A Simplified Trans-Septal Puncture Technique using a Needle Free Approach for Cryoablation of Atrial Fibrillation

Adam J Graham 1,*, Shohreh Honarbakhsh 2,*, Ben O’Brien 3,*, Richard J Schilling 4,*
PMCID: PMC5673294  PMID: 29250235

Abstract

Background:

Trans-septal puncture (TSP) is routinely performed during treatment of atrial fibrillation (AF) and other electrophysiology procedures. The purpose of this retrospective observational study was the analysis of a novel needle free technique employed to gain access to the left atrium (LA).

Methods:

The SafeSept Needle Free guidewire was delivered, using no needle, to the interatrial septum via a standard cryoablation sheath and dilator under transoesophageal guidance. The wire was then advanced into the LA with position confirmed by fluoroscopy. All cases were performed under general anaeasthetic (GA).

Results:

The novel procedure was performed in 43 patients (38 paroxysmal AF (88.4%) and 5 persistent AF (11.4%)) with 100% success rate in accessing the LA. The procedure times (51.70 ±18.18) and fluoroscopy times (2.75 ± 3.35) were recorded. There were no complications seen in the cohort.

Conclusion:

Using SafeSept Needle Free wire for TSP had a high success rate and resulted in no complications, with the potential to reduce procedure and fluoroscopy times.

Keywords: Atrial Fibrillation, Cryoablation, Trans-septal puncture, Needle-Free, Safe-Sept

Introduction

The origins of trans-septal puncture (TSP) lie in its use for left heart catheterization. However, studies for the evaluation of valve disease or percutaneous mitral balloon valvuloplasty indicate that it is now more commonly performed to access into the left atrium (LA) for catheter ablation of arrhythmia[1][2]. The original technique first described by Ross et al. in 1960[3] has evolved over the subsequent years. This means that while the use of ultrasound imaging has increased, the techniques and technologies used are not specifically designed for the purpose they are being put to today[4]. An increase in atrial fibrillation ablation procedures facilitated the development of specific techniques and the adoption of new technologies to assist TSP such as the RF needle[5]. However, it is well known that TSP is related to some life-threatening complications. This is especially true in the presence of challenging interatrial septal anatomy[2]. We sought to investigate whether a novel TSP puncture wire, the SafeSept Needle Free guidewire (SafeSeptTM Pressure Products, San Pedro, CA, USA), could be used through a cryoablation sheath to reduce procedural steps such as sheath and wire exchanges.

Patients and methods

In this retrospective observational study, a total of 43 consecutive patients undergoing pulmonary vein isolation using cryoballoon were analysed. TheTSP procedures were performed under general anesthetic using the SafeSept Needle Free guide wire (SafeSeptTM Pressure Products, San Pedro, CA, USA) with Transoesophageal (TOE) guidance without interrupting oral anticoagulation therapies.

Seldinger technique was used to gain femoral venous access and introduce a cryoablation sheath (Flexcath AdvanceTM Steerable Sheath, Medtronic, Minneapolis, USA) with the tip placed in the SVC under TOE guidance (no fluoroscopy was used for this part of the procedure). No problems were encountered advancing the sheath with the support of the guidewire. A second 7F sheath was used to introduce a pacing wire into the SVC to pace the phrenic nerve during right pulmonary vein freezes. The guide wire was withdrawn from the cryoablation sheath and a SafeSept wire advanced to the mid portion of the sheath and dilator. The SafeSept wire is a 150cm, 0.0315” nitinol guidewire with a sharp tip and a radio-opaque distal coil. When unsupported by sheath and dilator it has a pre-formed J-shaped tip, a feature designed to render it atraumatic when inside thevasculature [Figure 1]. A bend of 300 was placed on the cryoablation sheath and dilator and they were then slowly withdrawninferiorly under TOE guidance. When the tip of the dilator was seen to drop into the fossa ovalis the SafeSept wire was advanced and used to puncture the septum [Figure 2]. Heparin was given as per local protocol at this juncture. The position of the wire was confirmed on fluoroscopy as being in the left upper pulmonary vein. The sheath and dilator were then advanced through the septum under fluoroscopic guidance and the needle and dilator removed. The cryoablation procedure proceeded as described before[6].

