Abstract
Purpose
In the last two decades, thyroglobulin autoantibodies (TgAb) measurement has progressively switched from marker of thyroid autoimmunity to test associated with thyroglobulin (Tg) to verify the presence or absence of TgAb interference in the follow-up of patients with differentiated thyroid cancer. Of note, TgAb measurement is cumbersome: despite standardization against the International Reference Preparation MRC 65/93, several studies demonstrated high inter-method variability and wide variation in limits of detection and in reference intervals. Taking into account the above considerations, the main aim of the present study was the determination of TgAb upper reference limit (URL), according to the National Academy of Clinical Biochemistry guidelines, through the comparison of eleven commercial automated immunoassay platforms.
Methods
The sera of 120 healthy males, selected from a population survey in the province of Verona, Italy, were tested for TgAb concentration using eleven IMA applied on as many automated analyzers: AIA-2000 (AIA) and AIA-CL2400 (CL2), Tosoh Bioscience; Architect (ARC), Abbott Diagnostics; Advia Centaur XP (CEN) and Immulite 2000 XPi (IMM), Siemens Healthineers; Cobas 6000 (COB), Roche Diagnostics; Kryptor (KRY), Thermo Fisher Scientific BRAHMS, Liaison XL (LIA), Diasorin; Lumipulse G (LUM), Fujirebio; Maglumi 2000 Plus (MAG), Snibe and Phadia 250 (PHA), Phadia AB, Thermo Fisher Scientific. All assays were performed according to manufacturers’ instructions in six different laboratories in Friuli-Venezia Giulia and Veneto regions of Italy [Lab 1 (AIA), Lab 2 (CL2), Lab 3 (ARC, COB and LUM), Lab 4 (CEN, IMM, KRY and MAG), Lab 5 (LIA) and Lab 6 (PHA)]. Since TgAb values were not normally distributed, the experimental URL (e-URL) was established at 97.5 percentile according to the non-parametric method.
Results
TgAb e-URLs showed a significant inter-method variability. Considering the same method, e-URL was much lower than that suggested by manufacturers (m-URL), except for ARC and MAG. Correlation and linear regression were unsatisfactory. Consequently, the agreement between methods was poor, with significant bias in Bland–Altman plot.
Conclusions
Despite the efforts for harmonization, TgAb methods cannot be used interchangeably. Therefore, additional effort is required to improve analytical performance taking into consideration approved protocols and guidelines. Moreover, TgAb URL should be used with caution in the management of differentiated thyroid carcinoma patients since the presence and/or the degree of TgAb interference in Tg measurement has not yet been well defined.
Keywords: Autoimmune thyroid disease, Thyroglobulin autoantibodies, Upper reference limit, Immunoassay, Harmonization
Introduction
Human thyroglobulin (Tg) is a high molecular weight (660 kDa) soluble glycoprotein, typically stored within the follicular colloid of the thyroid, acting as the substrate for thyroid hormones (triiodothyronine, T3 and thyroxine, T4). As Tg is produced and utilized entirely by benign or differentiated malignant thyroid cells, it is considered a good tumor marker for patients with differentiated thyroid carcinoma (DTC) [1, 2] after removal of benign and malignant thyroid tissue by surgery and I131 ablation. Over the years, advances in assay technologies have led to important improvements in the analytical performances of Tg immunometric assays (IMAs); above all, the functional sensitivity (FS) of Tg IMAs has greatly improved: from 0.5 to 1.0 μg/L of the first generation IMAs to 0.05–0.10 μg/L of the second generation (2G) IMAs [3].
Nevertheless, the major limitation of 2G IMA testing is interference by serum Tg autoantibodies (TgAb) causing, as a rule, underestimation of Tg results and possibly masking disease recurrence [4–6]: it has been hypothesized that the complex between free Tg and endogenous TgAb prevents free Tg from binding to the capture and/or signal monoclonal antibody reagents and/or alternatively, endogenous TgAb binding to free Tg masks the epitopes recognized by monoclonal antibody reagents [5, 7].
Serum TgAb are reported to be present in about 25–30% of DTC patients depending of the assay used and the cut-off employed to classify samples as positive or negative [1, 7]. They are more frequent in females [8] and they are also present in about 60% of patients with autoimmune thyroid disease (AITD) [9]. On the basis of these considerations, the role of TgAb measurement has evolved from a marker of thyroid autoimmunity [10, 11] to a test associated with Tg to investigate TgAb interference [12]. Consequently, serum TgAb have evolved as a surrogate test for tumor marker replacing Tg determination by IMAs, in cases of analytical interference from TgAb [13, 14].
Of note, the measurement of TgAb could be cumbersome. Analytical limitations of serum TgAb assays have been reported in the context of thyroid autoimmunity diagnosis [9]. Despite standardization against the International Reference Preparation (IRP) MRC 65/93, several studies demonstrated a high variability in the analytical performances of different TgAb IMAs: large variation in limits of detection (LOD), FS, inter-method results, reference intervals with poor concordance between TgAb assays in patients with DTC [15–22]. The difficulty in standardization is in part due to the heterogeneous Tg immunoreactivity: differential splicing of Tg mRNA, various post-translational modifications, and alterations of biosynthesis regulation in thyroid tumor cells lead to exposure or masking of epitopes with resulting differences in Tg immunologic structure [23]. Besides Tg heterogeneity, assay discordance has also been assigned to various specificity of circulating TgAb in patient sera [6]. As a result, different TgAb values are obtained when the same serum is tested with different methods [15–22]. Finally, differences in assay reagents, above all the preparation of the antigen (Tg), definitely contribute to assay variability [1–12].
