Reflectance-mode interferometric near-infrared spectroscopy quantifies brain absorption, scattering, and blood flow index in vivo

Dawid Borycki 1, Oybek Kholiqov 1, Vivek J Srinivasan 1,*
PMCID: PMC5565174  NIHMSID: NIHMS893371  PMID: 28146535

Abstract

Interferometric near-infrared spectroscopy (iNIRS) is a new technique that measures time-of-flight- (TOF-) resolved autocorrelations in turbid media, enabling simultaneous estimation of optical and dynamical properties. Here, we demonstrate reflectance-mode iNIRS for noninvasive monitoring of a mouse brain in vivo. A method for more precise quantification with less static interference from superficial layers, based on separating static and dynamic components of the optical field autocorrelation, is presented. Absolute values of absorption, reduced scattering, and blood flow index (BFI) are measured, and changes in BFI and absorption are monitored during a hypercapnic challenge. Absorption changes from TOF-resolved iNIRS agree with absorption changes from continuous wave NIRS analysis, based on TOF-integrated light intensity changes, an effective path length, and the modified Beer–Lambert Law. Thus, iNIRS is a promising approach for quantitative and non-invasive monitoring of perfusion and optical properties in vivo.


Quantifying optical properties to infer chromophore concentrations [1] and quantifying scattered light dynamics to infer scatterer motion [2] are ongoing goals of near-infrared spectroscopy (NIRS) [3]. In biological tissue, the absorption coefficient, if measured at multiple wavelengths, yields oxy-hemoglobin concentration, de oxy-hemoglobin concentration, and oxygen saturation. Coherent intensity fluctuations of multiply scattered light, if analyzed with a model informed by optical properties, yield scatterer motion due to blood flow [4]. The combination of oxygen saturation, hemoglobin content, and blood flow provides tissue oxygen consumption.

Continuous wave (CW) NIRS measures reemitted light intensity and is the simplest and most widely used NIRS method [1,3]. However, accurate quantification of baseline optical properties using only light intensity requires additional assumptions about scattered path length, instrumentation loss, and extra cerebral confounds [3]. Hence, baseline oxygen saturation remains challenging to quantify robustly using CW NIRS. For baseline optical properties, time-domain (TD) and frequency-domain (FD) approaches [5,6] add an additional dimension to the measurement space to reduce the number of required assumptions.

To determine blood flow, diffuse correlation spectroscopy (DCS) quantifies temporal intensity fluctuations of coherent light scattered from dynamic tissue [2,4]. DCS critically depends on optical properties in two ways. First, since DCS is a CW approach, the relevant field autocorrelation is an average of field autocorrelations over all photon path lengths, weighted by the optical property-dependent photon time-of-flight (TOF) distribution [7,8]. Second, the total photon momentum transfer, which depends on the reduced scattering coefficient ( μs) in the diffusion regime, is required to relate dynamics to an effective diffusion coefficient or blood flow index (BFI). In addition, conventional DCS requires ergodicity since it derives field autocorrelations from intensity fluctuations [9].

Recently, interferometric NIRS (iNIRS), based on the TOF-resolved measurement of coherent light reemitted from a turbid medium, was introduced [10]. This approach is based on interferometry using a tunable laser to achieve path length resolution [11,12]. A two-dimensional measurement of reemitted light provides TOF-resolved autocorrelations, from which optical and dynamical properties can be derived [10]. The iNIRS method recently was shown to also provide TOF-resolved field autocorrelations, more directly related to scatterer dynamics than TOF-integrated intensity autocorrelations provided by DCS [13]. In this Letter, we validate iNIRS in reflectance mode using calibrated phantoms, and in the brain of a nude mouse, showing that iNIRS can simultaneously and non invasively quantify optical properties and BFI in vivo.

Here, we make use of the two-dimensional field autocorrelation, G1(iNIRS)(τs,τd)=Γrs(τs,td)Γrs(τs,td+τd)td, measured as a function of TOF (τs) and lag time (τd). Γrs(τs, td) is the TOF-resolved mutual coherence function of the reference and sample fields (Ur and Us, respectively): Γrs(τs,td)=Ur(ts,td)Us(ts+τs,td)ts.

