Abstract
Purpose
The purpose of this study was to characterize the Mobius AIRO Mobile CT System for localization and image‐guided proton therapy. This is the first known application of the AIRO for proton therapy.
Methods
Five CT images of a Catphan®504 phantom were acquired on the AIRO Mobile CT System, Varian EDGE radiosurgery system cone beam CT (CBCT), Philips Brilliance Big Bore 16 slice CT simulator, and Siemens SOMATOM Definition AS 20 slice CT simulator. DoseLAB software v.6.6 was utilized for image quality analysis. Modulation transfer function, scaling discrepancy, geometric distortion, spatial resolution, overall uniformity, minimum uniformity, contrast, high CNR, and maximum HU deviation were acquired. Low CNR was acquired manually using the CTP515 module. Localization accuracy and CT Dose Index were measured and compared to reported values on each imaging device. For treatment delivery systems (Edge and Mevion), the localization accuracy of the 3D imaging systems were compared to 2D imaging systems on each system.
Results
The AIRO spatial resolution was 0.21 lp mm−1 compared with 0.40 lp mm−1 for the Philips CT Simulator, 0.37 lp mm−1 for the Edge CBCT, and 0.35 lp mm−1 for the Siemens CT Simulator. AIRO/Siemens and AIRO/Philips differences exceeded 100% for scaling discrepancy (191.2% and 145.8%). The AIRO exhibited higher dose (>27 mGy) than the Philips CT Simulator. Localization accuracy (based on the MIMI phantom) was 0.6° and 0.5 mm. Localization accuracy (based on Stereophan) demonstrated maximum AIRO‐kV/kV shift differences of 0.1 mm in the x‐direction, 0.1 mm in the y‐direction, and 0.2 mm in the z‐direction.
Conclusions
The localization accuracy of AIRO was determined to be within 0.6° and 0.5 mm despite its slightly lower image quality overall compared to other CT imaging systems at our institution. Based on our study, the Mobile AIRO CT system can be utilized accurately and reliably for image‐guided proton therapy.
Keywords: AIRO, IGPT, mobile CT, proton therapy
1. Introduction
Advances in in‐room computed tomography (CT) scanners and cone‐beam technology have led to the proliferation of CT localization for image guided radiation therapy (IGRT).1, 2, 3, 4, 5, 6, 7 These imagers provide improved treatment accuracy over conventional orthogonal imaging allowing for increased precision in radiation delivery. Patient localization accuracy is particularly important in proton beam radiation therapy due to the sharp dose fall‐off compared to conventional x‐ray therapy. Unfortunately, due to the size and geometry of proton therapy units, imaging has largely been limited to orthogonal kV/kV x‐ray systems.8 Recently, people have reported on the use of in‐room CT scanners for proton beam radiation therapy localization.9, 10, 11 Some proton vendors are developing technologies for CBCT (Proteus®ONE, IBA, Belgium, HITACHI, Tokyo, Japan). To date, stand‐alone in‐room CT scanners utilized have had large footprints and are either not amenable to or cumbersome to use in the more compact proton therapy centers, such as the S250 (Mevion Medical Systems, Littleton, MA, USA).12 For these reasons, a small‐footprint mobile CT scanner, commonly used for image‐guided surgery, could be of great value for 3D image‐guided proton therapy (IGPT).
The AIRO Mobile CT System (Mobius Imaging LLC, Shirley, MA, USA) is a large bore (107 cm) helical 32 slice CT scanner historically utilized for intra‐operative imaging for spinal surgeries. Our institution is the first to acquire and clinically implement the AIRO Mobile CT System (AIRO) for IGPT. The AIRO's small footprint (W × L × H: 1.94 × 1.54 × 1.90 m) occupies 1.28 m2 of treatment floor space (in scan mode). When not in use, the AIRO is stored in the maze (Fig. 1) to avoid radiation‐induced damage to its sensitive electronics. The AIRO's motor‐controlled castors provide effortless transport from the storage location (in the maze hallway) to the scanning location.
