Joint system relaxometry (JSR) and Crámer‐Rao lower bound optimization of sequence parameters: A framework for enhanced precision of DESPOT T1 and T2 estimation

Rui Pedro AG Teixeira 1,2,, Shaihan J Malik 1, Joseph V Hajnal 1,2
PMCID: PMC5763350  PMID: 28303617

Abstract

Purpose

This study aims to increase the precision of single‐compartment DESPOT relaxometry by two means: (i) a joint system relaxometry (JSR) approach that estimates parameters in a single step using all available data; and (ii) optimizing acquisition parameters by deploying a robust design tool based on the Crámer‐Rao lower bound (CRLB).

Methods

Following the development of the analysis and design capabilities, phantom and four in vivo subject experiments were performed to compare directly the precision achieved with DESPOT and JSR estimation using published protocols and protocols designed using a proposed CRLB framework.

Results

Experimental data demonstrate JSR's ability to decrease relaxometry estimation variance. Phantom results show 72 to 77% improvement using the same data as conventional DESPOT. This is further improved to 81 to 87% using optimal parameters. Both experiments show systematic bias depending on the acquisition parameters used, which are shown to be highly reproducible and to vary with different magnetization transfer conditions.

Conclusions

Compared with DESPOT, JSR produces reproducible relaxation maps with improved precision. Further improvement was achieved using CRLB as a protocol design tool. With this combined approach, it is possible to achieve submillimeter maps of ρ,T1,T2, and B0 in an 11‐min examination, making the approach appealing for potential clinical use. Magn Reson Med 79:234–245, 2018. © 2017 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

Keywords: relaxometry, CRLB, DESPOT

INTRODUCTION

Despite the continued dominance of qualitative assessment in clinical applications of MRI, the ability to probe tissue relaxation times T1 and T2 has proven to be useful—particularly in detecting and understanding neurological diseases 1, 2. Unfortunately, “gold standard” methods for estimating T1 and T2 are 2D acquisitions based on spin‐echo (SE) methods, which require long acquisition times and have been shown to produce variable results across different studies 3, hindering their application in clinical settings 4. Gradient‐echo methods such as the variable flip angle (VFA) method 5, or driven equilibrium single pulse observation of T1 and T2 (DESPOT1/2) 6 were previously shown to generate 1‐mm3 3D maps of the human brain in clinically feasible times 6, 7. However, they have been reported to produce a systematic offset when compared with gold standard methods and, to the authors' knowledge, a common ground between both has yet to be found 3, 8.

The DESPOT1/2 approach makes use of spoiled gradient echo (SPGR) and balanced steady‐state free precession (bSSFP) sequences over a range of varying flip angles (FAs) at fixed repetition times (TRs), to estimate T1 and T2 of each measured voxel. The standard fitting procedure 6 obtains T1 from a linearized Ernst relationship using SPGR data alone, and then feeds the T1 value into a linearized bSSFP model to estimate T2  A DESPOT2 approach that avoids linearization of the equations (DESPOT‐FM) was also proposed, to estimate the underlying field map 9. Recently, multicomponent signal models have also been proposed 10, 11, and an interesting discussion regarding their estimation stability based on both Monte Carlo 12 and Cramér‐Rao lower bound (CRLB) methods 13 can be found elsewhere 12, 13, 14 (the CRLB is a statistical tool that predicts the minimum obtainable variance of the estimated model parameters given the noise properties of the measurement experiment 13, 15, 16). However, a limited amount of work has been presented 17, 18, 19, 20 regarding the evaluation of the single‐compartment model for the range of relaxation times found in the human brain. This study is concerned with optimizing the estimation precision, and thereby the efficiency of single‐compartment DESPOT‐style methods. It is divided into two main topics. First, a joint system relaxometry (JSR) approach is adopted, in which full (ie, not linearized) models for both SPGR and bSSFP signals are simultaneously evaluated, maximizing the information that is used to estimate the relaxation parameters. Second, following the work presented by Lankford and Does 13, a way to optimize the JSR based on the CRLB framework is proposed. This allows the selection of optimized acquisition protocols for tissues of interest. The resulting framework has been tested on a phantom and in vivo to determine the performance benefits that can be achieved.

THEORY

DESPOT Theory

Originally called variable nutation angle 21, the DESPOT nomenclature was first introduced by Homer and Beevers in 1985 21 and expanded to DESPOT1/2 by Deoni et al in 2003, when the use of the bSSFP sequence to map T2 information after T1 had been assessed was suggested 6. However, this ignores both T2 or T2* dependency of the SPGR signals as well as T1 information present in the bSSFP curves. Both effects can be taken into consideration by carefully stating the signal evolution for different acquisition parameters. The expected SPGR signal equation, SSPGR, adopted in this work, is given by 6, 22, 23 as follows:

SSPGR(α,TRSPGR)=1E1(TRSPGR)1E1(TRSPGR)cos(κα)sin(κα)eTESPGRT2eTESPGRT2ei2πΔΩTESPGR (1)

whereas the bSSFP signal, SbSSFP, is described by 22, 24, 25 as follows:

SSSFP(α,TRbSSFP)=[MxSS(α,TRbSSFP)+iMySS(α,TRbSSFP)]eTEbSSFPT2ei2πΔΩTEbSSFP (2)

where

MxSS(α,TRbSSFP)=(1E1(TRbSSFP))E2(TRbSSFP)sin(κα)sin(β)d (3)
MySS(α,TRbSSFP)=(1E1(TRbSSFP))(1E2(TRbSSFP)cos(β))sin(κα)d (4)

and

d=(1E1(TRbSSFP)cos(κα))(1E2(TRbSSFP)cos(β))E2(TRbSSFP)(E1(TRbSSFP)cos(κα))(E2(TRbSSFP)cos(β)) (5)

Equations [1] to [5] describe a full magnetization model, which we refer to as joint system relaxometry, to distinguish it from the commonly used sequential procedure. Here, TRSPGR, TESPGR and TRbSSFP, and TEbSSFP are the repetition and echo times of the SPGR and bSSFP sequences; T2 is defined by 1/ T2* = 1/T2 + 1/ T2; α is the FA modulated by a (dimensionless) transmit field inhomogeneity κ; E1(TRSPGR,bSSFP)=exp(TRSPGR,bSSFP/T1); and β=2πΔΩTRbSSFP+ϕRF is the off‐resonance dephasing angle in radians, resulting from frequency offset ΔΩz(Hz) plus the rotation induced by incrementing the phase, ϕRF, of the radiofrequency (RF) pulse at each TRbSSFP 22. Finite RF pulse duration ( TRF) was considered by following the correction proposed in 8, 25 and setting E2=exp((TRζβTRFTRbSSFP)/T2) with ζβ=ζcos2[β2(1(1ζ)TRFTRbSSFP)] and ζ0.680.125(1+TRFTRbSSFP)T2T1  Note that the two sequence types need not have the same repetition and echo times as each other. MxSS and MySS correspond, respectively, to bSSFP real and imaginary steady‐state components. In Equations [1] to [5] we adopted the convention that all signals are normalized by the proton density, M0 as all measured signals have a scanner‐dependent scaling factor that must be explicitly accommodated during fitting (see “Methods” section). Both SPGR (Eq. [1]) and bSSFP (Eqs. [(2), (3), (4), (5)]) signal equations can be immediately derived from the Bloch equations as described elsewhere 22, 25.