The manufacturers’ upper reference limit (URL) for TgAb, set up to identify patients with AITD but misleading for evaluation of TgAb interference in Tg assay, is another aspect to consider. Reference intervals are the most widely used tool for the interpretation of clinical laboratory results. The Clinical and Laboratory Standards Institute (CLSI) Expert Panel on Reference Values has provided guidelines for the determination of reliable reference intervals (EP28-A3c) [24]. They recommended the use of the direct method, which implies the enrolment of a healthy population of at least 120 individuals and the determination of 2.5th and 97.5th percentile for the lower reference limit and the URL, respectively. As regards thyroid antibodies (thyroid peroxidase antibodies—TPOAb and TgAb) for AITD diagnosis, the 2003 proposal of the National Academy of Clinical Biochemistry (NACB) recommends the use of a direct method and a reference group composed of 120 men younger than 30 years, biochemically euthyroid [i.e., with serum thyrotropin stimulating hormone (TSH), concentrations between 0.5 and 2.0 mIU/L], and without risk parameters (goiter, family history of AITD, or other autoimmune diseases) [25].
However, the definition of the TgAb URL remains a matter of debate, because of the problems in enrolling the appropriate reference group [25] and in the determination of TgAb cut-off suitable for the identification of assay interference and consequently for the use of TgAb as surrogate marker in the follow-up of DTC [12].
Taking into account the above considerations, the main aim of the present study was the determination of TgAb URL, according to the NACB guidelines, by the use of eleven commercial automated IMA platforms. A further aim of the study was to compare the analytical performances of the methods used, in an attempt to evaluate, whenever possible, their effectiveness in detecting TgAb interference.
Materials and methods
One hundred and twenty male subjects were selected from a population survey in the province of Verona, Italy, according to the NACB criteria [25]. All of them gave informed consent for their participation in the study. Their sera were tested for TgAb concentration by using eleven IMA methods applied in as many automated analyzers: AIA-2000 (AIA) and AIA-CL2400 (CL2), Tosoh Bioscience; Architect (ARC), Abbott Diagnostics; Advia Centaur XP (CEN) and Immulite 2000 XPi (IMM), Siemens Healthineers; Cobas 6000 (COB), Roche Diagnostics; Kryptor (KRY), Thermo Fisher Scientific BRAHMS, Liaison XL (LIA), Diasorin; Lumipulse G (LUM), Fujirebio; Maglumi 2000 Plus (MAG), Snibe and Phadia 250 (PHA), Phadia AB, Thermo Fisher Scientific. All assays were performed according to manufacturers’ instructions at six different laboratories in Friuli-Venezia Giulia and Veneto regions of Italy [Lab 1 (AIA), Lab 2 (CL2), Lab 3 (ARC, COB and LUM), Lab 4 (CEN, IMM, KRY and MAG), Lab 5 (LIA) and Lab 6 (PHA)]. The main features of the eleven methods are summarized in Table 1. All methods are standardized with the reference preparation (IRP MRC 65/93) and use International Units (IU), except for CEN and KRY whose results were initially expressed in Arbitrary Units and then converted in IU (Table 1). The normality of the distribution was assessed using the Shapiro–Wilk test. Since TgAb values were not normally distributed, the experimental URL (e-URL) was established at 97.5th according to the non-parametric percentile method (CLSI standard C28-A3c) [24]. Moreover, the non-parametric Kruskal–Wallis test and the Dunn’s multiple comparison test were used for comparing the median values of the eleven groups.
Table 1.
Method | Immunoassay principle | Tracer/enzyme | Assay type | Imprecision (%):intra-; inter-; total | LoDd (IU/mL) | LoQd (IU/mL) | Assay range (IU/mL) |
---|---|---|---|---|---|---|---|
AIA | FEIA | 4MUP/ Alkaline phosphatase |
NC | 4.3–5.1; nd; 5.5–6.0 |
0.12 | nd | 0.12–2000 |
ARC | CLIA | Acridinium esters | NC | 1.7–6.6b; nd; 2.7–8.2b |
0.07 | 0.31 | 0.07–1000 |
CENa | CLIA | Acridinium esters | C | 2.9–5.5; 1.8–2.0; 3.5–5.8 |
10 | 30e | 10–500 |
CL2 | CLEIA | Difurat® | NC | 5.1–5.5; 5.8–6.6; nd |
0.005 | nd | 0.005–2500 |
COB | ECLIA | Ruthenium derivatives | C | 1.3–5.6c; 2.1–8.7c; nd |
10 | nd | 10–4000 |
IMM | CLIA | Adamantyl dioxetane phosphate/Alkaline phosphatase | NC | 3.2–4.9; 4.6–5.8; nd |
2.2 | nd | 20–3000 |
KRYa | TRACE | Europium cryptate/ XL 665 |
C | 1.5–3.5; 6.8–20.0; nd |
10 | 33 | 10–850 |
LIA | CLIA | Isoluminol derivatives | NC | 2.3–3.2; 4.4–8.9; nd |
5 | 10 | 5–5000 |
LUM | CLEIA | AMPPD | NC | 1.8–4.6; nd; 2.5–5.3c |
5.152 | 5.152 | 5.152–3000 |
MAG | CLIA | ABEI | NC | 2.8–9.1; 5.2–9.8; nd |
10 | nd | 10–2800 |
PHA | FIA | 4-methyl-umbellipheryl-β-d-galactoside/β-galactosidase | NC | 3.3–5.6; 2.6–6.5; nd |
12 | nd | 12–4794 |
