Abstract
Background
Magnetic resonance-guided high intensity focused ultrasound (MR-HIFU) ablation is in development for minimally invasive treatment of breast cancer. Cost-effectiveness has not been assessed yet. An early health technology assessment was performed to estimate costs of MR-HIFU ablation, compared to breast conserving treatment (BCT).
Methods
An MR-HIFU treatment model using the dedicated MR-HIFU breast system (Sonalleve, Philips Healthcare) was developed. Input parameters (treatment steps and duration) were based on the analysis of questionnaire data from an expert panel. MR-HIFU experts assessed face validity of the model. Data collected by questionnaires were compared to published data of an MR-HIFU breast feasibility study. Treatment costs for tumours of 1 to 3 cm were calculated.
Results
The model structure was considered of acceptable face validity by consulted experts, and questionnaire data and published data were comparable. Costs of MR-HIFU ablation were higher than BCT costs. MR-HIFU best-case scenario costs exceeded BCT costs with approximately €1000. Cooling times and breathing correction contributed most to treatment costs.
Conclusions
MR-HIFU ablation is currently not a cost-effective alternative for BCT. MR-HIFU experience is limited, increasing uncertainty of estimations. The potential for cost-effectiveness increases if future research reduces treatment durations and might substantiate equal or improved results.
Keywords: Breast cancer, MR-HIFU, HTA, Cost-effectiveness, Breast-conserving therapy
Background
Breast cancer is the most common malignancy in women worldwide and its incidence is increasing [1, 2]. As a result of national screening programmes, most breast cancers are detected at an early stage [3]. Early stage breast cancer is usually treated with breast conserving therapy (BCT), which consists of lumpectomy combined with radiotherapy, followed by systemic therapy in patients deemed at high risk of metastases [4]. The overall prognosis after BCT is good, i.e. survival is similar to more radical mastectomy [5]. However, any surgical treatment always bears a risk of impaired cosmetic results and complications [6–8].
Currently, a shift towards non-surgical and less invasive treatment has been observed in several clinical trials, assessing the feasibility and efficacy of minimally invasive therapies [9–12]. One of these novel treatments is Magnetic Resonance guided High Intensity Focused Ultrasound ablation (MR-HIFU) [13]. Using focused ultrasound beams with a high power MRI-integrated HIFU systems heat breast tumours to high temperatures, inducing coagulation necrosis. Possible advantages of MR-HIFU ablation are a lower risk of complications such as infection and haemorrhage, improved cosmetic outcome and the possibility to offer the treatment in an outpatient setting without general anaesthesia. MRI-guidance is used for tumour visualization and temperature measurement during the procedures [13–15]. Initial clinical MR-HIFU studies report the treatment of approximately 122 malignant breast tumours, of which 77 were excised afterwards to assess histopathological response. The percentage of complete tumour ablation in these small feasibility studies varies from 16.7 to 90% [16–25].
Besides effectiveness, potential cost-effectiveness is a relevant aspect of introducing a new technique. MR-HIFU ablation will only have the potential to become a primary treatment in the future if its cost-effectiveness is acceptable compared to surgical treatment. Because its costs have not been assessed yet, the purpose of the current study was early health technology assessment. While assuming equal effectiveness of MR-HIFU and BCT, costs for treatment using MR-HIFU ablation compared to BCT were estimated. Additionally, the influence of several treatment-related features on these costs was assessed.
Methods
A decision tree model was developed to evaluate the additional costs of MR-HIFU ablation as a replacement of BCT for the treatment of early-stage breast cancer and to what extent these costs are influenced by several treatment-related features. Equal effectiveness of MR-HIFU ablation and BCT was assumed for these analyses, because of the limited amount of clinical data of MR-HIFU treatments [26, 27]. Model input data were collected in a systematic way. Where possible, parameter estimates were based on literature. For parameters that were not available in literature, a survey among experts was performed. MR-HIFU experts were asked to assess the validity of the model. In addition, the model input was validated by comparison of treatment duration estimates to a recent publication on the feasibility and safety of MR-HIFU ablation [28].
BCT and MR-HIFU scenarios
BCT was compared with MR-HIFU ablation. For BCT, treatment consisted of surgery with sentinel lymph node procedure, hospital admission, histopathological examination of excision specimen and adjuvant treatment in most cases. MR-HIFU treatment comprised a pre-treatment MRI scan, separate pre-treatment sentinel lymph node procedure, MR-HIFU ablation in day care setting and adjuvant treatment in most cases. Follow-up was not taken into account in this early health technology assessment (HTA).
Patients
MR-HIFU ablation was considered most suitable for patients with early stage breast cancer with a maximum diameter of three centimetres with no malignant foci at a larger distance than 1 cm from the tumour edge [29]. Additional exclusion criteria for MR-HIFU ablation were: ductal carcinoma in situ (DCIS) and lobular histological type, as both increased the risk of incomplete resection and ablation [30, 31], the presence of axillary lymph node metastases and all contra-indications for MRI. Furthermore, patients could be excluded from MR-HIFU treatment due to the following factors assessed on pre-treatment MRI: tumour not reachable for the ultrasound beam, or distance from tumour to skin or pectoral muscle < 1 cm [28].
Model
A model comprising the MR-HIFU treatment as performed with the dedicated MR-HIFU breast system (Sonalleve-based prototype, Philips Healthcare, Vantaa, Finland) was developed [32]. The model distinguished four separate phases: patient positioning on the MR-HIFU system, test phase (establishment of the exact treatment focus and treatment planning), therapeutic phase (the actual tumour ablation) and post-treatment phase (Fig. 1).
