Abstract
The aim of the present Monte Carlo study is to evaluate the variation of energy deposition in healthy tissues in the human eye which is irradiated by brachytherapy sources in comparison with the resultant dose increase in the gold nanoparticle (GNP)‐loaded choroidal melanoma. The effects of these nanoparticles on normal tissues are compared between and as two ophthalmic brachytherapy sources. Dose distribution in the tumor and healthy tissues has been taken into account for both brachytherapy sources. Also, in certain points of the eye, the ratio of the absorbed dose by the normal tissue in the presence of GNPs to the absorbed dose by the same point in the absence of GNPs has been calculated. In addition, differences of the absorbed dose in the tumor observed in the comparison of simple water phantom and actual simulated human eye in presence of GNPs are also a matter of interest that have been considered in the present work. The difference between the eye globe and the water phantom is more obvious for than that of the when the ophthalmic dosimetry is done in the presence of GNPs. Whenever these nanoparticles are utilized in enhancing the absorbed dose by the tumor, the use of brachytherapy source will greatly amplify the amount of dose enhancement factor (DEF) in the tumor site without inflicting much damage to healthy organs, when compared to the source. For instance, in the concentration of 30 mg GNPs, the difference amongst the calculated DEF for between these phantoms is 5.3%, while it is 2.45% for . Furthermore, in Monte Carlo studies of eye brachytherapy, more precise definition of the eye phantom instead of a water phantom will become increasingly important when we use as opposed to .
PACS number(s): 87.53.Jw, 87.85.Rs, 87.10.Rt
Keywords: brachytherapy, choroidal melanoma, MCNP5, gold nanoparticles, and
I. INTRODUCTION
Uveal melanoma is one of the primary ocular cancerous tumors which arises within the eyeball in the uvea involving the iris, ciliary body, or choroid. Although this kind of cancer is rare, it is the most common eye cancer in people who are middle‐aged or older. 1 , 2 , 3 , 4 Regarding the size and location of the tumor and also the rate of its progress, treatment of the choroidal melanoma is managed. Enucleation, local resection, and radiotherapy are the most common therapeutic processes for the treatment of choroidal melanoma. 5 , 6 , 7
Considering the sensitive tissues which the eye is involved in, ocular tumors present a therapeutic challenge. (8) Although the radiation is effective therapeutically, it can be harmful to healthy tissues. Plaque brachytherapy is the most widely used treatment, which aims to transfer the maximum amount of dose to the tumor while preventing dose absorption by normal tissue. 9 , 10 However, in the period of treatment both the tumor and the proximal normal tissues obey a certain pattern in radiation absorption; hence, in this therapeutic procedure minimizing the absorbed dose by the normal tissue is still one of the major concerns. Applying gold nanoparticles (GNPs) as radiation dose enhancers in combination with brachytherapy can be an effective method to reduce the radiation effects on healthy nearby tissues.
In the in vivo study by Hainfeld et al., 11 , 12 on the use of GNPs in mice which were irradiated by X‐ray photons, the results show that presence of GNPs in the tumor will cause more radiation dose in the cancerous cells than in those of the healthy tissues. In the in vitro study by Kong et al., (13) on the application of GNPs to enhance radiation cytotoxicity, the results show that radiotherapy killed more cancerous cells in the presence of GNPs than in the absence of these nanoparticles. Subsequently, GNP radiosensitization has been observed in more controlled in vitro irradiation of cells and plasmid DNA. 14 , 15 , 16
The idea of using gold nanoparticles in cancer therapy as a radiosensitizer is not a new one and several Monte Carlo in vitro and in vivo studies on the application of nanotechnology‐based cancer therapy have been performed. 17 , 18 , 19 , 20 However, only a handful of studies, such as our previous one, (21) have yet been conducted to investigate the effects of these nanoparticles on human eye tumors such as choroidal melanoma. If the choroidal melanoma tumor could be loaded with these nanoparticles in proper concentrations and dimensions, this would lead to a higher absorbed dose by the tumor during a shorter time. Since the eye is an extremely sensitive organ, the reduction of the period of treatment would decrease the absorbed dose by normal cells, resulting in a major reduction in the damage inflicted.
