Optical methods for quantitative and label-free sensing in living human tissues: principles, techniques, and applications

Robert H Wilson 1,*, Karthik Vishwanath 2, Mary-Ann Mycek 3
PMCID: PMC5560608  NIHMSID: NIHMS838222  PMID: 28824194

Abstract

We present an overview of quantitative and label-free optical methods used to characterize living biological tissues, with an emphasis on emerging applications in clinical tissue diagnostics. Specifically, this review focuses on diffuse optical spectroscopy, imaging, and tomography, optical coherence-based techniques, and non-linear optical methods for molecular imaging. The potential for non- or minimally-invasive assessment, quantitative diagnostics, and continuous monitoring enabled by these tissue-optics technologies provides significant promise for continued clinical translation.

Keywords: tissue optics, diffuse optical spectroscopy, laser speckle imaging, diffuse correlation spectroscopy, multi-photon microscopy, second harmonic generation, coherent anti-stokes Raman

1. Introduction

Measuring the interaction of light with biological tissues can tell us about tissue morphology and biological function on the microscopic, mesoscopic, and macroscopic levels. Light tissue interactions have been known for millennia, but over the past century (and particularly since the invention of the laser), researchers have developed a number of clinically compatible techniques in which light with prescribed properties (e.g., wavelength, frequency, coherence, spatial profile) is shone onto tissue, and the portion of this light that returns to the surface is detected as a function of these same variables. Interaction with the tissue perturbs the light in a manner that can be detected and quantified by optical instrumentation coupled with mathematical light transport models. This information can be employed to assess the health of tissue and characterize its response to perturbation.

One advantage of these methods is that they frequently interrogate endogenous optical contrast in tissues from, e.g., absorbing, scattering, or fluorescing tissue constituents. Therefore, many of these methods do not require an exogenous contrast agent (such as a fluorescent dye) to probe tissue structure and function. Another objective advantage of these techniques is that they are quantitative (to a first approximation), using light-tissue interaction models to extract perturbation-induced changes in concentrations of absorbers (e.g., hemoglobin, water, lipid), scatterers (e.g., collagen, cells, organelles), and fluorophores (e.g., collagen, NADH, FAD) within the tissue. For instance, these techniques can include hybrid photon-wave descriptions of diffuse light (Section 2), coherence- and interference-based effects (Sections 3 and 4), or modelled in terms of photon absorption and scattering (Section 4). These techniques span multiple spatial and temporal scales and have reached various points along the path toward clinical and/or commercial translation.

In this review, we will discuss three widely-used methods in tissue-optics (Table 1): Diffuse optical techniques (Section 2), Coherence-based techniques (Section 3), and Non-linear optical techniques (Section 4).

Table 1.

Overview of tissue-optics methods discussed in this review, including diffuse optical spectroscopy and tomography (DOS/DOT), laser speckle contrast imaging and diffuse correlation spectroscopy (LSCI/DCS), and non-linear optical (NLO) molecular sensing methods.

Technique Physical Process Light Source and Spectral Range Origin of Contrast Within Tissue Quantitative Endpoints
DOS/DOT Diffuse photon waves Broadband lamp or set of LEDs; typically visible to near-infrared Absorption and scattering Hemoglobin concentration and oxygenation; Scatterer concentration and morphology
LSCI/DCS Coherent waves Laser; typically near-infrared Particle motion Changes in blood flow velocity
Nonlinear Optical (NLO) Molecular Imaging Two-Photon Excitation Fluorescence (TPEF) Pulsed laser; typically far visible to near-infrared Molecular absorption Fluorophore concentration and spatial distribution; Fluorophore binding with microenvironment (using FLIM)
Second Harmonic Generation (SHG) Pulsed laser; typically far visible to near-infrared Non-centrosymmetric structures Concentration, spatial distribution, and orientation

2. Diffuse optical spectroscopy, imaging, and tomography

2.1: Diffuse Optical Spectroscopy (DOS)

Diffuse optical spectroscopy (DOS) techniques have been employed for over two decades to quantitatively characterize biological tissues in pre-clinical and clinical settings [17]. DOS methods typically interrogate endogenous contrast from absorption and scattering of light by tissue. Tissue components such as water, lipid, and oxygenated and deoxygenated hemoglobin (HbO2, Hb) have distinct light absorption profiles in the visible, near-infrared, and short-wave infrared regions of the electromagnetic spectrum [6]. The optical absorption coefficient μa(λ) of the tissue can be represented as a linear combination of the extinction coefficients εi(λ) of these chromophores, weighted by their concentrations ci within the tissue: μa(λ) = Σciεi(λ) [4]. For light that has undergone enough scattering events to be considered diffuse, the optical scattering coefficient μs(λ) of the tissue can be modeled with a power law of the form μs(λ) = A(λ/λo)-b, where A and b are referred to as the scattering amplitude and power, respectively, and λo is an arbitrary reference wavelength. In diffuse regimes, the scattering of the tissue is typically represented with the reduced scattering coefficient μs(λ) = μs(λ)(1-g(λ)), which is a combination of the scattering coefficient μs(λ) and the scattering anisotropy g(λ) (defined as the mean cosine of the scattering angle) [7].

The tissue absorption and scattering properties are obtained by illuminating the tissue with broadband light (e.g., using a tungsten-halogen lamp) or light at several discrete wavelength bands (e.g., using light emitting diodes) and detecting the light that is scattered back to the detector. The light is typically delivered and detected with fiber-optic probes separated by a controlled distance ρ that has a direct effect on the interrogated tissue volume. To correct for the spectral profiles of the light source and detector, the measured backscattered intensity from the tissue is calibrated against that obtained from a material with known optical properties (known as a “tissue phantom”) [8]. This calibrated measurement, known as the diffuse reflectance Rd(λ), can be directly related to μa(λ) and μs(λ) (and, consequently, to ci, A, and b) by using the diffusion approximation (DA) to the radiative transport equation. The DA is valid in regimes where the light has been scattered enough to be modeled as diffuse; this typically occurs when μsa > 10. In non-diffuse regimes, such as locations close to the source and wavelengths where absorption from blood is very high (e.g., ~400–450 nm), the DA is no longer valid and Monte Carlo simulations are often employed to model light transport instead [9]. The flow of the DOS procedure, from data collection to calibration to mathematical modeling for extraction of tissue properties, is shown in Fig. 2.

Figure 2.

Figure 2

Flow of DOS method. First, the measured backscattered intensity is obtained from the tissue sample and from a reference with known optical properties, enabling acquisition of the calibrated diffuse reflectance Rd of the tissue. Then, a mathematical model is employed to obtain the tissue absorption and scattering coefficients from the diffuse reflectance, and mathematical representations of μa(λ) and μs(λ) are employed to extract tissue absorber concentrations ci, scattering amplitude A, and scattering power b.

2.2: Diffuse Optical Tomography (DOT)

There are many different techniques for obtaining two-dimensional and three-dimensional maps of tissue properties using DOS measurements. These methods fall under the umbrella of Diffuse Optical Tomography (DOT). One common DOT approach (Fig. 3), which has been described by numerous groups [1012], is to obtain diffuse reflectance at a number of different source-detector separations ρ and use a numerical inversion algorithm to obtain spatial maps of ci, A, and b. These multi-fiber DOT probes have been widely employed in vivo for breast and brain imaging [10, 13], but they typically suffer from low spatial resolution. Recently, wide-field DOT methods, in which a projected field of light is directed onto the sample and a camera detects the backscattered signal, have been employed for reconstruction of optical properties with high (~1 mm) spatial resolution [14]. These methods typically modulate the light at several different spatial frequencies in order to rigorously decouple the absorption and scattering coefficients [8].

Figure 3.

