Abstract
The purpose of this study is to accurately and effectively automate the calculation of the water‐equivalent diameter () from 3D CT images for estimating the size‐specific dose. is the metric that characterizes the patient size and attenuation. In this study, was calculated for standard CTDI phantoms and patient images. Two types of phantom were used, one representing the head with a diameter of 16 cm and the other representing the body with a diameter of 32 cm. Images of 63 patients were also taken, 32 who had undergone a CT head examination and 31 who had undergone a CT thorax examination. There are three main parts to our algorithm for automated calculation. The first part is to read 3D images and convert the CT data into Hounsfield units (HU). The second part is to find the contour of the phantoms or patients automatically. And the third part is to automate the calculation of based on the automated contouring for every slice (). The results of this study show that the automated calculation of and the manual calculation are in good agreement for phantoms and patients. The differences between the automated calculation of and the manual calculation are less than 0.5%. The results of this study also show that the estimating of using (central slice along longitudinal axis) produces percentage differences of and , and estimating using produces percentage differences of and , for thorax and head examinations, respectively. From this study, the percentage differences between normalized size‐specific dose estimate for every slice () and are and for thorax and head examinations, respectively; between and are and for thorax and head examinations, respectively.
PACS number(s): 87.57.Q‐, 87.57.uq‐
Keywords: water‐equivalent diameter (), size‐specific dose estimates (SSDE), volume CT dose index (), patient dose
I. INTRODUCTION
CT dose is an important concept in the medical community, 1 , 2 , 3 , 4 especially due to the growing use of computed tomography (CT) examinations 5 , 6 and concern over the relatively high dose from CT compared with other modalities. 7 , 8 , 9 There is an ongoing need to estimate the radiation dose from CT examinations accurately. Currently, CT dose is characterized in terms of the CT Dose Index (CTDI) and dose‐length product (DLP). 10 , 11 CTDI values are widely used for quality assurance (12) and accreditation purposes. (13) However, CTDI was not intended to represent patient dose; 14 , 15 it was designed to approximate the average dose to the central slice of a cylindrical acrylic (polymethyl methaacrylate (PMMA)) phantom in a contiguous axial or helical examination. (4)
There are limitations to the CTDI values that are calculated using standard PMMA phantoms, which are typically 14 to 15 cm long and 16 cm in diameter to represent a patient's head or 32 cm in diameter to represent a patient's body. The first limitation is that the length of phantom is too short for measuring scattered radiation. 16 , 17 , 18 , 19 The second limitation is that the diameter of the phantom does not represent the variation in diameters among patients. 20 , 21 , 22 For a given CT technique, the patient dose decreases as patient size increases, due to the increased attenuation of the incident X‐ray beam. The third limitation is that the phantom is homogenous, whereas a patient is composed of many materials with different attenuation properties. 23 , 24 For these reasons it has been recognized that CTDI represents the output radiation of the CT scanner rather than the patient dose.
Regarding the first limitation, CTDI is often measured using a 10 cm ionization chamber, which is considered too short because it excludes any contribution from radiation scattered beyond the relatively short range of integration along the Z‐direction, and it tends to undervalue the cumulative dose at the center position, as reported by many investigators. 16 , 17 , 18 , 19 To overcome this limitation, the American Association of Physicists in Medicine (AAPM) issued report TG‐111 which recommended using a sufficiently long (e.g., at least 45 cm) phantom to accommodate the scattered radiation. (25) An alternative is to use a small‐volume ion chamber to directly measure the cumulative dose at any point by scanning the length of the phantom long enough to produce dose equilibrium in the center. 25 , 26
The second limitation has been evaluated by many authors who have reported that the dose to the patient depends on the size of patients. 20 , 27 , 28 Huda et al. (20) reported that the effective dose show an inverse correlation with increasing patient weight and size. The radiation dose was determined as a function of patient size (i.e., weight), ranging from young infants to overweight adults. However, AAPM provides methods that can be used to take into account the size of patient and, in 2011, introduced a new metric called the size‐specific dose estimates (SSDE).(29)In this report, patient size is estimated based on a geometrical parameter of the patient called the effective diameter (ED). Look‐up tables in the report provide conversion factors that can be applied to to calculate SSDE. Conversion factors are based on four different measurements: anterior‐posterior (AP), lateral (LAT), , and effective diameter. 29 , 30
The effective diameter is a simple physical measure of patient size, and does not account for patient composition and attenuation properties. 23 , 31 , 32 However, X‐ray attenuation is the fundamental physical parameter affecting the absorption of X‐rays and hence it determines the dose to the patient. (33) This limitation was addressed by AAPM report TG‐220. (34) In this report, AAPM used the concept of water‐equivalent diameter (). It had previously been proposed by Huda et al., (20) Wang et al., (23) and Toth et al. (35)
The value of can be calculated either by a semiautomated (i.e., with some manual intervention) or by a fully automated method. The purpose of this study was to calculate the water‐equivalent diameter () from CT images using a fully automated method. The resulting estimate should be sufficiently accurate for computing the size‐specific dose estimate (SSDE). The study also investigated the difference between the water‐equivalent diameter using different numbers of slices, n (giving ), and the water‐equivalent diameter using all slices ().
