Abstract
Purpose
To assess the uptake, accumulation, temporal stability, and spatial localization of isoflurane (ISO) in the C57BL/6 mouse, and to identify its potential interference with the detection of labeled cardiac progenitor cells using 19F MRI/MR spectroscopy (MRS).
Materials and Methods
Objectives are demonstrated using (a) in vitro ISO tests, (b) in vivo temporal accumulation/spatial localization C57BL/6 studies (n = 3), and (c) through injections of perfluoro‐crown‐ether (PFCE) labeled cardiac progenitor cells into femoral muscle areas of the murine hindlimb post‐mortem (n = 1) using 1H/19F MRI/MRS at 9.4 Tesla. Data were acquired using double‐gated spoiled gradient echo images and pulse‐acquire spectra. For the in vivo study, the temporal stability of ISO resonances was quantified using coefficient of variability (CV) (5 min) estimates.
Results
Two ISO resonances were observed in vivo that correspond to the ‐CF3 and ‐OCHF2 moieties. CV values ranged between 3.2 and 6.4% (‐CF3) and 6.4 and 11.2% (‐OCHF2). Reductions of the ISO dose (2.0 to 1.7%) at 80 min postinduction had insignificant effects on ISO signals (P = 0.23; P = 0.71). PFCE‐labeled cells exhibited a resonance at −16.25 ppm in vitro that did not overlap with the ISO resonances, a finding that is confirmed with MRS post‐mortem using injected, labeled cells. Based on 19F MRI, similar in vivo/post‐mortem ISO compartmentalization was also confirmed in peripheral and thoracic skeletal muscles.
Conclusion
Significant ISO accumulation was observed by 19F MRS in vivo with temporally stable signals over 90 min postinduction. ISO effects on PFCE labels are anticipated to be minimal but may be more prominent for perfluoropolyether or perfluorooctyl bromide labels.
Level of Evidence: 1
Technical Efficacy: Stage 1
J. MAGN. RESON. IMAGING 2017;45:1659–1667
Keywords: cell labeling, 19F MRI, isoflurane localization, contamination
Noninvasive imaging and tracking of labeled cells in vivo and their functional impact has taken an increasing role in recent years.1, 2, 3 Specifically, nanoparticles (NPs) containing perfluorocarbons (PFC) have enabled direct tracking and quantification of labeled cell populations using 19F MR spectroscopy (MRS) and MRI.4, 5 Correspondingly, labeling and in vivo MRI tracking of cells has been applied in small animal models of neurovascular disease,4, 6, 7, 8 exhibiting potential for translational work.9
Most preclinical studies are now routinely conducted using inhalational isoflurane (ISO),10, 11, 12, 13 the most commonly used volatile anesthetic owing to its minimal cardio‐depressive effects.12 Isoflurane (CF3CH2ClOCHF2) contains two groups (moieties) of 19F atoms, which give rise to two peaks (with chemical shifts of ‐4 and ‐10.3 ppm with respect to CFCl3) in the NMR spectrum. These peaks may interfere with the detection of fluorinated agents in vivo (chemical shift range of ‐126.6–9.1 ppm with respect to CFCl3 and ‐50–86 ppm with respect to trifluoroacetic acid [TFA]) due to potential spectral overlaps. Furthermore, despite the existence of numerous prior 19F MRI studies on the uptake and clearance characteristics of ISO in the brain in vivo,14, 15, 16, 17 the localization of ISO in cardiac thoracic and peripheral skeletal tissues, its bio‐distribution, accumulation, kinetics, and possible interference with 19F labels has not been investigated in detail. Additionally, only a few reported cardiovascular applications of fluorinated agents exist,18, 19 and investigations into the possible implications of the use of ISO have been speculative.20
Specifically, Flögel et al.18 and van Heeswijk et al.19 have, respectively, demonstrated successful cardiac and vascular 19F MR imaging in the mouse thorax following injection of PFC emulsions into the venous circulation. While Flögel et al.18 do not discuss the impact of the presence of ISO on the imaging of 19F emulsions, van Heeswijk et al.19 have shown minimal ISO accumulation of the anesthetic in the thoracic vascular space, and have quantified the relative spectral shifts from the injected agents. More recently, Fox et al.20 have argued that the use of ISO should have a minimal impact on the 19F MR images of PFC agents, and have suggested the use of selective radiofrequency (RF) excitation to ameliorate any potential issues; however, their discussion is speculative and lacks direct experimental evidence.
