Abstract
A new design of (Model IR‐Seed2) brachytherapy source has been manufactured recently at the Applied Radiation Research School, Nuclear Science and Technology Research Institute in Iran. The source consists of six resin beads (0.5 mm diameter) that are sealed in a cylindrical titanium capsule of 0.7 mm internal and 0.8 mm external diameters. This work aims to evaluate the dosimetric parameters of the newly designed source using experimental measurements and Monte Carlo (MC) simulations. Dosimetric characteristics (dose rate constant, radial dose function, and 2D and 1D anisotropy functions) of the IR‐Seed2 were determined using experimental measurements and MC simulations following the recommendations by the Task Group 43 (TG‐43U1) report of the American Association of Physicists in Medicine (AAPM). MC simulations were performed using the MCNP5 code in water and Plexiglas, and experimental measurements were carried out using thermoluminescent dosimeters (TLD‐GR207A) in Plexiglas phantoms. The measured dose to water in Plexiglas data were used for verification of the accuracy of the source and phantom geometry in the Monte Carlo simulations. The final MC simulated data to water in water were recommended for clinical applications. The MC calculated dose rate constant (Λ) of the IR‐Seed2 seed in water was found to be . Additionally, its radial dose function by line and point source approximations, and , calculated for distances from 0.1 cm to 7 cm. The values of at radial distances from 0.5 cm to 5 cm were measured in a Plexiglas phantom to be between 1.212 and 0.413. The calculated and measured of values for 2D anisotropy function, , were obtained for the radial distances ranging from 1.5 cm to 5 cm and angular range of 0°‐90° in a Plexiglas phantom. Also, the 2D anisotropy function was calculated in water for the clinical application. The results of these investigations show that the uncertainty of the experimental data is within between the measured and simulated data in Plexiglas. Based on these results, the MC‐simulated dosimetric parameters of the new source model in water are presented for its clinical applications in brachytherapy treatments.
PACS number(s): 87.56.bg
Keywords: (Model IR‐Seed2), brachytherapy, Monte Carlo simulation, thermoluminescent dosimeter, TG‐43U1
I. INTRODUCTION
and brachytherapy seeds share a great role in brachytherapy implants in various tumor sites such as the eye and prostate tumors. (1) In addition to the strong inverse‐square law reduction in gamma ray fluence at short distances, low‐energy photon emissions of sources lead to a rapid decrease in radiation dose with increasing distance. Therefore, this effect reduces the unnecessary radiation dose to normal tissue located beyond the tumor. Different designs of sources, such as IsoAid Advantage (IsoAid, LLC, Port Richey, FL), BEBIG model 125. So6 (Eckert & Ziegler BEBIG Inc., Mount Vernon, NY), and Best Medical model 2301 (Best Medical International, Springfield, VA) are commercially available. 2 , 3 Recently, a new design of (Model IR‐Seed2) brachytherapy seed has been produced for clinical applications by the Applied Radiation Research School, Nuclear Science and Technology Research Institute, Tehran, Iran. TG‐43U1 recommends that the dosimetric parameters of each new source model must be determined before its clinical application. (3)
The purpose of this study is to determine the TG‐43U1 recommended dosimetric parameters of the Model IR‐Seed2 source using Monte Carlo (MC) simulation and experimental measurements. The MC simulations were performed using the MCNP5 Monte Carlo code in water and Plexiglas phantom materials. The experimental procedures were performed by using TLD chips in Plexiglas phantoms. The results of these investigations were compared with the published data of other commercially available sources. 4 , 5
II. MATERIALS AND METHODS
A. IR‐Seed2 source
Figure 1 shows a schematic diagram of the IR‐Seed2 source. This source model consists of six spherical beads with percentage weight composition of: H, 8%; C, 90%; N, 0.3%; Cl, 0.7%; I, 1%. To access high activity up to 40 mCi, the IR‐Seed2 seed has been manufactured without X‐ray marker. Diameter of each bead is 0.5 mm and the beads are sealed within a titanium cylindrical tube with external diameter of 0.8 mm. The two ends of this tube are sealed using 0.65 mm thick titanium by a laser welding technique which is seen in radiography/CT images. ; ; and outer diameter of the source is 0.8 mm. The effective active length of the source is .
Figure 1.