4MUP 4-methyl-umbelliferyl phosphate, ABEI N-(aminobutil)-N-(ethyl)-isoluminol, AIA AIA-2000, Tosoh Bioscience, AMPPD alkaline phosphatase-spiroadamantyl-methoxy-phosphoryloxy-phenyl-dioxetane, ARC Architect, Abbott Diagnostics, C competitive immunoassay, CEN Advia Centaur XP, Siemens Healthineers, CL2 AIA CL-2400, Tosoh Bioscience, CLIA chemiluminescence immunoassay, CLEIA chemiluminescence enzyme immunoassay, COB Cobas 6000, Roche Diagnostics, ECLIA electrochemiluminescence immunoassay, FEIA fluorescence enzyme immunoassay, FIA fluoroimmunoassay, IMM Immulite 2000 XPi, Siemens Healthineers, KRY Kryptor, Thermo Fisher Scientific BRAHMS, LIA Liaison XL, Diasorin, LUM Lumipulse G, Fujirebio, MAG Maglumi 2000 Plus, Snibe, NC non-competitive immunoassay, nd not declared, PHA Phadia 250, Phadia AB, Thermo Fisher Scientific, TRACE time resolved amplified cryptate emission
aAll methods are standardized with the reference preparation MRC 65/93 and use International Units (IU/mL) except for Centaur and Kryptor which refer to a secondary standard and use Arbitrary Units (AU/mL); to obtain IU multiply for the conversion factor 2.8 (CEN) and 7.14 (KRY)
bPrecision defined by the NCCLS Protocol EP5-A [26]
cPrecision defined by the modified NCCLS Protocol EP5-A2 [27]
dLoD and LoQ defined by the CLSI protocol EP17-A [28]
eFunctional sensitivity defined as TgAb concentration with total CV ≤20%, determined for a period of two days using one lot of reagents and testing, by four instruments, multiple samples from normal patients
The inter-method variability was assessed considering the interquartile range (25th and 75th percentile). To compare the eleven methods, ARC was regarded as the reference assay since it showed a satisfactory combination between the LoD and the assay imprecision (Table 1). Correlation between assays was assessed by Spearman Rank correlation coefficient (r s); Passing-Bablok regression was applied to verify the linear association between methods, while agreement between assays was analyzed by Bland–Altman plot considering the difference between ARC and the other ten methods (AIA, CEN, CL2, COB, IMM, KRY, LIA, LUM, MAG and PHA). The difference between manufacturer’s URL (m-URL) and e-URL was expressed as the ratio between them in percentage (Delta% = |m-URL − e-URL|/m-URL × 100). A two-sided value of p < 0.05 was considered statistically significant. Statistical analyses were performed by GraphPad Prism Software, version 4.0 (San Diego, CA, USA) and MedCalc software, version 11.6 (Ostend, Belgium).
Results
TgAb results showed a relevant inter-method variability with wide interquartile ranges: the difference reached 48 times for the 25th percentile (minimum: 0.24 IU/mL and maximum: 11.5 IU/mL) and 30 times for 75th percentile (minimum: 0.59 IU/mL, maximum: 17.97 IU/mL) (Fig. 1) (Table 2).
Table 2.
Method | AIA | ARC | CENa | CL2 | COB | IMM | KRYa | LIA | LUM | MAG | PHA |
---|---|---|---|---|---|---|---|---|---|---|---|
No. | 120 | 120 | 120 | 120 | 120 | 120 | 120 | 120 | 120 | 120 | 120 |
Mean (95% CI) |
1.56 1.18–1.95 |
1.81 1.58–2.03 |
10.93 9.74–12.13 |
0.53 0.43–0.63 |
14.23 12.87–15.58 |
1.28 1.04–1.53 |
10.79 9.43–12.14 |
10.38 9.57–11.2 |
7.94 7.2–8.69 |
15.61 14.19–17.03 |
12.05 11.09–13.01 |
SD | 2.12 | 1.26 | 6.61 | 0.56 | 7.52 | 1.36 | 7.48 | 4.49 | 4.12 | 7.87 | 5.33 |
RSD | 1.36 | 0.7 | 0.6 | 1.06 | 0.53 | 1.05 | 0.69 | 0.43 | 0.52 | 0.50 | 0.44 |
Median (95% CI) |
0.91 0.8–1.08 |
1.51 1.39–1.61 |
9.61 8.81–10.6 |
0.4 0.32–0.46 |
12.56 11.93–12.84 |
1.03 0.81–1.21 |
8.41 7.27–10.49 |
9.4 9.11–9.81 |
6.87 6.36–7.58 |
13.35 12.38–15.07 |
11 10–11 |
25th–75th p | 0.63–1.48 | 1.21–1.92 | 6.96–13.83 | 0.24–0.59 | 11.5–13.61 | 0.36–1.65 | 5.04–15.15 | 8.66–10.42 | 5.63–8.64 | 10.05–17.97 | 10–13.5 |
2.5th–97.5th p | 0.36–6.75 | 0.73–5.14 | 1.64–26.25 | 0.1–2.25 | 10.31–41.15 | 0.02–4.23 | 1.22–29.94 | 7.50–21.79 | 5.15–20.56 | 10.05–34.23 | 6–24.5 |
Normal distribution | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 |
Mean, SD, median, percentiles are expressed in IU/mL
AIA AIA-2000, Tosoh Bioscience, ARC Architect, Abbott Diagnostics, CEN Advia Centaur XP, Siemens Healthineers, CI confidence intervals, CL2 AIA CL-2400, Tosoh Bioscience, COB Cobas 6000, Roche Diagnostics, IMM Immulite 2000 XPi, Siemens Healthineers, KRY Kryptor, Thermo Fisher Scientific BRAHMS, LIA Liaison XL, Diasorin, LUM Lumipulse G, Fujirebio, MAG Maglumi 2000 Plus, Snibe, No. number, p percentile, PHA Phadia 250, Phadia AB, Thermo Fisher Scientific, RSD relative standard deviation, SD standard deviation
aThe results obtained with Advia Centaur XP and Kryptor Compact Plus are expressed as arbitrary units (AU/mL); to obtain IU multiply for the conversion factor 2.8 (CEN) and 7.14 (KRY)
A statistically significant difference between medians was observed for all methods except for 11 pairs of the 45 combinations analyzed (Fig. 1) (Table 3).