The scope of this study is generalized on the accumulation of ISO (in striated skeletal and cardiac muscles). It extends beyond the two prominent prior 19F MRS/MRI publications that focused on cardiovascular applications,18, 19 whereby direct injections of labeled emulsions were achieved by means of the intravenous route. Their reported approach18, 19 exhibits bio‐distribution, delivery, temporal dynamics, and label dilution differences, compared with current state‐of‐the‐art schemes using either direct injections (adopted strategy in this study), or scaffold‐based, 19F labeled, cardiac progenitor stem cell administrations.
Therefore, the objectives of this study are to: (a) determine the spatial localization of ISO in the thoracic and peripheral skeletal muscle areas of the C57BL/6 mouse, and quantitatively assess its uptake, accumulation, and temporal stability within a period of 90 min postinduction, including its potential accumulation in the ventricular blood pool, and myocardium, (b) define the spectral characteristics of the ISO moieties in vivo, (c) identify the spectral characteristics of extensively used 19F labels, and assess their potential spectral overlaps and interference with respect to ISO, including the extensively used perfluoro‐crown‐ether (PFCE) label, using 19F MRS/MRI.
Materials and Methods
Animal Ethics
All procedures were approved by the local ethical review committee (U. Oxford), and conformed to the Animal (Scientific Procedures) Act 1986 (UK) incorporating European Directive 2010/63/EU.
MRI
All experiments were conducted on a 9.4Tesla (T) Agilent scanner equipped with a DirectDrive console and a 1 mT/m actively shielded gradient set (internal diameter = 60 mm) (Agilent Technologies, Santa Clara, CA).
Radiofrequency Coils
An eight‐rung, low‐pass quadrature birdcage coil (internal diameter = 34 mm), was constructed in‐house, tuned and matched at 375.8 MHz, and used for 1H/19F MRS/MRI. The broad frequency response of the coil permitted imaging on both the 1H and 19F nuclei. Post‐mortem imaging was conducted using a home‐built 80 × 40 mm2 butterfly coil.
Pulse Sequences
An autoshim method was used for in vivo mouse shimming on the 1H resonance, as described previously.21 In vivo 1H and 19F MR images were acquired in mice (thoracic area) using a segmented k‐space, prospectively electrocardiograph‐ (ECG‐) and respiratory‐gated (double‐gated) spoiled gradient echo (SPGR) sequence. A three‐dimensional (3D) SPGR sequence (ungated) was used for the post‐mortem imaging (hind‐limb skeletal muscle areas).
For MRS, the temporal evolution of ISO accumulation was monitored using nonlocalized, fully relaxed, pulse‐acquire spectra.
Phantom Studies
a) Aqueous sodium fluoride (NaF) phantoms at 50 mM [Sigma‐Aldrich, UK] (T1 = 1564 ms; T2 = 1257 ms) were used as reference markers in ISO studies for calibration and quantification. b) Pulse‐acquire spectra were collected (repetition time [TR] = 10 s; flip angle = 90 °; bandwidth [BW] = 10–20 kHz; number of excitations [NEX] = 1–10; 512–8192 points) using phantoms containing: (1) liquid ISO (ISO Flo, 100% w/w, Abbott Laboratories, UK), (2) PFC NPs (3.8–10 mg/mL) emulsified in deionized water and mixed in cell media solution (IMDM, Thermo Fisher Scientific, UK), and (3) 25 mM TFA, separately or in combination with NPs and PFCE‐labeled cells (TR = 20 s; flip angle = 90 °; NEX = 4; 512 points; BW = 20 kHz).
Cells
Nanoparticle Formulation
Particles were synthesized in accordance to previously published methodologies.22, 23 In brief, nanoparticles (∼200 nm in diameter) with entrapped PFCEs were synthesized using poly(lactic‐co‐glycolic acid) (PLGA). All particles were washed extensively with distillated water and lyophilized for 3–4 days. Prior 19F MRS characterization of PFCE NPs has confirmed a single spectral peak at ‐16.00 ppm (with respect to TFA).
