R G Holt 1, D Luo 2, N Gruver 1, D B Khismatullin 2,*
PMCID: PMC5495613  NIHMSID: NIHMS872152  PMID: 28453901

Summary

Background

Thromboelastography is widely used as a tool to assess the coagulation status of surgical patients. It allows observation of changes in material properties of whole blood, beginning with early stages of clot formation and ending with clot lysis. However, the contact activation of the coagulation cascade at surfaces of thromboelastographic systems leads to inherent variability and unreliability in predicting bleeding or thrombosis risks.

Objectives

To develop acoustic tweezing thromboelastometry as a noncontact method for perioperative assessment of blood coagulation.

Methods

Acoustic tweezing is used to levitate microliter drops of biopolymer and human blood samples. By quasi-statically changing the acoustic pressure we control the sample drop location and deformation. Sample size, deformation and location are determined by digital imaging at each pressure.

Results

Simple Newtonian liquid solutions maintain a constant, reversible location vs deformation curve. In contrast, the deformation/location curves for gelatin, alginate, whole blood, and blood plasma uniquely change as the samples solidify. Increasing elasticity causes the sample to deform less, leading to steeper stress/strain curves. By extracting a linear regime slope, we show that whole blood or blood plasma exhibits a unique slope profile as it begins to clot. By exposing blood samples to pro- or anti-thrombotic agents, the slope profile changes, allowing detection of hyper- or hypo-coagulable states.

Conclusions

We demonstrate that quasi-static acoustic tweezing can yield information about clotting onset, maturation, and strength. The advantages of small sample size, non-contact and rapid measurement make this technique desirable for real-time monitoring of blood coagulation.

Keywords: blood coagulation tests, rheology, acoustics, whole blood coagulation time, biopolymers

Introduction

The depletion of both pro- and anti-coagulants and other alterations in the coagulation system, known generically as coagulopathy, often develop as a result of a significant blood loss during trauma, major surgery, or post-partum hemorrhage; extracorporeal blood flow during cardiopulmonary bypass or hemodialysis; or disseminated intravascular coagulation caused by systemic activation of the coagulation system in sepsis and cancer. Coagulopathic patients are at high risk of both bleeding and thrombotic complications and require perioperative monitoring of their coagulation status. Plasma coagulation assays, e.g., prothrombin time (PT) and activated partial thromboplastin time (aPTT), are a standard way to monitor for coagulation disorders. However, they have low predictive power in patients with coagulopathy [1]. In this situation, the use of whole-blood coagulation tests becomes important.

Contact “pin-and-cup” methods such as thromboelastography (TEG), rotational thromboelastometry (ROTEM), and Sonoclot are currently available to measure the coagulation status of whole blood [24]. These methods measure temporal changes in the shear force between a disposable cup containing a 0.3 – 0.4 ml sample of whole blood and a pin immersed in the blood sample. Intrinsic pathway activators such as kaolin or ellagic acid are required to initiate coagulation using this approach. The “pin-and-cup” techniques accurately diagnose hyperfibrinolysis and are helpful but not reliable tools in screening for hypercoagulable states and transfusion guidance [5]. However, the contact of a blood sample with the pin and cup surfaces creates artificial conditions for blood coagulation, leading to substantial differences from the dynamics of hemostasis in the body. This inherent deficiency is an important reason behind poor standardization and high variability of these methods [6], their inability to determine disorders of primary hemostasis [7], unreliability in detection of impaired platelet function [8] and prediction of bleeding after major surgery [9], and a strong effect of heparin flush on thromboelastographic parameters leading to the necessity of discarding a large volume of blood before measurement [10]. Even with intrinsic pathway activators present, the coagulation process occurring in “pin-and-cup” devices remains slow. A significant amount of time (30–60 minutes) is required to obtain the results needed for diagnosis unless the extrinsic pathway activators (e.g., tissue factor) are used [3, 11] or when the disease state can be detected using only early parameters such as reaction and kinetic times.

Employing a containerless (air-contact) processing technique would eliminate the problems alluded to above. Ultrasonic acoustic levitation has been widely utilized for containerless materials processing [1218]. Its low interference with samples and relative simplicity, along with its ability to quasi-statically or dynamically deform the sample with no contact make it a desirable candidate for rheological applications [1416, 19]. In the present study the quasi-static technique, referred to as “quasi-static acoustic tweezing thromboelastometry” (QATT), is employed to measure rheology of biological gels and human blood.

Materials and Methods

Reagents and solutions

In the proof-of-concept experiments, three gel mixtures were used: 2,000,000-MW dextran from Leuconostoc spp. (Sigma-Aldrich, St. Louis, MO), 300-bloom gelatin from porcine skin (Sigma-Aldrich), and sodium alginate (Sigma-Aldrich). Gel solutions were prepared by hydrating dextran, gelatin, or sodium alginate in a small amount of distilled water for ten minutes, then adding boiling distilled water to achieve the desired concentration. Calcium carbonate CaCO3 (Sigma-Aldrich) in combination with 6% (w/w) D-(+)-Gluconic acid δ-lactone (GDL, Sigma-Aldrich) was used as a source of calcium ions to initiate gelation of sodium alginate. The molar ratio of a basic calcium ion to carboxyl was kept at 0.36. The sodium alginate solution was mixed and vortexed with the CaCO3 suspension for one minute. A fresh aqueous GDL solution was then added to the resulting mixture to initiate gelation.

