Abstract
Diffusion magnetic resonance imaging (MRI) is a standard imaging tool in clinical neurology, and is becoming increasingly important for neuroscience studies due to its ability to depict complex neuroanatomy (eg, white matter connectivity). Single‐shot echo‐planar imaging is currently the predominant formation method for diffusion MRI, but suffers from blurring, distortion, and low spatial resolution. A number of methods have been proposed to address these limitations and improve diffusion MRI acquisition. Here, the recent technical developments for image formation in diffusion MRI are reviewed. We discuss three areas of advance in diffusion MRI: improving image fidelity, accelerating acquisition, and increasing the signal‐to‐noise ratio.
Level of Evidence: 5
Technical Efficacy: Stage 1
J. MAGN. RESON. IMAGING 2017;46:646–662
Keywords: diffusion MRI, pulse sequence, image artifacts, accelerated imaging, high‐resolution imaging, navigation
Single‐shot echo planar imaging (SSH‐EPI) has been used as the standard image formation method for diffusion magnetic resonance imaging (MRI) on commercialized scanners for more than 20 years. This is mostly due to the fast acquisition speed of SSH‐EPI (100–200 msec per slice including diffusion preparation), which makes it fairly immune to subject motion and enables advanced diffusion protocols with a large number of diffusion directions and/or b‐values within reasonable scan times. However, SSH‐EPI is prone to several limitations, including image distortions due to B0 inhomogeneity at tissue/air interfaces and blurring, both of which place limitations on spatial resolution. High‐resolution diffusion MRI provides the ability to resolve fine‐scale structures, enabling detection of cortical anisotropy,1, 2 delineation of thin white matter tracts,3 and more accurate fiber tractography.4 Although parallel imaging has improved the data quality of SSH‐EPI, these problems still exist and become more pronounced at high field strength and high resolution.
Alternative acquisition schemes have been proposed to overcome the limitations of SSH‐EPI, including segmented‐EPI readout, non‐EPI trajectories, and reduced field of view (FOV). These methods have undergone rapid development in recent years, demonstrating significantly improved image quality compared to SSH‐EPI. Conventional 2D acquisition schemes suffer from long scan time and low signal‐to‐noise ratio (SNR) efficiency when acquiring high isotropic‐resolution diffusion MRI data with full brain coverage, which is increasingly needed in neuroscience studies. The recent development of simultaneous multislice techniques has dramatically changed this situation and diffusion MRI data can be acquired more rapidly and with higher SNR efficiency. Various 3D diffusion MRI acquisitions have also been developed, which have high SNR efficiency and can provide more accurate slice definitions than 2D acquisition. Several studies have reported high‐quality diffusion MRI data at ultrahigh field of 7T, which opens new possibilities for achieving higher spatial and angular resolution. Finally, developments for fast diffusion MRI using compressed sensing have been reported, which can further accelerate diffusion acquisition.
In this article we review recent developments in image formation methods for diffusion MRI and discuss how these are likely to be used in practice. We focus on three kinds of advances: improving image fidelity, accelerating acquisition, and increasing SNR. However, inevitably a method that impacts one of these metrics has consequences for the others, and we aim to describe these tradeoffs throughout the review. Some of the reviewed methods are fairly unique to diffusion imaging (eg, navigated correction of motion‐induced phase errors), while others have broader application (eg, simultaneous multislice imaging). For conciseness, we do not cover diffusion contrast mechanisms and use examples focusing on diffusion MRI of the brain.
Improving Image Fidelity
The two dominant image artifacts in SSH‐EPI are image blurring and distortion. The spatial resolution of SSH‐EPI is severely affected by the tissue decay (Fig. 1), which results in intense signal loss at the outermost edges of k‐space. Because outer k‐space corresponds to fine spatial detail (high spatial frequencies), this weighting introduces image blurring. Thus, a short readout window is important for reducing EPI blurring. Image distortion in SSH‐EPI mainly happens in regions with strong local magnetic field inhomogeneity (eg, tissue/air boundaries with fast susceptibility variation). This local field distortion can swamp the weak gradients used for phase encoding, resulting in misplaced signal that appears as image distortion (Fig. 2). The scale of distortion is determined by the speed of k‐space transversal along the phase‐encoding direction. Therefore, short echo spacing and undersampling, both of which enable faster traversal along the phase‐encoding direction, are desirable properties for improving SSH‐EPI. The application of parallel imaging5, 6 in SSH‐EPI has been very successful, although this method faces challenges from noise amplification, particularly for diffusion MRI. “Effective echo spacing” accounts for the reduced distortion in parallel imaging by dividing the acquired echo spacing by the acceleration factor, which gives the echo spacing that would have been required to achieve this level of distortion without acceleration (ie, any EPI scan with the same effective echo spacing will have the same distortion). Reduced FOV methods can reduce distortion without noise penalty, but is limited to small coverage.
Another source of distortion in diffusion MRI is the eddy currents induced within the conducting surfaces of the magnet due to fast gradient switching. Modern systems include gradient “pre‐emphasis” that can substantially reduce eddy currents, but these corrections struggle to compensate for strong diffusion‐encoding gradients with high slew‐rate. The twice‐refocused spin‐echo diffusion preparation7 can further reduce the effects of eddy currents, but at the cost of longer echo time. This is particularly problematic for high‐resolution scans at ultrahigh field, where this scheme requires ∼30 msec longer echo time (TE) than conventional spin‐echo diffusion preparation for 1 mm resolution scan, exacerbating the already problematic T 2 signal loss. Another approach is to correct the distortion in postprocessing.8 More recently, correction of distortion based on nonparametric modeling of diffusion signal with respect to diffusion‐encoding direction has demonstrated excellent results.9
Partial Fourier acquisition is often used to reduce the long echo time of SSH‐EPI, which is another challenge for diffusion MRI due to greater T 2 signal decay (lower SNR). This is particularly problematic for high‐resolution scans with longer echo time and smaller voxel size. Although partial Fourier reduces echo time, it increases the sensitivity to subject bulk motion during diffusion encoding,10 which induces echo shifting in k‐space.11 For strong motion, the echo is shifted toward the edge of k‐space (higher spatial frequency), violating the fundamental assumptions underlying partial Fourier and causing the reconstruction to fail. This leads to image intensity oscillations and signal loss.12 To alleviate this problem, adaptive partial Fourier reconstruction algorithms have been proposed.12, 13
Other common EPI artifacts that may appear in diffusion MRI images include Nyquist ghosting caused by hardware‐related odd–even echo misalignment and fat shifting along the phase‐encoding direction. Nyquist ghosting is typically corrected by measuring the k‐space shift between odd and even echoes using a reference scan and subsequently realigning the k‐space data. Reference‐less methods using image‐entropy as a selection metric have also demonstrated robust correction of Nyquist ghosting.14 Fat shifting occurs due to the difference of resonance frequencies between water and fat and the low bandwidth in the phase‐encoding direction. For example, because fat differs from water by 440 Hz at 3T, fat signal will be shifted by ∼1/3 FOV for an EPI acquisition with echo spacing of 0.8 msec = (1250 Hz)−1. To eliminate fat‐shifting artifacts, fat suppression is commonly implemented in EPI acquisition. Most fat suppression methods utilize special excitation schemes, including water‐only spectral spatial excitation, fat saturation (and spoiling), and inversion‐recovery preparation.15
A number of multishot acquisition techniques have also been proposed, for which subject motion must be carefully handled, typically using a k‐space “navigator.” Navigators provide an unaliased, low‐resolution image corresponding to a limited central k‐space region that is used to predict effects from subject motion. Navigators are usually acquired immediately before/after the imaging data or extracted from the imaging data directly (“self‐navigation”). Diffusion signals are intrinsically sensitive to subject motion because diffusion preparation gradients encode tiny (molecular) motions in signal phase. Even small subject motions (eg, cardiac pulsation, respiration) during diffusion preparation can lead to substantial spatially varying phase that is unrelated to diffusion. SSH‐EPI is immune to these phase errors because it captures the entire image in one shot, such that the phase of the signal can be discarded, while diffusive motion is reflected in the signal magnitude. By comparison, multishot image acquisitions must retain phase information in order to accurately combine across the different k‐space segments acquired in each shot. If motion‐induced phase is not corrected before combining multishot segments, images become severely corrupted (Fig. 3). To correct motion‐induced phase errors, it is common to acquire additional k‐space measurements that can be used as a low‐resolution navigator, using the phase of the navigator image to rectify the phase inconsistency between segments.
