. 2017 May 24;17(6):1204. doi: 10.3390/s17061204

A New Calibration Method for Commercial RGB-D Sensors

Walid Darwish 1, Shenjun Tang 1,2,3, Wenbin Li 1, Wu Chen 1,*
Editors: Zhizhong Kang, Jonathan Li, Cheng Wang
PMCID: PMC5492766  PMID: 28538695

Abstract

Commercial RGB-D sensors such as Kinect and Structure Sensors have been widely used in the game industry, where geometric fidelity is not of utmost importance. For applications in which high quality 3D is required, i.e., 3D building models of centimeter-level accuracy, accurate and reliable calibrations of these sensors are required. This paper presents a new model for calibrating the depth measurements of RGB-D sensors based on the structured light concept. Additionally, a new automatic method is proposed for the calibration of all RGB-D parameters, including internal calibration parameters for all cameras, the baseline between the infrared and RGB cameras, and the depth error model. When compared with traditional calibration methods, this new model shows a significant improvement in depth precision for both near and far ranges.

Keywords: RGB-D sensors, depth calibration, structure sensor, structured light concept

1. Introduction

In recent years, RGB-D sensors have attracted great attention in 3D modeling due to their low cost. Two major concepts, time of flight (ToF) and structured light (SL), are used in RGB-D sensors. Many devices released in the market are based on these concepts; for example, Kinect v1, Structure Sensor [1] and ASUS Xtion Pro Live [2] are based on the SL concept, while Kinect v2 [3] is based on the ToF concept [4] The RGB-D sensor consists of three different sensors: an RGB camera, an infrared (IR) camera and an IR projector. In Kinect and ASUS, all three sensors are manufactured in a fixed frame, whereas the Structure Sensor combines two IR sensors only and is designed to be attached to a portable device with an RGB camera.

Although RGB-D sensors were originally designed to be used for gaming purposes such as remote controlling, they have recently made an important contribution to surveying and navigation applications, such as building information modeling, indoor navigation, and indoor 3D modeling [5,6] Although the accuracy required for gaming applications is not high, to extend the use of RGB-D sensors to surveying-type applications, accurate calibration of the device’s geometric parameters (i.e., camera focal lengths and baselines between cameras) and modeling the sensor errors (i.e., lens distortions and systematic depth errors) are necessary to produce high quality-spatial data and 3D models (e.g., cm-level precision).

The calibration of consumer-grade depth sensors has been widely investigated since the release of the first-generation Kinect in 2010. Various calibration methods, particularly for the depth sensor, have been studied by different research groups [6,7,8,9,10,11]. The calibration methods can be divided into four main categories.

The first category assumes that the three sensors (RGB camera, IR camera, and IR projector) behave similarly and pinhole camera concept is applicable to them all. This method is typified by [8] which uses the disparity data delivered from the IR camera and projector with the RGB image and a photogrammetric bundle adjustment to calibrate the internal and external parameters of the three sensors. Conventional distortion models [12,13] for each sensor separately compensate for their distortion effects. As it is difficult to obtain the raw measurements from the IR projector, the author computed such data from the disparity and the approximated baseline between the IR camera and projector. The main limitation of this category is the dependency between the observations from the IR camera and projector and their calibration parameters.

The second category combines disparity image produced by the depth sensor, the image produced by the RGB camera, and an empirical mathematical model to eliminate the distortions of the IR camera [9]. The distortion model is based on the error behavior of the whole RGB-D sensor. Unlike the first category, which is valid for any RGB-D sensor, this one is restricted to Kinect v1. Other limitations include the lack of automation and rigor. The user has to select the points in the depth image manually, and the homography matrix is computed using only four corresponding points.

The third category is dealing with the refinement of the in-factory calibration procedure, where the manufacture parameters including the baseline between IR camera and projector as well as the standard depth were precisely calibrated along with the RGB camera. Zhang’s [14] method is such a type of calibration; the author used the maximum likelihood estimation to calibrate color internal parameters and the manufacture parameters for a depth sensor without any distortion effect. The main limitation of this method is the distortion parameters for both IR cameras and projectors, which are not estimated or compensated.

Finally, the fourth category is mainly dealing with the depth calibration of the RGB-D sensor [15,16], the method adopts a mathematical error model derived from observation equation of depth measurements. The method used the output depth information from depth sensor and an adequate experiment setup to obtain the true depth of each pixel to model each pixel’s depth error. The method is valid for both SL and ToF RGB-D sensor. Although the error model was precisely applying to the middle part of the depth image, the whole image’s depth error model can be achieved and applied to the whole depth image. The error model has demonstrated a significant improvement while scanning small objects.

