Abstract
Purpose
MR-compatible medical devices operate within the MR environment while benefitting from the superior anatomic information of MRI. Avoiding electromagnetic interference between such instrumentation and the MR system is crucial. In this work, various shielding configurations for positron emission tomography (PET) detectors were studied and analyzed regarding RF shielding effectiveness and gradient-induced eddy current performances. However, the results of this work apply to shielding considerations for any MR-compatible devices.
Methods
Six shielding enclosure configurations with various thicknesses, patterns and materials were designed: solid and segmented copper, phosphor bronze mesh (PBM), and carbon fiber composite (CFC). A series of tests were performed on RF shielding effectiveness and the gradient-induced eddy current.
Results
For the shielding effectiveness, the solid copper with various thickness and PBM configurations yield significantly better shielding effectiveness (>15 dB) compared to CFC and segmented configurations. For the gradient-induced eddy current performance, the solid copper shielding configurations with different thicknesses showed significantly worse results, up to a factor of 3.89 dB, compared to the segmented copper, PBM and the CFC configurations.
Conclusion
We have evaluated the RF shielding effectiveness and the gradient-induced eddy current artifacts of several shielding designs and only the PBM showed positive outcomes for both aspects.
Keywords: Faraday shielding enclosure, MR-compatible, gradient-induced eddy current, RF shielding effectiveness, PET/MR, PET detector
INTRODUCTION
MR systems provide exquisite high-resolution anatomic information 1–5. To benefit from the anatomical information of MR images, PET 6–9 or single-photon emission computed tomography (SPECT) 10,11, which provide sensitive visualization and quantification of bio-molecular pathways and signatures of disease in living subjects are combined with MRI for simultaneous hybrid imaging 6,7. In addition, there are a range of other medical devices that have been brought into an MR system 12–16.
Systems that simultaneously operate within an MRI system require special design considerations to avoid the electromagnetic (EM) interference between the systems 17–24. MR images are acquired by fast-oscillating EM fields in a strong static magnetic field environment. To avoid EM interference, the inserted systems typically encapsulate electronic components inside a shielding Faraday enclosure to preserve the performances of both MRI and the inserted device 22–29. Without sufficient shielding designs, intense radiofrequency (RF) fields in the MHz range could disturb the device electronics or, reciprocally, the noise generated from the device could corrupt the MRI image 21,22,26,30.
The main function of a shielding enclosure is to shield the RF field, while also minimizing any secondary artifact, such as the eddy current induced from the MR gradient fields. The primary mechanisms of EM interference (EMI) shielding are reflection and absorption 30,31. For reflection of the incident fields, the shield must have mobile charge carriers, which interact with the fields. Electrically conducting shields are preferred but not necessarily required. In addition, connectivity in the current path enhances the shielding effectiveness. The absorption loss is described by a function known as σrμr, where σr is the relative electrical conductivity and μr is the relative permeability. With a few assumptions, including high conductivity (σ ≫ ωε), thick shield material (t ≫ δ), and the intrinsic impedance mismatch (ηs ≪ η0), the reflection loss is independent of the shield thickness, while the absorption loss is proportional to the thickness of the shield. The shielding “effectiveness” is equal to the sum of both loss mechanisms, and may be expressed in terms of material parameters as follows:
| Eq.1 |
where σ is the conductivity of the shield, ω is the angular frequency of the RF field, ε is the relative permittivity of the shield, η0 is the intrinsic impedance of air, ηs is the intrinsic impedance of the shield, t is the thickness of the shield, and δ is the skin depth of the shield. From this expression it can be seen that metals with high conductivity are by far the most common materials for EM shielding. However, a changing magnetic flux from the time-varying gradient magnetic field induces eddy currents on any conductive surfaces based on Faraday’s law of induction, which in turn may cause resistive heating or ghosting artifacts in MRI images 32–36. The power dissipation of eddy currents can be calculated using the following equation37,
| Eq.2 |
where P is the power lost per unit mass (W/kg), BP is the peak magnetic field (T), d is the thickness of the sheet (m), f is the frequency (Hz), k is a constant equal to 1 for a thin sheet and 2 for a thin wire, ρ is the resistivity of the material (Ω·m), and D is the density of the material (kg/m3).
In turn, the eddy current generates magnetic fields that oppose the primary gradient fields due to Lenz’s law 38–40; hence the imaging objects do not experience the gradient field magnitudes that were programmed. Accordingly, this results in a distorted MR image during gradient-intense sequences such as echo planar imaging (EPI). Eddy currents can generate resistive heat and vibration as well.