Readout‐Segmented EPI
Readout‐segmented EPI (rs‐EPI) uses a series of EPI acquisitions to cover k‐space in a mosaic pattern,16 as depicted in Fig. 1c. By acquiring concatenated k‐space segments along the readout direction, rs‐EPI can achieve much shorter echo spacing (eg, echo spacing could be shortened from about 0.8 msec in SSH‐EPI to about 0.3 msec in rs‐EPI with 7 readout segments for diffusion scan at 1.2 mm isotropic‐resolution17), and hence considerably reduce geometric distortion and blurring.18, 19, 20, 21 The effective echo spacing can be further reduced via the combination with parallel imaging.5, 6 The resulting TE of rs‐EPI is also shorter than SSH‐EPI (eg, for 2 mm2 isotropic‐resolution scan and b = 1000s/mm2 at 3T, TE could be reduced from about 87 msec in SSH‐EPI to about 73 msec in rs‐EPI,17 which is predicted to result in an ∼25% SNR improvement). The improvements offered by the reduced echo spacing (reduced distortion, blurring, and TE) should be particularly beneficial at ultrahigh field, where tissue T 2 and are very short and field inhomogeneity is worse.22
As a multishot diffusion acquisition method, rs‐EPI requires navigation to remove motion‐induced phase errors. rs‐EPI acquires a continuous k‐space segment, enabling fast and robust correction using the nonlinear phase correction method.19, 20 By comparison, in conventional segmented EPI (see “Segmented EPI” section, below), the image reconstructed from each segment suffers from aliasing, which is difficult to correct robustly using the nonlinear phase correction. Iterative algorithms23 can address this problem, but at the cost of long computational time.
The nonlinear phase correction only works if the motion‐induced phase errors can be accurately extracted from the navigator. In case of severe subject motion during diffusion preparation, the navigator fails to correct the phase errors when the center of k‐space shifts out of the navigator acquisition window, making the motion information obtained from the navigator inaccurate. These corrupted data could severely degrade the image quality, and they cannot be simply removed as in Propeller24 (see “Propeller” section, below) because there is no redundancy between readout segments. Instead, severely corrupted segments can be detected using the navigator and replaced by reacquired segments at the same k‐space location.20
As with other multishot methods, the primary challenge of rs‐EPI is the longer scan times required to form each image volume. Many of the approaches to reducing scan times discussed in the section on acceleration of diffusion MRI are compatible with rs‐EPI, and both partial Fourier25 and simultaneously multislice17 have been proposed to reduce scan times. rs‐EPI has also been demonstrated at 7T22 and in conjunction with 3D multislab acquisitions.26
Superior data quality using rs‐EPI has been demonstrated compared to SSH‐EPI, especially in regions with strong susceptibility variation, such as the temporal lobes and brainstem.19, 20 Several clinical studies have investigated the performance of rs‐EPI for diagnostics, demonstrating high data fidelity and improved conspicuity of pathology, for applications including: breast cancer,27, 28, 29 liver tumors,30 pelvic,31, 32 and renal33 diseases, and pediatric neuropathology.34
Segmented EPI
Segmented EPI samples k‐space over multiple EPI trajectories with broadly spaced lines that interleave to cover the full k‐space, as depicted in Fig 1b. In early implementations, a 2D navigator was used to correct the motion‐induced phase errors.35, 36
Recently, the multiplexed sensitivity encoding (MUSE) method37 has been proposed to correct the motion‐induced phase errors in segmented EPI without acquiring 2D navigators. MUSE reconstruction consists of three components: first, a phase navigator is calculated for each segment using parallel imaging to fill in the missing samples in central k‐space; second, each segment is phase‐error‐corrected using this navigator; third, all segments are combined to form the final image (in practice, the second and third parts are calculated simultaneously). Several refinements of MUSE were proposed to correct the rigid motion between different segments (Fig. 5)38, 39 and prospectively detect and reject severely motion‐corrupted interleaves.40 Extensions of MUSE with robust partial Fourier reconstruction, simultaneous multislice,13 and 3D multislab acquisition have also been proposed.41
A disadvantage of the MUSE method is the limitation on the number of interleaves, since this equates to the acceleration factor in the first reconstruction step, which is constrained by the receiver coil design. A more recent method formulated the reconstruction as a low‐rank matrix completion process without explicitly estimating the motion‐induced phase errors.42 These navigator‐free methods improve the efficiency of segmented EPI acquisition (navigator acquisition takes 30–40 msec for each excitation) and reduce the specific absorption rate (SAR) (by ∼30% due to the removal of the refocusing pulse for the navigator echo), which may be critical at ultrahigh field.
Propeller
One approach to reducing image distortion and blurring uses a class of sequences known as fast spin echo (FSE, sometimes called TSE or RARE),43 which uses a series of refocusing pulses to create a train of spin echoes. This method can reduce artifacts by acquiring every k‐space line at the center of a spin echo, thus avoiding the phase accumulation that leads to distortion and blurring in EPI. The most widely used implementation of FSE in diffusion MRI is PROPELLER (Periodically Rotated Overlapping ParallEL Lines with Enhanced Reconstruction), which acquires a strip (“blade”) covering the center of the k‐space in each train of refocusing pulses.24, 44 Multiple rotated blades are acquired to fully sample a circular k‐space region over multiple shots, as shown in Fig 4a. A key strength of PROPELLER is that it is self‐navigating: all blades cover the k‐space center, which can be used to estimate the motion‐induced phase errors for each shot.
Several key challenges for PROPELLER have been identified and addressed, including motion sensitivity (due to the conditions for formation of spin‐echo trains), RF deposition (due to the large number of refocusing pulses in the FSE readout), and imaging speed (due to the multishot acquisition). Signal formation requires stable signal phase over the course of the spin‐echo train,45 which is disrupted by motion, resulting in signal decay and oscillates. Alternating the phase of the refocusing pulses between x and y axes has been demonstrated to stabilize the signal.24, 46, 47 To accelerate the acquisition and reduce RF deposition, Turboprop48 collects multiple gradient echoes between refocusing pulse pairs. Turboprop is in effect a gradient‐ and spin‐echo (GRASE49) sequence, providing a tradeoff between image distortion and speed. Going further in this direction, PROPELLER‐EPI50 acquires each blade using a single EPI readout, combining the self‐navigation of PROPELLER and the rapid acquisition and low‐SAR properties of EPI. PROPELLER‐EPI has greater blurring and distortion than the original FSE technique, which can be mitigated with parallel imaging.51 Alternatively, short‐axis PROPELLER‐EPI52 places the EPI readout along the short axis of the blade (Fig. 6), leading to short echo spacing, and therefore reduced blurring artifacts.
Spiral
The most commonly considered alternative to Cartesian sampling techniques like EPI are spiral trajectories, in which k‐space is traced in a radiating pattern rather than a line‐by‐line scan. Spiral imaging has the merit of intrinsic motion compensation through gradient‐moment‐nulling53 and efficient use of gradient power.54 Single‐shot spiral imaging acquires diffusion MRI data with similar efficiency as SSH‐EPI,55 but suffers from different artifacts due to off‐resonance and decay, both of which result in image blurring.
Similar to segmented EPI, spiral acquisitions can be acquired in multiple shots using interleaved acquisitions to achieve high spatial resolution and reduced image blurring. As with all multishot diffusion imaging methods, motion‐induced phase errors need to be corrected before combining interleaved spiral acquisitions. Variable density spirals (VDS) can be self‐navigating by sampling central k‐space densely with each interleave (Fig. 4b).56 Alternatively, constant density spirals can be used with a similar strategy as MUSE to extract motion navigators using a parallel imaging reconstruction.57 Spirals are also pseudo‐incoherent with respect to undersampling artifacts, which makes it a preferable sampling method for compressed sensing reconstruction,58 as discussed below. Spirals have also been extended to 3D acquisitions using thin slabs.59 The major challenge for spiral acquisition is image blurring, which compromises the spatial resolution that can be achieved. Although deblurring methods can alleviate this problem to some extent, their performance still needs to be improved in the presence of strong susceptibility variations or severe B0 inhomogeneity (eg, ultrahigh field).
Reduced FOV Methods
Similar to parallel imaging,5, 6 reduced field of view (rFOV) techniques reduce distortion and blurring by skipping phase‐encoding lines. In rFOV, aliasing is avoided by excluding signal from outside a limited volume within a region of extended tissue. Three main strategies for rFOV acquisition and their application in diffusion MRI are reviewed here: inner volume imaging, outer volume suppression, and multidimensional RF excitation.
