Abstract
Two semiportable metal air incinerators, each with a capacity of 1,000 to 2,200 standard ft3 of air per min, were constructed to sterilize infectious aerosols created for investigative work in a microbiological laboratory. Each unit has about the same air-handling capacity as a conventional air incinerator with a brick stack but costs only about one-third as much. The units are unique in that the burner housing and combustion chamber are air-tight and utilize a portion of the contaminated air stream to support combustion of fuel oil. Operation is continuous. Aerosols of liquid and dry suspensions of Bacillus subtilis var. niger spores and dry vegetative cells of Serratia marcescens were disseminated into the two incinerators to determine the conditions required for sterilization of contaminated air. With the latter organisms (concentration 2.03 × 107 cells/ft3 of air), a temperature of 525 F (274 C), measured at the firebox in front of the heat exchanger, was sufficient for sterilization. To sterilize 1.74 × 107 and 1.74 × 109 wet spores of B. subtilis per ft3, the required temperature ranged from 525 to 675 F (274 to 357 C) and 625 to 700 F (329 to 371 C), respectively. Air-sterilization temperature varied with each incinerator. This was because of innate differences of fabrication, different spore concentrations, and use of one or two burners With dry B. subtilis spores (1.86 × 108/ft3), a temperature of 700 F was required for sterilization. With dry spores, no difference was noted in the sterilization temperature for the two incinerators.
Full text
PDF





Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Barbeito M. S. Emergency disinfection of operating room and patient ward with beta-propiolactone. Hospitals. 1966 Mar 16;40(6):100–106. doi: 10.21236/ad0640342. [DOI] [PubMed] [Google Scholar]
- DECKER H. M., CITEK F. J., HARSTAD J. B., GROSS N. H., PIPER F. J. Time temperature studies of spore penetration through an electric air sterilizer. Appl Microbiol. 1954 Jan;2(1):33–36. doi: 10.1128/am.2.1.33-36.1954. [DOI] [PMC free article] [PubMed] [Google Scholar]
- GREMILLION G. G., MILLER L. F., BODMER G. A. An electric incinerator for sterilization of small volumes of air. Appl Microbiol. 1958 Jul;6(4):274–276. doi: 10.1128/am.6.4.274-276.1958. [DOI] [PMC free article] [PubMed] [Google Scholar]
- REITMAN M., FRANK M. A., Sr, ALG R., MILLER W. S. Modifications of the high-speed safety blendor. Appl Microbiol. 1954 May;2(3):173–173. doi: 10.1128/am.2.3.173-173.1954. [DOI] [PMC free article] [PubMed] [Google Scholar]
- WEDUM A. G. LABORATORY SAFETY IN RESEARCH WITH INFECTIOUS AEROSOLS. Public Health Rep. 1964 Jul;79:619–633. [PMC free article] [PubMed] [Google Scholar]