Figure 3

Schematics of common DOS (top) and DOT (bottom) measurement setups. In DOS, light is delivered to the tissue via a source fiber and backscattered photons are detected via a detector fiber separated some distance ρ from the source. This source-detector separation plays a key role in controlling the region of tissue sampled by the majority of the photons. DOS measurements provide the reflectance at different wavelengths, enabling detection of spectral features of tissue absorption and scattering. In DOT, reflectance measurements at multiple source-detector pairs are employed to enable reconstruction of a two-dimensional map of the reflectance (and hence, absorption and scattering) at each wavelength.

In order to reconstruct three-dimensional maps of tissue optical properties, an additional set of measurements typically must be performed in order to provide depth sectioning of the interrogated tissue volume. This extra variable is often wavelength [15], time [16], or spatial (or temporal) frequency [17] of the incident light. Acquiring surface images of the tissue at different wavelengths can provide depth-sectioning capabilities because the tissue absorption and scattering properties are heavily wavelength-dependent, so the mean penetration depth of the photons can vary significantly with wavelength [15]. Time-gating of the detected photons also serves as a means to separate the photons that travelled different total path lengths (and thus, sampled different depths) within the tissue before returning to the surface [16].

2.3: Photoacoustic tomography (or microscopy) (PAT/PAM)

Photoacoustic techniques combine the penetration depth of ultrasound with the functional specificity of tissue optical absorption contrast to provide quantitative and high-resolution functional and structural maps of tissue deeper than 1mm below the surface. Given that bulk of the reconstructive imaging in PAT and PAM rely on principles governing the propagation and detection of ultrasound waves in tissue and that intrinsic functional contrast in these techniques arises from the optically-induced thermal pressure waves – something that is directly dependent on the optical absorption spectrum of tissue as discussed above – a full review of PAT/PAM techniques will not be provided in this review. However, we would like to emphasize that this area remains one of the most active areas of research in the realm of biomedical optics mainly because of the tremendous potential these methods for in vivo medical imaging, as discussed in several recent reviews [1823].

2.3: Clinical applications of DOS and DOT: breast cancer

One common clinical application of DOS and DOT methods is the detection, characterization, and monitoring of breast cancer and its response to a variety of treatment mechanisms. Fig. 4 shows maps of tissue absorber concentrations (oxygenated hemoglobin, ctO2Hb; deoxygenated hemoglobin, ctHHb; water; lipid) of the right and left breast of a patient with cancer in the right breast [6]. The rightmost column is a map of a quantity known as the tissue optical index (TOI), defined by the equation TOI = (ctH2O/Lipid)(ctHHb) and determined empirically to be a combination of parameters that is particularly effective for highlighting contrast between tumors and surrounding normal tissue. The figure demonstrates the ability of DOS to reconstruct two-dimensional maps of tumor optical properties that correspond well to the spatial location and extent of the tumor (black dashed circle), as determined by palpation. These methods have also been employed to monitor the response of breast tumors to neoadjuvant chemotherapy (Fig. 5) and other types of treatment.

Figure 4.

Figure 4

DOT maps of concentrations of oxygenated hemoglobin (ctO2Hb), deoxygenated hemoglobin (ctHHb), water, and lipid in the breasts of a human subject with a cancerous tumor in the right breast. The rightmost column is a map of the tissue optical index (TOI), defined by TOI = (ctH2O/Lipid)(ctHHb). The changes in the measured optical parameters correspond well to the tumor region estimated by palpation (black dashed circle) ([6], reproduced with permission).

Figure 5.

Figure 5

DOT maps of tissue optical index (TOI), defined by TOI = (ctH2O/Lipid)(ctHHb), of the breast of a patient with a cancerous lesion (black elliptical region) at different time points throughout the process of neoadjuvant chemotherapy ([6], reproduced with permission). The areola (green elliptical region) is used as a fiducial marker. The TOI maps demonstrate the successful treatment of the patient, who exhibited a pathologic complete response to the chemotherapy.

2.4: Clinical applications of DOS and DOT: functional brain imaging

DOS techniques have also been employed for functional brain imaging. Since DOS methods typically use near-infrared light, which can penetrate deep beneath the tissue surface, in vivo human brain imaging can be achieved by using the appropriate combinations of wavelengths and probe source-detector separations [13]. One common DOS-based brain imaging experiment is the monitoring of the cerebral hemodynamic response to an external stimulus. Fig. 6 shows the experimental setup and results for such an experiment [24]. A probe containing a number of different sources and detectors is placed on the back of the head, enabling diffuse optical interrogation of the brain. As the brain responds to the stimulus, maps of changes in oxygenated hemoglobin concentration (Fig. 6(c)) are reconstructed from the optical data. Time courses of changes in the concentrations of oxygenated hemoglobin, deoxygenated hemoglobin, and total hemoglobin (Fig. 6(d)) can be calculated to characterize the hemodynamic response of the brain to the stimulus. This technique is of great potential use for quantitatively monitoring the brain’s response to conditions such as stroke, traumatic brain injury, and Alzheimer’s disease.

Figure 6.

Figure 6

In vivo diffuse optical measurement of functional activation of the human brain using a DOS setup (left) with multiple sources (red) and detectors (blue) placed on the back of the head. The DOS measurements provide spatial and temporal maps (a–c) of changes in hemoglobin concentration in the tissue in response to stimulus. Time courses (d) of changes in the concentrations of oxygenated hemoglobin (red), deoxygenated hemoglobin (blue), and total hemoglobin (green) can be calculated to characterize the brain’s response to the stimulus ([24], reproduced with permission).

3. Coherence-based sensing approaches

As discussed in the previous section, diffuse optical imaging techniques rely on quantifying spatio-temporal changes of the detected light intensity by connecting them to differences in optical absorption or scattering properties of media, at one or more wavelengths. The phase of the electromagnetic optical fields is not considered to play any discernible role in these diffuse optical measurements. However, in interferometric detection, it is the phase differences between optical fields that are detected as intensity fluctuations. Interferometric sensing using coherent (laser) light sources is an extremely active area of research in field of translational biomedical optical imaging and we describe two important techniques – laser speckle contrast imaging (LSCI) and diffuse correlation spectroscopy (DCS) – that have been used in a number of clinical studies and are beginning to emerge as commercial options.

Before beginning the discussion about LSCI and DCS, we wish to state we will specifically not be discussing another important and widely-used coherence-based optical imaging technique – optical coherence tomography or microscopy [2528] (OCT/OCM) – here. This decision to focus primarily on non-OCT based coherence based optical tissue sensing stems from the fact that OCT is a very well-established clinical technique that has previously been extensively reviewed [2938]. We also note that recent advances in Fourier and spectral domain Doppler-OCT have been developed to enable functional sensing of blood flow velocities within intact tissues – also termed optical microangiography – and are well-described elsewhere [3942].

3.1: Laser speckle contrast imaging (LSCI)

The formation of laser speckle is a well-known phenomenon, especially given the ubiquity of coherent laser sources today. Laser speckle is mathematically described as a complex pattern resulting from the interference of a large number of coherent photon fields scattered off a rough object and therefore having random phases at the image plane [43]. Although the speckle pattern itself appears chaotic and possess little semblance to any discernible features to the illuminated and imaged object, laser speckle has several quantifiable statistical features that can provide a wide range of information about spatio-temporal properties within the imaged scattering medium [4447]. Laser speckle contrast imaging (LSCI) for biomedical imaging was demonstrated nearly four decades ago as an experimentally simple technique capable of non-invasively imaging microvascular flow in retinal tissues [48].

