Summary
Purpose:
To investigate the robustness of four fitting methods for bi-component effective spin-spin T2 (T2*) relaxation time analysis of human patellar tendon.
Methods:
A three-dimensional (3D) cone ultra-short echo-time (UTE) sequence was performed on the knees of ten healthy volunteers at 3.0T. Four fitting methods incorporating either Gaussian or Rician noise distribution were used for voxel-by-voxel bi-component T2* analysis of the patellar tendon. The T2* for the short relaxing (T**,s) and long relaxing (T*2,l) water components and the fraction of the short relaxing water component (fs) were measured, and different fitting methods were compared using Friedman’s and Wilcoxon signed rank tests. A numerical simulation study was also performed to predict the accuracy and precision of bi-component T2* parameter estimation in tendon at different signal-to-noise ratios (SNR) levels.
Results:
The average T*2,s, T*2,l, fs of human patellar tendon were 1.5ms, 30ms, and 80% respectively. Incorporating different noise models and fitting methods influenced the measured bi-component T2* parameters. Fitting methods incorporating Rician noise were superior to traditional fitting methods for bi-component T2* analysis especially at lower SNR. fs and T*2,s were less sensitive than T*2,1 to noise at even moderate and low SNR. The result of the in-vivo bi-component T2* analysis of tendon agreed well with numerical simulations.
Conclusion:
Our study demonstrated the use of a 3D cone UTE sequence to perform in vivo voxel-by-voxel bi-component T2* analysis of human patellar tendon. Incorporating Rician noise was useful for improving bi-component T2* analysis especially at lower SNR.
Level of evidence:
IV.
Keywords: tendon, T2 relaxation time, bi-component, curve fitting, noise
Introduction
Various imaging methods including radiographs, ultrasound, computed tomography, and conventional and quantitative magnetic resonance (MR) imaging can be used to evaluate musculoskeletal tissues. However, only quantitative MR imaging can be used to provide information regarding tissue composition and microstructure. Thus, these imaging techniques play an important role in the non-invasive assessment of disease-related and treatment-related changes in cartilage, bone, meniscus, and tendon. Spin-spin (T2) and effective spin-spin (T2*) relaxation times are the most commonly used quantitative MR method used in musculoskeletal imaging and have been shown to useful for evaluating tissue composition and microstructure 1–4. However, changes in T2 and T2* are nonspecific and can be caused by multiple factors including hydration, macromolecular content, and tissue anisotropy with comparable changes occurring in disparate settings 5–9.
Bi-component T2 and T2* mapping techniques have been used to improve the specificity of T2 analysis by assessing the individual water components of musculoskeletal tissues 10–22. Bi-component T2 and T2* mapping methods have measured two distinct T2 components in cartilage assumed to represent short relaxing water bound to the macromolecular matrix and long relaxing bulk water 12–14,21,22. Bi-component T2* mapping methods have been used in cortical bone to differentiate between water bound to the organic matrix and free water in the Haversian systems 15–17, Bi-component T2 and T2* mapping methods have also been used in the meniscus to differentiate between macromolecular bound water and bulk water 18–20.
In tendon, recent studies using bi-component T2* mapping techniques have measured two distinct T2 components representing short relaxing water bound to the highly organized collagen fibers and long relaxing bulk water19. Chang et al. detected short and long relaxing water components in histologically normal human cadaveric Achilles tendon with a T2* of 1.8ms and 9.2ms and a fraction of 79.2% and 20.8% respectively23. The Authors later reported that bi-component T2* parameters did not change with tensile loading of the Achilles tendon 24. Jarus et al. compared single-component and bi-component T2* analysis of the Achilles tendon in human subjects and found that the bi-component analysis provided greater diagnostic performance for distinguishing between normal and pathologic tendon 25.
However, multiple factors may affect estimations of bi-component T2 and T2* parameters within musculoskeletal tissues. Bi-component T2 measurements have been shown to be strongly influenced by experimental details such as the field strength, gradient system, pulse sequence, imaging parameters, and specimen preparation26,27. Bouhrara et al. also demonstrated that accurate estimation of bi-component T2 parameters requires sufficient signal-to-noise (SNR) ratio and proper noise modeling which is particularly important when estimating pixel-by-pixel spatial variations across a tissue sample28. In their study, simulations and imaging experiments on ex vivo phantoms and bovine nasal cartilage specimens were used to compare multiple model fitting methods and to investigate the sensitivity and robustness of bi-component T2* measurements at different SNR levels. Their conclusion resided in the importance of incorporating Rician noise models in bi-component T2* analysis of cartilage.
However, questions remain regarding the effectiveness of using Rician noise models for estimating bi-component T2* parameters in in vivo musculoskeletal imaging. Furthermore, the applicability of these models for imaging tendon which has different tissue composition and microstructure than bovine nasal cartilage requires further investigation. In this study, we investigated the applicability of translating observations derived from simulations and ex vivo experiments into in-vivo tendon imaging to provide insight into the best fitting methods for use in bi-component T2* analysis. More specifically, our study compared different curve fitting methods incorporating different noise models for measuring bi-component T2* parameters of the patellar tendon in a numerical simulation study and in healthy volunteers using a three-dimensional (3D) cone ultra-short echo-time (UTE) sequence.
