. 2012 May 10;13(3):43–59. doi: 10.1120/jacmp.v13i3.3751

On the quantification of the dosimetric accuracy of collapsed cone convolution superposition (CCCS) algorithm for small lung volumes using IMRT

Oscar I Calvo 1, Alonso N Gutiérrez 1,, Sotirios Stathakis 1, Carlos Esquivel 1, Nikos Papanikolaou 1
PMCID: PMC5716560  PMID: 22584174

Abstract

Specialized techniques that make use of small field dosimetry are common practice in today's clinics. These new techniques represent a big challenge to the treatment planning systems due to the lack of lateral electronic equilibrium. Because of this, the necessity of planning systems to overcome such difficulties and provide an accurate representation of the true value is of significant importance. Pinnacle3 is one such planning system. During the IMRT optimization process, Pinnacle3 treatment planning system allows the user to specify a minimum segment size which results in multiple beams composed of several subsets of different widths. In this study, the accuracy of the engine dose calculation, collapsed cone convolution superposition algorithm (CCCS) used by Pinnacle3, was quantified by Monte Carlo simulations, ionization chamber, and Kodak extended dose range film (EDR2) measurements for 11 SBRT lung patients. Lesions were <3.0cm in maximal diameter and <27.0cm3 in volume. The Monte Carlo EGSnrc\BEAMnrc and EGS4\MCSIM were used in the comparison. The minimum segment size allowable during optimization had a direct impact on the number of monitor units calculated for each beam. Plans with the smallest minimum segment size (0.1cm2 to 2.0cm2) had the largest number of MUs. Although PTV coverage remained unaffected, the segment size did have an effect on the dose to the organs at risk. Pinnacle3‐calculated PTV mean doses were in agreement with Monte Carlo‐calculated mean doses to within 5.6% for all plans. On average, the mean dose difference between Monte Carlo and Pinnacle3 for all 88 plans was 1.38%. The largest discrepancy in maximum dose was 5.8%, and was noted for one of the plans using a minimum segment size of 0.1cm2. For minimum dose to the PTV, a maximum discrepancy between Monte Carlo and Pinnacle3 was noted of 12.5% for a plan using a 6.0cm2 minimum segment size. Agreement between point dose measurements and Pinnacle3‐calculated doses were on average within 0.7% in both phantoms. The profiles show a good agreement between Pinnacle3, Monte Carlo, and EDR2 film. The gamma index and the isodose lines support the result.

PACS number: 87.56.bd

Keywords: Monte Carlo, CCCS, small fields, SBRT

I. INTRODUCTION

The growing use of more specialized techniques, such as intracranial and extra‐cranial stereotactic radiotherapy (SRT) in which small fields with electronic disequilibrium are present, has been subject to investigation by many researchers.( 1 ) Intensity‐modulated radiotherapy (IMRT) plans make use of several superimposed small fields to achieve a highly conformal dose distribution to the target volume and spare healthy tissue.

IMRT treatment planning systems calculate an ideal intensity map for each field using optimization routines with physical dose objectives set by the user.( 2 ) Once the optimization completes, the optimized fluence for each field of the plan is realized into several small segments. These segments of small aperture are generated by the segmentation process that produces the delivery sequence of the multileaf collimator. Under such conditions of small field geometries, the electronic equilibrium can be lost, making it challenging for the dose calculation algorithm to accurately predict the dose, especially in the presence of tissue heterogeneities.( 1 )

IMRT planning utilizes iterative optimization techniques, during which the dose needs to be continuously calculated. To improve the efficiency and speed of such implementations, a pencil beam algorithm is typically used for the intermediate (iterative) dose calculation steps, thus introducing convergence errors.( 3 ) It is important to always perform a full 3D dose calculation at the conclusion of the optimization to evaluate the dose the patient will truly receive, especially since it is known that the pencil beam implementations are unreliable in the presence of tissue heterogeneities.

The traditional two‐step process of IMRT inverse planning may produce more segments and MUs than necessary—largely due to the conversion of the fluence profiles to deliverable MLC settings. However, with more advance techniques like direct machine parameter optimization (DMPO), the plan will not degrade because of the lack of post processing (no need of fluence profiles conversion to MLC settings).( 4 )

When calculating the dose in a low‐density medium such as lung, the use of narrow beams may produce significant perturbations that are energy and density dependent and ultimately affect the accuracy of the dose calculation. This problem is more pronounced when the TPS uses simple, one‐dimensional density scaling.( 5 8 ) The level of accuracy improves with the use of sophisticated treatment planning algorithms( 9 17 ) where multisource modeling is included, allowing a more accurate dose prediction for small fields and under non‐equilibrium conditions. ( 18 , 19 ) It has been shown that the accuracy of small field dosimetry is greatly improved when Monte Carlo simulations are employed, especially for beam sizes less than 3.0 by 3.0cm2 in inhomogeneous media.( 20 23 )

Currently, the model‐based dose calculation methods considered most accurate in radiotherapy are Monte Carlo transport and the convolution/superposition method.( 24 ) Researchers have reported various techniques and devices used for obtaining dosimetric input data for small and complicated photon beams.( 25 29 ) Francescon et al.( 30 ) reported that Pinnacle3 overestimates the dose by up to 8% for narrow 1.0 cm wide photon segments. Azcona et al.( 31 ) reported an overestimation of calculated dose between 4% and 14% for fields of widths between 1 and 3 cm.