The proposed JSR approach requires careful balance between SPGR and bSSFP signal intensities during readout periods. Although possible to compute the echo‐time decay from the magnetization before the RF pulse, using the magnetization immediately after the excitation pulse allows a simple exponential decay to be added to the steady‐state solution. This is contrary to the approach described by Deoni in 2009 (9), in which the steady‐state magnetization immediately before each RF pulse was considered. The eTE/T2 term is present in the SPGR and not in the bSSFP sequence model, which is a consequence of the latter's spin‐echo behavior as described in 26.

Equation [1] is only valid if perfect spoiling of the transverse magnetization is attained before each RF pulse. This is easily obtained for TRT1 however, RF and gradient spoiling methods need to be applied when TRT1. A good discussion of this issue is presented in 27, and the reported value of 50° as “stable” RF phase increment is assumed in this paper; no further correction to the apparent T1 obtained is applied. However, the obtained steady state is still TR,T1, T2 and FA dependent. To ensure that imperfect spoiling was kept under control, the SPGR sequence was modeled using the extended phase graph (EPG) algorithm 28, 29, 30, including attenuating effects associated with diffusion resulting from both imaging and spoiler gradients 28, 31. Choice of FA and TR was then constrained to ensure a peak error of less than 5% compared with the ideal Ernst regime for relevant T1 and T2 ranges.

Cramér‐Rao Lower Bound

Following the notation presented by Lankford and Does 13, considering a model g(x,θ), where x=[α1,α2,, αN]T is a vector containing N independent measurements (in the case of this paper, SPGR and bSSFP images at specific FA and TR) and θ=[θ1,θ2,, θM]T is a vector containing M model parameters (eg, M0, T1, T2, and ΔΩ (with M=4) then the variances of the elements of the unbiased estimate vector of the model parameters θ^ based on noisy measurements y=[y1,y2,, yN]T are described by the M × M covariance matrix Σθ^2 as follows:

Σθ^2 F1 (6)

where F is the Fisher information matrix. The CRLB, which is the lowest limit of Σθ^2 is obtained by taking the equality in Equation [6].

Assuming y follows a Gaussian noise distribution with standard deviation σ, F simplifies 13, 32 to:

Fj,k=i=1N1σi2giθjgiθk (7)

where j and k (j, k = 1, 2,…, M) are, respectively, the row and column indexes of F. The CRLB has been used previously to optimize experiment design in diffusion MRI 33 and quantitative magnetization transfer MRI 16. Following the methodology proposed in 16, we sought to enhance the estimation precision for multiple parameter estimates (eg, T1 and T2 by considering a subset, L of diagonal elements of Σθ^2 simultaneously, which corresponds to only considering the minimal obtainable variance of each parameter of interest.

To achieve equal relative precision (defined throughout this work as pθmCRLB=diag(Σθ^2)m/θm2 for each degree of freedom, the CRLB of each parameter is weighted by the inverse square of the parameter value. This means we can compute a root mean squared precision PrmsCRLB as follows:

PrmsCRLB=m=1Ldiag(Σθ^2)m   θm2=m=1LpθmCRLB (8)

It is important to notice that PrmsCRLB is computed for a specific parameter vector θ (eg, M0, T1, T2, and ΔΩ) However, we are interested in being able to evaluate PrmsCRLB over a wide range of different θ. Therefore, we compute PrmsCRLB over G different parameter vectors  θG=[θ1G,θ2G,,θMG] and the maximum value obtained is minimized, resulting in the following cost function, CFgrid, which is to be minimized as follows:

CFgrid=maxG{PrmsCRLB(θG)} (9)

CFgrid is conservative in that it always selects the worst‐case relative precision detected on the defined search space grid, and therefore minimizes the estimation variance for the worst case in all of the  θG considered.

METHODS

All simulations and offline postprocessing were performed using MATLAB 2016a (The MathWorks Inc, Natick, MA, USA). Equations [1] to [5] were normalized by M0 effectively defining them with respect to unit proton density. To fit the JSR model, which is the concatenation of the SPGR and bSSFP signal models established in Equations [1] to [5], to actual image data, it is necessary to take account of the strength of the signals found in each voxel. This is proportional to M0 scaled by an arbitrary position‐dependent complex gain  AeiϕA which can be attributed to receive sensitivity, including all sources of incidental phase.

JSR=AeiϕAM0[SSPGR1,,SSPGRNSPGR,SSSFP1,,SSSFPNbSSFP] (10)

Because it is not possible to separate the different contributions to this compound scaling factor, we define a complex “weighted proton density” ρ=AeiϕAM0=ρr+iρi This formulation keeps the bSSFP model and acquisitions everywhere differentiable, which is not the case in their characteristic stop‐band areas when using magnitude images, allows ΔΩ to be directly estimated from the image data, and retains a Gaussian image noise distribution. The SPGR acquisitions may contain additional phase factors, which might require an extra parameter to fit that is not needed for relaxation‐time determination. We therefore discard the SPGR image phase by taking the magnitude and enforcing ρSPGR=|ρ|=(ρr)2+(ρi)2 resulting in the final considered JSR model of Equation [11]:

JSRfinal=[|ρ|SSPGR1,,|ρ|SSPGRNSPGR,  Re(ρSSSFP1),Im(ρSSSFP1), ,Re(ρSSSFPNbSSFP),Im(ρSSSFPNbSSFP)] (11)

where real (Re) and imaginary (Im) signals are concatenated to keep the cost function real valued. For this approach, it is important that the signal‐to‐noise ratio (SNR) of each SPGR image set be sufficient to avoid significant Rician bias. This was confirmed empirically by using the MATLAB Kolmogorov‐Smirnov test to check for normal distributions on white matter (WM) and gray matter (GM) segmented SPGR images.

For both simulation and optimization, a grid of T1, T2 pairs were used to cover the relevant ranges for brain (obtained from pilot data acquired while setting up this study), from 600 to 1200 ms in steps of 25 ms for T1 and 25 to 80 ms in increments of 5 ms for T2. Although estimated, cerebral spinal fluid relaxation times have been excluded from the optimization, as it is difficult to guarantee correct spoiling of the magnetization and are, usually, of less clinical interest. Optimizing the measurement to estimate the higher T1, T2 values of cerebrospinal fluid (CSF) could diminish precision for brain tissue, while not achieving any valid measurement. The proton density was given the value of 10 in arbitrary units with zero phase (ie, ρr=10 and ρi= 0i). Although not necessary, for ease of guaranteeing equal sampling conditions between acquisitions such as bandwidth and geometrical distortions, TESPGR=TESSFP=0.5TRSSFP was imposed. Also, because of the short readout times used in this work, SPGR signal T2 deviations were neglected. This is justified, because, for typical frontal WM of T2=285 ms 34, induced signal deviations of not considering eTESPGR/T2 are less than 1%.

All images were obtained on a 3 Tesla (T) Philips Achieva‐TX system (Philips Healthcare, Best, Netherlands) with the manufacturer 32‐channel receive adult head coil and processed from k‐space raw data using the MRecon environment (Gyrotools LLC, Zurich, Switzerland). To assure equal RF pulse duration TRF for all acquired sequences, which allows a common finite pulse width correction 8, 25 to be used, the scanner software was modified to force a fixed pulse duration that is determined by the largest FA required with all other FA obtained by varying the pulse amplitude only. Nonlinearity of the RF amplifier was corrected by enforcing a vendor calibration step, which allows pre‐emphasis of the RF shape. The VFA data were sampled as a sagittal acquisition with a field of view (FOV) of 250 × 250 × 250 mm3 at 0.8‐mm3 isotropic resolution, fixed bandwidth of 959 Hz/pixel, and SENSE factor of 2 (anterior–posterior) and 2 (right–left) in both phase‐encode directions for both in vivo and phantom measurements. The SPGR spoiling area was kept at the default values optimized by the manufacturer software, which resulted in a total phase dispersion of approximately 7 radians per voxel in the readout direction. Nonselective excitation pulses were used for all 3D measurements. Correct knowledge of transmit field (B1) distribution is assumed as it can be assessed with mapping techniques 31, 35, 36, 37, which are not the main focus of this paper. The spatially dependent transmit scaling factor, k (defined in Eqs. [(1), (2), (3), (4), (5)]), was experimentally measured using the actual flip angle approach 37, with a TR1/TR2 = 25/125 ms and maximum allowed gradient spoiling between each TR. The FOV was set to 250 × 250 × 250 mm3 for an acquired isotropic resolution of 3.91 mm3, resulting in a total acquisition time of 2 min. The actual FA, αactual was then specified as kαprescribed in all calculations.

The parameter estimates, θJSR=[ρr,ρi, T1,T2 and ΔΩ] for each measured voxel are obtained through fitting Equation [11] on a least‐squares criteria using the MATLAB 2016a lsqnonlin routine. The objective function was defined as the sum of the square difference between the model and the measured signal intensity, and the stopping criteria were set as 1×1015 tolerance on the cost function value or a maximum of 500 iterations. The optimization's initial conditions were kept fixed for all estimated voxels and chosen as the expected average relaxation time in the entire brain from pilot data acquired while setting up this study. For all experimental estimations, each imaged voxel was independently processed on an Intel Xeon CPU E5‐2687W 0 at 3.10 GHz, parallelized to 16 cores, taking a total computational time of between 6 and 7 h, depending on the extracted brain volume.

For comparison purposes, linearized estimations of ρ1, T1 were obtained using the DESPOT1 approach as described in 6, and then ρ2, T2, and ΔΩ were estimated using DESPOT2‐FM 9 with a finite RF pulse correction 8, 25. Note that in this approach the proton density is separately estimated for each type of image, requiring the introduction of ρ1  and ρ2 As a point of reference for the optimized acquisition, a baseline DESPOT1/2 protocol was adapted from 9 and consists of two SPGR (DESPOT1 ‐ > FA = 4 ° and 18 °, TR = 6.2 ms) and four bSSFP (DESPOT2 ‐ > FA = 15 ° and 65 °, TR = 4.2 ms for ϕRF=π,0 rad measurements resulting in a total acquisition time of 9 min and 18 s.

CRLB Numerical Validation

Validation that the CRLB is able to accurately predict JSR estimation quality was performed with a Monte Carlo simulation using the previously defined values of ρ, T1, and T2. For each element of the resulting grid, 1 × 105 independent trials were generated. Gaussian distributed noise with zero mean and a standard deviation σ=0.02|ρ|  13was added to both real and imaginary parts of each signal.

For each T1 and T2 combination, the standard deviation σ of the Monte Carlo simulation as well as the predicted CRLB, σCRLB for both relaxation times were extracted. For comparison, each value was normalized by its respective relaxation‐time defining Monte Carlo and σCRLB2 precisions as pT1,2MC=(σT1,22/T1,22) d pT1,2CRLB=(σCLRB2T1,2/T1,22), respectively.

Phantom Validation

Phantom validation was performed by imaging an in‐house built spherical phantom filled with a 0.5% agarose, 0.9% NaCl, and 0.02‐mM MnCl2 solution, and primarily focused on two goals:

  1. Validation of estimation improvement based on JSR fitting approach; and

  2. Validation of the CRLB framework as a protocol design tool.

Comparison between the conventional two‐step fitting approach and the proposed JSR is performed using the baseline protocol. Both the DESPOT1/2 and JSR estimation maps were obtained using the same measured data with the procedure described previously. Reference relaxation values for the phantom were obtained using spin‐echo measurements within a single slice with a FOV of 250 × 250 mm2 and an acquired voxel size of 1.6 × 1.6 × 4 mm3. The T1 was mapped using single‐shot inversion recovery fast spin echo with k‐space filled from low to high frequencies (minimizing T2 effects) for inversion times of 107 to 1857 ms in increments of 125 ms. Measurements were separated by a 20‐s gap to guarantee full T1 recovery. The T2 was obtained by multi‐echo spin echo sampled at 32 echoes (each echo samples an image with different T2‐weighted contrast) with echo times ranging from 15 to 480 ms in increments of 15 ms and a fixed TR of 2000 ms. To minimize imperfect refocusing contributions, only the even echoes were used to estimate T2 decay, and both T1 and T2 maps were estimated based on a least‐square criteria against their expected mono‐exponential decay curves 3, 22. Although spin‐echo methods may sometimes be described as “gold standard,” in the present context we note that they provide a robust means of estimating relaxation parameters by a completely different approach to the one under investigation, and therefore provide an independent point of reference. However, this distinctiveness may also lead to discrepancy with the family of methods under investigation. The magnetization transfer ratio (MTR) was also measured using the standard Philips product sequence ‐ transverse acquisition with voxel size of 1 × 1 × 2 for a FOV of 224 × 168 × 120, sinc‐shaped preparation with 19.3‐ms duration, max amplitude of 12.2 µT, and off‐resonance of 1100 Hz.

To validate the proposed CRLB framework, a phantom‐specific VFA protocol was designed and the resulting parameter estimation was compared with the baseline protocol defined previously.

Optimizing JSR

This work proposed that the CRLB can be used as a criterion to select the optimal JSR acquisition parameters FA, TR, and ϕRF. To achieve this, Equation [9] was minimized making use of the pattern search optimization routine, implemented in the MATLAB 2016a optimtool. The standard T1 and T2 grid defined previously was combined with field offsets ranging from −125 to 125 in 1 Hz increments. This routine is not guaranteed to find global optima, so the optimization was simply repeated for 10 randomly distributed starting values for FA, TR, and ϕRF The global optimum, if not obtained, is not expected to dramatically improve the estimation quality compared with the solution achieved, as in practice, several different local minima are found whose cost function demonstrate similar performance.

A key constraint that must be considered is the total imaging time, as it is always possible to improve performance simply by imaging for longer (ie, at the expense of efficiency). Working at fixed resolution (taken to be the same for all of the contributing acquisitions) with only one average, the total imaging time is proportional to the sum of the repetition times of the individual sequences in the protocol. For the considered baseline protocol 9, Ttotal=29.2 ms was taken as a fixed upper time limit for all solutions explored. Within this constraint, the number of SPGR (NSPGR) and bSSFP (NbSSFP) measurements can be varied along with the TR, ϕRF and FA for each acquisition. Solutions can be calculated using different values of TR for each acquisition, however, early exploratory work showed that the optimal acquisitions converged for TRbSSFP=min{TR} thus, for simplicity, we constrained bSSFP to have minimum TR. This time depends on the resolution, and particularly FA (specific absorption rate constraints), but is otherwise the same for both SPGR and bSSFP on our system. Allowing for a suitable range of FAs, we set TRmin=4.2 ms which sets the maximum possible number of acquisitions in the permitted Ttotal. The TRSPGR is allowed to freely vary, subject to TRSPGR+NbSSFPTRbSSFPTtotal in which an explicit sum is written because both the number of SPGR sequences and their individual repetition times are varied.

With five unknowns to be estimated θJSR=[ρr,ρi, T1,T2 and ΔΩ] a minimum of five measurements are required for a fully determined solution. In addition, given that equal echo times are used for all acquisitions, a minimum of two bSSFP measurements is required to estimate the field map ΔΩ.

The optimal TRSPGR, ϕRF and FAs were therefore determined for different combinations of NbSSFP and NSPGR, as expressed in Table 1 (grayed‐out cells are excluded as infeasible).

Table 1.

Optimization Options Explored within the Time Constraint Explored

NbSSFP
1 2 3 4 5
NSPGR 1
2
3
4

Note: Gray areas were not considered in the optimization search, as they are infeasible.

In Vivo Validation

In vivo scans were acquired in four healthy volunteers (two male, two female, mean age 27 (min = 20, max = 31)), who gave written informed consent according to local ethics requirements.

Each subject was imaged using the complete baseline protocol and an optimized set of parameters were obtained by making use of the proposed CRLB optimization framework. The optimized parameters are listed in the “Results” section.

Before relaxation‐map estimation, all images were skull‐stripped and aligned using standard FSL bet and flirt tools (www.fsl.fmrib.ox.ac.uk) 38, 39, 40. An automatic segmentation was also performed on the T1‐weighted SPGR image using the FSL fast tool (38‐40).

Individual T1 and T2 histograms were calculated to compare the estimation performance between baseline and optimized protocols 3, 41.

RESULTS

CRLB Numerical Validation

Figure 1 shows a comparison of the JSR CRLB pT1,2CRLB and Monte Carlo–predicted precision pT1,2MC for the baseline protocol of the full grid of brain relaxation times. The relative percentage difference between pT1,2CRLB and pT1,2MC and defined as ϵT1MCCRLB=(pT1,2MCpT1,2CRLB)/pT1,2CRLB is below 1%, as can be observed in the right column of Figure 1.

Figure 1.

Figure 1

Comparison between σCRLB2 (left column) and Monte Carlo simulations (middle column) predicted precisions for different T1 and T2 relaxation‐time combinations. The right‐hand column displays the percent different between pT1CRLB and pT1MC predictions ( ϵT1MCCRLB=(pT1,2MCpT1,2CRLB)/pT1,2CRLB)

Phantom Validation

The phantom had T1 = 2058 ± 27 ms and T2 = 185 ± 1 ms, as determined using the reference spin‐echo sequences and MTR = 5.8% ± 0.7%. Phantom data were used to validate the hypothesis that JSR precision outperforms the standard DESPOT1/2 approach, and allowed the proposed CRLB framework to be tested. To achieve this, an extensive set of phantom‐specific optimized acquisition protocols (Table 1) were obtained using the CRLB framework, and the two best protocols are reported in Table 2, together with the baseline parameters for comparison. These three protocols were acquired in the same scanning session, and the respective T1 and T2 distributions inside a 10‐mL region at the center of the phantom are compared in the box plots of Figure 2. For reference, a blue dotted line is drawn that marks the T1 and T2 values obtained using spin echo–based estimation.

Table 2.

Phantom‐Specific Acquisition Protocols TRbSSFP=4.2ms

graphic file with name MRM-79-234-g007.jpg

Note: Color bars represent the fraction of time that SPGR (orange) and bSSFP (blue) occupy in each protocol. TRbSSFP=4.2 ms in all cases. The JSR‐optimized solutions were obtained for the expected T1 and T2 values and independently assessed using spin‐echo methods. For each protocol, the FAs are listed in the upper row, and the bottom row lists the repeat times, for the SPGR and the phase increments for the bSSFP sequences, as indicated by the adjacent labels. Only JSR 1:4 and 2:4 are displayed, as they were the best two performing solutions from the explored set (Table 1).

Figure 2.

Figure 2

Box and whisker plots comparing T1 and T2 distributions within a 10‐mL region of interest, estimated using each of the acquisition protocols defined in Table 2. Each acquisition protocol is identified by the number of SPGR:bSSFP measurements. Baseline corresponds to the acquisition protocol adapted from (9). Vertical lines show the average spin echo–measured T1 and T2. Parameter estimation by DESPOT was processing for the baseline protocol only or JSR, as indicated.

The conventional two‐step fitting approach on the baseline protocol results in interquartile range estimations of 1162 ms for DESPOT1 T1 and 111 ms for DESPOT2‐FM T2. Significant improvement is achieved by performing the proposed JSR approach on the same data, with interquartile ranges reduced to 323 ms for T1 and 26 ms for T2. Further improvement is obtained by making use of the proposed CRLB optimization framework. The JSR‐1:4 and JSR‐2:4 interquartile T1 ranges are reported as 220 and 200 ms (32 and 38% reductions), respectively. Similar improvement can be observed in T2 estimation, in which the JSR‐1:4 and JSR‐2:4 interquartile ranges are 14 and 15 ms (46 and 43% reductions), respectively.

The percentage deviation of the median value relative to the reference T1 spin‐echo measurement DESPOT1, JSR baseline, JSR 1:4, and JSR 2:4 demonstrate 15.1, 11.0, 3.3, and 4.4% bias, respectively. On the same note, the T2 percentage deviation is reported as 20.7% for DESPOT2‐FM, 10.9% for JSR baseline, −11.4% for JSR 1:4, and 9.2% for JSR 2:4.

Optimizing JSR

The SPGR acquisitions have been criticized for their sensitivity to imperfect spoiling of transverse magnetization before each excitation pulse 3, 36, 42. To mitigate this, EPG simulation of the SPGR signal response for the parameters of interest was compared with the signal intensity predicted by the Ernst equation. Figure 3 demonstrates the expected percentage signal error ϵ=(SspgrSspgrEPG)/SspgrEPG for both WM ( ϵWM) and GM ( ϵGM) as a function of different TRSPGR and FA αSPGR To avoid solutions with a WM error larger than ±5%, the maximum allowed FA was set to 15o (red line in Fig. 3).

Figure 3.

Figure 3

Percentage deviation error ϵ=(SspgrSspgrEPG)/SspgrEPG between EPG and Ernst model for both WM ( ϵWM ) and ( ϵGM) as functions of repeat time and FA for the SPGR sequence. The red dotted lines represent the maximum allowed SPGR FA that is consistent with ensuring the error for WM is kept below 5%.

As defined in the “Methods” section, optimized solutions were sought using the CRLB framework for the different combinations of SPGR and bSSFP measurements presented in Table 1. The results can be seen in Table 3, which is color‐coded based on the fraction of time spent on either SPGR (orange) or bSSFP (blue) acquisitions. Each bar shows the parameters obtained for each explored combination of NSPGR:NbSSFP measurements, and is rank‐ordered based on their expected CFgrid value (Eq. [9]; last column in Table 3). All explored protocols are restricted to a self‐imposed total time constraint of Ttotal = 29.2 ms.

Table 3.

Brain‐Specific Optimized JSR Acquisition Protocols TRbSSFP=4.2ms

graphic file with name MRM-79-234-g008.jpg

Note: Baseline shown for reference. The 2:4 acquisition protocol was selected for in vivo acquisition. From top to bottom, optimized protocols are rank‐ordered based on their attained worse‐case percentage precision (100 × CFgrid − eq. 9) values. Each row is color‐coded based on the fraction of time spent on either SPGR (orange) or bSSFP (blue) acquisitions. All obtained protocols are subject to a Ttotal = 29.2 ms constraint.

In Vivo Experiments

In vivo validation compared the estimation maps obtained using conventional DESPOT1/2 and JSR fitting methods, with data acquired using the baseline protocol. Based on Table 3, a brain‐optimized JSR protocol with two SPGR and four bSSFP measurements was also acquired. Given the similar performance between 1:4 and 2:4, we chose the 2:4 acquisition, as collecting two SPGR scans provides a more robust solution for subjects who might not be able to be so still. For reference, representative axial slices of the different acquisition parameters are demonstrated in Figure 4.

Figure 4.

Figure 4

A representative axial slice of in vivo acquired data from the baseline (top row) and optimal (bottom row) protocols. Acquisition parameters for each image are given in Table 3, listed in the same order as presented here.

Exemplar estimation maps can be seen in Figure 5. Note that DESPOT 1/2 relaxation parameter maps (left column in Fig. 5) are clearly more noisy that the corresponding JSR maps (columns 2 and 3 in Fig. 5). In addition, the off‐resonance map obtained with the two‐step fit is much more contaminated by explicit anatomy (particularly the CSF spaces) than the corresponding estimate from the JSR process.

Figure 5.

Figure 5

Comparison among conventional DESPOT1/2 approach (left column), JSR with baseline protocol (middle column), and JSR optimized protocol (right column) estimation maps.

Because of its high SNR and contrast, the adult optimized SPGR with 8.2‐ms TR was used after brain extraction to segment CSF, WM and GM, as described in the “Methods” section. Tissue‐specific relaxation histograms were then plotted for all of the acquisition schemes and are summarized in Figure 6.

Figure 6.

Figure 6

Tissue‐specific histograms for in vivo acquisition. Top row compares the distributions obtained for Healthy Volunteer 2 (HV2) with conventional DESPOT1/2 (blue), JSR with baseline parameters (red), and JSR with optimized acquisition parameters (yellow). Bottom row compares JSR‐optimized histograms for all imaged volunteers. Tissue segmentation was obtained using the FSL fast tool.

Figures 6a to 6d show the WM/GM T1 and WM/GM T2 pixel‐parameter‐value‐frequency distributions obtained on Healthy Volunteer 2 (HV2) using DESPOT1/2 (blue), JSR with baseline acquisition (red), and JSR with optimized acquisition (yellow) parameters. As with the phantom data, JSR produces narrower estimation distributions compared with the conventional DESPOT1/2 approach. However, the estimation mean relaxation‐time values show significant shifts toward lower relaxation time values.

Figures 6e to 6h show the WM/GM T1 and WM/GM T2 histograms obtained using the optimized JSR acquisition for all of the healthy volunteers, demonstrating the reproducibility of the proposed method.

DISCUSSION

This work aimed to improve the precision of DESPOT1/2 based on the JSR approach, in which all of the key parameters are estimated in a single calculation. Figure 2 shows 72%/77% decrease in interquartile range of T1/T2 estimation between conventional DESPOT and JSR estimation. This is further corroborated by the in vivo data, in which visual assessment (Fig. 5) shows that the JSR approach reduces the noise of the estimated maps, more specifically in deep GM areas where the SNR is intrinsically lower. Tissue‐specific distributions in Figures 6a to 6d show narrower distributions for JSR compared with VFA DESPOT1/2. This is primarily due to conventional DESPOT estimation being limited by the inherently lower SNR of the SPGR acquisition. The noise distribution of T1 then limits T2 assessment and does not make full use of the high bSSFP SNR. Our approach makes full use of bSSFP information, diminishing the estimation variance for both relaxation times. Although more narrow, there are systemic differences between the estimated relaxation distributions, which are dependent on the specific set of acquisition parameters used, but these are highly reproducible (Figs. 6e–6h). Numerical simulations using the described single‐pool model suggest that minimal to no difference on the average estimated values should occur between the estimation procedures, even when taking into account errors induced by neglecting T2. Indeed, not taking this extra decay into account induces higher JSR‐estimated T1 values (opposite of what is seen in Figs. 3 and 6), as the reduction in signal is accommodated as higher saturation of the SPGR curve. Another possible source of error is persistent, residual nonlinearities that may occur in the RF power amplifier. One possible solution is to follow the work presented by Lutti et al 43 and fix the RF amplifier power and vary the RF pulse duration to achieve different FAs. In our work, we opted to keep a fixed pulse duration and performed a vendor calibration step on each subject, to allow pre‐emphasis of the nonlinearities. This approach has the advantage of excitation off‐resonance deviations being maintained consistent between different measurements. Phantom data and in vivo discrepancies are most likely the result of magnetization transfer effects, which were not taken into account by the model used in this work, but are known to strongly affect the signal of bSSFP 44, 45, which is now used to jointly estimate T1 and T2. To corroborate this, the same phantom protocol was acquired on a gel phantom with 4% agarose solution (measured MTR of 20%). For this phantom, the bias between spin echo and VFA measurements was between 5 and 43% for T1, and −18% and 22% for T2. This is a larger variation than for the 0.5% agarose experiment summarized in Figure 2, in which the range of obtained bias is 4.4 to 15.1% for T1 and 11.4 to 20.7% for T2. In Figures 6e to 6h, the optimal JSR protocol tissue‐specific histograms for each of the healthy volunteers are overlaid for comparison. Excellent agreement is shown among all subjects for WM. Relaxation‐times histogram for GM show slightly larger deviations among subjects, although a good agreement is still present. This is most likely because of differences in the brain‐extraction step of our processing pipeline, although some of the discrepancy can also be attributed to genuine interindividual variance.

The in vivo T2 values show a significant underestimation when compared with other studies 46, 47, 48. However, this was not the case for the phantom experiment; in fact, failing to apply the TRF correction (8) for the phantom data resulted in an overestimation of the obtained T2 values by 15 to 20% compared with the ones displayed in Figure 2. This gives confidence that the correction is appropriate, at least for the phantom, which has a much lower MTR than adult brain 4, 49, although it is still not zero.

The second part of this work focused on the validation of the proposed CRLB framework as a protocol design tool. Numerical validations (Fig. 1) suggest good agreement between the proposed estimation procedure variance and the predicted CRLB (percentage difference < 1%), without affecting the estimation bias, giving confidence in both the fitting routine and the optimization tool. To make all results comparable, and especially because varying the total acquisition time changes the overall SNR in all scenarios, we optimized the SPGR and bSSFP measurements subject to a self‐imposed time constraint Ttotal = 29.2 ms, which matches the baseline protocol taken from the literature 9. Sensitivity of the SPGR measurement to imperfect spoiling was avoided by restricting the maximum allowed SPGR FA to 15o. This cut‐off guarantees that the expected WM spoiling inconsistencies are below 5%, whereas GM are kept bellow 8% for all optimization protocols explored. Although some solutions might benefit from allowing higher SPGR FA, this would imply either exploring optimal ways to weight Equation [9], given the EPG versus SPGR deviations, or using a better spoiling scheme such as the one presented in 50. Both solutions are outside the scope of this work, as the two best protocols obtained are expected to induce a signal bias of less than 3% for both WM and GM. Table 1 summarizes the different acquisition options explored in this paper; their respective cost function (CF) figures of merit are summarized in Table 3. The standard approach requires a minimum of two SPGR measurements to estimate T1; we note that performing JSR allows the acquisition to require only one SPGR measurement. This is interchangeable with increasing the number of SPGR measurements, provided that the same time is spent sampling SPGR information (Tables 2 and 3), which is further validated in Figure 2, where experimental data corroborate this interchangeability. Both optimal protocols result in a T1 and T2 interquartile‐range reduction of 81 and 87%, respectively, in comparison to the two‐step fitting process, and 32 and 43% relative to the JSR fitting using the baseline protocol. Allowing a single SPGR measurement could be of significant practical importance, as SPGR acquisitions are routinely used for clinical diagnosis. Therefore, full brain relaxometry can be achieved by adding four bSSFP measurements, which corresponds to less than 5‐min extension of the total scan time.

The JSR framework presented can help achieve optimal acquisition efficiency. In this study, full 3D 0.8‐mm3 isotropic relaxation maps of the human brain were obtained with a total acquisition time of 11 min and 18 s (9:18 for SPGR and bSSFP data, plus 2:00 for B1 field estimation). Although we restricted our acquisition time to a baseline protocol for comparison, it would be possible to trade off the JSR gain in precision for further reduction of the acquisition time.

The work presented here is a proof of concept and demonstrates the benefit of simultaneously using images with different signal responses to minimize variance in relaxometry estimation. Although neglected throughout this work, the JSR framework can be expanded to estimate T2' by allowing different echo times between acquisitions. In addition, it is feasible to include B1 field estimation directly from the VFA data by incorporating an inversion recovery sequence (eg, MPRAGE) into the joint system model. Investigation of such protocols showed that the CRLB design framework can be applied, but initial in vivo testing with JSR processing resulted in T1 values with higher systematic deviations for different acquisition parameters (eg, inversion time, RAGE block duration), compared with what is presented in this work. We attribute this to the very different RF conditions engendered by such sequences, causing a model bias as a result of MT effects that are not considered (data not shown). This is currently the subject of further investigation; in particular, the option to explicitly include MT in the JSR approach is an interesting future possibility. The JSR approach has a lot in common with the recently introduced fingerprinting concept (MRF) 51, 52, in that all acquired data are used to estimate the final relaxation and associated parameters. The relative merits of this optimized framework and MRF, both in terms of precision and efficiency, remain to be explored.

CONCLUSIONS

This work shows an immediate benefit of the proposed JSR analysis approach, compared with conventional DESPOT1/2, in producing relaxation maps with improved precision. It also shows that further improvement can be achieved by making use of the CRLB as a protocol design tool. With this approach, it is possible to achieve submillimeter maps of ρ,T1,T2, and B0 in an 11‐min examination, making the approach appealing for potential clinical use. For examinations in which the clinical protocol already includes SPGR, the additional time needed to achieve quantitative relaxometry maps is even smaller ( < 7 additional minutes, including B1 calibration in the example shown). As with all variable FA approaches, the absolute values of the relaxation times found tend to deviate systematically from spin echo–based measurements. However, high reproducibility combined with efficiency endow these methods with significant advantages for larger‐scale studies. The proposed framework enhances both precision and efficiency, further adding to their potential utility.

REFERENCES

  • 1. Meadowcroft MD, Mutic NJ, Bigler DC, Wang J‐L, Simmons Z, Connor JR, Yang QX. Histological‐MRI correlation in the primary motor cortex of patients with amyotrophic lateral sclerosis. JMRI 2015;41:665–675. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Kitzler H, Su J, Zeineh M, Harper‐Little C. Deficient MWF mapping in multiple sclerosis using 3D whole‐brain multi‐component relaxation MRI. NeuroImage 2012;59:2670–2677. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Stikov N, Boudreau M, Levesque IR, Tardif CL, Barral JK, Pike GB. On the accuracy of T1 mapping: searching for common ground. Magn Reson Med 2014;0:1–9. [DOI] [PubMed] [Google Scholar]
  • 4. Stanisz G, Odrobina E. T1, T2 relaxation and magnetization transfer in tissue at 3T. Magn Reson Med 2005;54:507–512. [DOI] [PubMed] [Google Scholar]
  • 5. Christensen KA, Grant DM, Schulman EM, Walling C. Optimal determination of relaxation times of fourier transform nuclear magnetic resonance. Determination of spin‐lattice relaxation times in chemically polarized species. J Phys Chem 1974;78:1971–1977. [Google Scholar]
  • 6. Deoni S, Rutt B, Peters T. Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady state. Magn Reson Med 2003;49:515–526. [DOI] [PubMed] [Google Scholar]
  • 7. Deoni S, Peters T, Rutt B. High‐resolution T1 and T2 mapping of the brain in a clinically acceptable time with DESPOT1 and DESPOT2. Magn Reson Med 2005;53:237–241. [DOI] [PubMed] [Google Scholar]
  • 8. Crooijmans H. Finite RF pulse correction on DESPOT2. Magn Reson Med 2011;65:858–862. [DOI] [PubMed] [Google Scholar]
  • 9. Deoni SSCLS. Transverse relaxation time (T2) mapping in the brain with offresonance correction using phase‐cycled steady‐state free precession imaging. J Magn Reson Imaging 2009;30:411–417. [DOI] [PubMed] [Google Scholar]
  • 10. Deoni S, Rutt B. Gleaning multicomponent T1 and T2 information from steady state imaging data. Magn Reson Med 2008;60:1372–1387. [DOI] [PubMed] [Google Scholar]
  • 11. Deoni S. One component? Two components? Three? The effect of including a nonexchanging “free” water component in multicomponent driven equilibrium single pulse. Magn Reson Med 2013;70:147–154. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Deoni SCLS, Kolind SH. Investigating the stability of mcDESPOT myelin water fraction values derived using a stochastic region contraction approach. Magn Reson Med 2014;73:161–169. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Lankford C, Does M. On the inherent precision of mcDESPOT. Magn Reson Med 2013;69:127–136. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Zhang J, Kolind S. Comparison of myelin water fraction from multiecho T2 decay curve and steady‐state methods. Magn Reson Med 2014;73:223–232. [DOI] [PubMed] [Google Scholar]
  • 15. Brihuega‐Moreno O. Optimization of diffusion measurements using Crámer Rao lower bound theory and its application to articular cartilage. Magn Reson Med 2003;50:1069–1076. [DOI] [PubMed] [Google Scholar]
  • 16. Cercignani M, Alexander D. Optimal acquisition schemes for in vivo quantitative magnetization transfer MRI. Magn Reson Med 2006;56:803–810. [DOI] [PubMed] [Google Scholar]
  • 17. Wang HZ, Riederer SJ, Lee JN. Optimizing the precision in T1 relaxation estimation using limited flip angles. Magn Reson Med 1987;5:399–416. [DOI] [PubMed] [Google Scholar]
  • 18. Deoni S. High‐resolution T1 mapping of the brain at 3T with driven equilibrium single pulse observation of T1 with high‐speed incorporation of RF field inhomogeneities (DESPOT1 HIFI). J Magn Reson Imaging 2007;26:1106–1111. [DOI] [PubMed] [Google Scholar]
  • 19. Helms G, Dathe H, Weiskopf N, Dechent P. Identification of signal bias in the variable flip angle method by linear display of the algebraic ernst equation. Magn Reson Med 2011;66:669–677. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Wood TC. Improved formulas for the two optimum VFA flip‐angles. Magn Reson Med 2015;74:1–3. [DOI] [PubMed] [Google Scholar]
  • 21. Homer J, Beevers M. Driven‐equilibrium single‐pulse observation of T1 relaxation. A reevaluation of a rapid “new” method for determining NMR spin‐lattice relaxation. J Magn Reson 1985;63:287–297. [Google Scholar]
  • 22. Haacke E, Brown R, Thompson M, Venkatesan R. Magnetic Resonance Imaging: Physical Principles and Sequence Design. New York: John Wiley & Sons; 1999. [Google Scholar]
  • 23. Deoni S, Ward H. Rapid T2 estimation with phase‐cycled variable nutation steady‐state free precession. Magn Reson Med 2004;52:435–439. [DOI] [PubMed] [Google Scholar]
  • 24. Freeman R, Hill HD. Phase and intensity anomalies in fourier transform NMR. J Magn Reson 1971;4:366–383. [Google Scholar]
  • 25. Bieri O, Scheffler K. SSFP signal with finite RF pulses. Magn Reson Med 2009;62:1232–1241. [DOI] [PubMed] [Google Scholar]
  • 26. Scheffler K, Hennig J. Is TrueFISP a gradient‐echo or a spin‐echo sequence? Magn Reson Med 2003;49:395–397. [DOI] [PubMed] [Google Scholar]
  • 27. Preibisch C, Deichmann R. Influence of RF spoiling on the stability and accuracy of T1 mapping based on spoiled FLASH with varying flip angles. Magn Reson Med 2009;61:125–135. [DOI] [PubMed] [Google Scholar]
  • 28. Weigel M. Extended phase graphs: Dephasing, RF pulses, and echoes—pure and simple. J Magn Reson Imaging 2014;41:266–295. [DOI] [PubMed] [Google Scholar]
  • 29. Malik S, Padormo F. Spatially resolved extended phase graphs: Modeling and design of multipulse sequences with parallel transmission. Magn Reson Med 2012;68:1481–1494. [DOI] [PubMed] [Google Scholar]
  • 30. Hennig J. Echoes‐how to generate, recognize, use or avoid them in MR‐imaging sequences. I: Fundamental and not so fundamental properties of spin echoes. Concepts Magn Reson 1991;3:125–143. [Google Scholar]
  • 31. Nehrke K. On the steady‐state properties of actual flip angle imaging (AFI). Magn Reson Med 2009;61:84–92. [DOI] [PubMed] [Google Scholar]
  • 32. Nielsen F. Cramer‐Rao Lower Bound and Information Geometry 2013. arXiv:1301.3578 [cs.IT].
  • 33. Alexander D. A general framework for experiment design in diffusion MRI and its application in measuring direct tissue microstructure features. Magn Reson Med 2008;60:439–448. [DOI] [PubMed] [Google Scholar]
  • 34. Gelman N, Gorell JM, Barker PB, Savage RM, Spickler EM, Windham JP, Knight RA. MR imaging of human brain at 3.0 T: preliminary report on transverse relaxation rates and relation to estimated iron content. Radiology 1999;210:759–767. [DOI] [PubMed] [Google Scholar]
  • 35. Nehrke K, Börnert P. DREAM—a novel approach for robust, ultrafast, multislice B1 mapping. Magn Reson Med 2012;68:1517–1526. [DOI] [PubMed] [Google Scholar]
  • 36. Yarnykh VL. Optimal radiofrequency and gradient spoiling for improved accuracy of T1 and B1 measurements using fast steady‐state techniques. Magn Reson Med 2010;63:1610–1626. [DOI] [PubMed] [Google Scholar]
  • 37. Yarnykh VL. Actual flip‐angle imaging in the pulsed steady state: a method for rapid three‐dimensional mapping of the transmitted radiofrequency field. Magn Reson Med 2007;57:192–200. [DOI] [PubMed] [Google Scholar]
  • 38. Woolrich MW, Jbabdi S, Patenaude B, Chappell M, Makni S, Behrens T, Beckmann C, Jenkinson M, Smith SM. Bayesian analysis of neuroimaging data in FSL. NeuroImage 2009;45:S173–S186. [DOI] [PubMed] [Google Scholar]
  • 39. Smith SM, Jenkinson M, Woolrich MW, et al. Advances in functional and structural MR image analysis and implementation as FSL. NeuroImage 2004;23:208–219. [DOI] [PubMed] [Google Scholar]
  • 40. Jenkinson M, Beckmann CF, Behrens TEJ, Woolrich MW, Smith SM. FSL. NeuroImage 2012;62:782–790. [DOI] [PubMed] [Google Scholar]
  • 41. Deoni S, Williams S, Jezzard P. Standardized structural magnetic resonance imaging in multicentre studies using quantitative T1 and T2 imaging at 1.5 T. NeuroImage 2008;40:662–671. [DOI] [PubMed] [Google Scholar]
  • 42. Heule R, Ganter C, Bieri O. Variable flip angle T1 mapping in the human brain with reduced T2 sensitivity Using fast radiofrequency‐spoiled gradient echo imaging. Magn Reson Med 2015;75:1413–1422. [DOI] [PubMed] [Google Scholar]
  • 43. Lutti A, Weiskopf N. Optimizing the accuracy of T1 mapping accounting for RF non‐linearities and spoiling characteristics in FLASH imaging. In Proceedings of the 21st Annual Meeting of ISMRM, Salt Lake City, Utah, USA, 2013. p. 2478.
  • 44. Gloor M, Scheffler K, Bieri O. Quantitative magnetization transfer imaging using balanced SSFP. Magn Reson Med 2008;60:691–700. [DOI] [PubMed] [Google Scholar]
  • 45. Bieri O, Scheffler K. On the origin of apparent low tissue signals in balanced SSFP. Magn Reson Med 2006;56:1067–1074. [DOI] [PubMed] [Google Scholar]
  • 46. Deoni S. Transverse relaxation time (T2) mapping in the brain with offresonance correction using phase‐cycled steady‐state free precession imaging. J Magn Reson Imaging 2009;30:411–417. [DOI] [PubMed] [Google Scholar]
  • 47. Ben‐Eliezer N, Sodickson DK, Block KT. Rapid and accurate T2 mapping from multi‐spin‐echo data using bloch‐simulation‐based reconstruction. Magn Reson Med 2015;73:809–817. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48. Jutras JD, Wachowicz K, De Zanche N. Analytical corrections of banding artifacts in driven equilibrium single pulse observation of T2 (DESPOT2). Magn Reson Med 2016;76:1790–1804. [DOI] [PubMed] [Google Scholar]
  • 49. Sled JG, Pike GB. Quantitative imaging of magnetization transfer exchange and relaxation properties in vivo using MRI. Magn Reson Med 2001;46:923–931. [DOI] [PubMed] [Google Scholar]
  • 50. Hess AT, Robson MD. Hexagonal gradient scheme with RF spoiling improves spoiling performance for high‐flip‐angle fast gradient echo imaging. Magn Reson Med 2016;0:1–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51. Ma D, Gulani V, Seiberlich N, Liu K. Magnetic resonance fingerprinting. Nature 2013;495:187–192. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52. Ma D, Hamilton J, Jiang Y, Seiberlich N, Griswold M. Fast 3D magnetic resonance fingerprinting (MRF) for whole brain coverage in less than 3 minutes. In Proceedings of the 24th Annual Meeting of ISMRM, Singapore, 2016. Abstract 3180.

Articles from Magnetic Resonance in Medicine are provided here courtesy of Wiley