II. MATERIALS AND METHODS
A. The images of phantoms and patients
The objects for the calculations were standard CTDI phantoms (Leeds Test Objects (LTO) Medical Imaging Phantoms, Boroughbridge, UK) and patients. The phantoms consisted of two PMMA cylinders, one cylinder representing the head (16 cm diameter), and when nested inside the other representing the body (32 cm). The length of each cylinder was 14 cm. The head phantom was scanned by a CT scan Siemens SOMATOM Sensation 64 (Siemens Medical Solutions, Malvern, PA), installed at Dr. Karyadi Hospital, Semarang, Indonesia. The scanning parameters were: head spiral for adult study, slice thickness 0.3 cm, voltage 120 kVp, exposure 370 mAs, reconstruction diameter 18.7 cm, convolution kernel H20s. The body phantom was scanned by the same scanner with: abdomen routine for adult study, slice thickness 0.3 cm, voltage 120 kVp, exposure 200 mAs, reconstruction diameter 35.0 cm, and convolution kernel B30f.
There were 63 patients, 32 who underwent a CT head examination and 31 who underwent a CT thorax examination. The patients who underwent CT head examinations comprised 20 females and 12 males, with ages between 16 and 77 years. Those who underwent CT thorax examinations comprised 13 females and 18 males aged between 13 and 85 years.
The patients were scanned by a CT scan Siemens SOMATOM Emotion 6 at the Prof. Dr. Margono Hospital, Purwokerto, Indonesia. The scanning parameters for the head were: slice thickness 0.4 cm, voltage 130 kVp, tube current 167 mA, rotation time 1500 ms, reconstruction diameter 20.0 cm, number of slices 24, and convolution kernel H31s. The scanning parameters for the thorax were: slice thickness 1.0 cm, voltage 130 kVp, tube current 35–133 mA, rotation time 600 ms, reconstruction diameter 28.1 cm, number of slices 29–37, and convolution kernel B20s.
B. The algorithm for automated calculation
In order to estimate , AAPM suggested that the ROI be defined to include all of the patient, but no peripheral objects. (34) A manually drawn ROI based on the patient contour could be used, but this technique requires more time and effort. Instead we implemented an automatically drawn contour around the patient.
Figure 1 shows the flow chart for the automated calculation. There were three main parts to the algorithm. The first part is to read the 3D images and convert the CT data into Hounsfield units (HU). The second part is to contour the patients or phantoms automatically. And the third part is to automate the calculation of based on the automated contouring for every slice.
Figure 1.
Flow chart for automated water‐equivalent diameter () calculation.
Some of the CT images were stored as CT data and some were stored in Hounsfield units (HU). A typical CT dataset ranges from 0 to 4095, although the real pixel values used only part of this range. Hounsfield units correspond to a scale from to The pixel value was displayed in the clinical application according to the average attenuation of the tissue on the Hounsfield scale. Water has an attenuation of 0 Hounsfield units (HU), air is , bone is typically HU or greater, and metallic implants are usually HU.
We applied a linear transformation to convert the CT data to Hounsfield units using Eq. (1), where S is the slope and I is the intercept. Normally, the value of slope and intercept are stored in the DICOM file itself. The tags are generally called the rescale slope and the rescale intercept, and typically have values of 1 and , respectively. Tag (0028, 1052) is for the rescale intercept and tag (0028, 1053) is for the rescale slope.
