Abstract
Background
Magnetic resonance is a major preclinical and clinical imaging modality ideally suited for longitudinal studies, e.g. in pharmacological developments. The lack of a proven platform that maintains an identical imaging protocol between preclinical and clinical platforms is solved with the construction of an animal scanner based on clinical hard- and software.
Methods
A small animal magnet and gradient system were connected to a clinical MR system. Several hardware components were either modified or built in-house to achieve compatibility. The clinical software was modified to account for the different field-of-view of a preclinical MR system. The established scanner was evaluated using clinical QA protocols, and platform compatibility for translational research was verified against clinical scanners of different field strength.
Results
The constructed animal scanner operates with the majority of clinical imaging sequences. Translational research is greatly facilitated as protocols can be shared between preclinical and clinical platforms. Hence, when maintaining sequences parameters, maximum similarity between pulses played out on a human or an animal system is maintained.
Conclusion
Coupling of a small animal magnet with a clinical MR system is a flexible, easy to use way to establish and advance translational imaging capability. It provides cost and labor efficient translational capability as no tedious sequence reprogramming between moieties is required and cross-platform compatibility of sequences facilitates multi-center studies.
Keywords: High-field MRI, Preclinical MRI, Translational platform
Background
Translational research—which may be defined as: “studies that are designed to address human or animal diseases including development of drugs and treatments but excluding studies carried out for regulatory purposes” [1, 2]—accounts for a large proportion of animal studies carried out annually. The European Union reports that almost 19% of animals used in studies addressed research and development in the fields of human medicine, veterinary medicine and dentistry [3]. Mice and rats were the most widely used animals with 61 and 14%, respectively; studied mainly due to their relative ease of breeding and housing, their similar basic biology and chemistry with humans and the wide availability of excellently characterized genetically engineered strains [4].
Since the adoption of the three Rs (replacement, reduction, refinement) principle [5] into the European Directive 210/63/EU, medical imaging has become a key technique in translational research as “it provides a unique opportunity for studying disease from onset in real time, in a quantitative way and non-invasively. It is the preferred method to monitor disease progression and response to treatment in small-animal models in basic and preclinical science and acts as a bridge between novel discoveries and clinical implementation in patient treatment.” [6] Magnetic resonance imaging (MRI) is the second most frequently used modality in preclinical imaging, accounting for about 23% of all examinations [7].
While small rodent imaging can be carried out on human MRI scanners, dedicated small animal systems display performance benefits such as higher temporal and spatial resolution [8]. However, these dedicated systems usually operate with vendor specific software and require the reimplementation of MR sequences to facilitate translational imaging studies. In addition to the tedious reprogramming of the MRI sequences in another programming environment, this approach is also prone to creating mismatching sequences and consequently creating experimental discrepancies, which reduce translational validity. That this is in fact problematic is obvious from early reports on—e.g. compare [9] for a discussion on errors in T2 measurements due to different sampling strategies and differences in commercial implementations of multi-echo sequences. While there are currently only a few dedicated vendors for preclinical MRI systems, new companies are emerging, hoping to capitalize on the potential market. However, due to cost and complexity, advanced animal imaging sequences may not be readily available on all systems and will differ in their implementation. Shrinking the gap between preclinical and clinical studies, while changing experimental parameters as little as possible, makes transposition of data easier. Thus the development of a dedicated small animal MRI machine using clinical software presents a major step in bridging this gap for truly translational research.
According to Tsui et al. [10], combining a clinical MRI with a preclinical magnet is one way to advance translational research. Major applications envisaged for an imaging platform established in this way are MRI/MRS investigations that require extensive translational work. Most prominent areas are drug discovery, e.g. in tumor research [11], cardiovascular imaging [10] and imaging of the central nervous system [12, 13].
Here, the construction and operation of a unique 9.4 T MRI scanner for small rodents, which is based on a clinical system but connected to a preclinical magnet and gradient coils, is presented. It operates with clinical software and allows execution of sequences compiled for the analogous human MRI scanner family. An initial report of this work has been presented at the 24th International Conference of the Society of Magnetic Resonance in Medicine [14].