In the positioning phase, the patient was positioned on the HIFU table, which was then placed in the MR scanner, and tumour reachability for the HIFU beams was checked. Next, sedation analgesia was administered and target definition was performed based on MR images. In the test phase the respiratory breathing pattern was tracked to correct for breathing artefacts during proton resonance frequency shift (PRFS) thermometry [15]. Additionally, in the test phase the exact location of the focal point was checked and adjusted if necessary. In the therapeutic phase sonications with a higher power than used in the test phase were applied. Using PRFS thermometry maps, the temperature rise in the targeted tissue was followed to ensure that ablative temperatures were reached. Test sonications were applied again after switching to the next treatment slice. In the post-treatment phase contrast-enhanced MRI was performed to evaluate treatment results and sedation analgesia was ceased. The patient was admitted to a clinical ward for observation during the next four hours. Modelling was performed in MATLAB (R2014a). The conceptual model was tested for face validity with MR-HIFU experts involved in the MR-HIFU ablation feasibility and safety study performed with the aforementioned HIFU breast system [28].
Model input data
For an overview of the model input parameters, see Table 1 and Table 2. To estimate the yearly number of early stage breast cancer (stage I and stage II tumours (≤3 cm in diameter)) patients, data from the Netherlands Comprehensive Cancer Organisation were used [33]. Several parameters concerned duration of treatment steps and probability of events related to treatment, e.g. repositioning a patient. These were based on a questionnaire filled in a by a team of (inter) national experts (physicians involved in breast cancer treatment, physicists, technicians and physicians with MR-HIFU experience). Durations were estimated by seven experts, probabilities by four of those seven experts.
Table 1.
Treatment phase | Parameter | Experts | Faesibility study [28] | |||
---|---|---|---|---|---|---|
Unit | Median | Min | Max | Median | ||
Positioning | Time patient verification | min | 15 | 10 | 25 | 11.5 |
Time verification reachability | min | 15 | 7 | 20 | 14.5 | |
Time target definition | min | 8 | 5 | 15 | n/a | |
Chance of repositioning | - | 0.30 | 0.10 | 0.75 | n/a | |
Test | Time to place navigator | min | 2 | 1 | 10 | 1 |
Time MRI scan | min | 2.5 | 1 | 15 | 2.5 | |
Time for treatment planning | min | 2 | 1 | 5 | n/a | |
Time to fill LUT | min | 2.75 | 0 | 5 | n/a | |
Test sonication and check focal point | min | 3 | 1 | 5 | 4 | |
Chance of adjustment focal point per coronal plane | - | 0.55 | 0.20 | 0.90 | n/a | |
Therapeutic | Time therapeutic phase | min | 0.50 | 0.25 | 0.75 | n/a |
Cooling time after each sonication | min | 3.5 | 1 | 10 | n/a | |
Chance of abortion per coronal plane | - | 0,10 | 0,05 | 0,20 | n/a | |
Chance of resonication per coronal plane | - | 0.20 | 0.10 | 0.30 | n/a | |
Post-treatment | Time clinical ward | min | 240 | 120 | 300 | n/a |
Chance of complications | - | 0.015 | 0.01 | 0.03 | n/a |
n/a not available
Table 2.
Cost category | Unit Price | Source |
---|---|---|
Diagnostics | ||
Pathology and evaluation | € 83 | Flobbe et al (2004) |
DBC sentinel node procedure | € 367.71 | Passantenprijslijst UMC Utrecht - Overige zorgproducten, 2014 |
Treatment items | ||
Estimated costs for using MR scanner and HIFU device | € 11/min | Overzicht tarieven onderlinge dienstverlening UMCG; 2016 |
Anesthetics costs | € 2.44 /min | Division of Vital Functions, UMC Utrecht, 2015 |
Costs clininal ward | € 0.44/min | CVZ, Handleiding voor kostenonderzoek, 2010; |
Costs technician | € 0.42/min | CVZ, Handleiding voor kostenonderzoek, 2010; CAO universitair medische centra 2015–2017 |
Costs nurse | € 0.42/min | CVZ, Handleiding voor kostenonderzoek, 2010; CAO universitair medische centra 2015–2017 |
Costs radiologist | € 1.33/min | CVZ, Handleiding voor kostenonderzoek, 2010; CAO universitair medische centra 2015–2017 |
DBC DCE MRI | € 331 | Dutch Healthcare Authority (NZA) 2012 |
Breast conserving therapy | € 1109 | https://www.zorginstituutnederland.nl/ |
Radiation therapy | € 3179 | Flobbe et al (2004); LPRM and NABON (2000); Slotman et al (2000) |
Chemo therapy | € 1044 | Flobbe et al (2004); Slotman et al (2000) |
Hormonal therapy | € 806 | Flobbe et al (2004); Slotman et al (2000) |
Specialist Visits | € 738 | Flobbe et al (2004); |
Specialist visits (adjuvant therapy) | € 1966 | Flobbe et al (2004); |
Hospital stay | € 1753 | Flobbe et al (2004); Oostenbrink et al (2000) |
To estimate the time that was needed for MR-HIFU treatment, a simplified tumour model was assumed with sphere-shaped tumours with a diameter of 1, 2 or 3 cm. A safety margin of 0.5 cm was added to the tumour size, resulting in spheres with a diameter of 2, 3 or 4 cm. The treatment cells of the HIFU device were considered cylindrically shaped with a diameter of 3, 6, 9 or 12 mm in the coronal plane and a height of 2, 4, 6 or 8 in the sagittal plane respectively. These are the approximate values provided by the dedicated MR-HIFU system [28, 32]. The number of sonications required for tumours of different diameters was approximated by assuming cylindrical shaped treatment cells covering a sphere shaped tumour. The number of sonications varied with the height and diameter of treatment cells, as follows: treatment cells with a diameter of 9 mm and height of 6 mm resulted in 17 sonications for tumours of 1 cm, 50 sonications for tumours of 2 cm and 110 sonications for tumours of 3 cm. Treatment cells with a diameter of 12 mm and height of 8 mm resulted in 9, 25, and 50 sonications respectively.