The AIRO's image quality characteristics have been reported by Weir et al. for surgical applications.13 Our study assessed the performance characteristics of the AIRO at our institution compared with IGRT systems used for intensity‐modulated radiation therapy (IMRT) and CT simulators at our institution for radiotherapy applications. Our current IGPT workflow involves CT simulation using a Philips Big Bore 16 Slice CT Simulator with routine orthogonal kV/kV image pairs (flat‐panel detectors) preceding each treatment fraction. The AIRO will be utilized for target localization and inter‐fraction adaptive treatment assessment. To our knowledge, our facility is the first compact proton therapy system to use a mobile CT scanner for IGPT.
2. Materials and methods
Mevion Medical Systems has developed software, Verity™ 3D, that uses CT images from any scanner to perform 3D image registrations for proton therapy. Due to the ease of motion and small footprint, we have selected the AIRO CT scanner for IGPT. The AIRO is sold commercially for surgical purposes by BrainLab (Munich, Germany). Mevion has developed an infrared camera tracking system to interface with the CT scanner to facilitate use of CT images for IGPT. This system references the 3D image set to the in‐room coordinates and planning CT via a reference frame, infrared camera, and CT scan. This reference frame attaches to the treatment couch and includes infrared markers to determine initial room coordinates as well as ceramic markers to reference the CT image to room coordinates. An additional infrared marker is rigidly attached to the gantry so that changes in the reference frame location in relation to the treatment room can be detected as the robotic couch moves between the initial treatment setup location and the CT imaging location. The AIRO unit was specifically adapted to use inside the proton vault by removing the integrated stand that supports the weight of surgical gurneys in order to avoid interference with the robotic couch (Fig. 1).
2.A. Image quality characterization and comparison to simulators and IGRT systems
Five CT images of a Catphan®504 phantom (The Phantom Laboratory, Salem, NY, USA) were acquired on the AIRO Mobile CT System, a Varian EDGE™ Radiosurgery System (EDGE) On Board Imager, a Philips Brilliance Big Bore 16 slice CT simulator, and a Siemens SOMATOM Definition AS 20 slice CT simulator. Medial and lateral lasers were used to align the phantom prior to image acquisition.
Image acquisition was performed with various protocols. On the AIRO, images were acquired with soft, standard, and sharp reconstruction kernels with the pre‐set clinical head protocol. The AIRO's pre‐set clinical protocols for head, thorax, abdomen, pelvis, shoulder are listed in Table 1. The sharp, standard, and soft kernels are noise filtering Gaussian smoothing kernels with σ = 0.00, 0.68, and 0.92, respectively.
Table 1.
Edge CBCT techniques | AIRO helical CT protocols | ||||
---|---|---|---|---|---|
Techniques | kVp | mAs | Protocols | kVp | mAs |
Head | 100 | 150 | Head | 120 | 326 |
Thorax | 125 | 268 | Thorax | 120 | 68 |
Pelvis | 125 | 1080 | Pelvis | 120 | 149 |
Pelvis Obese | 140 | 1681 | Abdomen | 120 | 149 |
Image Gently | 80 | 97 | Shoulder | 120 | 149 |
On the EDGE, images were acquired using the following pre‐set clinical techniques: head, thorax, pelvis, pelvis obese, and image gently (Table 1). On the Philips CT simulator, standard clinical protocols were used (120 kVp, 450 mA) with 1.5 mm, 2.0 mm, and 3.0 mm slice thickness. On the Siemens CT simulator images were acquired with standard brain (120 kV, 450 mAs), high res brain (120 kV, 580 mAs), abdomen (120 kV, 250 mAs), and thorax (120 kV, 140 mAs) protocols with 1.5 mm, 2.0 mm, and 3.0 mm slice thickness.
Multimodality comparisons were made using the clinical head protocols on the AIRO, Philips, and Siemens CT simulators and the pelvis technique on the Edge. The pelvis technique was used on the Edge because it uses full 360° gantry rotation and provides the highest image quality. The Edge head protocol involves a 200° gantry rotation and is configured for low dose rather than image quality.
