A novel plastic scintillator is developed for the application in the digital positron emission tomography (PET). The novelty of the concept lies in application of the 2-(4-styrylphenyl)benzoxazole as a wavelength shifter. The substance has not been used as scintillator dopant before. A dopant shifts the scintillation spectrum towards longer wavelengths making it more suitable for applications in scintillators of long strips geometry and light detection with digital silicon photomultipliers. These features open perspectives for the construction of the cost-effective and MRI-compatible PET scanner with the large field of view. In this article we present the synthesis method and characterize performance of the elaborated scintillator by determining its light emission spectrum, light emission efficiency, rising and decay time of the scintillation pulses and resulting timing resolution when applied in the positron emission tomography. The optimal concentration of the novel wavelength shifter was established by maximizing the light output and it was found to be 0.05 ‰ for cuboidal scintillator with dimensions of 14 mm x 14 mm x 20 mm.
Introduction
Plastic scintillators are commonly used as radiation detectors in particle and nuclear physics experiments. Low cost of manufacturing, possibility of forming in different shapes and opportunity to acquire relatively high light output renders plastic scintillators suitable for cost-effective construction of large size detectors [1]. Recently novel cost-effective method of the whole-body positron emission tomography was developed by the Jagiellonian Positron Emission Tomography Collaboration [2–5]. The main novelty of the method is the application of long strips of low-cost plastic scintillators instead of small crystals as detectors of annihilation gamma quanta [6–9]. To date, all commercial PET scanners use inorganic crystals as detectors of annihilation gamma quanta. Plastic scintillators were not applied because of their low probability of photoelectric effect and therefore lower detection probability compared to inorganic crystals [10, 11]. This disadvantage is compensated in the J-PET scanner by the large geometrical acceptance, superior timing properties, and utterly different method of determination of gamma quanta interaction place that is based on the time when scintillation pulse has been registered with photodetectors.
In the recent publications [2–4], it was shown that the PET based on plastic scintillators may even outperform the classical crystal PET also as regards sensitivity. TOF resolution for the detector of 30 cm length is better by about a factor of two with respect to the current TOF-PET tomographs characterized by typical field of views of about 16 cm. In the novel approach, developed by J-PET collaboration, timing of the photodetector signals is used instead of their amplitude information. To improve the sampling rates of the fast signals in the voltage domain, solely digital, front-end signal processing electronics was developed [12–14]. Compressing sensing [15, 16] and method based on the library of synchronized model signals [3, 17] are used for the hit-time and hit-position reconstruction. A single module of the J-PET detector consists of a scintillator strip read out at two ends by photomultipliers (left side of Fig 1) [2]. The gamma quantum hit position is determined from the difference of times of arrival of the light signals to the ends of the scintillator. The annihilation point (indicated as red dot) along the LOR (Line of Response) is calculated as , where c denotes a speed of light, and denote hit-times of gamma quanta in the upper and lower strip as indicated in the left side of Fig 1. Right side of Fig 1 illustrates one of the possible arrangements of scintillator strips in the J-PET tomograph. Such construction permits to enlarge the length of the field-of-view of the tomograph by increasing the length of the scintillator strips without rising of the cost of photomultipliers and electronics.
Fig 1. J-PET detection modules: A pair and a single layer.
(A) Left panel: scheme of the pair of J-PET detection modules. Such pair constitutes a simplest detection unit capable of registration both collinearly propagating quanta resulting from electron—positron annihilation. Times of light signals arrivals to photomultipliers are denoted as: —to left upper photomultiplier, —to right upper photomultiplier, —to left lower photomultiplier, —to right lower photomultiplier. Red dot indicates a place of annihilation, Δz denotes the distance between the point of interaction in scintillator and the center of the strip, ΔLOR denotes the position of annihilation along the line-of-response between two strips with respect to the centre of LOR. Both Δz and ΔLOR are determined based on the measured times , , , . (B) Right panel: schematic visualization of an example of a single detection layer of the J-PET detector. Each scintillator strip is aligned axially and read out at two ends by photomultipliers.
It permits also to combine PET and Magnetic Resonance Imaging (MRI) in a unique way [18, 19], complementary to presently applied PET/MRI solutions [20, 21] such that both metabolic and morphological images can be registered at the same place and at the same time. This is possible by inserting a non-magnetic plastic PET barrel inside a MRI tomograph. Such solution requires however usage of long strips of scintillators and silicon photomultipliers (SiPM) whose operation is not disturbed by the magnetic field. Application of the matrix readout of SiPM instead of vacuum tube photomultipliers would be also an advantage in terms of improvement of the time and spatial resolution. In the reference [4] it was shown that recording timestamps of several photons, at two ends of the scintillator strip, by means of SiPMs matrix may improve time and spatial resolution by more than factor of 1.5. The time resolution decreases however with the increase of the length of the scintillator strip due to the light attenuation. Bulk light attenuation length of plastic scintillators e.g. BC-420 (Saint Gobain) is claimed by the producer to be about 100 cm [22], however in practice the effective attenuation of light signals in strips with finite cross section is much lower. This is due to the fact that photons on a way from the emission point to the end of the scintillators undergoes many internal reflections and their optical length is much longer than the length measured along the scintillator. For example, for strips of BC-420 scintillator with cross section of 19 mm x 5 mm the attenuation length amounts to about 37 cm only [23].
