Abstract
Background
The influences of potential bone-to-implant contact (BIC) area (pBICA), BIC area (BICA), and three dimensional (3D) BIC percentage (3D BIC%; defined as BICA divided by pBICA) in relation to the implant length on initial implant stability were studied. Correlations between these parameters were also evaluated.
Methods
Implants with lengths of 8.5, 10, 11.5, and 13 mm were placed in artificial bone specimens to measure three indexes of the initial implant stability: insertion torque value (ITV), Periotest value (PTV), and implant stability quotient (ISQ). The implants and bone specimens were also scanned by microcomputed tomography, and the obtained images were imported into Mimics software to reconstruct the 3D models and calculate the parameters of 3D bone-to-implant contact including pBICA, BICA, and 3D BIC%. The Kruskal-Wallis test, Wilcoxon rank-sum test with Bonferroni adjustment, and Spearman correlations were applied for statistical and correlation analyses.
Results
The implant length affected ITV more than PTV and ISQ, and significantly affected pBICA, BICA, and 3D BIC%. A longer implant increased pBICA and BICA but decreased 3D BIC%. The Spearman coefficients were high (>0.78) for the correlations between the three 3D BIC parameters and the three indexes of the initial implant stability.
Conclusions
pBICA, BICA, and 3D BIC% are useful when deciding on treatment plans related to various implant lengths, since these 3D BIC parameters are predictive of the initial implant stability.
Keywords: Microcomputed tomography, Implant length, 3D Bic%, Insertion torque, Periotest, Implant stability quotient
Background
The initial stabilization of a dental implant is considered to be a crucial factor influencing implantation success [1] and the development of osseointegration between the implant and bone [2]. Inadequate initial implant stability may allow micromovement between the implant and bone that results in the formation of fibrous tissue ingrowth into the interface instead of osseointegration [3]. Especially for patients with jaw bone of poor quality, the dimensions of the implant (length or diameter) are important factors in anchoring the implant strongly in bone [4–6] and hence preventing its initial mobility.
A shorter implant will generally have less contact with the surrounding bone, which could result in lower initial implant stability [6–8]. However, the results from studies of the relationship between initial implant stability and implant length have been inconsistent, with Hong et al. [6] finding that the initial implant stability was influenced by the implant length, Degidi et al. [9] indicating that only a weak correlation existed between initial implant stability and implant length, and Ostman et al. [10] reporting that a longer implant is associated with a lower initial implant stability.
While increasing the implant length would be expected to increase the surface area of an implant, the additional (deeper) portion of the implant length is only in contact with trabecular bone, which has a highly porous structure. This means that the actual increase in contact area between the implant and bone might be less than expected for a longer implant. Therefore, if clinicians want to increase the initial implant stability by embedding the longer implant in bone for patients, the measurement of actual contact area between the implant and bone seems to be necessary, especially for the three-dimensional (3D) bone-to-implant contact (BIC) [8, 11].
Cone-bean computer tomography (CBCT) is widely accepted as a useful tool in the diagnosis of diseases, injuries, and defects in oral and maxillofacial regions [12]. The advantages of CBCT over the conventional computed tomography include it involving smaller and cheaper hardware, producing higher image accuracy focused on a localized area, rapid scan time, dose reduction, and smaller image artifacts when using artifact suppression algorithms supplied by the device manufacturers [13]. CBCT has been increasingly applied in treatment planning for computer assisted maxillofacial surgery and dental implantation [14], such as for determining the available dimension (e.g., height and width) and quality of alveolar bone [15, 16], identifying anatomical structures (e.g., inferior alveolar nerve, mental foramen, and maxillary sinus) [17] and assessing bone grafting [18, 19]. Although CBCT is useful for dentists and oral surgeons, limitations in its resolution and in the functionality of software supplied with CBCT devices mean that 3D BIC still cannot be obtained for use in diagnoses nowadays. However, recent developments in micro-level CBCT images and professional medical imaging technology has allowed areas of 3D BIC and surfaces of implant site to be evaluated as parameters correlated with the initial stability of an implant [8, 11, 20, 21].
