Abstract
An equation is derived for the TG43 geometry function, G(r,θ), of a linear brachytherapy source in terms of its active length. This equation is validated by comparison to published values. It is then used to calculate values of the geometry function for the Model seed, which is a segmented linear source.
PACS number(s): 87.53.–j
Keywords: brachytherapy, geometry function, palladium, TG43
I. DERIVATION
The TG43 formalism 1 expresses the dose rate distribution around a cylindrically symmetric brachytherapy source as separable into component functions.
(1) |
is the air kerma strength, Λ is the dose rate constant, G(r,θ) is the geometry function, and G(r,θ) is the anisotropy function.
The geometry function is normalized to a point 1 cm from the source center along a perpendicular bisector. Its purpose is to quantify the effect of the spatial arrangement of radioactive material on the dose distribution. If the source extent is assumed to be negligible, then the geometry function is modeled by the inverse square relationship. For a finite line source, the geometry function is calculated as follows:
(2) |
β is the angle subtended by the point and the source endpoints and L is the active source length.
The report of TG43 provides a tabulation of the geometry function for a 3‐mm line source and leaves it for the reader to calculate this function for other linear sources. The activity distribution of the Model seed (Theragenics Corporation, Norcross, GA) can be described as a 2.8 mm segmented line source with a 1.0 mm inactive region centered between two active regions of 0.9 mm length2 The geometry function of this source is not readily calculated from Eq. (2). The objective of the present work is to recast Eq. (2) into a function of r, θ, and L and calculate the geometry function of the Model seed. Recasting Eq. (2) requires geometric analysis of the source configuration.
Consider a linear source whose end points are labeled S and F as shown in Fig. 1. The midpoint, M, is located midway between S and F and serves as the origin of the coordinate system. Point P is located in a position which is arbitrary with respect to the source. Using the conventions of TG43, line segments connecting S, M, and F with P form angles of , with the long axis of the source. The magnitude of the angle ∠ SPF is designated as β.
If the line segment connecting M and P has length r, then the position of P can be referred to as P(r,θ). The relationship between , and can be found by recognizing that the sum of the angles in the triangle ΔASFP is it.
(3) |
If we treat the points S and P as diametrically opposed corners of a rectangle, this provides the means to eliminate and from the expression for β. The value of is expressed in terms of r and θ using the trigonometric definitions,
(4) |
(5) |
Thus
(6) |
(7) |
(8) |
The law of sines is used to express β n terms of the independent variables
(9) |
From the property of the sine function
(10) |
(11) |
Since the points S, A, and P form a right triangle, the pythagorean theorem is used to calculate the distance from S to P.
(12) |
Substituting Eq. (8) into Eq. (11),
(13) |
Since β is given in terms of r, θ, and the constant L; G(r, θ) can also be expressed in terms of r, θ, and L.
(14) |
Thus to tabulate the geometry function of a line source in terms of the polar coordinates, r and 4tH, we need only know the length of the source.
II. VALIDATION
The geometry function for the 3‐mm line source presented in TG43's Table V was used as a benchmark against which to validate Eq. (14). The line source geometry function presented in TG43 is presented as multiplied by . In this way, the tabulated value demonstrates the deviation of the behavior of the small line source from that of a true point source. It also presents a slowly varying function for which linear interpolation can be used to determine values intermediate to those presented in the table. It is noteworthy that in the general formalism for dose calculation presented in TG43's Eq. (2), the geometry function is used only as normalized to its value at the reference point. Thus this normalization is present in Table I such that the values presented are . For a true point source, each of the values in such a table would be unity.
Table I.
|
1 cm | 2 cm | 3 cm | 4 cm | 5 cm | 6 cm | 7 cm | 8 cm | 9 cm | ||
---|---|---|---|---|---|---|---|---|---|---|---|
|
1.1071 | 1.0306 | 1.0132 | 1.0100 | 1.0089 | 1.0084 | 1.0081 | 1.0079 | 1.0078 | 1.0077 | |
10° | 1.1023 | 1.0297 | 1.0129 | 1.0099 | 1.0088 | 1.0083 | 1.0081 | 1.0079 | 1.0078 | 1.0077 | |
20° | 1.0889 | 1.0269 | 1.0123 | 1.0096 | 1.0087 | 1.0082 | 1.0080 | 1.0078 | 1.0078 | 1.0077 | |
30° | 1.0694 | 1.0227 | 1.0112 | 1.0091 | 1.0084 | 1.0081 | 1.0079 | 1.0078 | 1.0077 | 1.0076 | |
40° | 1.0471 | 1.0176 | 1.0100 | 1.0086 | 1.0081 | 1.0079 | 1.0077 | 1.0077 | 1.0076 | 1.0076 | |
50° | 1.0252 | 1.0123 | 1.0087 | 1.0080 | 1.0078 | 1.0077 | 1.0076 | 1.0076 | 1.0075 | 1.0075 | |
60° | 1.0059 | 1.0074 | 1.0074 | 1.0075 | 1.0075 | 1.0075 | 1.0075 | 1.0075 | 1.0075 | 1.0075 | |
70° | 0.9912 | 1.0034 | 1.0064 | 1.0070 | 1.0072 | 1.0073 | 1.0073 | 1.0074 | 1.0074 | 1.0074 | |
80° | 0.9819 | 1.0009 | 1.0058 | 1.0067 | 1.0070 | 1.0072 | 1.0073 | 1.0073 | 1.0074 | 1.0074 | |
90° | 0.9788 | 1.0000 | 1.0056 | 1.0066 | 1.0070 | 1.0072 | 1.0072 | 1.0073 | 1.0073 | 1.0074 |
Note that Eq. (14) has a singularity, i.e., a point where its value is indeterminate [0/0], when the angle is zero. The values at this angle were determined by numerically evaluating the limit as the angle approached zero. In other words, a value was used to approximate zero, which was sufficiently small that the value of the geometry function (to the fourth decimal place) did not change when the angle was doubled or halved.