Inner volume imaging (IVI) uses orthogonal orientations for excitation and refocusing pulses such that only the overlapping regions of the excited volume and refocused volume create signal, enabling a reduced imaging FOV.60 In its original form, IVI was limited to single‐slice imaging because the refocusing pulse saturates parallel slices. A refinement of the IVI method places the refocusing pulse at a shallower angle to the excitation,61 which enables multiple slice acquisition, but requires gaps between slices. Alternatively, one can apply another refocusing pulse after the readout,62 returning the spins from the non‐imaged slices to the positive longitudinal axis, which enables contiguous interleaved multislice acquisition.
Outer volume suppression (OVS) suppresses signal from outside the imaging volume using spatially selective RF pulses followed by dephasing gradients.63 The OVS pulses are applied prior to the imaging acquisition, resulting in signal only from the nonsuppressed target region. OVS incurs increased SAR and longer scan time due to these suppression pulses, and is sensitive to RF transmit field inhomogeneity. Nevertheless, reduced FOV diffusion MRI with OVS has demonstrated superior structural details in spinal cord63 and pons64 compared to conventional SSH‐EPI. In combination with parallel imaging, OVS has been used to address the severe B0 inhomogeneity and short tissue T 2 value at ultrahigh field (Fig. 7).65, 66, 67 OVS has also been combined with SMS for high‐resolution diffusion MRI.68, 69
The third approach to reduced FOV uses 2D spatially selective RF pulse for excitation and a conventional slice‐selective 180° pulse for refocusing.70 The most common multidimensional pulses, using echo‐planar gradients, result in a periodic excitation profile. This profile places limits on the orientation and number of slices.71 A refinement of this method demonstrated the ability to simultaneously refocus two slices,72 doubling the number of slices that can be acquired in each scan. An alternate approach has been proposed that has virtually unlimited slice coverage, but which requires separate fat saturation.73 The long pulse durations associated with multidimensional excitations can be reduced using parallel transmission.74 Multidimensional excitation for rFOV diffusion imaging has been compared to the standard SSH‐EPI method, demonstrating clinical feasibility and improved conspicuity for spinal cord75, 76 and breast imaging.77, 78, 79
Accelerating Diffusion MRI Acquisitions
As noted above, SSH‐EPI is highly efficient during the signal readout period, providing all the spatial information for one slice in 20–40 msec. However, regardless of readout, diffusion MRI sequences are generally inefficient, with ≤50% of the sequence time dedicated to signal acquisition due to the need for a long diffusion preparation. In conventional 2D SSH‐EPI sequences, this inefficiency is compounded by the fact that each slice is encoded independently in series, such that the acquisition time per volume scales with the number of imaging slices. Assuming fixed coverage, increased spatial resolution requires more slices, further inflating volume scan time. This inefficiency results in a difficult tradeoff, particularly when a large number of diffusion directions are desired to improve the accuracy of angular information (eg, for diffusion tractography). When total scan time is limited, there is a fundamental tradeoff between spatial coverage, spatial resolution, and angular resolution (density of sampling in the diffusion‐encoding [directional] domain).
Several techniques have been introduced in the past few years that have the potential to dramatically reduce this tradeoff. Simultaneous multislice (SMS) techniques have provided the ability to acquire multiple diffusion‐encoded slices simultaneously, increasing the scan efficiency (as reflected in the number of slices acquired per unit time). Compressed sensing has shown the possibility to reconstruct MRI image from highly undersampled data,58 which can benefit diffusion scans with a large number of diffusion directions and/or multishot k‐space acquisition.
Simultaneous Multislice Imaging
The idea of exciting multiple slices simultaneously was proposed more than 20 years ago.80, 81 However, the first SMS techniques required multiple excitations to separate slices and did not reduce scan time. A key advance was made when it was realized that multichannel coil arrays enabled slice separation from a single acquisition through a parallel imaging formulation, thereby accelerating volume acquisition.82 This approach was subsequently extended to SSH‐EPI83 and demonstrated enabling high spatial‐angular resolution diffusion MRI.84, 85
A major challenge faced by SMS is noise amplification when gaps between slices are small. Coil profiles in general vary slowly across space, meaning that closely separated slices tend to have similar profiles (ie, the high‐signal region from one slice overlaps with the high‐signal region from another slice) (Fig 8b). This problem is the SMS manifestation of the “g‐factor” (the noise amplification for a given image voxel, which reflects coil configuration, acquisition protocol, and reconstruction algorithm) from conventional parallel imaging5 and is particularly challenging for diffusion MRI due to its low intrinsic SNR. The “blipped‐CAIPI” (controlled aliasing in parallel imaging) method reduces the g‐factor, representing a major improvement on SMS.85, 86 Blipped‐CAIPI introduces an apparent in‐plane shift between the excited slices, such that a given coil profile is spatially separated in the overlapping slices. In this case, the aliased voxels corresponding to two slices can be more easily separated because the coil profiles appear more distinct (ie, the high‐signal region from an unshifted slice overlaps with the low‐signal region from a shifted slice) (Fig. 8c). SMS‐EPI, in particular blipped‐CAIPI and its variants, has greatly improved the quality of diffusion imaging studies by reducing the tradeoff between spatial and angular resolution. Two high‐profile examples in brain imaging include the Human Connectome Project, where it has enabled a protocol with high spatial and angular resolution at several diffusion weighting “shells”87; and the UK Biobank Project, where it has enabled multiple shells with reasonable angular resolution in very limited scan time.88 Blipped‐CAIPI SMS‐EPI has also been incorporated with segmented‐EPI13, 17, 89 and reduced FOV (see above). An extension to 3D simultaneously multislab acquisition has also been reported.90
Image reconstruction to separate slices in SMS builds strongly on the existing literature in parallel imaging. There are two main categories of reconstruction methods for SMS data: SENSE‐GRAPPA91 and Slice‐GRAPPA.85 SENSE‐GRAPPA treats the overlapping slices as if they are neighboring in the phase encode direction space over a larger FOV, which casts the slice separation problem in a form that can be solved by conventional parallel imaging reconstructions. With the blipped‐CAIPI scheme, SENSE‐GRAPPA contains artifacts at the concatenation points, which have been avoided by additional zero‐padding92, 93, 94 or concatenating along the readout direction.95 By comparison, slice‐GRAPPA trains slice‐specific kernels that project k‐space data to one corresponding slice. Slice‐GRAPPA has been shown to be dependent on coil sensitivity rather than the image contrast,85 which is a desirable property for diffusion MRI, where the reconstruction is trained on data with no diffusion weighting. Several further refinements to SMS reconstruction have been proposed. The reconstruction kernel in Slice‐GRAPPA has been improved to reduce “leakage” between slices96 by training the kernel to block signal from all but one slice. This modification has been crucial for simultaneous slice and in‐plane acceleration, both of which play an important role in data quality. However, the interaction between these two accelerations is still challenging, given that both methods rely on multichannel coils in a similar manner.
RF pulses that excite multiple slices simultaneously are generally referred to as “multiband” (MB) pulses, since they deposit energy at several separate frequency bands. These pulses in general require a higher energy deposition, which is necessarily limited by patient safety considerations. A basic MB pulse is a superposition of multiple conventional RF pulses, leading to an N2 increase of the peak RF power for N simultaneously excited slices. This problem is particularly challenging for diffusion MRI due to the use of high‐energy 180° refocusing pulses. RF power can be reduced by optimizing the phases of the superimposed RF pulses97 or using a time‐shift scheme98; however, these methods only reduce the peak RF power, not the SAR level. The variable‐rate selective excitation (VERSE) algorithm99 can be used to improve MB pulses by modifying the gradients to slow down k‐space transversal speed during peak power deposition; however, VERSE suffers from RF profile distortion in the presence of strong field inhomogeneity. Another approach is PINS (power independent of number of slices) pulses,100 which undersample conventional single‐band pulses in a manner that excites equally separated slices while retaining power deposition comparable to a single‐slice excitation. Drawbacks of PINS pulse include poor slice profiles and limitations on the achievable slice orientation. Improvements include combination of PINS with more conventional MB pulses and methods that improve robustness against inhomogeneity.101, 102, 103 Parallel transmission (pTx) has also been explored for improving SMS acquisition. Using a full pTx‐MB model, a significant reduction on total RF power can be achieved.104 Alternatively, a dual‐ring RF array design105, 106 has been proposed, but suffers from the RF discontinuity between SMS slice stacks.
In summary, SMS enables full‐brain diffusion MRI with acquisition time of 3–7 seconds per volume (dependent on resolution and acceleration factor), enabling high spatial resolution and high angular resolution diffusion acquisition, provided reconstruction challenges such as compatibility with in‐plane acceleration can be addressed. The challenges from RF inhomogeneity and high SAR level are likely to be addressed by novel RF pulse design and pTx technique.
