. 2013 Sep 6;14(5):13–24. doi: 10.1120/jacmp.v14i5.3828

Evaluation of the sensitivity of two 3D diode array dosimetry systems to setup error for quality assurance (QA) of volumetric‐modulated arc therapy (VMAT)

Guangjun Li 1,2, Sen Bai 1,2, Nianyong Chen 1, Lansdale Henderson 3,, Kui Wu 1,2, Jianghong Xiao 1,2, Yingjie Zhang 1,2, Qingfeng Jiang 1,2, Xiaoqin Jiang 1,2
PMCID: PMC5714579  PMID: 24036856

Abstract

The purpose of this study is to evaluate the sensitivities of 3D diode arrays to setup error for patient‐specific quality assurance (QA) of volumetric‐modulated arc therapy (VMAT). Translational setup errors of ±1,±2, and ±3 mm in the RL, SI, and AP directions and rotational setup errors of ±1° and ±2° in the pitch, roll, and yaw directions were set up in two phantom systems, ArcCHECK and Delta4, with VMAT plans for 11 patients. Cone‐beam computed tomography (CBCT) followed by automatic correction using a HexaPOD 6D treatment couch ensured the position accuracy. Dose distributions of the two phantoms were compared in order to evaluate the agreement between calculated and measured values by using γ analysis with 3%/3 mm, 3%/2 mm, and 2%/2 mm criteria. To determine the impact on setup error for VMAT QA, we evaluated the sensitivity of results acquired by both 3D diode array systems to setup errors in translation and rotation. For the VMAT QA of all patients, the pass rate with the 3%/3 mm criteria exceeded 95% using either phantom. For setup errors of 3 mm and 2°, respectively, the pass rates with the 3%/3 mm criteria decreased by a maximum of 14.0% and 23.5% using ArcCHECK, and 14.4% and 5.0% using Delta4. Both systems are sensitive to setup error, and do not have mechanisms to account for setup errors in the software. The sensitivity of both VMAT QA systems was strongly dependent on the patient‐specific plan. The sensitivity of ArcCHECK to the rotational error was higher than that of Delta4. In order to achieve less than 3% mean pass rate reduction of VMAT plan QA with the 3%/3 mm criteria, a setup accuracy of 2 mm/1° and 2 mm/2° is required for ArcCheck and Delta4 devices, respectively. The cumulative effect of the combined 2 mm translational and 1° rotational errors caused 3.8% and 2.4% mean pass rates reduction with 3%/3 mm criteria, respectively, for ArcCHECK and Delta4 systems. For QA of VMAT plans for nasopharyngeal cancer (NPC) using the ArcCHECK system, the setup should be more accurate.

PACS numbers: 87.55.ne, 87.55.Qr, 87.55.km

Keywords: VMAT, setup error, patient‐specific QA, 3D diode array

I. INTRODUCTION

Volumetric‐modulated arc therapy (VMAT) is a new intensity‐modulated radiotherapy (IMRT) technology with single or multiple gantry arcs that achieves appropriate dose‐target conformity and permits critical organ sparing. VMAT delivers radiation via dynamic multileaf collimator (MLC) motion, and allows for variable dose rates, gantry speed modulation, and collimator rotation. (1) Thus far, VMAT has been used to treat various tumor sites, including head and neck, 2 , 3 , 4 , 5 lung, 5 , 6 , 7 , 8 prostate,( 4 , 5 , 9 , 10 , 11 , 12 ) rectum,( 5 , 13 ) cervix uteri, (14) spinal metastases,( 15 , 16 ) and brain metastases. (17) However, the dose calculation and the implementation of VMAT plans are highly complex. It is, therefore, essential to perform patient‐specific quality assurance (QA) of VMAT plans. (18)

Various types of dosimetry systems exist for dose verification, including gel dosimetry,( 19 , 20 ) water phantoms with film and ion chambers, (21) online 2D detector arrays, 22 , 23 , 24 , 25 Monte Carlo‐based frameworks, (26) and 3D diode arrays. 27 , 28 , 29 With the exception of the 3D diode arrays, these QA systems are limited either by single‐plane measurements of the dose distribution or increased off‐line processing time for measured data. (27) Two commercial 3D diode array systems, ArcCHECK (Sun Nuclear, Melbourne, FL) and Delta4 (ScandiDos AB, Uppsala, Sweden), were applied for dose verification in IMRT and VMAT.