Cost data
The average costs of BCT of the aforementioned patient population were based on a database comprising 1,345 breast cancer patients [34]. Hereby a weighting to the amount of women undergoing lumpectomy with or without adjuvant therapy, i.e. systemic therapy (hormonal therapy and chemotherapy) and radiotherapy, was done [35–37]. To estimate the costs for using the MR scanner and HIFU device, tariffs for MR procedures that were comparable in complexity, such as brain and heart MR imaging were used as a proxy [38]. Costs for the sentinel node procedure and contrast enhanced MRI were based on their tariffs [39]. Depreciation and maintenance costs of devices were incorporated in these tariffs. Costs of the additional MR-HIFU treatment components, e.g. sedation costs, were based on hospital specific rates. Costs of personnel present during the procedures was based on estimates of time needed multiplied by hourly costs, based on the guidance of the National Health Care Institute (Dutch: Zorginstituut Nederland) [40]. Costs of follow-up were not taken into account. Costs were indexed to 2014 by using consumer price index numbers [41].
Analysis
The costs of MR-HIFU ablation were based on the MR-HIFU submodel. The lowest and median estimates of time needed obtained with expert questionnaires were used to calculate the most optimistic (‘best case’) and less optimistic (‘median case’) MR-HIFU treatment scenarios. MR-HIFU treatment costs for tumour sizes of 1, 2 and 3 cm were calculated. This was done for treatment cells of 9 × 6 mm and 12 × 8 mm. These costs were compared to the average BCT costs. Tornado diagrams were constructed to describe the sensitivity of costs to parameter estimates.
Results
Patients
Taking all possible exclusion criteria for MR-HIFU treatment into account, the proportion of patients eligible for MR-HIFU treatment was 11.9% of all patients diagnosed with breast cancer (Table 3) [31, 42–52].
Table 3.
Inclusion criteria | Proportion (%) |
---|---|
Tumour ≤ 3 cm | 78.6 |
No lymph node metastasis | 65.0 |
No lobular subtype | 90.2 |
No EIC | 84.6 |
No previous surgery | 91.3 |
No renal insufficiency | 97.3 |
Not multifocal | 82.0 |
No BRCA mutation | 97.4 |
Tumour reachable | 66.3 |
Distance to skin ≥ 1 cm | 65.0 |
Eligible patientsa | 11.9 |
EIC extensive intraductal component
ai.e. all of the inclusion criteria present
Model
The structure of the model was considered of acceptable face validity by experts consulted. Model input parameters on analogous variables derived from actually observed data were comparable to the answers obtained through the questionnaires [28]. The duration and chances of occurrence of the different MR-HIFU treatment steps, compared to the data observed in the feasibility study, is shown in Table 1.
Costs of MR-HIFU ablation and BCT
The costs of MR-HIFU ablation for best and median case scenarios and two different cell sizes and the costs of BCT for tumours of 1, 2 or 3 cm are displayed in Table 4. The larger the treatment cell, the lower the MR-HIFU costs and the shorter the procedure time. For all variants, the costs of MR-HIFU ablation were higher than the costs of BCT. When using treatment cells size of 12 × 8 mm, the best case scenario costs of MR-HIFU ablation approached those of BCT.
Table 4.
MR-HIFU ablation | BCT | ||||||||
---|---|---|---|---|---|---|---|---|---|
Median case | Best case | ||||||||
Treatment cell size (mm) | Treatment cell size (mm) | ||||||||
6 × 9 | 12 × 8 | 6 × 9 | 12 × 8 | ||||||
Tumour size (mm) | Costs (€1000) | Time (h) | Costs (€1000) | Time (h) | Costs (€1000) | Time (h) | Costs (€1000) | Time (h) | Costs (€1000) |
10 | 11.5 | 4.4 | 10.0 | 2.8 | 8.5 | 1.1 | 8.2 | 0.8 | 7.1 |
20 | 15.5 | 8.8 | 12.5 | 5.4 | 9.1 | 1.9 | 8.6 | 1.3 | 8.1 |
30 | 23.3 | 17.4 | 15.5 | 8.8 | 10.5 | 3.4 | 9.1 | 1.9 | 8.1 |
Factors influencing treatment costs
Factors contributing most to the total treatment costs were: cooling time after each sonication, and time required for breathing correction. For 9 × 6 mm treatment cells the average sensitivity for cooling time over all tumour sizes was 64.6 ± 10.3%, for the breathing correction this was 28 ± 2.7%. For treatment cells of 12 × 8 mm this was 59.3 ± 10.8% and 29.1 ± 1.6% respectively. Changes in these two parameters had larger impact on the cost estimations of treatments with smaller treatment cells than with larger treatment cells. The sensitivity of each model input parameter on model output is shown in Table 5 in tornado diagrams. The sensitivity values for both 9 × 6 mm and 12 × 8 mm treatment cells are presented in respectively Fig. 2 and 3.
Table 5.