Cell Isolation
Cardiac progenitor cells comprising either collagenase‐trypsin (CT) or cardiosphere‐derived cells (CDC)] were isolated from adult, C57BL/6, GFP positive, mouse atria, using standard protocols.24, 25
Labeling
Cells were subsequently incubated with PFCE‐containing NPs (loading concentration of 10 mg/ml/million cells) for approximately 24 h before trypsinization, isolation, and washing. Cell pellets were suspended and maintained in IMDM media and were used for MRI, flow cytometry, or confocal microscopy (after fixation in 2% paraformaldehyde solution). Successful labeling was confirmed with a CyAn ADP flow cytometer (Beckman Coulter, Brea, CA) using control and labeled cell samples.
In Vivo and Post‐mortem Murine MRI/MRS
Physiology
To ascertain the temporal effects of anesthesia, ISO was administered in three C57BL/6 mice (male [mean ± SD], 24.6 ± 3.6 g, weight range = 25.0–27.2 g). The mice were induced using 4% ISO, and were maintained for up to 95 min with 1.5–2.0% ISO, mixed in 100% oxygen.
All animals were placed in a specially constructed cradle (www.civm.duhs.duke.edu), and were allowed to breathe freely throughout the study. A homeostatically controlled hot‐air system was used to maintain mouse body temperature at approximately 37 °C. ECG and breathing rates were monitored using a gating system.26 Heart rates were maintained at 382–565 beats/min.
One additional C57BL/6 mouse (male, weight = 32.4 g) was euthanized with an overdose of ISO (5%) and was subsequently injected in the femoral area of the right hindlimb with labeled CDC cells, transferred to a butterfly coil, and then re‐imaged post‐mortem.
1H Imaging
Axial cardiac 1H images (TR/TE = 2.98/1.49 ms; echo = 50%; flip angle = 15 °; NEX = 8; field of view [FOV] = 40 × 40 mm2; 10 slices; slice thickness (ST) = 1 mm; matrix = 128 × 128; acquisition time = 3.25 min; eight phase encoding steps per segment) were acquired with a 2D segmented k‐space pulse sequence. Correspondingly, post‐mortem skeletal muscle images (TR/TE = 2.33/1.18 ms; flip angle = 20 °; NEX = 1; 40 × 40 × 40 mm3; ST = 40 mm; and matrix = 128 × 128 × 128) were acquired with a 3D SPGR sequence.
19F MRI/MRS
Double‐gated axial 19F images of the mouse thorax were acquired (TR/TE = 9‐17/4.6–8.4 ms; echo = 50%; flip angle = 50 °; NEX = 128–512; FOV = 40 × 40 mm2; 1 slice; ST = 10 mm; matrix = 32 × 32; BW = 2–4 kHz; acquisition time = 2.42 min; and 8 phase encoding steps per segment). Spectral selection of the ISO and NP label resonant peaks was achieved using broadband excitations and narrowband acquisitions centered on the resonant frequency of interest. For skeletal muscle imaging, the acquisition parameters were: TR/TE = 17/8.4 ms; flip angle = 50 °; NEX = 512; FOV = 40 × 40 mm2; 1 slice (coronal and axial views); ST = 40 mm; matrix = 32 × 32; BW = 2 kHz; and acquisition time = 4.5 min.
NP‐labeled cells (skeletal muscle areas) and the temporal evolution of ISO accumulation (30 successive spectra acquired and averaged over 5‐min intervals in thoracic areas) were monitored using nonlocalized MR spectra (TR = 20 s; flip angle = 90 °; NEX = 2–8; 512 points; BW = 10 kHz) within a temporal window of 35–90 min postinduction.
Image Processing
Image and Spectral Analyses
19F images were imported and interpolated in ImageJ (NIH, Bethesda, MD) using bicubic splines to match the 1H matrix size. Thoracic 1H and 19F MRI were overlaid in ImageJ (opacity = 70%). For skeletal MRI, the outer hind limb boundaries were segmented in ImageJ by an expert user, and the segmented outline was overlaid on the 19F images. Spectra were processed in CSX (P. Barker, Johns Hopkins, USA) or using custom‐written IDL tools (Harris Geospatial, USA). For temporal ISO comparisons, peak integrals were estimated from gated, magnitude spectra.
Statistical Analyses
All results are reported as mean ± SD. The temporal constancy of the spectral areas for both ISO resonances is quantified using the coefficient of variability (CV = 100 × SD/Mean). Repeated‐measures, nonparametric, two‐tailed Student t‐tests, were also used (XLSTAT, Addinsoft, New York) to determine whether the ISO dose reduction (2.0% to 1.7%) caused significantly different changes in the areas of the spectral peaks (α = 5%).
