Evaluation of the Rapid Polymyxin NP Test for Polymyxin B Resistance Detection Using Enterobacter cloacae and Enterobacter aerogenes Isolates

Shelby Simar 1, Diane Sibley 1, Deborah Ashcraft 1, George Pankey 1,
Editor: Nathan A Ledeboer2
PMCID: PMC5625387  PMID: 28747375

ABSTRACT

Polymyxin resistance is an increasing problem worldwide. Currently, determining susceptibility to polymyxins is problematic and lengthy. Polymyxins diffuse poorly into agar, potentially giving inaccurate disk diffusion and Etest results. A rapid screening test (2 h) for the detection of polymyxin resistance in Enterobacteriaceae, developed by P. Nordmann and L. Poirel (rapid polymyxin NP test) in 2016, detects glucose metabolization in the presence of polymyxin E (PE) and PB via pH-induced color change. The sensitivity and specificity were 99.3 and 95.4%, respectively, with results obtained in ≤2 h. Our goal was to evaluate this test using PB against larger numbers of Enterobacter. A total of 143 nonduplicate Enterobacter isolates (102 E. cloacae complex, 41 E. aerogenes) were tested, including 136 collected from Ochsner Health System patients from March to May 2016 and 7 previously determined PB-resistant E. cloacae isolates from JMI Laboratories. MICs were determined via broth microdilution. For the rapid polymyxin NP test, a color change from orange to yellow is positive; a weak/no color change is deemed negative after 4 h. Of 143 Enterobacter isolates, 25 were determined to be PB resistant by broth microdilution (MIC > 2 μg/ml), including all 7 JMI isolates. Of these 25, 7 were positive by the rapid polymyxin NP test (included 3/7 JMI isolates). All 118 isolates determined to be PB susceptible by broth microdilution were NP test negative. The sensitivity and specificity for the rapid polymyxin NP test were 25 and 100%, respectively, compared to broth microdilution. Although the rapid polymyxin NP test is a much faster method (2 to 4 h) for polymyxin resistance determination compared to broth microdilution (16 to 20 h), our study indicates that it may be subject to limitations when testing Enterobacter.

KEYWORDS: microbial drug resistance, Enterobacter, polymyxin B, drug resistance

INTRODUCTION

The increase in antimicrobial resistance has created significant challenges for appropriate treatment. Rapid diagnostics for adequate antibiotic stewardship of antibiotics are needed (1).

With the 2016 discovery of the first plasmid-mediated colistin (polymyxin E) resistance gene, mcr-1, reported in the United States (2), it is now more important than ever to develop rapid methods to detect resistance to the polymyxins, the antibiotics of last resort for many multidrug-resistant Gram-negative bacteria. Susceptibility testing of polymyxins is challenging and time-consuming. These drugs diffuse poorly into agar, potentially giving inaccurate disk diffusion and Etest results (35). In addition, heteroresistant subpopulations found in Enterobacter species can cause “skipped wells” in broth microdilution, as well as growth inside the zone of inhibition with disk diffusion and Etest methods, leading to uninterpretable results (6). A rapid screening test (≤4 h) for detection of polymyxin resistance in Enterobacteriaceae, developed by P. Nordmann and L. Poirel (rapid polymixin NP test), detects glucose metabolization in the presence of a polymyxin via an orange-to-yellow color change of the pH indicator phenol red, signifying resistance to polymyxins (7). Their evaluation reported the test to be 99.3% sensitive and 95.4% specific for 200 Enterobacteriaceae isolates, including 22 Enterobacter (18 Enterobacter cloacae, 3 E. aerogenes, and 1 E. asburiae), compared to standard broth microdilution. Results were determined in <2 h. Nordmann and Poirel reported that their study was subject to several limitations, including the facts that (i) they did not assess the ability of the rapid polymyxin NP test to detect heteroresistant isolates with low polymyxin MICs (such isolates may be problematic to detect) and (ii) larger-scale studies by others were needed to fully evaluate the reliability of the new test's performance (7). The goal of the present study was to use this novel test with a larger collection of Enterobacter isolates, which are known to exhibit heteroresistance when performing susceptibility testing with polymyxins.

(Data from this study were presented at the 35th Annual Meeting of Southwestern Association of Clinical Microbiology, Fort Worth, TX, 7 to 10 September 2016.)

RESULTS

Rapid polymyxin NP tests were interpreted according to the original protocol: a color change from orange to yellow in the well containing the isolate suspension with polymyxin B indicated a positive result (polymyxin B resistance), and no or an incomplete color change indicated a negative result (polymyxin B susceptibility) (Fig. 1). Isolates that yielded a positive NP test result and a broth microdilution polymyxin B MIC > 2 μg/ml or a negative NP test and a polymyxin B MIC ≤ 2 μg/ml were considered in agreement (Table 1). Using these interpretations, 25/143 Enterobacter isolates were determined to be polymyxin B resistant by broth microdilution (24 E. cloacae complex, 1 E. aerogenes), which included the 7 previously determined polymyxin B-resistant JMI E. cloacae isolates. Seven of the 25 polymyxin B-resistant isolates were also NP test positive: 6 E. cloacae complex and 1 E. aerogenes (included 3/7 JMI isolates). Of these isolates, positive NP test results were observed at 1 h (n = 1), at 2 h (n = 2), at 3 h (n = 2), and at 4 h (n = 2). However, the remaining 18 polymyxin B-resistant E. cloacae complex isolates were NP test negative (false-negative results). Broth microdilution tests and NP tests (using fresh reagents) were repeated in parallel for these polymyxin B-resistant/NP test-negative isolates, and the results matched those of the original tests. All 118 Enterobacter isolates that were susceptible to polymyxin B by broth microdilution (MIC ≤ 2 μg/ml) produced a negative rapid polymyxin NP test result. There were no false-positive NP test results. Skipped wells in broth microdilution (Fig. 2) were seen with 22 E. cloacae complex isolates, all of which were NP test negative, and 11 of which were polymyxin B resistant; these tests were repeated and yielded the same MIC results. For our study, the test sensitivity and specificity were 25 and 100%, respectively, compared to broth microdilution. This is markedly different from Nordmann and Poirel's original findings of 100% sensitivity and specificity for the 22 Enterobacter isolates included in their study. We speculate that this difference may be attributable to the larger number of Enterobacter, including possible heteroresistant isolates, tested in our study.

FIG 1.

FIG 1

Results of the rapid polymyxin NP test. The first column consists of uninoculated wells (NaCl in place of isolate suspension) with or without polymyxin B (PB) in the indicator solution, and the second column consists of wells with the isolate suspension with or without polymyxin B in the indicator solution. Each square represents an individual test. The bottom square shows an incomplete color change in the bottom right corner (reported as a negative NP test or PB susceptible).

TABLE 1.

Polymyxin B broth microdilution and rapid polymyxin NP test results

PB broth microdilution (MIC)a Rapid polymyxin NP test (no. of results)
Positive Negative Total
PB resistant, non-WT (>2 μg/ml) 7 18 25
PB susceptible, WT (≤2 μg/ml) 0 118 118
Total 7 136 143
a

PB, polymyxin B; WT, wild type.

FIG 2.

FIG 2

Broth microdilution wells in polymyxin B MIC (μg/ml) panel. The results represent the positive growth control (POS; rightmost well); growth in 0.125 and 0.25 wells; no growth in 0.5, 1.0, and 2.0 wells (which represents the skipped-well phenomenon); and growth in 4.0, 8.0, and 16.0 wells. The test was determined invalid and must be repeated when multiple skipped wells are present.

DISCUSSION

The new rapid polymyxin NP test may be subject to important limitations when testing Enterobacter species. We agree with Nordmann and Poirel's speculation that heteroresistant isolates may be problematic to detect. The skipped wells we observed, a phenomenon often seen with Enterobacter in broth microdilution testing (6), as well as our false-negative (negative NP test result but resistant by broth microdilution), could possibly be attributed to isolate cluster-dependent heteroresistance described by Guérin et al. (8). Testing to determine cluster association and the genetic mechanism of polymyxin B heteroresistance for these isolates was not performed. In addition, a weak or incomplete color change was observed in 15 isolates, 4 of which were determined to be polymyxin B resistant by broth microdilution (Fig. 1). All isolates with a weak or incomplete color change were determined to be negative upon repetition of the rapid NP test (no color change). Personal communication with P. Nordmann confirmed that any intermediate color change should be interpreted as a negative NP test result.

Furthermore, the polymyxin B-resistant Enterobacter isolates with a positive NP test result did not show the complete color change (positive result) until 3 to 4 h, compared to ≤2 h reported by Nordmann and Poirel. Warming of the NP indicator solution in a 37°C water bath during setup did not decrease the time in which results were seen. Overall, the rapid polymyxin NP test was easy to perform and yielded results much faster (3 to 4 h) than traditional broth microdilution (16 to 20 h). Our study indicates that this test may be subject to important limitations when testing Enterobacter species. Further testing with Enterobacter on a larger scale is needed.

MATERIALS AND METHODS

Bacterial isolates, media, and antimicrobial susceptibility testing.

A total of 143 clinical Enterobacter isolates from individual patients (102 E. cloacae complex, 41 E. aerogenes) from urine (54%), wound/abscess (22%), respiratory system (7%), skin/tissue (5%), blood (3%), and other sources (9%) were tested. Isolates included 136 collected from Ochsner Health System patients from March to May 2016 and 7 genetically unique (as determined by Rep-PCR; Diversilab, bioMérieux, Inc., Durham, NC) polymyxin B-resistant E. cloacae isolates obtained from JMI Laboratories (North Liberty, IA). Identification of Enterobacter isolates was performed using matrix-assisted laser desorption/ionization-time of flight mass spectrometry (Bruker Daltonics, Inc., Billerica, MA). Isolates were subcultured to BBL Trypticase soy agar with 5% sheep blood (Becton-Dickinson Diagnostic Systems, Sparks, MD) and tested after overnight incubation at 35°C.

Broth microdilution was performed for determination of polymyxin B MICs in cation-adjusted Mueller-Hinton II broth (Becton-Dickinson Diagnostic Systems) according to Clinical and Laboratory Standards Institute guidelines (9). A direct colony suspension in Mueller-Hinton broth using 18 to 24 h growth was adjusted to the turbidity of a 0.5 McFarland standard (1 × 108 to 2 × 108 CFU/ml). The suspension was diluted with sterile saline, and the 96-well polystyrene panel MIC panel was inoculated to give a final concentration of approximately 5 × 105 CFU/ml in each well. Panels were incubated 16 to 20 h at 35°C in ambient air. Polymyxin B sulfate powder (Sigma-Aldrich Inc., St. Louis, MO) was used, and the concentration range tested was 0.125 to 128 μg/ml. Escherichia coli ATCC 25922 and Pseudomonas aeruginosa ATCC 27853 were included as control strains on each day of testing (10). In addition, two E. coli isolates (0346 and 0349), possessing the mcr-1 gene for colistin resistance, were obtained from the FDA-CDC Antimicrobial Resistance Isolate Bank and included as controls. In the absence of polymyxin B interpretations for the Enterobacteriaceae, MICs were interpreted according to the colistin epidemiological cutoff value for Enterobacteriaceae of ≤2 μg/ml (10). Broth microdilution and the rapid NP test were performed on the same day using the same suspension of organism.

Rapid polymyxin NP test.

