ABSTRACT
Screening of blood for human T-cell lymphotropic virus type 1 and type 2 (HTLV-1 and -2, respectively) is important to diagnose and prevent infection and ensure the safety of blood supplies. The Elecsys HTLV-I/II assay is a newly developed, electrochemiluminescence screening assay for the detection of HTLV-1/2 infection. The sensitivity and specificity of the Elecsys HTLV-I/II assay were determined using well-characterized HTLV-1/2-positive serum and plasma samples and routine diagnostic and blood donor samples expected to be HTLV negative, respectively. These results were compared with those for at least one of the following CE-marked assays at seven independent laboratories and the Roche Diagnostics facility in Penzberg, Germany: Abbott Architect rHTLV-I/II, Ortho Avioq HTLV-I/II Microelisa system, Abbott Prism HTLV-I/HTLV-II, and DiaSorin Murex HTLV I+II. Fujirebio INNO-LIA HTLV-I/II Score was used as a confirmatory assay. The Elecsys HTLV-I/II, Abbott Architect rHTLV-I/II, and Abbott Prism HTLV-I/HTLV-II assays detected all HTLV-1/2-positive samples (sensitivity, 100%). Sensitivity for Ortho Avioq HTLV-I/II was 98.63%. The Elecsys HTLV-I/II assay had a specificity of 99.95% in blood donor samples, which was comparable to results for the other assays (range, 99.91 to 100%). In routine diagnostic samples, the specificity of the Elecsys HTLV-I/II assay was 99.83%, compared with 99.70% for Abbott Architect rHTLV-I/II. Specificity for the Elecsys HTLV-I/II assay in potentially cross-reactive samples was 100%, compared with 99.0% for Ortho Avioq HTLV-I/II and 99.2% for DiaSorin Murex HTLV I+II. The Elecsys HTLV-I/II assay has the sensitivity and specificity to support its use as a routine screening assay for detecting HTLV infection.
KEYWORDS: Elecsys, HTLV-1, HTLV-2, blood screening, diagnosis, electrochemiluminescence, immunoassays, sensitivity, specificity
INTRODUCTION
Human T-cell lymphotropic virus (HTLV) is a cell-associated retrovirus, and four types have been identified to date. Type 1 (HTLV-1) is the only type that has been linked conclusively to disease. Infection with HTLV-1 is associated with the development of adult T-cell leukemia/lymphoma (ATL) and HTLV-1-associated myelopathy/tropical spastic paraparesis (HAM/TSP), with the former occurring in up to 5% and the latter in up to 3% of infected individuals (1, 2). HTLV-1 infection is also linked to a number of inflammatory conditions other than HAM/TSP (3). In contrast to HTLV-1, HTLV-2 has not been proven to cause disease (4). However, HTLV-2 infection has been associated with increased lymphocyte and platelet counts (5), an increased overall cancer mortality (6), sporadic reports of myelopathy (7), and a slightly increased risk of bacterial infections, particularly of the chest and bladder/kidney (8). The potential for HTLV-3 and HTLV-4 to cause disease is currently unknown (9).
HTLV-1 and HTLV-2 are the most prevalent types, but their global distribution is unusual and their prevalence is poorly understood. The most recent estimate suggests that 5 to 10 million people worldwide are infected with HTLV-1, but the actual number of infected individuals is likely to be much higher (10). HTLV-1 is highly endemic in the southwestern part of Japan; parts of sub-Saharan Africa, South America, and the Caribbean; and certain foci in the Middle East and Australo-Melanesia (10). In comparison, HTLV-2 is prevalent in indigenous populations of Africa, native inhabitants of Central and South America, and intravenous drug users in Europe and the United States (11, 12).
HTLV is transmitted through sexual intercourse, blood transfusions, organ transplantation, and needle reuse or from mother to child, predominantly during breastfeeding (13, 14). HTLV has also been shown to be acquired by humans through a bite from a simian T-lymphotropic virus type 1 (STLV-1)-infected nonhuman primate (15, 16). In the absence of a preventive vaccine and with limited treatment options for HTLV-related diseases, the focus is on prevention of viral transmission by diagnosing infected individuals and providing appropriate education. However, HTLV-1/2 infection remains underdiagnosed, partly because the majority of infections are asymptomatic.
Laboratory screening for HTLV-1/2 infection has become routine for blood donors in developed and some developing countries where the virus is endemic or where blood donors are considered to be at risk of infection (17, 18, 29). In low-seroprevalence populations, the positive predictive power of any assay is low and most positive results are false positives (4). For this reason, samples testing positive with screening tests should be tested with confirmatory assays. The development of screening assays with outstanding sensitivity and specificity would be valuable in improving HTLV-1/2 detection in the diagnostic setting, in the screening of blood donors, and for epidemiologic studies.