Methods
Bi-Component signal model and fitting methods
A bi-component exponential signal model was used to characterize the short (denoted as s) and long (denoted as l) relaxing water components in patellar tendon, given by
| (1) |
| (2) |
where S is the acquired signal at varying TEs with noise, SN is the underlying noise free signal, A is the combined factor of proton density and hardware gain, T*2,s, and T*2,l are T2* for the short and long relaxing water components respectively, and fs is the water fraction for the short water component, defined as the ratio of the short water content to the total water (short + long) content. ε(TE) is the noise term at the individual echo with noise between each echo assumed to be independent.
Four different fitting methods using the non-linear least square minimization of the acquired signal and theoretical model were compared. Under the simple assumption of Gaussian distributed noise of the image signal, the first method, M1, is the traditional approach where |S–SN| is minimized. Under the assumption of a Gaussian distributed noise with equal variance in both real and imaginary channel of the acquired signal, the noise of the magnitude image can be described using Rician distribution as
| (3) |
where σ is the noise standard deviation. I0 denotes the modified zero order Bessel function of the first kind. The second fitting method, M2, is minimization of based on McGibney’s scheme for correcting noise bias at low SNR29. Gudbjartsson et al.30 also introduced a simple correction scheme, M3, for minimization of . Bouhrara et.al. 28 recently introduced a fitting scheme, M4, with the minimization of |S − SR| for
| (4) |
where and I1 is the modified first order Bessel function of the first kind. In all the fitting methods, the image SNR was estimated for the first echo signal and defined as 28,29
| (5) |
where the noise standard deviation σ was determined as the mean signal of background regions of all images acquired at each TE divided by 31.
Simulation study
Numerical simulations were performed to compare accuracy and precision of bi-component T2* parameter estimation using four different fitting methods. Theoretical multi-echo spin-echo signal data were generated based on bi-exponential decay equations as described in Eq.(1) for the real part of the signal. To simulate the Rician noise, the imaginary part of signal was created using only the zero mean Gaussian noise with the same noise standard deviation as the real part, and the magnitude of the complex signal was used as the final theoretical signal28. The final signal was fitted with the four different fitting methods and corresponding bi-component T2* parameters were estimated for each individual method. A total of 10,000 instances of noise were added to the signal in Monte-Carlo simulations to measure the accuracy and precision of the parameter estimation for fs, T*2,s, and T*2,l. The performance measure included the percentage error, which was defined as the difference between the estimated mean and true parameter value divided by the true parameter value. The same simulation was repeated with different levels of added noise at SNR = 20, 50 and 90.
The set of model parameters used in the numerical simulations was chosen to mimic bi-component T2* parameters of tendon previously published in the literature24. Accordingly, parameters of the two water pools were chosen as follows: fs =80%, T*2,s = 1 ms, and T*2,l = 20 ms. The signals were generated using the experimental design of the in vivo protocol described in the next subsection.
In vivo study
The in-vivo study was performed in compliance with Health Insurance Portability and Accountability Act (HIPPA) regulations and with approval from the University Institutional Review Board. The study met the ethical standards of the Muscle, Tendon, and Ligament Journal 32. All subjects signed written informed consent prior to their participation in the study. The study group consisted of 10 healthy volunteers (6 males with an average age of 28 years and 4 females with an average age of 27 years) who had no history of prior knee pain, trauma, or surgery.
All subjects underwent an MR examination of the right knee on the same 3.0T scanner (Discovery MR750, GE Healthcare, Waukesha, WI) using an 8-channel phased-array extremity coil (InVivo, Orlando, FL). Foam padding was used to firmly secure the knee within the coil to minimize subject motion during the MR examination. The patellar tendon was imaged in the sagittal plane using a 3D gradient-echo-based multi-echo UTE sequence 33. The UTE sequence utilized a k-space sampling scheme of a center-out twisted 3D cone trajectory which allowed a minimal echo time (TE) of 0.03ms. A total of 16 echoes were acquired at TEs of 0.03, 0.1, 0.8, 1.6, 4.3, 6.0, 8.0, 10.0, 14.0, 16.0, 18.0, 20.0, 24.0, 26.0, 28.0, 30.0 ms based upon TE selection from a previous tendon study 24. For each acquisition, the same repetition time (TR) was used, and four echoes were acquired. Other imaging parameters included 40ms TR, 20° flip angle, 16cm field-of-view, ±150KHz readout bandwidth, 1.21ms readout length, 256×256 in-plane matrix, 3mm slice thickness, one excitation, and 10 slices covering the entire patellar tendon. The total imaging time was approximately 12 minutes.
All images were registered to the first echo using a rigid registration method implemented in Elastix software to correct for subject motion between scans 34. The patellar tendon of all 10 subjects on all sagittal images was manually segmented by an experienced research assistant under the supervision of a musculoskeletal radiologist using an in-house software program developed in Matlab (Matlab 2010b, MathWorks Inc, Natick, MA). A voxel-by-voxel signal curve fitting was performed to obtain 3D parameter maps for fs, T*2,s, and T*2,l. The tendon masks were then superimposed over the bi-component T2* maps to measure the mean and standard deviation of fs, T*2,s, and T*2,l of the entire patellar tendon for all subjects.
