Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2017 Jun 12.
Published in final edited form as: Eur J Nucl Med Mol Imaging. 2015 Jun 5;42(10):1530–1541. doi: 10.1007/s00259-015-3094-6

Test-retest reproducibility of the metabotropic glutamate receptor 5 ligand [18F]FPEB with bolus plus constant infusion in human

Eunkyung Park 1, Jenna M Sullivan 1, Beata Planeta 1, Jean-Dominique Gallezot 1, Keunpoong Lim 1, Shu-Fei Lin 1, Jim Ropchan 1, Timothy J McCarthy 3, Yu-Shin Ding 4, Evan D Morris 1,2, Wendol A Williams 1,2, Yiyun Huang 1, Richard E Carson 1
PMCID: PMC5467218  NIHMSID: NIHMS860864  PMID: 26044120

Abstract

Purpose

[18F]FPEB is a promising PET radioligand for the metabotropic glutamate receptor 5 (mGluR5), a potential target for the treatment of neuropsychiatric diseases. The purpose of this study was to evaluate the test-retest reproducibility of [18F]FPEB in the human brain.

Methods

Seven healthy male subjects were scanned twice, 3–11 weeks apart. Dynamic data were acquired using bolus plus infusion of 162±32 MBq [18F]FPEB. Four methods were used to estimate volume of distribution (VT): equilibrium analysis (EQ) using arterial (EQA) or venous input data (EQV), MA1, and two-tissue compartment model (2T). Binding potential (BPND) was also estimated using cerebellar white matter (CWM) or grey matter (CGM) as a reference region using EQ, 2T and MA1. Absolute test-retest variability (aTRV) of VT and BPND were calculated for each method. Venous blood measurements (CV) were compared with arterial input (CA) to examine their usability for EQ analysis.

Results

Regional VT estimated by the four methods displayed a high degree of agreement (r2 ranging from 0.83 to 0.99 between methods), although EQA and EQV overestimated VT by a mean of 9% and 7%, respectively, compared to 2T. Mean aTRV of VT were 11% by EQA, 12% by EQV, 14% by MA1 and 14% by 2T. Regional BPND also agreed well between methods and mean aTRV of BPND was 8–12% (CWM) and 7–9% (CGM). Venous and arterial blood concentrations of [18F]FPEB were well matched during equilibrium (CV=1.01·CA, r2=0.95).

Conclusion

[18F]FPEB binding shows good test-retest variability with minor differences between analysis methods. Venous blood can be used as an alternative for input function measurement instead of arterial blood in EQ analysis. Thus, [18F]FPEB is an excellent PET imaging tracer for mGluR5 in humans.

Keywords: [18F]FPEB, human, metabotropic glutamate receptor 5, positron emission tomography, reproducibility

INTRODUCTION

Glutamate is the most abundant excitatory neurotransmitter in the human brain, mediating 70–90% of synaptic transmission [1]. Physiological functions regulated by glutamate include brain development, learning and memory, and long-term potentiation [13]. Not surprisingly, dysfunction in the glutamate system is associated with a variety of pathologic conditions such as drug addiction, schizophrenia, anxiety, mood disorders, pain, epilepsy, stroke, Alzheimer’s disease, Parkinson’s disease, Huntington’s disease, and Fragile X syndrome [4, 5].

Glutamate receptors are classified into ionotropic receptors (iGluRs) and metabotropic receptors (mGluRs). iGluRs mediate fast excitatory neurotransmission and NMDA, AMPA, and kainate receptors fall into this category; while the mGluRs mediate slower, modulatory neurotransmission and are comprised of eight subtypes (mGluR1-8) organized into three groups (I, II, and III) based on their pharmacological and signal transduction properties. mGluRs are coupled to their associated iGluRs by a second messenger cascade and modulate function of iGluRs [6, 7]. iGluRs have received great attention as therapeutic targets for drug addiction and various forms of relapse-like behavior based on evidence from animal studies. However, iGluR antagonists have shown serious side effects in humans and interest has shifted to modulating glutamate transmission via mGluRs [5].

Group I mGluRs (mGluR1 and 5) are mainly located postsynaptically at the periphery of postsynaptic densities, while group II and III mGluRs are present primarily in the presynaptic terminals [1, 8]. mGluR5 is structurally linked to the NMDA receptor; it produces positive modulation of NMDA receptor function, and activation of NMDA receptors enhances mGluR5 responses [5, 6, 9]. This synergistic action activates a number of Ca2+-regulated proteins and leads to an increase in AMPA receptor activity [6, 9]. Selective mGluR5 ligands have been shown or are believed to have therapeutic potential for neuropsychiatric diseases through this receptor crosstalk. For example, mGluR5 agonists and positive allosteric modulators are believed to improve memory function in Alzheimer’s disease by increasing neuronal excitability and synaptic plasticity, while mGluR5 antagonists and negative allosteric modulators could be potential treatments for epilepsy and multiple sclerosis by way of their neuroprotective effects against glutamate excitotoxicity [4].

The ability to detect changes in mGluR5 availability in vivo could be valuable for elucidating functional and dysfunctional glutamate transmission in the brain and facilitating therapeutic drug development targeting mGluR5. [18F]FPEB is a promising PET radioligand that binds to mGluR5 with high specificity and selectivity [10, 11]. The purpose of this study was to evaluate test-retest reproducibility of [18F]FPEB binding using high resolution PET imaging in healthy humans. Kinetic modeling and equilibrium analysis methods to quantify [18F]FPEB binding were evaluated. Specifically, we tested if a shorter equilibrium measurement, instead of a 2-hour dynamic PET scan, was sufficient to derive reliable binding parameters. Venous input functions were also compared with arterial data for equilibrium analysis to examine their usability in estimating [18F]FPEB volume of distribution (VT).

MATERIALS AND METHODS

Subjects

Seven healthy male subjects (mean age 27 ± 7 y, range 19 – 38 y, body weight 83 ± 17 kg, body mass index 26 ± 4 kg/m2) were studied under a protocol approved by the Yale Human Investigation Committee, the Yale-New Haven Hospital Radiation Safety Committee, and the Yale University Radiation Safety Committee. Subjects were examined by a physician to exclude major medical conditions or neurologic disorders. Electrocardiography, complete blood counts, serum chemistries, thyroid function test, liver function test, urinalysis and urine toxicology screening were performed during screening. The Structured Clinical Interview for DSM-IV axis I Disorders (SCID-CV) [12] was administered to rule out axis I psychiatric disorders. Written informed consent was obtained from all subjects before the study after complete explanation of the study procedures.