In our clinical practice, we used ArcCHECK and Delta4 for QA of VMAT plans. During QA measurements, the phantoms were positioned such that the crosslines on the surface aligned with the room lasers. In practice, we noted that the registration errors (e.g., 2 mm) of the room lasers and the radiation isocenter of the linacs affected the phantom positioning error, and thereby obviously influenced the dose verification of VMAT. However, unlike most other commercial systems, the analysis software available in the ArcCHECK and Delta4 systems is unable to correct for positioning errors. Therefore, it is crucial to determine the sensitivity of the 3D detector arrays to setup error for QA of VMAT plans. In one case, Letourneau et al. (27) assessed the sensitivity of the prototype of the ArcCHECK dosimetry system to phantom translational setup error in the right‐left and anterior‐posterior directions and the results demonstrated that the diode array sensitivity to setup error is strongly dependent on the patient‐specific VMAT plans. However, the effect of the rotational setup error on ArcCHECK and other 3D detector arrays with various detector positions is not fully understood. In this study, we examined the sensitivities of ArcCHECK and Delta4 to translational and rotational setup errors in all directions for patient‐specific QA of VMAT plans.

II. MATERIALS AND METHODS

A. Patients' plan selection

Eleven patients requiring VMAT plans of differing complexity for cancers, including esophageal, prostate, cervix uteri, rectal, and nasopharyngeal cancer (NPC), were selected for this study. The VMAT plans were designed using a commercial 3D treatment planning system (Pinnacle v9.0, Philips Medical, Madison, WI) with a SmartArc optimization algorithm. (30) Patients' characteristics and planning states are summarized in Table 1. The plans for NPC had two full arcs with one control point per 4°, and the plans of other cancer sites had only one full arc with one control point per 4°. All TPS calculations in this study were done with a dose grid resolution of 2 mm.

Table 1.

Patient characteristics and planning states. Using the SIB technique, two or three dose levels are defined for each patient, except for patients with rectal cancer

Disease Site Number of Patients Dose Levels Monitor Units Average Field Width (mm) Average Leaf Travel (a) (mm) Average Leaf Travel Speed (b) (mm/s)
Esophageal 3 2
497±108
37±5
579±75
3.8±0.6
Prostate (c) 1 2 1316 51 848 5.4
Cervix uteri 1 2 906 49 818 5.0
Rectal 3 1
1259±136
82±8
610±57
4.2±0.3
NPC (d) 3 3
668±65
26±1
1418±121
4.5±0.3
a

Average leaf travel is calculated over all leaves, excluding leaves which remain closed over all treatment.

b

Average leaf travel speed is average leaf travel divided by delivery time.

c

The prostate plan is a whole pelvis and prostate boost plan.

d

The treatment volume for NPC included all sites: primary, upper, and lower neck with a single VMAT plan.

B. Measurement devices

Two 3D dosimetry systems, ArcCHECK and Delta4, were used for measurements. The ArcCHECK dosimetry system,( 27 , 31 ) consists of 1386 diodes, each with 0.8×0.8mm2 active measuring area, embedded in the cylindrical wall of the phantom. The Delta4 dosimetry system is based on two crossing arrays including 1069 diodes in a fixed cylindrical geometry, providing full coverage of the cross section for any beam direction.( 28 , 29 ) The spatial locations of the detectors are different between the two dosimetry systems. The dose distribution tested by ArcCHECK forms a cylindrical distribution with a diameter of 21 cm, typically positioned in the region surrounding the tumor target volume. In the Delta4 system, the dose distribution is measured on the two intersected perpendicular planes that cut through the tumor target volume.

C. Delivery and patient‐specific QA

The 11 VMAT plans were cast on the reference CT images of ArcCHECK and Delta4 phantoms and the dose distributions were recalculated. Both 3D diode arrays were placed on the HexaPOD 6D robotic treatment couch (Elekta, Crawley, UK) for measurements. Beam attenuation by the treatment couch has been considered in the plans by generating the couch's model from the contour of the couch, together with the density information. All tests were carried out using an Elekta Synergy accelerator at the nominal energy of 6 MV X‐rays with a 1 cm leaf width MLCi and an RTD 7.01 controller system (Elekta). Before recording QA measurements, we performed the quality control (QC) for the linac according to the TG142 report (32) ensuring the coincidence of lasers with the isocenter was within a radius of 1 mm. To minimize the setup error, cone‐beam computed tomography (CBCT) and a HexaPOD robotic treatment couch (HRTC) were used to set up the phantoms with 0.5 mm and 0.5° residual errors as prescribed in the studies by Sharpe et al. (33) and Meyer et al. (34) The reference CT images were acquired on a CT scanner with a slice thickness of 1 mm, and the resulting CBCTs possessed a voxel resolution of 0.5 mm in all three dimensions of the reconstructed images. Registration between the reference CT and CBCT was carried out automatically using an inbuilt method in XVI, namely gray value match.