Methods
Figure 1 shows a system overview containing the major MR hardware components. The majority of electronic components are from a standard Trio, a Tim System (Siemens Healthcare GmbH, Erlangen, Germany). However, several modifications were required and these are described in more detail below. The user interface is based on the Syngo® platform (Siemens Healthcare GmbH, Erlangen, Germany), and was also adapted to the modified imaging environment. The solution presented is based on modifying software parameters only and can potentially be upgraded to forthcoming new software baselines. Furthermore, it is capable of performing non-proton MR measurements, such as 23Na and 31P and is currently being extended to facilitated parallel transmission.
Hardware components
Magnet
As in-house research takes place on a 9.4 T human MRI machine, in order to facilitate translational studies it was desirable that the same field strength be used for the animal system. A suitable 9.4 T 210 ASZ (Varian, Inc. Palo Alto, USA) magnet was sourced, to fit into an existing animal MRI suite. It has a free bore of 210 mm and was initially configured with a double cryostat filled with liquid helium and liquid nitrogen, respectively. However, it was retrospectively fitted with a pulse tube cryocooler (Cryomech, Inc. Syracuse, USA) that achieves zero boil-off operation and consequently only required that the nitrogen cryostat be filled with helium gas. The original magnet supervision and emergency discharge unit supplied by the magnet vendor was maintained without integrating these into the clinical system. Although integration is technically feasible, the interface between the different supervision units is complex and would not significantly alleviate system handling.
Gradient and shim system
In addition to the inherent demand for high gradient strength and slew rate, the requirements for the gradient insert are mechanical compatibility with the magnet dimensions; an inner diameter suitable for measuring small animals, e.g. rats and mice; and, importantly, compatibility with the clinical gradient amplifiers. Compatibility with the clinical gradient amplifiers was found to be essential, as the gradient amplifiers of a standard clinical system are designed for the strong drive requirements of large volume human gradient coils. However, by disabling single stages of the multiple-stage H-bridge amplifier configuration, it was possible to reduce the maximum output power of the gradient amplifiers. Following these modifications, the current output capability of each gradient amplifier was reduced to a maximum of approx. 320 A—disabling the appropriate amplifier stages can be achieved by jumpers on the control circuitry—making them compatible with commercially available gradient inserts such as the BGA12S (Bruker Cooperation, Billerica, USA) or the SGRAD 205/120/HD/S (Varian, Inc. Palo Alto, USA). The later was chosen for the system described here. Further hardware modifications of the gradient amplifier system were made on the ohmic loss supervision and the maximum pulse length supervision, depending on the applied output current. This was accomplished by modifying the analogue supervision circuitry of the gradient amplifiers and required changing resistor values in the operational amplifier based integration circuits. In addition, the temperature of the gradient insert is monitored using integrated PT100 temperature sensors and a vendor supplied monitoring circuit connected to a novel interface unit that sends a temperature interlock signal to the clinical MR system (Fig. 2). As an additional safeguard, gas discharge units were connected in parallel to the gradient coils in order to protect them in case the gradient amplifiers exceeded the maximum voltage settings (e.g. in case of an emergency electric shutdown).
In the system used, the shim coils integrated into the gradient insert are driven by a standard power supply MXH-5-CO (Resonance Research Inc., Billerica, USA). Each of the five current sources has the capability to drive ± 10 A into the shim coils. A vendor supplied CAN-bus (Controller Area Network—a standardised serial bus system to reduce wire count) to serial interface was used to connect the shim amplifiers to the MR scanner.
RF chain and coils
The RF power amplifiers (barthel HF-Technik GmbH, Aachen, Germany) were specifically designed for the platform described here and consist of a narrowband amplifier for protons and a broadband amplifier for X-nuclei, each capable of delivering 1 kW peak power. The proton amplifier combines the output power of four independent 250 W units—a design option chosen to facilitate later extension of the system for parallel transmit operation. The X-nucleus amplifier covers a frequency range from 50 to 305 MHz, which allows imaging of the most biologically relevant nuclei, such as 13C, 17O, 23Na and 31P. It is important to note that the proton amplifier also covers the 19F frequency of approx. 376 MHz at 9.4 T, albeit with a slightly reduced peak output power. Both power amplifiers were integrated into the clinical MR platform through a CAN bus interface, which was developed by the manufacturer of the amplifiers.