In this article we describe results of investigations aiming at the development of a new scintillating material which would be characterized by the timing properties similar to the presently known plastic scintillators but whose photon emission spectrum would be extended towards longer wavelengths, thus providing lower light attenuation and more efficient matching to the quantum efficiency of the SiPM converters.
In the following sections we first describe method of manufacturing and composition of a novel material, further on referred to as J-PET scintillator. Subsequently optical properties and molecular weight of the J-PET scintillators will be discussed in view of its application for PET/MRI scanner. Next light output and characteristics of scintillation signals will be determined and compared to the properties of commercially available scintillators.
Materials and methods
Scintillator composition and the polymerization method
Plastic scintillators usually consist of three components: polymeric base, primary and secondary additives. The scintillation mechanism lies in energy transfer between those components. Firstly, polymer molecules are excited by the radiation e.g. gamma quanta. Next, this excitation energy is transferred to the primary additive via Förster mechanism and subsequently to the secondary additive—so called wavelength shifter—which de-excites via emission of photons in the wavelength range shifted towards longer values with respect to the emission spectrum of the primary additive [1].
Plastic scintillators discussed in this article were obtained via bulk polymerization of styrene or vinyltoluene (Sigma Aldrich, pure, about 50 ppm of polymerization inhibitor: 4-tert-buthylcatechol) [24, 25]. There are many ways of producing plastic scintillators from popular, widely available components like PPO, PTP or POPOP [26, 27].
We use 2-(4-styrylphenyl)benzoxazole as a wavelength shifter (WLS) [28]. This chemical compound is obtained in 3-step synthesis followed by the purification process by means of gel chromatography.
Scintillators described in this article, consist of polyvinyltoluene, 2 wt. % of 2,5-diphenyloxazole (Sigma Aldrich, 99%, suitable for scintillation) and various concentrations of 2-(4-styrylphenyl)benzoxazole. Chemical structures of the scintillator components are shown in Fig 2.
Fig 2. Components of the J-PET plastic scintillators.
The 2-(4-styrylphenyl)benzoxazole is a well known substance which was synthesized for the first time in 1960 by Adolf Emil Siegrist [29]. Since that time it was widely used in many fields e.g. as emitting material in organic light—emitting diodes (OLED) [30] and nonlinear optics (NLO) materials [31], however it was never used as a scintillator dopant so far.
The polymerization process was carried out in silylated glass ampoules. Scintillator dopants were dissolved in liquid purified monomer (styrene or vinyltoluene), next bulk polymerization reaction was initiated thermally. The detailed temperature schedule is shown in Fig 3. The maximal applied temperature reached 140°C and the whole process lasted 100 hours. Obtained scintillators are optically homogeneous. The main disadvantage is polymerization shrinkage covering about 20% of sample volume [32]. It may generate voids or cracks in the block of scintillator. However, when conducting the synthesis with the temperature schedule shown in Fig 3, we obtained pure and homogeneous scintillators characterized by high light output.
Fig 3. Thermal schedule for scintillator polymerization.
Experimental setup
Measurements of the J-PET scintillators light yield, light signals characteristics and time resolution achievable when applied to the positron emission tomography were carried out with the experimental setup presented in Fig 4. For the tests we used gamma quanta with energy of 511 keV originating from the positron-electron annihilation. 22Na radioactive isotope applied as a source of positrons was placed inside a lead collimator providing a well collimated beam of 511 keV gamma quanta with the spatial profile of about 1 mm [33]. The setup enabled to perform simultaneous measurements with the tested and reference scintillator. Coincident registration of signals in reference and tested detector and the 20 cm long slit of the collimator ensured reduction of the background from the 1.27 MeV quanta to the negligible level. For reference we used ternary plastic scintillator BC-420 [22]. The scintillators were cut to the rectangular shape with the dimensions of 14 x 14 x 20 mm3. The surfaces of the scintillators were polished. Such prepared samples were wrapped in the Vikuiti reflective foil [34] and placed between Hamamatsu R9800 photomultipliers as shown in Fig 4 [35]. The optical connection between scintillator and photomultiplier was provided by the EJ-550 optical gel [36].
Fig 4. Scheme of the experimental setup for the J-PET scintillators tests.