The voxel-by-voxel curve fitting was performed using an in-house T2*, mapping software program developed based on the non-linear least square “fmincon” function implemented in Matlab. The initial parameter guess T*2,s = 1 ms, T*2,1 =20 ms and fs =80% based on previous published tendon studies was used in the iterative optimization 24. The parameter searching space were [0ms, 20ms], [0ms, 100ms] and [0%, 100%] for T*2,s, and T*2,l, and fs respectively, reflecting a reasonable parameter range within the patellar tendon. The algorithm converging iterations were continued until the step size between the successive estimates was small (<10−8), the object function was small (<10−8), or total iteration of 1000 was achieved. The voxels failing the converging iteration criteria or converged at the parameter searching space boundaries were recorded as the outliers, and the corresponding bi-component T2* parametric values were substituted with the value of 0 or boundary values respectively. The outliers were excluded prior to the statistical analysis to avoid statistical bias, and the percentage of outliers was recorded for each subject. To demonstrate the quality of the curve fitting models for the acquired image data, quality of fit measures were compared for the mean signal from a homogenous ROI chosen on a sagittal slice in the central portion of the patellar tendon to avoid partial volume effect with adjacent tissues. The relative residual error defined as the percentage of the residual error to the signal at different echo times was used to measure the fitting quality of the models at individual echoes. The overall measure of goodness-of-fit was tested for all the four methods by using chi-square, χ2, value.
The SNR was calculated on all subjects and referred to as SNRnormal. To estimate the model fitting robustness at different SNR levels, a set of repeat scans were performed on one subject using the same imaging parameters except for a 1.0mm slice thickness to decrease image SNR (referred to as SNRlow). Another set of repeat scans were performed on the same subject using the same imaging parameters except for four excitations to increase image SNR (referred to as SNRhigh).
Statistical analysis
Statistical analysis was performed using Matlab. Friedman’s tests were used to compare the means and standard deviations of fs, T*2,s, and T*2,l of the entire patellar tendon of all subjects for the four fitting methods. The Friedman’s test is a non-parametric one-way ANOVA test used to compare more than two groups. For those bi-component T2* parameters in which there was a significant difference between the four models, Wilcoxon signed rank tests were used for pairwise comparison to determine if there was significant differences between any two of the four fitting methods. The Holm-Bonferroni correction method was used to adjust all p-values to account for comparison of multiple MR parameters within the knee joint 35.
Results
Figure 1 demonstrates the accuracy and precision of bi-component T2* parameter estimation when using four different fitting methods at different SNR levels in Monte-Carlo simulation. At low SNR (top row, SNR=20), M1, M3 and M4 provided more accurate estimation for fs (1.6%, 1.3% and 0.1% error) and (0.9%, 0.1% and −2.2% error) compared to that of M2. M2 provided biased estimation for fs (5.4% error) and T*2,s (4.3% error) and more estimation uncertainty (i.e. less precision) for both parameters indicated by a much broader spectrum. M3 (11.8% error) and M4 (−0.1% error) provided better estimation accuracy of T*2, l, M3 than M1 (44.6% error) and M2 (21.4% error) at low SNR. In addition, the peak at 100ms in the estimation of T*2, l indicates that all of the four fitting methods yield small amount of estimation outliers at low SNR. At moderate SNR (middle row, SNR=50), all four fitting methods provided accurate estimation for fs (0.4%, 0.6%, 0.1% and 0.1% error), T*2,s (0.5%, 0.4%, 0.1% and 0.1% error), and T*2, l (6.2%, −0.3%, 0.4% and 0.6% error), while M2 provided slightly larger estimation uncertainty. At high SNR (bottom row, SNR=90), all four fitting methods provided accurate estimation for fs (0.1%, 0.2%, 0% and 0% error), T*2,s (0.1%, 0.1%, −0.1% and 0% error) and T*2, l (1.7%, −0.5%, 0% and 0% error), while M2 provided slightly larger estimation uncertainty. Overall, the accuracy and precision of bi-component T2* parameter estimation increased with increasing level of SNR for all four fitting methods.
Figure 1.
Histogram of bi-component T2* parameter estimation in Monte-Carlo simulation at different SNR levels (Top row: SNR=20, Middle row: SNR=50, Bottom row: SNR=90). The dark arrows indicate the ground truth value for each individual parameter.
Figure 2 shows sagittal images at all 16 echoes acquired using the 3D cone UTE sequence. The mean SNR of the patellar tendon for all subjects was 54.5 with a standard deviation of 7.6 and a range between 41.3 and 70.4. There was a monotonic decay of the MR signal of the patellar tendon with much stronger decay occurring prior to 4.3ms than after 4.3ms. Other tissues including cartilage, bone marrow fat, and subcutaneous fat also demonstrated signal decay from the short echoes to the long echoes.
Figure 2.
Sagittal UTE images from the central portion of the patellar tendon of a 31-years-old healthy male volunteer. The white dash line delineates the region of patellar tendon after manual segmentation. The series of images demonstrate signal decay within the patellar tendon at different TEs from the shortest echo (TE=30μs) to the longest echo (TE=30ms).