(1) |
After converting the CT data to Hounsfield units, automated patient or phantom contouring was performed using an algorithm designed to produce accurate results with a relatively fast computation time. The algorithm uses a combination of basic segmentation techniques and specific information about the border of the patient body. The first step was thresholding, using a HU value of . This value was selected because the border between the patient and its surroundings is skin with HU values of approximately zero, while the surroundings outside the patient are air or other materials with HU values lower than . The thresholding produces binary images. However, thresholding alone was not be able to contour the patient completely because of the presence of other objects inside the patient with HU values lower than . To overcome this problem, we used edge detection to identify these objects and labeled them using their areas. The largest area identified was considered to be the border of the patient or phantom. (32) The steps of this automated contouring are shown in Fig. 2.
Figure 2.
Steps for the automatic contouring process: (a) original image, (b) result of thresholding, (c) image after edge detection, (d) result of labeling, (e) patient's border image, and (f) result of autocontouring.
After automated contouring, is calculated. The results of the automated contouring are used to crop the original images, and the area of the cropped image and the average HU value are calculated. is then calculated using Eq. (2), introduced by Wang et al. (23) and adopted by AAPM 220: (34)
(2) |
where is the area of the patient after cropping and is the average HU value of the patient or phantom. After the automated calculation is completed for one slice, it is continued for all the slices in the scan range. This gives the values for all slices, and the average .
C. Validation of
The results of the automated calculation were compared to those calculated manually where the contouring was done freehand. Comparisons were made for both head and body phantoms, and for patient thorax and head. The automated and manual calculations for the patients were then correlated by regression.
In addition to the automated and manual calculations, the values of were also calculated theoretically for the phantoms. The phantoms were made from PMMA material with , approximately 120 HU, 16 cm in diameter for head phantom and 32 cm in diameter for body phantom.
D. Comparisons of to
The calculation of water‐equivalent diameter using all slices () gives an accurate result. However, it is not practical clinically because it requires a long time for calculation, especially in the case of modern CT scanners that produce more than 500 images (i.e., slices) for every examination. We postulated that the value of could be estimated to an acceptable accuracy by using just a few (n) slices (). In this study, we investigated the effect of using just n slices (, and 9) to produce an estimate . The position of every slice used for calculating and is indicated in Fig. 3.
Figure 3.
Slice positions for calculating along the longitudinal axis.
The percentage difference (PD) between and was calculated using Eq. (3):
(3) |
E. Normalized and SSDE calculation
The goal of the calculation is to estimate the size‐specific dose (SSDE). The SSDE value is determined by two main factors, namely the characteristic of the patient in terms of and the output dose of CT scanner in terms of the volume CT dose index ().
The values were calculated using ImPACT CT Patient Dosimetry Calculator version 1.0.1a. The exposure parameters (tube voltage, tube current, rotation time, slice thickness, pitch, and the type of phantom), manufacturer name and the type of scanner in each patient were put into the ImPACT CT in order to calculate the . Because the tube current for thorax examination spanned a wide range (35 mA to 133 mA), we calculated the normalized () instead of for both thorax and head examinations. The unit of is mGy/100 mAs.
After determining the values of and , we calculated the value of normalized SSDE () using Eq. (4). The conversion factor from to is adapted from AAPM Task Group 220:
(4) |
The value of depends on the type of phantom. For this study, there were two phantoms, namely the head CTDI phantom (16 cm in diameter) and the body CTDI phantom (32 cm in diameter).
III. RESULTS
A. for phantoms
The results of contouring both automatically and manually are shown in Fig. 4. It can be seen visually that the results of automated contouring are similar to the manual contouring using free hand. The results of for the body and head phantoms using automated and manual calculations from the images and using theoretical calculations are shown in Table 1. The values for the body and head phantoms based on automated and manual calculations along the longitudinal axis are shown in Fig. 5.
Figure 4.
The first column shows the original images, the second column shows the results of autocontouring, and the third column shows the result of manual contouring. The first row shows images from the head phantom, the second row shows images from the patient head, and the third row shows images from the patient thorax.
Table 1.
The values for the head and body phantoms based on theoretical calculation, and automated and manual calculations from the CT image
Water‐equivalent Diameter () | ||||||
---|---|---|---|---|---|---|
Phantoms | Theoretical Calculation | Manual Calculation | Automated Calculation | |||
Body |
|
|
|
|||
| ||||||
| ||||||
| ||||||
| ||||||
Head |
|
|
|
|||
| ||||||
| ||||||
| ||||||
|
Figure 5.