The objective of this study was to determine the impact of implant length on the initial stability of implants. In addition, high-resolution microcomputed tomography (micro-CT) images were used to calculate the following three types of 3D BIC parameter and evaluate how these parameters influence the initial implant stability: the potential bone-to-implant contact area (pBICA) [22], which is the total exterior surface area of the dental implant inside the artificial bone specimen; the actual area of contact between the bone and implant (BICA); and the 3D BIC percentage (3D BIC%), which was calculated as BICA divided by pBICA [8, 11, 20].
Methods
Preparation of artificial bone specimens and dental implants
Rigid cellular polyurethane blocks (Sawbones, Vashon, WA, USA) representing trabecular bone with an elastic modulus of 137 MPa (model 1522–12) were attached to a 2-mm-thick synthetic cortical shells (model 3401–01) with an elastic modulus of 16.7 GPa (Fig. 1). The structure of cellular polyurethane artificial bone is porous, which simulated the human trabecular bone structure. Commercial dental implants with the same diameter (4 mm; ICE, 3i Implant Innovation, Palm Beach, FL, USA) and four different lengths were used in this study: 8.5, 10, 11.5, and 13 mm (models XFOS 485, XFOS 410, XFOS 411, and XFOS 413, respectively) (Fig. 1). Twenty specimens divided into four types of experimental group were prepared for measurements of initial implant stability indexes and 3D BIC parameters.
Measurement of implant initial stability
Pilot holes were drilled into each artificial bone specimen in accordance with the manufacturer’s instructions. The peak ITV was measured for each specimen using a customized torque-rotation machine consisting of a digital torque meter (TQ-8800, Lutron Electronic Enterprise, Taipei, Taiwan), electronic rotation motor, and X-Y table (Fig. 2a). The force when inserting a dental implant into the artificial bone specimen was 14 N. The setting up and procedure used with the experimental machine mainly followed the standard of ASTM F543.
Furthermore, PTV was measured by connecting a temporary abutment (implant temporary hexed cylinder, 3i Implant Innovation) and then measuring the mobility of the implant using the Periotest device (Medizintechnik Gulden, Bensheim, Germany) (Fig. 2b). The tip of the Periotest device was placed perpendicular to the abutment at a distance of 2 mm, and it impacted the implant four times per second for 4 s. The Periotest™ device measures the contact time between the implant and the tapping tip as the single and that single were then transformed to a special value called PTV [23]. In general, a shorter contact time presents a lower value of PTV which indicates that the implant in bone is more stable. The attached microcomputer converted the duration obtained from the measurement cycle to PTV on a scale from −8 (very stable) to +50 (extremely unstable) [23].
Finally, the wireless resonance frequency analyzer Osstell ISQ™ (Osstell ISQ, Osstell AB, Gothenborg, Sweden) was used to measure the ISQ value (Fig. 2c). The smart peg for external hex connection of implant (Type 1, Osstell AB) was placed on the top of the implants. The peg has a magnetic material attached to its upper part. When the probe of the Osstell ISQ instrument is near to the smartpeg, the peg is vibrated by magnetic pulses and then Osstell ISQ instrument can detect the resonance frequency and converted it into a unique value called ISQ. The resonance frequency of the abutment–implant system was assigned a value between 0 and 100 to represent ISQ, where a larger ISQ indicates a higher stability.
Measuring 3D BIC parameters
The measurement approach of 3D BIC parameters were the same with our previous study. [24] As stated by Stoppie et al.’s study, [25] a thin layer of metal artefact around an metallic implant was existing between the inserted implant and bone in the micro-CT images. [25] In order to prevent the problem of artefact, in this study, the artificial bone specimens and dental implants were scanned separately by micro-CT (SkyScan 1076, Skyscan, Aartselaar, Belgium) at the same resolution (17.2 × 17.2 × 17.2 μm3) but with different scanning voltages and currents (89 kV and 112 μA for dental implants, 49 kV and 149 μA for artificial bones). The obtained micro-CT images of the artificial bones and dental implants were imported into Mimics 15.0 (Materialise, Leuven, Belgium) to create the 3D models of artificial bone and dental implants individually.
For the measurement approach of BICA which is the actual area of contact between the bone and implant, the 3D model of dental implant was placed into the 3D model of artificial bone (Fig. 3). Then the intersection area, which is BICA, between the 3D models of implant and artificial bone can be measured by using “Boolean operation” in Mimics.