Based on these calibration algorithms, different calibration methods have been implemented and tested. Methods include the use of 1D [17] 2D [11], and 3D [18] calibration objects that work with the depth images directly; calibration of the manufacture parameters of the IR camera and projector [9,19]; or photogrammetric bundle adjustments used to model the systematic errors of the IR sensors [8,20]. To enhance the depth precision, additional depth error models are added to the calibration procedure [7,8,21,22,23]. All of these error models are used to compensate only for the distortion effect of the IR projector and camera. Other research works have been conducted to obtain the relative calibration between an RGB camera and an IR camera by accessing the IR camera [24,25,26]. This can achieve relatively high accuracy calibration parameters for a baseline between IR and RGB cameras, while the remaining limitation is that the distortion parameters for the IR camera cannot represent the full distortion effect for the depth sensor.

Two main issues faced by current RGB-D calibration procedures relate to the depth sensor. The first is the implementation of IR projector distortion along with the IR camera in the calibration procedure; the second is the correction of systematic errors resulting from the inaccuracy of in-factory calibration. This study addressed these issues using a two-step calibration procedure to calibrate all of the geometric parameters of RGB-D sensors. The first step was related to the joint calibration between the RGB and IR cameras, which was achieved by adopting the procedure discussed in [27] to compute the external baseline between the cameras and the distortion parameters of the RGB camera. The second step focused on the depth sensor calibration. First, the in-factory calibration parameters were updated to eliminate the systematic error resulting from the baseline between the IR camera and projector. Second, a combined distortion model was used to compensate for the distortion and systematic errors resulting from both the IR camera and projector. The experimental design and results are discussed in comparison with the conventional calibration method, and concluding remarks are made.

2. A Distortion Calibration Model for Depth Sensor

Numerous RGB-D sensors were released on the market based on a structured light concept consisting of an IR camera and an IR projector. In addition to IR sensors, an optional RGB camera acquires the color information of the observed scene. The IR projector emits a predefined pattern and the IR camera receives it [28]. The depth of the image is obtained by triangulation based on the distance between their camera and projector. Figure 1 shows the main elements of the sensors, which use the SL concept. Two sensors are shown: Kinect v1 and the Structure Sensor. The main difference between both sensors is the baseline between the IR camera and projector. The longer the sensor’s baseline, the longer working distance can be achieved. The working range of Kinect v1 is 0.80 m to 4.0 m, while it is 0.35 m to 3.5 m for Structure Sensor.

Figure 1.

Figure 1

Main elements of RGB-D sensors (right: Microsoft Kinect [3]; left: Structure Sensor [1]).

The principle of depth computation for RGB-D sensors is shown in Figure 2, where both IR sensors are integrated together to produce a pixel-shifted distance called disparity. The infrared projector pattern on a predefined plane (Z0) used in in-factory calibration [14] was stored in the sensor firmware. While capturing the real feature (Qi), both the IR standard projector pattern (xi,0c) and the real IR projector captured by IR camera (xic) are identified. The difference between both locations is called disparity di=xicxi,0c. Using the disparity value and the predefined configuration information, including the focal length (f) of the IR sensors, the baseline between the IR projector and camera (w), and the standard depth of the projected pattern (Z0) [5,29], we can compute the depth of the feature point (Qi ).

Figure 2.

Figure 2

Depth computation concept from RGB-D sensors.

Using the triangle similarity, the relationship between the standard pattern location (xi,0c), the real projector pattern location (xic) in the IR camera space and the IR projector location (xip) can be written as:

xi,0c=xip+fw/Z0 (1)
xic=xip+fw/Zi (2)

Applying the disparity di=xicxi,0c, the fundamental equation for computing the depth value for a feature point (Qi ) can be written as follows:

Zi=fwfw/Z0+di (3)

As noted previously, the disparity value is measured by the firmware of the RGB-D sensor. However, the output value by the sensor is the normalized value (din), which ranges from 0 to 2047. The sensor uses (α) and (β) as two linear factors to normalize the measured disparity (di). The normalized disparity is stated as din=1α(diβ). By substituting (din) in Equation (3) and combining all of the constants to assigned factors a and b, Equation (3) becomes:

Zi=1a+bdin (4)

where a and b are constants:

a=1Z0+βfwb=αfw (5)

The final coordinates Xi, Yi, and Zi for the acquired feature (Qi) are computed as:

Zi=1a+bdinXi=xicZifYi=yicZif (6)

The formula presented in Equation (4) with the factors a and b is called the manufacturer’s equation or the mapping function, which produces the depth information from the normalized disparity. The a and b coefficients are a function of the design parameters of the RGB-D sensor, which are the standard plane distance, the baseline, the focal length, and the linear parameters that convert the measured disparity to the normal disparity.

