Abstract
Cross-term spatiotemporal encoding (xSPEN) is a recently introduced imaging approach delivering single-scan 2D NMR images with unprecedented resilience to field inhomogeneities. The method relies on performing a pre-acquisition encoding and a subsequent image read out while using the disturbing frequency inhomogeneities as part of the image formation processes, rather than as artifacts to be overwhelmed by the application of external gradients. This study introduces the use of this new single-shot MRI technique as a diffusion-monitoring tool, for accessing regions that have hitherto been unapproachable by diffusion-weighted imaging (DWI) methods. In order to achieve this, xSPEN MRI’s intrinsic diffusion weighting effects are formulated using a customized, spatially-localized b-matrix analysis; with this, we devise a novel diffusion-weighting scheme that both exploits and overcomes xSPEN’s strong intrinsic weighting effects. The ability to provide reliable and robust diffusion maps in challenging head and brain regions, including the eyes and the optic nerves, is thus demonstrated in humans at 3T. New avenues for imaging other body regions are also briefly discussed.
Introduction
Pulsed field gradients enable nuclear magnetic resonance (NMR) to measure water’s diffusivity under in vivo conditions1–4. By measuring the extent and directionality of this random motion, diffusion-weighted and diffusion-tensor NMR imaging (DWI, DTI) provide unique opportunities for extracting structural insights about tissues in general, and about human organ morphology in particular1,5,6. Indeed the motion of water in tissues is heavily influenced by the latter’s composition, including its density and microstructure, and so mapping its diffusion can illuminate these properties in a non-invasive way. This information is usually extracted using variations of the original Stejskal-Tanner experiment (1), whereby a pair of bipolar gradients ±G d defining the displacements experienced by molecules along these gradients’ direction, are incorporated as independent parameters into a 3D NMR imaging (MRI) sequence. By varying the cumulative effects of these pulsed gradients as defined by a summarizing “b-matrix”1–6, it becomes possible to characterize the full tensorial diffusivity properties. Single-scan methods play a primary role in such in vivo measurements as, in multi-scan MRI acquisitions, motions will generally interfere with the gradients employed in the diffusion- and the position-encoding stages, corrupting the quantitativeness of the experiments7,8. Spin-echo echo-planar imaging (SE-EPI) techniques capable to deliver quality 2D imaging information in a single scan, are therefore widely used in preclinical and clinical diffusion studies9. Despite SE-EPI’s critical role in these studies10–12, this sequence is also prone to artifacts that limit its use to study diffusion in relatively large, homogeneous organs. Particularly deleterious are the effects of external field (ΔB o) and chemical shift heterogeneities, which can introduce severe geometrical distortions and corrupt the diffusivity arising from EPI-derived maps13,14. A number of alternatives based on radial and on rapid-acquisitions with refocused echoes (RARE) strategies, have been proposed for tackling these limitations15–19. Spatiotemporal encoding (SPEN) principles provide yet another alternative to achieve an enhanced immunity to these distortions under similar acquisition, sensitivity and resolution parameters conditions20–23. Spin-packets in SPEN can refocus the dephasing derived from local field inhomogeneities (T2 * effects) throughout the course of the acquisition, rather than at a single time as in conventional spin echoes. It has been shown that this makes out of SPEN a robust tool to map diffusion in preclinical24,25 and clinical settings26,27 as well as in functional MRI28,29, under conditions leading to severe image distortions from SE-EPI. Recently we have introduced an alternative modality dubbed cross-term SPEN (xSPEN30,31), which is not only endowed with these robust refocusing capabilities, but also free from ΔB o- or chemical-shift-driven misplacements. xSPEN provides this unprecedented resilience to field heterogeneities by recruiting –rather than overcoming– field inhomogeneities into the image encoding and decoding processes. To do so xSPEN relies on a continuously acting, linear background gradient, which adds as a source of line broadening to whatever field or shift distortions exist. In combination with frequency-swept pulses imparting a saddle-shaped phase profile onto the spin-packets30,31, a stationary-phase focal point is created that over the course of an acquisition performed under the constant action of the same inhomogeneous broadenings displaces over the object and thereby rasterizes the sought profile –delivering it free from distortions.
This works exploits these unique xSPEN capabilities, to map diffusion in areas that have hitherto proved inaccessible to single-shot studies. This mapping is complicated by the constant gradient employed by xSPEN MRI, which while capable to provide T2 *-free images for a large range of ΔB o inhomogeneities, also imparts a heavy diffusion weighting of its own. Such weightings can overwhelm the effects that the ±G d have to impose for monitoring motions along orthogonal axes, making it hard to obtain full anisotropic details –or even isotropically-weighted information– about water’s diffusivity. In order to perform such full tensorial mapping this study relies on a local b-matrix analysis that unifies the time- and space-dependent effects imposed by both the diffusion and the imaging gradients, and uses it to devise xSPEN-based pulsed-gradient spin-echo (PGSE) schemes that can sample diffusivity over a sufficiently large range of directions. The usefulness of these new tools is demonstrated in a series of preclinical and clinical imaging tests providing diffusivity information in human head and brain regions that are often unreachable by single-shot methods.
