Abstract
Raster-scan optoacoustic mesoscopy (RSOM), also termed photoacoustic mesoscopy, offers novel insights into vascular morphology and pathophysiological biomarkers of skin inflammation in vivo at depths unattainable by other optical imaging methods. Using ultra-wideband detection and focused ultrasound transducers, RSOM can achieve axial resolution of 4 micron and lateral resolution of 20 micron to depths of several millimeters. However, motion effects may deteriorate performance and reduce the effective resolution. To provide high-quality optoacoustic images in clinical measurements, we developed a motion correction algorithm for RSOM. The algorithm is based on observing disruptions of the ultrasound wave front generated by the vertical movement of the melanin layer at the skin surface. From the disrupted skin surface, a smooth synthetic surface is generated, and the offset between the two surfaces is used to correct for the relative position of the ultrasound detector. We test the algorithm in measurements of healthy and psoriatic human skin and achieve effective resolution up to 5-fold higher than before correction. We discuss the performance of the correction algorithm and its implications in the context of multispectral mesoscopy.
Introduction
Raster-scan optoacoustic (photoacoustic) mesoscopy (RSOM) is a biomedical optical imaging technique1, 2, capable of visualizing tissue morphology3, 4 and pathophysiological biomarkers of inflammatory diseases5 through several millimeters of depth while preserving high resolution. Exploiting ultra-wideband (UWB) ultrasound detectors the technique has recently demonstrated label free imaging and quantification of inflammatory biomarkers in psoriasis and eczema5. Moreover, multi-wavelength illumination and spectral unmixing allows visualization of a larger number of pathophysiological features, including tissue/blood oxygenation, melanin patterns6, or the bio-distribution of externally administered photo-absorbing labels7.
RSOM image formation is achieved by raster-scanning a focused ultrasound detector over the region of interest (ROI) to collect optoacoustic waves generated in the tissue in response to pulsed laser illumination. The focal point of the ultrasound detector lies slightly above the surface of the sample, and the detector collects ultrasound signals over an acceptance angle (aperture) defined by the area and focusing characteristics of the detecting element. The collected data are tomographically reconstructed to yield an image of the light absorbers within the skin. Data acquisition speed is determined by the pulse repetition rate of the laser, which cannot exceed the limits imposed by laser safety regulations8, 9. Generating a two-dimensional skin scan based on a grid of 250 × 250 points may require ~60 seconds when the laser has a pulse repetition rate (PRR) of 1 kHz; this scan time is longer when larger grids, multi-wavelength illumination, lower laser safety limits or data averaging are used1, 3–6, 8, 10–14. During this time, the subject’s movement, e.g. breathing motion at a respiratory rate of up to 30 times per minute15, 16, can alter the position of the ultrasound detector relative to the sample, affecting RSOM image quality8, 17. Restricting human motion during RSOM scanning is challenging. The RSOM scan head, which is connected to the main RSOM console by a semi-rigid arm, is usually attached to the ROI using suction or adhesive tape restricting lateral movement of the ROI. Nevertheless, motion effects are not entirely eliminated since the skin surface can possibly move relative to the acoustic axis of the detector due to possible deformation of the soft coupling media, such as water, acoustic gel, and thin, acoustically transparent membranes3, 4, 10, 17–22.
It is therefore important to explore image processing methods that can reduce the effects of such motion on RSOM imaging performance. We have previously developed an algorithm to correct for macroscopic motion in small animals. In this approach, mouse thorax cross-sections are clustered based on the cardiac cycle23. Clustering has also been employed to correct for motion artifacts due to the breathing cycle24. However, these clustering approaches are used to classify inter-frame motion of image frames, where each frame (image) is based on a reconstruction following a single laser pulse. RSOM, in contrast, involves tomographic reconstruction based on a set of recorded time signals and requires a motion correction approach targeting inner-frame motion.
