Abstract
Activating mutations in codon 12 and codon 13 of the KRAS (Kirsten rat sarcoma viral oncogene homolog) gene are implicated in the development of several human cancer types and influence their clinical evaluation, treatment and prognosis. Numerous different methods for KRAS genotyping are currently available displaying a wide range of sensitivities, time to answer and requirements for laboratory equipment and user skills. Here we present SensiScreen® KRAS exon 2 simplex and multiplex CE IVD assays, that use a novel real-time PCR-based method for KRAS mutation detection based on PentaBase’s proprietary DNA analogue technology and designed to work on standard real-time PCR instruments. By means of the included BaseBlocker™ technology, we show that SensiScreen® specifically amplifies the mutated alleles of interest with no or highly subdued amplification of the wild type allele. Furthermore, serial dilutions of mutant DNA in a wild type background demonstrate that all SensiScreen® assays display a limit of detection that falls within the range of 0.25–1%. Finally, in three different colorectal cancer patient populations, SensiScreen® assays confirmed the KRAS genotype previously determined by commonly used methods for KRAS mutation testing, and notably, in two of the populations, SensiScreen® identified additional mutant positive cases not detected by common methods.
Introduction
KRAS belongs to the RAS family of small GTPases involved in the coupling of signal transduction from surface receptors to many different targets that regulate diverse biological responses including cell growth, proliferation, differentiation and survival [1,2]. In cancer, the KRAS gene is typically altered by activating point mutations occurring in hot-spots, mainly in codons 12 and 13 of exon 2 (overall accounting for more than 80% of KRAS mutations) and, less frequently, in codons 59, 61, 117 and 146 [3–7]. The KRAS gene is mutated in several cancers, including adenocarcinomas occurring in the pancreas, lung, ovary and thyroid. However, it is in colorectal cancer (CRC) that KRAS mutations have recently acquired most clinical significance [4,8,9]. CRC is the third most diagnosed cancer in both men and women and it is one of the leading causes of cancer related mortality, accounting for more than 600.000 deaths worldwide annually [10–12]. Standard care for patients with metastatic CRC (mCRC) is surgery combined with adjuvant chemotherapy or radiotherapy. However, more recently, targeted therapies have been included in the treatment of selected patients with mCRC. In particular, several studies have clearly and extensively demonstrated that monoclonal antibodies (e.g. cetuximab and panitumumab) targeting the Epidermal Growth Factor Receptor (EGFR) may be effective in the treatment of mCRC patients with tumors showing a KRAS exon 2 wild type status [8,9,13]. In fact, activating KRAS mutations (occurring in about 30–40% of CRC) are associated with lack of response to EGFR (epidermal growth factor receptor) inhibitors and, at least for panitumumab, have been associated with a detrimental effect with respect to other chemotherapeutic regimens [14]. Based on this knowledge, the recommendations issued by the US Food and Drug Administration (FDA) and the European Medicines Agency (EMA) now mandate mutational testing of KRAS in mCRC patient tumors before the administration of EGFR inhibitors. Of note, especially in the case of rectal cancer, the tumor material available to laboratories for testing may be limited to only small biopsies, or tissues from partially treated patients containing only few cancer cells, thus making KRAS mutational assessment even more challenging. Therefore, sensitive mutation testing techniques have become important tools when selecting patients who will benefit from targeted anti-EGFR agents.
The response rates of CRC patients with apparently wild type KRAS tumors treated with anti-EGFR therapies range from only 40% to 60% [15]. The high numbers of non-responders could, at least in part, be due to intratumoral heterogeneity of KRAS mutations that are not included in the tumor areas analysed [16]. However, it may also reflect the fact that there are currently no specific requirements regarding the limit of detection (LOD) for the methods used for KRAS mutation testing, in spite of recent evidence indicating that mCRC patient tumors with as few as 1% KRAS mutated cancer cells fail to respond to anti-EGFR therapies [17]. Thus, direct sequencing (DS) still remains one of the most widely used methodologies for KRAS testing, in spite of its low LOD (10–50%) [18,19]. This is the case, even though several more sensitive commercial methods for KRAS mutation testing are now available. These include next-generation sequencing (NGS) methods with LODs of 1–10%, depending on platform and coverage [18,20–22], mass spectrometry-based methods with LODs down to 5% [23], and real-time PCR-based methods such as cobas® KRAS mutation test (Roche, LOD of 2.4–12.1%) [24], therascreen® KRAS RGQ PCR kit (Qiagen, LOD of 0.5–1%) [20], and Idylla™ KRAS Mutation Test (Biocartis, LOD ≥5–10%) [25]. Digital droplet PCR (ddPCR) is the latest addition to the technologies available for KRAS mutation testing recently demonstrating a LOD as low as 0.0005%, although a large amount of DNA was required to achieve these levels [26]. Thus, the most common methods used for KRAS mutation testing have very different LODs with possible implications for cancer treatment in tested patients. Furthermore, these methods are characterized by marked variations in handling times, time to answer, as well as differences in the skills and expertise required of the personnel performing the analyses, the latter being especially relevant with regard to NGS and ddPCR methods.