The Elecsys HTLV-I/II assay is a newly launched, double-antigen sandwich screening assay. The assay detects the immune response against the gp21 and p24 antigens (19). The immune response to p24 is one of the earliest detectable responses in HTLV-1/2 infection, and specific inclusion of this antigen in the Elecsys HTLV-I/II assay ensures that early infections are not missed (20). The objective of this study was to evaluate the clinical sensitivity and specificity of the Elecsys HTLV-I/II assay using well-characterized HTLV-1/2-positive serum and plasma samples and routine diagnostic and blood donor samples expected to be HTLV negative, respectively, in a head-to-head comparison with other commercially available, CE-marked HTLV-1/2 screening assays, and to assess its specificity using potentially cross-reactive samples.
RESULTS
Sensitivity.
The overall sensitivity of each assay is shown in Table 1. All samples were correctly identified as positive for HTLV-1/2 antibody by the Elecsys HTLV-I/II, Abbott Architect rHTLV-I/II, and Abbott Prism HTLV-I/HTLV-II assays (small sample size, n = 10) (sensitivity, 100%), while Ortho Avioq HTLV-I/II missed three HTLV-1-positive samples (sensitivity, 98.63%).
TABLE 1.
Cohort (geographic origin of samples)a | Elecsys HTLV-I/II |
Abbott Architect rHTLV-I/II |
Ortho Avioq HTLV-I/II |
Abbott Prism HTLV-I/HTLV-II |
||||
---|---|---|---|---|---|---|---|---|
No. tested | % sensitivity | No. tested | % sensitivity | No. tested | % sensitivity | No. tested | % sensitivity | |
Caribbean | 97 | 100 | 97 | 100 | ||||
South America | 134 | 100 | 134 | 100 | ||||
Europe/Middle East | 236 | 100 | 17 | 100 | 219 | 98.63b | ||
Africa | 3 | 100 | 3 | 100 | ||||
USA | 259 | 100 | 249 | 100 | 10 | 100 | ||
Japan | 420 | 100 | 420 | 100 | ||||
Overall sensitivity | 1,149 | 100 | 920 | 100 | 219 | 98.63 | 10 | 100 |
Countries within regions: Caribbean—Guadeloupe and Martinique; South America—Bolivia, Chile, Colombia, Ecuador, French Guiana, Honduras, and Peru; Europe/Middle East—France, Germany, Iran, Romania, and Spain; Africa—Ivory Coast, Morocco, and Senegal.
False-negative results were reported for three precharacterized HTLV-1-positive samples at the Paris laboratory.
Specificity.
The overall specificity in blood donor samples was 99.91% for the Abbott Architect rHTLV-I/II, 99.95% for the Elecsys HTLV-I/II (based on results for serum and plasma samples) and Ortho Avioq HTLV-I/II assays, and 100% for Abbott Prism HTLV-I/HTLV-II (Table 2). One sample from the Barcelona center was excluded from the specificity calculations, having given the following results: Elecsys HTLV-I/II reactive, Abbott Architect rHTLV-I/II nonreactive, and Fujirebio INNO-LIA HTLV-I/II score reactive (“env gp21 I/II” [“1+”], “env gp46 I/II,” “env gp46-I” [“+/−”]). No clarification by nucleic acid technology (NAT) testing or by a sequential bleed was possible due to anonymization of samples. The specificity results for the assays in blood donor samples at the individual laboratories are shown in Table S1 in the supplemental material.
TABLE 2.
