Skip to main content
Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
. 2017 Jun 23;55(7):2198–2203. doi: 10.1128/JCM.00603-17

Evaluation of Aptima Zika Virus Assay

Ping Ren a,, Daniel A Ortiz a, Ana C B Terzian b, Tatiana E Colombo b, Mauricio L Nogueira b, Nikos Vasilakis c, Michael J Loeffelholz a,
Editor: Yi-Wei Tangd
PMCID: PMC5483922  PMID: 28468854

ABSTRACT

The Zika virus (ZIKV) epidemic in the Americas poses a public health emergency that requires a swift response. Accurate and reliable ZIKV diagnostic tests serve as an important tool for limiting the spread of ZIKV infections. The Aptima Zika virus assay (Hologic, Marlborough, MA) performed on the automated Panther system is a rapid and high-throughput method for detecting ZIKV RNA using transcription-mediated amplification (TMA) technology. We evaluated the performance characteristics of the Aptima Zika virus assay on clinical serum and urine specimens (n = 124) from two different patient populations and samples spiked with ZIKV from three different lineages (n = 10). Compared to the real-time reverse transcription-PCR (rRT-PCR) reference method, the Aptima ZIKV assay detected ZIKV RNA with a diagnostic accuracy of 94.8% (95% confidence interval [CI], 89.4 to 97.6), a sensitivity of 94.7% (95% CI, 73.5 to 99.9), and a specificity of 94.8% (95% CI, 88.9 to 97.8). Similar results were obtained regardless of whether a serum or urine source was used. The limits of detection of the assay at a 95% detection probability were 11.5 genome copy equivalents (GCE)/ml (95% fiducial limits, 7.9 to 20.2) in serum and 17.9 GCE/ml (95% fiducial limits, 13.1 to 27.5) in urine. The Aptima Zika virus assay results were highly reproducible (99%), and no cross-reactivity was seen during the testing of a panel of 95 specimens with potentially interfering substances, such as clinically relevant bacteria, fungi, and viruses, including other flaviviruses. The excellent performance characteristics and the convenience of a fully automated testing system make the Aptima ZIKV assay an attractive choice for clinical laboratories detecting ZIKV RNA from serum and urine.

KEYWORDS: Aptima Zika virus assay, sensitivity, specificity

INTRODUCTION

Zika virus (ZIKV), a member of the Flaviviridae family, is named after the Ziika Forest in Uganda, Africa, from which it was first identified in 1947 (1). The virus is primarily transmitted between humans by Aedes spp. mosquitoes, the same vectors that also transmit dengue, yellow fever, and chikungunya viruses. Prior to 2007, only 13 cases of human ZIKV infections were reported, which is partially attributed to the mild nature of ZIKV infection (2). Up to 80% of individuals infected with ZIKV are asymptomatic, while symptomatic individuals display a broad range of symptoms that may include fever, headache, rash, arthralgia, conjunctivitis, and myalgia. However, complications of ZIKV infection have been shown to cause congenital microcephaly and trigger Guillain-Barré syndrome (3). As of February 2017, a total of 76 countries and territories reported cases of ZIKV transmission, with active ZIKV transmission still occurring in the Pacific islands and the Americas (4). The increasing geographical range of ZIKV stresses the importance of diagnostic laboratory tests in confirming infection and mitigating the spread of ZIKV infection.

A diagnosis of ZIKV infection is confirmed by the detection of viral components (e.g., viral RNA, viral proteins, or virus isolation) and/or the host humoral immune response. At early stages of infection (≤14 days after symptom onset), ZIKV RNA can be detected in whole blood (also serum and plasma) and urine samples from patients (5, 6). Real-time reverse transcription-PCR (rRT-PCR) is commonly used for ZIKV RNA detection because of its high sensitivity and specificity. The U.S. Centers for Disease Control and Prevention (CDC) Trioplex real-time RT-PCR assay was the first diagnostic test granted emergency use authorization (EUA) by the Food and Drug Administration (FDA) for the qualitative detection and differentiation of ZIKV RNA from that of dengue and chikungunya viruses (7). Additional EUA-approved ZIKV assays have since been developed, yet their diagnostic utility in a clinical setting has not been thoroughly evaluated.