The baseline between IR camera and IR projector has a great influence in depth precision, using the covariance error propagation concept to estimate the variance of the depth resulting from SL RGB-D sensor. Using Equation (3) to estimate the depth variance, it can be figured out that σz=z2(fw)1σd, where σz and σd are the precision of depth and disparity, respectively. For a certain SL RGB-D sensor, if the baseline (w) between IR camera and IR projector was increased to double, the precision of depth will be improved by 50% assuming all other variables were constants.

The disparity value can be computed from three constants, f, w, and Z0, and two measured quantities, xic and xip. The measured quantities are affected by the distortion of the IR camera and projector, respectively. The systematic error can be assumed to be a function of the distortion parameters. The general expression that combines the effect of the systematic error and the distortion effect can be written as follows:

dti=di+de (7)

where dti is the true disparity, di is the measured disparity, and de represents the error resulting from the effect of lens distortion and systematic error for the IR sensors.

The disparity error can be expressed as a function of the distortion effect for both the IR camera and projector for both tangential and radial distortions:

de=(δtang.c+δrad.c)(δtang.c+δrad.c) (8)

where: δtang.c and δtang.p are the tangential distortion effect for the IR camera and projector, respectively, and δrad.c and δrad.p are the radial distortion effect for the IR camera and projector, respectively.

The radial distortion quantifies the lens quality, which is caused by bending the ray linked object, image, and focal points. The Brown model [30] with two factors (K1 and K2) is applied to compensate for the radial distortion. The tangential other model describes the distortion resulting from the inaccurate location of the lens with respect to the focal point, and the effect is compensated for using another two factors (P1 and P2) [12,31]:

δtang.=P1((xt2+y2t)+2xt)+P2xtytδrad.=xt(K1(xt2+y2t)+K2(xt2+y2t)2) (9)

where P1 and P2 are the factors represent the tangential distortion model, K1 and K2 are the factors represent the radial distortion model and xt and yt are the free distortion coordinates of the image point

As the disparity value is computed from the horizontal shift from the projected pattern to the standard one, the relation between the measured point coordinate and the true point coordinate can be expressed as follows:

xm=xt+δtang.+δrad. (10)

where xm is the measured x coordinate for the sensor and xt is the true x coordinate for the sensor.

Inserting Equation (9) into Equation (8), the full disparity error model considering both IR camera and projector distortions can be expressed in Equation (11) with eight parameters:

de=(P1((xt2+y2t)+2xt2)+P2xtyt)c(P1((xt2+y2t)+2xt2)+P2xtyt)p+(xt(K1(xt2+y2t)+K2(xt2+y2t)2))c(xt(K1(xt2+y2t)+K2(xt2+y2t)2))p (11)

where p refers to the IR projector and c refers to the IR camera.

To further simplify Equation (11), we used two parameters (W1 and W2) to describe the tangential distortion by correcting the relative orientation between the IR camera and projector lenses and applied another two parameters (W3 and W4) to describe the combined the radial distortion for both the IR camera and the projector, which can be considered as one lens combining the overlaying IR camera and projector lenses. As the relative orientation between the camera and projector is well fixed (as a rigid body) and pre-calibrated (mapping function calibration), the y axis for IR camera and IR projector can be assumed to be identical. The third and fourth terms of Equation (11) represent the radial distortion in the x direction resulting from the IR camera and projector. Due to the unknown projector data, we used the gross combined radial distortion, known as Seidal aberrations [12] and the IR camera’s pixel location to assign x distortion effect. This gives the two constraints shown in Equation (12):

ytcytpδsensorrad.=xtcF(xtc,ytc,W3,W4) (12)

Equation (11) can then be simplified as:

de=3W1(xc2xp2)+ycW2(xcxp)+W3xc(xc2xp2)+W4xc(xc4xp4+2yc2(xc2xp2)) (13)