Table 3.
A. Kruskal–Wallis test | |
---|---|
No. of groups | 11 |
p Value | p < 0.0001 |
Do the medians vary significantly (p < 0.05) | Yes |
B. Dunn’s multiple comparison test | p value |
---|---|
AIA vs ARC | p > 0.05 |
AIA vs IMM | p > 0.05 |
ARC vs IMM | p > 0.05 |
CEN vs KRY | p > 0.05 |
CEN vs LIA | p > 0.05 |
CEN vs LUM | p > 0.05 |
CL2 vs IMM | p > 0.05 |
COB vs MAG | p > 0.05 |
KRY vs LIA | p > 0.05 |
KRY vs LUM | p > 0.05 |
LIA vs LUM | p > 0.05 |
A There is a significant difference between the medians of the eleven groups. B In the table are reported only the comparison of groups with non-significant differences
AIA AIA-2000, Tosoh Bioscience, ARC Architect, Abbott Diagnostics, CEN Advia Centaur XP, Siemens Healthineers, CI confidence intervals, CL2 AIA CL-2400, Tosoh Bioscience, COB Cobas 6000, Roche Diagnostics, IMM Immulite 2000 XPi, Siemens Healthineers, KRY Kryptor, Thermo Fisher Scientific BRAHMS, LIA Liaison XL, Diasorin, LUM, Lumipulse G, Fujirebio, MAG, Maglumi 2000 Plus, Snibe, No. number, PHA Phadia 250, Phadia AB, Thermo Fisher Scientific
e-URLs differed from one method to the other. Of note, within the same method, e-URL was much lower than m-URL, except for ARC and MAG, which showed similar values for both (Table 4).
Table 4.
Method | No. | m-URL (IU/mL) | e-URL (C.I. 90%) (IU/mL) | Delta (%) |
---|---|---|---|---|
AIA | 120 | 13.6 | 6.82 (5–15.7) | 49.85 |
ARC | 120 | 4.11 | 5.66 (3.29–11.64) | 37.71 |
CENa | 120 | 60 | 27.44 (21.9–52) | 54.27 |
CL2 | 120 | 6.8 | 2.63 (1.15–4.08) | 61.32 |
COB | 120 | 115 | 43.69 (21.16–70.88) | 62 |
IMM | 120 | 40 | 4.46 (3.26–9.41) | 88.85 |
KRYa | 120 | 33 | 32.23 (25.14–41.61) | 2.33 |
LIA | 120 | 100 | 24.93 (14.37–43.91) | 75.07 |
LUM | 120 | 55.4 | 21.07 (12.3–32.8) | 61.97 |
MAG | 120 | 30 | 34.23 (29.46–69.36) | 14.1 |
PHA | 120 | 60 | 25.93 (21–48) | 56.78 |
e-URL: 97.5th percentile; Delta = |m-URL − e-URL|/m-URL × 100
AIA AIA-2000, Tosoh Bioscience, ARC Architect, Abbott Diagnostics, CEN Advia Centaur XP, Siemens Healthineers, CI confidence intervals, CL2 AIA CL-2400, Tosoh Bioscience, COB Cobas 6000, Roche Diagnostics, e-URL experimental upper reference limit, IMM Immulite 2000 XPi, Siemens Healthineers, KRY Kryptor, Thermo Fisher Scientific BRAHMS, LIA Liaison XL, Diasorin, LUM Lumipulse G, Fujirebio, m-URL manufacturer upper reference limit, MAG Maglumi 2000 Plus, Snibe, No. number, PHA Phadia 250, Phadia AB, Thermo Fisher Scientific
aThe results obtained with Advia Centaur XP and Kryptor Compact Plus are expressed as arbitrary units (AU/mL); to obtain IU multiply for the conversion factor 2.8 (CEN) and 7.14 (KRY)
As regards the correlations between methods, r s ranged from 0.17 (ARC vs CEN) to 0.56 (ARC vs CL2) (Table 5). Using Passing-Bablok analysis, TgAb method comparison resulted in varying degrees of agreement with the reference method (ARC). Slopes were all far from 1 except for ARC vs AIA (slope = 1.15) and ARC vs CL2 (0.34) (Fig. 2) (Table 5); intercepts varied from −29.92 to 3.7, they were far from 0 except for ARC vs AIA (−0.75) and ARC vs CL2 (−0.15) (Fig. 2) (Table 5). Subsequently, a relevant positive or negative mean biases were observed by Bland–Altman analysis ranging from −115.8% (CL2 vs ARC) to 156.4% (MAG vs ARC). The best agreement was between AIA and ARC with a mean bias of −37% (Fig. 3) (Table 6).
Table 5.
Comparison | ARC vs AIA | ARC vs CEN | ARC vs CL2 | ARC vs COB | ARC vs IMM | ARC vs KRY | ARC vs LIA | ARC vs LUM | ARC vs MAG | ARC vs PHA |
---|---|---|---|---|---|---|---|---|---|---|
No. | 120 | 120 | 120 | 120 | 120 | 120 | 120 | 120 | 120 | 120 |
Slope (95% CI) |
1.15 (0.77–2.02) |
26.56 (16.24–63.2) |
0.34 (0.26–0.44) |
5.98 (0.32 – 8.12) |
1.92 (1.42–2.64) |
38.36 (21.22–94.83) |
5.45 (3.48–7.52) |
7.03 (5.27–10.57) |
17.99 (11.86–32.26) |
8.33 (5.77–13.33) |
y-Intercept (95% CI) | −0.75 (−1.92 to −0.26) |
−29.92 (−85.36 to −14.53) |
−0.15 (−0.29 to −0.03) |
3.7 (0.67–6.39) |
−1.9 (−2.96 to −1.17) |
−47.31 (−132.72 to −22.98) |
1.51 (−1.58 to 4.22) |
3.51 (−8.71 to −0.77) |
−13.14 (−33.37 to −4.24) |
−1.63 (−9.07 to 2.15) |
Equation | y = −0.75 + 1.15x | y = −29.92 + 26.56x | y = −0.15 + 0.34x | y = 3.7 + 5.98x | y = −1.9 + 1.92x | y = −47.31 + 38.36x | y = 1.51 + 5.45x | y = −3.51 + 7.03x | y = −13.14 + 17.99x | y = −1.63 + 8.33x |
r s (95% CI) | 0.41 (0.24–0.55) |
0.17 (−0.01–0.34) |
0.56 (0.42–0.67) |
0.49 (0.34–0.61) |
0.28 (0.1–0.43) |
0.19 (0.01–0.35) |
0.28 (0.11–0.44) |
0.18 (−0.01–0.35) |
0.23 (0.05–0.39) |
0.35 (0.18–0.49) |
The y-intercept is expressed as IU/mL
AIA, AIA-2000, Tosoh Bioscience, ARC Architect, Abbott Diagnostics, CEN Advia Centaur XP, Siemens Healthineers, CI confidence intervals, CL2 AIA CL-2400, Tosoh Bioscience, COB Cobas 6000, Roche Diagnostics, e-URL experimental upper reference limit, IMM Immulite 2000 XPi, Siemens Healthineers, KRY Kryptor, Thermo Fisher Scientific BRAHMS, LIA Liaison XL, Diasorin, LUM Lumipulse G, Fujirebio, m-URL manufacturer upper reference limit, MAG Maglumi 2000 Plus, Snibe, No. number, PHA Phadia 250, Phadia AB, Thermo Fisher Scientific, r s Spearman’s rank correlation coefficient
Table 6.
Agreement | AIA vs ARC | CEN vs ARC | CL2 vs ARC | COB vs ARC | IMM vs ARC | KRY vs ARC | LIA vs ARC | LUM vs ARC | MAG VS ARC | PHA vs ARC |
---|---|---|---|---|---|---|---|---|---|---|
No. | 120 | 120 | 120 | 120 | 120 | 120 | 120 | 120 | 120 | 120 |
Bias % (±1.96 SD) |
−37 (78.7 to −152.6) |
133.6 (219.2 – 48) |
−115.8 (−45.1 to −186.6) |
155.1 (194.3–115.9) |
−54.6 (106.8–216.1) |
125.8 (226.7–24.8) |
141.8 (189.5–94) |
125 (188.3–61.7) |
156.4 (201.7–111) |
147.2 (190.1–104.2) |
AIA AIA-2000, Tosoh Bioscience, ARC Architect, Abbott Diagnostics, CEN Advia Centaur XP, Siemens Healthineers, CI confidence intervals, CL2 AIA CL-2400, Tosoh Bioscience, COB Cobas 6000, Roche Diagnostics, e-URL experimental upper reference limit, IMM Immulite 2000 XPi, Siemens Healthineers, KRY Kryptor, Thermo Fisher Scientific BRAHMS, LIA Liaison XL, Diasorin, LUM Lumipulse G, Fujirebio, m-URL manufacturer upper reference limit, MAG Maglumi 2000 Plus, Snibe, No. number, PHA Phadia 250, Phadia AB, Thermo Fisher Scientific, r s Spearman’s rank correlation coefficient
Discussion
The determination of the cut-off for the definition of TgAb positivity is an important and controversial issue.
In this study, we have determined the TgAb URL in a reference group of male individuals, meticulously defined as being free of thyroid diseases, by eleven IMA methods, currently used in autoimmunology laboratories, and compared to each other. Actually, to our knowledge, no similar data are present in literature: in the past, other studies faced the same topic but with small numbers of different analytical methods, most of which are no longer in use [9, 15–22, 29].
The first relevant result of the present study was the demonstration of differences between TgAb URLs claimed in the package insert (m-URL) and those obtained in the male reference sample (e-URL): with the exception of ARC and MAG method, e-URLs were lower than those proposed by the manufacturers, the difference ranging from 2.33 to 88.85%. These results were similar to those described in two previous studies dealing with the definition of TPOAb reference limits, determined by several current IMA platforms [30, 31]. In our opinion, these discrepancies could be related to the lack of strict criteria in the selection of the subjects for the reference group. Specifically, racial differences could play some role, as most of the studies, sponsored by manufacturers, were performed in the geographical area of the production line and consequently difficult to reproduce in other settings. Moreover, the use of non-stringent criteria in the choice of subjects could have led to the enrolment of individuals with subclinical AITD, thus resulting in relatively high levels of TgAb causing the raise of the 97.5th percentile of the reference value distribution platforms [32–37].
The second relevant consideration that emerged from the present study was the variation of e-URLs according to the method used. The e-URL ranged from 2.25 (CL2) to 41.15 IU/mL (COB), with an approximately 18-fold variation, consistent with a previous paper which reported the same magnitude of variation using five IMA methods distinct from those considered in the present study (18). The difference between e-URLs supports concerns regarding inter-method variation [38]. Specifically, there were relevant differences between methods in terms of medians (31-fold) (p < 0.05, Kruskal–Wallis test) and interquartile ranges. These discrepancies were not expected and not easily explained; in fact, in recent decades, there have been significant improvements in harmonization between methods [39], resulting from the high level of automation of analytical procedures and the use of the same reference preparation (IRP MRC 65/93). Moreover, analytical imprecision seems not contribute to the above differences, as the values declared by the individual manufacturer were essentially overlapping (although obtained with different protocols, some of them standardized, some others not) and in general lower than 10% for both intra- and inter-assay imprecision (Table 1). Such discordance between TgAb assays could be attributed to various factors, including: (1) TgAb heterogeneity which is often independent to standardization efforts, and which implies different specificity for Tg antigen; (2) Tg interference and (3) differences in assay reagents, including solid phase material and the preparations of the antigen (Tg), which could affect the proper exposure of the immunodominant epitopes. Another important aspect to consider, to explain inter-method variability, was the diverse assay structures of the eleven IMA methods leading to a different LoD (Table 1) ranging from 0.005 to 12 IU/mL. Especially, a clear-cut discrepancy between methods with a LoD lower than 0.2 IU/mL (ARC, AIA and CL2) and methods with a LoD equal to or higher than 2 IU/mL was apparent.
To better evaluate the relationship between methods, ARC was chosen as the reference method on the basis of the best combination between LoD and imprecision (Table 1): the correlation of ARC with the other methods was not satisfactory, in line with the variability of the results, broadly described above. Passing-Bablok regression did not show a satisfactory agreement between assays. Furthermore, consistent with regression results, Bland–Altman plot highlighted a statistically significant positive or negative mean biases.
The lack of acceptable agreement between methods has relevant practical implications: clinicians have to use the same method to monitor TgAb concentration in the follow-up of DTC, on the other hand, laboratories must keep users timely informed about any modification in TgAb method to simplify re-baselining.
Despite the analysis of the data showed satisfactory analytical performances of some methods in terms of LoD, being able to measure also low levels of TgAb with adequate precision, the main limitation to this study lay in having contributed only indirectly to the debated question of TgAb analytical interference. In fact, the obtained results did not prove but only suggested the opportunity to choose the more sensitive and accurate latest generation methods for measuring TgAb, to better detect any false negative results even in patients with TgAb levels lower than the cut-off (the so-called “negative patient”). Therefore, according to these considerations, two different cut-offs for TgAb could be proposed, one for the diagnosis of AITD and one for the effects of TgAb on Tg measurement.
Conclusions
In spite of the attempt of harmonization, quantitative agreement between methods was generally not satisfactory and methods could not be used interchangeably. Therefore, additional standardization efforts are required to improve analytical performance, and biomedical industries are strongly invited to re-evaluate their assays taking into consideration CLSI approved protocols and guidelines.
Finally, as long as the relationship between TgAb concentration and interference in Tg measurement is not clearly defined, TgAb URL must be used with caution, taking into account that it is usually set for the diagnosis of AITD and not for the identification of potential interference in Tg assay.
Acknowledgements
We thank all participant centres for their excellent technical support and the industry’s manufacturers for providing reagents.
Abbreviations
- 2G
Second generation
- AIA
AIA-2000, Tosoh Bioscience
- ARC
Architect, Abbott Diagnostics
- CEN
Advia Centaur XP, Siemens Healthineers
- CL2
AIA-CL2400 (CL2), Tosoh Bioscience
- CLSI
Clinical and Laboratory Standards Institute
- COB
Cobas 6000, Roche Diagnostics
- DTC
Differentiated thyroid carcinoma
- FS
Functional sensitivity
- IMA
Immunometric assay
- IMM
Immulite 2000 XPi, Siemens Healthineers
- KRY
Kryptor, Thermo Fisher Scientific BRAHMS
- LIA
Liaison XL, Diasorin
- LOD
Limit of detection
- LUM
Lumipulse G, Fujirebio
- MAG
Maglumi 2000 Plus, Snibe
- NACB
National Academy of Clinical Biochemistry
- PHA
Phadia 250, Phadia AB, Thermo Fisher Scientific
- Tg
Thyroglobulin
- TgAb
Thyroglobulin autoantibodies
- TPOAb
Thyroid peroxidase antibodies
- URL
Upper reference limit
Compliance with ethical standards
Ethical approval
The Thyroid Study, conducted in the province of Verona, was approved by the local ethics committee.
Informed consent
Informed consent was obtained from all the participants in the study.
Conflict of interest
The authors declare that they have no conflict of interest.
