Abstract
Purpose
To accelerate high-resolution diffusion imaging, rotating single-shot acquisition (RoSA) with composite reconstruction is proposed. Acceleration was achieved by acquiring only one rotating single-shot blade per diffusion direction, and high-resolution diffusion-weighted (DW) images were reconstructed by utilizing similarities of neighboring DW images. Parallel imaging technique was implemented in RoSA to further improve the image quality and acquisition speed. RoSA performance was evaluated by simulation and human experiments.
Methods
A brain tensor phantom was developed to determine an optimal blade size and rotation angle by considering similarity in DW images, off-resonance effects, and k-space coverage. With the optimal parameters, RoSA MR pulse sequence and reconstruction algorithm were developed to acquire human brain data. For comparison, multi-shot echo-planar imaging (EPI) and conventional single-shot EPI sequences were performed with matched scan time, resolution, FOV, and diffusion directions.
Results
The simulation indicated an optimal blade size of 48×256 and a 30° rotation angle. For 1×1 mm2 in-plane resolution, RoSA was 12 times faster than the multi-shot acquisition with comparable image quality. With the same acquisition time as SS-EPI, RoSA provided superior image quality and minimum geometric distortion.
Conclusion
RoSA offers fast high-quality high-resolution diffusion images. The composite image reconstruction is model-free and compatible with various diffusion computation approaches including parametric and non-parametric analyses.
Keywords: high resolution, diffusion imaging, dMRI, fast, single shot, blade
Introduction
Single-shot echo-planar imaging (SS-EPI) is the most common sequence for diffusion-weighted imaging (DWI), a powerful tool for assessing the neuronal microstructures in vivo [1–3]. Although SS-EPI has gained popularity due to its fast acquisition, its spatial resolution is often limited [4], making it difficult to measure accurate diffusion properties in fine structures that require high spatial resolution [5–7], such as gyral projections [6,7], or penetrating fibers between basal ganglia [8]. Even with advances in parallel imaging, SS-EPI at a high resolution is limited by geometric distortion and a low signal-to-noise ratio (SNR).
Multi-shot techniques have been developed to improve the resolution by filling in the k-space with multiple readouts at the expense of possible motion artifacts and phase inconsistencies between shots [9–19]. In this category, periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) is a robust technique that is particularly suitable for high-resolution DWI in the presence of patient motion [12,13]. It uses fast spin-echo (FSE) readouts, which are free from EPI artifacts. In addition, the motion artifacts may be corrected by its self-navigation property as the center of the k-space is repeatedly sampled in each echo train with a rotating blade trajectory. The drawbacks of PROPELLER are that it is much slower than EPI and is quite specific absorption rate (SAR)-intensive. Therefore, PROPELLER EPI sequences were proposed to improve upon PROPELLER FSE by replacing FSE readouts with EPI [14,15]. Unlike conventional SS-EPI, short-axis PROPELLER EPI features short echo spacing, which greatly alleviates all of the aforementioned issues in SS-EPI, including geometric distortion [15]. Although the imaging speed of PROPELLER EPI is faster than that of PROPELLER FSE, it is still much slower than SS-EPI due to its multi-shot nature. Imaging speed is a common limitation for extensive use of multi-shot techniques, especially in applications that require a large number of diffusion directions [20–24].
To increase imaging speed for multi-shot DWI, acquiring only a subset of PROPELLER FSE blades for each DW image and a tensor model based reconstruction was suggested in [25]. A similar idea was proposed with a different reconstruction scheme to accelerate PROPELLER EPI, in which 10 diffusion directions are acquired with one unique blade per direction, and a full resolution DW image is reconstructed using all of the blades [26]. Although both studies [25,26] reported encouraging results with reduced data acquisition, these approaches lack the flexibility for extension to arbitrary diffusion direction schemes larger than 10 direction or to diffusion models other than Diffusion Tensor Imaging (DTI). Moreover, both methods use data from all diffusion directions to reconstruct a single DW image. Therefore, the similarities of neighboring DW images are either not considered [25] or only partly considered [26]. DW images contain a similar diffusion contrast when their diffusion directions are close in a q-space sphere. To our knowledge, this similarity feature among neighboring diffusion directions has not yet been explored as a means to reduce data acquisition and to improve diffusion imaging speed.
By taking advantage of similarities in DW images, we propose a rotating single-shot acquisition (RoSA) approach with model-free reconstruction to accelerate data acquisition for high-resolution diffusion imaging. In the RoSA scheme, one single-shot blade is acquired per diffusion direction. Upon reconstruction of any DW image, only a subset of DW images whose diffusion directions are close to the current DW image on the q-space sphere is used. Neighboring DW images are expected to become highly correlated when the diffusion signals are finely sampled on the q-space sphere. An image reconstruction technique used in highly constrained back-projection (HYPR) was adapted to reconstruct the RoSA. HYPR is a conventional non-iterative reconstruction technique for undersampled radial acquisition in time-resolved MRI in which correlations exist in a one-dimensional time domain [27], whereas correlations in the RoSA occur in a three-dimensional q-space. Parallel imaging is implemented in RoSA to further accelerate its acquisition and improve image quality. Here we demonstrate the feasibility of RoSA based on evaluation of its performance in both simulations and human experiments.
Methods
RoSA acquisition
During data acquisition, one EPI blade with a short axis in the frequency encoding direction in k-space was acquired for each diffusion direction in q-space. The EPI blade rotated as diffusion direction changed in a sorted sequence (Figure 1A,B). Herein, we demonstrate an intuitive heuristic spiral ordering scheme. As illustrated in Figure 1A, for a given number of total diffusion directions, they were first sampled on the half sphere in q-space to maximize the angular sampling resolution. Then, diffusion directions were sorted into a sequence that was closest to an Archimedean spherical spiral curve with corresponding blades periodically rotating in this order. Figure 1B shows rotating blades in k-space and Figure 1C shows corresponding DW images. To avoid the edge effect around the equator where only the upper hemisphere has diffusion directions, we took advantage of the antipodal symmetry property of q-space, and mirrored the diffusion directions to create a whole sphere distribution (Figure 1D).