DoseLAB software v.6.6 (Mobius Medical Systems LP, Houston, TX, USA) was used for image quality analysis. Modules CTP404, CTP528, and CTP486 were analyzed. The following image quality measures were assessed: modulation transfer function (MTF), scaling discrepancy, geometric distortion, spatial resolution, overall uniformity, minimum uniformity, contrast, CNR, and maximum HU deviation (Table 2).14 The mean and standard deviation of five or more scans are reported. The low contrast module (CTP515) was assessed on the AIRO, Philips, and Siemens scans. Low CNR was calculated using the 15.0 mm 1% Supra‐slice target. A region‐of‐interest (ROI) was dropped onto the Supra‐slice target and another on a uniform region of the phantom.
Table 2.
Image quality tests | Definition |
---|---|
Modulation Transfer Function (MTF)a | The modulation of multiple ROIs in the phantom with various line bar patterns. As spatial frequency increases, ROI's modulation decreases. The modulation is normalized to its highest value and plotted. |
Modulationa |
Where s the 90th percentile CT number in the ROI and s the 10th percentile CT number in the ROI. |
Scaling Discrepancya |
The maximum possible error in a measurement of distance. Where is the magnification of the image and is the length of the image's longest side. |
Geometric Distortiona | Verifies that correct distance measurement occurs in all regions of the CT image. |
Overall Uniformitya |
Where is the maximum 90th percentile CT number in all uniformity regions and is the minimum 10th percentile CT number in all uniformity regions |
Minimum Uniformitya |
The lowest uniformity of all ROIs. Uniformity is calculated as:
Where is the 90th percentile of the pixel CT number in the ROI and is the 10th percentile of the CT number in the ROI. |
Maximum HU Deviationa | The maximum deviation between mean HU value of an ROI subtracted from its defined reference HU absolute value of all calculated HU deviation on an image. |
Contrasta |
Where is the mean CT number in the region with greater signal and is the mean CT number in the region with lesser signal. |
High Contrast‐to‐Noise Ratioa |
Where is the standard deviation of the high contrast ROI, is the standard deviation of background, is the mean CT number of the high contrast ROI and is the mean CT number of background. |
Low Contrast‐to‐Noise Ratio13 |
Where is the standard deviation of the background ROI, is the mean CT number inside the low contrast ROI and is the mean CT number of the background ROI. |
Definitions derived from DoseLab v.6.6 User Manual.17
2.B. Localization accuracy
A Stereophan phantom (Sun Nuclear Corporation, Melbourne, FL, USA) was used to measure the localization accuracy of the Mevion‐AIRO system. The universal spacer insert and CT/MRI insert were inserted into the cylinder cavity. The phantom was leveled with the precision leveling stand. Treatment isocenter was placed at the BB target corresponding to the laser alignment marks on the outside of the phantom. The robotic couch was re‐positioned to acquire the AIRO images following the Mevion localization workflow. Once the images were acquired, the robotic couch was moved back to the initial treatment isocenter based on the robotic couch's coordinates. The software then performed a CT registration correlating the reference frame's ceramic fiducials to its infrared markers (relative to treatment isocenter). The images were then manually registered to the 3D planning CT. Successively, a kV/kV image pair was acquired and registered to the digitally reconstructed radiograph (DRR). Localization accuracy was defined as the difference between suggested shifts from the kV/kV pair versus the AIRO CT image set, where the kV/kV image was considered the gold standard. In addition, end‐to‐end tests were performed to illustrate both time and workflow for patient setup and target localization.
Additional localization accuracy tests were performed using the 6 degree‐of‐freedom MIMI Phantom (Standard Imaging, Inc. Middleton, WI, USA). The default Q‐fix overlay couch top was used. With the proton gantry positioned at 55° the reference frame was attached to the couch top and the couch was positioned to 270° to capture the reference frame. Known shifts were applied to the phantom separately including all translational directions (lateral, longitudinal, and vertical), all rotations (yaw, shift, and roll), and cumulatively (all 6 degrees of freedom). A reference kV/kV image set was acquired preceding the AIRO CT image. The couch was moved to the 3D imaging position and a 3D CT image was acquired. The image was sent to the Mevion treatment console and Verity, and then registered to the 3D planning CT. An initial manual rigid registration was performed followed by automated registration for fine tuning. Suggested shifts were applied. The robotic couch was then re‐positioned to standard imaging position and a 2D orthogonal kV x‐ray pair was acquired. Suggested shifts (kV/kV) were recorded. Localization accuracy was defined as the difference between suggested shifts and known shifts.
