Abstract
Purpose
The purpose of this study was to assess the dosimetric equivalence of magnetic resonance (MR)-generated synthetic CT (synCT) and simulation CT for treatment planning in radiotherapy of rectal cancer.
Methods
This study was conducted on eleven patients who underwent whole-body PET/MR and PET/CT examination in a prospective IRB-approved study. For each patient synCT was generated from Dixon MR using a model-based method. Standard treatment planning directives were used to create a four-field box (4F), an oblique four-field (O4F) and a volumetric modulated arc therapy (VMAT) plan on synCT for treatment of rectal cancer. The plans were recalculated on CT with the same monitor units (MUs) as that of synCT. Dose-volume metrics of planning target volume (PTV) and organs at risk (OARs) as well as gamma analysis of dose distributions were evaluated to quantify the difference between synCT and CT plans. All plans were calculated using the analytical anisotropic algorithm (AAA). The VMAT plans on synCT and CT were also calculated using the Acuros XB algorithm for comparison with the AAA calculation.
Results
Medians of absolute differences in PTV metrics between synCT and CT plans were 0.2%, 0.2% and 0.3% for 4F, O4F and VMAT respectively. No significant differences were observed in OAR dose metrics including bladder V40Gy, mean dose in bladder, bowel V45Gy and femoral head V30Gy in any techniques. Gamma analysis with 2%/2mm dose difference/distance to agreement criteria showed median passing rates of 99.8% (range: 98.5 to 100%), 99.9% (97.2 to 100%), and 99.9% (99.4 to 100%) for 4F, O4F and VMAT, respectively. Using Acuros XB dose calculation, 2%/2mm gamma analysis generated a passing rate of 99.2% (97.7 to 99.9%) for VMAT plans.
Conclusion
SynCT enabled dose calculation equivalent to conventional CT for treatment planning of 3D conformal treatment as well as VMAT of rectal cancer. The dosimetric agreement between synCT and CT calculated doses demonstrated the potential of MR-only treatment planning for rectal cancer using MR generated synCT.
Introduction
Colorectal cancer is the third most common cancer in American men and women. It is estimated that 39,910 new cases of rectal cancer will be diagnosed in United States in 2017 [1]. While colorectal cancer incidence in USA is declining overall, the cancer is increasing rapidly among young adults, and nearly one-third of rectal cancer patients are younger than age 55 [2]. Radiation therapy (RT) is an important adjunct in the treatment of rectal cancer [3, 4]. Preoperative chemoradiotherapy is considered as the standard of care for patients with stage II and III rectal cancer. Short-course preoperative radiotherapy appears to be effective in tumor control and reduce the risk of local recurrence [5].
Magnetic resonance imaging is the most accurate tool for the local staging of rectal cancer due to superior soft tissue visualization [6, 7]. Meanwhile, MR has shown great prognostic power for treatment outcome in rectal cancer [7–9], and become a powerful tool in selection of appropriate treatment [10]. Although MR is the modality of choice for rectal cancer, it has been rarely used solely in its radiotherapy workflow. Conventional radiotherapy employs computed tomography (CT) for treatment planning due to the geometrical accuracy and direct correlation with tissue electron density of CT images. To utilize MR for disease assessment and tumor delineation in RT, MR images are first registered to CT simulation images, which is susceptible to systematic uncertainties [11]. Recently, the MR-only radiotherapy is gaining momentum with emerging MR-LINAC technology [12–14]. The MR-only treatment planning can avoid potential error in MR-to-CT registration, and spare the cost and radiation from dedicated CT simulation. The emerging hybrid MR and linear accelerator technique [15] also motivate MR-only treatment planning to simplify the process of MR guidance in RT.
A CT image directly provides a map of electron density that is necessary for accurate dose calculation [16]. However, there is no simple conversion from MR intensity to electron density value. To address the issue, techniques that generate synthetic CT (synCT) images from MR data are being developed [14, 17], and have shown promising results in treatment planning including brain [18], liver [19], prostate [20], and pelvic tumors [21]. These methods typically can be classified as atlas-based and classification-based approaches. Atlas-based methods search a template MR from an atlas of paired MR and CT data that optimally aligns with patient MR images, and then use the corresponding atlas CT for dose calculation [17]. These methods still rely on image registration, and ignore potential differences in anatomy and electron densities between patients. Classification-based methods use segmentation approach to MR images into several tissue classes and assign bulky densities to form a pseudo-CT image [22, 23]. However, these methods have difficulty in differentiating between bone and air as both air and bone show low signal in MR images.
A hybrid method that combines image classification and atlas delineation [24] recently has been evaluated to generate synCT from MR for treatment planning in lung cancer [25]. In this study, we investigated the potential of the hybrid method to create synCT for MR-based treatment planning of rectal cancer for both 3D conformal RT (3DCRT) and volumetric modulated arc therapy (VMAT). We hypothesized that the MR-generated synCT could provide equivalent dose calculation as simulation CT in colon and rectum cancer where substantial tissue inhomogeneities are present.
Methods and materials
Patient imaging
Under an Institutional Review Board (IRB)-approved protocol, PET/CT and subsequent whole-body PET/MR examination were performed for a clinical imaging study. The data for this work were retrospectively obtained from eleven patients (9 female, 2 male, age range: 25–75 years old) in the study. All image data was de-identified by anonymization and analyzed retrospectively.