Treatment phase | Tumour size (mm)a | 10 | 20 | 30 | |||||||
---|---|---|---|---|---|---|---|---|---|---|---|
Treatment cell size (mm) | 9×6 | 12×8 | 9×6 | 12×8 | 9×6 | 12×8 | 9×6 | 12×8 | |||
Parameter | Mean | sd | Mean | sd | |||||||
Positioning | Time patient positioning | 0.5 | 1.6 | 0.1 | 0.3 | 0.0 | 0.1 | 0.2 | 0.3 | 0.7 | 0.8 |
Time verification reachability | 0.4 | 1.2 | 0.1 | 0.2 | 0.0 | 0.1 | 0.2 | 0.2 | 0.5 | 0.6 | |
Time target definition | 0.1 | 0.4 | 0.0 | 0.1 | 0.0 | 0.0 | 0.0 | 0.1 | 0.2 | 0.2 | |
Chance of repositioning | 0.5 | 1.6 | 0.1 | 0.3 | 0.0 | 0.1 | 0.2 | 0.3 | 0.7 | 0.8 | |
Test | Time MR scan | 7.7 | 8.9 | 1.2 | 4.1 | 0.5 | 1.2 | 3.1 | 4.0 | 4.7 | 3.9 |
Time to place navigator | 3.2 | 3.7 | 0.5 | 1.7 | 0.2 | 0.5 | 1.3 | 1.7 | 2.0 | 1.6 | |
Time to perform treatment planning per coronal plane | 0.6 | 0.7 | 0.1 | 0.3 | 0.0 | 0.1 | 0.2 | 0.3 | 0.4 | 0.3 | |
Time to fill LUT | 31.0 | 30.2 | 27.2 | 29.8 | 25.9 | 27.2 | 28.0 | 2.7 | 29.1 | 1.6 | |
Time to perform test sonications and verify focal point | 1.5 | 1.7 | 0.2 | 0.8 | 0.1 | 0.2 | 0.6 | 0.8 | 0.9 | 0.8 | |
Chance of readjustment focal point | 0.6 | 0.7 | 0.1 | 0.3 | 0.0 | 0.1 | 0.2 | 0.3 | 0.4 | 0.3 | |
Therapeutic | Time to perform therapeutic sonication | 0.2 | 0.1 | 0.2 | 0.2 | 0.2 | 0.2 | 0.2 | 0.0 | 0.2 | 0.1 |
Cooling time after each sonication | 52.8 | 47.8 | 69.2 | 60.9 | 71.8 | 69.2 | 64.6 | 10.3 | 59.3 | 10.8 | |
Chance of resonication per coronal plane | 0.8 | 0.7 | 1.1 | 1.0 | 1.1 | 1.1 | 1.0 | 0.2 | 0.9 | 0.2 | |
Post treatment | Time clinical ward | 0.2 | 0.5 | 0.0 | 0.1 | 0.0 | 0.0 | 0.1 | 0.1 | 0.2 | 0.3 |
Sensitivity (variance) was calculated by calculation of the swing square relatively to the total swing square. Hereby, the swing is the range of cost values for a given uncertainty
aA safety margin of 0.5 cm was added to the tumour size
Discussion
To our knowledge, this is the first study on the potential cost-effectiveness of MR-HIFU ablation of breast cancer. This early health technology assessment suggests that MR-HIFU ablation was more expensive than BCT. When larger treatment cells were assumed, the potential for MR-HIFU ablation to have comparable costs increased. The duration of certain treatment steps including cooling time after each sonication and the time needed to apply breathing correction, had most impact on MR-HIFU costs. Importantly, the analyses were performed under the assumption that MR-HIFU ablation and BCT are equally effective.
Due to the limited amount of MR-HIFU treatment data, the clinical effectiveness, complication rate and the effect on quality of life and cosmetic outcome is still scarce. Therefore, the effectiveness of MR-HIFU ablation was considered equal to BCT and complications were not taken into account. However, if MR-HIFU treatment would be optimized and surgical excision would be omitted in the future, surgical complications and breast deformation might occur less frequently. This is expected to have a favourable effect on quality of life and cosmetic outcome. Even if MR-HIFU ablation would be slightly less effective than BCT, some patients might still prefer MR-HIFU because of its favourable effect on cosmetic outcome or reduced risk of complications. Especially elderly patients with more comorbidities and shorter life expectancy may be interested in MR-HIFU ablation. Quality of life measures are usually incorporated in cost-effectiveness analyses as well, and hence better treatment associated utility scores would increase the potential for MR-HIFU to become cost-effective [53].
Our results indicate that in order to improve the cost-effectiveness of MR-HIFU ablation treatment time should be reduced. The currently used cooling times are applied to guarantee safety. If more clinical experience with MR-HIFU ablation is gained, shorter cooling times may possibly appear equally safe. Being able to measure the temperature in the surrounding (adipose) tissue in the breast may contribute by providing real time temperature measurements during cooling. Possible methods for thermometry in adipose tissue are T2-weighted thermometry [54, 55] or a hybrid method for thermometry in fat and adipose tissue at the same time [56, 57]. Implementation of these techniques will become possible in the near future. The time needed for breathing correction could be improved by using sedation that results in a more regular breathing pattern, or even obviates the need for breathing correction. Furthermore, the parameters of breathing correction could be made more flexible enabling sonications to start after a shorter period. To what extent this is possible should be clinically investigated first and is not expected in the near future. Another reason for the higher costs of MR-HIFU ablation is the separately performed sentinel lymph node procedure, which is incorporated in the operation for BCT.
Strong points of this study are that it provided the first data on the potential cost-effectiveness of MR-HIFU ablation and that MR-HIFU experts validated the applied treatment models. Due to the lack of empirical treatment data, other sources for model input were needed. Estimations of experts are the most accurate option in this case. Furthermore, the duration of several treatment elements was validated with the published MR-HIFU breast study in which the dedicated MR-HIFU breast system was used [28].