Reconstruction Window
For each diffusion direction to be reconstructed (green dot in Figure 1D), we searched its neighborhood to find a subset of DW blades each with a unique rotation angle (red dots in Figure 1D). Thus, on the q-space sphere each subset formed a reconstruction window (yellow discs in Figure 1D). Within the reconstruction window, “N” was defined as the number of diffusion directions, and “window size” as the distance between the two farthest points within the window (Figure 1E). The searching procedure was repeated till minimal window size was achieved for each diffusion direction, and high-resolution DW images were reconstructed with the windowed DW images through the composite reconstruction described below.
In theory, any diffusion direction in Figure 1A can be assigned with arbitrary rotating blade in Figure 1B. Such an assignment can be optimized to minimize the window size, therefore maximizing the similarity of DW images used for image reconstruction. To assess the goodness of the heuristic spiral-ordering scheme, we compared its averaged window size to those acquired from 1,000,000 trials of random assignments. Among the comparisons, the spiral-ordering scheme yielded the smallest overall window size (Supporting Figure S1).
Composite Reconstruction
The k-space blades within the reconstruction window were first combined to form a composite k-space data, which was then inversely Fourier-transformed (FT) to form a composite image. A final DW image was then reconstructed in the image space by constraining the composite image with the contrast provided by the current DW image. This reconstruction was adapted from HYPR [27] by replacing backprojection with FT. The full reconstruction comprised four computational steps that generated three intermediate images, Ii, IC, , and one final reconstructed image, , where i denotes the ith target diffusion direction and C denotes the composite image (Figure 2). Note that prior to the composite reconstruction, the k-space data were centered, regridded, and Nyquist ghost corrected (see below sections In-line Reference Scan and Image Preprocessing).
Step 1: Generate low-resolution DW images (Ii). The k-space blade of the target diffusion direction i (green dot) was first inversely fast Fourier-transformed (iFFT) to form a low-resolution DW image, Ii, followed by motion correction to a reference image b0 (i.e., b-value = 0 s/mm2) using a rigid body 6 DOF registration in FLIRT with correlation ratio as cost function (FSL).
Step 2: Generate high-resolution composite images (IC). The motion corrected k-space blades within the window were combined and density-corrected to form the composite k-space data followed by iFFT to yield the high-resolution composite image, IC.
Step 3: Generate scaling image ( ). The composite k-space data was cropped by the mask that matched the ith k-space blade and iFFT to yield the scaling image .
Step 4: Generate reconstructed image ( ). The high-resolution was reconstructed by tuning the composite image IC with the contrast specific to the target DW image. That is . As highlighted by arrows in Figure 2, the contrast that is unique to Ii is preserved in .
Simulations
To evaluate the RoSA approach, a high-quality, 1×1 mm2 in-plane resolution brain tensor phantom was developed using a diffusion dataset acquired with multi-shot readout segmented EPI sequence (RESOLVE). The tensor matrix of the brain was derived and could be used as a phantom for generating DW images for an arbitrary direction scheme. In the simulations, we generated DW images at 60 directions and undersampled the k-space data to form rotating blades using the spiral ordering scheme described above. We performed three simulations:
Simulation 1: The purpose of the first simulation was to demonstrate the performance of RoSA reconstruction in recovering full-resolution DW images and DTI fractional anisotropy (FA) without off-resonance effects. A blade size of 48×256 and 30 degree rotation angle for N=6 was used to generate the undersampled k-space. Full-resolution DW images were reconstructed with both RoSA reconstruction and simple zero-padding for comparison. The DTI metrics were estimated using a toolbox, DTIFIT, in FSL (http://fsl.fmrib.ox.ac.uk/fsl/fslwiki/) that fits a diffusion tensor model linearly at each voxel. Errors in FA and major eigenvector (EV1) were computed for quantitative comparisons. FA error was summarized as the 75th percentile of absolute difference to the ground truth. EV1 error was summarized as the 75th percentile of the angular difference (in degree) to the ground truth.
Simulation 2: The second simulation was performed to investigate the image quality in the presence of off-resonance effects with and without the parallel imaging technique. Parallel imaging is expected to reduce the geometric distortion associated with the off-resonance effects via reducing the field of view (FOV). We used the field map acquired for the same subject to simulate distortions in DW images, and used an echo spacing of 0.36 ms based on our gradient system (80 mT/m maximum amplitude and 200 T/m/s slew rate) and a blade size of 48×256 (readout×phase). The distortion was simulated in the image domain using the FUGUE tool in FSL where voxels were displaced based on the given field map. To simulate the effects of parallel imaging, we artificially reduced the echo spacing by the reduction factor (R), which would achieve the same effect as reducing the FOV in SENSitivity Encoding (SENSE) or GeneRalized Autocalibrating Partial Parallel Acquisition (GRAPPA) approaches. The distorted DW images were then fed to the RoSA pipeline for reconstruction, and DTI results were compared.
Simulation 3: The last simulation experiment was performed to investigate the optimal combination of blade width and N ranging from 2 to 12. In RoSA, the overall image quality is influenced by k-space coverage, and similarity among the DW images within the reconstruction window. The former is determined by blade width and N, and the latter is by N. We used FA error as a proxy for image quality to investigate combinations of blade width (32×256, 48×256, 64×256) and N. Similar to Simulation 2, the off-resonance effects were considered, and a reduction factor of R=3 was used for parallel imaging. Combinations with minimal FA errors were considered for subsequent human data acquisitions.