The Aptima Zika virus assay (Hologic, Marlborough, MA) obtained EUA approval from the FDA on June 17, 2016, for the qualitative detection of ZIKV RNA from human serum, plasma, and urine (collected alongside a patient-matched serum or plasma specimen) (8). The Aptima ZIKV assay uses transcription-mediated amplification (TMA) that targets two highly conserved sequences in the NS2 and NS4/NS5 regions of ZIKV RNA (9). The entire assay is performed in a single tube on an automated Panther system, and initial results are produced in 3.5 h. In this study, we evaluated the performance characteristics of the Aptima Zika virus assay on clinical and spiked specimens at the University of Texas Medical Branch (UTMB) clinical microbiology laboratory.

RESULTS

Diagnostic accuracy.

The diagnostic accuracy of the Aptima ZIKV assay was evaluated using clinical serum and urine specimens collected from two different patient populations, along with a subset of ZIKV-spiked samples.

In our service area, 104 clinical specimens from 70 patients met inclusion criteria for ZIKV RNA testing. Female patients comprised 84% (59/70) of the total test volume, and 67% (40/59) of the females were pregnant women. Thirty-eight patients displayed one or more symptoms of ZIKV infection such as headache, fever, rash, conjunctivitis, and joint or muscle pain. Among the 104 U.S. clinical samples tested by the Aptima ZIKV assay, 3 serum and 3 urine specimens tested positive for ZIKV RNA (Table 1). These positive specimens were paired serum and urine specimens from three patients who were symptomatic and had recent travel to an area with active ZIKV transmission (Dominican Republic, Barbados, and Guatemala). Only one pair of serum and urine samples out of these three paired Aptima ZIKV assay-positive samples correlated with rRT-PCR results. The diagnostic accuracy of the Aptima ZIKV assay in testing 104 specimens from our service area was 95.2% (95% confidence interval [CI], 89.0 to 98.2). The diagnostic accuracies were similar regardless of whether the sample was from a serum or urine source.

TABLE 1.

Aptima ZIKV assay results compared to rRT-PCR results in clinical and spiked specimens

Aptima ZIKV assay result rRT-PCR results (no.)
Positive Negative
Clinical specimens (United States)
    Serum (n = 70)
        Positive 1 2
        Negative 1 66
    Urine (n = 34)
        Positive 1 2
        Negative 0 31
Clinical specimens (Brazil)
    Serum (n = 20)
        Positive 10 2
        Negative 0 8
Spiked specimens
    Serum (n = 5)
        Positive 3 0
        Negative 0 2
    Urine (n = 5)
        Positive 3 0
        Negative 0 2

The utility of the Aptima ZIKV assay in a patient population from an area in which ZIKV is endemic was assessed using 20 serum samples from Brazilian residents who had previous ZIKV rRT-PCR testing results by a previously described method (5). All patients displayed two or more symptoms of ZIKV infection, and were tested within 6 days of symptom onset by rRT-PCR for ZIKV, dengue virus serotypes 1 to 4 (10), and chikungunya virus (11). Among the 20 samples received, 10 were positive and 10 were negative for ZIKV infection. All patient specimens were negative for dengue virus serotypes 1 to 4 and for chikungunya virus. The diagnostic accuracy of the Aptima ZIKV assay in the Brazilian patient population was 90% (95% CI, 68.7 to 98.4) (Table 1).

The ability of the Aptima ZIKV assay to detect Zika viruses of different lineages was assessed by using a panel of spiked ZIKV-negative specimens. A total of five serum and five urine samples were spiked with one of the following: Mexican ZIKV strain (Mex-81), Cambodian ZIKV strain (FSS13025), African ZIKV strain (DAK), dengue virus, or no virus. The Aptima ZIKV assay and the rRT-PCR reference method accurately identified all samples.

In summary, a total of 134 samples were tested by the Aptima ZIKV assay and the rRT-PCR reference method. Compared to the reference method, the Aptima ZIKV assay displayed a diagnostic accuracy of 94.8% (95% CI, 89.4 to 97.6) with a sensitivity of 94.7% (95% CI, 73.5 to 99.9) and a specificity of 94.8% (95% CI, 88.9 to 97.8) (Table 2).

TABLE 2.

Overall performance characteristics of the Aptima ZIKV assay (n = 134)

Aptima ZIKV assay result rRT-PCR results
Sensitivity (% [95% CI]) Specificity (% [95% CI]) Diagnostic accuracy (% [95% CI])
Positive Negative
Positive 18 6 94.7 (73.5–99.9) 94.8 (88.9–97.8) 94.8 (89.4–97.6)
Negative 1 109

Limits of detection.

