Abstract
Purpose
To simultaneously assess reproducibility of three MRI transverse relaxation parameters (R2’, R2* and R2) for brain tissue oxygenation mapping and to assess changes in these parameters with inhalation of gases that increase and decrease oxygenation, to identify the most sensitive parameter for imaging brain oxygenation.
Materials and Methods
48 healthy subjects (25M, ages 35±8 yrs) were scanned at 3.0T, each with one of four gases (mildly and strongly hypercapnic and hypoxic) administered in a challenge paradigm, using a gas delivery setup designed for patient use. Cerebral blood flow mapping with arterial spin labeling, and simultaneous R2’, R2* and R2 mapping with Gradient-Echo Sampling of Free Induction Decay and Echo (GESFIDE) were performed. Reproducibility in air and gas-induced changes were evaluated using non-parametric analysis with correction for multiple comparisons.
Results
Our gas delivery setup achieved stable gas challenges as shown by physiological monitoring. Test-retest variability of R2’, R2*, and R2 were found to be 0.24 s−1 (8.6% of mean), 0.24 s−1 (1.3% of mean), and 0.15 s−1 (1.0% of mean), respectively. Strong hypoxia produced the most conclusive oxygenation-driven relaxation change, inducing increases in R2’ (25±13%, p=0.03), R2* (5±2%, p=0.02), and R2 (2±2%, NS).
Conclusion
We benchmarked the intra-scan test-retest variability in GESFIDE-based transverse relaxation rate mapping. Using a reliable framework for gas challenge paradigms, we recommend strong hypoxia for validating oxygenation mapping methods, and the use of tissue R2’ change, instead of R2* or R2, as a metric for studying brain tissue oxygenation using transverse relaxation methods.
Keywords: Oxygenation, transverse relaxation, R2’, hypercapnia, hypoxia
Introduction
Brain tissue oxygenation mapping provides invaluable information for studying normal brain physiology and function (1, 2), and evaluating pathological conditions including cerebrovascular diseases and tumors (3, 4). Oxygenation can be assessed with several MRI-based methods, including quantification of transverse spin relaxation rates R2’, R2*, and R2 (reciprocals, respectively, of relaxation time constants T2’, T2*, and T2; related as R2* = R2 + R2’). These relaxation rates are sensitive to the susceptibility effect of paramagnetic deoxyhemoglobin within the imaging voxel, as described by the blood oxygenation level dependent (BOLD) principle (5). Relaxometry samples all tissue compartments, including blood, extra-cellular and intra-cellular water (6). Transverse relaxometry is insensitive to RF field inhomogeneity, and can offer vast improvements over measuring only signal magnitude.
Due to the relative simplicity of measuring R2*, it is the most common relaxation rate marker for susceptibility changes in tissues. Several studies have used R2* mapping to study brain tissue oxygenation changes in cerebrovascular pathologies (7–9), though comparison against positron emission tomography did not find a relationship between abnormal oxygen metabolism and increased R2* (10), and measurements are confounded by blood volume factors (11). Blood R2 is also used to measure oxygenation in the brain (12), though measurement of localized tissue R2 (13) is relatively limited, partly because it is affected by factors unrelated to oxygenation, including vasogenic edema (14), gliosis (15) and blood compartment changes (16). In theory, R2’ is independent of these factors and has been proposed as a better measure of tissue susceptibility changes due to oxygenation (17).
In this study, we aimed to concurrently assess the intra-scan air-breathing reproducibility of tissue R2’, R2*, and R2 in the human brain, and to quantify changes in these parameters when the subject is breathing different strengths of hypercapnic and hypoxic gases. Based on our results, we aimed to ascertain which relaxation parameter is most suitable for imaging brain oxygenation, both in terms of the magnitude and polarity of changes observed during gas challenges, and whether the changes can be definitively attributed to oxygenation and perfusion.
Materials and Methods
Subject Population
With approval from Stanford University's Institutional Review Board, 48 normal volunteers (25 male, ages 33±6 yrs; 23 female, ages 36±10 yrs) were scanned in this prospective study with informed consent and HIPAA compliance. All were screened for history of cerebrovascular disease, anemia, heart disease, or lung diseases that might affect blood and brain oxygenation. Volunteers were instructed to abstain from consuming caffeine for 12 hours before the scan, given the known effects of caffeine on cerebral blood flow (CBF) (18). Volunteers were not instructed to abstain from smoking prior to the scan, though all scans took place on a tobacco-free campus.
Imaging Protocol
All subjects were scanned at 3.0T (MR750, GE Healthcare, Waukesha WI). The imaging protocol consisted of three identical epochs (denoted as “Air 1”, “Air 2” and “Gas” in Figure 1). Air 1 and Air 2 epochs were performed consecutively during air breathing, while the Gas epoch followed a transition period of 2-3 minutes after commencing gas breathing, to ensure stable gas environment and subject physiology. Each epoch included measurements of oxygenation-driven relaxation and CBF as described below.
Figure 1.

Imaging protocol. Initial anatomical imaging was performed using a T1-weighted brain volume (BRAVO) sequence. During initial air breathing, reproducibility of oxygenation (measured using 2D GESFIDE) and CBF (using ASL) was performed in succession to test reproducibility. Following this, one of four gases was administered to perturb oxygenation, and following a 2-3 minute period to allow brain oxygenation levels to stabilize, a final set of oxygenation and CBF imaging was performed. At the start of each epoch, high-order quadratic shimming was performed to minimize extrinsic susceptibility effects.