According to the rapid polymyxin NP test protocol, a rapid polymyxin NP solution containing 6.25 g of cation-adjusted Mueller-Hinton II powder, 0.0125 g of phenol red (Sigma-Aldrich), and 225 ml of distilled water was prepared, and the solution was adjusted to a pH of 6.7 with 1 mol/liter HCl and then autoclaved at 121°C for 15 min. After cooling, 25 ml of filter-sterilized 10% anhydrous D(+)-glucose (Sigma-Aldrich) was added. The solution was made fresh each week, 250 ml each time, and it was prewarmed to 37°C before each use to prevent delays in bacterial growth and color change. Because of the volume of isolates being tested, the solution being used was kept warm in a water bath during setup of the tests. Stock solution of polymyxin B was added to a portion of the rapid polymyxin NP solution before each batch of isolates was tested, for a final polymyxin B concentration of 3.75 μg/ml in each applicable well. Each isolate tested was set up alongside uninoculated wells, using saline in place of a bacterial isolate suspension, with or without polymyxin B as built-in negative controls. In addition, a well containing the isolate suspension without polymyxin B was used as a built-in positive growth control. Bacterial suspensions were prepared in 10 ml of sterile 0.85% saline to obtain turbidity corresponding to a 3.0 to 3.5 McFarland standard. The final concentration of each isolate suspension per well was ∼108 CFU/ml. The same procedure was used for the positive (Proteus vulgaris ATCC 6380) and negative (E. coli ATCC 25922) controls, which were included each day of testing. In addition, two E. coli mcr-1 isolates (0346 and 0349) were included as controls. The tests were incubated uncovered in ambient air at 35°C and observed at 1, 2, 3, and 4 h for color changes.

ACKNOWLEDGMENTS

We thank Ronald Jones (JMI Laboratories, North Liberty, IA) for providing polymyxin B-resistant E. cloacae isolates and the FDA-CDC Antimicrobial Resistance Bank for providing the E. coli isolates carrying the mcr-1 gene.

This research received no specific grant from any funding agency in the public, commercial, or not-for-profit sectors.

REFERENCES

  • 1.Marston HD, Dixon DM, Knisely JM, Palmore TN, Fauci AS. 2016. Antimicrobial resistance. JAMA 316:1193–1204. doi: 10.1001/jama.2016.11764. [DOI] [PubMed] [Google Scholar]
  • 2.McGann P, Snesrud E, Maybank R, Corey B, Ong AC, Clifford R, Hinkle M, Whitman T, Lesho E, Schaecher KE. 2016. Escherichia coli harboring mcr-1 and blaCTX-M on a novel IncF plasmid: first report of mcr-1 in the United States. Antimicrob Agents Chemother 60:4420–4421. doi: 10.1128/AAC.01103-16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Hindler J, Humphries R. 2013. Colistin MIC variability by method for contemporary clinical isolates of multidrug-resistant gram-negative bacilli. J Clin Microbiol 51:1678–1684. doi: 10.1128/JCM.03385-12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Humphries R. 2015. Susceptibility testing of the polymyxins: where are we now? Pharmacotherapy 35:22–27. doi: 10.1002/phar.1505. [DOI] [PubMed] [Google Scholar]
  • 5.Poirel L, Jayol A, Nordmann P. 2017. Polymyxins: antibacterial activity, susceptibility testing, and resistance mechanisms encoded by plasmids or chromosomes. Clin Microbiol Rev 30:557–596. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Landman D, Salamera J, Quale J. 2013. Irreproducible and uninterpretable polymyxin B MICs for Enterobacter cloacae and Enterobacter aerogenes. J Clin Microbiol 51:4106–4111. doi: 10.1128/JCM.02129-13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Nordmann P, Jayol A, Poirel L. 2016. Rapid detection of polymyxin resistance in Enterobacteriaceae. Emerg Infect Dis 6:1038–1043. doi: 10.3201/eid2206.151840. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Guérin F, Isnard C, Sinel C, Morand P, Dhalluin A, Cattoir V, Giard J. 2016. Cluster-dependent colistin hetero-resistance in Enterobacter cloacae complex. J Antimicrob Chemother 71:3058–3061. doi: 10.1093/jac/dkw260. [DOI] [PubMed] [Google Scholar]
  • 9.Clinical and Laboratory Standards Institute. 2015. Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically; approved standard, 10th ed Document M07-A10. CLSI, Wayne, PA. [Google Scholar]
  • 10.Clinical and Laboratory Standards Institute. 2016. Performance standards for antimicrobial susceptibility testing, 26th ed, supplement M100S CLSI, Wayne, PA. [Google Scholar]