In contrast to human high field scanners, the animal system is equipped with a body-coil. This conventional system setup is appropriate since wavelength effects in small animal systems are less pronounced due to the smaller size of imaged objects. Consequently, it was necessary to develop both the body-coil as well as the corresponding transmit/receive (T/R) switch in house. The T/R unit consists of a transmit quadrature-hybrid, directional couplers for power supervision, PIN diode based T/R switches as well as low noise preamplifiers for the 0° and 90° path, respectively (Fig. 3a, c). The body-coil is a self-shielded, 8-rung, high-pass quadrature birdcage with a free inner diameter of 74 mm (Fig. 3b, d). Capacitors are 4.7 pF series 25 type (Voltronics Corp., Cazenovia, USA), trimmers for tuning, matching and balancing use 25 pF NMA_HW series from the same manufacturer and PIN diodes employed are DH80106-11 N (Cobham plc, Dorset, UK.). It is possible to operate the body-coil either in T/R mode or in transmit (TX) only mode by detuning it with PIN diodes and receiving with local receive only coils. It can also be detuned statically in case local transmitter coils are used for excitation.
Local coils, e.g. surface coil arrays, can be connected via two dedicated coil connectors—using the ODU MAC non-magnetic connector series (ODU GmbH & Co. KG, Mühldorf, Germany)—one on the front and one at the back of the magnet. Each connector supports up to eight receive channels with the front connector being equipped with a proton transmit and the back connector with an additional X transmit channel. The available X-nucleus imaging coil set has been described previously [15]. All coils are detected by their unique coil code and the system automatically sets PIN diode control and sequence parameters according to the connected coil systems, as known from clinical routine.
Animal handling and monitoring
Animal handling includes the table shown in Fig. 1 (Bruker Cooperation, Billerica, USA), which is mechanically interfaced to a rat or mouse animal bed (RAPID Biomedical GmbH, Rimpar, Germany) allowing the animal to be manually slid into the magnet. The animal bed includes a facemask for gas anesthesia, which is connected to a standard veterinarian vaporizer (A.M. Bickford, Inc., New York, USA) and a charcoal filter to remove unused anesthetic.
Animal supervision, allowing the acquisition of ECG, respiration and temperature, uses a commercially available monitoring system (Small Animal Instruments, Inc., New York, USA). The system provides a user configurable trigger signal that is interfaced to the MRI scanner to enable synchronized sequences using the ECG or breathing signals of the animal.
Software modifications
Due to both its use as a translational platform and the complexities involved in reprogramming, an important requirement of the animal MR scanner is that it can be operated with only minor tweaks to the measurement parameters and without having to modify major parts of its software. Ordinarily two major modifications would be required relating to field-of-view (FoV) settings of an animal system compared to a human scanner and hardware component differences. The latter was addressed by disabling software supervision of non-existing components, e.g. the patient bed. FoV settings were amended by introducing a 1:5 scaling of linear dimensions. By doing this, the FoV discrepancy was reduced to an extent that the clinical software worked with the smaller imaging dimensions of the animal system. The scaling was implemented by modifying gradient and shim sensitivities accordingly.
In addition, the system tune-up also required several parameter modifications. The most prominent being, definition of tune-up phantoms used with the animal system, modification of hardware supervision such as minimum flow of cooling water through the gradient insert, and a reduction of the transmit power limits to account for the less potential power amplifies in an animal environment. Modification of parameters is carried out by changing these in the measurement settings file of the clinical MR software.
Experiments
Several experiments were carried out after establishing system validity based on clinical quality assurance (QA) routines to validate the translational workflow. In all cases, compiled sequences were copied on both systems and data acquired using a clinical scanner as well as the animal platform is presented here. For comparative purposes, the basic MR sequences—spin-echo and gradient-echo—were employed as they are the building blocks for all advanced imaging sequences. All adjustments (frequency, reference TX power and shimming) were carried out using the system’s automated workflow.
A set of phantom images was acquired using a standard spin echo sequence with isotropic resolution. Details of the imaging parameters are shown in Table 1. The phantoms employed were a 170 mm diameter spherical plastic phantom filled with doped water (1.25 g NiSO4 + 5 g NaCl per 1000 g distilled water) having a T1 of approximately 350 ms at 3 T and an animal sized phantom (T1 ≈ 560 ms) of 40 mm diameter (0.0444 g MnCl2 + 0.0667 g NaCl per 1000 g distilled water) for 9.4 T. For this sequence, small signal RF output as well as the D/A converter output of the three gradient axes were monitored with an oscilloscope (Tektronix, Inc., Beaverton, USA). All images were acquired with the body-coil of the respective MRI system.