There have been several approaches to reduce the gradient-induced eddy currents while preserving the RF shielding effectiveness inside the MRI; one method is to reduce the shielding thickness, which directly reduces the induced surface current density. For good conductors (σ ≫ ωε) with thick shield material (t ≫ δ), the shielding effectiveness is dominantly dependent on reflection loss and slightly affected by thickness 41–44. Another way is to disturb the current loop path by segmenting or meshing the conductive surface so that only limited current loops are formed; the RF shielding effectiveness is still preserved by keeping the electrical connectivity throughout the paths 45–48. One example of the segmented design application is the RF shield placed between the RF coil and the gradient coil of commercial MR systems 45,48–50. Similarly, the wire mesh limits the induced eddy current loop based on the same principle; metal wire mesh has shown to be a good shield 50,51. Alternately, a lower conductivity shield material can be employed to reduce current induction. Although, metal is preferred for RF shields, carbon fiber composites (CFC) have been applied inside MRI systems for RF shielding and shown to reduce eddy current induction 30,52; in addition to the EM properties, its low density, high strength, high oxidation resistance and thermal stability are attractive as well.
In this paper we have designed various shielding enclosure configurations for detector modules in a PET insert for simultaneous PET/MR 53–53 and evaluated the RF shielding effectiveness and gradient-induced eddy current performances of these enclosures in several direct and indirect measurements with the goal to understand their potential effects on MR performance. To date there has not been any comprehensive work on shielding designs to be applied for PET detector modules inside an MR bore. However, the results of this work apply to shielding considerations for any MR-compatible devices.
METHODS
We have studied six shielding enclosures with various materials, thickness and patterns. The list of shielding enclosure properties is shown in Fig. 1.
Figure 1.
Overview of the custom-built shielding enclosure configuration under study. Fig 1a, 1b and 1c are Solid copper shielding cages with thicknesses of 36, 18 and 9 μm. Fig 1d, 1e and 1f are 36 μm segmented copper, 0.8 mm carbon fiber composite and 36 μm phoshpor bronze mesh shielding cages, respectively.
All shielding enclosures under study had the same dimensions for consistency. Each cage was a trapezoidal cage (bases: 6 and 8 cm, height: 4 cm, length: 20cm) (Fig. 1, bottom); six side plates were built and then assembled into an enclosure with electrical contacts and mechanical support to form a full Faraday cage. For the bare copper designs (Fig. 1(A, B and C)), different thicknesses (36 μm, 18 μm, 9 μm) of copper were printed on a 31 mil (0.79 mm) FR4 sheets. Then the five plates, excluding the top plate of the enclosure, were soldered together at the inner perimeter edge (the dotted line in Fig. 1, bottom), and electrically connected to the outer copper plane through plated vias. The top plate was mechanically held in place with brass screws and electrically connected with copper tapes (0.25″ width, 0.0014″ thick foil with 0.0021″ conductive acrylic adhesive) on the outer top edges. For the segmented copper (Fig. 1(D)), the segment pattern was based on the pattern from a double-sided RF shield 50. The 20 mil (0.51 mm) width slit was patterned on a copper plate with the maximum length of 20 cm and the smallest slit-to-slit pitch of 0.5 cm. The two segmented patterns were separated by a 31 mil (0.79 mm) FR4 sheet. The assembly method was the same as above. Then, CFC sheet (DragonPlate, NY, USA), with orthotropic laminates utilizing a twill weave at 0°/90° orientation, was computerized numerical control (CNC) machined to the reference shielding enclosure dimensions, mechanically assembled through brass screws, and electrically connected to the plates using copper tape on the outside edge between the plates (Fig. 1(E)). The copper tape was applied in short pieces to avoid creating a large copper loop for inducing eddy current. For the phosphor bronze mesh (PBM) shielding enclosure (Fig. 1(F)), the mesh sheet (36 μm wire diameter, 325 wires/in.) was laminated on to FR4 plates using epoxy with minimal heat and pressure. Then the assembly and electrical connection was achieved the same as for the CFC shielding enclosure.
To test the shielding effectiveness and the gradient-induced eddy current artifact of various shielding designs, bench tests and MRI experiments were performed.
For all of the bench (outside MRI) tests, a network analyzer (HP 4195A, USA) (Fig. 2(A)) along with custom-built instruments (Figs. 2(C, D)) were used. Then, a 3T MRI (GE MR750, GE, USA) system equipped with the built-in body coil consisting of a liquid-cooled 35 kW solid-state RF power amplifier and a gradient system driven by an amplitude system (amplitude 50 mT/m, slew rate 200 T/m/s) were used for assessing the performance in MRI (Fig 2(B)) along with custom-built instruments (Figs. 2(E)–(G)). Each shielding enclosure was individually placed radially 16 cm away from the MRI isocenter to mimic its realistic position for a 32cm diameter head PET inserted placed inside the MRI bore.
Figure 2.
Experimental materials for bench tests (Fig. 2a: network analzyer, 2c: magnetic field loops for shielding effectiveness measurements, and 2d: magnetic surface probe for eddy current measurement) and MRI experiments (Fig. 2b. 3T MRI, 2e: 127.68-MHz RF source, 2f: magnetic field detection loop, 2g: fluoroptic thermometer) of the different shielding configurations.
