Abstract
Because radiation is harmful to patients, it is important to reduce X-ray exposure in the clinic. For CT, reconstructions from sparse views or limited angle tomography are being used more frequently for low dose imaging. However, insufficient sampling data causes severe streak artifacts in images reconstructed using conventional methods. To solve this issue, various methods have recently been developed. In this paper, we improve a statistical iterative algorithm based on the minimization of the image total variation (TV) for sparse or limited projection views during CT image reconstruction. Considering the statistical nature of the projection data, the TV is performed under a penalized weighted least-squares (PWLS-TV) criterion. During implementation of the proposed method, the image reconstructed using the filtered back-projection (FBP) method is used as the initial value of the first iteration. Next, the feature refinement (FR) step is performed after each PWLS-TV iteration to extract the fine features lost in the TV minimization, which we refer to as ‘PWLS-TV-FR’.
Introduction
Although the clinical value of CT is unquestionable, radiation exposure is harmful to patients1, 2. To date, many hardware-based scanning protocols have been proposed to reduce the CT scan radiation dose3, 4. Many software-based image reconstruction techniques have also been used to handle low-dose data from available CT scanners lacking hardware modifications5–7. In the CT field, the radiation dose is usually reduced by lowering the X-ray tube current, lowering the X-ray tube voltage, or sampling fewer views or using limited-angle projection. However, lowering X-ray exposure or using sparse or limited views increases noise and data inconsistency. Without adequate processing, it is difficult to reconstruct high-quality diagnostic CT images using conventional algorithms. Generally, a traditional CT algorithm using insufficient data will lead to degraded reconstruction accuracy and severe image artifacts8. Exact reconstruction methods with noise suppression capabilities intended to support dose reduction have been widely investigated.
One of the main strategies for low-dose CT image reconstruction restores the line integrals from acquired low-dose projection data, and a transform-based method has been presented to address the noise properties of low-mAs CT sinogram data. Li et al. determined a nonlinear relationship between the variance and the mean of acquired low-mAs sinogram data9, which facilitates low-dose CT image reconstruction. Based on this relationship, a framework for image reconstruction using low-mAs sinogram data with penalized re-weighted least squares was investigated by Wang et al.10. A general series of sophisticated CT image reconstruction methods were reported by Fessler et al.11.
Reducing projection views or limiting the scan angle is another strategy that has been used for CT dose reduction, as reducing the number of projections will clearly reduce the radiation dose5. In this study, we focused on CT image reconstruction from sparse-view or limited-angle projection data. Many iterative algorithms have been used for CT image reconstruction, and the actual iterative algorithms are well established12. Among them, statistical iterative reconstruction (SIR), which is used to model measurement statistics and imaging geometry, can significantly maintain the image quality in cases of insufficient sampling data, especially compared with the filtered back-projection (FBP) reconstruction algorithm13. The cost function of SIR is usually composed of a data-fidelity term and a regularization term. The data-fidelity term models the measurement statistics to ensure successful SIR image reconstruction, and the regularization term shows the prior information for the desired image intended to regularize the solution14. The traditional regularization term usually produces unfavorable smoothing effects on edge regions. To address this issue, several edge-preserving regularization terms have been presented, such as total variation (TV) regularization with a piecewise constant assumption (PCA)15–17. Numerous investigations have shown that high-quality CT images can be reconstructed via TV minimization from sparse-view sampling data. However, PCA often produces noticeable patchy artifacts in reconstructed images5.
Recently, we developed a statistical interior tomography approach for CT reconstruction18. Its potential usefulness have been demonstrated for feature preservation of interior tomography with truncated projection measurements. However, this algorithm is a little bit complex, involving FBP with projection extension, PWLS-TV and feature refinement and the last component is not very well characterized with the CT interior tomography application. Moreover, the algorithm was applied to a scanning configuration which is not practically used (namely the interior problem).
In this paper, we extend this method to a more practical protocol, sparse-view and limited-angle image reconstruction. Furthermore, the FBP algorithm is adopted instead of the FBP-EP algorithm which is time-consuming. To remove the patchy artifacts from the TV-based methods, we improved a statistical iterative image reconstruction algorithm based on minimizing the image TV that is specifically performed using PWLS criteria (PWLS-TV), which models the statistical properties of the projection data. Furthermore, after each PWLS-TV iteration, a feature refinement step (FR) is performed to restore the desired structure details that are lost in the TV minimization. For simplicity, the presented method is termed “PWLS-TV-FR”. The proposed PWLS-TV-FR method tends to relieve the PCA requirement by statistical modeling and iterative feature refinement.