Table 1.
Sequence (System) | TR (ms) | TE (ms) | Slice (mm) | Matrix FoV (mm) | BW (Hz/Pixel) | FA (deg.) | Coil | |
---|---|---|---|---|---|---|---|---|
SE (9.4 T Animal) | 1000 | 10 | 0.5 | 128 × 128 64 × 64 |
221 | 90 | Body-boil | Phantom images (Figure 4) |
SE (3 T Human) | 1000 | 10 | 2.5 | 128 × 128 320 × 320 |
221 | 90 | Body-coil | |
GRE (9.4 T Animal) | 95 | 10 | 1 | 256 × 256 50 × 50 |
260 | 25 | Body-coil | Brain images (Figure 5) |
GRE (9.4 T Human) | 95 | 10 | 2 | 256 × 256 250 × 250 |
260 | 25 | 8ch pTX | |
TSE (9.4 T Animal) | 850 | 7 | 0.7 | 384 × 384 60 × 60 |
260 | 140 | Body-coil + 2 channel RX | Spinal cord (Figure 6) |
EPI (9.4 T Animal) | 3000 | 6.2 | 1 | 64 × 64 60 × 60 |
4882 | 35 | Body coil | EPI (Figure 7) |
Multi-GRE (9.4 T Animal) | 75 | 2.5, 5, 7.5, 10, 12,5 | 1 | 64 × 64 60 × 60 |
500 | 7 | Body coil | B0 hom. (Figure 8) |
AFI (9.4 T Animal) | 750 | 2.72 | 1 | 64 × 64 (60 × 60) |
330 | 60 | Body coil | B1 hom. (Figure 8) |
As a second test-case, images were also acquired with the vendor supplied gradient echo sequence at 9.4 T. Sequence parameters for the acquisition at the same field strength were kept identical, except for resolution and slice thickness, which were adapted to the desired FoV (compare Table 1). Due to the wavelength effect associated with the larger dimensions of the human body the 9.4 T human head acquisitions made use of a B1 + shimming approach [16]. A home-built 8-channel transceiver array was used as T/R coil [17] rather than the quadrature birdcage employed for the rat measurement.
Advanced imaging experiments were carried out on the animal scanner only to validate its performance using state of the art MR sequences. The first investigation images a rat spinal cord using a turbo-spin-echo sequence with gating on the animal breathing. The sequence employed is derived from a clinical T1 weighted spinal cord sequence. Modifications required for 9.4 T small animal imaging were different echo- and repetition time to account for the different field strength, an adaption of the FoV to the animal dimensions and a change of turbo-factor to allow gated acquisition without concatenations. The second analysis uses an echo planar readout for a single shot acquisition of the 40 mm diameter phantom described above to demonstrate gradient performance in the scope of combining clinical and non-clinical hardware. The EPI sequence was derived from a clinical sequence supplied by the vendor and sequence parameters optimized for the small animal system (compare Table 1).
Finally, quantitative performance measures were acquired to evaluate static and RF field homogeneity of the system. Based on the principles described in [18] B0 field maps were generated. The field map data was acquired using a standard multi-echo GRE sequence. Constant phase offsets were removed by subtracting the first echo from all following echoes. Resulting phase difference images were spatially unwrapped using FSL PRELUDE (FMRIB, Oxford, UK) [19]. Quantitative fieldmaps were estimated by linear regression of the unwrapped signal phase difference versus echo time difference. The actual flip angle method as implemented in [20] was used for plotting the flip angle distribution of the animal body coil.
Results
QA measurements
To establish initial operation and in order to perform a “sanity check” the clinical QA protocols of the MRI system were successfully carried out. They include: image orientation, RF amplifier linearity, phantom shimming, gradient cross term compensation, eddy current compensation, gradient delay compensation, gradient sensitivity, image artefact, fat saturation and short term field drift. The long-term field stability test was not carried out as the same magnet had already operated with different scanner hardware prior to being transformed into a translational system and there were no issues with field drifts.