Some limitations of the present study should also be acknowledged. First, MR-HIFU treatment duration may have been overestimated as a result of the lack of experience. As mentioned before, cooling times may be unnecessarily prolonged. Besides, no experience with total tumour ablation exists yet, possibly affecting some estimations. Second, treatment cells were assumed to be cylindrical shaped to enable calculations of the number of sonications required per tumour. The MR- HIFU breast system provides treatment cells with the shape of an oblate ellipse. This difference in shape may have affected the results. Third, the costs of BCT may have been overestimated, as this cost estimate also comprised patients with lobular carcinoma and positive axillary lymph nodes. However, patients with a tumour of the lobular subtype are considered ineligible for MR-HIFU treatment, as lobular breast cancer has a higher risk of incomplete resection and is consequently more expensive to treat. The same applies for patients with positive axillary lymph nodes. Fourth, our estimations were based on the dedicated MR-HIFU breast system used in our centre (Sonalleve-based prototype, Philips Healthcare, Vantaa, Finland). Other MR-HIFU systems exist and it is unclear if our results would be generalizable to these systems. Fifth, the cost data is based on the local health system. This means that the cost data cannot be extrapolated to other countries. However, we have generated a model that is usable internationally. Lastly, in the performed early HTA indirect costs were not taken into account. For example, the expected duration of absence at work will be longer after breast surgery than after MR-HIFU. As a result, the costs of BCT may have been underestimated in this study.
Several issues need to be addressed before MR-HIFU ablation can become a clinical treatment. First of all, the clinical efficacy of MR-HIFU has to be proven. MR-HIFU is competing with a very reliable and well-established treatment option, BCT. Clinical trials on MR-HIFU are therefore hard to perform and require many participants. Other concerns are that no surgical excision specimen is available after MR-HIFU treatment, the indication for adjuvant treatment needs to be assessed prior to treatment. Furthermore, margin status cannot be assessed, which may be resolved by frequent follow-up with MRI. In conclusion, MR-HFIU ablation is a novel treatment which is currently not ready for clinical implementation, but recent developments are promising for the future.
Conclusions
We tentatively conclude that MR-HIFU ablation currently is not a cost-effective alternative to BCT. The costs of MR-HIFU ablation are mostly affected by the long duration of certain treatment components, i.e. cooling time after sonications and the time needed to apply breathing correction. Furthermore, costs were influenced by the size of treatment cell used and decreased with larger treatment cell size. Being an early HTA analysis, the study had to be based on several assumptions and estimations, because the experience of MR-HIFU ablation is still quite limited. Therefore, our results may give important directions for future development of MR-HIFU ablation. Especially cooling time in between sonications and accurate breathing correction take relatively long and thus appear relevant targets for further innovation.
Acknowledgements
We thank Dr. D. Gianfelice for collaborating with us by sharing his MR-HIFU breast expertise.
Funding
This study has received funding from the Center for Translational Molecular Medicine (CTMM).
Availability of data and materials
The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.
Abbreviations
- BCT
Breast conserving treatment
- DCIS
Ductal carcinoma in situ
- HTA
Health technology assessment
- MR-HIFU
Magnetic resonance guided high intensity focused ultrasound ablation
- PRFS
Proton resonance frequency shift
Authors’ contributions
FMK and SEMH share first authorship. FMK wrote the manuscript and performed the clinically related analysis. SEMH performed most analyses and built all required models. TLF supervised the economical analyses and determined which economical analysis was just. CTWM supervised the MR-HIFU related analyses and determined the right MR-HIFU workflow. MAAJB was involved in het MR-HIFU analyses as well. EB and MJWG both performed economical analyses and supervised writing the manuscript. GHB supervised writing the manuscript and provided economical data and determined which economical methods were used. All authors read and approved the final manuscript
Ethics approval and consent to participate
Ethics approval for the clinical MR-HIFU breast study, of which data was used in this article, was provided by the Medical Ethical Committee of the University of Utrecht [28].
Consent for publication
Not applicable
Competing interests
The authors declare that they have no competing interests.
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.DeSantis C, Ma J, Bryan L, Jemal A. Breast cancer statistics, 2013. CA Cancer J Clin. 2014;64:52–62. doi: 10.3322/caac.21203. [DOI] [PubMed] [Google Scholar]
- 2.Siegel R, Naishadham D, Jemal A. Cancer statistics, 2013. CA Cancer J Clin. 2013;63:11–30. doi: 10.3322/caac.21166. [DOI] [PubMed] [Google Scholar]
- 3.Bastiaannet E, Liefers GJ, de Craen AJ, Kuppen PJ, van de Water W, Portielje JE, van der Geest LG, Janssen-Heijnen ML, Dekkers OM, van de Velde CJ, Westendorp RG. Breast cancer in elderly compared to younger patients in the Netherlands: stage at diagnosis, treatment and survival in 127,805 unselected patients. Breast Cancer Res Treat. 2010;124:801–807. doi: 10.1007/s10549-010-0898-8. [DOI] [PubMed] [Google Scholar]
- 4.Integraal Kankercentrum Nederland: Mammacarcinoom. Landelijke richtlijn, versie: 2.0. Oncoline; 2012.
