Skip to main content
Journal of Synchrotron Radiation logoLink to Journal of Synchrotron Radiation
. 2017 Aug 8;24(Pt 5):1078–1081. doi: 10.1107/S1600577517009869

Strategies for high-throughput focused-beam ptychography

Chris Jacobsen a,b,c,*, Junjing Deng a, Youssef Nashed d,
PMCID: PMC5580791  PMID: 28862631

X-ray ptychography is being utilized for a wide range of imaging experiments with a resolution beyond the limit of the X-ray optics used. Here, strategies for data sampling and for increasing imaging throughput when the specimen is at the focus of an X-ray beam are discussed, and the tradeoffs between large and small illumination spots are examined.

Keywords: ptychography, ptychographic resolution gain, high throughput

Abstract

X-ray ptychography is being utilized for a wide range of imaging experiments with a resolution beyond the limit of the X-ray optics used. Introducing a parameter for the ptychographic resolution gain G p (the ratio of the beam size over the achieved pixel size in the reconstructed image), strategies for data sampling and for increasing imaging throughput when the specimen is at the focus of an X-ray beam are considered. The tradeoffs between large and small illumination spots are examined.

1. Introduction  

Ptychography (Hoppe, 1969) involves the use of overlapping coherent illumination regions on a specimen and the collection of diffraction data from each illumination spot, followed by reconstruction of an image. Following initial experimental demonstrations of ptychography (Rodenburg et al., 2007; Thibault et al., 2008) and related methods (Chapman, 1996), X-ray ptychography is finding increased utilization in X-ray microscopy because it can be used to deliver amplitude and phase-contrast images beyond the resolution limit of the coherent beam size. It does so without the small isolated specimen limitations of X-ray coherent diffraction imaging (Miao et al., 1999) which are intrinsic to the use of finite support iterative phase retrieval (Fienup, 1978) unless one uses a spatially restricted coherent beam to satisfy the support constraint (Abbey et al., 2008) (which begins to look like ptychography if one scans the beam).

2. Discussion  

Because ptychography requires mostly coherent beams, and involves point-by-point scanning, it is usually regarded as a low-throughput imaging method even though several higher throughput examples exist (Guizar-Sicairos et al., 2014; Holler et al., 2014). We consider here the factors that can be used to increase that throughput, characterizing them in terms of the ptychographic resolution gain Inline graphic which we define as the ratio between the desired reconstructed image pixel size1 δ and the diameter d of the coherent beam spot that is scanned across the specimen, or

2.

In order to have the possibility of a square half-period pixel size of δ using an illumination wavelength λ, Fig. 1 shows that one must record first-order diffraction out to an angle (in the small-angle approximation of Inline graphic) of

2.

This requires coherent superposition between waves with a path length difference from the top and bottom edges of the illumination spot to a distant detector of Inline graphic, or a number m of wavelengths of longitudinal coherence (Spence et al., 2004; van der Veen & Pfeiffer, 2004; Enders et al., 2014) of

2.

where we have again used the small-angle approximation. Therefore the spectral bandwidth should be

2.

Illumination from the coherent illumination spot of diameter d will diffract out by a semi-angle2 of Inline graphic = Inline graphic to cover a diameter D on the detector of

2.

The signal from small features within the illumination spot is recorded on a detector with N pixels (each of width Δ) on a side, so that it subtends a semi-angle of

2.

The illumination spot will therefore be spread out over a number of detector pixels Inline graphic of

2.

where the final expression uses equations (1), (2) and (5). Finally, because the last two pixels on the detector must be able to record fringes caused by interference from points at the top and bottom of the illumination spot d with Nyquist sampling, we must have 2m pixels from the detector center to the edge or 4m pixels overall in each dimension. In other words, we find that the minimum number of detector pixels Inline graphic is given by

2.

where we have used equation (3). We note that ptychography can be performed with reduced detector sampling, though at a cost of image fidelity (Edo et al., 2013) or of finer real-space sampling (Batey et al., 2014; da Silva & Menzel, 2015) which effectively corresponds to larger values of the overlap factor o discussed below.

