. 2016 Mar 23;44(3):405–418. doi: 10.1177/0300060515607386

Comparison of oropharyngeal leak pressure and clinical performance of LMA ProSeal™ and i-gel® in adults: Meta-analysis and systematic review

Hye Won Shin 1,, Hae Na Yoo 1, Go Eun Bae 1, Jun Chul Chang 1, Min Kyung Park 1, Hae Seun You 1, Hyun Jung Kim 2, Hyung Sik Ahn 2
PMCID: PMC5536706  PMID: 27009026

Abstract

Background

A meta-analysis and systematic review of randomized controlled trials to compare the oropharyngeal leak pressure (OLP) and clinical performance of LMA ProSeal™ (Teleflex® Inc., Wayne, PA, USA) and i-gel® (Intersurgical Ltd, Wokingham, UK) in adults undergoing general anesthesia.

Methods

Searches of MEDLINE®, EMBASE®, CENTRAL, KoreaMed and Google Scholar® were performed. The primary objective was to compare OLP; secondary objectives included comparison of clinical performance and complications.

Results

Fourteen RCTs were included. OLP was significantly higher with LMA ProSeal™ than with i-gel® (mean difference [MD] −2.95 cmH2O; 95% confidence interval [CI] −4.30, −1.60). The i-gel® had shorter device insertion time (MD −3.01 s; 95% CI −5.80, −0.21), and lower incidences of blood on device after removal (risk ratio [RR] 0.32; 95% CI 0.18, 0.56) and sore throat (RR 0.56; 95% CI 0.35, 0.89) than LMA ProSeal™.

Conclusion

LMA ProSeal™ provides superior airway sealing compared to i-gel®.

Keywords: Airway sealing, equipment, i-gel®, laryngeal mask airway proseal, leak, meta-analysis

Introduction

Use of supraglottic airway (SGA) devices is increasingly common in clinical anesthesia.1 Second-generation SGAs including LMA ProSeal™ (Teleflex® Inc., Wayne, PA, USA) and i-gel® (Intersurgical Ltd, Wokingham, UK) were introduced in 2000 and 2007, respectively. These devices provide better airway sealing characteristics than classic LMA™, have an additional drainage tube for stomach decompression to reduce the risk of pulmonary aspiration, and are designed for use with spontaneous or positive pressure ventilation (PPV).2

Oropharyngeal leak pressure (OLP), measured by closing the expiratory valve of the anesthetic circle system at a fixed gas flow rate and noting the equilibrium airway pressure, is used to quantify the efficacy of airway sealing in SGA devices.3 Importantly, OLP indicates airway protection, successful SGA placement, and PPV.3,4 Several methods are used to quantify OLP, including audible noise detection, oral capnography, stethoscopic noise and manometric stability.3,4

The clinical performance and safety of both LMA ProSeal™ and i-gel® have been studied extensively,519 but reports vary as to which device offers superior OLP. Studies have shown LMA ProSeal™ to have comparable OLP to i-gel®,5,9,11,12,14 or significantly higher7,8,10,13,15,17 or lower19 OLP than i-gel®.

The present meta-analysis of published randomized controlled trials (RCTs) was performed to compare the clinical performance and airway-sealing characteristics, including OLP, of LMA ProSeal™ and i-gel® in adult patients undergoing general anesthesia.

Materials and methods

This meta-analysis was performed based upon the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statements.20

Literature search

The electronic databases MEDLINE®, EMBASE®, CENTRAL (Cochrane Central Register of Controlled Trials) and KoreaMed, as well as the web search engine Google Scholar®, were searched for eligible studies. All searches were conducted in April 2014 and updated in December 2014. The Medical Subject Heading search terms and text words included ‘LMA ProSeal’, ‘ProSeal LMA’, ‘PLMA’, and ‘i-gel’. The search was performed across all languages. The title and abstract of each paper were screened by two reviewers (H.W.S. and H.J.K.) and potentially relevant references retrieved.