D. Setup error simulation

The ArcCHECK and Delta4 phantoms were translated respectively in the right‐left (RL), anterior‐posterior (AP), and superior‐inferior (SI) directions by ±1,±2, and ±3 mm and rotated in the pitch, roll, and yaw directions by ±1° and ±2° using the 6D treatment couch. Figure 1 shows the definition of pitch, roll, and yaw used in this study. The 11 VMAT plans were separately delivered to the each of the two phantoms for dose verification; in total, 31 measurements (1 without positional error, 18 with translational errors, and 12 with rotational errors) were performed for each patient plan with one dosimetry system. For the combined reproducibility of setup and measurement, the above procedure was repeated three times for one rectal cancer case with the ArcCHECK and Delta4 systems, respectively. The intervals between the reproducibility tests were more than one month.

Figure 1.

Figure 1

Coordinate system used in the study. Arrows indicate positive rotation with respect to each axis.

We compared the measured dose distributions of each array with the calculated dose distributions generated by the planning system in order to analyze the effect of phantom‐specific setup errors on VMAT QA. We also analyze the cumulative effect of combined 2 mm translational and 1° rotational errors, which could be figured out by the quadratic summation method. (32) The pass rate of γ analysis was computed by comparing the calculated and measured dose distributions using 3%/3 mm, 3%/2 mm, and 2%/2 mm criteria, respectively. Diode readings, or “dose‐values,” lower than 10% of the highest diode signal were ignored in the analysis. These ignored readings reflect low‐dose and low‐gradient regions, typically located under the “jaws,” where the diode response is less reliable and the signal‐to‐noise ratio presents a concern. (35) The paired Student's t‐test was used for analysis of the results obtained from ArcCHECK and Delta4. All tests were two‐tailed with a p‐value of <0.05 considered the threshold for statistical significance. Statistical analysis was performed with the SPSS (v.14.0, Chicago, IL) program.

III. RESULTS

A. QA for VMAT patients' plans

QA results of the 11 VMAT plans tested with ArcCHECK and Delta4 are shown in Table 2. All pass rates of γ analysis with the 3%/3 mm criteria are higher than 95% for both diode arrays. Except for NPC, all pass rates of γ analysis with the 3%/2 mm and 2%/2 mm criteria are higher than 95% and 90%, respectively. The mean pass rate of γ analysis with the 2%/2 mm criteria for NPC by ArcCHECK was 84.7%. The lower results for NPC compared to the other cancer sites are due to the target volume complexity and the differences in the geometrical position of the diodes in ArcCHECK and Delta4 (see Fig. 2). A significant difference in the pass rate with the 3%/3mm criteria between the two systems (p=0.004) indicates that the pass rates of Delta4 are higher than those of ArcCHECK.

Table 2.

The QA results of 11 VMAT plans using ArcCHECK and Delta4 phantoms obtained without introducing any setup error

γ (%) a of ArcCHECK γ (%) a of Delta4
Cancer Site
3%/3mm
3%/2mm
2%/2mm
3%/3mm
3%/2mm
2%/2mm
Esophageal
98.5±0.3
96.8±0.4
91.2±0.7
99.7±0.3
98.2±1.0
95.3±1.7
Prostate 98.7 97.5 92.9 98.6 96.5 90.6
Cervix uteri 98.5 97.2 91.4 99.2 98.0 92.9
Rectal
98.8±0.9
97.2±1.8
93.6±2.4
99.4±0.4
97.8±1.2
95.8±1.0
NPC
95.6±0.8
92.5±0.9
84.7±1.0
98.5±0.5
95.3±1.9
90.1±2.0
a

Gamma (γ) results are the percentage of points passing the gamma criterion of 3%/3mm,3%/2mm, and 2%/2mm, respectively.

Figure 2.

Figure 2

The two systems measured the different section of the dose distribution because of the geometrical position of the diodes. ArcCHECK and Delta4 diodes positioned on the circular line and the crossline, respectively.

B. Sensitivities of two diode arrays to translational setup error

Figure 3 shows the impact of the translational setup error on 11 patient‐specific VMAT QA plans. Setup error was separately introduced in the RL, SI, and AP directions, and the impact was measured using ArcCHECK and Delta4. When the translational setup errors are ±1,±2, and ±3 mm, respectively, the pass rates of γ analysis with the 3%/3 mm criteria decreased by a maximum of 2.5%, 6.4%, and 14.0% for ArcCHECK and 2.5%, 6.9%, and 12.2% for Delta4 in the RL direction; 6.1%, 8.4%, and 13.4% for ArcCHECK and 1.6%, 6.3%, and 14.4% for Delta4 in the SI direction; 2.0%, 4.5%, and 9.5% for ArcCHECK and 1.7%, 5.1%, and 10.5% for Delta4 in the AP direction.