Imaging Sequences
At the start of the experiment, axial T1-weighted 3D SPGR BRAVO images (TR/TE/TI 9.2/3.7/400ms, flip angle 13°, voxel size 0.9×0.9×1.2mm3) were acquired for anatomic imaging and segmentation. At the start of each epoch, quadratic shim was performed to optimize magnetic field homogeneity, so that detected variation in relaxation rate can be attributed to intrinsic changes in tissue susceptibility.
Each epoch consisted of two sequences. Relaxometry was performed using the massively multi-echo 2D Gradient Echo Sampling of Free Induction Decay Echoes (GESFIDE) sequence (Figure 2) (19), which allowed simultaneous measurement of R2’, R2*, and R2. Parameters are: TESE/TR 100/2000ms, 40 echoes, TE 5-130ms, voxel size 1.9×1.9×1.5mm3, 14 slices with 1mm spacing and 34 mm-thick coverage superior to lateral ventricles, and scan duration 3min 30s. CBF was mapped using GE Healthcare product 3D pseudocontinuous arterial spin labeling (pcASL) sequence: TE/TR 10.6/4845ms, TL/PLD 1500/2025ms, stack-of-spirals fast spin-echo readout, resolution 3×3×4mm3, 36 slices with whole-brain coverage, and scan duration 4min 41s. Imaging parameters were chosen in accordance with current consensus guidelines (20).
Figure 2.
a) GESFIDE sequence's idealized signal evolution. Each circle represents a collected image, such that 40 echoes in total are acquired during one 3 min 30 s acquisition. b) Sequence diagram corresponding to the multiple echoes. c) Representative images corresponding to the echo times of the solid symbols on signal curve of a).
Gas Challenge Setup
Subjects were assigned to one of four gas groups, identified as mild hypercapnia (mHC: targeting change in end-tidal partial pressure of oxygen, ΔPetCO2=7mmHg), strong hypercapnia (sHC: targeting ΔPetCO2=10mmHg), mild hypoxia (mHO: targeting change in peripheral oxygen saturation, ΔSpO2=−5%), and strong hypoxia (sHO: targeting ΔSpO2=−10%). Target physiological changes were chosen to be comparable to reported values in literature. Observing from past experimentation with this mask, that delivered gases would be diluted by entrained room air by a factor of 1-2, we used the following premixed medical-grade gases to achieve desired conditions:
Medical air (control): 21% oxygen and 79% nitrogen;
Mild hypercapnia: 21% oxygen, 5% carbon dioxide, and 74% nitrogen;
Strong hypercapnia: 21% oxygen, 10% carbon dioxide, and 69% nitrogen;
Mild hypoxia: 14% oxygen and 86% nitrogen; and
Strong hypoxia: 10% oxygen and 90% nitrogen.
Our gas delivery setup prioritized comfort and safety, and was implemented based on its feasibility for in-clinic use in future studies. Compressed medical air and one of the four gases were delivered at approximately 12 L/min via a bubble humidifier (Hudson RCI, Teleflex Inc., Research Triangle Park NC) and standard oxygen therapy tubing to each subject, using an MR-safe high oxygen delivery face mask (Ceretec Hi-Ox, Ceretec Inc., Garden Grove CA), as shown in Figure 3. A sampling nasal cannula (4000F, Salter Labs, Arvin CA) was worn under the mask. A custom elastic strap (Figure 3) with quick-release buckles was worn during scan. As expected, leakage occurred and was dependent on subjects’ facial geometry. In addition, for subject comfort and safety, we used a mask with an anti-suffocation valve, which entrained room air if the subject emptied the reservoir bag during deep inhalation or hyperventilation.
Figure 3.
a) Gas challenge hardware setup, with the Ceretec Hi-Ox mask separated into its components, and arrows indicating direct of gas flow. The subject breathes through the facemask. The anti-suffocation valve permitted inflow of room air only when there is a negative pressure differential in the mask relative to the room environment. b) Photo of the customized elastic strap that improves facemask seal.
Physiological Monitoring
Extensive physiological recording was performed, including inspired and end-tidal oxygen fractional concentration and carbon dioxide partial pressure (finO2, fetO2, PinCO2 and PetCO2) via nasal cannula, heart rate and peripheral oxygen saturation (SpO2) via fingertip photoplethysmography, and systolic, diastolic, and mean arterial pressure (MAP) from a brachial blood pressure cuff. All measurements were recorded by a medical monitor (Medrad Veris 8600, Bayer Healthcare, Whippany NJ) at 1s intervals.
Data Analysis
All data analysis was performed using custom Matlab (MathWorks Inc., Natick MA) code, including functions from SPM8 (Wellcome Trust Centre for Neuroimaging, United Kingdom) and FSL (Oxford Centre for Functional MRI of the Brain, United Kingdom) software packages. To avoid introducing errors through co-registration, all GESFIDE images were analyzed in their native space, with T1-weighted anatomical images co-registered (using SPM8) to each GESFIDE data set before undergoing segmentation (using FSL) into three regions of interest (ROIs): gray matter (GM), white matter (WM), and cerebrospinal fluid (CSF). pcASL images and CBF maps automatically generated (20) by the scanner (MR750, GE Healthcare, Waukesha WI) were also co-registered to corresponding GESFIDE data sets. R2’ and R2 maps were calculated from GESFIDE images using mono-exponential fitting of echoes from the “A” and “B” portions of the time curve illustrated in Figure 2, as per the standard GESFIDE method for R2’ mapping (19). R2* was calculated from the “A” portion.
Intra-scan reproducibility of oxygenation-driven relaxation and CBF measurements was evaluated from the two air epochs, and assumed to be applicable to gas breathing. Data sets were excluded from analysis if at least one GESFIDE data set from an air epoch displayed noticeable motion artifact, which manifested as Gibbs ringing or noticeable low-spatial frequency R2’ variations in a pattern (e.g., bands) that did not correspond to anatomy.