The CT data were acquired on a PET/CT scanner (Biograph mCT; Siemens AG Healthcare) according to the standard clinical protocol. The free-breathing CT was acquired using a source of 100-140kVp for a volume with a pixel size of 1.52 × 1.52 mm2 or 1.37 × 1.37 mm2 and a slice thickness of 5.0 mm. The patients were subsequently scanned on a 3T whole-body PET/MR system (Biograph mMR; Siemens Healthcare) that integrates a PET detector with z-direction field of view of 25.8 cm in MR coils for hybrid imaging. In this study, the Dixon 2-point, 3-dimensional volume-interpolated (VIBE) sequence images [26] (TR/TE/Flip Angle: 3.6ms/2.46ms/10°) were acquired during patient breath-hold. The acquisition lasted for 19 sec and generated a 128-slice volume with 192 × 126 pixels in the coronal direction (resolution: 2.60 × 2.60 mm, slice thickness: 3.12mm).
Synthetic CT generation
The synCT was generated from the Dixon MR images using a model-based method that was developed to create linear attenuation coefficients (LACs) map from MR for PET attenuation correction [24]. Detail of this method can be found elsewhere [24]. Briefly, the Dixon MR images were first segmented into four tissue classes of air, fat, lung and soft tissue using a soft-tissue segmentation algorithm [27]. Subsequently, the method [24] updated bone structure by using an atlas-based method that registered patient MR images to an atlas. The atlas consisted of a set of aligned MR image and bone mask pairs. Instead of binary bone type, the bone masks were voxels of actual bone intensities. The bone mask paired with atlas MR that optimally matched with patient MR was transferred back to patient image space based on the registration, thereby, added bone into the tissue class map to form final LAC map.
The LAC map was at the PET energy level of 511 keV. To create synCT image from a standard CT simulator at 140 kVp, voxel-by-voxel mapping of the LAC image to a synCT was performed by using a linear conversion [28]: LAC = 0.096 × (1 × 10−3 × HU + 1) for HU ≤ 0; otherwise LAC = 0.096 × (1 + 6.40 × 10−4 × HU), where LAC is in unit of cm-1. The tissue type air, fat, and soft tissue had HU numbers of -1000, -110 and 70, respectively; and the bone that was transferred from a bone mask had HU in the range of 70 to 2661 [25].
Data preparation
To compare synCT and CT for treatment planning, synCT and CT images were expected to be perfectly aligned. We first performed an affine registration between the whole-body synCT and CT, and formatted both images into axial direction that is default setting for simulation CT. Subsequently, we cropped both volumes to the anatomy range typically requested by physicians in simulation for rectal cancer that covered from the interspace between vertebrate L3 and L4 to the slice below the perineum. The cropped volumes were then interpolated to be synCT and planning CT with an in-plane resolution of 1.25 × 1.25 mm2 and a slice thickness of 2.5 mm as required for RT planning. Finally, a deformable registration was performed between synCT and CT using the mutual information-based image registration in Velocity software (Velocity 3.2, Varian Medical Systems).
Treatment planning
Both registered synCT and CT were imported into a commercial treatment planning system Eclipse version 13.7 (Varian Medical Systems, Palo Alto, CA). Organs at risk (OARs) including small bowel, bladder, and femoral head were contoured for each patient on overlapped synCT and CT view. Body contour generated automatically from synCT in Eclipse was used as a common body structure for both synCT and CT images. A radiation oncologist specialized in gastro-intestinal disease contoured a hypothetical clinical target volume (CTV) in rectal region, representing typical rectal tumor. A planning target volume (PTV) was then created by adding 1 cm margin uniformly to the CTV.
Commonly-used treatment techniques including both 3DCRT and VMAT were planned on synCT to treat the rectal target volume. The planning used a typical clinical protocol that prescribed 50.4 Gy to PTV with a fractional dose of 1.8 Gy in 28 fractions. The conformal plan was implemented by a 4-field box technique (4F) that delivered static radiation fields at cardinal gantry angles of 0°, 90°, 180° and 270°. To evaluate the effect of bone content inside beams, an oblique 4-field radiation (O4F) which had gantry angles of 45°, 135°, 225° and 315° was also planned, as is often performed for prosthetic patients and recommended by TG-63 [29]. The VMAT plan used two 360° arc fields and optimized using Eclipse progressive resolution optimization (PRO 13.7.1) algorithm to fulfill dose constraints in the protocol.
The three plans on each patient were copied to CT data set for dose calculation with the same MUs used in the synCT plans. The analytical anisotropic algorithm (AAA 13.7) was used to calculated dose on both synCT and CT with heterogeneity corrections [30]. We also calculated VMAT plans on both images using Acuros XB algorithm (Acuro 13.1) to evaluate the impact of calculation algorithm on synCT-to-CT dosimetric comparison, as Acuros XB was reported to be superior in dose calculation than the AAA in heterogeneous phantom and various sites [31–35]. All volumetric dose calculations were set with a grid resolution of 2.5 × 2.5 × 2.5 mm3.