Figure 3.

Figure 3

The impact of translational setup errors on dosimetric verification of 11 VMAT plans using ArcCHECK and Delta4 phantoms in (a) RL, (b) SI, and (c) AP directions. The simulated translational setup errors are 1, 2, 3, ‐1, ‐2, and ‐3 mm, respectively. The decreased pass rates of γ analysis from the original results are assessed with the 3%/3 mm criteria.

To further test the difference between the two dosimetry systems in sensitivity to setup error, we compared all of their values for the reduction of γ analysis with the 3%/3 mm criteria in each direction. Significant differences in the pass rate of γ analysis in the RL and SI directions (p=0.019 and <0.001, respectively) indicate a higher sensitivity of ArcCHECK diodes than Delta4 diodes to translational setup error in both directions; however, only a nominal difference was observed in the AP direction between the two systems (p=0.074).

The tested results also indicated that the pass rate of γ analysis was most affected by translation in the RL and AP directions for NPC and esophageal cancer, but only affected by translation in the SI direction for prostate cancer. For ArcCHECK, the maximum standard deviations (error bars shown in Fig. 3) which were calculated for the three cases of each disease site for each setup error were 4.0%, 3.3%, and 3.1%, respectively, for NPC, esophageal, and rectal cancer; for Delta4 they were 3.8%, 4.5%, and 4.3%. These results indicate that the effects of translational setup errors on VMAT QA are strongly dependent on patient‐specific plans in spite of the same disease site.

C. Sensitivities of two diode arrays to rotational setup error

Figure 4 shows the impact of the rotational setup error for 11 patient‐specific VMAT QA. Setup error was separately introduced in the pitch, roll, and yaw directions, and the impact was measured using ArcCHECK and Delta4. When the rotational setup errors were ±1° and ±2°, respectively, the pass rates of γ analysis with the 3%/3 mm criteria decreased by a maximum of 5.5% and 9.9% for ArcCHECK and 2.5% and 5.0% for Delta4 in the pitch direction; 5.2% and 19.2% for ArcCHECK and 1.8% and 4.9% for Delta4 in the roll direction; and 8.4% and 23.5% for ArcCHECK and 1.7% and 4.9% for Delta4 in the yaw direction. Significant differences between the two systems in all rotation directions (p<0.001,=0.001, and <0.001 in the pitch, roll, and yaw directions, respectively), indicate that the configuration of the ArcCHECK system is more sensitive to rotational setup error than the Delta4 system when determining VMAT QA, and emphasize the importance of accurate rotational positioning during measurements using ArcCHECK.

Figure 4.

Figure 4

The impact of rotational setup errors on dosimetric verification of 11 VMAT plans using ArcCHECK and Delta4 phantoms in (a) pitch, (b) roll, and (c) yaw directions. The simulated rotational setup errors are 1°, 2°, ‐1°, and ‐2°, respectively. The decreased pass rates of γ analysis, compared to the original results, are assessed with the 3%/3 mm criteria.

From the results gathered by the two systems, we observed the greatest impact on the pass rate of γ analysis with the 3%/3 mm criteria in all directions for NPC and esophageal cancer. For ArcCHECK, the maximum standard deviations (error bars shown in Fig. 4) were 5.8%, 3.4%, and 2.5%, respectively, for NPC, esophageal, and rectal cancer; for Delta4 they were 1.1%, 2.2%, and 1.6%. These results indicate that the effects of the rotational setup errors on VMAT QA are strongly dependent on patient‐specific plans in spite of the same disease site.

D. Influence of setup error on the pass rate of γ analysis with various criteria

Table 3 shows the impact of setup errors in translation and rotation on the pass rate of γ analysis with various criteria attained by ArcCHECK and Delta4. Stricter gamma criteria resulted in a greater impact of the setup error on the pass rate of γ analysis. For a translational setup error of 3 mm, the pass rates of γ analysis with the 2%/2 mm criteria decreased by an average of 13.2% ±5.5% for ArcCHECK and by an average of 14.6%±6.7% for Delta4. For the rotational setup error of 2°, the pass rates of γ analysis with the 2%/2 mm criteria decreased by an average of 14.5%±6.6% for ArcCHECK and by an average of 7.0%±3.7% for Delta4.

Table 3.