Experimental implementation of LSCI involves shining a coherent laser light source to uniformly illuminate a region of interest that is focused using a camera and involves collection of the raw speckle image with no ambient light [49, 50]. Image analysis is performed to quantify blurring of the speckles recorded in the image by calculating a spatial contrast factor given by K=σμ where σ is the standard deviation and μ the mean of the intensity values recorded some set of pixels. These sets of pixels are usually selected as region of 7×7 neighbors in a square grid across all pixels in the acquired image as spatial averaging within a single acquisition, or across a series of images acquired as time-series (or video data) [49]. Mathematically, the contrast factor can be shown to be related to the speed of the moving particles causing the blurring of the speckle pattern [46, 51], which in biological tissues translates to a blood flow index. Although LSCI is experimentally simple to implement and has been shown to work even with low cost laser diodes for sources and web-cameras for detectors [52], there are several active and open areas of research including improving depth-sensitivity, developing improved theoretical models involved in the physics behind speckle formation, and in developing methods so that LSCI can generate quantitative blood flow maps [5357].

3.2: Diffuse correlation spectroscopy (DCS)

In DCS, optical fibre probes are used to launch a coherent source of light into the medium of interest and then measure intensity fluctuations across small areas (typically within diameters of 1–5 μm) several mm-cm from the incident source and at relatively high frequency (> 1–5 MHz). The observed intensity fluctuations represent speckle intensity fluctuations caused due to motion of scattering particles moving through diffuse photon trajectories leading from the source-detector. The temporal autocorrelation of these speckle intensity fluctuations is the normalized intensity-autocorrelation g2(τ) and is related to the normalized field-autocorrelation function g1(τ) through the Siegert relationship: g2(τ) = 1+β(g1(τ))2 [58]. The normalized field autocorrelation g1(τ)=G1(τ)/G1(0), where G1(τ) is related to the time-varying coherent electric field in the scattering and absorbing medium. G1(τ) is modeled as propagating through the turbid media via a DOS-like diffusion equation but having an attenuating factor that depends both on the tissue absorption as well as the mean-squared displacement of a scattering particle – which is a function of the autocorrelation delay τ [59, 60]. Thus, experimentally measured intensity-autocorrelation g2(τ) from DCS measurements, when supplemented with estimates of optical absorption and scattering properties at the laser wavelength (typically obtained from simultaneous or sequential DOS/DOT measurements), allow theoretical fitting of g2(τ) using expressions for G1(τ) [61, 62].

Experimentally, Figure 7 illustrates the schematic relation between the incident laser speckle, the intensity fluctuations measured and its autocorrelation to the experimental components of typical DCS instrument. The choice of the laser coherence length (i.e., the maximum distance within the tissue for which the laser is expected to sufficiently maintain its coherent properties) must be such that it is much longer than the average path length of photons diffusing from the source and the detector placed on the tissue. Most researchers use laser sources with coherence lengths longer than 10m. Detectors are small-area avalanche photodiodes running in photon counting mode whose digitized output is passed on to the autocorrelation device – usually a dedicated hardware board [61, 62].

Figure 7.

Figure 7

Illustration of DCS measurements in terms of speckle intensity fluctuations and experimental components required in measuring the autocorrelation trace. It is critical in DCS to capture only a few speckles and thus the detection probe is usually 4–5 μm. The laser also needs to have a coherence length that is much longer than the mean photon path from the source to the detector.

One burden with using DCS techniques has not only the need to have a long coherence length laser but also to require autocorrelator boards for fast computation of the numerically intensive calculation of the autocorrelation function, which usually adds cost to the instrument. However, recent developments in theoretical analysis of the autocorrelation decays indicate the possibility that blood flow information may be acquired measuring the autocorrelation for few initial delay times [63].

3.3: Clinical applications of LSCI

Applications of LSCI have focused on using this technique to detect blood flow in tissue that are transparent (ocular blood flow [64, 65]) or for tissue where the vasculature is more readily accessible from the surface (such as skin wound healing and open surgery [6671]). Figure 8 below shows one particularly novel application of LSCI to differentially image arteries and veins in the human eye by superposing an adaptive optics image of the retinal bed with a LSCI image [72], it was shown that LSCI could be used to compute a retinal blood flow index to quantify. The colour and contrast density of the merged image (Figure 8(C)) indicates mean blood flow rates for arteries in red and veins in grey.

Figure 8.

Figure 8

Representative merged figure from [72] (reproduced with permission). Laser speckle image (A) and adaptive optics image (B) were merged into (C).

3.3: Clinical applications of DCS

DCS has been more widely employed for a variety of clinical applications (relative to LSCI) for monitoring blood flow in the brain, skeletal muscle, bone, and during tumour growth [7383]. Tissue blood flow changes measured using DCS have also been quantitatively matched to blood flow measured using ultrasound, MRI and laser Doppler flow.

In the past, DCS instrumentation contained bulky electronic components and required significant logistical efforts for use in clinical settings. Recent efforts have focused on improving the form-factor of these systems. One such application of DCS (Figure 9) provides a proof-of-principle example where employed a hand held DCS instrument (source-detector separation of 2.5 cm) to optically scan women with clinically confirmed malignant breast lesions as they were seated in the clinic in a supine position [84]. Measurements were acquired both on the ipsilateral breast – containing the malignant lesion – and on the contralateral normal tissue. DCS measurements were obtained in a line-scan by moving the device across 10 cm. Although these studies used a priori knowledge about the location of the tumors to make the DCS measurements, the extracted blood flow values (relative to the repeated errors) are more than double the value measure normal sites (on either breast). a diffuse correlation tomographic approach could be designed with multi source detector configurations to increase spatial sensitivities to yield a 3D map of relative blood indices.

Figure 9.

Figure 9

Line-scans of BFI in (a) for ipsilateral cancerous breast and (b) for the normal contralateral breast, from [84] (reproduced with permission). Locations of the (T) and normal (N) regions were chosen using radiological data to guide probe placement. Mirror locations of the Tumor (MT) and normal (MN) regions (in the ipsilateral breast) are identified in the contralateral breast.

4. Non-linear optical methods for label-free molecular sensing and imaging in living biological tissues

The theoretical principles and experimental techniques of non-linear optics (NLO) are well known in physics and have been applied to numerous and diverse fields of study in the basic sciences and engineering [85]. In particular, a variety of NLO techniques have been applied to address fundamental and applied studies in biology and medicine [8689], as illustrated in Figure 10, including exciting applications of two-photon excitation fluorescence (TPEF) and second-harmonic generation (SHG) to label-free molecular sensing and imaging in clinical medicine [90, 91] and emerging in vivo applications of coherent anti-Stokes Raman scattering (CARS) [92].

Figure 10.

Figure 10

Schematic energy level diagrams for linear (one-photon excitation) fluorescence and non-linear optical processes including two-photon excitation fluorescence (TPEF), second harmonic generation (SHG), and coherent anti-Stokes Raman scattering (CARS).

Employing NLO techniques for medical applications presents some distinct advantages as well as challenges relative to approaches based on linear optics. For sensing and imaging applications in biological tissues, general advantages offered by NLO methods include the ability to interrogate specimens at greater tissue depths by using longer (near infrared, NIR) wavelengths that are subject to less optical loss from absorption and scattering, to optically section tissues and minimize out-of-focus photodamage while decreasing background, and to employ label-free sensing with molecular specificity. In general, the greatest challenges to implementing NLO-based technologies clinically include high equipment costs, complexity of operation (both for data acquisition and analysis), and potential regulatory hurdles for human studies.