Dosimetric evaluation
To assess whether synCT and CT were equivalent for treatment planning, dose-volume histogram (DVH) and associated metrics were assessed for target volume and OARs in each plan. Based on our clinical practice, OAR doses were evaluated as: V40Gy and mean dose for bladder, V45Gy for small bowel, and V30Gy for femoral head, where VxGy is the percentage volume of the OAR that receives dose ≥ x Gy. PTV dose was assessed by D100%, D95%, D2% and mean dose, where Dx% was percentage dose (relative to prescription dose) in the PTV that covered x% volume. Medians, first (Q1) and third quantiles (Q3) of the metrics were calculated from the eleven patients’ data. Mann-Whitney non-parametric U-test was performed to evaluate the differences of these metrics between synCT and CT plan using a same treatment technique. A significant level p = 0.01 was used for the multiple tests. Furthermore, the graphic technique assessing agreement between two measurements [36] was applied to test the equivalence of the metrics between synCT and CT. The limit of agreement [36] was defined as (Mean– 2STD, mean + 2STD) of the metric differences, where STD is standard deviation. A tolerance of 2% was selected as the criterion for acceptance.
Spatial dose distributions were compared by absolute dose differences between synCT and CT plans for dose grid points with a dose greater than 10% prescription dose. Mean and STD of the differences in a patient volume were calculated. Moreover, 3D gamma analysis of dose distributions between synCT and CT was performed at dose-difference and distance-to-agreement criteria of 1%/1mm and 2%/2mm with a 10% dose threshold. The analysis generated a gamma map where a gamma value < 1 indicated that the point had comparable dose. A passing rate was calculated from the gamma map as the percentage of dose points that met the criteria (i.e., gamma value < 1). All comparisons for dose distributions were performed on both AAA and Acuros XB calculated plans. Non-parametric U test with p = 0.01 was performed to compare the means of the volumetric dose differences between dose calculation algorithms.
Results
The volumes of the patient PTVs varied in a range of 759–1204 mm3 (mean ± std: 963 ± 148 mm3). Colorwash doses along with the PTV for a representative patient #5 is shown in Fig 1 for various techniques. All three plans accomplished the planning goal of 50.4Gy prescription dose covering 95% PTV. The VMAT plan showed smaller doses in the tissue surrounding PTV due to plan optimization using arc beams. While O4F showed less radiation in the femoral bone as expected than the 4F plan, O4F had more doses in the bladder and bowel regions.
Average DVH curves for PTV and OARs from all the patients are shown in Fig 2. The average DVH curves were almost identical between synCT and CT plans. Dose metrics derived from the DVHs are listed in Table 1. Regardless of treatment technique, U tests suggested no significant difference in any PTV metric between synCT and CT plan. The OAR dose metrics present considerable variations between patients, but none showed significant differences between synCT and CT. Metric differences against means of the metrics from synCT and CT in all three treatment techniques were plotted in Fig 3 for statistical agreement assessment. The mean difference for the dose metrics (PTV D100%, D95%, D2%, Dmean and Bladder Dmean) was -0.2% with 95% interval -1.2% to 0.8% (Fig 3A). Fig 3B showed the limit of agreement for volume metrics (Bladder V40Gy, Bowel V45Gy and Femoral Head V30Gy) was (-0.9%, 0.8%), indicating the agreement between synCT and CT doses.
Table 1. Comparison of dose metrics for PTV and OARs.
4F | O4F | VMAT | ||||||||
---|---|---|---|---|---|---|---|---|---|---|
synCT | CT | p-value | synCT | CT | p-value | synCT | CT | p-value | ||
PTV | D100% | 93.2 | 92.7 | 0.53 | 92.1 | 92.2 | 1.00 | 86.2 | 86.5 | 0.77 |
(92.7, 93.4) | (92.7, 93.2) | (91.1, 92.5) | (91.1, 92.5) | (82.9, 87.5) | (83.3, 87.7) | |||||
D95% | 100.0 | 100.0 | 0.45 | 100.0 | 100.0 | 0.11 | 100.0 | 100.0 | 1.00 | |
(100.0, 100.0) | (100.0, 100.2) | (100.0, 100.0) | (100.0, 100.6) | (100.0, 100.0) | (99.7, 100.5) | |||||
D2% | 105.3 | 106.1 | 0.72 | 107.9 | 107.9 | 0.72 | 107.1 | 107.4 | 0.16 | |
(104.9, 107.1) | (105.1, 107.0) | (106.7, 108.9) | (106.8, 109.4) | (106.7, 107.3) | (107.0, 108.1) | |||||
Dmean | 103.2 | 103.5 | 0.24 | 104.4 | 104.6 | 0.39 | 103.1 | 103.2 | 0.35 | |
(102.8, 103.4) | (103.1, 104.0) | (103.9, 104.6) | (103.8, 105.1) | (102.9, 103.3) | (103.0, 103.7) | |||||
Bladder | V40Gy | 59.0 | 59.5 | 0.95 | 94.6 | 94.7 | 1.00 | 42.2 | 42.3 | 0.95 |
(35.4, 73.2) | (35.4, 73.3) | (87.4, 100.0) | (87.3, 100.0) | (25.6, 48.7) | (26.0, 48.7) | |||||
Dmean | 85.4 | 86.0 | 0.90 | 100.9 | 101.2 | 0.95 | 69.7 | 69.9 | 0.87 | |
(75.4, 89.5) | (75.3, 89.5) | (97.6, 102.5) | (97.8, 102.5) | (63.0, 76.2) | (63.4, 76.4) | |||||
Bowel | V45Gy | 19.5 | 19.7 | 0.97 | 62.6 | 62.9 | 1.00 | 8.5 | 8.8 | 0.91 |
(7.4, 45.2) | (7.4, 45.2) | (38.8, 73.0) | (38.7, 72.9) | (4.3, 27.0) | (4.1, 26.9) | |||||
Femoral Head | V30Gy | 8.7 | 7.3 | 1.00 | 3.9 | 3.7 | 0.90 | 5.9 | 6.5 | 1.00 |
(2.2, 17.4) | (2.2, 16.5) | (2.2, 14.9) | (2.0, 15.0) | (3.4, 10.7) | (3.4, 10.9) |
Dmean: percentage mean dose relative to prescription dose
The synCT-to-CT differences in the dose metrics are plotted in Fig 4 for each plan technique. Medians of the absolute differences in all evaluated PTV metrics were 0.2% (Q1-Q3: 0.1–0.7%) for 4F, 0.2% (0.1–0.4%) for O4F and 0.3% (0.2–0.6%) for VMAT plan. Meanwhile, in all treatment techniques (total 30 plans), the differences in bladder V40Gy, bowel V45Gy and femoral head V30Gy were 0.1% (0–0.2%), 0.1% (0–0.3%), 0.2% (0–0.4%), respectively.