The remainder of the paper is organized as follows. Section II briefly reviews the CT imaging model and describes the TV-regularized PWLS criterion in the image domain and the FR step. The experimental setup and evaluation metrics are also presented in this section. In section III and IV, the experimental results on a digital phantom and using micro-CT are reported, followed by a conclusion section.
Methods
PWLS-TV minimization
Based on the statistical properties of the projection data, the iterative PWLS approach including a data-fidelity term and a penalty term has been applied to the image domain for dual-energy X-ray CT reconstruction19–21. But the PWLS method for low-dose X-ray CT reconstruction are flawed in the suppression of noise and artifacts. According to our previous study, the TV-regularized PWLS algorithm we referred to as “PWLS-TV” can address this issue, which can be solved using the steepest descent algorithm. Based on the image gradient magnitude sparseness22, TV of the image is one of the most commonly used techniques:
1 |
in which s and t are the index of the location of the attenuation coefficients of the desired image and α is a small constant used for keeping the equation differentiable with respect to image intensity, which is assigned a value of in this work.
Thus, given the current estimated image μ n, a new image μ n+1 can be calculated by minimizing the cost function of PWLS-TV, which can be described as follows:
2 |
in which ∇TV(μn) denotes the gradient of TV(μn), τn represents the gradient step size and β is a smoothing parameter balancing the degree of agreement between the estimated and the measured data. This can be computed as follows:
3 |
Feature refinement (FR)
High-diagnostic CT images reconstructed from limited-angle or sparse-view acquisitions are very important in clinical practice. However, due to the piecewise constant assumption, some details in the reconstruction using the TV minimization are likely to be lost; in particular, the residual image between two successive iterations in PWLS-TV minimization might contain useful structural information, especially when the dose is very low or the projection data are insufficient. Motivated by this discovery, an FR was performed after each PWLS-TV minimization iteration to extract the main features of the loss from the residual image. This is described in the following equation:
4 |
in which the symbol ⊗ denotes point-wise multiplication, P is a non-negative control notation described as . denotes a feature descriptor which is defined as:
5 |
in which the constants C is introduced for numerical stability. Local statistics σp, σq and σpq at pixel i are defined as follows:
6 |
7 |
8 |
in which , , pi and qi denote two local image patches with a size of centered at pixel i and extracted from and a degraded image obtained by a linear Gaussian filter from , respectively. The nature of the presented model structure descriptor is a structure-related map and includes a contrast variation component and a structure correlation component. The former calculates the reduction of contrast variation due to the degraded operation, and the latter quantifies the structural correction between the original and degraded images23. The value of each element in the feature descriptor image ft is in the interval [0, 1], and the larger the value, the more likely it belongs to part of the structure. In this paper, the proposed statistical iterative CT image reconstruction algorithm using the FR approach is referred to as “PWLS-TV-FR” and includes the three main steps shown in Fig. 1.
Animal preparation and anesthesia administration
All animal experiments described in this work were approved by the Institutional Ethical Committee of Animal Experimentation of Shenzhen Institutes of Advanced Technology (Chinese Academy of Sciences). The experiments (Certificate number: SIAT-IRB-150203-YGS-QWB-A0088) were carried out strictly in accordance with governmental and international guidelines on animal experimentation. All efforts were made to minimize the usage amount of animals and the suffering during experiments according to the request of Biosafety and Animal Ethics. The mouse was anesthetized by intraperitoneal injection of ketamine (70 mg/kg) and xylazine (7 mg/kg) cocktail during CT scanning.
Data availability
The datasets generated or analyzed during the current study are available from the corresponding author on reasonable request.
Experiments
To validate the performance of PWLS-TV-FR, we performed a digital simulation experiment and a physical experiment. We used digital phantom simulation data that spanned 120 views to verify the effect of PWLS-TV-FR on sparse-view data. Additionally, sparse-view anesthetized mouse data and limited-angle mouse bone sample data were acquired from a commercial CT scanner to demonstrate the feasibility of the proposed method in practical applications for sparse-view or limited-angle sampling data. For both phantom simulation and Micro-CT data, Table 1 shows the idea projection number needed for FBP and the projection number used for proposed method.
Table 1.
Projection number | Simulated data | Micro-CT data |
---|---|---|
Idea projection number of FBP | 1160 | 450 |
Used projection number of PWLS-TV-FR | 120 | 225 |
Digital phantom simulations
A head model of the digital XCAT phantom was chosen as experimental simulation data (as shown in Fig. 2) to evaluate the performance of the FR step after each PWLS-TV iteration. To simulate sparse sampling, only 120 views were spanned on a circular orbit of 360°, and the number of bins per view was 672. The distance from the X-ray source to the detector was 1040 mm, and the distance from the rotation center to the curved detector was 570 mm.
Micro-CT data
In this experiment, two scans were performed using a Micro-CT Skyscan1076 scanner with a normal-dose (70 kVp, 225 mAs) protocol. One scan was performed on an anesthetized mouse, in which 450 projections were uniformly collected over a 360° range, and another scan was performed on a mouse bone sample, in which 291 projections were uniformly collected over a 197.88° range, which can be considered limited-angle sampling. In this work, 450 projections of the anesthetized mouse were encoded from 0 to 449, and only even-numbered projections were selected for use as under-sampled data for a final total of 225 projections. Only the central slice projection was extracted, which is typical for fan-beam geometry. For each projection, 4000 detector elements were equally angularly distributed with a total length of 50.12 mm. The distance from the object to the curved detector was 165 mm, and the distance from the object to the X-ray source was 121 mm.
Performance evaluation
(1) In this work, the root mean squared error (RMSE) was used to evaluate reconstruction accuracy and is described as:
9 |
in which denotes the reconstructed image, denotes the true image, m indexes the pixels of the image and Q is the total number of pixels in the image.
(2) To evaluate the noise reduction capability of different reconstruction algorithms, the peak signal to noise ratio (PSNR) metric was adopted in this paper and is as follows:
10 |
in which denotes the reconstructed image, denotes the true image, and represents the maximum intensity value of .
Results
Digital phantom result
In this experiment, with sparse sampling data, images were reconstructed using three methods, FBP, PWLS-TV and PWLS-TV-FR, which are shown with partial magnification in Fig. 3. From Fig. 3, it is clear that there are severe streak artifacts in the image reconstructed using FBP, whereas the PWLS-TV results exhibit fewer artifacts but degraded details. Moreover, the image reconstructed using PWLS-TV-FR showed noticeable improvements over that reconstructed using PWLS-TV by preserving small-scale and low-contrast structures, as indicated by the arrows in the amplified areas. Figure 4 shows residual reconstructed images (shown in Fig. 3) resulting from the true image subtracted from the reconstructed image. The smaller the features of the residual image, the more accurate the reconstructed image. As is clear from Fig. 4, the residual reconstructed image using PWLS-TV-FR has the smallest features, which indicates that the PWLS-TV-FR method for sparse-view data retains more details than either the PWLS-TV or FBP methods.
In the process of updating images using the PWLS-TV or PWLS-TV-FR functions, the trend is that the image gradually becomes clearer and eventually reaches an extreme value as the number of iterations increases. However, as shown Fig. 5, whether it is evaluated by RMSE or by PSNR, the PWLS-TV-FR method always performs better than PWLS-TV. Figure 6 depicts the horizontal profiles of the reconstructions in Fig. 3 and the corresponding horizontal residual value. The result reconstructed using PWLS-TV-FR was closer to the reference profiles, and the residual values reconstructed using PWLS-TV-FR are much smaller than those using PWLS-TV; it is obvious that PWLS-TV-FR achieves better performance. Figure 7 shows a histogram of the RMSE and PSNR evaluation results for the reconstructions in Fig. 3. The red column represents the RMSE value, and the blue column represents the PSNR value. It is important that the RMSE value of PWLS-TV-FR is the minimum and the PSNR value of PWLS-TV-FR is the maximum, which verifies the validity of the PWLS-TV-FR algorithm for sparse-view digital phantom data.
Micro-CT result
Figures 8 and 9 show images and zoomed images reconstructed using FBP, PWLS-TV and PWLS-TV-FR. Figure 8 is a reconstruction using sparse-view sampling of the anesthetized mouse projection data with 225 views, which shows that the PWLS-TV-FR method can suppress severe artifacts and preserve fine tissue structure from sparse-view sampling data in the clinic. Figure 9 is the reconstruction using limited-angle sampling data of mouse bone sample projection data using a scanning angle of 197.88°. Image reconstructions using FBP showed severe streak artifacts resulting in low detail visibility. The images reconstructed using the other two algorithms show a lower level of streak artifacts. However, the features reconstructed using PWLS-TV-FR are clearer than those reconstructed by PWLS-TV. In conclusion, the experimental micro-CT data demonstrate that the presented method can recover structural information lost using the PWLS-TV method for sparse-view or limited-angle sampling data.
Conclusion and Discussion
To address the global use of high-diagnostic CT images obtained from sparse-view or limited-angle projection data, we improved on the statistical iterative algorithm by including an FR step. Two key components were included in this algorithm. One is a TV minimization model using PWLS criteria that account for the statistical nature of projections. The other is an FR step that is performed after each iteration of PWLS-TV to recover any missing fine features. The improved method is designed to relieve the piecewise constant assumption of TV-based methods for sparse-view or limited-angle CT image reconstruction. The reconstruction results from a digital phantom, anesthetized mouse and a mouse bone sample dataset show that the proposed PWLS-TV-FR approach is a significant improvement over the PWLS-TV method in terms of the preservation of structural details and the suppression of undesired patchy artifacts. This result demonstrates its potential to reduce CT radiation doses by decreasing the number of projection views or the rotation angle of the scan.
An iterative reconstruction algorithm usually involves numerous parameters, which can significantly impact reconstruction properties. In this work, we try our best to select the “optimal” parameters for different algorithms in order to make a fair comparison on their performance. For example, the smoothing parameter β balancing the degree of agreement between the estimated and the measured data in function (2) should be a small value to avoid data distortion, however, after testing, we found that is better. Furthermore, the image patch size in the feature descriptor designed to distinguish structure from noise and artifacts has to be carefully selected because the large one leads to better detail-detection ability, while the small one can benefit the computational load. According our previous study, the patch size of is a good choice for balancing the structure detection ability and computational efficiency. Additionally, the parameter C is introduced for numerical stability to avoid the condition that is very close to zero. Therefore, it is assigned as a small constant in our work.
Acknowledgements
The authors would like to thank the editor and anonymous reviewers for their constructive comments and suggestions. This work was supported in part by the National Natural Science Foundation of China (81401410, 81527804, 11575285), the Natural Science Foundation of Guangdong Province in China (2017A030313743), the Shenzhen Overseas High-Level Talent Peacock Team of China (KQTD2016053117113327), the Basic Research Program of Shenzhen in China (JCYJ20160608153434110, JCYJ20150831154213680), the Applied Science and Technology Project of Guangdong Province in China (2015B020233014), the Key Laboratory for Magnetic Resonance and Multimodality Imaging of Guangdong Province (2014B030301013).
Author Contributions
Zhanli Hu and Juan Gao carried out the experiments and drafted the manuscript. Na Zhang participated in the design of the study and revised the manuscript. Zhanli Hu and Dong Liang conceived the approach and design of the study. Yongfeng Yang, Xin Liu and Hairong Zheng gave valuable advice. All authors reviewed the manuscript.
Competing Interests
The authors declare that they have no competing interests.
Footnotes
Zhanli Hu, Juan Gao and Na Zhang contributed equally to this work.