Figure 3A shows the mean signal from a ROI in the central portion of the patellar tendon. All four fitting methods showed visually good fit. However, the overall goodness-of-fit was better (i.e. lower χ2 value) for M1(χ2=160), M3(χ2=140) and M4(χ2=145) than that of M2(χ2=389). Figure 3B demonstrates the relative residual errors in percentage for the four fitting methods. For all fitting methods, there was unbiased signal fitting prior to 5ms with no more than 5% residual error, while the fitting had relatively larger error and more uncertainty after 5ms. M1, M3 and M4 showed visually almost identical fitting quality with small relative residual errors at all echoes. M2 showed similar fitting quality as the other three methods prior to 5ms but slightly larger residual errors at the later echoes. Table I shows the mean and percentage of outliers of fs, T*2,s, and T*2, l of the patellar tendon of all subjects for all four fitting methods. M2 had the largest (13.53%) and M1 the smallest (6.05%) mean percentage of outliers. There was a significant difference between the four fitting methods for the mean fs (p = 0.001), T*2,s (p = 0.04), and T*2, l (p = 0.004) of the patellar tendon. M1 provided significantly higher mean fs (p<0.05) and T*2, l (p<0.05) than M3 and M4. M3 provided significantly higher mean fs (p<0.05) and T*2, l (p<0.05) than M4, while M2 provided significantly higher of T*2,s (p<0.05) than M1, M3 and M4.
Figure 3.

(a) The average signal from an ROI (white circle) in the patellar tendon shows a monotonically signal decay from 30μs to 30ms. Rapid signal decay is shown at TEs prior to 5ms followed by slow signal decay. All four model fitting methods show good fit to the data. (b) Fitting methods M1, M3 and M4 show almost identical fit quality with small relative residual errors. Fitting method M2 shows similar fit quality as the other three methods at TEs prior to 5ms, but slightly larger residual errors at later echoes.
Table I.
Mean and standard deviation and mean percentage of outliers of fs, T*2,s, and T*2,l of the patellar tendon of all 10 subjects for all four fitting methods.
| Outliers [%] | fs [%] | T*2,s [ms] | T*2,l [ms] | ||||
|---|---|---|---|---|---|---|---|
|
| |||||||
| mean | mean | SD | mean | SD | mean | SD | |
| M1 | 6.05 | 82.45 | 10.08 | 1.54 | 0.56 | 31.21 | 22.81 |
| M2 | 13.53 | 82.49 | 11.35 | 1.68 | 0.69 | 29.96 | 26.15 |
| M3 | 9.17 | 81.88 | 10.14 | 1.54 | 0.56 | 28.42 | 18.13 |
| M4 | 9.31 | 81.72 | 10.17 | 1.54 | 0.56 | 28.11 | 18.10 |
Table I also shows the standard deviations of fs, T*2,s, and T*2, l of the patellar tendon of all subjects for all four fitting methods. There was a significant difference between the four fitting methods for the standard deviation of T*2,s (p < 0.001), T*2, l (p < 0.0001), and fs (p < 0.001) of the patellar tendon. M2 provided significantly higher standard deviation of fs (p<0.05), T*2,s (p<0.05), and T*2, l (p<0.05) than M1, M3 and M4. M1 provided significantly higher standard deviation of T*2, l (p<0.05) than M3 and M4.
The SNR of the patellar tendon was 84.2, 48.3, and 21.3 for the SNRhigh, SNRnormal, and SNRlow images respectively. Figure 4, 5, and 6 show fs, T*2,s, and T*2, l maps of the central portion of the patellar tendon in one subject at different SNR levels. Tables 2 and 3 show the mean and standard deviations of fs, T*2,s and T*2, l of all patellar tendon voxels in one subject at different SNR levels. There was a dramatic increase of mean T*2, l with decrease in image SNR, while fs and T*2,s remained relatively the same. There was an increase in the standard deviation of fs, T*2,s, and T*2, l with decrease in image SNR. The standard deviation of fs, T*2,s, and T*2, l was much larger for M2 than for M1, M3 and M4.
Figure 4.
Sagittal fs (%) maps of the central portion of the patellar tendon in a 29-year-old healthy female volunteer show the noise performance at three different SNR levels for the four different fitting methods.
Figure 5.
A sagittal T*2,l (ms) maps of the central portion of the patellar tendon in a 29-year-old healthy female volunteer show the noise performance at three different SNR levels for the four different fitting methods.
Figure 6.
A sagittal T*2,l (ms) maps of the central portion of the patellar tendon in a 29-year-old healthy female volunteer show the noise performance at three different SNR levels for the four different fitting methods.
Table II.
Mean, fs, T*2,s, and T*2,l of all patellar tendon voxels at different SNR levels in a 29 year old healthy female volunteer for all four fitting methods.
| fs [%] | T*2,s [ms] | T*2,l [ms] | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
| ||||||||||||
| M1 | M2 | M3 | M4 | M1 | M2 | M3 | M4 | M1 | M2 | M3 | M4 | |
| SNRhigh | 81.68 | 81.99 | 81.32 | 81.31 | 1.68 | 1.88 | 1.68 | 1.67 | 25.53 | 27.06 | 20.53 | 20.48 |
| SNRnormal | 80.18 | 82.53 | 80.06 | 80.70 | 1.85 | 1.92 | 1.85 | 1.83 | 25.90 | 29.70 | 24.82 | 24.39 |
| SNRlow | 80.26 | 81.20 | 80.79 | 80.53 | 1.70 | 1.81 | 1.71 | 1.70 | 62.42 | 56.27 | 56.33 | 54.25 |
Table III.