Brain magnetic resonance imaging

Brain magnetic resonance (MR) scans were collected using a 3T scanner (Trio, Siemens Medical Systems, Erlangen, Germany) for registration with PET and region-of-interest (ROI) definition. MR scans were performed within 2–3 weeks of the first PET scan, except in one subject whose MR scan occurred 3 months before the first PET scan. A 3D T1-weighted gradient-echo (MPRAGE) with 1 mm3 isometric resolution (FA=7°, TE=3.34 ms, TI=1100 ms, and TR=2500 ms) was used.

Brain [18F]FPEB PET scanning

[18F]FPEB was synthesized using a previously published method [13]. Average specific activity was 195 ± 47 MBq/nmol (n = 14) at the end of synthesis. To evaluate test-retest variability, two scans were performed on separate days, 3 to 11 weeks apart. All procedures for the test and retest scans were identical. PET scans were acquired on a High Resolution Research Tomograph (HRRT) scanner (Siemens/CTI, Knoxville, TN), with spatial resolution (full width at half maximum) of 2–3 mm. A radial artery catheter was inserted at the wrist area to measure the metabolite-corrected input function. A venous catheter was inserted at the antecubital area on the same side as the arterial catheter for intravenous administration of [18F]FPEB. Another venous catheter was placed on the contralateral side for venous sampling to compare the venous blood-derived input function with the arterial blood-derived input function.

Before the scan, an optical motion-tracking tool was fastened to the subject’s head via a Lycra swim cap. A 6-min transmission scan using an orbiting 137Cs point-source was obtained for attenuation correction. Emission data were acquired in list mode for 120 min along with bolus plus infusion (B/I) administration of [18F]FPEB. The injected dose was 162 ± 32 MBq with specific activity of 145 ± 50 MBq/nmol at the time of injection (n = 14). The injected mass was 0.32 ± 0.14 μg. For the B/I paradigm, a Kbol value of 190 min was used based on our previous study [14]. For [18F]FPEB administration, PTFE tubing was used because [18F]FPEB was found to stick to typical plastic catheter tubing during the slow infusion, which prevented the full dose delivery of [18F]FPEB. Dynamic scan data were reconstructed with corrections for attenuation, normalization, randoms, scatter, deadtime, and motion (Vicra, NDI Systems, Waterloo, Ontario), using the ordered-subset expectation maximization (OSEM)-based MOLAR algorithm [15].

Input function measurement

The arterial input function was obtained from the time-activity curve (TAC) of the metabolite-corrected arterial concentration of [18F]FPEB (CA), with measurement of plasma free fraction (fp) and unchanged fraction of [18F]FPEB. Arterial blood samples were taken immediately prior to [18F]FPEB injection and during the scan. To measure fp, triplicate 300 μL aliquots of plasma separated from blood collected prior to tracer injection and mixed with the radiotracer were pipetted into ultrafiltration units and centrifuged at room temperature (20 min at 1228 g). Plasma and ultrafiltrate activities were counted, and fp was calculated as the ratio of ultrafiltrate activity to total plasma activity. The same procedure was applied to arterial and venous blood.

For the first 7 min postinjection, continuous arterial blood activity measurements were made with an automated blood counting system (PBS-101, Veenstra Instruments, Joure, The Netherlands). Thirteen individual blood samples were drawn at 3, 8, 12, 15, 20, 25, 30, 45, 60, 75, 90, 105, and 120 min after the administration of [18F]FPEB, and whole blood and plasma were counted. HPLC was performed at selected time points and the parent fraction was determined by the ratio of counts in the parent peak to the total counts collected. The parent fractions were fitted to a 3-parameter function (1 exponential plus a constant) bounded to be ≤ 1. The final metabolite-corrected arterial concentration of [18F]FPEB was calculated from the product of plasma radioactivity and the parent fractions.

Full arterial input functions were not available for retest scans of three subjects, because two subjects were unable to tolerate arterial lines and the other subject had failure in the automated blood counting system (arterial samples during the equilibrium period were available for this subject). For these 3 cases, the input functions used for equilibrium analysis using arterial data (EQA), two-tissue compartment model (2T), and multilinear analysis 1 (MA1) were calculated from each subject’s test data, scaled by the ratio of injected doses in the test and retest scans.

Venous input functions were obtained and compared with arterial input function, in order to determine if less invasive venous blood samples could be used instead of arterial ones when estimating VT using equilibrium analysis (EQ). The parent fractions were averaged from 90 to 120 min. The venous input function was obtained from the time-activity curve (TAC) of the metabolite-corrected venous concentration of [18F]FPEB (CV). This curve was only obtained for the latter portion of the curve, specifically from five individual blood samples manually drawn at 60, 75, 90, 105, and 120 min.

The relative difference in fp between arterial and venous blood was calculated as [(arterial fp − venous fp)/arterial fp] · 100 (%). The relative difference between CA and CV was calculated as [(CACV)/CA] · 100 (%).

ROI delineation

ROIs from the AAL (Anatomical Automatic Labeling for SPM2) template were used. In order to compute TACs for these ROIs, summed PET images from 0 to 10 min post-injection (p.i.) were registered to each subject’s T1-weighted MR images using a six-parameter mutual information algorithm (FLIRT, FSL 3.2, Analysis Group, FMRIB, Oxford, UK), which was then registered to the MR template by nonlinear transformation using the Bioimagesuite software (version 2.5; http://www.bioimagesuite.com). The ROIs were: caudate (16 cm3), cerebellar grey matter (CGM, 84 cm3), cerebellar white matter (CWM, 6 cm3), anterior cingulate cortex (22 cm3), posterior cingulate cortex (6 cm3), frontal cortex (256 cm3), hippocampus (15 cm3), occipital cortex (81 cm3), pallidum (5 cm3), parietal cortex (65 cm3), putamen (17 cm3), temporal cortex (172 cm3), and thalamus (17 cm3).

Kinetic analysis

The distribution volume (VT) of [18F]FPEB was estimated by four different analysis methods: 2 equilibrium analysis [16, 17] methods using arterial and venous blood data (EQA and EQV, respectively), 2T, and MA1 (t* = 30 min) [18]. With EQ, once equilibrium is achieved in the tissue and the blood, VT can be measured directly from the concentration ratio of a tracer in the tissue (CT) to that in plasma. VT was measured as the ratio of the average CT and CA values from 90 to 120 min p.i. (EQA,90–120). Different time windows were also evaluated, in order to test if earlier and/or shorter scans are feasible. Specifically, 30-min time windows were tested by shifting the starting time by 10 min down to 60 min (80–110 min, 70–100 min, and 60–90 min). Fifteen-minute time windows of 105–120 min, 90–105 min, 75–90 min, and 60–75 min were also tested. VT was also measured using venous data (EQV,90–120). The quality of equilibrium was assessed by calculating the rate of change (% change/h) for CT, CA, and CV.

In addition, VT corrected for fP (VT/fP) was calculated from the four analysis methods, in order to test if fP correction affected test-retest reliability.

Even though a validated reference region is not available for [18F]FPEB [19, 20], it is sometimes the case that calculation of an apparent binding potential using a region with a small quantity of specific binding can be useful. Therefore, apparent binding potential (BPND) was calculated as VTVT,CWM1 using CWM as a reference region. This region showed the lowest uptake of [18F]FPEB in our previous study [14]. BPND was also calculated using CGM as a reference region for comparison.

Test-retest reliability

To evaluate the test-retest reliability of [18F]FPEB binding parameters, percent test-retest variability (TRV) of VT, VT/fp, and BPND were calculated for each subject:

TRV(%)=(TRT)(T+RT2)100

where T and RT refer to values from the test and retest scans, respectively.

Statistics to evaluate test-retest reliability of VT, VT/fp, and BPND of [18F]FPEB were mean and standard deviation (SD) of TRV, and mean of absolute TRV (aTRV). Both relative and absolute TRV were evaluated since they have been used in the literature previously. Mean TRV estimates any systematic trend of [18F]FPEB binding values between the test and the retest scans; if binding values across the ROIs tend to be higher in the test scan than the retest, mean TRV will be positive. If there is no such trend, it will be 0. SD of TRV estimates uncertainty in the percent change between the two scans. Mean of aTRV combines both trend and variability of change in [18F]FPEB binding, and thus can be complex if there is a systematic trend. In the absence of such a trend, the mean of aTRV is approximately equal to the percent uncertainty in one single scan value, and is thus smaller than the SD of TRV, which is the percent uncertainty in the difference between the two scans. The intersubject variability was measured by the coefficient of variation (CoV) for all ROIs.

Statistical analysis

Paired t-tests were used to compare [18F]FPEB injection parameters, including injected dose, specific activity, injected mass, fp, VT, VT/fp and BPND between the test and the retest studies. A p value less than 0.05 was considered statistically significant.

RESULTS

Injection parameters were similar between the test and retest scans (Table 1). Mean tissue time-activity curves of all ROIs from the 14 test and retest scans are shown in Figure 1. High uptake areas were the anterior cingulate, frontal, temporal, parietal, and occipital cortices, hippocampus, caudate nucleus, and putamen. Medium uptake was observed in the posterior cingulate cortex, pallidum, and thalamus. Uptake was low in the CGM and the lowest in CWM.

Table 1.

Study parameters

Injection parameters Test Retest p value
Specific activity at the end of synthesis (MBq/nmol) 198 ± 40 192 ± 56 0.86
Specific activity at the time of injection (MBq/nmol) 155 ± 42 135 ± 59 0.55
Injection dose (MBq) 164 ± 27 161 ± 38 0.87
Injection mass (μg) 0.28 ± 0.08 0.36 ± 0.18 0.32

Numbers are mean ± standard deviation. n=7.

Fig. 1.

Fig. 1

Mean tissue time-activity curves expressed as standard uptake value (n = 14 scans of 7 subjects)

For most regions, equilibrium was reached by 90 min. In all regions except CWM, the rate of change of CT between 90 and 120 min was −5±9%/h (mean ± SD), which was smaller in high-uptake regions (mean of −3%/h) than medium-uptake regions (mean of −8%/h). For CWM, CT change rate was −15±26%/h.

Comparison of 2T, MA1, and EQ methods for VT calculation

Regional VT estimates derived from four different analysis methods are shown in Table 2. By any method, anterior cingulate cortex had the highest VT and cerebellar white matter had the lowest among the ROIs examined. Values of VT estimated by different methods agreed well with each other (Figure 2). For example, VT values estimated by MA1 had excellent correlation with those estimated by 2T (VT(MA1) = 0.99 · VT(2T), r2 = 0.99). Values of VT estimated by EQA,90–120 also had good correlation with those from 2T (VT(EQA,90–120) = 1.09 · VT(2T), r2 = 0.93), although EQA,90–120 overestimated VT in all ROIs by a mean of 9%. Values of VT estimated by EQV,90–120 had somewhat worse correlation with those by 2T (VT(EQV,90––120) = 1.07 · VT(2T), r2 = 0.83), and overestimated VT in all ROIs by a mean of 7%. The intersubject variability measured by mean CoV across the ROIs was smallest for 2T (13.7%) and MA1 (13.5%), and largest for EQV,90–120 (22.2%). Mean CoV for EQA,90–120 was 18.2%. Regional BPND values derived from three methods using CWM as reference region are shown in Table S1 and Figure S1 of the Online Resource.

Table 2.

VT and VT/fp results by different analysis methods.

VT VT/fp

ROI 2T MA1
(t*=30)
EQA
(90–120 min)
EQV
(90–120 min)
2T MA1
(t*=30)
EQA
(90–120 min)
EQV
(90–120 min)
anterior cingulate 30.4 ± 4.1 30.0 ± 3.7 31.8 ± 5.8 31.3 ± 6.9 569 ± 64 564 ± 66 599 ± 97 678 ± 144
posterior cingulate 16.2 ± 3.8 16.5 ± 3.8 19.1 ± 5.3 18.8 ± 5.8 309 ± 75 311 ± 77 362 ± 105 405 ± 106
frontal 23.2 ± 3.0 22.9 ± 2.7 25.8 ± 4.2 25.4 ± 5.5 434 ± 45 431 ± 43 486 ± 67 553 ± 125
temporal 27.3 ± 4.2 26.7 ± 3.7 28.6 ± 5.3 28.1 ± 6.5 514 ± 70 501 ± 59 537 ± 85 611 ± 143
parietal 23.2 ± 3.8 23.0 ± 3.5 25.9 ± 5.2 25.6 ± 6.4 438 ± 62 431 ± 53 486 ± 83 557 ± 147
occipital 22.4 ± 3.6 22.4 ± 3.5 25.3 ± 5.1 24.9 ± 6.3 423 ± 54 420 ± 54 475 ± 83 542 ± 142
hippocampus 23.4 ± 3.1 23.4 ± 3.0 23.3 ± 4.3 22.9 ± 4.9 448 ± 69 443 ± 70 440 ± 83 496 ± 100
caudate nucleus 25.1 ± 2.9 24.9 ± 3.0 27.2 ± 4.7 26.7 ± 5.3 478 ± 85 474 ± 86 516 ± 112 577 ± 89
putamen 26.1 ± 3.3 25.7 ± 3.0 28.9 ± 4.6 28.3 ± 5.3 492 ± 70 487 ± 70 546 ± 92 614 ± 100
pallidum 14.9 ± 2.0 14.7 ± 1.9 16.9 ± 2.9 16.7 ± 3.6 278 ± 37 277 ± 38 320 ± 51 362 ± 76
Thalamus 17.0 ± 1.9 17.1 ± 1.8 20.0 ± 3.0 19.7 ± 3.7 321 ± 36 322 ± 38 378 ± 56 427 ± 82
cerebellar grey matter 10.0 ± 1.4 9.9 ± 1.3 11.8 ± 2.0 11.6 ± 2.5 187 ± 28 188 ± 29 222 ± 41 251 ± 50
cerebellar white matter 5.7 ± 0.6 5.7 ± 0.6 6.7 ± 0.9 6.6 ± 1.2 107 ± 17 109 ± 17 127 ± 23 143 ± 23

n = 7

Numbers are mean ± standard deviation.

VT and VT/fp values were averaged across test and retest scans, and the mean and SD was calculated across the 7 subjects. Retest VT and VT/fp by 2T, MA1, and EQA of 3 subjects were estimated using the input function from the test study of each subject scaled by the ratio of the injection doses.

VT: total volume of distribution, fP: plasma free fraction, ROI: region of interest, 2T: two-tissue compartment model, MA1: multilinear analysis 1, t*: starting time, EQA: equilibrium analysis using arterial input function, EQV: equilibrium analysis using venous input function.

Fig. 2.

Fig. 2

Comparison of VT by different analysis methods. Each point represents regional VT of each subject (13 ROIs, 14 test and retest studies of 7 subjects). 2T: two-tissue compartment model, MA1: multilinear analysis 1, t*: starting time, EQA,90–120: equilibrium analysis using arterial input function from 90 to 120 min p.i., EQV,90–120: equilibrium analysis using venous input function from 90 to 120 min p.i. Solid lines in the figures are linear fits of data.

Test-retest variability (TRV) of VT

Mean TRV of VT ranged from −13% to −9% for 2T, from −12% to −9% for MA1, from −6% to −2% for EQA,90–120, and from 0% to 4% for EQV,90–120 across the ROIs (Table 3). EQA,90–120 and EQV,90–120 had the smallest mean TRV. Test VT values were lower than those of retest in most ROIs of 6/7 subjects analyzed by 2T and MA1, but this difference was not significant.

Table 3.

Mean Test-Retest Variability (TRV) of VT and VT/fp by different analysis methods.

Mean TRV of VT (%) Mean TRV of VT/fp (%)

ROI 2T MA1
(t*=30)
EQA
(90–120 min)
EQV
(90–120 min)
2T MA1
(t*=30)
EQA
(90–120 min)
EQV
(90–120 min)
anterior cingulate −10 −9 −3 3 −6 −5 2 5
posterior cingulate −11 −10 −4 1 −6 −6 0 4
frontal −11 −11 −4 2 −6 −6 0 4
temporal −9 −12 −4 2 −5 −7 1 4
parietal −10 −12 −5 1 −5 −7 0 4
occipital −12 −12 −5 1 −7 −8 −1 3
hippocampus −10 −10 −2 4 −6 −5 3 6
caudate nucleus −13 −12 −4 2 −8 −7 1 4
putamen −10 −11 −4 2 −5 −6 1 4
pallidum −12 −12 −6 0 −7 −7 −1 2
thalamus −10 −10 −4 2 −6 −6 0 4
cerebellar grey matter −10 −10 −5 1 −5 −5 −1 3
cerebellar white matter −10 −9 −4 1 −5 −4 0 3

Mean of all ROIs −11 −11 −4 2 −6 −6 0 4

n = 7

For 2T, MA1 and EQA, values of 3 subjects were calculated using the retest VT estimated using scaled input function from the test study of each subject by the injection dose.

TRV: test-retest variability, ROI: region of interest, 2T: two-tissue compartment model, MA1: multilinear analysis 1, t*: starting time, EQA: equilibrium analysis using arterial input function, EQV: equilibrium analysis using venous input function.

SD of TRV ranged from 13% to 18% for 2T, 11% to 16% for MA1, 11% to 18% for EQA,90–120, and 13% to 22% for EQV,90–120 (Table 4). Between methods there was little overall difference in SD of TRV. Mean aTRV ranged from 12% to 16% for 2T and MA1, and 9% to 12% for EQA,90–120, and 10% to 16% for EQV,90–120 (Table 5). The EQ methods thus provided slightly better results than the kinetic methods. TRV of BPND values are shown in the Table S2 of the Online Resource.

Table 4.

SD of Test-Retest Variability (TRV) of VT and VT/fp by different analysis methods.

SD of TRV of VT (%) SD of TRV of VT/fp (%)

ROI 2T MA1
(t*=30)
EQA
(90–120 min)
EQV
(90–120 min)
2T MA1
(t*=30)
EQA
(90–120 min)
EQV
(90–120 min)
anterior cingulate 17 16 14 13 20 20 20 33
posterior cingulate 17 16 14 16 18 18 19 34
frontal 16 16 14 13 20 19 20 32
temporal 17 14 13 13 21 18 20 34
parietal 13 11 11 14 17 19 17 35
occipital 14 13 12 15 17 16 18 35
hippocampus 16 16 13 13 21 21 20 35
caudate nucleus 13 12 12 15 16 15 17 36
putamen 15 14 13 14 19 19 20 34
pallidum 17 16 14 14 21 21 21 33
thalamus 15 15 14 14 17 18 19 32
cerebellar grey matter 13 12 13 15 16 16 19 35
cerebellar white matter 18 16 18 22 20 18 22 39

Mean of all ROIs 16 14 14 15 19 18 19 34

n = 7

For 2T, MA1 and EQA, values of 3 subjects were calculated using the retest VT estimated using scaled input function from the test study of each subject by the injection dose.

TRV: test-retest variability, SD: standard deviation, ROI: region of interest, 2T: two-tissue compartment model, MA1: multilinear analysis 1, t*: starting time, EQA: equilibrium analysis using arterial input function, EQV: equilibrium analysis using venous input function.

Table 5.

Mean absolute test-retest variability (aTRV) of VT and VT/fp by different analysis methods.

Mean aTRV of VT (%) Mean aTRV of VT/fp (%)

ROI 2T MA1
(t*=30)
EQA
(90–120 min)
EQV
(90–120 min)
2T MA1
(t*=30)
EQA
(90–120 min)
EQV
(90–120 min)
anterior cingulate 15 14 11 11 17 16 17 25
posterior cingulate 15 14 12 12 16 16 16 26
frontal 16 16 11 10 17 16 17 24
temporal 15 15 11 10 19 16 16 25
parietal 12 13 9 10 16 14 14 26
occipital 14 14 10 11 16 15 16 26
hippocampus 14 14 11 12 19 19 17 27
caudate nucleus 14 14 9 11 14 14 14 25
putamen 14 15 11 12 17 16 16 25
pallidum 16 16 12 12 17 17 17 25
thalamus 14 14 11 11 15 15 16 23
cerebellar grey matter 12 12 11 12 14 14 16 25
cerebellar white matter 14 13 11 16 14 13 16 27

Mean of all ROIs 14 14 11 12 16 15 16 25

n = 7

For 2T, MA1 and EQA, values of 3 subjects were calculated using the retest VT/fp estimated using scaled input function from the test study of each subject by the injection dose.)

aTRV: absolute test-retest variability, ROI: region of interest, 2T: two-tissue compartment model, MA1: multilinear analysis 1, t*: starting time, EQA: equilibrium analysis using arterial input function, EQV: equilibrium analysis using venous input function.

Plasma free fraction correction

Plasma free fraction (fp) values were available from 12 scans in 7 subjects for the arterial blood, while all 14 values were available for the venous blood. In the test scans, fp values were 5.3% ± 0.9% and 4.7% ± 0.9%, respectively, in the arterial and venous blood (p = 0.182), while arterial and venous fp values were 5.1% ± 0.3% and 4.8% ± 1.0%, respectively, in the retest scans (p = 0.723). The overall relative difference in fp between the arterial and venous blood was 5.0% ± 15.9% (data from 12 studies of 7 subjects). Mean TRV, SD of TRV, and mean of aTRV of fp were −6%, 14%, and 13% for arterial blood and −2%, 33%, and 25% for the venous blood, respectively.

When VT was corrected for fp (VT/fp), the intersubject variability was similar to that of VT. Mean CoV across the ROIs were 14.5% for 2T, 14.5% for MA1, 18.0% for EQA,90–120, and 21.1% for EQV,90–120. With VT/fp, mean TRV was improved for 2T, MA1, and EQA,90–120, across almost all ROIs, but SD of TRV and mean aTRV were increased (Tables 3 through 5).

Comparison of arterial and venous concentration of [18F]FPEB

The arterial and the venous plasma concentrations of [18F]FPEB reached a common level during equilibrium. The correlation was excellent and the venous concentration (CV) was 1% higher than the arterial concentration (CA) at 90–120 min p.i. (CV = 1.01 · CA, r2 = 0.95, Figure 3(a)). In fact, the agreement of the arterial and the venous concentrations occurred by ~ 60 min p.i. The relative differences were 2.8% ± 9% (mean ± SD) at 60–65 min p.i. and −0.3% ± 9% at 115–120 min p.i. (Figure 3(b)).

Fig. 3.

Fig. 3

Comparison of arterial versus venous concentration of [18F]FPEB during equilibrium. (a) Arterial-venous differences were calculated every 5 min from 60 to 120 min p.i. Data are from 11 scans in 7 subjects. (b) Each point represents metabolite-corrected plasma concentration of [18F]FPEB averaged from 90–120 min p.i.

EQ analysis using various time windows

Earlier time windows to estimate VT by EQA and EQV were evaluated to see if scanning could be moved earlier. Time windows of 80–110 min, 70–100 min, and 60–90 min were examined. With both EQA and EQV, VT estimated using 60–90 min p.i. data demonstrated excellent correlation with VT,90–120 (VT,60–90 = 1.00 · VT,90–120, r2 = 0.96 with EQA; VT,60–90 = 1.02 · VT,90–120, r2 = 0.98 with EQV). The percent difference between the 60–90 and 90–120 min values was −1% ± 7% with EQA and −3% ± 5% with EQV. Shorter time windows were also evaluated to test if scanning time could be shortened. Fifteen-minute time windows of 105–120 min, 90–105 min, 75–90 min, and 60–75 min were examined. The correlation with VT,90–120 was excellent even for VT,60–75 (VT,60–75 = 0.99 · VT,90–120, r2 = 0.97 with EQA, Figure 4; VT,60–75 = 1.01 · VT,90–120, r2 = 0.97 with EQV). The difference between VT 60–75 and VT 90–120 was 0% ± 6% with EQA and −3% ± 7% with EQV. Both earlier and shorter time windows did not affect the reproducibility of [18F]FPEB VT. For example, with EQA, the mean TRV, SD TRV and mean of aTRV were −2%, 16%, and 12% for VT,60–90 and −1%, 15%, and 12% for VT,60–75, compared to −4%, 14%, and 11% for VT,90–120. Also, the quality of VT images created using earlier and shorter time windows was comparable to those using 90–120 min data (Figure 5). Characteristics of BPND with shorter and earlier time windows are shown in Figure S2 of the Online Resource.

Fig. 4.

Fig. 4

VT estimation by the equilibrium method (EQA) using an earlier and shorter time window. Each point represents VT of each ROI of each subject (13 ROIs, 14 studies of 7 subjects).

Fig. 5.

Fig. 5

VT parametric images of a representative subject created with EQ analysis using data from 90 to 120 min (upper panel) and from 60 to 75 min (lower panel) p.i.

DISCUSSION

The purpose of this study was to evaluate the test-retest reproducibility of [18F]FPEB binding parameters in the human brain using high resolution PET with bolus plus constant infusion of the radiotracer. Equilibrium analysis using arterial plasma concentration (EQA) provided [18F]FPEB binding parameters with good reproducibility: mean aTRV of 11% for VT and 12% for BPND across the brain areas examined. The 2T and MA1 kinetic models produced slightly worse reproducibility than EQA. Regional VT values and their reproducibility obtained in the present study agreed well with the results of a previous study in human using bolus injections of [18F]FPEB [21]. Reproducibility of [18F]FPEB binding parameters appear to be better than that of [11C]ABP688, another mGluR5 radioligand widely used in human. High variability was reported with [11C]ABP688 in human, with mean aTRV of up to 20% for BPND [22], or from 23 to 73% [23] in each region. Results from test-retest studies in human have not been published to date with other mGluR5 radioligands such as [18F]SP203 or [18F]PSS232.

Regional VT values estimated by four different methods agreed well with each other. Compared to 2T, EQA, 90–120 and EQV, 90–120 overestimated VT in all ROIs by a mean of 9% and 7%, respectively. Values of BPND estimated by three different methods also agreed well with each other. Compared to 2T, EQ90–120 underestimated BPND by a mean of 11%. This underestimation of regional BPND results from overestimation of CBW VT by EQA,90–120 and EQV,90–120 (19% and 17%, respectively), which was larger than the mean overestimation of the other ROIs (9% and 7%, respectively). These result are consistent with our previous study [14].

In principle, VT corrected for plasma free fraction (fp) of a radioligand (VT/fp) is a more accurate estimate for receptor binding than VT itself, because VT values should be linearly proportional to fP. Correction for fP is particularly useful if there are between-group differences in fP and if the fP measurement is reliable. However, VT/fp is not always a more useful measure than VT, especially if additional noise is introduced by including fp, such as when this measurement is not reliable. The fp of [18F]FPEB is relatively low and [18F]FPEB appears to be a highly lipophilic (“sticky”) compound that adsorbs to the ultrafiltration devices used for fp analysis, as well as to certain infusion tubing. This may explain the larger intersubject variability of VT/fp compared to that of VT in this study.

As is often the case in multi-day studies, full arterial blood samples were not available for the retest scans of 3 subjects. For these scans, input functions were created by scaling the arterial curve from the test study by the ratio of the injection doses. The mean and SD of TRV of VT by 2T and MA1 were similar between these 3 subjects who used scaled input function for their retest studies (‘scaled input group’) and the 4 subjects who had original arterial input function for both test and retest studies (‘original input group’). However, mean TRV of VT by EQA,90–120 was larger in absolute value for the scaled input group compared to the original input group (−9% vs. −1%). This does not appear to be due to scaling, because the mean TRV of VT by EQV,90–120 was also larger for the former compared to the latter (7% vs. −2%; venous input functions were original for all subjects.).

The arterial and the venous plasma concentrations of [18F]FPEB reached a common level during equilibrium. The correlation between them was excellent and the venous concentrations were only 1% higher than the arterial concentration on average at 90–120 min. This result is different from that of our previous study, which observed a mean 7% lower concentration in the venous plasma [14]. The difference may have originated from two sources. Study subjects were different and those who participated in the previous study had tissue TACs that were mostly increasing during the 90–120 min period, while subjects in the present study had stable or somewhat decreasing TACs. In addition, the fitting method for metabolite analysis in the venous blood was simplified in the current analysis (see Methods).

Equilibrium was achieved in most regions. The mean negative rate of concentration change (−5±9%/h) between 90 to 120 min suggests that, in this population, the bolus portion of the tracer delivery (Kbol) might be slightly too large. Interestingly, VT from EQ methods showed better stability, with mean % difference of 60–90 vs. 90–120 min values of −1% and −3% for EQA and EQV, respectively. This is consistent with theoretical expectations, whereby the magnitude of bias of the tissue-to-plasma ratio compared to true VT in non-equilibrium conditions is nonlinearly related to plasma and tissue clearance rates [16]. Note that CWM showed poorer equilibrium, which is not surprising since it has the lowest VT value and also has smaller rate constants, as it is a white matter region. If CWM is used as a reference region, BPND values do not show the same stability as VT values, with mean % difference of 60–90 vs. 90–120 min values of +8% (Online Resources), consistent with larger percent reductions in the denominator (CWM) than the numerators. Moving the equilibrium period later than 120 min would likely allow the CWM to obtain adequate stability, if it is to be used as reference region in estimating BPND by EQ.