Fast Diffusion MRI With Compressed Sensing
The total acquisition time of a diffusion MRI exam is determined by the number of diffusion volumes (referred to as “q‐space samples”) and scan time for each volume (k‐space). Recently, compressed sensing (CS) methods have been applied to accelerated diffusion acquisition by exploring data sparsity in these two independent domains: q‐space and k‐space (Fig. 9).
Density of sampling in q‐space is a key determinant of data quality for diffusion MRI methods that aim to characterize the orientational microstructure of tissue; for example, to enable diffusion tractography (delineation of white matter pathways achieved by following the preferential direction of water diffusion from voxel‐to‐voxel) in the brain. This presents an important challenge for these methods because scan times are in general proportional to the number of q‐space samples acquired (since each direction represents one imaging volume). Recently, CS has been used to improve the reconstruction for acquisitions with a small number of q‐space samples (ie, q‐space undersampling, generally with a pseudorandom sampling pattern). For example, several groups have reported CS methods for reconstruction data acquired with a single b‐value (q‐space “shell”) to capture complex fiber architecture based on a multi‐tensor model107 or spherical ridgelets basis.108 Other CS methods aim to capture the additional information available from multiple q‐space shells using adaptive dictionaries or a sparsity prior on the diffusion probability density function.109, 110 These CS methods have reported high‐fidelity recovery of fiber orientation properties from data with undersampling factors ranging from three to eight, where higher accelerations are more readily achievable for simple models (eg, multi‐tensor fits) compared to less constrained models (eg, orientation distribution functions). Note that there are some other developments for CS diffusion MRI,111, 112, 113, 114 which are not discussed in the current review due to limited space.
In general, k‐space undersampling is only effective for reducing scan time in multishot methods, since 2D single‐shot techniques like SSH‐EPI acquire the entire k‐space required for one slice in a very brief time window. To improve the reconstruction fidelity with highly undersampled data, CS reconstructions have been proposed using sparsity priors on diffusion anisotropy,115 wavelet representations of the image data,116 and Gaussian mixture models of the diffusion process.117
Joint k‐q space accelerations, which undersample in both domains, have also been proposed.117, 118, 119 These methods use circulated or randomized k‐space undersampling patterns for different q‐space locations to increase the incoherence of signal aliasing, and used CS or low rank models to recover the underlying images. A critical step in the k‐q joint reconstruction is to correct the motion‐induced phase errors, which are inconsistent among q‐space samples. Navigator correction117, 119 and the MUSE scheme118 discussed in previous sections were applied to address this issue.
Increasing Image SNR
The third major challenge of diffusion MRI that we consider in this review is the intrinsic low SNR, which makes it particularly difficult to achieve high spatial resolution diffusion MRI data. Two dominant methods to enhance SNR of diffusion MRI without significantly increasing the scan time include using imaging sequences with higher SNR per unit time (efficiency) (eg, 3D acquisitions), and the use of ultrahigh field scanner (7T and higher). In this section, the developments of these two approaches for diffusion acquisition are discussed.
3D Acquisition
As detailed later on in this section, 3D acquisition methods are attractive because they enable imaging with high SNR efficiency. A further benefit of 3D imaging is the ability to define thin slices using gradient encoding (3D k‐space), which produces slices with similar quality of definition to in‐plane voxels, where 2D slices tend to have poor slice definition (eg, warping).
3D diffusion MRI is challenging because k‐space encoding in three dimensions can only be accomplished with a multishot approach, meaning that motion‐induced phase errors need to be corrected before combining across shots. In general, phase errors will vary spatially in all directions, implying that a full correction of these errors would require navigation in three dimensions; however, the acquisition of such a 3D navigator would require prohibitive scan time. Several approaches for avoiding the need for navigation in 3D imaging have been proposed. Driven equilibrium returns the magnetization to the longitudinal axis after diffusion preparation in order to enable spoiling of the phase error prior to the 3D readout.120 Although driven equilibrium avoids the profound image artifacts associated with phase cancellation, this method still has signal dependence on motion due to corruption of the longitudinal magnetization, which is highly problematic for quantification. Similarly, cardiac gating can reduce the effects of pulsatile motion in 3D imaging,121 but residual motion artifacts persist.
Many of the 3D acquisitions for diffusion imaging were first proposed in the context of diffusion weighted SSFP (DW‐SSFP),122 which is an intrinsically 3D sequence. Early work explored the use of 1D123 and 2D124, 125 navigation for correcting 3D data. Subsequent work proposed a rotated‐EPI acquisition scheme (TURBINE), which enabled a composite 3D navigator based on cardiac‐synchronized “batching” of 2D readouts.126 This approach was extended by including rigid‐body motion correction.127 However, a major outstanding challenge faced by DW‐SSFP methods is the disruption of signal formation by motion‐induced phase errors, which results in altered signal magnitude that confounds quantification.
Recently, development in this area has focused on 3D multislab acquisitions, in which a series of slabs are excited in sequence (similar to 2D multislice imaging but many thick slices), within which slices are defined using gradient encoding (ie, kz encoding as in 3D imaging). As a hybrid 2D/3D method, 3D multislab acquisition can achieve thin slices like 3D imaging while being compatible with the moderate TR (1–2 sec) associated with high SNR efficiency (Fig. 10a).26, 128 When a thin slab is applied, the motion‐induced phase error along the slice direction can be assumed to be slow varying, thus a 2D navigator can provide sufficiently good correction.59, 128 The use of a 2D navigator for phase error correction is a key enabler of 3D multislab diffusion acquisition, as it relieves the demanding requirement of 3D navigator acquisition. Several groups have demonstrated the use of 3D multislab acquisitions to achieve high‐resolution diffusion MRI data.26, 41, 128, 129, 130 A further challenge with 3D multislab acquisition is slab boundary artifacts, which arises because the magnetization experiences variable saturation at the overlap of slab profiles. Several reconstruction methods have been proposed to address this issue, with increasing degrees of sophistication (Fig. 10b).131, 132, 133 A recent study acquiring 3D multislab diffusion MRI data at 7T demonstrated high spatial resolution and high SNR,130 as shown in Fig. 10.
Ultrahigh Field
Because it provides greater signal (polarization of water), ultrahigh field strength enables higher intrinsic SNR, which has great potential for high resolution and/or high b‐value diffusion imaging. However, in order to fully benefit from the increased signal polarization, several challenges must be addressed, including: shorter T 2, which reduces signal level dramatically at the long echo times typical in diffusion MRI; the strong RF transmit field inhomogeneity, which is particularly problematic for refocusing pulses in spin‐echo sequences; increased B0 inhomogeneity, which is associated with EPI distortion; and significantly increased SAR level, also associated with refocusing and/or SMS pulses. Despite these challenges, several studies have demonstrated high‐quality diffusion MRI data at 7T.2, 68, 130, 134, 135 Due to shorter T 2, achieving short echo time is critical for realizing the SNR benefit at ultrahigh field136; this has been demonstrated through the combination of parallel imaging and partial Fourier acquisition.130, 135 The use of rs‐EPI17, 22 and reduced FOV66, 68 can further reduce the echo time, as well as reducing EPI distortion. RF transmit field inhomogeneity can be mitigated using dielectric pads,135 although this approach inevitably improves some regions while worsening homogeneity in others.137 A more comprehensive solution to transmit inhomogeneity would be to use parallel transmission, which can also reduce the SAR level. The combination of 3D multislab acquisition and 7T can further enhance the SNR performance.130 In all, diffusion MRI at ultrahigh field is a rapidly advancing research area. Given the intrinsic low‐SNR of diffusion MRI and the increasing desire for higher spatial resolution and b‐value, ultrahigh field is expected to play an important role in future diffusion MRI research.
Conclusion and Outlook
We have presented an overview of the major technical advances for diffusion MRI acquisition in recent years. Compared to standard 2D SSH‐EPI acquisition, these methods have demonstrated capabilities to enhance data fidelity, reduce acquisition time, and improve SNR. Some of these techniques (rs‐EPI, rFOV, PROPELLER) have already been applied in clinical settings with the aim of minimizing image distortion and blurring for improved diagnostics or disease monitoring in clinically feasible scan times. In contrast, basic neuroscience research applications typically require many diffusion directions with full brain coverage, for which accelerated acquisitions like SMS have already demonstrated major impact. For high isotropic resolution (eg, submillimeter) diffusion scans, 3D acquisition has advantages including the accurate slice profile and SNR efficiency. Finally, diffusion MRI is starting to take advantage of advances in ultrahigh field technology, including improved hardware and RF transmit techniques. While most of these methods have been developed in the brain, diffusion MRI throughout the body has become a major area of research focus, building on these advances for entirely new applications. Further improvement and combination of these techniques should enable more robust diffusion MRI acquisition with higher data quality, providing a powerful tool for new diagnostics, monitoring, and mechanistic studies.