2.C. Measured imaging dose
CT Dose Index for helical scans (CTDIvol) was calculated using a 16‐cm and 32‐cm cylindrical polymethylmetacrylate phantom (CIRS, Computerized Imaging Reference Systems, Inc., Norfolk, VA, Model 007 & 007A) with 100 mm pencil ionization chamber and electrometer (Raysafe Xi CT Detector, Unfors RaySafe, Inc., Cleveland, OH, USA) on the AIRO and the Philips CT Simulator. The ion chamber was placed at the 3 o'clock and the 12 o'clock positions. Three dose measurements were acquired at each position. Measured dose was then compared to the vendor‐reported CTDIvol. ACR requirements specify that dose must be within 20% of manufacturer‐reported specifications.15
3. Results
3.A. Image quality characterization and comparison to simulators and IGRT systems
3.A.1. AIRO reconstruction kernel effects on image quality
The AIRO image quality results for various kernels are listed in Table 3. The AIRO sharp kernel provided the best MTF (0.26 line pairs/mm). High CNR and spatial resolution (31.1% difference) decreased from sharp to soft kernel. Low CNR increased from sharp kernel to soft kernel (5.73–8.51). Overall uniformity and minimum uniformity were comparable across reconstruction kernels (<0.5% difference). Geometric distortion was equal for sharp and standard reconstruction kernels and <7% difference between sharp/soft and standard/soft kernels. Figure 2 illustrates that the MTF for the sharp kernel with small FOV exceeded that of the soft kernel.
Table 3.
Sharp | Standard | Soft | |
---|---|---|---|
Scaling discrepancy (mm): | 1.19 | 1.34 | 1.35 |
Geometric distortion (mm): | 0.14 | 0.14 | 0.15 |
Spatial resolution (50% MTF, lp/mm): | 0.26 | 0.21 | 0.19 |
Overall uniformity (%): | 98.85 | 98.94 | 98.99 |
Minimum uniformity (%): | 99.06 | 99.19 | 99.31 |
Contrast (%): | 10.68 | 10.56 | 10.41 |
High CNR: | 30.34 | 21.45 | 17.49 |
Low CNR: | 5.73 | 7.12 | 8.51 |
Maximum HU deviation (HU): | 58.98 | 57.50 | 57.68 |
3.A.1. Comparison to simulators and IGRT systems
Figure 3 illustrates that the MTF and spatial resolution of the AIRO (0.21 lp/mm) was lower than the Philips CT simulator (0.40 lp/mm), Siemens CT simulator (0.35 lp/mm), and Edge CBCT (0.37 lp/mm). Conversely, the Philips CT simulator exhibited superior MTF and spatial resolution compared to all modalities (Table 4 and Fig. 3). The AIRO had comparable minimum and overall uniformity when compared to the Philips CT simulator, Siemens CT simulator or the Edge (<0.5% difference, Table 4). The Philips high CNR was superior to the AIRO with a 3.65 difference between the two values (18.9% difference). The Siemens high CNR was superior to the AIRO with a 7.02 difference between the two values (28.1% difference). Philips maximum HU deviation was comparable to the AIRO (17.0% difference). Siemens maximum HU deviation was larger than the AIRO's maximum HU deviation (43.5% difference). The Edge demonstrated the smallest HU deviation (21.66 HU, 90.6% difference from AIRO).
Table 4.