Results
xSPEN MRI
To better understand the challenges that xSPEN imaging poses to diffusion measurements, it is convenient to briefly review the features that distinguish this methodology from its SPEN predecessors. SPEN relies on a progressive excitation/inversion and refocusing of the spins, achieved by applying a frequency-swept radio frequency (RF) pulse lasting for a time T e and acting whilst in the presence of an encoding gradient G e 32,33. Whether this RF pulse is used for an excitation or inversion, the result is a spatially parabolic phase profile. Assuming for concreteness swept 180° inversion pulses applied while in the presence of a y-axis gradient, this phase can be written, within an unimportant constant, as
1 |
where R is a sweep rate defined by γG y FOV/T e, and FOV defines the targeted field-of-view along the y-axis being encoded. The quadratic coefficient of the parabolic phase in Eq. [1] defines the spatial extent of the spins emitting at any given moment, as signal emission will be dominated by spins positioned at the apex of this phase parabola. To probe the full FOV this stationary point is displaced, by applying an additional acquisition gradient G a over an acquisition time T a. Extending this 1D rasterization into a single-shot 2D MRI experiment requires encoding a second (e.g., x-) axis, something that is usually achieved by a conventional oscillating readout (RO) gradient. Fourier transform (FT) along the RO axis followed by a magnitude or super-resolved calculation, then delivers the final 2D image20,34,35.
The xSPEN pulse sequence takes Eq. [1] one step further, by melding into it the kind of encoding used in so-called “ultrafast” single-shot 2D NMR spectroscopy36,37. In this experiment the initial spin excitation is combined with two frequency-swept inversion pulses, acting in unison with a pair of bipolar gradients. This replaces the parabolic phase introduced in Eq. [1] by a bilinear phase encoding, proportional to both the chemical shifts Ωi of the targeted sites as well as on the spins’ positions along the axis of the bipolar gradient: , with C = 4T e /FOV a spatiotemporal constant under control and ϕ 0 a position-independent phase. Having imposed such encoding, the application of G a during acquisition leads to the generation of site-specific gradient echoes, enabling the acquisition of arbitrary nD NMR correlations in a single scan. In xSPEN, this spectroscopy-oriented approach is converted onto an imaging one by replacing the chemical shifts Ωis by a spatially-dependent frequency. This inhomogeneous frequency broadening can be imparted, for instance, by activating a constant G z along the slice-select z-axis (Fig. 1a). Considering an unknown term δω(r) coming from field heterogeneities and chemical shifts that adds to this z gradient, leads to an encoding phase profile
2 |
is now defined as a function of the bandwidth BW characterizing the swept RF that is applied while in the presence of the two encoding gradients: , with L z a nominal slice thickness. The hyperbolic encoding phase y.z dominating Eq. [2] has the unique feature that it allows the sum of the “encoding” frequencies to eventually decode a distortion-less y image. In other words, G z and δω become both the encoding mechanism of the experiment, as well as the decoding tool revealing the undistorted positions of the spins along the y-axis –regardless of the size of the field distortion30,31. Figure 1c exemplifies this by presenting results collected on a phantom incorporating a titanium screw and a Lego piece; clearly xSPEN provides more faithful representations of these objects than any of the remaining single-scan counterparts.
Alternative diffusion-monitoring strategies
The application of a continuous G z over the course of both the encoding and the decoding processes, imparts a heavy diffusion weighting into xSPEN MRI. This weighting is further complicated by xSPEN’s progressive spatial inversion/observation of the spins’ throughout the encoding and acquisition processes, which make these diffusion losses dependent on the y position being decoded. Understanding these effects requires a framework capable of computing the diffusion-derived signal attenuation produced by gradients such as G z/δω and G y, as well as accounting for the frequency (i.e., spatially) progressive nature of the RF encoding and readout processes25–27,38. Supporting Information A describes the formalism developed to estimate these effects, and to calculate diffusivity maps in single-shot xSPEN MRI. Based on those analytical and numerical derivations, Fig. 2 describes in further detail the challenges that the usual, variable-G d PGSE scheme, faces in the retrieval of xSPEN diffusivity data. To visualize these challenges the upper row in Fig. 2 depicts the -tensors arising in various DTI schemes, summarizing them in terms of the strength and direction that the eigenvector associated with the largest eigenvalue (b max), exhibits within an imaging-based frame. Notice that whereas in techniques like EPI the weak effects of the imaging gradients allow one to explore the full range of necessary b-values and directions by suitably varying the strengths and directions of a conventional PGSE block, the continuous G z gradient employed in xSPEN strongly biases the range of b-elements and directions that can be sampled. For instance, when considering bipolar diffusion modules whose directions are uniformly spread over the x-y plane, the dominant b-matrix eigenvector ends up subtending a cone in the (x, y, z) sub-space –even though components solely in the first two of these axes should have differed from zero (Fig. 2b, gray). Details of this behavior are further illustrated in Extended Data Figure S1, which describes how these principal eigenvectors –as well as other aspects of the full -tensor– vary for different positions along the imaged axis over the course of an xSPEN acquisition. These strong b-modulations may not impede diffusivity measurements based on xSPEN PGSE schemes if dealing with preclinical scanners, where strong gradients capable of spreading the b-eigenvectors over a sufficiently large range of directions are available. Figure 3 illustrates this with axial slices recorded on an ex-vivo rat head using xSPEN and SE-EPI pulse sequences using the standard PGSE scheme in Fig. 1a. Both SE-EPI and xSPEN methods provide similar Apparent Diffusion Coefficient (ADC) maps for the brains, even if distortions for other regions in the head are noticeable in the EPI (yellow arrows, Fig. 3b). Still, this flexibility will not be available for the ca. ten-fold weaker pulsed gradients available in conventional clinical scanners. Not even the sequence shown in Fig. 1b, incorporating a second bipolar diffusion-weighting block39 placed on the far ends of both inversion pulses and furnishing higher b-values for a given maximal G d strength, will provide a sufficient range of accessible b-values under clinical scanning conditions.