RF shielding effectiveness
For testing the shielding effectiveness at 127.76 MHz on the bench, two identical magnetic field loops (Fig. 2(C)) were built; one of the loops was placed inside the shielding enclosure and the other loop was placed outside the enclosure. Each loop was a shielded magnetic field loop, which was only sensitive to RF magnetic fields and insensitive to electric fields. The coil pair was used with a “S21 measurement” (transmission loss) 55 calibrated at 127.76 MHz with bandwidth of 20 MHz with no shield. Each shield is then placed between the 2 coils.
For the MR experiment, two experiments were performed. A 127.68 MHz RF field source was built using a miniature 42.56 MHz clock chip (FOX electronics, FL, USA) powered by a non-magnetic battery (Powerstream, UT, USA). The 42.56 MHz square wave was driven into a series of resonant LC circuits tuned to the 3rd harmonic, which was 127.68 MHz. This RF source was then measured with the 3T MRI system’s body coil receiver tuned with a center frequency at 127.7 MHz with a bandwidth of ±31.25kHz; the experiment began with the 127.68 MHz source unshielded as a reference and then again inside each of the shielded enclosure configurations (Fig. 2(E)). When placing the RF source inside the shielding enclosure, to avoid a frequency peak shift, the source was not placed too close to the shielding planes inside the enclosure. Then, a single magnetic field detection loop (Fig. 2(F)), which has the same design as in Fig 2(C), was placed inside each shielding enclosure and was connected to an oscilloscope for detecting the induced voltage from the RF-intense MRI fast spin echo (FSE) pulse inside the enclosure (Fig 2(D)).
Gradient-induced eddy current investigation
At the bench, the standard eddy current evaluation tests 56 were performed on different shielding enclosure configurations. By using a high inductance magnetic surface probe (25mH) and an impedance analyzer (Fig. 2(D)), we can perform a S11 measurement 55 to analyze the magnitude of eddy currents induced at different frequencies. To test a range of frequencies appropriate for the gradients used in an actual MRI study, the S11 measurement was acquired at a range of frequencies from 0.5 kHz to 10 kHz in steps of 500 Hz. Then, the resolution bandwidth and the source amplitude were set to 100Hz and 15 dBm, respectively, to increase the receive sensitivity and lower the noise. Three sets of S11 data were measured and the mean and standard deviation (as error bars) were plotted respect to the frequency. Shields with less eddy current induce smaller change in coil impedance. This change indirectly indicates the magnitude of the opposing magnetic field due to the induced eddy current.
In the MRI system, two experiments were performed. To test the gradient induced eddy current performance, the gradient-intense EPI sequence (Table 1) images of a 17.5 cm diameter spherical agar phantom were acquired. To assess the eddy current induced ghosting artifact from the EPI MR images, 2.5 cm2 region-of-interest (ROI) in ghosting and background noise and 15 cm2 ROI in signal were measured, Then, the ghost level intensity (GLI) in the images was calculated as following 57:
Table 1.
MR pulse sequence parameters
| Sequence | TR (ms) | TE (ms) | Voxel (mm) | Matrix | Flip angle (°) | Number of average (NEX) | Echo train length (ETL) | Pixel bandwidth (Hz/Pixel) |
|---|---|---|---|---|---|---|---|---|
| FSE | 3000 | 71 | 0.86 × 0.86 × 4.00 | 256 × 256 | 90 | 1 | 1 | 122.1 |
| EPI | 2000 | 30 | 3.44 × 3.44 × 4.00 | 64 × 64 | 90 | 1 | 1 | 7812.5 |
Then, the temperature of the shielding enclosure was steadily measured for 8 minutes using a fluoroptic thermometer (Luxtron 790, LumaSense, CA) (Fig. 2(G)) with the EPI sequence running (Table 1). Four temperature sensors of the fluoroptic thermometer were placed on four side plates of each of the shielding enclosures and the average of the four temperatures were analyzed.
RESULTS
RF shielding effectiveness
The S21 measurements at the bench are shown in Fig. 3. As expected, the solid copper shielding enclosures with 36, 18 and 9 μm thickness showed the best shielding effectiveness of −43.10±1.72, −42.47±3.36, and −42.27±2.62 dB, respectively. The thickness of solid copper did not affect the shielding effectiveness due to the skin depth effect 44. PBM also showed a comparable shielding effectiveness performance of −37.55±3.55 dB. However, the CFC result of −29.15±1.51 dB was not as effective and the segmented copper result of −10.98±0.66 dB showed the worst shielding performance.
Figure 3.
S21 measurement at 127.8 MHz at the bench and FSE pulse detection measurement inside the MRI system
The field attenuation results of the RF pulse voltage readout at 90° and 180° flip angle MR FSE sequence are shown in circles and triangles, respectively (Fig. 3). For both bench and MRI results, the ‘No shield’ case was set as the reference and the relative attenuation levels of bench and MRI results were closely matched.