PM denotes photomultiplier. As oscilloscope a Serial Data Analyser (Lecroy SDA6000A) was used. It allowed to collect waveforms of four photomultipliers’ signals simultaneously with the sampling interval of 100 ps.
Results and discussion
In this section basic properties of the J-PET scintillators are presented and compared to properties of the state-of-the art plastic scintillator BC-420 which is used in the first J-PET scanner [22].
Emission spectra
Photo-induced emission spectrum of a thin sample of J-PET scintillator was measured with spectrophotometer FluoroLog-3 (Horiba Jobin-Yvon). Measurements were carried out in reflecting mode using R928P PMT and continuous wave xenon source.
The obtained emission spectrum of the J-PET scintillator is indicated as solid black line in Figs 5 and 6. One can notice that compared to the BC-420 scintillator (red dashed line) the J-PET scintillator emits photons in a significantly broader wavelength range. In particular with respect to the BC-420, the J-PET emission spectrum is extended towards longer wavelengths enabling more effective registration of the scintillation light when using SiPMs whose quantum efficiency is increasing (up to about 450 nm) with the increasing wavelength as indicated by pink dashed-dotted line in Fig 5.
Fig 5. Emission spectra of scintillators and quantum efficieny of photomultipliers.
Emission spectra of the J-PET (solid line) and BC-420 (dashed line) scintillators [22] superimposed on the quantum efficiency dependence on photons wavelength for typical vacuum tube photomultiplier with bialcali window (dotted line) [35] and silicon photomultipliers (dashed-dotted line) [35]. Maximum of emission for BC-420 scintillator is placed at wavelength of 393 nm while the maximum of the J-PET scintillator at 403 nm. The emission spectra are normalized in amplitude.
Fig 6. Emission spectra of BC-420, J-PET scintillator and scintillators’ absorption coefficient μeff [37, 38].
The emission spectra are normalized in amplitude.
Emission of photons with larger wavelength is beneficial also in view of the light attenuation. In Fig 6 emission spectra for the BC-420 and the J-PET scintillators are compared to the effective light absorption coefficient (blue dotted line) (μeff) determined for the BC-420 scintillator strips with rectangular cross section of 7 mm x 19 mm [37]. It was established based on the absorption coefficient of pure polystyrene [38] and accounting for losses due to the transport of photons through the scintillator bar [37]. The decrease of the value of absorption coefficient with the growth of the wavelength indicates that the scintillation light for the J-PET scintillator will be less attenuated (with attenuation factor proportional to exp(−μeff) with respect to the BC-420 since the J-PET scintillator spectrum is extended toward higher wavelengths.
Molecular weight
Temperature schedule applied for the synthesis of scintillator influences not only its homogeneity but also its molecular mass. In our case, pure polystyrene samples were prepared using temperature schedule shown in Fig 3, and their molecular weight measured by means of capillary viscometry is equal to about (2 × 105) u. As it is shown in article [39] light output is changing significantly with the polystyrene molecular weight below about 105 u, and it is constant above this value. Thus large molecular weight of the J-PET scintillator ensures the light output stability and repeatability.
Light yield
Light yield defined as a number of emitted photons per unit of deposited energy is one of the most important characteristics of scintillators. In case of scintillator introduced in this article, it was determined relative to the known light yield of the BC-420 material. For this purpose both: tested and reference scintillators were irradiated simultaneously with the 511 keV gamma quanta. Measurements were conducted by means of the experimental setup described in section Materials and Methods (Fig 4). Exemplary distributions of charge of registered signals are shown in Fig 7. Plastic scintillators are composed mainly of hydrogen and carbon. Due to the low atomic number of these elements, the 511 keV gamma quanta interacts with electrons of the material structure predominantly via Compton effect. Therefore the observed charge distribution is continuous and does not show photoelectric maximum typical in the case of the measurements with crystal scintillators. The charge of the registered signal is proportional to the number of scintillation photons which in turn is proportional to the energy transferred by the interacting gamma quantum to the electrons in the scintillator. For the fixed energy of 511 keV gamma quanta the maximum possible energy transfer is well defined and is equal to 341 keV. Thus by the comparison of the maximum charge of signals from the J-PET and BC-420 we may determine the light yield of J-PET scintillator relative to the light yield of BC-420.
Fig 7. Charge spectra measured with the J-PET and BC-420 scintillator.
PM1, PM2, PM3, PM4 used in the legend denote photomultipliers indicated in Fig 3. In the left side charge spectra of the J-PET scintillator (0.05J-PET or 0.5J-PET) measured with photomultipliers PM1 and PM2 are compared to the spectra of BC-420 scintillator measured with the same pair of photomultipliers. The number in the name of the J-PET scintillator indicates the concentration of WLS in ‰. In the right side spectra of the J-PET scintillators are compared to spectra of BC-420 measured with the same pair of photomultipliers: PM3 and PM4. The difference between the spectra for the low values of charge are due to the fact that PM1 and PM3 were used in the trigger. The differences at large charge (at the Compton edge) are due to the differences in the light yield.