Sophisticated Monte Carlo codes are becoming available in parallel to the emerging use of small field radiotherapy techniques.( 32 ) By explicitly modeling the particle transport, complete Monte Carlo simulations are expected to result in the highest dose calculation accuracy( 33 ) and would become the standard for future planning systems. However, due to the long time required for a MC simulation, clinical implementation is currently not possible.

The quantification of the dose calculation accuracy for treatment planning systems is a matter of high importance in radiotherapy, as inaccuracies may lead to patient complications. This study has a two‐fold purpose: first, the dose calculated using the collapsed cone convolution superposition method used in Pinnacle3 TPS is compared against Monte Carlo dose calculations, ion chamber, and EDR2 film measurements; second, the effect of the smallest segment size allowed during optimization is examined in terms of dose calculation accuracy, and dose to organs at risk and PTV coverage. Although several studies exist that directly compare the CCCS dose calculations against Monte Carlo calculations, there has not been a study to date that includes the above stated purposes.

II. MATERIALS AND METHODS

A.1 Patient selection

Eleven patients (n=11) were randomly chosen for this study. All the patients were treated using a stereotactic body radiation therapy (SBRT) technique. The patient selections were restricted such that the tumor size was <3.0cm in maximum diameter and the volume was <27.0cc. The location and target volumes of each patient are shown in Table 1.

Table 1.

PTV location and volume size for all patients.

Location Volume Size
Patient 1 Mid right
1.4cm3
Patient 2 Lower right
5.0cm3
Patient 3 Upper left
11.9cm3
Patient 4 Mid right
19.1cm3
Patient 5 Lower right
13.6cm3
Patient 6 Mid right
7.7cm3
Patient 7 Upper right
8.0cm3
Patient 8 Upper left
22.4cm3
Patient 9 Mid upper‐right
17.0cm3
Patient 10 Lower left
10.5cm3
Patient 11 Lower left
9.1cm3

All patients were scanned in the supine position using a 16 slice GE LightSpeed (GE Medical, Waukesha WI), and immobilized using the Body Pro‐Lok system (CIVCO Medical Solutions, Iowa City IA). A single radiation oncologist delineated the gross tumor volume (GTV) and organs at risk (OAR) using a free breathing scan. CT images and structures were exported to Pinnacle3 (Philips Medical, Fitchburg WI) treatment planning system through DICOM and DICOM‐RT.

A.2 Patients' treatment plans

For each patient, eight optimized plans were created varying the minimum allowable segment size, while maintaining the same optimization dose constraints. The dose prescribed to the target was 45.0 Gy in three fractions (see Table 2). Five coplanar beams were used in all cases; these were optimally placed in order to avoid critical organs. All plans were optimized for delivery using a Varian 2100EX (Varian Medical Systems, Palo Alto, CA) 6 MV photon beam equipped with a 120 Millennium multileaf collimator (MLC). The minimum allowable segment sizes were varied from 0.1cm2 to 6.0cm2 (0.1, 0.25, 1.0, 2.0, 3.0, 4.0, 5.0, and 6.0cm2). Smaller segment sizes allow for higher spatial resolution of the optimized fluence map and, thus, permit greater flexibility during optimization. However, the use of smaller segment sizes could introduce unwanted electronic disequilibrium effects. On the other hand, by limiting the maximum size to 6.0cm2, the optimization is more restricted and might fail to provide an optimal solution. It should be noted that the minimum size of the segment also affects the number of segments created and total number of monitor units (MU) required for the delivery.

Table 2.

Optimization parameters used for the PTV and OARs in Pinnacle3 planning system.

Organ Volume (%) Dose(Gy) Type of Constraint
Target 45 Min Dose
Lung (Isp) 90 10 Max DVH
Lung (Isp) 50 20 Max DVH
Lung (Isp) 45 10 Max DVH
Lung (Cntrl) 90 1 Max DVH
Lung (Cntrl) 50 2.5 Max DVH
Lung (Cntrl) 10 5 Max Dose
Spinal Cord 10 10 Max DVH
Trachea 50 5 Max DVH
Trachea 10 10 Max DVH
Esophagus 10 5 Max DVH
Heart 10 5 Max Dose

The direct machine parameter optimization (DMPO) within Pinnacle3 was used during optimization for all plans. Utilizing the DMPO option eliminates the need for optimization conversion and, beams' fluence generated correspond to deliverable ones.

After optimization, a final dose calculation using the collapsed cone convolution superposition (CCCS) algorithm( 9 ) was performed. Dose grid size used for calculations was 0.2 cm by 0.2 cm by 0.2 cm. Special care was taken during the definition of the dose grid extent in order to ensure complete enclosure of all OARs.