CWM was selected as a reference region in our study, because it showed the lowest VT (5.7 ± 0.6 with kinetic modeling and 6.7 ± 0.9 with EQA). In another [18F]FPEB study in human [21], pons was used to calculate BPND, since it was the region with the lowest VT in that study (~ 5 mL/cm3). CGM VT was higher than the CWM VT values, indicating that there is either specific binding in the CGM, or that the non-displaceable volume of distribution VND is different between the two regions. If the VND is the same between the two regions, then BPND in CGM is 0.8 ± 0.1 by 2T, 0.7 ± 0.1 by MA1 and 0.8 ± 0.2 with EQA. An in vitro binding study revealed small but positive specific binding of [18F]FPEB in the cerebellar tissue of human as well as rat and rhesus monkey [19]. In vivo blocking study is not yet available for [18F]FPEB. Similarly, for [11C]ABP688 studies [2426], CGM has been used as a reference region, but studies indicate the presence of specific binding of [11C]ABP688 in CGM [20, 27, 28]. Thus, based on our findings and other published data, it appears that there is specific mGluR5 binding in the cerebellum. However, it still may be useful to estimate BPND using the cerebellum as a reference region, so long as investigators clearly spell out their assumptions in its use. In this case, a careful balance must be considered as to whether to use CGM or CWM as a reference region. For example, with [18F]FPEB, selecting CWM as a reference region may include the least possible specific binding but have poorer test-retest variability than selecting CGM for EQ analysis, due to slower kinetics in the white matter than the grey matter as well as noise effects due to region size differences (see Online Resource Table S2 and S3). To avoid this issue, using VT as an outcome measure may be advantageous, as it is not influenced by the specific binding in the cerebellum. However, VT is more sensitive to errors in the input function. BPND may be a better measure if there is no difference in specific binding between groups or patient states in the reference region. From a test-retest reproducibility point of view, [18F]FPEB BPND was better than VT with 2T and MA1 analyses, while BPND was equivalent to VT with EQ. Further, since a portion of the VT value reflects non-displaceable binding (VND), comparable TRV of BPND and VT means greater signal-to-noise contrast for BPND, since 100% of the BPND value is the receptor “signal” (assuming the reference region is correct).

Earlier as well as shorter scanning showed promising results with [18F]FPEB, in that the 60–75 min results were comparable to those of 90–120 min. The correlation of [18F]FPEB binding (VT and BPND by EQ analysis) was excellent and the test-retest reproducibility was similar between those two time windows. Earlier and shorter scanning could bring more convenience for patients, reduce motion issues during scanning, and enhance efficiency of the scanning schedule. However, it may be important to validate the optimal equilibrium in each patient population.

In this study, using a bolus plus constant infusion, we demonstrated test-retest reproducibility of [18F]FPEB to be very good and in agreement between venous and arterial blood measurements during equilibrium. Considering this test-retest reproducibility and the simplicity of scanning and analysis at equilibrium that does not require arterial blood sampling, EQ is considered to be the method of choice to estimate mGluR5 availability in the human brain using [18F]FPEB.

Supplementary Material

Online Resource

Acknowledgments

The authors thank the staff of the Yale University PET Center for their technical expertise and support.

This study was supported by the Yale Pfizer Bioimaging Alliance. This publication was also made possible by CTSA Grant Number UL1 RR024139 from the National Center for Research Resources (NCRR), a component of the National Institutes of Health (NIH), and NIH roadmap for Medical Research.

Footnotes

CONFLICT OF INTEREST

Authors report no conflict of interest relevant to this article.

References

  • 1.Meldrum BS. Glutamate as a neurotransmitter in the brain: review of physiology and pathology. J Nutr. 2000;130:1007S–15S. doi: 10.1093/jn/130.4.1007S. [DOI] [PubMed] [Google Scholar]
  • 2.Riedel G, Platt B, Micheau J. Glutamate receptor function in learning and memory. Behav Brain Res. 2003;140:1–47. doi: 10.1016/s0166-4328(02)00272-3. [DOI] [PubMed] [Google Scholar]
  • 3.Fitzjohn SM, Bashir ZI. Metabotropic glutamate receptor-dependent synaptic plasticity. In: Gereau RW, Swanson GT, editors. The Receptors: The Glutamate Receptors. Totowa, NJ: Humana Press; 2008. pp. 509–28. [Google Scholar]
  • 4.Benarroch EE. Metabotropic glutamate receptors: synaptic modulators and therapeutic targets for neurologic disease. Neurology. 2008;70:964–8. doi: 10.1212/01.wnl.0000306315.03021.2a. [DOI] [PubMed] [Google Scholar]
  • 5.Olive MF. Metabotropic glutamate receptor ligands as potential therapeutics for addiction. Curr Drug Abuse Rev. 2009;2:83–98. doi: 10.2174/1874473710902010083. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Bruno V, Battaglia G, Copani A, D’Onofrio M, Di Iorio P, De Blasi A, et al. Metabotropic glutamate receptor subtypes as targets for neuroprotective drugs. J Cereb Blood Flow Metab. 2001;21:1013–33. doi: 10.1097/00004647-200109000-00001. [DOI] [PubMed] [Google Scholar]
  • 7.Bird MK, Lawrence AJ. Group I metabotropic glutamate receptors: involvement in drug-seeking and drug-induced plasticity. Curr Mol Pharmacol. 2009;2:83–94. doi: 10.2174/1874467210902010083. [DOI] [PubMed] [Google Scholar]
  • 8.Lujan R, Nusser Z, Roberts JD, Shigemoto R, Somogyi P. Perisynaptic location of metabotropic glutamate receptors mGluR1 and mGluR5 on dendrites and dendritic spines in the rat hippocampus. Eur J Neurosci. 1996;8:1488–500. doi: 10.1111/j.1460-9568.1996.tb01611.x. [DOI] [PubMed] [Google Scholar]
  • 9.De Blasi A, Conn PJ, Pin J, Nicoletti F. Molecular determinants of metabotropic glutamate receptor signaling. Trends Pharmacol Sci. 2001;22:114–20. doi: 10.1016/s0165-6147(00)01635-7. [DOI] [PubMed] [Google Scholar]
  • 10.Patel S, Ndubizu O, Hamill T, Chaudhary A, Burns HD, Hargreaves R, et al. Screening cascade and development of potential Positron Emission Tomography radiotracers for mGluR5: in vitro and in vivo characterization. Mol Imaging Biol. 2005;7:314–23. doi: 10.1007/s11307-005-0005-4. [DOI] [PubMed] [Google Scholar]
  • 11.Wang JQ, Tueckmantel W, Zhu A, Pellegrino D, Brownell AL. Synthesis and preliminary biological evaluation of 3-[(18)F]fluoro-5-(2-pyridinylethynyl)benzonitrile as a PET radiotracer for imaging metabotropic glutamate receptor subtype 5. Synapse. 2007;61:951–61. doi: 10.1002/syn.20445. [DOI] [PubMed] [Google Scholar]
  • 12.First MB, Spitzer RL, Gibbon M, Williams JBW. Structured Clinical Interview for DSM-IV Axis I Disorders, Clinician Version (SCID-CV) Washington, D.C.: American Psychiatric Press, Inc.; 1996. [Google Scholar]
  • 13.Lim K, Labaree D, Li S, Huang Y. Preparation of the metabotropic glutamate receptor 5 (mGluR5) PET tracer [F]FPEB for human use: An automated radiosynthesis and a novel one-pot synthesis of its radiolabeling precursor. Appl Radiat Isot. 2014;94C:349–54. doi: 10.1016/j.apradiso.2014.09.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Sullivan JM, Lim K, Labaree D, Lin SF, McCarthy TJ, Seibyl JP, et al. Kinetic analysis of the metabotropic glutamate subtype 5 tracer [(18)F]FPEB in bolus and bolus-plus-constant-infusion studies in humans. J Cereb Blood Flow Metab. 2013;33:532–41. doi: 10.1038/jcbfm.2012.195. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Carson RE, Barker WC, Liow J-S, Johnson CA. Design of a motion-compensation OSEM list-mode algorithm for resolution-recovery reconstruction for the HRRT. IEEE. 2003:3281–5. 2003. [Google Scholar]
  • 16.Carson RE, Channing MA, Blasberg RG, Dunn BB, Cohen RM, Rice KC, et al. Comparison of bolus and infusion methods for receptor quantitation: application to [18F]cyclofoxy and positron emission tomography. J Cereb Blood Flow Metab. 1993;13:24–42. doi: 10.1038/jcbfm.1993.6. [DOI] [PubMed] [Google Scholar]
  • 17.Lassen NA. Neuroreceptor quantitation in vivo by the steady-state principle using constant infusion or bolus injection of radioactive tracers. J Cereb Blood Flow Metab. 1992;12:709–16. doi: 10.1038/jcbfm.1992.101. [DOI] [PubMed] [Google Scholar]
  • 18.Ichise M, Toyama H, Innis RB, Carson RE. Strategies to improve neuroreceptor parameter estimation by linear regression analysis. J Cereb Blood Flow Metab. 2002;22:1271–81. doi: 10.1097/00004647-200210000-00015. [DOI] [PubMed] [Google Scholar]
  • 19.Patel S, Hamill TG, Connolly B, Jagoda E, Li W, Gibson RE. Species differences in mGluR5 binding sites in mammalian central nervous system determined using in vitro binding with [18F]F-PEB. Nucl Med Biol. 2007;34:1009–17. doi: 10.1016/j.nucmedbio.2007.07.009. [DOI] [PubMed] [Google Scholar]
  • 20.DeLorenzo C, Milak MS, Brennan KG, Kumar JS, Mann JJ, Parsey RV. In vivo positron emission tomography imaging with [(11)C]ABP688: binding variability and specificity for the metabotropic glutamate receptor subtype 5 in baboons. Eur J Nucl Med Mol Imaging. 2011;38:1083–94. doi: 10.1007/s00259-010-1723-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Wong DF, Waterhouse R, Kuwabara H, Kim J, Brasic JR, Chamroonrat W, et al. 18F-FPEB, a PET radiopharmaceutical for quantifying metabotropic glutamate 5 receptors: a first-in-human study of radiochemical safety, biokinetics, and radiation dosimetry. J Nucl Med. 2013;54:388–96. doi: 10.2967/jnumed.112.107995. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Kuwabara H, Chamroonrat W, Mathews W, Waterhouse R, Brasic J, Guevara MR, et al. Evaluation of 11C-ABP688 and 18F-FPEB for imaging mGluR5 receptors in the human brain. J Nucl Med. 2011;52:390. [Google Scholar]
  • 23.DeLorenzo C, Kumar JS, Mann JJ, Parsey RV. In vivo variation in metabotropic glutamate receptor subtype 5 binding using positron emission tomography and [11C]ABP688. J Cereb Blood Flow Metab. 2011;31:2169–80. doi: 10.1038/jcbfm.2011.105. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Martinez D, Slifstein M, Nabulsi N, Grassetti A, Urban NB, Perez A, et al. Imaging glutamate homeostasis in cocaine addiction with the metabotropic glutamate receptor 5 positron emission tomography radiotracer [(11)C]ABP688 and magnetic resonance spectroscopy. Biol Psychiatry. 2014;75:165–71. doi: 10.1016/j.biopsych.2013.06.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Milella MS, Marengo L, Larcher K, Fotros A, Dagher A, Rosa-Neto P, et al. Limbic system mGluR5 availability in cocaine dependent subjects: a high-resolution PET [(11)C]ABP688 study. Neuroimage. 2014;98:195–202. doi: 10.1016/j.neuroimage.2014.04.061. [DOI] [PubMed] [Google Scholar]
  • 26.Deschwanden A, Karolewicz B, Feyissa AM, Treyer V, Ametamey SM, Johayem A, et al. Reduced metabotropic glutamate receptor 5 density in major depression determined by [(11)C]ABP688 PET and postmortem study. Am J Psychiatry. 2011;168:727–34. doi: 10.1176/appi.ajp.2011.09111607. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.DeLorenzo C, DellaGioia N, Bloch M, Sanacora G, Nabulsi N, Abdallah C, et al. In Vivo Ketamine-Induced Changes in [11C]ABP688 Binding to Metabotropic Glutamate Receptor Subtype 5. Biol Psychiatry. 2014 doi: 10.1016/j.biopsych.2014.06.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Kagedal M, Cselenyi Z, Nyberg S, Raboisson P, Stahle L, Stenkrona P, et al. A positron emission tomography study in healthy volunteers to estimate mGluR5 receptor occupancy of AZD2066 - estimating occupancy in the absence of a reference region. Neuroimage. 2013;82:160–9. doi: 10.1016/j.neuroimage.2013.05.006. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Online Resource

RESOURCES