The measurements were conducted for series of eight scintillators each with different mass fraction of the wavelength shifter 2-(4-styrylphenyl)benzoxazole varying from 0 ‰ to 0.5 ‰. The upper and lower panels of Fig 7 show examples of spectra obtained for the mass fractions of WLS equal to 0.05 ‰ and 0.5 ‰, respectively. In order to minimize instrumental uncertainties due to the possible miscalibration of photomultipliers we have performed tests for each sample twice, exchanging the J-PET and BC-420 scintillators in the setup (see Fig 4). In the left panel of Fig 7 the spectra for J-PET and BC-420 scintillators connected to PM1 and PM2 are shown while the right panel shows spectra registered when J-PET and BC-420 scintillator were connected to PM3 and PM4.
Spectra registered by each photomultiplier for BC-420 and the J-PET scintillator were compared to each other. The comparison was carried out using two methods.
The first method, described in details in reference [2], involves calculating the scaling factor by which the charge from BC-420 signals need to be multiplied that the J-PET and BC-420 charge spectra fit together. This scaling factor: LR = QJPET/QBC−420 is equal to the ratio of light outputs of these scintillators. We have determined the value of LR for each photomultiplier separately, and a mean value of obtained results is shown in the second column of Table 1.
Table 1. Relative and absolute light output of the J-PET scintillators as a function of the WLS concentration.
WLS concentration [‰] | LR (Method 1) | LR (Method 2) | Relative light output | Light output [photons/MeV] |
---|---|---|---|---|
0 | 0.636 ± 0.020 | 0.625 ± 0.002 | 0.625 ± 0.002 | 6397 ± 25 |
0.025 | 0.939 ± 0.008 | 0.938 ± 0.016 | 0.939 ± 0.007 | 9613 ± 72 |
0.05 | 1.009 ± 0.018 | 1.006 ± 0.021 | 1.008 ± 0.014 | 10318 ± 140 |
0.1 | 0.968 ± 0.015 | 0.967 ± 0.019 | 0.967 ± 0.019 | 9902 ± 200 |
0.2 | 0.965 ± 0.034 | 0.962 ± 0.031 | 0.962 ± 0.031 | 9853 ± 320 |
0.3 | 0.913 ± 0.017 | 0.912 ± 0.023 | 0.913 ± 0.014 | 9346 ± 140 |
0.4 | 0.910 ± 0.016 | 0.910 ± 0.035 | 0.910 ± 0.015 | 9317 ± 150 |
0.5 | 0.895 ± 0.023 | 0.897 ± 0.031 | 0.896 ± 0.018 | 9172 ± 190 |
In the second method we calculated LR as a ratio of the middles of the Compton edge of the charge spectra for the J-PET and BC-420 scintillators (centre of the right edges of spectra). The middle of the Compton edge was determined by fitting to the edge of the charge spectrum a Novosybirsk function [40]. The result obtained using this method is shown in the third column of Table 1.
Results obtained by both methods are statistically consistent within error bars. In Table 1 a total error including statistical and systematic uncertainties is shown.
Statistical uncertainties of fitting in both methods are of the order of one thousandth of the light output value. They are much smaller in comparison to the systematic uncertainties coming from differences caused by measurements with different photomultipliers. Differences resulting from the use of different fitting methods are also negligible comparing to uncertainties related to measurements with different photomultipliers.
Fourth column of Table 1 includes results obtained as weighted average of light output values determined by Method 1 and Method 2. In the fifth column absolute values of light output are given. They are calculated basing of results from the fourth column and light output of BC-420 scintillator equal to 10240 photons/MeV [22].
The final result is shown in Fig 8 which indicates that the light output of the J-PET scintillator is the largest for 2-(4-styrylphenyl)benzoxazole concentration of 0.05 ‰. It corresponds to 10318 ± 139 photons per MeV. This result is similar to the light output value of BC-420 scintillator which is equal to 10240 photons/MeV [22]. In case of lower and larger concentrations of WLS, light output gradually decreases. WLS concentrations lower than 0.05 ‰ are insufficient to enable the maximal efficiency. For the higher wavelength shifter concentrations the light yield decreases due to the quenching effect [41].
Fig 8. Relative and absolute light output of the J-PET plastic scintillators.
According to Adadurov et al. [41], optimal concentration of fluorescent additives in plastic scintillators is not a fixed value characterizing the material. Optimization of the scintillator composition has to be done for particular shape and size of the detector.
In the article [41], optimization of commercial wavelength shifter (POPOP) concentration was carried out, however shape and size of tested samples differ from scintillators described in this article, so it is impossible to compare quantitatively concentration and effectiveness of POPOP and our wavelength shifter. Nevertheless, effects observed in both experiments, like unsaturation or concentration quenching are similar.