Results
Typical nonlocalized pulse‐acquire 19F spectra of: (a) an ISO phantom, (b) a 75 mM TFA reference standard, (c) a 25 mM TFA phantom in the presence of an NP‐labeled cell phantom, and (d) ISO accumulation in thoracic areas in the in vivo mouse, are, respectively, shown in Figure 1A–D. For graphical presentations, all spectra are referenced to TFA at 0 ppm. Estimated in vitro linewidths are 92 Hz (ISO) and 210 Hz (75 mM TFA). More importantly, Figure 1D demonstrates that NP‐labeled CT cells exhibit a PFCE resonance at ‐16.00 ppm that does not overlap with the ISO resonances. Two distinct ISO resonances are observed in vivo (Fig. 1D) that, respectively, correspond to the ‐CF3, and ‐OCHF2 moieties.14, 17, 19 Noted also is a downfield spectral shift of the ISO resonances in vivo by approximately ‐4 ppm, and an increase in the spectral linewidth of the ‐CF3 peak to 500 Hz. The spectral separation of the two ISO resonances in vivo is ∼6.3 ppm (Table 1). Taken together, Figure 1C, D indicates that PFCE‐labeled CT cells exhibit spectral separations of ‐11.25 and ‐4.95 ppm from the ‐CF3 and ‐OCHF2 resonances of ISO, respectively.
Table 1.
19F Label | Cell Type | Chemical Shift with respect to CFCl3 [with respect to TFA] (ppm) |
---|---|---|
Poly(lactic‐co‐glycolic acid) perfluoro‐crown‐ether (PLGA‐PFCE) |
Human dendritic cells (DC)/macrophages/monocytes/endothelial cells Cardiac progenitor cells (CDC, CT) |
−91.8 [∼ −14.75–−16.00] (This study) |
Perfluoropolyether (PFPE) | Primary T‐cells/neural stem cells/primary human DCs/ |
−58 [18.55] −90.7 [−14.55] −93.2 [−16.65] |
Perfluorooctylbromide (PFOB) | Stem/progenitor cells/smooth muscle cells/macrophages |
−63.7 [26.7] −81.82 [−5.27] −117.5 [−40.95] −122 [−45.45] −126.6 [−50.05] |
Perfluorodecalin (PFD) | Monocytes‐macrophages |
−27.4 [49.15] −32.0 [44.55] −38.7 [37.85] −49.6 [26.95] −98.2 [−21.65] |
Trans‐bis‐perfluorobutyl ethylene (F‐44E) | Inflammatory cells – immune cells |
9.1 [85.65] −23.7 [52.85] −33.4 [43.15] −35.0 [41.55] |
ISO (ppm) | ||
‐CF3 moiety (in vivo) | ‐ | −80.55 [−4] |
‐OCHF2 moiety (in vivo) | ‐ | −86.85 [−10.3] |
TFA | ||
Aqueous solution | ‐ | −76.55 [0] |
NaF | ||
Aqueous solution | ‐ | −121.5 [−44.95] |
KF | ||
Aqueous solution | ‐ | −125.3 [−46.75] |
Also listed are the spectral shifts of the ISO moieties and of the aqueous reference standards TFA, NaF, and potassium fluoride (KF). Cardiac progenitor cells (CDC or CT) used in this study exhibit respective downfield shifts in the ranges of −10 to −11 and −4 to −4.95 ppm with respect to the two ISO moieties.
Figure 2 shows the in vivo temporal evolution of the nonlocalized, pulse‐acquire ISO signals (corresponding to the ‐CF3 and ‐OCHF2 moieties) at 5 min intervals from a typical mouse. To assess any changes in the in vivo 19F spectrum of ISO with time or dose, the areas of both ISO peaks were quantified using CSX, and the ISO dosage was decreased from 2.0% to 1.7% at t = 80 min postinduction (see Fig. 3).
The CV of the ISO peaks for each of the studied mice and for the entire cohort (during the time period of 50–90 min postinduction) were: (a) 6.4, 3.9, 3.2, and 3.9% (‐CF3), and (b) 11.2, 6, 8.5, and 6.4% (‐OCHF2).