Study selection

Prospective RCTs that compared LMA ProSeal™ and i-gel® for general anesthesia in patients aged >18 years were included in the analysis. Studies were selected according to predetermined inclusion criteria by two independent reviewers (H.N.Y. and G.E.B.). Any discrepancies were resolved through discussion or consultation with a third independent investigator (H.S.A).

Data extraction

Data were extracted independently and in duplicate by two reviewers (G.E.B. and H.S.Y.) and were recorded using a predefined form that included: name of the first author; year of publication; total number of patients studied; OLP; time required for device insertion; rate of insertion on the first attempt without assistance; fiber-optic view of the glottis (glottis visualization); ease of gastric tube insertion; incidence of blood on the device after removal; and incidence of patient sore throat. The primary objective was to compare OLP between the two devices; secondary objectives were to compare their clinical performance and rate of complications. Attempts were made to contact the authors of studies that had insufficient or missing data; if attempts were unsuccessful, data were extrapolated from the study text or tables to obtain the target information.

Risk of bias assessment

The quality of the RCTs was independently assessed by two authors (H.N.Y. and M.K.P.) using the risk-of-bias tool in RevMan version 5.2 (The Cochrane Collaboration, London, UK). Quality was evaluated using the following potential sources of bias: random sequence generation; allocation concealment; blinding; incomplete outcome data; selective outcome reporting; other sources of bias. The methodology for each RCT was graded as ‘high,’ ‘low’ or ‘unclear’, to reflect either a high, low or uncertain risk of bias, respectively.

Statistical analyses

RevMan 5.2 software was used for statistical analyses. The mean difference (MD) with 95% confidence interval (CI) was computed for continuous variables; risk ratio (RR) with corresponding 95% CI was calculated for dichotomous outcome data. Statistical heterogeneity was estimated using the I2 statistic, which was deemed significant when I2 >50%. Due to the relatively small number of RCTs and the resulting clinical heterogeneity in our meta-analysis, the Mantel–Haenszel or inverse variance random effects model was used instead of the fixed effect model. In the absence of heterogeneity, a Mantel–Haenszel or inverse variance fixed effects model was used.21 Subgroup analysis for OLP was performed to determine the influence of the use of neuromuscular blocker (NMB; without or with NMB) and type of surgery (non-laparoscopic or laparoscopic). Sensitivity analysis was performed for OLP to evaluate the sequential effect of excluding studies. Subgroup analysis for device insertion time was performed according to the use of NMB (without or with NMB) and study publication year (2009–2012 or 2013–2014). Differences were considered statically significant if P < 0.05. Publication bias was assessed by visual inspection of funnel plots. If the funnel plot was visually asymmetrical, the Egger’s linear regression test was used.

Results

The initial electronic publication search identified 699 potential studies (267 from MEDLINE®, 282 from EMBASE®, 136 from CENTRAL, 12 from KoreaMed and 2 from Google Scholar®). After exclusions, the analysis included 14 RCTs518 published between 2009 and 2014, comprising 1104 patients (545 with LMA ProSeal™ and 559 with i-gel®). No records were obtained from ClinicalTrials.gov. The study selection strategy is shown in Figure 1.

Figure 1.

Figure 1.

Process for inclusion of randomized controlled trials in the meta-analysis to compare the oropharyngeal leak pressure, clinical performance and rate of complications of LMA ProSeal™ and i-gel®.

The studies included in this analysis originated from eight countries (Austria,13 Belgium,12 China,10 Germany,16 India,5,6,8,15,17,18 Japan,9 Republic of Korea11,14 and the UK7). Patients had undergone various modes of surgery, including laparoscopic,10,11,15,17 gynecological, orthopedic and ambulatory surgery. Methods used to evaluate OLP included audible noise,5,8,15,17 stethoscopic noise14,17 and manometric stability.79,1113,15 Intracuff LMA ProSeal™ pressures were maintained at 30 cmH2O6 or 60 cmH2O.5,713,15,16 The studies included spontaneously breathing anesthetized patients without the use of NMB5,7,9,12,13 and paralyzed anesthetized patients with the use of NMB8,10,11,14,15,17 during anesthesia. Details of studies included in the analysis are shown in Table 1.