The minimum light output of about 6400 photons/MeV is observed as expected for the scintillator made solely of polyvinyltoluene and primary fluor PPO.
Rise and decay times of light signals
As a main timing characteristics of the J-PET scintillator we determine the rise and decay time of the light signals and compare them to the corresponding characteristics of the BC-420 scintillator. The rise time values were determined calculating time differences between 10% and 90% of the signals amplitude on the leading edge: t10−90. Fig 9 shows shapes of signals registered by photomultipliers for the 0.05J-PET and BC-420 scintillators. All average signals were normalized to amplitude equal to 1 V.
Fig 9. Average signals measured with detectors consisting of the BC-420 or the J-PET scintillators and Hamamatsu R9800 photomultiplier.
The shown shapes result from averaging of 105 signals using method described in references [17]. Signals were normalized to the amplitude equal to 1V.
The shape of signals shown in Fig 9 is a convolution of the temporal density distribution of photons emitted by the scintillator, shape of the single photo-electron signal generated by photomultiplier, transit time spread (TTS) and the broadening of signal due to the finite bandwidth of the oscilloscope. Thus in general the rise time of the light signal (Tscintillator) may be extracted from the rise time of the measured electric signal (Tobserved) using the following formula:
(1) |
where Tphotomultiplier, Toscilloscope and TTS denote the contribution to the observed rise time due to the shape of the single photo-electron signal, oscilloscope bandwidth and transit time spread, respectively. Teff denotes the overall effective contributions due to the possible deviation of the nominal values of the above mentioned properties from the values provided by the manufacturers. Tphotomultiplier is equal to 1 ns [35], Toscilloscope was calculated as a ratio 350/bandwidth, and is equal to 0.07 ns while TTS value is 0.27 ns [35]. Mean rise time of the electric signals: t10−90 is equal to 1.22 ± 0.02 ns for BC-420 and 1.24 ± 0.02 ns for the J-PET scintillator, respectively. The rise time of the BC-420 scintillator is equal to 0.5 ns [22]. This implies that within the measurement uncertainties the rise time of the scintillation in the J-PET scintillator is equal to the rise time of the BC-420 scintillator and amounts to 0.50 ns.
Although the signal rise time of BC-420 and the J-PET scintillators are almost equal, as it is visible in Fig 8, there is a slight difference in the decay time in both kinds of scintillators. In order to estimate a decay time of the signal we compare the measured average signals with the theoretical formula accounting for the scintillator emission spectrum and the signal modifications due to the electronic circuit and photomultiplier. The influence of the electronics is clearly visible in Fig 8 as oscillations of the signals at the trailing edge. In ternary plastic scintillators the distribution of the time of the photon emission followed by the interaction of the gamma quantum at time Θ is given by the Formula: described in [42, 43].
(2) |
where the Gaussian term with the standard deviation σ reflects the rate of energy transfer to the primary solute, whereas tr and td denote the average time of the energy transfer to the wavelength shifter, and decay time of the final light emission, respectively. K stands for the normalization constant.
The influence of the electronics on the signal was parametrised effectively by the following formula:
(3) |
where N, α, φ, Δ and δ denote fitting parameters and fGauss denotes the Gauss function.
Averaged signals registered by one photomultiplier connected to BC-420 and J-PET scintillator with the fitted function being a sum of equations given by Formulas (2) and (3) are shown in Fig 10.
Fig 10. Averaged signals from photomultiplier connected to BC-420 scintillator (top) and to J-PET scintillator (bottom).
Red dashed line indicates result of fitting a function given by the sum of Formulas (2) and (3). Signals were normalized to the amplitude equal to 1V.
Decay time values determined for BC-420 and the J-PET scintillators are equal to 1.49 ± 0.02 ns for BC-420 and 1.91 ± 0.03 ns for the J-PET scintillator, respectively. The value determined for the BC-420 scintillator agrees well with the decay time given by the manufacturer, equal to 1.5 ns [22]. It is worth to mention that the determined decay time of the J-PET scintillator is comparable with the decay time of commercial scintillators, e.g. BC-404, widely used as fast counting detector, with decay time of 1.8 ns and BC-408 with decay time equal to 2.1 ns, respectively.
In Table 2 basic properties of the J-PET and BC scintillators are compared. Light output of J-PET scintillator is equal to light output of commercially available scintillators. Rise and decay time values of the novel scintillator described in this article, are similar to BC scintillators. Maximal wavelength of J-PET scintillator emission and the shape of emission spectra enables the use of the scintillators in novel PET/MRI scanner.
Table 2. Properties of J-PET scintillator and of exemplary plastic scintillators manufactured by Saint Gobain [22].