Furthermore, no statistically significant changes in mean spectral areas of the ISO peaks (P = 0.23, P = 0.71) were observed pre‐ and postreduction of the ISO dose (t = 60–80 min versus t = 80–90 min).
In vivo localization of ISO was confirmed using 19F MRI, as shown in Figure 3. Axial 19F MR images (ST = 10 mm) of the two ISO moieties acquired using narrowband acquisitions are shown along with corresponding 1H images (only 8 of the 10 acquired slices, ST = 1 mm). To better demonstrate the anatomical localization of the ISO signal, merged 1H/19F images are also shown (Fig. 3C). For merging, a 1H image spatially located within the excited 19F imaging slab was selected. As it can be seen from the images, the 19F signal from the ISO peaks localizes in the skeletal muscle and possibly subcutaneous fat of the thorax, whereby both ISO resonances exhibit similar localization. The lack of 19F signals from the ventricular blood or myocardium is also noted.
Figure 4 shows post‐mortem nonlocalized pulse‐acquire 19F spectra acquired immediately before and after injection of CDC cells labeled with PFC NPs. The spectra were acquired ∼7 h post‐mortem in a mouse that was injected in the femoral area of the right leg with ∼2.5 million CDC cells labeled with PFCE NPs. Indicated (arrow) is the resonance corresponding to the PFCE label, which does not overlap with the ISO resonances. Shown also are coronal 19F MR images corresponding to each of the two ISO resonances, and the corresponding 1H images. An axial 19F MR image corresponding to the 19F label is also shown where no observable contribution from ISO is observed.
Discussion
This study demonstrates ISO accumulation in thoracic areas of the mouse. The results also demonstrate that the presence of ISO does not constitute a problem for the detection and quantification of PFCE in cardiac labeled cells.20 Herein, the two expected peaks of the NMR spectrum of ISO were observed both in vitro and in vivo.
ISO exhibited chemical shifts of ‐4.0 (CF3) and ‐10.3 ppm (‐OCHF2) in C57BL/6 mice in vivo with respect to the TFA resonance at 0 ppm, confirming previous reports in rabbits, and C57BL/6 and BALB/c mice.14, 19, 27 The downfield shift of ISO in vivo has been attributed to the chemical structure and the microenvironment in which ISO resides.14 It is also of interest to note that the chemical shift of the ‐CF3 resonance is either co‐resonant with (in vitro) or close to (in vivo) the TFA resonance. Consequently, TFA reference phantoms cannot be used in ISO/NP label studies because the signal overlap can lead to inaccuracies in calibration and quantification. Instead, NaF (‐44.95 ppm) or potassium fluoride (‐46.75 ppm) phantoms may be preferred because their chemical shifts do not overlap with those of the ISO moieties.
Substantial accumulation of ISO was observed in the live mouse as documented by 19F MRS. Prior work in vivo reports that the uptake and equilibration of ISO and sevoflurane in the brain is extremely fast (of the order of minutes),15, 17, 28 compared with the ISO release which appears to be extremely slow (of the order of 72–98 h).14, 15, 28
The results presented in this study show constancy and temporal stability of the evoked ISO spectra from anesthetized mice within time periods spanning 35–90 min postinduction. These findings are reinforced by the lack of statistically significant differences in the ISO signal amplitude following dose decreases. Additionally, the post‐mortem study confirmed persistence of ISO in thoracic and skeletal muscle areas for at least 7 h post‐euthanasia, consistent with prior literature.14, 15, 16 Thus, longitudinal studies with repeated imaging sessions may be affected by ISO accumulation. This issue has been overcome in prior work with the use of alternative anesthetics.20, 29
Noted also were significant increases in the linewidths of both ISO peaks attributed to either temporal, or physiologically dependent factors. From the physiological standpoint, the acquired spectra represent nonlocalized, spatiotemporal averaged signals that may be prone to: (a) beat‐to‐beat variations, (b) overall effects from ISO distribution (as a result of intraspecies variations in weight, body muscle/fat composition), and (c) heterogeneity of signal contribution arising from multiple organs and multiple T2 entities. From the technical standpoint, primary effects may be attributed to: (a) shimming, (b) B0/B1 effects (leading to flip‐angle dependencies and off‐resonance effects), and (c) the inability to resolve multiplet patterns from the respective spectral splitting of the two ISO moieties, ultimately leading to two distinct, broadened, peaks. Certainly, the in vivo ISO peak intensities are substantially larger than that of the PFCE spectral intensity arising from the labeled cells.