Acknowledgments
Contract grant sponsor: Marie Curie Initial Training Network program; contract grant number: FP7‐PEOPLE‐2012‐ITN‐316716 (to W.W.); Contract grant sponsor: Wellcome Trust fellowships; contract grant number: 091509/Z/10/Z (to K.M.)
The copyright line for this article was changed on 7 March 2017 after original online publication.
Contributor Information
Wenchuan Wu, Email: wenchuan.wu@ndcn.ox.ac.uk.
Karla L. Miller, Email: karla@fmrib.ox.ac.uk.
References
- 1. McNab JA, Polimeni JR, Wang R, et al. Surface based analysis of diffusion orientation for identifying architectonic domains in the in vivo human cortex. Neuroimage 2013;69:87–100. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Kleinnijenhuis M, van Mourik T, Norris DG, Ruiter DJ, van Cappellen van Walsum AM, Barth M. Diffusion tensor characteristics of gyrencephaly using high resolution diffusion MRI in vivo at 7T. NeuroImage 2015;109:378–387. [DOI] [PubMed] [Google Scholar]
- 3. Miller KL, Stagg CJ, Douaud G, et al. Diffusion imaging of whole, post‐mortem human brains on a clinical MRI scanner. Neuroimage 2011;57:167–181. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Sotiropoulos SN, Jbabdi S, Xu J, et al. Advances in diffusion MRI acquisition and processing in the Human Connectome Project. NeuroImage 2013;80:125–143. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med 1999;42:952–962. [PubMed] [Google Scholar]
- 6. Griswold MA, Jakob PM, Heidemann RM, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med 2002;47:1202–1210. [DOI] [PubMed] [Google Scholar]
- 7. Reese T, Heid O, Weisskoff R, Wedeen V. Reduction of eddy‐current‐induced distortion in diffusion MRI using a twice‐refocused spin echo. Magn Reson Med 2003;49:177–182. [DOI] [PubMed] [Google Scholar]
- 8. Jezzard P, Barnett AS, Pierpaoli C. Characterization of and correction for eddy current artifacts in echo planar diffusion imaging. Magn Reson Med 1998;39:801–812. [DOI] [PubMed] [Google Scholar]
- 9. Andersson JLR, Sotiropoulos SN. An integrated approach to correction for off‐resonance effects and subject movement in diffusion MR imaging. Neuroimage 2016;125:1063–1078. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Robson MD, Porter DA. Reconstruction as a source of artifact in nongated single‐shot diffusion‐weighted EPI. Magn Reson Imaging 2005;23:899–905. [DOI] [PubMed] [Google Scholar]
- 11. Anderson AW, Gore JC. Analysis and correction of motion artifacts in diffusion weighted imaging. Magn Reson Med 1994;32:379–387. [DOI] [PubMed] [Google Scholar]
- 12. Storey P, Frigo FJ, Hinks RS, et al. Partial k‐space reconstruction in single‐shot diffusion‐weighted echo‐planar imaging. Magn Reson Med 2007;57:614–619. [DOI] [PubMed] [Google Scholar]
- 13. Chang HC, Guhaniyogi S, Chen NK. Interleaved diffusion‐weighted improved by adaptive partial‐Fourier and multiband multiplexed sensitivity‐encoding reconstruction. Magn Reson Med 2015;73:1872–1884. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Clare S. Iterative Nyquist ghost correction for single and multi‐shot EPI using an entropy measure. In: Proc 11th Annual Meeting ISMRM, Toronto; 2003. p 1041.
- 15. Delfaut EM, Beltran J, Johnson G, Rousseau J, Marchandise X, Cotten A. Fat suppression in MR imaging: techniques and pitfalls. Radiographics 1999;19:373–382. [DOI] [PubMed] [Google Scholar]
- 16. Robson MD, Anderson AW, Gore JC. Diffusion‐weighted multiple shot echo planar imaging of humans without navigation. Magn Reson Med 1997;38:82–88. [DOI] [PubMed] [Google Scholar]
- 17. Frost R, Jezzard P, Douaud Gl, Clare S, Porter DA, Miller KL. Scan time reduction for readout‐segmented EPI using simultaneous multislice acceleration: Diffusion‐weighted imaging at 3 and 7 Tesla. Magn Reson Med 2015;74:136–149. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Porter D, Mueller E. Multi‐shot diffusion‐weighted EPI with readout mosaic segmentation and 2D navigator correction. In: Proc 12th Annual Meeting of ISMRM, Kyoto; 2004. p 442.
- 19. Holdsworth SJ, Skare S, Newbould RD, Guzmann R, Blevins NH, Bammer R. Readout‐segmented EPI for rapid high resolution diffusion imaging at 3T. Eur J Radiol 2008;65:36–46. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Porter DA, Heidemann RM. High resolution diffusion‐weighted imaging using readout‐segmented echo‐planar imaging, parallel imaging and a two‐dimensional navigator‐based reacquisition. Magn Reson Med 2009;62:468–475. [DOI] [PubMed] [Google Scholar]
- 21. Holdsworth SJ, Skare S, Newbould RD, Bammer R. Robust GRAPPA‐accelerated diffusion‐weighted readout‐segmented (RS)‐EPI. Magn Reson Med 2009;62:1629–1640. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Heidemann RM, Porter DA, Anwander A, et al. Diffusion imaging in humans at 7T using readout‐segmented EPI and GRAPPA. Magn Reson Med 2010;64:9–14. [DOI] [PubMed] [Google Scholar]
- 23. Liu C, Moseley ME, Bammer R. Simultaneous phase correction and SENSE reconstruction for navigated multi‐shot DWI with non‐Carteian k‐space sampling. Magn Reson Med 2005;54:1412–1422. [DOI] [PubMed] [Google Scholar]
- 24. Pipe JG, Farthing VG, Forbes KP. Multishot diffusion‐weighted FSE using PROPELLER MRI. Magn Reson Med 2002;47:42–52. [DOI] [PubMed] [Google Scholar]
- 25. Frost R, Porter DA, Miller KL, Jezzard P. Implementation and assessment of diffusion‐weighted partial Fourier readout‐segmented echo‐planar imaging. Magn Reson Med 2012;68:441–451. [DOI] [PubMed] [Google Scholar]
- 26. Frost R, Miller KL, Tijssen RHN, Porter DA, Jezzard P. 3D multi‐slab diffusion‐weighted readout‐segmented EPI with real‐time cardiac‐reordered k‐space acquisition. Magn Reson Med 2014;72:1565–1579. [DOI] [PubMed] [Google Scholar]
- 27. Bogner W, Pinker‐Domenig K, Bickel H, et al. Readout‐segmented echo‐planar imaging improves the diagnostic performance of diffusion‐weighted MR breast examinations at 3.0 T. Radiology 2012;263:64–76. [DOI] [PubMed] [Google Scholar]
- 28. Bogner W, Pinker K, Zaric O, et al. Bilateral diffusion‐weighted MR imaging of breast tumors with submillimeter resolution using readout‐segmented echo‐planar imaging at 7 T. Radiology 2015;274:74–84. [DOI] [PubMed] [Google Scholar]
- 29. Wisner DJ, Rogers N, Deshpande VS, et al. High‐resolution diffusion‐weighted imaging for the separation of benign from malignant BI‐RADS 4/5 lesions found on breast MRI at 3T. J Magn Reson Imaging 2014;40:674–681. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Tokoro H, Fujinaga Y, Ohya A, et al. Usefulness of free‐breathing readout‐segmented echo‐planar imaging (RESOLVE) for detection of malignant liver tumors: Comparison with single‐shot echo‐planar imaging (SS‐EPI). Eur J Radiol 2014;83:1728–1733. [DOI] [PubMed] [Google Scholar]
- 31. Foltz WD, Porter DA, Simeonov A, et al. Readout‐segmented echo‐planar diffusion‐weighted imaging improves geometric performance for image‐guided radiation therapy of pelvic tumors. Radiother Oncol 2014;117:525–531. [DOI] [PubMed] [Google Scholar]
- 32. Thian YL, Xie W, Porter DA, Weileng Ang B. Readout‐segmented Echo‐planar imaging for diffusion‐weighted imaging in the pelvis at 3T—A feasibility study. Acad Radiol 2014;21:531–537. [DOI] [PubMed] [Google Scholar]
- 33. Friedli I, Crowe LA, Viallon M, et al. Improvement of renal diffusion‐weighted magnetic resonance imaging with readout‐segmented echo‐planar imaging at 3T. Magn Reson Imaging 2015;33:701–708. [DOI] [PubMed] [Google Scholar]