Here we illustrate the utility of NLO methods for label-free molecular sensing and imaging in living biological tissues to address critical regulatory issues in tissue engineering and regenerative medicine (TE/RM). The fields of TE/RM have advanced to the point where tissues damaged by disease or trauma can be replaced with a living engineered tissue construct manufactured from the patient’s own freshly-harvested primary cells [93]. Generally, engineered tissue manufacturing processes take several days, are strictly controlled by regulatory agencies to ensure cell-based device effectiveness and patient safety, and require rigorous quality control testing on the day of product release to determine the most robust tissue construct to implant into the patient. One of the challenges facing the commercialization and clinical translation of TE/RM technologies is the unmet need for reliable, rapid quality control methods to assess construct viability without destroying or compromising the living, cell-based device and that yield quantified parameters employable as release criteria. Methods currently employed for the assessment of TE/RM products have serious limitations. For example, tissue histology is time-consuming and requires an invasive and destructive tissue biopsy, while biochemical assays of tissue culture media are non-invasive, but lack sensitivity by reporting parameters integrated over the entire construct without providing spatially localized information. An ideal quality control method would rapidly and non-invasively provide a quantitative, spatially localized assessment of tissue construct viability prior to product release and would be translatable to a clinical environment for in vivo assessment of the tissue construct post-implantation in patients.

Optical sensing methods based on endogenous tissue fluorescence have been suggested to characterize the structure and function of living, cell-based, tissue-engineered devices fabricated for implantation in clinical applications [94]. Rather than employing destructive cell viability assays or exogenous fluorescent dyes that are invasive to the cell manufacturing process, the label-free methods proposed target biologically- and metabolically-relevant endogenous cellular metabolic co-factors and extracellular structural proteins via NLO microscopic imaging techniques, including TPEF and SHG [95, 96]. Endogenous fluorescent biomolecules that report on cellular metabolism and oxygen consumption include nicotinamide adenine dinucleotide (phosphate) (NAD(P)H) and flavin adenine dinucleotide (FAD), whose concentrations are related to cellular glucose uptake and energy consumption and production in metabolic pathways. Fluorescence lifetime imaging (FLIM) provides further information about the protein binding states of cellular coenzymes [96]. Optical signals from endogenous structural proteins including keratin, elastin, and collagen report on the morphology and composition of the extracellular matrix, with SHG providing an excellent means for revealing collagen in biological tissues.

An illustration of using an NLO-based approach to characterize morphology and biochemistry in living engineered tissues is shown in Figure 11 [91]. While histological assessment would destroy the tissue construct, NLO microscopic imaging could be performed non-invasively via optical sectioning in 3-dimensions (3D) in the living tissue (left) under environmentally controlled and sterile conditions. 3D spatially localized information about cellular metabolic function, cellular spatial organization, and tissue microenvironment could be extracted from the optical datasets via quantitative image analysis algorithms applied to cross-sectional (middle) and en-face (right) images. For example, the tissue’s layered structure was visible in the destructive histology section (left) as well as the optically cross-sectioned image of endogenous cellular TPEF (NAD(P)H, cyan) with overlaid scaffold SHG (collagen, blue) (middle). Cellular viability information, primarily contained within TPEF signals from metabolic co-factors NAD(P)H and FAD, was obtained by optically sectioning thin cellular layers and using SHG signals to correct for extracellular matrix components (right). The acquired optical signals enabled tissue viability assessment via metrics of cellular spatial organization and metabolic function.

Figure 11.

Figure 11

NLO imaging can be employed for non-invasive, label-free assessment of cell viability in living tissue engineered constructs prior to surgical implantation in patients to repair damaged tissues. The living tissue construct (illustrated by the histology cube at left) is imaged microscopically via NLO techniques in three dimensions, generating both cross-sectional (middle) and en-face (right) images. Two-photon excited fluorescence (TPEF) microscopy, fluorescence lifetime imaging microscopy (FLIM), and second harmonic generation (SHG) microscopic images are analyzed via quantitative algorithms to assess local tissue viability, including cellular metabolic function, spatial organization, and cellular microenvironment. (Figure reprinted from [91] with permission.)

For these studies, cells and tissues were cultured from distinct primary human patient donors to intentionally introduce the intra-patient variability anticipated in clinical use [91]. Additionally, realistic stressing conditions anticipated during the biofabrication process were introduced to intentionally compromise cell viability. These experimental design features were introduced to challenge the reliability and robustness of the NLO approach, which successfully differentiated control from stressed constructs. This study was the first report demonstrating the ability of label-free NLO microscopy to characterize the viability of living cell-based devices manufactured with primary human cells. Thus, label-free NLO microscopic imaging was found to non-destructively provide reliable and quantitative release criteria for engineered human cell devices.

Thus, the microscopic detection of TPEF, FLIM, and SHG from endogenous constituents in living engineered tissue constructs provides information-rich multi-dimensional optical datasets that can be analyzed quantitatively to extract parameters linked to pre-implantation construct structure and function and, hence, the likelihood of in vivo graft success. In contrast to existing methods, these NLO-based methods for quality control enable rapid, quantitative, non-destructive, label-free, and spatially-resolved assessment of engineered tissue construct morphology and viability. Thus, such NLO-based measures could serve as the basis for reliable manufacturing release criteria for cell-based tissue-engineered constructs prior to human implantation, thereby addressing a critical regulatory need in regenerative medicine.

Summary, perspectives, and conclusions

In this review, we have presented several key capabilities of diffuse, coherent, and non-linear optical techniques for characterization of living human tissues. These methods provide quantitative, non- and minimally-invasive means of assessing tissue structure and function, through physical parameters such as tissue absorber and scatterer concentrations, velocity of blood flow, and fluorescence from intracellular and extracellular tissue components.

The techniques discussed in this paper are able to probe a wide range of spatial and temporal scales. DOS/DOT and LSCI/DCS can provide sub-mm spatial resolution for quantifying optical properties on the so-called “mesoscopic” length scale (between microscopic and macroscopic). Certain versions of these techniques can also be performed very rapidly, at upwards of 50 Hz, to obtain information about pulsatile hemodynamics in living tissue [97]. Some DOS/DOT techniques, such as those involving streak cameras, can even enable near-ps scale temporal resolution to measure photon time-of-flight distributions in tissue [98, 99]. NLO methods provide microscopic spatial resolution for imaging of structures such as cells and collagen fibers in vivo [91]. Certain NLO techniques, such as those involving fluorescence lifetime, can also provide sub-ns temporal resolution to quantify the dynamic behavior of fluorescent molecules in living tissue [91].

In terms of depth sectioning within tissue, DOS/DOT, LSCI/DCS, and NLO techniques are all typically associated with somewhat different spatial scales in the z-dimension. Diffuse methods such as DOS/DOT and DCS typically penetrate on the order of cm into the tissue, although the penetration depth is a strong function of probe source-detector separation and wavelength of incident light. Indeed, DOT methods typically employ a large number of different source-detector separations to enable depth-sectioning of several cm of tissue. By contrast, LSCI and NLO methods primarily interrogate the first few hundred microns to 1 mm beneath the tissue. NLO techniques are capable of sectioning with a high spatial resolution (dz ~ 10 microns) along the z-axis, suitable for separating the fluorescence of the superficial keratinized layer from that of the cellular layers in skin or a tissue-engineered construct (Figure 11). LSCI is not typically employed for depth-sectioning, although LSCI technology has been employed to characterize the penetration depth of photons in turbid media [100].

One major trend in tissue optics over the past decade has been the development of spectroscopy and imaging methods that combine multiple modalities to provide a more complete characterization of living tissue. An example of such a technique is the combination of diffuse optical methods (DOS/DOT) with coherence-based methods (LSCI/DCS) to provide information about tissue blood flow and oxygenation in tandem [101]. These combined flow/oxygenation technologies have the potential to help decouple flow-based hemodynamic parameters (e.g., reperfusion following injury or ischemia) from metabolism-based hemodynamic parameters (e.g., rate of oxygen consumption by tissue) [101]. When these types of multi-modal measurements are coupled with advances in imaging speed to acquire data at a rapid frame rate (e.g., >50 Hz) [97], the pulsatile nature of these hemodynamic parameters can also be characterized, providing information about vascular reactivity and compliance. These rapid, multi-modal imaging systems may enable the empirical development of more comprehensive models of the response of living tissue to different perturbations, providing a valuable inroad to understanding how the body responds to, and recovers from, trauma and disease.