Standard deviations of, fs, T*2,s, and T*2,l of all patellar tendon voxels at different SNR levels in a 29 year old healthy female volunteer for all four fitting methods.
| fs [%] | T*2,s [ms] | T*2,l [ms] | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
| ||||||||||||
| M1 | M2 | M3 | M4 | M1 | M2 | M3 | M4 | M1 | M2 | M3 | M4 | |
| SNRhigh | 10.58 | 11.21 | 10.61 | 10.60 | 0.47 | 0.62 | 0.47 | 0.47 | 19.21 | 28.32 | 18.28 | 18.46 |
| SNRnormal | 11.18 | 12.97 | 11.27 | 11.32 | 0.66 | 0.71 | 0.66 | 0.66 | 24.28 | 33.40 | 23.77 | 23.41 |
| SNRlow | 15.24 | 16.32 | 15.51 | 15.64 | 0.73 | 0.76 | 0.73 | 0.73 | 29.34 | 39.67 | 29.96 | 30.28 |
Discussion
Various imaging methods including ultrasound and conventional MR imaging have been used to evaluate tendon 36. However, tendon and other musculoskeletal tissues containing abundant highly organized collagen fibers are difficult to assess using quantitative MR techniques due to their extremely rapid signal decay. New MR techniques including ultra-short echo-time (UTE) imaging 10,37,38, variable echo time (VTE) imaging 25,39 and zero echo-time (ZTE) imaging 40,41 have been recently developed to capture the rapidly decaying signal within musculoskeletal tissues. The UTE sequence used in our study utilized a center-out twisted 3D cone k-space trajectory which has been shown to increase sampling uniformity, decrease undersampling artifact, and provide better SNR performance when compared to traditional 2D and 3D radial UTE sequences33. The sequence has had successful applications in sodium imaging42,43 and ex-vivo Achilles tendon imaging24. In our study, the 3D cone UTE sequence provided complete anatomic coverage of the patellar tendon in human subjects at 3.0T with clinically feasible spatial resolutions and scan times and with an SNR of 54.5, which has been shown to be adequate to accurately estimate bi-component T2* parameters from previous literature 28 and the results of our simulation study.
Recent studies have demonstrated that bi-component exponential signal models are better suited than single-component models for evaluating tendon19,23–25. Bi-component models have detected two distinct T2 components within tendon representing fast relaxing water bound to the highly organized collagen fibers and slow relaxing bulk water. The T2* of the two water components in tendon have varied in different studies between 0.3 ms and 1.3 ms for the short component and 8.2ms and 20.4ms for the long component 19,23–25. The differences in measured T2* parameters is likely due to multiple factors including different field strengths, vender platforms, sequence designs, imaging parameters, imaging conditions, and tendon types. Our study reported a short T2* of around 1.5ms and a long T2* of around 28 ms which is close to the values of 0.9 ms and 20.4 ms reported by Chang et al.24. This may be due to the fact that both studies used similar 3D Cone UTE sequences although the sequence used by Chang et al. to evaluate ex-vivo Achilles tendon in human cadavers had a scan time of 60 minutes which was much longer than the 12 minute scan time of our sequence.
Bi-component T2* analysis has been shown to provide better diagnostic performance than single-component T2* analysis for distinguishing between patients with Achilles tendinopathy and asymptomatic volunteers 44. Furthermore, previous studies have found a significant increase in short T2* in pathologic tendon when compared to normal tendon, which is specifically due to disruption of the highly organized collagen fiber network, but no significant difference in long T2* 25,39. Accurate estimations of bi-component T2* parameters requires sufficient SNR and proper noise modeling which is particularly important when estimating pixel-by-pixel spatial variations across a tissue sample 28. Both our simulation and in-vivo studies showed that T*2, l in tendon has much greater spatial variation, as indicated by the standard deviation of image voxels, and much higher sensitivity to noise when compared to T*2,s and fs. Thus, the diagnostic performance of T*2, l for detecting tendon pathology could be substantially reduced by both low image SNR and by large spatial variations (i.e. uncertainty) in the setting of normal or high image SNR21. One explanation for the large uncertainty and high sensitivity to noise of T*2, l is the relatively small fraction of the long T2* component in tendon which provides little MR signal at later echoes. More accurate estimations of T*2, l may require a larger number of later acquired echoes, use of multiple excitations, or further optimization of echo number and echo spacing. However, our results in both simulation and in-vivo study also showed that fs and T*2,s measurements are relatively insensitive to image SNR which suggests that they may serve as robust MR parameters for evaluating tendon composition and microstructure. Our study has demonstrated that bi-component T2* measurements are influenced by the method used for curve fitting. Previous studies performed by Karlsen et al.45 and Raya et al.46 have described methods for single-component T2 analysis at low image SNR with incorporation of the Rician noise distribution. Recently, a simulation study performed by Bouhrara et al. extended the work to bi-component T2* analysis where comparison of noise performance among different fitting methods was conducted using Monte-Carlo and Cramer-Rao lower bound (CRLB) simulations28. Their results showed that the M1 and M2 fitting methods provided higher estimation bias and greater estimation uncertainty than the M3 and M4 fitting methods. Our numerical simulation and in-vivo study evaluating the patellar tendon in human subjects showed similar findings with the M2 fitting method having the greatest uncertainty which could be attributed to the fact that M2 yields neither the Rician nor Gaussian noise distribution at low SNR30,45. The M3 and M4 fitting methods in our study provided nearly identical performance at all SNR conditions which agrees with the simulation results from Bouhrara et al.28 and Karlsen et al.45. Thus, proper modeling of image noise may improve the accuracy of bi-component T2* measurements, especially for the long T2* component which is most severely affected by image noise. Since the M3 and M4 fitting methods are almost equally effective and M3 is less computational expensive, our general recommendation is that the M3 fitting methods should be used for in-vivo bi-component T2* analysis of tendon in human subjects.