Regarding the radiosensitizing properties of gold nanoparticles, dose enhancement in the tumor is expected when the GNP‐loaded choroidal melanoma is locally irradiated with brachytherapy sources. However, considering the healthy tissues and the different seed models which are used in the eye plaque therapy, the variation of energy deposition in normal tissues in comparison with the resultant dose increase in the tumor by diverse sources is of outmost importance in the investigation of GNP effects on ophthalmic brachytherapy dosimetry. Some studies have been carried out through making a dosimetry comparison between and . 3 , 22 These studies reported higher absorbed dose by the tumor for versus for an equivalent radiation time. However, the focus of the present study is to know what changes will occur in absorbed dose by the healthy tissues compared to the dose increase in the tumor when the nanoparticle‐induced tumor is irradiated with and .
Here, a rigorous comparison between low‐energy photon sources and within COMS eye plaques was provided in the study of the effects of GNPs in ophthalmic brachytherapy. We also studied the difference between water phantom and human eye globe in the presence and absence of these nanoparticles for both the mentioned sources. The mean absorbed dose to the apex of the tumor, as well as to other critical points in both water and eye phantoms, were calculated in order to evaluate the effects of both sources and identify the most efficient source with the least amount of induced cell damage to critical points and healthy tissue in the presence of GNPs.
To this end, the eye globe was simulated precisely by considering different parts of the eye and their components. Here, both water and eye phantoms were simulated with MCNP5 code in which the water and eye phantoms were nearly identical owing to the fact that water was taken as the composition of the simulated eye globe, albeit by considering accurate geometry of the noted organ (Fig. 1).
The geometric characteristic of this phantom is in accordance with the eye of an adult. GNPs 50 nm in diameter were chosen, since they portray maximum uptake in mammalian cells. (23) Dosimetric characteristics of a single source for and were utilized to validate the accuracy of the Monte‐Carlo simulation technique. Fully‐loaded 16 mm COMS standard eye‐plaque with source (model 6711, GE Healthcare/Oncura, Arlington Heights, IL), and source (model 200, TheraSeed, Theragenics Corp., Buford, GA) were the focus of this study.
II. MATERIALS AND METHODS
The present Monte Carlo simulations are carried out using MCNP5 code. (24) The mentioned code benefits from a three‐dimensional heterogeneous geometry for both photons and electrons situated within the energy range of 1 KeV to 1 GeV; moreover, the libraries incorporated within MCNP5 are based upon the 8th release of ENDF/B‐VI. (25) tally following histories and F6 tally following histories were used in this study to perform the phantom dosimetry and the air kerma simulations respectively, in order to achieve a relative statistical error of less than 1%.
F6 tally, which estimates the deposited energy equivalent to the collisional kerma, is used to score kerma on terms of MeV/g per photon in the cells. The dose in air can be calculated directly using the F6 tally option of MCNP to calculate the air kerma rate. tally is determined based on full charged particle transport physics and it estimates the average deposited energy in units of MeV. The energy absorbed by a given cell must be divided by the mass of the cell in order to achieve the appropriate units: . It can be modified by a constant multiplier () to get dose values in units of Gy.
The current Monte Carlo study was conducted with two phantom test cases (water phantom and complete simulated human eye), in which the ophthalmic brachytherapy dosimetry was evaluated in both cases, in the presence and in the absence of gold nanoparticles using the full loaded eye plaque containing and sources. We simulated the human eye globe in a manner similar to that in our previous work. (21)
In order to define complete human eye globe, its components have been simulated using interface among different shapes with specific geometry and characteristics. Hence, all the details relating to the simulated phantoms are based upon the mentioned report. Furthermore, in addition to the 16 mm COMS eye plaque loaded with 13 seeds, a 16 mm eye plaque containing 13 seeds with the exact same coordinates was defined. The center of the simulated eye globe coincided with the origin.
Rows of cubic voxels with individual volumes of 0.05 cm3 were utilized to evaluate the depth dose of the and radionuclide sources within both water and eye phantoms. The mentioned voxels were assumed to be arranged on the central axis of the eye. An equator temporal eye melanoma with a height of 0.5 cm is considered in the present study. The COMS‐style eye plaque with the diameter of 16 mm was modeled on the equator temporal periphery to the eyeball.
It is a scientifically proven fact that the optimum GNP uptake within cancerous cells occurs when the noted substances are roughly 50 nm in diameter; (23) hence, the study benefits from the deployment of 50 nm GNPs within the tumor with varying concentrations of 7 mg, 10 mg, 18 mg, and 30 mg. Given the aforementioned fact, the tumor is latticed by identical cubes each with a volume of 0.1 cm3.