Schematic diagram of the Model IR‐Seed2 brachytherapy source.
The radioactive material is uniformly adsorbed on the surface of each resin bead. The activities of the beads within the source are distributed in a symmetrical fashion and they are arranged as follows: each of the outermost beads were 2.2 mCi, the next two beads were 2.3 mCi, and the remaining two beads (closest to the center of the source) were 1.7 mCi.
B. TG‐43U1 dose calculation formalism
Characteristics of the Model IR‐Seed2 source were determined according to the recommendations of the Task Group 43 (TG‐43U1) by the American Association of Physicists in Medicine (AAPM). 3 , 4 Following this protocol, the spatial dose rate distribution around a sealed brachytherapy source can be determined using the following formula:
| (1) |
where Λ is the dose rate constant at a reference point of is the air kerma strength of the source, is the geometry function, is the line radial dose function, and is the 2D anisotropy function. The above quantities are discussed in detail in the AAPM TG‐43 report. (4)
The air kerma strength, , was calculated using the recommended equation below:
| (2) |
Due to the low energy of the photons from and small range of secondary electrons produced by photons emitted from the source, it was assumed in the Monte Carlo calculations that all electrons generated by the photon collisions are absorbed locally, so it was assumed that dose is equal to kerma at all points of interest.
C. Thermoluminescent dosimetry
The dose distribution around the source was measured to water in Plexiglas phantoms using TLD GR207A chips with dimensions of . These chips are also known as 7‐LiF:Mg,Cu,P thermoluminescent dosimeters (Fimel, Vélizy, France). These chips were annealed following the procedure recommended by the manufacturer (240°C, 10 min). (6)
In order to reduce the effect of statistical fluctuations in the measured data, the averages from several TLD chips were used as the representative of the data for a given point. However, the differences of the mass and physical geometries of the TLD chips may lead to some variation in their responses. In order to practically eliminate these effects, a correction factor was introduced for each TLD chip by calculating the ratio of the measured responses of the individual chip to the average response from the entire batch, when all were exposed to the same dose. Each correction factor is referred to as the element correction coefficient (ECC).
C.1 Calibration
The ECCs of the TLDs and the dose‐response curve were acquired via irradiation by 6 MV and 120 kVp X‐ray beams, respectively. The dose‐response curve was obtained using doses ranging from 50 cGy to 300 cGy. This dose range was selected to cover the dose range that was used in this research project. However, the TLD calibration factor (ε: response per cGy) was determined using a 120 kVp X‐ray beam. Figure 2 displays the TLD dose‐response curve.
Figure 2.

TLD calibration curve with an 120 kVp X‐ray beam.
The irradiated TLDs were read out by an LTM TLD reader (Fimel, Vélizy, France). The following equation was used to calculate the dose rate per air‐kerma strength from the TLD responses for each point irradiated in the phantom:
| (3) |
where was the absorbed dose rate at the start of the irradiation at a point (r, θ), R was the TLD response that had been corrected for physical differences between the TLD chips and the background using the predetermined chip factors, 1 , 7 , 8 T was the irradiation time (hours), and ε was the calibration factor for the TLD response (nC/cGy). To obtain the calibration factor, 18 TLD chips were placed in a custom‐designed Plexiglas slab phantom in a kilovoltage X‐ray (120 kVp) field that was calibrated using the TRS 398 protocol of the International Atomic Energy Agency (IAEA). (9) E(r) was the correction factor for the energy dependence of the TLDs between the calibration beam and the photons, which is equal to unity in this study. (10) The parameter d(T) is a correction factor which was used to account for source decay during the exposure. (7) is the nonlinearity correction factor of the TLD response for the given dose. Responses of the TLDs were linear within the range of the doses used in this project.
Two Plexiglas phantoms with dimensions of were used to measure radial dose function g(r), dose rate constant (Λ) and 2D anisotropy function of the source. (11) The thickness of each phantom was 1 cm. The overall phantom setup for the experimental procedures were composed of 15 slabs of Plexiglas. The central slab for each phantom was machined to house the TLDs. The experimental setup for measurement of the anisotropy function and radial dose function are shown in Figs. 3 and 4. The error propagation of the experimental data is shown in Table 1. The seed geometry error is due to differences of air gap distance between spheres and resin bead diameters in Monte Carlo simulation and reality.