For the measurement approach of pBICA which is the total exterior surface area of the dental implant within the artificial bone, that can also be calculated in Mimics. Finally, 3D BIC% can be calculated as BICA divided by pBICA. The three BIC parameters were measured four times in different positions for each dental implant.
Statistical analysis
The values of the three indexes of the primary initial stability (i.e., ITV, PTV, and ISQ) and the three types of 3D BIC parameter (i.e., pBICA, BICA, and 3D BIC%) were summarized as median and interquartile range (IQR) values. The Kruskal-Wallis test was used to compare how the three indexes of primary initial stability and the 3D BIC parameters differed among the implants with four diameters. In addition, post-hoc pairwise comparisons were conducted using the Wilcoxon rank-sum test with Bonferroni adjustment, with the cutoff for statistical significance set at 0.00833 (0.05/6). Spearman correlation coefficients (R 2; also called coefficients of determination) were used to evaluate significant correlations between the primary implant stability indexes and the 3D BIC parameters. All statistical analyses were performed using SPSS (version 19, IBM Corporation, Armonk, NY, USA).
Results
Implant length versus initial implant stability
The median, IQR, maximum, and minimum values of the ITV, PTV, and ISQ initial implant stability indexes are listed in Table 1. All three indexes were significantly influenced (P < 0.05) by the implant length. However, post-hoc pairwise comparisons revealed that the implant length affected ITV more than PTV and ISQ (Table 1). The R 2 values for the correlations between the implant length and the initial implant stability indexes were 0.82–0.97. The correlation was stronger (R 2 = 0.97) for ITV than for PTV and ISQ (Fig. 4a).
Table 1.
Length (mm) | ITV (N.cm) | PTV | ISQ | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Median* | IQR | Max | Min | Median* | IQR | Max | Min | Median* | IQR | Max | Min | |
8.5 | 32.70a | 1.8 | 34.8 | 31.2 | −3.2 a | 0.45 | −2.9 | −3.85 | 78.00 a | 0.5 | 80 | 75 |
10 | 39.60b | 5.3 | 43 | 35.2 | −4.2 ab | 0.9 | −3.4 | −4.8 | 82.00ab | 3 | 83 | 80 |
11.5 | 44.80bc | 11.1 | 52.9 | 39.3 | −4.5ac | 0.75 | −3.8 | −5.05 | 84.00b | 2 | 85 | 83 |
13 | 56.40c | 5 | 66.7 | 50.7 | −4.6bc | 0.35 | −4.35 | −5 | 85.00b | 1 | 85.5 | 83 |
P† | <0.001 | 0.019 | 0.002 |
†Kruskal-Wallis test
* Post hoc pairwise comparisons were conducted by exact Mann-Whitney U-Test with the Bonferroni adjustment; medians with the same letter a,b,c,d are not significantly different at the 0.00833 (0.05/6) level
Implant length versus 3D BIC parameters
Table 2 lists the median, IQR, maximum, and minimum values of the 3D BIC parameters, and indicates that pBICA, BICA, and 3D BIC% differed significantly (P < 0.05) with the implant length. According to the findings of post-hoc pairwise comparisons, the implant length affected pBICA and BICA more than 3D BIC% (Table 2). There were linear correlations between 3D BIC parameters and the implant length, all with R 2 values higher than 0.97 (Fig. 4b) but with differing slopes. A longer implant increased pBICA and BICA but decreased 3D BIC% (Fig. 4b).
Table 2.
Length (mm) | pBICA (mm2) | BICA (mm2) | 3D BIC% | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Median* | IQR | Max | Min | Median* | IQR | Max | Min | Median* | IQR | Max | Min | |
8.5 | 117.35 a | 0.06 | 117.39 | 117.31 | 67.97 a | 2.19 | 70.33 | 66.57 | 57.91 a | 1.87 | 59.93 | 56.76 |
10 | 139.62 b | 0.05 | 139.64 | 139.57 | 78.08 b | 2.25 | 79.35 | 76.31 | 55.91 a | 1.62 | 56.02 | 54.65 |
11.5 | 165.52 c | 0.08 | 165.56 | 165.47 | 89.10 c | 0.84 | 89.38 | 88.21 | 53.83 b | 0.49 | 54 | 53.31 |
13 | 183.48 d | 0.06 | 183.54 | 183.46 | 94.16 d | 2.22 | 96.13 | 92.62 | 51.31 c | 1.22 | 52.39 | 50.46 |
P† | <0.001 | <0.001 | 0.001 |
†Kruskal-Wallis test
* Post hoc pairwise comparisons were conducted by exact Mann-Whitney U-Test with the Bonferroni adjustment; medians with the same letter a,b,c,d are not significantly different at the 0.00833 (0.05/6) level
3D BIC parameters versus initial implant stability
Fig. 5 shows the correlations between 3D BIC parameters and indexes of the initial implant stability, all of which had R 2 values higher than 0.78. In general, the R 2 values were higher for the correlations of 3D BIC parameters with ITV and ISQ (all >0.89) than for those between 3D BIC parameters and PTV.