Translational sequence measurement
Phantom images, along with the acquired pulse sequence traces for the first k-space line are shown in Fig. 4. Both images show the homogeneous contrast distribution that is expected for single compartment phantoms. The 3 T phantom image, however, displays some remaining non-uniformity in image intensity. The sequence diagrams directly measured using an oscilloscope show identical timing observable through same locations of excitation pulses (same TE), same RF excitation pulse envelope, and similar gradient waveform and crusher gradients.
In vivo translational imaging
Figure 5 shows a single axial slice of a rat brain and a healthy human volunteer brain obtained using a multi-slice gradient recalled echo sequence, with the translational platform of two 9.4 T MRI systems—the animal system described here and a 9.4 T human system.
Exemplary investigation: rat spinal cord
Figure 6 shows sagittal images of a rat spinal cord focusing on the T8/T9 position, which is located directly above the animal’s lung. To eliminate respiratory motion, the acquisitions were gated using a trigger signal derived from the animal monitoring system. Despite the fact that subcutaneous fat yields a hyperintense signal—the fatty tissue is located directly underneath the receiver surface coil-array—the intervertebral discs are clearly visible.
Performance evaluation: animal system
Figure 7 shows a five slice EPI acquisition of a spherical phantom. The images show a tolerable degree of geometric distortions and a slight Nyquist N/2 ghosting artefact.
In Fig. 8 central slice homogeneity of the static B0 field and the RF B1 field are given. Both show reasonable uniformity comparable with those of other commercial animal systems, e.g. compare [21] for a plot of field homogeneities in a 7 T system.
Discussion
Translational sequence measurement
Interestingly, the 3 T image of a water phantom shows some non-uniformity in image intensity, which is not visible in the 9.4 T animal scanner acquisitions. Based on previous experience in high-field measurements, the ring like patterns appear to be wavelength effects caused by constructive and destructive interferences of the RF fields. Consequently, the spin-echo acquisition was repeated using an oil phantom (Bayol oil with a dielectric constant εr ≈ 2.4 [22]). The acquired image is shown in Fig. 9. The ring-like intensity modulation is no longer present which clearly validates the wavelength assumption. This was reconfirmed by measuring the dielectric constant of both water based phantoms using a dielectric assessment kit (Schmid & Partner Engineering AG, Zurich, Switzerland). The dielectric constant was found to be 79 in both cases and in agreement with results from literature [23]. This results in the larger 3 T phantom having a diameter of approximately 0.63 wavelengths and the smaller 9.4 T animal phantom having a diameter of 0.48 wavelengths. Thus, although operating at a lower frequency, the 3 T phantom is electrically larger than the 9.4 T animal sized phantom.
The gradient waveforms differ slightly due to the different electrical properties of the gradient coils (slew rate and maximum amplitude) as well as different physical FoVs. The same holds for the amplitude of the RF signal as the body-coil of the clinical systems requires significantly higher drive power to achieve the selected flip angle compared to the small diameter but high fill-factor animal system setup.
Translational imaging result
The images show the expected different contrasts since longitudinal and transversal relaxation times between animals and humans differ, despite measuring at the same main magnetic field strength. An explanation may be due to differences in the structural organisation of human and animal brains [24] resulting in different MR properties. For reference, Table 2 shows numeric values for brain grey and white matter relaxation times at 9.4 T assembled from the literature [25, 26].
Table 2.
Grey matter | White matter | |||
---|---|---|---|---|
T1 (ms) | T2 (ms) | T1 (ms) | T2 (ms) | |
Human | 2002 ± 105 | 35 ± 3 | 1425 ± 48 | 29 ± 2 |
Animal | 2097 ± 68 | 42 ± 1.6 | 1660 ± 79 | 37 ± 2 |
General discussion
General applicability and operational reliability was demonstrated with the spinal cord acquisition of a rat in vivo using a turbo-spin echo sequence, which is considered to be a workhorse in clinical imaging protocols [27]. Other studies that have been carried out with the system described here include diffusion measurements [28, 29], phase contrast investigations [30] as well as X-nucleus experiments [15].
An extension of the preclinical scanner currently being implemented is an extension into parallel transmission capability. Driving the available four power amplifiers modules for the 1H frequency independently will enable parallel excitation experiments. The envisaged applications include sequence optimization for the 9.4 T human scanner as well as intrinsic parallel-transmit applications, e.g. anatomy-specific excitation to reduce image encoding needs, arbitrary region of interest chemical shift imaging or selective spin tagging. The use of parallel-transmit to mitigate B1 inhomogeneities induced by wavelength effects is not envisaged, or required, on the animal 9.4 T scanner.