These values are found to be in agreement with the corresponding entries in TG43's Table V to within 0.1% at all points except where the discrepancy is 8.3%. At this point, the result of the present work is in agreement with the correction to TG43 provided by Rivard. 3 Equation (14) is seen to predict results consistent with values known to be accurate.
III. APPLICATION
To calculate the geometry function of the Model seed using Eq. (14), the calculation takes three steps. First, the geometry function is calculated for the 2.8 mm outer length and the result is scaled by the outer length. Second, the geometry function is calculated for the 1.0 mm inner length and the result is scaled by the inner length. Finally, the scaled results are subtracted and normalized to the reference point. The resulting values are presented in Table II.
Table II.
|
1 cm | 2 cm | 3 cm | 4 cm | 5 cm | 6 cm | 7 cm | ||
---|---|---|---|---|---|---|---|---|---|
|
1.1376 | 1.0397 | 1.0171 | 1.0129 | 1.0115 | 1.0109 | 1.0105 | 1.0103 | |
10° | 1.1316 | 1.0384 | 1.0168 | 1.0128 | 1.0114 | 1.0108 | 1.0105 | 1.0103 | |
20° | 1.1146 | 1.0348 | 1.0159 | 1.0124 | 1.0112 | 1.0107 | 1.0104 | 1.0102 | |
30° | 1.0898 | 1.0294 | 1.0146 | 1.0119 | 1.0109 | 1.0105 | 1.0102 | 1.0101 | |
40° | 1.0613 | 1.0228 | 1.0130 | 1.0111 | 1.0105 | 1.0102 | 1.0100 | 1.0099 | |
50° | 1.0329 | 1.0159 | 1.0113 | 1.0104 | 1.0101 | 1.0099 | 1.0099 | 1.0098 | |
60° | 1.0079 | 1.0096 | 1.0097 | 1.0097 | 1.0097 | 1.0097 | 1.0097 | 1.0097 | |
70° | 0.9886 | 1.0044 | 1.0084 | 1.0091 | 1.0093 | 1.0095 | 1.0095 | 1.0096 | |
80° | 0.9764 | 1.0011 | 1.0075 | 1.0087 | 1.0091 | 1.0093 | 1.0094 | 1.0095 | |
90° | 0.9723 | 1.0000 | 1.0072 | 1.0086 | 1.0091 | 1.0093 | 1.0094 | 1.0095 |
Before subtraction, the component geometry functions are scaled by segment length, to reflect a uniform linear activity density. Why these segments should be scaled by length can be understood by recognizing that the normalized geometry function is a multiplicative term in the calculation of dose. For an idealized source with a uniform linear activity density of unity, a unit dose rate constant, no self‐attenuation, and a distance‐independent radial dose function, the dose would be calculated as the normalized geometry function multiplied by length. Since the superposition principle applies to the quantity absorbed dose, the dose delivered by individual source components are added and subtracted as those components are introduced and removed. Since, for a source of uniform linear activity density, the length scales the activity; it also scales the dose and thus the net‐effective normalized geometry function.
The normalized geometry function is one component of the TG43 formalism used to estimate the dose at points surrounding a Model source. Improvements in the accuracy of the geometry function translate into improvements in the estimation of dose. The magnitude of the improvement is position dependent. Improvements in of up to 3% are seen as compared to treating this segmented source as a 3 mm uniform line source. Up to 14% are seen when compared to use of a point source approximation.
References
- 1. Nath Ravinder et al., “Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group 43,” Med. Phys. 22, 209–234 (1985). [DOI] [PubMed] [Google Scholar]
- 2. Fontenia Doracy P. et al., “Diode dosimetry of model 200 seed in water phantom,” Med. Phys. 21, 818 (1984). [DOI] [PubMed] [Google Scholar]
- 3. Rivard Mark J., “Comment on ‘Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group 43,’” Med. Phys. 26, 2514 (1999). [DOI] [PubMed] [Google Scholar]