Table 1.

Characteristics of randomized controlled trials comparing LMA ProSeal™ and i-gel® for oropharyngeal leak pressure, clinical performance and rate of complications.

First author, year n
Type of surgery Neuromuscular blocker Ventilation Outcome variables Cuff pressure cmH2O OLP measurement method
i-gel® LMA ProSeal™
Kini G, 20145 24 24 Elective short surgery (30–120 min) None Spontaneous Insertion time, effective seal, fiber-optic view, ease of gastric tube, sore throat 60 Audible noise
Das A, 20146 30 30 Elective day surgery. Excluded difficult airway and BMI > 35 kg/m2 Atracurium Controlled Hemodynamics, stress response, ease of insertion, insertion time, number of attempts, complications 30 Not checked
Bosley NJ, 20147 51 47 Elective surgery. Excluded difficult airway and BMI > 40 kg/m2 None Spontaneous and controlled in non-paralyzed patients Ease of insertion, insertion time, ventilator performance, leak pressure and compliance 60 Manometric stability
Chauhan G, 20138 40 40 Elective surgery. Excluded difficult airway and BMI > 25 kg/m2 Rocuronium Controlled Ease of insertion, insertion attempt, fiber-optic assessment, airway sealing pressure, ease of gastric tube placement, complications 60 Manometric stability, audible noise
Hayashi K, 20139 50 50 Elective surgery None Spontaneous Insertion time, success rate at first attempt, necessity of finger insertion, leak pressure, success rate of gastric tube placement, complications 60 Manometric stability
Shi YB, 201310 30 30 Elective laparoscopic gynecological surgery. Excluded difficult airway and BMI > 35 kg/m2 Rocuronium Controlled Insertion time, airway sealing pressure, complications 60 Not reported
Jeon WJ, 201211 15 15 Elective laparoscopic gynecological surgery. Excluded difficult airway and BMI > 35 kg/m2 Rocuronium Controlled Insertion time, leak pressure, number of attempts 60 Manometric stability
van Zundert TC, 201212 50 50 Elective peripheral or superficial surgery. Excluded difficult airway and BMI > 35 kg/m2 None Spontaneous Ease of insertion, anatomical position, OLP, change in OLP 60 Manometric stability, audible noise
Gasteiger L, 201013 75 76 Elective gynecological or orthopedic surgery. Excluded difficult airway and BMI > 35 kg/m2 None Spontaneous and controlled in non-paralyzed patients Insertion success rate, insertion time, OLP 60 Manometric stability
Shin WJ, 201014 64 53 Elective orthopedic surgery. Excluded difficult airway and BMI > 35 kg/m2 Rocuronium Controlled Hemodynamics, airway leak pressure, leak volume, success rate, complications Unclear Stethoscopic noise
Sharma B, 201015 30 30 Elective laparoscopic cholecystectomy. Excluded difficult airway and BMI > 35 kg/m2 Vecuronium Controlled Insertion time, easy insertion, gastric tube insertion, dynamic compliance, OLP, airway resistance, work of breathing, minute ventilation, fiber-optic positioning, adverse events. 60 Manometric stability, audible noise,
Heuer JF, 200916 40 40 Elective ambulatory surgery. Excluded difficult airway None Controlled Insertion time, easy insertion, tightness, patient comfort, respiratory morbidity 60 Not checked
Singh I, 200917 30 30 Elective orthopedic surgery and laparoscopic cholecystectomy. Excluded difficult airway Rocuronium Controlled Airway sealing pressure, ease of insertion, success rate of insertion, ease of gastric tube placement, complications. Unclear Manometric stability, audible noise, stethoscopic noise
Trivedi V, 200918 30 30 Elective surgery under general anesthesia Vecuronium Controlled Insertion time, Aldrete recovery score, complications Unclear Not checked

BMI, body mass index; OLP, oropharyngeal leak pressure.

All studies mentioned randomization, but only seven68,1113,15 included details of concealed allocation. However, the operator inserting the device and the OLP assessors were not blinded in any of the studies (due to the impossibility of blinding their use). Risk of bias in individual studies is summarized in Figure 2. There were no funnel asymmetries in OLP, time required for device insertion, insertion on the first attempt without assistance, fiber-optic view of the glottis, ease of gastric tube insertion, blood on device after removal or sore throat (data not shown).

Figure 2.

Figure 2.

Risk of bias for randomized controlled trials comparing oropharyngeal leak pressure, clinical performance and rate of complications of LMA ProSeal™ and i-gel®. The color version of this figure is available at: http://imr.sagepub.com.

Data from RCTs that quantified OLP5,715,17 indicated significantly lower OLP with i-gel® compared to LMA ProSeal™ (MD −2.95 cmH2O; I2= 71%; P < 0.0001) with high heterogeneity. Subgroup analyses revealed significantly lower OLP with i-gel® compared with LMA ProSeal™ with the use of NMB and laparoscopic surgery (P < 0.0001 and I2= 0% for both analyses; Figure 3A and Table 2). There were no between-subgroup differences in OLP with respect to use of NMB and type of surgery (Table 2). Sensitivity analyses revealed no interactions for OLP. There was no funnel plot asymmetry.

Figure 3.

Figure 3.

Forest plot comparing LMA ProSeal™ and i-gel® for (a) oropharyngeal leak pressure (cmH2O) stratified according to the use of neuromuscular blockade; (b) device insertion time (s) stratified according to publication year of studies. CI, confidence interval; I2, I-square heterogeneity statistic; IV, inverse variance.

Table 2.

Subgroup meta-analysis for oropharyngeal leak pressure with LMA ProSeal™ and i-gel®.

Subgroup differences
Oropharyngeal leak pressure MD 95% CI I2 Statistical significance I2 Statistical significance
Total −2.95 −4.30, −1.60 71% P < 0.0001
Without NMB −3.04 −6.31, 0.23 87% NS 0% NS
With NMB −2.84 −3.74, −1.97 0% P < 0.0001
Non-laparoscopic surgery −3.03 −5.04, −1.02 81% P < 0.0003 0% NS
Laparoscopic surgery −2.85 −4.17, −1.52 0% P < 0.0001

MD, mean difference; CI, confidence interval; I2, l-square heterogeneity statistic; NMB, neuromuscular blocker; NS, not statistically significant (P ≥ 0.05).

Device insertion time5,6,813,15,16,18 was significantly shorter for i-gel® than for LMA ProSeal™, with high heterogeneity (MD −3.01 s; I2= 97%; P = 0.03). Subgroup analysis indicated significantly shorter insertion time for i-gel® than for LMA ProSeal™ in studies published in 2013–2014, with sustained high heterogeneity (MD −6.20 s; I2= 96%; P < 0.00001; Figure 3B). Subgroup analyses revealed significant differences based on study publication year (P = 0.002) but not on use of NMB. There was no funnel plot asymmetry.

Blood on the device after removal510,12,14,15,17 and sore throat510,12,14,15,18 were significantly more common with LMA ProSeal™ than with i-gel® (for blood RR 0.32, I2= 0%, P < 0.0001; for sore throat RR 0.56, I2= 18%, P = 0.01; Figures 4A and 4B). There were no between-device differences with respect to insertion on the first attempt without assistance57,9,11-18, fiber-optic view of the glottis5,8,12,15 or ease of gastric tube insertion.810,15,17

Figure 4.

Figure 4.

Forest plot comparing LMA ProSeal™ and i-gel® for (a) blood on device after removal; (b) sore throat. CI, confidence interval; I2, I-square heterogeneity statistic; M–H, Mantel–Haenszel.

Discussion

The present meta-analysis indicated that i-gel® results in lower OLP, shorter insertion times, lower incidences of blood on device after removal, and sore throat, than LMA ProSeal™.