Figure 3.

Schematic diagram of the experimental setup for measurement of radial dose function.
Figure 4.

Schematic diagram of the experimental setup for measurement of anisotropy function.
Table 1.
Uncertainty determination for the experimental measurements using TLDs and Monte Carlo simulation
| Component | TLD Uncertainties | |||
|---|---|---|---|---|
| Type A | Type B | |||
| Repetitive TLD measurements | 4.4% | |||
| TLD dose calibration | 1.41% | |||
| Source to TLD position | 1.1% | |||
| Correction of energy dependence of LiF | 5.5% | |||
| Quadrature combination | 4.4% | 5.8% | ||
| Total combined standard uncertainty | 7.2% | |||
| Monte Carlo Uncertainties | ||||
| Component |
|
|
||
| Statistics | 0.05% | 0.07% | ||
| Photoionizationa(3) | 1.5% | 4.5% | ||
| Cross‐section (2.3%) (3) | ||||
| Seed geometry(3) | 2.0% | 2.0% | ||
| Source energy spectruma(3) | 0.1% | 0.3% | ||
| Quadrature sum | 2.5% | 4.5% | ||
a On the transverse plane.
D. Monte Carlo calculations
Version 5 of the MCNP code developed by Los Alamos National Laboratory (Los Alamos, NM) was used to perform the simulations for these investigations. (12) There are several different tally types available in the MCNP5 code for scoring diverse physical characteristics.(12)*F4 tally was used to determine the energy flux in MeV/cm2, which could be converted to absorbed dose by applying suitable coefficients. The photon spectrum and photoionization in this simulation were extracted from TG‐43U1 report. In these calculations, the titanium characteristic X‐ray production was suppressed with the energy cutoff . (13) In the Monte Carlo calculation, according to the cross‐section data in the Monte Carlo library, the effect of self‐adsorptions has been automatically considered. Simulations were performed to calculate absorbed dose to water in Plexiglas in order to provide data comparable with the TLD measurements. Once the Monte Carlo calculations in Plexiglas were shown to be in agreement with the TLD data (within experimental uncertainty), the calculations were performed to water in water to obtain data for clinical applications per TG‐43 protocol.
A spherical water phantom of 30 cm diameter (with an atomic ratio of 2:1 for H:O and ) was modeled. The phantom size in this simulation was comparable with the experimental setup (i.e., ). The composition of the Plexiglas was H, 8%; C, 60%; and O, 32%, with a mass density of 1.19 g/cm3. (14) To calculate its dosimetric parameters, the seed was simulated in the center of the phantom and the simulations were performed for radial distances of r = 0.1, 0.2, …, 7 cm away from the source and at polar angles relative to the longitudinal axis of the seed from 0° to 90° with 5° increments.
III. RESULTS
A. Dose rate constant
The dose rate constant, Λ, of the source calculated in water was found to be (Table 2). Table 2 also presents a comparison of dose rate constant obtained for this source model with those for the other commercially available brachytherapy sources.
Table 2.
Comparison of the calculated dose rate constant, Λ, obtained in this study and some of those reported in the literature
| Source Type | Method | Medium |
Dose‐Rate Constant
|
|
|---|---|---|---|---|
| IR‐Seed2 | Monte Carlo (MCNP5) | Liquid water |
|
|
| Amersham 6702 (3) | Monte Carlo | Liquid water | 1.036 | |
| IBt 1251L(16) | Monte Carlo | Liquid water | 1.038 | |
| MBI SL‐125/SH‐125(16) | Monte Carlo | Liquid water | 0.953 |
B. Radial dose function
Figure 5(a) shows a comparison between the calculated and measured radial dose functions of the IR‐Seed2 in a Plexiglas phantom. The uncertainties of the measured and calculated data were and , respectively. A good agreement (within experimental uncertainties) between the measured and calculated values can be seen in this figure. Figure 5(b) shows a comparison between the calculated of the Model IR‐Seed2 seed with three other available brachytherapy sources. 3 , 15 The values of and in Plexiglas and water phantoms are listed in Table 3.
Figure 5.

A comparison (a) between the measured and calculated radial dose function of the Model IR‐Seed2 source in a Plexiglas phantom. The solid line represents the fifth‐order polynominal fit to the data of the present study. The error bars represent ; (b) shows a comparison of the calculated radial dose function of the IR‐Seed2 source in water with three other brachytherapy sources. The line represents the fifth‐order polynominal fit to the data of the present study. The error bars represent .
Table 3.
Calculated and measurement radial dose function, and values for line and point‐source approximation, of the IR‐Seed2 brachytherapy source in Plexiglas and water
| Distance From Source Center r (cm) |
|
|
|
g P(r)Monte Carlo (Water) | |||
|---|---|---|---|---|---|---|---|
| 0.1 | ‐ | 1.164 | 1.126 | 0.900 | |||
| 0.2 | ‐ | 1.176 | 1.148 | 1.010 | |||
| 0.3 | ‐ | 1.192 | 1.167 | 1.085 | |||
| 0.4 | ‐ | 1.178 | 1.157 | 1.082 | |||
| 0.5 | 1.212 | 1.149 | 1.132 | 1.094 | |||
| 1 | 1.000 | 1.000 | 1.000 | 1.000 | |||
| 1.5 | 0.923 | 0.929 | 0.945 | 0.950 | |||
| 2 | 0.833 | 0.851 | 0.876 | 0.883 | |||
| 2.5 | ‐ | 0.770 | 0.799 | 0.806 | |||
| 3 | 0.682 | 0.679 | 0.722 | 0.729 | |||
| 3.5 | ‐ | 0.604 | 0.648 | 0.655 | |||
| 4 | 0.543 | 0.527 | 0.579 | 0.585 | |||
| 4.5 | ‐ | 0.457 | 0.512 | 0.517 | |||
| 5 | 0.413 | 0.399 | 0.452 | 0.457 | |||
| 5.5 | ‐ | 0.349 | 0.398 | 0.402 | |||
| 6 | ‐ | 0.300 | 0.349 | 0.353 | |||
| 6.5 | ‐ | 0.260 | 0.306 | 0.309 | |||
| 7 | ‐ | 0.225 | 0.267 | 0.270 |
The measurements indicate up to 5% differences in radial dose function values for the IR‐Seed2 seed at distance of 0.5 cm compared to the MC simulated data due to the high dose gradient. The radial dose function in water for clinical application was fitted to a fifth order polynomial function as follows:
| (4) |
where .
C. Anisotropy function
The anisotropy function, , of the IR‐Seed2 was measured and calculated at 30° and 5° intervals, respectively, at radial distances of 1.5, 2, 3, and 5 cm relative to the center of the seed and polar angles (θ) ranging from 0° to 90°. The results can be seen in Table 4. Table 5 shows the MC‐calculated data in water ranging from 0.5 cm to 7 cm. Figure 6 shows a comparison between the calculated and measured at various angles in the Plexiglas phantom. The uncertainties of the measured and calculated data were and , respectively. Because of the limited size of the TLD‐GR207A chips and the uncertainty of the experiments during the multifold TLD measurements, this large variation in dose cannot be fully tracked. Figure 7 presents the variations in the calculated in water at various distances.
Table 4.