A.3 Monte Carlo calculations

Monte Carlo calculations for each of the optimized fluence plans were performed using both patient geometry and two phantom geometries. CT scans from each patient and the CT scans of the two phantoms were converted to Monte Carlo geometry. A total of 3 by 88=264 Monte Carlo dose calculations were performed. Monte Carlo calculations were performed in two steps. First, a phase space file of the linear accelerator was created using the EGSnrc/BEAMnrc system.( 34 ) Geometry and material composition used in the simulation were based on the specifications of the linear accelerator treatment head as provided by the manufacturer. The cutoff energies used for the simulations were 700 keV for electrons and 10 keV for photons. Energy thresholds for X‐ray production (AE) and for bremsstrahlung production (AP) were 700 keV and 10 keV, respectively. Results of the LINAC simulation were in agreement (±1% or within 1 mm) with the measured PDD curves and the dose profiles at various depths in water obtained with a 0.125 cc PTW‐Freiburg ion chamber. After the Monte Carlo commissioning of the linear accelerator, a phase space file was created before the jaws in order to be used in the dose calculation simulations (Fig. 1).

Figure 1.

Figure 1

Monte Carlo vs. measurements for several field sizes: a) PDDs, and b) profiles.

The Monte Carlo dose calculations were performed using the EGS4/MCSIM user code. The EGS4/MCSIM code system can accurately calculate patient dose distributions by simulating the accelerator head leakage, MLC leaf leakage and scatter, and the effect of beam modifiers such as collimator jaws, wedges, and blocks. Furthermore, calculation times are 10–30 times faster than other widely available general‐purpose Monte Carlo codes.( 35 )

The EGS4/MCSIM code was used in this work due to its extended functionality. The code is capable of calculating dose in a patient given the intensity map (ODM) or the Radiation Therapy Plan (RTP) file, which includes patient setup parameters and beam and leaf‐sequence information. The ODM files generated by Pinnacle3 were used as input for EGS4/MCSIM. Based on the dose per incident particle, as derived from calculations using calibration conditions, EGS4/MCSIM can calculate the absolute dose to the patient. Moreover, if contours exist in the patient geometry from computed tomography (CT) images, doses to each organ may be calculated and DVHs generated for all contoured organs.

A. 4 Ion chamber and film measurements

For each of the resulting 88 plans, two IMRT QA verification plans were created using a homogenous and heterogeneous phantom (Fig. 2). The heterogeneous phantom (Standard Imaging, Middleton WI) was used in order to mimic the patient geometry during radiation delivery. Each phantom was structured so that ionization chamber could be incorporated, as well as an extended dose radiographic (EDR2) film (Kodak Inc., Rochester NY). The film was placed within the phantom to measure the planar dose distribution. Verification plans were calculated and delivered with all of the beams positioned in planned angles. This resulted in 176 verification plans.

Figure 2.

Figure 2

Setup geometries for film and ionization chamber measurements: (a) Plexiglass homogeneous phantom, and (b) StandardImaging heterogeneous phantom.

A vented cylindrical ionization chamber (PTW N31003, New York NY) with a sensitive volume of 0.3 cc was used for all measurements. Measured point doses from the ion chamber were compared against their respective calculated point dose from the TPS. The film dose measurements were compared to their respective calculated planar doses exported from Pinnacle3. Film calibration was performed by exposing films to a dose range of 0.1 to 4.0 Gy using a step wedge with known dose values at the center of each step.( 36 ) All films were scanned on a Vidar 16‐bit scanner (VIDAR Systems Corporation Herndon, VA) and analyzed using the RIT113v5 (Radiological Imaging Technology, Colorado Springs, CO), software platform.

III. RESULTS

A.1 Effect of minimum segment size on plan quality

The minimum segment size allowable during optimization had a direct impact on the number of monitor units calculated for each beam. Plans with the smallest minimum segment size (0.1cm2 to 2.0cm2) had the largest number of MUs per field (Table 3). Dose statistics for the planning target volumes of all of the patients are shown in Table 4. Small variations in the PTV dose statistics were noted, leading to the conclusion that there is no significant dependence of the PTV coverage on the minimum allowable segment size.

Table 3.

Mean MU per field as a function of minimum segment size for each patient calculated using the CCCS algorithm.

Minimum Segment Field Size
0.1 0.25 1.0 2.0 3.0 4.0 5.0 6.0
Patient 1 363 378 357 381 294 270 263 258
Patient 2 325 324 313 307 288 275 282 271
Patient 3 362 366 372 320 318 301 295 287
Patient 4 391 391 402 462 355 321 314 285
Patient 5 365 368 347 371 320 314 303 298
Patient 6 334 335 324 405 294 288 287 274
Patient 7 344 345 334 308 304 300 289 276
Patient 8 359 357 344 433 327 318 313 298
Patient 9 368 341 358 365 337 336 336 321
Patient 10 355 356 357 366 311 303 299 282
Patient 11 368 370 332 401 326 324 314 298
Mean 358 357 349 374 316 305 300 286

Table 4.

Pinnacle3 PTV dose statistics for all patients as a function of segment size.

Segment Size Minimum Dose (cGy) Maximum Dose (cGy) Mean Dose (cGy) Dose STD (cGy)
0.1cm2
3978.8 4791.2 4544.7 115.4
0.25cm2
3964.6 4619.1 4429.5 99.0
1.0cm2
4019.1 4590.7 4467.4 90.0
2.0cm2
4069.9 4656.0 4467.3 77.5
3.0cm2
4016.1 4600.0 4438.3 81.6
4.0cm2
4023.5 4603.0 4441.4 81.7
5.0cm2
4025.7 4612.3 4443.6 83.5
6.0cm2
4004.8 4623.0 4431.2 86.6

Although PTV coverage remained unaffected, the segment size did have an effect on the dose to the organs at risk. When segment sizes are large, the minimum segment shape is comparable to the aperture size and thereby limits the spatial resolution of the deliverable fluence. This limitation affects the sparing of critical structures, particularly those adjacent to the target volume. The dose to the organs at risk was shown to increase overall for very small and large minimum segment sizes allowable (Table 5).

Table 5.

Mean (cGy) and maximum dose (cGy) to the lung and spinal cord for each patient plan calculated.

# Site
0.1cm2
0.25cm2
1cm2
2cm2
3cm2
4cm2
5cm2
6cm2
1 Lung Mean 349.0 344.3 348.7 350.2 356.7 346.9 348.3 351.4
Max 4500.2 4482.7 4551.6 4613.3 4589.2 4579.8 4593 4629.2
Spinal Mean 589.2 613.8 612 672 539.7 706.3 717.2 586.6
Cord Max 787.8 852.5 721 920.1 620.7 836.9 851.6 661.4
2 Lung Mean 494.3 499 506.8 520.3 493.2 502.4 501.4 501.5
Max 4533.3 4803.1 4890.4 4639.7 4532.7 4595 4592.2 4603.8
Spinal Mean 74.6 86.7 88.7 70.2 70.9 72.2 71.6 73.7
Cord Max 237.5 290.6 320.3 185.5 226.6 241.1 218.1 225.9
3 Lung Mean 420.3 481.3 503.8 511.9 504 504.9 508.8 515.5
Max 5748.8 4643.9 4628.4 4601.8 4607.8 4591.9 4589.4 4598.6
Spinal Mean 17.2 62.9 52.2 43.6 45.04 42.8 39.4 39.8
Cord Max 32.4 128 87.9 87.1 125.7 81.5 73 75.5
4 Lung Mean 516 516.2 522.1 537.8 524.9 526.4 526.7 521.3
Max 4636.2 4636. 8 4672.2 4844.5 4678.9 4680.7 4699.4 4729.6
Spinal Mean 680.9 684.6 684.7 612.6 686 686.6 679.3 676.2
Cord Max 1285.3 1287.7 1302.7 1556.9 1300.7 1274.4 1278 1257.6
5 Lung Mean 262.3 265.8 265.8 266.2 270.9 267.7 266.7 268.8
Max 4495.8 4494.6 4494.6 4541.8 4536.9 4540.7 4557.7 4570.2
Spinal Mean 7.3 7.4 7.4 8.6 8.2 8 7.7 8
Cord Max 15.7 15.7 15.7 17.3 163 15.8 16.1 16.5
6 Lung Mean 334.5 337.8 335.1 323.5 335.3 338 338.4 342.1
Max 4555.8 4610.2 4568.7 4606.3 4550.7 4556.1 4561.6 4552.2
Spinal Mean 286 286.8 316.2 182.4 281.6 285.2 283.5 314.1
Cord Max 763.9 676.7 767.4 697.3 742.8 729 722.2 745.4
7 Lung Mean 293.2 294.7 296.4 301.9 299.7 298.6 299.2 299.5
Max 4573.8 4591.4 4617.3 4625.4 4592.7 4601.3 4610 4576.1
Spinal Mean 66.4 66.1 66.8 65.8 61.5 60.8 59 57.2
Cord Max 154.1 153.2 154.4 147.4 141 139.2 131.8 127.7
8 Lung Mean 229.5 229 231.2 234.1 231.8 230.9 231 232.7
Max 4624.3 4598.5 4603.3 4709.6 4615.1 4618.6 4637 4626.8
Spinal Mean 68.1 69 69.9 57.9 69.3 68.4 67.6 75.3
Cord Max 405.4 405.7 417.4 373.3 415.4 403.9 395.3 425.3
9 Lung Mean 584.2 571 584.2 606.1 583.7 586.1 586.4 590.6
Max 1587.1 4464.6 4569.2 4609.1 4544.4 4554.8 4559.2 4554.4
Spinal Mean 214.3 317.4 216.5 174.8 212.8 213.2 212.8 224.5
Cord Max 461.3 583.8 468.7 422.9 467.1 457.3 453.9 440.4
10 Lung Mean 314.8 314.7 315.6 321.9 313.8 314.5 314.9 313.7
Max 4629.6 4626.1 4645.8 4719 4667.6 4659 4668.2 4636.2
Spinal Mean 642.5 641.4 633.7 598.1 639.7 635.3 634.8 653.9
Cord Max 1163.4 1163.2 1137.3 1120.1 1139.1 1142.9 1144.3 1131.8
11 Lung Mean 362.6 362.4 364 607.4 367.4 368.9 367.2 368.9
Max 4555.9 4553.5 4548.5 4678.6 4568.2 4559 4566.5 4543
Spinal Mean 369.4 369.1 364 639.8 361.6 364 365 379.3
Cord Max 572.4 572.1 571.5 748.5 573 574 573.4 574

Small differences were noted in the DVHs as the minimum segment sizes were varied for the target volumes (Fig. 3). Larger discrepancies were witnessed in the DVHs for the organs at risk with the exception of the lung. The DVH of the lung for each patient was not significantly different among the various optimized plans. Different results were observed for the spinal cord, esophagus, and trachea DVHs in which it can be observed that the segment size had an influence in the dose. The location of the tumor relative to the heart for the sample patient makes it difficult to draw any conclusive results. Similar results can be observed for the rest of the patients. Figure 3 shows the DVHs of all eight plans for patient 1 for: (1) the PTV, (2) the lung, (3) spinal cord, (4) the esophagus, (5) the trachea, and (6) the heart.

Figure 3.

Figure 3

Dose volume histogram (DVH) of all eight plans for patient 1 for: (a) the PTV, (b) the lung, (c) spinal cord, (d) the esophagus, (e) the trachea, and (f) the heart.

A.2 Pinnacle3 and Monte Carlo simulations comparison

A.2.1 Patient dose calculation comparison

All the Monte Carlo dose calculations had a maximum dose uncertainty of less than 1%. The average minimum, maximum, and mean doses to the PTV for all of the patients were extracted from both Pinnacle3 and Monte Carlo. The maximum and mean doses to the organs at risk were also reported (Table 6). Pinnacle3‐calculated PTV mean doses were in agreement to Monte Carlo‐calculated mean doses to within 5.6% for all plans. On average, the mean dose difference between Monte Carlo and Pinnacle3 for all 88 plans was 1.38%.

Table 6.

Percent differences of mean dose values between Pinnacle3 and MC calculation as a function of minimum segment size for the target and specific organs at risk.

PTV Lung Spinal Cord
Plan Min Ave Max Ave Mean Ave Max Ave Mean Ave Max Ave Mean Ave
0.10cm2
2.1±6.2
2.9±2.6
1.3±2.1
6.0±3
2.3±3.1
0.7±23.5
15.9±24.4
0.25cm2
0.3±8.9
1.0±3.6
0.5±5.5
3.9±4.3
0.1±7.2
12.4±25.3
18.2±30
1cm2
4.0±6.2
3.7±2.1
2.7±2.9
6.4±3.4
3.5±3.1
11.0±25.5
16.4±30.6
2cm2
0.1±10.4
2.6±1.9
0.4±2.4
7.1±2.9
1.9±3.8
10.6±37.9
32.4±41
3cm2
2.4±7.7
3.5±0.7
1.6±2.8
5.7±1.6
1.8±3.1
2.1±28
18.8±25.4
4cm2
1.3±7.3
3.1±1.7
1.4±2.9
5.0±2.7
2.0±3.8
4.4±21.9
20.2±23.9
5cm2
2.9±8.1
3.6±0.8
2.0±2.8
5.4±1.7
2.1±3.7
9.8±26.8
20.9±24.4
6cm2
3.8±8.7
2.9±1.5
2.1±3.5
4.8±2.0
2.4±4.1
14.0±30.8
21.7±28.8
Mean
2.01±6.82
2.91±1.41
4.95±2.42
5.54±2.16
1.99±3.08
8.13±24.3
19.31±24.81

Higher discrepancies were observed for the minimum and maximum doses calculated. The largest discrepancy in maximum dose was 5.8% and was noted for one of the plans using a minimum segment size of 1.0cm2. For the minimum dose to the PTV, a maximum discrepancy between Monte Carlo and Pinnacle3 was noted of 12.5% for a plan using a 6.0cm2 minimum segment size. Average differences for maximum and minimum doses to the PTV were 2.91 and 2.01%, respectively. Similar trends were observed for the organs at risk. It is important to mention that larger percentage discrepancies observed for some organs at risk are partly due to the fact that the absolute dose value is small. Hence, a relatively small absolute dose difference will lead to a relatively high percentage difference.

A DVH comparison for the PTV, spinal cord, and lung between Pinnacle3 and Monte Carlo is presented in Fig. 4. The figure presents the comparison for three different segment sizes (0.25cm2, 3cm2, and 4cm2). The Pinnacle3 PTV coverage was in agreement to the Monte Carlo calculation for the 0.25cm2 and 4cm2 plans. Monte Carlo showed better coverage of the PTV for the 2cm2 plan. The spinal cord and the lung showed similar coverage for all plans. The Monte Carlo calculation gave a lower dose to the spinal cord compared to the Pinnacle3 plan, and the lung received a larger dose with the Monte Carlo calculation as compared to the Pinnacle3 plan.

Figure 4.

Figure 4

Dose volume histogram (DVH) comparison as a function of segment size between Pinnacle3 and Monte Carlo calculations: (a) 0.25cm2, (b) 2cm2, and (c) 4cm2.

The Pinnacle3 dose grid was exported and compared against the one calculated with Monte Carlo. The comparison between Monte Carlo and Pinnacle3 planar isodose distributions were performed on the RITv5 software. An in‐house software was developed in MATLAB (v7.8 R2009a, The MathWorks, Natick, MA) to convert 3D dose distributions from Monte Carlo calculations into RITv5 readable format. Comparison was then possible on a slice‐by‐slice basis. The coronal, sagittal or transverse planes intersecting the beam isocenter were chosen for comparison. Representative isodose line comparisons are shown in Figs. 5 and 6 for different patients and minimum segment sizes. The corresponding gamma index, as well as horizontal and vertical profiles for these comparisons using a criterion of DTA 2 mm/percent dose difference 3%, is shown in Figs. 5 and 6. The profiles show a good agreement between Pinnacle3 and Monte Carlo. The gamma index and the isodose lines support the result.

Figure 5.

Figure 5

Dose comparison between MC and CCCS for patient 1 and for segment sizes 0.25cm2 and 4cm2, respectively: (a) and (e) isodose distributions; (b) and (f) horizontal profiles; (c) and (g) gamma images; (d) and (h) vertical profiles.

Figure 6.

Figure 6

Dose comparison between MC and CCCS for patient 2 and for segment sizes 0.25cm2 and 4cm2, respectively: (a) and (e) isodose distributions; (b) and (f) horizontal profiles; (c) and (g) gamma images; (d) and (h) vertical profiles.

A.3 Ion chamber and film measurements

Using both phantoms, measurements were acquired using a PTW 0.3 cc ionization chamber and radiographic EDR2 film. Point absolute dose measurements using the ionization chamber were made for all plans. Agreement between point dose measurements and Pinnacle3 calculated doses were on average within 0.7% in both phantoms (see Table 7). Coronal planar doses at the level of the film were also exported. The dose to the center of the planar dose was compared between measurements and calculations. The agreement in this case was inferior to the point measurements. This is probably due to uncertainties in the film calibration curve and inaccuracy in determining the “center” of each step. Observing the absolute film doses at the center between the solid and lung phantom measurements and the respective TPS calculations, it is clear that better agreement is achieved in the case where the solid phantom was used. This is due to the difference in the phantoms composition and to the inhomogeneities embedded in the lung phantom (Solid Acrylic (Virtual Water, StandardImaging, Middleton, WI)) and two wood slabs lung tissue equivalent. This heterogeneous phantom gives the closest representation of a real clinical case in comparison to the solid water phantom in which case the difference in the results gives a significant representation of the actual plan delivered in a patient. These results are summarized in Table 7

Table 7.

Percent differences of the average values between Pinnacle3, ion chamber, and film.

Solid Phantom Lung Phantom
Plan Ion Chamber vs. Pinnacle Film vs. Pinnacle Ion Chamber vs. Pinnacle Film vs. Pinnacle
0.1cm2
0.7±5.6
0.7±7.8
0.7±5.9
5.7±9.8
0.25cm2
0.7±6.3
2±7.7
0.6±6
2.6±16.6
1.0cm2
0.6±6.5
5.5±10.3
0.6±8.4
4.6±13.1
2.0cm2
0.5±7
1.6±8.8
1±7
6.8±10.6
3.0cm2
0.1±5.9
2.4±5.8
0.4±6.5
5.4±10.3
4.0cm2
0.1±6.1
1.6±6.9
0.3±6.3
6.2±10
5.0cm2
0.1±7.6
0.5±8
0.4±6.7
6.9±10
6.0cm2
0.4±7.7
1±7.8
0.2±7.1
6.7±10.2

Planar doses exported from Pinnacle3 were compared against their respective film measurements. The isodose distribution, gamma index using 2 mm DTA and 3% dose difference, and profiles were used for the evaluation of the TPS performance (Figs. 7 and 8). The figures show a good agreement between Pinnacle3 and film measurements in the high‐dose region, in which case the result is confirmed with the isodose lines and gamma index. Small differences can be seen in the low‐dose regions which can be explained by the reasons stated above.

Figure 7.

Figure 7

Dose comparison between film (EDR2) and CCCS for patient 1 and segment size 0.25cm2 for the solid water phantom (a–d) and the lung phantom (e–f): (a) and (e) isodose distributions; (b) and (f) horizontal profiles; (c) and (g) gamma images; (d) and (h) vertical profiles.

Figure 8.

Figure 8

Dose comparison between film (EDR2) and CCCS for patient 1 and segment size 5cm2 for the solid water phantom (a–d) and the lung phantom (e–f): (a) and (e) isodose distributions; (b) and (f) horizontal profiles; (c) and (g) gamma images; (d) and (h) vertical profiles.

IV. DISCUSSION

The average values of the percent difference between Pinnacle3 and MC for the minimum, maximum, and mean doses characterizes the ability of Pinnacle3 to accurately spare organs at risk while at the same time cover the PTV. The results showed that Pinnacle3 is able to match the predictions made by the Monte Carlo simulations to within a few percent differences with the understanding that structures such as the spinal cord, trachea, and esophagus—typically small in volume—may in some circumstances be misleading. This is partly due to small differences in the conversion of the contours from Pinnacle3 to Monte Carlo phantoms because of voxel size limitations which create potential discrepancies in the calculated volume. However, overall, high‐dose regions were accurately predicted by Pinnacle3.

Superimpositions of the DVHs showed that Pinnacle3 is able to obtain good coverage of the PTV independent of the segment size used. For the organs at risk, the TPS showed good agreement in the coverage of the total lung; as for the spinal cord, the location of the tumor influenced the dose statistics. Pinnacle3 is able to deliver the prescribed dose to the PTV independent of the segment size used indicating that the planning system is able to model the difficulties associated with small fields.

The Monte Carlo codes EGSnrc\BEAMnrc and EGS4\MCSIM were used to calculate the particle transport to obtain the absorbed dose with the help of the ODMs by Pinnacle3. These codes were used because they have shown that they provide the feasibility and flexibility to give accurate results. The number of histories used was chosen to provide a standard deviation of less than 2%. The improved Monte Carlo statistics lead to confirmation of good agreement among Pinnacle3, MC and ion chamber measurements as shown in Table 7.