The RF signal from the 127.7 MHz source inside the various enclosure was detected by the body coil (Fig. 4). When no shield was present, the raw peak value was the maximum as expected. For solid copper shielding enclosures, regardless of the copper thicknesses (36, 18 and 9 μm), almost no RF signals (−26.11±1.18, −26.02±1.17, and −26.18±1.22 dB, respectively) were detected by the body coil. However, when a segmented copper or a CFC was present, some RF fields leaked (−10.89±1.91 and −12.30±2.02 dB, respectively) from inside the shielding enclosures and were detected by the body coil, indicated by the peaks at the 127.7 MHz. The PBM showed −26.08 dB shielding effectiveness, which was comparable to that of the solid copper enclosures at the Larmor frequency.
Figure 4.
Shielding measurements using the 127.7 MHz RF source
Gradient-induced eddy-current artifact investigation
The S11 attenuation magnitude result was significantly higher with the solid copper configurations compared to those of the other configurations (Fig. 5), which were close to the values observed from the “No shield” configuration. As the frequency decreased, the S11 magnitude decreased as well. The results below 1.5 kHz were noisy.
Figure 5.
S11 measurements of the shields from 0.5 kHz to 10 kHz.
The ghosting level intensities of EPI images were acquired for various shielding enclosures (Fig 6). As shown in the plot, EPI images with solid copper enclosures showed high ghosting artifact intensity, and the intensity is higher for thicker copper, while other enclosures showed negligible artifact. The worst ghosting artifact was shown when imaging the axial slice with frequency encode direction of left-right and phase encode direction of anterior-posterior. This is the case where the maximum gradient field is applied on the largest conductive area (top and bottom planes of the shielding enclosures, see Fig. 1). In other cases (sagittal and coronal orientations) where the gradient field was applied to the side plates, the ghosting artifact was not as high as that of the axial ghosting artifacts.
Figure 6.
(Left) Schematics of the shielding enclosure at different orientations with phase encoding and frequency encoding directions. (Right) Ghosting level intensity (%) on EPI images for axial, sagittal and coronal orientations.
Temperature rose when the EPI sequence was running due to the Joule heating from the induced eddy current (Fig. 7). The temperature drift with the 36 μm solid copper resulted in the highest temperature rise rate of 0.29 °C/min, and 18 and 9 μm solid copper enclosures showed minor temperature rise of 0.08 and 0.09 °C/min rate, respectively. Other shielding enclosures showed negligible temperature drift.
Figure 7.
Temperature measurement of the shields while running the EPI pulse sequence.
From the MRI experiments, the segmented copper, CFC and PBM configurations showed negligible eddy current artifacts.
DISCUSSION
The aim of this work was to evaluate the shielding effectiveness and eddy current performance of several shielding configurations inside the MRI. Several shielding designs with various thicknesses, patterns and materials were developed and studies were performed outside and inside the MRI.
Before performing experiments, simple theoretical calculations with ideal assumptions were performed using equations 1 and 2 in order to understand the eddy current induction mechanisms in the shielding. However, as these equations are only valid with these ideal assumptions, the theoretical calculation and experimental measurements can have mismatches. For example, the multilayer design of the PCB with conductors and inductors and the different shielding patterns including the segment, mesh and weave are difficult to model in the theoretical calculation, but these can affect the shielding and eddy current induction performance43,51,58.
The shielding thicknesses (36, 18 and 9 μm) for solid copper designs were thicker than the skin depth (5.8 μm) of copper at 127.76 MHz. Thus, all three were sufficient for shielding the RF field at 127.76 MHz (Figs. 3 and 4). Despite the slightly higher absorption in the thicker copper; the shielding effectiveness is mostly dominated by reflection rather than absorption, so the three RF shielding effectiveness values were not significantly different.
On the other hand, as the shielding thickness and the EM frequency are proportional to the eddy current power loss (Eq. 2), the ghosting artifact and the temperature increase during the gradient-intense EPI sequence were significantly higher for the thicker solid copper compared to the thinner copper (Figs. 6 and 7). On the bench using a network analyzer (Fig. 5), the eddy current experiment was performed in a range of frequencies from 0.5 kHz to 10 kHz expected for actual gradients in an MR study. The eddy current performance was worse (higher S11 magnitude) at higher frequencies due to increased eddy current induction with high EM field frequencies. Below 1.5 kHz, the data was noisy and S11 measurement was at the noise level. This is due to the limitation of the network analyzer with narrow bandwidth and also the result of low eddy current induction at low frequencies. However, most importantly, the segmented copper, CFC and PBM configurations showed significantly lower S11 magnitudes compared to the solid copper configurations.