Statistical Analysis
Wilcoxon signed-rank test was used to test for significance in changes in all physiological and imaging measurements between epochs, such as between first and second air epochs, or between air and gas epochs. Wilcoxon rank-sum test was used in pairwise comparisons between gas groups. Multiple comparisons were corrected using the Holm-Bonferroni method, which controls family-wise error rate. The significance level of 0.05 was used for all tests. For reproducibility analysis, agreement between the medical air epoch measurements was measured using Lin's concordance correlation coefficient ρc (21) and Bland-Altman analysis.
Results
Subject Population
Several datasets with motion artifacts were discarded. The final data set consisted of 38 subjects (18 male, ages 34±7 yrs; 20 female, ages 34±9 yrs) for intra-scan reproducibility analysis, and 40 subjects (21 male, ages 33±7 yrs; 19 female, ages 34±9 yrs) for gas challenge analysis. Among reproducibility analysis data sets, 5 subjects were discarded as they were scanned on an outdated protocol with only one air epoch, and 5 others were discarded due to motion artifact during at least one air epoch. Among gas challenge data sets, 8 from the following groups had motion artifacts that visibly corrupted the images: mild hypoxia (2 subjects), strong hypercapnia (3 subjects), and strong hypoxia (3 subjects).
Gas Challenge Setup
Real-time recordings of SpO2, PinCO2, finO2, PetCO2 and fetO2 show that both normoxia and normocapnia were sustained during the air epochs, and that stable hypercapnia and hypoxia were indeed achieved during gas challenges (Figure 4 and Supplemental Figure S1). As expected, measured finO2 and PinCO2 deviated from the administered gas mixtures due to room air entrainment. Therefore, the achieved changes in end-tidal gas content were lower than expected for the gas mixtures: ΔPetCO2 of 10±3 mmHg (sHC) and 7±4 mmHg (mHC), and ΔSpO2 of −6±3% (mHO) and −11±3% (sHO). Recordings of heart rate, respiratory rate, and blood pressure also confirm that subject physiology was stable during air breathing, and changed in a consistent and expected manner during gas challenges (Figure 4 and Supplemental Figure S1).
Figure 4.
Changes in physiological measurements due to each gas challenge: strong hypercapnia (sHC), mild hypercapnia (mHC), mild hypoxia (mHO), and strong hypoxia (sHO).
Pairwise comparisons of changes in physiological measurements were performed between the four gas conditions (Table 2). Notable findings include: 1) ΔfetO2 was the only physiological change that significantly (p<0.05) differentiated all pairs of groups, indicating its sensitivity to both hypercapnic and hypoxic challenges; 2) strong hypercapnia more significantly increased heart rate and diastolic blood pressure compared to other challenges. Full results are shown in Table 2.
Table 2.
p-values for pairwise comparisons, between gas conditions (four gas challenges), of changes in physiological measurements.
| ΔSpO2 | ΔPinCO2 | ΔfinO2 | ΔPetCO2 | ΔfetO2 | ΔHeart Rate | ΔResp Rate | ΔSys BP | ΔDias BP | ΔMAP | |
|---|---|---|---|---|---|---|---|---|---|---|
| mHC-sHC | 0.49 | 0.16 | 0.001# | 0.03# | 0.001# | 0.01# | 0.94 | 1.00 | 0.02# | 0.73 |
| mHO-sHC | 0.001# | 0.001# | 0.001# | 0.001# | 0.001# | 0.01# | 1.00 | 1.00 | 0.003# | 0.45 |
| mHO-mHC | <0.001# | <0.001# | <0.001# | <0.001# | <0.001# | 0.99 | 1.00 | 1.00 | 0.83 | 1.00 |
| sHO-sHC | 0.001# | 0.001# | 0.001# | 0.001# | 0.001# | 0.99 | 1.00 | 1.00 | 0.04# | 0.45 |
| sHO-mHC | <0.001# | <0.001# | <0.001# | <0.001# | <0.001# | 0.05 | 0.94 | 1.00 | 1.00 | 1.00 |
| sHO-mHO | 0.02# | 0.52 | 0.001# | 0.06 | 0.001# | 0.09 | 1.00 | 1.00 | 1.00 | 1.00 |
indicates p<0.05.
All tests were corrected for multiple comparisons using the Holm-Bonferroni method.
Reproducibility
CBF was highly reproducible (Figure 5), with Lin's concordance correlation coefficient ρc of 0.95 in GM and 0.93 in WM. Among the relaxation rates, R2* and R2 had very high reproducibility (Figure 5), with ρc > 0.90 for GM and WM. R2’ had lower correlation (ρc=0.69 and 0.81 in GM and WM, respectively). The test-retest (Air 2 minus Air 1) differences between the two scans had overall standard deviation 3.6 mL/min/100 g or 6.7% (of mean value) for CBF, 0.24 s−1 or 8.6% for R2’, 0.24 s−1 or 1.3% for R2*, and 0.15 s−1 or 1.0% for R2. Only CBF experienced a significant (p=0.021) offset in the test-retest difference, but it was very small (0.97 mL/min/100 g).
Figure 5.
Correlation and Bland-Altman plots for CBF, R2’, R2* and R2 in air. R2’ had relatively lower reproducibility compared to R2* and R2, with the standard deviation of intra-scan difference of approximately 8.6% of mean. CBF had high reproducibility, though our measured baseline GM/WM CBF ratio was 157±6%—lower than previously reported pcASL measurements with different imaging parameters (39, 40).
