Abstract
Purpose
To a) achieve cardiac 19F-Magnetic Resonance Imaging (MRI) of perfluoro-crown-ether (PFCE) labeled cardiac progenitor stem cells (CPCs) and bone-derived bone marrow macrophages, b) determine label concentration and cellular load limits, and c) achieve spectroscopic and image-based quantification.
Methods
Theoretical simulations and experimental comparisons of spoiled-gradient echo (SPGR), rapid acquisition with relaxation enhancement (RARE), and steady state at free precession (SSFP) pulse sequences, and phantom validations, were conducted using 19F MRI/Magnetic Resonance Spectroscopy (MRS) at 9.4 T. Successful cell labeling was confirmed using flow cytometry and confocal microscopy. For CPC and macrophage concentration quantification, in vitro and post-mortem cardiac validations were pursued with the use of the transfection agent FuGENE. Feasibility of fast imaging is demonstrated in murine cardiac acquisitions in vivo, and in post-mortem murine skeletal and cardiac applications.
Results
SPGR/SSFP proved favorable imaging sequences yielding good signal-to-noise ratio values. Confocal microscopy confirmed heterogeneity of cellular label uptake in CPCs. 19F MRI indicated lack of additional benefits upon label concentrations above 7.5–10 mg/ml/million cells. The minimum detectable CPC load was ~500k (~10k/voxel) in two-dimensional (2D) acquisitions (3–5 min) using the butterfly coil. Additionally, absolute 19F based concentration and intensity estimates (trifluoroacetic-acid solutions, macrophages, and labeled CPCs in vitro and post-CPC injections in the post-mortem state) scaled linearly with fluorine concentrations. Fast, quantitative cardiac 19F-MRI was demonstrated with SPGR/SSFP and MRS acquisitions spanning 3–5 min, using a butterfly coil.
Conclusion
The developed methodologies achieved in vivo cardiac 19F of exogenously injected labeled CPCs for the first time, accelerating imaging to a total acquisition of a few minutes, providing evidence for their potential for possible translational work.
Introduction
Implantation of stem cells (SCs) has provided a methodological pathway that promises cardiac tissue regeneration and structural and functional improvements following injury. The basic approach of SC therapy involves the direct transplantation of cells, followed by their migration, differentiation, and proliferation, ultimately attaining homing and engraftment. However, while the feasibility of SC technologies has been proven, efficacy is still in question [1].
Within the realm of SC therapies, non-invasive imaging and tracking of labeled SCs, and their functional impact, has taken a prominent role in recent years. The visualization of the implanted SCs to define optimal therapy strategies (dose, timing, delivery) using pre-labeled cells with fluorescent probes [2], transduced expression of fluorescent proteins [3], or iron oxide particles (MPIOs) [4], and their assessment for temporal label persistence, has become a subject of intense research. Over the past decade, nanoparticles (NPs) containing perfluoro-crown-ethers (PFCE) have led to direct tracking and quantification of exogenously labeled cell populations [5, 6, 7, 8] with 19F magnetic resonance imaging (MRI).
Despite the implementation of 19F MRI early on in the development of MRI, exploitation efforts had languished until recent years [5, 6, 9, 10]. The resurgence of interest in 19F imaging arose further to initiatives in molecular imaging, and capitalized on the exogenously injected fluorine’s 100% abundance, and the high relative sensitivity and contrast with respect to the 1H nucleus. The lack of endogenous fluorine provides fluorinated labels an added advantage as tracking agents. Consequently, the technique has found applicability in cellular labeling and tracking applications in vivo [5, 11], with potential for translational value [12].
Furthermore, prior applications were confined to either direct injections of neural SCs [13], immune cells [6, 7, 9], hematopoietic SCs [14], or on direct intravascular administrations of NP emulsions [15, 16, 17, 18, 19, 11]. Correspondingly, there have been no prior reported 19F MRI preclinical applications in normal or infarcted hearts using exogenously administered, labeled progenitor SCs, while direct applications of other types of SCs in the rodent heart have been limited [14].