- 5.Hwang ES, Lichtensztajn DY, Gomez SL, Fowble B, Clarke CA. Survival after lumpectomy and mastectomy for early stage invasive breast cancer: the effect of age and hormone receptor status. Cancer. 2013;119:1402–1411. doi: 10.1002/cncr.27795. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Bajaj AK, Kon PS, Oberg KC, Miles DA. Aesthetic outcomes in patients undergoing breast conservation therapy for the treatment of localized breast cancer. Plast Reconstr Surg. 2004;114:1442–1449. doi: 10.1097/01.PRS.0000138813.64478.A7. [DOI] [PubMed] [Google Scholar]
- 7.Schmitz KH, Speck RM, Rye SA, DiSipio T, Hayes SC. Prevalence of breast cancer treatment sequelae over 6 years of follow-up: the Pulling Through Study. Cancer. 2012;118:2217–2225. doi: 10.1002/cncr.27474. [DOI] [PubMed] [Google Scholar]
- 8.Hennigs A, Hartmann B, Rauch G, Golatta M, Tabatabai P, Domschke C, Schott S, Schutz F, Sohn C, Heil J. Long-term objective esthetic outcome after breast-conserving therapy. Breast Cancer Res Treat. 2015;153:345–351. doi: 10.1007/s10549-015-3540-y. [DOI] [PubMed] [Google Scholar]
- 9.Postma EL, van Hillegersberg R, Daniel BL, Merckel LG, Verkooijen HM, van den Bosch MA. MRI-guided ablation of breast cancer: where do we stand today? J Magn Reson Imaging. 2011;34:254–261. doi: 10.1002/jmri.22599. [DOI] [PubMed] [Google Scholar]
- 10.Waaijer L, Kreb DL, Fernandez Gallardo MA, Van Rossum PS, Postma EL, Koelemij R, Van Diest PJ, Klaessens JH, Witkamp AJ, Van Hillegersberg R. Radiofrequency ablation of small breast tumours: evaluation of a novel bipolar cool-tip application. Eur J Surg Oncol. 2014;40:1222–1229. doi: 10.1016/j.ejso.2014.07.031. [DOI] [PubMed] [Google Scholar]
- 11.Dowlatshahi K, Francescatti DS, Bloom KJ. Laser therapy for small breast cancers. Am J Surg. 2002;184:359–363. doi: 10.1016/S0002-9610(02)00942-X. [DOI] [PubMed] [Google Scholar]
- 12.Manenti G, Perretta T, Gaspari E, Pistolese CA, Scarano L, Cossu E, Bonanno E, Buonomo OC, Petrella G, Simonetti G, Masala S. Percutaneous local ablation of unifocal subclinical breast cancer: clinical experience and preliminary results of cryotherapy. Eur Radiol. 2011;21:2344–2353. doi: 10.1007/s00330-011-2179-2. [DOI] [PubMed] [Google Scholar]
- 13.Jolesz FA. MRI-guided focused ultrasound surgery. Annu Rev Med. 2009;60:417–430. doi: 10.1146/annurev.med.60.041707.170303. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Schmitz AC, Gianfelice D, Daniel BL, Mali WP, van den Bosch MA. Image-guided focused ultrasound ablation of breast cancer: current status, challenges, and future directions. Eur Radiol. 2008;18:1431–1441. doi: 10.1007/s00330-008-0906-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Rieke V, Butts Pauly K. MR thermometry. J Magn Reson Imaging. 2008;27:376–390. doi: 10.1002/jmri.21265. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Furusawa H, Namba K, Nakahara H, Tanaka C, Yasuda Y, Hirabara E, Imahariyama M, Komaki K. The evolving non-surgical ablation of breast cancer: MR guided focused ultrasound (MRgFUS) Breast Cancer. 2007;14:55–58. doi: 10.2325/jbcs.14.55. [DOI] [PubMed] [Google Scholar]
- 17.Furusawa H, Namba K, Thomsen S, Akiyama F, Bendet A, Tanaka C, Yasuda Y, Nakahara H. Magnetic resonance-guided focused ultrasound surgery of breast cancer: reliability and effectiveness. J Am Coll Surg. 2006;203:54–63. doi: 10.1016/j.jamcollsurg.2006.04.002. [DOI] [PubMed] [Google Scholar]
- 18.Gianfelice D, Khiat A, Amara M, Belblidia A, Boulanger Y. MR imaging-guided focused ultrasound surgery of breast cancer: correlation of dynamic contrast-enhanced MRI with histopathologic findings. Breast Cancer Res Treat. 2003;82:93–101. doi: 10.1023/B:BREA.0000003956.11376.5b. [DOI] [PubMed] [Google Scholar]
- 19.Gianfelice D, Khiat A, Amara M, Belblidia A, Boulanger Y. MR imaging-guided focused US ablation of breast cancer: histopathologic assessment of effectiveness-- initial experience. Radiology. 2003;227:849–855. doi: 10.1148/radiol.2281012163. [DOI] [PubMed] [Google Scholar]
- 20.Gianfelice D, Khiat A, Boulanger Y, Amara M, Belblidia A. Feasibility of magnetic resonance imaging-guided focused ultrasound surgery as an adjunct to tamoxifen therapy in high-risk surgical patients with breast carcinoma. J Vasc Interv Radiol. 2003;14:1275–1282. doi: 10.1097/01.RVI.0000092900.73329.A2. [DOI] [PubMed] [Google Scholar]
- 21.Huber PE, Jenne JW, Rastert R, Simiantonakis I, Sinn HP, Strittmatter HJ, von Fournier D, Wannenmacher MF, Debus J. A new noninvasive approach in breast cancer therapy using magnetic resonance imaging-guided focused ultrasound surgery. Cancer Res. 2001;61:8441–8447. [PubMed] [Google Scholar]
- 22.Khiat A, Gianfelice D, Amara M, Boulanger Y. Influence of post-treatment delay on the evaluation of the response to focused ultrasound surgery of breast cancer by dynamic contrast enhanced MRI. Br J Radiol. 2006;79:308–314. doi: 10.1259/bjr/23046051. [DOI] [PubMed] [Google Scholar]
- 23.Marincola BC, Napoli A, Pediconi F, Di Mare L, Palla C, Telesca M, Miglio E, Amabile MI, Marangi G, d’Amati G, et al. Initial clinical experience of non-invasive treatment of Magnetic Resonance guided high intensity focused Ultrasound (MRgFUS) for focal breast cancer. Journal of Therapeutic Ultrasound. 2014;2:A16. doi: 10.1186/2050-5736-2-S1-A16. [DOI] [Google Scholar]
- 24.Napoli A, Anzidei M, Ciolina F, Marotta E, Cavallo Marincola B, Brachetti G, Di Mare L, Cartocci G, Boni F, Noce V, et al. MR-guided high-intensity focused ultrasound: current status of an emerging technology. Cardiovasc Intervent Radiol. 2013;36:1190–1203. doi: 10.1007/s00270-013-0592-4. [DOI] [PubMed] [Google Scholar]
- 25.Zippel DB, Papa MZ. The use of MR imaging guided focused ultrasound in breast cancer patients; a preliminary phase one study and review. Breast Cancer. 2005;12:32–38. doi: 10.2325/jbcs.12.32. [DOI] [PubMed] [Google Scholar]
- 26.Chapman AM, Taylor CA, Girling AJ. Early HTA to Inform Medical Device Development Decisions - The Headroom Method. In: Roa Romero L. (eds) XIIIMediterranean Conference on Medical and Biological Engineering and Computing 2013. IFMBE Proceedings, vol 41. Springer, Cham; 2014.