Human Experiments
A healthy volunteer was scanned at a 3 T system (MAGNETOM Prisma, Siemens Healthcare, Erlangen, Germany) with a 32-channel head coil. A prototype of RoSA sequence, combined with GRAPPA [28], was used for acquisition; Stejskal-Tanner pulsed-gradient spin-echo (PGSE) was used for diffusion weighting. The diffusion parameters were 4 b0s and 60 diffusion directions at b-value = 1000 s/mm2. RoSA parameters were: GRAPPA reduction factor R=3, matrix/blade size = 48×256, rotation angle = 30° (N=6), TE = 53 ms, FOV = 256 mm, 7/8 partial Fourier k-space sampling along the phase encoding (PE) direction, slice thickness = 2 mm, 10 slices, echo spacing = 0.36 ms, 2 averages. Total acquisition time was 2 min. The study was approved by our institutional review board, and informed consent was obtained for human studies.
For comparison, conventional SS-EPI diffusion data were acquired for the same volunteer at an in-plane resolution of 1×1 mm2 and 2×2 mm2 with a total acquisition time of 2 min and 1 min, respectively. Multi-shot sequence (RESOLVE) was performed at 1×1 mm2 with 13 shots and a total acquisition time of 24 min. Echo spacing and TE are listed in Figure 7B. Other acquisition parameters were matched with RoSA including GRAPPA reduction factor, slice thickness, slice number, FOV, partial k-space coverage, and number of diffusion directions.
In-line Reference Scan and Nyquist Ghost Correction
Nyquist or FOV/2 ghosts commonly exist in EPI due to misalignment of the odd and even echoes caused by time delays, eddy currents, and static field inhomogeneity during EPI readout [29]. This problem was not fully addressed in previous PROPELLER EPI-based techniques [14,15]. In contrast to conventional single reference approaches [30,31], an in-line reference scan in the EPI pulse sequence was embeded immediately after every 90° excitation with only 3 readout lines (2 odds and 1 even) [32], which rotated the same way as the readout blade. Thus, the misalignment can be computed for each blade and the ghost can be corrected accordingly (Supporting Figure S2).
Image Preprocessing prior to RoSA reconstruction
Following Nyquist ghost correction, ramp-sampling correction was then carried out with sinc interpolation [33]. The Nyquist and ramp-sampling corrected k-space data were then GRAPPA reconstructed followed with homodyne reconstruction [34] to recover the partial k-space sampling along the PE direction and produce a full k-space blade (48×256). At this point, one can obtain an image at full resolution in the PE direction. Because the homodyne reconstruction clears the phase in the image domain, this procedure also centers the k-space peaks between the blades, a crucial step for subsequent data regridding [35]. According to the angulation of individual blades, the echo-centered full blades were regridded to Cartesian grid points using Kaiser-Bessel kernel [36] with an inherent post-compensation density correction [37]. The Kaiser–Bessel kernel is a smooth low-pass window function whose tradeoffs between main lobe width and side lobe levels of the spectral leakage can be controlled. Such characteristics have been shown to provide a good interpolation quality that is superior to both nearest-neighbor and linear interpolation and is computational inexpensive. The preprocessed low resolution DW images were then sent to the composite reconstruction. The apodization induced by the convolution kernel from regridding was corrected afterwards. In parallel, b0 images were reconstructed similar to PROPELLER, as described in [15].
Results
Simulations
Simulation 1: A tensor phantom without off-resonance effects was generated (Figure 3A). DW images and FA of the simple zero-padding approach and the RoSA reconstruction were compared with the ground truth in Figure 3A. With only 1/8 k-space data, the zero-padded images were blurred (Figure 3B). In the RoSA reconstruction, however, high-resolution DW images and FA were recovered and were similar to the ground truth (Figure 3C). The 75th percentile of absolute FA error in the RoSA reconstruction was 0.02, which was six times smaller than 0.12 in zero-padding. In addition, the 75th percentile of EV1 error was 8° in RoSA and 32° in zero-padding. In RoSA, EV1 had higher error mostly in the cerebrospinal fluid where water diffuses isotropically.
Simulation 2: In the presence of off-resonance effects due to field inhomogeneity (ΔB0), RoSA FA showed blurring in the tip of the frontal lobes as highlighted by the red circle in Figure 4. The off-resonance caused image distortions in the low-resolution DW images (Ii), and the rotation of blades produced blurring and image errors. The errors reduced significantly by using parallel imaging R=3 (Figure 4, right column, bottom), and gyral fiber stalks were better delineated (Figure 4, right column, middle).
Simulation 3: The tradeoff between N and k-space coverage is shown in Figure 5 without off-resonance effects. For a fixed blade size (32×256), a reconstruction window comprising six blades (i.e., N=6) had a smaller window size compared to N=12 (Figure 5, yellow discs). Although with full k-space coverage, N=12 produced larger image errors than N=6 (Figure 5, DWI error and FA error). This demonstrates that a smaller window size, albeit partial k-space coverage, improves similarity among DW images within the reconstruction window; and hence, increases reconstructed image quality.
The optimal N for a given blade size can be identified via investigating image errors. FA errors are calculated for different combinations of blade size, number of blades in each reconstruction window (N), and the off-resonance effects (ΔB0). While the smallest global error occurred at 64×256 blade with N=4, the smallest local error in high ΔB0 region occurred at 48×256 with N=6 (Supporting Figure S3). Taking into account that TE was 9ms longer in a 64×256 blade, we moved forward with a blade size of 48×256 and N=6 for our human experiments.
Human Experiments
Rotating k-space blades (48×256) are shown in Figure 6A for six diffusion directions. The corresponding DW images via zero padding showed expected low resolution along the frequency encoding direction (Figure 6B). The RoSA reconstructed DW images (Figure 6C) not only recovered the image resolution along the frequency direction, but also maintained the diffusion contrast. Matching diffusion contrast can be appreciated in the RESOLVE DW images (Figure 6D) obtained with matched diffusion encoding scheme. While acquiring less k-space data by a factor of 5.3, RoSA produced comparable image quality, resolution, and diffusion contrast to that of RESOLVE by utilizing neighborhood similarities.