The values for the head and body phantoms based on automated and manual calculations along the longitudinal axis.
The standard errors in the automated and manual methods are very small, indicating that the methods are precise. The theoretical result falls within the estimated errors (or just outside, in the case of the automated head phantom), indicating that the methods are accurate. The percentage differences the automated calculations to the manual calculations are less than 0.4%. The results indicate that the accuracy of automated calculation of is very good.
B. for patients
The automated and manual for the patient thorax and head are shown in Fig. 6. The resulting linear correlation coefficients between the automated and manual calculations gives for both the thorax and the head, and a slope of 1.008 for thorax and 1.003 for head. The manual values for are and for the thorax and the head, respectively. The automated values for are and for the thorax and the head, respectively. The percentage differences between the automated and manual calculations of are less than 0.5%, which is very small compared to the permitted tolerance, which is 10%. (33) The results indicate that the automated calculations of from patient images are very close to the manual calculations.
Figure 6.
Graph of using automatic and manual calculations for (a) the thorax and (b) the patient head.
C. along the longitudinal axis
values along the longitudinal axis for the thorax and head for one of the patients are shown in Fig. 7. It shows that in the area of the lung, values are lower than the surroundings, because the main composition of the lung is air with HU value around . Figure 7(b) shows that values in the head region are relatively flat, except in the area close to the apex of head.
Figure 7.
Profiles of values along longitudinal axis for (a) thorax and (b) head.
AAPM TG‐220 recommended that water‐equivalent diameter be calculated as the average for every slice along the longitudinal axis (). The average of for thorax and head examinations are listed in Table 2, which shows that the standard deviation of for head examinations is relatively small (). On the other hand, the standard deviation of for thorax examination is relatively high (). Table 2 also indicates the list of based on the gender. It can be seen that there is no significant difference of between male and female patients in the thorax region, but in the head region the value for female patients is slightly higher (about 1 cm) than for male patients.
Table 2.
The values for thorax and head
|
||||
---|---|---|---|---|
Sex | Thorax | Head | ||
Male |
|
|
||
Female |
|
|
||
All |
|
|
Percentage differences between the average and are listed in Table 3. It shows that in the head region, the percentage difference between calculated from the central slice and the average is . It indicates that the value of calculated using the central slice is always greater than . The percentage differences decrease when the number of slices (n) increases. The percentage difference for slices is or about 1%. On the other hand, in the thorax region, the average percentage difference between using the central slice and is relatively small (), but there is a high standard deviation (). The standard deviation decreases with increasing of number of slices used (n). At , the standard deviation of percentage difference is very small ().
Table 3.
Percentage differences of to the average for thorax and head
Percentage Differences of
|
||||||||||
---|---|---|---|---|---|---|---|---|---|---|
Body Part | 1 slice | 3 slices | 5 slices | 7 slices | 9 slices | |||||
Thorax |
|
|
|
|
|
|||||
Head |
|
|
|
|
|
D. Normalized and SSDE
The normalized () values were 9.30 mGy/100 mAs and 25.10 mGy/100 mAs for thorax and head examination, respectively. The normalized SSDE () values based on a manual calculation of are and for thorax and head, respectively. The values based on an automated calculation of are and for thorax and head, respectively. The percentage differences of between the automated and manual calculations are less than 0.5%. As indicated in Fig. 8, the values depend on . It can be seen that the decreases exponentially with increasing .
Figure 8.
Normalized SSDE vs. for (a) thorax and (b) head.
The average for thorax and head examinations is listed in Table 4. It shows that the standard deviation of for head examination is smaller () than for thorax examination ().
Table 4.
The normalized SSDE values for thorax and head
|
||||
---|---|---|---|---|
Sex | Thorax | Head | ||
Male |
|
|
||
Female |
|
|
||
All |
|
|
The percentage differences between with and 9 slices to the are shown in Table 5. The percentage differences between with slices and the average are very small ( and for the thorax and head examinations, respectively).
Table 5.
Percentage differences of to the average of for thorax and head
Percentage Differences of
|
||||||||||
---|---|---|---|---|---|---|---|---|---|---|
Body Part | 1 slice | 3 slices | 5 slices | 7 slices | 9 slices | |||||
Thorax |
|
|
|
|
|
|||||
Head |
|
|
|
|
|
IV. DISCUSSION
The aim of this study was to calculate the water‐equivalent diameter () automatically from axial CT images. A number of authors have reported the calculation of the patient diameter. Cristianson et al. (28) calculated the effective diameter of patients automatically, but the calculation was performed on scout images using an adaptive threshold algorithm with a threshold set to 30% of the maximum pixel value. The limitation of that study, as reviewed by Pourjabbar et al., (30) is that the calculation of patient diameter only in the lateral or in the anterior‐posterior (AP) direction resulted in a high variability. In addition, the calculation of patient diameter from a scout image must be performed carefully. The patient centering must be set correctly, otherwise it can produce a magnified or minified image. Wang et al. (23) showed that placing a 30 cm diameter water cylinder 5 cm and 9 cm closer to the X‐ray tube led to an overestimation of by 4.5% and 9.9%, respectively. In contrast, the calculation of the diameter for axial images requires no patient centering to attain high accuracy.
Patient contouring is very important in defining the patient. Care should be taken in defining the patient to include the entire patient, with only minimal portions of other attenuating materials such as the patient table. Including large amounts of the table would lead to an overestimate of , particularly for small pediatric patients. Li et al. (36) showed that the CT table can contribute up to 12% of the total attenuation for small objects, and Wang et al. (23) reported that the patient table contributes up to 45.7% for a 7.7 cm wide phantom. Using our algorithm minimized the inclusion of parts outside the object (table).
Our method allows to be calculated automatically and with high accuracy. Accuracy is very important, since an overestimation of will lead to an underestimation of SSDE, and vice versa. Additionally, in our study, the percentage differences of the automated from manual calculations were less than 0.5%. The calculation using our algorithm is fast: using a standard notebook (Intel(R) Celeron CPU 1005M, 1.90 GHz, installed memory RAM 2.0 GB, and 32‐bit operating system), the automated calculation for each slice took less than 1 s.
Because patient dimension and attenuation can vary considerably along the longitudinal axis, the water‐equivalent diameter be calculated for every slice or for all positions along that axis (). However, a previous study suggested that could be estimated from the central image in the scan range. Leng et al. (24) reported that the average of correlated extremely well with a from a central image in the scan range. The important finding from this study is that the value of the central image approaches the value of average of only for thorax examinations. In the case of head examinations, however, the value of the central image is always greater than the average of , with a percentage difference of . The values using slices produce a relatively small percentage difference ( and for thorax and head examinations, respectively), compared to for both thorax and head examinations.
There were several limitations in this study. First, the automated calculation was done to the patient images both for the full anatomy images and for the truncated anatomy images in the same way, without any correction for the latter images. In fact, the truncated anatomy images are missing part of the patient, and hence the calculation of is an underestimate, leading in turn to an overestimate of the SSDE. Truncation correction methods are very important to increase the accuracy of the measurement, especially in the thorax where the more peripheral distribution of attenuating tissue results in a proportionately larger truncation effect than in the more uniformly distributed abdominal tissue. (37) Second, this algorithm was only applied to images that were taken using a fixed tube current (FTC). Nevertheless, the automated contouring achieved good precision and accuracy for all slices, even though they had different noise levels.
V. CONCLUSIONS
We have proposed and tested a new algorithm for the automated calculation of water‐equivalent diameter (). The percentage differences of the automated results from the manual calculations were less than 0.5%, which is promising for the implementation of this method in the clinical environment. We found that the average value of water‐equivalent diameter from all slices () could be estimated accurately using slices only, distributed along the longitudinal axis (). Using nine slices, the percentage difference between and was about 1%.
ACKNOWLEDGMENTS
The authors are grateful for the funding of this work by the Research and Innovation Program, Bandung Institute of Technology (ITB), No. 237h/I1.C01/PL/2015 and No. 006n/I1.C01/ PL/2016. The authors would like to thank Mr. Masdi from Prof. Dr. Margono Hospital, Purwokerto, Indonesia.
COPYRIGHT
This work is licensed under a Creative Commons Attribution 3.0 Unported License.
