Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2018 Jul 1.
Published in final edited form as: J Arthroplasty. 2017 Feb 7;32(7):2268–2273. doi: 10.1016/j.arth.2017.01.054

Are instrumented knee forces representative of a larger population of cruciate-retaining total knee replacements?

Ryan Freed a,*, Jacqueline C Simon a,*, Christopher B Knowlton a,b, Diego A Orozco Villaseñor a,b,1, Markus A Wimmer a, Hannah J Lundberg a
PMCID: PMC5469705  NIHMSID: NIHMS850277  PMID: 28262455

Abstract

Background

It is not known if the loads and motions reported for instrumented knees are generalizable to a larger population of total knee replacement (TKR) patients. The purpose of this study is to: 1) report axial implant force data for chair and stair activities for a population of cruciate-retaining TKR patients and 2) compare the population forces to those measured with instrumented TKRs.

Methods

Twenty-three subjects with a cruciate-retaining TKR underwent motion analysis during stair ascending, stair descending, chair sitting, and chair rising activities after informed consent in this IRB approved study. Axial TKR forces were calculated using a previously validated computational model. Differences between the mean and variability of population instrumented TKR peak forces and force impulses were tested using t-tests and Levene’s test.

Results

Peak axial forces were 3.06, 2.74, 2.65, and 2.60 kN for stair ascent, stair descent, chair rising, and chair sitting, respectively. Force impulses were 123.3, 123.4, 153.5, and 154.0 kN*% activity cycle for stair ascent, stair descent, chair sitting, and chair rising, respectively. Population TKR and instrumented TKR peak forces were different for stair ascent (p=0.03) and stair descent (p=0.03) in the second half of the activity cycles. The variability of the peak forces and impulses were not different (p=0.106 to p=0.99)

Conclusion

The forces and motions presented in this study represent cruciate-retaining TKR patients and could be used for displacement-driven knee wear testing. The forces are similar to those in the literature from instrumented prostheses of an ultra-congruent cruciate-sacrificing TKR.

Keywords: total knee replacement, axial force, musculoskeletal model, stair and chair activities

Introduction

Total knee replacement (TKR) is a widely successful procedure for the treatment of end-stage osteoarthritis. Rates of implantation are increasing dramatically [14], and to such an extent that even small rates of failure can become large numbers of revision surgeries [5]. Thus, it is important to continually improve the longevity of TKRs.

An important tool for preclinical testing of TKR design enhancements is total knee wear simulation. Wear simulators require knowledge of the loads and motions experienced by the knee during daily living. Recently, with the availability of data from instrumented total knees, there has been a lot of interest surrounding knee simulator standards and their physiological relevance. The International Organization for Standardization (ISO) released an update to the displacement standard for knee wear testing [6]. In addition, the American Society for Testing and Materials (ASTM) initiated a workgroup to evaluate how multiple activities should be incorporated into a new wear testing standard for TKRs [7] and published a guide of TKR loading profiles under different activities for force control knee wear testing [8]. Recommendations for standardized testing waveforms for multiple activities were recently made from instrumented knee loads and motions measured during level walking, stair ascending and descending, standing up and sitting down, jogging on a treadmill, knee bend, and one-legged stance [9]. The data has been used in wear simulations [1012]. Incorporating multiple activities into knee wear testing is essential, as a study of TKR patient activity levels and wear has shown that nearly 30% of wear may be attributed to chair and stair activities [13].

Motion waveforms for primary walking kinematics (knee flexion angle) are available from two groups, Bergmann et al. and D’Lima et al [14,15]. Knee load waveforms recommended for knee wear testing for walking, stair, and chair activities have been reported as the average of eight patients by one group [9]. The instrumented implant used by Bergmann et al. is an ultra-congruent cruciate-sacrificing design [14]. It is not known if the loads and motions reported for instrumented knees are generalizable to a larger population of TKR patients, or what effect using inputs that are not population specific for testing of a particular TKR design will have on wear testing.

The purpose of this study is to: 1) to report axial implant force data for chair and stair activities for a population of TKR patients implanted with a modern cruciate-retaining implant design and 2) compare the population forces to patient-specific forces measured with instrumented total knee replacements. This study will help determine if force waveform recommendations are representative of a larger population of TKR patients providing insight into considerations for wear simulation, numerical modeling, and implant design.

Material and methods

Twenty-three subjects (10M/13F) with a NexGen cruciate-retaining total knee system (Zimmer Inc., Warsaw, IN) were tested after informed consent in this IRB approved study. The average age was 61.7±7.0 years old, BMI was 30.9±7.7, and measurements were taken at least 12 months post-op (41.8±29.7). All measurements reported as mean ± standard deviation (SD). All participants were able to function independently without assistive devices and were active in the home and/or workplace.

Motion testing took place in the Rush Motion Analysis Laboratory. A four-camera optoelectronic system (Qualysis, Gothenburg, Sweden) tracked marker motion. Ground reaction forces were recorded using an embedded force plate (Bertec, Columbus, OH). Knee flexion-extension angles, anterior-posterior femoral translation, and internal-external femoral rotation angles were measured using the point cluster technique and marker set [16,17]. Inverse dynamics were used to calculate the external moments and intersegmental forces at the knee from the kinematics and ground reaction forces [18].

During motion testing, subjects performed three trials each of stair ascent, stair descent, chair sitting, and chair rising activities. For stair ascent and descent, a three-step staircase unit (8.5 in step height) was positioned with the middle step on a force plate. Subjects took two normal walking steps to approach the staircase before ascending and were instructed to continue walking away from the staircase after descending. Kinematics were recorded for all steps; external moments were calculated from steps one to two for stair ascent and from steps two to one for stair descent. Subjects did not use a handrail for either stair activity. For chair rising and sitting, an armless chair with adjustable seat height was strategically placed at the perimeter of the force plate so that only the tested foot was detected. The seat height was adjusted so that the patient’s knee was flexed to 90 degrees when in the seated position.

For stair ascent and stair descent activities, the activity cycle started when force was first registered on the force plate. The end of the activity cycle was defined at the point of maximum knee extension after the patient had placed their foot on the subsequent step, indicating the end of swing phase. For chair rising and sitting, the patient first lifted their foot off the force plate and then placed it back down before starting the activity. This allowed for the start of the activity cycle to be defined when force was first registered on the force plate. The end of the chair rising and sitting activity was defined when the knee reached maximum extension and flexion, respectively.

To calculate axial forces during the activities, we used a previously developed numerical model [19] which calculates a solution space of three-dimensional contact forces for both medial and lateral compartments of the tibial plateau. Quasi-static equilibrium is solved for by balancing internal and external forces and moments at 100 instances of an activity cycle. Internal moments and forces are transferred across the knee joint via contact forces, muscle forces, and soft tissue forces, while external forces and moments are transferred to the body via ground reaction forces. Inputs to the model include subject kinematics and kinetics, maximum physiological lower limb muscle forces from a musculoskeletal model [20,21], and a contact path between the tibia and femur. The relationship between medial and lateral knee forces was assumed to vary with the external knee adduction moment [22,23]. The initial relationship between the medial and lateral knee forces and external knee adduction moment was assumed to be the same as what we have previously reported for level walking [22]. If forces were not obtained by the knee model, the relationship was iteratively varied until force solutions were obtained. Knee kinematics were used to calculate the tibiofemoral contact path with previously developed software [24]. The contribution of passive structures to moment equilibrium was included for the frontal and transverse planes if the sum of the muscle moments was not great enough to balance the external moments. The parametric model calculates a range of contact force solutions resulting from variation of relative muscle activation levels. The model has been extensively validated against instrumented total knee axial forces in a blinded and unblinded fashion [22,25].

Axial force profiles for each activity will be reported as the mean of all solutions at each of 100 instances of the activity cycle, averaged across all available trials for each activity. Results for the entire population of TKR patients will be reported as the mean ± 1 SD of all subject means. Peak forces (two for each half of the activity cycle for stair activities and one for chair activities) and force impulse will be calculated for each activity. Force impulse is a single value for each activity equal to the area under the loading curve during an activity cycle. To compare our results to instrumented TKRs, individual subject curves were downloaded from the knee joint test loads found online at www.orthoload.com [9,14]. All available trials were used (8 for stair activities, 7 for chair activities). Impulse and peak forces between the model and instrumented TKR will be compared using independent samples t-tests. Variability of the impulse and peak forces will be compared using Levene’s test for Equality of Variances.

Results

Of the 23 subjects, data collection of stair activities was successful for 22 subjects for stair ascent, and all 23 subjects for stair descent. Data was successfully collected for 21 subjects for chair activities. Three subjects were excluded because their BMI was greater than 40. Higher BMIs result in higher measurement error associated with skin markers; a cutoff BMI of 40 was chosen as it is the definition of class 3 obesity [26]. The knee model was not able to calculate force curves for any of the activities for one individual. For another, the model was not able to calculate forces during chair sitting. In all, knee joint axial forces were obtained for 18 subjects for stair ascent, 19 subjects for stair descent, 17 subjects for chair rising, and 16 subjects for chair sitting.

Knee flexion angles, femoral internal rotation angles, and femoral anterior-posterior displacement during all activities for all processed trials can be seen in Figure 1. Flexion, adduction, and internal rotation external moments for the same are depicted in Figure 2. Moments were normalized to percent body weight times height (%BW*height). Activity forces were reported in kilonewtons (kN) for comparison to the OrthoLoad data recommended for knee simulator wear testing. Force curves were time-normalized, therefore force impulse was reported as a factor of percent activity cycle (kN*% activity cycle).

Figure 1.

Figure 1

The knee flexion angle, femoral internal rotation angle, and femoral anterior displacement for (A) stair ascent (n=18), (B) stair descent (n=19), (C) rising from a chair (n=17), and (D) sitting down on a chair (n=16). Data is reported as mean ± 1 standard deviation, time normalized to percent activity cycle. Coordinate system is as follows: negative extension positive flexion (−Ext/+Flex), negative external rotation of the femur positive internal rotation of the femur (−ER/+IR), and negative posterior displacement of the femur positive displacement of the femur (−Post/+Ant Displ).

Figure 2.

Figure 2

The average knee flexion, adduction, and internal rotation moments normalized to percent body weight times height (%BW*HT) and time normalized to percent activity cycle for (A) stair ascent, (B) stair descent, (C) rising from a chair, and (D) sitting down on a chair. Data is reported as mean ± 1 standard deviation. Coordinate system is as follows: negative extension positive flexion (−Ext/+Flex), negative adduction positive abduction (−Add/+Abd), and negative external rotation positive internal rotation (−ER/+IR) moments about the knee.

Stair ascending and descending had a characteristic waveform with two local maxima, one in the first half of the activity cycle and one in the second half of the activity cycle (Figure 3A–B). Stair ascent peak forces were 3.06 ± 0.67 kN in the first half and 1.97 ± 0.62 kN in the second half of the activity cycle (Table 1). Stair descent peak forces were 2.74 ± 0.70 kN in the first half and 2.31 ± 0.64 kN in the second half of the activity cycle. Force impulses over the entire activity were 123.3 ± 24.2 kN*% activity cycle and 123.4 ± 33.8 kN*% activity cycle for stair ascent and descent respectively (Table 2).

Figure 3.

Figure 3

Axial loading during (A) stair ascent, (B) stair descent, (C) chair rising, and (D) chair sitting. Data is reported as mean ± 1 standard deviation time normalized to percent activity cycle. Knee model output is compared to measured forces from instrumented knee implants reported by Bergmann et al. 2014 which can be found at www.orthoload.com.

Table 1.

Peak forces for the study population TKRs and instrumented TKRs. Results are reported as mean ± 1 standard deviation. Independent samples t-tests were used to determine significance of differences in the means of the peak forces. Levene’s inequality of variances test was used to test for differences in the variability of the peak forces.

Activity Study Population TKRs Instrumented TKRs t test Inequality of Variances
(kN) (kN) p-value f p-value
Chair Rising 2.65 ± 0.84 2.48 ± 0.62 0.663 0.819 0.381
Chair Sitting 2.60 ± 0.77 2.44 ± 0.71 0.655 0 0.99
Stair Ascent
Peak 1 3.06 ± 0.67 2.77 ± 0.57 0.358 0.032 0.861
Peak 2 1.97 ± 0.62 2.70 ± 0.58 0.030 0.101 0.755
Stair Descent
Peak 1 2.74 ± 0.70 2.89 ± 0.62 0.626 0.638 0.438
Peak 2 2.31 ± 0.64 3.01 ± 0.59 0.030 0.434 0.520

Table 2.

Force impulse for study population TKRs and instrumented TKRs. Results are reported as mean ± 1 standard deviation. Independent samples t-tests were used to determine significance of differences in the means of force impulses. Levene’s inequality of variances tests were used to test for differences in the variability of force impulses.

Activity Study Population TKRs Instrumented TKRs t test Inequality of Variances
(kN*% Activity Cycle) (kN*% Activity Cycle) p-value f p-value
Chair Rising 153.5 ± 41.2 137.4 ± 25.3 0.386 3.02 0.106
Chair Sitting 154.0 ± 36.0 129.3 ± 26.3 0.137 0.382 0.545
Stair Ascent 123.3 ± 24.2 131.8 ± 22.5 0.452 0.073 0.791
Stair Descent 123.4 ± 33.8 146.6 ± 23.7 0.116 0.6 0.449

The initial medial-lateral force distribution ratio (originally tuned for level walking) did not result in solutions for chair activities and was interactively changed to obtain more solutions. The final medial-lateral weight distribution ratio was a constant 50%—meaning an equal percentage of load passed medially and laterally through the knee joint. For chair rising and sitting, the peak forces were 2.65 ± 0.84 kN and 2.60 ± 0.77 kN (Figure 3C–D and Table 1) and the force impulses were 153.5 ± 41.2 kN*% activity cycle and 154.0 ± 36.0 kN*% activity cycle respectively (Table 2).

Mean values of peak forces and force impulses were comparable between study population TKR and instrumented (OrthoLoad) data (Tables 1 and 2). The only significant differences between the study population TKR peak forces and impulses and the instrumented TKR measured peak forces and impulses were for the second peak stair ascent force (p=0.03) and the second peak stair descent force (p=0.03). There were no differences in the variability of the peak forces and impulses between the study population TKR and instrumented TKR measurements as measured by Levene’s test (p=0.106 to p=0.99).

Discussion

The first purpose of this study was to report axial implant forces for chair and stair activities for a population of TKR patients implanted with a modern cruciate-retaining implant design. Axial forces reached an average of 3.1 kN and 2.7 kN during stair ascent and descent, and an average of 2.7 kN and 2.6 kN for chair rising and sitting, respectively. Normalized to body weight, the average peak forces are 3.3 ± 0.5 BW, 3.0 ± 0.4 BW, 2.9 ± 0.8 BW, 2.9 ± 0.6 BW during stair ascent, stair descent, chair sitting, and chair rising, respectively.

The second purpose of this study was to compare the study population TKR kinetics to instrumented TKR kinetics. The only statistically significant differences were that the peak axial forces in the second half of the activity cycle of stair ascent and descent were lower for the study population TKRs than the instrumented TKRs. Levene’s test did not reveal any differences in the variability of the force waveforms between the study population TKRs and instrumented TKRs. Although there were no differences in most peak forces, force variability, and force impulses, the force waveforms appeared visually different between the two prosthesis designs for stair descent and chair sitting. For stair descent, the calculated cruciate-retaining TKR forces were lower than the instrumented knee ultra-congruent cruciate-sacrificing TKR measured forces throughout the activity cycle. Differences in implant design have previously been shown to affect stair climbing kinematics [27], which could in turn result in different forces. In addition, force differences may be due to the test protocol. The subjects in our study did not use handrails to descend stairs where the instrumented subjects may have. Differences in the test staircase may also have affected the axial force output, as the step height in this study is about 0.5 inches taller than the step height of the staircase used in the Bergmann study. For chair sitting, the calculated cruciate-retaining TKR forces were greater than instrumented knee ultra-congruent cruciate-sacrificing TKR measured forces during the first half of the activity cycle. There were a few differences in the chair setup between our laboratory and the OrthoLoad data. Our chair testing setup used a standardized chair with a subject-specific chair height and no armrests, while the videos from OrthoLoad testing suggest that various chairs were used for testing with the instrumented TKR.

This study had several limitations. First, we could not obtain solutions for axial forces with the computational model for a few trials. The model failed to calculate forces during chair sitting for one subject and during all activities for another subject. An examination of the knee kinematics for these trials revealed large offsets for anterior-posterior displacement and internal-external rotation. This is likely an artifact due to errors associated with skin movement in maker-based gait analysis. Second, the model relies on knowledge of the ratio between medial and lateral axial forces to the total axial force passing through the knee. A prior validation study showed that the ratio between medial and lateral axial forces significantly affected force prediction results. The study also identified strategies for proper tuning of this parameter which we used in this study [28]. Although previous investigations have found that the medial-lateral force ratio is correlated with the external knee adduction moment [23,29], it is uncertain as to whether this relationship holds true for all activities and whether it may be influenced by patient characteristics such as knee alignment [30,31]. Specifically, for chair sitting and rising the medial-lateral force ratio of 50%, indicating equal force transfer between the medial and lateral sides of the tibial plateau, was necessary to obtain forces with the computational model. A previous study of instrument prostheses, although with a more constrained prosthesis design, also found the percentage of force transmitted medially through the knee to be lower for chair activities than stair and walking activities [32,31].

Recently, Rasnick et al. predicted the knee joint stair ascent forces for five patients implanted with a posterior stabilized TKR [33]. Average resultant peak contact forces were 2.8 BW and 3.9 BW for the peaks in the first and second half of the stair ascent activity cycle. The value of the first peak compares well to our study but the value of the second is considerably higher than our findings and of that found in instrumented TKRs. Additionally, the force waveform had the opposite pattern where the second peak is higher than the first peak. Our peak forces in this study also compare well to values previously reported for the eKnee instrumented TKR, a cruciate-retaining design. Peak eKnee forces have been reported as ~3.6 BW, 3.5 BW, 1.9 BW, and 2.5 BW for stair ascent, stair descent, chair sitting, and chair rising, respectively [34].

Conclusions

We present forces and motions for stair and chair activities representative of a population of TKR patients implanted with a cruciate-retaining prosthesis design which could be used for displacement-driven knee wear testing. The forces are similar to those in the literature from instrumented prostheses of an ultra-congruent cruciate-sacrificing TKR. Although the variability in this patient cohort was not greater than that of the instrumented knee patients, solely using the recommended waveforms as force inputs may not represent wear of this implant type due to differences in knee kinematics, especially secondary kinematics (anterior-posterior translation and internal-external rotation). Although studies have investigated the difference in wear due to prosthesis design, it is unknown how using inputs for wear testing from populations of TKR patients implanted with different prostheses will affect the outcome of wear tests. However since wear tested simulator components continue to not be representative of wear measured on retrieved TKRs [35,36], work should continue towards defining accurate wear simulation inputs generalizable to multiple TKR patient cohorts who participate in a variety of daily activities.

Supplementary Material

1
2
3
4
5
6

Acknowledgments

We thank Connor Tobin and Steven Mell for technical assistance. This work was supported by the National Institutes of Health (grant numbers R01 AR059843, MAW, R03 AR052039, MAW, and F32 AR057297, HJL), a University of Illinois FMC Education Fund Fellowship (CBK), and the Rush Arthritis and Orthopedics Institute.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • 1.Kurtz S, Ong K, Lau E, Mowat F, Halpern M. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am. 2007;89:780–5. doi: 10.2106/JBJS.F.00222. [DOI] [PubMed] [Google Scholar]
  • 2.Kurtz SM, Ong KL, Lau E, Bozic KJ. Impact of the economic downturn on total joint replacement demand in the United States: Updated projections to 2021. Journal of Bone and Joint Surgery - American Volume. 2014;96:624–30. doi: 10.2106/JBJS.M.00285. [DOI] [PubMed] [Google Scholar]
  • 3.Losina E, Katz JN. Total knee arthroplasty on the rise in younger patients: Are we sure that past performance will guarantee future success? Arthritis Rheum. 2012;64:339–41. doi: 10.1002/art.33371. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.NJR England and Wales. National Joint Registry for England and Wales: Surgical data to 31st December 2011. 2012.
  • 5.Hamilton DF, Howie CR, Burnett R, Simpson AHRW, Patton JT. Dealing with the predicted increase in demand for revision total knee arthroplasty. Bone Joint J. 2015;97-B:723–8. doi: 10.1302/0301-620X.97B6.35185. [DOI] [PubMed] [Google Scholar]
  • 6.ISO 14243-3. Implants for surgery -- wear of total knee-joint prostheses -- part 3: loading and displacement parameters for wear-testing machines with displacement control and corresponding environmental conditions for test. ISO 2014.
  • 7.ASTM WK31128. New test method for physiological durability testing standard - testing for knee activities of daily living. Draft Under Development 2010.
  • 8.ASTM F3141 - 15. Standard Guide for Total Knee Replacement Loading Profiles. ASTM International; 2016. [DOI] [Google Scholar]
  • 9.Bergmann G, Bender A, Graichen F, Dymke J, Rohlmann A, Trepczynski A, et al. Standardized Loads Acting in Knee Implants. PLoS ONE. 2014;9:e86035. doi: 10.1371/journal.pone.0086035. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Schwiesau J, Schilling C, Kaddick C, Utzschneider S, Jansson V, Fritz B, et al. Definition and evaluation of testing scenarios for knee wear simulation under conditions of highly demanding daily activities. Medical Engineering & Physics. 2013;35:591–600. doi: 10.1016/j.medengphy.2012.07.003. [DOI] [PubMed] [Google Scholar]
  • 11.Schwiesau J, Fritz B, Kutzner I, Bergmann G, Grupp TM. CR TKA UHMWPE Wear Tested after Artificial Aging of the Vitamin E Treated Gliding Component by Simulating Daily Patient Activities. BioMed Research International. 2014;2014:1–9. doi: 10.1155/2014/567374. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Reinders J, Sonntag R, Vot L, Gibney C, Nowack M, Kretzer JP. Wear Testing of Moderate Activities of Daily Living Using In Vivo Measured Knee Joint Loading. PLOS ONE. 2015;10:e0123155. doi: 10.1371/journal.pone.0123155. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Orozco DA, Wimmer MA. Development of a Multi-Activity Protocol for TKR Wear Assessment. Transactions of the Annual Meeting of the Orthopaedic Research Society; 2011. p. 1109. [Google Scholar]
  • 14.Bergmann G, editor. OrthoLoad. 2008. [Google Scholar]
  • 15.D’Lima DD, Patil S, Steklov N, Slamin JE, Colwell CW., Jr The Chitranjan Ranawat Award: in vivo knee forces after total knee arthroplasty. Clinical Orthopaedics and Related Research. 2005;440:45–9. doi: 10.1097/01.blo.0000186559.62942.8c. [DOI] [PubMed] [Google Scholar]
  • 16.Alexander EJ, Andriacchi TP. Correcting for deformation in skin-based marker systems. Journal of Biomechanics. 2001;34:355–361. doi: 10.1016/s0021-9290(00)00192-5. [DOI] [PubMed] [Google Scholar]
  • 17.Andriacchi T, Alexander E, Toney M, Dyrby C, Sum J. A point cluster method for in vivo motion analysis: applied to a study of knee kinematics. J Biomech Eng. 1998;120:743–9. doi: 10.1115/1.2834888. [DOI] [PubMed] [Google Scholar]
  • 18.Andriacchi TP, Johnson TS, Hurwitz DE, Natarajan RN. Basic Orthopaedic Biomechanics and Mechano-Biology. Vol. 3. Philadelphia: Lippincott Williams & Wilkins; 2005. Musculoskeletal Dynamics, Locomotion, and Clinical Applications; pp. 1–1. [Google Scholar]
  • 19.Lundberg HJ, Foucher KC, Wimmer MA. A parametric approach to numerical modeling of TKR contact forces. J Biomech. 2009;42:541–5. doi: 10.1016/j.jbiomech.2008.11.030. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Delp SL, Loan JP, Hoy MG, Zajac FE, Topp EL, Rosen JM. An interactive graphics-based model of the lower extremity to study orthopaedic surgical procedures. IEEE Transactions on Biomedical Engineering. 1990;37:757–67. doi: 10.1109/10.102791. [DOI] [PubMed] [Google Scholar]
  • 21.Delp SL, Anderson FC, Arnold AS, Loan P, Habib A, John CT, et al. OpenSim: open-source software to create and analyze dynamic simulations of movement. IEEE Trans Biomed Eng. 2007;54:1940–50. doi: 10.1109/TBME.2007.901024. [DOI] [PubMed] [Google Scholar]
  • 22.Lundberg HJ, Knowlton C, Wimmer MA. Fine tuning total knee replacement contact force prediction algorithms using blinded model validation. J Biomech Eng. 2013:135. doi: 10.1115/1.4023388. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Zhao D, Banks SA, Mitchell KH, D’Lima DD, Colwell CW, Jr, Fregly BJ. Correlation between the knee adduction torque and medial contact force for a variety of gait patterns. J Orthop Res. 2007;25:789–97. doi: 10.1002/jor.20379. [DOI] [PubMed] [Google Scholar]
  • 24.Lundberg HJ, Swanson A, Knowlton C, Inoue N, Wimmer MA. Methods for locating the tibio-femoral contact pathway in total knee replacements using marker-based gait analysis and standard radiography. Iowa Orthop J. 2014;34:94–101. [PMC free article] [PubMed] [Google Scholar]
  • 25.Lundberg HJ, Foucher KC, Andriacchi TP, Wimmer MA. Direct comparison of measured and calculated total knee replacement force envelopes during walking in the presence of normal and abnormal gait patterns. J Biomech. 2012;45:990–6. doi: 10.1016/j.jbiomech.2012.01.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Physical status: the use and interpretation of anthropometry. Report of a WHO Expert Committee. World Health Organization Technical Report Series. 1995;854:1–452. [PubMed] [Google Scholar]
  • 27.Andriacchi TP, Galante JO, Fermier RW. The Influence of Total Knee-Replacement Design on Walking and Stair-Climbing. J Bone Joint Surg Am. 1982;64-A:1328–35. [PubMed] [Google Scholar]
  • 28.Lundberg HJ, Knowlton C, Wimmer MA. Fine Tuning Total Knee Replacement Contact Force Prediction Algorithms Using Blinded Model Validation. Journal of Biomechanical Engineering. 2013;135:21015. doi: 10.1115/1.4023388. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Kutzner I, Trepczynski A, Heller MO, Bergmann G. Knee adduction moment and medial contact force-facts about their correlation during gait. PLoS ONE. 2013:8. doi: 10.1371/journal.pone.0081036. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Halder A, Kutzner I, Graichen F, Heinlein B, Beier A, Bergmann G. Influence of limb alignment on mediolateral loading in total knee replacement: In vivo measurements in five patients. Journal of Bone and Joint Surgery - Series A. 2012;94:1023–9. doi: 10.2106/JBJS.K.00927. [DOI] [PubMed] [Google Scholar]
  • 31.Trepczynski A, Kutzner I, Bergmann G, Taylor WR, Heller MO. Modulation of the relationship between external knee adduction moments and medial joint contact forces across subjects and activities. Arthritis and Rheumatology. 2014;66:1218–27. doi: 10.1002/art.38374. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Varadarajan KM, Moynihan AL, D’Lima D, Colwell CW, Li G. In vivo contact kinematics and contact forces of the knee after total knee arthroplasty during dynamic weight-bearing activities. J Biomech. 2008;41:2159–68. doi: 10.1016/j.jbiomech.2008.04.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Rasnick R, Standifird T, Reinbolt JA, Cates HE, Zhang S. Knee Joint Loads and Surrounding Muscle Forces during Stair Ascent in Patients with Total Knee Replacement. PLoS One. 2016:11. doi: 10.1371/journal.pone.0156282. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Mündermann A, Dyrby CO, D’Lima DD, Colwell J, Andriacchi TP. In vivo knee loading characteristics during activities of daily living as measured by an instrumented total knee replacement. Journal of Orthopaedic Research. 2008;26:1167–72. doi: 10.1002/jor.20655. [DOI] [PubMed] [Google Scholar]
  • 35.Harman MK, DesJardins J, Benson L, Banks SA, LaBerge M, Hodge WA. Comparison of polyethylene tibial insert damage from in vivo function and in vitro wear simulation. Journal of Orthopaedic Research. 2009;27:540–8. doi: 10.1002/jor.20743. [DOI] [PubMed] [Google Scholar]
  • 36.Orozco Villasenor D, Wimmer MA. Wear Scar Similarities between Retrieved and Simulator-Tested Polyethylene TKR Components - An Artificial Neural Network Approach. BioMed Res Int. 2016 doi: 10.1155/2016/2071945. In Press. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

1
2
3
4
5
6

RESOURCES