Given these circumstances, we investigated alternative strategies for enabling reliable human diffusivity measurements using xSPEN. To this end an imaging scheme was devised, whereby the low-bandwidth xSPEN dimension was decoded twice, in two independent single-shot experiments that alternated the gradients employed along the RO and SS axes (Fig. 1b, purple labels). As a result of this the dominating b-weight associated with the SS axis also alternated among different (e.g., axial and sagittal) orientations. Figures 2c and 2d illustrate how this leads to two sets of orthogonal conic shapes in the sampling of the -tensor space, rotated by 90° about y. While not involving the comprehensive exploration of the tensorial space that is possible in SE-EPI40, these two orthogonal xSPEN acquisitions lead to a sufficiently wide sampling to reliably measure the diffusion parameters. The lower panels in Fig. 2 demonstrate this with a series of fractional anisotropy (FA) and ADC plots employed to assess the reliability of the different -space sampling schemes. These plots summarize a series of simulations where various 3D media (2601 “tissues”) were assumed, each of them having a random proton density and realistic FA and ADC values (see Supporting Information and figure captions for further details). The signals arising from these synthetic “tissues” under the action of the different gradient schemes and pulse sequences were then calculated, and employed to estimate FA and ADC values on the basis of , where {D ij}i,j=1-3 represent the diffusion tensor elements41. r 2-values were then calculated against the ground truth FAs and ADCs known for the various “tissues”, and from these the reliability of the various approaches was estimated. The most reliable ADC/FA assessments arose in all cases from SE-EPI (Fig. 2a); by contrast, a similar PGSE-based strategy gave unreliable results when incorporated into the original xSPEN sequence (Fig. 2b). Switching to two orthogonal acquisitions where the roles of the RO and SS axes in the xSPEN process are swapped and their outcomes processed in a combined analysis increased this considerably (Fig. 2c); the reliability of this double-cone -space acquisition could be further enhanced by incorporating into the fits, independent b o-samplings (i.e., samplings with all G ds set to null) for each of the imaging schemes (Fig. 2d). The Extended Data Figure S2 illustrates an experimental validation of the resulting b o-including “double-cone” approach, conducted on a water phantom in a clinical 3 T MRI machine, showing essentially the same reliability as EPI-based maps. This latter gradient scheme was adopted for the human ADC and DTI xSPEN mapping.
Diffusion in humans by xSPEN MRI
Figure 4 compares SE-EPI and xSPEN results obtained on two representative human head slices, showing the b o magnitude images, ADC, FA and cFA (color-coded FA) maps to which each experiment leads. For an upper, relatively homogeneous brain slice (Fig. 4a), both methods show similar tissue contrasts and diffusivity information. Still, the FA maps show a slightly stronger contrast for SE-EPI vis-à-vis xSPEN –perhaps reflecting the different ADC mapping accuracies of the two methods as discussed in the context of Fig. 2. For a lower brain slice (Fig. 4b), however, the SE-EPI data suffers from evident inhomogeneity distortions, clearly seen in the sinus regions and in the shapes of the eyes (yellow arrows). Interestingly, these are not only reflected in the SE-EPI b o images but also translate in diffusion map artifacts –for instance, in false anisotropies that SE-EPI describes in the center of the eyes’s vitreous cFA maps.