REFERENCES
- 1. Brenner DJ, Elliston CD, Hal EJ, Berdon WE. Estimated risks of radiation induced fatal cancer from pediatric CT. Am J Roentgenol. 2000;176(2):289–96. [DOI] [PubMed] [Google Scholar]
- 2. Yu L, Liu W, Leng S et al. Radiation dose reduction in computed tomography: techniques and future perspective. Imaging Med. 2009;1(1):65–84. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Ruan C, Yukihara EG, Clouse WJ, Gasparian PB, Ahmad S. Determination of multislice computed tomography dose index (CTDI) using optically stimulated luminescence technology. Med Phys. 2010;37(7):3560–68. [DOI] [PubMed] [Google Scholar]
- 4. Kalender WA. Dose in x‐ray computed tomography. Phys Med Biol. 2014;59(3):R129–R150. [DOI] [PubMed] [Google Scholar]
- 5. International Atomic Energy Agency . Dose reduction in CT while maintaining diagnostic confidence: a feasibility/demonstration study. IAEA‐TECDOC‐1621. Vienna: IAEA; 2009. [Google Scholar]
- 6. International Atomic Energy Agency . Status of computed tomography: dosimetry for wide cone beam scanners. Human Health Reports No.5. Vienna: IAEA; 2011. [Google Scholar]
- 7. Slovis TL. CT and computed radiography: the picture are great, but is the radiation dose greater than required? Am J Roentgenol. 2002;179(1):39–41. [DOI] [PubMed] [Google Scholar]
- 8. Dawson P. Patient dose in multislice CT: why is it increasing and does it matter? Br J Radiol. 2004;77(Spec No. 1):S10–S13. [DOI] [PubMed] [Google Scholar]
- 9. Bauhs JA, Vrieze TJ, Primak AN, Bruesewitz MR, McCollough CH. CT dosimetry: comparison of measurement techniques and devices. RadioGraphics. 2008;28(1):245–53. [DOI] [PubMed] [Google Scholar]
- 10. Kim S, Song H, Samei E, Yin F, Yoshizumi TT. Computed tomography dose index and dose length product for cone‐beam CT: Monte Carlo simulations. J Appl Clin Med Phys. 2011;12(2):84–95. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Huda W and Mettler FA. Volume CT dose index and dose‐length product displayed during CT: what good are they? Radiology. 2011;258(1):236–42. [DOI] [PubMed] [Google Scholar]
- 12. International Atomic Energy Agency . Quality assurance programme for computed tomography: diagnostic and therapy applications. Human Health Series No. 19. Vienna: IAEA; 2012. [Google Scholar]
- 13. American Association of Physicists in Medicine . The measurement, reporting, and management of radiation dose in CT. Report No. 96. Report of the Task Group 23 of the Diagnostic Imaging Council CT Committee. College Park, MD: AAPM; 2008. [Google Scholar]
- 14. McCollough CH, Leng S, Yu L, Cody DD, Boone JM, McNitt‐Gray MF. CT dose index and patient dose: they are not the same thing. Radiology. 2011;259(2):311–16. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Brink JA and Morin RL. Size‐specific dose estimation for CT: how should it be used and what does it mean? Radiology. 2012;265(3):666–68. [DOI] [PubMed] [Google Scholar]
- 16. Dixon RL. A new look at CT dose measurement: beyond CTDI. Med Phys. 2003;30(6):1272–80. [DOI] [PubMed] [Google Scholar]
- 17. Mori S, Endo M, Nishizawa K et al. Enlarged longitudinal dose profiles in cone‐beam CT and the need for modified dosimetry. Med Phys. 2005;32(4):1061–69. [DOI] [PubMed] [Google Scholar]
- 18. Boone JM. The trouble with CTDI100. Med Phys. 2007;34(4):1364–71. [DOI] [PubMed] [Google Scholar]
- 19. Geleijns J, SalvadoArtell M, deBruin PW, Mather R, Muramatsu Y, McNitt‐Gray MF. Computed tomography dose assessment for a 160 mm wide, 320 detector row, cone beam CT scanner. Phys Med Biol. 2009;54(10):3141–59. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Huda W, Scalzetti EM, Roskopf M. Effective doses to patients undergoing thoracic computed tomography examinations. Med Phys. 2000;27(5):838–44. [DOI] [PubMed] [Google Scholar]
- 21. Li X, Samei E, Segars WP, Sturgeon GM, Colsher JG, Frush DP. Patient‐specific dose estimation for pediatric chest CT. Med Phys. 2008;35(12):5821–28. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Christner JA, Braun NN, Jacobsen MC, Carter RE, Kofler JM, McCollough CH. Size‐specific dose estimates for adult patients at CT of the torso. Radiology. 2012;265(3):841–47. [DOI] [PubMed] [Google Scholar]
- 23. Wang J, Duan X, Christner JA, Leng S, Yu L, McCollough CH. Attenuation‐based estimation of patient size for the purpose of size specific dose estimation in CT. Part I. Development and validation of methods using the CT image. Med Phys. 2012;39(11):6764–71. [DOI] [PubMed] [Google Scholar]
- 24. Leng S, Shiung M, Duan X, Yu L, Zhang Y, McCollough CH. Size specific dose estimation in abdominal CT: impact of longitudinal variations in patient size. Presented at the 55th AAPM annual meeting, Indianapolis, 2013.