A potential risk of SGA use is incomplete airway sealing, which may cause gastric insufflation; inflation of airways at pressures above 20cmH2O can induce opening of the esophageal sphincter.22 Case reports have noted regurgitation and aspiration in patients with both LMA ProSeal™ and i-gel® during anesthesia.2325 However, a cadaver study reported fast drainage of esophageal fluid using SGAs with gastric channels.26

Airway sealing in SGA is characterized by OLP as assessed via an audible noise from the mouth or in the neck using a stethoscope, sampling of end-tidal carbon dioxide in the mouth or manometer equilibrium pressure at fixed fresh gas flow rates. OLP is also referred to as airway sealing pressure or airway leak pressure.3 All four OLP evaluation methods provide similar OLP values, with good correlation in children,4 and the manometric stability test has been shown to accurately measure OLP in adults.3 An airway sealing study using a cadaver aspiration model reported that the lack of an inflatable cuff may reduce the airway sealing ability of i-gel® compared with that of LMA ProSeal™.26

Other factors that may affect OLP include the use of NMB, intra-abdominal pressure during surgery and intracuff pressure of the SGA device.15,26,27 In our meta-analysis, the substantial overall heterogeneity (I2= 71%) was reduced by subgroup analysis based on NMB use (I2= 0%) and laparoscopic surgery (I2= 0%). Our findings suggest that OLP may be variable during surgery without NMB and non-laparoscopic surgery.

Device insertion time was shorter for i-gel® than for LMA ProSeal™ in the studies published in 2013–2014. There appears to be a preference for i-gel® over LMA ProSeal™,2,28 possibly due to the convenience of a disposable device, ease of insertion by stiff bite block and the natural oropharyngeal curvature of i-gel® compared with LMA ProSeal™.28 Device insertion time showed high heterogeneity after subgroup analysis with use of NMB and publication year; this was possibly due to differences in measurement standards among the studies included in our analysis.

It is possible that the fiber-optic view is better with i-gel® than with LMA ProSeal™ due to interference from folding of the LMA ProSeal™ cuff after insertion, but the absence of a between-group difference in this parameter suggests that both devices might function similarly as a conduit during airway management. The ease of gastric tube insertion was similar with each device in our review. The esophageal drain tube of i-gel® is smaller than that of LMA ProSeal™ (12 F versus 16 F for size 4, respectively).25 Correct SGA positioning is important to prevent gastric aspiration; the i-gel®, with its good positional stability, may be superior to LMA ProSeal™.28 The gastric channels of both devices allow early identification of regurgitation and prompt response to prevent aspiration.2,28 The inflated cuff of LMA ProSeal™ may contribute to the higher incidence of sore throat seen with this device compared with i-gel® (which has no cuff).

Meta-analyses comparing LMA Proseal™ and i-gel® have reported similar OLP for both devices.29,30 This is in contrast to our findings, which showed that LMA ProSeal™ provided higher OLP than i-gel®. This disparity may be due to differences in data collection. OLP is also referred to as airway sealing pressure and airway leak pressure.3,4 We included ‘OLP’, ‘airway sealing pressure’ and ‘airway leak pressure’ as search terms, but other studies searched only for ‘OLP’.29,30 Subgroup analysis for OLP including ‘OLP’, ‘oropharyngeal seal pressure’ and ‘airway sealing pressure’ as search terms found that second-generation LMAs (ProSeal™, Supreme™) had lower OLP than i-gel®.30 This partially incomplete search strategy would have omitted several studies that were included in the present meta-analysis.10,1618

There are many situations in which SGA devices are required to maintain high OLP against increased intra-abdominal pressure in laparoscopic surgery, obese patients and patients with restrictive lung disease. A meta-analysis of pediatric studies found higher OLP with i-gel® than with LMA ProSeal™.19 This contradictory finding may be explained by the lack of dorsal cuffs in sizes 1.5–2.5 for LMA ProSeal™.4,19 Anesthetists must weigh up the clinical performance and airway sealing safety of SGAs in clinical practice. The LMA ProSeal™ is regarded as a choice for airway sealing in adults that has a good safety profile, but i-gel® is preferred for pediatric procedures because it has a good safety profile in children.519

A limitation of this review is the clinical heterogeneity without power analysis or sample-size determination of the included studies. Other limitations are the performance and detection bias arising from the impossibility of blinding to device insertion, measurement of OLP and clinical performances.

In conclusion, our findings are that LMA ProSeal™ provides superior airway sealing (higher OLP) compared to i-gel®, while i-gel® offers rapid insertion time, and lower incidences of blood on the device after removal and sore throat compared to LMA ProSeal™ in anesthetized adult patients.

Declaration of conflicting interest

The authors declare that there are no conflicts of interest.

Funding

This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.

References

  • 1.Jolliffe L, Jackson I. Airway management in the outpatient setting: new devices and techniques. Curr Opin Anaesthesiol 2008; 21: 719–722. [DOI] [PubMed] [Google Scholar]
  • 2.Kim YH. Pulmonary aspiration associated with supraglottic airways: ProSeal laryngeal mask airway and I-Gel™. Korean J Anesthesiol 2012; 63: 489–490. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Keller C, Brimacombe JR, Keller K, et al. Comparison of four methods for assessing airway sealing pressure with the laryngeal mask airway in adult patients. Br J Anaesth 1999; 82: 286–287. [DOI] [PubMed] [Google Scholar]
  • 4.Lopez-Gil M, Brimacombe J, Keller C. A comparison of four methods for assessing oropharyngeal leak pressure with the laryngeal mask airway (LMA) in paediatric patients. Paediatr Anaesth 2001; 11: 319–321. [DOI] [PubMed] [Google Scholar]
  • 5.Kini G, Devanna GM, Mukkapati KR, et al. Comparison of I-gel with proseal LMA in adult patients undergoing elective surgical procedures under general anesthesia without paralysis: A prospective randomized study. J Anaesthesiol Clin Pharmacol 2014; 30: 183–187. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Das A, Majumdar S, Mukherjee A, et al. I-gel™ in ambulatory surgery: a comparison with LMA-ProSeal™ in paralyzed anaesthetized patients. J Clin Diagn Res 2014; 8: 80–84. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Bosley NJ, Burrows LA, Bhayani S, et al. A randomised comparison of the performance of ProSeal laryngeal mask airway with the i-gel for spontaneous and controlled ventilation during routine anaesthesia in European population. J Anesth Clin Res 2014; 5: 459–459. [Google Scholar]
  • 8.Chauhan G, Nayar P, Seth A, et al. Comparison of clinical performance of the I-gel with LMA ProSeal. J Anaesthesiol Clin Pharmacol 2013; 29: 56–60. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Hayashi K, Suzuki A, Kunisawa T, et al. A comparison of the single-use i-gel with the reusable laryngeal mask airway proseal in anesthetized adult patients in Japanese population. Masui 2013; 62: 134–139. [in Japanese, English Abstract]. [PubMed] [Google Scholar]
  • 10.Shi YB, Zuo MZ, Du XH, et al. Comparison of the efficacy of different types of laryngeal mask airways in patients undergoing laparoscopic gynecological surgery. Zhonghua Yi Xue Za Zhi 2013; 93: 1978–1980. [in Chinese, English Abstract]. [PubMed] [Google Scholar]
  • 11.Jeon WJ, Cho SY, Baek JS, et al. Comparison of the proseal LMA and intersurgical I-gel during gynecological laparoscopy. Korean J Anesthesiol 2012; 63: 510–514. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.van Zundert TC, Brimacombe JR. Similar oropharyngeal leak pressure during anaesthesia with i-gel™, LMA-ProSeal™ and LMA-Supreme™ laryngeal masks. Acta Anaesth Belg 2012; 63: 35–41. [PubMed] [Google Scholar]
  • 13.Gasteiger L, Brimacombe J, Perkhofer D, et al. Comparison of guided insertion of the LMA ProSeal™ vs the i-gel™. Anaesthesia 2010; 65: 913–916. [DOI] [PubMed] [Google Scholar]
  • 14.Shin WJ, Cheong YS, Yang HS, et al. The supraglottic airway I-gel in comparison with ProSeal larygngeal mask airway and classic laryngeal mask airway in anaesthetized patients. Eur J Anaesthesiol 2010; 27: 598–601. [DOI] [PubMed] [Google Scholar]
  • 15.Sharma B, Sehgal R, Sahai C, et al. PLMA vs. I-gel: a comparative evaluation of respiratory mechanics in laparoscopic cholecystectomy. J Anaesthesiol Clin Pharmacol 2010; 26: 451–457. [PMC free article] [PubMed] [Google Scholar]
  • 16.Heuer JF, Stiller M, Rathgeber J, et al. Evaluation of the new supraglottic airway devices Ambu auraonce and intersurgical i-gel Positioning, sealing, patient comfort and airway morbidity. Anaesthesist 2009; 58: 813–820. [in German, English Abstract]. [DOI] [PubMed] [Google Scholar]
  • 17.Singh I, Gupta M, Tandon M. Comparison of clinical performance of I-gel with LMA-Proseal in elective surgeries. Indian J Anaesth 2009; 53: 302–305. [PMC free article] [PubMed] [Google Scholar]
  • 18.Trivedi V, Patil B. A clinical comparative study of evaluation of proseal LMA V/S I-GEL for ease of insertion and hemodynamic stability; a study of 60 cases. Internet J Anesthesiol 2009; 27. [Google Scholar]
  • 19.Maitra S, Baidya DK, Bhattacharjee S, et al. Evaluation of i-gel(™) airway in children: a meta-analysis. Paediatr Anaesth 2014; 24: 1072–1079. [DOI] [PubMed] [Google Scholar]
  • 20.Liberati A, Altman DG, Tetzlaff J, et al. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: explanation and elaboration. PLoS Med 2009; 6: e1000100–e1000100. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Higgins JP and Green S. Cochrane handbook for systemic reviews of interventions. The Cochrane Collaboration 2011; Version 5.1.0.
  • 22.Devitt JH, Wenstone R, Noel AG, et al. The laryngeal mask airway and positive-pressure ventilation. Anesthesiology 1994; 80: 550–555. [DOI] [PubMed] [Google Scholar]
  • 23.Barker P, Langton JA, Murphy PJ, et al. Regurgitation of gastric contents during general anesthesia using the laryngeal mask airway. Br J Anaesth 1992; 69: 314–315. [DOI] [PubMed] [Google Scholar]
  • 24.Liew G, John B, Ahmed S. Aspiration recognition with an i-gel airway. Anaesthesia 2008; 63: 786–786. [DOI] [PubMed] [Google Scholar]
  • 25.Schmidbauer W, Bercker S, Volk T, et al. Oesophageal seal of the novel supralaryngeal airway device I-Gel in comparison with the laryngeal mask airways classic and ProSeal using a cadaver model. Br J Anaesth 2009; 102: 135–139. [DOI] [PubMed] [Google Scholar]
  • 26.Levitan RM, Kinkle WC. Initial anatomic investigations of the I-gel airway: a novel supraglottic airway without inflatable cuff. Anaesthesia 2005; 60: 1022–1026. [DOI] [PubMed] [Google Scholar]
  • 27.Goldmann K, Hoch N, Wulf H. Influence of neuromuscular blockade on the airway leak pressure of the ProSeal laryngeal mask airway. Anasthesiol Intensivmed Notfallmed Schmerzther 2006; 41: 228–232. [in German, English Abstract]. [DOI] [PubMed] [Google Scholar]
  • 28.Gibbison B, Cook TM, Seller C. Case series: protection from aspiration and failure of protection from aspiration with the i-gel airway. Br J Anaesth 2008; 100: 415–417. [DOI] [PubMed] [Google Scholar]
  • 29.Park SK, Choi GJ, Choi YS, et al. Comparison of the i-gel and the laryngeal mask airway proseal during general anesthesia: a systematic review and meta-analysis. PLoS One 2015; 10: e0119469–e0119469. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.de Montblanc J, Ruscio L, Mazoit JX, et al. A systematic review and meta-analysis of the i-gel(®) vs laryngeal mask airway in adults. Anaesthesia 2014; 69: 1151–1162. [DOI] [PubMed] [Google Scholar]

Articles from The Journal of International Medical Research are provided here courtesy of SAGE Publications