Measured and calculated anisotropy function values of the Model IR‐Seed2 source in a Plexiglas phantom
| Angle θ (degrees) |
Measured
|
Calculated
|
|||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
r=2 cm | r=3 cm |
|
||||||
| 0 | 0.282 | 0.319 | 0.348 | 0.400 | 0.344 | 0.373 | 0.435 | 0.487 | |||||
| 5 | ‐ | ‐ | ‐ | ‐ | 0.355 | 0.394 | 0.443 | 0.502 | |||||
| 10 | ‐ | ‐ | ‐ | 0.417 | 0.449 | 0.499 | 0.548 | ||||||
| 15 | ‐ | ‐ | ‐ | ‐ | 0.507 | 0.535 | 0.573 | 0.610 | |||||
| 20 | ‐ | ‐ | ‐ | ‐ | 0.595 | 0.614 | 0.642 | 0.671 | |||||
| 25 | ‐ | ‐ | ‐ | ‐ | 0.670 | 0.685 | 0.705 | 0.726 | |||||
| 30 | 0.757 | 0.767 | 0.726 | 0.751 | 0.738 | 0.748 | 0.761 | 0.776 | |||||
| 35 | ‐ | ‐ | ‐ | ‐ | 0.787 | 0.792 | 0.803 | 0.813 | |||||
| 40 | ‐ | ‐ | ‐ | ‐ | 0.829 | 0.833 | 0.840 | 0.847 | |||||
| 45 | ‐ | ‐ | ‐ | ‐ | 0.862 | 0.866 | 0.874 | 0.877 | |||||
| 50 | ‐ | ‐ | ‐ | ‐ | 0.894 | 0.896 | 0.901 | 0.904 | |||||
| 55 | ‐ | ‐ | ‐ | ‐ | 0.921 | 0.922 | 0.928 | 0.929 | |||||
| 60 | 0.894 | 0.994 | 0.891 | 0.892 | 0.946 | 0.944 | 0.947 | 0.948 | |||||
| 65 | ‐ | ‐ | ‐ | ‐ | 0.965 | 0.964 | 0.966 | 0.967 | |||||
| 70 | ‐ | ‐ | ‐ | ‐ | 0.981 | 0.978 | 0.980 | 0.980 | |||||
| 75 | ‐ | ‐ | ‐ | ‐ | 0.991 | 0.989 | 0.992 | 0.990 | |||||
| 80 | ‐ | ‐ | ‐ | ‐ | 0.998 | 0.995 | 0.997 | 0.994 | |||||
| 85 | ‐ | ‐ | ‐ | ‐ | 0.999 | 0.998 | 0.999 | 0.999 | |||||
| 90 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | |||||
|
|
0.857 | 0.849 | 0.841 | 0.863 | 0.859 | 0.860 | 0.858 | 0.878 | |||||
Table 5.
MC‐calculated anisotropy function values of the Model IR‐Seed2 brachytherapy source in water
| Angle (degrees) | 0.5 cm | 1 cm | 1.5 cm | 2 cm | 3 cm | 4 cm | 5 cm | 7 cm | |
|---|---|---|---|---|---|---|---|---|---|
| 0 | 0.230 | 0.289 | 0.344 | 0.373 | 0.435 | 0.473 | 0.487 | 0.505 | |
| 5 | 0.229 | 0.305 | 0.355 | 0.394 | 0.443 | 0.476 | 0.502 | 0.531 | |
| 10 | 0.277 | 0.366 | 0.417 | 0.449 | 0.499 | 0.523 | 0.548 | 0.570 | |
| 15 | 0.403 | 0.466 | 0.507 | 0.535 | 0.573 | 0.590 | 0.610 | 0.630 | |
| 20 | 0.529 | 0.566 | 0.595 | 0.614 | 0.642 | 0.658 | 0.671 | 0.684 | |
| 25 | 0.636 | 0.651 | 0.670 | 0.685 | 0.705 | 0.714 | 0.726 | 0.737 | |
| 30 | 0.716 | 0.725 | 0.738 | 0.748 | 0.761 | 0.764 | 0.776 | 0.780 | |
| 35 | 0.776 | 0.778 | 0.787 | 0.792 | 0.803 | 0.807 | 0.813 | 0.818 | |
| 40 | 0.825 | 0.822 | 0.829 | 0.833 | 0.840 | 0.842 | 0.847 | 0.850 | |
| 45 | 0.865 | 0.858 | 0.862 | 0.866 | 0.874 | 0.871 | 0.877 | 0.881 | |
| 50 | 0.899 | 0.891 | 0.894 | 0.896 | 0.901 | 0.899 | 0.904 | 0.906 | |
| 55 | 0.929 | 0.921 | 0.921 | 0.922 | 0.928 | 0.921 | 0.929 | 0.927 | |
| 60 | 0.952 | 0.945 | 0.946 | 0.944 | 0.947 | 0.942 | 0.948 | 0.945 | |
| 65 | 0.971 | 0.966 | 0.965 | 0.964 | 0.966 | 0.964 | 0.967 | 0.968 | |
| 70 | 0.982 | 0.981 | 0.981 | 0.978 | 0.980 | 0.978 | 0.980 | 0.981 | |
| 75 | 0.990 | 0.992 | 0.991 | 0.989 | 0.992 | 0.985 | 0.990 | 0.988 | |
| 80 | 0.995 | 0.997 | 0.998 | 0.995 | 0.997 | 0.992 | 0.994 | 0.995 | |
| 85 | 0.999 | 0.999 | 0.999 | 0.998 | 0.999 | 0.992 | 0.999 | 0.997 | |
| 90 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | |
|
|
0.898 | 0.849 | 0.844 | 0.850 | 0.856 | 0.857 | 0.860 | 0.862 |
Figure 6.

A comparison of the measured and calculated anisotropy function of the IR‐Seed2 source in a Plexiglas phantom at 2 cm, 3 cm, and 5 cm distances. The line show a 4th polynominal fit to the data.
Figure 7.

The variation of the calculated anisotropy function of the Model IR‐Seed2 source in water at distances ranging from 0.5 cm to 5 cm.
A comparison between the calculated anisotropy function of the new source at 5 cm radii from the axis of the seed in water with previously published data for three other brachytherapy sources is shown in Fig. 8. (15)
Figure 8.

Comparison of the MC calculated anisotropy functions of the Model IR‐Seed2 source in water with the three other brachytherapy sources. The error bars on the data for the IR‐Seed2 source are .
IV. DISCUSSION
This study was performed to evaluate the dosimetric parameters of a newly designed brachytherapy source (IR‐Seed2) using TLD measurements and MC calculations in water and Plexiglas phantoms. These determinations were performed in accordance with the TG‐43U1 recommendations. (3) The results of these investigations were compared with the published data by the Task Group 43 report. 3 , 15 The value of the dose rate constant, Λ, in water was calculated for clinical applications and was found to be , which is close to the dose‐rate constant of source, for 6702, (3) MBI, (15) and IBt (15) models. The calculated dosimetric parameters of the source are within of the published data for those brachytherapy sources. In this project, high precision was obtained in the results of the simulations, due to simulating large numbers of histories (up to ) and using variance reduction methods.
The radial dose function, , of the IR‐Seed2 seed was measured using TLDs and MC simulations were also performed in water and Plexiglas phantoms. Figure 5(a) shows an excellent agreement (within experimental uncertainty) between the measured and calculated values in the Plexiglas phantom. This agreement validates the accuracy of the source and phantom geometry used in the MC simulations in the range of 0.1 cm to 7 cm.
The 2D anisotropy function, , of the IR‐Seed2 was measured and calculated. The measured for this source in Plexiglas is in good agreement (within experimental uncertainty) with the calculated values in the same medium (Fig. 6). The discrepancy between the measured and calculated anisotropy functions at different distances of 2 cm, 3 cm, and 5 cm is within experimental uncertainty (). Figure 8 displays a discrepancy of between the MC calculated anisotropy function of the IR‐Seed2 source to water in water with the published data of the 6702 , IBt Model 1251L, and Mills Biopharmaceuticals (Mills Biopharmaceuticals Inc., Oklahoma City, OK) model SL‐125/SH‐125 sources at a distance of 5 cm. 3 , 15 This figure indicates larger anisotropy functions at small angles for the International Brachytherapy InterSource125, which can be attributed to the absence of end caps on that source model. The values of the calculated and measured anisotropy functions, and anisotropy factors for the IR‐Seed2 source are listed in Tables 4 and 5.
These results show that the higher discrepancies between the calculated and measured values are related to those points which are located in the longitudinal plane of the seed. These discrepancies are due to self‐adsorption and oblique filtration of the radiation in the encapsulating material. The IR‐Seed2 has thicker end caps than the other listed sources in this project.
Overall, the discrepancies between the calculated and measured values of g(r) and in the distances close to the seed are due to the high gradient dose and relatively large dimensions of TLDs compared to the phantom and seed sizes.
In IR‐Seed2 source, the radioactive material is nonuniformly adsorbed on the surfaces of the resin beads, unlike the other commercially available brachytherapy sources which have been distributed uniformly. Ignoring this property, this source can be used clinically for catheter‐based brachytherapy applications.
V. CONCLUSION
The dosimetric parameters of the Model IR‐Seed2 source have been determined experimentally and theoretically based on the TG‐43U1 recommendations.3 A complete set of both measured and calculated data was presented herein for this source in water and Plexiglas phantoms.
The MC simulation results to water in water are recommended for clinical application of this source model. The data of IR‐Seed2 are comparable to other commercially available brachytherapy sources. Based on these acceptable results, the IR‐Seed2 source is being used just for eye plaque and brain implants that are catheter‐ or applicator‐based implants for their localization.
The design of this source is in progress in terms of X‐ray marker use, and of its application in prostate brachytherapy implantation.
ACKNOWLEDGMENTS
The authors would like to thank the staff of Radiotherapy Physics Department and Cancer Research Centre for their assistance in performing this project.
COPYRIGHT
This work is licensed under a Creative Commons Attribution 3.0 Unported License.
Supporting information
Supplementary Material
REFERENCES
- 1. Gearheart DM, Drogin A, Sowards K, Meigooni AS, Ibbott GS. Dosimetric characteristics of a new 125I brachytherapy source. Med Phys. 2000;27(10):2278–85. [DOI] [PubMed] [Google Scholar]
- 2. Meigooni AS, Hayes JL, Zhang H, Sowards K. Experimental and theoretical determination of dosimetric characteristics of IsoAid ADVANTAGE 125I brachytherapy source. Med Phys. 2002;29(9):2152–58. [DOI] [PubMed] [Google Scholar]
- 3. Rivard MJ, Coursey BM, DeWerd LA et al. Update of AAPM Task Group No. 43 Report: a revised AAPM protocol for brachytherapy dose calculations. Med Phys. 2004;31(3):633–74. [DOI] [PubMed] [Google Scholar]
- 4. Nath R, Anderson LL, Luxton G, Weaver KA, Williamson JF, Meigooni AS. Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No. 43. American Association of Physicists in Medicine. Med Phys. 1995;22(2):209–34. [DOI] [PubMed] [Google Scholar]
- 5. Meigooni AS, Williamson JF, Nath R. Single source dosimetry for interstitial brachytherapy. In: Williamson JF, Thomadsen BR, Nath R, editors. Brachytherapy Physics. Madison, WI: Medical Physics Publishing Corporation;1995. p.210–33. [Google Scholar]
- 6. Perni5ka F and McLean ID. Technical Reports Series 457. Dosimetry in diagnostic radiology: an international code of practice. Vienna: IAEA; 2007. [Google Scholar]
- 7. Meigooni AS, Mishra V, Panth H, Williamson J. Instrumentation and dosimeter‐size artifacts in quantitative thermoluminescent dosimetry of low‐dose fields. Med Phys. 1995;22(5):555–61. [DOI] [PubMed] [Google Scholar]
- 8. Meigooni AS, Bharucha Z, Yoe‐Sein M, Sowards K. Dosimetric characteristics of the Best double‐wall 103Pd brachytherapy source. Med Phys. 2001;28(12):2568–75. [DOI] [PubMed] [Google Scholar]
- 9. International Atomic Energy Agency . Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based standards of absorbed dose to water. Technical reports series No. 398. Vienna: IAEA; 2001. [Google Scholar]
- 10. Muench PJ, Meigooni AS, Nath R, McLaughlin WL. Photon energy dependence of the sensitivity of radiochromic film and comparison with silver halide film and LiF TLDs used for brachytherapy dosimetry. Med Phys. 1991;18(4):769–75. [DOI] [PubMed] [Google Scholar]
- 11. Meigooni AS, Gearheart DM, Sowards K. Experimental determination of dosimetric characteristics of Best 125I brachytherapy source. Med Phys. 2000;27(9):2168–73. [DOI] [PubMed] [Google Scholar]
- 12. MCNP‐5Monte Carlo Team . MCNP — a general Monte Carlo N‐particle transport code, version 5. Los Alamos, NM: Los Alamos National Laboratory; 2003. [Google Scholar]
- 13. Popescu CC, Wise J, Sowards K, Meigooni AS, Ibbott GS. Dosimetric characteristic of the PharmaSeed model BT‐125‐I source. Med Phys. 2000;27(9):2174–81. [DOI] [PubMed] [Google Scholar]
- 14. Saidi P, Sadeghi M, Shirazi A, Tenreiro C. Monte Carlo calculation of dosimetry parameters for the IR08‐103Pd brachytherapy source. Med Phys. 2010;37(6):2509–15. [DOI] [PubMed] [Google Scholar]