- 27.Ijzerman MJ, Steuten LM. Early assessment of medical technologies to inform product development and market access: a review of methods and applications. Appl Health Econ Health Policy. 2011;9:331–347. doi: 10.2165/11593380-000000000-00000. [DOI] [PubMed] [Google Scholar]
- 28.Merckel LG, Knuttel FM, Deckers R, van Dalen T, Schubert G, Peters NH, Weits T, van Diest PJ, Mali WP, Vaessen PH, et al. First clinical experience with a dedicated MRI-guided high-intensity focused ultrasound system for breast cancer ablation. Eur Radiol. 2016;26(11):4037–46. doi: 10.1007/s00330-016-4222-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Faverly DR, Hendriks JH, Holland R. Breast carcinomas of limited extent: frequency, radiologic-pathologic characteristics, and surgical margin requirements. Cancer. 2001;91:647–659. doi: 10.1002/1097-0142(20010215)91:4<647::AID-CNCR1053>3.0.CO;2-Z. [DOI] [PubMed] [Google Scholar]
- 30.Pleijhuis RG, Kwast AB, Jansen L, de Vries J, Lanting R, Bart J, Wiggers T, van Dam GM, Siesling S. A validated web-based nomogram for predicting positive surgical margins following breast-conserving surgery as a preoperative tool for clinical decision-making. Breast. 2013;22:773–779. doi: 10.1016/j.breast.2013.01.010. [DOI] [PubMed] [Google Scholar]
- 31.Jung W, Kang E, Kim SM, Kim D, Hwang Y, Sun Y, Yom CK, Kim SW. Factors Associated with Re-excision after Breast-Conserving Surgery for Early-Stage Breast Cancer. J Breast Cancer. 2012;15:412–419. doi: 10.4048/jbc.2012.15.4.412. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Merckel LG, Bartels LW, Kohler MO, van den Bongard HJ, Deckers R, Mali WP, Binkert CA, Moonen CT, Gilhuijs KG, van den Bosch MA. MR-guided high-intensity focused ultrasound ablation of breast cancer with a dedicated breast platform. Cardiovasc Intervent Radiol. 2013;36:292–301. doi: 10.1007/s00270-012-0526-6. [DOI] [PubMed] [Google Scholar]
- 33.Landelijke evaluatie van bevolkingsonderzoek naar borstkanker in Nederland 1990 – 2011/2012 [http://www.rivm.nl/dsresource?objectid=e50f46e6-c439-4f9e-9383-a972c8327793&type=org&disposition=inline]
- 34.Treatment of breast cancer in IKNO region, disaggregated into tumour size. [http://www.cijfersoverkanker.nl]
- 35.Flobbe K, Kessels AG, Severens JL, Beets GL, de Koning HJ, von Meyenfeld MF, van Engelshoven JM. Costs and effects of ultrasonography in the evaluation of palpable breast masses. Int J Technol Assess Health Care. 2004;20:440–448. doi: 10.1017/S0266462304001333. [DOI] [PubMed] [Google Scholar]
- 36.LPRM and NABON . Behandelingsrichtlijnen radiotherapie van het operabele mammacarcinoom na mamma-amputatie en okselkliertoilet in Nederland. 2000. pp. 1–13. [Google Scholar]
- 37.Oostenbrink JB, Bouwmans CAM, Koopmanschap MA, Rutten FFH. Handleiding voor kostenonderzoek, methoden en standaard kostprijzen voor economische evaluaties in de gezondheidszorg. 2004. [Google Scholar]
- 38.Universitair Medisch Centrum Groningen. Tarieven onderlinge dienstverlening. 2016
- 39.Passantenprijslijst UMC Utrecht - Overige zorgproducten.