Translational and rotational setup errors result in a decrease in the pass rate of γ analysis with various criteria on the average and standard deviation

ArcCHECK
Delta4
Setup Error
3%/3mm
3%/2mm
2%/2mm
3%/3mm
3%/2mm
2%/2mm
Translation
1 mm
0.7±1.0
1.4±1.5
2.0±2.0
0.6±0.6
1.6±1.5
2.1±2.2
2 mm
2.8±1.7
5.0±2.9
6.7±3.5
2.3±1.9
5.4±3.3
7.3±4.2
3 mm
6.9±3.3
10.0±4.9
13.2±5.5
6.1±3.8
11.1±5.5
14.6±6.7
Rotation
1.8±1.9
3.7±2.9
4.9±3.7
0.5±0.8
1.5±1.5
2.5±2.6
8.6±4.7
12.1±5.5
14.5±6.6
2.4±1.7
5.0±2.9
7.0±3.7

E. Combined reproducibility of setup and measurement

Figure 5 shows the standard deviation of pass rates with the 3%/3 mm criteria for the reproducibility tests for one rectal cancer case with the ArcCHECK and Delta4 systems, respectively. The mean standard deviations were 0.59% for ArcCHECK and 0.44% for Delta4 for all reproducibility tests in this case, and the mean standard deviations for 1, 2, and 3 mm translational setup errors, respectively, were 0.32%, 0.65%, and 0.83% for ArcCHECK and 0.12%, 0.41%, and 0.88% for Delta4; for 1° and 2° rotational setup errors, respectively, they were 0.41% and 0.75% for ArcCheck and 0.29% and 0.54% for Delta4.

Figure 5.

Figure 5

The standard deviation of pass rates with the 3%/3 mm criteria for the reproducibility tests for one rectal cancer case with the ArcCHECK and Delta4 systems, respectively. The tests contain all simulated (a) translational and (b) rotational setup errors for this case.

F. Cumulative effect of both translational and rotational error

Figure 6 shows the cumulative impact of 2 mm translational and 1° rotational setup errors for the 11 patient‐specific VMAT QA using ArcCHECK and Delta4, respectively. For ArcCHECK system, the average decreased pass rates of γ analysis with the 3%/3 mm criteria in all directions were 3.4%, 3.1%, 3.3%, 2.9%, and 5.6%, respectively, for esophageal, prostate, cervix uteri, rectal, and nasopharyngeal cancer, and for Delta4 the average decreased pass rates were 2.9%, 3.3%, 1.7%, 1.4%, and 3.0%, respectively.

Figure 6.

Figure 6

The cumulative impact of 2 mm translational and 1° rotational errors was evaluated for the pass rate reduction with 3%/3 mm criteria for both ArcCHECK and Delta4 systems. The calculated average and standard deviation of the pass rate reduction contained the translational and rotational errors of combinations in all directions.

IV. DISCUSSION

Patient‐specific dosimetric verification has been indispensable for IMRT QA. Furthermore, achieving accurate QA results is critical in detecting discrepancies between delivery and planning. The characteristics of Delta4, as reported by Korreman et al., (36) were determined by comparing consecutive deliveries of the same plan. The results indicated a strong agreement in all cases for the accumulated dose with dose deviations <1% for all measurement points and cases. Letourneau et al. (27) assessed the combined reproducibility of the ArcCHECK dosimeter system response and the linear accelerator (Elekta Synergy) for VMAT with the repeat delivery of the head and neck plans. The results demonstrated strong performance and stability of both systems. In our study, we tested the combined reproducibility of setup and measurement. The mean standard deviations were 0.59% (0.06%‐1.27%) for ArcCHECK and 0. 44% (0.06%‐1.06%) for Delta4, which shows good reproducibility. However, the results were slightly worse than the reports above because the pass rate altered more significantly as the setup error increased.

Dosimetric verification in the study indicated that the QA of VMAT assessed by both ArcCHECK and Delta4 met therapeutic quality requirements. However, it has been noted that the pass rate of γ analysis for Delta4 was higher than ArcCHECK, referring to the results for zero setup errors (p=0.004). The reasons for the differences are as follows. First, ArcCHECK and Delta4 dosimetry systems have very different spatial locations of diode detectors. Thus, each system measures a different section of the total dose distribution and samples different dose gradients (see Fig. 2). Compared to Delta4, dose distributions of ArcCHECK are more complex. Moreover, ArcCHECK is more sensitive to the dose delivery errors and the angular discretization effect because of the different diode locations. (37) Second, the dose error normalization was different. To establish the percent dose error normalization value, a different method was used for each device, given the different arrangement of the detectors. The Delta4 results were normalized at the isocenter, while the ArcCHECK results were normalized to the maximum measured dose in the detector ring. Moreover, the 10% dose cutoff threshold, below which the voxel is excluded from analysis, may not mean the same on the ArcCHECK diode surface as it does on the Delta4 diode planes. (37)