- 25. American Association of Physicists in Medicine . Comprehensive methodology for the evaluation of radiation dose in x‐ray computed tomography. A new measurement paradigm based on a unified theory for axial, helical, fan‐beam, and cone‐beam scanning with or without longitudinal translation of the patient table. AAPM Report No. 111. Report of the AAMP Task Group 111: The Future of CT Dosimetry. College Park, MD: AAPM; 2010. [Google Scholar]
- 26. Dixon RL and Ballard AC. Experimental validation of a versatile system of CT dosimetry using a conventional ion chamber: beyond CTDI100. Med Phys. 2007;34(8):3399–413. [DOI] [PubMed] [Google Scholar]
- 27. Israel GM, Cicchiello L, Brink J, Huda W. Patient size and radiation exposure in thoracic, pelvic, and abdominal CT examinations performed with automatic exposure control. Am J Roentgenol. 2010;195(6)1342–46. [DOI] [PubMed] [Google Scholar]
- 28. Christianson O, Li X, Frush D, Samei E. Automated size‐specific CT dose monitoring program: assessing variability in CT dose. Med Phys. 2012;39(11):7131–39. [DOI] [PubMed] [Google Scholar]
- 29. American Association of Physicists in Medicine . Size‐specific dose estimates (SSDE) in pediatric and adult body CT examinations. Report No. 204. Report of the Task Group 204. College Park, MD: AAPM; 2011. [Google Scholar]
- 30. Pourjabbar S, Singh S, Padole A, Saini A, Blake MA, Kalra MK. Size‐specific dose estimates: localizer or transverse abdominal computed tomography images? World J Radiol. 2014;6(5):210–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Wang J, Christner JA, Duan Y, Leng S, Yu L, McCollough CH. Attenuation‐based estimation of patient size for the purpose of size specific dose estimation in CT. Part II. Implementation on abdomen and thorax phantoms using cross sectional CT images and scanned projection radiograph images. Med Phys. 2012;39(11):6772–78. [DOI] [PubMed] [Google Scholar]
- 32. Anam C, Haryanto F, Widita R, Arif I. Automated estimation of patient's size from 3D image of patient for size‐specific dose estimates (SSDE). Adv Sci Eng Med. 2015;7(10):892–96. [Google Scholar]
- 33. Bostani M, McMillan K, Lu P et al. Attenuation‐based size metric for estimating organ dose to patients undergoing tube current modulated CT exams. Med Phys. 2015;42(2):958–68. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. American Association of Physicists in Medicine . Use of water equivalent diameter for calculating patient size and size‐specific dose estimates (SSDE) in CT. Report No. 220. Report of the AAPM Task Group 220. College Park, MD: AAPM; 2014. [PMC free article] [PubMed] [Google Scholar]
- 35. Toth T, Ge Z, Daly MP. The influence of patient centering on CT dose and image noise. Med Phys. 2007;34(7):3093–101. [DOI] [PubMed] [Google Scholar]
- 36. Li B, Behrman RH, Norbash AM. Comparison of topogram‐based body size indices for CT dose consideration and scan protocol optimization. Med Phys. 2012;39(6):3456–65. [DOI] [PubMed] [Google Scholar]
- 37. Ikuta I, Warden GI, Andriole KP, Khorasani R, Sodickson A. Estimating patient dose from x‐ray tube output metrics: automated measurement of patient size from CT images enables large‐scale size‐specific dose estimates. Radiology. 2014;270(2):472–80. [DOI] [PMC free article] [PubMed] [Google Scholar]