Figure 1.

Figure 1

Data sampling in ptychography. The specimen is illuminated by beam spots of diameter d with divergence Inline graphic, leading to an incident illumination diameter D on the detector. The desired half-period pixel width δ (the limit of resolution) implies a diffraction numerical aperture Inline graphic out to the edges of the detector.

Achieving high resolution in ptychography requires the detection of scattering out to large angles, which is determined in part by the specimen’s optical properties at the chosen X-ray wavelength and the number of photons Inline graphic used to illuminate each pixel (Glaeser, 1971; Sayre et al., 1977; Schropp & Schroer, 2010). While high resolution can also be aided by having high-spatial-frequency content in the illuminating beam if a larger focal spot (larger value of Inline graphic) is used (Guizar-Sicairos et al., 2012), we assume that the dominating factor is the required photon density per area, Inline graphic = Inline graphic, independent of the ptychographic spatial resolution gain Inline graphic. If a is the center-to-center position increment between illumination spots of diameter d, it is recommended to use an overlap factor Inline graphic = Inline graphic with Inline graphic ≃ 0.6 for robust reconstructions (Bunk et al., 2008) [this parameter has also been used in connection with continuous scan ptychography (Deng et al., 2015a )]. Imaging of a square area A then requires the use of a number Inline graphic of illumination spots given by

2.

if one ignores incomplete illumination at the edges. Imaging the area A then requires a net photon count Inline graphic of

2.

where we have used equation (9) to arrive at the final result. If the source delivers a coherent flux of Inline graphic photons per second, the exposure time Inline graphic for each of the Inline graphic illumination spots (Inline graphic equals the transit time per distance a in the case of continuous scanning) can be found from Inline graphic = Inline graphic, leading to an exposure time per illumination spot of

2.

Because the coherent flux Inline graphic is an intrinsic property of the source [with the caveat that one can obtain higher coherent flux from a broad-band source as indicated by equation (4)], it is unchanged by the size d of the spot into which the coherent flux is delivered; that is, Inline graphic does not depend on the ptychographic spatial resolution gain Inline graphic.

From the above, we see that decreasing the ptychographic spatial resolution gain Inline graphic (that is, using smaller coherent illumination spots d) has several implications:

(i) Decreasing Inline graphic means detectors with fewer pixels Inline graphic can be used, as can be seen directly from equation (8).

(ii) Decreasing Inline graphic leads to relaxed requirements for the monochromaticity Inline graphic in the illumination, as can be seen from equation (4). When using large-bandwidth X-ray sources, this could allow one to use multilayer monochromators with bandpass Inline graphicInline graphic as compared with crystal monochromators with bandpass Inline graphicInline graphic, thus leading to a usable flux gain increase of Inline graphic. This has already been demonstrated as a route to increased throughput in ptychography (Enders et al., 2014).

(iii) Decreasing the gain Inline graphic has the effect of increasing the fractional number of detector pixels within which the incident beam is recorded, as can be seen from equation (7) divided by Inline graphic. Since the incident beam is often much stronger than the fraction of the beam scattered by the sample, putting the incident beam into a larger fraction of detector pixels reduces demands on the dynamic range of the detector. This is in contrast to the case of far-field coherent diffraction imaging of weakly scattering objects like biological cells, where the dynamic range required of the detector can be in excess of Inline graphic:1 so that multiple detector exposures must be acquired with various exposure times and direct-beam absorber positions (Shapiro et al., 2005).

(iv) When smaller illumination spots d are used with smaller ptychographic gain Inline graphic, one has the option of simultaneously acquiring high-spatial-resolution scanning microscope images using other contrast modes, such as fluorescence from elemental content (Schropp et al., 2012; Deng et al., 2015b , 2017).

(v) Decreasing Inline graphic is associated with shorter exposure times per diffraction pattern recording, as can be see from equation (11). This means that higher detector frame rates are required if one uses smaller values of Inline graphic. However, the net ‘information rate’ to be transferred by the detector of total pixels per second, or Inline graphic, is not affected as can be seen by the fact that both Inline graphic [from the square of equation (8)] and the exposure time Inline graphic [from equation (11)] depend on Inline graphic, thus canceling out any dependence on Inline graphic. In other words, the total amount of data to be saved or transferred is the same in optimized experiments, independent of the ptychographic gain Inline graphic. We have not accounted for factors such as thermal noise or amplifier readout noise which might be present in charge-integrating detectors versus photon-counting detectors.

(vi) Finally, if there is an overhead time associated with the collection of signal from each of the Inline graphic illumination spots, reducing Inline graphic and thus increasing Inline graphic [equation (9)] will lead to longer scan times. However, continuous-scan methods largely remove this overhead time between pixels (Pelz et al., 2014; Deng et al., 2015a ); in raster scans, the overhead between scan lines depends both on the step distance (favoring small Inline graphic) and data transfer overheads (favoring large Inline graphic).

These advantages of using small illumination spots or small values of ptychographic spatial resolution gain Inline graphic can be compelling for certain applications. Finally, we note that increased experimental throughput should be coupled with fast computing; one example is to use efficient parallelization schemes for ptychographic image reconstruction (Nashed et al., 2014).

Acknowledgments

This work was supported by the US Department of Energy (DOE) Office of Science under Contract No. DE-AC02-06CH11357, and by the National Institutes for Health under grant R01 GM104530. We thank John Damoulakis and A. J. F. Levi for discussions regarding some applications for which higher throughput would be desired in X-ray ptychography.

Footnotes

1

Of course, the achieved image resolution might be poorer than the reconstructed image pixel size δ due to factors such as illumination strength and intrinsic specimen scattering as will be noted below.

2

If one instead assumes a circular focusing optic with a numerical aperture of Inline graphic, then d refers to the diameter of the Airy disk of the focus; the numerical factor of 1.22 which is slightly modified depending on central stops, or non-circular optics, plays only a minor role in the discussion that follows.

References

  1. Abbey, B., Nugent, K. A., Williams, G. J., Clark, J. N., Peele, A. G., Pfeifer, M. A., de Jonge, M. & McNulty, I. (2008). Nat. Phys. 4, 394–398.
  2. Batey, D. J., Edo, T. B., Rau, C., Wagner, U., Pešić, Z. D., Waigh, T. A. & Rodenburg, J. M. (2014). Phys. Rev. A, 89, 043812.
  3. Bunk, O., Dierolf, M., Kynde, S., Johnson, I., Marti, O. & Pfeiffer, F. (2008). Ultramicroscopy, 108, 481–487. [DOI] [PubMed]
  4. Chapman, H. N. (1996). Ultramicroscopy, 66, 153–172.
  5. Deng, J., Nashed, Y. S. G., Chen, S., Phillips, N. W., Peterka, T., Ross, R., Vogt, S., Jacobsen, C. & Vine, D. J. (2015a). Opt. Express, 23, 5438–5451. [DOI] [PMC free article] [PubMed]
  6. Deng, J., Vine, D. J., Chen, S., Jin, Q., Nashed, Y. S. G., Peterka, T., Vogt, S. & Jacobsen, C. (2017). Sci. Rep. 7, 445. [DOI] [PMC free article] [PubMed]
  7. Deng, J., Vine, D. J., Chen, S., Nashed, Y. S. G., Jin, Q., Phillips, N. W., Peterka, T., Ross, R., Vogt, S. & Jacobsen, C. J. (2015b). Proc. Natl Acad. Sci. USA, 112, 2314–2319. [DOI] [PMC free article] [PubMed]
  8. Edo, T. B., Batey, D. J., Maiden, A. M., Rau, C., Wagner, U., Pešić, Z. D., Waigh, T. A. & Rodenburg, J. M. (2013). Phys. Rev. A, 87, 053850.
  9. Enders, B., Dierolf, M., Cloetens, P., Stockmar, M., Pfeiffer, F. & Thibault, P. (2014). Appl. Phys. Lett. 104, 171104.
  10. Fienup, J. (1978). Opt. Lett. 3, 27–29. [DOI] [PubMed]
  11. Glaeser, R. M. (1971). J. Ultrastruct. Res. 36, 466–482. [DOI] [PubMed]
  12. Guizar-Sicairos, M., Holler, M., Diaz, A., Vila-Comamala, J., Bunk, O. & Menzel, A. (2012). Phys. Rev. B, 86, 100103.
  13. Guizar-Sicairos, M., Johnson, I., Diaz, A., Holler, M., Karvinen, P., Stadler, H.-C., Dinapoli, R., Bunk, O. & Menzel, A. (2014). Opt. Express, 22, 14859–14870. [DOI] [PubMed]
  14. Holler, M., Diaz, A., Guizar-Sicairos, M., Karvinen, P., Färm, E., Härkönen, E., Ritala, M., Menzel, A., Raabe, J. & Bunk, O. (2014). Sci. Rep. 4, 3857. [DOI] [PMC free article] [PubMed]
  15. Hoppe, W. (1969). Acta Cryst. A25, 495–501.
  16. Miao, J., Charalambous, P., Kirz, J. & Sayre, D. (1999). Nature (London), 400, 342–344.
  17. Nashed, Y. S., Vine, D. J., Peterka, T., Deng, J., Ross, R. & Jacobsen, C. (2014). Opt. Express, 22, 32082. [DOI] [PMC free article] [PubMed]
  18. Pelz, P. M., Guizar-Sicairos, M., Thibault, P., Johnson, I., Holler, M. & Menzel, A. (2014). Appl. Phys. Lett. 105, 251101.
  19. Rodenburg, J., Hurst, A., Cullis, A., Dobson, B., Pfeiffer, F., Bunk, O., David, C., Jefimovs, K. & Johnson, I. (2007). Phys. Rev. Lett. 98, 034801. [DOI] [PubMed]
  20. Sayre, D., Kirz, J., Feder, R., Kim, D. M. & Spiller, E. (1977). Ultramicroscopy, 2, 337–349. [DOI] [PubMed]
  21. Schropp, A., Hoppe, R., Patommel, J., Samberg, D., Seiboth, F., Stephan, S., Wellenreuther, G., Falkenberg, G. & Schroer, C. G. (2012). Appl. Phys. Lett. 100, 253112.
  22. Schropp, A. & Schroer, C. G. (2010). New J. Phys. 12, 035016.
  23. Shapiro, D., Thibault, P., Beetz, T., Elser, V., Howells, M., Jacobsen, C., Kirz, J., Lima, E., Miao, H., Neiman, A. M. & Sayre, D. (2005). Proc. Natl Acad. Sci. 102, 15343–15346. [DOI] [PMC free article] [PubMed]
  24. Silva, J. C. da & Menzel, A. (2015). Opt. Express, 23, 33812–33821. [DOI] [PubMed]
  25. Spence, J. C. H., Weierstall, U. & Howells, M. R. (2004). Ultramicroscopy, 101, 149–152. [DOI] [PubMed]
  26. Thibault, P., Dierolf, M., Menzel, A., Bunk, O., David, C. & Pfeiffer, F. (2008). Science, 321, 379–382. [DOI] [PubMed]
  27. Veen, F. van der & Pfeiffer, F. (2004). J. Phys. Condens. Matter, 16, 5003–5030.

Articles from Journal of Synchrotron Radiation are provided here courtesy of International Union of Crystallography

RESOURCES