A light path can be envisioned as a series of scattering events, either from stationary tissue (static) or from blood (dynamic), where each scattering event has an associated momentum transfer q and phase shift δr · q [14], where δr is scatterer displacement. Thus, the instantaneous TOF-resolved scattered optical field from the sample can be written as Us(τs, td) = Uc(τs) + Uf(τs, td) where Uc is the static field contribution, arising from paths with static scattering events or dynamic scattering events with negligible phase shift, while Uf is the dynamic field contribution, arising from paths with dynamic scattering events with a phase shift. The static contribution is understood to be nonergodic over a certain time scale [9]. Hence, the autocorrelation measured in iNIRS becomes [13]

G1(iNIRS)(τs,τd)=[Ic(τs)+G1,f(τs,τd)]I0(τs), (1)

where Ic(τs) = |Uc(τs)|2 is the static distribution of time-of-flight (DTOF), G1,f(τs,τd)=Uf(τs,td)Uf(τs,td+τd)td is the dynamic autocorrelation, I0(τs)is the instrument response function (IRF), If(τs) = G1,f(τs, 0)is the dynamic DTOF, and “*” denotes convolution with respect to τs.

The static (Ic, slowly decorrelating) and dynamic (G1,f, rapidly decorrelating) components of the field autocorrelation function may be associated with different photon paths. The ratio of static to dynamic components should diminish for larger TOFs, as momentum transfer from dynamic scattering is more probable for long paths than short paths. By independently assessing these two components, these paths may be independently probed. This new approach of correlation gating is used here to improve the accuracy of iNIRS analysis. In our in vivo experiment, some photons may propagate through tissue without experiencing dynamic scattering events. Such static photons may not pass through blood, may not be diffuse, and thus should be excluded from an analysis that assumes diffuse photons and an ergodic, uniformly absorbing medium. By using correlation gating to exclude Ic and analyze If or G1,f, it may be possible to better satisfy the assumptions of the MBLL and diffusion theory for an ergodic, uniformly absorbing medium. By evaluating Eq. (1) at τd = 0, we obtain that Is(iNIRS)(τs)=G1(iNIRS)(τs,0)=Is(τs)I0(τs), or the temporal point spread function (TPSF) is equal to the DTOF, Is(τs) = Ic(τs)+If(τs), convolved with the IRF. Alternatively, Is(iNIRS)(τs) could have been obtained through conventional temporal speckle averaging. Since G1,f(τs, τd) decorrelates with increasing τd, the static TPSF can be determined by Ic(iNIRS)(τs)=limτdG1(iNIRS)(τs,τd)=Ic(τs)I0(τs). Practically, the asymptotic limit is replaced by a time scale, τd = τd,c, that experimentally defines the boundary between static and dynamic components. The TOF-dependent decay rate ξ(τs) of G1,f(τs,τd) relates to the effective diffusion coefficient of light scatterers. Last, by subtracting Ic(iNIRS)(τs),G1,f(iNIRS)(τs,τd)=G1,f(τs,τd)I0(τs) yields the dynamic TPSF: If(iNIRS)(τs)=G1,f(iNIRS)(τs,0)=If(τs)I0(τs).

The spectral density, 2Re[𝒲rs(ν, td)], related to Γrs(τs, td)by a Fourier transform [10], is measured over time td and ν (conjugate variable to τs) by a MachZehnder interferometer with a narrow linewidth, rapidly tunable distributed feedback (DFB) laser [Fig. 1]. Light is divided into reference and sample arms [Fig. 1(A)]. The collimated sample beam irradiates the turbid medium, and the diffusively reflected light is collected at a distance ρ away. The light paths are combined with a fiber coupler (780HP) and detected by a differential detector. The digitized electronic signal approximates 2Re[𝒲rs]. Since the laser is swept in frequency, short photon paths produce smaller beat frequencies than long paths, provided that the reference arm TOF is shorter than the shortest sample arm TOF [Fig. 1(A)]. TOF is thus encoded as electronic frequency and can be decoded by re sampling and inverse Fourier transformation [10] to yield the complex Γrs time series [Fig. 1(B)].