In this work, we aimed to develop to our knowledge the first motion correction algorithm for RSOM. We hypothesize that by tracking the skin surface based on melanin content at each scan point, we could generate a synthetic surface, representative of skin motion. We further postulated that we could use this synthetic surface to correct for sample movement in the axial direction of the transducer during data acquisition, significantly reducing motion artifacts. To test this hypothesis, we developed a motion extraction and correction algorithm and measured a benign nevus with high-frequency RSOM (20–180 MHz). We interrogated the performance of the algorithm by studying motion artifact removal in single capillary loops, achieving effective resolution up to 5-fold better than before correction. We measured healthy and psoriatic skin with clinical RSOM (14–126 MHz), and demonstrate the correction of breathing motion, which allows for reliable imaging in a clinical context. Finally, we measured blood oxygenation and melanin content in healthy human skin with multispectral RSOM (14–126 MHz). We examined the performance of the algorithm by tracing the level of blood oxygenation along single vessels, and show that motion correction is necessary for correcting corrupted readings of blood oxygenation.
Methods
RSOM and the motion problem
RSOM involves raster-scanning a focused ultrasound detector over the ROI and recording the ultrasound waves generated in tissue in response to pulsed laser illumination (Fig. 1a). The collected data are tomographically reconstructed into an image using three-dimensional beam forming1, 6, 14. Typically, the ultrasound detector is scanned along a fast-scanning axis (fs-axis) and a slow-scanning axis (ss-axis) (Fig. 1a). One-dimensional time signals (making up the so-called “A-scan”) are recorded at each scan point, collecting ultrasound signal generated within the cone-shaped sensitivity field of the focused detector. One-dimensional scanning of the detector along the fs-axis generates a two-dimensional sinogram called the “B-scan”, and scanning along the ss-axis generates a stack of B-scans making up a three-dimensional sinogram.
Assuming an acquisition step size of ∆ and an ROI spanning lfs along the fs-axis, acquisition speed along the fs-axis is vfs = PRR ∆ and along the ss-axis is vss = vfs ∆/lfs. Since the step size is much smaller than the dimensions of the ROI, vfs ≫ vss. Clinical RSOM typically involves vfs = 15 mm/s and vss = 0.045 mm/s. These speeds are such that patient motion can alter the axial position of the transducer relative to the ROI between successive B-scans (Fig. 1b). Consequently, two-dimensional sinograms along the ss-axis appear discontinuous and rippled (Fig. 1d), whereas cross-sections along the fs-axis appear continuous and smooth (Fig. 1c). Such changes in transducer height above the ROI are not taken into account by RSOM reconstruction algorithms published to date1, 3, 4, 14, 17.
The motion correction algorithm
The algorithm is based on the observation that the ultrasound wave front generated by the melanin-containing layers of the skin (primarily the basal layer of the epidermis and the stratum corneum) show vertical disruptions in the three-dimensional sinogram. According to the Huygens-Fresnel principle, an ultrasound wave emitted from a continuous surface will generate a secondary wave front with a continuous footprint in the three-dimensional sinogram. Movement during RSOM measurement distorts this footprint and generates discontinuities in the three-dimensional sinogram (Fig. 1d). The motion correction algorithm described here detects disruptions of the stratum corneum/stratum basale (hereafter “skin surface”) and corrects for this motion by creating an artificial continuous surface. The primary assumption in this algorithm is that the skin surface is continuous on a mesoscopic scale, i.e. on the order of several micrometers.
The motion correction algorithm proceeds in two steps. First, an algorithm is applied to detect the discontinuous skin surface in the three-dimensional sinogram. Second, an artificial, smooth, continuous skin surface is generated based on the detected surface, and the difference between the two surfaces is taken to indicate the vertical movement of the skin with respect to the detector surface. Based on this vertical movement, the focal point of the detector is adjusted during image reconstruction.