Agreement between point dose measurements and Pinnacle3 calculated doses were, on average, within 0.7% in both phantom geometries. Better agreement was observed between Pinnacle3 and the solid phantom as opposed to the inhomogeneous phantom. This is due probably to the difference in the phantom's composition and to the inhomogeneities embedded in the lung phantom. Using the inhomogeneous phantom, however, provides a closer representation of real patient anatomy.

V. CONCLUSIONS

Agreement between Pinnacle3 and Monte Carlo was within ±2% for the target mean dose, while higher discrepancies (up to 4%) were observed for the minimum and maximum doses. For the ipsilateral lung, a difference in the maximum dose of up to 7%, and 2% for the mean dose was noted. Small differences in the total coverage of the PTV were found for all cases as a function of segment size. Good agreement between Pinnacle3 and MC using dose profiles, isodose distributions, and Gamma Index analysis verified the results—particularly in the high dose region.

REFERENCES

  • 1. Das IJ, Ding GX, Ahnesjö A. Small fields: nonequilibrium radiation dosimetry. Med Phys, 2008;35(1):206–15. [DOI] [PubMed] [Google Scholar]
  • 2. Lydon JM. Theoretical and experimental validation of treatment planning for narrow MLC defined photon fields. Phys Med Biol. 2005;50(11):2701–14. [DOI] [PubMed] [Google Scholar]
  • 3. Papanikolaou N and Stathakis S. Dose‐calculation algorithms in the context of inhomogeneity corrections for high energy photon beams. Med Phys. 2009;36(10):4765–75. [DOI] [PubMed] [Google Scholar]
  • 4. Hårdemark B, et al. Direct machine parameter optimization with RayMachine in Pinnacle3. RaySearch White Paper. Stockholm, Sweden: RaySearch Laboratories AB; 2003. [Google Scholar]
  • 5. Chetty IJ, Charland PM, Tyagi N, McShan DL, Fraass BA, Bielajew AF. Photon beam relative dose validation of the DPM Monte Carlo code in lung‐equivalent media. Med Phys. 2003;30(4):563–73. [DOI] [PubMed] [Google Scholar]
  • 6. Saitoh H, Fujisaki T, Sakai R, Kunieda E. Dose distribution of narrow beam irradiation for small lung tumor. Int J Radiat Oncol Biol Phys. 2002;53(5):1380–87. [DOI] [PubMed] [Google Scholar]
  • 7. Al‐Hallaq HA, Reft CS, Roeske JC. The dosimetric effects of tissue heterogeneities in intensity‐modulated radiation therapy (IMRT) of the head and neck. Phys Med Biol. 2006;51(5):1145–56. [DOI] [PubMed] [Google Scholar]
  • 8. Jones AO and Das IJ. Comparison of inhomogeneity correction algorithms in small photon fields. Med Phys. 2005;32(3):766–76. [DOI] [PubMed] [Google Scholar]
  • 9. Papanikolaou N, Mackie TR, Gehring M, Fairbanks J. Clinical implementation of a convolution based algorithm for 3D treatment planning [abstract]. Int J Radiat Oncol Biol Phys. 1995;32(971):302. [Google Scholar]
  • 10. Tillikainen L, Helminen H, Torsti T, et al. A 3D pencil‐beam‐based superposition algorithm for photon dose calculation in heterogeneous media. Phys Med Biol. 2008;53(14):3821–39. [DOI] [PubMed] [Google Scholar]
  • 11. Craig J, Oliver M, Gladwish A, Mulligan M, Chen J, Wong E. Commissioning a fast Monte Carlo dose calculation algorithm for lung cancer treatment planning. J Appl Clin Med Phys. 2008;9(2):82–97. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Gagné IM and Zavgorodni S. Evaluation of the analytical anisotropic algorithm in an extreme water‐lung interface phantom using Monte Carlo dose calculations. J Appl Clin Med Phys. 2006;8(1):33–46. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Garcia‐Vicente F, Minarmbres A, Jerez I, Modolell I, Pérez L, Torres JJ. Experimental validation tests of fast Fourier transform convolution and multigrid superposition algorithms for dose calculation in low density media. Radiother Oncol. 2003;67(2):239–49. [DOI] [PubMed] [Google Scholar]
  • 14. Kunieda E, Deloar HM, Kishitani N, et al. Variation of dose distribution of stereotactic radiotherapy for small‐volume lung tumors under different respiratory conditions. Phys Med. 2008;24(4):204–11. [DOI] [PubMed] [Google Scholar]
  • 15. Scholz C, Shulze C, Oelfke U, Bortfield T. Development and clinical application of a fast superposition algorithm in radiation therapy. Radiother Oncol. 2003;69(1):79–90. [DOI] [PubMed] [Google Scholar]
  • 16. Bragg CM and Conway J. Dosimetric verification of the anisotropic analytical algorithm for radiotherapy treatment planning. Radiother Oncol. 2006;81(3):315–23. [DOI] [PubMed] [Google Scholar]
