Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2018 Sep 1.
Published in final edited form as: Optom Vis Sci. 2017 Sep;94(9):886–893. doi: 10.1097/OPX.0000000000001107

Aberrometry Repeatability and Agreement with Autorefraction

Mylan T Nguyen 1, David A Berntsen 1
PMCID: PMC5578873  NIHMSID: NIHMS886524  PMID: 28727613

Abstract

Significance

Commercially-available aberrometers are essential to clinical studies evaluating refractive error and image quality. The Discovery System is a promising clinical instrument that allows investigators to export aberration data for research and analysis purposes. An assessment of the Discovery System’s performance is essential to the interpretation of the data obtained.

Purpose

To determine the between-visit repeatability of refractive error and higher-order aberration measurements with the Discovery System, and to examine between-instrument agreement of refractive error measurements with the Discovery System and Grand Seiko WAM-5500 open-field autorefractor.

Methods

Cycloplegic refractive error values from the Discovery System (over a 3-mm pupil) and the Grand Seiko autorefractor were converted to power vectors (M, J0, and J45) and averaged. Zernike coefficients were also calculated by the Discovery System over a 6-mm pupil through the 6th radial order. Between-visit repeatability and agreement were evaluated using Bland-Altman difference versus mean plots. T-tests compared each mean difference (bias) to zero and the 95% limits of agreement (LoA) were calculated.

Results

Twenty-five young adults with a mean (±SD) cycloplegic spherical-equivalent refractive error of −2.91 ± 1.85 D (range: −6.96 to +0.74) were enrolled. There were no significant between-visit differences with the Discovery System for M, J0, J45, 3rd through 6th order RMS, HORMS, or spherical aberration (all p > 0.30), and repeatability for defocus and HORMS were ±0.31 D and ±0.095 μm, respectively, for a 6-mm pupil. At a 3-mm pupil, the Discovery System on average measured slightly more positive values than the Grand Seiko for M (0.28 D), J0 (0.11 D), and J45 (0.12 D; all p <0.005).

Conclusions

The Discovery System was very repeatable and would be an appropriate instrument to measure cycloplegic refractive error and higher-order aberration changes in adults. Small differences in refractive error were found between the Discovery System and Grand Seiko.

Keywords: between-visit repeatability, aberrometry, lower-order aberrations, higher-order aberrations, between-instrument agreement, Hartmann screen


Measurements of both lower-order and higher-order aberrations are critical in a wide range of clinical studies evaluating refractive error and image quality, such as studies of myopia and keratoconus.15 Grand Seiko autorefractors (Grand Seiko Co., Hiroshima, Japan), also manufactured under the name Shin-Nippon, are frequently used in clinical studies to objectively measure refractive error because of their well-documented repeatability; however, the Grand Seiko autorefractor does not measure higher-order aberrations.612 Hartmann-Shack aberrometry has been widely used in clinical research over the last several decades.13, 14 The Complete Ophthalmic Analysis System (COAS; AMO Wavefront Sciences, Albuquerque, New Mexico), a commercial Hartmann-Shack aberrometer, has been widely used in clinical research to simultaneously measure both lower-order (refractive error) and higher-order aberrations. The repeatability of the COAS was documented and refractive error measurements had been validated against the Grand Seiko autorefractor; however, the COAS is no longer commercially available.1518 There is a need for a commercially available aberrometer with accurate refraction measurements and known repeatability in clinical research. Having access to these data allows for more detailed analysis such as calculating metrics of image quality.

The Discovery System (Innovative Visual Systems, Elmhurst, Illinois) is a new commercially-available wavefront aberrometer. Unlike the COAS, which uses a Hartmann-Shack approach,13, 19 the Discovery System utilizes a Hartmann-screen design. This design does not utilize a lenslet array and has a high-dynamic range that eliminates the need to pre-correct defocus as is done by the COAS. The instrument is capable of calculating Zernike coefficients up to the 10th radial order over a range of pupil sizes. The purpose of this study is to assess the between-visit repeatability of refractive error (lower-order) and higher-order aberration measurements made with the Discovery System, and to examine the between-instrument agreement of refractive error measurements between the Discovery System and the Grand Seiko WAM-5500 open-field autorefractor.

METHODS

Adults between the ages of 18 and 39 were recruited to participate in two study visits separated by no more than 14 days. The study protocol was approved by the University of Houston Committee for the Protection of Human Subjects and followed the guidelines of the Declaration of Helsinki. Study subjects were provided a written informed consent document that they signed prior to enrollment. All subjects had no history of ocular disease, ocular trauma, ocular surgery, or systemic diseases that could cause refractive error variation (e.g., diabetes). Rigid gas permeable contact lens wearers were excluded and soft contact lens wearers were instructed to wear glasses on the days of their study visits.