Our study has several limitations. First, our study focused only on the patellar tendon and included only a small number of healthy volunteers. In addition, statistical analysis could not be performed to more thoroughly evaluate the influence of image SNR on bi-component T2* measurements. Although we attempted to acquire SNRhigh scans on multiple subjects to perform statistical analysis, only one individual was able to tolerate the more than 45 minute four excitation 3D Cones UTE sequence without significant motion artifact. In addition, our study did not use phantoms to investigate bias of the different fitting methods relative to true parameter values or clinical patients with patellar tendinopathy to investigate whether the choice of fitting method and image SNR influences the diagnostic performance for distinguishing between normal and pathologic tendon.
In conclusion, our study has demonstrated the use of a 3D cone UTE sequence to perform in-vivo voxel-by-voxel bi-component T2* analysis of the patellar tendon in human subjects at 3.0T with clinically feasible spatial resolutions and scan times. Our simulation and in vivo results have shown that incorporating Rician noise is helpful for improving bi-component T2* analysis of tendon especially at lower SNR. However, additional studies in patients with patellar tendinopathy and other tendon disorders are needed to determine whether bi-component T2* analysis can be used to detect tendon pathology and investigate the influence of the fitting method and image SNR on the identification and quantification of tendon disease.
Footnotes
Conflict of interest
The Author have no financial or personal relationships with other people or organizations that could inappropriately influence their work.
References
- 1.Dunn TC, Lu Y, Jin H, Ries MD, Majumdar S. T2 relaxation time of cartilage at MR imaging: comparison with severity of knee osteoarthritis. Radiology. 2004;232(2):592–598. doi: 10.1148/radiol.2322030976. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Patten C, Meyer RA, Fleckenstein JL. T2 mapping of muscle. Semin Musculoskelet Radiol. 2003;7(4):297–305. doi: 10.1055/s-2004-815677. [DOI] [PubMed] [Google Scholar]
- 3.Baum T, Joseph GB, Karampinos DC, Jungmann PM, Link TM, Bauer JS. Cartilage and meniscal T2 relaxation time as non-invasive biomarker for knee osteoarthritis and cartilage repair procedures. Osteoarthritis Cartilage. 2013;21(10):1474–1484. doi: 10.1016/j.joca.2013.07.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Du J, Hermida JC, Diaz E, et al. Assessment of cortical bone with clinical and ultrashort echo time sequences. Magn Reson Med. 2013;70(3):697–704. doi: 10.1002/mrm.24497. [DOI] [PubMed] [Google Scholar]
- 5.Liess C, Lusse S, Karger N, Heller M, Gluer CC. Detection of changes in cartilage water content using MRI T2-mapping in vivo. Osteoarthritis and cartilage/OARS, Osteoarthritis Research Society. 2002;10(12):907–913. doi: 10.1053/joca.2002.0847. [DOI] [PubMed] [Google Scholar]
- 6.Watrin-Pinzano A, Ruaud JP, Olivier P, et al. Effect of proteoglycan depletion on T2 mapping in rat patellar cartilage. Radiology. 2005;234(1):162–170. doi: 10.1148/radiol.2341030394. [DOI] [PubMed] [Google Scholar]
- 7.Nishioka H, Hirose J, Nakamura E, et al. T1rho and T2 mapping reveal the in vivo extracellular matrix of articular cartilage. Journal of magnetic resonance imaging: JMRI. 2012;35(1):147–155. doi: 10.1002/jmri.22811. [DOI] [PubMed] [Google Scholar]
- 8.Mosher TJ, Smith H, Dardzinski BJ, Schmithorst VJ, Smith MB. MR imaging and T2 mapping of femoral cartilage: in vivo determination of the magic angle effect. AJR American journal of roentgenology. 2001;177(3):665–669. doi: 10.2214/ajr.177.3.1770665. [DOI] [PubMed] [Google Scholar]
- 9.Goodwin DW, Wadghiri YZ, Zhu H, Vinton CJ, Smith ED, Dunn JF. Macroscopic structure of articular cartilage of the tibial plateau: influence of a characteristic matrix architecture on MRI appearance. AJR American journal of roentgenology. 2004;182(2):311–318. doi: 10.2214/ajr.182.2.1820311. [DOI] [PubMed] [Google Scholar]
- 10.Du J, Diaz E, Carl M, Bae W, Chung CB, Bydder GM. Ultrashort echo time imaging with bicomponent analysis. Magnetic resonance in medicine: official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2012;67(3):645–649. doi: 10.1002/mrm.23047. [DOI] [PubMed] [Google Scholar]
