. 2017 Jun 3;6(5):331–336. doi: 10.1302/2046-3758.65.BJR-2016-0221.R1

Photofunctionalised Ti6Al4V implants enhance early phase osseointegration

R Yamauchi 1,, T Itabashi 2, K Wada 1, T Tanaka 1, G Kumagai 1, Y Ishibashi 1
PMCID: PMC5457646  PMID: 28522447

Abstract

Objectives

Ultraviolet (UV) light-mediated photofunctionalisation is known to improve osseointegration of pure titanium (Ti). However, histological examination of titanium alloy (Ti6Al4V), which is frequently applied in orthopaedic and dental surgery, has not yet been performed. This study examined the osseointegration of photofunctionalised Ti6Al4V implants.

Methods

Ti and Ti6Al4V implants were treated with UV light, and the chemical composition and contact angle on the surfaces were evaluated to confirm photofunctionalisation. The implants were inserted into femurs in rats, and the rats were killed two or four weeks after the surgery. For histomorphometric analysis, both the bone–implant contact (BIC) ratio and the bone volume (BV) ratio were calculated from histological analysis and microcomputed tomography data.

Results

The amount of carbon and the contact angle on both implants were significantly reduced after UV irradiation. The BIC ratios for both UV light-treated implants significantly increased at two weeks, but there was no significant difference at four weeks. There was no significant difference in the BV ratios between the UV light-treated and control implants at two or four weeks.

Conclusions

This study suggests that photofunctionalisation of Ti6Al4V implants, similar to that of Ti implants, may promotes osseointegration in early but not in the late phase of osseointegration.

Cite this article: R. Yamauchi, T. Itabashi, K. Wada, T. Tanaka, G. Kumagai, Y. Ishibashi. Photofunctionalised Ti6Al4V implants enhance early phase osseointegration. Bone Joint Res 2017;6:331–336. DOI: 10.1302/2046-3758.65.BJR-2016-0221.R1.

Keywords: Ultraviolet, Osseointegration, Titanium alloy, Bone-implant contact, Hydrophilicity

Article focus

  • To investigate the effect of ultraviolet light-mediated photofunctionalisation on Ti6Al4V implants

Key messages

  • Photofunctionalisation reduces the amount of carbon on the Ti6Al4V surface

  • The surface of UV-treated Ti6Al4V implants enhances establishment of osseointegration in the early healing stage, but not in the late stage.

Strengths and limitations

  • This study indicated the effect of photofunctionalisation on Ti6Al4V surface for the first time.

  • Implant biomechanical tests were not performed, and the number of specimens in each group was small

Introduction

Pure titanium (Ti) and titanium alloy (Ti6Al4V) implants are often used as prosthetic devices in orthopaedic and dental surgery because of their advantages, which include corrosion resistance and good biocompatibility.1 The stiffness of Ti6Al4V is higher than that of Ti, and orthopaedic implants are often composed of the former because of the high degree of weight-bearing required from orthopaedic implants.2 The ageing population is increasing in size and the numbers of compromised hosts due to osteoporosis, dialysis, and the use of steroids have increased; in turn, the resulting poor integration between the bone and implant surface often causes severe complications. Therefore, it is important to minimise the time required for early and strong osseointegration.

Implant anchorage is successful when the bone is deposited directly onto the Ti surface without the intervention of soft or connective tissue.3 Early and strong establishment of osseointegration is of great clinical interest and benefit, but the total bone–implant contact (BIC) ratio remains at 50% to 65%, which is far below the ideal of 100%.4 A reduced BIC ratio is reportedly caused by biological ageing secondary to time-dependent biological degradation of the Ti surface.5 Ti surfaces constantly absorb hydrocarbons from the atmosphere, and water and cleaning solutions after the implants are manufactured.6 Some studies have reported that the initial affinity for osteoblasts and the amount of bone–Ti integration are influenced by the amount of carbon on the surface.5,7,8 Additionally, the absorption of hydrocarbons leads to an increase in hydrophobicity on the implant surface. Many studies have reported that surface wettability is an important property for cell behaviour and that cell attachment onto hydrophobic surfaces tends to be weaker than that onto hydrophilic surfaces.9-11