FA, EV1, and colormap are shown in Figure 7A for different methods: RoSA, RESOLVE, and conventional SS-EPI with matched resolution (1×1×2 mm3). Critical acquisition parameters are listed in Figure 7B. The DTI results from a shorter version of the RESOLVE sequence (i.e., RESOLVE-short) was added for comparisons with the matched scan time. RESOLVE-short was achieved by reducing the diffusion encoding directions from 60 to 6. Similar to Figure 6, RoSA produced comparable image quality in computed DTI metrics to RESOLVE. DTI images of SS-EPI and RESOLVE-short presented excessive noise due to long TE (101 ms) and severe undersampled q-space, respectively. With an advantage of shorter TE (~53 ms), RoSA and RESOLVE were able to show fine structures within the internal capsule (Figure 7A, top row, red arrows), and clear gyral stalks. In addition, images of RoSA, RESOLVE, and RESOLVE-short had minimum geometric distortions benefited from parallel imaging and very short echo spacing (~0.36 ms). In contrast, despite parallel imaging, long echo spacing (1.15 ms) in the SS-EPI sequence caused visible geometric distortions in the frontal lobes due to off-resonance effects of local field inhomogeneity (Figure 7A, 2nd and 3rd row, blue arrows). Further, orientations of major eigenvectors showed visible random variations using the RESOLVE-short and SS-EPI sequence (Figure 7A, bottom zoomed-in view), whereas major eigenvectors using RoSA and RESOLVE showed ordered alignments along the gyral stalks and perpendicular projections inside grey matter (Figure 7A, bottom row, yellow arrows). While RoSA and SS-EPI had same scan times, for the same brain coverage, the acquisition time of RESOLVE was 12 times longer.
Discussion
In this paper, we demonstrated the feasibility of RoSA, an innovative technique for fast high-resolution diffusion imaging. RoSA acquires one rotating blade per diffusion direction, and the high-resolution DW images are reconstructed with model-free composite reconstruction that takes advantage of the similarities between neighboring DW images. Similar to PROPELLER [12,38], RoSA has inherent navigation capabilities provided by the rotating blade trajectory that repeatedly samples the center of the k-space. As implemented with short-axis EPI readout, it also maintains the advantages of short-axis EPI over conventional SS-EPI in its reduced distortion, improved effective resolution in the phase-encoding direction associated with decay, and that geometric distortions do not scale with the target image resolution [15]. As a 2D acquisition technique, it can be easily combined with other fast-imaging techniques applied in the slice direction to further improve its imaging speed, such as the simultaneous multi-slice (SMS) [39] and generalized Slider method (gSlider-SMS) [40].
High-resolution diffusion imaging offers advantages in clinical diagnosis and basic research. For example, recurrent cholesteatoma may be detected earlier while the lesion is still small [41]; and small and short fibers may be detected to aid accuracy in white matter tractography: penetrating fibers between the space of extreme and external capsule [42], subcortical arcuate fibers [7], and gyral stalk projection fibers [5,6]. While conventional neuroimaging modalities such as T1-weighted and T2-weighted imaging routinely provide isotropic 1 mm resolution for whole-brain coverage, diffusion MRI is still restricted by coarse voxel sizes or very limited brain coverage. Compared to the low-resolution SS-EPI (2×2×2 mm3), the high-resolution FA map of RoSA (1×1×2 mm3) shows better separation between the external (Figure 8A, yellow arrows) and extreme capsules (Figure 8A, white arrows) and clearer fine structures in Figure 8B (white arrows). Such high-resolution images are currently available by the RESOLVE sequence in Siemens platforms, but with very limited brain coverage due to long scan time. With matched data quality shown in Figures 6–8, the RoSA approach provides 12 times faster imaging speed.
The efficiency in RoSA is achieved by utilizing the similarity in DW images whose directions are close. The closeness of DW images in each sliding-window has a direct impact on the image quality. That is for each diffusion direction, its “unique neighbors” (neighbors with unique rotation blades) cannot be far away. However, finding the global optimal solution in pairing diffusion directions and rotating blades is computationally challenging. We proposed an intuitive heuristic spiral ordering scheme and assigned rotating blades to a diffusion direction in this order. A similar problem was previously proposed and set up as a traveling salesman problem [26]. Only 10 total diffusion directions, however, were considered in that study where the window length (i.e., N) was constrained to be the same as the total number of directions. Such a fixed reconstruction window is difficult to generalize to other direction schemes with a flexible N. While a better scheme may exist and has yet to be explored, the current RoSA scheme provides desired flexibility and demonstrates to achieve comparable data quality.
The composite image is a key component in HYPR, an imaging technique that explores temporal redundancy in time-resolved MRI with radial sampling [27]. The composite image comprises a set of correlated frames that provide high-resolution structural information, and is used to constrain the reconstruction of an undersampled frame. In RoSA, the q-space redundancy can be viewed as analogous to temporal redundancy as both have correlated information in the 4th dimension (time domain vs. q-space). Therefore, the HYPR method can be readily adapted to reconstruct RoSA with a slight modification. In particular, as radial lines were replaced with rotating blades, we reconstruct the composite image by combining the data in k-space with density correction instead of filtered backprojection in image space. Projection reconstruction techniques have advantages over k-space methods with respect to diminished motion artifacts [43]. As rotating blades are self-navigating and a registration step is taken before combining in k-space, it is as robust to motion artifact as the PROPELLER technique [12]. Thus, by combining rotating blades with HYPR, RoSA preserves the merits of both.
The image quality of RoSA can be optimized by further improving k-space coverage and the off-resonance effects. As demonstrated in Simulation 3, wedges of k-space are not fully covered in order to guarantee a small window size (Figure 5,6). Although the effective resolution may be reduced due to the missing blades, a study showed that collecting every other blade only slightly decreases the SNR (from 18.5 to 18.2) with a mildly increased full-width half-maximum (FWHM) of the point spread function (PSF) (1.09 to 1.18) and non-significant artifacts [44]. Alternatively, an iterative reconstruction could be employed to further improve the situation [45].
Off-resonance artifact is a common problem in EPI-based acquisition techniques. Compared to conventional SS-EPI, echo spacing is reduced in RoSA and off-resonance artifacts are improved. In addition, parallel imaging was incorporated to further reduce the artifacts by a factor of R as has been implemented in the previous PROPELLER EPI-based techniques [46–47]. Note that depending on R and the geometric factor (g-factor), in parallel imaging, noise may be amplified. Because the HYPR reconstruction is a linear operation system ( ), the noise amplification of parallel imaging in RoSA ( ) is expected to be directly proportional to IC, which is reconstructed similarly to PROPELLER. PROPELLER accelerated via parallel imaging, SENSE and GRAPPA, has been studied and documented in [48] and [49] respectively. Alternatively, ultimate image quality may be achieved by a post-processing approach. The off-resonance field ∆B0 can be obtained via a set of reversed phase-encoding blades (i.e., an additional set of b0s acquired with reversed phase-encoding) similar to that described in [50] as implemented in FSL TOPUP command [51]. Using the off-resonance fieldmap, the TOPUP command produces corrected images and the artifacts can be corrected at the blade level.
Other common artifacts in EPI-based imaging are eddy current and chemical shift artifacts. For RoSA, the eddy current arising from the diffusion gradients may produce eddy current artifacts/distortions in DW images of individual blades. Since diffusion directions within a reconstruction window are very close, the eddy current effect is expected to be very similar. Therefore, it could be corrected in the same way as correcting conventional Cartesian DW images. Minimal eddy current effects were observed in this study; thus, no correction was performed. In addition, while not observed in this study, ghosting artifact related to eddy current can be present and cannot be corrected with current Nyquist ghost correction method as reference lines precede the diffusion gradients. Such eddy-current related ghosting may be corrected at the level of image reconstruction described in [52]. The chemical shift may produce “fat-rim” artifacts in DW images when fat suppression pulses such as Spectral Attenuated Inversion Recovery (SPAIR) and Spectral Presaturation with Inversion Recovery (SPIR) fail. While fat rim is easier to detect in routine Cartesian encoding DW images, the presence of residual fat signal is a challenge for radially acquired sequences as the fat rim is shifted radially, leading to blurring in the reconstructed images. Alternatively, Dixon fat suppression can be applied at the cost of some increase in TR and scan time. Dixon is a robust fat suppression method that estimates the fat signal instead of suppressing it [53,54]. Specifically, in a recent work [55], Dixon is successfully combined with PROPELLER; thus, could be implemented to improve fat suppression in RoSA.
While the RoSA relies on the similarity feature in diffusion imaging, its image reconstruction is model-free, which is independent of diffusion models or underlying assumptions. Because of high popularity, DTI was presented in this work. Nevertheless, RoSA supports all diffusion analyses including parametric diffusion modeling [24,56–58] and non-parametric q-space approaches [22,23,59,60]. A feature in RoSA is that the more diffusion directions, the better the reconstruction, as a smaller window size can be achieved and DW images become more correlated. This feature is consistent with the current trend of diffusion MRI where larger numbers of directions are desired. Notably, 60–80 diffusion directions are recommended in [62], and the human connectome project (HCP) standardized 90-direction schemes [61].
Conclusion
RoSA offers high resolution and high image quality with more than a 10-fold improvement in imaging efficiency. The imaging speed is achieved by utilizing similarities in neighboring DW images, whereas the high image quality is achieved by the nature of short-axis EPI blade and composite reconstruction. The RoSA approach is versatile and compatible with various diffusion analyses or modeling approaches. In addition, as a 2D acquisition technique, it has the flexibility of being combined with simultaneous multi-slice acceleration approaches.
Supplementary Material
Acknowledgments
This work was supported by Grants NIH R01AG053993, Dartmouth Synergy, and Indiana Alzheimer Disease Center pilot grant. The authors thank Drs. Andy L. Alexander and Charles A. Mistretta for inspiring discussions.
References
- 1.Basser PJ, Mattiello J, LeBihan D. MR diffusion tensor spectroscopy and imaging. Biophys J. 1994 Jan;66(1):259–67. doi: 10.1016/S0006-3495(94)80775-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Le Bihan D, Breton E, Lallemand D, Aubin ML, Vignaud J, Laval-Jeantet M. Separation of diffusion and perfusion in intravoxel incoherent motion MR imaging. Radiology. 1988 Aug;168(2):497–505. doi: 10.1148/radiology.168.2.3393671. [DOI] [PubMed] [Google Scholar]
- 3.Moseley ME, Cohen Y, Kucharczyk J, Mintorovitch J, Asgari HS, Wendland MF, Tsuruda J, Norman D. Diffusion-weighted MR imaging of anisotropic water diffusion in cat central nervous system. Radiology. 1990 Aug;176(2):439–45. doi: 10.1148/radiology.176.2.2367658. [DOI] [PubMed] [Google Scholar]
- 4.Farzaneh F, Riederer SJ, Pelc NJ. Analysis of T2 limitations and off-resonance effects on spatial resolution and artifacts in echo-planar imaging. Magn Reson Med. 1990 Apr;14(1):123–39. doi: 10.1002/mrm.1910140112. [DOI] [PubMed] [Google Scholar]
- 5.Miller KL, Stagg CJ, Douaud G, Jbabdi S, Smith SM, Behrens TE, Jenkinson M, Chance SA, Esiri MM, Voets NL, Jenkinson N, Aziz TZ, Turner MR, Johansen-Berg H, McNab JA. Diffusion imaging of whole, post-mortem human brains on a clinical MRI scanner. Neuroimage. 2011 Jul 1;57(1):167–81. doi: 10.1016/j.neuroimage.2011.03.070. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.McNab JA, Polimeni JR, Wang R, Augustinack JC, Fujimoto K, Stevens A, Triantafyllou C, Janssens T, Farivar R, Folkerth RD, Vanduffel W, Wald LL. Surface based analysis of diffusion orientation for identifying architectonic domains in the in vivo human cortex. Neuroimage. 2013 Apr 1;69:87–100. doi: 10.1016/j.neuroimage.2012.11.065. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Song AW, Chang HC, Petty C, Guidon A, Chen NK. Improved delineation of short cortical association fibers and gray/white matter boundary using whole-brain three-dimensional diffusion tensor imaging at submillimeter spatial resolution. Brain Connect. 2014 Nov;4(9):636–40. doi: 10.1089/brain.2014.0270. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Chang HC, Sundman M, Petit L, Guhaniyogi S, Chu ML, Petty C, Song AW, Chen NK. Human brain diffusion tensor imaging at submillimeter isotropic resolution on a 3Tesla clinical MRI scanner. Neuroimage. 2015 Sep;118:667–75. doi: 10.1016/j.neuroimage.2015.06.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Atkinson D, Porter DA, Hill DL, Calamante F, Connelly A. Sampling and reconstruction effects due to motion in diffusion-weighted interleaved echo planar imaging. Magn Reson Med. 2000 Jul;44(1):101–9. doi: 10.1002/1522-2594(200007)44:1<101::aid-mrm15>3.0.co;2-s. [DOI] [PubMed] [Google Scholar]
- 10.Atkinson D, Counsell S, Hajnal JV, Batchelor PG, Hill DL, Larkman DJ. Nonlinear phase correction of navigated multi-coil diffusion images. Magn Reson Med. 2006 Nov;56(5):1135–9. doi: 10.1002/mrm.21046. [DOI] [PubMed] [Google Scholar]
- 11.Bammer R, Stollberger R, Augustin M, Simbrunner J, Offenbacher H, Kooijman H, Ropele S, Kapeller P, Wach P, Ebner F, Fazekas F. Diffusion-weighted imaging with navigated interleaved echo-planar imaging and a conventional gradient system. Radiology. 1999 Jun;211(3):799–806. doi: 10.1148/radiology.211.3.r99jn15799. [DOI] [PubMed] [Google Scholar]
- 12.Pipe JG, Farthing VG, Forbes KP. Multi-shot diffusion-weighted FSE using PROPELLER MRI. Magn Reson Med. 2002 Jan;47(1):42–52. doi: 10.1002/mrm.10014. [DOI] [PubMed] [Google Scholar]
- 13.Pipe JG, Zwart N. Turboprop: improved PROPELLER imaging. Magn Reson Med. 2006 Feb;55(2):380–5. doi: 10.1002/mrm.20768. [DOI] [PubMed] [Google Scholar]
- 14.Wang FN, Huang TY, Lin FH, Chuang TC, Chen NK, Chung HW, Chen CY, Kwong KK. PROPELLER EPI: an MRI technique suitable for diffusion tensor imaging at high field strength with reduced geometric distortions. Magn Reson Med. 2005 Nov;54(5):1232–40. doi: 10.1002/mrm.20677. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Skare S, Newbould RD, Clayton DB, Bammer R. Propeller EPI in the other direction. Magn Reson Med. 2006 Jun;55(6):1298–307. doi: 10.1002/mrm.20890. [DOI] [PubMed] [Google Scholar]
- 16.Skare S, Newbould RD, Clayton DB, Albers GW, Nagle S, Bammer R. Clinical multi-shot DW-EPI through parallel imaging with considerations of susceptibility, motion, and noise. Magn Reson Med. 2007 May;57(5):881–90. doi: 10.1002/mrm.21176. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Porter DA, Heidemann RM. High resolution diffusion-weighted imaging using readout-segmented echo-planar imaging, parallel imaging and a two-dimensional navigator-based reacquisition. Magn Reson Med. 2009 Aug;62(2):468–75. doi: 10.1002/mrm.22024. [DOI] [PubMed] [Google Scholar]
- 18.Li Z, Pipe JG, Lee CY, Debbins JP, Karis JP, Huo D. X-PROP: a fast and robust diffusion-weighted propeller technique. Magn Reson Med. 2011 Aug;66(2):341–7. doi: 10.1002/mrm.23033. [DOI] [PubMed] [Google Scholar]
- 19.Jeong HK, Gore JC, Anderson AW. High-resolution human diffusion tensor imaging using 2-D navigated multi-shot SENSE EPI at 7 T. Magn Reson Med. 2013 Mar 1;69(3):793–802. doi: 10.1002/mrm.24320. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Tuch DS. Q-ball imaging. Magn Reson Med. 2004 Dec;52(6):1358–72. doi: 10.1002/mrm.20279. [DOI] [PubMed] [Google Scholar]
- 21.Tournier JD, Calamante F, Gadian DG, Connelly A. Direct estimation of the fiber orientation density function from diffusion-weighted MRI data using spherical deconvolution. Neuroimage. 2004 Nov;23(3):1176–85. doi: 10.1016/j.neuroimage.2004.07.037. [DOI] [PubMed] [Google Scholar]
- 22.Wu YC, Alexander AL. Hybrid diffusion imaging. Neuroimage. 2007 Jul 1;36(3):617–29. doi: 10.1016/j.neuroimage.2007.02.050. Epub 2007 Mar 24. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Wedeen VJ, Wang RP, Schmahmann JD, Benner T, Tseng WY, Dai G, Pandya DN, Hagmann P, D’Arceuil H, de Crespigny AJ. Diffusion spectrum magnetic resonance imaging (DSI) tractography of crossing fibers. Neuroimage. 2008 Jul 15;41(4):1267–77. doi: 10.1016/j.neuroimage.2008.03.036. [DOI] [PubMed] [Google Scholar]
- 24.Zhang H, Schneider T, Wheeler-Kingshott CA, Alexander DC. NODDI: practical in vivo neurite orientation dispersion and density imaging of the human brain. Neuroimage. 2012 Jul 16;61(4):1000–16. doi: 10.1016/j.neuroimage.2012.03.072. [DOI] [PubMed] [Google Scholar]
- 25.Cheryauka AB, Lee JN, Samsonov AA, Defrise M, Gullberg GT. MRI diffusion tensor reconstruction with PROPELLER data acquisition. Magn Reson Imaging. 2004 Feb;22(2):139–48. doi: 10.1016/j.mri.2003.08.001. [DOI] [PubMed] [Google Scholar]
- 26.Engstrom M, Nordell A, Nordell B, Skare S. Fast diffusion imaging using undersampled PROPELLER EPI. Proceedings of the 16th Annual Meeting ISMRM; Toronto, ON, Canada. 2008; p. 415. [Google Scholar]
- 27.Mistretta CA, Wieben O, Velikina J, Block W, Perry J, Wu Y, Johnson K, Wu Y. Highly constrained backprojection for time-resolved MRI. Magn Reson Med. 2006 Jan;55(1):30–40. doi: 10.1002/mrm.20772. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A. Generalized autocalibrating partially parallel acquisitions (GRAPPA) Magn Reson Med. 2002 Jun;47(6):1202–10. doi: 10.1002/mrm.10171. [DOI] [PubMed] [Google Scholar]
- 29.Bruder H, Fischer H, Reinfelder HE, Schmitt F. Image reconstruction for echo planar imaging with nonequidistant k-space sampling. Magn Reson Med. 1992 Feb;23(2):311–23. doi: 10.1002/mrm.1910230211. [DOI] [PubMed] [Google Scholar]
- 30.Reeder S, Atalar E, McVeigh E. Echo delay estimation in echo planar reference scans: implications for artifact reduction. Proceedings of the 4th Annual Meeting of ISMRM; New York, NY, USA. 1996; p. 1479. [Google Scholar]
- 31.Jesmanowicz A, Wong E, Hyde J. Phase correction for EPI using internal reference lines. Proceedings of the 12th Annual Meeting of SMRM; New York, NY, USA. 1993; p. 1239. [Google Scholar]
- 32.Oliver Heid. Method for the phase correction of nuclear magnetic resonance signals. 6043651 A. US. 2000 Mar
- 33.O’Sullivan JD. A fast sinc function gridding algorithm for fourier inversion in computer tomography. IEEE Trans Med Imaging. 1985;4(4):200–7. doi: 10.1109/TMI.1985.4307723. [DOI] [PubMed] [Google Scholar]
- 34.Pauly J. Partial k-space reconstruction. http//users.fmrib.ox.ac.uk/~karla/reading_group/lecture_notes/Recon_Pauly_read.pdf.
- 35.Peters DC, Derbyshire JA, McVeigh ER. Centering the projection reconstruction trajectory: reducing gradient delay errors. Magn Reson Med. 2003;50:1–6. doi: 10.1002/mrm.10501. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Beatty PJ, Nishimura DG, Pauly JM. Rapid gridding reconstruction with a minimal oversampling ratio. IEEE Trans Med Imaging. 2005 Jun;24(6):799–808. doi: 10.1109/TMI.2005.848376. [DOI] [PubMed] [Google Scholar]
- 37.Jackson JI, Meyer CH, Nishimura DG, Macovski A. Selection of a convolution function for Fourier inversion using gridding [computerized tomography application] IEEE Trans Med Imaging. 1991;10:473–478. doi: 10.1109/42.97598. [DOI] [PubMed] [Google Scholar]
- 38.Pipe JG. Motion correction with PROPELLER MRI: application to head motion and free-breathing cardiac imaging. Magn Reson Med. 1999 Nov;42(5):963–9. doi: 10.1002/(sici)1522-2594(199911)42:5<963::aid-mrm17>3.0.co;2-l. [DOI] [PubMed] [Google Scholar]
- 39.Setsompop K, Cohen-Adad J, Gagoski BA, Raij T, Yendiki A, Keil B, Wedeen VJ, Wald LL. Improving diffusion MRI using simultaneous multi-slice echo planar imaging. Neuroimage. 2012 Oct 15;63(1):569–80. doi: 10.1016/j.neuroimage.2012.06.033. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Setsompop K, Stockmann J, Fan Q, Witzel T, Wald LL. Generalized SLIce Dithered Enhanced Resolution Simultaneous MultiSlice (gSlider-SMS) to increase volume encoding, SNR and partition profile fidelity in high-resolution diffusion imaging. Proceedings of the 24th Annual Meeting ISMRM; Singapore. 2016; p. 607. [Google Scholar]
- 41.Schwartz KM, Lane JI, Bolster BD, Jr, Neff BA. The utility of diffusion-weighted imaging for cholesteatoma evaluation. AJNR Am J Neuroradiol. 2011 Mar;32(3):430–6. doi: 10.3174/ajnr.A2129. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Chang HC, Sundman M, Petit L, Guhaniyogi S, Chu ML, Petty C, Song AW, Chen NK. Human brain diffusion tensor imaging at submillimeter isotropic resolution on a 3Tesla clinical MRI scanner. Neuroimage. 2015 Sep;118:667–75. doi: 10.1016/j.neuroimage.2015.06.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Glover GH, Pauly JM. Projection reconstruction techniques for reduction of motion effects in MRI. Magn Reson Med. 1992 Dec;28(2):275–89. doi: 10.1002/mrm.1910280209. [DOI] [PubMed] [Google Scholar]
- 44.Arfanakis K, Tamhane AA, Pipe JG, Anastasio MA. k-space undersampling in PROPELLER imaging. Magn Reson Med. 2005 Mar;53(3):675–83. doi: 10.1002/mrm.20380. [DOI] [PubMed] [Google Scholar]
- 45.Tamhane AA, Arfanakis K, Anastasio M, Guo X, Vannier M, Gao JH. Rapid PROPELLER-MRI: a combination of iterative reconstruction and under-sampling. J Magn Reson Imaging. 2012 Nov;36(5):1241–7. doi: 10.1002/jmri.23720. [DOI] [PubMed] [Google Scholar]
- 46.Chuang TC, Huang TY, Lin FH, Wang FN, Juan CJ, Chung HW, Chen CY, Kwong KK. PROPELLER-EPI with parallel imaging using a circularly symmetric phased-array RF coil at 3.0 T: application to high-resolution diffusion tensor imaging. Magn Reson Med. 2006 Dec;56(6):1352–8. doi: 10.1002/mrm.21064. [DOI] [PubMed] [Google Scholar]
- 47.Skare S, Newbould RD, Nordell A, Holdsworth SJ, Bammer R. An auto-calibrated, angularly continuous, two-dimensional GRAPPA kernel for propeller trajectories. Magn Reson Med. 2008 Dec;60(6):1457–65. doi: 10.1002/mrm.21788. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Chang Y, Pipe JG, Karis JP, Gibbs WN, Zwart NR, Schär M. The effects of SENSE on PROPELLER imaging. Magn Reson Med. 2015 Dec;74(6):1598–608. doi: 10.1002/mrm.25557. [DOI] [PubMed] [Google Scholar]
- 49.Skare S, Newbould RD, Nordell A, Holdsworth SJ, Bammer R. An auto-calibrated, angularly continuous, two-dimensional GRAPPA kernel for propeller trajectories. Magn Reson Med. 2008 Dec;60(6):1457–65. doi: 10.1002/mrm.21788. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Andersson JL, Skare S, Ashburner J. How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging. Neuroimage. 2003 Oct;20(2):870–88. doi: 10.1016/S1053-8119(03)00336-7. [DOI] [PubMed] [Google Scholar]
- 51.Smith SM, Jenkinson M, Woolrich MW, Beckmann CF, Behrens TE, Johansen-Berg H, Bannister PR, De Luca M, Drobnjak I, Flitney DE, Niazy RK, Saunders J, Vickers J, Zhang Y, De Stefano N, Brady JM, Matthews PM. Advances in functional and structural MR image analysis and implementation as FSL. Neuroimage. 2004;23(Suppl 1):S208–19. doi: 10.1016/j.neuroimage.2004.07.051. [DOI] [PubMed] [Google Scholar]
- 52.Buonocore MH, Gao L. Ghost artifact reduction for echo planar imaging using image phase correction. Magn Reson Med. 1997 Jul;38(1):89–100. doi: 10.1002/mrm.1910380114. [DOI] [PubMed] [Google Scholar]
- 53.Dixon WT. Simple proton spectroscopic imaging. Radiology. 1984 Oct;153(1):189–94. doi: 10.1148/radiology.153.1.6089263. [DOI] [PubMed] [Google Scholar]
- 54.Sepponen RE, Sipponen JT, Tanttu JI. A method for chemical shift imaging: demonstration of bone marrow involvement with proton chemical shift imaging. J Comput Assist Tomogr. 1984;8:585–587. doi: 10.1097/00004728-198408000-00001. [DOI] [PubMed] [Google Scholar]
- 55.Schär M, Eggers H, Zwart NR, Chang Y, Bakhru A, Pipe JG. Dixon water-fat separation in PROPELLER MRI acquired with two interleaved echoes. Magn Reson Med. 2016 Feb;75(2):718–28. doi: 10.1002/mrm.25656. [DOI] [PubMed] [Google Scholar]
- 56.Assaf Y, Basser PJ. Composite hindered and restricted model of diffusion (CHARMED) MR imaging of the human brain. Neuroimage. 2005 Aug 1;27(1):48–58. doi: 10.1016/j.neuroimage.2005.03.042. [DOI] [PubMed] [Google Scholar]
- 57.Alexander DC, Hubbard PL, Hall MG, Moore EA, Ptito M, Parker GJ, Dyrby TB. Orientationally invariant indices of axon diameter and density from diffusion MRI. Neuroimage. 2010 Oct 1;52(4):1374–89. doi: 10.1016/j.neuroimage.2010.05.043. [DOI] [PubMed] [Google Scholar]
- 58.Wang Y, Wang Q, Haldar JP, Yeh FC, Xie M, Sun P, Tu TW, Trinkaus K, Klein RS, Cross AH, Song SK. Quantification of increased cellularity during inflammatory demyelination. Brain. 2011 Dec;134(Pt 12):3590–601. doi: 10.1093/brain/awr307. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Wu YC, Field AS, Alexander AL. Computation of diffusion function measures in q-space using magnetic resonance hybrid diffusion imaging. IEEE Trans Med Imaging. 2008 Jun;27(6):858–65. doi: 10.1109/TMI.2008.922696. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Jensen JH, Helpern JA. MRI quantification of non-Gaussian water diffusion by kurtosis analysis. NMR Biomed. 2010 Aug;23(7):698–710. doi: 10.1002/nbm.1518. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Sotiropoulos SN, Jbabdi S, Xu J, Andersson JL, Moeller S, Auerbach EJ, Glasser MF, Hernandez M, Sapiro G, Jenkinson M, Feinberg DA, Yacoub E, Lenglet C, Van Essen DC, Ugurbil K, Behrens TE, WU-Minn HCP Consortium Advances in diffusion MRI acquisition and processing in the Human Connectome Project. Neuroimage. 2013 Oct 15;80:125–43. doi: 10.1016/j.neuroimage.2013.05.057. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Jones DK, Knösche TR, Turner R. White matter integrity, fiber count, and other fallacies: the do’s and don’ts of diffusion MRI. Neuroimage. 2013 Jun;73:239–54. doi: 10.1016/j.neuroimage.2012.06.081. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.