Prior efforts attempted to optimize fluorine acquisitions in terms of speed, evoked MR signal, and cellular detectability [11], by focusing on spectroscopy [17, 18, 19] or on dedicated pulse sequences [20–26], and by selecting imaging parameters that elicited maximum signal responses, despite the prohibitively long acquisition times. To our knowledge, there is no prior 19F MRI study on the use of labeled cardiac progenitor cells (CPCs) (previously used to show efficacy of regeneration and cardiac functional improvements [27]). Certainly, lacking are also detailed relaxometric studies in these cells post-labeling.
We present a comprehensive methodology that applies 19F MRI aiming to achieve: a) fast imaging of PFCE-labeled CPCs within clinically relevant times (of the order of a few minutes) in the in vivo mouse, b) determination of detection limits of label cellular load with clinically applicable surface and volume coils, and c) spectroscopy and image-based quantification validated in phantoms, CPCs, labeled bone-marrow-derived murine macrophages, and in the post-mortem mouse. The stated objectives were investigated based on theoretical and simulation comparisons of pulse sequence performances, in vitro relaxation value characterization of PFCE-labeled CPCs, experimental concentration validations, and post-mortem and in vivo applicability of the imaging approach in the cardiac and skeletal muscles of the C57BL/6 mouse.
Materials and methods
Animal ethics
All procedures were in accordance with the Home Office (UK) guidelines under The Animals (Scientific Procedures) Act, 1986 (Permit Number: PIL30/3322), the European Animal Research Directive, and with local institutional guidelines. All surgery and live animal imaging was performed under isoflurane (ISO) anesthesia, and all efforts were made to minimize suffering. Animals were euthanized using cervical dislocation.
Cardiac progenitor stem cells and bone marrow-derived macrophages
Nanoparticle synthesis
Particles were synthesized in accordance to previously published methodologies [7]. All particles were then extensively washed with distilled water and lyophilized for 2–3 d. For the cell labeling, nanoparticles were synthesized with the addition of fluorescent dye (Atto647, ATTO-TEC, GmbH, Germany) to the organic phase. Prior 19F Magnetic Resonance Spectroscopy (MRS) characterization of PFCE NP labels confirmed the presence of a single spectral peak at -91.8 ppm (with respect to CFCl3) [11].
Cell isolation
Cardiac progenitor cells (CPCs, comprising either cardiosphere-derived (CDC) or collagenase-trypsin (CT)) were isolated from adult, C57BL/6, green fluorescent protein (GFP) positive or GFP negative, mouse atria, using standard protocols [28], and were maintained in Iscove’s Modified Dulbecco’s Medium (IMDM) media (Thermo Fisher Scientific, UK). CDCs have been previously used to show efficacy of regeneration and cardiac functional improvements [27], while CTs have been recently shown to express similar cardiac phenotypic characteristics to CDCs [29]. Additionally, bone marrow-derived macrophages were cultured from bone marrow harvested from C57BL/6 mouse hindlimbs. Bone marrow cells were washed with phosphate buffer solution (PBS, Sigma-Aldrich, UK), passed through a cell strainer to produce a single cell suspension, and differentiated for a week in petri-dishes in Dulbecco’s modified eagle’s medium (DMEM) that contained L-cell conditioned media. At that point, adherent bone-marrow derived macrophages were harvested. They were subsequently washed with PBS and were re-suspended in pellets in improved minimal essential medium (OPTIMEM).
Cellular culture and labeling
Cells were plated in IMDM solutions and incubated with PLGA-PFCE-Atto647-containing NPs for approximately 24 h before isolation and pelleting. Addition of a fluorescent dye (Atto647) allowed independent flow cytometry and confocal microscopy validation studies. Cell pellet suspensions (CPCs or macrophages) were maintained in media (IMDM or OPTIMEM) and were subsequently used for MRI, flow cytometry, or confocal microscopy, after fixation in 2% methanol-free paraformaldehyde solution (Thermo Scientific Pierce, UK) mixed with PBS (1:7 v/v).
FuGENE labeling
The commercially available DNA transfection agent FuGENE (Promega, Madison, WI, USA) was used to label the CPCs (both CT and CDC cells in separate cultures) with the NPs (using 25 μl of FuGENE in ~106 cells). FuGENE was pre-mixed and incubated with the NPs before cell transfection for ~20 min. Cells were then labeled overnight [30].