A dosimetric analysis for a VMAT plan is shown in Fig 5. Fig 5C shows absolute dose difference of the slice between synCT and CT plan with a scale from 0 to 1 Gy, on which the dose differences is less than 0.2 Gy (i.e., <0.4% prescription dose) in most part of the body. The points with dose differences greater than 1 Gy were present at the edges of the body and PTV. The gamma map from 2%/2mm gamma analysis of the slice is shown in Fig 5D with a gamma value scale from 0 to 1, indicating almost all the points had equivalent doses between synCT and CT plan.
Fig 6 compares volumetric averages of absolute synCT-to-CT dose differences for each patient. Medians of the dose difference averages were 0.21 Gy (0.16–0.29 Gy), 0.18Gy (0.16–0.23Gy) and 0.13Gy (0.11–0.17Gy) for 4F, O4F and VMAT, respectively. 1%/1mm gamma analysis also showed that the passing rate from VMAT is ~2% higher than those from 3DCRTs (Table 2). Regardless of planning techniques, the median passing rates between synCT and CT plans (Table 2) were higher than 96% and 99% for 1%/1mm and 2%/2mm criteria, respectively.
Table 2. Passing rates of gamma analysis for synCT and CT-calculated dose distributions for each patient.
4F | OBL 4F | VMAT | ||||
---|---|---|---|---|---|---|
Patient ID | 1%/1mm | 2%/2mm | 1%/1mm | 2%/2mm | 1%/1mm | 2%/2mm |
1 | 99.4 | 100.0 | 97.5 | 99.9 | 99.0 | 99.9 |
2 | 97.9 | 99.8 | 99.1 | 100.0 | 98.9 | 100.0 |
3 | 97.0 | 99.7 | 96.7 | 99.9 | 98.0 | 99.9 |
4 | 93.1 | 99.7 | 96.6 | 99.7 | 98.8 | 99.9 |
5 | 99.0 | 100.0 | 99.7 | 100.0 | 100.0 | 100.0 |
6 | 87.9 | 98.5 | 94.6 | 99.7 | 97.4 | 99.7 |
7 | 99.6 | 99.9 | 100.0 | 100.0 | 100.0 | 100.0 |
8 | 90.6 | 99.8 | 98.4 | 100.0 | 99.9 | 100.0 |
9 | 88.1 | 99.0 | 85.4 | 97.2 | 93.7 | 99.6 |
10 | 98.1 | 100.0 | 96.5 | 100.0 | 99.0 | 100.0 |
11 | 92.5 | 99.6 | 95.5 | 98.7 | 91.9 | 99.4 |
median (range) | 97.0 (87.9–99.6) | 99.8 (98.5–100.0) | 96.7 (85.4–100.0) | 99.9 (97.2–100) | 98.9 (91.9–100.0) | 99.9 (99.4–100.0) |
We re-calculated the VMAT plans on synCT and CT using the Acuros XB algorithm that may improve patient dose estimation [34]. Table 3 summarizes dose differences and gamma analysis of Acuros XB-calculated VMAT plans. Using Auros XB, 2%/2mm gamma analysis of VMAT plans showed 99% pass rating between synCT and CT, along with a median of the dose difference averages 0.20 Gy (~0.4%, range: 0.13–0.64 Gy), suggesting the clinical agreement between synCT and CT plans, irrespective of dose calculation algorithm.
Table 3. Comparison of dose distributions between VMAT plans using AAA and Acuros XB calculations on synCT and CT.