Aita et al3 reported the use of ultraviolet (UV) light-mediated photofunctionalisation of Ti surfaces to overcome biological ageing. This innovative technology enhanced osseointegration, and the BIC ratio of photofunctionalised implants was increased to a near maximum level of 98.2% in their rat model. The effect of photofunctionalisation on Ti6Al4V surfaces was demonstrated in vitro and enhanced both bioactivity and osteoconductivity.13 Photofunctionalised Ti implants are already being used in dentistry, and clinical outcome studies have shown that the use of photofunctionalisation results in a high success rate and decreased healing time.12 Currently used dental and orthopaedic Ti implants were developed based on the concept of bone–implant integration.3 However, in vivo bone histomorphometric parameters, such as the BIC ratio on photofunctionalised Ti6Al4V, remain unclear. The purpose of this study was to examine the effect of UV light-mediated photofunctionalisation on Ti6Al4V implants.

Materials and Methods

Photofunctionalisation of Ti and Ti6Al4V implants

The implants used in this study were made from pure Ti and Ti6Al4V (diameter 2 mm, length 15 mm; the average values of surface roughness were 0.66 and 0.34, respectively), and provided by B. Braun Aesculap Japan Co., Ltd (Tokyo, Japan). Half of the implants were treated with UV irradiation for 15 minutes using a photo device (TheraBeam Affinity; Ushio Inc., Tokyo, Japan) at an intensity of 3 mW/cm2. The light source mounted in the TheraBeam Affinity is a low-pressure mercury (Hg) lamp, which emits predominantly 254 nm UV light. The implants were divided into the following four groups of five: pure Ti implants without UV irradiation (Ti(−)); pure Ti implants with UV irradiation (Ti(+)); Ti6Al4V without UV irradiation (64Ti(−)); and Ti6Al4V with UV irradiation (64Ti(+)).

X-ray photoelectron spectroscopy

To confirm the removal of carbon from the implant surfaces after UV irradiation, the chemical composition of the four groups was evaluated by electron spectroscopy for chemical analysis. This analysis was performed by X-ray photoelectron spectroscopy (XPS) (PHI Quantera SXM; ULVAC-PHI, Inc., Kanagawa, Japan). The analysis spot, analysed at 7.5mm, was the centre of the implants, and the spot size was 300 μm.

Contact angle analysis

To examine the change in wettability on the implant surfaces after UV irradiation, the hydrophilicity and hydrophobicity were evaluated by measuring the contact angle with 1 μL of water at room temperature immediately, and at 0 minutes, 30 minutes, one hour, three hours, six hours, two weeks, and four weeks after UV irradiation. The measurements were performed on the surfaces by the sessile drop technique, taken from the θ/2 method.14 After a drop of water was placed on the implant surface, the height (a) and contact diameter (b) of the drop were calculated by image analysis software (ImageJ v.1.48; National Institutes of Health, Bethesda, Maryland). Using these measurements, the contact angle (θ) was calculated with the following formula:14

θ=2tan1(2a/b)

The surface shows hydrophilicity when the contact angle is low.

Surgery

The study protocol was approved by the Animal Research Committee of Hirosaki University, and all experimentation was performed in accordance with the Rules for Animal Experimentation of Hirosaki University. Eight-week-old male Sprague–Dawley rats were used for the animal experiments. The rats were anaesthetised with 1% to 2% isoflurane. Both hind limbs were shaved, and the skin and fascia layers were opened separately. The flat aspects of the distal femurs were exposed and selected for implantation. The right and left distal femurs were drilled using a 2 mm-diameter drill. The implants with UV irradiation were inserted into the right femur holes, and the implants without UV irradiation were inserted into the left femur holes. After implant placement, the skin and fascia were closed.