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Linton, O. W., A, F. & Mettler, J. In American Journal of Roentgenology181 321–329 (2003). [DOI] [PubMed]
- 2.Hu Z, et al. Geometric calibration of a micro-CT system and performance for insect imaging. IEEE transactions on information technology in biomedicine: a publication of the IEEE Engineering in Medicine and Biology Society. 2011;15:655–660. doi: 10.1109/TITB.2011.2159012. [DOI] [PubMed] [Google Scholar]
- 3.McCollough CH, et al. Strategies for Reducing Radiation Dose in CT. Radiol. Clin. N. Am. 2009;47 doi: 10.1016/j.rcl.2008.10.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Hu ZL, Zou J, Gui JB, Zheng HR, Xia D. Region-of-interest reconstruction for a cone-beam dental CT with a circular trajectory. Nucl Instrum Meth A. 2013;708:39–45. doi: 10.1016/j.nima.2013.01.003. [DOI] [Google Scholar]
- 5.Liu Y, Ma J, Fan Y, Liang Z. Adaptive-weighted total variation minimization for sparse data toward low-dose x-ray computed tomography image reconstruction. Phys Med Biol. 2012;57:7923–7956. doi: 10.1088/0031-9155/57/23/7923. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Hu Z, et al. Image reconstruction from few-view CT data by gradient-domain dictionary learning. J Xray Sci Technol. 2016;24:627–638. doi: 10.3233/XST-160579. [DOI] [PubMed] [Google Scholar]
- 7.Hu ZL, Liang D, Xia D, Zheng HR. Compressive sampling in computed tomography: Method and application. Nucl Instrum Meth A. 2014;748:26–32. doi: 10.1016/j.nima.2014.02.026. [DOI] [Google Scholar]
- 8.Wang G, Yu H. The meaning of interior tomography. Phys Med Biol. 2013;58:R161–186. doi: 10.1088/0031-9155/58/16/R161. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Li TF, et al. Nonlinear sinogram smoothing for low-dose X-ray CT. IEEE Trans. Nucl. Sci. 2004;51:2505–2513. doi: 10.1109/TNS.2004.834824. [DOI] [Google Scholar]
- 10.Wang J, Li T, Lu H, Liang Z. penalized weighted Least-Squares approach to sinogram noise reduction and image reconstruction for low-dose X-Ray computed Tomography. IEEE Trans Med Imaging. 2006;25(10):1272–1283. doi: 10.1109/TMI.2006.882141. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Elbakri, L. A. & Fessler, J. A. In Medical Imaging 2003 Conference. 1839–1850 (Spie-Int Soc Optical Engineering, 2003).
- 12.Sidky EY, Kao C-M, Pan X. Accurate image reconstruction from few-views and limited-angle data in divergent-beam CT. Journal of X-Ray Science and Technology. 2006;14:119–139. [Google Scholar]
- 13.Lauzier PT, Chen GH. Characterization of statistical prior image constrained compressed sensing (PICCS): II. Application to dose reduction. Medical physics. 2013;40 doi: 10.1118/1.4773866. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Niu S, et al. Sparse-view x-ray CT reconstruction via total generalized variation regularization. Phys Med Biol. 2014;59:2997–3017. doi: 10.1088/0031-9155/59/12/2997. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Panin VY, Zeng GL, Gullberg GT. Total variation regulated EM algorithm. IEEE Trans. Nucl. Sci. 1999;46:2202–2210. doi: 10.1109/23.819305. [DOI] [Google Scholar]
- 16.Ma JH, Feng QJ, Feng YQ, Huang J, Chen WF. Generalized Gibbs priors based positron emission tomography reconstruction. Comput. Biol. Med. 2010;40:565–571. doi: 10.1016/j.compbiomed.2010.03.012. [DOI] [PubMed] [Google Scholar]
- 17.Ma JH, et al. Iterative image reconstruction for cerebral perfusion CT using a pre-contrast scan induced edge-preserving prior. Phys Med Biol. 2012;57:7519–7542. doi: 10.1088/0031-9155/57/22/7519. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Hu Z, et al. A feature refinement approach for statistical interior CT reconstruction. Phys Med Biol. 2016;61:5311–5334. doi: 10.1088/0031-9155/61/14/5311. [DOI] [PubMed] [Google Scholar]
- 19.Wang J, et al. An experimental study on the noise properties of x-ray CT sinogram data in Radon space. Phys Med Biol. 2008;53:3327–3341. doi: 10.1088/0031-9155/53/12/018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Ma JH, et al. Variance analysis of x-ray CT sinograms in the presence of electronic noise background. Medical physics. 2012;39:4051–4065. doi: 10.1118/1.4722751. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Sukovic P, Clinthorne NH. Penalized weighted least-squares image reconstruction for dual energy X-ray transmission tomography. Ieee Transactions on Medical Imaging. 2000;19:1075–1081. doi: 10.1109/42.896783. [DOI] [PubMed] [Google Scholar]
- 22.Candes EJ, Romberg J, Tao T. Robust uncertainty principles: Exact signal reconstruction from highly incomplete frequency information. IEEE Trans. Inf. Theory. 2006;52:489–509. doi: 10.1109/TIT.2005.862083. [DOI] [Google Scholar]
- 23.Yuemin Z, Jianhua L, Qiegen L. Adaptive Image Decomposition by Improved Bilateral Filter. International Journal of Computer Applications. 2011;23:16–22. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The datasets generated or analyzed during the current study are available from the corresponding author on reasonable request.