Parameterb | Elecsys HTLV-I/II |
Abbott Architect rHTLV-I/II serum | Abbott Prism HTLV-I/HTLV-II serum | Ortho Avioq HTLV-I/II serum | ||
---|---|---|---|---|---|---|
Total | Serum | Plasma | ||||
No. of samples | ||||||
Total | 11,574a | 9,550a | 2,024 | 5,737a | 3,813 | 2,024 |
Negative | 11,568 | 9,546 | 2,022 | 5,734 | 3,812 | 2,022 |
With ≥1 s/co | ||||||
IRpos | 9 | 8 | 1 | 6 | 5 | 3 |
IRfpos | 6 | 6 | 0 | 5 | 5 | 2 |
RRpos | 9 | 8 | 1 | 6 | 0 | 2 |
RRfpos | 6 | 6 | 0 | 5 | 0 | 1 |
No. of samples with HTLV-1/2 immunoblot assay result/total no. tested | ||||||
Positive | 0/9 | 0/8 | 0/1 | 0/6 | 0/0 | 0/3 |
Indeterminate | 3/9 | 2/8 | 1/1 | 1/6 | 0/0 | 1/3 |
Negative | 6/9 | 6/8 | 0/1 | 5/6 | 0/0 | 2/3 |
% specificity IR ≥ 1 s/co | 99.95 | 99.94 | 100 | 99.91 | 99.87 | 99.90 |
95% confidence limit (two-sided, IR ≥ 1) | 99.89–99.98 | 99.86–99.98 | 99.82–100 | 99.80–99.97 | 99.69–99.96 | 99.64–99.99 |
% specificity RR ≥ 1 s/co | 99.95 | 99.94 | 100 | 99.91 | 100 | 99.95 |
95% confidence limit (two-sided, RR ≥ 1) | 99.89–99.98 | 99.86–99.98 | 99.82–100 | 99.80–99.97 | 99.90–100 | 99.72–100 |
One sample excluded from specificity calculations: Elecsys HTLV-I/II reactive; Abbott Architect rHTLV-I/II nonreactive; Fujirebio INNO-LIA HTLV I/II Score reactive, “env gp21 1/2” (“1+”), “env gp46 1/2,” “env gp46-I” (“+/−”). No clarification by nucleic acid technology (NAT) testing or by a sequential bleed was possible due to anonymization of samples.
Total, number of analyzed samples; negative, number of true-negative samples (excluding indeterminate, positive, and false-negative samples); IR, initially reactive; IRpos, initially reactive sample, including true-positive samples; IRfpos, initially reactive sample, not including true-positive samples; RR, repeatedly reactive; RRpos, repeatedly reactive sample, including true-positive samples; RRfpos, repeatedly reactive sample, not including true-positive samples; s/co, signal/cutoff ratio.
The overall specificity in routine diagnostic samples was 99.70% for Abbott Architect rHTLV-I/II (the only comparator assay used in this part of the study) and 99.83% for the Elecsys HTLV-I/II assay (Table 3). The specificity results for the assays in routine diagnostic samples at the individual laboratories are shown in Table S2.
TABLE 3.
Parametera | Elecsys HTLV-I/II serum | Abbott Architect rHTLV-I/II serum |
---|---|---|
No. of samples | ||
Total | 2,399 | 2,399 |
Negative | 2,336 | 2,336 |
With ≥1 s/co | ||
IRpos | 66 | 68 |
IRfpos | 4 | 7 |
RRpos | 66 | 68 |
RRfpos | 4 | 7 |
No. of samples with HTLV-1/2 immunoblot assay result/total no. tested | ||
Positive | 59/66 | 59/68 |
Indeterminate | 3/66 | 2/68 |
Negative | 4/66 | 7/68 |
% specificity IR ≥ 1 s/co | 99.83 | 99.70 |
95% confidence limit (two-sided, IR ≥ 1) | 99.56–99.95 | 99.38–99.88 |
% specificity RR ≥ 1 s/co | 99.83 | 99.70 |
95% confidence limit (two-sided, RR ≥ 1) | 99.56–99.95 | 99.38–99.88 |
Total, number of analyzed samples; negative, number of true-negative samples (excluding indeterminate, positive, and false-negative samples); IR, initially reactive; IRpos, initially reactive sample, including true-positive samples; IRfpos, initially reactive sample, not including true-positive samples; RR, repeatedly reactive; RRpos, repeatedly reactive sample, including true-positive samples; RRfpos, repeatedly reactive sample, not including true-positive samples; s/co, signal/cutoff ratio.
There were no notable differences in specificity results across the individual laboratories.
Cross-reactivity/interference.
The Elecsys HTLV-I/II assay was shown to be unaffected by icterus (bilirubin, ≤1,129 μmol/liter or ≤66 mg/dl), hemolysis (hemoglobin [Hb], ≤0.3 mmol/liter or ≤0.5 g/dl), lipemia (intralipid, ≤2,000 mg/dl), and biotin (≤246 nmol/liter or ≤60 ng/ml) during development (data not shown).
In the investigation of possible cross-reacting factors at the laboratories in Paris and Penzberg, no false-positive result was found with the Elecsys HTLV-I/II assay (specificity, 100%). At the Paris laboratory, the Ortho Avioq HTLV-I/II assay showed one false-positive result for a precharacterized hepatitis B virus (HBV)-positive sample (specificity, 99.0%; signal/cutoff ratio [s/co] initial reactive [IR] = 1.22; repeats = 1.22/0.64). At the Penzberg laboratory, the DiaSorin Murex HTLV I+II assay showed one false-positive result for a precharacterized rubella virus-positive sample (specificity, 99.2%; s/co IR = 1.99; repeats = 1.89/0.28).