Equation (13) is the fundamental equation that describes the depth sensor distortion with only four parameters. As xcxp=dt and xc+xp=2xcdt, we have:

xc2xp2=dt(2xcdt)xc2+xp2=dt(2xcdt)+2(xcdt)2 (14)

Finally, the full distortion model for the RGB-D sensor can be given as:

dei=[W1W2W3W4]T(3dti(2xidti)2yidtixi(dti(2xidti))xi[(dti(2xidti)+2(xidti)2)(dti(2xidti))+2yi2dti(2xidti)]) (15)

In Equation (15), the distortion model uses four parameters that consider both the radial and tangential distortions for both IR camera and projector lenses.

3. RGB-D Sensor Calibration

The calibration procedure was divided into two steps, with the first step handling the calibration of the RGB and IR cameras’ geometric parameters, including the focal length, principal point, and distortion parameters, and calibrated the RGB-IR baseline. The second step deals with the depth sensor calibration, whereby the manufacturer’s parameters and the proposed combined distortion model parameters are computed. In this section, we illustrate the methodology we used to achieve the two calibration steps.

3.1. RGB-D Joint Calibration

The methodology for jointly calibrating the RGB and IR cameras is indicated in the following chart (Figure 3). It started from stereo images produced by the RGB and IR cameras while the IR projector was switched off. After acquiring the stereo images, automatic corner extraction for a conventional A3 checkerboard was performed to extract the image points for each image. Using the ground coordinates of the checkerboard and extracted image points, a pinhole camera model was applied based on the calibration method discussed in [27] with five distortion parameters for each camera (three [K1 K2 K3] represented the tangential distortion and two [P1 P2] represented the radial distortion). After we optimized the external and internal parameters for both cameras, the global refinement was applied to enhance the estimated geometric parameters for each camera as well as the baseline between the RGB and IR cameras using Equation (16) [32]:

minnNmM(pmn(Pmn[RnTn][K])color2+pmn(Pmn[RnTn][K])IR2) (16)

where N is total number of images, M is the total number of points, Pmn is the point pixel coordinates, Pmn is the ground point coordinates, Rn and Tn are the rotation and translation matrix and K is the camera intrinsic matrix

Figure 3.

Figure 3

RGB-IR cameras baseline calibration methodology.

3.2. Error and Distortion Model

The parameters calibrated in this step included the manufacturer constants a and b and the distortion parameters for depth sensors indicated in Equation (15). In this stage, the IR projector was switched on and the information produced by the depth sensor, presented in normal disparity and depth images, were acquired. After that, the recovery of manufacturer constants was estimated by applying Equation (4). Using the true depth and normal disparity, the manufacturer constants were calibrated using the least squares method, and the normal disparity was corrected based on calibrated a and b. The proposed distortion model was adopted to model the remaining distortion error resulted from the difference between true disparity and normal disparity. Using least square method with a weight matrix based on depth information, Equation (15) was solved and W’s parameters were computed. The methodology of the error model and distortion effect is introduced in Figure 4.

Figure 4.

Figure 4

Computation methodology of depth sensor distortion model parameters.

4. Experimental Design and Data Collection

Our experiments were designed to achieve the full calibration parameters for the RGB-D sensor. The calibration parameters were divided into three sets. The first set was the baseline between the RGB and IR cameras with the RGB camera’s internal parameters. The second set was the calibration of the mapping function (Equation (4)), which could consume the systematic error resulting from the baseline between the IR camera and projector, the standard depth, and the focal length of the depth sensor. The third set contained the distortion model parameters, which corrected the relative distortion resulting from the IR camera and projector. Dealing with the true value of observed plane, each sensor was attached to IPad device and setup on a movable tripod, a four control points were staked on each IPad. Using 0.50-m interval and starting from 0.50 to 3.00 m, six stations were identified. For each station, the control points in IPad were acquired using a total station, the distance between the sensor and observed plane was computed for the designated stations. Between the six stations, we collected several true depth images to enrich the observed data set. Based on the previously mentioned three steps, the data were collected in two phases. In the first phase, we collected a stereo-pairs image of the RGB and IR cameras for an ordinary chessboard to achieve the first set of calibration parameters. In the second phase, we collected pairs of depth and disparity images for a planar surface that we used to calibrate the depth sensor for the calibration parameters of sets two and three. We conducted the experiments on two different Structure Sensors. Table 1 shows the number of images collected to calibrate each sensor for two phases. Sensor 1 was attached to an iPad Air with serial number (S.N. 26779), while Sensor 2 was attached to an iPad Air 2 with serial number (S.N. 27414).