Comparison of measurements during air breathing also allows quantification of inter-subject variation. We found large baseline variation in the first air epoch measurements across subjects for CBF (19% [GM] and 18% [WM] standard deviation) and R2’ (18% [GM] and 9% [WM]), and less so for R2* (3.9% [GM] and 2.8% [WM]) and R2 (4.4% [GM] and 2.1% [WM]). Inter-subject variation in the second air epoch measurements was similar. Two physiological measurements, fetO2 and PetCO2, showed significant change (p<0.01 for both) between the air epochs, though the changes were extremely small (−0.3% for fetO2 and +0.2mmHg for PetCO2), and is a result of the small variances due to the stable gas environment that allowed detection of these minimal changes. A full summary is presented in Supplemental Table 1.
Gas Challenges
Robust and significant (p=0.03) CBF increases were observed during hypercapnic challenges (Table 1). In GM, strong hypercapnia augmented CBF by 21±12%, and mild hypercapnia by 19±24%. We observed no significant change in CBF during hypoxia (3±5% with p=0.52 and 2±6% with p=1.00 in GM during mild and strong hypoxia, respectively). Changes in WM CBF were similar but had smaller mean and standard deviation compared to GM. Figure 6 illustrates two representative data sets during strong hypercapnia and hypoxia.
Table 1.
Mean and standard deviation for absolute and percentage ΔCBF, ΔR2′, ΔR2*, and ΔR2 in GM, along with p values assessing the significance of the values.
| sHC | mHC | mHO | sHO | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Mean±SD | p | Mean±SD | p | Mean±SD | p | Mean±SD | p | ||
| ΔCBF | mL/100 g/min | 14±9 | 0.03# | 10±7 | 0.03# | 2±3 | 0.52 | 1±4 | 1.00 |
| % | 21±12 | 0.03# | 19±24 | 0.03# | 3±5 | 0.42 | 2±6 | 1.00 | |
| ΔR2′ | s−1 | −0.02±0.26 | 1.00 | −0.08±0.30 | 1.00 | 0.28±0.31 | 0.16 | 0.53±0.24 | 0.03# |
| % | −1±12 | 1.00 | −3±15 | 1.00 | 13±13 | 0.16 | 25±13 | 0.03# | |
| ΔR2* | s−1 | −0.28±0.26 | 0.12 | −0.28±0.39 | 0.24 | 0.37±0.33 | 0.12 | 0.80±0.40 | 0.02# |
| % | −2±2 | 0.12 | −2±2 | 0.19 | 2±2 | 0.12 | 5±2 | 0.02# | |
| ΔR2 | s−1 | −0.26±0.14 | 0.03# | −0.20±0.17 | 0.03# | 0.09±0.13 | 0.14 | 0.27±0.25 | 0.10 |
| % | −2±1 | 0.03# | −1±1 | 0.03# | 1±1 | 0.14 | 2±2 | 0.10 | |
indicates p<0.05.
All tests were corrected for multiple comparison using the Holm-Bonferroni method.
Figure 6.
Representative data sets from a) strong hypercapnic and b) strong hypoxic gas challenges. Despite the low SNR of the GESFIDE data sets, we observe no reduction in R2’ in spite of CBF augmentation during hypercapnia, and elevation in R2’ with no change in CBF during hypoxia.
Gas-induced R2’ changes are shown in Figure 7 and Table 1. In GM, the hypercapnic challenges did not show significant changes (sHC: ΔR2’=−0.02±0.26 s−1, p=1.00; mHC: ΔR2’=−0.08±0.30 s−1, p=1.00), though the direction of the decrease would be consistent with increased tissue oxygenation. The two hypoxic challenges produced also produced dose-dependent R2’ increases (mHO: ΔR2’=0.28±0.31 s−1, p=0.16; sHO: ΔR2’=0.53±0.24 s−1, p=0.03). ΔR2* displayed expected polarities for both hypercapnic and hypoxic challenges, and was significant (p=0.02) for strong hypoxia. ΔR2 had expected polarity and dose-dependent magnitude in all challenges, and was only significant (p=0.03 for both) in hypercapnia.
Figure 7.
Bar charts illustrating ΔCBF, ΔR2’, ΔR2*, and ΔR2 (a: absolute change; b: percentage change) with error bars indicating standard deviation.
Relative percentage changes in measurements (denoted as %ΔCBF, %ΔR2’, Table 1) show similar trends, but also shows that R2’ experienced greater relative change than R2* and R2 in hypoxia (mHO: %ΔR2’ 13±13%, %ΔR2* 2±2%, and %ΔR2 1±1%; sHO: %ΔR2’ 25±13%, %ΔR2* 5±2%, and %ΔR2 2±2%). Measurements in WM show similar trends, though with smaller values and reduced statistical significance.
Pairwise comparison was performed for each measurement between all gas challenges. In GM, notable findings include: 1) there was no statistically significant difference between strong and mild hypercapnia, or between strong and mild hypoxia, although ΔR2’ magnitude suggested dose-dependence; 2) strong hypoxia induced a significant (p=0.01) increase in tissue ΔR2’ relative to the hypercapnic challenges; 3) tissue ΔR2* and ΔR2 were more sensitive to differences between gas challenges. Also, for hypercapnia, changes in R2* were largely accounted for by changes in R2. Full results are shown in Table 3.
Table 3.
All p-values for pairwise comparison between gas groups, for absolute and relative percentage ΔCBF, ΔR2′, ΔR2* and ΔR2 measurements in GM.
| ΔCBF | ΔR2′ | ΔR2* | ΔR2 | |||||
|---|---|---|---|---|---|---|---|---|
| Absolute | Relative | Absolute | Relative | Absolute | Relative | Absolute | Relative | |
| mHC vs. sHC | 0.77 | 0.42 | 0.57 | 0.62 | 0.85 | 0.85 | 0.43 | 0.43 |
| mHO vs. sHC | 0.01# | 0.01# | 0.07 | 0.10 | 0.01# | 0.01# | 0.00# | 0.00# |
| mHO vs. mHC | 0.01# | 0.01# | 0.07 | 0.10 | 0.01# | 0.01# | 0.00# | 0.00# |
| sHO vs. sHC | 0.01# | 0.01# | 0.01# | 0.01# | 0.00# | 0.00# | 0.00# | 0.00# |
| sHO vs. mHC | 0.01# | 0.01# | 0.00# | 0.01# | 0.00# | 0.00# | 0.00# | 0.00# |
| sHO vs. mHO | 0.77 | 0.73 | 0.13 | 0.18 | 0.05 | 0.09 | 0.21 | 0.21 |
indicates p<0.05.
All tests were corrected for multiple comparison using the Holm-Bonferroni method.
Discussion
This study successfully measured the intra-scan air-breathing reproducibility of CBF and all three transverse relaxation rates simultaneously, as well as the impact of different extents of hypercapnia and hypoxia on these measurements, using clinically friendly sequences and equipment. In doing so, we have produced valuable benchmarking information for future studies, supported some prior publications, and underlined many practical considerations in clinical applications.
Rigorous physiological monitoring in this study validated our gas challenge setup, which produced stable normoxic normocapnia and desired extents of hypercapnia and hypoxia. Intra-scan reproducibility studies indicate the minimum changes that can be confidently attributed to an external perturbation in an individual subject during a challenge experiment. During the air epochs, measured differences in CBF and relaxation parameters are attributed to fluctuations in unobserved physiological variables and noise from the imaging sequences. Our measured CBF variability was comparable to a previous study (22), while there are no published simultaneous assessment of the reproducibility of these three transverse relaxation rates. Our GESFIDE relaxometry method fits R2* and R2*B from the measured signal curve, and then calculates R2’ and R2. As a result, fitting error is additively propagated to both R2’ and R2 in identical ways. R2’ was found to have the lowest reproducibility, though we should note that ASL measurements of CBF had comparable variability. In addition, absolute test-retest differences in R2’ and R2* were comparable, so the difference in percentage reproducibility originated from different magnitudes of baseline measurement, with baseline R2’ being about 20-25% of R2* and R2. Moreover, despite our attempt to obtain a homogenous population by limiting enrolment to relatively young, normal volunteers, we also observed large inter-subject variation, which is also similar or lower than that of CBF measurements. Therefore, researchers investigating R2’ as a biomarker need to quantify and report reproducibility, and use intra-subject control data (e.g., contralateral measurements) wherever possible.
According to Yablonskiy and Haacke's quantitative BOLD (qBOLD) biophysical model (23), R2’ is modeled as proportional to the product of deoxygenated blood volume (DBV) and SO2. Therefore, if we were to calculate SO2 from independent measurements of R2’ and DBV, with intra- or inter-scan variability denoted by standard deviations δR2’% and δDBV% respectively, the corresponding variability in SO2 will range between min{δR2’, δDBV}% and (δR2’+ δDBV)%. In other words, the intra-scan test-retest difference of SO2 calculated using GESFIDE-measured R2’ will be at least 8.6% of the mean SO2 measurement. δDBV may be approximated with δCBV, the variability of cerebral blood volume (CBV). In literature, intra-scan test-retest δCBV has been measured to be around 22% in the whole brain for Dynamic Susceptibility Contrast (DSC) MRI at different field strengths (24), and to be 4-5% (25) but regionally highly variable (26) for the non-contrast vascular-space-occupancy (VASO) method.
Our measurements of hypercapnia-induced CBF increase were mostly consistent with values reported in literature (22, 27, 28), though ΔCBF per mmHg of ΔPetCO2 was unexpectedly lower in strong hypercapnia than in mild hypercapnia. On the other hand, prior studies reported a wide range of CBF changes in acute hypoxia, from small reductions to ~20% increases (29–31). In our work, we observed no significant CBF changes (p=0.05 and 1.00 for mild and strong hypoxia respectively), possibly due to the short duration of hypoxia.
During hypoxia, we observed positive and increasing ΔR2’ with increasing strength of challenge. This is consistent with theory. According to the BOLD principle, Yablonskiy and Haacke's biophysical model (23), and the Grubb relation between CBF and CBV (32), we expect R2’ to increase with CBF increase if SO2 remains constant, and to increase with SO2 decrease if CBF remains constant. Our observation was that the hypoxic challenge may have achieved the latter condition, where tissue SO2 decreases due to reduction in oxygen supply without concomitant CBF change.
However, our hypercapnic data found no significant R2’ changes (all p>0.05), possibly due to competing blood volume and oxygenation effects. During hypercapnia, CBF increases without any metabolic increases, leading to oversupply of oxygenated blood, increase in tissue SO2, and reduction in R2’. Concurrently, increasing CBF is expected to increase CBV via the Grubb relationship, leading to an increase in the blood volume term in the Yablonskiy and Haacke model, thereby reducing R2’. Since hypercapnia produced opposing effects of increased SO2 and increased blood volume, we believe that it is a suboptimal oxygenation challenge. An additional confounding factor is the Bohr effect during hypercapnia, where the lower pH would reduce SO2. Without blood sampling, we cannot quantify this competing effect. An alternative gas challenge for increasing tissue oxygenation is hyperoxia, which has been reported to cause insignificant reduction in CBF with appropriate T1 correction for ASL-based measurements (33), removing one of the opposing effects and resulting in a net R2’ reduction. We plan to test this challenge with our setup in the future.
Blood ΔR2 also experiences these competing effects (16), so we expect brain tissue ΔR2 to face similar problems. Our tissue ΔR2 measurements in hypercapnia suggest that ΔR2 is more weighted by SO2 change than ΔR2’, but ΔR2 in hypoxia was smaller than ΔR2’ despite the absence of blood volume change. This suggests that the oxygenation-based component of tissue R2 has a more complicated model than R2’. Due to the lack of theoretical foundation for tissue R2 modeling, it is limited in utility.
Additionally, due to the sensitivity of tissue R2 to non-oxygenation factors, baseline R2 is limited as a biomarker in the presence of pathology. Though this study is performed in normal volunteers, the findings contribute towards the knowledge base needed for translating relaxometry-based oxygenation mapping to patients.
By magnitude alone, ΔR2* is most sensitive, producing strongly polar responses to all four gas challenges. However, due to the dependence of ΔR2* (=ΔR2’ + ΔR2) on ΔR2, it is similarly poorly modeled and has limited utility in patients with brain pathologies.
Lastly, we note that percentage ΔR2’ was by far the largest during hypoxia. This suggests that relative R2’ change is potentially a good parameter for studying oxygenation changes in challenge paradigms where blood volume changes are either negligible or quantified to isolate the effect of oxygenation change. This is particularly true in focal changes, such as lesions in ischemic stroke, where pathology is likely to impact R2 and therefore R2*.
While we had a relatively large sample size, we did not perform multiple gas challenges in the same individual, so we cannot attribute inter-group differences to gas challenges alone. It is possible, though unlikely, that physiology of subjects could have been unbalanced at the group level, potentially confounding our results. We did not measure CBV change in this study, so we could not estimate its effect on spin relaxation. A non-contrast and therefore repeatable measurement of relative CBV is theoretically possible using the VASO technique (34). In pathologies, tissue composition may also change, altering the baseline measurements of relaxation rates and potentially affecting the relaxation rate changes due to challenges. Therefore, separate CBV quantitation is crucial in pathologies.
Additionally, in our study, we selected GESFIDE imaging parameters to prioritize high spatial resolution and short echo time spacing. The small voxel size resulted in low spatial SNR in the R2’ measurements, so that analysis was performed for large ROIs at group level, instead of voxel-wise in individual subjects, where imaging parameters would need to be optimized. Finally, we did not perform T1 mapping or correction in calculating CBF from pcASL data during different gas epochs and across subjects. Various modeling and experimental studies (35–38) have shown that blood T1 is strongly dependent on hematocrit, weakly but non-negligibly dependent on deoxyhemoglobin concentration, and strongly dependent on the amount of dissolved molecular oxygen (O2). The lack of hematocrit correction in this study likely added noise to our results. Although the amount of dissolved O2 is already low during air-breathing and would not be increased by either hypercapnia or hypoxia, the corresponding deoxyhemoglobin concentration changes may have biased the calculated CBF values. Specifically, during hypoxia, actual CBF may have been slightly higher than reported. This effect competes against the impact of decreasing oxygenation on transverse relaxation rates.
In summary, this study utilized a reliable setup for gas challenge paradigms, assessed the intra-scan reproducibility of transverse relaxation rate measurements for the first time, and demonstrated changes in these parameters with hypercapnia and hypoxia in a large cohort. From our measurements, we recommend the use of strong hypoxia for studying oxygenation mapping methods in healthy volunteers over hypercapnia, which creates competing effects and potentially inconclusive measurements. We also recommend the use of R2’ change, over R2* or R2, for studying brain tissue oxygenation. Future studies should concentrate on improving R2’ mapping sequences to evaluate focal oxygenation changes in individuals. T1 quantification should also be performed for quantitative oxygenation studies involving ASL, relaxometry, and any strong gas challenge.
Supplementary Material
Supplemental Figure S1: Physiological measurements during both baseline air epochs (“Air 1” and “Air 2”) and each gas challenge state (strong hypercapnia [sHC], mild hypercapnia [mHC], mild hypoxia [mHO], and strong hypoxia [sHO]).
Acknowledgments
We would like to acknowledge the support of Anne Sawyer, Gary Glover, Jarrett Rosenberg, and Thomas Brosnan at the Lucas Imaging Center, and John Hahesy at Stanford Health Care.
Grant support:
NIH 1R01-NS66506, P41-EB015891, R21-NS087491
References
- 1.Marchal G, Rioux P, Petit-Taboue M-C, et al. Regional cerebral oxygen consumption, blood flow, and blood volume in healthy human aging. Arch Neurol. 1992;49:1013–1020. doi: 10.1001/archneur.1992.00530340029014. [DOI] [PubMed] [Google Scholar]
- 2.Hoge RD, Atkinson J, Gill B, Crelier G, Marrett S, Pike B. Flow-metabolism regulation during brain activation and respiratory manipulations. Int Congr Ser. 2002;1235:33–38. [Google Scholar]
- 3.Kidwell CS, Alger JR, Saver JL. Beyond mismatch: evolving paradigms in imaging the ischemic penumbra with multimodal magnetic resonance imaging. Stroke. 2003;34:2729–35. doi: 10.1161/01.STR.0000097608.38779.CC. [DOI] [PubMed] [Google Scholar]
- 4.Brown JM, Wilson WR. Exploiting tumour hypoxia in cancer treatment. Nat Rev Cancer. 2004;4:437–47. doi: 10.1038/nrc1367. [DOI] [PubMed] [Google Scholar]
- 5.He X, Yablonskiy DA. Quantitative BOLD: mapping of human cerebral deoxygenated blood volume and oxygen extraction fraction: default state. Magn Reson Med. 2007;57:115–26. doi: 10.1002/mrm.21108. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Yablonskiy DA, Sukstanskii AL, He X. Blood oxygenation level-dependent (BOLD)-based techniques for the quantification of brain hemodynamic and metabolic properties - theoretical models and experimental approaches. NMR Biomed. 2013;26:963–86. doi: 10.1002/nbm.2839. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Tamura H, Hatazawa J, Toyoshima H, Shimosegawa E, Okudera T. Detection of Deoxygenation-Related Signal Change in Acute Ischemic Stroke Patients by T2*-Weighted Magnetic Resonance Imaging. Stroke. 2002;33:967–971. doi: 10.1161/01.str.0000013672.70986.e2. [DOI] [PubMed] [Google Scholar]
- 8.Wardlaw JM, Von Heijne A. Increased oxygen extraction demonstrated on gradient echo (T2*) imaging in a patient with acute ischaemic stroke. Cerebrovasc Dis. 2006;22:456–458. doi: 10.1159/000095384. [DOI] [PubMed] [Google Scholar]
- 9.Morita N, Harada M, Uno M, et al. Ischemic findings of T2*-weighted 3-tesla MRI in acute stroke patients. Cerebrovasc Dis. 2008;26:367–375. doi: 10.1159/000151640. [DOI] [PubMed] [Google Scholar]
- 10.Donswijk ML, Jones PS, Guadagno JV, et al. T2*-weighted MRI versus oxygen extraction fraction PET in acute stroke. Cerebrovasc Dis. 2009;28:306–313. doi: 10.1159/000229017. [DOI] [PubMed] [Google Scholar]
- 11.Christen T, Lemasson B, Pannetier N, et al. Is T2* Enough to Assess Oxygenation? Quantitative Blood Oxygen Level-Dependent Analysis in Brain Tumor. Radiology. 2012;262:495–502. doi: 10.1148/radiol.11110518. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Lu H, Ge Y. Quantitative evaluation of oxygenation in venous vessels using T2-Relaxation-Under-Spin-Tagging MRI. Magn Reson Med. 2008;60:357–63. doi: 10.1002/mrm.21627. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Wedegärtner U, Kooijman H, Andreas T, Beindorff N, Hecher K, Adam G. T2 and T2*measurements of fetal brain oxygenation during hypoxia with MRI at 3T: Correlation with fetal arterial blood oxygen saturation. Eur Radiol. 2010;20:121–127. doi: 10.1007/s00330-009-1513-4. [DOI] [PubMed] [Google Scholar]
- 14.Loubinoux I, Volk A, Borredon J, et al. Spreading of Vasogenic Edema and Cytotoxic Edema Assessed by Quantitative Diffusion and T2 Magnetic Resonance Imaging. Stroke. 1997;28:419–427. doi: 10.1161/01.str.28.2.419. [DOI] [PubMed] [Google Scholar]
- 15.Lansberg MG, Thijs VN, Brien MWO, et al. Evolution of Apparent Diffusion Coef cient, Diffusion-weighted, and T2-weighted Signal Intensity of Acute Stroke. Stroke. 2001;22:637–644. [PMC free article] [PubMed] [Google Scholar]
- 16.Anderson CM, Kaufman MJ, Lowen SB, Rohan M, Renshaw PF, Teicher MH. Brain T2 relaxation times correlate with regional cerebral blood volume. Magn Reson Mater Physics, Biol Med. 2005;18:3–6. doi: 10.1007/s10334-004-0076-2. [DOI] [PubMed] [Google Scholar]
- 17.Yang X, Cao J, Wang X, Li X, Xu Y, Jiang X. Evaluation of Renal Oxygenation in Rat by Using R2’ at 3-T Magnetic Resonance. Initial Observation. Acad Radiol. 2008;15:912–918. doi: 10.1016/j.acra.2008.01.015. [DOI] [PubMed] [Google Scholar]
- 18.Xu F, Liu P, Pekar JJ, Lu H. Does acute caffeine ingestion alter brain metabolism in young adults? Neuroimage. 2015;110:39–47. doi: 10.1016/j.neuroimage.2015.01.046. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Ni WW, Christen T, Zun Z, Zaharchuk G. Comparison of R2’ measurement methods in the normal brain at 3 tesla. Magn Reson Med. 2015;73:1228–1236. doi: 10.1002/mrm.25232. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Alsop DC, Detre JA, Golay X, et al. Recommended implementation of arterial spin-labeled perfusion MRI for clinical applications: A consensus of the ISMRM perfusion study group and the european consortium for ASL in dementia. Magn Reson Med. 2015;73:102–116. doi: 10.1002/mrm.25197. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Lin LI-K. A Concordance Correlation-Coefficient to Evaluate Reproducibility. Biometrics. 1989;45:255–268. [PubMed] [Google Scholar]
- 22.Heijtel DFR, Mutsaerts HJMM, Bakker E, et al. Accuracy and precision of pseudo-continuous arterial spin labeling perfusion during baseline and hypercapnia: A head-to-head comparison with (15)O H2O positron emission tomography. Neuroimage. 2014;92:182–192. doi: 10.1016/j.neuroimage.2014.02.011. [DOI] [PubMed] [Google Scholar]
- 23.Yablonskiy DA, Haacke EM. Theory of NMR signal behavior in magnetically inhomogeneous tissues: the static dephasing regime. Magn Reson Med. 1994;32:749–763. doi: 10.1002/mrm.1910320610. [DOI] [PubMed] [Google Scholar]
- 24.Grandin CB, Bol A, Smith AM, Michel C, Cosnard G. Absolute CBF and CBV measurements by MRI bolus tracking before and after acetazolamide challenge: repeatabilily and comparison with PET in humans. Neuroimage. 2005;26:525–35. doi: 10.1016/j.neuroimage.2005.02.028. [DOI] [PubMed] [Google Scholar]
- 25.Uh J, Lin A-L, Lee K, Liu P, Fox P, Lu H. Validation of VASO cerebral blood volume measurement with positron emission tomography. Magn Reson Med. 2011;65:744–749. doi: 10.1002/mrm.22667. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Landman BA, Huang AJ, Gifford A, et al. Multi-parametric neuroimaging reproducibility: A 3-T resource study. Neuroimage. 2011;54:2854–2866. doi: 10.1016/j.neuroimage.2010.11.047. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Faraco CC, Strother MK, Siero JC, et al. The cumulative influence of hyperoxia and hypercapnia on blood oxygenation and R2*. J Cereb Blood Flow Metab. 2015 Jan;35:2032–2042. doi: 10.1038/jcbfm.2015.168. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Tancredi FB, Lajoie I, Hoge RD. Test-retest reliability of cerebral blood flow and blood oxygenation level-dependent responses to hypercapnia and hyperoxia using dual-echo pseudo-continuous arterial spin labeling and step changes in the fractional composition of inspired gases. J Magn Reson Imaging. 2015;42:1144–1157. doi: 10.1002/jmri.24878. [DOI] [PubMed] [Google Scholar]
- 29.Peebles KC, Richards AM, Celi L, McGrattan K, Murrell CJ, Ainslie PN. Human cerebral arteriovenous vasoactive exchange during alterations in arterial blood gases. J Appl Physiol. 2008;105:1060–1068. doi: 10.1152/japplphysiol.90613.2008. [DOI] [PubMed] [Google Scholar]
- 30.Xu F, Liu P, Pascual JM, Xiao G, Lu H. Effect of hypoxia and hyperoxia on cerebral blood flow, blood oxygenation, and oxidative metabolism. J Cereb Blood Flow Metab. 2012;32:1909–18. doi: 10.1038/jcbfm.2012.93. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Vestergaard MB, Lindberg U, Aachmann-andersen NJ, et al. Acute hypoxia increases the cerebral metabolic rate –a magnetic resonance imaging study. J Cereb Blood Flow Metab. 2015 doi: 10.1177/0271678X15606460. in press.:10.1177/0271678X15606460. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Grubb RL, Raichle ME, Eichling JO, Ter-Pogossian MM. The Effects of Changes in PaCO2 Cerebral Blood Volume, Blood Flow, and Vascular Mean Transit Time. Stroke. 1974;5:630–639. doi: 10.1161/01.str.5.5.630. [DOI] [PubMed] [Google Scholar]
- 33.Zaharchuk G, Martin AJ, Dillon WP. Noninvasive imaging of quantitative cerebral blood flow changes during 100% oxygen inhalation using arterial spin-labeling MR imaging. AJNR Am J Neuroradiol. 2008;29:663–7. doi: 10.3174/ajnr.A0896. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Lu H, Law M, Johnson G, Ge Y, Van Zijl PCM, Helpern JA. Novel approach to the measurement of absolute cerebral blood volume using vascular-space-occupancy magnetic resonance imaging. Magn Reson Med. 2005;54:1403–1411. doi: 10.1002/mrm.20705. [DOI] [PubMed] [Google Scholar]
- 35.Silvennoinen MJ, Kettunen MI, Kauppinen R a. Effects of hematocrit and oxygen saturation level on blood spin-lattice relaxation. Magn Reson Med. 2003;49:568–71. doi: 10.1002/mrm.10370. [DOI] [PubMed] [Google Scholar]
- 36.Blockley NP, Jiang L, Gardener a. G, Ludman CN, Francis ST, Gowland PA. Field strength dependence of R1 and R2* relaxivities of human whole blood to ProHance, Vasovist, and deoxyhemoglobin. Magn Reson Med. 2008;60:1313–20. doi: 10.1002/mrm.21792. [DOI] [PubMed] [Google Scholar]
- 37.Hales PW, Kirkham FJ, Clark CA. A general model to calculate the spin-lattice (T1) relaxation time of blood, accounting for haematocrit, oxygen saturation and magnetic field strength. J Cereb Blood Flow Metab. 2016;36:370–374. doi: 10.1177/0271678X15605856. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Li W, Grgac K, Huang A, Yadav N, Qin Q, van Zijl PCM. Quantitative theory for the longitudinal relaxation time of blood water. Magn Reson Med. 2016;76:270–281. doi: 10.1002/mrm.25875. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Xu G, Rowley H a, Wu G, et al. Reliability and precision of pseudo-continuous arterial spin labeling perfusion MRI on 3.0 T and comparison with 15O-water PET in elderly subjects at risk for Alzheimer's disease. NMR Biomed. 2010;23:286–93. doi: 10.1002/nbm.1462. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Wu B, Lou X, Wu X, Ma L. Intra- and interscanner reliability and reproducibility of 3D whole-brain pseudo-continuous arterial spin-labeling MR perfusion at 3T. J Magn Reson Imaging. 2014;39:402–409. doi: 10.1002/jmri.24175. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supplemental Figure S1: Physiological measurements during both baseline air epochs (“Air 1” and “Air 2”) and each gas challenge state (strong hypercapnia [sHC], mild hypercapnia [mHC], mild hypoxia [mHO], and strong hypoxia [sHO]).






