Abstract
The FDA recently expanded the approved use of Dexcom’s G5 Mobile continuous glucose monitoring (CGM) system to allow for diabetes treatment decisions. This decision is expected to reduce the burden of SMBG testing and increase the adoption and persistent use of CGM. The safety of nonadjunctive CGM use was questioned because of sporadic large discrepancies between CGM and SMBG values. These data were viewed in the context of complaints found in the FDA MAUDE database and social media postings. This commentary provides additional perspective on the inferences that can be drawn from these reports and the risk of nonadjunctive use of CGM data.
Keywords: continuous glucose monitoring, nonadjunctive use, self-monitored blood glucose, FDA decisions
I read with interest the recent commentary regarding the safety of nonadjunctive use of continuous glucose monitoring (CGM) data for insulin dosing,1 and the now-resolved question2 of whether this use should be included among the labeled indications for the Dexcom G5 Mobile CGM System. Although I agree with some of the points made, I feel that several issues regarding device accuracy and the way in which CGM data are presented and used have not been sufficiently addressed.
Nonadjunctive Use of CGM
First, the accuracy of self-monitored blood glucose (SMBG) data deserves further scrutiny, and the possibility (and consequences) of egregiously inaccurate SMBG readings deserve acknowledgement. Current-generation SMBG devices perform well under ideal conditions.3 In the real world, however, deviations from best handwashing practices are common,4 and even under supervised conditions, many patient-owned SMBG devices do not meet ISO 15197 criteria.5 There are tens of thousands of inaccuracy complaints in the MAUDE database against meters—most of which are unsubstantiated, highlighting one of the limitations of the database.
Second, the focus on inaccuracy of selected CGM readings neglects the quantitative and qualitative differences between SMBG and CGM data. From the standpoint of data quantity, the default CGM screen view gives the current numerical estimate and 36 previous glucose values (shown as a trend graph) over the past 3 hours. In routine use, screen views occur an average of 29 ± 18 times per day.6 The availability of comprehensive glucose data allows CGM users to base management decisions on patterns, rather than just the current point estimate of glucose concentration. Rate of change (ROC) arrows provided on the CGM display may have a greater impact than the glucose value on the need to dose insulin and in determining insulin dose amount,7 and specific recommendations for basing treatment decisions on these ROC arrows have been published.8
Even when CGM data were limited to adjunctive-only use, off-label use was common: in one survey, half of the CGM users indicated that they would treat a nighttime low glucose alert without a confirmatory fingerstick, and 34% indicated that they would dose insulin for hyperglycemia without SMBG confirmation.7 A separate survey9 revealed that 69% of Dexcom CGM users regularly use CGM alone to adjust bolus insulin doses. None of these patients had the advantage of training on when, how, or when not to use CGM for dosing decisions due to the lack of a labeled indication.
Given the quantity and manner in which CGM data are presented, the premise that any particular CGM reading could form the basis of a treatment decision may not be fully justified. CGM data are interpreted in context and CGM users, whether they use the data adjunctively or nonadjunctively, also have the benefit of alerts which mitigate risks from transient inaccurate values. Further, CGM requires daily calibrations; each calibration can provide a user with reassurance or result in a potentially reportable complaint.
Third and most important, there is a mismatch between the question posed and the anecdotes reported. The question of whether requiring fingersticks impacts safety among CGM users were resolved by the REPLACE-BG study, results of which were recently published.10 Because the REPLACE-BG results may not be generalizable to youth, or to individuals with hypoglycemia unawareness, further studies are needed; however, simulations presented at the FDA advisory board meeting11 suggest that hypoglycemia-unaware individuals may particularly benefit from using the comprehensive CGM data.
CGM for Insulin Dosing
There are three main approaches to glucose monitoring to guide insulin dosing and diabetes management decisions. First, the decisions may be entirely based on fingerstick values. The low frequency of SMBG testing reported in the T1D Exchange Clinic Registry12 and resultant long between-test intervals limit the value of this approach. Problematic hypoglycemia and sustained hyperglycemia remain common in insulin users and data from many outcome studies13 and registries14 suggest that getting more people using CGM will improve glycemic control. The second approach in which CGM data are used as an adjunct to fingersticks maintains the requirement for frequent, painful, and burdensome blood glucose testing.15 This requirement likely contributes to the current, low penetration rate of CGM. The third approach—the option to use CGM data nonadjunctively—reflects current use and offers several benefits. I anticipate a reduction in the burden of fingerstick testing, increased adoption and persistent use of the technology, and development of training materials to discuss when and how best to use CGM.