The limits of detection (LoD) of the Aptima ZIKV assay were determined by serial dilutions of ZIKV stock (Asian lineage; FSS13025) in ZIKV-negative serum and urine. Final urine dilutions were mixed with urine transport media to simulate clinical collection practices. Several replicates were tested at each dilution to calculate the lowest concentration reliably detected. The LoD of the Aptima ZIKV assay were 11.5 genome copy equivalents (GCE)/ml (95% fiducial limits, 7.9 to 20.2) in serum, and 17.9 GCE/ml (95% fiducial limits, 13.1 to 27.5) in urine at a 95% detection probability (Table 3).

TABLE 3.

Serial dilutions of ZIKV stock (2.18 × 109 GCE/ml) in serum and urine for limits of detection

Dilution factor Presumptive GCE/ml Serum
Urine
No. of samples No. (%) positive 95% CI No. of samples No. (%) positive 95% CI
1:100 21,800,000 1 1 (100) 2.5–100 1 1 (100) 2.5–100
1:1,000 2,180,000 1 1 (100) 2.5–100 1 1 (100) 2.5–100
1:10,000 218,000 1 1 (100) 2.5–100 1 1 (100) 2.5–100
1:100,000 21,800 1 1 (100) 2.5–100 1 1 (100) 2.5–100
1:1,000,000 2,180 1 1 (100) 2.5–100 1 1 (100) 2.5–100
1:10,000,000 218 20 20 (100) 83–100 20 20 (100) 83–100
1:100,000,000 21.8 50 50 (100) 93–100 50 49 (98) 89–100
1:200,000,000 10.9 50 46 (92) 81–98 50 42 (84) 71–93
1:1,000,000,000 2.18 50 33 (66) 51–79 50 12 (24) 13–38
1:2,000,000,000 1.09 50 9 (18) 9–31 50 3 (6) 1–17
1:20,000,000,000 0.109 50 1 (2) 0.1–11 50 0 (0) 0–7
1:200,000,000,000 0.0109 30 0 (0) 0–12 30 0 (0) 0–12

CI, confidence interval.

Analytical specificity.

The specificity of the Aptima ZIKV assay was challenged in the presence of interfering substances to determine the analytical specificity. Clinical sera and urine from 35 patients positive for Chlamydia trachomatis, Neisseria gonorrhoeae, hepatitis C virus (HCV), hepatitis B virus (HBV), BK virus, cytomegalovirus (CMV), or human immunodeficiency virus (HIV) nucleic acid produced expected results in the presence or absence of spiked ZIKV. In addition, 46 samples spiked with yeast, bacteria, or closely related arboviruses (dengue, Kunjin, yellow fever, and chikungunya viruses) did not affect the outcome of the Aptima ZIKA assay. Collectively, the analytical sensitivity of the Aptima ZIKV assay was 100%, with no cross-reactivity or interference seen in the 81 samples tested.

Reproducibility.

The reproducibility within the laboratory and between laboratories was tested using two panels. The intralaboratory panel consisted of 40 positive and 45 negative samples, while the interlaboratory panel contained 18 positive and 22 negative samples. The intra- and interlaboratory reproducibility results were 100% and 97.5%, respectively. The single discrepancy was reported on a urine specimen sent to an outside laboratory.

DISCUSSION

The ZIKV epidemic continues to be problematic in many countries. As a response, many companies have developed their own diagnostic assays for detecting ZIKV infection. However, limited clinical data are available for these new assays, which makes the decision of choosing a diagnostic assay to implement challenging.

In this study, we evaluated the performance characteristics of the Aptima ZIKV assay compared to the rRT-PCR reference method. Among the 134 specimens tested by both assays, the Aptima ZIKV assay displayed excellent diagnostic accuracy, sensitivity, and specificity, which were 94.8%, 94.7%, and 94.8%, respectively. A total of seven discrepant results were found between the two assays. Although additional testing of these specimens was not performed to elucidate false-positive or false-negative results, some of the discrepant results may be attributed to the LoD of the two assays.

A critical component of the ZIKV RNA assays is their ability to detect low levels of viral RNA at the initial and final stages of viremia or viruria. The CDC reported ZIKV LoD for the Trioplex rRT-PCR assay of 2.45 × 103 GCE/ml in serum and 4.64 × 103 GCE/ml in urine for either small- (0.2-ml input) or large-volume (1-ml input) extractions (http://www.cdc.gov/zika/pdfs/trioplex-real-time-rt-pcr-assay-instructions-for-use.pdf). Compared to the Trioplex rRT-PCR, we found the Aptima ZIKV assay to be >100-fold more sensitive with LoD of 11.5 GCE/ml and 17.9 GCE/ml in serum and urine, respectively. This indicates that the six Aptima ZIKV assay positive and rRT-PCR negative results may be true positives. Out of these six discrepant results, four were paired serum and urine samples from two patients, which also suggests that these patients had true ZIKV infections. In addition, convalescent-phase serum obtained from one of these patients tested positive for ZIKV IgM antibodies, thus confirming the original results of the Aptima ZIKV assay, indicating that the Aptima ZIKV assay can detect ZIKV infections that would otherwise be missed by rRT-PCR.