Figure 5 compares another set of axial, coronal and sagittal human head images, showing SE-EPI and xSPEN ADC maps collected at 4 mm isotropic resolution to probe diffusivity on head regions challenged by field inhomogeneity effects. Some of these regions are highlighted in the Figure by yellow arrows and include (a) the eyes, (b, c and e) the nose and nasal cavity, (d and g) the cerebellum, (f) the brain stem and (h) the tongue area. In SE-EPI these regions are partially or fully distorted and the ensuing ADC maps convey limited useful information, whereas xSPEN makes them clearly accessible. Worth comparing among these sets are the results observed for the vitreous humor, which in the center of the eyes include average ADCs of 3.7 ± 0.2 × 10−3 mm2/s and 3.6 ± 0.4 × 10−3 mm2/s for xSPEN and SE-EPI, respectively. These values are in good agreement with each other, as well as with literature reports42.
Figure 6 shows an additional example collected at a higher, 3 mm isotropic resolution, showing zoomed-in DTI maps of the head’s anterior region. This folding-free zoom-in is possible at no cost in the sequence’s complexity, and it helps to highlight features associated with the frontal lobes and eyes (Fig. 6a). It also shows how xSPEN could help to target the diffusivity of the optical nerves, for which xSPEN FA and ADC maps were collected at a slight tilt (Fig. 6b) that allowed us to observe regions with distinct diffusion values in the center of these nerves. These features include a more mobile core with average ADC and FA values of 1.5 ± 0.3 × 10−3 mm2/s and 0.25 ± 0.09 respectively, and a more slowly diffusing periphery with average ADC and FA values of 1.1 ± 0.3 × 10−3 mm2/s and 0.28 ± 0.10 respectively. Judging by the corresponding T1-weighted images, these data are affected by minimal artifacts despite the strong air-interface ΔB os associated to these tissue regions. These ADC values are in agreement with literature reports based on fast spin-echo and on EPI measurements (0.8–1.4 × 10−3 mm2/s), even if reported FA values for optic nerves have been usually higher (0.39–0.64)43,44. The origin of this variation is under investigation.
Discussion and Conclusions
Different diffusion-monitoring variants have been explored over the years as diffusion-monitoring alternatives to EPI, particularly within the context of overcoming heterogeneous field environments. These include multi-shot schemes liable to motion-derived errors, as well as single-shot acquisitions based on radial samplings or on the use of multiple rf-driven spin echoes15–19. Recently introduced single-shot 2D SPEN methods can achieve many of these goals20,35,45–47, yet if overwhelmed by inhomogeneities even these methods are liable to yield corrupted images reflecting ΔB o displacements. xSPEN by contrast shows unprecedented resilience to in-plane field inhomogeneities, with the sole distortions identified in their single-scan 2D images arising from limitations in the slice selection process31. xSPEN’s immunity to field inhomogeneities stems from its reliance on a constantly active frequency broadening mechanism that incorporates background inhomogeneities into the image-formation process. This use of a continuous gradient throughout the encoding and acquisition, however, results in a strong intrinsic b-weighting that needs to be accounted for when attempting quantitative diffusion measurements. In order to enable such accounting, the sequence’s diffusion effects were quantitatively evaluated using a customized, spatially-localized b-matrix analysis. With the aid of this formalism and of numerical simulations on “synthetic tissues”, a diffusion-weighting scheme was devised that overcomes xSPEN’s original limitations. Such scheme operates by exchanging, in independent measurements, the roles that orthogonal gradients play as the broadening mechanism enabling the xSPEN image formation. Phantom experiments validated the quantitativeness of the resulting “double-cone” b-sampling approach; when the diffusion gradients were not sufficiently intense, a double PGSE block could be readily incorporated into the sequence. When implemented on human volunteers (Figs 4–6) the ensuing double-PGSE, double-cone scheme improved the ADC and FA depiction afforded by SE-EPI, providing reliable maps of isotropic and anisotropic diffusivity for the brain stem, cerebellum, mouth and the ocular regions. Thanks to its reliance on frequency swept pulses xSPEN also provided a built-in “zooming” capability that allowed us to focus on the frontal brain region, where ADC mapping of the optical nerve was demonstrated in all three orientations –always evidencing a good, rounded shape and faithful mapping of the nerves’ diffusivity characteristics to their original locations.
Despite these features, xSPEN still exhibits a number of limitations that remain to be overcome. Foremost among these is its SNR limitation; while we have managed to overcome these for the original SPEN experiment using super-resolution procedures48,49, similar procedures remain to be devised for xSPEN. In their absence, we could not explore higher human spatial resolutions than 3 mm isotropic. An additional limitation of xSPEN compared to other single-shot methods –included its SPEN predecessors– is its higher SAR. This results from xSPEN’s use of two frequency-swept inversion pulses, and it led to SAR reaching ca. 90% of its maximum allowed (normal mode) value when performing the multi-slice experiments illustrated above. Approaches to minimize this problem’s impact are also under investigation. Despite these two main limitations we believe that xSPEN’s ability to target challenging inhomogeneity-dominated regions, of the kind that are usually inaccessible by traditional single-shot methods, can open new opportunities in basic and in clinical investigations. This immunity to field heterogeneities was here demonstrated for various head regions of healthy volunteers; we are exploring what new avenues can be opened when imaging other body regions, as well as tissues proximate to metal implants including spine, mouth and dental MRI.