In this study, we report the development of new real-time PCR-based assays for KRAS mutation detection, called SensiScreen® KRAS EXON 2 simplex and multiplex, CE IVD. SensiScreen® assays are based on PentaBase’s proprietary DNA analogue technology [27,28], and are designed to work on standard real-time PCR instruments. SensiScreen® was developed with the aim of combining high sensitivity with ease-of-use, short handling time, and short time to answer. Furthermore, the SensiScreen® multiplex version allows for fast screening of the 9 most common KRAS mutations (7 single- and 2 double-mutations) with minimal use of input DNA. We compared, SensiScreen® assays in different cohorts of CRC patients with DS, mutant-enriched PCR (ME-PCR), therascreen® and cobas® genotyping methods. In addition to confirming all mutations found by the alternative methods, SensiScreen® assays identified additional KRAS exon 2 mutated cases in two out of three patient populations studied.
Materials and methods
Construction of model templates
For the development of SensiScreen®, seven plasmids comprising sequences with 7 different KRAS exon 2 mutations (G12A: c.35G>C, G12D: c.35G>A, G12R: c.34G>C, G12C: c.34G>T, G12S: c.34G>A, G12V: c.35G>T and G13D: c.38G>A) were constructed. A 221 bp fragment of the KRAS gene surrounding the codon 12/13 hotspot was cloned from a wild type (WT) genomic DNA template using the primers Cloning forward and Cloning reverse [29] (S1 Table). The fragment was inserted into the pCR4TOPO-vector according to the manufacturer’s instructions (Invitrogen, Carlsbad, CA, USA). This WT plasmid was used as template in a series of site-directed mutagenesis reactions using designed mutagenesis primers (PentaBase ApS, Odense, Denmark) and the QuikChange Lightning Site-Directed Mutagenesis Kit (Agilent Technologies, Santa Clara, CA, USA) for introduction of the 7 most common KRAS exon 2 mutations. The resulting plasmids were validated by sequencing using the vector specific primers M13rev and M13fwd. The plasmids were purified using GenElute Plasmid Miniprep Kit (Sigma-Aldrich, St. Louis, MO, USA) and linearized with the restriction enzyme NotI (New England Biolabs, Ipswich, MA, USA). The quality of the linearized plasmids was evaluated by agarose gel electrophoresis and quantitated using a Nanodrop (Thermo Scientific Waltham, MA, USA) spectrophotometer.
Cell lines
Genomic DNA from six tumor cell lines, SW480, SW1116, MICOL29, T84, CAL-62 and A549, containing 6 different KRAS mutations were used for sensitivity studies (S2 Table). The cell lines were subcultured in appropriate media according to the manufacturer’s instructions and genomic DNA was isolated using the QIAmp Mini kit (Qiagen, Chatsworth, CA, USA). SW480 (catalogue number: CCL-228™), SW1116 (catalogue number: CCL-233™), T84 (catalogue number: CCL-248™) and A549 (catalogue number: CCL-185™) were obtained from ATCC®; CAL-62 (catalogue number: ACC-448™) was obtained from DSMZ® and MICOL29 was kindly provided by Fondazione IRCCS Istituto Nazionale dei Tumori, Milan, Italy.
SensiScreen® assay development
SensiScreen® real-time PCR assays were developed to specifically amplify few copies of mutated DNA with a single-nucleotide polymorphism in a large WT background. In order to achieve this, we took advantage of PentaBase’s intercalating nucleic acid modifications called pentabases™ that allow for the creation of oligonucleotides that, among other features, show increased affinity and specificity towards their target sequence [27,28]. Thus, SensiScreen® include primers, hydrolysis probes and blocking oligonucleotides containing pentabases™, which are designated SuPrimers™, HydrolEasy™ probes and BaseBlockers™, respectively. Different oligo designs containing different combinations of pentabases™ were tested and evaluated for: specificity of BaseBlockers™, signal-to-noise ratio of HydrolEasy™ probes, and PCR efficiency of SuPrimers™. Furthermore, to obtain the highest possible specificity towards the selected mutations, we evaluated different mutation-specific primer designs all with the particular sequence variation in the 3’ end of the primer, and with either no additional mismatch or one of the three possible nucleotide variations in position -3 or -4 relative to the 3’-end. All SensiScreen® assays (7 simplex and 2 multiplex) also include an internal control assay targeting the CYP17A1 gene that does not interfere with amplification by the primary assays (data not shown).
Sensitivity studies
In order to evaluate the LOD of SensiScreen®, we performed serial dilutions of mutated DNA from cell lines and plasmid DNA (pCR4TOPO-G12C), respectively, in a background of WT DNA (Promega, Madison,WI, USA) totalling 50 ng genomic input DNA. Eight different concentrations of mutated DNA (10%, 5%, 1%, 0.5%, 0.1%, 0.05%, 0.01% and 0%) were tested by real-time PCR, using the SensiScreen® protocol described below.
Patient samples
Inclusion criteria for the study were a diagnosis of CRC following surgical resection and the availability of at least 3 different tissue blocks containing cancer cells for each patient. The patients had undergone routine surgery and clinico-pathological assessment of their tumors as part of their standard clinical care. The tumor materials used in the study were surplus to requirements for routine testing. All materials were anonymized. This study was approved by the Institutional Ethical Committee of the Institute of Pathology of Locarno (Switzerland), by the Central Denmark Region Committee on Health Research Ethics, and by the Danish Data Protection Agency. All procedures were performed in accordance with the ethical standards of the Helsinki Declaration.
Tissue analyses
Formalin-fixed paraffin-embedded (FFPE) tumor blocks were analysed for quality and tumor content. A single representative tumor block from each patient, containing at least 70% neoplastic cells (if necessary after macrodissection to enrich tumor cell content), was selected. Genomic DNA was extracted from six 7 μm-thick serial sections of each FFPE block using the QIAmp Mini kit (cohort 1) and QIAsymphony (cohort 2+3) (Qiagen) according to the manufacturer’s instructions.