Properties | BC-420 | BC-404 | BC-408 | J-PET |
---|---|---|---|---|
Light output, [% Anthracene] | 64 | 68 | 64 | 64 |
Rise time [ns] | 0.5 | 0.7 | 0.9 | 0.5 |
Decay time [ns] | 1.5 | 1.8 | 2.1 | 1.9 |
Peak emission wavelength [nm] | 391 | 408 | 425 | 404 |
H:C ratio | 1.102 | 1.107 | 1.104 | 1.104 |
Conclusions
In this article properties of novel J-PET plastic scintillator made of polyvinyltoluene base, doped with 2,5-diphenyloxazole and 2-(4-styrylphenyl)benzoxazole as a wavelength shifter were presented. The purpose of the development was the elaboration of scintillator with optical properties allowing for more efficient transport of photons in long strip geometry, their more efficient registration with SiPM array with respect to the presently available plastic scintillators and, at the same time, with the superior timing characteristics. The novelty of the elaborated scintillator lies in the use of 2-(4-styrylphenyl)benzoxazole as a second additive—wavelength shifter. The substance has been used for the first time as a scintillator dopant. The new scintillator was manufactured via bulk polymerization of vinyltoluene and the optimal concentration of the 2-(4-styrylphenyl)benzoxazole was set by maximizing the light output for particular size and shape of the scintillator. It was shown that the maximum light output of the new scintillator is achieved when concentration of WLS is equal to 0.05 ‰. The light yield for the optimum 2-(4-styrylphenyl)benzoxazole concentration was established to be equal to about 104 photons per MeV. The rise- and decay-times of the developed scintillator were determined to be 0.5 ns and 1.9 ns, respectively. It was shown that when applied to the positron emission tomography it provides time resolution as obtainable with the BC-420 plastic scintillators. Moreover, the emission spectrum of the developed scintillator is extended towards longer wavelengths with respect to the typical commercial plastic scintillators as e.g. BC-420. It is therefore better matched to the quantum efficiency of silicon photomultipliers and lowers the light attenuation. These features of the developed scintillator make it suitable for the application in the construction of the PET insert for the simultaneous PET and MR imaging [18, 19].
Acknowledgments
We acknowledge technical and administrative support of A. Heczko, M. Kajetanowicz, W. Migdal, and the financial support by the Polish National Centre for Development and Research through grant LIDER/274/L-6/14/NCBR/2015, INNOTECH-K1/IN1/64/159174/NCBR/12, National Science Centre through grant No. 2016/21/B/ST2/01222, the Foundation for Polish Science through MPD programme, the EU and MSHE Grant No. POIG.02.03.00-161 00- 013/09 and Marian Smoluchowski Krakow Research Consortium “Matter-Energy-Future” and Polish Ministry of Science and Higher Education through grant: 7150/E-338/M/2015. BCH acknowledges gratefully the Austrian Science Fund (FWFP 26783).
Data Availability
All relevant data are within the paper.
Funding Statement
This work was supported by INNOTECH-K1/IN1/64/159174/NCBR/12 (http://www.ncbr.gov.pl/), LIDER/274/L-6/14/NCBR/2015 (http://www.ncbr.gov.pl/), POIG.02.03.00-161 00- 013/09 (http://www.fnp.org.pl/), 7150/E-338/M/2015 (http://www.nauka.gov.pl/), 2016/21/B/ST2/01222 (https://www.ncn.gov.pl/), the Foundation for Polish Science through MPD programme, Marian Smoluchowski Krakow Research Consortium “Matter-Energy-Future,” and the Austrian Science Fund (FWFP 26783). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1. Buvat I, Grupen C. Handbook of Particle Detection and Imaging. Berlin: Springer; 2012. [Google Scholar]
- 2. Moskal P, et al. Test of a single module of the J-PET scanner based on plastic scintillators. Nucl Instrum Meth. 2014;A764:317–321. doi: 10.1016/j.nima.2014.07.052 [Google Scholar]
- 3. Moskal P, et al. A novel method for the line-of-response and time-of-flight reconstruction in TOF-PET detectors based on a library of synchronized model signals. Nucl Instrum Meth. 2015;A775:54–62. doi: 10.1016/j.nima.2014.12.005 [Google Scholar]
- 4. Moskal P, et al. Time resolution of the plastic scintillator strips with matrix photomultiplier readout for J-PET tomograph. Phys Med Biol. 2016;61:2025 doi: 10.1088/0031-9155/61/5/2025 [DOI] [PubMed] [Google Scholar]