- 11.Reiter DA, Lin PC, Fishbein KW, Spencer RG. Multicomponent T2 relaxation analysis in cartilage. Magnetic resonance in medicine: official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine. 2009;61(4):803–809. doi: 10.1002/mrm.21926. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Reiter DA, Roque RA, Lin PC, Doty SB, Pleshko N, Spencer RG. Improved specificity of cartilage matrix evaluation using multiexponential transverse relaxation analysis applied to pathomimetically degraded cartilage. NMR in biomedicine. 2011;24(10):1286–1294. doi: 10.1002/nbm.1690. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Liu F, Choi KW, Samsonov A, et al. Articular Cartilage of the Human Knee Joint: In Vivo Multicomponent T2 Analysis at 3.0 T. Radiology. 2015:142201. doi: 10.1148/radiol.2015142201. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Liu F, Chaudhary R, Hurley SA, et al. Rapid multicomponent T2 analysis of the articular cartilage of the human knee joint at 3.0T. J Magn Reson Imaging. 2014;39(5):1191–1197. doi: 10.1002/jmri.24290. [DOI] [PubMed] [Google Scholar]
- 15.Biswas R, Bae W, Diaz E, et al. Ultrashort echo time (UTE) imaging with bi-component analysis: bound and free water evaluation of bovine cortical bone subject to sequential drying. Bone. 2012;50(3):749–755. doi: 10.1016/j.bone.2011.11.029. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Bae WC, Chen PC, Chung CB, Masuda K, D’Lima D, Du J. Quantitative ultrashort echo time (UTE) MRI of human cortical bone: correlation with porosity and biomechanical properties. Journal of bone and mineral research: the official journal of the American Society for Bone and Mineral Research. 2012;27(4):848–857. doi: 10.1002/jbmr.1535. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Li S, Chang EY, Bae WC, et al. Ultrashort echo time bi-component analysis of cortical bone-a field dependence study. Magn Reson Med. 2014;71(3):1075–1081. doi: 10.1002/mrm.24769. [DOI] [PubMed] [Google Scholar]
- 18.Juras V, Apprich S, Zbýň S, et al. Quantitative MRI analysis of menisci using biexponential T2 * fitting with a variable echo time sequence. Magn Reson Med. 2013 doi: 10.1002/mrm.24760. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Diaz E, Chung CB, Bae WC, et al. Ultrashort echo time spectroscopic imaging (UTESI): an efficient method for quantifying bound and free water. NMR Biomed. 2012;25(1):161–168. doi: 10.1002/nbm.1728. [DOI] [PubMed] [Google Scholar]
- 20.Liu F, Samsonov A, Wilson JJ, Blankenbaker DG, Block WF, Kijowski R. Rapid in vivo multicomponent T2 mapping of human knee menisci. J Magn Reson Imaging. 2015 doi: 10.1002/jmri.24901. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Williams A, Qian Y, Bear D, Chu CR. Assessing degeneration of human articular cartilage with ultra-short echo time (UTE) T2* mapping. Osteoarthritis and cartilage/OARS, Osteoarthritis Research Society. 2010;18(4):539–546. doi: 10.1016/j.joca.2010.02.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Shao H, Chang EY, Pauli C, et al. UTE bi-component analysis of T2* relaxation in articular cartilage. Osteoarthritis and cartilage/OARS, Osteoarthritis Research Society. 2016;24(2):364–373. doi: 10.1016/j.joca.2015.08.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Chang EY, Du J, Statum S, Pauli C, Chung CB. Quantitative bi-component T2* analysis of histologically normal Achilles tendons. Muscles Ligaments Tendons J. 2015;5(2):58–62. [PMC free article] [PubMed] [Google Scholar]
- 24.Chang EY, Du J, Iwasaki K, et al. Single- and Bi-component T2* analysis of tendon before and during tensile loading, using UTE sequences. J Magn Reson Imaging. 2015;42(1):114–120. doi: 10.1002/jmri.24758. [DOI] [PubMed] [Google Scholar]
- 25.Juras V, Apprich S, Szomolanyi P, Bieri O, Deligianni X, Trattnig S. Bi-exponential T2 analysis of healthy and diseased Achilles tendons: an in vivo preliminary magnetic resonance study and correlation with clinical score. European radiology. 2013;23(10):2814–2822. doi: 10.1007/s00330-013-2897-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Zheng S, Xia Y. On the measurement of multi-component T2 relaxation in cartilage by MR spectroscopy and imaging. Magnetic resonance imaging. 2010;28(4):537–545. doi: 10.1016/j.mri.2009.12.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Wang N, Xia Y. Dependencies of multi-component T2 and T1ρ relaxation on the anisotropy of collagen fibrils in bovine nasal cartilage. J Magn Reson. 2011;212(1):124–132. doi: 10.1016/j.jmr.2011.06.031. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Bouhrara M, Reiter DA, Celik H, et al. Incorporation of rician noise in the analysis of biexponential transverse relaxation in cartilage using a multiple gradient echo sequence at 3 and 7 tesla. Magn Reson Med. 2015;73(1):352–366. doi: 10.1002/mrm.25111. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.McGibney G, Smith MR. An unbiased signal-to-noise ratio measure for magnetic resonance images. Medical physics. 1993;20(4):1077–1078. doi: 10.1118/1.597004. [DOI] [PubMed] [Google Scholar]