It should be noted that a 7 T system with a clinical interface has been commercially available but has been discontinued, so that the presented animal MR scanner is the only truly translational platform available at the moment. Also the commercial 7 T system required reprogramming of the software for each release of a new clinical software baseline which could be avoided with the approach presented here. Finally, the authors have already presented X-nuclei capability of their system which has not been feasible with the prior system.
Conclusions
The successful use of a preclinical MRI scanner integrated with a clinical user interface has been demonstrated. With the translational platform of this machine and standard clinical MRI scanners, multicenter studies become easier as pre-clinical and clinical work can be carried out at different locations but with identical pulse sequences [31]. This is a crucial advantage as multiple sites are now able to share the financial burden of high field systems. Additionally, it facilitates the translation back into today’s clinical field strength, as the platform is also compatible with existing 1.5 T or 3 T installations.
Authors’ contributions
JF developed the animal platform, built several hardware components carried out the translational measurements and wrote the manuscript; AAC built the animal body coil and several mechanical parts; CHC carried out integration of all x-nucleus components and their validation; SS carried out the QA evaluation. NJS conceived the system, coordinated and oversaw the study. All authors read and approved the final manuscript.
Acknowledgements
The authors thank A. Potthast and U. Fontius for numerous tips and tricks on debugging MR hard- and software as well as S. Widmaier, H. Lehr and R. Dolde who shared their experience with the ClinScan® system. Animal handling and care by M. Schöneck and help in acquiring the static field maps by M. Schwerter are gratefully acknowledged.
Competing interests
NJS, AAC and JF are co-founders of Affinity Imaging GmbH.
Availability of data and materials
The datasets generated and/or analyzed during the current study are not publicly available due data protection regulations in the clinical trial consent form but are available from the corresponding author on reasonable request.
Consent for publication
Not applicable.
Ethics approval and consent to participate
In the animal studies presented, all institutional and national guidelines for the care and use of laboratory animals were observed. The Landesumweltamt NRW was notified about all studies involving animals.
In vivo human studies were carried out within the framework of a clinical trial approved by the state authorities as well as a local ethics committee (Ethik-Kommission der RWTH Aachen), under observation of the Declaration of Helsinki and after having obtained written consent. The clinical trial is registered under CIV-11-05-000364 in the EUDAMED database. Consent to publish for anonymized images has been obtained.
Funding
The authors gratefully acknowledge funding by the German Federal Ministry of Education and Research under Grant Number 13N9121 as well as financial support by Siemens Healthcare GmbH. Neither institution had any influence on the study design or on collection, analysis, and interpretation of data and in writing the manuscript.
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abbreviations
- AFI
actual flip angle imaging
- CAN
controller area network
- EPI
echo-planar imaging
- FoV
field-of-view
- MR
magnetic resonance
- MRI
magnetic resonance imaging
- PIN
positive–intrinsic–negative
- QA
quality assurance
- RF
radio frequency
- TR
transmit–receive
- TX
transmit
Contributor Information
Jörg Felder, Phone: +49 2461 61-1908, Email: j.felder@fz-juelich.de.
A. Avdo Celik, Email: a.celik@fz-juelich.de.
Chang-Hoon Choi, Email: c.choi@fz-juelich.de.
Stefan Schwan, Email: s.schwan@fz-juelich.de.
N. Jon Shah, Email: n.j.shah@fz-juelich.de.
References
- 1.Great B, Home O. Annual statistics of scientific procedures on living animals Great Britain 2015 2016. https://nls.ldls.org.uk/welcome.html?ark:/81055/vdc_100034090191.0x000001. [DOI] [PubMed]
- 2.Rubio DM, Schoenbaum EE, Lee LS, Schteingart DE, Marantz PR, Anderson KE, et al. Defining translational research: implications for training. Acad Med. 2010;85(3):470–475. doi: 10.1097/ACM.0b013e3181ccd618. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Commision E. Commission staff working document accompanying document to the report from the commission to the council and the European parliament seventh report on the statistics on the number of animals used for experimental and other scientific purposes in the member states of the European Union SWD/2013/0497 final: directorate-general for the environment; 2013. http://eur-lex.europa.eu/legal-content/EN/TXT/?uri=SWD:2013:0497:FIN.