Discussion
It has long been considered that using implants that are as long as possible will maximize the initial implant stability [4], since this will increase the implant surface area and BIC. However, the additional length of the implant will only be in contact with trabecular bone, and so this will limit the real increase in the BIC surface area. The better way to confirm of how much BIC area will be added by embeding longer implant in bone would be to obtain quantitative 3D BIC data. Therefore, the present study combined a micro-CT system with professional medical image-processing software to develop 3D BIC parameters, and evaluated them in relationship to the initial implant stability of implants with various lengths.
The results of this study indicated that ITV is the index of the initial implant stability that seems to be most closely related to changes in implant length. A longer implant is considered to enhance ITV by providing a larger surface area of the implant in contact with trabecular bone, which might provide a greater mechanical resistance against the applied torque. Althoug insertion torque is a destructive measurement that can not be practised in clinic repeatedly, a high ITV indicates that the extent of micromotion between the implant and bone is reduced [26] and that the implant is mechanically more stable [27]. Not only implant length, but bone density [28]and cortical bone thickness [28] has been found by other studies could increase ITV.
The other indexes of the initial implant stability (PTV and ISQ) did not vary clearly with the implant length. These findings are consistent with Degidi et al. [9] reporting a weak correlation between ISQ and implant length, and implant length has lower influence on ISQ as compared to implant diameter. Even though there are different opinions regarding whether the implant length influences ISQ [6], those previous in vitro experiments used solid artificial polyurethane bone block to mimic the trabecular bone. Since it is known that trabecular bone has a porous structure unlike cortical bone, the use of solid specimens to mimic trabecular bone might exaggerate the contact area between implant and bone if longer implant was placed. In the present study, the use of porous trabecular bone specimens showed that there were only significant differences in ISQ values between implants with lengths of 8.5 and 13 mm and of 8.5 and 11.5 mm.
The 3D BIC parameter measurements indicated that pBICA was highest for the longest implant (13 mm long). This is expected since the value of pBICA is directly correlated with the area of the implant surface, and so a longer implant should exhibit a larger surface area of implant. Also, a longer implant also has a larger area of the implant surface in contact with porous trabecular bone, and that would also increase BICA.
However, it was also apparent that using a longer implant reduced 3D BIC%. This difference relative to the findings for BICA and pBICA might be related to 3D BIC% being defined as BICA divided by pBICA. Increasing the implant length only increases the area of the implant surface in contact with trabecular bone, which has a porous structure, and so the values of pBICA and BICA will not increase proportionally, resulting in the change in 3D BIC% differing from those in pBICA and BICA.
Linear regressions between 3D BIC parameters and initial implant stability indexes were evaluated in this study (Fig. 5). The Spearman coefficient was highest (R 2 > 0.98) for the correlation between 3D BIC% and ITV, which is consistent with Capparé et al. [7] finding that a significant linear correlation was existed between the two-dimensional (2D) BIC% and ITV. It appears that either 3D BIC% or 2D BIC% might be usefully assessed from the variation in ITV. For the linear regression models related to pBICA, the value of R 2 was high (>0.9) for the correlation between pBICA and ITV or ISQ, which is consistent with the findings of Hong et al. [6]. They found that ITV and ISQ were proportional to the total surface area of the implant fixture, with R 2 also higher than 0.9. Nevertheless, in our study R 2 was also high for the linear relationship between BICA and ITV (R 2 > 0.9) or ISQ (R 2 > 0.95). It seems that both the total implant surface area and the implant surface area in contact with bone may be good indexes for predicting the initial stability of a dental implant.
The main limitation of this study was the use of artificial foam bone specimens as testing samples, rather than real human bone. The use of artificial bone samples has some advantages, such as consistent porosities of the cancellous bone models as well as the material properties being similar to those of human bone, which avoids the differences between individual human subjects. However, the use of cadavers still needs to be considered in future studies. Another limitation was the smallness of the sample size, comprising only five specimens in each group. Future cadaveric studies might be considered to include more specimens in order to reduce errors caused by individual differences.
Conclusions
Within the limitations of this study, it is possible to draw the following conclusions. Firstly the implant length had a greater effect on ITV than on ISQ and PTV, and pBICA and BICA increased while 3D BIC% decreased for longer implants. The R 2 values in linear regression analyses were higher for the correlations between 3D BIC parameters and implant length than for those between indexes of the primary implant stability and the implant length. Finally, the goodness of fit (as reflected by R 2) was better for linear relationships between 3D BIC parameters and ITV or ISQ. That might reveal that all three 3D BIC parameters (i.e., pBICA, BICA and 3D BIC%) are reflected in variations in ITV and ISQ, which may be helpful in the treatment plans that include the possibility of selecting longer implants.
Acknowledgements
Not applicable.
Funding
This research was supported by the following government and academic organizations: Chang Gung Medical Foundation (CMRPG8E0301), China Medical University (CMU105-S-02), and Ministry of Science and Technology (MOST 105–2628-E-039-001-MY2) in Taiwan.
Availability of data and materials
The data that support the findings of this study are available from the first and corresponding authors upon reasonable request.
Abbreviations
- 3D BIC%
3D BIC percentage
- BIC
Bone-to-implant contact
- BICA
BIC area;
- CBCT
Cone-bean computer tomography
- IQR
Interquartile range
- ISQ
Implant stability quotient
- ITV
Insertion torque value
- micro-CT
Microcomputed tomography
- pBICA
Potential BIC area
- PTV
Periotest value
- R2
Correlation coefficients
Authors’ contributions
JH and HH designed the study, and analyzed and interpreted the data as well as also were major contributors in drafting article. AYW and LF contributed in providing the resources for the data measurement and the important information related to clinical relevance as well as interpretation of the results. All authors read and approved the final manuscript.
Ethics approval and consent to participate
Not applicable.
Consent for publication
Not applicable.
Competing interests
The authors declare that they have no competing interests.
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Contributor Information
Jui-Ting Hsu, Email: jthsu@mail.cmu.edu.tw.
Aaron Yu-Jen Wu, Email: aaronyujen.wu@gmail.com.
Lih-Jyh Fuh, Email: ljfuh@mail.cmu.edu.tw.
Heng-Li Huang, Phone: 1-886-4-22053366, Email: henleyh@gmail.com.
References
- 1.Scarano A, Degidi M, Iezzi G, Petrone G, Piattelli A. Correlation between implant stability quotient and bone-implant contact: a retrospective histological and Histomorphometrical study of seven titanium implants retrieved from humans. Clin Implant Dent Relat Res. 2006;8(4):218–222. doi: 10.1111/j.1708-8208.2006.00022.x. [DOI] [PubMed] [Google Scholar]
- 2.Mathieu V, Vayron R, Richard G, Lambert G, Naili S, Meningaud J-P, Haiat G. Biomechanical determinants of the stability of dental implants: influence of the bone–implant interface properties. J Biomech. 2014;47(1):3–13. doi: 10.1016/j.jbiomech.2013.09.021. [DOI] [PubMed] [Google Scholar]
- 3.Duyck J, Vandamme K, Geris L, Van Oosterwyck H, De Cooman M, Vandersloten J, Puers R, Naert I. The influence of micro-motion on the tissue differentiation around immediately loaded cylindrical turned titanium implants. Arch Oral Biol. 2006;51(1):1–9. doi: 10.1016/j.archoralbio.2005.04.003. [DOI] [PubMed] [Google Scholar]
- 4.Winkler S, Morris HF, Ochi S. Implant survival to 36 months as related to length and diameter. Ann Periodontol. 2000;5(1):22–31. doi: 10.1902/annals.2000.5.1.22. [DOI] [PubMed] [Google Scholar]
- 5.Olate S, Lyrio MCN, de Moraes M, Mazzonetto R, Moreira RWF. Influence of diameter and length of implant on early dental implant failure. J Oral Maxillofac Surg. 2010;68(2):414–419. doi: 10.1016/j.joms.2009.10.002. [DOI] [PubMed] [Google Scholar]
- 6.Hong J, Lim YJ, Park SO. Quantitative biomechanical analysis of the influence of the cortical bone and implant length on primary stability. Clin Oral Implants Res. 2012;23(10):1193–1197. doi: 10.1111/j.1600-0501.2011.02285.x. [DOI] [PubMed] [Google Scholar]
- 7.Capparé P, Vinci R, Di Stefano DA, Traini T, Pantaleo G, Gherlone EF, Gastaldi G. Correlation between initial BIC and the insertion torque/depth integral recorded with an instantaneous torque-measuring implant motor: an in vivo study. Clin Implant Dent Relat Res. 2015;17(2):e613–e620. doi: 10.1111/cid.12294. [DOI] [PubMed] [Google Scholar]
- 8.Huang H-L, Tsai M-T, Su K-C, Li Y-F, Hsu J-T, Chang C-H, Fuh L-J, Wu AY-J. Relation between initial implant stability quotient and bone-implant contact percentage: an in vitro model study. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2013;116(5):e356–e361. doi: 10.1016/j.oooo.2012.01.037. [DOI] [PubMed] [Google Scholar]
- 9.Degidi M, Daprile G, Piattelli A. Primary stability determination by means of insertion torque and RFA in a sample of 4,135 implants. Clin Implant Dent Relat Res. 2012;14(4):501–507. doi: 10.1111/j.1708-8208.2010.00302.x. [DOI] [PubMed] [Google Scholar]
- 10.Ostman P, Hellman M, Wendelhag I, Sennerby L. Resonance frequency analysis measurements of implants at placement surgery. Int J Prosthodont. 2005;19(1):77–83. [PubMed] [Google Scholar]
- 11.Hsu J-T, Huang H-L, Chang C-H, Tsai M-T, Hung W-C, Fuh L-J. Relationship of three-dimensional bone-to-implant contact to primary implant stability and peri-implant bone strain in immediate loading: microcomputed tomographic and in vitro analyses. Int J Oral Maxillofac Implants. 2013;28(2):367–374. doi: 10.11607/jomi.2407. [DOI] [PubMed] [Google Scholar]
- 12.De Vos W, Casselman J, Swennen G. Cone-beam computerized tomography (CBCT) imaging of the oral and maxillofacial region: a systematic review of the literature. Int J Oral Maxillofac Surg. 2009;38(6):609–625. doi: 10.1016/j.ijom.2009.02.028. [DOI] [PubMed] [Google Scholar]
- 13.Scarfe WC, Farman AG, Sukovic P. Clinical applications of cone-beam computed tomography in dental practice. J Can Dent Assoc. 2006;72(1):75–80. [PubMed] [Google Scholar]
- 14.Ito K, Gomi Y, Sato S, Arai Y, Shinoda K. Clinical application of a new compact CT system to assess 3-D images for the preoperative treatment planning of implants in the posterior mandible. Clin Oral Implants Res. 2001;12(5):539–542. doi: 10.1034/j.1600-0501.2001.120516.x. [DOI] [PubMed] [Google Scholar]
- 15.Sumer A, Caliskan A, Uzun C, Karoz T, Sumer M, Cankaya S. The evaluation of palatal bone thickness for implant insertion with cone beam computed tomography. Int J Oral Maxillofac Surg. 2016;45(2):216–220. doi: 10.1016/j.ijom.2015.09.012. [DOI] [PubMed] [Google Scholar]
- 16.Zhang W, Skrypczak A, Weltman R. Anterior maxilla alveolar ridge dimension and morphology measurement by cone beam computerized tomography (CBCT) for immediate implant treatment planning. BMC oral health. 2015;15(1):65. doi: 10.1186/s12903-015-0055-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Korkmaz Y, Kayıpmaz S, Senel F, Atasoy K, Gumrukcu Z. Does additional cone beam computed tomography decrease the risk of inferior alveolar nerve injury in high-risk cases undergoing third molar surgery? Does CBCT decrease the risk of IAN injury? Int J Oral Maxillofac Surg. 2017;46(5):628–635. doi: 10.1016/j.ijom.2017.01.001. [DOI] [PubMed] [Google Scholar]
- 18.Benavides E, Rios HF, Ganz SD, An C-H, Resnik R, Reardon GT, Feldman SJ, Mah JK, Hatcher D, Kim M-J. Use of cone beam computed tomography in implant dentistry: the international congress of oral Implantologists consensus report. Implant Dent. 2012;21(2):78–86. doi: 10.1097/ID.0b013e31824885b5. [DOI] [PubMed] [Google Scholar]
- 19.Kang Y-H, Kim H-M, Byun J-H, Kim U-K, Sung I-Y, Cho Y-C, Park B-W. Stability of simultaneously placed dental implants with autologous bone grafts harvested from the iliac crest or intraoral jaw bone. BMC oral health. 2015;15(1):172. doi: 10.1186/s12903-015-0156-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Hsu J-T, Huang H-L, Tsai M-T, Wu A-J, Tu M-G, Fuh L-J. Effects of the 3D bone-to-implant contact and bone stiffness on the initial stability of a dental implant: micro-CT and resonance frequency analyses. Int J Oral Maxillofac Surg. 2013;42(2):276–280. doi: 10.1016/j.ijom.2012.07.002. [DOI] [PubMed] [Google Scholar]
- 21.Möhlhenrich SC, Heussen N, Loberg C, Goloborodko E, Hölzle F, Modabber A. Three-dimensional evaluation of implant bed preparation and the influence on primary implant stability after using 2 different surgical techniques. J Oral Maxillofac Surg. 2015;73(9):1723–1732. doi: 10.1016/j.joms.2015.03.071. [DOI] [PubMed] [Google Scholar]
- 22.Schicho K, Kastner J, Klingesberger R, Seemann R, Enislidis G, Undt G, Wanschitz F, Figl M, Wagner A, Ewers R. Surface area analysis of dental implants using micro-computed tomography. Clin Oral Implants Res. 2007;18(4):459–464. doi: 10.1111/j.1600-0501.2007.01338.x. [DOI] [PubMed] [Google Scholar]
- 23.Atsumi M, Park S-H, Wang H-L. Methods used to assess implant stability: current status. Int J Oral Maxillofac Implants. 2007;22(5):743–754. [PubMed] [Google Scholar]
- 24.Hsu J-T, Shen Y-W, Kuo C-W, Wang R-T, Fuh L-J, Huang H-L. Impacts of 3D bone-to-implant contact and implant diameter on primary stability of dental implant. J Formos Med Assoc. 2017;116(8):582–590. doi: 10.1016/j.jfma.2017.05.005. [DOI] [PubMed] [Google Scholar]
- 25.Stoppie N, Waerden JP, Jansen JA, Duyck J, Wevers M, Naert IE. Validation of microfocus computed tomography in the evaluation of bone implant specimens. Clin Implant Dent Relat Res. 2005;7(2):87–94. doi: 10.1111/j.1708-8208.2005.tb00051.x. [DOI] [PubMed] [Google Scholar]
- 26.Trisi P, De Benedittis S, Perfetti G, Berardi D. Primary stability, insertion torque and bone density of cylindric implant ad modum Branemark: is there a relationship? An in vitro study. Clin Oral Implants Res. 2011;22(5):567–570. doi: 10.1111/j.1600-0501.2010.02036.x. [DOI] [PubMed] [Google Scholar]
- 27.Ito Y, Sato D, Yoneda S, Ito D, Kondo H, Kasugai S. Relevance of resonance frequency analysis to evaluate dental implant stability: simulation and histomorphometrical animal experiments. Clin Oral Implants Res. 2008;19(1):9–14. doi: 10.1111/j.1600-0501.2007.01419.x. [DOI] [PubMed] [Google Scholar]
- 28.Hsu JT, Fuh LJ, Tu MG, Li YF, Chen KT, Huang HL. The effects of cortical bone thickness and trabecular bone strength on noninvasive measures of the implant primary stability using synthetic bone models. Clin Implant Dent Relat Res. 2013;15(2):251–261. doi: 10.1111/j.1708-8208.2011.00349.x. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data that support the findings of this study are available from the first and corresponding authors upon reasonable request.