Figure 1. Schematic view of the use of the Needle Free SafeSept wire to cross the atrial septum.

Figure 1.

Figure 2. Transoesophageal echocardiography views demonstratetenting of the inter-atrial septum (left) and subsequent advancement into the left atrium (right).

Figure 2.

Procedure time was defined as time from initial venous puncture to removal of sheaths and measured in minutes. Duration of fluoroscopy was also measured in minutes. Patients were monitored overnight and followed up in clinic 3 months post procedure. They were assessed for standard major and minor complications of AF ablation during these times.

Normal left ventricular function was defined as an ejection fraction above 55%; with normal valvular function defined as no pathological stenosis or regurgitation. Out of those with abnormal LV function all were mildly impaired.

Results

From July 2015 to April 2016 the SafeSept Needle Free guidewire was used to perform TSP in 43 consecutive patients. Baseline characteristics are demonstrated in [Table 1].

Table 1. Demonstrates the baseline characteristics.

^Ischaemic heart disease

SafeSept Needle Free
Age y, mean ± SD 59.5 ±10.8
Sex M (%) 34 (79)
Hypertension, n (%) 13 (30)
IHD^, n (%) 2 (4.7)
Diabetes, n (%) 2 (4.7)
Echocardiography
Normal LV function n (%) 40 (93)
Normal valvular function n (%) 41 (95.3)
^Ischaemic heart disease

The majority of patients in the cohort were undergoing ablation for paroxysmal atrial fibrillation (PAF); 38 PAF, 88.4% and 5 early persistent AF, 11.6% [Table 2]. None had had a prior TSP performed. Most patients had preserved left ventricular systolic function (40/43, 93%; [Table 2]).

Table 2. Demonstrates the indication for catheter ablation, extra procedures performed and the echocardiographic features.

ˆCavotricuspid isthmus *Electrophysiology study

Indications SafeSept Needle Free
Persistent AF, n (%) 5 (11.6)
Paroxysmal AF n (%) 38 (81.4)
Extra Procedures
CTI^ Line n (%) 11 (25.6)
EPS* n (%) 2 (4.7)
^Cavotricuspid isthmus
*Electrophysiology study

In all patients, the TSP was successful with no complications. Procedure data is shown in [Table 3].

Table 3. Demonstrates the procedural data.

SafeSept Needle Free
Successful TSP n (%) 43 (100)
Procedure time min ± SD 51.70 ±18.18
Fluoroscopy time min ± SD 2.75 ± 3.35
Complications n (%) 0

Discussion

Given the expansion in electrophysiology procedures a variety of techniques designed to improve efficacy and safety of the TSP have been employed[7]. Radio frequency delivery to the inter-atrial septum has been applied with success in cases in which standard techniques have failed [8][5]. A recent study has demonstrated a benefit to deep inspiratory maneuvers[2].

Our study demonstrates for the first time that a specifically designed trans-septal wire can be used without a trans-septal needle to successfully perform TSP; with LA access achieved in all patients. The use of a SafeSept guide wire has been previously described [9],[10], with the larger of these cohorts consisting of 210 patients. These groups’ technique differs from that employed here, using a different model of the guide wire that requires introduction via the lumen of a standard Brockenbrough needle. Successful performance of the TSP occurred in 97.6% the larger cohort and in all 19 patients in the smaller study.

In this cohort of patients, we experienced no complications and this indicates the technique to be safe in this group. Lower fluoroscopy times when compared to those reported in other studies[6], could be due to no fluoroscopy being used until the LA is accessed. This improvement also applied to total procedure time[6], with the simpler work-flow, avoiding the need to exchange sheaths and wires making the procedure more efficient. It may also improve safety, as sheath exchange is one point in the procedure where there is the potential to produce air embolism or lose position of the wires/sheaths.

Limitations

This study is limited due to it being a retrospective analysis. A larger randomized controlled study would be necessary to conclusively evaluate its validity. Furthermore, we have measured fluoroscopy time and procedure time, which are influenced by the length of the procedure post TSP. Any further evaluation of this technique should focus on fluoroscopy time and procedure duration from RFV sheath insertion to TSP.

Conclusions

Our technique, whilst requiring further evaluation, offers promise of streamlining the trans-septal puncture and potentially reducing procedure and fluoroscopy times with no obvious compromise to safety.

Conflict of Interests

None.

Disclosures

None.

References

  • 1.Korkmaz Sule, Demirkan Burcu, Güray Yeşim, Yılmaz Mehmet Birhan, Aksu Tolga, Saşmaz Hatice. Acute and long-term follow-up results of percutaneous mitral balloon valvuloplasty: a single-center study. Anadolu Kardiyol Derg. 2011 Sep;11 (6):515–20. doi: 10.5152/akd.2011.136. [DOI] [PubMed] [Google Scholar]
  • 2.Aksu Tolga, Guler Tumer Erdem, Yalin Kivanc, Golcuk Sukriye Ebru, Ozcan Kazim Serhan, Guler Niyazi. A Novel Deep Inspiration Maneuver for Difficult Transseptal Puncture. Am. J. Cardiol. 2017 Feb 01;119 (3):428–433. doi: 10.1016/j.amjcard.2016.10.023. [DOI] [PubMed] [Google Scholar]
  • 3.ROSS J. Transeptal left heart catheterization: a new method of left atrial puncture. Ann. Surg. 1959 Mar;149 (3):395–401. doi: 10.1097/00000658-195903000-00009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Erden İsmail, Erden Emine Çakcak, Golcuk Ebru, Aksu Tolga, Yalin Kıvanç, Güler Tümer Erdem, Özcan Kazım Serhan, Turan Burak. Impact of transesophageal echocardiography during transseptal puncture on atrial fibrillation ablation. J Arrhythm. 2016 Jun;32 (3):170–5. doi: 10.1016/j.joa.2015.12.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Sherman Warren, Lee Paul, Hartley Amanda, Love Barry. Transatrial septal catheterization using a new radiofrequency probe. Catheter Cardiovasc Interv. 2005 Sep;66 (1):14–7. doi: 10.1002/ccd.20362. [DOI] [PubMed] [Google Scholar]
  • 6.Kuck Karl-Heinz, Brugada Josep, Fürnkranz Alexander, Metzner Andreas, Ouyang Feifan, Chun K R Julian, Elvan Arif, Arentz Thomas, Bestehorn Kurt, Pocock Stuart J, Albenque Jean-Paul, Tondo Claudio. Cryoballoon or Radiofrequency Ablation for Paroxysmal Atrial Fibrillation. N. Engl. J. Med. 2016 Jun 09;374 (23):2235–45. doi: 10.1056/NEJMoa1602014. [DOI] [PubMed] [Google Scholar]
  • 7.MJ Earley. How to perform a transseptal puncture. Heart [Internet]. 2008;95:85–92. doi: 10.1136/hrt.2007.135939. [DOI] [PubMed] [Google Scholar]
  • 8.Knecht Sébastien, Jaïs Pierre, Nault Isabelle, Wright Matthew, Matsuo Seiichiro, Madaffari Antonio, Lellouche Nicolas, O'Neill Mark D, Derval Nicolas, Deplagne Antoine, Bordachar Pierre, Sacher Frederic, Hocini Mélèze, Clémenty Jacques, Haïssaguerre Michel. Radiofrequency puncture of the fossa ovalis for resistant transseptal access. Circ Arrhythm Electrophysiol. 2008 Aug;1 (3):169–74. doi: 10.1161/CIRCEP.108.788000. [DOI] [PubMed] [Google Scholar]
  • 9.Wadehra Vineet, Buxton Alfred E, Antoniadis Antonios P, McCready James W, Redpath Calum J, Segal Oliver R, Rowland Edward, Lowe Martin D, Lambiase Pier D, Chow Anthony W C. The use of a novel nitinol guidewire to facilitate transseptal puncture and left atrial catheterization for catheter ablation procedures. Europace. 2011 Oct;13 (10):1401–5. doi: 10.1093/europace/eur155. [DOI] [PubMed] [Google Scholar]
  • 10.De Ponti Roberto, Marazzi Raffaella, Picciolo Giuseppe, Salerno-Uriarte Jorge A. Use of a novel sharp-tip, J-shaped guidewire to facilitate transseptal catheterization. Europace. 2010 May;12 (5):668–73. doi: 10.1093/europace/euq060. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Atrial Fibrillation are provided here courtesy of CardioFront, LLC

RESOURCES