Here, we have presented a review of several optical technologies (DOS/DOT, LSCI/DCS, and NLO methods) that are frequently employed for quantitative characterization of living biological tissue. Since these techniques rely on endogenous contrast arising from the subsurface tissue constituents, they are label-free and cause minimal perturbation to the structures being investigated. The quantitative nature of these technologies lends itself well to the development of standardized, objective, optical-parameter-based metrics for tissue classification. Recent progression toward devices with smaller form factors and greater ease of use has facilitated gradual translation of biomedical photonic technologies into the clinic, paving the way for optical techniques to provide valuable assistance for a variety of tissue diagnostic and monitoring applications.

Acknowledgments

This work was supported in part by U. S. National Institutes of Health (R44 DE021935-02, 1 TL1 TR 1415-1).

References

  • 1.Brown JQ, Vishwanath K, Palmer GM, Ramanujam N. Advances in quantitative UV-visible spectroscopy for clinical and pre-clinical application in cancer. Curr Opin Biotechnol. 2009;20:119–131. doi: 10.1016/j.copbio.2009.02.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Kanick SC, Leest Cvd, Djamin RS, Janssens AM, Hoogsteden HC, Sterenborg HJCM, Amelink A, Aerts JGJV. Characterization of mediastinal lymph node physiology in vivo by optical spectroscopy during endoscopic ultrasound-guided fine needle aspiration. J Thorac Oncol. 2010;5:981–987. doi: 10.1097/jto.0b013e3181ddbc0e. [DOI] [PubMed] [Google Scholar]
  • 3.Lloyd WR, Wilson RH, Lee SY, Chandra M, McKenna B, Simeone D, Scheiman J, Mycek M-A. In vivo optical spectroscopy for improved detection of pancreatic adenocarcinoma: a feasibility study. Biomed Opt Express. 2014;5:9–15. doi: 10.1364/BOE.5.000009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Zonios G, Perelman LT, Backman V, Manoharan R, Fitzmaurice M, Dam JV, Feld MS. Diffuse reflectance spectroscopy of human adenomatous colon polyps in vivo. Appl Opt. 1999;38:6628–6637. doi: 10.1364/ao.38.006628. [DOI] [PubMed] [Google Scholar]
  • 5.Georgakoudi I, Jacobson BC, Dam JV, Backman V, Wallace MB, Müller MG, Zhang Q, Badizadegan K, Sun D, Thomas GA, Perelman LT, Feld MS. Fluorescence, reflectance, and light-scattering spectroscopy for evaluating dysplasia in patients with Barrett’s esophagus. Gastroenterol. 2001;120:1620–1629. doi: 10.1053/gast.2001.24842. [DOI] [PubMed] [Google Scholar]
  • 6.O’Sullivan TD, Cerussi AE, Cuccia DJ, Tromberg BJ. Diffuse optical imaging using spatially and temporally modulated light. J Biomed Opt. 2012;17:071311. doi: 10.1117/1.JBO.17.7.071311. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Jacques SL. Optical properties of biological tissues: a review. Phys Med Biol. 2013;58:R37–R61. doi: 10.1088/0031-9155/58/11/R37. [DOI] [PubMed] [Google Scholar]
  • 8.Cuccia DJ, Bevilacqua F, Durkin AJ, Ayers FR, Tromberg BJ. Quantitation and mapping of tissue optical properties using modulated imaging. J Biomed Opt. 2009;14:024012. doi: 10.1117/1.3088140. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Wilson RH, Mycek M-A. Models of light propagation in human tissue applied to cancer diagnostics. Technol Cancer Res Treat. 2011;10:121–134. doi: 10.7785/tcrt.2012.500187. [DOI] [PubMed] [Google Scholar]
  • 10.Choe R, Corlu A, Lee K, Durduran T, Konecky SD, Grosicka-Koptyra M, Arridge SR, Czerniecki BJ, Fraker DL, DeMichele A, Chance B, Rosen MA, Yodh AG. Diffuse optical tomography of breast cancer during neoadjuvant chemotherapy: a case study with comparison to MRI. Med Phys. 2004;32:1128–1139. doi: 10.1118/1.1869612. [DOI] [PubMed] [Google Scholar]
  • 11.Wang Z-M, Panasyuk GY, Markel VA, Schotland JC. Experimental demonstration of an analytic method for image reconstruction in optical diffusion tomography with large data sets. Opt Lett. 2005;30:3338–3340. doi: 10.1364/ol.30.003338. [DOI] [PubMed] [Google Scholar]
  • 12.Boas DA, Brooks DH, Miller EL, DiMarzio CA, Kilmer M, Gaudette RJ, Zhang Q. Imaging the body with diffuse optical tomography. IEEE Signal Processing Magazine. 2001;18:57–75. [Google Scholar]
  • 13.Eggebrecht AT, Ferradal SL, Robichaux-Viehoever A, Hassanpour MS, Dehghani H, Snyder AZ, Hershey T, Culver JP. Mapping distributed brain function and networks with diffuse optical tomography. Nature Photon. 2014;8:448–454. doi: 10.1038/nphoton.2014.107. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Mazhar A, Dell S, Cuccia DJ, Gioux S, Durkin AJ, Frangioni JV, Tromberg BJ. Wavelength optimization for rapid chromophore mapping using spatial frequency domain imaging. J Biomed Opt. 2010;15:061716. doi: 10.1117/1.3523373. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Konecky SD, Wilson RH, Hagen N, Mazhar A, Tkacyzk TS, Frostig RD, Tromberg BJ. Hyperspectral optical tomography of intrinsic signals in the rat cortex. Neurophoton. 2015;2:045003. doi: 10.1117/1.NPh.2.4.045003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Chen J, Venugopal V, Intes X. Monte Carlo based method for fluorescence tomographic imaging with lifetime multiplexing using time gates. Biomed Opt Express. 2011;2:871–886. doi: 10.1364/BOE.2.000871. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Konecky SD, Mazhar A, Cuccia D, Durkin AJ, Schotland JC, Tromberg BJ. Quantitative optical tomography of sub-surface heterogeneities using spatially modulated structured light. Opt Express. 2009;17:14780–14790. doi: 10.1364/oe.17.014780. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Ntziachristos V. Going deeper than microscopy: the optical imaging frontier in biology. Nat Methods. 2010;7:603–614. doi: 10.1038/nmeth.1483. [DOI] [PubMed] [Google Scholar]
  • 19.Yao J, Wang LV. Photoacoustic tomography: fundamentals, advances and prospects. Contrast Media Mol Imaging. 2011;6:332–345. doi: 10.1002/cmmi.443. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Wang LV, Hu S. Photoacoustic tomography: in vivo imaging from organelles to organs. Science. 2012;335:1458–1462. doi: 10.1126/science.1216210. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Allen J, Howell K. Microvascular imaging: techniques and opportunities for clinical physiological measurements. Physiol Meas. 2014;35:R91–R141. doi: 10.1088/0967-3334/35/7/R91. [DOI] [PubMed] [Google Scholar]
  • 22.Wang LV, Gao L. Photoacoustic microscopy and computed tomography: from bench to bedside. Annu Rev Biomed Eng. 2014;16:155–185. doi: 10.1146/annurev-bioeng-071813-104553. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Taruttis A, van Dam GM, Ntziachristos V. Mesoscopic and macroscopic optoacoustic imaging of cancer. Cancer Res. 2015;75:1548–1559. doi: 10.1158/0008-5472.CAN-14-2522. [DOI] [PubMed] [Google Scholar]
  • 24.Zeff BW, White BR, Dehghani H, Schlaggar BL, Culver JP. Retinotopic mapping of adult human visual cortex with high-density diffuse optical tomography. Proc Natl Acad Sci USA. 2007;104:12169–12174. doi: 10.1073/pnas.0611266104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Schmitt JM. Optical coherence tomography (OCT): A review. IEEE J Sel Top Quant Electron. 1999;5:1205–1215. [Google Scholar]
  • 26.Drexler W, Liu M, Kumar A, Kamali T, Unterhuber A, Leitgeb RA. Optical coherence tomography today: speed, contrast, and multimodality. J Biomed Opt. 2014;19:071412. doi: 10.1117/1.JBO.19.7.071412. [DOI] [PubMed] [Google Scholar]
  • 27.Schiffman JS, Patel NB, Cruz RA, Tang RA. Optical Coherence Tomography for the Radiologist. Neuroimaging Clin N Am. 2015;25:367–382. doi: 10.1016/j.nic.2015.05.002. [DOI] [PubMed] [Google Scholar]
  • 28.Kim J, Brown W, Maher JR, Levinson H, Wax A. Functional optical coherence tomography: principles and progress. Phys Med Biol. 2015;60:R211–R237. doi: 10.1088/0031-9155/60/10/R211. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Welzel J. Optical coherence tomography in dermatology: a review. Skin Research and Technology. 2001;7:1–9. doi: 10.1034/j.1600-0846.2001.007001001.x. [DOI] [PubMed] [Google Scholar]
  • 30.Chen J, Lee L. Clinical applications and new developments of optical coherence tomography: an evidence-based review. Clin Exp Optom. 2007;90:317–335. doi: 10.1111/j.1444-0938.2007.00151.x. [DOI] [PubMed] [Google Scholar]
  • 31.Stifter D. Beyond biomedicine: a review of alternative applications and developments for optical coherence tomography. Applied Physics B-Lasers and Optics. 2007;88:337–357. [Google Scholar]
  • 32.Pizurica A, Jovanov L, Huysmans B, Zlokolica V, De Keyser P, Dhaenens F, Philips W. Multiresolution Denoising for Optical Coherence Tomography: A Review and Evaluation. Current Medical Imaging Reviews. 2008;4:270–284. [Google Scholar]
  • 33.Ramos JLB, Li Y, Huang D. Clinical and research applications of anterior segment optical coherence tomography - a review. Clinical and Experimental Ophthalmology. 2009;37:81–89. doi: 10.1111/j.1442-9071.2008.01823.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Petzold A, de Boer JF, Schippling S, Vermersch P, Kardon R, Green A, Calabresi PA, Polman C. Optical coherence tomography in multiple sclerosis: a systematic review and meta-analysis. Lancet Neurology. 2010;9:921–932. doi: 10.1016/S1474-4422(10)70168-X. [DOI] [PubMed] [Google Scholar]
  • 35.Wong JJ, Chen TC, Shen LQ, Pasquale LR. Macular Imaging for Glaucoma Using Spectral-domain Optical Coherence Tomography: A Review. Seminars in Ophthalmology. 2012;27:160–166. doi: 10.3109/08820538.2012.712734. [DOI] [PubMed] [Google Scholar]
  • 36.Zivelonghi C, Ghione M, Kilickesmez K, Loureiro RE, Foin N, Lindsay A, de Silva R, Ribichini F, Vassanelli C, Di Mario C. Intracoronary optical coherence tomography: a review of clinical applications. Journal of Cardiovascular Medicine. 2014;15:543–553. doi: 10.2459/JCM.0000000000000032. [DOI] [PubMed] [Google Scholar]
  • 37.Kohli D, Shah TU, Zfass AM. Optical Coherence Tomography for the Diagnosis and Staging of Barrett’s Esophagus, Buried Glands, and Early Esophageal Cancer: A Systematic Review and Meta-analysis. Am J Gastroenterol. 2015;110:S702. [Google Scholar]
  • 38.Jorge E, Baptista R, Calisto J, Faria H, Monteiro P, Pan M, Pego M. Optical coherence tomography of the pulmonary arteries: A systematic review. J Cardiol. 2016;67:6–14. doi: 10.1016/j.jjcc.2015.09.024. [DOI] [PubMed] [Google Scholar]
  • 39.Agrawal R, Xin W, Keane PA, Chhablani J, Agarwal A. Optical coherence tomography angiography: a non-invasive tool to image end-arterial system. Expert Rev Med Devices. 2016;13:519–521. doi: 10.1080/17434440.2016.1186540. [DOI] [PubMed] [Google Scholar]
  • 40.Mahmud MS, Cadotte DW, Vuong B, Sun C, Luk TW, Mariampillai A, Yang VX. Review of speckle and phase variance optical coherence tomography to visualize microvascular networks. J Biomed Opt. 2013;18:50901. doi: 10.1117/1.JBO.18.5.050901. [DOI] [PubMed] [Google Scholar]
  • 41.Reif R, Wang RK. Label-free imaging of blood vessel morphology with capillary resolution using optical microangiography. Quant Imaging Med Surg. 2012;2:207–212. doi: 10.3978/j.issn.2223-4292.2012.08.01. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Wang RK. Optical Microangiography: A Label Free 3D Imaging Technology to Visualize and Quantify Blood Circulations within Tissue Beds in vivo. IEEE J Sel Top Quantum Electron. 2010;16:545–554. doi: 10.1109/JSTQE.2009.2033609. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Goodman JW. Some Fundamental Properties of Speckle. J Opt Soc Am. 1976;66:1145–1150. [Google Scholar]
  • 44.Harris M. Light-Field Fluctuations in-Space and Time. Contemporary Physics. 1995;36:215–233. [Google Scholar]
  • 45.Duncan DD, Kirkpatrick SJ, Wang RKK. Statistics of local speckle contrast. J Opt Soc Am A. 2008;25:9–15. doi: 10.1364/josaa.25.000009. [DOI] [PubMed] [Google Scholar]
  • 46.Briers D, Duncan DD, Hirst E, Kirkpatrick SJ, Larsson M, Steenbergen W, Stromberg T, Thompson OB. Laser speckle contrast imaging: theoretical and practical limitations. J Biomed Opt. 2013;18:066018. doi: 10.1117/1.JBO.18.6.066018. [DOI] [PubMed] [Google Scholar]
  • 47.Duncan DD, Kirkpatrick SJ, Gladish JC, Hurst SA. Laser Speckle Contrast Imaging for the Quantitative Assessment of Flow. Dynamics and Fluctuations in Biomedical Photonics Vi. 2009;7176 [Google Scholar]
  • 48.Fercher AF, Briers JD. Flow Visualization by Means of Single-Exposure Speckle Photography. Optics Communications. 1981;37:326–330. [Google Scholar]
  • 49.Boas DA, Dunn AK. Laser speckle contrast imaging in biomedical optics. J Biomed Opt. 2010;15:011109. doi: 10.1117/1.3285504. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Ponticorvo A, Dunn AK. How to Build a Laser Speckle Contrast Imaging (LSCI) System to Monitor Blood Flow. Journal of Visual Experiments. 2010;45 doi: 10.3791/2004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Yuan S, Devor A, Boas DA, Dunn AK. Determination of optimal exposure time for imaging of blood flow changes with laser speckle contrast imaging. Appl Opt. 2005;44:1823–1830. doi: 10.1364/ao.44.001823. [DOI] [PubMed] [Google Scholar]
  • 52.Richards LM, Kazmi SMS, Davis JL, Olin KE, Dunn AK. Low-cost laser speckle contrast imaging of blood flow using a webcam. Biomed Opt Express. 2013;4:2269–2283. doi: 10.1364/BOE.4.002269. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Davis MA, Kazmi SMS, Dunn AK. Imaging depth and multiple scattering in laser speckle contrast imaging. J Biomed Opt. 2014;19:086001. doi: 10.1117/1.JBO.19.8.086001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Duncan DD, Kirkpatrick SJ. Can laser speckle flowmetry be made a quantitative tool? J Opt Soc Am A. 2008;25:2088–2094. doi: 10.1364/josaa.25.002088. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Kazmi SMS, Richards LM, Schrandt CJ, Davis MA, Dunn AK. Expanding applications, accuracy, and interpretation of laser speckle contrast imaging of cerebral blood flow. J Cereb Blood Flow Metab. 2015;35:1076–1084. doi: 10.1038/jcbfm.2015.84. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Parthasarathy AB, Weber EL, Richards LM, Fox DJ, Dunn AK. Laser speckle contrast imaging of cerebral blood flow in humans during neurosurgery: a pilot clinical study. J Biomed Opt. 2010;15:066030. doi: 10.1117/1.3526368. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Strong AJ, Bezzina EL, Anderson PBJ, Boutelle MG, Hopwood SE, Dunn AK. Evaluation of laser speckle flowmetry for imaging cortical perfusion in experimental stroke studies: quantitation of perfusion and detection of peri-infarct depolarisations. J Cerebr Blood Flow Metab. 2006;26:645–653. doi: 10.1038/sj.jcbfm.9600240. [DOI] [PubMed] [Google Scholar]
  • 58.Yu G, Durduran T, Zhou C, Cheng R, Yodh AG. Near-infrared diffuse correlation spectroscopy (DCS) for assessment of tissue blood flow. In: Boas DA, Pitris C, Ramanujam N, editors. Handbook of Biomedical Optics. Taylor & Francis; 2011. [Google Scholar]
  • 59.Boas DA, Campbell LE, Yodh AG. Scattering and Imaging with Diffusing Temporal Field Correlations. Phys Rev Lett. 1995;75:1855–1858. doi: 10.1103/PhysRevLett.75.1855. [DOI] [PubMed] [Google Scholar]
  • 60.Boas DA, Yodh AG. Spatially varying dynamical properties of turbid media probed with diffusing temporal light correlation. J Opt Soc Am A. 1997;14:192–215. [Google Scholar]
  • 61.Buckley EM, Parthasarathy AB, Grant PE, Yodh AG, Franceschini MA. Diffuse correlation spectroscopy for measurement of cerebral blood flow: future prospects. Neurophotonics. 2014;1 doi: 10.1117/1.NPh.1.1.011009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Durduran T, Yodh AG. Diffuse correlation spectroscopy for non-invasive, micro-vascular cerebral blood flow measurement. Neuroimage. 2014;85(Pt 1):51–63. doi: 10.1016/j.neuroimage.2013.06.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Baker WB, Parthasarathy AB, Busch DR, Mesquita RC, Greenberg JH, Yodh AG. Modified Beer-Lambert law for blood flow. Biomed Opt Express. 2014;5:4053–4075. doi: 10.1364/BOE.5.004053. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Himori N, Kunikata H, Shiga Y, Omodaka K, Maruyama K, Takahashi H, Nakazawa T. The association between systemic oxidative stress and ocular blood flow in patients with normal-tension glaucoma. Graefes Arch Clin Exp Ophthalmol. 2016;254:333–341. doi: 10.1007/s00417-015-3203-z. [DOI] [PubMed] [Google Scholar]
  • 65.Kunikata H, Aizawa N, Kudo M, Mugikura S, Nitta F, Morimoto R, Iwakura Y, Ono Y, Satoh F, Takahashi H, Ito S, Takahashi S, Nakazawa T. Relationship of ocular microcirculation, measured by laser speckle flowgraphy, and silent brain infarction in primary aldosteronism. PLoS One. 2015;10:e0117452. doi: 10.1371/journal.pone.0117452. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 66.Della Rossa A, D’Ascanio A, Barsotti S, Stagnaro C, Mosca M. Post-occlusive reactive hyperaemia (POHR) in systemic sclerosis: very early disease (VEDOSS) represents a separate entity compared to established disease. Scand J Rheumatol. 2016:1–4. doi: 10.3109/03009742.2015.1127411. [DOI] [PubMed] [Google Scholar]
  • 67.Eriksson S, Nilsson J, Lindell G, Sturesson C. Laser speckle contrast imaging for intraoperative assessment of liver microcirculation: a clinical pilot study. Med Devices (Auckl) 2014;7:257–261. doi: 10.2147/MDER.S63393. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68.Ideguchi M, Kajiwara K, Yoshikawa K, Goto H, Sugimoto K, Inoue T, Nomura S, Suzuki M. Avoidance of ischemic complications after resection of a brain lesion based on intraoperative real-time recognition of the vasculature using laser speckle flow imaging. J Neurosurg. 2016:1–7. doi: 10.3171/2016.1.JNS152067. [DOI] [PubMed] [Google Scholar]
  • 69.Richards LM, Towle EL, Fox DJ, Jr, Dunn AK. Intraoperative laser speckle contrast imaging with retrospective motion correction for quantitative assessment of cerebral blood flow. Neurophotonics. 2014;1:015006. doi: 10.1117/1.NPh.1.1.015006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70.Ringkamp M, Wooten M, Carson BS, Sr, Lim M, Hartke T, Guarnieri M. Laser speckle imaging to improve clinical outcomes for patients with trigeminal neuralgia undergoing radiofrequency thermocoagulation. J Neurosurg. 2016;124:422–428. doi: 10.3171/2015.1.JNS14408. [DOI] [PubMed] [Google Scholar]
  • 71.Ruaro B, Sulli A, Smith V, Paolino S, Pizzorni C, Cutolo M. Short-term follow-up of digital ulcers by laser speckle contrast analysis in systemic sclerosis patients. Microvasc Res. 2015;101:82–85. doi: 10.1016/j.mvr.2015.06.009. [DOI] [PubMed] [Google Scholar]
  • 72.Iwase T, Ra E, Yamamoto K, Kaneko H, Ito Y, Terasaki H. Differences of Retinal Blood Flow Between Arteries and Veins Determined by Laser Speckle Flowgraphy in Healthy Subjects. Medicine (Baltimore) 2015;94:e1256. doi: 10.1097/MD.0000000000001256. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Yu G, Floyd TF, Durduran T, Zhou C, Wang J, Detre JA, Yodh AG. Validation of diffuse correlation spectroscopy for muscle blood flow with concurrent arterial spin labeled perfusion MRI. Opt Express. 2007;15:1064–1075. doi: 10.1364/oe.15.001064. [DOI] [PubMed] [Google Scholar]
  • 74.Carp SA, Roche-Labarbe N, Franceschini MA, Srinivasan VJ, Sakadzic S, Boas DA. Due to intravascular multiple sequential scattering, Diffuse Correlation Spectroscopy of tissue primarily measures relative red blood cell motion within vessels. Biomed Opt Express. 2011;2:2047–2054. doi: 10.1364/BOE.2.002047. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 75.Lin Y, He L, Shang Y, Yu G. Noncontact diffuse correlation spectroscopy for noninvasive deep tissue blood flow measurement. J Biomed Opt. 2012;17:010502. doi: 10.1117/1.JBO.17.1.010502. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76.Li T, Lin Y, Shang Y, He L, Huang C, Szabunio M, Yu G. Simultaneous measurement of deep tissue blood flow and oxygenation using noncontact diffuse correlation spectroscopy flow-oximeter. Sci Rep. 2013;3:1358. doi: 10.1038/srep01358. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 77.Shang Y, Gurley K, Yu G. Diffuse Correlation Spectroscopy (DCS) for Assessment of Tissue Blood Flow in Skeletal Muscle: Recent Progress. Anat Physiol. 2013;3:128. doi: 10.4172/2161-0940.1000128. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 78.Verdecchia K, Diop M, Lee TY, St Lawrence K. Quantifying the cerebral metabolic rate of oxygen by combining diffuse correlation spectroscopy and time-resolved near-infrared spectroscopy. J Biomed Opt. 2013;18:27007. doi: 10.1117/1.JBO.18.2.027007. [DOI] [PubMed] [Google Scholar]
  • 79.Shang Y, Li T, Chen L, Lin Y, Toborek M, Yu G. Extraction of diffuse correlation spectroscopy flow index by integration of th-order linear model with Monte Carlo simulation. Appl Phys Lett. 2014;104:193703. doi: 10.1063/1.4876216. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 80.Binzoni T, Martelli F. Assessing the reliability of diffuse correlation spectroscopy models on noise-free analytical Monte Carlo data. Appl Opt. 2015;54:5320–5326. doi: 10.1364/AO.54.005320. [DOI] [PubMed] [Google Scholar]
  • 81.Buckley EM, Miller BF, Golinski JM, Sadeghian H, McAllister LM, Vangel M, Ayata C, Meehan WP, III, Franceschini MA, Whalen MJ. Decreased microvascular cerebral blood flow assessed by diffuse correlation spectroscopy after repetitive concussions in mice. J Cereb Blood Flow Metab. 2015;35:1995–2000. doi: 10.1038/jcbfm.2015.161. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 82.Han S, Hoffman MD, Proctor AR, Vella JB, Mannoh EA, Barber NE, Kim HJ, Jung KW, Benoit DS, Choe R. Non-Invasive Monitoring of Temporal and Spatial Blood Flow during Bone Graft Healing Using Diffuse Correlation Spectroscopy. PLoS One. 2015;10:e0143891. doi: 10.1371/journal.pone.0143891. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 83.Li Z, Baker WB, Parthasarathy AB, Ko TS, Wang D, Schenkel S, Durduran T, Li G, Yodh AG. Calibration of diffuse correlation spectroscopy blood flow index with venous-occlusion diffuse optical spectroscopy in skeletal muscle. J Biomed Opt. 2015;20:125005. doi: 10.1117/1.JBO.20.12.125005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 84.Choe R, Putt ME, Carlile PM, Durduran T, Giammarco JM, Busch DR, Jung KW, Czerniecki BJ, Tchou J, Feldman MD, Mies C, Rosen MA, Schnall MD, DeMichele A, Yodh AG. Optically Measured Microvascular Blood Flow Contrast of Malignant Breast Tumors. PLoS ONE. 2014;9:e99683. doi: 10.1371/journal.pone.0099683. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 85.Rottwit K, Tidemand-Lichtenberg P. Nonlinear Optics: Principles and Applications (Optical Sciences and Applications of Light) CRC Press; 2014. [Google Scholar]
  • 86.Zipfel WR, Williams RM, Webb WW. Nonlinear magic: multiphoton microscopy in the biosciences. Nature Biotechnology. 2003;21:1369–1377. doi: 10.1038/nbt899. [DOI] [PubMed] [Google Scholar]
  • 87.Evans CL, Xie XS. Coherent anti-Stokes Raman scattering microscopy: chemical imaging for biology and medicine. Annu Rev Anal Chem. 2008;1(2008):883–909. doi: 10.1146/annurev.anchem.1.031207.112754. [DOI] [PubMed] [Google Scholar]
  • 88.Masters BR, So PTC. Handbook of Biomedical Nonlinear Optical Microscopy. Oxford University Press; 2008. [Google Scholar]
  • 89.Min W, Freudiger CW, Lu S, Xie XS. Coherent nonlinear optical imaging: beyond fluorescence microscopy. Annual Review of Physical Chemistry. 2011;62:507–530. doi: 10.1146/annurev.physchem.012809.103512. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 90.König K. Clinical multiphoton tomography. J Biophoton. 2008;1:13–23. doi: 10.1002/jbio.200710022. [DOI] [PubMed] [Google Scholar]
  • 91.Chen L-C, Lloyd WR, Kuo S, Kim HM, Marcelo CL, Feinberg SE, Mycek M-A. The potential of label-free nonlinear optical molecular microscopy to non-invasively characterize the viability of engineered human tissue constructs. Biomaterials. 2014;35:6667–6676. doi: 10.1016/j.biomaterials.2014.04.080. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 92.Weinigel M, Breunig HG, Darvin ME, Klemp M, Röwert-Huber J, Lademann J, König K. Impact of refractive index mismatches on coherent anti Stokes Raman scattering and multiphoton autofluorescence tomography of human skin in vivo. Phys Med Biol. 2015;60:6881–6899. doi: 10.1088/0031-9155/60/17/6881. [DOI] [PubMed] [Google Scholar]
  • 93.Lee MH, Arcidiacono JA, Bilek AM, Wille JJ, Hamill CA, Wonnacott KM, Wells MA, Oh SS. Considerations for tissue-engineered and regenerative medicine product development prior to clinical trials in the United States. Tissue Engineering Part B: Reviews. 2010;16:41–54. doi: 10.1089/ten.TEB.2009.0449. [DOI] [PubMed] [Google Scholar]
  • 94.Lloyd WR, Chen L-C, Mycek M-A. Fluorescence Spectroscopy. In: Morgan SP, Rose FR, Matcher SJ, editors. Optical Techniques in Regenerative Medicine. Ch.7. CRC Press – Taylor & Francis Group; 2013. pp. 171–204. [Google Scholar]
  • 95.Chang T, Zimmerley MS, Quinn KP, Lamarre-Jouenne I, Kaplan DL, Beaurepaire E, Georgakoudi I. Non-invasive monitoring of cell metabolism and lipid production in 3D engineered human adipose tissues using label-free multiphoton microscopy. Biomaterials. 2013;34:8607–8616. doi: 10.1016/j.biomaterials.2013.07.066. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 96.Chen L-C, Lloyd WR, Chang C-W, Sud D, Mycek M-A. Fluorescence lifetime imaging microscopy for quantitative biological imaging. In: Sluder G, Wolf DE, editors. Methods in Cell Biology Digital Microscopy 4th Edition. Elsevier; San Diego: 2013. [DOI] [PubMed] [Google Scholar]
  • 97.Nadeau KP, Rice TB, Durkin AJ, Tromberg BJ. Multifrequency synthesis and extraction using square wave projection patterns for quantitative tissue imaging. J Biomed Opt. 2015;20:116005. doi: 10.1117/1.JBO.20.11.116005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 98.Andersson-Engels S, Berg R, Persson A, Svanberg S. Multispectral tissue characterization with time-resolved detection of diffusely scattered white light. Opt Lett. 1993;18:1697–1699. doi: 10.1364/ol.18.001697. [DOI] [PubMed] [Google Scholar]
  • 99.Zint CV, Uhring W, Torregrossa M, Cunin B, Poulet P. Streak camera: a multidetector for diffuse optical tomography. Appl Opt. 2003;42:3313–3320. doi: 10.1364/ao.42.003313. [DOI] [PubMed] [Google Scholar]
  • 100.Sidorov IS, Miridonov SV, Nippolainen E, Kamshilin AA. Estimation of light penetration depth in turbid media using laser speckles. Opt Express. 2012;20:13692–13701. doi: 10.1364/OE.20.013692. [DOI] [PubMed] [Google Scholar]
  • 101.Roche-Labarbe N, Fenoglio A, Radhakrishnan H, Kocienski-Filip M, Carp SA, Dubb J, Boas DA, Grant PE, Franceschini MA. Somatosensory evoked changes in cerebral oxygen consumption measured non-invasively in premature neonates. Neuroimage. 2014;85:279–286. doi: 10.1016/j.neuroimage.2013.01.035. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES