N, Kishi S, An KM, Sim KW, Chae BY, Kim JK, Bong-Ki Ryu BK, Shimanoe K. Electrochemical detection of volatile organic compounds using a Na3Zr2Si2PO12/Bi2Cu0.1V0.9O5.35 heterojunction device. Electrochim Acta. 2011;56:7484–7490. [Google Scholar]; d) Vastarella W, Nicastri R. Enzyme/semiconductor nanoclusters combined systems for novel amperometric biosensors. Talanta. 2005;66:627–633. doi: 10.1016/j.talanta.2004.12.007. [DOI] [PubMed] [Google Scholar]; e) Chen T, Liu Q, Zhou Z, Wang Y. The fabrication and gassensing characteristics of the formaldehyde gas sensors with high sensitivity. Sens Actuators B Chem. 2008;131:301–305. [Google Scholar]; f) Wang J, Zhang P, Qi JQ, Yao PJ. Silicon-based micro-gas sensors for detecting formaldehyde. Sens Actuators B Chem. 2009;136:399–404. [Google Scholar]; g) Dai H, Gong L, Xu G, Li X, Zhang S, Lin Y, Zeng B, Yang C, Chen G. An electrochemical impedimetric sensor based on biomimetic electrospun nanofibers for formaldehyde. Analyst. 2015;140:582–589. doi: 10.1039/c4an02021a. [DOI] [PubMed] [Google Scholar]
  • 11.Ali MB, Gonchar M, Gayda G, Paryzhak S, Maaref M, Jaffrezic-Renault N, Korpan Y. Formaldehyde-sensitive sensor based on recombinant formaldehyde dehydrogenase using capacitance versus voltage measurements. Biosens Bioelectron. 2007;22:2790–2795. doi: 10.1016/j.bios.2006.10.002. [DOI] [PubMed] [Google Scholar]
  • 12.a) Friedfeld S, Fraser M, Lancaster D, Leleux D, Rehle D, Tittel F. Field intercomparison of a novel optical sensor for formaldehyde quantification. Geophys Res Lett. 2000;27:2093–2096. [PubMed] [Google Scholar]; b) Kudo H, Suzuki Y, Gessei T, Takahashi D, Arakawa T, Mitsubayashi K. Biochemical gas sensor (biosniffer) for ultrahigh-sensitive gaseous formaldehyde monitoring. Biosens Bioelectron. 2010;26:854–858. doi: 10.1016/j.bios.2010.07.099. [DOI] [PubMed] [Google Scholar]
  • 13.a) Dirksen JA, Duval K, Ring TA. NiO thin-film formaldehyde gas sensor. Sens Actuators B Chem. 2001;80:106–115. [Google Scholar]; b) Lv P, Tang ZA, Yu J, Zhang FT, Wei GF, Huang ZX, Hu Y. Study on a micro-gas sensor with SnO2–NiO sensitive film for in-door formaldehyde detection. Sens Actuators B Chem. 2008;132:74–80. [Google Scholar]; c) Wang J, Liu L, Cong SY, Qi JQ, Xu BK. An enrichment method to detect low concentration formaldehyde. Sens Actuators B Chem. 2008;134:1010–1015. [Google Scholar]
  • 14.Qin X, Wang R, Tsow F, Forzani E, Xian X, Tao N. Colorimetric chemical sensing platform for real-time monitoring of indoor formaldehyde. IEEE Sens J. 2015;15:1545–1551. [Google Scholar]
  • 15.Nyholm L. Electrochemical techniques for lab-on-a-chip applications. Analyst. 2005;130:599–605. doi: 10.1039/b415004j. [DOI] [PubMed] [Google Scholar]
  • 16.Sadik OA, Aluoch AO, Zhou A. Status of biomolecular recognition using electrochemical techniques. Biosens Bioelectron. 2009;24:2749–2765. doi: 10.1016/j.bios.2008.10.003. [DOI] [PubMed] [Google Scholar]
  • 17.Persson B, Hedlund J, Jornvall H. Medium-and short-chain dehydrogenase/reductase gene and protein families. Cell Mol Life Sci. 2008;65:3879. doi: 10.1007/s00018-008-8587-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Tanaka N, Kusakabe Y, Ito K, Yoshimoto T, Nakamura KT. Crystal structure of formaldehyde dehydrogenase from Pseudomonas putida: the structural origin of the tightly bound cofactor in nicotinoprotein dehydrogenases. J Mol Biol. 2002;324:519. doi: 10.1016/s0022-2836(02)01066-5. [DOI] [PubMed] [Google Scholar]