- 34. Yeom KW, Holdsworth SJ, Van AT, et al. Comparison of readout‐segmented echo‐planar imaging (EPI) and single‐shot epi in clinical application of diffusion‐weighted imaging of the pediatric brain. AJR Am J Roentgenol 2013;200:437–443. [DOI] [PubMed] [Google Scholar]
- 35. Butts K, deCrespigny A, Pauly JM, Moseley M. Diffusion‐weighted interleaved echo‐planar imaging with a pair of orthogonal navigator echoes. Magn Reson Med 1996;35:763–770. [DOI] [PubMed] [Google Scholar]
- 36. Atkinson D, Counsell S, Hajnal JV, Batchelor PG, Hill DL, Larkman DJ. Nonlinear phase correction of navigated multi‐coil diffusion images. Magn Reson Med 2006;56:1135–1139. [DOI] [PubMed] [Google Scholar]
- 37. Chen Nk, Guidon A, Chang HC, Song AW. A robust multi‐shot scan strategy for high‐resolution diffusion weighted MRI enabled by multiplexed sensitivity‐encoding (MUSE). NeuroImage 2013;72:41–47. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Guhaniyogi S, Chu ML, Chang HC, Song AW, Chen Nk. Motion immune diffusion imaging using augmented MUSE for high‐resolution multi‐shot EPI. Magn Reson Med 2015;652:639–652. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Herbst M, Zahneisen B, Knowles B, Zaitsev M, Ernst T. Prospective motion correction of segmented diffusion weighted EPI. Magn Reson Med 2015;74:1675–1681. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Zhang Z, Huang F, Ma X, Xie S, Guo H. Self‐feeding MUSE: A robust method for high resolution diffusion imaging using interleaved EPI. NeuroImage 2015;105:552–560. [DOI] [PubMed] [Google Scholar]
- 41. Chang HC, Sundman M, Petit L, et al. Human brain diffusion tensor imaging at submillimeter isotropic resolution on a 3Tesla clinical MRI scanner. NeuroImage 2015;118:667–675. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Mani M, Jacob M, Kelley D, Magnotta V. Multi‐shot sensitivity‐encoded diffusion data recovery using structured low‐rank matrix completion (MUSSELS). Magn Reson Med 2016. doi:10.1002/mrm.26382 [Epub ahead of print]. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. Hennig J, Nauerth A, Friedburg H. RARE imaging: a fast imaging method for clinical MR. Magn Reson Med 1986;3:823–833. [DOI] [PubMed] [Google Scholar]
- 44. Pipe JG. Motion correction with PROPELLER MRI: application to head motion and free‐breathing cardiac imaging. Magn Reson Med 1999;42:963–969. [DOI] [PubMed] [Google Scholar]
- 45. Meiboom S, Gill D. Modified spin‐echo method for measuring nuclear relaxation times. Rev Sci Instrum 1958;29:688–691. [Google Scholar]
- 46. Levitt MH, Freeman R. Compensation for pulse imperfections in NMR spin‐echo experiments. J Magn Reson 1981;43:65–80. [Google Scholar]
- 47. Le Roux P. Non‐CPMG fast spin echo with full signal. J Magn Reson 2002;155:278–292. [DOI] [PubMed] [Google Scholar]
- 48. Pipe JG, Zwart N. Turboprop: Improved PROPELLER imaging. Magn Reson Med 2006;55:380–385. [DOI] [PubMed] [Google Scholar]
- 49. Oshio K, Feinberg DA. GRASE (gradient‐ and spin‐echo) imaging: a novel fast MRI technique. Magn Reson Med 1991;20:344–349. [DOI] [PubMed] [Google Scholar]
- 50. Wang F‐N, Huang T‐Y, Lin F‐H, et al. PROPELLER EPI: An MRI technique suitable for diffusion tensor imaging at high field strength with reduced geometric distortions. Magn Reson Med 2005;54:1232–1240. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51. Chuang T‐C, Huang T‐Y, Lin F‐H, et al. PROPELLER‐EPI with parallel imaging using a circularly symmetric phased‐array RF coil at 3.0 T: Application to high‐resolution diffusion tensor imaging. Magn Reson Med 2006;56:1352–1358. [DOI] [PubMed] [Google Scholar]
- 52. Skare S, Newbould RD, Clayton DB, Bammer R. Propeller EPI in the other direction. Magn Reson Med 2006;55:1298–1307. [DOI] [PubMed] [Google Scholar]
- 53. Meyer CH, Hu BS, Nishimura DG, Macovski A. Fast spiral coronary‐artery imaging. Magn Reson Med 1992;28:202–213. [DOI] [PubMed] [Google Scholar]
- 54. Glover G. Basic and advanced concepts of spiral imaging. International Society for Magnetic Resonance in Medicine Fast MRI Workshop, Asilomar, CA; 1997. p 115‐119.
- 55. Bammer R, Glover G, Moseley M. Diffusion tensor spiral imaging. In: Proc 10th Annual Meeting of ISMRM, Honolulu; 2002. p 1111.
- 56. Liu C, Bammer R, Kim D‐H, Moseley ME. Self‐navigated interleaved spiral (SNAILS): application to high‐resolution diffusion tensor imaging. Magn Reson Med 2004;52:1388–1396. [DOI] [PubMed] [Google Scholar]
- 57. Guo H, Ma X, Zhang Z, Zhang B, Yuan C, Huang F. POCS‐enhanced inherent correction of motion‐induced phase errors (POCS‐ICE) for high‐resolution multishot diffusion MRI. Magn Reson Med 2016;75:169–180. [DOI] [PubMed] [Google Scholar]
- 58. Lustig M, Donoho DL, Santos JM, Pauly JM. Compressed sensing MRI. IEEE Signal Process Mag 2008;25:72–82. [Google Scholar]
- 59. Frank LR, Jung Y, Inati S, Tyszka JM, Wong EC. High efficiency, low distortion 3D diffusion tensor imaging with variable density spiral fast spin echoes (3D DW VDS RARE). NeuroImage 2010;49:1510–1523. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60. Feinberg DA, Hoenninger J, Crooks L, Kaufman L, Watts J, Arakawa M. Inner volume MR imaging: technical concepts and their application. Radiology 1985;156:743–747. [DOI] [PubMed] [Google Scholar]
- 61. Wheeler‐Kingshott CAM, Parker GJM, Symms MR, et al. ADC mapping of the human optic nerve: Increased resolution, coverage, and reliability with CSF‐suppressed ZOOM‐EPI. Magn Reson Med 2002;47:24–31. [DOI] [PubMed] [Google Scholar]
- 62. Jeong EK, Kim SE, Guo J, Kholmovski EG, Parker DL. High‐resolution DTI with 2D interleaved multislice reduced FOV single‐shot diffusion‐weighted EPI (2D ss‐rFOV‐DWEPI). Magn Reson Med 2005;54:1575–1579. [DOI] [PubMed] [Google Scholar]
- 63. Wilm BJ, Svensson J, Henning A, Pruessmann KP, Boesiger P, Kollias SS. Reduced field‐of‐view MRI using outer volume suppression for spinal cord diffusion imaging. Magn Reson Med 2007;57:625–630. [DOI] [PubMed] [Google Scholar]
- 64. Karampinos DC, Van AT, Olivero WC, Georgiadis JG, Sutton BP. High‐resolution diffusion tensor imaging of the human pons with a reduced field‐of‐view, multishot, variable‐density, spiral acquisition at 3 T. Magn Reson Med 2009;62:1007–1016. [DOI] [PubMed] [Google Scholar]
- 65. von Morze C, Kelley DaC, Shepherd TM, Banerjee S, Xu D, Hess CP. Reduced field‐of‐view diffusion‐weighted imaging of the brain at 7 T. Magn Reson Imaging 2010;28:1541–1545. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66. Heidemann RM, Anwander A, Feiweier T, Knosche TR, Turner R. k‐space and q‐space: Combining ultra‐high spatial and angular resolution in diffusion imaging using ZOOPPA at 7T. NeuroImage 2012;60:967–978. [DOI] [PubMed] [Google Scholar]
- 67. Wargo CJ, Gore JC. Localized high‐resolution DTI of the human midbrain using single‐shot EPI, parallel imaging, and outer‐volume suppression at 7T. Magn Reson Imaging 2013;31:810–819. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68. Eichner C, Setsompop K, Koopmans PJ, et al. Slice accelerated diffusion‐weighted imaging at ultra‐high field strength. Magn Reson Med 2014;71:1518–1525. [DOI] [PubMed] [Google Scholar]
- 69. Setsompop K, Bilgic B, Nummenmaa A, et al. SLIce Dithered Enhanced Resolution Simultaneous MultiSlice (SLIDER‐SMS) for high resolution (700 μm) diffusion imaging of the human brain. In: Proc 23rd Annual Meeting ISMRM, Toronto; 2015. p 339.