In the whole blood studies, Cytochalasin D (Sigma-Aldrich) dissolved in phosphate-buffered saline (PBS, Thermo Fisher Scientific, Waltham, MA) was used as an anti-platelet agent. Cytochalasin D inhibits platelet function by blocking polymerization of actin filaments. It is used in ROTEM to measure the functional level of Fibrinogen and diagnose Fibrin/Fibrinogen deficiency [20]. Recombinant human tissue factor (TF, R&D Systems, Minneapolis, MN) in PBS was used as a pro-coagulant. In the blood plasma studies, human Fibrinogen (Enzyme Research Laboratories, South Bend, IN) and Gly-Pro-Arg-Pro (GPRP, Sigma-Aldrich), both dissolved in PBS, were used as pro- and anti-thrombotic agents, respectively.

Preparation of blood samples

Whole blood samples were collected from 30 healthy male and female volunteers with age ranging from 22–39 using protocol 520566-13 approved by the Institutional Review Board of Tulane University. Each volunteer donated 29.9 ml blood, which transferred into Vacutainer tubes (BD, Franklin Lakes, NY) containing sodium citrate (blue cap, one 4.5 ml tube and two 2.7 ml tubes), EDTA (pink cap, three 6 ml tubes), or K2EDTA (purple cap, one 2.0 ml tube). One blue-cap and one purple-cap tubes were used for standard coagulation tests (PT, INR, APTT, fibrinogen, and platelet concentration) conducted at the Coagulation Laboratory of Tulane University Hospital. The remaining tubes were used in acoustic tweezing experiments.

For half the experiments no pre-treatment to initiate coagulation of the collected blood (whether citrated or EDTA-treated) was done. For the remaining experiments, 180 μl of citrated- or EDTA-treated blood was mixed with 20 μl of 0.2 M calcium chloride CaCl2 (Sigma-Aldrich) within 4 hours of blood collection to start the coagulation cascade. To test the effect of coagulation activators or anti-platelet agents, 16 EDTA-treated blood samples were mixed respectively with TF (8 samples) or Cytochalasin D (8 samples) 10 minutes before the sample re-calcification. The final concentration of TF and Cytochalasin D in blood was 13 pg/mL and 7.6 μg/mL, respectively. Immediately after pre-treatment, a 4 μL drop was deployed into the tweezing apparatus to begin the deformation experiments. This process was repeated until the desired number of experiments was performed.

Commercial factor assay control plasma (FACT), which is citrated blood plasma pooled from 30 or more normal human donors, was purchased from George King Bio-Medical (Overland Park, KS). 300 μL aliquots of FACT were stored in −80°C. To initiate coagulation, the aliquots were thawed in a 37°C water bath and then mixed with 0.2 M CaCl2 (22.5 μL of FACT per 2.5 μL CaCl2) immediately before tweezing. To determine the effect of pro- or anti-thrombotic agents on plasma coagulation, FACT samples were mixed with Fibrinogen or GPRP 10 minutes before the recalcification procedure. The final concentration of Fibrinogen in plasma, accounting for the intrinsic concentration of Fibrinogen (269 mg/dL) was 610 mg/dL. Note that the normal range for Fibrinogen concentration is 150–400 mg/dL [21]. The GPRP concentration was 340 mg/dL (8 mM).

Acoustic tweezing apparatus and quasi-static technique

The apparatus for acoustic tweezing, also known as an acoustic levitator, was custom fabricated. Figure 1A shows a schematic of the electrical and mechanical systems used to levitate samples. The transducer consists of two 3.175-mm thick piezoelectric discs (Channel Industries, Santa Barbara, CA) and aluminum bottom mass and horn. The working frequency of the transducer is nominally 30 kHz, requiring slight retuning to compensate for temperature shifts. The transducer and the reflector (an aluminum cylinder) were mounted either a full or half wavelength apart, and the assembly could be optionally inserted into a custom fabricated and sealed environmental chamber for pressure, temperature and humidity control or a 3-D positioning system custom built using parts bought from Thorlabs (Newton, NJ).

Figure 1.

Figure 1

(A) Schematic of the acoustic tweezing device. (B) Sequence of photos of a whole blood sample levitating in the device under increasing acoustic pressure (from left to right). The levitator frequency was 29.4 kHz, the field of view of each image is 5mm × 5mm, and the elapsed time between images is 3s.

The 30 kHz sinusoidal input signal was generated by a function synthesizer (Agilent 33220A, Santa Clara, CA) and amplified (Krohn-Hite 7500, Brockton, MA) before being sent to the transducer, whose resulting vibration creates an acoustic standing wave in the air gap between the transducer and the reflector.