Table 1.

Data description for an RGB-D sensor (Structure Sensor).

Sensor Phase 1 Phase 2
IR RGB Disparity Depth
1 53 90
2 59 44

5. Calibration Results

The phase 1 data for each sensor were processed to compute the calibrated baseline between the RGB and IR cameras, while the phase 2 data were processed to calibrate the depth sensor. Table 2 and Table 3 are the results of phase1 for sensors 1 and 2. The output data are the internal parameters for the RGB and IR cameras. The internal parameters are represented as camera focal lengths (Fx, Fy) in pixels and principal point (Cx, Cy) in pixels and five distortion vectors (K1, K2, P1, P2, and K3), where K’s are introduced to consume the radial distortion effect and P’s are presented to eliminate the tangential distortion effect. The IR-RGB camera baseline is expressed in six parameters, including three translations (dx, dy, and dz) in mm and three rotation Euler angles (Rx, Ry, and Rz) in radians.

Table 2.

The internal calibration parameters for both sensors.

Parameter Fx (Pixels) Fy (Pixels) Cx (Pixels) Cy (Pixels) K1 K2 K3 P1 P2
Sensor 1 RGB Camera 592.59 ± 2.23 590.48 ± 2.04 305.69 ± 1.84 235.92 ± 1.30 0.019 ± 0.012 1.144 ± 0.091 −3.580 ± 0.291 −0.004 ± 0.001 −0.015 ± 0.002
IR Camera 592.49 ± 2.17 591.22 ± 2.02 301.02 ± 1.76 239.9 ± 1.26 −0.13 ± 0.013 1.219 ± 0.131 −3.305 ± 0.476 0.000 ± 0.000 −0.014 ± 0.001
Sensor 2 RGB Camera 579.23 ± 2.22 580.29 ± 2.18 330.41 ± 1.43 242.63 ± 1.15 0.061 ± 0.014 0.744 ± 0.15 −3.607 ± 0.512 0.005 ± 0.001 0.006 ± 0.001
IR Camera 568.57 ± 2.15 570.35 ± 2.11 323.83 ± 1.37 247.41 ± 1.16 −0.05 ± 0.018 0.343 ± 0.248 −0.982 ± 0.993 0.003 ± 0.001 0.003 ± 0.001

Table 3.

The baseline calibration parameters for both sensors.

Parameter Sensor 1 Sensor 2
dx (mm) −37.997 ± 0.154 −32.603 ± 0.101
dy (mm) −3.861 ± 0.117 0.431 ± 0.091
dz (mm) −23.170 ± 0.656 −22.479 ± 0.473
Rx (rad) 0.0079 ± 0.0018 0.0110 ± 0.0021
Ry (rad) −0.0024 ± 0.0025 −0.0044 ± 0.0028
Rz (rad) −0.0053 ± 0.0002 0.0013 ± 0.0003

Table 4 shows the default parameters used by the firmware of the structure sensor, stated as the internal parameters for the depth sensor and color camera. The focal length and principal point for both sensors are the same, while the distortion parameters for both RGB and IR cameras are set are to zero.

Table 4.

The preset calibration parameters for the Structure Sensor.

Parameter RGB Camera IR Camera
Fx (pixels) 566.80 566.80
Fy (pixels) 566.80 566.80
Cx (pixels) 320 320
Cy (pixels) 240 240

After calibrating the baseline between the RGB and IR cameras and the internal parameters for the RGB camera, the phase 2 data were processed to calibrate the depth sensor. The two steps, including the mapping function calibration and distortion model, were conducted to deliver a high-precision depth measurement from the sensor. Table 5 shows the calibration result for mapping the function calibration. a and b are the mapping parameters mentioned in Equation (4).

Table 5.

a and b values before and after calibration for both sensors

Sensor In-Factory Calibrated Value Calibrated Value
a b a b
1 −3.38807 × 10−6 3.82665 × 10−3 −3.42936 × 10−6 3.86688 × 10−3
2 −3.38649 × 10−6 3.82538 × 10−3 −3.34912 × 10−6 3.78253 × 10−3

After computing the calibrated mapping function, we modified the measured depth and disparity information to correct the systematic error resulting from the mapping function error, then continued to compute the distortion model parameters. Figure 5 shows the distortion model parameters set for both Structure Sensors.

Figure 5.

Figure 5

Distortion parameters for both sensors. (a) is W1; (b) is W2; (c) is W3; (d) is W4.

Figure 5 shows the main conclusion for the calibration procedure. Although the distortion parameters (W1, W2, W3, and W4) change with the measured distance, after 2.50 m, each distortion model parameter tends to be the same value. This means that for the full calibration parameters, it could be sufficient to collect the depth data with corresponding disparity images up to 2.50 m.

6. Accuracy Assessment of the Calibration Models

To examine our calibration methodology as well as the distortion model performance on the depth accuracy, we captured a new dataset for each sensor and applied the calibration results. To examine the IR-RGB baseline calibration, two images (depth and RGB) were collected and aligned using the calibrated parameters. Figure 6 shows the effect of the calibration parameters.

Figure 6.

Figure 6

Alignment result for IR-RGB camera baseline (a) after applying calibration; (b) before applying calibration).

To examine the performance of the depth calibration parameters, including the calibration of the mapping function and the distortion model, we collected several depth images for a planar surface and applied our calibration parameters. The examination criterion was based on the same procedure illustrated in [8,9]. Compared with the planes resulted from the total station, the RMSE of the fitted plane surface was used to describe the measured depth precision. Figure 7 and Figure 8 show the depth precision performance evaluation for sensors 1 and 2, respectively. The left side introduces the full range performance, and the right side zooms in to display the near range.

Figure 7.

Figure 7

Depth precision variation with measured depth for Sensor 1. (a) the full range; (b) zoomed-in display of near range.

Figure 8.

Figure 8

Depth precision variation with measured depth Sensor 2. (a) the full range; (b) zoomed-in display of near range.

In addition to the planar surfaces assessment, two perpendicular planar surfaces (part of wall and ceiling) were captured using one of the calibrated sensors, the average distance between the sensor and the observed planes is 2.00 m (with minimum of 1.20 m and maximum of 3.00 m). The data were processed using the default parameters provided by the manufacturer and processed again after applying our distortion model.

Figure 9 shows the point cloud resulted from the RGB-D sensor before and after applying our distortion model. It is clearly seen that the warp in the wall was removed after calibration also the distortion on the depth image corners was significantly compensated. Comparison between the computed angle before and after applying our distortion model was shown in Table 6. Using different threshold for RANSAC to extract planes, the recovered average angle using our method is 89.897 ± 0.37 compared to 90.812 ± 7.17 using the default depth.

Figure 9.

Figure 9

Point cloud of two perpendicular planes (blue color: default depth; red color: modeled depth): highlighted black dashed circles shows the significant impact of the calibration method on the point cloud quality.

Table 6.

Recovered angle between two perpendicular planes using our calibration method and default calibration.

RANSAC Threshold (m) Recovered Angle (Degrees)
Default Depth Modeled Depth
0.001 79.8288 89.8004
0.002 99.8740 89.3294
0.005 91.5966 89.9098
0.010 92.2871 90.2850
0.020 90.4728 90.1596

Comparing our results with those given in [9] our calibration method achieved nearly similar accuracy for the near range and a significant improvement in accuracy for the far range. However, our calibration method is simpler than the method given in [9] and the error model is based on a mathematical concept of the lens distortion effect.

7. Conclusions and Future Work

In this study, we propose a new method for calibrating the RGB-D sensor. This method can be applied to either three fixed cameras such as Kinect or separate systems like the Structure Sensor. The method was fully automated for both steps, which included calibrating the external RGB-IR baseline and modeling the distortion and depth error for a depth sensor. Based on the structured light concept, we also proposed a new distortion error model to compensate for the systematic and distortion effects for the RGB-D sensor. Better accuracy could be achieved for both the near and far ranges, compared with traditional calibration methods that used ordinary stereo calibration to produce a distortion model for a depth sensor or applied an empirical distortion model. Traditional calibration procedures can be used to achieve a significant improvement in depth accuracy for the near range (up to 50% accuracy improvement), which was already achieved in this procedure. For a far range, the traditional methods cannot be used to achieve any significant improvement in depth precision compared with our method; we could achieve an accuracy improvement of around 40% for the far range. The RGB-D sensor can extend up to 9.0 m in full depth range. Future research should focus on improving the depth uncertainty result from a small baseline between IR sensors to achieve a better depth precision for a far range. Calibration parameter stability over time and different light conditions must also be examined.

Acknowledgments

The work described in this paper was substantially supported by a grant from The National Key Research and Development Program of China (No. 2016YFB0501803).

Author Contributions

W.D. and W.C. conceived and designed the experiments, W.D., S.T. and W.L. performed the experiments and analyzed the data. W.D. wrote the article, and W.C. provided editorial assistance.

Conflicts of Interest

The authors declare no conflict of interest.

References

  • 1.Occipital. [(accessed on 22 November 2015)]; Available online: http://structure.io/
  • 2.Xtion PRO LIVE. [(accessed on 5 April 2016)]; Available online: https://www.asus.com/3D-Sensor/Xtion_PRO_LIVE/
  • 3.Meet Kinect for Windows. [(accessed on 15 January 2016)]; Available online: https://dev.windows.com/en-us/kinect.
  • 4.Gokturk S.B., Yalcin H., Bamji C. A time-of-flight depth sensor-system description, issues and solutions; Proceedings of the Conference on Computer Vision and Pattern Recognition Workshop; Washington, DC, USA. 27 June–2 July 2004. [Google Scholar]
  • 5.Yamazoe H., Habe H., Mitsugami I., Yagi Y. Easy depth sensor calibration; Proceedings of the 21st International Conference on Pattern Recognition (ICPR); Tsukuba, Japan. 11–15 November 2012. [Google Scholar]
  • 6.Wang K., Zhang G., Bao H. Robust 3D reconstruction with an RGB-D camera. IEEE Trans. Image Process. 2014;23:4893–4906. doi: 10.1109/TIP.2014.2352851. [DOI] [PubMed] [Google Scholar]
  • 7.Haggag H., Hossny M., Filippidisy D., Creighton D., Nahavandi S., Puriy V. Measuring depth accuracy in RGBD cameras; Proceedings of the 7th International Conference on Signal Processing and Communication Systems (ICSPCS); Carrara, Australia. 16–18 December 2013. [Google Scholar]
  • 8.Chow J.C.K., Lichti D.D. Photogrammetric bundle adjustment with self-calibration of the primesense 3D camera technology: Microsoft Kinect. IEEE Access. 2013;22:465–474. doi: 10.1109/ACCESS.2013.2271860. [DOI] [Google Scholar]
  • 9.Herrera C.D., Kannala J., Heikkila J. Joint depth and color camera calibration with distortion correction. IEEE Trans. Pattern Anal. Mach. Intell. 2012;34:2058–2064. doi: 10.1109/TPAMI.2012.125. [DOI] [PubMed] [Google Scholar]
  • 10.Wang Y.-T., Shen C.-A., Yang J.-S. Calibrated Kinect sensors for robot simultaneous localization and mapping; Proceedings of the 2014 19th International Conference on Methods and Models in Automation and Robotics (MMAR); Miedzyzdroje, Poland. 2–5 September 2014. [Google Scholar]
  • 11.Shibo L., Qing Z. A New approach to calibrate range image and color image from Kinect; Proceedings of the 4th International Conference on Intelligent Human-Machine Systems and Cybernetics; Nanchang, China. 26–27 August 2012. [Google Scholar]
  • 12.Fryer J.G. Camera calibration in non-topographic photogrammetry. In: Karara H., editor. Non-Topographic Photogrammetry. American Society for Photogrammetry and Remote Sensing; Bethesda, MD, USA: 1989. pp. 59–70. [Google Scholar]
  • 13.Zhang Z. A flexible new technique for camera calibration. IEEE Trans. Pattern Anal. Mach. Intell. 2000;22:1330–1334. doi: 10.1109/34.888718. [DOI] [Google Scholar]
  • 14.Zhang C., Zhang Z. Calibration between depth and color sensors for commodity depth cameras; Proceedings of the 2011 IEEE International Conference on Multimedia and Expo (ICME); Barcelona, Spain. 11–15 July 2011. [Google Scholar]
  • 15.Pagliari D., Pinto L. Calibration of kinect for Xbox one and comparison between the two generations of Microsoft sensors. Sensors. 2015;15:27569–27589. doi: 10.3390/s151127569. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Lachat E., Macher H., Landes T., Grussenmeyer P. Assessment and calibration of a RGB-D camera (kinect v2 sensor) towards a potential use for close-range 3D modeling. Remote Sens. 2015;7:13070–13097. doi: 10.3390/rs71013070. [DOI] [Google Scholar]
  • 17.Liu W., Fan Y., Zhong Z., Lei T. A new method for calibrating depth and color camera pair based on Kinect; Proceedings of the 2012 International Conference on Audio, Language and Image Processing (ICALIP); Shanghai, China. 16–18 July 2012. [Google Scholar]
  • 18.Gui P., Ye Q., Chen H., Zhang T., Yang C. Accurately calibrate Kinect sensor using indoor control field; Proceedings of the 2014 3rd International Workshop on Earth Observation and Remote Sensing Applications (EORSA); Changsha, China. 11–14 June 2014. [Google Scholar]
  • 19.Raposo C., Barreto J.P., Nunes U. Fast and accurate calibration of a Kinect sensor; Proceedings of the 2013 International Conference on 3D Vision-3DV; Seattle, WA, USA. 29 June–1 July 2013. [Google Scholar]
  • 20.Davoodianidaliki M., Saadatseresht M. calibrate kinect to use in computer vision, simplified and precise; Proceedings of the 8th Iranian Conference on Machine Vision and Image Processing (MVIP); Zanjan, Iran. 10–12 September 2013. [Google Scholar]
  • 21.Mallick T., Das P.P., Majumdar A.K. Characterizations of noise in Kinect depth images: A review. IEEE Sens. J. 2014;14:1731–1740. doi: 10.1109/JSEN.2014.2309987. [DOI] [Google Scholar]
  • 22.Khoshelham K., Elberink S.O. Accuracy and resolution of kinect depth data for indoor mapping applications. Sensor. 2012;12:1437–1454. doi: 10.3390/s120201437. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Filippo B., Pretto A., Emanuele M. Unsupervised intrinsic and extrinsic calibration of a camera-depth sensor couple; Proceedings of the 2014 IEEE International Conference on Robotics and Automation (ICRA); Hong Kong, China. 31 May–7 June 2014. [Google Scholar]
  • 24.Macknojia R., Chávez-Aragón A., Payeur P., Laganière R. Calibration of a network of kinect sensors for robotic inspection over a large workspace; Proceedings of the IEEE Workshop on Robot Vision (WORV); Clearwater Beach, FL, USA. 15–17 January 2013. [Google Scholar]
  • 25.Ju Z., Wang Y., Zeng W., Chen S., Liu H. Depth and RGB image alignment for hand gesture segmentation using Kinect; Proceedings of the 2013 International Conference on Machine Learning and Cybernetics; Tianjin, China. 14–17 July 2013. [Google Scholar]
  • 26.Kim J.H., Choi J.S., Koo B.K. Calibration of multi-kinect and multi-camera setup for full 3D reconstruction; Proceedings of the 44th International Symposium on Robotics (ISR); Seoul, Korea. 24–26 October 2013. [Google Scholar]
  • 27.Morvan Y. Acquisition, Compression and Rendering of Depth and Texture for Multi-View Video. [(accessed on 1 June 2015)]; Available online: http://www.epixea.com/research/multi-view-coding-thesis.html.
  • 28.Jia W., Yi W.-J., Saniie J., Oruklu E. 3D image reconstruction and human body tracking using stereo vision and Kinect technology; Proceedings of the 2012 IEEE International Conference on Electro/Information Technology (EIT); Indianapolis, IN, USA. 6–8 May 2012; pp. 1–4. [Google Scholar]
  • 29.Khoshelham K. Accuracy analysis of kinect depth data; Proceedings of the ISPRS Calgary 2011 Workshop; Calgary, AB, Canada. 29–31 August 2011; pp. 133–138. [Google Scholar]
  • 30.Heikkila J., Silven O. A Four-step camera calibration procedure with implicit image correction; Proceedings of the IEEE Computer Society Conference on Computer Vision and Pattern Recognition; San Juan, Puerto Rico. 17–19 June 1997. [Google Scholar]
  • 31.Fryer J.G., Brown D.C. Lens distortion for close-range photogrammetry. Photogramm. Eng. Remote Sens. 1986;52:51–58. [Google Scholar]
  • 32.Tang S., Zhu Q., Chen W., Darwish W., Wu B., Hu H., Chen M. Enhanced RGB-D mapping method for detailed 3D indoor and outdoor modeling. Sensors. 2016;16:1589. doi: 10.3390/s16101589. [DOI] [PMC free article] [PubMed] [Google Scholar]