The first step of the algorithm, which is the more challenging, involves detecting the skin surface in the three-dimensional sinogram. We have developed two approaches depending on the tissue being examined. When the ROI lies in an area of hairless skin (e.g. palm of the hand), the stratum corneum is roughly segmented using a two-dimensional parabolic slab, reflecting the fact that optoacoustic signal from the stratum corneum is displaced by several hundred microns from the signal emitted from underlying vasculature (Fig. 1e). Within the segmented slab, the surface is identified based on the maximum value within each A-scan. When the ROI lies in an area of hairy skin (e.g. lower arm), we use a different approach because the stratum corneum is not sufficiently separated in space from the underlying stratum basale and dermal papillae (Fig. 1f). Nonetheless, the basal layer of the epidermis leaves a distinctive footprint in the three-dimensional sinogram: ultrasound signals emitted by the stratum basale induce shear waves within the acoustic lens of the detector, which in turn stimulate lateral modes in the piezo element of the detector, generating a characteristic low-frequency footprint approximately 600 µm below the skin surface (Fig. 1g). Normally, this low-frequency band is filtered out before reconstruction1, 3, 4, 14, 17, but our algorithm exploits it to generate an image of the skin surface. An exponential bandpass filter from 1–3 MHz is applied to the raw data, which are roughly segmented by time to identify the interval containing the highest peak (Fig. 1f). The maximum in the low-frequency band is determined for each A-scan, resulting in a two-dimensional map of the motion-corrupted melanin layer (Fig. 2a).
The second step in the algorithm is the same regardless of whether the skin ROI is hairy or hairless. A moving average filter is applied to the disrupted surface, correcting the entire three-dimensional sinogram and creating an artificially smoothed surface (Fig. 2b). In our experiments, the moving average filter performed best when it spanned 200 pixels along the ss-axis and 10 pixels along the fs-axis. Therefore these span dimensions were used in all experiments described below. The two-dimensional map between the motion-corrupted and artificially smoothed surfaces was calculated to provide the correction for detector z-position during reconstruction.
In vivo validation of the correction algorithm
The performance of the motion correction algorithm was assessed using in vivo data collected with three RSOM set-ups based on a spherically focused ultrasound detector with acoustic lens. Human patient procedures were approved by the Ethics Committee of the Technical University of Munich, and all patients provided informed consent. All measurements were performed in compliance with laser safety standards for single laser pulses, pulse trains, and the average power9. A derivation of formulas defining maximum exposure limits in optoacoustic mesoscopy have been derived recently8.
The high-frequency RSOM set-up (hereafter “RSOM100”) was used to image small vessels in a benign nevus on the lower arm of a healthy man with skin type II. The detector had a detection band of 20–180 MHz, an active aperture of 1.5 mm and an f-number of 1.1. To ensure fast, high-resolution scans, a single-wavelength laser emitting pulses of 532-nm light lasting <1 ns at repetition rates up to 2 kHz was used (Wedge HB532, Bright Solutions SRL, Italy).
A clinical RSOM set-up (hereafter “RSOM55”), in which the scan head was attached to a portable clinical imaging platform5, was used to image healthy skin as well as a psoriasis plaque on the arm of an individual diagnosed with psoriasis by physicians at the Klinikum rechts der Isar/Technical University of Munich. The plaque was located on the long head of the triceps brachii, approximately 5 cm above the olecranon. The healthy skin ROI was also located on the long head of the triceps brachii, approximately 4 cm away from the psoriasis plaque. The RSOM55 detector had a detection band of 14–126 MHz, an active aperture of 3 mm, and an f-number of 1. The laser was the same as in the RSOM100 set-up.
During image reconstruction of data collected with the RSOM100 or RSOM55 set-ups, the reconstruction band was split up into low- and high-frequency sub-bands, which were 20-60 and 60-180 MHz for RSOM100 or 14–42 and 42–126 MHz for RSOM55. The two bands were equalized, colored green (high-frequency) or red (low-frequency), and overlaid in final images. Reconstruction using frequency sub-bands improves visualization of higher frequency signals emitted by smaller structures in the tissue1, 5, 17.
The third experimental set-up, multispectral RSOM (hereafter “MSOM”), was used to measure blood oxygenation in the lower arm of a healthy man with skin type III. The detector was the same as in the RSOM55 set-up but with a detection band of 14–126 MHz. Tissue was excited using a tunable-wavelength OPO laser emitting 6-ns pulses in the spectral range from 420 to 2300 nm at a pulse repetition rate of 100 Hz (SpitLight OPO, InnoLas Laser GmbH, Krailling, Germany). Other aspects of MSOM imaging, reconstruction, and unmixing were described in a previous publication6.
Parameters used to evaluate the correction algorithm
Uncorrected and corrected reconstructions obtained using the RSOM100 and RSOM55 set-ups were compared in terms of their contrast-to-noise ratio (CNR) as a function of depth. CNR was defined as
1 |
where is the maximal amplitude within the image plane at depth z k, and noise max is the background signal based on the maximal amplitude of the image above the skin surface. The term represents the three-dimensional reconstruction, where ; ; and ; and N x, N y, and N z are the numbers of voxels along the x-, y-, and z-axes. The plot of as a function of depth should show smaller peaks for motion-corrupted data than for data appropriately motion-corrected.
We also compared uncorrected and corrected RSOM100 images in terms of effective resolution calculated from the full width at half maximum (FWHM) of optoacoustic peaks along line profiles corresponding to single capillary loops in the skin. The top part of capillary loops appears as a circular absorber in RSOM images5.
We compared uncorrected and corrected MSOM images in terms of blood vessel visualization and blood oxygenation measurement. For blood oxygenation analysis, the same vessel was segmented in motion-corrupted and -corrected three-dimensional reconstructions, and blood oxygenation was measured along the vessel. The segmented vessel was oriented primarily along the y-axis. Blood oxygenation at position y j′ was averaged across the diameter of the vessel using the equation
2 |
where is the voxel-to-voxel ratio of oxygenated hemoglobin to total hemoglobin (oxyhemoglobin plus deoxyhemoglobin). The vessel was segmented by thresholding the reconstructed images at 20% of the maximal value:
3 |
where H(x) is the Heaviside step function and describes a cuboid in the reconstructed volume that contains the vessel of interest. if voxel amplitude is greater than 20% of the maximum amplitude within the cuboid; otherwise, . Blood oxygenation level along the length of blood vessels was compared before and after motion correction. Motion during MSOM data acquisition should give rise to abrupt or large fluctuations in that do not overlap with vessel bifurcations.
Data availability
The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request and with permission of Technical University of Munich and the Helmholtz Zentrum München.
Results
Motion correction in high-resolution RSOM100
Figure 2 illustrates the performance of the motion correction algorithm when imaging a benign nevus on the lower arm of a healthy volunteer using the RSOM100 set-up. The algorithm identified a rippled surface in the raw data, indicating micromotion in the z-direction on the order of several micrometers between successive B-scans (Fig. 2a). The skin surface appeared disrupted only along the ss-axis, not along the fs-axis. The motion-corrected skin surface was continuous and smooth, with no offset between neighboring B-scans. Figure 2b displays the vertical maximum amplitude projections (MAPs) of motion-corrupted and motion-corrected imaging volumes. Blurring was visible in the uncorrected image, especially in the high-frequency (green) reconstruction. Figure 2c plots mean CNR with respect to depth based on the vertical MAPs. Motion correction led to higher CNR in the stratum corneum as well as superficial horizontal plexus; the CNR increase over uncorrected data was greater at high frequencies (up to 5 dB) than at low ones (up to 2.5 dB). In addition to this quantitative increase in CNR, motion correction led to significant motion artifact removal in the epidermal-dermal junction and the dermis in lateral MAPs (Fig. 2d,e). The increase in effective resolution was particularly apparent at the epidermal-dermal junction (Fig. 2d): the uncorrected MAP showed an unstructured pattern, while the motion-corrected MAP showed a structured pattern of capillary loops. The zoomed-in region (500 µm × 500 µm) in Fig. 2d revealed no individual capillary loops before motion correction, whereas more than a dozen loops were visible after correction. Similarly, microvessels in the papillary dermis were resolved only after motion correction. Inset images in Fig. 2e suggest that micromotion affected small vessels much more than large ones.
To quantify the improvement in effective resolution associated with the motion correction algorithm, line profiles were traced through neighboring capillary loops and analyzed (Fig. 2f,g). While the uncorrected image showed several neighboring capillaries as a lateral tube without clear peaks in the line profile, the corrected image showed four distinctive capillary loops, each appearing as a resolved peak with FWHM values of 36–41 µm. Peaks 3 and 4 combined to form a plateau with FWHM of 189 µm. A walk-through of the motion-corrected three-dimensional volume of the ROI illustrates high resolution of vascular networks at various depths (Supplementary Movie S1).
Motion correction with clinical RSOM55
Figure 3a reveals sinusoidal modulation of healthy and psoriatic skin surfaces at approximately 15 cycles per minute (17–18 cycles per measurement), corresponding to the patient’s breathing. This sinusoidal movement was eliminated by the motion correction algorithm, allowing the visualization of inflammation biomarkers typical of psoriatic skin, including thickening of the epidermis, broadening of capillary loops within the dermal papillae, and increased blood volume fraction in dermal vasculature (Fig. 3b–d). Motion correction significantly improved image quality: capillary loops in the psoriasis plaque appeared blurred before motion correction and well resolved afterwards (Fig. 3c). Perhaps the greatest improvement in image quality was observed in the microvasculature of the superficial horizontal plexus (Fig. 3d). Images of both healthy and psoriatic skin revealed a structured network of smaller vessels (green) and larger ones (red) running through the dermis. Smaller vessels were completely blurred before motion correction but well resolved afterwards.
Motion correction with MSOM
By measuring three-dimensional absorption maps at multiple wavelengths, MSOM provides functional information based on the spectral signature of intrinsic chromophores. Figure 4 shows the impact of motion on the visualization of vasculature as well as the quantification of blood oxygenation in the lower arm of a healthy volunteer. Figure 4a shows that without motion correction, the skin surface appeared strongly rippled, even along the fs-axis, reflecting the prolonged measurement time needed for MSOM data acquisition. The motion-corrected surface was smooth and continuous along both the fs- and ss-axes. Figure 4b highlights two areas (ellipses 1 and 2) where single blood vessels appeared as a cloud without clear boundaries prior to motion correction but as well-resolved vessels afterwards. Another vessel (ellipse 3) appeared as a double structure before motion correction, and the “shadow vessel” disappeared after correction (Fig. 4b,c).
In addition to affecting blood vessel visualization in MSOM, motion also interfered with measurement of blood oxygenation. The vessel within ellipse 3 in Fig. 4b was segmented and mean blood oxygenation was measured along its length (Fig. 4c). Blood oxygenation level in the uncorrected image fluctuated sharply between 55% and 90%, and the fluctuations did not correlate with vessel bifurcations (gray bands). Blood oxygenation level in the motion-corrected image showed smaller, less abrupt variations, ranging from 55% to 70%. In addition, regions of vessel bifurcation were associated with changes in blood oxygenation.
Discussion
In this study, we present to our knowledge the first motion correction algorithm for RSOM, enabling motion artifact removal with effective resolution up to 5-fold better than conventional RSOM. We show that by detecting the disrupted skin surface in raw data and comparing it to an artificially smoothed, continuous surface, we can reliably correct for motion effects during image reconstruction in optoacoustic dermoscopy. We validate the algorithm using high-frequency RSOM100, clinical RSOM55, and functional MSOM set-ups.
The RSOM100 set-up is particularly vulnerable to motion artifacts because it can offer higher resolution of 4 µm axially and 20 µm laterally1, 17 at slower acquisition speeds of ~80 pixels per second8. We have shown that our motion correction algorithm can provide unparalleled RSOM100 imaging of capillary loops in the epidermal-dermal junction as well as the superficial horizontal plexus.
Clinical RSOM55, although in principal less vulnerable to motion artifacts than RSOM100 because it images at lower resolutions of 10 µm axially and 40 µm laterally and faster acquisition speeds of ~140 pixels per second5, 17, must perform under challenging conditions. Various parts of the body, some difficult to hold still, must be imaged in the clinic, including lower extremities, trunk, elbow and scalp25, and most individuals being scanned are elderly or unwell and may have difficulty controlling their breathing or trembling. Imaging under these conditions is more demanding than imaging the palm3, 4, 10, 12, 26 or lower arm8, 11 of healthy individuals under laboratory conditions. We have shown that our motion correction algorithm can correct for strong breathing motion enabling quantification of disease biomarkers even under challenging conditions.
Functional MSOM is extremely sensitive to motion because it acquires only 3.4 pixels per second8, meaning that a scan of 4 mm × 2 mm takes approximately 13 minutes6, at a lateral resolution of approximately 55 µm27. Each of the datasets acquired at different wavelengths shows unique motion artifacts in the uncorrected multi-wavelength reconstructions. Thus, not only contrast is affected by motion but the spectral signature of individual image pixels as well. Simple co-registration of the data based on Euclidean transformation is not sufficient to perfectly co-register each pixel in inter-wavelength reconstructions of uncorrected datasets. We have shown that without motion correction, MSOM not only fails to resolve anatomical structures but also generates highly unreliable blood oxygenation measurements.
The motion correction algorithm presented in this work was designed for optoacoustic mesoscopy, which requires tomographic reconstruction. In comparison, optoacoustic microscopy, which does not need 3D reconstruction, will not benefit from the proposed algorithm significantly since motion artifacts are restricted to the axial dimension and do not propagate to the lateral dimension. Yet optoacoustic mesoscopy outperforms optoacoustic microscopy in imaging depth. The imaging depth in optoacoustic microscopy is limited by the high scattering coefficient of human skin, which is approximately twice as strong compared to subcutaneous tissue, which in turn is stronger than muscle tissue28. At a wavelength of 532 nm the reduced scattering coefficient of human skin measures approximately µs’ = 50 cm−1, severely limiting the possibility to optically focus light beyond a few hundred micron. Favazza et al. have performed a direct comparison of optoacoustic micrsocopy and mesocopy and showed that microscopy can resolve only capillaries and small vessels of the upper superficial horizontal plexus, whereas mesoscopy can observe all vasculature within the dermis29. We have recently shown that optoacoustic mesoscopy is even capable of measuring all vasculare structures within the dermis from the small capillaries in the epidermal-dermal junction to the larger vessels of the deep horizontal plexus in the dermis8.
The algorithm presented here works by identifying the footprint of the skin surface in three-dimensional sinograms, which is facilitated by the strongly absorbing stratum basale. In the case of humans or animals lacking an absorbing layer close to the skin surface, such a layer can easily be created by dissolving ink in the acoustic coupling agent, by slightly coloring the skin surface with a marker, or by introducing absorbing molecules in the coupling membrane. This is analogous to marking the skin surface with a fiducial marker in optical coherence tomography30. Thus, the motion correction algorithm described here is applicable to a wide range of preclinical and clinical RSOM studies.
This motion correction algorithm will help provide the accuracy and reliability needed for realizing the full clinical potential of RSOM. This technique, already optimized in terms of frequency and excitation energy3, 4, 8, has shown the ability to measure angiogenesis in preclinical models of melanoma tumors, and phenotypic biomarkers of inflammation in human skin5, 17. The multispectral capability of MSOM can provide readings of dermal blood oxygenation level and melanin content6. The algorithm presented in this work will allow for a wide range of interesting RSOM-based dermatology studies in the near future.
Electronic supplementary material
Acknowledgements
This project has received funding from the European Union’s Horizon 2020 research and innovation program under grant agreement no. 687866 (INNODERM). Juan Aguirre acknowledges funding from the European Marie Curie IEF fellowships - project HIFI (no. 327391). We thank Dr. A. Chapin Rodriguez for help revising the manuscript.
Author Contributions
M.S. developed the imaging systems, designed and performed the experiments, developed the motion correction algorithm, processed the data, and wrote the paper. N.G. recruited the patients and helped performing the experiments. K.E. wrote the ethical protocol and provided conceptual input related to patient studies. J.A. developed the imaging systems, helped performing the experiments, and provided conceptual input. V.N. provided conceptual input, supervised and led the research, and wrote the paper. All authors contributed to the refinement of the article.
Competing Interests
V.N. and M.S. have a financial interest in iThera Medical GmbH, which, however, did not support this work.
Footnotes
Electronic supplementary material
Supplementary information accompanies this paper at doi:10.1038/s41598-017-11277-y
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Omar M, Soliman D, Gateau J, Ntziachristos V. Ultrawideband reflection-mode optoacoustic mesoscopy. Optics letters. 2014;39:3911–3914. doi: 10.1364/OL.39.003911. [DOI] [PubMed] [Google Scholar]
- 2.Taruttis A, Ntziachristos V. Advances in real-time multispectral optoacoustic imaging and its applications. Nat Photonics. 2015;9:219–227. doi: 10.1038/nphoton.2015.29. [DOI] [Google Scholar]
- 3.Schwarz M, Omar M, Buehler A, Aguirre J, Ntziachristos V. Implications of Ultrasound Frequency in Optoacoustic Mesoscopy of the Skin. IEEE transactions on medical imaging. 2014;34:672–677. doi: 10.1109/TMI.2014.2365239. [DOI] [PubMed] [Google Scholar]
- 4.Aguirre J, et al. Broadband mesoscopic optoacoustic tomography reveals skin layers. Optics letters. 2014;39 doi: 10.1364/OL.39.006297. [DOI] [PubMed] [Google Scholar]
- 5.Aguirre J, et al. Precision assessment of label-free psoriasis biomarkers with ultra-broadband optoacoustic mesoscopy. Nature Biomedical Engineering. 2017;1 doi: 10.1038/s41551-017-0068. [DOI] [Google Scholar]
- 6.Schwarz M, Buehler A, Aguirre J, Ntziachristos V. Three-dimensional multispectral optoacoustic mesoscopy reveals melanin and blood oxygenation in human skin in vivo. Journal of biophotonics. 2016;9:55–60. doi: 10.1002/jbio.201500247. [DOI] [PubMed] [Google Scholar]
- 7.Beziere N, et al. Optoacoustic imaging and staging of inflammation in a murine model of arthritis. Arthritis & rheumatology. 2014;66:2071–2078. doi: 10.1002/art.38642. [DOI] [PubMed] [Google Scholar]
- 8.Schwarz M, et al. Optoacoustic Dermoscopy of the Human Skin: Tuning Excitation Energy for Optimal Detection Bandwidth With Fast and Deep Imaging in vivo. IEEE transactions on medical imaging. 2017;36:1287–1296. doi: 10.1109/TMI.2017.2664142. [DOI] [PubMed] [Google Scholar]
- 9.American National Standards Institute & The Laser Institute of America. American National Standard for safe use of lasers: approved March 16, 2007. (The Laser Institute of America, 2007).
- 10.Bost W, Lemor R, Fournelle M. Optoacoustic imaging of subcutaneous microvasculature with a class one laser. IEEE transactions on medical imaging. 2014;33:1900–1904. doi: 10.1109/TMI.2014.2326552. [DOI] [PubMed] [Google Scholar]
- 11.Favazza CP, Jassim O, Cornelius LA, Wang LV. In vivo photoacoustic microscopy of human cutaneous microvasculature and a nevus. Journal of biomedical optics. 2011;16 doi: 10.1117/1.3528661. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Zabihian B, et al. In vivo dual-modality photoacoustic and optical coherence tomography imaging of human dermatological pathologies. Biomed Opt Express. 2015;6 doi: 10.1364/BOE.6.003163. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Zhang EZ, Laufer JG, Pedley RB, Beard PC. In vivo high-resolution 3D photoacoustic imaging of superficial vascular anatomy. Physics in medicine and biology. 2009;54:1035–1046. doi: 10.1088/0031-9155/54/4/014. [DOI] [PubMed] [Google Scholar]
- 14.Omar M, Gateau J, Ntziachristos V. Raster-scan optoacoustic mesoscopy in the 25-125 MHz range. Optics letters. 2013;38:2472–2474. doi: 10.1364/OL.38.002472. [DOI] [PubMed] [Google Scholar]
- 15.Barrett, K. E., Barman, S. M., Boitano, S. & Brooks, H. Ganong’s Review of Medical Physiology, 24th Edition. p. 619 (Mcgraw-hill, 2012).
- 16.Rodriguez-Molinero A, Narvaiza L, Ruiz J, Galvez-Barron C. Normal respiratory rate and peripheral blood oxygen saturation in the elderly population. Journal of the American Geriatrics Society. 2013;61:2238–2240. doi: 10.1111/jgs.12580. [DOI] [PubMed] [Google Scholar]
- 17.Omar M, Schwarz M, Soliman D, Symvoulidis P, Ntziachristos V. Pushing the Optical Imaging Limits of Cancer with Multi-Frequency-Band Raster-Scan Optoacoustic Mesoscopy (RSOM) Neoplasia. 2015;17:208–214. doi: 10.1016/j.neo.2014.12.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Schwarz M, Buehler A, Ntziachristos V. Isotropic high resolution optoacoustic imaging with linear detector arrays in bi-directional scanning. Journal of biophotonics. 2015;8:60–70. doi: 10.1002/jbio.201400021. [DOI] [PubMed] [Google Scholar]
- 19.Vionnet L, et al. 24 MHz Scanner for Optoacoustic Imaging of Skin and Burn. IEEE transactions on medical imaging. 2013;33:535–545. doi: 10.1109/TMI.2013.2289930. [DOI] [PubMed] [Google Scholar]
- 20.Buehler A, Dean-Ben XL, Claussen J, Ntziachristos V, Razansky D. Three-dimensional optoacoustic tomography at video rate. Optics express. 2012;20:22712–22719. doi: 10.1364/OE.20.022712. [DOI] [PubMed] [Google Scholar]
- 21.Buehler A, Herzog E, Razansky D, Ntziachristos V. Video rate optoacoustic tomography of mouse kidney perfusion. Optics letters. 2010;35:2475–2477. doi: 10.1364/OL.35.002475. [DOI] [PubMed] [Google Scholar]
- 22.Buehler A, Kacprowicz M, Taruttis A, Ntziachristos V. Real-time handheld multispectral optoacoustic imaging. Optics letters. 2013;38:1404–1406. doi: 10.1364/OL.38.001404. [DOI] [PubMed] [Google Scholar]
- 23.Taruttis, A., Claussen, J., Razansky, D. & Ntziachristos, V. Motion clustering for deblurring multispectral optoacoustic tomography images of the mouse heart. Journal of biomedical optics17, doi:10.1117/1.Jbo.17.1.016009 (2012). [DOI] [PubMed]
- 24.Xia J, Chen W, Maslov K, Anastasio MA, Wang LV. Retrospective respiration-gated whole-body photoacoustic computed tomography of mice. Journal of biomedical optics. 2014;19 doi: 10.1117/1.JBO.19.1.016003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Farber EM, Nall ML. The natural history of psoriasis in 5,600 patients. Dermatologica. 1974;148:1–18. doi: 10.1159/000251595. [DOI] [PubMed] [Google Scholar]
- 26.Zhang HF, Maslov K, Stoica G, Wang LV. Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging. Nature biotechnology. 2006;24:848–851. doi: 10.1038/nbt1220. [DOI] [PubMed] [Google Scholar]
- 27.Schwarz M, Aguirre J, Soliman D, Buehler A, Ntziachristos V. Unmixing chromophores in human skin with a 3D multispectral optoacoustic mesoscopy system. 2016;9708 [Google Scholar]
- 28.Simpson CR, Kohl M, Essenpreis M, Cope M. Near-infrared optical properties of ex vivo human skin and subcutaneous tissues measured using the Monte Carlo inversion technique. Physics in medicine and biology. 1998;43:2465–2478. doi: 10.1088/0031-9155/43/9/003. [DOI] [PubMed] [Google Scholar]
- 29.Favazza CP, et al. In vivo multiscale photoacoustic microscopy of human skin. Proc. SPIE. 2011;7899 doi: 10.1117/12.875969. [DOI] [Google Scholar]
- 30.Liew YM, McLaughlin RA, Wood FM, Sampson DD. Motion correction of in vivo three-dimensional optical coherence tomography of human skin using a fiducial marker. Biomed Opt Express. 2012;3:1774–1786. doi: 10.1364/BOE.3.001774. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request and with permission of Technical University of Munich and the Helmholtz Zentrum München.