The enhanced analytical sensitivity of the Aptima ZIKV assay may be a result of the differences in the initial testing volume. The Aptima ZIKV assay requires 0.5 ml of serum or urine for testing, while the rRT-PCR reference method tests only a small fraction of the initial 0.2-ml sample volume. Increasing the initial sample volume for extraction, input RNA for amplification, or input cDNA for RT-PCR testing may increase the sensitivity of the rRT-PCR reference method. This has recently been demonstrated by Stone et al. (12), who showed that increasing the sample input volume can improve the sensitivities of ZIKV rRT-PCR assays by >10-fold.

The main limitation in this study was the lack of ZIKV-positive patient specimens. ZIKV viremia and viruria is transient and only reliably detected within 2 weeks of symptoms or exposure (5, 6). This narrow detection window and the low risk of ZIKV infection in our service area limited the availability of ZIKV-positive specimens. As an alternative, ZIKV-negative patient specimens were spiked with ZIKV stock to enhance the robustness of the Aptima ZIKV assay evaluation. Currently, there is a proposal to test whole blood versus serum, as the ZIKV seems to last longer in whole blood, which expands the window of detection significantly (13, 14). Thus, collecting whole blood from pregnant women with concerns about ZIKV infection could be considered in the future in order to increase testing sensitivity.

The results of this study demonstrate the utility of the Aptima ZIKV assay, but the automated testing platform demonstrates the practicality of performing this assay in a clinical setting. The Aptima ZIKA assay runs on the fully automated Panther system, which performs sample preparation, extraction, amplification, detection, and results analysis. Up to 120 samples can be loaded at a time, with the ability to continuously load samples on the instrument as samples are processed. The instrument can also be interfaced with the hospital laboratory information system (LIS) to automatically post results to patients' charts. These features allow clinical laboratories to perform high-volume testing and produce faster results with less hands-on time.

In conclusion, the Aptima ZIKV assay run on the Panther system is a rapid and high-throughput method that accurately and reliably detects ZIKV RNA in patient serum and urine specimens. The automated system and enhanced analytical sensitivity of the Aptima ZIKV assay are significant advantages over the rRT-PCR reference method.

MATERIALS AND METHODS

Viruses, bacteria, and fungi.

The Zika virus (Asian lineage, strain FSS13025) used for the limit of detection assay was obtained from Pei-Yong Shi's laboratory at UTMB. The Zika virus Mexico (Mex-81) and Africa (DAK) strains, and the dengue virus used for diagnostic accuracy assessment were obtained from Richard Pyle's laboratory at UTMB. Dengue, Kunjin, yellow fever, and chikungunya virus vaccine strains were provided by Scott Weaver at UTMB. All other bacterial and fungal isolates were collected from the UTMB clinical microbiology laboratory.

Clinical and spiked specimens.

All 105 clinical samples accepted for ZIKV RNA rRT-PCR were collected within 10 days of potential ZIKV exposure or symptoms and also met two thirds of the CDC's clinical and epidemiological testing criteria, which include symptoms, pregnancy, and exposure risk (see https://www.cdc.gov/zika/symptoms/index.html). Similar criteria were applied to serum samples (n = 20) collected from Faculdade de Medicina de São José do Rio Preto (FAMERP), São José do Rio Preto, Brazil. Spiked specimens were prepared by adding one of the Zika virus lineages (Mex-81, Cambodia strain [FSS13025], or DAK) or dengue virus into ZIKV-negative serum and urine. Users performing the Aptima ZIKV assay were blind to the identity of all specimens.

rRT-PCR.

The rRT-PCR reference method was performed by The Texas Department of State Health Services using the CDC Trioplex rRT-PCR method or by UTMB personnel according to the protocol previously described by Lanciotti et al. (5). At UTMB, viral RNA was extracted from serum and urine specimens using the QIAamp viral RNA mini kit (Qiagen, Hilden, Germany). rRT-PCR was performed on the CFX96 real-time PCR detection system (Bio-Rad, Hercules, CA) using cDNA prepared from extracted RNA with the iScript cDNA synthesis kit (Bio-Rad, Hercules, CA). Serum or urine spiked with ZIKV was prepared alongside patient samples and used as the positive control.

Aptima Zika virus assay.

Serum and urine samples were tested using the Aptima ZIKV assay (Hologic, Marlborough, MA) on the automated Panther system (Hologic, Marlborough, MA) according to the manufacturer's instructions. The Aptima ZIKV assay amplifies two highly conserved sequences in the NS2 and NS4/NS5 regions by transcription-mediated amplification (TMA). The entire assay is performed in a single tube and involves three main steps: sample preparation, ZIKV RNA target amplification by TMA, and detection of the amplification products (amplicons) by a hybridization protection assay (HPA). The chemiluminescence produced by hybridized probes is measured in relative light units and reported as positive or negative. Each test is equipped with an internal control (IC) to monitor nucleic acid capture, amplification, and detection. A field application specialist from Hologic provided training for the Aptima ZIKV assay on the Panther system at the UTMB laboratory.

Limits of detection.

ZIKV stock (Asian lineage, strain FSS13025) with 2.18 × 109 GCE/ml was diluted 1:100 in pooled ZIKV-negative serum and urine, and serial diluted to reach a final concentration of 1:100,000,000. Final urine dilutions were mixed with Aptima urine transport media at the ratio of 1:1 prior to testing. Multiple replicates of the lower dilutions were performed to reliably determine the limits of detection. The 95% detection probability of the GCE/ml was determined using probit analysis (see www.medcalc.org/calc/diagnostic_test.php).

Analytical specificity.

Interfering substances included residual serum and urine from patients testing positive for bacterial or viral nucleic acid; serum and urine spiked with 1 × 106 CFU/ml of commonly isolated bacteria and fungi; and serum and urine spiked with the closely related arboviruses dengue, Kunjin, yellow fever, and chikungunya (Table 4). Each sample was split into two aliquots. One aliquot was tested directly by the Aptima ZIKV assay whereas the second aliquot was spiked with ZIKV at 50 GCE/ml prior to testing.

TABLE 4.

Interfering substances for analytical specificity analysis

Nucleic acid Spiked organism
Chlamydia trachomatis Candida albicans
Neisseria gonorrhoeae Candida glabrata
BK virus Candida krusei
Cytomegalovirus Cryptococcus neoformans/C. gattii
Hepatitis B virus Enterococcus faecalis
Hepatitis C virus Escherichia coli
HIV Staphylococcus aureus
Streptococcus pneumoniae
Chikungunya virus
Dengue 2 virus
Kunjin virus
Yellow fever virus

Reproducibility.

A replicate panel (40 positive, 45 negative) was made by spiking ZIKV stock into pooled ZIKV-negative serum and urine. Replicate panel tests were performed a week apart with different operators on two separate Panther systems. A second replicate panel containing 20 serum (10 positive and 10 negative) and 20 urine (8 positive and 12 negative) samples was tested using the Aptima ZIKV assay at UTMB and in a second outside clinical laboratory.

Ethics statement.

This study was conducted under the research protocol “Evaluation/Verification of Assay and Instrument Systems Using Human Specimens” approved by the UTMB Institutional Review Board (protocol 08-182).

ACKNOWLEDGMENTS

We thank Richard Pyle, Pei-Yong Shi, and Scott Weaver at UTMB for kindly providing viral stocks; Kathleen Lawless from University Hospital at San Antonio, TX, for sharing Aptima ZIKA assay validation results; and Kimberly Bufton from UTMB for performing the blinded reproducibility results on the second Panther instrument.

M.L.N. is supported by FAPESP (grant 13/21719-3) and CNPq-FAPESP PPSUS (grant 16/15021-1). M.L.N. is a CNPq Research Fellow. A.C.B.T. is a FAPESP Fellow and is supported by FAPESP (grant 15/12295-0). Hologic provided Aptima ZIKV assay reagents. Hologic also provided travel support for P.R., D.A.O., and M.J.L.

REFERENCES

  • 1.Dick GW, Kitchen SF, Haddow AJ. 1952. Zika virus. I. Isolations and serological specificity. Trans R Soc Trop Med Hyg 46:509–520. doi: 10.1016/0035-9203(52)90042-4. [DOI] [PubMed] [Google Scholar]
  • 2.Petersen LR, Jamieson DJ, Powers AM, Honein MA. 2016. Zika virus. N Engl J Med 374:1552–1563. doi: 10.1056/NEJMra1602113. [DOI] [PubMed] [Google Scholar]
  • 3.Parra B, Lizarazo J, Jiménez-Arango JA, Zea-Vera AF, González-Manrique G, Vargas J, Angarita JA, Zuñiga G, Lopez-Gonzalez R, Beltran CL, Rizcala KH, Morales MT, Pacheco O, Ospina ML, Kumar A, Cornblath DR, Muñoz LS, Osorio L, Barreras P, Pardo CA. 2016. Guillain-Barré syndrome associated with Zika virus infection in Colombia. N Engl J Med 375:1513–1523. doi: 10.1056/NEJMoa1605564. [DOI] [PubMed] [Google Scholar]
  • 4.World Health Organization (WHO). 2017. Zika Situation Report, February 2, 2017. http://apps.who.int/iris/bitstream/10665/254507/1/zikasitrep2Feb17-eng.pdf?ua=1 Accessed 11 May 2017.
  • 5.Lanciotti RS, Kosoy OL, Laven JJ, Velez JO, Lambert AJ, Johnson AJ, Stanfield SM, Duffy MR. 2008. Generic and serologic properties of Zika virus associated with an epidemic, Yap State, Micronesia, 2007. Emerg Infect Dis 14:1232–1239. doi: 10.3201/eid1408.080287. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Bingham AM, Cone M, Mock V, Heberlein-Larson L, Stanek D, Blackmore C, Likos A. 2016. Comparison of test results for Zika virus RNA in urine, serum, and saliva specimens from persons with travel-associated Zika virus disease—Florida, 2016. MMWR Morb Mortal Wkly Rep 65:475–478. doi: 10.15585/MMWR.mm6518e2. [DOI] [PubMed] [Google Scholar]
  • 7.Food and Drug Administration. 2016. Emergency use authorizations. Trioplex real-time RT-PCR assay (CDC). https://www.fda.gov/MedicalDevices/Safety/EmergencySituations/ucm161496#zika Accessed 11 May 2017.
  • 8.Food and Drug Administration. 2016. Emergency use authorizations. Aptima Zika Virus Assay (Hologic, Inc). https://www.fda.gov/MedicalDevices/Safety/EmergencySituations/ucm161496#zika Accessed 11 May 2017.
  • 9.Hologic, Inc. 2016. Aptima Zika virus assay package insert. http://www.hologic.com/package-inserts/clinical-diagnostics/assays-and-tests/aptima-zika-virus-assay Accessed 11 May 2017.
  • 10.De Morais Bronzoni RV, Bakeittu FG, Ribeiro Nogueira RM, Nunes M, Moraes Figueiredo LT. 2005. Duplex reverse transcription-PCR followed by nested PCR assays for detection and identification of Brazilian alphaviruses and flaviviruses. J Clin Microbiol 43:696–702. doi: 10.1128/JCM.43.2.696-702.2005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Lanciotti RS, Kosoy OL, Laven JJ, Panella AJ, Velez JO, Lambert AJ, Campbell GL. 2007. Chikungunya virus in US travelers returning from India, 2006. Emerg Infect Dis 13:764–767. doi: 10.3201/eid1305.070015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Stone M, Lanteri MC, Bakkour S, Deng X, Galel SA, Linnen JM, Muñoz-Jordán JL, Lanciotti RS, Rios M, Gallian P, Musso D, Levi JE, Sabino EC, Coffey LL, Busch MP. 2017. Relative analytical sensitivity of donor nucleic acid amplification technology screening and diagnostic real-time polymerase chain reaction assays for detection of Zika virus RNA. Transfusion 57:734–747. doi: 10.1111/trf.14031. [DOI] [PubMed] [Google Scholar]
  • 13.Lustig Y, Mendelson E, Paran N, Melamed S, Schwartz E. 2016. Detection of Zika virus RNA in whole blood of imported Zika virus disease cases up to 2 months after symptom onset, Israel, December 2015 to April 2016. Euro Surveill 21(26). http://www.eurosurveillance.org/ViewArticle.aspx?ArticleId=22515. [DOI] [PubMed] [Google Scholar]
  • 14.Murray KO, Gorchakov R, Carlson AR, Berry R, Lai L, Natrajan M, Garcia MN, Correa A, Patel SM, Aagaard K, Mulligan MJ. 2017. Prolonged detection of Zika virus in vaginal secretions and whole blood. Emerg Infect Dis 23:99–101. doi: 10.3201/eid2301.161394. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES