Abstract
Optical microscopy remains a fundamental tool for modern biological discovery owing to its excellent spatial resolution and versatile contrast in visualizing cellular and sub-cellular structures. Yet, the time domain is paramount for the observation of biological dynamics in living systems. Commonly, acquisition of microscopy data involves scanning of a spherically- or cylindrically-focused light beam across the imaged volume, which significantly limits temporal resolution in 3D. Additional complications arise from intense light scattering of biological tissues, further restraining the effective penetration depth and field of view of optical microscopy techniques. To overcome these limitations, we devised a fast optoacoustic micro-tomography (OMT) approach based on simultaneous acquisition of 3D image data with a high-density hemispherical ultrasound array having effective detection bandwidth beyond 25 MHz. We demonstrate fast three-dimensional imaging of freely-swimming zebrafish larvae, achieving 3D imaging speed of 100 volumes per second with isotropic spatial resolution approaching the dimensions of large cells across a field of view exceeding 50mm3. As opposed to other microscopy techniques based on optical contrast, OMT resolves optical absorption acoustically using unfocused light excitation. Thus, no penetration barriers are imposed by light scattering in deep tissues, suggesting it as a powerful approach for multi-scale functional and molecular imaging applications.
Introduction
Microscopy refers to a myriad of methods aimed at resolving structures invisible to the naked eye. The angular resolution of the human eye is estimated to be approximately 1 arcminute, i.e., points separated by less than ~75 μm cannot be distinguished via normal vision at the closest focusing distance (~25 cm)1. This resolution barrier can be readily overcome with different microscopic techniques based on electromagnetic radiation2, electron beams3 or ultrasound4. Of particular importance is optical microscopy, which is arguably the most widespread modality in biomedical research owing to its unique ability to non-invasively and dynamically interrogate in vivo biology with molecular- and cell-specific contrast. The recently introduced super-resolution techniques have further extended the spatial resolution beyond the optical diffraction limit5, 6. Nonetheless, light scattering still imposes hard limits on the penetration depth of optical microscopy into biological tissues7.
Optoacoustic (photoacoustic) imaging has emerged as a promising modality capable of high-resolution imaging beyond the penetration limits of optical microscopy8. Since optoacoustics can resolve optical absorption contrast acoustically, the spatial resolution typically scales with 1/200 of the imaging depth9, enabling microscopic imaging at depths beyond several millimeters. Acoustic-resolution optoacoustic microscopy is usually implemented analogously as acoustic microscopy, i.e., by raster-scanning a spherically-focused ultrasound transducer to form an image. The axial (depth) resolution in this case scales with the acoustic detection bandwidth10 and is ultimately limited by acoustic attenuation11. On the other hand, the lateral resolution is further conditioned by the numerical aperture of the transducer12–14. Numerous implementations have been reported that use point-by-point scanning of focused transducers with detection bandwidths from several MHz up to the GHz range, corresponding to acoustic spatial resolution down to micrometer scales15, 16.
The advantages of a raster-scan-based configuration stem from the ease of implementation. Multiple systems for high-resolution mapping of the optical absorption at diffuse light depths have been reported17–22. However, techniques employing focused ultrasound detectors are often afflicted by severe out-of-focus23 and limited-view24 imaging artifacts. The focal zone of the detector also restricts the field-of-view in the depth direction that can efficiently be covered with each scan. Most importantly, implementation of real-time imaging remains challenging with raster scanning approaches, in particular when it comes to three-dimensional (3D) image acquisition. Overall, it has been shown that accurate and fast optoacoustic imaging performance implies simultaneous tomographic acquisition of 3D data using broad angular (tomographic) coverage around the imaged object25. In this regard, sampling of data using dense spherical matrix array configurations represents a convenient acquisition geometry, which has demonstrated superb in vivo imaging performance in small animals and human studies26–30. However, implementation of high-frequency spherical arrays providing detection bandwidths beyond 10 MHz has remained a technological challenge.
Herein, we introduce a novel tomographic design that enables accurate real-time 3D imaging of optical absorption with microscopic resolution, approaching single cell dimensions at depths beyond the limits imposed by light scattering.