Cohort 1: FFPE samples from 100 patients with histologically confirmed CRC collected from 1996 to 2009 were retrospectively analysed for KRAS exon 2 mutations by DS, ME-PCR and SensiScreen® simplex and multiplex. All tumors were colorectal adenocarcinomas, diagnosed at the Institute of Pathology in Locarno, Switzerland (S3 Table).
Cohort 2: FFPE samples from 79 patients with histologically confirmed CRC were retrospectively analysed for KRAS exon 2 mutations by DS, therascreen® (Qiagen) and SensiScreen® simplex. All tumors were colorectal adenocarcinomas, diagnosed at Aarhus University Hospital, Denmark (S4 Table).
Cohort 3: FPPE samples from 283 patients with histologically confirmed CRC were retrospectively analysed for KRAS exon 2 mutations by cobas® (Roche, Basel, Switzerland) and SensiScreen® multiplex. Cases only found to be mutated with SensiScreen® Multiplex were in addition analysed with SensiScreen® simplex. All tumors were colorectal adenocarcinomas, diagnosed at Aarhus University Hospital, Denmark (S5 Table).
Mutational analysis by direct sequencing
The mutational analyses of KRAS exon 2 by DS were performed at the Institute of Pathology in Locarno (Switzerland) using the primers listed in S1 Table. Samples were subjected to automated sequencing on an ABI PRISM 3130 Genetic Analyzer (Applied Biosystems, Foster City, CA, USA), and evaluated with Sequencing Navigator Software (Applied Biosystems). All mutated cases were confirmed twice with independent PCR reactions.
Mutational analysis by ME-PCR
Analyses of KRAS exon 2 mutations by ME-PCR were performed at the Institute of Pathology in Locarno (Switzerland) essentially as described previously [30,31]. ME-PCR combines PCR amplification with restriction enzyme digestion to enrich for mutant alleles, while WT alleles are eliminated by digestion [32–34]. ME-PCR products were subsequently subjected to automated sequencing on an ABI PRISM 3130 Genetic Analyzer. All mutated cases were confirmed twice with independent PCR reactions.
Mutational analysis by therascreen®
The analysis of KRAS mutations in exon 2, codon 12/13 and 61 using the therascreen® KRAS test (Qiagen) was performed at the Department of Pathology, Aarhus (Denmark). DNA was extracted from FFPE tissue using the QiaSymphony (Qiagen) according to the manufacturer’s instructions. The DNA concentration was measured (Implen, Munich, Germany) and the DNA was diluted to a concentration of 5 ng/μl. Five μl were added to each reaction tube. The analysis was subsequently run on a Rotor-Gene Q (Qiagen).
Mutational analyses by cobas® KRAS mutation test
A Cobas 480Z (Roche) was used for the testing of KRAS mutations in exon 2 codons 12/13 and 61. The analysis was performed at the Department of Pathology, Aarhus. Extraction and measurement of the DNA concentration was performed as described above (therascreen®). The DNA was diluted to 5 ng/μl. Twenty-five μl were added to each reaction tube.
Mutational analysis by SensiScreen®
SensiScreen® development and clinical validation was performed in 25 μl reactions in KAPA Probe Fast qPCR 2x Master Mix (KAPA Biosystems, Wilmington, MA, USA) on a Rotor-Gene 6000 (Corbett Research, Mortlake, NSW, Australia), a CFX96 (Bio-rad Laboratories, Hercules, CA, USA) and a Mx3005P qPCR system (Stratagene, CA, USA). The reaction mixture contained 300–900 nM of each primer, 200 nM of each probe, and 1000–5000 nM WT BaseBlocker (WTB). qPCR (quantitative PCR) was performed using either 50 ng of cell line DNA or 50 ng WT DNA (Promega G304A) in the absence or presence of approximately 1000 copies of plasmid DNA containing the indicated mutations. The thermocycling conditions used were: 2 min of initial activation of the hotstart taq-polymerase at 95°C, followed by 45 cycles of a 2-step PCR with a 15 sec denaturation step at 94°C and 60 sec extension step at 60°C. Fluorescence was measured at the end of each extension step.
The qPCR threshold cycle (Ct) of normalized fluorescence was used for the evaluation of the data. Ct is defined as the number of cycles where a fluorescence signal crosses the threshold. In order to make data analysis independent of the type of instrument used, the threshold was defined as 10% of the signal strength of the reference assay at cycle 45 (S1 Fig). For all valid samples (23 < Ctref < 36), a ΔCt value was calculated by taking the Ct value of the mutation-specific assay and subtracting the Ct value of the reference assay:
Patient samples analysed by SensiScreen® were regarded as positive for a given mutation if the Ctmutation was ≤ 38 and ΔCt was ≤9.
Results
SensiScreen® assay design and specificity
SensiScreen® KRAS EXON2 simplex and multiplex CE IVD (hereinafter referred to as SensiScreen® simplex and multiplex) are real-time PCR-based assays developed by PentaBase (Denmark) for the detection of KRAS exon 2 mutations. SensiScreen® detects nine different mutations (seven single and two double nucleotide polymorphisms) in exon 2 codons 12 and 13 of the KRAS gene (Table 1) and is certified in accordance to the EU guidelines 98/79/EC Medical equipment for in vitro diagnostics. The simplex version includes a reference assay and 7 different mutation-specific assays, while the multiplex version includes a reference assay and 2 different mutation-specific assays (Table 1). The reference assay comprises a set of allele-independent primers and a green fluorescent HydrolEasy™ probe, while the mutation-specific assays include an allele-independent forward primer, allele-specific reverse primers, a green fluorescent HydrolEasy™ probe and a BaseBlocker™ (Fig 1 and S1 Table). In addition, all assays include an internal control assay comprising a set of allele-independent primers and a yellow fluorescent HydrolEasy™ probe targeting the CYP17A1 gene. SensiScreen® includes standard DNA oligonucleotides as well as DNA oligonucleotides modified with pentabases™ which are flat hetero aromatic molecules inserted into the backbone of the oligonucleotide via a linker [27,28]. The presence of pentabases™ increases both the affinity, sensitivity and specificity of the oligonucleotide towards the target sequence [27,28]. A central feature of SensiScreen® is the inclusion of a BaseBlocker™ that comprises several pentabases™ and is designed to bind strongly and specifically to WT DNA with little or no affinity towards mutated DNA (Fig 1). Thus, by combining the use of a BaseBlocker™ with allele-specific primers, SensiScreen® qPCR leads to highly specific amplification of mutated DNA whereas amplification of WT DNA is strongly impaired (Fig 2).
Table 1. KRAS mutations detected by SensiScreen® KRAS exon 2 simplex and multiplex.
KRAS CDS mutation | KRAS AA substitution | COSMIC ID | SensiScreen multiplex assay |
---|---|---|---|
c.35G>C | Gly12Ala (G12A) | COSM522 | Multiplex 2 |
c.35G>A | Gly12Asp (G12D) | COSM521 | Multiplex 2 |
c.34G>C | Gly12Arg (G12R) | COSM518 | Multiplex 1 |
c.34G>T | Gly12Cys (G12C) | COSM516 | Multiplex 1 |
c.34G>A | Gly12Ser (G12S) | COSM517 | Multiplex 1 |
c.35G>T | Gly12Val (G12V) | COSM520 | Multiplex 1 |
c.38G>A | Gly13Asp (G13D) | COSM532 | Multiplex 2 |
c.34_35GG>TT | Gly12Phe (G12F) | COSM512 | Multiplex 1 |
c.34_35GG>AT | Gly12Ile (G12I) | COSM34144 | Multiplex 1 |
KRAS, Kirsten rat sarcoma viral oncogene homolog; CDS, COSMIC database sequence; AA, amino acids; COSMIC, catalogue of somatic mutations in cancer; COSMIC ID, COSMIC identification number.
SensiScreen® assay sensitivity
We sought to develop an assay that was not only highly specific for amplification of mutant DNA but also sensitive enough to be used on samples that contain very low amounts of mutated DNA. Thus, to evaluate the sensitivity of SensiScreen® simplex and multiplex assays, we performed serial dilutions of mutant cell line or plasmid DNA in a WT background. To evaluate the robustness of SensiScreen®, the sensitivity studies were performed on two different real-time PCR systems, the Rotor-Gene 6000 (Corbett Research) (Figs 3 and 4, Table 2) and the MyGo Pro (IT-IS Life Science Ltd., Mahon, Ireland) (S2 Fig; S3 Fig and Table 2).
Table 2. Sensitivity and PCR efficiency of SensiScreen® simplex and multiplex assays determined by serial dilutions of mutated DNA in a wild type background.
Rotor-Gene | MyGo Pro | |||||
---|---|---|---|---|---|---|
Assay | LOD* | ΔCt WT^ | Efficiency (%)¤ | LOD* | ΔCt WT^ | Efficiency (%)¤ |
G12A Spx | 0.5% | No Signal | 88 | 0.5% | No Signal | 96 |
G12D Spx | 0.5% | 12.4 | 94 | 0.5% | 16.6 | 99 |
G12R Spx | 0.5% | No Signal | 97 | 0.5% | No Signal | 97 |
G12C Spx | 0.5% | 13.6 | 86 | 0.5% | 13.8 | 98 |
G12S Spx | 0.5% | 14.5 | 89 | 0.5% | No Signal | 96 |
G12V Spx | 0.5% | 19.3 | 101 | 0.5% | No Signal | 95 |
G13D Spx | 0.5% | 14.2 | 85 | 0.25% | 14.7 | 106 |
G12A Mpx 2 | 0.5% | No Signal | 86 | 0.5% | 12.6 | 102 |
G12D Mpx 2 | 1.0% | 13.85 | 74 | 1.0% | No Signal | 96 |
G12R Mpx 1 | 0.5% | No Signal | 97 | 1.0% | No Signal | 80 |
G12C Mpx 1 | 0.5% | No Signal | 92 | 1.0% | No Signal | 91 |
G12S Mpx 1 | 1.0% | 15.6 | 87 | 1.0% | 17.4 | 96 |
G12V Mpx 1 | 0.5% | 16 | 92 | 0.5% | No Signal | 93 |
G13D Mpx 2 | 0.25% | 12.4 | 100 | 0.25% | No Signal | 109 |
LOD, limit of detection; WT, wild type; SPX, simplex assay; MPX, multiplex assay.
* Limit of detection (LOD) determined using a ΔCt between reference and assay of 9.
^ ΔCt calculated as the lowest possible difference in Ct between reference and assay of the duplicate PCR reactions performed using the 0% dilution point.
¤ PCR efficiency calculated using the 10%, 5%, 1% and 0.5% dilution points.
SensiScreen® determines if a sample is positive or negative for the specific mutation(s) by measuring the difference in cycle threshold (ΔCt) of the reference assay (Ctreference) and the cycle threshold of the mutation-specific assay (Ctmutation). The ΔCt cut-off should be set at a value that will avoid false positive results from amplification of WT samples and at the same time identify as many true positive samples as possible. Since a ΔCt of 12.4 was the lowest possible value identified when using WT DNA as template, we considered a ΔCt of 9 as the optimal value for subsequent analysis of assay sensitivities.
Interestingly, using a ΔCt of 9, the LODs of the different SensiScreen® assays were very similar on the two real-time PCR instruments. Thus, all SensiScreen® simplex assays except G12R Simplex (LOD less than 1.0% on the MyGo Pro instrument) were found to display a LOD of less than 0.5%. Similarly, using the same ΔCt of 9, SensiScreen® multiplex assays showed LODs from 0.25% (G13D) to 1% on both instruments.
SensiScreen® assays identify more patients with KRAS mutations than commonly used methods
To validate SensiScreen® assays in the clinical setting, we used SensiScreen® to retrospectively analyse FFPE DNA from three different patient populations with histologically confirmed CRC that were previously analysed by common KRAS mutation tests. In general, when including both Ct and Delta Ct values in the analysis of the patient samples, there was a good separation of the wild type and mutant groups in all three cohorts (Fig 5). Using the analysis settings described in Table 3, in all three populations, SensiScreen® confirmed the same mutations already identified by the common method(s) used in each population (Table 3; S6 Table, S7 Table and S8 Table). Thus, in cohort 1, both SensiScreen® KRAS EXON 2 simplex and multiplex assays confirmed the 28 and 43 mutational cases identified by DS and ME-PCR, respectively (Fig 5, Table 3 and S6 Table). In cohort 2, SensiScreen® KRAS EXON 2 simplex confirmed the 17 and 19 mutational cases identified by DS and therascreen®, respectively (Table 3 and S7 Table). Finally, in cohort 3, SensiScreen® KRAS EXON 2 multiplex confirmed the 87 mutational cases identified by cobas® (Table 3 and S8 Table). Interestingly, however, SensiScreen® assays in addition identified one new mutational case in cohort 1, and 6 new mutational cases in cohort 3. Additional mutational cases identified by SensiScreen® were confirmed SensiScreen® simplex or multiplex when possible.
Table 3. Summary of cohort data.
Comparison of SensiScreen® to common methods for KRAS exon 2 mutation testing | |||||||
---|---|---|---|---|---|---|---|
Common method | Cohort no | Total patient samples | Mutated cases identified by | Additional mutated cases identified by SensiScreen® | |||
Common method | SensiScreen® | no | % | ||||
KRAS exon 2 simplex | Direct sequencing | 1 | 100 | 28 | 44 | 16 |
57% |
ME-PCR | 1 | 100 | 43 | 44 | 1 |
2% | |
Direct sequencing | 2 | 79 | 17 | 19 | 2 | 12% | |
therascreen® | 2 | 79 | 19 | 19 | 0 |
0% | |
KRAS exon 2 multiplex | cobas® | 3 | 283 | 87 | 93 | 6 | 7% |
KRAS, Kirsten rat sarcoma viral oncogene homolog; no: number; ME-PCR: mutant-enriched PCR.
Discussion
The assessment of KRAS mutational status has acquired high relevance in recent years and, at the same time, has become a challenge. Although originally considered as an alteration occurring in early phases of developing CRC, KRAS mutations may also appear first in the metastatic lesion, probably due to the selection of a clone already present in the primary tumor. The fraction of tumoral DNA in the sample is highly dependent on which tissue block the pathologist selects for further analysis [35], and tumor heterogeneity may significantly complicate this pathological evaluation. Furthermore, tumoral cells can be polysomic, and the allele with the KRAS WT sequence can be overrepresented with respect to the allele carrying the KRAS mutation. Finally, the availability of either only small biopsy samples with few cancer cells or tumors already treated prior to surgical resection (as is often the case with rectal carcinomas), presents an additional challenge for diagnostic laboratories. It is clear that the availability of methodologies that are both highly sensitive and specific may make a significant contribution to the successful evaluation of the KRAS mutational status in the routine clinical setting. An incorrect KRAS mutational analysis may lead either to the use of inappropriate targeted therapy or to a delay in the most suitable treatment, both of which may have serious consequences for the patients involved. Furthermore, it has been proposed that patients with tumors characterized by a low level of KRAS mutated DNA may also be resistant to the administration of EGFR-targeted therapies, although the precise threshold of mutant allele required for EGFR-targeted therapy resistance has not yet been established [17,36]. Fortunately, recently developed methods for KRAS mutation testing, including SensiScreen®, can detect highly diluted mutant DNA. Thus, we show here, that SensiScreen®, a novel real-time PCR based assay for detection of KRAS exon 2 mutations, is extremely sensitive and is able to find more mutation positive patients than the commonly used methods we compared it with.
One of the enhanced qualities of SensiScreen® is its specificity. This is mainly related to the inclusion of a blocking oligonucleotide (BaseBlocker™) that is designed to bind strongly and specifically to WT DNA in order to inhibit its amplification. Thus, compared with standard ARMS PCR, the BaseBlocker™ provides a large part of the allele specificity of the assay that is normally solely dependent on the mutation-specific primers. This not only leads to a better overall specificity of the SensiScreen® assays, but also allows for the use of primers where the efficiency is not compromised by primer specificity design strategies such as additional mismatches. In line with this, most SensiScreen® assays display a sensitivity that is close to the theoretical limit of a PCR with 100% efficiency and, importantly, is associated with no or very low amplification of the WT template. Thus, by performing serial dilutions of mutant DNA, we show that SensiScreen® simplex and multiplex assays can safely detect as little as 0.25–1% mutant DNA in a WT background. Finally, the fact that the assays perform very similar on both the air temperature-controlled slow ramping Rotor-Gene machine and the pure silver block fast ramping MyGo Pro machine underlines the robustness of the SensiScreen® assays.