- 40.Zorginstituut Nederland . Richtlijn voor het uitvoeren van economische evaluaties in de gezondheidszorg. 2016. [Google Scholar]
- 41.Zorginstituut Nederland . Richtlijn voor het uitvoeren van economische evaluaties in de gezondheidszorg. 2016. [Google Scholar]
- 42.Anglian Breast Cancer Study Group Prevalence and penetrance of BRCA1 and BRCA2 mutations in a population-based series of breast cancer cases. Anglian Breast Cancer Study Group. Br J Cancer. 2000;83:1301–1308. doi: 10.1054/bjoc.2000.1407. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Berg WA, Zhang Z, Lehrer D, Jong RA, Pisano ED, Barr RG, Bohm-Velez M, Mahoney MC, Evans WP, 3rd, Larsen LH, et al. Detection of breast cancer with addition of annual screening ultrasound or a single screening MRI to mammography in women with elevated breast cancer risk. JAMA. 2012;307:1394–1404. doi: 10.1001/jama.2012.388. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Chung A, Gangi A, Amersi F, Bose S, Zhang X, Giuliano A. Impact of Consensus Guidelines by the Society of Surgical Oncology and the American Society for Radiation Oncology on Margins for Breast-Conserving Surgery in Stages 1 and 2 Invasive Breast Cancer. Ann Surg Oncol. 2015;22(Suppl 3):S422–7. doi: 10.1245/s10434-015-4829-0. [DOI] [PubMed] [Google Scholar]
- 45.Clemons M, Danson S, Hamilton T, Goss P. Locoregionally recurrent breast cancer: incidence, risk factors and survival. Cancer Treat Rev. 2001;27:67–82. doi: 10.1053/ctrv.2000.0204. [DOI] [PubMed] [Google Scholar]
- 46.Darby S, McGale P, Correa C, Taylor C, Arriagada R, Clarke M, Cutter D, Davies C, Ewertz M, Godwin J, et al. Effect of radiotherapy after breast-conserving surgery on 10-year recurrence and 15-year breast cancer death: meta-analysis of individual patient data for 10,801 women in 17 randomised trials. Lancet. 2011;378:1707–1716. doi: 10.1016/S0140-6736(11)61629-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Houssami N, Ciatto S, Macaskill P, Lord SJ, Warren RM, Dixon JM, Irwig L. Accuracy and surgical impact of magnetic resonance imaging in breast cancer staging: systematic review and meta-analysis in detection of multifocal and multicentric cancer. J Clin Oncol. 2008;26:3248–3258. doi: 10.1200/JCO.2007.15.2108. [DOI] [PubMed] [Google Scholar]
- 48.Mook S, Schmidt MK, Rutgers EJ, van de Velde AO, Visser O, Rutgers SM, Armstrong N, van’t Veer LJ, Ravdin PM. Calibration and discriminatory accuracy of prognosis calculation for breast cancer with the online Adjuvant! program: a hospital-based retrospective cohort study. Lancet Oncol. 2009;10:1070–1076. doi: 10.1016/S1470-2045(09)70254-2. [DOI] [PubMed] [Google Scholar]
- 49.Moran MS, Schnitt SJ, Giuliano AE, Harris JR, Khan SA, Horton J, Klimberg S, Chavez-MacGregor M, Freedman G, Houssami N, et al. Society of Surgical Oncology-American Society for Radiation Oncology consensus guideline on margins for breast-conserving surgery with whole-breast irradiation in stages I and II invasive breast cancer. J Clin Oncol. 2014;32:1507–1515. doi: 10.1200/JCO.2013.53.3935. [DOI] [PubMed] [Google Scholar]
- 50.Newman B, Mu H, Butler LM, Millikan RC, Moorman PG, King MC. Frequency of breast cancer attributable to BRCA1 in a population-based series of American women. JAMA. 1998;279:915–921. doi: 10.1001/jama.279.12.915. [DOI] [PubMed] [Google Scholar]
- 51.Plana MN, Carreira C, Muriel A, Chiva M, Abraira V, Emparanza JI, Bonfill X, Zamora J. Magnetic resonance imaging in the preoperative assessment of patients with primary breast cancer: systematic review of diagnostic accuracy and meta-analysis. Eur Radiol. 2012;22:26–38. doi: 10.1007/s00330-011-2238-8. [DOI] [PubMed] [Google Scholar]
- 52.Saadatmand S, Bretveld R, Siesling S, Tilanus-Linthorst MM. Influence of tumour stage at breast cancer detection on survival in modern times: population based study in 173 797 patients. BMJ. 2015;351:h4901. doi: 10.1136/bmj.h4901. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Torrance GW. Measurement of health state utilities for economic appraisal. J Health Econ. 1986;5:1–30. doi: 10.1016/0167-6296(86)90020-2. [DOI] [PubMed] [Google Scholar]
- 54.Baron P, Ries M, Deckers R, de Greef M, Tanttu J, Kohler M, Viergever MA, Moonen CT, Bartels LW. In vivo T2 -based MR thermometry in adipose tissue layers for high-intensity focused ultrasound near-field monitoring. Magn Reson Med. 2014;72:1057–1064. doi: 10.1002/mrm.25025. [DOI] [PubMed] [Google Scholar]
- 55.Baron P, Deckers R, Knuttel FM, Bartels LW. T1 and T2 temperature dependence of female human breast adipose tissue at 1.5 T: groundwork for monitoring thermal therapies in the breast. NMR Biomed. 2015;28:1463–1470. doi: 10.1002/nbm.3410. [DOI] [PubMed] [Google Scholar]
- 56.Diakite M, Odeen H, Todd N, Payne A, Parker DL. Toward real-time temperature monitoring in fat and aqueous tissue during magnetic resonance-guided high-intensity focused ultrasound using a three-dimensional proton resonance frequency T method. Magn Reson Med. 2014;72(1):178–87. doi: 10.1002/mrm.24900. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Todd N, Diakite M, Payne A, Parker DL. In vivo evaluation of multi-echo hybrid PRF/T1 approach for temperature monitoring during breast MR-guided focused ultrasound surgery treatments. Magn Reson Med. 2014;72:793–799. doi: 10.1002/mrm.24976. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.