References
- 1.Evans C, Tennant S, Perros P. Thyroglobulin in differentiated thyroid cancer. Clin Chim Acta. 2015;444:310–317. doi: 10.1016/j.cca.2014.10.035. [DOI] [PubMed] [Google Scholar]
- 2.Grebe SK. Diagnosis and management of thyroid carcinoma: focus on serum thyroglobulin. Expert Rev Endocrinol Metab. 2009;4:25–43. doi: 10.1586/17446651.4.1.25. [DOI] [Google Scholar]
- 3.Spencer C, Petrovic I, Fatemi S, LoPresti J. Serum thyroglobulin (Tg) monitoring of patients with differentiated thyroid cancer using sensitive (second-generation) immunometric assays can be disrupted by false-negative and false-positive serum thyroglobulin autoantibody misclassifications. J Clin Endocrinol Metab. 2014;99:4589–4599. doi: 10.1210/jc.2014-1203. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Giovanella L, Feldt-Rasmussen U, Verburg FA, et al. Thyroglobulin measurement by highly sensitive assays: focus on laboratory challenges. Clin Chem Lab Med. 2015;53:1301–1314. doi: 10.1515/cclm-2014-0813. [DOI] [PubMed] [Google Scholar]
- 5.Locsei Z, Szabolcs I, Rácz K, et al. Serum thyroglobulin antibody levels within or near to the reference range may interfere with thyroglobulin measurement. Biochem Med. 2012;22:365–370. doi: 10.11613/BM.2012.038. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Spencer C, Petrovic I, Fatemi S. Current thyroglobulin autoantibody (TgAb) assays often fail to detect interfering TgAb that can result in the reporting of falsely low/undetectable serum Tg IMA values for patients with differentiated thyroid cancer. J Clin Endocrinol Metab. 2011;96:1283–1291. doi: 10.1210/jc.2010-2762. [DOI] [PubMed] [Google Scholar]
- 7.Görges R, Maniecki M, Jentzen W, et al. Development and clinical impact of thyroglobulin antibodies in patients with differentiated thyroid carcinoma during the first 3 years after thyroidectomy. Eur J Endocrinol. 2005;153:49–55. doi: 10.1530/eje.1.01940. [DOI] [PubMed] [Google Scholar]
- 8.Spencer CA, Takeuchi M, Kazarosyan M, et al. Serum thyroglobulin autoantibodies: prevalence, influence on serum thyroglobulin measurement, and prognostic significance in patients with differentiated thyroid carcinoma. J Clin Endocrinol Metab. 1998;83:1121–1127. doi: 10.1210/jcem.83.4.4683. [DOI] [PubMed] [Google Scholar]
- 9.Tozzoli R, Bizzaro N, Tonutti E, et al. Italian Society of Laboratory Medicine (SIMeL) Study Group on the Laboratory Diagnosis of Autoimmune Diseases. Immunoassay of anti-thyroid autoantibodies: high analytical variability in second generation methods. Clin Chem Lab Med. 2002;40:568–573. doi: 10.1515/CCLM.2002.098. [DOI] [PubMed] [Google Scholar]
- 10.Hollowell JG, Staehling NW, Flanders WD, et al. Serum TSH, T(4), and thyroid antibodies in the United States population (1988–1994): National Health and Nutrition Examination Survey (NHANES III) J Clin Endocrinol Metab. 2002;87:489–499. doi: 10.1210/jcem.87.2.8182. [DOI] [PubMed] [Google Scholar]
- 11.Latrofa F, Ricci D, Montanelli L, et al. Lymphocytic thyroiditis on histology correlates with serum thyroglobulin autoantibodies in patients with papillary thyroid carcinoma: impact on detection of serum thyroglobulin. J Clin Endocrinol Metab. 2012;97:2380–2387. doi: 10.1210/jc.2011-2812. [DOI] [PubMed] [Google Scholar]
- 12.Spencer C, Fatemi S. Thyroglobulin antibody (TgAb) methods—Strengths, pitfalls and clinical utility for monitoring TgAb-positive patients with differentiated thyroid cancer. Best Pract Res Clin Endocrinol Metab. 2013;27:701–712. doi: 10.1016/j.beem.2013.07.003. [DOI] [PubMed] [Google Scholar]
- 13.Chiovato L, Latrofa F, Braverman LE, et al. Disappearance of humoral thyroid autoimmunity after complete removal of thyroid antigens. Ann Intern Med. 2003;139:346–351. doi: 10.7326/0003-4819-139-5_Part_1-200309020-00010. [DOI] [PubMed] [Google Scholar]
- 14.Verburg FA, Luster M, Cupini C, et al. Implications of thyroglobulin antibody positivity in patients with differentiated thyroid cancer: a clinical position statement. Thyroid. 2013;23:1211–1225. doi: 10.1089/thy.2012.0606. [DOI] [PubMed] [Google Scholar]
- 15.Bohuslavizki KH, vom Baur E E, Weger B, et al. Evaluation of chemiluminescence immunoassays for detecting thyroglobulin (Tg) and thyroid peroxidase (TPO) autoantibodies using the IMMULITE 2000 system. Clin Lab. 2000;46:23–31. [PubMed] [Google Scholar]
- 16.Gilmour J, Brownlee Y, Foster P, et al. The quantitative measurement of autoantibodies to thyroglobulin and thyroid peroxidase by automated microparticle based immunoassays in Hashimoto’s disease, Graves’ disease and a follow-up study on postpartum thyroid disease. Clin Lab. 2000;46:57–61. [PubMed] [Google Scholar]
- 17.Sapin R, d’Herbomez M, Gasser F, et al. Increased sensitivity of a new assay for anti-thyroglobulin antibody detection in patients with autoimmune thyroid disease. Clin Biochem. 2003;36:611–616. doi: 10.1016/S0009-9120(03)00114-0. [DOI] [PubMed] [Google Scholar]
- 18.La’ulu SL, SLev PR, Roberts WL. Performance characteristics of 5 automated thyroglobulin and thyroperoxidase autoantibody assays. Clin Chim Acta. 2007;376:88–95. doi: 10.1016/j.cca.2006.07.018. [DOI] [PubMed] [Google Scholar]