Therefore, reducing the shielding thickness diminishes the induced eddy current; however, this does not completely remove the artifact, as we have observed slight ghosting artifacts (Fig. 6) and temperature increment (Fig. 7) for thin solid copper enclosures. Instead, as mentioned earlier, there are other methods to remove the induced eddy current. By adding a precise design of segmented or meshed patterns, the current loop area can be reduced while conserving the electrical connectivity for sufficient shielding. The RF shield between the RF coil and the gradient coil inside a typical MRI system is an example of this. Although, the custom-built segmented copper was motivated by these considerations, the shielding effectiveness result was poor in our experiment (Figs. 3 and 4). While the segmentation definitely mitigates the induced eddy current issues, depending on the segmentation pattern, undesirable resonances or RF leak into the slots may result. If the slots in the segment pattern are long, they behave like a “slot aperture”, which affects the shielding effectiveness of the enclosure. The longest slot aperture in our design was ~20 cm, which is ~1/12 of the wavelength. To improve the shielding performance, the longest aperture length should be reduced down to at least 1/20 of the wavelength (~12 cm) for sufficient RF shielding efficiency 58,59.
For an alternative method, CFC was tested as its continuous fibers have been found to be an excellent EM shielding material due to the low surface impedance and high reflectivity in the frequency range from 0.3 MHz to 1.5 GHz 30. There also has been successful attempts in other groups of using the CFC as a shielding material for their PET Inserts 52. Conversely, the result in our experiment showed poor RF shielding performance, which could be due to the structural parameters of the design, such as, the carbon fiber orientation of individual layers, fiber distribution and the number of layers, which can affect the shielding and eddy current performance 60. The current CFC prototype comprises orthotropic laminates utilizing a 0°/90° twill weave; the shielding performance could be enhanced by a quasi-isotropic layup having a −45°/45° weave laminations 61.
Even though significant ghosting artifacts and temperature rise were observed for the solid copper shielding enclosures, one thing to note here is that these are only caused by the gradient-intense high-duty-cycle EPI sequences, but not in other sequences, such as, gradient echo or FSE sequences.
The bench and MRI results of shielding effectiveness were comparable (Fig. 3); nonetheless, the slight mismatches could be due to the dynamic range difference between the network analyzer and the MRI RF receiver; the MRI body coil has a much higher dynamic range compared to that of the network analyzer.
To fully satisfy the MR-compatible requirements of MR-inserted medical devices, all of the electronic circuit parts of the device are required to be carefully examined for MR susceptibility. As the parts inside the enclosure satisfy the MR susceptibility requirements, the appropriate shielding configuration will allow mutual EM interference-free operation inside the MRI.
For potential applications, if more than one shielding enclosure is placed inside the bore, the eddy current performance would proportionally depend on the amount of the integrated amount of conductive materials inside the bore, and also on the orientation of the shielding enclosures.
As the number of medical devices that are inserted into MR systems increases, considerations of proper shielding is a critical factor for simultaneous measurements with an MR system with low mutual interference. We believe some of the studies in this paper will be highly relevant to these goals.
CONCLUSION
The RF shielding effectiveness and the gradient-induced eddy current performance of several shielding configurations inside a 3T MR system were evaluated. Several shielding designs with various thicknesses, patterns and materials were developed and tested inside and outside the MRI. Solid copper shielding enclosures and PBM showed outstanding RF shielding performance, while significant RF leaks were observed in the segmented copper and the CFC configurations. On the other hand, the solid copper shielding designs showed poor eddy current performance and the others showed negligible eddy current artifacts. Therefore, PBM, which shields the RF field effectively while blocking the eddy current induced by the gradient fields, showed the best overall performance. Although we plan to incorporate these findings into enclosure design for a RF-penetrable PET insert system for simultaneous PET/MR 53–53, these results can be applied to any other electronic devices inserted into an MR system.
Acknowledgments
This work was supported in part by NIH-NIBIB grants R01 EB01946501 and F31 EB020504.
Footnotes
Institution information
Molecular Imaging Instrumentation Lab, Stanford University, Alway Building M-001, 300 Pasteur Dr., Stanford, CA 94305
References
- 1.Henderson RG. Nuclear magnetic resonance imaging: a review. Journal of the Royal Society of Medicine. 1983;76(3):206–212. doi: 10.1177/014107688307600312. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.De Wilde JP, Rivers AW, Price DL. A review of the current use of magnetic resonance imaging in pregnancy and safety implications for the fetus. Progress in Biophysics and Molecular Biology. 2005;87(2–3):335–353. doi: 10.1016/j.pbiomolbio.2004.08.010. [DOI] [PubMed] [Google Scholar]
- 3.Logothetis NK. What we can do and what we cannot do with fMRI. Nature. 2008;453(7197):869–878. doi: 10.1038/nature06976. [DOI] [PubMed] [Google Scholar]
- 4.Marcu CB, Beek AM, van Rossum AC. Clinical applications of cardiovascular magnetic resonance imaging. CMAJ : Canadian Medical Association Journal. 2006;175(8):911–917. doi: 10.1503/cmaj.060566. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Jensen MC, Brant-Zawadzki MN, Obuchowski N, Modic MT, Malkasian D, Ross JS. Magnetic Resonance Imaging of the Lumbar Spine in People without Back Pain. New England Journal of Medicine. 1994;331(2):69–73. doi: 10.1056/NEJM199407143310201. [DOI] [PubMed] [Google Scholar]
- 6.Stefaan V, Paul KM. PET-MRI: a review of challenges and solutions in the development of integrated multimodality imaging. Physics in Medicine and Biology. 2015;60(4):R115. doi: 10.1088/0031-9155/60/4/R115. [DOI] [PubMed] [Google Scholar]
- 7.Wehrl HF, Sauter AW, Divine MR, Pichler BJ. Combined PET/MR: A Technology Becomes Mature. J Nucl Med. 2015;56(2):165–168. doi: 10.2967/jnumed.114.150318. [DOI] [PubMed] [Google Scholar]
- 8.Bailey DL, Pichler BJ, Gückel B, Barthel H, Beer AJ, Bremerich J, Czernin J, Drzezga A, Franzius C, Goh V, et al. Combined PET/MRI: Multi-modality Multi-parametric Imaging Is Here. Molecular Imaging and Biology. 2015;17(5):595–608. doi: 10.1007/s11307-015-0886-9. [DOI] [PubMed] [Google Scholar]
- 9.Kolb A, Sauter AW, Eriksson LA, Vandenbrouke A, Liu C-C, Levin CS, Pichler BJ, Rafecas M. Shine-through in PET/MRI: effects of the magnetic field on positron range and subsequent image artifacts. Journal of Nuclear Medicine. 2015 doi: 10.2967/jnumed.114.147637. [DOI] [PubMed] [Google Scholar]
- 10.Histed SN, Lindenberg ML, Mena E, Turkbey B, Choyke PL, Kurdziel KA. Review of Functional/Anatomic Imaging in Oncology. Nuclear Medicine Communications. 2012;33(4):349–361. doi: 10.1097/MNM.0b013e32834ec8a5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.de Rosales RTM. Potential clinical applications of bimodal PET-MRI or SPECT-MRI agents. Journal of Labelled Compounds and Radiopharmaceuticals. 2014;57(4):298–303. doi: 10.1002/jlcr.3154. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Bhandiwad AR, Cummings KW, Crowley M, Woodard PK. Cardiovascular magnetic resonance with an MR compatible pacemaker. Journal of Cardiovascular Magnetic Resonance. 2013;15(1):1–4. doi: 10.1186/1532-429X-15-18. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Kalin R, Stanton MS. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. Pacing Clin Electrophysiol. 2005:28. doi: 10.1111/j.1540-8159.2005.50024.x. [DOI] [PubMed] [Google Scholar]
- 14.Wilkoff BL, Bello D, Taborsky M. Magnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environment. Heart Rhythm. 2011:8. doi: 10.1016/j.hrthm.2010.10.002. [DOI] [PubMed] [Google Scholar]
- 15.Roguin A, Zviman MM, Meininger GR, Rodrigues ER, Dickfeld TM, Bluemke DA, Lardo A, Berger RD, Calkins H, Halperin HR. Modern Pacemaker and Implantable Cardioverter/Defibrillator Systems Can Be Magnetic Resonance Imaging Safe. Circulation. 2004;110(5):475–482. doi: 10.1161/01.CIR.0000137121.28722.33. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Wilkoff BL, Bello D, Taborsky M, Vymazal J, Kanal E, Heuer H, Hecking K, Johnson WB, Young W, Ramza B, et al. Magnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environment. Heart Rhythm. 2011;8(1):65–73. doi: 10.1016/j.hrthm.2010.10.002. [DOI] [PubMed] [Google Scholar]
- 17.Kalin RON, Stanton MS. Current Clinical Issues for MRI Scanning of Pacemaker and Defibrillator Patients. Pacing and Clinical Electrophysiology. 2005;28(4):326–328. doi: 10.1111/j.1540-8159.2005.50024.x. [DOI] [PubMed] [Google Scholar]
- 18.Madigan JD, Choudhri AF, Chen J, Spotnitz HM, Oz MC, Edwards N. Surgical Management of the Patient with an Implanted Cardiac Device: Implications of Electromagnetic Interference. Annals of Surgery. 1999;230(5) doi: 10.1097/00000658-199911000-00005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Duppenbecker PM, Wehner J, Renz W, Lodomez S, Truhn D, Marsden PK, Schulz V. Gradient transparent RF housing for simultaneous PET/MRI using carbon fiber composites. 2012:3478–3480. Oct 27 2012–Nov 3. [Google Scholar]
- 20.Lee BJ, Olcott PD, Key JoH, Grant AM, Chen-Ming C, Levin CS. Studies of electromagnetic interference of PET detector insert for simultaneous PET/MRI. 2013:1–3. Oct. 27 2013-Nov. 2. [Google Scholar]
- 21.Yamamoto S, Watabe H, Kanai Y, Aoki M, Sugiyama E, Watabe T, Imaizumi M, Shimosegawa E, Hatazawa J. Interference between PET and MRI sub-systems in a silicon-photomultiplier-based PET/MRI system. Physics in Medicine and Biology. 2011;56(13):4147. doi: 10.1088/0031-9155/56/13/026. [DOI] [PubMed] [Google Scholar]
- 22.Randal BS, Keyvan F, Yiping S, Paul KM, Joanne T, Paul ES, Steve W, John B, Simon RC. A study of artefacts in simultaneous PET and MR imaging using a prototype MR compatible PET scanner. Physics in Medicine and Biology. 1999;44(8):2015. doi: 10.1088/0031-9155/44/8/312. [DOI] [PubMed] [Google Scholar]
- 23.Grant AM, Lee BJ, Chang C-M, Levin CS. Simultaneous PET/MR imaging with a radio frequency-penetrable PET insert. Med Physics. 2017;44:112–120. doi: 10.1002/mp.12031. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Levin CS, Maramraju SH, Khalighi MM, Deller TW, Delso G, Jansen F. Design features and mutual compatibility studies of the time-of-flight PET capable GE SIGNA PET/MR system. IEEE Trans Med Imaging. 2016;35(8):1907–14. doi: 10.1109/TMI.2016.2537811. [DOI] [PubMed] [Google Scholar]
- 25.Peng BJ, Wu Y, Cherry SR, Walton JH. New shielding configurations for a simultaneous PET/MRI scanner at 7T. Journal of Magnetic Resonance. 2014;239:50–56. doi: 10.1016/j.jmr.2013.10.027. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Maramraju SH, Smith SD, Rescia S, Stoll S, Budassi M, Vaska P, Woody C, Schlyer D. Electromagnetic Interactions in a Shielded PET/MRI System for Simultaneous PET/MR Imaging in 9.4 T: Evaluation and Results. IEEE Transactions on Nuclear Science. 2012;59(5):1892–1899. [Google Scholar]
- 27.Ewing JR, Warner R, Helpern JA. A cylindrically symmetric magnetic shield for a large-bore 3.0 Tesla magnet. Magnetic Resonance in Medicine. 1993;29(3):398–401. doi: 10.1002/mrm.1910290319. [DOI] [PubMed] [Google Scholar]
- 28.Brumwell DA, Perttu JS, Kroll MW, Nelson RS. Shield for implantable cardioverter defibrillator. Google Patents; 1998 [Google Scholar]
- 29.Hassler BA, Donders AP. Implantable medical device with multi-layered ceramic enclosure. Google Patents. 1995 [Google Scholar]
- 30.Chung DDL. Electromagnetic interference shielding effectiveness of carbon materials. Carbon. 2001;39(2):279–285. [Google Scholar]
- 31.Peng BJ, Walton JH, Cherry SR, Willig-Onwuachi J. Studies of the interactions of an MRI system with the shielding in a combined PET/MRI scanner. Physics in Medicine and Biology. 2010;55(1):265–80. doi: 10.1088/0031-9155/55/1/016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Jezzard P, Barnett AS, Pierpaoli C. Characterization of and correction for eddy current artifacts in echo planar diffusion imaging. Magn Reson Med. 1998;39(5):801–12. doi: 10.1002/mrm.1910390518. [DOI] [PubMed] [Google Scholar]
- 33.Schmitt F. The Gradient System. 10th Annual meeting of ISMRM. Volume Weekend Educational Course Syllabus; Honolulu, HI. 2013; pp. 477–486. [Google Scholar]
- 34.Jezzard P, Barnett AS, Pierpaoli C. Characterization of and correction for eddy current artifacts in echo planar diffusion imaging. Magnetic Resonance in Medicine. 1998;39(5):801–812. doi: 10.1002/mrm.1910390518. [DOI] [PubMed] [Google Scholar]
- 35.Wang W, Eisenberg SR. A three-dimensional finite element method for computing magnetically induced currents in tissues. IEEE Transactions on Magnetics. 1994;30(6):5015–5023. [Google Scholar]
- 36.Alecci M, Jezzard P. Characterization and reduction of gradient-induced eddy currents in the RF shield of a TEM resonator. Magnetic Resonance in Medicine. 2002;48(2):404–407. doi: 10.1002/mrm.10226. [DOI] [PubMed] [Google Scholar]
- 37.Fiorillo F. Characterization and measurement of magnetic materials. Academic Press; 2004. [Google Scholar]
- 38.Hahn KD, Johnson EM, Brokken A, Baldwin S. Eddy current damping of a magnet moving through a pipe. American Journal of Physics. 1998;66(12):1066–1076. [Google Scholar]
- 39.Graf H, Lauer UA, Schick F. Eddy current induction in extended metallic parts as a source of considerable torsional moment. Journal of Magnetic Resonance Imaging. 2006;23(4):585–590. doi: 10.1002/jmri.20539. [DOI] [PubMed] [Google Scholar]
- 40.Ahn C, Cho Z. Analysis of eddy currents in nuclear magnetic resonance imaging. Magnetic resonance in medicine. 1991;17(1):149–163. doi: 10.1002/mrm.1910170119. [DOI] [PubMed] [Google Scholar]
- 41.Lee B, Woo W, Park H, Hahm H, Wu J, Kim M. Influence of aspect ratio and skin effect on EMI shielding of coating materials fabricated with carbon nanofiber/PVDF. Journal of materials science. 2002;37(9):1839–1843. [Google Scholar]
- 42.Yang S, Lozano K, Lomeli A, Foltz HD, Jones R. Electromagnetic interference shielding effectiveness of carbon nanofiber/LCP composites. Composites Part A: applied science and manufacturing. 2005;36(5):691–697. [Google Scholar]
- 43.Schulz RB, Plantz V, Brush D. Shielding theory and practice. IEEE Transactions on Electromagnetic Compatibility. 1988;30(3):187–201. [Google Scholar]
- 44.Wheeler HA. Formulas for the skin effect. Proceedings of the IRE. 1942;30(9):412–424. [Google Scholar]
- 45.DeMeester GD, Morich MA, Amor WH. Combined self shielded gradient coil and shimset. Google Patents. 1994 [Google Scholar]
- 46.Amor WH, Alden JS, DeMeester GD, Gruden JL, Ling J. Shim tray with reduced heat conduction and forced cooling. Google Patents. 1998 [Google Scholar]
- 47.Richard MA, Zou X, Morich MA. Low eddy current radio frequency shield for magnetic resonance imaging. Google Patents. 1997 [Google Scholar]
- 48.Weyers DJ, Liu Q. Shielding apparatus for magnetic resonance imaging. Google Patents. 2006 [Google Scholar]
- 49.Frederick PS, Zimmermann WA, Roemer PB General Electric Company, assignee. Double-sided RF shield for RF coil contained within gradient coils used in high speed NMR imaging. Vol. 8. United States: 1994. p. 5. [Google Scholar]
- 50.Weyers DQL. An RF Shield Comparative Study of Different Materials and Types. 2004. p. 1624. [Google Scholar]
- 51.Casey KF. Electromagnetic shielding behavior of wire-mesh screens. IEEE Transactions on Electromagnetic Compatibility. 1988;30(3):298–306. [Google Scholar]
- 52.Düppenbecker PM, Wehner J, Renz W, Lodomez S, Truhn D, Marsden PK, Schulz V. Gradient transparent RF housing for simultaneous PET/MRI using carbon fiber composites. IEEE; 2012. pp. 3478–3480. [Google Scholar]
- 53.Chang C-M, Grant A, Lee B, Levin C. Preliminary PET performance evaluation of an RF field-penetrable brain-sized PET insert for simultaneous PET/MR imaging. J Nucl Med. 2015;56(supplement 3):99–99. [Google Scholar]
- 54.Olcott PD, Kim E, Hong KJ, Lee BJ, Grant AM, Chang C-M, Glover G, SLC Prototype positron emission tomography insert with electro-optical signal transmission for simultaneous operation with MRI. Physics in Medicine and Biology. 2015;60(9):3459. doi: 10.1088/0031-9155/60/9/3459. [DOI] [PubMed] [Google Scholar]
- 55.S-Parameters and Impedance Transformation. Radio Frequency Integrated Circuits and Technologies. Berlin, Heidelberg: Springer Berlin Heidelberg; 2007. pp. 57–78. [Google Scholar]
- 56.García-Martín J, Gómez-Gil J, Vázquez-Sánchez E. Non-Destructive Techniques Based on Eddy Current Testing. Sensors. 2011;11(3):2525. doi: 10.3390/s110302525. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.van der Zwaag W, Marques JP, Lei H, Just N, Kober T, Gruetter R. Minimization of Nyquist ghosting for echo-planar imaging at ultra-high fields based on a “negative readout gradient” strategy. Journal of Magnetic Resonance Imaging. 2009;30(5):1171–1178. doi: 10.1002/jmri.21951. [DOI] [PubMed] [Google Scholar]
- 58.Kraft CH. Modeling leakage through finite apertures with TLM. Aug 22–26, 1994. pp. 73–76. [Google Scholar]
- 59.Celozzi S, Lovat G, Araneo R. Electromagnetic shielding. Wiley Online Library; 2008. [Google Scholar]
- 60.Cheng J, Qiu J, Takagi T, Uchimoto T, Hu N. Numerical analysis of correlation between fibre orientation and eddy current testing signals of carbon-fibre reinforced polymer composites. International Journal of Applied Electromagnetics and Mechanics. 2012;39(1–4):251–259. [Google Scholar]
- 61.Rea S, Linton D, Orr E, McConnell J. Electromagnetic shielding properties of carbon fibre composites in avionic systems. Mikrotalasna revija. 2005;11(1):29–32. [Google Scholar]