Methods
Preclinical scans
Animal experiments were preapproved by Weizmann’s Animal Care and Use Committee (protocol 01500312-3) and all procedures were carried in accordance with the guidelines of this IACUC. Weizmann’s animal program is accredited by the Association for Assessment and Accreditation of Laboratory Animal Care (AAALAC), by the US NIH’s Office of Laboratory Animal Welfare, and by Israel’s Ministry of Health. Phantom and ex-vivo whole-head rat experiments were performed on a DD2 7 T/110 mm horizontal magnet scanner (Agilent Technologies, Santa Clara, CA) using a quadrature-coil probe. For these scans the xSPEN pulse sequence shown in Fig. 1a was used and compared against a SE-EPI sequence with similarly structured timing and gradient strengths. These sequences ran in and were processed within Agilent’s VNMRJ 3.2 software environment. For the preclinical anatomical references, scanner-provided fast spin-echo multi-shot (SEMS) sequences were used. Scanning parameters for the Titanium-Lego phantom included TE ≈ 50 ms, T a = 22.0 ms, FOV = 32 × 32 mm2, resolution = 0.5 × 0.5 mm2, 3.0 mm slice. Number of averages: SEMS = 1, SE-EPI = 1 SPEN = 1 and xSPEN = 2. TR: SEMS = 2 s, SE-EPI = 8 s, SPEN = 8 s, xSPEN = 8 s. Additional parameters: SPEN’s encoding bandwidth BW = 18.0 kHz, G y = 1.33 G/cm; xSPEN’s encoding bandwidth BW = 5.8 kHz, G y = 0.21 G/cm, G z = 2.27 G/cm. The DWI parameters for all phantom experiments were δ = 3 ms, ∆ = 12 ms, maximum diffusion gradients G d = 35 G/cm applied in three orthogonal directions leading to a b max = 870 s/mm2, six b max-scaling values (0, 0.25, 0.63, 0.77, 0.89 and 1.0). For the ex vivo DWI experiments, xSPEN and SE-EPI parameters were δ = 5 ms, ∆ = 13 ms, maximum diffusion gradients G d = 26 G/cm applied in three orthogonal directions leading to a b max = 1370 s/mm2, six b max-scaling values (0, 0.25, 0.63, 0.77, 0.89 and 1.0). Common scanning parameters included TR = 5 s, FOV = 32 × 32 mm2, resolution = 0.5 × 0.5 mm2, 3.0 mm slice thickness. Number of averages: SEMS = 2, SE-EPI = 4, xSPEN = 4. Additional parameters: for SE-EPI TE = 40 ms and T a = 22 ms; for xSPEN middle TE = 62 ms, T p = T a/2 = 11.0 ms, encoding bandwidth BW = 3.6 kHz, G y = 0.13 G/cm, G z = 1.42 G/cm.
Clinical scans
Phantom and in vivo human head diffusion maps were acquired at 3 T using a Siemens TrioTIM scanner (Erlangen, Germany) equipped with a 32-channels head coil. Three subjects (males aged 26, 28 and 30) were scanned for improving the optic nerve and ocular globe diffusional statistics. All these experiments were approved by the Internal Review Board WOMC-0091-11 of the Wolfson Medical Center (Holon, Israel) and by Weizmann’s IRB 141-1 protocol; all experiments were performed in accordance with these institutions relevant guidelines and regulations, and images were collected after obtaining suitable informed consents from all the participants. The phantom used consisted of a plastic bottle containing 1.9 L of water with 3.75 g NiSO4*6H2O + 5 g NaCl. For these scans the diffusion xSPEN pulse sequence in Fig. 1b was used, and compared against a double bipolar diffusion SE-EPI sequence39. For the simulations (Figs 2 and S1), the water phantom experiments (Fig. S2) and the lower-resolution human head scans (Figs 4 and 5), the b max used for both xSPEN and SE-EPI had a nominal 1000 s/mm2 value. Additional xSPEN diffusion parameters included δ = 16 ms, ∆ = 57 ms, G d = 3.26 G/cm. xSPEN scanning parameters for these data were: 48 slices, one nominal |b| = b max value, T p = T a/2 = 10.9 ms, G y = 0.027 G/cm, G z = 1.24 G/cm, chirp bandwidth BW = 2.2 kHz, FOV = 185 × 193 mm2 (PExRO), resolution = 4.0 mm3 isotropic, TR = 25 s, SAR = 90%. Number of averages: SE-EPI = 1, xSPEN = 2. Other SE-EPI scanning parameters: TE = 80 ms, TR = 10 s, FOV = 184 × 256 mm2 (PExRO), resolution = 4.0 mm3 isotropic. For the higher, 3.0 mm3 isotropic resolution head scans (Fig. 6) the b max-value for both xSPEN and SE-EPI was 800 s/mm2. Additional xSPEN parameters: δ = 17.2 ms, ∆ = 49.1 ms, G d = 2.95 G/cm, 32 slices, one b-value, TR = 15 s, FOV = 90 × 96 mm2 (PExRO), T p = T a/2 = 7.75 ms, G y = 0.0504 G/cm, G z = 1.52 G/cm, chirp bandwidth BW = 2 kHz, SAR = 90%. Additional scanning parameters: Number of averages: SE-EPI = 1, xSPEN = 4. Other SE-EPI scanning parameters: TE = 84 ms, TR = 10 s, FOV = 180 × 252 mm2 (PExRO), resolution = 3.0 mm3 isotropic. T1- and T2-weighted scans are also included as anatomical references.