DISCUSSION
This study demonstrates that the Elecsys HTLV-I/II assay has the sensitivity (100%) and specificity (99.95% in blood donor samples, 99.83% in routine diagnostic samples), regardless of the geographic origin of the samples, virus type, or location of the testing laboratory, to support its use as a routine screening assay worldwide.
The Abbott Architect rHTLV-I/II and Abbott Prism HTLV-I/HTLV-II assays also identified all precharacterized HTLV-1/2-positive samples correctly (sensitivity, 100%). These findings are consistent with previously reported results. In an earlier study of 406 precharacterized HTLV-1/2-positive samples, Abbott Architect rHTLV-I/II had a sensitivity of 100% (21, 22). Similarly, in a study of 714 confirmed HTLV-1/2-positive samples, a sensitivity of 100% was reported for Abbott Prism HTLV-I/HTLV-II (23). Based on 636 precharacterized HTLV-1/2-positive samples, the sensitivity of Ortho Avioq HTLV-I/II has been reported previously as 100% (24). This compares with a sensitivity of 98.63% for Ortho Avioq HTLV-I/II in the present study (three false-negative results at the Paris laboratory).
Overall, the Elecsys HTLV-I/II assay (99.95%) had specificity comparable to that of Ortho Avioq HTLV-I/II (99.95%), numerically lower specificity than Abbott Prism HTLV-I/HTLV-II (100%), and numerically higher specificity than Abbott Architect rHTLV-I/II (99.91%) when tested in blood donor samples. The specificity of the Elecsys HTLV-I/II assay (99.83%) was numerically higher than that of Abbott Architect rHTLV-I/II (99.70%) when tested in routine diagnostic samples. The confidence intervals of the specificity data for all four screening assays were overlapping, and therefore, these numerical differences were not statistically significant. All assays showed consistent specificity regardless of the location of the testing laboratory. The results for the comparator assays were consistent with previous data. In a study of 5,646 blood donor samples, Abbott Architect rHTLV-I/II had a specificity of 99.95%, compared with 99.86% when tested in 692 routine diagnostic samples (21, 22). Ortho Avioq HTLV-I/II had a specificity of 99.95% in a study of 11,415 blood donor samples (24), while Abbott Prism HTLV-I/HTLV-II had a specificity of 99.93% in a study of 21,943 blood donor samples (23).
In certain geographic regions, particularly in Africa, which represents the largest area where HTLV-1 infection is endemic, there is a wide genetic variability of HTLV-1 genotypes and a large pattern of indeterminate serology exists (25, 26). Unfortunately, the number of samples available for analysis from African regions during this study was limited. In addition, it was not possible to obtain genotypic information for specific samples due to the unavailability of whole blood in order to extract peripheral blood mononuclear cells (PBMCs). Therefore, future studies might aim to investigate the performance of HTLV screening assays in a larger series of samples from African individuals, including those with an HTLV-1 Gag-indeterminate pattern (HGIP) or new Western blot (NWB) profile, in order to assess the performance of these assays in this extremely important demographic group.
The Elecsys HTLV-I/II assay is a quick (assay time, 18 min), one-step (double-antigen sandwich) assay, and the simplicity of the methodology and speed of the assay offer clear benefits in terms of the number of samples that can be processed by a testing facility within a given time. Furthermore, the antigen combination in the Elecsys HTLV-I/II assay was not associated with any cross-reactivity in our study.
In summary, the Elecsys HTLV-I/II assay is a reliable test for screening for HTLV-I/II infection in the blood and organ donation settings and for diagnosis where there is suspicion of HTLV-associated diseases, such as ATL, HAM/TSP, and undiagnosed myelopathy.
MATERIALS AND METHODS
Participating laboratories and compliance.
Seven independent laboratories from six countries (Austria [Innsbruck], France [Paris], Germany [Hagen and Augsburg], Japan [Nagasaki], Portugal [Porto], and Spain [Barcelona]) and the Roche Diagnostics facility at Penzberg in Germany participated in the study (Table 4). The study was conducted in compliance with relevant directives of the European Union (EU) parliament and EU council. Ethical approval for the study was obtained from independent review bodies when required by national regulations, and waivers from ethical committees were obtained by each laboratory participating in the study, as necessary. All samples were deidentified or coded prior to use, and therefore, individual patient data were not available for this study.
TABLE 4.
Laboratory | Comparator assay(s) used | Platform for the Elecsys HTLV-I/II assay |
---|---|---|
Central Institute for Blood Transfusion and Immunology, University Hospital, Innsbruck, Austria | Abbott Architect rHTLV-I/II | Modular Analytics E170 |
Ortho Avioq HTLV-I/II Microelisa system | ||
DRK Blutspendedienst West, Hagen, Germany | Abbott Prism HTLV-I/HTLV-II | cobas e 602 |
Labor Schottdorf MVZ, Augsburg, Germany | Abbott Architect rHTLV-I/II | Modular Analytics E170 |
Instituto Português do Sangue e da Transplantaçăo, Porto, Portugal | Abbott Architect rHTLV-I/II | cobas e 601 |
Banc de Sang i Teixits, Barcelona, Spain | Abbott Architect rHTLV-I/II | cobas e 411 |
Institut National de la Transfusion Sanguine, Paris, France | Ortho Avioq HTLV-I/II Microelisa system | cobas e 411 |
Nagasaki University Hospital, Nagasaki, Japan | Abbott Architect rHTLV-I/II | cobas e 411 |
Fujirebio Lumipulse HTLV-I | ||
Fujirebio Serodia-HTLV-I PA | ||
Roche Diagnostics, Penzberg, Germany | Abbott Architect rHTLV-I/II (performed at Microcoat Biotechnologie GmbH, Bernried, Germany) | Modular Analytics E170 |
Abbott Prism HTLV-I/HTLV-II (data provided by sample provider) | cobas e 411 | |
DiaSorin Murex HTLV I+II | cobas e 601 |
Assays.
The Elecsys HTLV-I/II assay (Roche Diagnostics International Ltd., Rotkreuz, Switzerland) is an electrochemiluminescence immunoassay for use on the cobas e analyzers. The three principal comparator screening assays were Ortho Avioq HTLV-I/II Microelisa system (Avioq Inc., Durham, NC, USA), which is an enzyme-linked immunosorbent assay, and Abbott Prism HTLV-I/HTLV-II (Abbott Laboratories, Abbott Park, IL, USA) and Abbott Architect rHTLV-I/II (Abbott Laboratories, Wiesbaden, Germany), which are chemiluminescent immunoassays.
Each laboratory evaluated the Elecsys HTLV-I/II assay and at least one comparator assay (Table 4). All assays were performed according to the manufacturers' instructions. All laboratories used the master pilot lot of the Elecsys HTLV-I/II assay. They were provided with Elecsys HTLV PreciControl and Elecsys HTLV-I/II CalSet (both from single lots) for control and calibration purposes. The confirmatory test used in the specificity evaluation was Fujirebio INNO-LIA HTLV-I/II Score (Fujirebio Diagnostics Inc., Malvern, PA, USA) at all laboratories. In the evaluation of cross-reacting factors, the comparator screening assay at the Penzberg facility was DiaSorin Murex HTLV I+II (DiaSorin, Saluggia, Italy), which is an enzyme-linked immunosorbent assay. Key features of the screening assays are summarized in Table 5.
TABLE 5.
Assay name (reference[s]) | Mfr | Assay type | Analyzer system | Key principle of assay | Testing time (min) |
---|---|---|---|---|---|
Elecsys HTLV-I/II (19) | Roche Diagnostics | One-step double-antigen sandwich chemiluminescent immunoassay | Modular Analytics E170 or cobas e 411, cobas e 601, cobas e 602 | Detects antibodies to viral recombinant antigens gp21 and p24 | 18 |
Abbott Architect rHTLV-I/II (21) | Abbott Laboratories | Two-step chemiluminescent immunoassay | Architect i system | Detects antibodies to viral synthetic peptide gp46 and recombinant antigen gp21 | 29 |
Ortho Avioq HTLV-I/II Microelisa system (24) | Avioq Inc. | ELISA | Microelisa plate reader | Detects antibodies to viral antigens (purified viral lysate) and recombinant HTLV-1 p21E antigen | 150 |
Abbott Prism HTLV-I/HTLV-II (23, 27) | Abbott Laboratories | Three-step sandwich chemiluminescent immunoassay | Abbott Prism system | Detects antibodies to HTLV-1 and HTLV-2 antigens | 54 |
DiaSorin Murex HTLV I+II (28) | DiaSorin | ELISA | Microelisa plate reader | Detects antibodies to HTLV-1 and HTLV-2 antigens | 90 |
Abbreviations: Mfr, manufacturer; ELISA, enzyme-linked immunosorbent assay.
Samples used for sensitivity analysis.
Assay sensitivity was evaluated at the Roche Diagnostics facility at Penzberg using frozen serum samples provided by Slieagen, USA; the American Red Cross, USA; Universidad Peruana Cayetano Heredia, Peru; Laboratoire CERBA, France; ProMedDx LLC, USA; Trina Bioreactives AG, Switzerland; and the University of Tokyo, Japan (906 samples). The national reference center for human immunodeficiency virus (HIV), hepatitis B virus (HBV), and hepatitis C virus (HCV) in transfusion (INTS), Paris, France (219 samples), and the laboratory in Barcelona, Spain (24 samples), also analyzed samples. All samples (n = 1,149) were previously characterized as positive for HTLV-1/2 (926 HTLV-1 positive, 200 HTLV-2 positive, and 23 HTLV-1/2 positive but not subtyped). Assay sensitivity was calculated as the percentage of precharacterized HTLV-1/2-positive samples identified as positive by the assay.
Samples used for specificity analysis.
At most laboratories, assay specificity was assessed in fresh serum and EDTA plasma samples provided by the individual laboratory (Barcelona used frozen samples and Nagasaki used frozen and fresh samples). All were leftover samples from routine diagnostic requests and blood donations. Assay specificity was tested in 11,575 samples from blood donors and 2,399 routine diagnostic samples that had previously been identified as negative for HTLV-1/2. Assay specificity was evaluated using blood donor samples at four laboratories (Innsbruck [4,048 samples], Hagen [3,813 samples], Porto [2,166 samples], and Barcelona [1,548 samples]) and routine diagnostic samples at two laboratories (Augsburg [1,500 samples] and Nagasaki [899 samples]). Assay specificity was calculated as the percentage of true HTLV-negative samples that tested negative with the assay.
Samples used to test for cross-reacting factors.
Assays were also tested for interference using 222 frozen serum samples containing potentially cross-reacting factors. These samples had been characterized previously as positive for hepatitis A virus (HAV), HBV, HCV, HIV, herpes simplex virus, rubella virus, Epstein-Barr virus, Escherichia coli, elevated rheumatoid factor (>200 IU/ml), and various autoimmune disorders. Testing for potential cross-reacting factors was conducted at the Roche Diagnostics facility at Penzberg (122 samples) and at the INTS laboratory in Paris (100 samples).
Methods and analyses.
For all screening assays, determinations were performed as single measurements. In the evaluation of specificity, samples giving an initial reactive (IR) result were retested in duplicate and were considered to be repeatedly reactive (RR) if either of the retest results had a signal/cutoff (s/co) ratio of ≥1.00. The exception to this was the laboratory at Hagen in Germany, where IR samples were retested with a single determination. All positive samples in the specificity evaluation were subjected to confirmatory testing with HTLV immunoblotting. A sample was considered to be true positive, indeterminate, or true negative depending on the result of the HTLV immunoblot assay.
Supplementary Material
ACKNOWLEDGMENTS
We thank Satomi Sonoyama for technical assistance at the Nagasaki laboratory. Medical writing assistance was provided by Elements Communications Ltd., Westerham, United Kingdom.
Syria Laperche, Silvia Sauleda, Maria Piron, Annelies Mühlbacher, Harald Schennach, Lucinda Queirós, Naoki Uno, Katsunori Yanagihara, Roland Imdahl, Ariann Hey, and Toshiki Watanabe declare no conflicts of interest. Markus Klinkicht, Walter Melchior, and Peter Muench are currently employed by Roche Diagnostics GmbH (Penzberg, Germany). Volkmar Schottstedt is currently self-employed (Head of Schottstedt Lab-Consulting, Iserlohn, Germany). He has no further conflicts of interest to declare.
This study and the medical writing assistance by Elements Communications Ltd. were funded by Roche Diagnostics GmbH (Penzberg, Germany).
Footnotes
Supplemental material for this article may be found at https://doi.org/10.1128/JCM.00169-17.
REFERENCES
- 1.Tsukasaki K, Hermine O, Bazarbachi A, Ratner L, Ramos JC, Harrington W Jr, O'Mahony D, Janik JE, Bittencourt AL, Taylor GP, Yamaguchi K, Utsunomiya A, Tobinai K, Watanabe T. 2009. Definition, prognostic factors, treatment, and response criteria of adult T-cell leukemia-lymphoma: a proposal from an international consensus meeting. J Clin Oncol 27:453–459. doi: 10.1200/JCO.2008.18.2428. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Gessain A, Mahieux R. 2012. Tropical spastic paraparesis and HTLV-1 associated myelopathy: clinical, epidemiological, virological and therapeutic aspects. Rev Neurol (Paris) 168:257–269. doi: 10.1016/j.neurol.2011.12.006. [DOI] [PubMed] [Google Scholar]
- 3.Martin F, Taylor GP, Jacobson S. 2014. Inflammatory manifestations of HTLV-1 and their therapeutic options. Expert Rev Clin Immunol 10:1531–1546. doi: 10.1586/1744666X.2014.966690. [DOI] [PubMed] [Google Scholar]
- 4.Morrison BJ, Labo N, Miley WJ, Whitby D. 2015. Serodiagnosis for tumor viruses. Semin Oncol 42:191–206. doi: 10.1053/j.seminoncol.2014.12.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Bartman MT, Kaidarova Z, Hirschkorn D, Sacher RA, Fridey J, Garratty G, Gibble J, Smith JW, Newman B, Yeo AE, Murphy EL, HTLV Outcomes Study (HOST) Investigators. 2008. Long-term increases in lymphocytes and platelets in human T-lymphotropic virus type II infection. Blood 112:3995–4002. doi: 10.1182/blood-2008-05-155960. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Biswas HH, Kaidarova Z, Garratty G, Gibble JW, Newman BH, Smith JW, Ziman A, Fridey JL, Sacher RA, Murphy EL, HTLV Outcomes Study. 2010. Increased all-cause and cancer mortality in HTLV-II infection. J Acquir Immune Defic Syndr 54:290–296. doi: 10.1097/QAI.0b013e3181cc5481. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Araujo A, Hall WW. 2004. Human T-lymphotropic virus type II and neurological disease. Ann Neurol 56:10–19. doi: 10.1002/ana.20126. [DOI] [PubMed] [Google Scholar]
- 8.Murphy EL, Wang B, Sacher RA, Fridey J, Smith JW, Nass CC, Newman B, Ownby HE, Garratty G, Hutching ST, Schreiber GB. 2004. Respiratory and urinary tract infections, arthritis, and asthma associated with HTLV-I and HTLV-II infection. Emerg Infect Dis 10:109–116. doi: 10.3201/eid1001.020714. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.LeBreton M, Switzer WM, Djoko CF, Gillis A, Jia H, Sturgeon MM, Shankar A, Zheng H, Nkeunen G, Tamoufe U, Nana A, Le Doux Diffo J, Tafon B, Kiyang J, Schneider BS, Burke DS, Wolfe ND. 2014. A gorilla reservoir for human T-lymphotropic virus type 4. Emerg Microbes Infect 3:e7. doi: 10.1038/emi.2014.7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Gessain A, Cassar O. 2012. Epidemiological aspects and world distribution of HTLV-1 infection. Front Microbiol 3:388. doi: 10.3389/fmicb.2012.00388. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Zella D, Cavicchini A, Salemi M, Casoli C, Lori F, Achilli G, Cattaneo E, Landini V, Bertazzoni U. 1993. Molecular characterization of two isolates of human T cell leukaemia virus type II from Italian drug abusers and comparison of genome structure with other isolates. J Gen Virol 74:437–444. doi: 10.1099/0022-1317-74-3-437. [DOI] [PubMed] [Google Scholar]
- 12.Roucoux DF, Murphy EL. 2004. The epidemiology and disease outcomes of human T-lymphotropic virus type II. AIDS Rev 6:144–154. [PubMed] [Google Scholar]
- 13.Pique C, Jones KS. 2012. Pathways of cell-cell transmission of HTLV-1. Front Microbiol 3:378. doi: 10.3389/fmicb.2012.00378. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Hino S. 2011. Establishment of the milk-borne transmission as a key factor for the peculiar endemicity of human T-lymphotropic virus type 1 (HTLV-1): the ATL Prevention Program Nagasaki. Proc Jpn Acad Ser B Phys Biol Sci 87:152–166. doi: 10.2183/pjab.87.152. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Kazanji M, Mouinga-Ondémé A, Lekana-Douki-Etenna S, Caron M, Makuwa M, Mahieux R, Gessain A. 2015. Origin of HTLV-1 in hunters of nonhuman primates in Central Africa. J Infect Dis 211:361–365. doi: 10.1093/infdis/jiu464. [DOI] [PubMed] [Google Scholar]
- 16.Filippone C, Betsem E, Tortevoye P, Cassar O, Bassot S, Froment A, Fontanet A, Gessain A. 2015. A severe bite from a nonhuman primate is a major risk factor for HTLV-1 infection in hunters from central Africa. Clin Infect Dis 60:1667–1676. doi: 10.1093/cid/civ145. [DOI] [PubMed] [Google Scholar]
- 17.Malm K, Kjerstadius T, Andersson S. 2010. Evaluation of a new screening assay for HTLV-1 and -2 antibodies for large-scale use. J Med Virol 82:1606–1611. doi: 10.1002/jmv.21867. [DOI] [PubMed] [Google Scholar]
- 18.Xie J, Ge S, Zhang Y, Lin Y, Ni H, Zhang J, Chen C. 2015. The prevalence of human T-lymphotropic virus infection among blood donors in southeast China, 2004-2013. PLoS Negl Trop Dis 9:e0003685. doi: 10.1371/journal.pntd.0003685. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Roche Diagnostics. 2015. Elecsys HTLV-I/II fact sheet. Roche Diagnostics, Indianapolis, IN: http://www.cobas.com/content/dam/cobas_com/pdf/product/Elecsys-HTLV-I-II/BrandBridge%20Elecsys_HTLV-I_II_FS.pdf Accessed 29 June 2016. [Google Scholar]
- 20.Manns A, Murphy EL, Wilks R, Haynes G, Figueroa JP, Hanchard B, Barnett M, Drummond J, Waters D, Cerney M, Seals JR, Alexander SS, Lee H, Blattner WA. 1991. Detection of early human T-cell lymphotropic virus type I antibody patterns during seroconversion among transfusion recipients. Blood 77:896–905. [PubMed] [Google Scholar]
- 21.Abbott Laboratories. 2008. Architect rHTLV-I/II. Package insert. Abbott Laboratories, Abbott Park, IL: http://www.ilexmedical.com/files/PDF/HTLV_ARC.pdf Accessed 29 June 2016. [Google Scholar]
- 22.Kapprell HP, Stieler M, Oer M, Goller A, Hausmann M, Schochetman G, Devare SG, Qiu X. 2010. Evaluation of a new third-generation ARCHITECT rHTLV-I/II assay for blood screening and diagnosis. Diagn Microbiol Infect Dis 67:61–69. doi: 10.1016/j.diagmicrobio.2009.12.021. [DOI] [PubMed] [Google Scholar]
- 23.Abbott Laboratories. 2007. Abbott Prism HTLV-I/HTLV-II. Product information. Abbott Laboratories, Abbott Park, IL: http://www.fda.gov/downloads/biologicsbloodvaccines/ucm091045.pdf Accessed 29 June 2016. [Google Scholar]
- 24.Avioq Inc. 2015. Avioq HTLV-I/II Microelisa system. Package insert. Avioq Inc., Research Triangle Park, NC: http://www.fda.gov/downloads/BiologicsBloodVaccines/BloodBloodProducts/ApprovedProducts/LicensedProductsBLAs/BloodDonorScreening/InfectiousDisease/UCM298517.pdf Accessed 29 June 2016. [Google Scholar]
- 25.Mauclère P, Le Hesran JY, Mahieux R, Salla R, Mfoupouendoun J, Abada ET, Millan J, de Thé G, Gessain A. 1997. Demographic, ethnic, and geographic differences between human T cell lymphotropic virus (HTLV) type I-seropositive carriers and persons with HTLV-I Gag-indeterminate Western blots in Central Africa. J Infect Dis 176:505–509. doi: 10.1086/514071. [DOI] [PubMed] [Google Scholar]
- 26.Filippone C, Bassot S, Betsem E, Tortevoye P, Guillotte M, Mercereau-Puijalon O, Plancoulaine S, Calattini S, Gessain A. 2012. A new and frequent human T-cell leukemia virus indeterminate Western blot pattern: epidemiological determinants and PCR results in central African inhabitants. J Clin Microbiol 50:1663–1672. doi: 10.1128/JCM.06540-11. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Andersson S, Thorstensson R, Ramirez KG, Krook A, von Sydow M, Dias F, Biberfeld G. 1999. Comparative evaluation of 14 immunoassays for detection of antibodies to the human T-lymphotropic virus types I and II using panels of sera from Sweden and West Africa. Transfusion 39:845–851. doi: 10.1046/j.1537-2995.1999.39080845.x. [DOI] [PubMed] [Google Scholar]
- 28.DiaSorin. 2015. Murex HTLV I+II. Product information. DiaSorin, Saluggia, Italy: http://www.annardx.com/productos/images/productos/banco-de-sangre/infecciosas/ifuk-en-8e2204-0979440434.pdf Accessed 11 October 2016. [Google Scholar]
- 29.Murphy EL. 2016. Infection with human T-lymphotropic virus types-1 and -2 (HTLV-1 and -2): implications for blood transfusion safety. Transfus Clin Biol 23:13–19. doi: 10.1016/j.tracli.2015.12.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data