Fig. 1.

Fig. 1

Reflectance-mode iNIRS. (A) A Mach–Zehnder interferometer with a frequency tuned, temporally coherent light source measures a spectral interference signal (2Re[𝒲rs(ν)]) produced by optical fields from the reference and sample arms. DFB, distributed feedback laser centered at λ0 = 855 nm with a tuning range of ~60 pm and static coherence length of >100 m; APP, anamorphic prism pair; OI, optical isolator; FOPC, fiber optical polarization controller; DBD, dual balanced detector; L1–L4, lenses; PC, personal computer. (B) TOF-resolved fluctuations of Γrs, obtained by inverse Fourier transformation of 2Re[𝒲rs], determine both TPSFs and autocorrelations. (C) In analysis I, TPSFs are used to obtain optical properties. (D) In analysis II, TOF-resolved field autocorrelations yield scatterer dynamics, using the reduced scattering coefficient and DWS theory. (E) Optical properties of fluid phantoms, in which varying amounts of India Ink and Intralipid 20% were mixed in water, were obtained using Eq. (2) with n = 1.33 and ρ = 7.5 mm and then compared to true values. (F) Phantom dynamics were extracted by fitting TOF-dependent autocorrelations with DWS theory and compared with DCS analysis of TOF-integrated autocorrelations (note that the probability of dynamic scattering, α, is unity for these samples).

In analysis I [Fig. 1(C)], the resulting TPSFs are fit to the expression for TOF-resolved diffuse reflectance from a semi infinite turbid medium [15], ℛ(ρ, τs), to extract optical properties (μa and μs). For perpendicularly incident collimated light,

R(ρ,τs)=exp[-μacτs/n-ρ2/2σt2]4πσt2τs3/24πDc/n{z0exp[-z02ϰ(τs)]+(z0+2zb)exp[-(z0+2zb)2ϰ(τs)]}, (2)

where c denotes the speed of light, D=13μs is the diffusion coefficient, z0 = 3D denotes the depth of the positive image source, while zb = 2D(1 + R)/(1 − R) is the extrapolated boundary corresponding to the negative image source located at z = −z0 − 2zb, and R = −1.4399n−2 + 0.7099n−1 + 0.6681 + 0.0636n denotes the reflection coefficient [16], which depends on the refractive index n, and ϰ(τs) = 4Dcτs/n, whereas σt2=σs2+σd2+2Dcτs/n incorporates the extended source and detector collimated modes, described by Gaussian intensity distributions with experimentally determined variances of σs2=0.09mm2 and σd2=0.04mm2, respectively [10].

To incorporate finite TOF resolution, ℛ(ρ, τs)is then convolved with the IRF. The shape of the IRF is determined from a free-space transmission measurement. To obtain the reflectance-mode IRF, the transmission-mode IRF is then shifted along the TOF axis, so its peak coincides with τs = 0. This reference point is obtained from the peak location of the TPSF acquired for ρ = 0 and corrected based on Monte Carlo simulations. The fitting window starts at τs = 35 ps to exclude some nondiffuse, early light and continues into the tails as long as the peak-normalized TPSF values are larger than 0.1. Fitting results (based on N = 40,000 iNIRS signals acquired with a laser sweep duration of ~10 μs at 50 kHz with ~25 mW incident power) in fluid phantoms [Fig. 1(E)] support the capability to quantify optical properties within the physiological range.

