Skip to main content
NIST Author Manuscripts logoLink to NIST Author Manuscripts
. Author manuscript; available in PMC: 2017 Sep 29.
Published in final edited form as: Metrologia. 2012;49(Technical Suppl):06006. doi: 10.1088/0026-1394/49/1A/06006

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NIST, USA and the BIPM in medium-energy x-rays

D T Burns 1, C Kessler 1, M O’Brien 2, R Minniti 2
PMCID: PMC5621609  NIHMSID: NIHMS905960  PMID: 28966399

Abstract

A key comparison has been made between the air-kerma standards of the NIST, USA and the BIPM in the medium-energy x–ray range. The results show the standards to be in agreement at the level of the standard uncertainty of the comparison of 3.8 parts in 103, except at 250 kV where the difference is 1.5 times the standard uncertainty. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database.

1. Introduction

An indirect comparison has been made between the air-kerma standards of the National Institute of Standards and Technology (NIST), USA, and the Bureau International des Poids et Mesures (BIPM) in the x-ray range from 100 kV to 250 kV. Three cavity ionization chambers were used as transfer instruments. The measurements at the BIPM took place in May 2016 using the reference conditions recommended by the Consultative Committee for Ionizing Radiation (CCRI) (CCEMRI 1972). Final results were supplied by the NIST in November 2016.

2. Determination of the air-kerma rate

For a free-air ionization chamber standard with measuring volume V, the air-kerma rate is determined by the relation

K.=IρairVWaire11-gairiki (1)

where ρair is the density of air under reference conditions, I is the ionization current under the same conditions, Wair is the mean energy expended by an electron of charge e to produce an ion pair in air, gair is the fraction of the initial electron energy lost through radiative processes in air, and Π ki is the product of the correction factors to be applied to the standard.

The values used for the physical constants ρair and Wair/e are given in Table 1. For use with this dry-air value for ρair, the ionization current I must be corrected for humidity and for the difference between the density of the air of the measuring volume at the time of measurement and the value given in the table1.

Table 1.

Physical constants used in the determination of the air-kerma rate

Constant Value uia
ρairb 1.2930 kg m−3 0.0001
Wair/e 33.97 J C−1 0.0015
a

ui is the relative standard uncertainty.

b

Density of dry air at T0 = 273.15 K and P0 = 101.325 kPa adopted at both laboratories.

3. Details of the standards

Both free-air chamber standards are of the conventional parallel-plate design. The measuring volume V is defined by the diameter of the chamber aperture and the length of the collecting region. The BIPM air-kerma standard is described in Boutillon (1978) and the changes made to certain correction factors in Burns (2004), Burns et al (2009) and the references therein. The NIST Wyckoff-Attix standard is described in Wyckoff and Attix (1957) and Lamperti and O’Brien (2001) and was previously compared with the BIPM standard in an indirect comparison carried out in 2003, the results of which are reported in Burns and O’Brien (2006). The main dimensions, the measuring volume and the polarizing voltage for each standard are shown in Table 2.

Table 2.

Main characteristics of the standards

Standard BIPM M-01 NIST Wyckoff-Attix
Aperture diameter/mm 9.939 9.999
Air path length/mm 281.5 308
Collecting length/mm 60.004 100.8
Electrode separation/mm 180 200
Collector width/mm 200 268
Measuring volume/mm3 4655.4 7915
Polarizing voltage/V +4000 −5000

4. The transfer instruments

4.1 Determination of the calibration coefficient for a transfer instrument

The air-kerma calibration coefficient NK for a transfer instrument is given by the relation

NK=K.Itr (2)

where is the air-kerma rate determined by the standard using (1) and Itr is the ionization current measured by the transfer instrument and the associated current-measuring system. The current Itr is corrected to the standard conditions of air temperature and pressure chosen for the comparison (T = 293.15 K, P = 101.325 kPa). No humidity correction has been applied to the current measured using the transfer instruments, on the basis that both measurement laboratories are operated with a relative humidity in the range from 35 % to 65 %.