- 30.Gudbjartsson H, Patz S. The Rician distribution of noisy MRI data. Magn Reson Med. 1995;34(6):910–914. doi: 10.1002/mrm.1910340618. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Sijbers J, den Dekker AJ. Maximum likelihood estimation of signal amplitude and noise variance from MR data. Magnetic resonance in medicine: official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2004;51(3):586–594. doi: 10.1002/mrm.10728. [DOI] [PubMed] [Google Scholar]
- 32.Padulo J, Oliva F, Frizziero A, Maffulli N. Muscles, Ligaments and Tendons Journal - Basic principles and recommendations in clinical and field Science Research: 2016 Update. MLTJ. 2016;6(1):1–5. doi: 10.11138/mltj/2016.6.1.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Gurney PT, Hargreaves BA, Nishimura DG. Design and analysis of a practical 3D cones trajectory. Magn Reson Med. 2006;55(3):575–582. doi: 10.1002/mrm.20796. [DOI] [PubMed] [Google Scholar]
- 34.Klein S, Staring M, Murphy K, Viergever MA, Pluim JP. elastix: a toolbox for intensity-based medical image registration. IEEE Trans Med Imaging. 2010;29(1):196–205. doi: 10.1109/TMI.2009.2035616. [DOI] [PubMed] [Google Scholar]
- 35.Holm S. A Simple Sequentially Rejective Multiple Test Procedure. Scand J Stat. 1979;6(2):65–70. [Google Scholar]
- 36.Weinreb JH, Sheth C, Apostolakos J, et al. Tendon structure, disease, and imaging. Muscles Ligaments Tendons J. 2014;4(1):66–73. [PMC free article] [PubMed] [Google Scholar]
- 37.Du J, Carl M, Diaz E, et al. Ultrashort TE T1rho (UTE T1rho) imaging of the Achilles tendon and meniscus. Magn Reson Med. 2010;64(3):834–842. doi: 10.1002/mrm.22474. [DOI] [PubMed] [Google Scholar]
- 38.Du J, Bydder M, Takahashi AM, Carl M, Chung CB, Bydder GM. Short T2 contrast with three-dimensional ultrashort echo time imaging. Magnetic resonance imaging. 2011;29(4):470–482. doi: 10.1016/j.mri.2010.11.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Juras V, Zbyn S, Pressl C, et al. Regional variations of T2* in healthy and pathologic achilles tendon in vivo at 7 Tesla: preliminary results. Magn Reson Med. 2012;68(5):1607–1613. doi: 10.1002/mrm.24136. [DOI] [PubMed] [Google Scholar]
- 40.Weiger M, Brunner DO, Dietrich BE, Müller CF, Pruessmann KP. ZTE imaging in humans. Magn Reson Med. 2013;70(2):328–332. doi: 10.1002/mrm.24816. [DOI] [PubMed] [Google Scholar]
- 41.Larson PE, Han M, Krug R, et al. Ultrashort echo time and zero echo time MRI at 7T. MAGMA. 2015 doi: 10.1007/s10334-015-0509-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Staroswiecki E, Bangerter NK, Gurney PT, Grafendorfer T, Gold GE, Hargreaves BA. In vivo sodium imaging of human patellar cartilage with a 3D cones sequence at 3 T and 7 T. Journal of magnetic resonance imaging: JMRI. 2010;32(2):446–451. doi: 10.1002/jmri.22191. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Lu A, Atkinson IC, Claiborne TC, Damen FC, Thulborn KR. Quantitative sodium imaging with a flexible twisted projection pulse sequence. Magn Reson Med. 2010;63(6):1583–1593. doi: 10.1002/mrm.22381. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Grosse U, Syha R, Hein T, et al. Diagnostic value of T1 and T2 * relaxation times and off-resonance saturation effects in the evaluation of Achilles tendinopathy by MRI at 3T. Journal of magnetic resonance imaging: JMRI. 2015;41(4):964–973. doi: 10.1002/jmri.24657. [DOI] [PubMed] [Google Scholar]
- 45.Karlsen OT, Verhagen R, Bovée WM. Parameter estimation from Rician-distributed data sets using a maximum likelihood estimator: application to T1 and perfusion measurements. Magn Reson Med. 1999;41(3):614–623. doi: 10.1002/(sici)1522-2594(199903)41:3<614::aid-mrm26>3.0.co;2-1. [DOI] [PubMed] [Google Scholar]
- 46.Raya JG, Dietrich O, Horng A, Weber J, Reiser MF, Glaser C. T2 measurement in articular cartilage: impact of the fitting method on accuracy and precision at low SNR. Magnetic resonance in medicine: official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2010;63(1):181–193. doi: 10.1002/mrm.22178. [DOI] [PubMed] [Google Scholar]