Electronic supplementary material
Acknowledgements
We are grateful to Dr. Sagit Shushan (Wolfson Medical Center) and the Weizmann MRI team (Edna Furman-Haran, Fanny Attar and Nachum Stern) for assistance in the human scans, and to Dr. Qingjia Bao (Weizmann) for assistance in the preclinical experiments. ZZ thanks Israel’s Council of Higher Education and to the Koshland Foundation for partial postdoctoral fellowships. Financial support from the Israel Science Foundation grant 795/13, the EU through ERC-2016-PoC grant #751106, Minerva funding (#712277) from the Federal German Ministry for Education and Research, the Kimmel Institute for Magnetic Resonance and the generosity of the Perlman Family Foundation, are also acknowledged.
Author Contributions
E.S., G.L., Z.Z. & L.F. designed research; E.S., G.L., Z.Z. acquired the data; E.S., G.L. contributed new tools and analyzed data; E.S. & L.F. wrote the paper.
Competing Interests
The authors declare that they have no competing interests.
Footnotes
Eddy Solomon and Gilad Liberman contributed equally to this work.
Electronic supplementary material
Supplementary information accompanies this paper at 10.1038/s41598-017-17947-1.
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Callaghan, P. T. Principles of nuclear magnetic resonance microscopy. Corrected edn, (Clarendon Press, 1993).
- 2.Price WS. Pulsed‐field gradient nuclear magnetic resonance as a tool for studying translational diffusion: Part 1. Basic theory. Concepts in Magnetic Resonance Part A. 1997;9:299–336. doi: 10.1002/(SICI)1099-0534(1997)9:5<299::AID-CMR2>3.0.CO;2-U. [DOI] [Google Scholar]
- 3.Price WS. Pulsed‐field gradient nuclear magnetic resonance as a tool for studying translational diffusion: Part II. Experimental aspects. Concepts in Magnetic Resonance Part A. 1998;10:197–237. doi: 10.1002/(SICI)1099-0534(1998)10:4<197::AID-CMR1>3.0.CO;2-S. [DOI] [Google Scholar]
- 4.Stejskal EO, Tanner JE. Spin Diffusion Measurements: Spin Echoes in the Presence of a Time-Dependent Field Gradient. Journal of Chemical Physics. 1965;42:288–292. doi: 10.1063/1.1695690. [DOI] [Google Scholar]
- 5.Basser PJ, Mattiello J, LeBihan D. MR diffusion tensor spectroscopy and imaging. Biophys J. 1994;66:259–267. doi: 10.1016/S0006-3495(94)80775-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Le Bihan D, et al. Diffusion tensor imaging: Concepts and applications. J Magn Reson. 2001;13:534–546. doi: 10.1002/jmri.1076. [DOI] [PubMed] [Google Scholar]
- 7.Norris DG, Driesel W. Online motion correction for diffusion-weighted imaging using navigator echoes: Application to RARE imaging without sensitivity loss. Magn Reson Med. 2001;45:729–733. doi: 10.1002/mrm.1099. [DOI] [PubMed] [Google Scholar]
- 8.Rohde GK, Barnett AS, Basser PJ, Marenco S, Pierpaoli C. Comprehensive approach for correction of motion and distortion in diffusion-weighted MRI. Magn Reson Med. 2004;51:103–114. doi: 10.1002/mrm.10677. [DOI] [PubMed] [Google Scholar]
- 9.Mansfield P. Multi-Planar Image-Formation Using Nmr SpinEchoes. J Phys C Solid State. 1977;10:L55–L58. doi: 10.1088/0022-3719/10/3/004. [DOI] [Google Scholar]
- 10.Mukherjee P, Chung SW, Berman JI, Hess CP, Henry RG. Diffusion tensor MR imaging and fiber tractography: technical considerations. AJNR Am J Neuroradiol. 2008;29:843–852. doi: 10.3174/ajnr.A1052. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Schwarzbauer C, Mildner T, Heinke W, Brett M, Deichmann R. Dual echo EPI–the method of choice for fMRI in the presence of magnetic field inhomogeneities? Neuroimage. 2010;49:316–326. doi: 10.1016/j.neuroimage.2009.08.032. [DOI] [PubMed] [Google Scholar]
- 12.van de Looij Y, Kunz N, Huppi P, Gruetter R, Sizonenko S. Diffusion Tensor Echo Planar Imaging Using Surface Coil Transceiver with a Semiadiabatic RF Pulse Sequence at 14.1T. Magn Reson Med. 2011;65:732–737. doi: 10.1002/mrm.22656. [DOI] [PubMed] [Google Scholar]
- 13.Bernstein, M. A., King, K. F. & Zhou, X. J. Handbook of MRI pulse sequences. (Elsevier Academic, 2004).
- 14.Farzaneh F, Riederer SJ, Pelc NJ. Analysis of T2 limitations and off-resonance effects on spatial resolution and artifacts in echo-planar imaging. Magn Reson Med. 1990;14:123–139. doi: 10.1002/mrm.1910140112. [DOI] [PubMed] [Google Scholar]
- 15.Bammer R, Holdsworth SJ, Veldhuis WB, Skare ST. New methods in diffusion-weighted and diffusion tensor imaging. Magn Reson Imaging Clin N Am. 2009;17:175–204. doi: 10.1016/j.mric.2009.01.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Knoll F, et al. A model-based reconstruction for undersampled radial spin-echo DTI with variational penalties on the diffusion tensor. NMR Biomed. 2015;28:353–366. doi: 10.1002/nbm.3258. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Lee CY, Li Z, Pipe JG, Debbins JP. Turboprop+: enhanced Turboprop diffusion-weighted imaging with a new phase correction. Magn Reson Med. 2013;70:497–503. doi: 10.1002/mrm.24481. [DOI] [PubMed] [Google Scholar]
- 18.Sotiropoulos SN, et al. Advances in diffusion MRI acquisition and processing in the Human Connectome Project. Neuroimage. 2013;80:125–143. doi: 10.1016/j.neuroimage.2013.05.057. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Taviani V, et al. High-resolution diffusion-weighted imaging of the breast with multiband 2D radiofrequency pulses and a generalized parallel imaging reconstruction. Magn Reson Med. 2017;77:209–220. doi: 10.1002/mrm.26110. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Ben-Eliezer N, Shrot Y, Frydman L. High-definition, single-scan 2D MRI in inhomogeneous fields using spatial encoding methods. Magn Reson Imaging. 2009;28:77–86. doi: 10.1016/j.mri.2009.05.026. [DOI] [PubMed] [Google Scholar]
- 21.Chamberlain R, et al. RASER: a new ultrafast magnetic resonance imaging method. Magn Reson Med. 2007;58:794–799. doi: 10.1002/mrm.21396. [DOI] [PubMed] [Google Scholar]
- 22.Chen Y, et al. Partial Fourier transform reconstruction for single-shot MRI with linear frequency-swept excitation. Magn Reson Med. 2013;69:1326–1336. doi: 10.1002/mrm.24366. [DOI] [PubMed] [Google Scholar]
- 23.Schmidt R, Seginer A, Frydman L. Interleaved multishot imaging by spatiotemporal encoding: A fast, self-referenced method for high-definition diffusion and functional MRI. Magn Reson Med. 2016;75:1935–1948. doi: 10.1002/mrm.25742. [DOI] [PubMed] [Google Scholar]
- 24.Solomon E, et al. Major mouse placental compartments revealed by diffusion-weighted MRI, contrast-enhanced MRI, and fluorescence imaging. Proc Natl Acad Sci USA. 2014;111:10353–10358. doi: 10.1073/pnas.1401695111. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Solomon E, Liberman G, Nissan N, Frydman L. Robust diffusion tensor imaging by spatiotemporal encoding: Principles and in vivo demonstrations. Magn Reson Med. 2017;77:1124–1133. doi: 10.1002/mrm.26197. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Solomon E, et al. Overcoming limitations in diffusion-weighted MRI of breast by spatio-temporal encoding. Magn Reson Med. 2015;73:2163–2173. doi: 10.1002/mrm.25344. [DOI] [PubMed] [Google Scholar]
- 27.Solomon E, Shemesh N, Frydman L. Diffusion weighted MRI by spatiotemporal encoding: analytical description and in vivo validations. J Magn Reson. 2013;232:76–86. doi: 10.1016/j.jmr.2013.02.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Ben-Eliezer N, Goerke U, Ugurbil K, Frydman L. Functional MRI using super-resolved spatiotemporal encoding. Magn Reson Imaging. 2012;30:1401–1408. doi: 10.1016/j.mri.2012.05.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Goerke U, Garwood M, Ugurbil K. Functional magnetic resonance imaging using RASER. Neuroimage. 2011;54:350–360. doi: 10.1016/j.neuroimage.2010.08.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Paquin R, Pelupessy P, Bodenhausen G. Cross-encoded magnetic resonance imaging in inhomogeneous fields. J Magn Reson. 2009;201:199–204. doi: 10.1016/j.jmr.2009.09.008. [DOI] [PubMed] [Google Scholar]