- 5. Smyrski J, et al. Measurement of gamma quantum interaction point in plastic scintillator with WLS strips. Nucl Instrum Meth. 2017;A851:39–42. doi: 10.1016/j.nima.2017.01.045 [Google Scholar]
- 6. Conti M. Focus on time-of-flight PET: the benefits of improved time resolution. European Journal of Nuclear Medicine and Molecular Imaging. 2011. June;38(6):1147–1157. Available from: http://dx.doi.org/10.1007/s00259-010-1711-y. [DOI] [PubMed] [Google Scholar]
- 7. Moses WW. Recent advances and future advances in time-of-flight PET. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 2007;580(2):919–924. Imaging 2006. Available from: http://www.sciencedirect.com/science/article/pii/S0168900207012909. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Conti M. State of the art and challenges of time-of-flight PET. Physica Medica. 2009;25(1):1–11. Available from: http://www.sciencedirect.com/science/article/pii/S1120179708000975. doi: 10.1016/j.ejmp.2008.10.001 [DOI] [PubMed] [Google Scholar]
- 9. Karp JS, Surti S, Daube-Witherspoon ME, Muehllehner G. The benefit of time-of-flight in PET imaging: Experimental and clinical results. Journal of Nuclear Medicine. 2008;49:462–470. doi: 10.2967/jnumed.107.044834 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Humm JL, Rosenfeld A, Del Guerra A. From PET detectors to PET scanners. European Journal of Nuclear Medicine and Molecular Imaging. 2003. November;30(11):1574–1597. Available from: https://doi.org/10.1007/s00259-003-1266-2. [DOI] [PubMed] [Google Scholar]
- 11. D T. Physical principles and technology of clinical PET imaging. European Journal of Nuclear Medicine and Molecular Imaging. 2004. March;33(2):133–145. [PubMed] [Google Scholar]
- 12. Paĺka M, Strzempek P, Korcyl G, Bednarski T, NiedĹşwiecki S, Biaas P, et al. Multichannel FPGA based MVT system for high precision time (20 ps RMS) and charge measurement. Journal of Instrumentation. 2017;12(08):P08001 Available from: http://stacks.iop.org/1748-0221/12/i=08/a=P08001. [Google Scholar]
- 13. Krzemień W, et al. Overview of the software architecture and data flow for the J-PET tomography device. Acta Phys Polon. 2016;B47:561. [Google Scholar]
- 14. Korcyl G, et al. Sampling FEE and Trigger-less DAQ for the J-PET Scanner. Acta Phys Polon. 2016;B47:491 doi: 10.5506/APhysPolB.47.491 [Google Scholar]
- 15. Raczyński L, et al. Novel method for hit-position reconstruction using voltage signals in plastic scintillators and its application to Positron Emission Tomography. Nucl Instrum Meth. 2014;A764:186–192. [Google Scholar]
- 16. Raczyński L, et al. Compressive Sensing of Signals Generated in Plastic Scintillators in a Novel J-PET Instrument. Nucl Instrum Meth. 2015;A786:105–112. [Google Scholar]
- 17. Moskal P, et al. Hit Time and Hit Position Reconstruction in the J-PET Detector Based on a Library of Averaged Model Signals. Acta Phys Polon. 2015;A127:1495–1499. doi: 10.12693/APhysPolA.127.1495 [Google Scholar]
- 18.Moskal P. A hybrid tof-pet/mri tomograph. Google Patents; 2016. US Patent App. 14/915,259. Available from: https://www.google.ch/patents/US20160209483.
- 19.Głowacz B, Moskal P, Zieliński M. Mobile tof-pet insert. Google Patents; 2017. WO Patent App. PCT/PL2016/000,082. Available from: https://encrypted.google.com/patents/WO2017010896A1?cl=sv.
- 20. Slomka PJ, Pan T, Germano G. Recent Advances and Future Progress in PET Instrumentation. Seminars in Nuclear Medicine. 2016;46(1):5–19. Recent Advances and Future Perspectives in Nuclear Medicine. Available from: http://www.sciencedirect.com/science/article/pii/S0001299815001075. doi: 10.1053/j.semnuclmed.2015.09.006 [DOI] [PubMed] [Google Scholar]
- 21. Vandenberghe S, Marsden PK. PET-MRI: a review of challenges and solutions in the development of integrated multimodality imaging. Physics in Medicine and Biology. 2015;60(4):R115 Available from: http://stacks.iop.org/0031-9155/60/i=4/a=R115. [DOI] [PubMed] [Google Scholar]
- 22.Saint-Gobain; 2017. Accessed: 20.06.2017. http://www/crystals.saint-gobain.com/.
- 23.Tomasz B. The energy calibration and time synchronization of the modular scintillator detector system for the Time-of-Flight Positron Emission Tomography, Doctoral Dissertation. Krakow: Jagiellonian University; 2016.