QATT focuses on the deformed shapes that samples take when subjected to an acoustic pressure varying slowly enough to maintain mechanical equilibrium. In a typical experiment, a sample drop is levitated, and the simultaneous vertical location and drop shape deformation (aspect ratio) are obtained via imaging. The technique relies on the fact that, as the acoustic pressure amplitude changes (pressure sweep) but remains relatively low, the location and aspect ratio of a sample drop are uniquely determined by its material properties and size. The pressure sweep is completed in 30 seconds or less, which is much shorter than blood clotting time. In the low pressure amplitude, gravity-controlled limit in which the location changes appreciably, the location is a measure of the acoustic stress applied to the drop to lift it, while the aspect ratio is a measure of the strain resulting from the applied acoustic stress. Therefore, within this limit, a location vs aspect ratio curve represents an effective stress/strain curve for a given material.

Experimental procedure

Samples (4 μL nominally) were deployed manually into a pressure minimum of the standing wave using a gastight 100 μL glass syringe (Hamilton 7656, Reno, NV) with a polytetrafluoroethylene-coated stainless steel blunt-tipped needle (Hamilton 8646). Every experiment began with tuning the transducer to near resonance, and then injecting a sample into a pressure node of the standing wave. Sample deformation was induced by varying the standing wave pressure amplitude. This was accomplished either by varying the amplifier input voltage at a fixed frequency, or varying the frequency at a fixed voltage input. Sample images were obtained by a Retiga 1300 digital camera (QImaging, Surrey, Canada) or an acA1920-25um camera (Basler, Ahrensburg, Germany) at regular intervals. The spatial resolution of the images was 0.012 mm. Figure 1B shows typical shapes of a whole blood sample during an acoustic tweezing experiment. As seen in this figure, an increase in the deformation of the sample correlates with an increase in its vertical position.

Quantification of drop size, deformation and location

Location and shape deformation of tweezing samples were obtained by analyzing the image sequences using a custom program written in MATLAB (Mathworks, Natick, MA) which relied on the MATLAB image processing toolbox. The analysis began with edge detection using a modified Canny method [22]. The “blob analysis” tools within MATLAB were then used to find the centroid of the drop and quantify deformation as an aspect ratio (width b/height a, cf. Fig. 1B). Location was measured as a vertical distance from the sample centroid to a fixed location on the apparatus. Slopes of the initial portion of location vs aspect ratio curves were obtained by linear regression, and quantified by calculating the angle of inclination to the horizontal (aspect ratio) axis. Hence, for a line of slope m, we refer to an angle θ = arctan(m). For convenience, we refer to the slope angle vs time curves (Figs. 4B, 6,7) as ‘tweezographs’. Some of the tweezographs (Fig. 8) have been normalized based on θ0, the slope angle at time 0.

Figure 4.

Figure 4

Location vs aspect ratio curves of a 1.0mm radius drop of citrated whole blood at 0, 12, 20, 24, and 26 minutes. The blood sample was not exposed to CaCl2.

Figure 6.

Figure 6

(A) Tweezographs of EDTA-treated whole blood with added CaCl2 exposed to 0.9% saline (8 drops), TF (8 drops), or cytochalasin D (8 drops). (B) Effect of TF on θ at selected times; (C) Effect of Cytochalasin D on θ at selected times. *p < 0.05.

Figure 7.

Figure 7

(A) Tweezographs of citrated control plasma (FACT) with added CaCl2 exposed to 0.9% saline (9 drops), Fibrinogen (9 drops), or GPRP (9 drops). (B) Effect of Fibrinogen and GPRP on θ at 5, 6, and 7 minutes. **p < 0.01, ***p < 0.001.

Statistical analysis

25 volunteers’ data were analyzed for the results reported in Figs. 1, 5 and 6. 12–15 volunteers’ blood were subject to TF and Cytochalasin D, as reported in Fig. 7. The results were evaluated with t-test and ANOVA using GraphPad Prism version 5.0.2 (GraphPad Software, La Jolla, CA). The statistical data were presented as mean ± standard error of the mean (SEM). Statistically significant differences were set at p < 0.05 (95% confidence).

Figure 5.

Figure 5

Tweezographs of (A) 39 drops of citrated control and 47 drops of citrated + CaCl2 whole blood; (B) 44 drops of EDTA control and 50 drops of EDTA + CaCl2 whole blood.

Results

Newtonian fluids

Generally, for viscoelastic fluids such as human blood, the sample drop shape is a function of bulk viscosity and elasticity. These two properties have different time profiles during coagulation, potentially complicating the analysis and interpretation of raw measurement data. Blood coagulation is associated with an increase in blood elasticity (firmness), but its effect on blood viscosity is less clear and most likely transient. Thus, for coagulation analysis, it is desirable to have an elastometric technique, which is highly sensitive to changes in blood elasticity but not blood viscosity. To test whether QATT satisfies this requirement, we first conducted experiments with a dextran-in-water solution. This solution is a non-reacting Newtonian fluid [23] where the bulk viscosity increases with polymer concentration, the bulk elasticity is zero, and the restoring force for deformation is just the surface tension. As seen in Fig. 2A, the aspect ratio vs location curves of 5% dextran solution remain constant and reversible during 10 minutes of acoustic tweezing. The solution concentration has negligible effect on the curves (Fig. 2B), indicating that QATT is insensitive to changes in bulk viscosity.

Figure 2.

Figure 2

Location vs aspect ratio for dextran samples obtained by increasing and decreasing pressure amplitude. (A) A 0.95 mm radius drop of 5% dextran at 0, 6, 10 minutes. (B) Single drops of 1%, 3% and 5% dextran at 2 minutes. The drop radii were nominally 0.95mm.

A very different behavior is seen when the material under study undergoes a state change from a liquid to a solid. When gelatin or alginate are diluted in water, they form hydrogels characterized by much higher bulk elasticity than the initial solutions of these polymers. The location vs aspect ratio curves plotted in Fig. 3 demonstrate that QATT is sensitive to changes in bulk elasticity occurring during gelation of these materials. For both gelatin and alginate, the sample location increases approximately linearly with aspect ratio until reaching the gravity-controlled limit where the drop deforms more readily than its location increases. In the linear regime, the slope of the location vs aspect ratio curve for gelatin increases with time until 13 minutes later, when the sample is fully gelled (Fig. 3A). Figure 3B shows the concentration dependence of gelatin drops at 2 minutes into an experiment, when they are partially gelled. The slope in the linear region of the location vs aspect ratio curves increases with increasing gelatin concentration. Alginate, in contrast to porcine gels, will not gel without exposure to calcium ions. Thus, it behaves similar to citrated or EDTA-treated blood, which do not coagulate until exposure to calcium ions (recalcification). Figure 3C shows the location vs aspect ratio curves for alginate at selected times up to 28 minutes after initiation of gelation by a GDL solution. The linear response region is restricted to the data in the bounding box in the lower left corner of Fig. 3C. After performing linear regression on the data in the box, where the location is well below the gravity-controlled limit, we obtained the “tweezograph”, Fig. 3D, which plots the linear slope angle θ vs time. From the tweezograph we can clearly detect changes in gel firmness, as quantified by angle θ, due to an increase in polymer concentration. Additionally, we see that the firmness dramatically increases for 4% alginate at about 4 minutes after the alginate drop injection into the acoustic tweezing device. After about 14 minutes θ begins to level off with increasing time, indicating the approach to the fully gelled state. We identify these three distinct regions as Stage I (initial gelation or coagulation), Stage II (rapid gelation or coagulation), and Stage III (convergence to fully gelled or coagulated) in Fig. 3D to facilitate our discussion below. Taken together, Fig. 3 confirms that we may take θ as a measure of the sample firmness, and the method itself is capable of measuring time-dependent changes in the firmness of reacting samples. The “tweezograph” of the time dependence of θ displays three distinct stages of clotting kinetics.

Figure 3.

Figure 3

Location vs aspect ratio for porcine gelatin and alginate samples obtained by increasing and decreasing pressure amplitude. (A) 0.90mm radius drop of 3% gelatin at increasing times. (B) 0.90, 0.89 and 0.86mm drops of 2%, 3% and 4% gelatin at 2 minutes. (C) Location vs aspect ratio curves of a 4% alginate drop with radius of 0.98mm from 0 to 28 minutes. The box indicates the portion of the data for which we performed linear regression for location vs aspect ratio to obtain θ. (D) Tweezograph (θ vs time) of 4 drops (nominal radii of 0.98 mm) each of 3% and 4 % alginate from 0 to 26 minutes.

Blood

Healthy subjects studied (thirteen male and seventeen female) had the values of their coagulation parameters within normal ranges: PT of 10.53 ± 0.12 s, INR of 0.995 ± 0.017, APTT of 28.40 ± 0.94 s, fibrinogen of 290 ± 12 mg/dL, and 260.0 ± 8.0 platelets/ml.

Figure 4 shows the typical aspect ratio vs location curves for all whole blood samples. As with alginate, for short times (<20 minutes) the firmness increases with time, but the drops return to the same initial deformation and location. After 20 minutes, the coagulated drops return to a permanently deformed state at the lowest location. As the cyclic loading and unloading process continues this permanent deformation itself continues to increase with time.

Figure 5 shows the tweezographs for citrated acid (A) and EDTA (B) treated groups. Remarkably, all the curves (except EDTA without CaCl2 in Fig. 6B) show the same general features as the results for gelatin and alginate already shown in Fig. 3. At short times (Stage I), θ starts between 52° and 56° (citrated: 54.77±0.88, citrated+CaCl2: 55.92±1.78, EDTA:52.83±0.53, EDTA+CaCl2: 53.78±1.00), increasing gradually as coagulation proceeds for a time whose value depends on whether the coagulation is aided by the addition of free calcium ions via introduction of CaCl2. Following this initial mild increase in firmness, there follows a period of rapid increase (Stage II) in θ, which happens much sooner for CaCl2 treated blood in both citrated and EDTA samples. This period of rapid increase is then followed by a leveling off of the firmness increase (Stage III), approaching an asymptotic value for long times. It is noteworthy that the coagulation process converges at long times to roughly the same Stage III path, with values for θ reaching 82° – 84° (citrated: 81.72±0.91, citrated+CaCl2: 82.20±0.38, EDTA+CaCl2: 83.51±0.52) at 25 minutes when the experiments were halted.

Blood samples treated with EDTA without CaCl2 (Fig. 5B) exhibit a prolonged Stage I slow growth in firmness. These data were only shown up to 14 minutes, because many of these more slowly coagulating samples developed a ‘scab’ on the surface after 14 minutes. While scab formation is a natural outcome, and our method is certainly capable of detecting its formation, its presence only in the EDTA without CaCl2 group makes comparison to samples without scab formation difficult.

Figure 6 illustrates the effect of TF and Cytochalasin D on whole blood coagulation in the tweezing device. Samples treated with TF are immediately stiffer (TF: initial θ =57.3±1.63°, untreated: 52.32±1.64° in Fig. 6A), and, since they show almost immediate Stage II rapid coagulation, continue to be stiffer than the untreated group throughout the untreated group’s Stage I and II coagulation, as seen in Fig. 6B. On the other hand, samples treated with Cytochalasin D (Fig. 6C) display a longer Stage I with a delayed onset of Stage II, and thus remain less stiff relative to the untreated group. By 25 minutes, both treated and untreated samples converge to a similar Stage III firmness (untreated: 85.46±0.27°, TF: 85.51±0.60°, Cytochalasin-D: 84.79±0.49°).

Blood plasma samples show a similar three-phase coagulation process in their tweezographs (Fig. 7). As seen in Fig. 7A, high Fibrinogen plasma produces much stiffer clots (θ/θ0 =1.10) than FACT (θ/θ0 =1.05) or GPRP treated plasma (θ/θ0 =1.02). According to Fig. 7B, a significant difference in clot firmness between high Fibrinogen plasma and FACT was already observed at 5 minutes of sample tweezing (FACT: θ/θ0 = 1.02±0.002, high Fibrinogen: 1.04±0.01), while GPRP treated plasma showed a significant decrease in clot firmness as compared to FACT starting at 7 minutes (FACT: 1.03±0.002, GPRP: 1.02±0.003). Using FACT samples from the same batch, we obtained that the coefficient of variation (CV) for 12-minute QATT coagulation (θ/θ0) measurement ranged from 0.4% to 0.8%.

Discussion

We report here the development of an acoustic tweezing technique for acquiring soft biological material properties, and its initial application to gels and blood. The tweezograph analysis introduced in Figs. 4 and 6 illustrates the ability of our technique to detect time-dependent changes in elasticity of soft materials as they undergo solidification or in blood as it undergoes coagulation. The results establish that θ, the quantitative measure in the tweezograph, is robust and sensitive to the changes characteristic of such processes. When applied to whole blood (Fig. 6) or blood plasma (Fig. 7), our method can distinguish between normal, hyper and hypo-coagulable states in less than 10 minutes.

The coagulation status of whole blood in coagulopathic patients and patients requiring blood product administration is routinely assessed by thromboelastographic tests. In these techniques, a cuvette with blood and a suspended pin are forced into oscillatory rotation (TEG, ROTEM) or translation (Sonoclot) relative to each other. The clot that forms induces changes in the motion of the pin or cuvette, and the amplitude of the motion is plotted vs time for both clot formation and lysis [2]. As an example, in TEG, the envelope of the peak amplitude forms a rising sigmoidal curve (the “trace”). The trace is initially zero when no fibrin links the pin and cup – the time from beginning of a TEG experiment to the first non-zero trace is called the “Reaction time” or “R time”, which is a measure of the time to form a fibrin network between cup and pin, typically 5–10 minutes when no data is obtained from TEG. The apparent amplitude at long times (“MA” in TEG) is a measure of firmness, and the rise to MA is characterized by its angle with the horizontal time axis (“angle α” in TEG) at a chosen value (22mm in TEG) of the induced amplitude. The time to achieve this 22mm amplitude after the R time is reached is called the “Kinetic time” or “K time” in TEG, typically 1–3 minutes.

It is useful to compare our method and results with thromboelastographic tests. Our tweezograph is analogous to the trace in TEG. Because our sample size is so small, 4 μL compared to 360 μL in TEG, the R time in our method is shorter than our first measurement as indicated by Figs. 57. Short R times (80 – 140s) can only be achieved in TEG using additive clotting factors (“TEG-ACT”). TEG needs R time plus K time (minimum 3 min TEG-ACT, maximum 13 min TEG) in order to obtain data with which to distinguish coagulation states. By contrast, our method can immediately detect differences in coagulation states induced by tissue factor (Fig. 6) or requires 5 minutes to detect the effects of pro-thrombotic agents such as fibrinogen (Fig. 7). The anti-thrombotic agent, e.g., GPRP, effects become pronounced in 7 minutes (Fig. 7).

The constant oscillatory shear that TEG, ROTEM, and Sonoclot apply to a blood clot continuously disrupts the clot structure, leading to two competing processes: clotting and mechanical lysis. The trace obtained reflects the balance between these processes, with clotting dominating at early times and lysis dominating at later times. Importantly, it is unclear whether the lysis occurs in the bulk of the blood sample or at the surfaces of the pin or the cup. In contrast, our quasi-static technique is both noncontact and less destructive than the pin-and-cup methods, i.e., it measures clotting throughout the blood sample without causing significant mechanical disruption. As a result, as can be seen in Figs. 57, our method does not artificially limit the clot firmness at some arbitrary time.

Three stages are identified on our graphical data (cf. Fig. 3D). The transition from Stage I and Stage II occurs when the fibrin network begins to connect the surface and bulk of the blood sample, leading to acceleration of the firmness increase. By the end of Stage II, these connections fully populate the sample volume. The transition to Stage III is the manifestation of a cross-linked fibrin network, causing both slowing of the firmness increase and permanent deformation, as observed in Fig. 4.

There are many improvements possible and many avenues of application open to us in the future. Of particular interest is the development of the oscillatory technique [15, 16, 19], which allows inference of viscosity and surface tension of whole blood in the liquid state within a few seconds. As the blood coagulates, this technique will yield complementary information about firmness, while simultaneously allowing inference of energy dissipation in the solid state.

Small sample volume and non-contact, rapid measurement make the proposed method well suited for perioperative monitoring of blood coagulation in patients (especially, pediatric patients) with coagulation disorders such as hemophilia [24] and thrombophilia [25], critical care patients [2629], and patients with other diseases that cause the coagulation system dysfunction [3034]. This method could be a valuable tool for transfusion guidance and stored blood products analysis [3537]. The small sample volume required opens up the possibility of whole blood testing at home by patients with manageable diseases, who are on antithrombotic medication [38].

Essentials.

  • Blood coagulation measurement during contact with an artificial surface leads to unreliable data.

  • Acoustic tweezing thromboelastometry is a novel non-contact method for coagulation monitoring.

  • This method detects differences in the blood coagulation state within 10 minutes.

  • Coagulation data were obtained using a much smaller sample volume (4 μL) than currently used.

Acknowledgments

The authors thank J. Buell, H. Murad, and H. Yu for help with blood experiments. This study has been financially supported by American Heart Association Grant-in-Aid 13GRNT17200013, U.S. National Science Foundation grants 1438537 and 1438569, and in part by U.S. National Institutes of Health grant GM104940, which funds the Louisiana Clinical and Translational Science Center. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

Footnotes

Addendum

R. G. Holt and D. B. Khismatullin were responsible for the study design and interpretation of the data. D. Luo and N. Gruver acquired the gel data. D. Luo acquired the blood data. R. G. Holt and D. Luo wrote the manuscript. D. B. Khismatullin, R. G. Holt and D. Luo reviewed and revised the manuscript. All authors approved the final version of the manuscript.

Disclosure of conflict of interests

D. B. Khismatullin and R. G. Holt have a patent “PCT/US14/55559 - Apparatus, Systems & Methods for Non-Contact Rheological Measurements of Biological Materials” pending.

Other authors have nothing to disclose.

References

  • 1.Tripodi A, Mannucci PM. The coagulopathy of chronic liver disease. N Engl J Med. 2011;365:147–56. doi: 10.1056/NEJMra1011170. [DOI] [PubMed] [Google Scholar]
  • 2.Luddington RJ. Thrombelastography/thromboelastometry. Clin Lab Haematol. 2005;27:81–90. doi: 10.1111/j.1365-2257.2005.00681.x. [DOI] [PubMed] [Google Scholar]
  • 3.Collins S, MacIntyre C, Hewer I. Thromboelastography: Clinical Application, Interpretation, and Transfusion Management. AANA J. 2016;84:129–34. [PubMed] [Google Scholar]
  • 4.Saleem A, Blifeld C, Saleh SA, Yawn DH, Mace ML, Schwartz M, Crawford ES. Viscoelastic measurement of clot formation: a new test of platelet function. Ann Clin Lab Sci. 1983;13:115–24. [PubMed] [Google Scholar]
  • 5.Bolliger D, Seeberger MD, Tanaka KA. Principles and practice of thromboelastography in clinical coagulation management and transfusion practice. Transfus Med Rev. 2012;26:1–13. doi: 10.1016/j.tmrv.2011.07.005. [DOI] [PubMed] [Google Scholar]
  • 6.Chen A, Teruya J. Global hemostasis testing thromboelastography: old technology, new applications. Clin Lab Med. 2009;29:391–407. doi: 10.1016/j.cll.2009.04.003. [DOI] [PubMed] [Google Scholar]
  • 7.Lang T, von Depka M. Possibilities and limitations of thrombelastometry/thromboelastography. Hamostaseologie. 2006;26:S21– S9. [PubMed] [Google Scholar]
  • 8.Carr ME., Jr In vitro assessment of platelet function. Transfus Med Rev. 1997;11:106–15. doi: 10.1053/tm.1997.0110106. [DOI] [PubMed] [Google Scholar]
  • 9.Wang JS, Lin CY, Hung WT, O’Connor MF, Thisted RA, Lee BK, Karp RB, Yang MW. Thromboelastogram fails to predict postoperative hemorrhage in cardiac patients. Ann Thorac Surg. 1992;53:435–9. doi: 10.1016/0003-4975(92)90264-5. [DOI] [PubMed] [Google Scholar]
  • 10.Durila M, Kalincik T, Pacakova Z, Cvachovec K. Discard volume necessary for elimination of heparin flush effect on thromboelastography. Blood Coagul Fibrinolysis. 2010;21:192–5. doi: 10.1097/MBC.0b013e3283338c0d. [DOI] [PubMed] [Google Scholar]
  • 11.Chitlur M, Sorensen B, Rivard GE, Young G, Ingerslev J, Othman M, Nugent D, Kenet G, Escobar M, Lusher J. Standardization of thromboelastography: a report from the TEG-ROTEM working group. Haemophilia. 2011;17:532–7. doi: 10.1111/j.1365-2516.2010.02451.x. [DOI] [PubMed] [Google Scholar]
  • 12.Marston PL. Shape Oscillation and Static Deformation of Drops and Bubbles Driven by Modulated Radiation Stresses - Theory. J Acoust Soc Am. 1980;67:15–26. doi: 10.1121/1.383798. [DOI] [Google Scholar]
  • 13.Tian YR, Holt RG, Apfel RE. Deformation and Location of an Acoustically Levitated Liquid-Drop. Journal of the Acoustical Society of America. 1993;93:3096–104. [Google Scholar]
  • 14.Tian Y, Holt RG, Apfel RE. A New Method for Measuring Liquid Surface-Tension with Acoustic Levitation. Review of Scientific Instruments. 1995;66:3349–54. [Google Scholar]
  • 15.Tian Y, Holt RG, Apfel RE. Investigations of Liquid Surface Rheology of Surfactant Solutions by Droplet Shape Oscillations - Theory. Physics of Fluids. 1995;7:2938–49. doi: 10.1006/jcis.1996.4698. [DOI] [PubMed] [Google Scholar]
  • 16.Tian Y, Holt RG, Apfel RE. Investigation of liquid surface rheology of surfactant solutions by droplet shape oscillations: Experiments. Journal of Colloid and Interface Science. 1997;187:1–10. doi: 10.1006/jcis.1996.4698. [DOI] [PubMed] [Google Scholar]
  • 17.McDaniel JG, Holt RG. Measurement of aqueous foam rheology by acoustic levitation. Physical Review E. 2000;61:R2204–R7. [Google Scholar]
  • 18.Ran W, Saylor JR, Holt RG. Improved particle scavenging by a combination of ultrasonics and water sprays. J Aerosol Sci. 2014;67:104–18. doi: 10.1016/j.jaerosci.2013.08.009. [DOI] [Google Scholar]
  • 19.Khismatullin DB, Nadim A. Shape oscillations of a viscoelastic drop. Phys Rev E Stat Nonlin Soft Matter Phys. 2001;63:061508. doi: 10.1103/PhysRevE.63.061508. [DOI] [PubMed] [Google Scholar]
  • 20.Solomon C, Schochl H, Ranucci M, Schlimp CJ. Can the Viscoelastic Parameter alpha-Angle Distinguish Fibrinogen from Platelet Deficiency and Guide Fibrinogen Supplementation? Anesth Analg. 2015;121:289–301. doi: 10.1213/ANE.0000000000000738. [DOI] [PubMed] [Google Scholar]
  • 21.Wehinger H, Klinge O, Alexandrakis E, Schurmann J, Witt J, Seydewitz HH. Hereditary hypofibrinogenemia with fibrinogen storage in the liver. Eur J Pediatr. 1983;141:109–12. doi: 10.1007/BF00496800. [DOI] [PubMed] [Google Scholar]
  • 22.Canny J. A computational approach to edge detection. IEEE Trans Pattern Anal Mach Intell. 1986;8:679–98. [PubMed] [Google Scholar]
  • 23.Tirtaatmadja V, Dunstan DE, Boger DV. Rheology of dextran solutions. Journal of Non-Newtonian Fluid Mechanics. 2001;97:295–301. doi: 10.1016/S0377-0257(00)00226-3. [DOI] [Google Scholar]
  • 24.Chitlur M, Young G. Global assays in hemophilia. Semin Hematol. 2016;53:40–5. doi: 10.1053/j.seminhematol.2015.10.001. [DOI] [PubMed] [Google Scholar]
  • 25.Jennings I, Cooper P. Screening for thrombophilia: a laboratory perspective. Br J Biomed Sci. 2003;60:39–51. doi: 10.1080/09674845.2003.11783675. [DOI] [PubMed] [Google Scholar]
  • 26.Herbstreit F, Winter EM, Peters J, Hartmann M. Monitoring of haemostasis in liver transplantation: comparison of laboratory based and point of care tests. Anaesthesia. 2010;65:44–9. doi: 10.1111/j.1365-2044.2009.06159.x. [DOI] [PubMed] [Google Scholar]
  • 27.Jensen E, Andreasson S, Bengtsson A, Berggren H, Ekroth R, Larsson LE, Ouchterlony J. Changes in hemostasis during pediatric heart surgery: impact of a biocompatible heparin-coated perfusion system. Ann Thorac Surg. 2004;77:962–7. doi: 10.1016/j.athoracsur.2003.09.028. [DOI] [PubMed] [Google Scholar]
  • 28.Merlani PG, Chenaud C, Cottini S, Reber G, Garnerin P, de Moerloose P, Ricou B. Point of care management of heparin administration after heart surgery: A randomized, controlled trial. Intensive Care Med. 2006;32:1357–64. doi: 10.1007/s00134-006-0220-8. [DOI] [PubMed] [Google Scholar]
  • 29.Nascimento B, Al Mahoos M, Callum J, Capone A, Pacher J, Tien H, Rizoli S. Vitamin K-dependent coagulation factor deficiency in trauma: a comparative analysis between international normalized ratio and thromboelastography. Transfusion. 2012;52:7–13. doi: 10.1111/j.1537-2995.2011.03237.x. [DOI] [PubMed] [Google Scholar]
  • 30.Loeffen R, Spronk HM, ten Cate H. The impact of blood coagulability on atherosclerosis and cardiovascular disease. J Thromb Haemost. 2012;10:1207–16. doi: 10.1111/j.1538-7836.2012.04782.x. [DOI] [PubMed] [Google Scholar]
  • 31.Borissoff JI, Heeneman S, Kilinc E, Kassak P, Van Oerle R, Winckers K, Govers-Riemslag JW, Hamulyak K, Hackeng TM, Daemen MJ, ten Cate H, Spronk HM. Early atherosclerosis exhibits an enhanced procoagulant state. Circulation. 2010;122:821–30. doi: 10.1161/CIRCULATIONAHA.109.907121. [DOI] [PubMed] [Google Scholar]
  • 32.Marcucci R, Sofi F, Fedi S, Lari B, Sestini I, Cellai AP, Pulli R, Pratesi G, Pratesi C, Gensini GF, Abbate R. Thrombophilic risk factors in patients with severe carotid atherosclerosis. J Thromb Haemost. 2005;3:502–7. doi: 10.1111/j.1538-7836.2005.01173.x. [DOI] [PubMed] [Google Scholar]
  • 33.Banga JD. Coagulation and fibrinolysis in diabetes. Semin Vasc Med. 2002;2:75–86. doi: 10.1055/s-2002-23098. [DOI] [PubMed] [Google Scholar]
  • 34.Noubouossie D, Key NS, Ataga KI. Coagulation abnormalities of sickle cell disease: Relationship with clinical outcomes and the effect of disease modifying therapies. Blood Rev. 2016;30:245–56. doi: 10.1016/j.blre.2015.12.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Karkouti K, McCluskey SA, Callum J, Freedman J, Selby R, Timoumi T, Roy D, Rao V. Evaluation of a novel transfusion algorithm employing point-of-care coagulation assays in cardiac surgery: a retrospective cohort study with interrupted time-series analysis. Anesthesiology. 2015;122:560–70. doi: 10.1097/ALN.0000000000000556. [DOI] [PubMed] [Google Scholar]
  • 36.Cardigan R, Van der Meer PF, Pergande C, Cookson P, Baumann-Baretti B, Cancelas JA, Devine D, Gulliksson H, Vassallo R, de Wildt-Eggen J. Coagulation factor content of plasma produced from whole blood stored for 24 hours at ambient temperature: results from an international multicenter BEST Collaborative study. Transfusion. 2011;51(Suppl 1):50S–7S. doi: 10.1111/j.1537-2995.2010.02963.x. [DOI] [PubMed] [Google Scholar]
  • 37.Scott E, Puca K, Heraly J, Gottschall J, Friedman K. Evaluation and comparison of coagulation factor activity in fresh-frozen plasma and 24-hour plasma at thaw and after 120 hours of 1 to 6 degrees C storage. Transfusion. 2009;49:1584–91. doi: 10.1111/j.1537-2995.2009.02198.x. [DOI] [PubMed] [Google Scholar]
  • 38.Salmela B, Joutsi-Korhonen L, Armstrong E, Lassila R. Active online assessment of patients using new oral anticoagulants: bleeding risk, compliance, and coagulation analysis. Semin Thromb Hemost. 2012;38:23–30. doi: 10.1055/s-0031-1300948. [DOI] [PubMed] [Google Scholar]

Articles from Journal of thrombosis and haemostasis : JTH are provided here courtesy of Health Research Alliance manuscript submission