  • 17. Boyer AL and Mok EC. Calculation of photon dose distributions in an inhomogeneous medium using convolutions. Med Phys. 1986;13(4):503–09. [DOI] [PubMed] [Google Scholar]
  • 18. Zhu TC and Bjärngard BE. The fraction of photons undergoing head scatter in x‐ray beams. Phys Med Biol. 1995;40(6):1127–34. [DOI] [PubMed] [Google Scholar]
  • 19. Zhu TC and Bjärngard BE. The head‐scatter factor for small field sizes. Med Phys. 1994;21(1):65–68. [DOI] [PubMed] [Google Scholar]
  • 20. Kijewski PK, Bjärngard BE, Petti PL. Monte Carlo calculations of scatter dose for small field sizes in a 60Co beam. Med Phys. 1986;13(1):74–77. [DOI] [PubMed] [Google Scholar]
  • 21. Ding GX, Duggan DM, Coffey CW. Commissioning stereotactic radiosurgery beams using both experimental and theoretical methods. Phys Med Biol. 2006;51(10):2549–66. [DOI] [PubMed] [Google Scholar]
  • 22. Ding GX. Dose discrepancies between Monte Carlo calculations and measurements in the buildup region for a high‐energy photon beam. Med Phys. 2002;29(11):2459–63. [DOI] [PubMed] [Google Scholar]
  • 23. Ahnesjö A. Collimator scatter in photon therapy beams. Med Phys. 1995;22(3):267–78. [DOI] [PubMed] [Google Scholar]
  • 24. Naqvi SA, Earl MA, Shepard DM. Convolution/superposition using the Monte Carlo method. Phys Med Biol. 2003;48(14):2101–21. [DOI] [PubMed] [Google Scholar]
  • 25. Paskalev KA, Seuntjens JP, Patrocinio HJ, Podgorsak EB. Physical aspects of dynamic stereotactic radiosurgery with very small photon beams (1.5 and 3 mm in diameter). Med Phys. 2003;30(2):111–18. [DOI] [PubMed] [Google Scholar]
  • 26. Sánchez‐Doblado F, Capote R, Leal A, et al. Micro ionization chamber for reference dosimetry in IMRT verification: clinical implications on OAR dosimetric errors. Phys Med Biol. 2005;50(5):959–70. [DOI] [PubMed] [Google Scholar]
  • 27. Sánchez‐Doblado F, Andreo P, Capote R, et al. Ionization chamber dosimetry of small photon fields: a Monte Carlo study on stopping‐power ratios for radiosurgery and IMRT beams. Phys Med Biol. 2003;48(14):2081–99. [DOI] [PubMed] [Google Scholar]
  • 28. Crop F, Reynaert N, Pittomvils G, et al. Monte Carlo modeling of the ModuLeaf miniature MLC for small field dosimetry and quality assurance of the clinical treatment planning system. Phys Med Biol. 2007;52(11):3275–90. [DOI] [PubMed] [Google Scholar]
  • 29. Araki F, Ikegarni T, Ishidoya T, Kubo HD. Measurements of Gamma‐Knife helmet output factors using a radiophotoluminescent glass rod dosimeter and a diode detector. Med Phys. 2003;30(8):1976–81. [DOI] [PubMed] [Google Scholar]
  • 30. Francescon P, Cora S, Chiovati P. Dose verification of an IMRT planning system with BEAM EGS4‐based Monte Carlo code. Med Phys. 2003;30(2):144–57. [DOI] [PubMed] [Google Scholar]
  • 31. Azcona JD, Siochi RA, Azinovic I. Quality assurance in IMRT: importance of the transmission through the jaws for an accurate calculation of absolute doses and relative distributions. Med Phys. 2002;29(3):269–74. [DOI] [PubMed] [Google Scholar]
  • 32. De Vlamynck K, Palmans H, Verhaegen F, De Wagter C, De Neve W, Thierens H. Dose measurements compared with Monte Carlo simulations of narrow 6 MV multileaf collimator shaped photon beams. Med Phys. 1999;26(9):1874–82. [DOI] [PubMed] [Google Scholar]
  • 33. Vanderstraeten B, Reynaert N, Paelinek L, et al. Accuracy of patient dose calculatin for lung IMRT: a comparison of Monte Carlo, convolution/superposition, and pencil beam computations. Med Phys. 2006;33(9):3149–58. [DOI] [PubMed] [Google Scholar]
  • 34. Kawrakow I and Rogers DWO. The EGSnrc Code System: Monte Carlo simulation of electron and photon transport. NRC Technical Report PIRS–701. Ottawa, Ontario: National Research Council of Canada; 2000. [Google Scholar]
  • 35. Ma CM, Li JS, Pawlicki P, et al. MCSIM a Monte Carlo dose calculation tool for radiation therapy. In: Yi BY, Aln SD, Choi EK, Ha SW. (eds.), Proceedings of the XIVth International Conference on the Use of Computers in Radiation Therapy (ICCR). Seoul, Korea: Jeong Publishing; 2004. p. 515–19. [Google Scholar]
  • 36. Liu Y, Shi C, Tynan P, Papanikolaou N. Dosimetric characteristics of dual‐layer multileaf collimation for small‐field and intensity‐modulated radiation therapy applications. J Appl Clin Med Phys. 2008;9(2):15–29. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Applied Clinical Medical Physics are provided here courtesy of Wiley