To derive a comparison result from the calibration coefficients NK,BIPM and NK,NMI measured, respectively, at the BIPM and at a national metrology institute (NMI), differences in the radiation qualities must be taken into account. Normally, each quality used for the comparison has the same nominal generating potential at each institute, but the half-value layers (HVLs) may differ. A radiation quality correction factor kQ is derived for each comparison quality Q. This corrects the calibration coefficient NK,NMI determined at the NMI into one that applies at the ‘equivalent’ BIPM quality and is derived by interpolation of the NK,NMI values in terms of log(HVL). The comparison result at each quality is then taken as

RK,NMI=kQNK,NMINK,BIPM (3)

In practice, the half-value layers normally differ by only a small amount and kQ is close to unity.

4.2 Details of the transfer instruments

Three spherical cavity ionization chambers belonging to the NIST were used as transfer instruments for the comparison, the same three chambers used in the 2003 comparison. Their main characteristics are given in Table 3.

Table 3.

Main characteristics of the transfer chambers

Chamber type Shonka-Wyckoff Shonka-Wyckoff Exradin A3
Serial number 2022 2023 260
Geometry spherical spherical spherical
External diameter/mm 19.1 19.1 19.5
Wall material C552 C552 C552
Wall thickness/mm 0.25 0.25 0.25
Nominal volume/cm3 3.6 3.6 3.6
Reference point centre of sphere centre of sphere centre of sphere
Polarizing potential/V −300a −300a −300a
a

Potential of the chamber wall with respect to the central electrode.

5. Calibration at the BIPM

5.1 The BIPM irradiation facility and reference radiation qualities

The BIPM medium-energy x-ray laboratory houses a high-stability generator and a tungsten-anode x-ray tube with a 3 mm beryllium window. An aluminium filter of thickness 2.228 mm is added (for all radiation qualities) to compensate for the decrease in attenuation that occurred when the original BIPM x-ray tube (with an aluminium window of approximately 3 mm) was replaced in June 2004. Two voltage dividers monitor the tube voltage and a voltage-to-frequency converter combined with data transfer by optical fibre measures the anode current. No transmission monitor is used. For a given radiation quality, the standard uncertainty of the distribution of repeat air-kerma rate determinations over many months is better than 3 parts in 104. The radiation qualities used in the range from 100 kV to 250 kV are those recommended by the CCRI (CCEMRI 1972) and are given in Table 4.

Table 4.

Characteristics of the BIPM reference radiation qualities

Radiation quality 100 kV 135 kV 180 kV 250 kV
Generating potential/kV 100 135 180 250
Inherent Be filtration/mm 3 3 3 3
Additional Al filtration/mm 3.431 2.228 2.228 2.228
Additional Cu filtration/mm - 0.232 0.485 1.570
Al HVL/mm 4.030 - - -
Cu HVL/mm 0.149 0.489 0.977 2.484
(μ/ρ)air a/cm2 g−1 0.290 0.190 0.162 0.137
BIPM/mGy s−1 0.50 0.50 0.50 0.50
a

Measured at the BIPM for an air path length of 280 mm.

The irradiation area is temperature controlled at around 20 °C and is stable over the duration of a calibration to around 0.2 °C. Two calibrated thermistors measure the temperature of the ambient air and the air inside the BIPM standard (which is controlled at 25 °C). Air pressure is measured by means of a calibrated barometer positioned at the height of the beam axis. The relative humidity is controlled within the range 40 % to 50 %.

5.2 The BIPM standard and correction factors

The reference plane for the BIPM standard was positioned at 1200 mm from the radiation source, with a reproducibility of 0.03 mm. The standard was aligned on the beam axis to an estimated uncertainty of 0.1 mm. The beam diameter in the reference plane is 98 mm for all radiation qualities. During the calibration of the transfer chambers, measurements using the BIPM standard were made using positive polarity only. A correction factor of 1.00015 is applied to correct for the known polarity effect in the standard. The leakage current for the BIPM standard, relative to the ionization current, was measured to be around 1 part in 104.

The correction factors applied to the ionization current measured at each radiation quality using the BIPM standard, together with their associated uncertainties, are given in Table 5. The factor ka corrects for the attenuation of the x-ray fluence along the air path between the reference plane and the centre of the collecting volume. It is evaluated using the measured mass attenuation coefficients for air given in Table 4. In practice, the values used for ka take account of the temperature and pressure of the air in the standard. Ionization current measurements (both for the standard and for transfer chambers) are also corrected for changes in air attenuation arising from variations in the temperature and pressure of the ambient air between the radiation source and the reference plane.

Table 5.

Correction factors for the BIPM M-01 standard

Radiation quality 100 kV 135 kV 180 kV 250 kV uiA uiB
Air attenuation kaa 1.0099 1.0065 1.0055 1.0047 0.0002 0.0001
Photon scatter ksc 0.9952 0.9959 0.9964 0.9974 - 0.0003
Fluorescence kfl 0.9985 0.9992 0.9994 0.9999 - 0.0003
Electron loss ke 1.0000 1.0015 1.0047 1.0085 - 0.0005
Ion recombination ks 1.0010 1.0010 1.0010 1.0010 0.0002 0.0001
Polarity kpol 1.0002 1.0002 1.0002 1.0002 0.0001 -
Field distortion kd 1.0000 1.0000 1.0000 1.0000 - 0.0007
Diaphragm correction kdia 0.9995 0.9993 0.9991 0.9980 - 0.0003
Wall transmission kp 1.0000 1.0000 0.9999 0.9988 0.0001 -
Humidity kh 0.9980 0.9980 0.9980 0.9980 - 0.0003
Radiative loss 1 – gair 0.9999 0.9999 0.9998 0.9997 - 0.0001
a

Values for the BIPM reference conditions of 293.15 K and 101.325 kPa; each measurement is corrected using the air density measured at the time.

5.3 Transfer chamber positioning and calibration at the BIPM

The reference point for each chamber was positioned in the reference plane (1200 mm from the radiation source), with a reproducibility of 0.03 mm. Each transfer chamber was aligned on the beam axis to an estimated uncertainty of 0.1 mm. The leakage current was measured before and after each series of ionization current measurements and a correction made using the mean value. The relative leakage current for each transfer chamber was below 1 part in 104.

For each transfer chamber and at each radiation quality, two sets of seven measurements were made, each measurement with integration time 60 s. The relative standard uncertainty of the mean ionization current for each pair was below 2 parts in 104. Repeat calibrations for all three chambers (after repositioning) showed a reproducibility below 2 parts in 104, which is included in Table 11 for the short-term reproducibility of the calibration coefficients determined at the BIPM.

Table 11.

Uncertainties associated with the calibration of the transfer chambers

Institute BIPM NIST
Relative standard uncertainty uiA uiB uiA uiB
std 0.0004 0.0019 0.0025 0.0028
Positioning of transfer chamber 0.0001 - - 0.0001
Itr 0.0002 0.0002 0.0003 0.0006
Short-term reproducibility 0.0002 - - a -
NK,std 0.0005 0.0019 0.0026 0.0029
a

The reproducibility of the NIST transfer chamber calibrations over the duration of the comparison is implicitly included in str in Table 9 and consequently in str,comp in Table 12.

6. Calibration at the NIST

6.1 The NIST irradiation facility and reference radiation qualities

The medium-energy x-ray facility at the NIST comprises a constant-potential generator and a tungsten-anode x-ray tube with an inherent filtration of 3 mm beryllium. No transmission monitor is used. The characteristics of the NIST realization of the CCRI comparison qualities (CCEMRI 1972) are given in Table 6.

Table 6.

Characteristics of the NIST reference radiation qualities

Radiation quality 100 kV 135 kV 180 kV 250 kV
Generating potential/kV 100 135 180 250
Additional Al filtration/mm 3.248 1.060 3.842 3.842
Additional Cu filtration/mm - 0.265 0.482 1.618
Al HVL/mm 3.943 - - -
Cu HVL/mm 0.149 0.496 1.003 2.502
(μ/ρ)air a/cm2 g−1 0.409 0.225 0.212 0.149
NIST/mGy s−1 0.97 0.96 1.19 1.50
a

Measured at the NIST for an air path length of 308 mm.

A calibrated thermistor was used to measure the ambient air temperature. Air pressure was recorded using a calibrated barometer positioned at the approximate height of the beam axis. The relative humidity in the NIST measurement area was controlled in the range from 35 % to 55 %.

6.2 The NIST standard and correction factors

The defining plane for the NIST standard was positioned at 1000 mm from the radiation source, with a reproducibility of 0.1 mm. The standard was aligned on the beam axis to an estimated uncertainty of 0.1 mm. The beam diameter in the reference plane is 70 mm for all radiation qualities. During the calibration of the transfer chambers, measurements using the NIST standard were made using negative polarity only. No polarity correction has been applied. An uncertainty component of 1 part in 103 is included for the polarity effect. The relative leakage current was measured to be around 1 part in 104.

The correction factors applied to the ionization current measured at each radiation quality using the NIST standard, together with their associated uncertainties, are given in Table 7. The correction factor ka is evaluated using the measured mass attenuation coefficients for air given in Table 6; note that these values are for the NIST reference temperature of 295.15 K. In practice, the values used for ka take account of the temperature and pressure of the air in the standard at the time of the measurements.

Table 7.

Correction factors for the NIST standard

Radiation quality 100 kV 135 kV 180 kV 250 kV uiA uiB
Air attenuation kaa 1.0152 1.0083 1.0079 1.0055 0.0012 0.0002
Photon scatter ksc 0.9942 0.9952 0.9958 0.9969 - 0.0007
Fluorescence kfl 0.9981 0.9991 0.9995 0.9999 - 0.0003
Electron loss ke 1.0000 1.0006 1.0027 1.0055 - 0.0005
Ion recombination ks 1.0004 1.0004 1.0004 1.0004 0.001 -
Polarity kpol 1.0000 1.0000 1.0000 1.0000 0.001 -
Field distortion kd 1.0015 1.0015 1.0015 1.0015 - 0.002
Aperture edge transmission kl 1.0000 1.0000 1.0000 1.0000 - 0.0004
Wall transmission kp 1.0000 1.0000 1.0000 1.0000 - 0.0001
Humidity kh 0.9980 0.9980 0.9980 0.9980 - 0.0003
Radiative loss 1 – gair 1.0000 1.0000 1.0000 1.0000 - 0.0001
a

Values for the NIST reference conditions of 295.15 K and 101.325 kPa; each measurement is corrected using the air density measured at the time.

6.3 Transfer chamber positioning and calibration at the NIST

The reference point for each transfer chamber was positioned at the reference distance (at the NIST 1000 mm from the radiation source), with a reproducibility of 0.1 mm. Alignment on the beam axis was to an estimated uncertainty of 0.1 mm. The leakage current was measured before and after each series of ionization current measurements and a correction made using the mean value. The relative leakage current for all three transfer chambers was less than 1 part in 104. The relative standard uncertainty of the calibration coefficient for each transfer chamber at each radiation quality was typically 3 parts in 104.

7. Additional corrections to transfer chamber measurements

7.1 Ion recombination, polarity, radial non-uniformity, distance and field size

As can be seen from Tables 4 and 6, the air-kerma rates at the NIST are up to three times higher than those at the BIPM. Thus volume recombination effects will be greater for the transfer chamber calibrations at the NIST, although no recombination corrections have been applied at either laboratory. Additional measurements at the BIPM using the Exradin chamber at different air-kerma rates showed the effect to be not more than 2 parts in 104 when increasing from 0.2 mGy s−1 to 0.6 mGy s−1. Based on these results, an uncertainty component of 5 parts in 104 is included in Table 12.

Table 12.

Uncertainties associated with the comparison results

Relative standard uncertainty uiA uiB
NK,NIST/NK,BIPM 0.0026 0.0025a
Ion recombination in transfer chambers - 0.0005
Field size/distance - 0.0005
kQ - 0.0004
Transfer chambers str,comp - 0.0009
RK,NIST uc = 0.0038
a

Takes account of correlation in type B uncertainties as noted in Section 9.

Each transfer chamber was used with the same polarity at each laboratory and so no corrections are applied for polarity effects in the transfer chambers. No correction is applied at either laboratory for the radial non-uniformity of the radiation fields. For the spherical chambers used, the effective diameter2 is similar to the diameter of the free-air chamber apertures used and the relative correction required would be less than 1 part in 104. No additional uncertainty component is included.

The reference distance is 1000 mm at the NIST and 1200 mm at the BIPM, and the field diameter is smaller, 70 mm at the NIST and 98 mm at the BIPM. It is known that transfer chambers respond to scattered radiation in a way that free-air chambers do not, so that calibration coefficients can show some sensitivity to field size. However, the difference in field size is not large and the magnitude of field-size effects, at least for thimble chambers, is relatively small. An uncertainty component of 5 parts in 104 is introduced in Table 12 for this effect.

7.2 Radiation quality correction factors kQ

As noted in Section 4.1, slight differences in radiation qualities might require a correction factor kQ, depending on the energy response of the transfer chamber. From Tables 4 and 6 it is evident that the BIPM and NIST radiation qualities are reasonably matched in terms of HVL, and the energy dependence of the transfer chambers is sufficiently small, such that the required correction is at most 5 parts in 104 with a mean value of 2 parts in 104. Consequently, kQ is taken to be unity for all chambers and qualities, with a standard uncertainty of 4 parts in 104 included in Table 12.

8. Comparison results

The calibration coefficients NK,NIST and NK,BIPM for the transfer chambers are presented in Table 8. The values NK,NIST measured before and after the measurements at the BIPM give rise to the relative standard uncertainties str,1, str,2 and str,3 for the three chambers, which represent the uncertainty in NK arising from transfer chamber stability.

Table 8.

Calibration coefficients for the transfer chambers

Radiation quality 100 kV 135 kV 180 kV 250 kV
Shonka 2022
NK,NIST (pre-comp)/Gy μC−1 8.341 8.415 8.541 8.661
NK,NIST (post-comp)/Gy μC−1 8.329 8.394 8.518 8.636
str,1 (relative) a 0.0009 0.0016 0.0018 0.0019
NK,BIPM/Gy μC−1 8.360 8.435 8.553 8.698
Shonka 2023
NK,NIST (pre-comp)/Gy μC−1 8.420 8.464 8.581 8.692
NK,NIST (post-comp)/Gy μC−1 8.415 8.458 8.569 8.675
str,2 (relative) a 0.0004 0.0005 0.0009 0.0013
NK,BIPM/Gy μC−1 8.446 8.498 8.605 8.739
Exradin A3 260
NK,NIST (pre-comp)/Gy μC−1 7.941 8.023 8.117 8.180
NK,NIST (post-comp)/Gy μC−1 7.944 8.030 8.112 8.181
str,3 (relative) a 0.0002 0.0006 0.0004 0.0001
NK,BIPM/Gy μC−1 7.943 8.042 8.131 8.223
a

For each pre-post pair of NK,NIST values with half-difference d, the standard uncertainty of the mean is taken to be str,i = d/√(n–1.4), where the term (n–1.4) is found empirically to be a better choice than (n–1) to estimate the standard uncertainty for low values of n. For n = 2, str,i = 1.3d.

For each chamber at each radiation quality, the mean of the NIST results before and after the BIPM measurements is used to evaluate the comparison results NK,NIST/NK,BIPM given in Table 9. The final results RK,NIST in Table 9 are evaluated as the mean for the three transfer chambers. For each quality, the corresponding uncertainty str is the standard uncertainty of this mean (using again the choice (n–1.4) introduced in the footnote to Table 8), or taken as

Table 9.

Comparison results

Radiation quality 100 kV 135 kV 180 kV 250 kV
NK,NIST/NK,BIPM using Shonka 2022 0.9970 0.9964 0.9973 0.9943
NK,NIST/NK,BIPM using Shonka 2023 0.9966 0.9956 0.9965 0.9936
NK,NIST/NK,BIPM using Exradin A3 260 0.9999 0.9981 0.9980 0.9948
str 0.0012 0.0008 0.0007 0.0008
RK,NIST 0.9978 0.9967 0.9973 0.9942
Previous results for RK,NIST 1.0030 1.0002 1.0021 1.0004
str=(str,12+str,22+str,32)/3 (4)

if this is larger (on the basis that the agreement between the comparison results for transfer chambers should, on average, not be better than their combined stability estimated using str,1, str,2 and str,3 from Table 8). The mean value of str for the four qualities, str,comp = 0.0009, is a global representation of the comparison uncertainty arising from the transfer chambers and is included in Table 12.

Also given in Table 9 are the results of the previous comparison of the NIST and BIPM standards (Burns and O’Brien 2006), revised for the published changes made to the BIPM standard in 2003 (Burns 2004) and in 2009 (Burns et al 2009).

9. Uncertainties

The uncertainties associated with the primary standards are listed in Table 10 and those for the transfer chamber calibrations in Table 11. The combined standard uncertainty uc for the comparison results RK,NIST is presented in Table 12. This combined uncertainty takes into account correlation in the type B uncertainties associated with the physical constants and the humidity correction. Correlation in the values for ke, ksc and kfl, derived from Monte Carlo calculations in each laboratory, are taken into account in an approximate way by assuming half of the uncertainty value for each factor at each laboratory. This is consistent with the analysis of the results of BIPM comparisons in medium-energy x-rays in terms of degrees of equivalence described in Burns (2003).

Table 10.

Uncertainties associated with the standards

Standard BIPM NIST
Relative standard uncertainty uiA uiB uiA uiB
Ionization current 0.0002 0.0002 0.0017 0.0006
Volume 0.0001 0.0005 0.0004 0.0001
Positioning 0.0001 0.0001 - 0.0001
Correction factors (excl. kh) 0.0003 0.0010 0.0019 0.0023
Humidity kh - 0.0003 - 0.0003
Physical constants - 0.0015 - 0.0015
std 0.0004 0.0019 0.0025 0.0028

10. Discussion

The comparison results presented in Table 9 show the NIST and BIPM standards to be in agreement at the level of the standard uncertainty of the comparison of 3.8 parts in 103, except at 250 kV where the difference is 1.5 times the standard uncertainty. There is nothing in the Monte Carlo corrections for ke, ksc and kfl to indicate why the result is lower at 250 kV. A set of BIPM calculations for the NIST standard gives values that are within 2 parts in 104 for all three parameters at all four energies. Furthermore, the NIST corrections for aperture and wall transmission are unity and if any correction was indeed required it would further reduce the NIST standard at 250 kV.

The only remaining parameter known to have an impact on the energy dependence is the air attenuation correction. It is notable that the NIST values for (μ/ρ)air in Table 6 (unchanged from the values used for the 2003 comparison) are higher, particularly at 100 kV, than those measured at the BIPM (given in Table 4) and at other NMIs with similar beam qualities in terms of the HVL in copper. It can be shown that the use of the BIPM (μ/ρ)air values for the NIST standard yields comparison results RK,NIST of 0.994 to 0.995 for all four qualities. While better in terms of energy dependence, these results are farther from unity (although still within two standard uncertainties).

The present results are lower than those obtained during the 2003 comparison, shown in the final row of Table 9, by 4 to 6 parts in 103. The results from 2003 have been corrected for the changes made to the BIPM standard in the intervening period. During this period the NIST standard has been reduced by 1 part in 103 arising from a different treatment of the polarity effect. The remaining reduction in the results of 3 to 5 parts in 103 between 2003 and the present comparison remains unexplained.

11. Degrees of Equivalence

The analysis of the results of BIPM comparisons in medium-energy x-rays in terms of degrees of equivalence is described in Burns (2003). Following a decision of the CCRI, the BIPM determination of the air-kerma rate is taken as the key comparison reference value, for each of the CCRI radiation qualities. It follows that for each laboratory i having a BIPM comparison result xi with combined standard uncertainty ui, the degree of equivalence with respect to the reference value is the relative difference Di = (KiKBIPM,i)/KBIPM,i = xi – 1 and its expanded uncertainty Ui = 2 ui. The results for Di and Ui, expressed in mGy/Gy and including those of the present comparison, are shown in Table 13 and in Figure 1, which include the linked results of the corresponding regional key comparisons APMP.RI(I)-K3 (Lee et al 2008) and SIM.RI.(I)– K3 (O’Brien et al 2015).

Table 13.

Degrees of equivalence. For each laboratory i, the degree of equivalence with respect to the key comparison reference value is the difference Di and its expanded uncertainty Ui Tables formatted as they appear in the BIPM key comparison database; Inline graphic indicates paricipation in Inline graphic in Inline graphic and Inline graphic in Inline graphic.

100 kV 135 kV 180 kV 250 kV





Lab i Di Ui Di Ui Di Ui Di Ui
I(mGy/Gy) I(mGy/Gy) I(mGy/Gy) I(mGy/Gy)





NPL −0.1 6.4 0.5 6.4 1.3 6.4 −0.7 6.4





NLM 4.3 6.0 1.9 6.0 2.0 6.0 0.3 6.0





LNE-LNHB 0.4 7.8 1.2 7.8 −0.1 7.8 −2.0 7.8





GUM 0.9 56 3.8 56 3.5 5.6 4.5 5.6





ARPANSA 3.7 7.6 5.6 7.6 6.0 7.6 5.3 7.6





MKEH −0.4 6.8 0.9 6.8 0.4 6.8 0.4 6.8





VNIIM 1.4 3.6 1.8 3.6 2.6 3.6 2.6 3.6





PTB 2.7 5.0 4.5 5.0 4.9 5.0 5.5 5.0





ENEA 3.9 6.2 4.2 6.2 7.3 6.2 5.6 6.2





BEV 3.2 6.4 4.7 6.4 4.1 6.4 1.1 6.4





NRC 3.1 6.6 2.3 6.6 1.3 6.6 0.4 6.6





NMIJ −0.8 6.2 −1.4 6.2 −2.4 6.2 −3.7 6.2





VSL −1.0 6.4 −0.4 6.4 0.0 6.4 −2.1 6.4





NIST −2.2 7.6 −3.3 7.6 −2.7 7.6 −5.8 7.6





INER 3.7 7.8 4.3 7.8 6.0 7.8 5.5 7.8





Nuc. Malaysia 14.2 12.8 16.2 12.8 15.0 12.8 15.9 12.8





DMSc −3.1 13.4 4.2 13.4 9.6 13.4 13.0 13.4





BARC 8.5 15.2 14.8 15.2





NMISA 4.5 5.6 2.0 5.6 4.8 5.6 7.5 5.6





KRISS −8.4 5.2 1.1 5.2 6.6 5.2 7.6 5.2





IAEA 4.3 7.4 9.2 7.4 13.1 7.4 14.0 7.4





CNEA −6.0 14,3 1.1 14.3 2.1 14.3 1.4 14.3





LMNRI/IRD −9.5 12.1 −9.4 12.1 −8.0 12.1 −8.5 12.1





ININ −9.3 16.1 −12.1 16.1 −11.1 16.1 −12.0 16.1

Figure 1.

Figure 1

Degrees of equivalence for each laboratory i with respect to the key comparison reference value. Results to the left are for the ongoing international comparison BIPM.RI(I)-K3, those in the middle section for the regional comparison APMP.RI(I)-K3 and those to the right for the regional comparison SIM.RI(I)-K3.

When required, the degree of equivalence between two laboratories i and j can be evaluated as the difference Dij = DiDj and its expanded uncertainty Uij = 2uij, both expressed in mGy/Gy. In evaluating uij, account should be taken of correlation between ui and uj (Burns 2003).

12. Conclusions

The key comparison BIPM.RI(I)-K3 for the determination of air kerma in medium-energy x–rays shows the NIST and BIPM standards to be in agreement at the level of the standard uncertainty of the comparison of 3.8 parts in 103, except at 250 kV where the difference is 1.5 times the standard uncertainty. These results are 4 to 6 parts in 103 lower than those obtained for the previous comparison in 2003.

Tables and graphs of degrees of equivalence, including those for the NIST, are presented for entry in the BIPM key comparison database. Note that the data presented in the tables, while correct at the time of publication of the present report, become out of date as laboratories make new comparisons with the BIPM. The formal results under the CIPM MRA are those available in the BIPM key comparison database (KCDB 2016).

Footnotes

1

For an air temperature T ~ 293 K, pressure P and relative humidity ~50 % in the measuring volume, the correction for air density involves a temperature correction T/T0, a pressure correction P0/P and a humidity correction kh = 0.9980. At the BIPM, a factor 1.0002 is also included to account for the compressibility of dry air between T ~ 293 K and T0 = 273.15 K.

2

It can be shown that beam non-uniformity for spherical chambers can be treated as for a flat chamber (or a free-air chamber aperture) by replacing the true diameter with an effective diameter that is smaller by the factor √(2/5).

References

  1. Boutillon M. Mesure de l’exposition au BIPM dans le domaine des rayons X de 100 à 250 kV. Rapport BIPM-78/3 1978 [Google Scholar]
  2. Burns DT. Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays. Metrologia. 2003;40(Tech Suppl):06036. [Google Scholar]
  3. Burns DT. Changes to the BIPM primary air-kerma standards for x-rays. Metrologia. 2004;41:L3. doi: 10.1088/0026-1394/49/1A/06006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Burns DT, O’Brien M. Comparison of the NIST and BIPM standards for air kerma in medium-energy x-rays. J Res Natl Inst Stand Technol. 2006;111:385–91. doi: 10.6028/jres.111.028. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Burns DT, Kessler C. Diaphragm correction factors for free-air chamber standards for air kerma in x-rays. Phys Med Biol. 2009;54:2737–45. doi: 10.1088/0031-9155/54/9/009. [DOI] [PubMed] [Google Scholar]
  6. Burns DT, Kessler C, Allisy PJ. Re-evaluation of the BIPM international standards for air kerma in x-rays. Metrologia. 2009;46:L21–23. [Google Scholar]
  7. CCEMRI. Qualités de rayonnement Comité Consultatif pour les Étalons de Mesures des Rayonnements Ionisants (Section I) 2nd meeting; 1972. pp. R15–16. [Google Scholar]
  8. KCDB. The BIPM key comparison database is available. 2016 online at http://kcdb.bipm.org/
  9. Lamperti P, O’Brien M. NIST Measurement Services: Calibration of X-ray and gamma-ray measuring instruments. NIST Special Publication. 2001:250–58. [Google Scholar]
  10. Lee JH, Hwang WS, Kotler LH, Webb DV, Büermann L, Burns DT, Takeyeddin M, Shaha VV, Srimanoroth S, Meghzifene A, Hah SH, Chun KJ, Kadni TB, Takata N, Msimang Z. APMP/TCRI key comparison report of measurement of air kerma for medium-energy x-rays (APMP.RI(I)-K3) Metrologia. 2008;45(Tech Suppl):06012. [Google Scholar]
  11. O’Brien M, McCaffrey J, Shen H, Saraví M, Stefanic A, Montaño Ortiz G, Carlos M, da Silva C, Álvarez J, Tovar V. SIM.RI(I)-K3 comparison of calibration coefficients at radiotherapy level for orthovoltage x-ray beams. Metrologia. 2015;52(Tech Suppl):06015. [Google Scholar]
  12. Wyckoff HO, Attix FH. National Bureau of Standards Handbook. 1957 Design of free-air ionization chambers; p. 64. [Google Scholar]

RESOURCES