The dose enhancement factor (DEF) has been calculated to compare the effects of the presence of GNPs in brachytherapy for the noted configurations of sources and phantoms. DEF is the ratio of the absorbed dose by the tumor when it is loaded with GNPs to the absorbed dose by the tumor without these nanoparticles. Also, the dose distribution in different points of the eye has been determined for two mention sources in both eye and water phantoms. Validity of dosimetry computations for both the simulated sources were investigated by parameterized calculations of TG‐43 dosimetry parameters such as air kerma strength, dose rate constant, and radial dose function (RDF), which were then compared with the reported results of Thomson et al. and Rivard et al. 3 , 26 For RDF calculations each of the brachytherapy sources were simulated within a water phantom 30 cm3 in volume. Furthermore, for the calculations related to dose falloff, toroid tally cells (torus‐shaped cells) with major radii in the range of 0.5 cm to 10 cm were simulated.
III. RESULTS
A. Calculations of TG‐43 parameter
The calculated RDF in the present work, as reported by Thomson et al., (3) and by TG‐43 (26) are shown in Table 1 and Fig. 2. Also, the calculated dose rate constant in this work, as reported by Thomson et al., and by TG‐43 are shown in Table 2. The results show excellent agreement. It is also noteworthy to mention that all the referred TG‐43 dosimetry parameters were calculated for in our previous work. (21)
Table 1.
Radial Dose Function, g(r) | |||
---|---|---|---|
Distance from source (cm) | This work | Thomson et al. (3) | TG‐43 (26) |
0.05 | 0.251 | 0.249 | —– |
0.06 | 0.414 | 0.401 | —– |
0.07 | 0.580 | 0.561 | —– |
0.08 | 0.735 | 0.704 | —– |
0.09 | 0.864 | 0.826 | —– |
0.1 | 0.962 | 0.929 | 0.911 |
0.15 | 1.263 | 1.226 | 1.21 |
0.2 | 1.363 | 1.34 | —– |
0.25 | 1.383 | 1.381 | 1.37 |
0.3 | 1.381 | 1.39 | 1.38 |
0.4 | 1.332 | 1.365 | 1.36 |
0.5 | 1.292 | 1.31 | 1.3 |
0.6 | 1.213 | 1.245 | —– |
0.7 | 1.183 | 1.18 | —– |
0.75 | 1.149 | 1.148 | 1.15 |
0.8 | 1.106 | 1.113 | —– |
0.9 | 1.042 | 1.059 | —– |
1 | 1.001 | 1.001 | 1 |
1.5 | 0.763 | 0.742 | 0.749 |
2 | 0.576 | 0.552 | 0.555 |
2.5 | 0.426 | 0.407 | 0.41 |
3 | 0.315 | 0.298 | 0.302 |
3.5 | 0.231 | 0.219 | 0.223 |
4 | 0.170 | 0.16 | 0.163 |
4.5 | 0.124 | 0.117 | —– |
5 | 0.092 | 0.0865 | 0.0887 |
5.5 | 0.067 | 0.0635 | —– |
6 | 0.050 | 0.0469 | 0.0482 |
6.5 | 0.037 | 0.0346 | —– |
7 | 0.027 | 0.0256 | 0.0262 |
7.5 | 0.021 | 0.0193 | —– |
8 | 0.015 | 0.0147 | —– |
8.5 | 0.012 | 0.0112 | —– |
9 | 0.009 | 0.0084 | —– |
9.5 | 0.007 | 0.0064 | —– |
10 | 0.005 | 0.0051 | 0.00615 |
Table 2.
B. Dosimetry calculations
Figure 3 shows the depth‐dose curve in the plaque central axis direction for source in the fully‐loaded 16 mm COMS eye plaque in water phantom where the dose is presented relative to the dose at the tumor apex. The results have been compared with those reported by Thomson et al., (3) showing excellent agreement.
The dose to points of interest has been calculated in the water phantom and eye globe for both and sources and the results are reported in Table 3. Having yielded a relative statistical uncertainty less than 1% with the maximum percentage being apparent in the opposite side of the eye and the minimal amount of the noted factor observable in the sclera, the study has tabulated the full body of prescription points within the aforementioned table. The results indicate that in a healthy tissue such as the lens the difference in the calculated absorbed dose between water and eye phantoms in varying concentrations of GNPs is around 19% for the source and 14% for the source. The deviances for both phantoms in the presence of GNPs are greater when compared to those that were in the absence of the mentioned substance.
Table 3.
