Polymer-free dual drug-eluting stents evaluated in a porcine model

Bin Zhang 1, Bo Zheng 1, Xingang Wang 1, Qiuping Shi 1, Jia Jia 1, Yong Huo 1, Chunshui Pan 2,3, Jingyan Han 2,3,, Ming Chen 1,
PMCID: PMC5558731  PMID: 28810900

Abstract

Background

Although drug-eluting stents have dramatically reduced the rates of restenosis and target lesion revascularization, they are associated with stent thrombosis (ST), a catastrophic event likely due to delayed endothelialization and chronic inflammation caused by the polymer and the metal scaffolds. To increase the safety and efficacy of stents, polymer-free dual drug-eluting stents (DDES) have been developed.

Methods

A total 160 stents (Bare-metal stents (BMS), polymer-free probucol stents (PrES), sirolimus-eluting stents (SES) and DDES) were randomly implanted in the coronary arteries of 80 pigs. 14, 28, 90 and 191 days after implantation, QCA and OCT evaluations were performed in 20 pigs respectively, and the arteries were harvested for scanning electron microscope (SEM), histomorphology, histopathology analyses and for the relative expression of CD31, CD34 and CD133 on mRNA and protein levels.

Results

At the 14-day time point, there were significant differences in the strut rate coverage (p = 0.011), with greater coverage in the PrES than in the SES group (53.2%vs. 20.3%, p = 0.002). As well, there were no significant differences in the expression of CD31, CD34 and CD133 between groups in mRNA and protein level.

Conclusions

DDES were as safe as BMS and SES, but they did not further improve the endothelialization of the stented coronary arteries in the porcine model.

Keywords: Stent thrombosis, Endothelialization, Probucol, Optical coherence tomography, Porcine

Background

Coronary heart disease is one of the leading causes of mortality and morbidity worldwide. Noninvasive percutaneous coronary intervention (PCI) with stenting has become the standard of care, but delayed endothelialization and chronic inflammation caused by the polymer and the metal scaffolds lead to the development of stent restenosis (SR) and stent thrombosis (ST), the most important clinical problems following PCI. In an effort to decrease the incidence and consequences of these complications, bioresorbable coronary scaffolds have been developed albeit their use did not further improve short-term outcomes [1, 2].Polymer-free dual drug-eluting stents (DDES) may be a more effective approach.

Sirolimus (rapamycin) and its derivatives are still the most used antiproliferative agents in contemporary drug-eluting stents despite the fact they are associated with delayed endothelialization of the stents because of the inhibition of endothelial cell (EC) and endothelial progenitor cell (EPC) proliferation and differentiation [3, 4]. Indeed, autopsy studies have shown that the best morphometric predictor of late ST was the ratio of uncovered struts relative to the total stent struts, which is a marker of stent endothelialization [5]. Probucol is not only a potent antioxidant and lipid-lowering drug, it is also effective in reducing the risk of restenosis and the incidence of major adverse cardiac events after PCI [6]. Indeed, several studies, both in animals and humans, have shown some improvement in the rate of endothelialization of the stented arteries by probucol [79]. However, few preclinical studies have focused on the endothelialization of stented arteries associated with novel polymer-free DDESs. Consequently, the aim of the present study was to evaluate the efficacy and safety of the polymer-free stent-based delivery system of probucol alone compared with rapamycin, and probucol in combination with rapamycin in a porcine model of coronary artery stenting.

Methods

Stents

The stents platform and drug loading were described previously [10]. Four types of stents were used in this study shared the same metallic platform: 316 L stainless steel bare-metal stents (BMS), polymer-free probucol-eluting stents (PrES) loaded with 0.8 μg/mm2 probucol, polymer-free sirolimus-eluting stents (SES) loaded with 1.6 μg/mm2 sirolimus and the novel polymer-free DDES coated with 1.6 μg/mm2 sirolimus and 0.8 μg/mm2 probucol. All stents were purchased from Lepu Medical (Beijing, China).

Animals

The study was approved by the Institutional Animal Care and Use Committee of Peking University First Hospital, and conformed to the National Institutes of Health Guide for the Care and Use of Laboratory Animals. Chinese miniature swines of either sex (25 kg to 35 kg, 6 to 8 months of age) were provided by the China Agricultural University. All animals were fed a standard laboratory chow diet without lipid throughout the course of the study. In order to decrease the incidence of acute thrombosis, premedication with 300 mg aspirin (Bayer, Germany) and 75 mg clopidogrel (Xin Li Tai Pharmaceutical Co Ltd., China) was administered before stent implantation for one day. Aspirin (100 mg/d) and clopidogrel (75 mg/d) were then given until sacrificed.

Stent implantation and experimental groups

A total of 160 stents (BMS, PrES, SES and DDES) were randomly implanted in the left anterior descending, circumflex or right coronary arteries of 80 pigs (two types of stents per two arteries per pig) (Table 1). The anesthesia and stent implantation procedures have been described previously [11].

Table 1.

Stents distribution in four endpoints included in the analysis

14 day 28 day 90 day 191 day
LAD LCX RCA LAD LCX RCA LAD LCX RCA LAD LCX RCA
BMS 7 2 0 3 5 1 5 3 0 4 1 4
PrES 5 5 0 5 4 0 3 4 1 4 5 0
SES 2 5 2 5 2 1 3 3 2 5 4 0
DDES 5 4 1 4 4 0 5 3 0 5 1 3

BMS Bare-Metal Stents, PrES Probucol-Eluting Stents, SES Sirolimus-Eluting Stents, DDES Dual Drug-Eluting Stents, LAD Left Anterior Descending, LCX Left Circumflex, RCA Right Coronary Arteries

Quantitative coronary angiography (QCA) evaluation

At 14, 28, 90 and 180 days following stent implantation, repeat QCA were performed in 20 animals using the CAAS 5.9 QCA Software respectively (PIEMEDICAL IMAGINE, The Netherlands). Coronary artery measurements included baseline vessel diameter, minimal lumen diameter immediately after implantation (I-MLD), reference vessel diameter (RVD), MLD at repeat angiography (R-MLD) and lumen loss (LL).

Optical coherence tomography (OCT) imaging and evaluation

The OCT evaluation of the stented vessels was performed 14 and 28 days immediately after the repeat QCA (n = 7 in each group) at a pullback speed of 2.5 cm/s using a commercially available OCT system (C7-XR Dragonfly System, St. Jude Medical, USA). Cross-sectional OCT images were analyzed by frame measurements. Lumen area, internal elastic lamina area, total struts and covered struts were assessed. Neointimal area (NA) and the ratio of covered struts (CSR) were then calculated [5].

Histopathological evaluation

At 14, 28, 90 and 180-days following stent implantation, twenty animals in each group were sacrificed, and the hearts were removed and perfused with heparin saline for five minutes at a pressure of 100 mmHg (1 mmHg = 0.0133 kPa). The stented vessel segments were separated into three pieces on ice, and the segment with the higher restenosis rate (according to the QCA results) was fixed with 10% formaldehyde solution and embedded in methyl methacrylate plastic prepared by the Peking University School of Stomatology for histological examination [11]. The other segments were stored in freezing tubes and stored at −80 °C for subsequent molecular biology study (western blot and real time-polymerase chain reaction (PCR)).

Scanning electron microscopy (SEM) image acquisition and evaluation

Scanning electron microscopy was performed on longitudinal sections harvested 14 and 28 days following the OCT imaging. The hearts were removed and perfused as described above. The stented vessel segments were separated and fixed with 3% buffered glutaraldehyde and 1% buffered osmium tetroxide (n = 1 for each group). The samples were then dehydrated in a series of ethanol baths (50%, 75% and 100%), dried in liquid CO2 in a critical point dryer (72.8 atm, 31 °C), and sputter-coated for 3 min at 15 mA with Au. En face SEM images (H-450, Hitachi, Japan) were acquired at low magnification (×18 magnification) to evaluate the overall neointimal coverage of the stents, and at various high magnifications to identify the composition of the tissue covering the surfaces of the stents. Endothelial cells were identified as sheets of closely connected monolayer cells with a spindle or polygonal shape [12].

Western blot and real time PCR evaluation

All the samples stored for the molecular biology study were divided into two parts, one to extract the total protein content for Western blot analysis and the other to extract the total RNA content for Real time PCR. At the four time points (14, 28, 90 and 180 days following stent implantation), the stented segments of all stent types were evaluated for the expression of CD31, CD34 and CD133. Protein extracts were fractionated on SDS ployacrylamide gels, transferred to PVDF membranes, incubated with a purified affinity polyclonal antibody to CD31 (Earthox: E021010, USA), CD34 (Santa Cruz Biotechnology: ab80508, USA), CD133 (Abcam: ab109366, USA) and glyceraldehyde-3-phosphate dehydrogenase (GAPDH) (Earthox: E021010, USA), washed and incubated with a second antibody. Signals were detected using ECL chemiluminesence detection system, and autoradiographic signal were quantified by densitometry. The relative expression level of the markers was determined by the comparison of band intensities on bolts to GAPDH, a housekeeping protein employed as an internal control. Total RNA of each stented segment was isolated using an RNA extraction Kit (Roche: 12,033,674,001, Germany).

Total cDNA, synthesized by reverse transcription from total RNA using Reverse Transcription Kit (Roche: 05081955001, Germany), was amplified by PCR for 35 cycles at 50 °C for 2 min, 95 °C for 10 min, and 58 °C for 1 min using SYBR Green/ROX qPCR Master Mix (Roche: 06924204001, Germany) on Fluorescent Quantitative PCR (Opticon 2 MJ Research, Bio-Rad, USA). The PCR primer sets for CD31 (5′-CACCGAGGTCTGGGAACAAA-3′ and 5′-CTGCGGTCCTAAGTCCCATC-3′), CD34 (5′-GTCTTGGCCAACGGAACAGA-3’and 5′-GTCTTCGCCCAGCCTTTCTC-3′), CD133 (5′-AATGCCTCTGGTGGGGCTTA-3′ and 5′-CACCAAGAGGGAAACGGCAA-3′) and GAPDH (5′-CGATGGTGAAGGTCGGAGTG-3′ and 5′-TGCCGTGGGTGGAATCATAC-3′) were used for amplification.

Statistical analysis

Values in normal distribution were expressed as mean ± standard deviation and non-normal distributions were expressed as median and interquartile range. Group imaging, histological and integral data were analyzed with a one-way ANOVA test (normal distribution values) or Kruskal-Wallis rank test (non-normal distribution values). N represents the number of stents of different types or samples. Comparisons within and between groups were performed using paired t test (normal distribution values) and Wilcoxon’s signed-rank test (non-normal distribution values). The value of p < 0.05 was considered statistically significant. All statistical analyses were performed with SPSS 19.0 system software (IBM, USA).

Results

A total of 160 stents were successfully implanted in the coronary arteries of 80 pigs. Overall, 7 animals died during the course of the study; 2 of respiratory failure (1 in the 14-day group and 1 in the 28-day group) which was most likely due to the anesthesia, 3 of post-operative puncture site bleeding (2 in the 28-day group and 1 in the 90-day group), 1 of pneumonia (at 71 days follow-up) and 1 of ileus (at 180 days follow-up). None of the animals showed any evidence of ST or myocardial infarction. The latter 2 animals were included in the histological analysis. All other animals survived the intended study follow-up period without any angiographic ST or clinical complications. Stents distribution included in the analysis was showed in Table 1.

Quantitative coronary angiography evaluation

From baseline, the RVD, expansion ratio and the LL remained similar among the groups at all the four endpoints (Fig. 1 and Table 2).

Fig. 1.

Fig. 1

Lumen loss (LL) at all four endpoints. QCA: Quantitative Coronary Angiography; LL: Lumen Loss; BMS: Bare-Metal Stents; PrES: Probucol-Eluting Stents; SES: Sirolimus-Eluting Stents; DDES: Dual Drug-Eluting Stents

Table 2.

Quantitative coronary angiography evaluation at all the four endpoints

RVD (mm) P ER P LL (mm) P DS% P
14 day
BMS (n = 9) 2.60 ± 0.22 0.061 1.13 ± 0.14 0.067 0.32 ± 0.11 0.655 12.43 ± 4.50 0.429
PrES (n = 10) 2.41 ± 0.31 1.25 ± 0.17 0.43 ± 0.29 19.12 ± 14.87
SES (n = 9) 2.62 ± 0.28 1.17 ± 0.08 0.34 ± 0.21 12.99 ± 7.61
DDES(n = 10) 2.36 ± 0.16 1.27 ± 0.09 0.36 ± 0.18 15.46 ± 8.07
28 day
BMS (n = 8) 2.55 ± 0.39 0.994 1.19 ± 0.19 0.946 0.41 ± 0.39 0.409 18.26 ± 20.64 0.365
PrES (n = 8) 2.56 ± 0.31 1.16 ± 0.09 0.28 ± 0.21 10.34 ± 6.93
SES (n = 8) 2.55 ± 0.19 1.18 ± 0.12 0.18 ± 0.15 6.94 ± 5.76
DDES(n = 8) 2.52 ± 0.38 1.20 ± 0.14 0.37 ± 0.32 15.90 ± 15.14
90 day
BMS (n = 8) 2.37 ± 0.30 0.945 1.26 ± 0.13 0.662 0.76 ± 0.36 0.767 32.71 ± 16.40 0.809
PrES (n = 8) 2.30 ± 0.22 1.29 ± 0.08 0.85 ± 0.44 37.19 ± 18.71
SES (n = 8) 2.30 ± 0.27 1.29 ± 0.10 0.83 ± 0.34 35.45 ± 15.47
DDES(n = 8) 2.34 ± 0.18 1.24 ± 0.10 0.96 ± 0.35 41.66 ± 15.79
191 day
BMS (n = 9) 2.41 ± 0.30 0.335 1.28 ± 0.16 0.368 0.34 ± 0.31 0.079 15.76 ± 16.68 0.811
PrES (n = 9) 2.31 ± 0.29 1.24 ± 0.14 0.62 ± 0.34 27.44 ± 16.22
SES (n = 9) 2.31 ± 0.19 1.30 ± 0.10 0.41 ± 0.25 18.01 ± 11.28
DDES(n = 9) 2.52 ± 0.31 1.19 ± 0.11 0.72 ± 0.36 29.70 ± 15.11

RVD reference vessel diameter, ER expansion ratio, LL lumen loss, DS% diameter stenosis %, BMS Bare-Metal Stents, PrES Probucol-Eluting Stents, SES Sirolimus-Eluting Stents, DDES Dual Drug-Eluting Stents

Optical coherence tomography evaluation

At 14- and 28-day time points, there were no significant differences in the NA between the four groups (Fig. 2a,b and Table 3), with no obvious ST, stent malapposition, dissection or prolapsed. At the 14-day time point, there was significant difference in the CSR between the four groups (p = 0.011), the CSR in the PrES group was greater than in SES group (53.2%, interquartile: 39.0% -81.0% vs. 20.3%, interquartile: 13.5% -26.0%, p = 0.002) (Fig. 2c and Table 3). However, a similar tendency was not observed at the 28-day time point.

Fig. 2.

Fig. 2

a Neointimal area (NA) at 14-day and 28-day by optical coherence tomography. OCT: Optical Coherence Tomography; NA: Neointimal Area; BMS: Bare-Metal Stents; PrES: Probucol-Eluting Stents; SES: Sirolimus-Eluting Stents; DDES: Dual Drug-Eluting Stents. b Covered struts rate (CSR) at 14-day and 28-day by optical coherence tomography. *represents CSR in PrES group (53.2%, interquartile: 39.0%–81.0%) was statistically higher than in the SES group (20.3%, interquartile: 13.5%–26.0%), (P = 0.002). OCT: Optical Coherence Tomography; CSR: Covered Struts Rate; BMS: Bare-Metal Stents; PrES: Probucol-Eluting Stents; SES: Sirolimus-Eluting Stents; DDES: Dual Drug-Eluting Stents

Table 3.

Optical Coherence Tomography Evaluation at 14 and 28-day

NA (mm2) P CSR% P
14 day
BMS (n = 8) 1.46 (1.35,1.69) 0.708 37.67 (26.74,50.65) 0.011
PrES (n = 9) 1.56 (1.35,1.88) 53.24 (38.99,81.05)
SES (n = 7) 1.52 (1.16,1.61) 20.27 (13.49,25.95)
DDES(n = 8) 1.38 (1.34,1.56) 40.06 (14.12,50.16)
28 day
BMS (n = 7) 2.57 (1.75,2.98) 0.300 93.64 (89.07,97.79) 0.644
PrES (n = 7) 2.23 (1.57,3.15) 92.77 (50.00,98.23)
SES (n = 7) 1.91 (1.49,2.30) 85.15 (84.94,100.00)
DDES(n = 6) 1.68 (1.54,2.40) 88.22 (64.24,94.56)

NA Neointimal Area, CSR% Ratio of Covered Struts, BMS Bare-Metal Stents, PrES Probucol-Eluting Stents, SES Sirolimus-Eluting Stents, DDES Dual Drug-Eluting Stents

Histology evaluation

At the four time points, there were no significant differences in the NA, and injury, inflammation and endothelialization scores between the four groups (Fig. 3).

Fig. 3.

Fig. 3

Neointimal area (NA) at 14-day, 28-day, 90-day and 191-day by hematoxylin and eosin staining. NA: Neointimal Area; BMS: Bare-Metal Stents; PrES: Probucol-Eluting Stents; SES: Sirolimus-Eluting Stents; DDES: Dual Drug-Eluting Stents

Scanning electron microscope evaluation

At the 14- and 28-day time points, the rate of covered struts was observed at low magnification and the rate of covered struts by endothelial cells was observed at high magnification. The former was ranked as PrES > DDES > BMS > SES, and the latter was ranked as PrES > BMS > DDES > SES at 14 days (Fig. 4a). At 28 days, the rate of covered struts observed at low magnification was ranked as BMS > SES > PrES > DDES, while it was ranked as PrES > DDES > SES > BMS at high magnification (Fig. 4b).

Fig. 4.

Fig. 4

a Scanning electron microscopy (SEM) image at 14-day (×18). BMS: Bare-Metal Stents; PrES: Probucol-Eluting Stents; SES: Sirolimus-Eluting Stents; DDES: Dual Drug-Eluting Stents. b Scanning electron microscopy (SEM) image at 28-day (×18). BMS: Bare-Metal Stents; PrES: Probucol-Eluting Stents; SES: Sirolimus-Eluting Stents; DDES: Dual Drug-Eluting Stents. c Scanning electron microscopy (SEM) image at 28-day in the BMS group (×130). Black rectangle represents the region of d. d Scanning electron microscopy (SEM) image at 28-day in BMS group (×600). Black circle represents smooth muscle cells without extracellular matrix and endothelial cells. White circle represents the site completely covered by endothelial cells. Hexagrams represent extracellular matrix. e Scanning electron microscopy (SEM) image at 28-day in the PrES group (×18 left) and ×1000 right). White rectangles in the left image represent the area of the right image that is covered by endothelial cells, except for a small amount of extracellular matrix

Western blot and real time polymerase chain reaction

At the four time points, there were no significant differences in the expression of CD31, CD34 and CD133 between the four groups in terms of mRNA (Fig. 5) and protein level (Fig. 6).

Fig. 5.

Fig. 5

mRNA levels of CD31,CD34 and CD133 at 14-day, 28-day, 90-day and 191-day. BMS: Bare-Metal Stents; PrES: Probucol-Eluting Stents; SES: Sirolimus-Eluting Stents; DDES: Dual Drug-Eluting Stents

Fig. 6.

Fig. 6

Protein expression of CD31, CD34 and CD133 at 14-day, 28-day, 90-day and 191-day. BMS: Bare-Metal Stents; PrES: Probucol-Eluting Stents; SES: Sirolimus-Eluting Stents; DDES: Dual Drug-Eluting Stents; NV: Normal Vessel

Discussion

Improvements in metal scaffolds, polymers and antiproliferative drugs of second generation DES have resulted in increased biocompatibility. However, they are still associated with a high incidence of SR and ST and, although polymer-free DES has been shown to be safe and efficacious in both animals [1315] and humans [1621], they are not routinely used in contemporary clinical practice. Animals study can effectively evaluate the safety of stents, but not their effectiveness. Clinical trials conducted over the past years were not large enough and lacked control groups to conclusively demonstrate the efficacy of polymer-free DES. Because of this, interventional cardiologists are often reluctant to use them despite their potential benefits. There are a few studies that have demonstrated that the antiproliferative drugs contained in the polymers could theoretically be lost in the process of transportation and release, reducing their long-term safety and short-term effectiveness. A coating of rapamycin or its derivatives on the surface of stents greatly reduces the rate of SR and target vessel revascularization, but not the occurrence of ST, especially late events. Because delayed endothelialization is believe to play a key role in the occurrence of late ST, research has attempted to improve the stent platform and the drug loading as well as drugs that may improve stent endothelialization by virtue of their antioxidant and anti-inflammatory properties. Probucol is such a drug, and several large clinical studies showed the non-inferiority of the polymer-free probucol/rapamycin DDES compared with the first and second generation polymer-based DES on the efficacy and safety [2224]. However, a multi-center clinical study conducted in China showed that the same stent was associated with a higher incidence of restenosis and target vessels revascularization than the biodegradable polymer DES used as controls [25]. On the other hand, an animal study showed that succinobucol (a probucol analogue) eluting stents significantly increased the neointimal area of the stented vessel segment compared with rapamycin-eluting stents [26]. Thus, the efficacy and safety of this type of stent need to be further evaluated. The goal of the present study was to evaluate the safety of this novel DDES using a porcine model of coronary stenting. As well, we also evaluated whether the DDES could improve stent endothelialization.

Results of our study show that, irrespective of stent type, all animals had similar expansion ratio and injury scores at each time points, reflecting comparable vessel injury. Also, the lumen loss evaluated by QCA, the neointimal area evaluated by OCT and the histomorphology analysis assessing the neointimal proliferation level of the stented vessel segment were similar between all animals at all time points. On the one hand, our results show that the novel DDES has a similar anti-proliferative capability with the commercially available SES. However, the SES and DDES groups showed a similar anti-proliferative capability with the BMS group in each time point. This result may be due to the small sample size of each group.

Delayed endothelialization is considered to be a major underlying mechanism of ST, and numerous studies have evaluated the strut coverage level as a measure of reendothelialization. However, it is important to note that the struts that were completely covered did not represent the EC covering the entire lumen. These OCT studies could not distinguish if the coverage of the stent struts was by EC or other cell type. The coverage might not have the functionality of EC, especially its anti-thrombotic property [27]. Although some study has suggested that the strut coverage level might be associated with endothelial function [28], the evidence presented is unclear.

Our OCT results show that, at the 14-day time point, the strut coverage percentage in the PrES group was significantly higher than that observed in the SES group. Also, it had the highest percentage of EC coverage under SEM. However, such difference did not translate into mRNA and protein expression. At the 28-day time point, there was no significant difference between the groups with respect to strut coverage percentage, but it seemed to be higher in the BMS group than in the other groups, as observed by SEM. At high magnification, it compared with the PrES group, although the struts were almost completely covered in the BMS group with large part of the covering being only a layer of extracellular matrix, not EC (Fig. 4c). Conversely, although the struts were incompletely covered in the PrES group, the majority of the covering was by EC (Fig. 4d). There was also no obvious connection between the strut coverage observed by OCT and EC coverage assessed by HE staining at high magnification (Fig. 7).

Fig. 7.

Fig. 7

a The strut has been covered as observed by optical coherence tomography and at low magnification by hematoxylin and eosin (HE) staining. Black rectangles in the left image represent the area of the right image that the covering was endothelial cells in high magnification; b The strut has been covered as observed by optical coherence tomography and at low magnification by hematoxylin and eosin staining. Black rectangles in the left image represent the area of the right image that the covering was not endothelial cells in high magnification; c The strut has not been covered as observed by optical coherence tomography and at low magnification with hematoxylin and eosin staining. Black rectangles in the left image represent the area of the right image that the covering was not endothelial cells in high magnification; d The strut has been covered as observed at low magnification with hematoxylin and eosin staining, but black rectangles in the left image represent the area of the right image that the covering was not exactly the same

There was no significant difference in the expression of both EC marker CD31 and EPC markers CD34 and CD133 in the mRNA and protein level between groups at all the four time points. In the histopathological analysis, there was also no significant difference in the endothelialization scores between groups at all the four time points. Overall, our results suggest that the local release of rapamycin at low concentration used in the SES and DDES groups did not significantly reduce the local expression of EC and EPC. These results are in agreement with those of previous in vitro studies [29, 30].

Although rapamycin can inhibit the differentiation, proliferation, migration, adhesion and endothelial nitric oxide synthase expression of EPC [31], its effect on the homing of EPC at the site of the stented vessel segment might be different from previous studies. Also, the local release of probucol in the PrES and DDES groups showed a tendency to improve the short-term endothelialization when compared with the other groups in qualitative SEM analysis, although it was not detected in mRNA and protein level. Small sample sizes and the sampling method of the stented segments might have played a role. Lastly, at the time points of 90 and 180 days, obvious angiogenesis in the neointima was occasionally observed, resulting in individual higher injury scores as assessed by HE staining (Fig. 8). Although this phenomenon has never been reported in similar animal models, it is relatively common in studies of atherosclerotic plaques and in-stent neoatherosclerotic lesions [32, 33] and is often associated with late adverse events. In the present study, the development of this phenomenon interfered with the detection of the marker of endothelialization, likely due to the overlapping of endothelial cells and EPC at the lumen surface and angiogenesis. Consequently, this phenomenon could allow researchers to use this model for unstable atherosclerotic lesions and in-stent neoatherosclerotic lesions.

Fig. 8.

Fig. 8

Angiogenesis in the neointima. # represents the venule in the neointima and * represents the arteriole in the neointima

Conclusion

Polymer-free dual drug-eluting stents are as safe as the commercially available BMS and SES, but they do not result in further improvement of the endothelialization of the stented porcine coronary arteries.

Acknowledgements

No.

Funding

This work was supported by the Chinese High-tech R&D (863) Program 2011AA03010.

Availability of data and materials

The datasets used and analysed during the current study available from the corresponding author on reasonable request.

Abbreviations

BMS

Bare-metal stent

CSR

Ratio of covered struts

DDES

Dual drug-eluting stent

EC

Endothelial cell

EPC

Endothelial progenitor cell

NA

Neointimal area

OCT

Optical coherence tomography

PCI

Percutaneous coronary intervention

PCR

Polymerase chain reaction

PrES

Probucol-eluting stent

QCA

Quantitative coronary angiography

SEM

Scanning electron microscope

SES

Sirolimus-eluting stents

SR

Stent restenosis

ST

Stent thrombosis

Authors’ contributions

ZBI, Analysis and interpretation of data, drafting of the manuscript, revising the manuscript critically for important intellectual content. ZBO, Analysis and interpretation of data, revising the manuscript critically for important intellectual content. WXG, Analysis and interpretation of data, revising the manuscript critically for important intellectual content. SQP, Acquisition of data, analysis and interpretation of data, revising the manuscript critically for important intellectual content. JJ, Acquisition of data, analysis and interpretation of data, revising the manuscript critically for important intellectual content. HY, Conception and design, final approval of the manuscript submitted. PCS, Analysis and interpretation of data, revising the manuscript critically for important intellectual content. HJY, Conception and design, analysis and interpretation of data, revising the manuscript critically for important intellectual content, final approval of the manuscript submitted. CM, Conception and design, analysis and interpretation of data, revising the manuscript critically for important intellectual content, final approval of the manuscript submitted. All authors have read and approved the manuscript.

Ethics approval

The study was approved by the Institutional Animal Care and Use Committee of Peking University First Hospital with the reference number 201106, and conformed to the National Institutes of Health Guide for the Care and Use of Laboratory Animals.

Consent for publication

Not Applicable

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Contributor Information

Jingyan Han, Phone: 86-10-82802996, Email: hanjingyan@bjmu.edu.cn.

Ming Chen, Phone: 86-10-83575727, Email: cm6141@sina.com.

References

  • 1.Ellis SG, Kereiakes DJ, Metzger DC, Caputo RP, Rizik DG, Teirstein PS, et al. Everolimus-eluting bioresorbable scaffolds for coronary artery disease. N Engl J Med. 2015;373:1905–1915. doi: 10.1056/NEJMoa1509038. [DOI] [PubMed] [Google Scholar]
  • 2.Serruys PW, Chevalier B, Dudek D, Cequier A, Carrié D, Iniguez A, et al. A bioresorbable everolimus-eluting scaffold versus a metallic everolimus-eluting stent for ischaemic heart disease caused by de-novo native coronary artery lesions (ABSORB II): an interim 1-year analysis of clinical and procedural secondary outcomes from a randomized controlled trial. Lancet. 2015;385:43–54. doi: 10.1016/S0140-6736(14)61455-0. [DOI] [PubMed] [Google Scholar]
  • 3.Viñals F, Chambard JC, Pouysségur J. p70 S6 kinase-mediated protein synthesis is a critical step for vascular endothelial cell proliferation. J Biol Chem. 1999;274:26776–26782. doi: 10.1074/jbc.274.38.26776. [DOI] [PubMed] [Google Scholar]
  • 4.Chen TG, Chen JZ, Wang XX. Effects of rapamycin on number activity and eNOS of endothelial progenitor cells from peripheral blood. Cell Prolif. 2006;39:117–125. doi: 10.1111/j.1365-2184.2006.00375.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Finn AV, Joner M, Nakazawa G, Kolodgie F, Newell J, John MC, et al. Pathological correlates of late drug-eluting stent thrombosis: strut coverage as a marker of endothelialization. Circulation. 2007;115:2435–2441. doi: 10.1161/CIRCULATIONAHA.107.693739. [DOI] [PubMed] [Google Scholar]
  • 6.Liu J, Li M, Lu H, Qiao W, Xi D, Luo T, et al. Effects of probucol on restenosis after percutaneous coronary intervention: a systematic review and meta-analysis. PLoS One. 2015;10 doi: 10.1371/journal.pone.0124021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Lau AK, Leichtweis SB, Hume P, Mashima R, Hou JY, Chaufour X, et al. Probucol promotes functional reendothelialization in balloon-injured rabbit aortas. Circulation. 2003;107:2031–2036. doi: 10.1161/01.CIR.0000062682.40051.43. [DOI] [PubMed] [Google Scholar]
  • 8.Keaney JF, Jr, Xu A, Cunningham D, Jackson T, Frei B, Vita JA. Dietary probucol preserves endothelial function in cholesterol-fed rabbits by limiting vascular oxidative stress and superoxide generation. J Clin Invest. 1995;95:2520–2529. doi: 10.1172/JCI117953. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Tagawa T, Urabe Y, Kimura Y, Suzuki S, Ono H, Takeda K. Long-term treatment with probucol improves endothelial function in patients with coronary artery disease. Hypertens Res. 2004;27:311–318. doi: 10.1291/hypres.27.311. [DOI] [PubMed] [Google Scholar]
  • 10.Yu M, Xu B, Kandzari DE, Wu Y, Yan H, Chen J, et al. First report of a novel polymer-free dual-drug eluting stent in de novo coronary artery disease: results of the first in human BICARE trial. Catheter Cardiovasc Interv. 2014;83:405–411. doi: 10.1002/ccd.25129. [DOI] [PubMed] [Google Scholar]
  • 11.Zhang B, Chen M, Zheng B, Wang XG, Wang XT, Fan YY, et al. A novel high nitrogen nickel-free coronary stents system: evaluation in a porcine model. Biomed Environ Sci. 2014;27:289–294. doi: 10.3967/bes2014.051. [DOI] [PubMed] [Google Scholar]
  • 12.Mills JS, N'diaye CS, Yow E, Urtz M, Povsic TJ, Greenfield JC, et al. Preliminary observations using optical coherence tomography to assess neointimal coverage of a metal stent in a porcine model. Cardiovasc Revasc Med. 2009;10:229–235. doi: 10.1016/j.carrev.2008.11.003. [DOI] [PubMed] [Google Scholar]
  • 13.John MC, Wessely R, Kastrati A, Schömig A, Joner M, Uchihashi M, et al. Differential healing responses in polymer- and nonpolymer-based sirolimus-eluting stents. JACC Cardiovasc Interv. 2008;1:535–544. doi: 10.1016/j.jcin.2008.08.004. [DOI] [PubMed] [Google Scholar]
  • 14.Tada N, Virmani R, Grant G, Bartlett L, Black A, Clavijo C, et al. Polymer-free biolimus a9-coated stent demonstrates more sustained intimal inhibition, improved healing, and reduced inflammation compared with a polymer-coated sirolimus-eluting cypher stent in a porcine model. Circ Cardiovasc Interv. 2010;3:174–183. doi: 10.1161/CIRCINTERVENTIONS.109.877522. [DOI] [PubMed] [Google Scholar]
  • 15.Chen M, Zheng B, Wu Z, Peng HY, Wang XG, Zhang B, et al. Efficacy and safetyof a novel nano-porous polymer-free sirolimus-eluting stent in pigs. Chin Med J. 2013;126:4731–4735. [PubMed] [Google Scholar]
  • 16.Costa JR, Jr, Abizaid A, Costa R, Feres F, Tanajura LF, Abizaid A, et al. Preliminary results of the hydroxyapatite nonpolymer-based sirolimus-eluting stent for the treatment ofsingle de novo coronary lesions a first-in-human analysis of a third-generation drug-eluting stent system. JACC Cardiovasc Interv. 2008;1:545–551. doi: 10.1016/j.jcin.2008.07.003. [DOI] [PubMed] [Google Scholar]
  • 17.Navarese EP, Kowalewski M, Cortese B, Kandzari D, Dias S, Wojakowski W, et al. Short and long-term safety and efficacy of polymer-free vs. durable polymer drug-eluting stents. A comprehensive meta-analysis of randomized trials including 6178 patients. Atherosclerosis. 2014;233:224–231. doi: 10.1016/j.atherosclerosis.2013.12.024. [DOI] [PubMed] [Google Scholar]
  • 18.Byrne RA, Kufner S, Tiroch K, Massberg S, Laugwitz KL, Birkmeier A, et al. Randomised trial of three rapamycin-elutingstents with different coating strategies for the reduction of coronary restenosis: 2-year follow-up results. Heart. 2009;95:1489–1494. doi: 10.1136/hrt.2009.172379. [DOI] [PubMed] [Google Scholar]
  • 19.King L, Byrne RA, Mehilli J, Schömig A, Kastrati A, Pache J. Five-yearclinical outcomes of a polymer-free sirolimus-eluting stent versus a permanent polymer paclitaxel-eluting stent: final results of the intracoronary stented and angiographic restenosis - test equivalence between two drug-eluting stents (ISAR-TEST) trial. Catheter Cardiovasc Interv. 2013;81:E23–E28. doi: 10.1002/ccd.24375. [DOI] [PubMed] [Google Scholar]
  • 20.Stiermaier T, Heinz A, Schloma D, Kleinertz K, Dänschel W, Erbs S, et al. Five-year clinical follow-up of a randomized comparison of a polymer-free sirolimus-eluting stent versus a polymer-based paclitaxel-eluting stent in patients with diabetes mellitus (LIPSIA Yukon trial) Catheter Cardiovasc Interv. 2014;83:418–424. doi: 10.1002/ccd.25131. [DOI] [PubMed] [Google Scholar]
  • 21.Kufner S, Sorges J, Mehilli J, Cassese S, Repp J, Wiebe J, et al. Randomized Trial of Polymer-Free Sirolimus- and Probucol-Eluting Stents Versus Durable Polymer Zotarolimus-Eluting Stents: 5-YearResults of the ISAR-TEST-5 Trial. JACC Cardiovasc Interv. 2016;9:784–792. doi: 10.1016/j.jcin.2016.01.009. [DOI] [PubMed] [Google Scholar]
  • 22.Byrne RA, Mehilli J, Iijima R, Schulz S, Pache J, Seyfarth M, et al. A polymer-free dual drug-eluting stent in patients with coronary artery disease: a randomized trial vs. polymer-based drug-eluting stents. Eur Heart J. 2009;30:923–931. doi: 10.1093/eurheartj/ehp044. [DOI] [PubMed] [Google Scholar]
  • 23.Byrne RA, Kastrati A, Tiroch K, Schulz S, Pache J, Pinieck S, et al. 2-year clinical and angiographic outcomes from a randomized trial of polymer-free dual drug-eluting stents versus polymer-based Cypher and Endeavor [corrected] drug-eluting stents. J Am Coll Cardiol. 2010;55:2536–2543. doi: 10.1016/j.jacc.2010.03.020. [DOI] [PubMed] [Google Scholar]
  • 24.Massberg S, Byrne RA, Kastrati A, Schulz S, Pache J, Hausleiter J, et al. Polymer-free sirolimus- and probucol-eluting versus new generation zotarolimus-eluting stents in coronary artery disease: the Intracoronary Stented and Angiographic Results: Test Efficacy of Sirolimus- and Probucol-Eluting Versus Zotarolimus-Eluting Stents (ISAR-TEST 5) Circulation. 2011;124:624–632. doi: 10.1161/CIRCULATIONAHA.111.026732. [DOI] [PubMed] [Google Scholar]
  • 25.Chen SL, Ye F, Zhang JJ, Zou JJ, Qian XS, Li F, et al. Real polymer-free sirolimus- and probucol-eluting versus biodegradable polymer sirolimus-eluting stents for obstructive coronary artery disease: DKPLUS-Wave 1, a multicenter, randomized, prospective trial. Cardiovasc Ther. 2013;31:193–200. doi: 10.1111/j.1755-5922.2012.00319.x. [DOI] [PubMed] [Google Scholar]
  • 26.Watt J, Kennedy S, McCormick C, Agbani EO, McPhaden A, Mullen A, et al. Succinobucol-eluting stents increase neointimal thickening and peri-strut inflammation in a porcine coronary model. Catheter Cardiovasc Interv. 2013;81:698–708. doi: 10.1002/ccd.24473. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Otsuka F, Finn AV, Yazdani SK, Nakano M, Kolodgie FD, Virmani R. The importance of the endothelium in atherothrombosis and coronary stented. Nat Rev Cardiol. 2012;9:439–453. doi: 10.1038/nrcardio.2012.64. [DOI] [PubMed] [Google Scholar]
  • 28.Murase S, Suzuki Y, Yamaguchi T, Matsuda O, Murata A, Ito T. The relationship between re-endothelialization and endothelial function after DES implantation: comparison between paclitaxcel eluting stent and zotarolims eluting stent. Catheter Cardiovasc Interv. 2014;83:412–417. doi: 10.1002/ccd.25140. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Fukuda D, Sata M, Tanaka K, Nagai R. Potent inhibitory effect of sirolimus on circulating vascular progenitor cells. Circulation. 2005;111:926–931. doi: 10.1161/01.CIR.0000155612.47040.17. [DOI] [PubMed] [Google Scholar]
  • 30.Zhang P, Huang L, Zhu GX, Cui B, Song MB, Zhou YP, et al. Sirolimus inhibits the differentiation, proliferation and migration of endothelial progenitor cells in vitro. Zhonghua Xin Xue Guan Bing Za Zhi. 2006;34:1021–1025. [PubMed] [Google Scholar]
  • 31.Chen TG, Chen JZ, Wang XX. Effects of rapamycin on number activity and eNOS ofendothelial progenitor cells from peripheral blood. Cell Prolif. 2006;39:117–125. doi: 10.1111/j.1365-2184.2006.00375.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Ho-Tin-Noé B, Michel JB. Initiation of angiogenesis in atherosclerosis: smoothmuscle cells as mediators of the angiogenic response to atheroma formation. Trends Cardiovasc Med. 2011;21:183–187. doi: 10.1016/j.tcm.2012.05.007. [DOI] [PubMed] [Google Scholar]
  • 33.Jaipersad AS, Lip GY, Silverman S, Shantsila E. The role of monocytes inangiogenesis and atherosclerosis. J Am Coll Cardiol. 2014;63:1–11. doi: 10.1016/j.jacc.2013.09.019. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The datasets used and analysed during the current study available from the corresponding author on reasonable request.