Philips | Airo | Siemens | Edgea | |
---|---|---|---|---|
Scaling discrepancy (mm): | 0.21 | 1.34 | 0.03 | 0.86 |
Geometric distortion (mm): | 0.11 | 0.14 | 0.11 | 0.13 |
Spatial resolution (50% MTF, lp/mm): | 0.40 | 0.21 | 0.35 | 0.37 |
Overall uniformity (%): | 99.21 | 98.94 | 99.37 | 98.93 |
Minimum uniformity (%): | 99.25 | 99.19 | 99.37 | 99.26 |
Contrast (%): | 10.30 | 10.56 | 12.03 | 10.79 |
High CNR: | 21.14 | 21.45 | 28.47 | 23.36 |
Low CNR: | 2.00 | 7.12 | 2.46 | – |
Maximum HU deviation (HU): | 48.50 | 57.5 | 89.49 | 21.66 |
EDGE CBCT images were acquired with the Pelvis protocol.
AIRO/Philips image quality differences were within 3% for overall uniformity, minimum uniformity, and contrast and high CNR. AIRO/Philips geometric distortion difference was 0.03 mm (24%). AIRO/Siemens high CNR differences were 28.1% and geometric distortion differences were 24%. AIRO/Siemens differences exceeded 100% for scaling discrepancy (191.2%). The AIRO low CNR was 7.12 for the standard reconstruction kernel. The Philips and Siemens CT Simulator low CNR were 2.00 and 2.46, respectively. Refer to Table 5 for HU constancy values for the AIRO, Philips CT Simulator, Siemens CT Simulator and Edge CBCT (Module CTP404).
Table 5.
Philips (HU) | Edge (HU) | Airo (HU) | Siemens (HU) | True (HU) | |
---|---|---|---|---|---|
Acrylic | 132.44 | 123.08 | 108.15 | 129.67 | 100 |
Air | −969.86 | −999.41 | −978.49 | −1023.34 | −1000 |
Polystyrene | −22.69 | −36.62 | −47.83 | −52.26 | −35 |
LDPE | −79.01 | −94.88 | −102.08 | −119.95 | −100 |
PMP | −167.91 | −186.91 | −191.44 | −223.49 | −200 |
Teflon | 941.50 | 968.34 | 932.32 | 1079.30 | 990 |
CTP404 | 103.64 | 106.06 | 92.99 | 96.00 | 100 |
3.B. Localization accuracy of Mevion‐AIRO system
Localization accuracy results using the Stereophan phantom (based on two trials) demonstrated maximum AIRO‐kV/kV shift differences of 0.1 mm in the x‐direction, 0.1 mm in the y‐direction, and 0.2 mm in the z‐direction. (Rotations were not included in the analysis because the target was a single BB inside of the phantom.) The Mimi Phantom was used to characterize the rotational corrections. The RMS values were between 0.22 mm and 0.28 mm. Localization accuracy (based on the MIMI phantom) was within 0.6°. Registration (kV/kV) was within 0.5° and 0.5 mm from 3D/3D registrations. These results included robotic couch motion from the CT scanner imaging position to the standard kV imaging isocenter.
The localization accuracy tests also included a workflow and time analysis to determine the optimal location of the CT scanner in the treatment room and to gauge the time required for a CT scan. The resultant workflow required one therapist to bring the CT scanner into the room while an additional therapist performed the initial patient setup. An additional 5–10 min were needed to acquire a CT scan.
3.C. CTDIvol
Measured CTDIvol for the AIRO, and Philips CT Simulator are illustrated in Table 6. The AIRO had the highest dose especially for the head protocol which gave 27.14 mGy more dose than the Philips CT Simulator. The standard helical AIRO head protocol (120 kV, 240 mA, 326 mAs (effective), 1.92 s rotation time, 1.06 mm slice collimation, 1.0 mm slice width, 1.415 pitch factor, soft kernel) gave 64.7 mGy CTDIvol. Measured CTDIvol was 72.05 mGy (Table 6). The helical AIRO abdomen protocol (120 kV, 110 mA, 149 mAs (effective), 1.92 s rotation time, 1.06 mm slice collimation, 1.0 mm slice width, 1.415 pitch factor, standard kernel) gave 22.3 mGy CTDIvol. Measured CTDIvol was 26.08 mGy (Table 6). The Philips clinical 1.5 mm head protocol (120 kV, 339 mA, 580 mAs (effective), 0.75 s rotation time, 1.5 mm slice thickness, 0.438 pitch factor) and clinical 3.0 mm body protocol (120 kV, 488 mA, 450 mAs (effective), 0.75 s rotation time, 3.0 mm slice thickness, 0.813 pitch factor) was used. CTDIw was also measured on the EDGE (pelvis protocol, Table 1). Dose measurements were 15.6 mGy for head and 55.3 mGy for abdomen. In comparison, Amer et al. reported CBDIw (cone beam CT dose index) values of 1.6 mGy, 6 mGy and 25 mGy for head, lung, and pelvis CBCT protocols, respectively, on an Elekta Synergy treatment machine.16 For the AIRO and the Philips Simulators, dose was within the ACR requirements of 20% of manufacturer‐specifications.
Table 6.
CTDIvol (mGy) | ||
---|---|---|
Airo | Philips | |
HEAD | 72.05 | 44.91 |
ABDOMEN | 26.08 | 24.09 |
4. Discussion
This study describes the characterization and commissioning of the AIRO mobile CT system for IGPT. Our characterization of the AIRO's image quality demonstrated that the spatial resolution of the AIRO was lower than our other simulators and IGRT systems. Weir et al. reported similar findings about the AIRO indicating its poor spatial resolution when compared to a Siemens Sensation 64 Slice CT scanner. Weir et al. also reported ring artifacts which could not be eliminated. We found similar ring artifacts present when phantoms had widths at or near the width of the detector (32 cm). Ring artifacts were most noticeable at the phantom/air interface in cylindrical phantoms. These artifacts were not present in anthropomorphic phantoms, thus we conclude that ring artifacts should not appear on patient scans. Image quality is protocol and reconstruction kernel dependent. We also found a large scaling discrepancy in comparison to the Edge and simulators. We attribute this difference to the AIRO's large bore. Scaling discrepancy is calculated based on the image's magnification and the length of the longest side of the image.17 One last potential disadvantage of the AIRO for daily IGRT is the additional patient dose. We measured AIRO's CTDIvol in the range of 26.1–72.1 mGy. These doses are slightly higher than the values measured for the Edge CBCT (15.6–55.3 mGy), and may need to be taken into account if multiple CT datasets are acquired on a daily basis.
Although the AIRO's image quality is worse than CT simulators and the Edge CBCT, the AIRO's localization accuracy was within 0.6 mm and 0.5° measured using the MIMI phantom. Additionally, we demonstrated that the localization accuracy of the Mevion‐AIRO system using the Stereophan phantom was within 0.2 mm when compared with orthogonal kV/kV pairs. This accuracy is comparable to other commonly used IGPT modalities. Landry et al. reported 1.4 mm localization accuracy for a gantry‐mounted CBCT system18 and Xu et al. 2016) reported the accuracy of the Edge localization kV/kV pair to be within ±0.3 mm and ±0.3°.19 It is also important to note that the AIRO's localization accuracy included the motion of the robotic couch from treatment isocenter to the CT location. We suggest that CT scanner QA include additional robotic couch tests to insure the AIRO's localization accuracy is not adversely affected by robotic couch movement. Based on this phantom study and initial commissioning, the AIRO can be used for accurate and reliable IGPT.
Although other facilities have acquired mobile helical CT systems, such as the BodyTom (Neurologica Corporation, Danvers, MA, USA) and the Aquilion LB (Toshiba), our proton therapy facility is the first to acquire the AIRO Mobile CT System for proton treatment localization.9, 10, 11 We are also the first to use this technology for patient localization in a small‐compact proton facility where space limitations exist. Based on our study, the AIRO Mobile CT System can be utilized for accurate and safe image‐guided proton therapy. At the time of writing this manuscript, we have commissioned the system and begun clinical use of the system for treatment localization. Our preliminary findings are that an additional 3–5 min of treatment time are required to bring the scanner into the treatment room. Future work will involve a time study for patient localization and evaluation scanning as well as characterization of the CT images as they relate to stopping power and CT number constancy for use in adaptive proton therapy.
Conclusions
This work evaluated various image quality, localization accuracy, and dosimetric metrics of the AIRO and compared them to other common modalities. Based on our findings, we recommend that the AIRO Mobile CT System can be applied clinically for safe and accurate image‐guided proton therapy.
conflict of interest
There are no conflicts of interest.
References
- 1. Wen N, Li H, Song K, et al. Characteristics of a novel treatment system for linear accelerator‐based stereotactic radiosurgery. J Appl Clin Med Phys. 2015;16:5313. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Beltran C, Pai Panandiker A, Krasin M, Merchant T. Daily image‐guided localization for neuroblastoma. J Appl Clin Med Phys. 2010;11:3388. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Alvarado R, Booth J, Bromley R, Gustafasson H. An investigation of image guidance dose for breast radiotherapy. J Appl Clin Med Phys. 2013;14:4085. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Fu L, Perera H, Ying X, Yu Y. Importance of CBCT setup verification for optical‐guided frameless radiosurgery. J Appl Clin Med Phys. 2014;15:4487. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Sonier M, Chu W, Lalani N, Erler D, Cheung P, Korol P. Evaluation of kidney motion and target localization in abdominal SBRT patients. J Appl Clin Med Phys. 2016;17:6406. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Stanley D, Papanikolaou N, Gutierrez A. An evaluation of the stability of image‐quality parameters of Varian on‐board imaging (OBI) and EPID imaging systems. J Appl Clin Med Phys. 2015;16:5088. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Unq N, Wee L, Hackett S, Jones A, Lim T, Harper C. Comparison of low‐dose, half‐rotation, cone‐beam CT with electronic portal imaging device for registration of fiducial markers during prostate radiotherapy. J Appl Clin Med Phys. 2013;14:4249. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Yin F‐F, Wong J, Balter J, et al. AAPM Therapy Imaging Committee Task Group 104, Report No.104;2009.
- 9. Li H, Wu R, Poenisch F, et al. Clinical implementation of in room mobile CT for image guided proton therapy. Med Phys. 2016;43:3451. [Google Scholar]
- 10. Pham R, Sun B, Zhao T, et al. Development of dose‐based image guided proton therapy workflow. Med Phys. 2016;43:3452. [Google Scholar]
- 11. Moriya S, Tachibana H, Hotta K, et al. Effectiveness of daily CT‐based three‐dimensional image guided and adaptive proton therapy. Med Phys. 2016;43:3419. [Google Scholar]
- 12. Zhao T, Sun B, Grantham K, et al. Commissioning and initial experience with the first clinical gantry‐mounted proton therapy system. J Appl Clin Med Phys. 2016;17:24–40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Weir VJ, Zhang J, Bruner AP. Dosimetric characterization and image quality evaluation of the AIRO mobile CT scanner. J XRay Sci Technol. 2015;23:373–381. [DOI] [PubMed] [Google Scholar]
- 14. DoseLab User Manual Version 6.6: Houston, TX: Mobius medical systems, LP; 2014. [Google Scholar]
- 15. Cody DD, Pfeiffer D, McNitt‐Gray MF, Ruckdeschel TG, Strauss KJ, Wilcox P. 2012 Computed Tomography: Quality Control Manual. American College of Radiology;2012. [Google Scholar]
- 16. Amer A, Marchant T, Sykes J, Czajka J, Moore C. Imaging doses from the Elekta Synergy X‐ray cone beam CT system. Br J Radiol. 2007;80:476–482. [DOI] [PubMed] [Google Scholar]
- 17. Lau SKM, Zhao X, Carmona R, et al. Frameless single‐isocenter modulated sterotactic radiosurgery for simultaneous treatment of multiple intracranial metastases. Transl Cancer Res. 2014;3:383–390. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Landry G, Nijhuis R, Dedes G, et al. Investigating CT to CBCT image registration for head and neck proton therapy as a tool for daily dose recalculation. Med Phys. 2015;42:1354–1366. [DOI] [PubMed] [Google Scholar]
- 19. Xu H, Chetty I, Wen N. Analysis of systematic errors with 2D/3D image registration for target localization and treatment delivery in stereotactic radiosurgery. Med Phys. 2016;43:3339. [DOI] [PMC free article] [PubMed] [Google Scholar]