Results
Figure 1 shows the experimental concept of the fast optoacoustic micro-tomography (OMT) based on simultaneous acquisition of 3D image data with a high-density hemispherical ultrasound array having effective detection bandwidth beyond 25 MHz (see online methods for a detailed description). Such implementation is challenged by the need to arrange a large number of high-frequency ultrasound detection elements with sufficient solid angular coverage around the imaged sample, ideally exceeding 180°25. Our proposed matrix detection array was designed by arranging 510 piezocomposite elements with 1 mm diameter over 7 planar segments in a “flower” geometry (Fig. 1a). The individual segments were subsequently bent to form a hemispherical surface with 15 mm radius covering an angle of 180° (solid angle 2π). The central segment further has a 3.8 mm diameter aperture for delivery of the excitation light.
Basic resolution characterization was first done with an agar phantom containing ~20 µm diameter absorbing polyethylene microspheres. Figure 2a shows the maximum intensity projections (MIPs) along the z and y directions for the 3D imaging dataset corresponding to a single position of the array (‘hot’ colormap) as well as the entire raster-scanned area (‘bone’ colormap). One may see that the array can effectively render a 3D image from an approximate volume of 50 mm3 for each laser pulse. A zoom-in of a microsphere located approximately at the center of the spherical array is also displayed in Fig. 2a. The zoom-in region was reconstructed with a voxel size of 5 × 5 × 5 µm3. It is shown that the microsphere was reconstructed with a FWHM <40 µm along all three Cartesian dimensions (one-dimensional (1D) profiles for the x (red), y (blue) and z (green) directions are shown in Fig. 2a). This corresponds to an approximately isotropic resolution of <35 µm estimated as the square difference between the measured FWHM and the actual diameter of the microsphere.
It is also important to highlight that the reconstructed microspheres are not affected by limited-view artefacts due to the broad angular coverage of the array. Limited-view artefacts are often manifested as negative ‘shadows’ in the reconstructed absorbers, which ultimately affect the visibility of elongated structures24. As comparison, Fig. 2b shows the lateral cross-section of a reconstructed microsphere in the center of the array when all the 510 detection elements are considered (left, full 180° angular coverage) along with the equivalent image when only the 256 inner elements are taken (right, ~90° angular coverage). As expected, negative image values are much less prominent with the broad 180° angular coverage, although limited detection bandwidth of the transducer elements may further contribute to generation of the negative values in the images.
A zebrafish larva embedded in an agarose phantom was also imaged post mortem in the OMT system. Figure 3 shows the resulting images. Specifically, the top (MIP along the z direction) and lateral (MIP along the x direction) views are displayed in Fig. 3a. Major anatomical features are readily resolvable in the images. The high spatial resolution of the system further enables resolving single cells. Indeed, a closer look at the larva data color-coded for depth reveals individual melanophores containing highly-absorbing eumelanin, whose appearance correlates well with the corresponding bright-field optical microscopy images shown in Fig. 3b.
Although zebrafish larvae are optically transparent and thus, in principle, can be efficiently visualized by optical microscopy, OMT has clear advantage in terms of acoustic spatial resolution that is not affected by light scattering in deep tissues as well as the large volumetric field of view and the imaging speed in 3D. This can be better appreciated in Fig. 4, which demonstrates the four-dimensional (4D) visualization capability of a freely-swimming fish. Here, two sequences of four images separated by 10 ms are displayed. In the sequence shown in Fig. 4a the fish swims slowly and its movements can be easily tracked from frame to frame. Note that the system enables constant monitoring of the freely behaving animal in the entire volumetric region of interest determined by the effective field of view of the system. The second sequence (Fig. 4b) shows abrupt movements of the fish with inter-frame motion exceeding 1 mm. Since volumetric tracking at such a high frame rate and field of view is not possible with any existing optical microscopy approach, OMT offers new capacities for imaging biological dynamics with microscopic resolution. The fish motion is best visualized in a movie file available in the on-line version of the journal (Supplementary Video 1).
The deep-tissue imaging capabilities of OMT are further illustrated in Fig. 5, where the same hemispherical array was used for rapid 3D scanning of large-scale microvasculature networks in a living athymic nude-Foxn1nu mouse. Specifically, the left lateral side in the abdominal region of the mouse was imaged. Figure 5 displays the MIP images along the z and x directions respectively. Both images are color-coded for depth. It is shown that vascular structures located at least 4 mm deep into highly absorbing and scattering mammalian tissues can be visualized with the suggested OMT approach. The broad 180° tomographic coverage of the hemispherical array averts image artifacts associated with limited-view acquisition geometries24, thus enabling clear visualization of the descending microvasculature propagating at steep angles with respect to the skin surface (Fig. 5, right).
Discussion and Conclusions
The presented results suggest that the newly developed optoacoustic micro-tomography (OMT) method can acoustically resolve optical absorption contrast with microscopic resolution at a high volumetric frame rate, i.e. in 4D. As opposed to other optical microscopy methods that employ cylindrically or spherically focused light beams, light scattering does not impose penetration barriers for OMT. Indeed, OMT uses unfocused light illumination to image optical absorption contrast with acoustic resolution. This enables the acquisition of 3D tomographic data from the entire imaged volume for each laser pulse, resulting in large fields of view exceeding 50 mm3 and volumetric imaging frame rates of 100 Hz, an unprecedented performance among other microscopy methods. OMT may further be extended into imaging spectroscopic dimension by using light at different wavelengths, thus offering five dimensional (5D, real-time multispectral three-dimensional) imaging capabilities31.
The presented novel implementation represents an important technological step in terms of the ultrasound sensing performance. The high angular aperture (180°) of the spherical array guarantees broad tomographic coverage, leading to more accurate reconstructions without limited-view artefacts commonly present in systems with lower angular coverage24, 25. In addition, previous designs of volumetric optoacoustic tomography have employed spherical array designs at significantly lower frequency range, which resulted in an inferior spatial resolution in the order of 200 µm32. These important novel features have enabled here the visualization of microvascular networks, including descending vascular structures, with ~35 μm resolution and penetration depth of 4 mm, thus opening new prospects for imaging of dynamic microcirculation and neovascularization.
Improving the resolution and sensitivity appears to be an important next step for this technology, which would necessitate the development of ultrasonic arrays with larger detection bandwidth. Increasing the temporal resolution can readily be accomplished by employing pulsed lasers with higher repetition frequencies, yet would also imply significantly higher data rates to be matched with faster digital acquisition electronics. On the one hand, increasing the number of ultrasound sensing elements and their detection bandwidth would naturally result in larger fields of view and/or better spatial resolution. However, such developments would again necessitate speeding up data acquisition/processing capacities or otherwise development of sophisticated sparse signal representation and compressed sensing approaches33–35. In addition, tomographic reconstruction algorithms must be improved to account for acoustic mismatches and attenuation11, 36 as well as for the frequency response of ultrasound transducers37, 38.
Indeed, OMT brings the powerful capabilities of 4D and 5D optoacoustics much closer to the single cell dimensions, which may enable a wealth of new applications in imaging of multi-scale biological dynamics. One promising direction is the study of freely behaving animals and organs whose fast motion precludes the application of conventional imaging and microscopy methods lacking fast 3D imaging capabilities in large tissue volumes. Note that the typical fields of view of state-of-the-art intravital microscopy methods are limited to volumes in the 1 mm3 range39, 40. Another potential biomedical application that may benefit from the newly discovered OMT capabilities is neuroimaging. The recently showcased imaging of genetically encoded calcium indicators (GEGIs) with optoacoustics has set the stage for direct measurements of large-scale neural activity with high spatio-temporal resolution41. OMT is ideal for those studies as it can differentiate individual cells with very high temporal resolution, which is sufficient for detecting calcium transients or, potentially, even faster voltage signaling using voltage-sensitive indicators42, 43.
In conclusion, the optoacoustic micro-tomography approach introduced in this work is poised to become an important new tool for biomedical research as it brings a new level of spatiotemporal resolution and imaging depth performance not attainable with other optical microscopy modalities.
Online Methods
Manufacture of the high-frequency hemi-spherical matrix detection array
The current work aims at devising a new generation of optoacoustic micro-tomography (OMT) system capable of rendering high-resolution (microscopic) 3D reconstructions in real time while also minimizing image artifacts associated with the limited-view tomographic coverage. Such implementation is challenged by the need to arrange a large number of high-frequency ultrasound detection elements with sufficient solid angular coverage around the imaged sample, ideally >180°25. In addition, real-time imaging prohibits signal averaging, which imposes demanding requirements on the detection sensitivity of piezoelectric sensing elements (proportional to their size). The proposed matrix detection array was custom-made by Sonaxis SAS (Besancon, France) by arranging 510 piezocomposite elements with 1 mm diameter over 7 planar segments in a “flower” geometry (Fig. 1a). The individual segments were subsequently bent to form a hemispherical surface with 15 mm radius covering an angle of 180° (solid angle 2π). The central segment has a 3.8 mm diameter aperture for light delivery. The electrodes of the piezoelectric elements were sputtered after shaping the segments and the transducer was eventually coated with a thin layer of gold for noise reduction and also to provide a reflecting surface enabling redirecting the scattered light back onto the sample surface.
Optoacoustic microtomography (OMT) set-up
The entire OMT imaging system is further illustrated in Fig. 1b. Optical excitation is provided with a short-pulsed (<10 ns) optical parametric oscillator (OPO)-based laser (Innolas Laser GmbH, Krailling, Germany) guided via a custom-made 3.8 mm fiber bundle through a central cavity of the transducer array. In this way, an unfocused Gaussian excitation beam with an approximate diameter of 6 mm (at full width at half maximum (FWHM)) is created on the surface of the imaged object placed near the geometrical focus of the spherical array geometry. A high-speed parallel acquisition electronic system that can simultaneously provide time-resolved sampling of all the 510 channels at 125 megasamples per second was further designed (Falkenstein Mikrosysteme GmbH, Taufkirchen, Germany). For each laser pulse, the data from all the channels is transmitted to a personal computer via 1 Gbit/s Ethernet interface for further processing and real-time three dimensional (3D) image reconstruction using a filtered 3D back-projection algorithm44. In this way, the volumetric image acquisition is effectively limited to 100 Hz by the pulse repetition rate of the laser. The setup further allows for a raster-scanning of the transducer array in the x-y plane in order to cover larger regions at the expense of temporal resolution (Fig. 1b).
Calibration of the tomographic detection positions
As the system was designed to provide tomographic reconstructions with spatial resolution in the 30 µm range, location of the individual detection elements in 3D needs to be known precisely. While accurate angular location was ensured by sputtering of the electrodes through a mask to the bent segments, the radial positions of the elements cannot be guaranteed, particularly for an array composed of multiple segments. In fact, any radial inaccuracies in positioning generally have a much larger influence in the achievable resolution than errors in the lateral position of the elements, as depicted in Fig. 1c. For instance, when reconstructing optoacoustic images with the regular back-projection algorithm, a radial uncertainty Δr = 30 µm in the detection location position would readily result in a Δx r = 30 µm error in the position of the back-projected arc for a given instant t (radius ct, being c the speed of sound), which corresponds to the error in the position of the reconstructed absorbers and a consequent resolution deterioration. On the other hand, the same position uncertainty in the angular direction (angular uncertanity of Δϕ.~ 0.1° for the given radius of the spherical array R = 15 mm) would only result in a maximum error of Δx ϕ ~ 5 µm for the effective radius of the imaged field of view of approximately R FV ~ 2.5 mm (dotted circle in Fig. 1c).
In order to achieve the best radial position accuracy, calibration was performed with two alternative approaches (Fig. 1d and e). In the first approach, a 13 mm diameter sphere made of steel was positioned at the geometrical center of the array and pulse-echo measurements of the time-of-flight were performed for each element. In the second approach, a ~20 μm diameter polyethylene microsphere (Cospheric BKPMS 20–27 µm) located at the center of the array was illuminated with <10 ns duration laser pulses and the time-of-flight of the optoacoustic responses was measured for each detection element. In both cases, the array was immersed in water, whose temperature was constantly kept at 22 °C during the measurements. The radial position of the elements was then estimated from the time-of-flight measurements by considering the known speed of sound in water for the given temperature45, 46.