- 31.Zhang Z, Seginer A, Frydman L. Single-scan MRI with exceptional resilience to field heterogeneities. Magn Reson Med. 2017;77:623–634. doi: 10.1002/mrm.26145. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Kunz D. Use of Frequency-Modulated Radiofrequency Pulses in Mr Imaging Experiments. Magn Reson Med. 1986;3:377–384. doi: 10.1002/mrm.1910030303. [DOI] [PubMed] [Google Scholar]
- 33.Pipe JG. Spatial Encoding and Reconstruction in Mri with Quadratic Phase Profiles. Magn Reson Med. 1995;33:24–33. doi: 10.1002/mrm.1910330105. [DOI] [PubMed] [Google Scholar]
- 34.Chen L, et al. An aliasing artifacts reducing approach with random undersampling for spatiotemporally encoded single-shot MRI. J Magn Reson. 2013;237:115–124. doi: 10.1016/j.jmr.2013.10.005. [DOI] [PubMed] [Google Scholar]
- 35.Schmidt R, Frydman L. In vivo 3D spatial/1D spectral imaging by spatiotemporal encoding: A new single-shot experimental and processing approach. Magn Reson Med. 2013;70:382–391. doi: 10.1002/mrm.24470. [DOI] [PubMed] [Google Scholar]
- 36.Frydman L, Lupulescu A, Scherf T. Principles and features of single-scan two-dimensional NMR spectroscopy. Journal of the American Chemical Society. 2003;125:9204–9217. doi: 10.1021/ja030055b. [DOI] [PubMed] [Google Scholar]
- 37.Frydman L, Scherf T, Lupulescu A. The acquisition of multidimensional NMR spectra within a single scan. Proc Natl Acad Sci USA. 2002;99:15858–15862. doi: 10.1073/pnas.252644399. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Shrot Y, Frydman L. The effects of molecular diffusion in ultrafast two-dimensional nuclear magnetic resonance. J Chem Phys. 2008;128:164513. doi: 10.1063/1.2890969. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Reese TG, Heid O, Weisskoff RM, Wedeen VJ. Reduction of eddy-current-induced distortion in diffusion MRI using a twice-refocused spin echo. Magn Reson Med. 2003;49:177–182. doi: 10.1002/mrm.10308. [DOI] [PubMed] [Google Scholar]
- 40.Topgaard D. Multidimensional diffusion MRI. J Magn Reson. 2017;275:98–113. doi: 10.1016/j.jmr.2016.12.007. [DOI] [PubMed] [Google Scholar]
- 41.Basser PJ, Jones DK. Diffusion-tensor MRI: theory, experimental design and data analysis - a technical review. NMR Biomed. 2002;15:456–467. doi: 10.1002/nbm.783. [DOI] [PubMed] [Google Scholar]
- 42.Meral I, Bilgili Y. Diffusion changes in the vitreous humor of the eye during aging. AJNR Am J Neuroradiol. 2011;32:1563–1566. doi: 10.3174/ajnr.A2543. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Chabert S, Molko N, Cointepas Y, Le Roux P, Le Bihan D. Diffusion tensor imaging of the human optic nerve using a non-CPMG fast spin echo sequence. J Magn Reson Imaging. 2005;22:307–310. doi: 10.1002/jmri.20383. [DOI] [PubMed] [Google Scholar]
- 44.Jeong HK, et al. Improved diffusion tensor imaging of the optic nerve using multishot two-dimensional navigated acquisitions. Magn Reson Med. 2015;74:953–963. doi: 10.1002/mrm.25469. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Liberman G, Frydman L. Reducing SAR requirements in multislice volumetric single-shot spatiotemporal MRI by two-dimensional RF pulses. Magn Reson Med. 2017;77:1959–1965. doi: 10.1002/mrm.26270. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Marhabaie S, Bodenhausen G, Pelupessy P. The effects of molecular diffusion in spatially encoded magnetic resonance imaging. J Magn Reson. 2016;273:98–104. doi: 10.1016/j.jmr.2016.10.010. [DOI] [PubMed] [Google Scholar]
- 47.Solomon E, et al. Removing silicone artifacts in diffusion-weighted breast MRI by means of shift-resolved spatiotemporally encoding. Magn Reson Med. 2015;75:2064–2071. doi: 10.1002/mrm.25757. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Ben-Eliezer N, Irani M, Frydman L. Super-resolved spatially encoded single-scan 2D MRI. Magn Reson Med. 2010;63:1594–1600. doi: 10.1002/mrm.22377. [DOI] [PubMed] [Google Scholar]
- 49.Seginer A, Schmidt R, Leftin A, Solomon E, Frydman L. Referenceless reconstruction of spatiotemporally encoded imaging data: principles and applications to real-time MRI. Magn Reson Med. 2014;72:1687–1695. doi: 10.1002/mrm.25084. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.