Water | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Location |
|
|
|
|
|
|
|
|
|
|
||||||||||
Sclera | 138.97 | 217.97 | 165.45 | 241.12 | 165.56 | 241.00 | 165.86 | 235.10 | 166.16 | 235.86 | ||||||||||
Apex | 41.03 | 64.31 | 79.78 | 104.18 | 95.22 | 123.00 | 138.29 | 169.24 | 201.48 | 235.67 | ||||||||||
Center of eye | 13.85 | 19.28 | 14.09 | 16.86 | 14.01 | 16.80 | 14.01 | 16.78 | 14.10 | 16.76 | ||||||||||
Opposite side | 2.91 | 2.62 | 3.22 | 2.12 | 3.20 | 2.13 | 3.20 | 2.37 | 3.24 | 2.48 | ||||||||||
Optic nerve | 4.99 | 5.05 | 5.07 | 4.59 | 5.07 | 4.59 | 5.07 | 4.74 | 5.08 | 4.93 | ||||||||||
Lens | 9.54 | 11.92 | 10.72 | 11.41 | 10.77 | 11.40 | 10.77 | 11.34 | 10.73 | 11.39 | ||||||||||
Macula | 6.97 | 7.68 | 7.72 | 7.48 | 7.71 | 7.48 | 7.71 | 7.52 | 7.77 | 7.47 | ||||||||||
Eye | ||||||||||||||||||||
Sclera | 174.89 | 263.62 | 184.90 | 265.00 | 185.18 | 259.47 | 185.53 | 265.26 | 185.81 | 265.95 | ||||||||||
Apex | 43.12 | 64.37 | 80.50 | 105.00 | 96.21 | 122.39 | 138.46 | 173.59 | 200.69 | 241.61 | ||||||||||
Center of eye | 14.76 | 18.82 | 13.46 | 16.20 | 13.48 | 16.25 | 13.47 | 16.67 | 13.53 | 16.66 | ||||||||||
Opposite side | 3.75 | 2.89 | 3.41 | 2.41 | 3.39 | 2.95 | 3.39 | 2.69 | 3.38 | 2.69 | ||||||||||
Optic nerve | 4.81 | 4.95 | 4.96 | 4.42 | 4.97 | 4.48 | 4.97 | 4.60 | 4.96 | 4.59 | ||||||||||
Lens | 8.55 | 10.23 | 8.74 | 9.71 | 8.65 | 9.71 | 8.65 | 9.73 | 8.62 | 9.73 | ||||||||||
Macula | 7.14 | 7.24 | 7.61 | 7.11 | 7.64 | 7.39 | 7.64 | 7.41 | 7.65 | 7.40 |
The dose enhancement factor (DEF) has been calculated for both and sources. This calculation has been done for different concentration of GNPs within the water phantom and compared with the calculated DEF for in Fig. 4. As is seen in this figure, the resultant DEF in the apex of the tumor for GNP concentrations of 30, 18, 10, and 7 mg/g are equal to 4.91, 3.37, 2.32, 1.94, respectively, for source, while being 3.66, 2.63, 1.91, and 1.62, respectively, for the source. The apparent deviances for the mentioned sources are greater for higher concentrations. A comparison between water phantom and eye globe in calculation of dose enhancement factor for is shown in Fig. 5. The difference between these phantoms is greater for higher concentrations. For instance, in the concentration of 30 mg GNPs, the calculated difference in the DEF at the apex of the tumor between water phantom and eye phantom is about 2.45%, while the noted deviance is about 0.7% in the concentration of 7 mg GNPs. The tally cells have been placed along the central axis of the plaque, starting from the sclera near the plaque to the sclera opposite side of the plaque. The points located on the transverse plane at distances of 0 to 0.5 cm indicate the center of voxels that were defined within the GNP‐loaded tumor. The points beyond 0.5 cm were in the healthy tissues. The DEF falloffs beyond 0.5 cm in both 4, 5 show the least effects of the presence of GNPs within the tumor on healthy tissues.
IV. DISCUSSION
Regarding the radiation dose enhancement properties of GNPs, studies have indicated that in radiotherapy of cancerous cells, the presence of these nanoparticles in the tumor site will locally increase the absorbed dose. Delivering the required tumor dose in a shorter period of time will result in lower doses to the normal tissues. In our previous study, (21) the dose enhancement properties of GNPs on choroidal melanoma for source were investigated. Given the differences in the properties of ophthalmic brachytherapy sources, the effects of the presence of GNPs in the tumor site in minimizing the damage done to normal tissue for each of these sources can be further investigated and compared. The results of this work have been quick to point out that in brachytherapy of choroidal melanoma, the source has granted us a greater DEF in the presence of GNPs when compared with the source. It is also noteworthy that the dosimetry calculations for the eye in the presence of GNPs report no notable differences for the absorbed dose in the healthy tissue in comparison with the cases where GNPs were absent for each of the noted sources. In reference to the noted evidence in brachytherapy of eye melanoma in presence of these particles, it is more efficient to utilize the source for it grants a higher DEF when compared with the source. The importance of defining an actual eye phantom instead of the water phantom in Monte Carlo studies has been another factor gaining much attention in recent years. The results signify that calculations of the resultant DEF in the tumor for the brachytherapy of eye melanoma utilizing a source show no striking difference when the actual eye model and the water phantom are compared. This accords with the previous study, (21) where the results pointed out that the use of the actual eye phantom in the Monte Carlo study of the dosimetry in eye melanoma with source, whilst GNPs were present, was of utmost importance. Furthermore, the results indicate that the deviances for both water and eye phantoms in the presence of GNPs are greater when compared to those that were in the absence of the mentioned substance. Regarding the numerous sources that are employed in the radiotherapy of the eye melanoma, the possible effects of these sources when accompanied by GNPs, regarding altering the period of treatment, is a matter of heated debate and requires more comprehensive investigations.
V. CONCLUSIONS
This work indicates that the presence of GNPs as a method of dose enhancement in treatment of eye tumors grants a higher DEF for the brachytherapy source as opposed to that of the source. Previous studies have been quick to point out that in the period of brachytherapy treatment of the eye (in the absence of GNPs), the source generates a greater dose in the tumor site. Since the results designate that the difference between absorbed dose in normal tissue, and in the presence of GNPs, is negligible when the two sources are compared, it is safe to say that the use of the source along with GNPs would yield a higher DEF in the tumor site of the eye melanoma. The previous study has also specified that ever‐more‐accurate definition of the actual eye phantom, instead of the water phantom, is an absolute necessity for precise dosimetry of the eye melanoma when examination utilizing the source is intended. However, even so, the dosimetry results obtained for the source in both the water and the eye phantoms are roughly similar; thus, in brachytherapy studies of the eye involving Monte Carlo methods, designation of a water phantom as an alternative for the eye phantom causes no observable difference in the results, and efficiently replicates those of the eye phantom.
COPYRIGHT
This work is licensed under a Creative Commons Attribution 4.0 International License.
REFERENCES
- 1. Patel M, Smyth E, Chapman PB, et al. Therapeutic implications of the emerging molecular biology of uveal melanoma. Clin Cancer Res. 2011;17(8):2087–100. [DOI] [PubMed] [Google Scholar]
- 2. Warren RM. Prognosis of malignant melanomas of the choroid and ciliary body. Cur Concepts Ophthalmol. 1974;4:158–67. [Google Scholar]
- 3. Thomson RM, Taylor RE, Rogers DW. Monte Carlo dosimetry for 125I and 103Pd eye plaque brachytherapy. Med Phys. 2008;35(12):5530–43. [DOI] [PubMed] [Google Scholar]
- 4. Nag S, Quivey JM, Earle JD, et al. The American Brachytherapy Society recommendations for brachytherapy of uveal melanomas. Int J Radiat Oncol Biol Phys. 2003;56(2):544–55. [DOI] [PubMed] [Google Scholar]
- 5. Damato B. Developments in the management of uveal melanoma. Clin Exp Ophthalmol. 2004;32(6):639–47. [DOI] [PubMed] [Google Scholar]
- 6. Augsburger JJ. Size classification of posterior uveal malignant melanomas. In: Yearbook of Ophthalmology, 1993. St. Louis: Mosby; 1993:155–65. [Google Scholar]
- 7. Simpson ER, Gallie B, Laperrierre N, et al. The American Brachytherapy Society consensus guidelines for plaque brachytherapy of uveal melanoma and retinoblastoma. Brachytherapy. 2014;13(1):1–14. [DOI] [PubMed] [Google Scholar]
- 8. Stannard C, Sauerwein W, Maree G, Lecuona K. Radiotherapy for ocular tumors. Eye. 2013;27(2):119–27. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Yoriyaz H, Sanchez A, dos Santos A. A new human eye model for ophthalmic brachytherapy dosimetry. Radiat Prot Dosimetry. 2005;115(1–4):316–19. [DOI] [PubMed] [Google Scholar]
- 10. Mourão AP and de Campos TP. Radiodosimetric considerations on ocular brachytherapy with iodine‐125 and ruthenium/rhodium‐106. Radiol Bras. 2009;42(1):43–48. [Google Scholar]
- 11. Hainfeld JF, Slatkin DN, Focella TM, Smilowitz HM. Gold nanoparticles: a new X‐ray contrast agent. Br J Radiol. 2006;79(939):248–53. [DOI] [PubMed] [Google Scholar]
- 12. Hainfeld JF, Slatkin DN, Smilowitz HM. The use of gold nanoparticles to enhance radiotherapy in mice. Phys Med Biol. 2004;49(18):N309–15. [DOI] [PubMed] [Google Scholar]
- 13. Kong T, Zeng J, Wang X, et al. Enhancement of radiation cytotoxicity in breast‐cancer cells by localized attachment of gold nanoparticles. Small. 2008;4(9):1537–43. [DOI] [PubMed] [Google Scholar]
- 14. Butterworth KT, Wyer JA, Brennan‐Fournet M, et al. Variation of strand break yield for plasmid DNA irradiated with high‐Z metal nanoparticles. Radiat Res. 2008;170(3):381–87. [DOI] [PubMed] [Google Scholar]
- 15. Foley EA, Carter JD, Shan F, Guo T. Enhanced relaxation of nanoparticle‐bound supercoiled DNA in X‐ray radiation. Chem Commun. 2005;(25):3192–94. [DOI] [PubMed] [Google Scholar]
- 16. Zheng Y, Hunting DJ, Ayotte P, Sanche L. Radiosensitization of DNA by gold nanoparticles irradiated with high‐energy electrons. Radiat Res. 2008;169(1):19–27. [DOI] [PubMed] [Google Scholar]
- 17. Cho SH. Estimation of tumor dose enhancement due to gold nanoparticles during typical radiation treatment: a preliminary Monte Carlo Study. Phys Med Biol. 2005;50(15):N163–73. [DOI] [PubMed] [Google Scholar]
- 18. Ngwa W, Korideck H, Kassis AI, et al. In vitro radiosensitization by gold nanoparticles during continuous low‐dose‐rate gamma irradiation with I‐125 brachytherapy seeds. Nanomedicine. 2013;9(1):25–27. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Chow JCL, Leung MKK, Fahey S, Chithrani DB, Jafray DA. Monte Carlo simulation on low‐energy electrons from gold nanoparticle in radiotherapy. J Phys Conf Ser. (2012);341(1):012012. [Google Scholar]
- 20. Huang X, Jain PK, El‐Sayed IH, El‐Sayed MA. Gold nanoparticles: interesting optical properties and recent applications in cancer diagnostic and therapy. Nanomedicine. 2007;2(5):681–93. [DOI] [PubMed] [Google Scholar]
- 21. Asadi S, Vaez‐Zadeh M, Masoudi F, Rahmani F, Knaup C, Meigooni AS. Gold nanoparticle‐based brachytherapy enhancement in choroidal melanoma using a full Monte Carlo model of the human eye. J Appl Clin Med Phys. 2015:16(5):1–13. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. American Brachytherapy Society — Ophthalmic Oncology Task Force . The American Brachytherapy Society consensus guidelines for plaque brachytherapy of uveal melanoma and retinoblastoma. Brachytherapy. 2014;13(1):1–14. [DOI] [PubMed] [Google Scholar]
- 23. Jain S, Hirst DG, O'Sullivan JM. Gold nanoparticles as novel agents for cancer therapy. Br J Radiol. 2012;85(1010):101–13. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. X‐5 Monte Carlo team . MCNP — a general Monte Carlo N‐particle transport code. Ver.5. Los Alamos (NM): Los Alamos National Laboratory; 2003. [Google Scholar]
- 25. Cross Section Evaluation Working Group. ENDF/B‐VI summary documentation. ENDF‐201. Brookhaven National Laboratory Report BNL‐NCS‐17541. 8th edition Upton, NY: National Nuclear Data Center; 2000. [Google Scholar]
- 26. Rivard MJ, Coursey BM, DeWerd LA, et al. Update of AAPM Task Group No.43 Report: a revised AAPM protocol for brachytherapy dose calculations. Med Phys. 2004;31(3):633–74. [DOI] [PubMed] [Google Scholar]
- 27. Taylor RE, Yegin G, Rogers DW. Benchmarking brachydose: voxel based EGSnrc Monte Carlo calculations of TG‐43 dosimetry parameters. Med Phys. 2007;34(2):445–57. [DOI] [PubMed] [Google Scholar]