  • 19.Achmann S, Hermann M, Hilbrig F, Jérôme V, Hämmerle M, Freitag R, Moos R. Direct detection of formaldehyde in air by a novel NAD+ and glutathione-independent formaldehyde dehydrogenase-based biosensor. Talanta. 2008;75:786–791. doi: 10.1016/j.talanta.2007.12.015. [DOI] [PubMed] [Google Scholar]
  • 20.Demkiv O, Smutok O, Paryzhak S, Gayda G, Sultanov Y, Guschin D, Shkil H, Schuhmann W, Gonchar M. Reagentless amperometric formaldehyde-selective biosensors based on the recombinant yeast formaldehyde dehydrogenase. Talanta. 2008;76:837–846. doi: 10.1016/j.talanta.2008.04.040. [DOI] [PubMed] [Google Scholar]
  • 21.Shimomura T, Itoh T, Sumiya T, Mizukami F, Ono M. Electrochemical biosensor for the detection of formaldehyde based on enzyme immobilization in mesoporous silica materials. Sens Actuators B Chem. 2008;135:268–275. [Google Scholar]
  • 22.Bareket L, Rephaeli A, Berkovitch G, Nudelman A, Rishpon J. Carbon nanotubes based electrochemical biosensor for detection of formaldehyde released from a cancer cell line treated with formaldehyde releasing anticancer prodrugs. Bioelectrochemistry. 2010;77:94–99. doi: 10.1016/j.bioelechem.2009.06.016. [DOI] [PubMed] [Google Scholar]
  • 23.Niroula J, Premaratne G, Shojaee SA, Lucca DA, Krishnan S. Combined Covalent and Noncovalent Carboxylations of Carbon Nanotubes for Sensitivity Enhancement of Clinical Immunosensors. Chem Comm. 2016;52:13039–13042. doi: 10.1039/c6cc07022a. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.a) Krishnan S, Armstrong FA. Order-of-magnitude enhancement of an enzymatic hydrogen-air fuel cell based on pyrenyl carbon nanostructures. Chem Sci. 2012;3:1015–1023. [Google Scholar]; b) Kamin RA, Wilson GS. Rotating ring-disk enzyme electrode for biocatalysis kinetic studies and characterization of the immobilized enzyme layer. Anal Chem. 1980;52:1198–1205. [Google Scholar]; c) Gooding JJ, Praig VG, Hall EAH. Platinum-catalyzed enzyme electrodes immobilized on gold using self-assembled layers. Anal Chem. 1998;70:2396–2402. doi: 10.1021/ac971035t. [DOI] [PubMed] [Google Scholar]
  • 25.Baños CE, Silva MA. A novel clean-up method for urine analysis of low-molecular mass aldehydes by capillary electrophoresis with laser-induced fluorescence detection. J Chromatogr B. 2011;879:1412–1418. doi: 10.1016/j.jchromb.2010.10.033. [DOI] [PubMed] [Google Scholar]
  • 26.Bartlett PN, Whitaker RG. Strategies for the development of amperometric enzyme electrodes. Biosensors. 1987;3:359–79. doi: 10.1016/0265-928x(87)80018-4. [DOI] [PubMed] [Google Scholar]
  • 27.Karachevtsev VA, Stepanian SG, Glamazda AY, Karachevtsev MV, Eremenko VV, Lytvyn OS, Adamowicz L. Noncovalent interaction of single-walled carbon nanotubes with 1-pyrenebutanoic acid succinimide ester and glucoseoxidase. J Phys Chem C. 2011;115:21072. [Google Scholar]
  • 28.Majeed S, Filiz V, Shishatskiy S, Wind J, Abetz C, Abetz V. Pyrene-POSS nanohybrid as a dispersant for carbon nanotubes in solvents of various polarities: its synthesis and application in the preparation of a composite membrane. Nanoscale Res Lett. 2012;7:1. doi: 10.1186/1556-276X-7-296. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Zhang M, Smith A, Gorski W. Carbon nanotube-chitosan system for electrochemical sensing based on dehydrogenase enzymes. Anal Chem. 2004;76:5045–5050. doi: 10.1021/ac049519u. [DOI] [PubMed] [Google Scholar]