- 70. Rieseberg S, Frahm J, Finsterbusch J. Two‐dimensional spatially‐selective RF excitation pulses in echo‐planar imaging. Magn Reson Med 2002;47:1186–1193. [DOI] [PubMed] [Google Scholar]
- 71. Saritas EU, Cunningham CH, Lee JH, Han ET, Nishimura DG. DWI of the spinal cord with reduced FOV single‐shot EPI. Magn Reson Med 2008;60:468–473. [DOI] [PubMed] [Google Scholar]
- 72. Saritas EU, Lee D, Cukur T, Shankaranarayanan A, Nishimura DG. Hadamard slice encoding for reduced‐FOV diffusion‐weighted imaging. Magn Reson Med 2014;72:1277–1290. [DOI] [PubMed] [Google Scholar]
- 73. Finsterbusch J. High‐resolution diffusion tensor imaging with inner field‐of‐view EPI. J Magn Reson Imaging 2009;29:987–993. [DOI] [PubMed] [Google Scholar]
- 74. Schneider JT, Kalayciyan R, Haas M, et al. Inner‐volume imaging in vivo using three‐dimensional parallel spatially selective excitation. Magn Reson Med 2013;69:1367–1378. [DOI] [PubMed] [Google Scholar]
- 75. Zaharchuk G, Saritas EU, Andre JB, et al. Reduced field‐of‐view diffusion imaging of the human spinal cord: comparison with conventional single‐shot echo‐planar imaging. AJNR Am J Neuroradiol 2011;32:813–820. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76. Andre JB, Zaharchuk G, Saritas E, et al. Clinical evaluation of reduced field‐of‐view diffusion‐weighted imaging of the cervical and thoracic spine and spinal cord. AJNR Am J Neuroradiol 2012;33:1860–1866. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77. Park JY, Shin HJ, Shin KC, et al. Comparison of readout segmented echo planar imaging (EPI) and EPI with reduced field‐of‐view diffusion‐weighted imaging at 3T in patients with breast cancer. J Magn Reson Imaging 2015;42:1679–1688. [DOI] [PubMed] [Google Scholar]
- 78. Singer L, Wilmes LJ, Saritas EU, et al. High‐resolution diffusion‐weighted magnetic resonance imaging in patients with locally advanced breast cancer. Acad Radiol 2012;19:526–534. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79. Wilmes LJ, McLaughlin RL, Newitt DC, et al. High‐resolution diffusion‐weighted imaging for monitoring breast cancer treatment response. Acad Radiol 2013;20:581–589. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80. Souza SP, Szumowski J, Dumoulin CL, Plewes DP, Glover G. Sima — Simultaneous Multislice Acquisition of MR Images by Hadamard‐Encoded Excitation. J Comput Assist Tomogr 1988;12:1026–1030. [PubMed] [Google Scholar]
- 81. Muller S. Multifrequency selective RF pulses for multislice MR imaging. Magn Reson Med 1988;6:364–371. [DOI] [PubMed] [Google Scholar]
- 82. Larkman DJ, Hajnal JV, Herlihy AH, Coutts GA, Young IR, Ehnholm G. Use of multicoil arrays for separation of signal from multiple slices simultaneously excited. J Magn Reson Imaging 2001;13:313–317. [DOI] [PubMed] [Google Scholar]
- 83. Nunes R, Hajnal J, Golay X, Larkman D. Simultaneous slice excitation and reconstruction for single shot EPI. In: Proc 14th Annual Meeting ISMRM, Seattle; 2006. p 293.
- 84. Feinberg DA, Moeller S, Smith SM, et al. Multiplexed echo planar imaging for sub‐second whole brain fMRI and fast diffusion imaging. PLoS One 2010;5:e15710. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85. Setsompop K, Gagoski BA, Polimeni JR, Witzel T, Wedeen VJ, Wald LL. Blipped‐controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g‐factor penalty. Magn Reson Med 2012;67:1210–1224. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86. Setsompop K, Cohen‐Adad J, Gagoski BA, et al. Improving diffusion MRI using simultaneous multi‐slice echo planar imaging. NeuroImage 2012;63:569–580. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87. Sotiropoulos SN, Jbabdi S, Xu J, et al. Advances in diffusion MRI acquisition and processing in the Human Connectome Project. Neuroimage 2013;80:125–143. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 88. Miller KL, Alfaro‐Almagro F, Bangerter NK, et al. Multimodal population brain imaging in the UK Biobank prospective epidemiological study. Nat Neurosci 2016;19:1523–1536. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89. Dai E, Ma X, Zhang Z, Yuan C, Guo H. Simultaneous multislice accelerated interleaved EPI DWI using generalized blipped‐CAIPI acquisition and 3D k‐space reconstruction. Magn Reson Med 2016;71:1–13. [DOI] [PubMed] [Google Scholar]
- 90. Frost R, Jezzard P, Porter DA, Tijssen R, Miller K. Simultaneous multi‐slab acquisition in 3D multislab diffusion‐weighted readout‐segmented echo‐planar imaging. In: Proc 21st Annual Meeting ISMRM, Salt Lake City; 2013. p 3176.
- 91. Blaimer M, Breuer FA, Seiberlich N, et al. Accelerated volumetric MRI with a SENSE/GRAPPA combination. J Magn Reson Imaging 2006;24:444–450. [DOI] [PubMed] [Google Scholar]
- 92. Stäb D, Ritter CO, Breuer FA, Weng AM, Hahn D, Köstler H. CAIPIRINHA accelerated SSFP imaging. Magn Reson Med 2011;65:157–164. [DOI] [PubMed] [Google Scholar]
- 93. Blaimer M, Choli M, Jakob PM, Griswold MA, Breuer FA. Multiband phase‐constrained parallel MRI. Magn Reson Med 2013;69:974–980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94. Zhu K, Kerr A, Pauly J. Autocalibrating CAIPIRINHA: reformulating CAIPIRINHA as a 3D problem. In: Proc 20th Annual Meeting ISMRM, Melbourne; 2012. p 518.
- 95. Moeller S, Vu A, Auerbach E, Ugurbil K, Yacoub E. RO extended FOV SENSE/GRAPPA for multiband imaging with FOV shift. In: Proc 22nd Annual Meeting ISMRM, Milan; 2014. p 4396.
- 96. Cauley SF, Polimeni JR, Bhat H, Wald LL, Setsompop K. Interslice leakage artifact reduction technique for simultaneous multislice acquisitions. Magn Reson Med 2014;72:93–102. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 97. Wong E. Optimized phase schedules for minimizing peak RF power in simultaneous multi‐slice RF excitation pulses. In: Proc 20th Annual Meeting ISMRM, Melbourne; 2012. p 2209.
- 98. Auerbach EJ, Xu J, Yacoub E, Moeller S, Ugurbil K. Multiband accelerated spin‐echo echo planar imaging with reduced peak RF power using time‐shifted RF pulses. Magn Reson Med 2013;69:1261–1267. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 99. Conolly S, Nishimura D, Macovski A, Glover G. Variable‐rate selective excitation. J Magn Reson (1969) 1988;78:440–458. [Google Scholar]
- 100. Norris DG, Koopmans PJ, Boyacioglu R, Barth M. Power independent of number of slices (PINS) radiofrequency pulses for low‐power simultaneous multislice excitation. Magn Reson Med 2011;66:1234–1240. [DOI] [PubMed] [Google Scholar]
- 101. Eichner C, Wald LL, Setsompop K. A low power radiofrequency pulse for simultaneous multislice excitation and refocusing. Magn Reson Med 2014;72:949–958. [DOI] [PubMed] [Google Scholar]
- 102. Sharma A, Holdsworth S, O'Halloran R, et al. kT‐PINS RF pulses for low‐power field inhomogeneity‐compensated multislice excitation. In: Proc 21st Annual Meeting ISMRM, Salt Lake City; 2013. p 73.
- 103. Feldman RE, Islam HM, Xu J, Balchandani P. A SEmi‐Adiabatic matched‐phase spin echo (SEAMS) PINS pulse‐pair for B1‐insensitive simultaneous multislice imaging. Magn Reson Med 2016;75:709–717. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104. Wu X, Schmitter S, Auerbach EJ, Moeller S, Ugurbil K, Van De Moortele PF. Simultaneous multislice multiband parallel radiofrequency excitation with independent slice‐specific transmit B1 homogenization. Magn Reson Med 2013;70:630–638. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 105. Poser BA, Anderson RJ, Guérin B, et al. Simultaneous multislice excitation by parallel transmission. Magn Reson Med 2014;71:1416–1427. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 106. Wu X, Tian J, Schmitter S, Vaughan JT, Ugurbil K, Van de Moortele PF. Distributing coil elements in three dimensions enhances parallel transmission multiband RF performance: A simulation study in the human brain at 7 Tesla. Magn Reson Med 2016;2472:2464–2472. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 107. Landman BA, Bogovic JA, Wan H, ElShahaby FEZ, Bazin PL, Prince JL. Resolution of crossing fibers with constrained compressed sensing using diffusion tensor MRI. NeuroImage 2012;59:2175–2186. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 108. Michailovich O, Rathi Y, Dolui S. Spatially regularized compressed sensing of diffusion MRI data. Psychiatry Interpers Biol Process 2010;1:1100–1115. [Google Scholar]
- 109. Menzel MI, Tan ET, Khare K, et al. Accelerated diffusion spectrum imaging in the human brain using compressed sensing. Magn Reson Med 2011;66:1226–1233. [DOI] [PubMed] [Google Scholar]
- 110. Bilgic B, Setsompop K, Cohen‐Adad J, Yendiki A, Wald LL, Adalsteinsson E. Accelerated diffusion spectrum imaging with compressed sensing using adaptive dictionaries. Magn Reson Med 2012;68:1747–1754. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 111. Merlet S, Caruyer E, Deriche R. Parametric dictionary learning for modeling EAP and ODF in diffusion MRI. Medical image computing and computer‐assisted intervention: MICCAI International Conference on Medical Image Computing and Computer‐Assisted Intervention 2012;15:10–17. [DOI] [PubMed] [Google Scholar]
- 112. Duarte‐Carvajalino JM, Lenglet C, Xu J, et al. Estimation of the CSA‐ODF using Bayesian compressed sensing of multi‐shell HARDI. Magn Reson Med 2014;72:1471–1485. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 113. Ning L, Setsompop K, Michailovich O, et al. A joint compressed‐sensing and super‐resolution approach for very high‐resolution diffusion imaging. NeuroImage 2016;125:386–400. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 114. Rathi Y, Michailovich O, Setsompop K, Bouix S, Shenton ME, Westin C‐F. Sparse multi‐shell diffusion imaging International Conference on Medical Image Computing and Computer‐Assisted Intervention. Berlin: Springer; 2011. p 58‐65. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 115. Shi X, Ma X, Wu W, Huang F, Yuan C, Guo H. Parallel imaging and compressed sensing combined framework for accelerating high‐resolution diffusion tensor imaging using inter‐image correlation. Magn Reson Med 2015;73:1775–1785. [DOI] [PubMed] [Google Scholar]
- 116. Wu Y, Zhu Y‐J, Tang Q‐T, et al. Accelerated MR diffusion tensor imaging using distributed compressed sensing. Magn Reson Med 2014;71:763–772. [DOI] [PubMed] [Google Scholar]
- 117. Mani M, Jacob M, Guidon A, Magnotta V, Zhong J. Acceleration of high angular and spatial resolution diffusion imaging using compressed sensing with multichannel spiral data. Magn Reson Med 2015;73:126–138. [DOI] [PubMed] [Google Scholar]
- 118. Chao T‐C, Chiou J‐G, Maier SE, Madore B. Fast diffusion imaging with high angular resolution. Magn Reson Med 2016. doi:10.1002/mrm.26163 [Epub ahead of print]. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 119. Gao H, Li L, Zhang K, Zhou W, Hu X. PCLR: Phase‐constrained low‐rank model for compressive diffusion‐weighted MRI. Magn Reson Med 2014;72:1330–1341. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 120. Jeong EK, Kim SE, Parker DL. High‐resolution diffusion‐weighted 3D MRI, using diffusion‐weighted driven‐equilibrium (DW‐DE) and multishot segmented 3D‐SSFP without navigator echoes. Magn Reson Med 2003;50:821–829. [DOI] [PubMed] [Google Scholar]
- 121. Golay X, Jiang H, van Zijl PC, Mori S. High‐resolution isotropic 3D diffusion tensor imaging of the human brain. Magn Reson Med 2002;47:837–843. [DOI] [PubMed] [Google Scholar]
- 122. Buxton RB. The diffusion sensitivity of fast steady‐state free precession imaging. Magn Reson Med 1993;29:235–243. [DOI] [PubMed] [Google Scholar]
- 123. Bosak E, Harvey PR. Navigator motion correction of diffusion weighted 3D SSFP imaging. Magn Reson Mater Phys Biol Med 2001;12:167–176. [DOI] [PubMed] [Google Scholar]
- 124. Miller KL, Hargreaves BA, Gold GE, Pauly JM. Steady‐state diffusion‐weighted imaging of in vivo knee cartilage. Magn Reson Med 2004;51:394–398. [DOI] [PubMed] [Google Scholar]
- 125. Miller KL, Pauly JM. Nonlinear phase correction for navigated diffusion imaging. Magn Reson Med 2003;50:343–353. [DOI] [PubMed] [Google Scholar]
- 126. McNab JA, Gallichan D, Miller KL. 3D steady‐state diffusion‐weighted imaging with trajectory using radially batched internal navigator echoes (TURBINE). Magn Reson Med 2010;63:235–242. [DOI] [PubMed] [Google Scholar]
- 127. O'Halloran RL, Aksoy M, Van AT, Bammer R. 3D isotropic high‐resolution diffusion‐weighted MRI of the whole brain with a motion‐corrected steady‐state free precession sequence. Magn Reson Med 2013;70:466–478. [DOI] [PubMed] [Google Scholar]
- 128. Engström M, Skare S. Diffusion‐weighted 3D multislab echo planar imaging for high signal‐to‐noise ratio efficiency and isotropic image resolution. Magn Reson Med 2013;70:1507–1514. [DOI] [PubMed] [Google Scholar]
- 129. Van AT, Hernando D, Sutton BP. Motion‐induced phase error estimation and correction in 3D diffusion tensor imaging. IEEE Trans Med Imaging 2011;30:1933–1940. [DOI] [PubMed] [Google Scholar]
- 130. Wu W, Poser BA, Douaud G, et al. High‐resolution diffusion MRI at 7T using a three‐dimensional multi‐slab acquisition. NeuroImage 2016;143:1–14. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 131. Van AT, Aksoy M, Holdsworth SJ, Kopeinigg D, Vos SB, Bammer R. Slab profile encoding (PEN) for minimizing slab boundary artifact in three‐dimensional diffusion‐weighted multislab acquisition. Magn Reson Med 2014;71:1–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 132. Wu W, Koopmans PJ, Frost R, Miller KL. Reducing slab boundary artifacts in three‐dimensional multislab diffusion MRI using nonlinear inversion for slab profile encoding (NPEN). Magn Reson Med 2016;76:1183–1195. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 133. Engström M, Martensson M, Avventi E, Skare S. On the signal‐to‐noise ratio efficiency and slab‐banding artifacts in three‐dimensional multislab diffusion‐weighted echo‐planar imaging. Magn Reson Med 2015;73:718–725. [DOI] [PubMed] [Google Scholar]
- 134. Jeong HK, Gore JC, Anderson AW. High‐resolution human diffusion tensor imaging using 2‐D navigated multishot SENSE EPI at 7 T. Magn Reson Med 2013;69:793–802. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 135. Vu AT, Auerbach E, Lenglet C, et al. High resolution whole brain diffusion imaging at 7T for the Human Connectome Project. NeuroImage 2015;122:318–331. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 136. Ugurbil K, Xu J, Auerbach EJ, et al. Pushing spatial and temporal resolution for functional and diffusion MRI in the Human Connectome Project. Neuroimage 2013;80:80–104. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 137. Teeuwisse WM, Brink WM, Webb AG. Quantitative assessment of the effects of high‐permittivity pads in 7 Tesla MRI of the brain. Magn Reson Med 2012;67:1285–1293. [DOI] [PubMed] [Google Scholar]