Histological analysis

Two or four weeks after the surgery, the rats were killed by intraperitoneal injection of pentobarbital, and the femurs were harvested. As in previous studies15 the specimens were embedded in methyl methacrylate without decalcification. Embedded specimens were cut perpendicular to the short axis of the implants using a microtome. Each section was stained with Villanueva–Goldner (Figs 3a to 3h and 4a to 4h) and observed by light microscopy (BZ-X700; Keyence Corp., Osaka, Japan). In each histological slice, the BIC ratio for each group was calculated by the digital image analysis software (ImageJ v.1.48). The BIC ratio was calculated as the length of bone in direct contact with the surface of the implant divided by the total length of the implant, multiplied by 100 (%). The bone in direct contact was defined as the interface at which bone tissue was located within 20 mm of the implant surface without intervention of soft tissue.

graphic file with name bonejointres-06-331-g003.jpg

Light microscope images (magnification ×4 and ×10, higher magnification images of the boxed areas) two and four weeks after the implantation. The micrographs show the bone response for the (A, C, E, G) Ti(−) and (B, D, F, H) Ti(+) implants. Scale bar: 200 μm. (I) BIC ratios for Ti(−) and Ti(+) were calculated. Results are shown as mean percentage ± standard deviation. p < 0.05, *Ti(−) vs Ti(+), **2 vs 4 w, †Ti(−) 2 w vs Ti(+) 4 w (n = 5).

graphic file with name bonejointres-06-331-g004.jpg

Light microscope images (magnification ×4 and ×10, higher magnification images of the boxed areas) two and four weeks after implantation. The micrographs show the bone response for (A, C, E, G) 64Ti(−) and (B, D, F, H) 64Ti(+) implants. Scale bar: 200 μm. (I) BIC ratios for 64Ti(−) and 64Ti(+) were calculated. Results are shown as mean percentage ± standard deviation. p < 0.05, *64Ti(−) vs 64Ti(+), †64Ti(−) 2 w vs 64Ti(+) 4 w (n = 5).

Bone formation around implants

The specimens were fixed in 10% buffered formalin and analysed using microcomputed tomography. Three-dimensional bone morphometric analysis was performed (Fig. 5b). The bone volume (BV) ratio was defined as the ratio of the mineralised bone volume within 100 μm from the implant surface. The BV ratio was calculated as the bone occupancy in the area of interest divided by the total area of interest, multiplied by 100%.

graphic file with name bonejointres-06-331-g005.jpg

BV ratio was calculated using micro CT. a) The image shows a representative cross-sectional microcomputed tomography slice in the perpendicular direction to the longitudinal axis of the implant. Scale bar: 1 mm. b) The image shows a representative three-dimensional computed tomography image around the implant. Scale bar: 2000 μm. BV ratios for c) Ti and d) Ti6Al4V were calculated. Results are shown as mean percentage and standard deviation. p < 0.05, *UV(−) vs UV(+), ††64Ti(−) 2 w vs 64Ti(−) 4 w (n = 5).

Statistical analysis

A paired t-test was performed to determine differences in XPS findings and in the contact angle between the UV(−) group and UV(+) group. One-way analysis of variance with Tukey’s post hoc test was performed to determine the differences in BIC ratios and BV ratios. Statistical analyses were performed using SPSS version 12.0J (SPSS Inc., Chicago, Illinois), and p values of < 0.05 were considered significant.

Results

X-ray photoelectron spectroscopy

In the XPS analysis, the surface of Ti had mean atomic percentage values of 46.4% for carbon and 33.4% for oxygen before UV irradiation, and 12.0% for carbon and 52.6% for oxygen after UV irradiation. The surface of Ti6Al4V had mean values of 46.8% for carbon and 33.4% for oxygen before UV irradiation, and 9.2% for carbon and 52.6% for oxygen after UV irradiation. The amount of carbon on the Ti and Ti6Al4V surfaces was significantly reduced after UV irradiation, whereas the amount of oxygen significantly increased after UV irradiation (Fig. 1).

graphic file with name bonejointres-06-331-g001.jpg

Atomic percentage on the Ti and Ti6Al4V surfaces as evaluated by XPS. (A, B) On the Ti and Ti6Al4V surfaces, the amount of carbon (C) was significantly reduced and the amount of oxygen (O) had significantly increased after UV irradiation. *p < 0.05, statistically significant difference between before and after UV irradiation (n = 5).

Contact angle analysis

The contact angle indicated that both Ti and Ti6Al4V had a hydrophobic surface before UV irradiation, with a mean contact angle of 64.3° and 72.3°, respectively. After UV irradiation, both surfaces became hydrophilic, with a contact angle of 11.1° and 6.0° degrees, respectively. The contact angle of both surfaces increased and the hydrophilicity decreased with time. There were significant differences between the UV(−) and UV(+) groups immediately, and at 0 and 30 minutes, one, three and six hours, and two weeks. However, there was no significant difference between the UV(−) and UV(+) groups at four weeks (Fig. 2).

graphic file with name bonejointres-06-331-g002.jpg

Hydrophilicity changed after UV irradiation as evaluated by measurement of the contact angle. a) Photographic images showed that a drop of water was deposited onto the Ti and Ti6Al4V surfaces. b) The contact angle of the Ti and Ti6Al4V surfaces exhibited hydrophobicity before UV irradiation and then exhibited hydrophilicity after UV irradiation. The contact angle of both surfaces showed an age-dependent reduction in hydrophilicity. *p < 0.05 (paired t-test), *Ti immediately vs 0 minutes, 30 minutes, 1 hour, 3 hours, 6 hours, and 2 weeks after UV irradiation; †Ti6Al4V or immediately vs 0 and 30 minutes, one, three and six hours and two weeks after UV irradiation (n = 3).

Histological analysis

In the Ti groups, the mean BIC ratio was 39.8% for Ti(−) and 56.8% for Ti(+) at two weeks, and 61.6% for 64Ti(−) and 80.7% for 64Ti(+) at four weeks. There was a significant difference in the BIC ratio between Ti(−) and Ti(+) at two weeks, but no significant difference at four weeks (Fig. 3). In the Ti6Al4V groups, the mean BIC ratio was 44.4% for 64Ti(−) and 65.0% for 64Ti(+) at two weeks, and 58.6% for 64Ti(−) and 76.3% for 64Ti(+) at four weeks. Similarly, there was a significant difference in the BIC ratio between 64Ti(−) and 64Ti(+) at two weeks, but no significant difference at four weeks (Fig. 4).

Bone formation around samples

In the control groups, the mean BV ratio was 61.6% for Ti(−) and 55.9% for 64Ti(−) at two weeks, and 74.5% for Ti(−) and 80.8% for 64Ti(−) at four weeks. In the UV-treated groups, the mean BV ratio was 58.0% for Ti(+) and 55.0% for 64Ti(+) at two weeks, and 86.4% for Ti(+) and 76.0% for 64Ti(+) at four weeks. There were no significant differences between Ti(−) and Ti(+) or between 64Ti(−) and 64Ti(+) at two or four weeks (Fig. 5).

Discussion

This study demonstrated that photofunctionalisation reduced the amount of carbon on Ti and Ti6Al4V surfaces, and changed the surfaces from hydrophobic to hydrophilic. The BIC ratios of both surfaces at two weeks were significantly higher in the UV-treated than the control groups, however, there was no significant difference between the UV-treated and control groups at four weeks.

Some studies have demonstrated that removal of carbon deposition from a Ti surface enhances the bioactivity and osseointegration of Ti.16 Hydroxyl groups increase the wettability and hydrophilicity on the Ti surface, which enhances the initial attachment, proliferation, and differentiation of osteoblasts.17-19 The results of Ti6Al4V in this study are similar to those in previous reports.3,5,13 The previous studies indicated that there were no differences in surface morphology before and after UV irradiation.3,20 This study suggests that the carbon deposition on the Ti and Ti6Al4V surfaces was removed after UV irradiation, increasing the hydroxyl groups and hydrophilicity on both surfaces.

In one study, photofunctionalisation of Ti implants increased the BIC ratio for UV-treated implants 2.5-fold at two weeks and 1.9-fold at four weeks compared with control implants.3 However, no reports have described the BIC ratio for UV-treated Ti6Al4V. In this study, the BIC ratio for UV-treated Ti6Al4V implants at two weeks was significantly higher than that of control implants, but the BIC ratio for UV-treated implants at four weeks was not significantly different. Previous studies reported that photofunctionalisation of Ti and Ti6Al4V surfaces enhanced attachment, spread, proliferation, and differentiation of osteoblasts in vitro.3,13 The present study suggests that photofunctionalisation enhances the bioactivity and osteoconductivity of both surfaces. Notably, the BIC ratios were similar between the UV-treated groups at two weeks and control groups at four weeks, clearly demonstrating an acceleration of osseointegration. Thus, photofunctionalisation of Ti and Ti6Al4V implants enhances osseointegration in the early phase. A previous study reported that after a healing period of 12 weeks, there was no difference in the BIC ratio for Ti implants between the UV-treated and control groups.21 However, when using implants for fracture operations, earlier osseointegration is important.

The present study indicated no significant differences in the BV ratio between UV-treated and control implants at two and four weeks. In one study, the BV for UV-treated implants at two weeks was almost equal to that of control implants, although UV-treated implants were stronger and more resistant to the forces of mechanical stress than control implants.22 Photofunctionalisation enhances osteogenesis around the implant, increases interfacial bone deposition on Ti, and improves the marginal bone seal and support.23 It has been suggested that photofunctionalisation induces denser cortical bone formation and a stiffer bone connection compared with control implants.22

The study had some limitations. First, implant biomechanical tests to assess the biomechanical strength of bone–implant integration were not performed. Previous studies showed that the push-in values for photofunctionalised Ti and Ti6Al4V implants were significantly higher than those for control implants.3,13 Second, only five specimens were included in each group, making it difficult to come to a statistical conclusion. Third, to analyse peri-implant bone regeneration, fluorochrome staining was not performed. This method can be used as a dynamic evaluation of bone regeneration. Further studies are needed to confirm the biomechanical strength and the fluorochrome staining evaluation of the implants used in this study.

In conclusion, the present study revealed the effect of photofunctionalisation on Ti6Al4V in vivo. As was seen by Yamamura et al18 the surface of UV-treated implants demonstrated significant differences in chemical properties and wettability. Additionally, photofunctionalisation induced an accelerated increase in the BIC ratios in the early healing stage. This technology may enable early load bearing and decrease morbidity by application to orthopaedic implants.

Footnotes

ICMJE Conflicts of Interest: None declared

Funding Statement

This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.

References

  • 1. Liu X, Chu PK, Ding C. Surface modification of titanium, titanium alloys, and related materials for biomedical applications. Mater Sci Eng Rep 2004;47:49-121. [Google Scholar]
  • 2. Navarro M, Michiardi A, Castaño O, Planell JA. Biomaterials in orthopaedics. J R Soc Interface 2008;5:1137-1158. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Aita H, Hori N, Takeuchi M, et al. The effect of ultraviolet functionalization of titanium on integration with bone. Biomaterials 2009;30:1015-1025. [DOI] [PubMed] [Google Scholar]
  • 4. Ogawa T, Nishimura I. Different bone integration profiles of turned and acid-etched implants associated with modulated expression of extracellular matrix genes. Int J Oral Maxillofac Implants 2003;18:200-210. [PubMed] [Google Scholar]
  • 5. Att W, Hori N, Takeuchi M, et al. Time-dependent degradation of titanium osteoconductivity: an implication of biological aging of implant materials. Biomaterials 2009;30:5352-5363. [DOI] [PubMed] [Google Scholar]
  • 6. Kilpadi DV, Lemons JE, Liu J, et al. Cleaning and heat-treatment effects on unalloyed titanium implant surfaces. Int J Oral Maxillofac Implants 2000;15:219-230. [PubMed] [Google Scholar]
  • 7. Hori N, Att W, Ueno T, et al. Age-dependent degradation of the protein adsorption capacity of titanium. J Dent Res 2009;88:663-667. [DOI] [PubMed] [Google Scholar]
  • 8. Hayashi R, Ueno T, Migita S, et al. Hydrocarbon deposition attenuates osteoblast activity on titanium. J Dent Res 2014;93:698-703. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Wei J, Yoshinari M, Takemoto S, et al. Adhesion of mouse fibroblasts on hexamethyldisiloxane surfaces with wide range of wettability. J Biomed Mater Res B Appl Biomater 2007;81:66-75. [DOI] [PubMed] [Google Scholar]
  • 10. Altankov G, Grinnell F, Groth T. Studies on the biocompatibility of materials: fibroblast reorganization of substratum-bound fibronectin on surfaces varying in wettability. J Biomed Mater Res 1996;30:385-391. [DOI] [PubMed] [Google Scholar]
  • 11. Groth T, Altankov G. Studies on cell-biomaterial interaction: role of tyrosine phosphorylation during fibroblast spreading on surfaces varying in wettability. Biomaterials 1996;17:1227-1234. [DOI] [PubMed] [Google Scholar]
  • 12. Funato A, Yamada M, Ogawa T. Success rate, healing time, and implant stability of photofunctionalized dental implants. Int J Oral Maxillofac Implants 2013;28:1261-1271. [DOI] [PubMed] [Google Scholar]
  • 13. Minamikawa H, Ikeda T, Att W, et al. Photofunctionalization increases the bioactivity and osteoconductivity of the titanium alloy Ti6Al4V. J Biomed Mater Res A 2014;102:3618-3630. [DOI] [PubMed] [Google Scholar]
  • 14. Yang M, Lin S. A method for correcting the contact angle from the u/2 method. Colloids Surf A Physicochem Eng Asp 2003;220:199-210. [Google Scholar]
  • 15. Yamazaki M, Yamada M, Ishizaki K, Sakurai K. Ultraviolet-C irradiation to titanium implants increases peri-implant bone formation without impeding mineralization in a rabbit femur model. Acta Odontol Scand 2015;73:302-311. [DOI] [PubMed] [Google Scholar]
  • 16. Hayashi R, Ueno T, Migita S, et al. Hydrocarbon deposition attenuates osteoblast activity on titanium. J Dent Res 2014;93:698-703. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Arima Y, Iwata H. Effect of wettability and surface functional groups on protein adsorption and cell adhesion using well-defined mixed self-assembled monolayers. Biomaterials 2007;28:3074-3082. [DOI] [PubMed] [Google Scholar]
  • 18. Yamamura K, Miura T, Kou I, et al. Influence of various superhydrophilic treatments of titanium on the initial attachment, proliferation, and differentiation of osteoblast-like cells. Dent Mater J 2015;34:120-127. [DOI] [PubMed] [Google Scholar]
  • 19. Sartoretto SC, Alves AT, Resende RF, et al. Early osseointegration driven by the surface chemistry and wettability of dental implants. J Appl Oral Sci 2015;23:279-287. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Sawase T, Jimbo R, Baba K, et al. Photo-induced hydrophilicity enhances initial cell behavior and early bone apposition. Clin Oral Implants Res 2008;19:491-496. [DOI] [PubMed] [Google Scholar]
  • 21. Hayashi M, Jimbo R, Xue Y, et al. Photocatalytically induced hydrophilicity influences bone remodelling at longer healing periods: a rabbit study. Clin Oral Implants Res 2014;25:749-754. [DOI] [PubMed] [Google Scholar]
  • 22. Hirota M, Tanaka M, Ishijima M, et al. Effect of photofunctionalization on Ti6Al4V screw stability placed in segmental bone defects in rat femurs. J Oral Maxillofac Surg 2016;74:861.e1-861.e16. [DOI] [PubMed] [Google Scholar]
  • 23. Pyo SW, Park YB, Moon HS, Lee JH, Ogawa T. Photofunctionalization enhances bone-implant contact, dynamics of interfacial osteogenesis, marginal bone seal, and removal torque value of implants: a dog jawbone study. Implant Dent 2013;22:666-675. [DOI] [PubMed] [Google Scholar]

Articles from Bone & Joint Research are provided here courtesy of British Editorial Society of Bone and Joint Surgery

RESOURCES