  • 30.Wood MW, Nordone SK, Vaden SL, Breitschwerdt EB. Assessment of urine solute and matrix effects on the performance of an enzyme-linked immunosorbent assay for measurement of interleukin-6 in dog urine. J Vet Diagn Invest. 2011;23:316–320. doi: 10.1177/104063871102300219. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Krishnan S, Walgama C. Electrocatalytic features of a heme protein attached to polymer-functionalized magnetic nanoparticles. Anal Chem. 2013;85:11420–11426. doi: 10.1021/ac402421z. [DOI] [PubMed] [Google Scholar]
  • 32.Bryan T, Luo X, Bueno PR, Davis JJ. An optimised electrochemical biosensor for the label-free detection of C-reactive protein in blood. Biosens Bioelectron. 2013;39:94–98. doi: 10.1016/j.bios.2012.06.051. [DOI] [PubMed] [Google Scholar]
  • 33.a) Lankelma J, Nie Z, Carrilho E, Whitesides GM. Paper-based analytical device for electrochemical flow-injection analysis of glucose in urine. Anal Chem. 2012;84:4147–4152. doi: 10.1021/ac3003648. [DOI] [PubMed] [Google Scholar]; b) Wang J, Krause R, Block K, Musameh M, Mulchandani A, Schöning MJ. Flow injection amperometric detection of OP nerve agents based on an organophosphorus–hydrolase biosensor detector. Biosens Bioelectron. 2003;18:255–260. doi: 10.1016/s0956-5663(02)00178-1. [DOI] [PubMed] [Google Scholar]; c) Cox JA, Gray T, Kulkarni KR. Stable modified electrodes for flow-injection amperometry: application to the determination of thiocyanate. Anal Chem. 1988;60:1710–1713. [Google Scholar]
  • 34.Španěl P, Smith D, Holland TA, Singary WA, Elder JB. Analysis of formaldehyde in the headspace of urine from blad-der and prostate cancer patients using selected ion flow tube mass spectrometry. Rapid Commun Mass Spectrom. 1999;13:1354–1359. doi: 10.1002/(SICI)1097-0231(19990730)13:14<1354::AID-RCM641>3.0.CO;2-J. [DOI] [PubMed] [Google Scholar]
  • 35.Schutte H, Flossorf J, Sahm H, Kula MR. Immobilized formaldehyde-metabolizing enzymes from Hansenula polymorpha for removal and control of airborne formaldehyde. J Eur Biochem. 1976;62:151. doi: 10.1016/j.jbiotec.2011.03.026. [DOI] [PubMed] [Google Scholar]
  • 36.Allais JJ, Louktibi A, Baratti J. Oxidation of methanol by the yeast, Pichia pastoris, purification and properties of the formaldehyde dehydrogenase. Agric Biol Chem. 1983;47:1509. [Google Scholar]
  • 37.Baerends RJS, Sulter GJ, Jeffries TW, Cregg JM, Veenhuis M. Molecular characterization of the Hansenula polymorphaFLD1 gene encoding formaldehyde dehydrogenase. Yeast. 2002;19:37. doi: 10.1002/yea.805. [DOI] [PubMed] [Google Scholar]
  • 38.Rickert J, Göpel W, Beck W, Jung G, Heiduschka P. A ‘mixed’self-assembled monolayer for an impedimetric immunosensor. Biosens Bioelectron. 1996;11:757–768. doi: 10.1016/0956-5663(96)85927-6. [DOI] [PubMed] [Google Scholar]
  • 39.Korpan Y, Dzyadevich S, Arkhipova V, El’skaya A, Gonchar M, Gibson T, Jaffrezic-Renault N, Martelet C, Soldatkin A. Enzyme-based electrochemical sensors for formaldehyde detection. Sensor Mater. 2000;12:79–87. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

supplement
NIHMS863828-supplement.docx (644.5KB, docx)

RESOURCES