From the biophysical standpoint, it is expected that the observed linewidth changes are attributed primarily to shortened T2 or values, or to poor B0 homogeneity over the entire thorax or skeletal muscle areas. The adopted MRS approach and inferences drawn are certainly based on nonlocalized spectra, spanning the entire FOV of the RF coil. Correspondingly, T2 reductions are expected to be most likely an end‐result of ISO binding/complexation. The broad (nonlocalized) spectral peaks observed in vivo (compared with linewidths of 50–70 Hz reported by Wyrwicz et al.14 in rabbit brains) indicate the existence of either suboptimal shimming, or ultrashort T2 moieties and/or a heterogeneous, multifocal pattern of ISO accumulation in lipid, blood, and other tissue areas. Our results are in agreement with Xu et al.15 who have presented evidence for ultrashort T2 anesthetic moieties (<3 ms) in the brain in rats as a result of intracranial skeletal muscle/fat accumulations (14), despite the lack of significant lipid accumulations/contributions as reported in a recent study by van Heeswijk et al.19
Given the chemical shift difference between the ISO resonance of the ‐OCHF2 moiety with respect to the PFCE resonance of cardiac progenitor‐labeled cells, and in conjunction with the selective acquisition bandwidths used in 19F MRI, we have shown negligible ISO effects on the imaging of 19F signals arising from PFCE labels in vitro, and in post‐mortem skeletal muscle imaging in the mouse. Combined selective excitation/suppression schemes can be used to excite the PFC label and minimize baseline drifts due to possible contaminating effects from ‐OCHF2. This is particularly important given the multifold 19F signal difference of labels in cells compared with the ISO moieties, particularly considering the expected submillimolar label concentrations in cells.29, 30, 31 The downside of selective excitation and prepulse suppression is the prolongation of TE. Nevertheless, in the case of the PFCE NPs, such effects are anticipated to be minimal, given their long T2 responses in the range of 50–600 ms.7
However, with the use of other commonly used 19F labels, such as perfluoropolyether (PFPE), perfluorodecalin (PFD), trans‐bis‐perfluorobutyl ethylene (F‐44E), or perfluorooctyl bromide (PFOB),23, 32 there is a significant potential for overlap of these labels and the ISO peaks, thereby confounding imaging and quantification. Despite the availability of other anesthetics and the possibility to ameliorate potential interference issues with alternative injectable anesthetic regimens,8 achievement of a stable dose level over the prolonged periods of MRI/MRS (injections and top‐ups) is challenging.
Despite the technical complexity of this work, there are several limitations in the study design. For example, despite the consistency of the effects of ISO observed in the male mice of this study, and in preliminary studies in female mice (results not shown), the sample sizes were small. Therefore, possible sex differences in the ISO effects need to be accounted for in large‐sample studies. Furthermore, the reported effects of temporal accumulation of ISO and its biodistribution, as well as the spectral response of the PFCE NPs must be assessed in models of pathology, including myocardial infarction and skeletal myopathies. Of more importance may be also the generalizability of the study and its extension to larger animal models (e.g., rats or pigs) in larger bore, lower‐field magnets. However, migration of the work may impose further challenges in terms of the separability of the ISO and PFCE NP spectra. Thus, a more appropriate study design may also include the characterization of the T1 relaxation values for both anesthetic moieties, and the selective inversion of the undesired spectral peaks in conjunction with the selective excitation of the peaks corresponding to the labels.
In conclusion, accumulation of ISO was observed in the mouse with signals that are temporally stable over 90 min post‐induction, as documented by 19F MRS. The temporal persistence of ISO spans multiple hours. Given the spectral separation of the ISO peaks from the PFCE resonance, and the selective acquisition bandwidths often used in 19F MRI, the anticipated signal overlap/contamination effects of ISO on PFCE labels are minimal, but may be more prominent for PFPE or PFOB labels. Correspondingly, ISO still remains the primary choice for 1H and multinuclear imaging studies.
Acknowledgments
The project leading to this study has received funding (C.C.) from the European Union's Horizon 2020 research and innovation program/ under the Marie Sklodowska‐Curie grant agreement, BHF grant, and European Research Council Grant (M.S.). J.E.S. acknowledges support from the BHF Centre of Regenerative Medicine, Oxford, and the BHF Centre of Research Excellence, Oxford. We are grateful to Professor C. Lygate for his constructive comments on the manuscript. We also thank Ms. A. Vernet for her support with the animal experiments and Mrs. S. Malandraki‐Miller for her original support with progenitor stem cell cultures. Conflicts of Interest: There are no conflicts of interest. Disclosures: There are no disclosures.
References
- 1. Ahrens ET, Flores R, Xu H, Morel PA. In vivo imaging platform for tracking immunotherapeutic cells. Nat Biotechnol 2005;23:983–987. [DOI] [PubMed] [Google Scholar]
- 2. Ruiz‐Cabello J, Barnett BP, Bottomley PA, Bulte JWM. Fluorine (19F) MRS and MRI in biomedicine. NMR Biomed 2011;24:114–129. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Jacoby C, Temme S, Mayenfels F, et al. Probing different perfluorocarbons for in vivo inflammation imaging by 19F MRI: image reconstruction, biological half‐lives and sensitivity. NMR Biomed 2014;27:261–271. [DOI] [PubMed] [Google Scholar]
- 4. Srinivas M, Herschap A, Ahrens ET Fiqdoe CG, de Vries IJ. 19F MRI for quantitative in vivo cell tracking. Trends Biotechnol 2010;28:363–370. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Gomes RSM, das Neves RP, Cochlin L, et al. Efficient pro‐survival/angiogenic miRNA delivery by an MRI‐detectable nanomaterial. ACS Nano 2013;7:3362–3372. [DOI] [PubMed] [Google Scholar]
- 6. Ahrens ET, Zhong J. In vivo MRI cell tracking using perfluorocarbon probes and fluorine‐19 detection. NMR Biomed 2013;26:860–871. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Srinivas M, Boehm‐Sturm P, Figdor CG, de Vries IJ, Hoehn M. Labeling cells for in vivo tracking using 19F MRI. Biomaterials 2012;33:8830–8840. [DOI] [PubMed] [Google Scholar]
- 8. Srinivas M, Boehm‐Sturm P, Aswendt M, et al. In vivo 19F MRI for cell tracking. J Vis Exp 2013;81:e50802. doi: 10.3791/50802. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Ahrens ET, Helfer BM, O'Hanlon CF, Schirda C. Clinical cell therapy imaging using a perfluorocarbon tracer and fluorine‐19 MRI. Magn Reson Med 2014;72:1696–1701. doi: 10.1002/mrm.25454. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Kober F, Iltis I, Cozzone PJ, Bernard M. Cine‐MRI assessment of cardiac function in mice anesthetized with ketamine/xylazine and isoflurane. MAGMA 2004;17:157–161. [DOI] [PubMed] [Google Scholar]
- 11. Hildebrandt I, Su H, Weber WA. Anesthesia and other considerations for in vivo imaging of small animals. ILAR J 2008;49:17–26. [DOI] [PubMed] [Google Scholar]
- 12. Constantinides C, Mean R, Janssen BJ. Effects of isoflurane anesthesia on the cardiovascular function of the C57BL/6 mouse. ILAR J 2011;52:e21–e31. [PMC free article] [PubMed] [Google Scholar]
- 13. Constantinides C, Murphy K. Molecular and integrative physiological effects of isoflurane anesthesia: the paradigm of cardiovascular studies in rodents using magnetic resonance imaging. Front Cardiovasc Med 2016;3:23. doi: 10.3389/fcvm.2016.00023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Wyrwicz AM, Pszenny MH, Tillman PC, Gordon RE, Martin PA. Noninvasive observations of fluorinated anesthesics in rabbit brain by fluorine‐19 nuclear magnetic resonance. Science 1983;222:428–430. [DOI] [PubMed] [Google Scholar]
- 15. Xu Y, Tang P, Zhang W, Firestone L, Winter PM. Fluorine‐19 nuclear magnetic resonance imaging and spectroscopy of sevoflurane uptake, distribution, and elimination in rat brain. Anesthesiology 1995;83:766–774. [DOI] [PubMed] [Google Scholar]
- 16. Venkatasubramanian PN, Shen YJ, Wyrwicz AM. Characterization of the cerebral distribution of general anesthetics in vivo by two‐dimensional 19F chemical shift imaging. Magn Reson Med 1996;35:625–630. [DOI] [PubMed] [Google Scholar]
- 17. Terekhov M, Scholz A, Wolf U, Rivoire J, Schreiber W. Fluorinated anesthetics uptake kinetic investigation on large animal model using 19F MRS/MRI. In: Proceedings of the 16th Annual Meeting of ISMRM, Toronto, Canada, 2008. (abstract 1736).
