Abstract
Objective
The aim of this pilot study was to compare spirometric values obtained with different types of spirometers, spirometers of same type, and repeated measurements with the same spirometer in a pulmonary function laboratory setting.
Results
12 healthy volunteers performed spirometry on four hot-wire (SensorMedics), two ultrasonic (Spirare) and one wedge-bellows (Vitalograph S) spirometers, according to ATS/ERS (American Thoracic Society/European Respiratory Society) guidelines. Spirometric values were compared using linear mixed models analysis with a random intercept for subjects and a fixed effect for type of spirometer used. Confidence intervals and p values were adjusted for multiple comparisons. Mean ± SD (L) values for hot-wire, ultrasonic and wedge-bellows spirometers for FVC (forced vital capacity) were 4.02 ± 0.66, 3.69 ± 0.61 and 3.93 ± 0.69, and for FEV1 (forced expiratory volume in one second) 3.06 ± 0.44, 2.95 ± 0.44 and 3.10 ± 0.49. Significant differences were found between hot-wire and ultrasonic and between wedge-bellows and ultrasonic spirometers for FVC and FEV1, and between hot-wire and wedge-bellows spirometers for FVC but not for FEV1. There were no significant differences between spirometers of same type, and low mean differences in repeated measurements for all spirometers included. In conclusion, the pilot study shows systematically higher values for FVC and FEV1 for hot-wire and wedge-bellows compared to ultrasonic spirometers.
Keywords: Spirometry, Spirometer comparison, Ultrasonic, Hot-wire, Wedge-bellows
Introduction
Spirometry is an important tool to assist diagnosis, detect severity of lung disease, follow disease development, determine effects of and changes in treatment, and to assess preoperative risk [1]. Thus, accurate and reliable results from spirometry testing are necessary for optimal treatment of most patients with lung disease. Multiple factors are known to influence lung function testing, including conditions involving the patient, the instructor and the equipment used [2–7]. With the variety of spirometric equipment used across and within laboratories, differences in measurements between spirometers can be a challenge in patient diagnostics and follow-up, and in research studies. Furthermore, despite meeting the ATS/ERS recommendations when tested with a standard forcing function like a waveform generator, spirometers may exhibit differences in measurements when applied to patients in the clinics. Our aim was therefore to compare the spirometer-subject system performance for different spirometers in a pulmonary function laboratory setting. We compared spirometric values from healthy subjects obtained with different types of spirometers, spirometers of same type, and repeated measurements with the same spirometer.
Main text
Methods
We performed a quality assurance study at the Respiratory Physiology Laboratory, Haukeland University Hospital, Bergen, where 12 healthy volunteers performed spirometry on 7 different spirometers within a period of 14 days in 2012. The study was remit assessed by the Regional Ethics Committee (REK Vest; http://helseforskning.etikkom.no, Norwegian Ministry of Education and Research), which classified the study as quality assurance (REK Vest #2016/1552). The spirometers used were four hot-wire (Vmax Encore 22D, Carefusion, Vmax Encore 22D, Vmax Encore 22, Vmax Spectra 229, SensorMedics, referred to as HW1-HW4), two ultrasonic (Spirare SPS320 sensors, Diagnostica AS, referred to as US1 and US2), and one wedge-bellows (Vitalograph S, Vitalograph™ Limited, referred to as WB). All spirometers meet the ATS/ERS recommendations. The hot-wire and wedge-bellows spirometers were calibrated and verified daily with 3-L (hot-wire) or 1-L (wedge-bellows) certified calibration syringes. The ultrasonic spirometers were pre-calibrated from the company, but volume-checked in quality control mode. All spirometers were further volume-checked by testing 3-L calibration syringes in patient mode, where the calibration syringe was completely discharged in one second. Spirometry was conducted by four experienced instructors according to ATS/ERS guidelines [8]. The tests were distributed equally among the four instructors, and both instructors and order of spirometers tested were randomized. Differences in repeated measurements with a spirometer were defined as differences between best and second best measurement in one test. Spirometric values were compared using linear mixed models analysis with a random intercept for subjects and a fixed effect for the type of spirometer used. Repeated covariance structure was set to unstructured to allow for heteroscedasticity. 95% confidence intervals and p values were adjusted for multiple comparisons (Bonferroni correction). Bland–Altman plots with 95% limits of agreement were used to demonstrate systematic differences [9]. Statistical analysis was performed using IBM SPSS Statistics, version 24. The level of significance was set at 0.05. Figures were made using Graphpad Prism, version 6.0.