In all tests of the setup error simulation, the two 3D diode arrays exhibited extreme sensitivity to translational and rotational setup errors in all axes for patient‐specific QA results of VMAT plans, while exhibiting strong dependence on the patient‐specific plan. In general, we have observed an impact of the translational setup error on the QA results of complex VMAT plans and target volumes such as NPC, a cancer contained in the upper and lower neck regions. We have also observed a marked influence of the rotational setup error on the QA results of VMAT plans with long target volumes, such as esophageal cancer. In this paper, we tested three cases for each of the sites of NPC, esophageal, and rectal cancers. Despite the same cancer site amongst case triplets, a difference in sensitivity to setup errors was observed due to the variation of patient‐specific plans. Letourneau et al. (27) assessed the sensitivity of the prototype ArcCHECK dosimetry system for phantom setup error after CBCT image‐guided setup. Letourneau and colleagues found specifically that the diodes' sensitivity to setup error in the RL and AP directions were highly plan‐dependent; the direction of the steepest dose gradients for a given plan did not necessarily correspond with the direction of the phantom setup.

In addition, the results in Figs. 3 and 4 for both sets of setup errors indicate some differences in the order of 3%–4% reduction rates between the negative and positive setup errors. The differences are mainly due to the following two reasons. Firstly, the residual setup errors up to 0.5 mm and 0.5° were still present when CBCT and HRTC were used. Secondly, plan‐specific features along each translational and rotational setup directions were different, such as the dose gradient orientation.

On the other hand, the respective sensitivities to setup error of ArcCHECK and Delta4 were not uniform. Though the diode arrays demonstrated similar sensitivity to translational setup error, ArcCHECK was slightly more sensitive than Delta4 for the gamma criteria 3%/3 mm, likely measuring a section of the dose distribution with more dose gradients. In addition, compared to the translational setup error, ArcCHECK diodes were more sensitive to the rotational setup error than Delta4, due to the difference in spatial locations between the two 3D diode arrays. For the same rotational setup error, the diodes of the ArcCHECK system shift a greater distance than those of Delta4.

The AAPM TG‐142 report recommended that the tolerance of laser localization was 1.5 mm for IMRT. (32) For both ArcCHECK and Delta4 systems, 1° rotational error could cause an approximate error of 2 mm on the surface of the phantoms. Therefore, the cumulative effect of the combined 2 mm translational and 1° rotational errors was evaluated, and the average pass rates reduction with the 3%/3 mm criteria were 3.8% and 2.4% for ArcCHECK and Delta4 systems, respectively. The cumulative effect for ArcCHECK system was more obvious than Delta4 system mainly due to the higher sensitivity of ArcCHECK to the rotational error. Especially for NPC tests using ArcCHECK, the cumulative effect was quite obvious; thus, the setup of ArcCHECK should be more accurate than the other cancer sites. In addition, because of the difference in the sensitivity between the two systems, their setup accuracy could be different. As shown in Table 3, in order to achieve less than 3% mean pass rate reduction of VMAT plan QA with the 3%/3 mm criteria, a setup accuracy of 2 mm/1° and 2 mm/2° are required for ArcCheck and Delta4 devices, respectively.

V. CONCLUSIONS

In this study, both the ArcCHECK and Delta4 diode arrays showed high sensitivity to setup errors, and sensitivity of both systems is strongly dependent on patient‐specific plans. The sensitivity of ArcCHECK to the rotational error was higher than that of Delta4. In order to achieve less than 3% mean pass rate reduction of VMAT plan QA with the 3%/3 mm criteria, a setup accuracy of 2 mm/1° and 2 mm/2° is required for ArcCheck and Delta4 devices, respectively. The cumulative effect of the combined 2 mm translational and 1° rotational errors caused 3.8% and 2.4% mean pass rates reduction with 3%/3 mm criteria, respectively, for ArcCHECK and Delta4 systems. For QA of VMAT plans for NPC using the ArcCHECK system, the setup should be more accurate.

ACKNOWLEDGMENTS

Guangjun Li, S en Bai, and Nianyong Chen contributed equally to this work. The Delta4 dosimetry system was provided by Beijing HGPT Technology & Trade Co. Ltd. This work was partially supported by the National Natural Science Foundation of China (Grant No. 81101697).

Supporting information

Supplementary Material

REFERENCES

  • 1. Otto K. Volumetric modulated arc therapy: IMRT in a single arc. Med Phys. 2008;35(1):310–17. [DOI] [PubMed] [Google Scholar]
  • 2. Verbakel WF, Cuijpers JP, Hoffmans D, Bieker M, Slotman BJ, Senan S. Volumetric intensity‐modulated arc therapy vs. conventional IMRT in head‐and‐neck cancer: a comparative planning and dosimetric study. Int J Radiat Oncol Biol Phys. 2009;74(1):252–59. [DOI] [PubMed] [Google Scholar]
  • 3. Bertelsen A, Hansen CR, Johansen J, Brink C. Single arc volumetric modulated arc therapy of head and neck cancer. Radiother Oncol. 2010;95(2):142–48. [DOI] [PubMed] [Google Scholar]
  • 4. Yu C, Li X, Ma L, Chen D, et al. Clinical implementation of intensity‐modulated arc therapy. Int J Radiat Oncol Biol Phys. 2002;53(2):453–63. [DOI] [PubMed] [Google Scholar]
  • 5. Cao D, Holmes T, Afghan M, Shepard D. Comparison of plan quality provided by intensity‐modulated arc therapy and helical tomotherapy. Int J Radiat Oncol Biol Phys. 2007;69(1):240–50. [DOI] [PubMed] [Google Scholar]
  • 6. Bedford J, Nordmark Hansen V, McNair H, et al. Treatment of lung cancer using volumetric modulated arc therapy and image guidance: a case study. Acta Oncol. 2008;47(7):1438–43. [DOI] [PubMed] [Google Scholar]
  • 7. Verbakel WF, Senan S, Cuijpers JP, Slotman BJ, Lagerwaard FJ. Rapid delivery of stereotactic radiotherapy for peripheral lung tumors using volumetric intensity‐modulated arcs. Radiother Oncol. 2009;93(1):122–24. [DOI] [PubMed] [Google Scholar]
  • 8. McGrath SD, Matuszak MM, Yan D, Kestin LL, Martinez AA, Grills IS. Volumetric modulated arc therapy for delivery of hypofractionated stereotactic lung radiotherapy: a dosimetric and treatment efficiency analysis. Radiother Oncol. 2010;95(2):153–57. [DOI] [PubMed] [Google Scholar]
  • 9. Ma L, Yu C, Earl M, et al. Optimized intensity‐modulated arc therapy for prostate cancer treatment. Int J Cancer. 2001;96(6):379–84. [DOI] [PubMed] [Google Scholar]
  • 10. Palma D, Vollans E, James K, et al. Volumetric modulated arc therapy for delivery of prostate radiotherapy: comparison with intensity‐modulated radiotherapy and three‐dimensional conformal radiotherapy. Int J Radiat Oncol Biol Phys. 2008;72(4):996–1001. [DOI] [PubMed] [Google Scholar]
  • 11. Wolff D, Stieler F, Welzel G, et al. Volumetric modulated arc therapy (VMAT) vs. serial tomotherapy, step‐and‐shoot IMRT and 3‐D‐conformal RT for treatment of prostate cancer. Radiother Oncol. 2009;93(2):226–33. [DOI] [PubMed] [Google Scholar]
  • 12. Guckenberger M, Richter A, Krieger T, Wilbert J, Baier K, Flentje M. Is a single arc sufficient in volumetric‐modulated arc therapy (VMAT) for complex‐shaped target volumes? Radiother Oncol. 2009;93(2):259–65. [DOI] [PubMed] [Google Scholar]
  • 13. Duthoy W, De Gersem W, Vergote K, et al. Clinical implementation of intensity‐modulated arc therapy (IMAT) for rectal cancer. Int J Radiat Oncol Biol Phys. 2004;60(3):794–806. [DOI] [PubMed] [Google Scholar]
  • 14. Cozzi L, Dinshaw KA, Shrivastava SK, et al. A treatment planning study comparing volumetric arc modulation with Rapid Arc and fixed field IMRT for cervix uteri radiotherapy. Radiother Oncol. 2008;89(2):180–91. [DOI] [PubMed] [Google Scholar]
  • 15. Matuszak M, Sui H, Yan D. Potential impact of volumetric modulated arc therapy on the planning and delivery of radiation therapy. Int J Radiat Oncol Biol Phys. 2008;72(1 Suppl):S651. [Google Scholar]
  • 16. Kuijper IT, Dahele M, Senan S, Verbakel W. Volumetric modulated arc therapy versus conventional intensity modulated radiation therapy for stereotactic spine radiotherapy: a planning study and early clinical data. Radiother Oncol. 2010;94(2):224–28. [DOI] [PubMed] [Google Scholar]
  • 17. Clark GM, Popple RA, Young PE, Fiveash JB. Feasibility of single‐isocenter volumetric modulated arc radiosurgery for treatment of multiple brain metastases. Int J Radiat Oncol Biol Phys. 2010;76(1):296–302. [DOI] [PubMed] [Google Scholar]
  • 18. Bortfeld T, Webb S. Single‐arc IMRT? Phys Med Biol. 2009;54(1):N9–N20. [DOI] [PubMed] [Google Scholar]
  • 19. Low D, Dempsey J, Venkatesan R, et al. Evaluation of polymer gels and MRI as a 3‐D dosimeter for intensity‐modulated radiation therapy. Med Phys. 1999;26(8):1542–51. [DOI] [PubMed] [Google Scholar]
  • 20. Vergote K, De Deene Y, Duthoy W, et al. Validation and application of polymer gel dosimetry for the dose verification of an intensity‐modulated arc therapy (IMAT) treatment. Phys Med Biol. 2004;49(2):287–305. [DOI] [PubMed] [Google Scholar]
  • 21. Pallotta S, Marrazzo L, Bucciolini M. Design and implementation of a water phantom for IMRT, arc therapy, and tomotherapy dose distribution measurements. Med Phys. 2007;34(10):3724–31. [DOI] [PubMed] [Google Scholar]
  • 22. Jursinic P and Nelms B. A 2‐D diode array and analysis software for verification of intensity modulated radiation therapy delivery. Med Phys. 2003;30(5):870–79. [DOI] [PubMed] [Google Scholar]
  • 23. Spezi E, Angelini A, Romani F, Ferri A. Characterization of a 2D ion chamber array for the verification of radiotherapy treatments. Phys Med Biol. 2005;50(14):3361–73. [DOI] [PubMed] [Google Scholar]
  • 24. Poppe B, Blechschmidt A, Djouguela A, et al. Two‐dimensional ionization chamber arrays for IMRT plan verification. Med Phys. 2006;33(4):1005–15. [DOI] [PubMed] [Google Scholar]
  • 25. Jursinic PA, Sharma R, Reuter J. MapCHECK used for rotational IMRT measurements: step‐and‐shoot, Tomo Therapy, RapidArc. Med Phys. 2010;37(6):2837–46. [DOI] [PubMed] [Google Scholar]
  • 26. Bush K, Townson R, Zavgorodni S. Monte Carlo simulation of RapidArc radiotherapy delivery. Phys Med Biol. 2008;53(19):N359–N370. [DOI] [PubMed] [Google Scholar]
  • 27. Letourneau D, Publicover J, Kozelka J, Moseley DJ, Jaffray DA. Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy. Med Phys. 2009;36(5):1813–21. [DOI] [PubMed] [Google Scholar]
  • 28. Bedford J, Lee Y, Wai P, South C, Warrington A. Evaluation of the Delta4 phantom for IMRT and VMAT verification. Phys Med Biol. 2009;54(9):N167–N176. [DOI] [PubMed] [Google Scholar]
  • 29. Sadagopan R, Bencomo J, Martin R, Nilsson G, Matzen T, Balter P. Characterization and clinical evaluation of a novel IMRT quality assurance system. J Appl Clin Med Phys. 2009;10(2):104–19. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Bzdusek K, Friberger H, Eriksson K, Hardemark B, Robinson D, Kaus M. Development and evaluation of an efficient approach to volumetric arc therapy planning. Med Phys. 2009;36(6):2328–39. [DOI] [PubMed] [Google Scholar]
  • 31. Yan G, Lu B, Kozelka J, Liu C, Li JG. Calibration of a novel four‐dimensional diode array. Med Phys. 2010;37(1):108–15. [DOI] [PubMed] [Google Scholar]
  • 32. Klein EE, Hanley J, Bayouth J, et al. Task Group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36(9):4197–212. [DOI] [PubMed] [Google Scholar]
  • 33. Sharpe MB, Moseley DJ, Purdie TG, Islam M, Siewerdsen J, Jaffray D. The stability of mechanical calibration for a kV cone beam computed tomography system integrated with linear accelerator. Med Phys. 2006;33(1):136–44. [DOI] [PubMed] [Google Scholar]
  • 34. Meyer J, Wilbert J, Baier K, et al. Positioning accuracy of cone‐beam computed tomography in combination with a HexaPOD robot treatment table. Int J Radiat Oncol Biol Phys. 2007;67(4):1220–28. [DOI] [PubMed] [Google Scholar]
  • 35. Basran PS and Woo MK. An analysis of tolerance levels in IMRT quality assurance procedures. Med Phys. 2008;35(6):2300–07. [DOI] [PubMed] [Google Scholar]
  • 36. Korreman S, Medin J, Kjaer‐Kristoffersen F. Dosimetric verification of RapidArc treatment delivery. Acta Oncol. 2009;48(2):185–91. [DOI] [PubMed] [Google Scholar]
  • 37. Feygelman V, Zhang G, Stevens C, Nelms BE. Evaluation of a new VMAT QA device, or the “X” and “O” array geometries. J Appl Clin Med Phys. 2011;12(2):146–68. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Material


Articles from Journal of Applied Clinical Medical Physics are provided here courtesy of Wiley

RESOURCES