The sensitivity studies furthermore illustrate the LOD potential of the SensiScreen® technology. Because of the subdued amplification of the WT template, at least for some of the SensiScreen® assays, the ΔCt could potentially be expanded to achieve a LOD down to 0.01% (G12V simplex). Since 50 ng of human genomic DNA corresponds to roughly 16,000 copies, the 0.01% dilution point in theory is equal to just 1.6 copies (i.e. 1–2 copies) of mutant DNA. While the clinical relevance of such a low amount of mutated DNA in primary tumor biopsies is unknown, this degree of sensitivity could be highly relevant when analysing liquid biopsies.
In line with the demonstrated high sensitivity of SensiScreen®, when comparing with common methodologies using three separate CRC cohorts, we found that SensiScreen® not only confirmed the mutated samples identified by the common methods, but also identified additional novel mutant cases. This was especially the case when comparing to DS even though the percentages of mutations detected by DS were in accordance with the literature [37,38]. However, also when comparing to the very sensitive ME-PCR method SensiScreen® was able to identify an additional mutant case. Finally, in the largest cohort SensiScreen® multiplex was able to identify additional mutant cases compared to cobas®. With regard to the cases classified as mutated only by SensiScreen®, since we could not confirm these mutations with other methods, both the simplex and the multiplex version were used to confirm the results when possible.
On a general note, we found in the clinical studies that the WT signals had lower Ct and ΔCt values and thus were closer to the cut off than what we found in the sensitivity studies using cell lines and plasmids. We believe that these putative WT signals are related by an unknown mechanism to the FFPE preparation. This assumption is supported by preliminary studies using SensiScreen® on non-FFPE liquid biopsies in which WT signals generally show higher ΔCt values (data not shown). More tests on liquid biopsies will be needed to confirm this observation, which will allow for the possibility of using a higher ΔCt cut-off of SensiScreen® assays and, thus will lead to a potentially even higher sensitivity for the assays, something that is of particular importance when analysing liquid biopsies. Another point to note is that in several cases, due to low concentration of DNA in the sample (reference >Ct 30), we found that the Ct of the mutation-specific assays and multiplexes was just above 38 while the ΔCt was below 9. Some of these cases could potentially be mutation positive if more material was used, but unfortunately, in most cases it was not possible to repeat the analyses with more DNA due to lack of material.
In addition to its high sensitivity, the advantages of SensiScreen® include a fast turn-around time (less than 2 hours, from PCR preparation to the final result), the same cut-off values valid for all the mutation types (rendering the assay highly user-friendly) and flexibility including the possibility to use whatever real-time PCR machine is available (the results of discrepant cases have been tested using different real-time PCR instruments). The multiplex version represents an innovative solution. Thus, it has as high a sensitivity as the simplex version, and since as little as 5 ng genomic DNA can be used as input, this enables the laboratories to perform a fast screening using only 15 ng of genomic DNA (3 tubes). This allows for a 2-step approach of initial screening with the multiplex assay, followed by confirmation of the result with the simplex assay together with the precise characterization of the mutation. This is a robust and rapid workflow enabling test laboratories to characterize KRAS exon 2 status in a few hours.
In conclusion, we have developed a new real-time based assay capable of identifying more patient tumors with mutations in KRAS exon 2 (codons 12 and 13) compared with methods commonly used in diagnostic laboratories for detection of mutations in this gene. SensiScreen® has a high specificity and sensitivity, is time saving, and can also be applied in other cancer types, particularly in regards to tumors in which KRAS mutational status analysis is more challenging, e.g. lung adenocarcinomas that are characterized by low numbers of cancer cells in the available cytological or histological samples.
Supporting information
Acknowledgments
The authors declare that there is none to list in the acknowledgments that contributed to the work but do not meet the authorship criteria for publication in PLOSONE.
Data Availability
All relevant data are within the paper and its Supporting Information files.
Funding Statement
Eurostars provided support in the form of salaries for authors MB, CA, UBC, MF, AR, MHK & JL, but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. PentaBase ApS provided salaries for UBC, CA and MB and took part in the designing of the SensiScreen products, but not the testing and analysis of patient material nor in the conclusions of the performance of SensiScreen against competing technologies. The specific roles of these authors are articulated in the ‘author contributions’ section. The funders had no additional role in study design, data collection and analysis, decision to publish, or preparation of the manuscript. Support for this study was also provided by Danish Agency for Science Technology and Innovation [E!7034 ONSET diagnostics] and Fondazione Ticinese per la Ricerca contro il Cancro. The Institute of Pathology, Locarno, Switzerland, and Aarhus University Hospital, Aarhus, Denmark performed some analyses using their own funding.
References
- 1.Bar-Sagi D and Hall A. Ras and Rho GTPases: a family reunion. Cell. 2000;103: 227–38. [DOI] [PubMed] [Google Scholar]
- 2.Malumbres M and Barbacid M. RAS oncogenes: the first 30 years. Nat Rev Cancer. 2003;3: 459–465 (2003). doi: 10.1038/nrc1097 [DOI] [PubMed] [Google Scholar]
- 3.Edkins S, O'Meara S, Parker A, Stevens C, Reis M, Jones S et al. Recurrent KRAS codon 146 mutations in human colorectal cancer. Cancer Biol Thr. 2009;5: 928–932. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Chang YY, Lin PC, Lin HH, Lin JK, Chen WS, Jiang JK et al. Mutation spectra of RAS gene amily in colorectal cancer. Am J Surg. 2016;212: 537–544.e3. doi: 10.1016/j.amjsurg.2016.02.013 [DOI] [PubMed] [Google Scholar]
- 5.Samowitz WS, Slattery ML, Sweeney C, Herrick J, Wolff RK, Albertsen H. APC mutations and other genetic and epigenetic changes in colon cancer. Mol Cancer Res. 2007;5(2): 165–170. doi: 10.1158/1541-7786.MCR-06-0398 [DOI] [PubMed] [Google Scholar]
- 6.Kudryavtseva AV, Lipatova AV, Zaretsky AR, Moskalev AA, Fedorova MS, Rasskazova AS et al. Important molecular genetic markers of colorectal cancer. Oncotarget. 2016;7: 53959 53983. doi: 10.18632/oncotarget.9796 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Graham DM, Coyle VM, Kennedy RD, Wilson RH. Molecular subtypes and personalized therapy in metastatic colorectal cancer. Curr Colorectal Cancer Rep. 2016;12: 141–150. doi: 10.1007/s11888-016-0312-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Kocarnik JM, Shiovitz S, Phipps AI. Molecular phenotypes of colorectal cancer and potential clinical applications. Gastroenterol Rep. 2015;1–8. doi: 10.1093/gastro/gov046 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Dienstmann R and Tabernero J. Spectrum of gene mutations in colorectal cancer. Cancer J. 2016;22: 149–155. doi: 10.1097/PPO.0000000000000191 [DOI] [PubMed] [Google Scholar]
- 10.Ferlay J, Parkin DM, Steliarova-Foucher E. Estimates of cancer incidence and mortality in Europe in 2008. Eur J Cancer. 2010;46: 765–781. doi: 10.1016/j.ejca.2009.12.014 [DOI] [PubMed] [Google Scholar]
- 11.Gulbake A, Jain A, Jain A, Jain A, Jain SK. Insight to drug delivery aspects for colorectal cancer. World J Gastroenterol. 2016;22: 582–599. doi: 10.3748/wjg.v22.i2.582 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Available from: http://www.europacolon.com/.
- 13.Xu W, Jing H, Zhang F. Epidermal growth factor receptor-targeted therapy in colorectal cancer. Front Biosci (Landmark Ed). 2016;21: 410–8. [DOI] [PubMed] [Google Scholar]
- 14.Loupakis F, Pollina L, Stasi I, Ruzzo A, Scartozzi M, Santini D et al. PTEN expression and KRAS mutations on primary tumors and metastases in the prediction of benefit from cetuximab plus irinotecan for patients with metastatic colorectal cancer. J Clin Oncol. 2009;27: 2622–2629. doi: 10.1200/JCO.2008.20.2796 [DOI] [PubMed] [Google Scholar]
- 15.Linardou H, Dahabreh IJ, Kanaloupiti D, Siannis F, Bafaloukos D, Kosmidis P et al. Assessment of somatic k-RAS mutations as a mechanism associated with resistance to EGFR targeted agents: a systematic review and meta-analysis of studies in advanced non-small-cell lung cancer and metastatic colorectal cancer. Lancet Oncol. 2008;9: 962–972. doi: 10.1016/S1470-2045(08)70206-7 [DOI] [PubMed] [Google Scholar]
- 16.Diaz LA Jr, Williams RT, Wu J, Kinde I, Hecht JR, Berlin J et al. The molecular evolution of acquired resistance to targeted EGFR blockade in colorectal cancers. Nature. 2012;486: 537–540. doi: 10.1038/nature11219 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Laurent-Puig P, Pekin D, Normand C, Kotsopoulos SK, Nizard P, Perez-Toralla K et al. Clinical relevance of KRAS-mutated subclones detected with picodroplet digital PCR in advanced colorectal cancer treated with anti-EGFR therapy. Clin Cancer Res. 2015;21: 1087–1097. doi: 10.1158/1078-0432.CCR-14-0983 [DOI] [PubMed] [Google Scholar]
- 18.Tsiatis AC, Norris-Kirby A, Rich RG, Hafez MJ, Gocke CD, Eshleman JR et al. Comparison of Sanger sequencing, pyrosequencing, and melting curve analysis for the detection of KRAS mutations: diagnostic and clinical implications. J Mol Diagn. 2010;12: 425–32. doi: 10.2353/jmoldx.2010.090188 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Davidson CJ, Zeringer E, Champion KJ, Gauthier MP, Wang F, Boonyaratanakornkit J et al. Improving the limit of detection for Sanger sequencing: a comparison of methodologies for KRAS variant detection. Biotechniques. 2012;53: 182–188. doi: 10.2144/000113913 [DOI] [PubMed] [Google Scholar]
- 20.Jancik S, Drabek J, Berkovcova J, Xu YZ, Stankova M, Klein J et al. A comparison of direct sequencing, pyrosequencing, high resolution melting analysis, theraScreen DxS, and the K-ras stripassay for detecting KRAS mutations in non small cell lung carcinomas. J Exp Clin Cancer Res. 2012;31: 79 doi: 10.1186/1756-9966-31-79 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Simen BB, Yin L, Goswami CP, Davis KO, Bajaj R, Gong JZ et al. Validation of a next generation-sequencing cancer panel for use in the clinical laboratory. Arch Pathol Lab Med. 2015;139: 508–517. doi: 10.5858/arpa.2013-0710-OA [DOI] [PubMed] [Google Scholar]
- 22.Gao J, Wu H, Wang L, Zhang H, Duan H, Lu J et al. Validation of targeted next-generation sequencing for RAS mutation detection in FFPE colorectal cancer tissues: comparison with Sanger sequencing and ARMS Scorpion real-time PCR. BMJ Open. 2016;6: e009532 doi: 10.1136/bmjopen-2015-009532 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.MassArray Dx colon panel product sheet available from: http://agenabio.com/wp content/uploads/2016/10/51-20001R2.0_Oncology-Solutions-Brochure-102016.pdf. [Google Scholar]
- 24.Lee S. Performance characteristics of a PCR assay for the detection of KRAS mutations in formalin-fixed paraffin-embedded tissue samples of non-small cell lung cancer. J Mol Biomark. 2015;6. [Google Scholar]
- 25.Idylla KRAS mutation test (CE IVD) brochure available from: https://media.biocartis.com/biocartis/documents/TECH_FICHE-KRAS-IVD-01122016.pdf
- 26.Pekin D, Skhiri Y, Baret JC, Le Corre D, Mazutis L, Salem CB et al. Quantitative and sensitive detection of rare mutations using droplet based microfluidics. Lab Chip. 2011;11: 2156–2166. doi: 10.1039/c1lc20128j [DOI] [PubMed] [Google Scholar]
- 27.Christensen UB and Pedersen EB. Intercalating nucleic acids containing insertions of 1-O (1-pyrenylmethyl)glycerol: stabilisation of dsDNA and discrimination of DNA over RNA. Nucleic Acids Res. 2002;30: 4918–4925. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Christensen UB, Wamberg M, El-Essawy FA, Ismail Ael-H, Nielsen CB, Filichev VV et al. Intercalating nucleic acids: the influence of linker length and intercalator type on their duplex stabilities. Nucleosides Nucleotides and Nucleic Acids. 2004;23: 207–225. [DOI] [PubMed] [Google Scholar]
- 29.Winder T, Mündlein A, Rhomberg S, Dirschmid K, Hartmann BL, Knauer M et al. Different types of K-Ras mutations are conversely associated with overall survival in patients with colorectal cancer. Oncol Rep. 2009;21: 223–230. [DOI] [PubMed] [Google Scholar]
- 30.Nollau P, Moser C, Weinland G and Wagener C. Detection of K-ras mutations in stools of patients with colorectal cancer by mutant-enriched PCR. Int J Cancer 1996;66: 332–336. doi: 10.1002/(SICI)1097-0215(19960503)66:3<332::AID-IJC11>3.0.CO;2-D [DOI] [PubMed] [Google Scholar]
- 31.Bearzatto A, Conte D, Frattini M, Zaffaroni N, Andriani F, Balestra D et al. p16 INK4A hypermethylation detected by fluorescent methylation specific PCR in plasmas from non small cell lung cancer. Clin Cancer Res. 2002;8: 3782–3787. [PubMed] [Google Scholar]
- 32.Chen J and Viola MV. A method to detect RAS point mutations in small subpopulations of cells. Anal Biochem. 1991;195: 51–56. [DOI] [PubMed] [Google Scholar]
- 33.Kahn SM, Jiang W, Culbertson TA, Weinstein IB, Williams GM, Tomita N et al. Rapid and sensitive nonradioactive detection of mutant K-ras genes via ‘enriched’ PCR amplification. Oncogene. 1991;6: 1079–83. [PubMed] [Google Scholar]
- 34.Mitsudomi T, Viallet J, Mulshine JL, Linnoila RI, Minna JD, Gazdar AF. Mutations of RAS genes distinguish a subset of non-small-cell lung cancer cell lines from small-cell lung cancer cell lines. Oncogene. 1991;6: 1353–62. [PubMed] [Google Scholar]
- 35.Van Krieken JH, Jung A, Kirchner T, Carneiro F, Seruca R, Bosman FT et al. KRAS mutation testing for predicting response to anti EGFR therapy for colorectal carcinoma: Proposal for an European quality assurance program. Virchows Arch. 2008;453: 417–431. doi: 10.1007/s00428-008-0665-y [DOI] [PubMed] [Google Scholar]
- 36.Molinari F, Felcioni L, Buscarino M, De Dosso S, Buttitta F, Malatesta S et al. Increased detection sensitivity for KRAS mutations enhances the prediction of anti-EGFR monoclonal antibody resistance in metastatic colorectal cancer. Clin Cancer Res. 2011;17: 4901–4914. doi: 10.1158/1078-0432.CCR-10-3137 [DOI] [PubMed] [Google Scholar]
- 37.Altimari A, de Biase D, De Maglio G, Gruppioni E, Capizzi E, Degiovanni A et al. 454 next generation-sequencing outperforms allele-specific PCR, sanger sequencing, and pyrosequencing for routine KRAS mutation analysis of formalin-fixed, paraffin embedded samples. Onco Targets Ther. 2013;6: 1057–1064. doi: 10.2147/OTT.S42369 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Tsiatis AC, Norris-Kirby A, Rich RG, Hafez MJ, Gocke CD, Eshleman JR et al. Comparison of Sanger sequencing, pyrosequencing, and melting curve analysis for the detection of KRAS mutations: diagnostic and clinical implications. J Mol Diagn. 2010;12: 425–32. doi: 10.2353/jmoldx.2010.090188 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All relevant data are within the paper and its Supporting Information files.