While the author presented interesting data and interpretations, and we agree that the nonadjunctive use of Dexcom CGM is not without risk, I disagree with the recommendation for redundant glucose measurements and maintenance of the status quo requiring fingersticks for all diabetes management decisions. I believe this approach is overly conservative, and fails to account for the current realities around CGM use.
Footnotes
Abbreviations: CGM, continuous glucose monitoring; MAUDE, manufacturer and user facility device experience; ROC, rate of change; SG, sensor glucose; SMBG, self-monitored blood glucose.
Declaration of Conflicting Interests: The author(s) declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: The author is an employee of, and owns stock in, Dexcom, Inc.
Funding: The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: The author is an employee and shareholder of: Dexcom, Inc.
References
- 1. Shapiro AR. Nonadjunctive use of continuous glucose monitors for insulin dosing: is it safe? J Diabetes Sci Technol. 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. US Food and Drug Administration. FDA expands indication for continuous glucose monitoring system, first to replace fingerstick testing for diabetes treatment decisions [press release]. 2016. Available at: http://www.fda.gov/NewsEvents/Newsroom/PressAnnouncements/ucm534056.htm. Accessed January 11, 2017.
- 3. Bedini JL, Wallace JF, Pardo S, Petruschke T. Performance evaluation of three blood glucose monitoring systems using ISO 15197: 2013 accuracy criteria, consensus and surveillance error grid analyses, and insulin dosing error modeling in a hospital setting. J Diabetes Sci Technol. 2016;10(1):85-92. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Hortensius J, van der Bijl JJ, Kleefstra N, Houweling ST, Bilo HJ. Self-monitoring of blood glucose: professional advice and daily practice of patients with diabetes. Diabetes Educ. 2012;38(1):101-107. [DOI] [PubMed] [Google Scholar]
- 5. Boettcher C, Dost A, Wudy SA, et al. Accuracy of blood glucose meters for self-monitoring affects glucose control and hypoglycemia rate in children and adolescents with type 1 diabetes. Diabetes Technol Ther. 2015;17(4):275-282. [DOI] [PubMed] [Google Scholar]
- 6. Nakamura K, Walker T, Balo A. Patient data from RT-CGM suggests use of threshold alerts impacts glycemic control. Diabetes. 2016;65(S1):A223-224. [Google Scholar]
- 7. Pettus J, Price DA, Edelman SV. How patients with type 1 diabetes translate translate continuous glucose monitoring data into diabetes management decisions. Endocr Pract. 2015;21:613-620. [DOI] [PubMed] [Google Scholar]
- 8. Pettus J, Edelman SV. Recommendations for using real-time continuous glucose monitoring (rtCGM) data for insulin adjustments in type 1 diabetes. J Diabetes Sci Technol. 2017; 11(1):138-147. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. FDA advisory panel votes to recommend non-adjunctive use of Dexcom G5 Mobile CGM. Diabetes Technol Ther. 2016;18(8):512-516. [DOI] [PubMed] [Google Scholar]
- 10. Aleppo G, Ruedy KJ, Riddlesworth TD, et al. REPLACE-BG: A randomized trial comparing continuous glucose monitoring with and without routine blood glucose monitoring in well-controlled adults with type 1 diabetes. Diabetes Care. 2017;40(4):538–545 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Dexcom G5 mobile continuous glucose monitoring (CGM) system for non-adjunctive management of diabetes. Presentation to: Clinical Chemistry and Clinical Toxicology Devices Panel; July 21, 2016 Available at: http://www.fda.gov/downloads/AdvisoryCommittees/CommitteesMeetingMaterials/MedicalDevices/MedicalDevicesAdvisoryCommittee/ClinicalChemistryandClinicalToxicologyDevicesPanel/UCM512789.pdf. Accessed January 11, 2017. [Google Scholar]
- 12. Miller KM, Foster NC, Beck RW, et al. Current state of type 1 diabetes treatment in the U.S.: updated data from the T1D Exchange clinic registry. Diabetes Care. 2015;38(6):971-978. [DOI] [PubMed] [Google Scholar]
- 13. Soupal J, Petruzelkova L, Flekac M, et al. Comparison of different treatment modalities for type 1 diabetes, including sensor-augmented insulin regimens, in 52 weeks of follow-up: a COMISAIR study. Diabetes Technol Ther. 2016;18(9):532-538. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Foster NC, Miller KM, Tamborlane WV, Bergenstal RM, Beck RW, T1D Exchange Clinic Network. Continuous glucose monitoring in patients with type 1 diabetes using insulin injections. Diabetes Care. 2016;39(6):e81-82. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Edelman SV. Regulation catches up to reality: nonadjunctive use of continuous glucose monitoring data. J Diabetes Sci Technol. 2017;11(1):160-164. [DOI] [PMC free article] [PubMed] [Google Scholar]