- 24.Kaplon Ĺ, et al. Plastic scintillators for positron emission tomography obtained by the bulk polymerization method. Bio-Algorithms and Med-Systems. 2014;10(1):27–31. [Google Scholar]
- 25. Wieczorek A, et al. A pilot study of the novel J-PET plastic scintillator with 2-(4-styrylphenyl)benzoxazole as a wavelength shifter. Acta Phys Polon. 2015;A127:1487–1490. doi: 10.12693/APhysPolA.127.1487 [Google Scholar]
- 26. B BJ. he theory and practice of scintillation counting. Oxford: Pergamon Press: 321–353; 1964. [Google Scholar]
- 27. Vasil’chenko VG, Lapshin VG, Peresypkin AI, Konstantinchenko AA, Pyshchev AI, Shershukov VM, et al. New results on radiation damage studies of plastic scintillators. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 1996;369(1):55–61. Available from: http://www.sciencedirect.com/science/article/pii/0168900295007741. [Google Scholar]
- 28.Wieczorek A, Danel A, Uchacz T, Moskal P. Use of 2-(4-styrylphenyl)benzoxazole and plastic scintillator. Google Patents; 2017. EP Patent App. EP20,150,744,711. Available from: https://encrypted.google.com/patents/EP3189523A1?cl=ja.
- 29. Siegrist AE. Uber eine neue Synthese zur darstellung heterocyclisch substituierter Stilbenverbindungen, die Anil-Sythese. Helvetica Chimica Acta. 1967;50(3):906–957. Available from: http://dx.doi.org/10.1002/hlca.19670500318. [Google Scholar]
- 30. Chung-Wen K, et al. Organic Light-Emitting Diodes Based on 2-(Stilben-4-yl)benzoxazole Derivatives: An Implication on the Emission Mechanism. Chem Mater. 2001;13(7):2441–2446. doi: 10.1021/cm010199u [Google Scholar]
- 31. Bobrovnikov YA, Vernigor EM, Zverev GM, Luk’yanets EA, Martynov AD, Khrolova OR. Effective transformation of the second harmonic of a ruby laser into stimulated emission in the range 400 470 nm in solutions of stilbenyloxazoles. Journal of Applied Spectroscopy. 1970. August;13:1009–1011. doi: 10.1007/BF00611129 [Google Scholar]
- 32.Britton EC, Le FWJ. Making polymerized vinyl aromatic articles. Google Patents; 1941. US Patent 2,256,160. Available from: https://www.google.com/patents/US2256160.
- 33. Kubicz E, et al. Beam profile investigation of the new collimator system for the J-PET detector. Acta Phys Polon. 2016;B47:537 doi: 10.5506/APhysPolB.47.537 [Google Scholar]
- 34.Vikuiti; 2017. Accessed: 20.06.2017. http://www.3M.com/Vikuiti/.
- 35.Hamamatsu; 2017. Accessed: 20.06.2017. http://www.hamamatsu.com.
- 36.Eljen Technology; 2017. Accessed: 20.06.2017. http://www.eljentechnology.com.
- 37. Kowalski P, et al. Multiple scattering and accidental coincidences in the J-PET detector simulated using GATE package. Acta Phys Polon. 2015;A127:1505–1512. doi: 10.12693/APhysPolA.127.1505 [Google Scholar]
- 38. Senchyshyn V, Lebedev V, Adadurov A, Budagov J, Chirikov-Zorin I. Accounting for self-absorption in calculation of light collection in plastic scintillators. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 2006;566(2):286–293. Available from: http://www.sciencedirect.com/science/article/pii/S0168900206011946. [Google Scholar]
- 39. Funt BL, Hetherington A. The influence of chain length on the luminescent output of plastic scintillators. The International Journal of Applied Radiation and Isotopes. 1959;4(3):189–195. Available from: http://www.sciencedirect.com/science/article/pii/0020708X59901929. [Google Scholar]
- 40. Ikeda H, Satpathy A, Ahn BS, Aulchenko VM, Bondar AE, Cheon BG, et al. A detailed test of the CsI(Ta) calorimeter for BELLE with photon beams of energy between 20 MeV and 5.4 GeV. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 2000;441(3):401–426. Available from: http://www.sciencedirect.com/science/article/pii/S0168900299009924. [Google Scholar]
- 41. Adadurov AF, Zhmurin PN, Lebedev VN, Titskaya VD. Optimizing concentration of shifter additive for plastic scintillators of different size. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 2009;599(2):167–170. Available from: http://www.sciencedirect.com/science/article/pii/S0168900208016343. [Google Scholar]
- 42. Moszyński M, Bengtson B. Light pulse shapes from plastic scintillators. Nuclear Instruments and Methods. 1977;142(3):417–434. Available from: http://www.sciencedirect.com/science/article/pii/0029554X77906760. [Google Scholar]
- 43. Moszyński M, Bengtson B. Status of timing with plastic scintillation detectors. Nuclear Instruments and Methods. 1979;158(Supplement C):1–31. Available from: http://www.sciencedirect.com/science/article/pii/S0029554X79901708. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All relevant data are within the paper.