Given the optical properties, sample dynamics can be obtained from analysis II [Fig. 1(D)]. According to diffusing wave spectroscopy (DWS) theory, the TOF-resolved normalized field autocorrelation is [7] g1(DWS)(τs,τd)=exp[-ξ(τs)τd], where ξ(τs)=2k2αDBμscτs/n is the TOF-resolved decay rate, where k is the medium wave number, αDB is the BFI, the product of the probability of dynamic scattering, α, and the effective diffusion coefficient, DB [4]. Thus, the field fluctuation rate increases with photon path length, l = s/n, and μs. In iNIRS, both are known, so the sample dynamics, quantified αDB, can be directly determined by fitting g1,f(iNIRS)(τs,τd)=G1,f(iNIRS)(τs,τd)/G1,f(iNIRS)(τs,0) with DWS theory. This procedure neglects the possibility of multiple decay rates, weighted by the DTOF, within the IRF window (~45 ps FWHM). For fluid phantoms [Fig. 1(F)], we obtained αDB = (2.33 ± 0.42) × 10−12 m2 s−1 and αDB = (2.04 ± 0.26) × 10−12 m2 s−1 for samples with varying absorption and scattering, respectively, in agreement with prior results [10]. TOF-integrated autocorrelations were fit with correlation diffusion Equation solutions for reflectance as in DCS [4]. The average DCS αDB value is shown in Fig. 1(F) as a dotted line.

Next, iNIRS was performed in a nude mouse (male, SKH-1E Charles River) brain during a hypercapnic challenge in vivo. Anesthesia was induced using a mixture of isoflurane, medical air, and oxygen. The mouse was placed on a heating pad and immobilized in a stereotactic frame. Isoflurane was maintained at 1.2% v/v during data acquisition to maintain vasoreactivity. After recording the TPSF for ρ = 0, the source and detector light beams were separated by a distance ρ = 7.5 mm between the parietal cortices of the right and left hemispheres. The CO2 supply was changed from 0% (normocapnia) to 7.5% (hypercapnia) and then back to 0%. During each stage, measurements containing N = 40,000 iNIRS signals (0.8 s per measurement) were recorded at intervals of ~17 s. Last, the reference arm signal alone was acquired. All experimental procedures and protocols were reviewed and approved by the UC Davis Institutional Animal Care and Use Committee.

In vivo, Is(iNIRS)(τs) and If(iNIRS)(τs) [Fig. 2(A)] were fit with ℛ(ρ, τs) *I0(τs) The static component was estimated at τd,c = 0.8 ms. While this cutoff is reasonable for the mouse brain [17], more general fitting approaches are also possible to estimate the static component. To determine the baseline scattering properties, fitting was first performed on the data sets for FiCO2 = 0% with n = 1.4 and both μa and μs as free parameters. Next, for consistency with the CW NIRS approach described below, scattering was fixed to the average baseline value of μs=0.831mm-1 for Is and μs=0.854mm-1 for If Then, the fitting procedure was repeated for all time points to determine absorption coefficients. The average baseline absorption was μa = 2.62 × 10−2 mm−1 for Is and μa = 2.67 × 10−2 mm−1 for If [Fig. 2(B)].

Fig. 2.

Fig. 2

TPSFs from a mouse brain in vivo. (A) The normocapnic TPSF with resolved static and dynamic components, determined from the autocorrelation (example at τs = 35 ps in inset). The static component is significant for short TOFs only, suggesting more superficial paths. For brevity, the superscript (iNIRS) was omitted in the inset. (B) The asymptotic slope of the hypercapnic dynamic TPSF is larger than the normocapnic one due to increased absorption during hypercapnia (Fig. 3).

To compare conventional CW NIRS with iNIRS, unnormalized total (s) and dynamic (f) TPSFs were integrated over τs The resulting TOF-integrated intensities, ℐs(t) and ℐf(t) with t denoting the time at which each data set was recorded), yielded temporal optical density changes ΔODs,f(t)=ln[Is,f(0)]-ln[Is,f(t)]. The baseline intensities Is,f(0) were computed by averaging ℐs(t) or ℐf(t) over the first ~10 minutes of the experiment. Given the optical density changes, absorption changes were determined from the modified Beer–Lambert law (MBLL): Δμa(s,f)(t)=ΔODs,f(t)/l¯s,f, in which s,f is the average path length [18], determined as the centroid of the total (s = 31.03 mm) or dynamic (f = 31.24 mm) TPSFs. Absorption changes, Δμa, determined using either the MBLL (CW NIRS) or Eq. (2) (iNIRS), are compared in Fig. 3. Note that iNIRS analysis is based on the TPSF shape alone, while the MBLL uses the detected intensity. Both methods show an absorption increase linked to increases in blood oxygenation and blood volume. We hypothesize that the so-called “static” component may fluctuate slowly over minutes during the experiment, leading to changes in ℐs(t) which CW NIRS analysis incorrectly attributes to absorption changes [Figs. 3(A) and 3(C)]. Importantly, correlation gating isolates the dynamic intensity and improves agreement between methods [Figs. 3(B) and 3(D)].

Fig. 3.

Fig. 3

Cross validation of absorption changes determined by iNIRS and CW NIRS in vivo. (A) The total TOF-integrated TPSF includes static signals and thus leads to some discrepancy between iNIRS and CW NIRS. This discrepancy is reduced when the static component is excluded. (B) Parametric plots of absorption changes based on (C) total and (D) dynamic intensity show that agreement is better for dynamic intensity (p < 0.001 for a paired t-test of distances from the line of equality).

Field autocorrelation decay rates, ξ(τs), were used to determine BFI (αDB) via DWS theory and then compared to DCS analysis. As shown in Fig. 4(A), averaged normocapnic decay rates (DRs) of total [ G1(iNIRS)] and dynamic [ G1,f(iNIRS)] autocorrelations increase with TOF. At short TOFs the static component reduces the DR of the total autocorrelation and worsens the R2 value of the exponential fit (inset). Except for the short TOF range, where DTOF weighting within the IRF window [Eq. (1)] increases DRs, dynamic autocorrelation DRs are almost linear with TOF as predicted by DWS theory. Therefore, time courses of dynamic autocorrelation DRs [Fig. 4(C)] differ depending on TOF, but BFIs agree for sufficiently large TOF [Fig. 4(D)]. Figures 4(B) and 4(D) suggest that vasodilation during hypercapnia increases cerebral blood flow. Moreover, BFIs determined by TOF-resolved iNIRS analysis and DCS analysis agree.

Fig. 4.

Fig. 4

BFI quantification by iNIRS in vivo. (A) Averaged normocapnic autocorrelation DRs increase with photon path length. For each TOF, R2 values for exponential fits to field autocorrelations, both with and without correlation gating (“dynamic” and “total” in inset, respectively), were converted to z-scores by Fisher transformation. Paired t-tests on the z-scores revealed that correlation gating increased R2 values for all TOFs (p < 0.001). (B)–(D) Averaged hypercapnic DRs are larger than their normocapnic counterparts due to the transient increase in BFI. Shaded regions in (A) and (B) denote standard deviations.

In summary, we demonstrated reflectance-mode iNIRS for continuous monitoring of a mouse brain in vivo during a hypercapnic challenge. Since iNIRS can measure the mean path length, iNIRS can quantify CW NIRS based on the MBLL. At the same time, iNIRS provides absorption and scattering coefficients based on fitting by diffusion theory and flow indices based on fitting by diffusing wave spectroscopy. Though iNIRS BFI values are in agreement with prior DCS measurements in mice [19], further cross validation is warranted. Ongoing work will optimize detection sensitivity and penetration depth for brains of larger species and compare iNIRS to conventional DCS. iNIRS integrates capabilities of NIRS and DCS into a single modality and may be a promising approach for a quantitative, noninvasive oxygenation and metabolism monitoring system in vivo.

Acknowledgments

Funding. National Institutes of Health (NIH) (R01NS094681).

Footnotes

OCIS codes: (030.6600) Statistical optics; (030.1640) Coherence; (160.4760) Optical properties; (170.6480) Spectroscopy, speckle.