- 4.Brockmann MA, Kemmling A, Groden C. Current issues and perspectives in small rodent magnetic resonance imaging using clinical MRI scanners. Methods. 2007;43(1):79–87. doi: 10.1016/j.ymeth.2007.07.001. [DOI] [PubMed] [Google Scholar]
- 5.Russell WMS, Burch RL. The principles of humane experimental technique. London: Methuen; 1959. p. 238. [Google Scholar]
- 6.Cunha L, Horvath I, Ferreira S, Lemos J, Costa P, Vieira D, et al. Preclinical imaging: an essential ally in modern biosciences. Mol Diagn Ther. 2014;18(2):153–173. doi: 10.1007/s40291-013-0062-3. [DOI] [PubMed] [Google Scholar]
- 7.Comley J. In vivo preclinical imaging: an essential tool in translational research. Drug Discov World. 2011;12(Summer):58–71. [Google Scholar]
- 8.Jakob P. Small animal magnetic resonance imaging: basic principles, instrumentation and practical issue. In: Kiessling F, Pichler JB, editors. Small animal imaging: basics and practical guide. Berlin: Springer Berlin Heidelberg; 2011. pp. 151–164. [Google Scholar]
- 9.Poon CS, Henkelman RM. Practical T2 quantitation for clinical applications. J Magn Reson Imaging. 1992;2(5):541–553. doi: 10.1002/jmri.1880020512. [DOI] [PubMed] [Google Scholar]
- 10.Tsui BM, Kraitchman DL. Recent advances in small-animal cardiovascular imaging. J Nucl Med. 2009;50(5):667–670. doi: 10.2967/jnumed.108.058479. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Moffat BA, Galban CJ, Rehemtulla A. Advanced MRI: translation from animal to human in brain tumor research. Neuroimaging Clin N Am. 2009;19(4):517–526. doi: 10.1016/j.nic.2009.08.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Fox GB, Chin CL, Luo F, Day M, Cox BF. Translational neuroimaging of the CNS: novel pathways to drug development. Mol Interv. 2009;9(6):302–313. doi: 10.1124/mi.9.6.6. [DOI] [PubMed] [Google Scholar]
- 13.Sakoglu U, Upadhyay J, Chin CL, Chandran P, Baker SJ, Cole TB, et al. Paradigm shift in translational neuroimaging of CNS disorders. Biochem Pharmacol. 2011;81(12):1374–1387. doi: 10.1016/j.bcp.2010.12.029. [DOI] [PubMed] [Google Scholar]
- 14.Felder J, Choi CH, Schwan S, Celik AA, Yun SD, da Silva NA, et al., editors. 9.4 T animal scanner for translation research with binary compatibility to human scanner and clinical UI. In: Proceedings 24th scientific meeting, international society for magnetic resonance in medicine; 2016 May. Singapore: Proceedings 24th scientific meeting, international society for magnetic resonance in medicine.
- 15.Choi CH, Ha Y, Veeraiah P, Felder J, Mollenhoff K, Shah NJ. Design and implementation of a simple multinuclear MRI system for ultra high-field imaging of animals. J Magn Reson. 2016;273:28–32. doi: 10.1016/j.jmr.2016.10.007. [DOI] [PubMed] [Google Scholar]
- 16.Hoffmann J, Shajan G, Scheffler K, Pohmann R, editors. RF Shimming Capabilities at 9.4 Tesla using a 16-channel Dual-Row Array. In: Proceedings 21st scientific meeting, international society for magnetic resonance in medicine; 2013 April 20–26. Salt Lake City: Proceedings 21st scientific meeting, international society for magnetic resonance in medicine.
- 17.Felder J, Hong S, Celik A, Park J, Geschewski F, Jeong H, et al., editors. RF coil array for accelerated excitation in three dimensions. In: Proceedings 20th scientific meeting, international society for magnetic resonance in medicine; 2012 May. Melbourne: Proceedings 20th scientific meeting, international society for magnetic resonance in medicine.
- 18.De Graaf RA. In vivo NMR spectroscopy: principles and techniques, 2nd edn. Chichester: John Wiley & Sons; 2007. xxi, p. 570, p. 8 of plates p.
- 19.Jenkinson M. Fast, automated, N-dimensional phase-unwrapping algorithm. Magn Reson Med. 2003;49(1):193–197. doi: 10.1002/mrm.10354. [DOI] [PubMed] [Google Scholar]
- 20.Brenner D, Vahedipour K, Stöcker T, Shah NJ, editors. Gradient cycled actual flip angle imaging (GC-AFI). In: Proceedings 26th annual scientific meeting of the European society for magnetic resonance in medicine and biology; 2009 October 1–3. Antalya, Turkey.
- 21.Wehrl HF, Judenhofer MS, Thielscher A, Martirosian P, Schick F, Pichler BJ. Assessment of MR compatibility of a PET insert developed for simultaneous multiparametric PET/MR imaging on an animal system operating at 7 T. Magn Reson Med. 2011;65(1):269–279. doi: 10.1002/mrm.22591. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Tofts PS, Barker GJ, Dean TL, Gallagher H, Gregory AP, Clarke RN. A low dielectric constant customized phantom design to measure RF coil nonuniformity. Magn Reson Imaging. 1997;15(1):69–75. doi: 10.1016/S0730-725X(96)00220-2. [DOI] [PubMed] [Google Scholar]
- 23.Meissner T, Wentz FJ. The complex dielectric constant of pure and sea water from microwave satellite observations. IEEE Trans Geosci Remote Sens. 2004;42(9):1836–1849. doi: 10.1109/TGRS.2004.831888. [DOI] [Google Scholar]
- 24.Deelchand DK, Van de Moortele PF, Adriany G, Iltis I, Andersen P, Strupp JP, et al. In vivo 1H NMR spectroscopy of the human brain at 9.4 T: initial results. J Magn Reson. 2010;206(1):74–80. doi: 10.1016/j.jmr.2010.06.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.de Graaf RA, Brown PB, McIntyre S, Nixon TW, Behar KL, Rothman DL. High magnetic field water and metabolite proton T1 and T2 relaxation in rat brain in vivo. Magn Reson Med. 2006;56(2):386–394. doi: 10.1002/mrm.20946. [DOI] [PubMed] [Google Scholar]
- 26.Zhu J, Klarhöfer M, Santini F, Scheffler K, Bieri O, editors. Relaxation measurements in brain tissue at field strengths between 0.35T and 9.4 T. In: Proceedings 22nd scientific meeting, international society for magnetic resonance in medicine; 2014 May. Milan: Proceedings 22nd scientific meeting, international society for magnetic resonance in medicine.
- 27.Kraff O, Quick HH. 7T: Physics, safety, and potential clinical applications. J Magn Reson Imaging. 2017;46(6):1573–1589. doi: 10.1002/jmri.25723. [DOI] [PubMed] [Google Scholar]
- 28.Bastiani M, Oros-Peusquens AM, Seehaus A, Brenner D, Mollenhoff K, Celik A, et al. Automatic segmentation of human cortical layer-complexes and architectural areas using ex vivo diffusion MRI and its validation. Front Neurosci. 2016;10:487. doi: 10.3389/fnins.2016.00487. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Roebroeck A, Oros-Peusquens AM, Brenner D, Moellenhoff K, Celik A, Felder J, et al., editors. Human cortical layers detected with diffusion MRI at 9.4 T. In: Proceedings 18th annual meeting of the organization for human brain mapping; 2012 June. Beijing: Proceedings 18th annual meeting of the organization for human brain mapping.
- 30.Oros-Peusquens A, Hirsch S, Felder J, Celik A, Cremer M, Shah N, editors. Phase contrast in the post mortem rat brain: Comparison with T2* and histology. In: Proceedings 17th scientific meeting, international society for magnetic resonance in medicine; 2009 Apri. Honolulu: Proceedings 17th scientific meeting, international society for magnetic resonance in medicine.
- 31.Mason GF, Krystal JH. MR spectroscopy: its potential role for drug development for the treatment of psychiatric diseases. NMR Biomed. 2006;19(6):690–701. doi: 10.1002/nbm.1080. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The datasets generated and/or analyzed during the current study are not publicly available due data protection regulations in the clinical trial consent form but are available from the corresponding author on reasonable request.