Comparison between plans | Median (Q1, Q3) of dose difference average (Gy) | Median (range) of gamma passing rates (%) | ||
---|---|---|---|---|
1%/1mm | 2%/2mm | |||
synCT.vs. CT | AAA | 0.13 (0.11, 0.17) | 98.9 (91.9–100.0) | 99.9 (99.4–100.0) |
Acuros XB | 0.20 (0.16, 0.23) | 96.3 (89.9–98.4) | 99.2 (97.7–99.9) | |
AAA.vs. Acuros XB | CT | 0.38 (0.35, 0.40) | 93.4 (90.6–95.7) | 99.7 (99.7–99.9) |
SynCT | 0.38 (0.35, 0.41) | 93.6 (87.8–94.7) | 99.8 (99.6–99.9) |
Table 3 also compares doses between AAA and Acuros XB on a same image. On either synCT or CT, 2%/2mm gamma analysis on doses between two algorithms had >99% passing rates, indicating their equivalence for clinical dose calculation. 1%/1mm gamma analysis showed ~93% passing rates between AAA and Acuros XB, which were slightly smaller than those (> 95%) from dose comparisons between synCT and CT. The means of AAA-to-Acuros XB dose difference averages were ~0.4Gy (~0.8%). Mann-Whitney U tests on the dose difference averages further demonstrated that synCT-to-CT dose differences using either algorithm were significantly (p < 0.001) smaller than AAA-to-Acuros XB dose differences on either image.
Discussion
This study showed that synCT generated from Dixon-MR using an automated model-based method provided accurate and equivalent dose calculation as conventional CT for treatment planning of rectal cancer. The study performed treatment planning on synCT as a MR-only workflow, and then recalculated the plans on CT for evaluation. Dosimetric agreement between synCT and CT was demonstrated by the similarities in spatial dose distributions as well as PTV and OARs dose metrics. The results suggest that MR-generated synCT can support a MR-only workflow for radiotherapy of rectal cancer by providing electron density map like conventional CT for accurate dose calculation.
The MR generated synCT has been evaluated for treatment planning of VMAT for brain [18] and liver tumors [19], IMRT for head and neck cancer [37], prostate tumor [20] and lung cancer [25]. Recently, a commercial synCT software has been evaluated for use in prostate radiotherapy [38]. In this study, we demonstrated that MR generated synCT was sufficient for planning of both 3DCRTand VMAT of rectal cancer. The clinical equivalence between synCT and CT plan was assessed by dose metrics that physicians use to evaluate a plan for rectal cancer treatment. Voxelwise dose differences further indicated the close matching of dose distributions from the two modalities. Fig 5 shows that the voxels having a relatively high dose difference mostly located at the high dose-gradient edge regions where the dose differences were sensitive to possible misalignment between synCT and CT. The spatial error was accounted for by a distance-to-agreement criterion in gamma analysis which is widely used for comparison of dose distributions [39]. The high passing rates of gamma analysis in Table 2 indicated that synCT calculated doses at these voxels, although exhibiting dose discrepancies, agreed with CT calculated dose distribution with clinically-used dose-difference/distance-to-agreement criteria. Acuros XB was found to provide more accurate dose calculation than AAA near air-tissue interface [34, 35, 40]. We thus evaluated AAA and Acuros XB dose calculations in this study as substantial air-tissue interface presents in rectum. Gamma analysis in Table 3 showed that synCT and CT calculated doses well agreed with each other for both Acuros XB and AAA. Acuros XB is now gradually gaining use in clinic for dose calculation. Table 3 showed that the dose discrepancy between AAA and Acuros XB calculation was greater than the difference between synCT and CT dose using either algorithm. These further suggested the clinical equivalence of synCT and CT for radiation dose calculation.
A hybrid method [24] was used in this study to generate synCT from MR images. The method avoids the difficulty of differentiating between bone and air in the classification-based method and the uncertainty in registration of soft tissue in the atlas-based method. The method still utilizes registration between patient MR images and atlas for bone delineation. The registration has been refined in the method by first a landmark matching and then an intensity-based deformable registration since the atlas is only for bone [24]. Nevertheless, the bone could be prone to registration uncertainty and can lead to dose calculation error when irradiating through bone. A VMAT plan delivered IMRT fields during a continuous gantry rotation. Compared with 4F conformal RT, a much smaller portion of the VMAT MU irradiated through pelvic bone. Therefore, as indicated in the Table 2, VMAT showed a slightly higher passing rate than the 3D plans for 1%/1mm gamma analysis. Despite the difference in 1%/1mm gamma analysis, our results showed that the dose differences between synCT and CT were small and clinically insignificant regardless of the treatment techniques. Recently, a synCT generation method has been reported to identify pelvic bone by using a pelvic bone shape model [21]. Similarly, bone identification from the shape model was prone to matching error, but clinically, the generated synCT was promising for treatment planning.
Image registration between synCT and CT was performed in this study to alleviate potential tissue displacement from PET/MRI to PET/CT acquisition. The uncertainty in the registration could lead to minor misalignment near body contour and organ edges between synCT and CT images, which may amplify the dose differences at high dose-gradient area (Fig 5C). Additionally, we observed differences in the level of rectal and bladder filling between MR and CT images. These may add extra errors in current results of dosimetric comparison between synCT and CT plans. Practically, use of the synCT in a MR-only workflow will avoid the CT scan, possible artifacts in CT images, and the uncertainty of MR/CT registration.
The synCT was generated by assigning a HU number to a voxel based on its MR-derived tissue type. It ignored the variation of HU numbers within a tissue type and between patients. Various studies have shown that radiation dose deposition is relatively insensitive to tissue HU number [16, 20, 41]. It was reported that HU differences smaller than 34 HU would not affect dosimetric result [16, 42]. In addition, synCT should be geometrically accurate to represent patient anatomy for dose calculation. However, MR image is known to have geometrical distortion. The distortion depends on numerous sources including the MR scanner, field strength, pulse sequences and image FOV, and the magnitude could be substantially greater than desired spatial accuracy for a radiation simulator [43]. Geometrical distortion for the rectum near scanner isocenter may be insignificant, but could be problematic at body surface for an oversize patient. Fortunately, phantoms and distortion correction methods considering specific requirements for radiotherapy are increasingly being provided by vendors [43, 44]. Various studies have shown that the distortions can be reduced to the acceptable level for radiation treatment in many sites [14, 45]. Undoubtedly, comprehensive quality assurance of MR geometrical distortion is an essential prerequisite for clinical application of MR-generated synthetic CT.
As the emergence of MR-Linac technique and development of MR-guided radiotherapy, a MR-only treatment planning will spare the extra effort of CT simulation and simplify treatment workflow with cost effectiveness. Implementation of MR-only treatment planning using synCT is being evaluated clinically in several institutions [12, 19, 45, 46]. This study comprehensively evaluated the synthetic CT for treatment planning in eleven patients. It is limited by the relatively small number of patients, but nevertheless supports further development of a MR-only treatment planning for rectal cancer.
Conclusion
Treatment planning on MR generated synCT agreed very well with the dose calculated on CT for 3DCRT and VMAT planning for rectal cancer treatment. Mean dosimetric difference between synCT and CT was less than 0.4% for any treatment technique. The study demonstrated the potential of MR-only treatment planning for colorectal cancer thus streamlining the patient care.
Data Availability
All relevant data are within the paper.
Funding Statement
The authors received no specific funding for this work.
References
- 1.Siegel RL, Miller KD, Fedewa SA, Ahnen DJ, Meester RG, Barzi A, et al. Colorectal cancer statistics, 2017. CA Cancer J Clin. 2017. doi: 10.3322/caac.21395 . [DOI] [PubMed] [Google Scholar]
- 2.Siegel RL, Fedewa SA, Anderson WF, Miller KD, Ma J, Rosenberg PS, et al. Colorectal Cancer Incidence Patterns in the United States, 1974–2013. J Natl Cancer Inst. 2017;109(8). doi: 10.1093/jnci/djw322 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Kye BH, Cho HM. Overview of radiation therapy for treating rectal cancer. Ann Coloproctol. 2014;30(4):165–74. doi: 10.3393/ac.2014.30.4.165 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Sauer R, Becker H, Hohenberger W, Rodel C, Wittekind C, Fietkau R, et al. Preoperative versus postoperative chemoradiotherapy for rectal cancer. N Engl J Med. 2004;351(17):1731–40. doi: 10.1056/NEJMoa040694 . [DOI] [PubMed] [Google Scholar]
- 5.Chen C, Sun P, Rong J, Weng HW, Dai QS, Ye S. Short Course Radiation in the Treatment of Localized Rectal Cancer: A Systematic Review and Meta-Analysis. Sci Rep. 2015;5:10953 doi: 10.1038/srep10953 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Group MS. Diagnostic accuracy of preoperative magnetic resonance imaging in predicting curative resection of rectal cancer: prospective observational study. BMJ. 2006;333(7572):779 doi: 10.1136/bmj.38937.646400.55 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Jhaveri KS, Hosseini-Nik H. MRI of Rectal Cancer: An Overview and Update on Recent Advances. AJR Am J Roentgenol. 2015;205(1):W42–55. doi: 10.2214/AJR.14.14201 . [DOI] [PubMed] [Google Scholar]
- 8.Kulkarni T, Gollins S, Maw A, Hobson P, Byrne R, Widdowson D. Magnetic resonance imaging in rectal cancer downstaged using neoadjuvant chemoradiation: accuracy of prediction of tumour stage and circumferential resection margin status. Colorectal Dis. 2008;10(5):479–89. doi: 10.1111/j.1463-1318.2007.01451.x . [DOI] [PubMed] [Google Scholar]
- 9.Brown G, Radcliffe AG, Newcombe RG, Dallimore NS, Bourne MW, Williams GT. Preoperative assessment of prognostic factors in rectal cancer using high-resolution magnetic resonance imaging. Br J Surg. 2003;90(3):355–64. doi: 10.1002/bjs.4034 . [DOI] [PubMed] [Google Scholar]
- 10.Taylor FG, Quirke P, Heald RJ, Moran B, Blomqvist L, Swift I, et al. Preoperative high-resolution magnetic resonance imaging can identify good prognosis stage I, II, and III rectal cancer best managed by surgery alone: a prospective, multicenter, European study. Ann Surg. 2011;253(4):711–9. doi: 10.1097/SLA.0b013e31820b8d52 . [DOI] [PubMed] [Google Scholar]
- 11.Ulin K, Urie MM, Cherlow JM. Results of a multi-institutional benchmark test for cranial CT/MR image registration. Int J Radiat Oncol Biol Phys. 2010;77(5):1584–9. doi: 10.1016/j.ijrobp.2009.10.017 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Paulson ES, Crijns SP, Keller BM, Wang J, Schmidt MA, Coutts G, et al. Consensus opinion on MRI simulation for external beam radiation treatment planning. Radiother Oncol. 2016;121(2):187–92. doi: 10.1016/j.radonc.2016.09.018 . [DOI] [PubMed] [Google Scholar]
- 13.Glide-Hurst CK, Wen N, Hearshen D, Kim J, Pantelic M, Zhao B, et al. Initial clinical experience with a radiation oncology dedicated open 1.0T MR-simulation. J Appl Clin Med Phys. 2015;16(2):5201 doi: 10.1120/jacmp.v16i2.5201 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Schmidt MA, Payne GS. Radiotherapy planning using MRI. Phys Med Biol. 2015;60(22):R323–61. doi: 10.1088/0031-9155/60/22/R323 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Lagendijk JJ, Raaymakers BW, Raaijmakers AJ, Overweg J, Brown KJ, Kerkhof EM, et al. MRI/linac integration. Radiother Oncol. 2008;86(1):25–9. doi: 10.1016/j.radonc.2007.10.034 . [DOI] [PubMed] [Google Scholar]
- 16.Das IJ, Cheng CW, Cao M, Johnstone PA. Computed tomography imaging parameters for inhomogeneity correction in radiation treatment planning. J Med Phys. 2016;41(1):3–11. doi: 10.4103/0971-6203.177277 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Dowling JA, Lambert J, Parker J, Salvado O, Fripp J, Capp A, et al. An atlas-based electron density mapping method for magnetic resonance imaging (MRI)-alone treatment planning and adaptive MRI-based prostate radiation therapy. Int J Radiat Oncol Biol Phys. 2012;83(1):e5–11. doi: 10.1016/j.ijrobp.2011.11.056 . [DOI] [PubMed] [Google Scholar]
- 18.Paradis E, Cao Y, Lawrence TS, Tsien C, Feng M, Vineberg K, et al. Assessing the Dosimetric Accuracy of Magnetic Resonance-Generated Synthetic CT Images for Focal Brain VMAT Radiation Therapy. Int J Radiat Oncol Biol Phys. 2015;93(5):1154–61. doi: 10.1016/j.ijrobp.2015.08.049 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Bredfeldt JS, Liu L, Feng M, Cao Y, Balter JM. Synthetic CT for MRI-based liver stereotactic body radiotherapy treatment planning. Phys Med Biol. 2017;62(8):2922–34. doi: 10.1088/1361-6560/aa5059 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Kim J, Garbarino K, Schultz L, Levin K, Movsas B, Siddiqui MS, et al. Dosimetric evaluation of synthetic CT relative to bulk density assignment-based magnetic resonance-only approaches for prostate radiotherapy. Radiat Oncol. 2015;10:239 doi: 10.1186/s13014-015-0549-7 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Liu L, Cao Y, Fessler JA, Jolly S, Balter JM. A female pelvic bone shape model for air/bone separation in support of synthetic CT generation for radiation therapy. Phys Med Biol. 2016;61(1):169–82. doi: 10.1088/0031-9155/61/1/169 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Johansson A, Karlsson M, Nyholm T. CT substitute derived from MRI sequences with ultrashort echo time. Med Phys. 2011;38(5):2708–14. doi: 10.1118/1.3578928 . [DOI] [PubMed] [Google Scholar]
- 23.Hsu SH, Cao Y, Huang K, Feng M, Balter JM. Investigation of a method for generating synthetic CT models from MRI scans of the head and neck for radiation therapy. Phys Med Biol. 2013;58(23):8419–35. doi: 10.1088/0031-9155/58/23/8419 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Paulus DH, Quick HH, Geppert C, Fenchel M, Zhan Y, Hermosillo G, et al. Whole-Body PET/MR Imaging: Quantitative Evaluation of a Novel Model-Based MR Attenuation Correction Method Including Bone. J Nucl Med. 2015;56(7):1061–6. doi: 10.2967/jnumed.115.156000 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Wang H, Chandarana H, Block KT, Vahle T, Fenchel M, Das IJ. Dosimetric evaluation of synthetic CT for magnetic resonance-only based radiotherapy planning of lung cancer. Radiat Oncol. 2017;12(1):108 doi: 10.1186/s13014-017-0845-5 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Dixon WT. Simple proton spectroscopic imaging. Radiology. 1984;153(1):189–94. doi: 10.1148/radiology.153.1.6089263 . [DOI] [PubMed] [Google Scholar]
- 27.Schulz V, Torres-Espallardo I, Renisch S, Hu Z, Ojha N, Bornert P, et al. Automatic, three-segment, MR-based attenuation correction for whole-body PET/MR data. Eur J Nucl Med Mol Imaging. 2011;38(1):138–52. doi: 10.1007/s00259-010-1603-1 . [DOI] [PubMed] [Google Scholar]
- 28.Bai C, Shao L, Da Silva AJ, Zhao Z. A generalized model for the conversion from CT numbers to linear attenuation coefficients. IEEE Transactions on Nuclear Science. 2003;50(5):1510–5. doi: 10.1109/tns.2003.817281 [Google Scholar]
- 29.Reft C, Alecu R, Das IJ, Gerbi BJ, Keall P, Lief E, et al. Dosimetric considerations for patients with HIP prostheses undergoing pelvic irradiation. Report of the AAPM Radiation Therapy Committee Task Group 63. Med Phys. 2003;30(6):1162–82. doi: 10.1118/1.1565113 . [DOI] [PubMed] [Google Scholar]
- 30.Van Esch A, Tillikainen L, Pyykkonen J, Tenhunen M, Helminen H, Siljamaki S, et al. Testing of the analytical anisotropic algorithm for photon dose calculation. Med Phys. 2006;33(11):4130–48. doi: 10.1118/1.2358333 . [DOI] [PubMed] [Google Scholar]
- 31.Bush K, Gagne IM, Zavgorodni S, Ansbacher W, Beckham W. Dosimetric validation of Acuros XB with Monte Carlo methods for photon dose calculations. Med Phys. 2011;38(4):2208–21. doi: 10.1118/1.3567146 . [DOI] [PubMed] [Google Scholar]
- 32.Fogliata A, Nicolini G, Clivio A, Vanetti E, Cozzi L. Dosimetric evaluation of Acuros XB Advanced Dose Calculation algorithm in heterogeneous media. Radiat Oncol. 2011;6:82 doi: 10.1186/1748-717X-6-82 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Fogliata A, Nicolini G, Clivio A, Vanetti E, Cozzi L. On the dosimetric impact of inhomogeneity management in the Acuros XB algorithm for breast treatment. Radiat Oncol. 2011;6:103 doi: 10.1186/1748-717X-6-103 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Fogliata A, Nicolini G, Clivio A, Vanetti E, Cozzi L. Critical appraisal of Acuros XB and Anisotropic Analytic Algorithm dose calculation in advanced non-small-cell lung cancer treatments. Int J Radiat Oncol Biol Phys. 2012;83(5):1587–95. doi: 10.1016/j.ijrobp.2011.10.078 . [DOI] [PubMed] [Google Scholar]
- 35.Han T, Followill D, Mikell J, Repchak R, Molineu A, Howell R, et al. Dosimetric impact of Acuros XB deterministic radiation transport algorithm for heterogeneous dose calculation in lung cancer. Med Phys. 2013;40(5):051710 doi: 10.1118/1.4802216 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10. . [PubMed] [Google Scholar]
- 37.Chin AL, Lin A, Anamalayil S, Teo BK. Feasibility and limitations of bulk density assignment in MRI for head and neck IMRT treatment planning. J Appl Clin Med Phys. 2014;15(5):4851 doi: 10.1120/jacmp.v15i5.4851 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Tyagi N, Fontenla S, Zhang J, Cloutier M, Kadbi M, Mechalakos J, et al. Dosimetric and workflow evaluation of first commercial synthetic CT software for clinical use in pelvis. Phys Med Biol. 2017;62(8):2961–75. doi: 10.1088/1361-6560/aa5452 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25(5):656–61. doi: 10.1118/1.598248 . [DOI] [PubMed] [Google Scholar]
- 40.Huang B, Wu L, Lin P, Chen C. Dose calculation of Acuros XB and Anisotropic Analytical Algorithm in lung stereotactic body radiotherapy treatment with flattening filter free beams and the potential role of calculation grid size. Radiat Oncol. 2015;10:53 doi: 10.1186/s13014-015-0357-0 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Cozzi L, Fogliata A, Buffa F, Bieri S. Dosimetric impact of computed tomography calibration on a commercial treatment planning system for external radiation therapy. Radiother Oncol. 1998;48(3):335–8. . [DOI] [PubMed] [Google Scholar]
- 42.Yoo S, Yin FF. Dosimetric feasibility of cone-beam CT-based treatment planning compared to CT-based treatment planning. Int J Radiat Oncol Biol Phys. 2006;66(5):1553–61. doi: 10.1016/j.ijrobp.2006.08.031 . [DOI] [PubMed] [Google Scholar]
- 43.Weygand J, Fuller CD, Ibbott GS, Mohamed AS, Ding Y, Yang J, et al. Spatial Precision in Magnetic Resonance Imaging-Guided Radiation Therapy: The Role of Geometric Distortion. Int J Radiat Oncol Biol Phys. 2016;95(4):1304–16. doi: 10.1016/j.ijrobp.2016.02.059 . [DOI] [PubMed] [Google Scholar]
- 44.Sun J, Dowling J, Pichler P, Menk F, Rivest-Henault D, Lambert J, et al. MRI simulation: end-to-end testing for prostate radiation therapy using geometric pelvic MRI phantoms. Phys Med Biol. 2015;60(8):3097–109. doi: 10.1088/0031-9155/60/8/3097 . [DOI] [PubMed] [Google Scholar]
- 45.Jonsson JH, Karlsson MG, Karlsson M, Nyholm T. Treatment planning using MRI data: an analysis of the dose calculation accuracy for different treatment regions. Radiat Oncol. 2010;5:62 doi: 10.1186/1748-717X-5-62 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Kapanen M, Collan J, Beule A, Seppala T, Saarilahti K, Tenhunen M. Commissioning of MRI-only based treatment planning procedure for external beam radiotherapy of prostate. Magn Reson Med. 2013;70(1):127–35. doi: 10.1002/mrm.24459 . [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All relevant data are within the paper.