References

  • 1.Jöbsis FF. Science. 1977;198:1264. doi: 10.1126/science.929199. [DOI] [PubMed] [Google Scholar]
  • 2.Durduran T, Yodh AG. NeuroImage. 2014;85:51. doi: 10.1016/j.neuroimage.2013.06.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Scholkmann F, Kleiser S, Metz AJ, Zimmermann R, Pavia JM, Wolf U, Wolf M. NeuroImage. 2014;85:6. doi: 10.1016/j.neuroimage.2013.05.004. [DOI] [PubMed] [Google Scholar]
  • 4.Yu G, Durduran T, Zhou C, Cheng R, Yodh AG. Near-Infrared Diffuse Correlation Spectroscopy for Assessment of Tissue Blood Flow. CRC Press; 2011. [Google Scholar]
  • 5.Torricelli A, Contini D, Pifferi A, Caffini M, Re R, Zucchelli L, Spinelli L. NeuroImage. 2014;85:28. doi: 10.1016/j.neuroimage.2013.05.106. [DOI] [PubMed] [Google Scholar]
  • 6.Fantini S, Franceschini MA, Fishkin JB, Barbieri B, Gratton E. Appl Opt. 1994;33:5204. doi: 10.1364/AO.33.005204. [DOI] [PubMed] [Google Scholar]
  • 7.Pine DJ, Weitz DA, Chaikin PM, Herbolzheimer E. Phys Rev Lett. 1988;60:1134. doi: 10.1103/PhysRevLett.60.1134. [DOI] [PubMed] [Google Scholar]
  • 8.Yodh AG, Kaplan PD, Pine DJ. Phys Rev B. 1990;42:4744. doi: 10.1103/physrevb.42.4744. [DOI] [PubMed] [Google Scholar]
  • 9.Parthasarathy AB, Tom WJ, Gopal A, Zhang X, Dunn AK. Opt Express. 2008;16:1975. doi: 10.1364/oe.16.001975. [DOI] [PubMed] [Google Scholar]
  • 10.Borycki D, Kholiqov O, Chong SP, Srinivasan VJ. Opt Express. 2016;24:329. doi: 10.1364/OE.24.000329. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Tualle JM, Nghiêm HL, Cheikh M, Ettori D, Tinet E, Avrillier S. J Opt Soc Am A. 2006;23:1452. doi: 10.1364/josaa.23.001452. [DOI] [PubMed] [Google Scholar]
  • 12.Mei L, Somesfalean G, Svanberg S. Biomed Opt Express. 2014;5:2810. doi: 10.1364/BOE.5.002810. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Borycki D, Kholiqov O, Srinivasan VJ. Optica. 2016;3:1471. doi: 10.1364/OPTICA.3.001471. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Bonner R, Nossal R. Appl Opt. 1981;20:2097. doi: 10.1364/AO.20.002097. [DOI] [PubMed] [Google Scholar]
  • 15.Kienle A, Patterson MS. J Opt Soc Am A. 1997;14:246. doi: 10.1364/josaa.14.000246. [DOI] [PubMed] [Google Scholar]
  • 16.Contini D, Martelli F, Zaccanti G. Appl Opt. 1997;36:4587. doi: 10.1364/ao.36.004587. [DOI] [PubMed] [Google Scholar]
  • 17.Kazmi SMS, Wu RK, Dunn AK. Opt Lett. 2015;40:3643. doi: 10.1364/OL.40.003643. [DOI] [PubMed] [Google Scholar]
  • 18.Durduran T, Choe R, Baker WB, Yodh AG. Rep Prog Phys. 2010;73:076701. doi: 10.1088/0034-4885/73/7/076701. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Buckley EM, Miller BF, Golinski JM, Sadeghian H, McAllister LM, Vangel M, Ayata C, Meehan WP, III, Franceschini MA, Whalen MJ. J Cereb Blood Flow Metab. 2015;35:1995. doi: 10.1038/jcbfm.2015.161. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES