Abstract
Purpose
Pulse height spectroscopy has been used by investigators to deduce the imaging properties of scintillators. Pulse height spectra (PHS) are used to compute the Swank factor, which describes the variation in scintillator light output per x‐ray interaction. The spread in PHS measured below the K‐edge is related to the optical component of the Swank factor, i.e., variations in light escape efficiency from different depths of x‐ray interaction in the scintillator, denoted . Optimizing scintillators for medical imaging applications requires understanding of these optical properties, as they determine tradeoffs between parameters such as x‐ray absorption, light yield, and spatial resolution. This work develops a model for PHS acquisition such that the effect of measurement uncertainty can be removed. This method allows to be quantified on an absolute scale and permits more accurate estimation of the optical Swank factor of scintillators.
Methods
The pulse height spectroscopy acquisition chain was modeled as a linear system of stochastic gain stages. Analytical expressions were derived for signal and noise propagation through the PHS chain, accounting for deterministic and stochastic aspects of x‐ray absorption, scintillation, and light detection with a photomultiplier tube. The derived expressions were used to calculate PHS of thallium‐doped cesium iodide (CsI) scintillators using parameters that were measured, calculated, or known from literature. PHS were measured at 25 and 32 keV of CsI samples designed with an optically reflective or absorptive backing, with or without a fiber‐optic faceplate (FOP), and with thicknesses ranging from 150–1000 μm. Measured PHS were compared with calculated PHS, then light escape model parameters were varied until measured and modeled results reached agreement. Resulting estimates of were used to calculate each scintillator's optical Swank factor.
Results
For scintillators of the same optical design, only minor differences in light escape efficiency were observed between samples with different thickness. As thickness increased, escape efficiency decreased by up to 20% for interactions furthest away from light collection. Optical design (i.e., backing and FOP) predominantly affected the magnitude and relative variation in . Depending on interaction depth and scintillator thickness, samples with an absorptive backing and FOP were estimated to yield 4.1–13.4 photons/keV. Samples with a reflective backing and FOP yielded 10.4–18.4 keV−1, while those with a reflective backing and no FOP yielded 29.5–52.0 keV−1. Optical Swank factors were approximately 0.9 and near‐unity in samples featuring an absorptive or reflective backing, respectively.
Conclusions
This work uses a modeling approach to remove the noise introduced by the measurement apparatus from measured PHS. This method allows absolute quantification of and more accurate estimation of the optical Swank factor of scintillators. The method was applied to CsI scintillators with different thickness and optical design, and determined that optical design more strongly affects and Swank factor than differences in CsI thickness. Despite large variations in between optical designs, the Swank factor of all evaluated samples is above 0.9. Information provided by this methodology can help validate Monte Carlo simulations of structured CsI and optimize scintillator design for x‐ray imaging applications.
Keywords: cesium iodide, digital radiography, indirect detection, pulse height spectroscopy, Swank factor
1. Introduction
Flat panel imagers (FPIs) are widely used in medical imaging systems, and are generally classified as either indirect or direct.1 Indirect FPIs use scintillators to convert energy of x‐ray interactions into bursts of optical light, and collect this light to form images using a 2D array of a‐Si:H photodiodes and thin‐film transistors. Direct FPIs replace the scintillator and photodiode components with a biased x‐ray photoconductor, which directly converts the incident x‐rays into a charge image for readout.
Swank2, 3 showed that fluctuations in the conversion gain of x‐ray converters (i.e., scintillators or photoconductors) adds to the noise of such imaging systems. This phenomenon is studied using pulse height spectroscopy,4 which characterizes the probability distribution of the number of quanta output by the converter per x‐ray interaction. Pulse height spectroscopy has been used by investigators to evaluate x‐ray conversion in x‐ray image intensifiers,2, 5 scintillators used in screen‐film,6, 7, 8, 9 computed radiography10 and digital radiography systems,11, 12, 13, 14, 15, 16 photoconductors used in direct FPIs such as a‐Se,17, 18, 19 and photon counting detectors for mammography20, 21 and computed tomography.22, 23
The spread in a pulse height spectrum is quantified by the Swank factor A s , which plays a large role in determining the converter's ability to efficiently transfer large‐area information. This aspect of performance is represented by the zero‐frequency detective quantum efficiency DQE(0). In an energy‐integrating detector, DQE(0) is calculated by the product of the converter's x‐ray quantum efficiency A Q and A s 2, 3:
(1) |
The Swank factor of a scintillator may be separated into an optical Swank factor (I OPD ) and x‐ray Swank factor (I AED ), according to2, 3:
(2) |
The x‐ray Swank factor I AED describes variations in the energy deposited in the scintillator per x‐ray interaction. These variations not only affect DQE(0) of energy‐integrating detectors, but are also an important consideration in the precision of energy measurements in single‐photon counting detectors.24 Factors that degrade I AED include spread in the incident x‐ray energy distribution and the escape of characteristic fluorescence or Compton scattered photons from the scintillator.3, 25 Hajdok et al.25 and more recently Yun et al.26 have described comprehensive analytical models which consider these effects and allow calculation of I AED in x‐ray converters.
Below the K‐edge of scintillators typically used in indirect FPIs (e.g., CsI:Tl), x‐ray interaction is dominated by the photoelectric effect.27 For monoenergetic x‐rays below the K‐edge, I AED approaches unity and A s is determined solely by I OPD , which depends on random variations in x‐ray conversion gain and the inherent optical properties of the scintillator. Consequently, pulse height spectra (PHS) measured under these conditions may be interpreted to assess scintillator optical properties, such as how efficiently light escapes from different depths of x‐ray interaction.12 The light escape properties of a scintillator may be influenced by optical design factors such as an optically absorptive or reflective backing to improve different aspects of performance (e.g., high spatial resolution or high light output, respectively).28 Understanding how design factors influence scintillator optical properties is crucial for indirect detector optimization. Drangova and Rowlands8 first demonstrated this by showing how reflective and absorptive backings in various powder phosphors change PHS shape and affect the Swank factor and DQE(0).
Previous work by Lubinsky et al.12 showed how PHS can be analyzed to estimate the depth dependence of relative light escape from a scintillator. Their approach has improved the understanding of the Swank factor and optical transport within scintillators, although it has two inherent limitations: (a) it is susceptible to errors from uncertainty in PHS measurement and; (b) no information on the absolute light escape efficiency is determined. Other investigators, such as Badano et al.,14, 29 Freed et al.30, 31 and Sharma et al.32 have used Monte Carlo simulations to investigate scintillator imaging properties, using experimental data such as Swank factor to validate simulation parameters. Such simulations are advantageous for estimating scintillator properties that are not readily studied experimentally, such as depth‐dependent blur,33 but rely heavily on measured data for validation. Accurate estimation of depth‐dependent light escape efficiency and Swank factor directly from measurements of PHS will be useful for further validation of Monte Carlo simulations of scintillator optics, though to the best of our knowledge such methods do not currently exist. A more complete understanding of light propagation in scintillators will facilitate their optimization for medical imaging applications, as these properties affect tradeoffs between parameters such as x‐ray absorption, light yield, and spatial resolution.
In the present work, we develop a method to remove the effect of noise introduced by the PHS measurement apparatus and derive the absolute depth‐dependent light escape efficiency of a scintillator, which enables more accurate estimation of its optical Swank factor. In our approach, the PHS are modeled by accounting for the stochastic gain processes of x‐ray absorption, scintillation and optical detection with a photomultiplier tube (PMT). This is achieved using analytical expressions for signal and noise propagation through each stage of the pulse height spectroscopy chain. Measured PHS of various thallium‐doped CsI scintillators, which we hereafter refer to as CsI, are compared with modeled PHS to deduce their depth‐dependent light escape efficiency. This analysis is used to investigate how scintillator thickness and optical design affect light escape efficiency and the Swank factor of CsI scintillators used in flat panel detectors.
2. Materials and methods
2.A. Analytical model of the noise‐free pulse height spectrum:
An analytical expression for PHS was derived based on x‐ray absorption, scintillation and light escape processes as a function of x‐ray interaction depth. Here, these processes are treated as deterministic, that is, noiseless, however we extend our analysis in Section 2.B.. to determine how statistical variations in these processes also affect PHS.
We consider the case where x‐rays are monoenergetic and below the scintillator K‐edge. We assume that x‐ray attenuation in the scintillator is solely due to photoelectric interactions, which is a reasonable approximation in high‐Z scintillators (e.g., CsI). For example, Compton scatter events occur in less than 1.5% of x‐ray interactions in CsI below the K‐edge of I.27 Coherent scatter also occurs in less than 10% of x‐ray interactions at these energies, however, these events do not deposit energy in the scintillator and thus do not contribute to PHS; therefore, they are ignored.
The detector geometry used in our analysis, shown in Fig. 1, matches the typical geometry of experimental measurements, where x‐rays are orthogonally incident to a scintillator with thickness L.
2.A.1. X‐ray absorption and conversion gain
For N 0 monoenergetic x‐rays incident on a scintillator with photoelectric linear attenuation coefficient μ, the number absorbed from depth 0 to z is:
(3) |
and the total number of absorption events is:
(4) |
Each interaction results in scintillation with a conversion gain of , which is determined by the absorbed energy E and the energy required to produce one scintillation photon W:
(5) |
We assume E is equal to the incident x‐ray energy, which is reasonable in the case we have considered here where: (a) x‐ray energy is below the scintillator K‐edge, such that there are no variations in energy absorption due to the generation and escape of K‐fluorescence and; (b) variations in E due to Compton scatter are improbable and can be neglected.27 More detailed models25, 26 should be considered in cases where characteristic fluorescence or Compton interactions cannot be avoided or neglected, e.g., with low‐Z scintillators.4
2.A.2. Optical photon escape
A fraction of the optical photons generated by an absorbed x‐ray escape the scintillator with probability , which depends on the depth of interaction z. For consistency with previous work, we refer to this fraction as the escape efficiency12, 34 . Others have referred to this quantity as the optical escape fraction15, 29 and optical collection efficiency.31 Escape efficiency is expected to increase with depth, e.g., interactions closer to optical collection will have higher escape efficiency than those occurring further away. The number of photons that escape the scintillator per x‐ray interaction, , is the product of the conversion gain and the depth‐dependent escape efficiency:
(6) |
As illustrated in Fig. 2(a), light escape from different depths of the scintillator is weighted by non‐uniform x‐ray absorption, which results in a probability density function of represented generically in Fig. 2(b). The following analysis was used to derive an expression for this function, denoted as .
2.A.3. Derivation of
The PHS experiment consists of a series of N T observations of the number n escaping photons per absorption event. We denote the probability by the cumulative distribution function which is represented generically in Fig. 2(c). The probability density function is related to this by:
(7) |
The product of the total number of observed events and the probability is equal to the number of absorption events in the scintillator from depth 0 to z:
(8) |
Applying the fundamental theorem of calculus to Eq. (7), we may write:
(9) |
Noting that is a function of z, we apply the chain rule to derive:
(10) |
Rearranging Eq. (10) and substituting into Eq. (9) yields:
(11) |
The numerator and denominator on the right side of Eq. (11) are determined by recalling Eqs. (8) and (6), respectively:
(12) |
Note that φ describes photons−1, which is unitless. Eq. (12) can be written in terms of in cases where the relation (6) between and z is invertible, using:
(13) |
2.A.4. Calculating PHS from : linear
Equation (12) can be used to calculate PHS when a particular form of is assumed. As an example, we consider the case where is linear with depth:
(14) |
although cases of non‐linear may also be treated using the same approach. Inverting Eq. (14) yields:
(15) |
Differentiating with respect to z results in:
(16) |
Substituting Eqs. (15) and (16) into Eq. (12) yields the final expression:
(17) |
This analytical expression can be used to calculate the noise‐free (i.e., no quantum or electronic noise) PHS of a scintillator as a function of , for the case of linear . The minimum number of photons escaping the scintillator is , and the maximum number ; therefore, is given by Eq. (17) over the range , and is zero elsewhere. In this example, Eq. (17) shows that and are sufficient to calculate φ if , μ and L are known.
Calculations of φ may be compared to measured PHS by noting that experimental observations sort φ into bins of width ΔN, which is a measurement system parameter expressed as a number of photons. The bounds of the j‐th bin of the PHS are [jΔN, (j+1)ΔN] , where j is a non‐negative integer and the event count in the bin is:
(18) |
2.B. Modeling measured PHS with noise: linear system analysis
Cascaded linear system analysis was used to determine how is degraded by stochastic effects in the PHS measurement chain shown in Fig. 3. The stages of x‐ray conversion gain, light escape, detection by a PMT and dynode amplification were modeled as a cascade of stochastic amplification processes.
The signal and noise propagation through each stage was determined using35, 36:
(19) |
(20) |
where and represent the mean numbers of input and output quanta, and denote the mean and variance of the stochastic gain, and and are the variances in N in and N out , respectively. Detailed expressions for the mean and variance in the number of quanta at each stage of Fig. 3 are given in Appendix A. Briefly, the total variance in measurements of , which includes noise due to the measurement apparatus, was derived to be:
(21) |
where denotes the average gain of each PMT dynode and represents the mean quantum efficiency of the PMT to scintillation light, defined by:
(22) |
where I(λ) represents the emission spectrum of the scintillator and η(λ) is the spectral response of the photocathode.
Figure 4 illustrates conceptually how the measurement chain degrades the PHS. Shown in Fig. 4(a) is a representation of a noise‐free PHS, which is determined by calculating and sorting into bins using Eqs. (17) and (1), respectively. In Fig. 4(b), the total stochastic noise in x‐ray conversion gain, light escape, PMT detection and dynode amplification is introduced to degrade the PHS. This is calculated using Gaussian distributions centered at each bin, with variances determined by Eq. (4). The contributions of these distributions to each bin are summed, as shown in Fig. 4(c), to model the measured PHS.
2.C. PHS measurement and calibration
Commercial CsI scintillator samples were obtained from Hamamatsu Photonics K.K, with design types FOS HR, FOS HL and GPXS. FOS designs have CsI deposited on a fiber‐optic faceplate (FOP) and are coated with an optically absorptive or reflective backing in HR or HL type, respectively.37 The FOP is always in contact with the optical sensor, as shown in Fig. 5(a). In the GPXS structure, CsI is deposited on a highly reflective aluminum substrate, as shown in Fig. 5(b), with a transparent protective coating on the surface to be coupled to the optical sensor. The nominal thickness of the FOS and GPXS samples ranged from 150 to 1000 μm.
PHS were measured at energies below the K‐edges of Cs and I using the apparatus sketched in Fig. 6, with acquisition parameters reported in Table 1. Monoenergetic x‐rays at 25.2 keV and 32.1 keV were generated via K α fluorescence of tin (Sn) and barium (Ba) metallic targets, with K β emissions of the targets filtered by silver (Ag) and I, respectively. The monochromatic x‐rays were collimated using a lead pinhole to expose the samples, which were coupled to a PMT (R1450, Hamamatsu Photonics K.K.) for light collection. The PMT output was fed to a charge amplifier (A1424, CAEN S.p.A.) with a decay time constant of 50 μs. The output of the charge amplifier was fed to a fast digitizer (V1724, CAEN S.p.A.) for pulse height analysis using a digital trapezoid shaper with 10 μs rise time and 16 μs peaking time.38 Shaper rise time was made substantially longer than the temporal decay time constants of CsI39 (primary decays τ 1 = 0.68 μs #bib63.7% and τ 2 = 3.34 μs #bib36.1%) to maximize the collection of light photons generated from each x‐ray interaction. X‐ray source intensity was limited (~1000 counts/s) in the experiments to reduce the probability of pulse pileup resulting from the long shaping time. The total acquisition time for each spectrum was less than 5 minutes.
Table 1.
PMT Gain | 2–4 × 106 |
Charge amplifier sensitivity | 3–10 mV/pC |
Charge amplifier decay | 50 μs |
Trapezoid shaping (rise time) | 10 μs |
Peaking time | 16 μs |
Abscissae of measured PHS were calibrated to relate measurement channel number to the number of optical photons detected per x‐ray interaction (i.e., ) using methodology described in detail by Watt et al.10 The calibration apparatus is shown in Fig. 7, in which a green LED source was pulsed in 0.75–2 μs intervals toward the PMT to produce PHS of the detected light bursts. Calibration measurements were acquired with parameters reported in Table 1.
The duration of LED pulses was varied to control the number of photons reaching the PMT photocathode. Photon detection is Poisson distributed, therefore, the variance in PHS channel number is related to the peak PHS channel number k p by:
(23) |
The constant a represents the system's constant of proportionality between channel number and generated photoelectrons, which was determined by the slope of the best‐fit line between and k p for the ensemble of LED PHS.
Dynode amplification adds noise to the PMT input signal, due to the statistical nature of each dynode's gain δ.4, 10, 12 We derive expressions that describe how noise propagates in a multi‐dynode PMT in Appendix B. We show that, given certain reasonable assumptions, the noise contribution from dynode amplification reduces to a multiplicative noise factor NF given by:
(24) |
This factor scales the variance of the PMT input, therefore the measured PHS variance was divided by NF prior to the calculation of a.
Once determined, a was multiplied by the effective quantum efficiency of the PMT to CsI scintillation light. A final correction factor accounting for light transmission through the PMT window40 was applied to the product of a and to determine the relationship between channel number (i.e., measured PHS abscissae) and the number of optical photons exiting the CsI scintillators.
2.D. Deriving depth‐dependent light escape from measured PHS
PHS of various CsI scintillators were calculated using the method outlined in Section 2.B.. and illustrated by Fig. 4. The mean conversion gain in CsI was assumed to be 58.8 keV−1 (W = 17 eV) based on literature,4, 40, 41, 42, 43, 44 and mean dynode gain was determined from the k‐th root of PMT gain (k = 10 in R1450 PMT). Previous investigations have shown that light escape in CsI varies approximately linearly with depth.12, 29, 31 Following these results, was modeled using
(25) |
making and the only variable parameters in PHS model calculations.
Various combinations of and were used to calculate PHS, and the process was repeated until modeled and measured PHS reached agreement. The quality of agreement was quantified by the objective function
(26) |
The values of and that minimized F were used in Eq. (8) to calculate the light escape efficiency as a function of depth in each sample.
2.E. Measured estimate vs. true optical Swank factor I OPD
An estimate of the Swank factor A s , which is equivalent to I OPD in our experiments, can be calculated directly from the measured PHS using:2, 3
(27) |
where m i represents the i‐th moment of the spectrum; however, this method underestimates the true optical Swank factor of the scintillator, as the PHS are degraded by noise introduced by the measurement apparatus.5, 12 A more accurate estimate of I OPD was obtained for each sample using the following method: First an estimate of was obtained using the iterative methods outlined in the previous section. Then the corresponding light escape parameters were used in Eqs. (17) and (4) to calculate the scintillator's PHS under ideal measurement conditions, i.e., = 100% and . Finally, the moments of the resulting PHS were used to calculate the true I OPD without noise contributions from the measurement apparatus.
3. Results
3.A. Modeling PHS with noise: linear system analysis
Figure 8 shows calculated PHS of a scintillator using the expressions (17) and (4) and methods described in Sections 2.A.. and 2.B.. B. Shown in Fig. 8(a) is the noise‐free PHS for a 600 μm CsI scintillator with 32 keV x‐ray energy and an example set of escape efficiency parameters. The x‐axis position and width of φ are determined by the values and range of , respectively, while the shape of the distribution depends on x‐ray attenuation and scintillator thickness.
Figure 8(b) shows how φ is degraded when stochastic effects in photon creation and light escape are introduced. The plotted spectrum represents the PHS measured with an idealized system, where the detection efficiency of the PMT is unity and dynode gain approaches infinity. Following Eq. (4), the total variance in measurements of is:
(28) |
The calculated PHS shown in Fig. 8(c) represents a realistic example of the measured PHS, which includes the additional stochastic effects of finite PMT detection efficiency and noise associated with finite dynode gain. The total variance in measurements of is given by Eq. (4). Although is reduced from unity in Fig. 8(c), the PHS abscissa represents the number of photons detected in a calibrated measurement, which does not change; however, the low efficiency of PMT detection broadens the measured PHS substantially according to the inverse relationship between and in Eq. (4).
3.B. Light output calibration
The measured data for PHS calibration are shown in Fig. 9. Figure 9(a) shows the measured PHS with 1–2 μs LED pulses, which demonstrate a linear relationship between pulse length (i.e., number of incident photons) and peak channel number k p , with a small offset due to the finite response time of the LED. Figure 9(b) shows the variation in channel variance as a function of k p for the various LED pulse lengths. The slope of the best‐fit line between k p and , also shown in Fig. 9(b), was the constant of proportionality a = 29.6 channels per photoelectron. Figure 9(c) shows the emission spectrum of CsI and the quantum efficiency of our PMT, which were used to calculate = 9.3% following Eq. (5). The product of a and was corrected for light transmission (92%) through the PMT window to relate 3.06 measurement channels to each optical photon emitted from CsI.
3.C. Measured and modeled PHS of CsI scintillators
Measured PHS of the 600 μm thick FOS (HR and HL) and GPXS scintillators are shown in Fig. 10. Their comparison shows that light yield per x‐ray interaction varied by almost an order of magnitude between different scintillator optical designs. The FOS HR sample emitted the fewest number of photons per interaction, due to the optically absorptive backing and light attenuation by the FOP.37 The reflective backing used in the FOS HL sample improved this yield by a factor of 3, by enabling light generated deep within the scintillator to be reflected back toward the optical detector. The GPXS design yielded approximately three times more light than FOS HL and nine times more than FOS HR, due to both its highly reflective backing and the absence of an FOP.
Figure 10 also shows a comparison between measured PHS and the best‐fit PHS model results after iterative fitting. Modeled PHS showed reasonable agreement with all measurements. All estimates of and from measurements at 25 and 32 keV were within 7% root‐mean‐square (RMS) error.
Figure 11 shows the estimates of in the 600 μm samples determined through PHS fitting. Light escape efficiency ranged from 6.9 to 20.7% #bib20.1 to 31.1% and 57.5 to 81.4% from the back (z = 0) to the front (z = L) of the FOS HR, FOS HL, and GPXS samples, respectively. The low light, right‐skewed PHS of FOS HR is due to the low values of with relatively large variations over x‐ray interaction depth. Similarly, the brighter and more symmetric PHS observed in FOS HL and GPXS corresponds to higher values of with smaller depth‐dependent variations.
Table 2 reports the estimated light escape efficiency minima and maxima for all of the investigated samples. These data are also scaled by the inherent W of CsI to the unit of keV−1, and provide upper and lower bounds to each scintillator's effective conversion gain, i.e., the number of photons emitted per unit energy of excitation.
Table 2.
Back | Front | ||||
---|---|---|---|---|---|
|
keV−1 |
|
keV−1 | ||
FOS 150 HR | 0.073 | 4.3 | 0.228 | 13.4 | |
FOS 600 HR | 0.069 | 4.1 | 0.207 | 12.2 | |
FOS 200 HL | 0.222 | 13.1 | 0.313 | 18.4 | |
FOS 600 HL | 0.201 | 11.8 | 0.311 | 18.3 | |
FOS 1000 HL | 0.176 | 10.4 | 0.293 | 17.2 | |
GPXS 200 | 0.600 | 35.3 | 0.810 | 47.6 | |
GPXS 600 | 0.575 | 33.8 | 0.814 | 47.9 | |
GPXS 1000 | 0.501 | 29.5 | 0.884 | 52.0 |
The data reported in Fig. 11 and Table 2 can be used to compare the evaluated samples as a function of thickness or optical design. For the same optical design, thicker samples are shown to yield less light, with greater variation in compared to thinner samples. The amount of light escaping from interactions furthest from the optical collection is reduced by 5.5% #bib20.7%, and 16.5% from the thinnest to the thickest samples of FOS HR, FOS HL, and GPXS, respectively. In comparison, scintillator optical design is shown to be the dominant factor in determining the values of , which varied up to an order of magnitude between samples.
3.D. Optical Swank factor
Table 3 reports measured estimates of the optical Swank factor which were computed directly from measured PHS, as well as the estimated true optical Swank factor (I OPD ) determined from PHS calculated with measurement noise removed. Optical design (i.e., backing and FOP) is shown to be the dominant factor affecting I OPD , which is estimated to be greater than 0.9 in all samples. The FOS HL and GPXS scintillators, which exhibit smaller variations in , are shown to have I OPD approaching unity. This result agrees well with estimates that others have provided through experiments5, 11, 12 and Monte Carlo simulations15, 32, 45 of CsI. The FOS HR scintillators, despite having poorer light escape properties, maintain high I OPD of approximately 0.9.
Table 3.
Measured estimate of the optical Swank factor (measurement noise included) | True optical Swank factor I OPD (measurement effects removed) | |||
---|---|---|---|---|
25 keV | 32 keV | 25 keV | 32 keV | |
FOS 150 HR | 0.828 | 0.844 | 0.900 | 0.903 |
FOS 600 HR | 0.847 | 0.821 | 0.911 | 0.906 |
FOS 200 HL | 0.958 | 0.948 | 0.988 | 0.988 |
FOS 600 HL | 0.944 | 0.936 | 0.985 | 0.983 |
FOS 1000 HL | 0.940 | 0.936 | 0.985 | 0.979 |
GPXS 200 | 0.968 | 0.964 | 0.998 | 0.995 |
GPXS 600 | 0.966 | 0.954 | 0.999 | 0.998 |
GPXS 1000 | 0.955 | 0.946 | 0.984 | 0.976 |
4. Discussion
4.A. Light transport in CsI scintillators: optical design
Results shown in Fig. 11 and Table 2 indicated that detector optical design parameters, i.e., backing and FOP, are the dominant factors determining the light escape properties of the CsI samples. Samples featuring a reflective backing and no FOP yielded appreciably more light than samples with an absorptive backing, FOP, or both. While it is desirable to maximize light yield (i.e., signal) to overcome electronic noise limitations in FPI readout, other imaging performance parameters need to be taken into consideration. For example, Zhao et al.11 showed that the absorptive optical backing of the FOS HR design reduces overall light yield to enhance spatial resolution. Alternatively, an FOP not only attenuates visible light, but also attenuates undetected x‐rays to minimize radiation absorption or damage in underlying electronics, e.g., CMOS sensor.37
4.B. Light transport in CsI scintillators: thickness
Results in Tables 2 and 3 showed how thickness variations affect light escape and the Swank factor of CsI scintillators. In each of the sample designs, large differences in thickness (e.g., from 200 to 1000 μm) only resulted in minor changes in . This indicates that light photons escape efficiently from CsI layers with thickness up to 1 mm. The small variations in exhibited by CsI scintillators result in optical Swank factors that approach unity.5, 11, 12, 13, 14, 32 In contrast, isotropic powder phosphors such as Gd2O2S:Tb, are known to exhibit larger variations in and lower Swank factors.6, 7, 8, 15 This inherent advantage of CsI can be used to improve the DQE(0) of indirect FPIs, particularly in high‐energy applications such as cone‐beam CT. Using a thicker CsI layer (up to 1 mm) for these applications can enhance DQE(0) by improving x‐ray absorption while maintaining a high Swank factor.13
4.C. Intrinsic x‐ray conversion gain: W of CsI
All parameters used in PHS modeling were either well‐known or directly measured, with the exception of the energy required to create one optical photon, W. Extensive efforts4, 40, 41, 42, 43, 44 have been devoted to estimate the intrinsic conversion gain in CsI. The gain value used in the present work (58.8 keV−1, or W = 17 eV) is a conservative estimate based on previous investigations, which reported W values ranging from 13.3 to 22 eV with an average of approximately 17 eV. The choice of W directly affects the absolute values of light escape efficiency deduced from the PHS. This should be considered when comparing scintillators without well‐defined W values. In this investigation, W was assumed to be constant between samples, which were prepared by the same manufacturer.
4.D. Precision in estimating and and
The quality of agreement between measured and modeled PHS is most sensitive to the value of used in calculations, and is less sensitive to . Consequently, measured PHS estimate more precisely than . Figure 12 shows an example of how the agreement between measured PHS and modeled PHS changes when one fitting parameter (i.e., or ) is varied while the other is fixed. Figure 12(a) shows PHS calculated when is fixed and is varied by 10% of its estimated value in the FOS 600 HL scintillator. Different values of shift the position of modeled PHS peaks, which strongly affects the agreement with measurement. Figure 12(b) shows the PHS calculated with fixed and 10% variation in . Differences in elongate the upper tails of modeled PHS, but the change is much less in comparison with the effect of .
Modeled PHS change unequally with respect to and due to our measurement geometry (Fig. 1) and exponential x‐ray attenuation in the scintillator. At low energies, e.g., 25 keV, most x‐ray interactions occur near the entrance plane (z = 0). These events, which have lower escape efficiency of light, contribute most to the PHS. Conversely, only a small portion of x‐ray interactions occur near the exit (z = L), and their contribution to the PHS is much less. This asymmetry is most evident in thick or high‐attenuating scintillators, where the difference in x‐ray absorption at z = 0 and z = L is the largest.
Escape efficiency estimates may also be affected by nonlinearity in , which is not considered in this work. Previous estimates of light escape efficiency in CsI using measured PHS12 have shown that it varies approximately linearly with depth. Monte Carlo simulations by Badano et al.29 and Freed et al.31 have also shown that in CsI can be approximated by a linear function. PHS interpretation with non‐linear is possible with the presented model using the approach described in Section 2.A.. This may be explored in future work; however, its impact on imaging performance parameters, e.g., Swank factor, is expected to be small.
5. Conclusions
In this paper, we developed a new approach to deriving depth‐dependent light escape efficiency and more accurately estimating the true optical Swank factor of scintillators from measured PHS, using a model that removes the effects of added noise in the measurement chain. This method was used to investigate CsI scintillators with various thickness and optical design used in indirect FPIs. Our results showed that light escape varies almost an order of magnitude between scintillators with different optical designs, and varies considerably less with changes in CsI thickness. After removing the effect of measurement noise, the optical Swank factor of the scintillators with the lowest light escape properties still exceeds 0.9, and approaches unity in scintillators with reflective backings. Estimates of the depth‐dependent light escape efficiency and Swank factor provided by the present work can help validate Monte Carlo simulations of structured CsI and can be used to optimize scintillator design for different x‐ray imaging applications.
Conflicts of interest
The authors have no conflicts of interest to disclose.
Acknowledgments
Financial support from the NIH (R01 EB002655) was gratefully acknowledged. The authors thank Hamamatsu Photonics, K.K. for providing samples and equipment, Winston G Ji for early work on the theory, and thank Drs. Amir Goldan and Yue‐Houng Hu, David Scaduto, James Scheuermann, and Jann Stavro for helpful discussions.
Appendix A. Variance in the number of optical photons detected per X‐ray interaction
A.1.
Here, we propagate the mean and variance in the number of quanta generated by an x‐ray (absorbed at a given depth z) through the stochastic amplification stages of PHS shown in Fig. 3. This analysis determines how each stage contributes to the total variance in measurements of given by Eq. (4).
A.1.1. Scintillation
The absorbed x‐ray is converted to N ph optical photons through conversion with mean gain . The variance in N ph is assumed to be proportional to by a constant γ 2:
(A1) |
(A2) |
A.1.2. Optical photon escape
Photon escape from the scintillator is modeled as a binary selection process,46 which is a special case of a stochastic amplification stage, wherein the gain in each quantum is either 1 or 0 for successful or unsuccessful escape, respectively. The mean gain of this stage is the escape efficiency , which corresponds to the interaction depth under consideration, and its variance is given by . The mean number of escaped photons and the variance in N phe are:
(A3) |
(A4) |
(A5) |
(A6) |
A.1.3. Optical photon detection
Optical photon interaction with the PMT photocathode is also modeled as a binary selection process, with average gain equal to the effective quantum efficiency defined in Eq. (5). The mean number of generated photoelectrons and variance in N e is:
(A7) |
(A8) |
(A9) |
(A10) |
A.1.4. Dynode amplification
The amplification of photoelectrons over k dynodes of a PMT is modeled as a serial cascade of k stochastic gain stages, each with mean gain and gain variance .4, 47, 48 The entire dynode chain has average gain , with variance . The mean number of output photoelectrons and the variance in N em are given by:
(A11) |
(A12) |
In Appendix B, we derive expressions for variance transfer in a multi‐dynode PMT to determine the form of . Briefly, is determined by two assumptions made in this work: (a) We consider the number of photons generated in the scintillator to be Poisson distributed, i.e., γ 2 = 1. This is an approximate but reasonable assumption in the study of various scintillators, including CsI.49 (b) We also consider the variance in each dynode's gain to be Poisson distributed, i.e., , which is a common approximation made in PMT literature.4, 47 Given these assumptions, reduces to
(A13) |
and following Eq. (B19) derived in Appendix B:
(A14) |
A.1.5. Pulse height analysis
In measured (and calibrated) PHS, is scaled by and to represent the number of optical photons escaping the scintillator :
(A15) |
However, the total variance in this quantity does not reduce to but rather to:
(A16) |
(A17) |
In cases where the total dynode gain is much greater than 1, which is true in our experiments (Table 1):
(A18) |
which is the relation (4) used in all calculations. In addition, Eq. (A18) shows that the multiplicative noise factor NF (24) described in Section 2.C. arises in the case that PMT input variance and dynode gain variance are Poisson distributed.
Appendix B. Variance transfer through a multi‐dynode photomultiplier tube
B.1.
In this Appendix, we derive expressions for the mean and variance in the number of photoelectrons at each dynode of a photomultiplier tube. A PMT with k dynodes is modeled as a serial cascade of k stochastic gain stages, with mean and variance transfer determined by Eqs. (2) and (3). We first derive a general expression for variance transfer over k dynodes, making no assumptions of the PMT input variance or dynode gain variance. We then derive an expression for the case where both the PMT input variance and dynode gain variance are Poisson distributed, which are the assumptions applied in this work.
Consider an average of photoelectrons generated at a PMT photocathode with variance . When amplified by a dynode with mean gain and gain variance , the mean and variance in the number of photoelectrons after amplification are:
(B1) |
(B2) |
When these photoelectrons are amplified by a second identical dynode, the mean and variance in the resulting number of photoelectrons are:
(B3) |
(B4) |
Repeating this process for a third dynode gives:
(B5) |
(B6) |
For k iterations of this process, i.e., amplification over k dynodes, the pattern emerges that:
(B7) |
(B8) |
In the case that amplification at each dynode is Poisson distributed, and:
(B9) |
(B10) |
(B11) |
(B12) |
Further, if the number of photoelectrons at the PMT input is Poisson distributed, and:
(B13) |
(B14) |
(B15) |
(B16) |
Factoring out from Eq. (B16) gives:
(B17) |
Noting the geometric series:
(B18) |
we simplify Eq. (B17) to:
(B19) |
Equation (B19) shows that for Poisson distributed input variance and dynode gain, the variance in the output of a multi‐dynode PMT is directly proportional to the variance in its input, and thus can be described by a multiplicative factor.
References
- 1. Rowlands JA, Yorkston J. Flat panel detectors for digital radiography. In: Beutel J, Kundel HL, Van Metter RL, eds. Handbook of Medical Imaging. Vol. 1. Bellingham, Washington DC: SPIE Publications; 2000:223–328. [Google Scholar]
- 2. Swank RK. Measurement of absorption and noise in an x‐ray image intensifier. J Appl Phys. 1974;45:3673–3678. [Google Scholar]
- 3. Swank RK. Absorption and noise in x‐ray phosphors. J Appl Phys. 1973;44:4199–4203. [Google Scholar]
- 4. Knoll GF. Radiation Detection and Measurement, 3rd edn. Hoboken, NJ: John Wiley & Sons; 2000. [Google Scholar]
- 5. Rowlands JA, Taylor KW. Absorption and noise in cesium iodide x‐ray image intensifiers. Med Phys. 1983;10:786–795. [DOI] [PubMed] [Google Scholar]
- 6. Trauernicht DP, Van Metter R. The Measurement Of Conversion Noise In X‐Ray Intensifying Screens. Proc SPIE. 1988;0914:100–116. [Google Scholar]
- 7. Ginzburg A, Dick CE. Image information transfer properties of x‐ray intensifying screens in the energy range from 17 to 320 keV. Med Phys. 1993;20:1013–1021. [DOI] [PubMed] [Google Scholar]
- 8. Drangova M, Rowlands JA. Optical factors affecting the detective quantum efficiency of radiographic screens. Med Phys. 1986;13:150–157. [DOI] [PubMed] [Google Scholar]
- 9. Dick CE, Motz JW. Image information transfer properties of x‐ray fluorescent screens. Med Phys. 1981;8:337–346. [DOI] [PubMed] [Google Scholar]
- 10. Watt KN, Yan K, DeCrescenzo G, Rowlands JA. The physics of computed radiography: measurements of pulse height spectra of photostimulable phosphor screens using prompt luminescence. Med Phys. 2005;32:3589–3598. [DOI] [PubMed] [Google Scholar]
- 11. Zhao W, Ristic G, Rowlands JA. X‐ray imaging performance of structured cesium iodide scintillators. Med Phys. 2004;31:2594. [DOI] [PubMed] [Google Scholar]
- 12. Lubinsky AR, Zhao W, Ristic G, Rowlands JA. Screen optics effects on detective quantum efficiency in digital radiography: zero‐frequency effects. Med Phys. 2006;33:1499. [DOI] [PubMed] [Google Scholar]
- 13. Howansky A, Peng B, Suzuki K, Yamashita M, Lubinsky AR, Zhao W. Investigation of the screen optics of thick CsI(Tl) detectors. Proc SPIE. 2015;9412:94120G. [Google Scholar]
- 14. Badano A, Kyprianou IS, Jennings RJ, Sempau J. Anisotropic imaging performance in breast tomosynthesis. Med Phys. 2007;34:4076. [DOI] [PubMed] [Google Scholar]
- 15. Badano A, Kyprianou IS, Sempau J. Anisotropic imaging performance in indirect x‐ray imaging detectors. Med Phys. 2006;33:2698–2713. [DOI] [PubMed] [Google Scholar]
- 16. Nillius P, Klamra W, Sibczynski P, Sharma D, Danielsson M, Badano A. Light output measurements and computational models of microcolumnar CsI scintillators for x‐ray imaging. Med Phys. 2015;42:600–605. [DOI] [PubMed] [Google Scholar]
- 17. Blevis IM, Hunt DC, Rowlands JA. X‐ray imaging using amorphous selenium: determination of Swank factor by pulse height spectroscopy. Med Phys. 1998;25:638–641. [DOI] [PubMed] [Google Scholar]
- 18. Lui BJM, Hunt DC, Reznik A, Tanioka K, Rowlands JA. X‐ray imaging with amorphous selenium: pulse height measurements of avalanche gain fluctuations. Med Phys. 2006;33:3183–3192. [DOI] [PubMed] [Google Scholar]
- 19. Rowlands JA, DeCrescenzo G, Araj N. X‐ray imaging using amorphous selenium: determination of x‐ray sensitivity by pulse height spectroscopy. Med Phys. 1992;19:1065–1069. [DOI] [PubMed] [Google Scholar]
- 20. Fredenberg E, Lundqvist M, Cederström B, Åslund M, Danielsson M. Energy resolution of a photon‐counting silicon strip detector. Nucl Instrum Methods Phys Res, Sect A. 2010;613:156–162. [Google Scholar]
- 21. Stavro J, Goldan AH, Zhao W. SWAD: inherent photon counting performance of amorphous selenium multi‐well avalanche detector. Proc SPIE. 2016;9783:97833Q‐1–97833Q‐8. [Google Scholar]
- 22. Iwanczyk JS, Nygård E, Meirav O, et al. Photon counting energy dispersive detector arrays for X‐ray imaging. IEEE Trans Nucl Sci. 2009;56:535–542. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Xu C, Danielsson M, Bornefalk H. Evaluation of energy loss and charge sharing in cadmium telluride detectors for photon‐counting computed tomography. IEEE Trans Nucl Sci. 2011;58:614–625. [Google Scholar]
- 24. Tanguay J, Kim HK, Cunningham IA. The role of x‐ray Swank factor in energy‐resolving photon‐counting imaging. Med Phys. 2010;37:6205. [DOI] [PubMed] [Google Scholar]
- 25. Hajdok G, Yao J, Battista JJ, Cunningham IA. Signal and noise transfer properties of photoelectric interactions in diagnostic x‐ray imaging detectors. Med Phys. 2006;33:3601–3620. [DOI] [PubMed] [Google Scholar]
- 26. Yun S, Kim HK, Youn H, Tanguay J, Cunningham IA. Analytic model of energy‐absorption response functions in compound x‐ray detector materials. IEEE Trans Med Imaging. 2013;32:1819–1828. [DOI] [PubMed] [Google Scholar]
- 27. XCOM . Photon cross section database (version 1.5). National Institute of Standards and Technology; 2010.
- 28. Nagarkar VV, Gupta TK, Miller SR, Klugerman Y, Squillante MR, Entine G. Structured CsI(Tl) scintillators for X‐ray imaging applications. IEEE Trans Nucl Sci. 1998;45:492–496. [Google Scholar]
- 29. Badano A, Gagne RM, Gallas BD, Jennings RJ, Boswell JS, Myers KJ. Lubberts effect in columnar phosphors. Med Phys. 2004;31:3122. [DOI] [PubMed] [Google Scholar]
- 30. Freed M, Miller S, Tang K, Badano A. Experimental validation of Monte Carlo (MANTIS) simulated x‐ray response of columnar CsI scintillator screens. Med Phys. 2009;36:4944. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Freed M, Park S, Badano A. A fast, angle‐dependent, analytical model of CsI detector response for optimization of 3D x‐ray breast imaging systems. Med Phys. 2010;37:2593–2605. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Sharma D, Badano A. Validation of columnar CsI x‐ray detector responses obtained with hybridMANTIS, a CPU‐GPU Monte Carlo code for coupled x‐ray, electron, and optical transport. Med Phys. 2013;40:031907. [DOI] [PubMed] [Google Scholar]
- 33. Lubberts G. Random noise produced by x‐ray fluorescent screens. J Opt Soc Am. 1968;58:1475–1483. [Google Scholar]
- 34. Lubinsky AR, Zhao W, Suzuki K. Screen optics effects on DQE in digital radiography: spatial frequency effects. Proc SPIE. 2008;6913:69134F. [Google Scholar]
- 35. Zweig HJ. Detective quantum efficiency of photodetectors with some amplification mechanisms. J Opt Soc Am. 1965;55:525–528. [Google Scholar]
- 36. Rabbani M, Shaw R, Van Metter R. Detective quantum efficiency of imaging systems with amplifying and scattering mechanisms. J Opt Soc Am A. 1987;4:895–901. [DOI] [PubMed] [Google Scholar]
- 37. FOS (Fiber Optic Plate with Scintillator) for Digital X‐ray Imaging. Hamamatsu: Hamamatsu; 1996.
- 38. Jordanov VT, Knoll GF. Digital synthesis of pulse shapes in real time for high resolution radiation spectroscopy. Nucl Instrum Methods Phys Res, Sect A. 1994;345:337–345. [Google Scholar]
- 39. Valentine JD, Wehe DK, Knoll GF, Moss CE. Temperature dependence of CsI(T1) absolute scintillation yield. IEEE Trans Nucl Sci. 1993;40.4:1267–1274. [Google Scholar]
- 40. Holl I, Lorenz E, Mageras G. A measurement of the light yield of common inorganic scintillators. IEEE Trans Nucl Sci. 1988;35:105–109. [Google Scholar]
- 41. Schotanus P, Kamermans R, Dorenbos P. Scintillation characteristics of pure and Tl‐doped CsI crystals. IEEE Trans Nucl Sci. 1990;37:177–182. [Google Scholar]
- 42. Carel WEvE. Inorganic scintillators in medical imaging. Phys Med Biol. 2002;47:R85. [DOI] [PubMed] [Google Scholar]
- 43. Onyshchenko GM, Nagornaya LL, Bondar VG, et al. Comparative light yield measurements of oxide and alkali halide scintillators. Nucl Instrum Methods Phys Res Sect A. 2005;537:394–396. [Google Scholar]
- 44. Shinichi S, Hiroko T, Kiwamu S, Mitsuhiro M, Eido S. Average energies required per scintillation photon and energy resolutions in NaI(Tl) and CsI(Tl) crystals for gamma rays. Jpn J Appl Phys. 2006;45:6420. [Google Scholar]
- 45. Hajdok G, Battista JJ, Cunningham IA. Fundamental x‐ray interaction limits in diagnostic imaging detectors: frequency‐dependent Swank noise. Med Phys. 2008;35:3194. [DOI] [PubMed] [Google Scholar]
- 46. Cunningham IA, Westmore MS, Fenster A. A spatial‐frequency dependent quantum accounting diagram and detective quantum efficiency model of signal and noise propagation in cascaded imaging systems. Med Phys. 1994;21:417–427. [DOI] [PubMed] [Google Scholar]
- 47. Engstrom RW. RCA Photomultiplier Handbook. Lancaster: RCA Corporation; 1980. [Google Scholar]
- 48. Photomultiplier Handbook. Lancaster, PA: Burle Industries Inc. and references therein; 1989. [Google Scholar]
- 49. Moses WW. Current trends in scintillator detectors and materials. Nucl Instrum Methods Phys Res Sect A. 2002;487:123–128. [Google Scholar]