Resolution characterization
In order to characterize the resolution and the field of view of the system, a clear agar phantom containing ~20 µm diameter absorbing polyethylene microspheres (Cospheric BKPMS 20–27 µm) was imaged. The wavelength of the laser was set to 720 nm, corresponding to the maximum energy per pulse of 25 mJ within the tunable wavelength range. The microspheres were placed approximately around the center of the spherical array geometry. The array was then raster-scanned along an area of 8 × 8 mm2 with a step of 0.5 mm in the x and y directions in order to acquire signals from microspheres distributed across the entire phantom. For each position of the detection array, a 3D image corresponding to a volume of 3 × 3 × 2 mm3 (300 × 300 × 200 voxels) was tomographically reconstructed with a graphics processing unit (GPU)-based 3D back-projection algorithm44, 47. Prior to reconstruction, the optoacoustic signals were band-pass filtered with cut-off frequencies between 1 and 35 MHz. The resulting images for different positions of the array were superimposed (added up) to render a combined image of a larger region. The field of view was estimated as the volume for which the amplitude of the reconstructed spheres for a given position of the transducer is higher than 50% of the maximum amplitude achieved for a sphere located in the center of the field of view. The field of view estimated in this manner corresponded to an approximately spherical (isotropic) volume of 2 mm radius.
Biological imaging experiments
The performance of OMT for imaging biological specimens was further tested by 3D recordings of a freely-swimming zebrafish larva and a mouse vasculature in vivo. For this, the per-pulse energy density in the imaged region was kept at approximately 20 mJ/cm2. All procedures involving animal care and experimentation were conducted in full compliance with the institutional guidelines of the Helmholtz Center Munich and with approval from the Government District of Upper Bavaria.
First, we tested the anatomical imaging capacity of the system. For this, a 6 days-post-fertilization (dpf) wild-type zebrafish larvae was euthanized with 250 mg/l of MS-222 and embedded into a low-melting-point agarose phantom. Since the temporal imaging resolution does not play a crucial role when imaging specimens post mortem, the spherical array was raster-scanned along an area of 5 × 2 mm2 with a step of 0.5 mm along the x and y directions in order to acquire tomographic data from a larger region. The laser wavelength was set to 720 nm. The acquired signals were also averaged 100 times to improve the signal-to-noise ratio (SNR). Image reconstruction of the scanned area was performed with the same procedure described in section 2D. A three-dimensional median filter with kernel size 3 × 3 × 3 was eventually applied to the reconstructed images.
In the second experiment, two freely-swimming zebrafish larvae were imaged by exploiting the high-frame-rate imaging performance of OMT in 3D. For this, the array was oriented upwards and the entire hemispherical cap detector was filled with agar to guarantee acoustic coupling. The agar block contained a 4 mm diameter cavity filled with water located approximately at the center of the hemispherical array geometry, where the larvae were allowed to swim. The detection array remained stationary during the experiment and the imaging was performed at the highest pulse repetition rate of the laser (100 Hz) without employing signal averaging. The laser wavelength was set to 720 nm. For each laser pulse, a cubic image volume of 64 mm3 (180 × 180 × 180 voxels) was reconstructed with the same 3D back-projection algorithm. The optoacoustic signals in this case were band-pass filtering with cut-off frequencies between 1 and 25 MHz for noise reduction. A 3D median filter with kernel size 3 × 3 × 3 was similarly applied to the images.
Finally, a twelve-week-old female athymic nude-Foxn1nu mouse (Harlan Laboratories LTD, Itingen, Switzerland) was imaged to test the deep-tissue imaging capabilities of OMT in mammalian tissues. The mouse was fixed to a holder and immersed in a temperature-controlled water tank, as described elsewhere48. The water temperature was maintained at 32 °C during the experiment. The laser wavelength was set to 800 nm. The spherical array was raster-scanned along an area of 5.5 × 3 mm2 with a step of 0.5 mm along the x and y directions. The acquired signals were also averaged 100 times to improve the signal-to-noise ratio (SNR). Image reconstruction of the scanned area was performed with the same procedure described in section 2D. In this case, a cubic volume of 96 mm3 (160 × 160 × 240 voxels) was reconstructed for each laser pulse. A three-dimensional median filter with kernel size 3 × 3 × 3 was applied to the reconstructed images, which were further normalized with an exponential function to correct for light attenuation49.
Electronic supplementary material
Acknowledgements
The authors acknowledge support from the European Research Council Consolidator grant ERC-2015-CoG-682379 and the US National Institutes of Health grant R21-EY026382.
Author Contributions
X.L.D.B. and D.R. designed the optoacoustic imaging system. H.L.S. provided the animal models. X.L.D.B. performed the experiments and analyzed the data. D.R. supervised the study. All authors discussed the results and contributed to writing the manuscript.
Competing Interests
The authors declare that they have no competing interests.
Footnotes
Electronic supplementary material
Supplementary information accompanies this paper at doi:10.1038/s41598-017-06554-9
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Yanoff, M. M. D. & Duker, J. S., MDt. Ophthalmology (Elsevier, 2012).
- 2.Mertz, J. Introduction to optical microscopy. Vol. 138 (Roberts, 2010).
- 3.Egerton, R. F. Physical principles of electron microscopy: an introduction to TEM, SEM, and AEM. (Springer Science & Business Media, 2006).
- 4.Maev, R. G. Acoustic microscopy: Fundamentals and applications. (John Wiley & Sons, 2008).
- 5.Huang B, Bates M, Zhuang X. Super resolution fluorescence microscopy. Annual review of biochemistry. 2009;78:993. doi: 10.1146/annurev.biochem.77.061906.092014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Godin AG, Lounis B, Cognet L. Super-resolution microscopy approaches for live cell imaging. Biophys J. 2014;107:1777–1784. doi: 10.1016/j.bpj.2014.08.028. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Ntziachristos V. Going deeper than microscopy: the optical imaging frontier in biology. Nature methods. 2010;7:603–614. doi: 10.1038/nmeth.1483. [DOI] [PubMed] [Google Scholar]
- 8.Wang XD, et al. Noninvasive laser-induced photoacoustic tomography for structural and functional in vivo imaging of the brain. Nat Biotechnol. 2003;21:803–806. doi: 10.1038/nbt839. [DOI] [PubMed] [Google Scholar]
- 9.Wang LHV, Hu S. Photoacoustic Tomography: In Vivo Imaging from Organelles to Organs. Science. 2012;335:1458–1462. doi: 10.1126/science.1216210. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Yao J, Wang LV. Sensitivity of photoacoustic microscopy. Photoacoustics. 2014;2:87–101. doi: 10.1016/j.pacs.2014.04.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Dean-Ben XL, Razansky D, Ntziachristos V. The effects of acoustic attenuation in optoacoustic signals. Physics in medicine and biology. 2011;56:6129–6148. doi: 10.1088/0031-9155/56/18/021. [DOI] [PubMed] [Google Scholar]
- 12.Estrada H, Turner J, Kneipp M, Razansky D. Real-time optoacoustic brain microscopy with hybrid optical and acoustic resolution. Laser Phys Lett. 2014;11:045601. doi: 10.1088/1612-2011/11/4/045601. [DOI] [Google Scholar]
- 13.Yao J, et al. High-speed label-free functional photoacoustic microscopy of mouse brain in action. Nature methods. 2015;12:407–410. doi: 10.1038/nmeth.3336. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Kim, J. Y., Lee, C., Park, K., Lim, G. & Kim, C. Fast optical-resolution photoacoustic microscopy using a 2-axis water-proofing MEMS scanner. Sci Rep-Uk5, doi:10.1038/Srep07932 (2015). [DOI] [PMC free article] [PubMed]
- 15.Bost, W. et al. In Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE. 5883–5886 (IEEE).
- 16.Strohm EM, Berndl ES, Kolios MC. High frequency label-free photoacoustic microscopy of single cells. Photoacoustics. 2013;1:49–53. doi: 10.1016/j.pacs.2013.08.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Bost W, Lemor R, Fournelle M. Optoacoustic imaging of subcutaneous microvasculature with a class one laser. IEEE transactions on medical imaging. 2014;33:1900–1904. doi: 10.1109/TMI.2014.2326552. [DOI] [PubMed] [Google Scholar]
- 18.Aguirre J, et al. Broadband mesoscopic optoacoustic tomography reveals skin layers. Opt Lett. 2014;39:6297–6300. doi: 10.1364/OL.39.006297. [DOI] [PubMed] [Google Scholar]
- 19.Strohm EM, Moore MJ, Kolios MC. Single cell photoacoustic microscopy: a review. Ieee J Sel Top Quant. 2016;22:1–15. doi: 10.1109/JSTQE.2015.2497323. [DOI] [Google Scholar]
- 20.Omar M, Gateau J, Ntziachristos V. Raster-scan optoacoustic mesoscopy in the 25–125 MHz range. Opt Lett. 2013;38:2472–2474. doi: 10.1364/OL.38.002472. [DOI] [PubMed] [Google Scholar]
- 21.Estrada, H., Sobol, E., Baum, O. & Razansky, D. Hybrid optoacoustic and ultrasound biomicroscopy monitors’ laser-induced tissue modifications and magnetite nanoparticle impregnation. Laser Phys Lett11, doi:10.1088/1612-2011/11/12/125601 (2014).
- 22.Dean-Ben, X. L., Pang, G. A., de Espinosa, F. M. & Razansky, D. Non-contact optoacoustic imaging with focused air-coupled transducers. Appl Phys Lett107, doi:10.1063/1.4928123 (2015).
- 23.Buehler A, Dean-Ben XL, Razansky D, Ntziachristos V. Volumetric Optoacoustic Imaging With Multi-Bandwidth Deconvolution. IEEE transactions on medical imaging. 2014;33:814–821. doi: 10.1109/TMI.2013.2282173. [DOI] [PubMed] [Google Scholar]
- 24.Deán-Ben XL, Razansky D. On the link between the speckle free nature of optoacoustics and visibility of structures in limited-view tomography. Photoacoustics. 2016;4:133–140. doi: 10.1016/j.pacs.2016.10.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Xu Y, Wang LV, Ambartsoumian G, Kuchment P. Reconstructions in limited-view thermoacoustic tomography. Medical physics. 2004;31:724–733. doi: 10.1118/1.1644531. [DOI] [PubMed] [Google Scholar]
- 26.Brecht, H. P. et al. Whole-body three-dimensional optoacoustic tomography system for small animals. Journal of biomedical optics14, doi:10.1117/1.3259361 (2009). [DOI] [PMC free article] [PubMed]
- 27.Buehler A, Dean-Ben XL, Claussen J, Ntziachristos V, Razansky D. Three-dimensional optoacoustic tomography at video rate. Optics express. 2012;20:22712–22719. doi: 10.1364/OE.20.022712. [DOI] [PubMed] [Google Scholar]
- 28.Xiang L, Wang B, Ji L, Jiang H. 4-D photoacoustic tomography. Sci Rep-Uk. 2013;3:1113. doi: 10.1038/srep01113. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Kruger, R. A. et al. Dedicated 3D photoacoustic breast imaging. Medical physics40, doi:10.1118/1.4824317 (2013). [DOI] [PMC free article] [PubMed]
- 30.Fehm TF, Dean-Ben XL, Ford SJ, Razansky D. In vivo whole-body optoacoustic scanner with real-time volumetric imaging capacity. Optica. 2016;3:1153–1159. doi: 10.1364/OPTICA.3.001153. [DOI] [Google Scholar]
- 31.Dean-Ben, X. L. & Razansky, D. Adding fifth dimension to optoacoustic imaging: volumetric time-resolved spectrally enriched tomography. Light-Sci Appl3, doi:10.1038/lsa.2014.18 (2014).
- 32.Dean-Ben, X. L., Bay, E. & Razansky, D. Functional optoacoustic imaging of moving objects using microsecond-delay acquisition of multispectral three-dimensional tomographic data. Sci Rep-Uk4, doi:10.1038/Srep05878 (2014). [DOI] [PMC free article] [PubMed]
- 33.Meng J, et al. High-speed, sparse-sampling three-dimensional photoacoustic computed tomography in vivo based on principal component analysis. Journal of biomedical optics. 2016;21:076007–076007. doi: 10.1117/1.JBO.21.7.076007. [DOI] [PubMed] [Google Scholar]
- 34.Arridge S, et al. Accelerated high-resolution photoacoustic tomography via compressed sensing. Physics in medicine and biology. 2016;61:8908. doi: 10.1088/1361-6560/61/24/8908. [DOI] [PubMed] [Google Scholar]
- 35.Sandbichler M, Krahmer F, Berer T, Burgholzer P, Haltmeier M. A novel compressed sensing scheme for photoacoustic tomography. Siam J Appl Math. 2015;75:2475–2494. doi: 10.1137/141001408. [DOI] [Google Scholar]
- 36.Dean-Ben, X. L., Ntziachristos, V. & Razansky, D. Effects of small variations of speed of sound in optoacoustic tomographic imaging. Medical physics41, doi:10.1118/1.4875691 (2014). [DOI] [PubMed]
- 37.Rosenthal A, Ntziachristos V, Razansky D. Model-based optoacoustic inversion with arbitrary-shape detectors. Medical physics. 2011;38:4285–4295. doi: 10.1118/1.3589141. [DOI] [PubMed] [Google Scholar]
- 38.Mitsuhashi K, Wang K, Anastasio MA. Investigation of the far-field approximation for modeling a transducer’s spatial impulse response in photoacoustic computed tomography. Photoacoustics. 2014;2:21–32. doi: 10.1016/j.pacs.2013.11.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Kong LJ, et al. Continuous volumetric imaging via an optical phase-locked ultrasound lens. Nature methods. 2015;12:759–U166. doi: 10.1038/nmeth.3476. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Bouchard MB, et al. Swept confocally-aligned planar excitation (SCAPE) microscopy for high-speed volumetric imaging of behaving organisms. Nat Photonics. 2015;9:113–119. doi: 10.1038/nphoton.2014.323. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Deán-Ben XL, et al. Functional optoacoustic neuro-tomography for scalable whole-brain monitoring of calcium indicators. Light: Science & Applications. 2016;5:e16201. doi: 10.1038/lsa.2016.201. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Looger LL, Griesbeck O. Genetically encoded neural activity indicators. Current Opinion in Neurobiology. 2012;22:18–23. doi: 10.1016/j.conb.2011.10.024. [DOI] [PubMed] [Google Scholar]
- 43.Lin MZ, Schnitzer MJ. Genetically encoded indicators of neuronal activity. Nature neuroscience. 2016;19:1142–1153. doi: 10.1038/nn.4359. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Dean-Ben XL, Ozbek A, Razansky D. Volumetric Real-Time Tracking of Peripheral Human Vasculature With GPU-Accelerated Three-Dimensional Optoacoustic Tomography. IEEE transactions on medical imaging. 2013;32:2050–2055. doi: 10.1109/TMI.2013.2272079. [DOI] [PubMed] [Google Scholar]
- 45.Marczak W. Water as a standard in the measurements of speed of sound in liquids. the Journal of the Acoustical Society of America. 1997;102:2776–2779. doi: 10.1121/1.420332. [DOI] [Google Scholar]
- 46.Lubbers J, Graaff R. A simple and accurate formula for the sound velocity in water. Ultrasound in medicine & biology. 1998;24:1065–1068. doi: 10.1016/S0301-5629(98)00091-X. [DOI] [PubMed] [Google Scholar]
- 47.Xu, M. H. & Wang, L. H. V. Universal back-projection algorithm for photoacoustic computed tomography. Phys Rev E71, doi:10.1103/Physreve.71.016706 (2005). [DOI] [PubMed]
- 48.Dean-Ben, X. L., Fehm, T. F., Ford, S. J., Gottschalk, S. & Razansky, D. Spiral volumetric optoacoustic tomography visualizes multi-scale dynamics in mice. Light-Sci Appl6, doi:10.1038/lsa.2016.247 (2017). [DOI] [PMC free article] [PubMed]
- 49.Laufer J, Zhang E, Raivich G, Beard P. Three-dimensional noninvasive imaging of the vasculature in the mouse brain using a high resolution photoacoustic scanner. Applied optics. 2009;48:D299–D306. doi: 10.1364/AO.48.00D299. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.