Results
Spirometric values were compared for the different groups of spirometers; hot-wire (HW), ultrasonic (US), and wedge-bellows (WB), and results for FVC and FEV1 are shown by Bland–Altman plots in Fig. 1. Mean ± SD (L) values for HW, US and WB spirometers for FVC were 4.02 ± 0.66, 3.69 ± 0.61 and 3.93 ± 0.69, and for FEV1 3.06 ± 0.44, 2.95 ± 0.44 and 3.10 ± 0.49, respectively. Linear mixed models analysis demonstrated significant differences between HW and US for FVC (p < 0.001) and FEV1 (p < 0.001), between WB and US for FVC (p < 0.001) and FEV1 (p < 0.001), and between HW and WB for FVC (p = 0.046), but not for FEV1 (p = 0.430). Mean differences, mean relative differences and Bland–Altman 95% limits of agreement for FVC and FEV1 are shown in Table 1. No significant differences were found between same type of spirometer (see Table 1). Mean ± SD (L) values for HW, US and WB spirometers for FEV6 were 3.91 ± 0.63, 3.63 ± 0.58 and 3.83 ± 0.65, with significant differences between HW and US (p < 0.001) and between WB and US (p < 0.01), but not between HW and WB (p = 0.104). Mean ± SD (L/min) values for PEF were 478 ± 79, 489 ± 99 and 479 ± 77, respectively (no significant differences). Mean differences with Bonferroni corrected 95% confidence intervals and p values from the linear mixed models analyses are shown for FVC, FEV1 and FEV6 for comparisons of different types of spirometers and spirometers of same type (Fig. 2). For differences between best and second best measurement in one test, mean differences were ranging from 0.03 to 0.06 L (0.63–1.51%) for FVC and from 0.03 to 0.05 L (0.84–1.70%) for FEV1 for the different spirometers (Table 1). Results from testing 3-L syringes in patient mode, showed an average FVC of 3.05 L for Spirare sensors, 3.30 L for the Vitalograph, and 3.37 L for the hot-wire spirometers.
Table 1.
Spirometers | FVC | FEV1 | ||||
---|---|---|---|---|---|---|
Mean diff. (L) | Mean rel. diff. (%) | Bland–Altman 95% lim. agr. | Mean diff. (L) | Mean rel. diff. (%) | Bland–Altman 95% lim. agr. | |
HW–US | 0.33*** | 8.99 | (0.09, 0.58) | 0.11*** | 3.59 | (− 0.03, 0.24) |
WB–US | 0.24*** | 6.56 | (0.02, 0.47) | 0.15*** | 5.13 | (− 0.06, 0.36) |
HW–WB | 0.09* | 2.28 | (− 0.14, 0.31) | − 0.05 n.s. | − 1.46 | (− 0.25, 0.16) |
HW1–US1 | 0.35*** | 9.52 | (0.01, 0.69) | 0.12** | 4.00 | (− 0.05, 0.29) |
HW1–US2 | 0.36*** | 9.64 | (− 0.004, 0.72) | 0.18** | 5.99 | (− 0.04, 0.39) |
HW2–US1 | 0.32*** | 8.53 | (0.03, 0.60) | 0.04 n.s. | 1.29 | (− 0.21, 0.29) |
HW2–US2 | 0.32*** | 8.65 | (0.06, 0.57) | 0.09 n.s. | 3.22 | (− 0.14, 0.33) |
HW3–US1 | 0.40*** | 10.79 | (− 0.02, 0.81) | 0.14* | 4.67 | (− 0.13, 0.41) |
HW3–US2 | 0.40*** | 10.91 | (0.03, 0.77) | 0.20** | 6.67 | (− 0.07, 0.46) |
HW4–US1 | 0.25** | 6.86 | (− 0.06, 0.57) | 0.02 n.s. | 0.53 | (− 0.16, 0.19) |
HW4–US2 | 0.26*** | 6.98 | (− 0.02, 0.54) | 0.07 n.s. | 2.45 | (− 0.15, 0.30) |
WB–US1 | 0.24*** | 6.50 | (− 0.002, 0.48) | 0.12* | 4.14 | (− 0.10, 0.34) |
WB–US2 | 0.24*** | 6.62 | (0.02, 0.47) | 0.18** | 6.13 | (− 0.05, 0.40) |
HW1–WB | 0.11 n.s. | 2.84 | (− 0.19, 0.41) | − 0.004 n.s. | − 0.13 | (− 0.24, 0.23) |
HW2–WB | 0.08 n.s | 1.91 | (− 0.18, 0.33) | − 0.09 n.s. | − 2.74 | (− 0.37, 0.20) |
HW3–WB | 0.16 n.s. | 4.02 | (− 0.26, 0.58) | 0.02 n.s. | 0.51 | (− 0.31, 0.33) |
HW4–WB | 0.01 n.s. | 0.34 | (− 0.21, 0.24) | − 0.11* | − 3.46 | (− 0.30, 0.09) |
HW1–HW2 | 0.04 n.s. | 0.91 | (− 0.31, 0.38) | 0.08 n.s. | 2.68 | (− 0.25, 0.41) |
HW1–HW3 | − 0.05 n.s. | − 1.14 | (− 0.43, 0.34) | − 0.02 n.s. | − 0.64 | (− 0.31, 0.27) |
HW1–HW4 | 0.10 n.s. | 2.49 | (− 0.21, 0.41) | 0.10 n.s | 3.45 | (− 0.14, 0.35) |
HW2–HW3 | − 0.08 n.s. | − 2.04 | (− 0.47, 0.30) | − 0.10 n.s. | − 3.23 | (− 0.36, 0.16) |
HW2–HW4 | 0.06 n.s. | 1.56 | (− 0.17, 0.29) | 0.02 n.s. | 0.75 | (− 0.19, 0.23) |
HW3–HW4 | 0.15 n.s. | 3.67 | (− 0.25, 0.54) | 0.12 n.s | 4.12 | (− 0.14, 0.38) |
US1–US2 | 0.004 n.s. | 0.11 | (− 0.1, 0.13) | 0.06 n.s. | 1.91 | (− 0.09, 0.20) |
Repeatability | FVC | FEV1 | ||||
---|---|---|---|---|---|---|
Mean diff. (L) | Mean rel. diff. (%) | ATS/ERS approved | Mean diff. (L) | Mean rel. diff. (%) | ATS/ERS approved | |
HW1 | 0.05 | 1.31 | Yes | 0.05 | 1.70 | Yes |
HW2 | 0.03 | 0.63 | Yes | 0.03 | 1.03 | Yes |
HW3 | 0.06 | 1.51 | Yes | 0.04 | 1.30 | Yes |
HW4 | 0.04 | 1.09 | Yes | 0.03 | 1.15 | Yes |
US1 | 0.04 | 1.07 | Yes | 0.03 | 1.07 | Yes |
US2 | 0.05 | 1.35 | Yes | 0.05 | 1.62 | Yes |
WB | 0.03 | 0.64 | Yes | 0.03 | 0.84 | Yes |
Repeatability: difference between best and second best test, Mean difference: e.g. HW–US, Mean relative difference: e.g. HW–US/US*100
diff difference, HW hot-wire, lim. agr. limits of agreement, n.s. not significant, rel relative, US ultrasonic, WB wedge-bellows
* p < 0.05, ** p < 0.01,*** p < 0.001 (Bonferroni-corrected from linear mixed models, n = 12)
Discussion
The observed differences in measurements for FVC and FEV1 between the different types of spirometers could be caused by several factors. The SensorMedics instruments are flow-measuring devices with a mass flow sensor based on Kelvin-sensed hot-wire anemometer principles, the Spirare sensors use two-way ultrasound transit time to measure the speed of the airflow, and the Vitalograph measures volume directly by the use of a wedge-bellows [10]. The hot-wire and wedge-bellows spirometers were calibrated and verified daily with different calibration syringes, which may affect the accuracy of the spirometer [11]. The ultrasonic spirometers were pre-calibrated from the company, a procedure previously proved sufficient to retain long-term accuracy in comparable instruments [12].
Spirare sensors apply a fixed BTPS (body temperature and pressure, saturated) correction factor of 1.02 to exhaled air, while the hot-wire instruments apply real-time BTPS correction in which exhalation temperature is measured continuously. The Vitalograph has a fixed BTPS correction factor of 1.09 at 22 °C (used in this study), with additional manual correction for ambient conditions outside of normal range (similar to real-time correction). Results from testing 3-L syringes in patient mode for the different spirometers were within expected values based on the spirometers different BTPS correction systems. When ultrasonic sensors are used for testing patients, exhaled air is measured close to the mouth and there are no physical obstacles in the air channel. Thus, one could argue that variable cooling of air does not affect the results considerably. However, the real-time BTPS correction applied by the hot-wire spirometers also takes into account variations in ambient temperature, pressure and humidity, possibly leading to increased accuracy.
There were also differences in the technique for performing spirometry. For the hot-wire spirometers, 3–4 times of tidal volume measurements were performed before maximal inspiration and forced expiration. For ultrasonic and wedge-bellows spirometers, maximal inspiration was performed before the patient connected to the mouthpiece. There was a tendency to increased difference with increased lung volumes for FVC (see Bland–Altman plots) and FEV6 for hot-wire compared and wedge-bellows compared to ultrasonic, but not for hot-wire compared to wedge-bellows spirometers. Furthermore, there was only a small difference in FVC for hot-wire compared to wedge-bellows spirometers, and no difference in FEV6. These small differences could be caused by dissimilarities in the technique for performing spirometry, while the larger proportional biases for hot-wire and wedge-bellows compared to ultrasonic spirometers are more likely to be caused by differences in measurements principles between the spirometers.
Patient and instructor variability may also interfere with the measurements [3, 4]. Nevertheless, the systematically higher values for FVC and FEV1 for hot-wire and wedge-bellows compared to ultrasonic spirometers suggest that there are actual differences in measurements between the different types of instruments. Similar, but smaller differences have been observed in previous studies of other hot-wire and ultrasonic spirometers [5, 6]. A strength of our pilot study is the inclusion of the Vitalograph, which is considered a gold standard in spirometry testing as it measures volume directly [13]. The Vitalograph has also shown agreement with other types of spirometers, like the pneumotachograph [14].
In conclusion, the pilot study shows systematically higher values for FVC and FEV1 for hot-wire and wedge-bellows compared to ultrasonic spirometers. Technicians and physicians involved in lung function testing and interpretation should be aware of the possible inter-variability between spirometers. The findings should be investigated in larger data sets including patients, instructors and spirometers within and across laboratories. The impact of spirometer inter-variability on conclusions regarding the patient’s diagnosis and treatment should be explored. Furthermore, the results warrant discussion on standardization of BTPS correction in order to improve agreement between spirometers.
Limitations
The pilot study demonstrates differences in measurements between spirometers for a small data set of 12 healthy individuals. In order to draw conclusions, larger cohorts including different groups of patients with a broad range of spirometric values should be investigated. Spirometry is a physiological test, where the testing procedure and results are influenced by a number of different factors. We have not tested the performance of the spirometers with a waveform generator, as this equipment is not available at our laboratory. However, all spirometers were tested by the manufacturer and shown to meet the ATS/ERS recommendations. Moreover, we could have tested each subject at several different time points for each spirometer to assess the variability within each subject. The instructor conducting the tests adds another well-known source of variability. In larger comparison studies, it is not possible for one instructor to perform all tests. A randomized design for instructors in addition to order of spirometers tested is therefore crucial. Differences in techniques for performing spirometry for the different spirometers may also influence the results, and make it difficult to distinguish between differences in measurements caused by the technique compared the detection itself. Finally, in our mixed models analysis, we have not included adjustments for sex, age and height, as this is a small quality assurance study where these data were not available. These characteristics, together with size of lung volume, are also shown to affect the magnitude of the bias in spirometer comparison studies [5], and should be included in future studies.
Authors’ contributions
MA, LS and IH designed the study. LS was responsible for the data collection. MA, IH and TE did the statistical analysis. MA, LS, SL, TE and IH interpreted data. MA, LS, SL, TE and IH participated in writing the manuscript. All authors read and approved the final manuscript.
Acknowledgements
We would like to thank the volunteers and instructors participating in this pilot study.
Competing interests
The authors declare that they have no competing interests.
Availability of data and materials
The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.
Consent for publication
Not applicable.
Ethics approval and consent to participate
The study was remit assessed by the Regional Ethics Committee (REK Vest; http://helseforskning.etikkom.no, Norwegian Ministry of Education and Research), which classified the study as quality assurance (REK Vest #2016/1552), without the need for written informed consent.
Funding
There was no funding for the study.
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abbreviations
- ATPS
ambient temperature and pressure, saturated
- ATS
American Thoracic Society
- BTPS
body temperature and pressure, saturated
- ERS
European Respiratory Society
- FEV1
forced expiratory volume in one second
- FEV6
forced expiratory volume in 6 second
- FVC
forced vital capacity
- HW
hot-wire
- PEF
peak expiratory flow
- US
ultrasonic
- WB
wedge-bellows
Contributor Information
Marit E. Aardal, Email: marit@heisenberg.no
Lene L. Svendsen, Email: lene.lovendahl.svendsen@helse-bergen.no
Sverre Lehmann, Email: sverre.lehmann@helse-bergen.no.
Tomas M. Eagan, Email: tomas.eagan@uib.no
Ingvild Haaland, Phone: +47 99462558, Email: Ingvild.Haaland@uib.no.
References
- 1.Lange NE, Mulholland M, Kreider ME. Spirometry: don’t blow it! Chest. 2009;136:608–614. doi: 10.1378/chest.08-2315. [DOI] [PubMed] [Google Scholar]
- 2.Miller MR, Crapo R, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A, Enright P, van der Grinten CP, Gustafsson P, et al. General considerations for lung function testing. Eur Respir J. 2005;26:153–161. doi: 10.1183/09031936.05.00034505. [DOI] [PubMed] [Google Scholar]
- 3.Dillard TA, Reyes J, Johnson M, Baker R. Patient-based studies of agreement between spirometers. Respir Care. 2003;48:587–588. [PubMed] [Google Scholar]
- 4.Akhtar R, Wilson A. A comparison of spirometry in general practice and a pulmonary function laboratory. Prim Care Respir J. 2005;14:215–220. doi: 10.1016/j.pcrj.2004.12.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Bridevaux PO, Dupuis-Lozeron E, Schindler C, Keidel D, Gerbase MW, Probst-Hensch NM, Bettschart R, Burdet L, Pons M, Rothe T, et al. Spirometer replacement and serial lung function measurements in population studies: results from the SAPALDIA Study. Am J Epidemiol. 2015;181:752–761. doi: 10.1093/aje/kwu352. [DOI] [PubMed] [Google Scholar]
- 6.Gerbase MW, Dupuis-Lozeron E, Schindler C, Keidel D, Bridevaux PO, Kriemler S, Probst-Hensch NM, Rochat T, Kunzli N. Agreement between spirometers: a challenge in the follow-up of patients and populations? Respiration. 2013;85:505–514. doi: 10.1159/000346649. [DOI] [PubMed] [Google Scholar]
- 7.Liistro G, Vanwelde C, Vincken W, Vandevoorde J, Verleden G, Buffels J. Technical and functional assessment of 10 office spirometers: a multicenter comparative study. Chest. 2006;130:657–665. doi: 10.1378/chest.130.3.657. [DOI] [PubMed] [Google Scholar]
- 8.Miller MR, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A, Crapo R, Enright P, van der Grinten CP, Gustafsson P, et al. Standardisation of spirometry. Eur Respir J. 2005;26:319–338. doi: 10.1183/09031936.05.00034805. [DOI] [PubMed] [Google Scholar]
- 9.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–310. doi: 10.1016/S0140-6736(86)90837-8. [DOI] [PubMed] [Google Scholar]
- 10.de Jongh F. Spirometers. Breathe. 2008;4:251–254. [Google Scholar]
- 11.Madsen F. Validation of spirometer calibration syringes. Scand J Clin Lab Invest. 2012;72:608–613. doi: 10.3109/00365513.2012.723739. [DOI] [PubMed] [Google Scholar]
- 12.Skloot GS, Edwards NT, Enright PL. Four-year calibration stability of the EasyOne portable spirometer. Respir Care. 2010;55:873–877. [PubMed] [Google Scholar]
- 13.Moore VC. Spirometry: step by step. Breathe. 2012;8:232–240. doi: 10.1183/20734735.0021711. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Marshall M, Jackson J, Cooper BG. Does it matter which type of spirometer is used to measure FEV1, FVC and VC? Thorax. 1999;54(Suppl 3):A78. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.