- 18. Flögel U, Ding Z, Hardung H, et al. In vivo monitoring of inflammation after cardiac and cerebral ischemia by fluorine magnetic resonance imaging. Circulation 2008;118:140–148. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. van Heeswijk RB, Pilloud Y, Flögel U, Schwitter J, Stuber M. Fluorine‐19 magnetic resonance angiography of the mouse. PLoS One 2012;7:e432236. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Fox MS, Gaudet JM, Foster PJ. Fluorine‐19 MRI contrast agents for cell tracking and lung imaging. Magn Reson Insights 2015;8(Suppl 1):53–67. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Schneider J, Jezzard P. Development of a rapid, automated shim approach for cardiac MR in mice and in vivo. In: Proceedings of the 14th Annual Meeting of ISMRM, Seattle, Washington, 2006. (abstract 1177).
- 22. Cruz LJ, Tacken PJ, Fokkink R, et al. Targeted PLGA nano‐ but not microparticles specifically deliver antigen to human dendritic cells via DC‐SIGN in vitro. J Control Release 2010;1;144:118–126. doi: 10.1016/j.jconrel.2010.02.013. [DOI] [PubMed] [Google Scholar]
- 23. Srinivas M, Cruz LJ, Bonetto F, Heerschap A, Figdor CG, de Vries IJM. Customizable, multi‐functional fluorocarbon nanoparticles for quantitative in vivo imaging using 19F MRI and optical imaging. Biomaterials 2010;31:7070–7077. [DOI] [PubMed] [Google Scholar]
- 24. Malandraki‐Miller S, Tyser R, Riley PR, Carr CA. Comparative characterization of cardiac atrial progenitor cell populations for use in cell therapy. Heart 2014;100:(Suppl 4) A14. doi:10.1136/heartjnl-2014-306916.41. [Google Scholar]
- 25. Tan SC, Gomes RS, Yeoh KK, et al. Preconditioning of cardiosphere‐derived cells with hypoxia or prolyl‐4‐hydroxylase inhibitors increases stemness and decreases reliance on oxidative metabolism. Cell Transplant 2016;25:35–53. doi: 10.3727/096368915X687697. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Cassidy PJ, Schneider JE, Grieve SM, Lygate C, Neubauer S, Clarke K. Assessment of motion gating strategies for mouse magnetic resonance at high magnetic fields. J Magn Reson Imaging 2004;19:229–237. [DOI] [PubMed] [Google Scholar]
- 27. Cortes‐Remirez M, Rizzorno G, Veni G, Gach HM. Interference of fluorinated anesthesia on 19F MRS of fluorinated drug metabolism in liver. In: Proceedings of the 17th Annual Meeting of ISMRM, Honolulu, Hawaii, 2009. (abstract 2361).
- 28. Litt L, Lockhart S, Cohen Y, et al. In vivo 19F nuclear magnetic resonance brain studies of halothane, isoflurane, and desflurane. Ann N Y Acad Sci 1991;625:707–724. [DOI] [PubMed] [Google Scholar]
- 29. Srinivas M, Turner MS, Janjic JM, Morel PA, Laidlaw DH, Ahrens ET. In vivo cytometry of antigen‐specific T cells using 19F MRI. Magn Reson Med 2009;62:747–753. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Srinivas M, Morel PA, Ernst LA, Laidlaw DH, Ahrens ET. Fluorine‐19 MRI for visualization and quantification of cell migration in a diabetes model. Magn Reson Med 2007;58:725–734. [DOI] [PubMed] [Google Scholar]
- 31. Chen J, Lanza GM, Wickline SA. Quantitative magnetic resonance fluorine imaging: today and tomorrow. Wiley Interdiscip Rev Nanomed Nanobiotechnol 2010;2:431–440. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Morawski AM, Winter PM, Yu C, et al. Quantitative “Magnetic resonance immunohistochemistry” with ligand‐targeted 19F nanoparticles. Magn Reson Med 2004;52:1255–1262. [DOI] [PubMed] [Google Scholar]