Measurements

Five consecutive cycloplegic measurements of the right eye were made with the Discovery System at two separate visits (Visit 1 and Visit 2). Subjects fixated the red fixation spot within the instrument. The Discovery System captures a single measurement with each button press, and each measurement was made after instructing the subject to blink twice and open their eye. Measurements were taken at the same time of day (±1 hour) during the two visits and all subjects were asked to present to visits with no contact lenses on. Ten consecutive cycloplegic measurements of the right eye were also made using the Grand Seiko WAM-5500 at Visit 2. Instrument order at the second visit was randomized. Measurements were made under cycloplegia (1% tropicamide) using an established cycloplegia protocol frequently used in both adult and pediatric clinical research.20, 21 The Discovery System does not fog the internal target, making cycloplegia important to ensure relaxed accommodation. Measurements began 30 minutes after the first of two drops of 1% tropicamide that were separated by five minutes. All subjects wore a patch over the left eye to ensure accurate fixation by the right eye during measurements.

Prior to each measurement, the examiner inspected the subject’s alignment to ensure each subject was positioned with their chin on the chin rest and their forehead against the forehead rest. The measurement axis of the Discovery System was aligned with the right eye’s line of sight by instructing the subject to fixate on the instrument’s internal fixation light. Before each measurement, subjects were instructed to blink twice and then hold their eye open. Between each measurement, the examiner re-evaluated each subject’s alignment. For measurements with the Grand Seiko WAM-5500, subjects fixated a red LED spot of light on the wall directly in front of them.

Data Analysis

Zernike coefficients were calculated by the Discovery System over a 6-mm pupil diameter through the 6th radial order. The measurements were averaged and exported for further analysis. Refractive error values were exported from the Discovery System (sphere, cylinder, and axis) and the Grand Seiko autorefractor. The Discovery System utilized only the second-order aberration values to compute the refractive error displayed. For the Discovery System, refractive error values were obtained for the 6-mm diameter over which all Zernike coefficients were calculated and for a 3-mm pupil diameter to allow for comparable conditions when comparing to refractive error measurements made by the Grand Seiko autorefractor’s 2.3-mm measurement beam diameter.22 A 3-mm pupil is the smallest diameter for which the Discovery System displays refractive error data (sphere, cylinder, and axis). All refractive error values were converted to power vectors (M, J0, and J45) using previously described methods and averaged.23 The root mean square (RMS) wavefront error was calculated for third-, fourth-, fifth-, sixth-order terms, and total higher-order RMS (HORMS) was calculated using the 3rd through 6th order Zernike terms.

Statistical analyses were performed using STATA 12 (Stata Corp., College Station, TX). Between-visit repeatability and between-instrument agreement were evaluated using Bland-Altman difference versus mean plots.24 Differences between the two visits for M, J0, J45, third-order RMS, fourth-order RMS, fifth-order RMS, sixth-order RMS, HORMS, vertical trefoil ( C3-3), vertical coma ( C3-1), horizontal coma ( C31), oblique trefoil ( C33), oblique secondary astigmatism ( C4-2), spherical aberration ( C40), and vertical secondary astigmatism ( C42) were calculated. Differences between the two instruments for M, J0, and J45 were also calculated. The mean of the differences describes the bias. A t-test (α = 0.05) was used to compare each mean difference to zero. When a Shapiro-Wilk test indicated lack of normality, a non-parametric sign test was used instead. A Pearson correlation or, when a non-parametric test was appropriate, a Spearman correlation was used to assess the relationship between the differences and means for each plot. The 95% limits of agreement (LoA) were calculated for each comparison as the mean difference ± 1.96 × standard deviation of the differences. Exact 95% confidence intervals for the upper and lower limits of agreement were also calculated.25

Sample Size Calculation

When designing this study, the only previous repeatability study using the Discovery System reported the repeatability of internal aberrations (whole eye aberrations minus corneal aberrations) as opposed to whole eye aberrations.26 The largest between-visit standard deviation reported for internal eye refraction components (M, J0, and J45) was for defocus using a 4.5 mm pupil (0.24 D = 0.177 μm). Assuming a two-sided alpha of 0.05 and 80% power, a sample size of 8 subjects was needed to detect a 0.25 D difference in M, J0, and J45 (the greatest contributors of whole eye aberrations). We also repeated the calculation after inflating the standard deviation by 50% to 0.36 D to account for any potential increased variability in our study, which resulted in a sample size of 17 subjects. A total of 25 subjects were enrolled in this study.

RESULTS

Twenty-five adults between the ages of 22 and 29 completed both visits. The median time between Visit 1 and Visit 2 was 5 days (range: 1 to 14 days). The mean (±SD) age at Visit 1 was 23.4 ± 1.7 years. The mean (±SD) cycloplegic spherical-equivalent refractive error and cylinder as measured by the Grand Seiko WAM-5500 at Visit 2 were −2.91 ± 1.85 D (range: −6.96 to +0.74) and −0.54 ± 0.35 D (range: −1.38 to 0.00), respectively. Of the 25 subjects in the study, 18 (72%) were female.

Between-Visit Repeatability

The bias (mean difference between Visit 1 and Visit 2), repeatability of power vector components (M, J0, J45) measured with the Discovery System, and confidence intervals for the limits of agreement for a 6-mm and 3-mm pupil are shown in Table 1. There were no significant differences between visits for M, J0, or J45 (all p > 0.71). The difference versus mean plots for a 6-mm pupil for M, J0, J45 are shown in Figure 1. The 95% LoA (repeatability) of measurements of J0 and J45 were both less than a quarter of a diopter while the repeatability of the defocus term (M) was ±0.31 D for a 6-mm pupil. For a 3-mm pupil, similar results were obtained for J0 and J45. The repeatability of the defocus (M) increased to ±0.35 D. The bias, repeatability of the higher-order aberration measurements, and confidence intervals for the limits of agreement for a 6-mm pupil are shown in Table 2. The bias was not significantly different from zero for HORMS, third-, fourth-, fifth-, sixth-order RMS, vertical coma, horizontal coma, oblique secondary astigmatism, spherical aberration, or vertical secondary astigmatism (all p > 0.25). The Discovery System measured on average +0.022 microns more vertical trefoil and −0.014 microns less oblique trefoil at the first visit compared to the second visit (both p < 0.05); while statistically significant, these biases are very small clinically. Difference versus mean plots for higher-order aberrations are shown in Figure 2. Between-visit repeatability of HORMS was ±0.095 μm. For the spherical aberration term and the other RMS values calculated by radial order, repeatability ranged from ±0.029 to ±0.106 μm.

Table 1.

Between-visit repeatability of refractive error measurements (diopters) for the Discovery System

Bias* Repeatability** Upper LoA (CI)*** Lower LoA (CI)***
6mm pupil
M −0.01 ± 0.16 ± 0.31 0.29 (0.42,0.23) −0.32 (−0.26, −0.45)
J0 0.00 ± 0.07 ± 0.13 0.12 (0.18,0.10) −0.13 (−0.11, −0.19)
J45 0.00 ± 0.07 ± 0.14 0.14 (0.20,0.11) −0.13 (−0.10, −0.19)
3mm pupil
M −0.02 ± 0.04 ± 0.35 0.32 (0.08,0.03) −0.37 (−0.08, −0.12)
J0 −0.01 ± −0.09 ± 0.18 0.17 (0.24,0.13) −0.18 (−0.15, −0.26)
J45 0.02 ± 0.06 ± 0.12 0.14 (0.20,0.12) −0.10 (−0.08, −0.15)
*

Mean (±SD) difference between two visits (Visit 1 – Visit 2)

**

1.96 × SD of differences between visits (Limits of Agreement = LoA)

***

95% confidence interval (CI) is in parenthesis for the upper and the lower 95% LoA

Figure 1.

Figure 1

Between-visit repeatability of refractive error measurements using the Discovery System. Difference versus mean plots for repeated measurements of central spherical-equivalent defocus (A), J0 (B), J45 (C) measured at two separate visits for a 6-mm pupil. The solid line represents the mean difference between the two visits (bias). The dashed line represents the 95% limits of agreement.

Table 2.

Between-visit repeatability of higher-order aberration measurements (microns) for the Discovery System

Bias* Repeatability ** Upper LoA (CI)*** Lower LoA (CI)***
6mm
HORMS 0.002 ± 0.049 ± 0.095 0.098 (0.138,0.078) −0.093 (−0.073, −0.133)
Third-Order RMS 0.002 ± 0.054 ± 0.106 0.108 (0.152,0.086) −0.104 (−0.082, −0.149)
Fourth-Order RMS 0.006 ± 0.033 ± 0.064 0.070 (0.097,0.057) −0.057 (−0.044, −0.084)
Fifth-Order RMS 0.003 ± 0.016 ± 0.032 0.035 (0.048,0.028) −0.029 (−0.022, −0.042)
Sixth-Order RMS 0.003 ± 0.015 ± 0.029 0.032 (0.044,0.026) −0.026 (−0.020, −0.038)
C3−3 (Vertical Trefoil) 0.022 ± 0.045 ± 0.088 0.110 (0.147,0.092) −0.065 (−0.047, −0.102)
C3−1 (Vertical Coma) −0.007 ± 0.074 ± 0.146 0.140 (0.201,0.109) −0.153 (−0.123, −0.215)
C31 (Horizontal Coma) −0.002 ± 0.056 ± 0.109 0.108 (0.154,0.085) −0.111 (−0.088, −0.157)
C33 (Oblique Trefoil) −0.014 ± 0.028 ± 0.056 0.042 (0.065,0.030) −0.070 (−0.058, −0.093)
C4−2 (Oblique 2°Astig.) 0.002 ± 0.019 ± 0.038 0.040 (0.056,0.032) −0.036 (−0.028, −0.052)
C40 (Spherical Aberration) 0.005 ± 0.036 ± 0.070 0.074 (0.104,0.060) −0.065 (−0.050, −0.094)
C42 (Vertical 2° Astig.) −0.007 ± 0.028 ± 0.054 0.048 (0.071,0.036) −0.061 (−0.050, −0.084)
*

Mean (±SD) difference between two visits (Visit 1 – Visit 2)

**

1.96 × SD of differences between visits (Limits of Agreement = LoA)

***

95% confidence interval (CI) is in parenthesis for the upper and the lower 95% LoA

Figure 2.

Figure 2

Between-visit repeatability of higher-order aberration measurements (6-mm pupil diameter) using the Discovery System. Difference versus mean plots for repeated measurements of HORMS (A), third-order RMS (B), fourth-order RMS (C), fifth-order RMS (D), sixth-order RMS (E), and spherical aberration (F) measured at two separate visits. The solid line represents the mean difference between the two visits (bias). The dashed line represents the 95% limits of agreement.

Between-Instrument Agreement

The bias, agreement of cycloplegic refractive error components (M, J0, J45) between the Discovery System and Grand Seiko WAM-5500, and confidence intervals for the limits of agreement for a 6-mm and 3-mm pupil are shown in Table 3. For a 6-mm pupil, the analysis did not find any evidence indicating that there was systematic bias for measurement of defocus (M) between instruments (p = 0.45). For J0 and J45, the Discovery System on average measured 0.14 and 0.13 D higher than the Grand Seiko at a 6-mm pupil, respectively (both p < 0.001). However, for a 3-mm pupil which is more similar to the measurement beam diameter of the Grand Seiko, the Discovery System on average measured more positive values that the Grand Seiko for defocus (0.28 D), J0 (0.11 D) and J45 (0.12 D; all p <0.005). The difference versus mean plots for refractive error measurements for a 3-mm pupil are shown in Figure 3. The 95% LoA between instruments was greater for defocus than for the astigmatism terms.

Table 3.

Refractive error agreement (diopters) between the Discovery System and the Grand Seiko WAM-5500

Bias* Repeatability ** Upper LoA (CI)*** Lower LoA (CI)***
6mm pupil
M 0.07 ± 0.43 ± 0.85 0.91 (1.27,0.74) −0.78 (−0.60, −1.14)
J0 0.14 ± 0.16 ± 0.31 0.46 (0.59,0.39) −0.17 (−0.10, −0.30)
J45 0.13 ± 0.15 ± 0.29 0.41 (0.54,0.36) −0.16 (−0.10, −0.28)
3mm pupil
M 0.28 ± 0.39 ± 0.77 1.05 (1.37,0.89) −0.49 (−0.33, −0.82)
J0 0.11 ± 0.17 ± 0.34 0.45 (0.60,0.38) −0.23 (−0.16, −0.37)
J45 0.12 ± 0.15 ± 0.30 0.42 (0.55,0.35) −0.19 (−0.12, −0.31)
*

Mean (±SD) difference between the two instruments (Discovery System - Grand Seiko)

**

1.96 × SD of differences between the two instruments (Limits of Agreement = LoA)

***

95% confidence interval (CI) is in parenthesis for the upper and the lower 95% LoA

Figure 3.

Figure 3

Refractive error agreement between the Discovery System (for a 3-mm pupil) and the Grand Seiko WAM-5500. Difference versus mean plots for repeated measurements of central spherical-equivalent defocus (A), J0 (B), J45 (C) measured with each instrument. The solid line represents the mean difference between the two instruments (bias). The dashed line represents the 95% limits of agreement.

DISCUSSION

Determining the repeatability of instruments to be use in clinical research under normal conditions to establish its validity in terms of measurement limitations is essential. Because of the increasing public health risk associated with myopia, many clinical research studies are evaluating possible therapeutic devices for slowing the rate of myopia progression in children.1, 3, 27 These studies require the longitudinal assessment of refractive error changes, which necessitates clear knowledge of instrument validity and repeatability in children. Additionally, aberrometry measurements are commonly made when there is an interest in image quality, such as in studies evaluating refractive error, contact lenses for myopia control, and corneal diseases such as keratoconus. Higher-order aberrations are significantly increased in patients with keratoconus and degrade the retinal image in these patients.2, 28 Studies utilizing aberrometry are being done to design custom treatments for keratoconus.29 Objective measurements of image quality are useful in a variety of clinical research applications but require a robust instrument that is research-friendly and repeatable.

Refractive Error Repeatability and Agreement

Several studies have evaluated the repeatability of non-cycloplegic refractive error measurements using Grand Seiko autorefractors and have found the repeatability to be between ±0.50 D to ±0.86 D.68, 1012 As expected, better repeatability has been reported under cycloplegic conditions (range: ±0.12 D to ±0.48 D). In our study, the repeatability of the Discovery System (±0.34 D) was found to be comparable to the repeatability of the Grand Seiko autorefractor under cycloplegic conditions in adult subjects.6, 9, 12

The Grand Seiko obtains measurements using a 2.3-mm measurement beam. In contrast, the minimum pupil diameter for which the Discovery System displays refractive error data (sphere, cylinder, and axis) is 3 mm. Therefore, an analysis of measurement agreement between the two instruments was conducted for both a 6-mm pupil and a 3-mm pupil. With a 3-mm pupil, all power vector components of refractive error measurements (M, J0, and J45) were statistically different between the two instruments, with the Discovery System measuring +0.28 D more defocus than the Grand Seiko. By contrast, this study failed to find any evidence of bias for spherical-equivalent defocus (M) between the Discovery System across a 6-mm pupil and Grand Seiko. While both instruments are very repeatable, the between-instrument 95% LoA of ±0.77 D indicate that the use of the Discovery System interchangeably with the Grand Seiko for defocus measurements is not advisable. The bias between the two instruments for J0 and J45 astigmatism was statistically significant with the Discovery System measuring slightly more of each across a 3-mm pupil (0.11 D and 0.12 D, respectively). Although these differences are clinically small, the differences indicate that the astigmatic components are not interchangeable between the instruments. That being said, the repeatability of astigmatism measurements on the Discovery System across visits was very good (J0: ±0.13 D; J45: ±0.14 D).

Repeatability of Higher-order Aberrations

In addition to lower-order aberrations (refractive error), higher-order aberrations also influence retinal image quality. Although not correctable with traditional ophthalmic corrections like glasses or contact lenses, higher-order aberrations can have a meaningful influence on vision, especially in conditions with elevated higher-order aberrations such as keratoconus.3033 In patients treated with laser-assisted in situ keratomileusis (LASIK), common post-treatment visual complaints are glare and halos, which are correlated with increases in post-treatment higher-order aberrations.3438 Optical corrections currently being studied for myopia control like orthokeratology increase higher-order aberrations;39 therefore, measuring aberrations with these treatments is important when evaluating their influence on image quality. Repeatable clinical instruments for measuring changes in higher-order aberrations due to refractive procedures and contact lenses are important.

The Discovery System showed very repeatable results for measures of higher-order aberrations (HORMS 95% LoA = ±0.095 μm) that were similar to a previous study that assessed the repeatability of HORMS measurements in an adult population made at different visits on the same day using the COAS aberrometer (95% LoA = ±0.132 μm).40 Differences in HORMS that lie outside the 95% LoA have a 95% probability of being a true change in HORMS between visits and 5% probability of being due to instrument measurement error. Therefore, a difference in HORMS between visits that is greater than 0.095 microns will have a 95% probability of being a true change in HORMS.

In evaluating the Discovery System’s ability to detect a change in both normal and abnormal populations, we reviewed the normal aberration levels in different populations reported in the literature. A study by Salmon and van de Pol established normal-eye RMS values.41 The mean values for HORMS, third-, fourth-, fifth-, sixth-order RMS, and spherical aberration in our study were all within one standard deviation of their study. The mean (±SD) HORMS in our study was 0.431 ± 0.123 microns compared to the mean HORMS of 0.327 ± 0.130 microns found in their study of subjects between the ages of 20 and 70.41 The mean HORMS in our study was also within one standard deviation of the HORMS reported by Karimian et al. (0.35 ± 0.12 microns) in a study of normal myopic subjects with an age range (18–43 years) that more closely resembles our subjects.42 In contrast, eyes with abnormal corneal conditions like keratoconus have higher than average HORMS. A study in patients with keratoconus fitted with a spherical-equivalent scleral lens measured between 0.20 to 0.90 microns of residual HORMS. Ten out of the 14 eyes measured had residual HORMS values greater than 0.4 microns.29 Changes in HORMS after procedures such as orthokeratology are generally on the order of more than 0.15 to 0.20 microns.39, 43 Additionally, in a normal population, the average change in HORMS seen with age from 20 year olds to 50 years olds is 0.109 microns for a 6mm pupil.44 Based on the repeatability found in this study, the Discovery System is capable of detecting 0.10 micron changes in HORMS in adult eyes under cycloplegia.

The Discovery System was also repeatable when HORMS was decomposed to assess each radial order individually. One previous study found good repeatability for measurements of spherical aberration using the Discovery System (intra-session 95% LoA = ±0.083 μm; between-session 95% LoA = ±0.051 μm); however, that study evaluated the repeatability of internal aberrations.26 In our study, we evaluated total ocular aberrations between measurements made on different days and found the between-visit repeatability to be similar (95% LoA = ±0.070 μm).

To put RMS wavefront error into more clinically meaningful terms, RMS can be converted from microns to equivalent diopters of defocus using a previously published formula.45 The repeatability for the HORMS (±0.095 μm) is equivalent to ±0.07 D of defocus, and the repeatability for the spherical aberration term (±0.070 μm) is equivalent to ±0.05 D of defocus, both of which are clinically small.

A potential limitation of our study is that we did not investigate the repeatability of the instrument under non-cycloplegic conditions. The purpose of the study was to assess the repeatability of the instrument; therefore, controlling the potential impact of accommodation was essential. Repeatability would likely be worse under non-cycloplegic conditions, which would be consistent with repeatability studies of autorefraction.12 Another limitation is that repeatability was assessed in a normal population. It is possible that the repeatability of the instrument could be different in highly-aberrated eyes or in children. However, a review of studies assessing the repeatability of autorefraction in children using instruments with a similar point-and-click capture method to the Discovery suggests similar repeatability to an adult population.46, 47 A study evaluating repeatability of the Shin Nippon SRW-5000 in children conducted by Chat et al. found differences in repeatability between children that were cyclopleged versus those that were not (Defocus 95% LoA cyclopleged: ±0.35 D vs non-cyclopleged: ±0.45 D). The study suggests that the cycloplegic state of the subject is a critical variable that affects repeatability.46 The repeatability in non-cyclopleged children in their study is well within the range of repeatability found across studies in non-cyclopleged adults using the Grand Seiko/Shin Nippon autorefractor (95% LoA range: ±0.43 to ± 0.86 D).7, 8, 10, 48 Additionally, it is possible that repeatability could be influenced by highly-aberrated eyes or when measuring refractive errors outside the range of those included in the current study. Overall, although studies of autorefraction do not show differences in repeatability between children and adults, future studies assessing the Discovery System’s validity should focus on assessing the instrument’s repeatability in highly-aberrated eyes and in children.

In conclusion, the Discovery System was found to be very repeatable. The instrument allows users to measure ocular aberrations and export the full Zernike data, making it a promising research instrument. Though, on average, there were minimal differences in refractive error measured between the Grand Seiko and Discovery, the 95% LoA for refractive error measurements between instruments suggest that one should not use refractive error values interchangeably between these instruments. The good repeatability of the Discovery System when measuring both refractive error and higher-order aberrations indicates that it would be an appropriate instrument to track changes in refractive error and visual quality in longitudinal studies in adults.

Acknowledgments

Grant support: NIH/NEI T35-EY07088

These data were presented, in part, at the 2016 Association for Research in Vision and Ophthalmology (ARVO) annual meeting in Seattle, WA.

References

  • 1.Cheng X, Xu J, Chehab K, et al. Soft contact lenses with positive spherical aberration for myopia control. Optom Vis Sci. 2016;93:353–66. doi: 10.1097/OPX.0000000000000773. [DOI] [PubMed] [Google Scholar]
  • 2.Colak HN, Kantarci FA, Yildirim A, et al. Comparison of corneal topographic measurements and high order aberrations in keratoconus and normal eyes. Cont Lens Anterior Eye. 2016;39:380–4. doi: 10.1016/j.clae.2016.06.005. [DOI] [PubMed] [Google Scholar]
  • 3.Khan MS, Humayun S, Fawad A, et al. Comparison of higher order aberrations in patients with various refractive errors. Pak J Med Sci. 2015;31:812–5. doi: 10.12669/pjms.314.7538. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Ortiz-Toquero S, Rodriguez G, de Juan V, et al. Repeatability of wavefront aberration measurements with a placido-based topographer in normal and keratoconic eyes. J Refract Surg. 2016;32:338–44. doi: 10.3928/1081597X-20160121-04. [DOI] [PubMed] [Google Scholar]
  • 5.Paune J, Thivent S, Armengol J, et al. Changes in peripheral refraction, higher-order aberrations, and accommodative lag with a radial refractive gradient contact lens in young myopes. Eye Contact Lens. 2016;42:380–7. doi: 10.1097/ICL.0000000000000222. [DOI] [PubMed] [Google Scholar]
  • 6.Bailey MD, Twa MD, Mitchell GL, et al. Repeatability of autorefraction and axial length measurements after laser in situ keratomileusis. J Cataract Refract Surg. 2005;31:1025–34. doi: 10.1016/j.jcrs.2004.12.040. [DOI] [PubMed] [Google Scholar]
  • 7.Cleary G, Spalton DJ, Patel PM, et al. Diagnostic accuracy and variability of autorefraction by the Tracey Visual Function Analyzer and the Shin-Nippon NVision-K 5001 in relation to subjective refraction. Ophthalmic Physiol Opt. 2009;29:173–81. doi: 10.1111/j.1475-1313.2008.00627.x. [DOI] [PubMed] [Google Scholar]
  • 8.Davies LN, Mallen EA, Wolffsohn JS, et al. Clinical evaluation of the Shin-Nippon NVision-K 5001/Grand Seiko WR-5100K autorefractor. Optom Vis Sci. 2003;80:320–4. doi: 10.1097/00006324-200304000-00011. [DOI] [PubMed] [Google Scholar]
  • 9.Moore KE, Berntsen DA. Central and peripheral autorefraction repeatability in normal eyes. Optom Vis Sci. 2014;91:1106–12. doi: 10.1097/OPX.0000000000000351. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Sheppard AL, Davies LN. Clinical evaluation of the Grand Seiko Auto Ref/Keratometer WAM-5500. Ophthalmic Physiol Opt. 2010;30:143–51. doi: 10.1111/j.1475-1313.2009.00701.x. [DOI] [PubMed] [Google Scholar]
  • 11.Wolffsohn JS, Davies LN, Naroo SA, et al. Evaluation of an open-field autorefractor’s ability to measure refraction and hence potential to assess objective accommodation in pseudophakes. Br J Ophthalmol. 2011;95:498–501. doi: 10.1136/bjo.2010.185009. [DOI] [PubMed] [Google Scholar]
  • 12.Zadnik K, Mutti DO, Adams AJ. The repeatability of measurement of the ocular components. Invest Ophthalmol Vis Sci. 1992;33:2325–33. [PubMed] [Google Scholar]
  • 13.Liang J, Grimm B, Goelz S, et al. Objective measurement of wave aberrations of the human eye with the use of a Hartmann-Shack wave-front sensor. J Opt Soc Am (A) 1994;11:1949–57. doi: 10.1364/josaa.11.001949. [DOI] [PubMed] [Google Scholar]
  • 14.Thibos LN. Principles of Hartmann-Shack aberrometry. J Refract Surg. 2000;16:S563–5. doi: 10.3928/1081-597X-20000901-14. [DOI] [PubMed] [Google Scholar]
  • 15.Cheng X, Himebaugh NL, Kollbaum PS, et al. Validation of a clinical Shack-Hartmann aberrometer. Optom Vis Sci. 2003;80:587–95. doi: 10.1097/00006324-200308000-00013. [DOI] [PubMed] [Google Scholar]
  • 16.Cheng X, Himebaugh NL, Kollbaum PS, et al. Test-retest reliability of clinical Shack-Hartmann measurements. Invest Ophthalmol Vis Sci. 2004;45:351–60. doi: 10.1167/iovs.03-0265. [DOI] [PubMed] [Google Scholar]
  • 17.Berntsen DA, Mutti DO, Zadnik K. Validation of aberrometry-based relative peripheral refraction measurements. Ophthalmic Physiol Opt. 2008;28:83–90. doi: 10.1111/j.1475-1313.2007.00535.x. [DOI] [PubMed] [Google Scholar]
  • 18.Salmon TO, West RW, Gasser W, et al. Measurement of refractive errors in young myopes using the COAS Shack-Hartmann aberrometer. Optom Vis Sci. 2003;80:6–14. doi: 10.1097/00006324-200301000-00003. [DOI] [PubMed] [Google Scholar]
  • 19.Salmon TO, Thibos LN, Bradley A. Comparison of the eye’s wave-front aberration measured psychophysically and with the Shack-Hartmann wave-front sensor. J Opt Soc Am (A) 1998;15:2457–65. doi: 10.1364/josaa.15.002457. [DOI] [PubMed] [Google Scholar]
  • 20.Egashira SM, Kish LL, Twelker JD, et al. Comparison of cyclopentolate versus tropicamide cycloplegia in children. Optom Vis Sci. 1993;70:1019–26. doi: 10.1097/00006324-199312000-00005. [DOI] [PubMed] [Google Scholar]
  • 21.Manny RE, Hussein M, Scheiman M, et al. Tropicamide (1%): an effective cycloplegic agent for myopic children. Invest Ophthalmol Vis Sci. 2001;42:1728–35. [PubMed] [Google Scholar]
  • 22.Fedtke C, Ehrmann K, Holden BA. A review of peripheral refraction techniques. Optom Vis Sci. 2009;86:429–46. doi: 10.1097/OPX.0b013e31819fa727. [DOI] [PubMed] [Google Scholar]
  • 23.Thibos LN, Wheeler W, Horner D. Power vectors: an application of Fourier analysis to the description and statistical analysis of refractive error. Optom Vis Sci. 1997;74:367–75. doi: 10.1097/00006324-199706000-00019. [DOI] [PubMed] [Google Scholar]
  • 24.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10. [PubMed] [Google Scholar]
  • 25.Carkeet A. Exact parametric confidence intervals for Bland-Altman limits of agreement. Optom Vis Sci. 2015;92:e71–80. doi: 10.1097/OPX.0000000000000513. [DOI] [PubMed] [Google Scholar]
  • 26.Gifford P, Swarbrick HA. Repeatability of internal aberrometry with a new simultaneous capture aberrometer/corneal topographer. Optom Vis Sci. 2012;89:929–38. doi: 10.1097/OPX.0b013e31825017c4. [DOI] [PubMed] [Google Scholar]
  • 27.Paune J, Thivent S, Armengol J, et al. Changes in peripheral refraction, higher-order aberrations, and accommodative lag with a radial refractive gradient contact lens in young myopes. Eye Contact Lens. 2016;42:380–7. doi: 10.1097/ICL.0000000000000222. [DOI] [PubMed] [Google Scholar]
  • 28.Delgado S, Velazco J, Delgado Pelayo RM, et al. Correlation of higher order aberrations in the anterior corneal surface and degree of keratoconus measured with a Scheimpflug camera. Arch Soc Esp Oftalmol. 2016;91:316–9. doi: 10.1016/j.oftal.2016.01.014. [DOI] [PubMed] [Google Scholar]
  • 29.Marsack JD, Ravikumar A, Nguyen C, et al. Wavefront-guided scleral lens correction in keratoconus. Optom Vis Sci. 2014;91:1221–30. doi: 10.1097/OPX.0000000000000275. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Cheng X, Bradley A, Ravikumar S, et al. Visual impact of Zernike and Seidel forms of monochromatic aberrations. Optom Vis Sci. 2010;87:300–12. doi: 10.1097/OPX.0b013e3181d95217. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Kingston AC, Cox IG. Population spherical aberration: associations with ametropia, age, corneal curvature, and image quality. Clin Ophthalmol. 2013;7:933–8. doi: 10.2147/OPTH.S44056. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Villegas EA, Alcon E, Artal P. Impact of positive coupling of the eye’s trefoil and coma in retinal image quality and visual acuity. J Opt Soc Am (A) 2012;29:1667–72. doi: 10.1364/JOSAA.29.001667. [DOI] [PubMed] [Google Scholar]
  • 33.Xu R, Bradley A, Lopez Gil N, et al. Modelling the effects of secondary spherical aberration on refractive error, image quality and depth of focus. Ophthalmic Physiol Opt. 2015;35:28–38. doi: 10.1111/opo.12185. [DOI] [PubMed] [Google Scholar]
  • 34.Al-Zeraid FM, Osuagwu UL. Induced higher-order aberrations after laser in situ keratomileusis (LASIK) performed with wavefront-guided intralase femtosecond laser in moderate to high astigmatism. BMC Ophthalmol. 2016;16:29. doi: 10.1186/s12886-016-0205-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Khan MS, Humayun S, Fawad A, et al. Effect of wavefront optimized LASIK on higher order aberrations in myopic patients. Pak J Med Sci. 2015;31:1223–6. doi: 10.12669/pjms.315.7683. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Liu TX, Chen YT, Dan TT, et al. Four-year follow-up of corneal aberrations and visual functions of myopic patients after laser in situ keratomileusis. Pak J Med Sci. 2015;31:1453–6. doi: 10.12669/pjms.316.8338. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Ganesh S, Gupta R. Comparison of visual and refractive outcomes following femtosecond laser- assisted lasik with smile in patients with myopia or myopic astigmatism. J Refract Surg. 2014;30:590–6. doi: 10.3928/1081597X-20140814-02. [DOI] [PubMed] [Google Scholar]
  • 38.Kung JS, Manche EE. Quality of vision after wavefront-guided or wavefront-optimized LASIK: A prospective randomized contralateral eye study. J Refract Surg. 2016;32:230–6. doi: 10.3928/1081597X-20151230-01. [DOI] [PubMed] [Google Scholar]
  • 39.Berntsen DA, Barr JT, Mitchell GL. The effect of overnight contact lens corneal reshaping on higher-order aberrations and best-corrected visual acuity. Optom Vis Sci. 2005;82:490–7. doi: 10.1097/01.opx.0000168586.36165.bb. [DOI] [PubMed] [Google Scholar]
  • 40.Berntsen DA, BJT, Mitchell GL. Repeatability of higher-order aberration measurements using the complete ophthalmic analysis system (COAS) Optom Vis Sci. 2003;80:117. [Google Scholar]
  • 41.Salmon TO, van de Pol C. Normal-eye Zernike coefficients and root-mean-square wavefront errors. J Cataract Refract Surg. 2006;32:2064–74. doi: 10.1016/j.jcrs.2006.07.022. [DOI] [PubMed] [Google Scholar]
  • 42.Karimian F, Feizi S, Doozande A. Higher-order aberrations in myopic eyes. J Ophthalmic Vis Res. 2010;5:3–9. [PMC free article] [PubMed] [Google Scholar]
  • 43.Hiraoka T, Okamoto C, Ishii Y, et al. Time course of changes in ocular higher-order aberrations and contrast sensitivity after overnight orthokeratology. Invest Ophthalmol Vis Sci. 2008;49:4314–20. doi: 10.1167/iovs.07-1586. [DOI] [PubMed] [Google Scholar]
  • 44.Applegate RA, Donnelly WJ, 3rd, Marsack JD, et al. Three-dimensional relationship between high-order root-mean-square wavefront error, pupil diameter, and aging. J Opt Soc Am (A) 2007;24:578–87. doi: 10.1364/josaa.24.000578. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Thibos LN, Hong X, Bradley A, et al. Statistical variation of aberration structure and image quality in a normal population of healthy eyes. J Opt Soc Am (A) 2002;19:2329–48. doi: 10.1364/josaa.19.002329. [DOI] [PubMed] [Google Scholar]
  • 46.Chat SW, Edwards MH. Clinical evaluation of the Shin-Nippon SRW-5000 autorefractor in children. Ophthalmic Physiol Opt. 2001;21:87–100. [PubMed] [Google Scholar]
  • 47.Lee TT, Cho P. Repeatability of relative peripheral refraction in untreated and orthokeratology-treated eyes. Optom Vis Sci. 2012;89:1477–86. doi: 10.1097/OPX.0b013e31826912cd. [DOI] [PubMed] [Google Scholar]
  • 48.Mallen EA, Wolffsohn JS, Gilmartin B, et al. Clinical evaluation of the Shin-Nippon SRW-5000 autorefractor in adults. Ophthalmic Physiol Opt. 2001;21:101–7. [PubMed] [Google Scholar]

RESOURCES