- 19.Krahn J, Dembinski T. Thyroglobulin and anti-thyroglobulin assays in thyroid cancer monitoring. Clin Biochem. 2009;42:416–419. doi: 10.1016/j.clinbiochem.2008.12.017. [DOI] [PubMed] [Google Scholar]
- 20.Taylor KP, Parkington D, Bradbury S, et al. Concordance between thyroglobulin antibody assays. Ann Clin Biochem. 2011;48:367–369. doi: 10.1258/acb.2011.010248. [DOI] [PubMed] [Google Scholar]
- 21.Pickett AJ, Jones M, Evans C. Causes of discordance between thyroglobulin antibody assays. Ann Clin Biochem. 2012;49:463–467. doi: 10.1258/acb.2012.012008. [DOI] [PubMed] [Google Scholar]
- 22.Lindberg B, Svensson J, Ericsson UB, et al. Comparison of some different methods for analysis of thyroid autoantibodies: importance of thyroglobulin autoantibodies. Thyroid. 2001;11:265–269. doi: 10.1089/105072501750159697. [DOI] [PubMed] [Google Scholar]
- 23.Schulz R, Bethäuser H, Stempka L, et al. Evidence for immunological differences between circulating and thyroid-derived thyroglobulin in men. Eur J Clin Invest. 1989;19:459–463. doi: 10.1111/j.1365-2362.1989.tb00259.x. [DOI] [PubMed] [Google Scholar]
- 24.CLSI (2008) Defining, establishing and verifying reference intervals in the clinical laboratory; Approved guideline, 3rd edn, vol 28, No. 30, CLSI document, Wayne
- 25.Baloch Z, Carayon P, Comte-Devolx B, et al. Laboratory medicine practice guidelines. Laboratory support for the diagnosis and monitoring of thyroid disease. Thyroid. 2003;13:3–126. doi: 10.1089/105072503321086962. [DOI] [PubMed] [Google Scholar]
- 26.NCCLS (1999) Evaluation of precision performance of clinical chemistry devices; approved guideline, vol 19, No. 2. NCCLS document EP5-A, Wayne
- 27.NCCLS (2004) Evaluation of Precision Performance of Quantitative measurement methods; Approved Guideline, 2nd edn, vol 24, No. 25. NCCLS document EP5-A2, Wayne
- 28.NCCLS (2004) Protocols for determination of limits of detection and limits of quantitation: approved guideline, 2nd edn, vol 24, No. 34. NCCLS publication EP 17-A, NCCLS, Wayne
- 29.Spencer CA, Bergoglio LM, Kazarosyan M, et al. Clinical impact of thyroglobulin (Tg) and Tg autoantibody method differences on the management of patients with differentiated thyroid carcinomas. J Clin Endocrinol Metab. 2005;90:5566–5575. doi: 10.1210/jc.2005-0671. [DOI] [PubMed] [Google Scholar]
- 30.Tozzoli R, D’Aurizio F, Ferrari A, et al. The upper reference limit for thyroid peroxidase autoantibodies is method-dependent: a collaborative study with biomedical industries. Clin Chim Acta. 2016;452:61–65. doi: 10.1016/j.cca.2015.10.025. [DOI] [PubMed] [Google Scholar]
- 31.D’Aurizio F, Metus P, Polizzi Anselmo A, et al. Establishment of the upper reference limit for thyroid peroxidase autoantibodies according to the guidelines proposed by the National Academy of Clinical Biochemistry: comparison of five different automated methods. Autoimmun Highlights. 2015;6:31–37. doi: 10.1007/s13317-015-0070-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Tozzoli R, Giavarina D, Villalta D, et al. Definition of reference limits for autoantibodies to thyroid peroxidase and thyroglobulin in a large population of outpatients using an indirect method based on current data. Arch Pathol Lab Med. 2008;132:1924–1928. doi: 10.5858/132.12.1924. [DOI] [PubMed] [Google Scholar]
- 33.Jensen EA, Petersen PH, Blaabjerg O, et al. Establishment of reference distributions and decision values for thyroid antibodies against thyroid peroxidase (TPOAb), thyroglobulin (TgAb) and the thyrotropin receptor (TRAb) Clin Chem Lab Med. 2006;44:991–998. doi: 10.1515/CCLM.2006.166. [DOI] [PubMed] [Google Scholar]
- 34.Kaloumenoum I, Mastorakos G, Alevizaki M, et al. Thyroid autoimmunity in schoolchildren in an area with long-standing iodine sufficiency: correlation with gender, pubertal stage and maternal thyroid autoimmunità. Thyroid. 2008;18:747–754. doi: 10.1089/thy.2007.0370. [DOI] [PubMed] [Google Scholar]
- 35.Taubner K, Schubert G, Pulzer F, et al. Serum concentrations of anti-thyroid peroxidase and anti-thyroglobulin antibodies in children and adolescents without apparent thyroid disorders. Clin Biochem. 2014;47:3–7. doi: 10.1016/j.clinbiochem.2013.09.017. [DOI] [PubMed] [Google Scholar]
- 36.Feldt-Rasmussen U, Hoier-Madsen M, Bech K, et al. Anti-thyroid peroxidase antibodies in thyroid disorders and non-thyroid autoimmune diseases. Autoimmunity. 1991;9:245–254. doi: 10.3109/08916939109007650. [DOI] [PubMed] [Google Scholar]
- 37.Mariotti S, Chiovato L, Franceschi C, Pinchera A. Thyroid autoimmunity and aging. Exp Gerontol. 1998;33:535–541. doi: 10.1016/S0531-5565(98)00030-8. [DOI] [PubMed] [Google Scholar]
- 38.Wood WG, Hanke R. Comparability of method results and performance in a national external quality assessment scheme between 1993 and 2003 using thyroid associated antibodies as examples. Clin Lab. 2004;50:209–221. [PubMed] [Google Scholar]
- 39.D’Aurizio F, Tozzoli R, Villalta D, et al. Immunoassay of thyroid peroxidase autoantibodies: diagnostic performance in automated third generation methods. Clin Chem Lab Med. 2015;53:415–421. doi: 10.1515/cclm-2014-0545. [DOI] [PubMed] [Google Scholar]