Confirmation of cellular label uptake and viability
Successful labeling was confirmed with a CyAn ADP flow cytometer (Beckman Coulter, USA) using control and labeled cell samples. Cellular viability was determined with the Trypan Blue exclusion assay, directly after labeling, and at the completion of MRI studies (wherever applicable), using a cell counter.
High-content epifluorescence imaging
Live cells were plated in 6-well plates and stained with Calcein (CellTrace™ Calcein Red-Orange, ThermoFisher Scientific, UK) and Hoechst (ThermoFisher Scientific, UK) for cytoplasmic and nuclear high-content imaging (Operetta, Perkin-Elmer, UK). Fluorescence was assessed based on Atto647. The Operetta’s Harmony software was used for image analyses. Imaging was based on a randomized field analysis methodology that covered each of the studied wells.
MRI/MRS
All experiments were conducted on a 9.4 T Agilent scanner equipped with a DirectDrive console and a 1000 mT/m actively shielded gradient set (internal diameter = 60 mm) (Agilent Technologies, USA). For comparative tests (pulse sequences, radiofrequency (RF) coils), the same acquisition parameters and total acquisition times were used.
Aqueous and cellular phantoms
Twenty-four cylindrical (15–50 ml) phantoms (1–100 mM), containing trifluoro-acetic acid (TFA), PFCE NPs, and labeled CPCs or macrophages, were used to test RF coil responses at 9.4 T (without, and with the use of adiabatic excitations), determine detection limits (TFA, NP solutions mixed in water and IMDM, labeled cells), and for image- and spectroscopy-based quantification.
RF coils
Coils comprising an eight-rung, low-pass, quadrature birdcage (diameter = 34 mm), an 40×20 mm2 butterfly constructed on a 28 mm diameter former, and a 5 (diameter) ×8 (length) mm2 solenoid prototypes, were constructed. All coils were tuned at 375.88 MHz (fluorine resonance), and were matched to 50 Ω. The broad frequency response of the coils (3dB range spanning a few tens of kHz) permitted imaging on both the 1H and 19F nuclei.
Spoiled gradient echo (SPGR) imaging allowed surface coil comparisons (without, and with the use of adiabatic excitation, as described below), using aqueous TFA phantoms (range of concentrations = 0.5–100 mM) with the following imaging parameters: TR = 6.3 ms, TE = 3.2 ms, flip angles = 50° (non-adiabatic), and 190° and 205° (adiabatic) (with comparable power settings for non-adiabatic and adiabatic acquisitions), NEX = 8, BW = 6 kHz, field-of-view (FOV) = 40×40 mm2, slice thicknesses (ST) = 2 mm, matrix = 32×32. Mid-axial profiles were obtained from reconstructed images and compared.
The butterfly coil was used for murine studies owing to its increased B1 detection sensitivity. The increased B1 homogeneity of the birdcage and solenoid coils allowed the a) assessment of the minimum detectable number of 19F atoms (and the fluorine content of the PLGA-NPs), and b) the estimation of relaxation times.
Adiabatic excitation
Phantom studies (six Eppendorf vials containing aqueous TFA solutions at 5–10 mM) were conducted without (Gaussian RF pulse excitation) and with the use of hyperbolic adiabatic full-passage (HS-AFP) RF pulses [31] using the butterfly coils to determine ultimate concentration detection limits and RF B1 penetration. Since HS-AFP pulses are non-ideal in achieving proper slice selection, three-dimensional (3D) adiabatic acquisitions were performed, whereby excitation of a thick tissue slab was accompanied by imaging of only a relatively smaller tissue area (relevant to the injected labeled CPCs).
Adiabatic pulse parameters were chosen appropriately to achieve adiabaticity, yet maintaining power to comparable levels as those used for Gaussian (non-adiabatic) excitations [31] (1H MRI: TR/TE = 13.6/1.72 ms, flip angle = 50° (versus 180° for HS-AFP adiabatic excitation), NEX = 16, BW = 50 kHz, FOV = 40×40 mm2, ST = 3 mm, matrix = 128×128; 19F MRI: TR/TE = 8.7/4.4 ms, flip angle = 50° or 190/205° for HS-AFP, NEX = 8 or 1248, BW = 4 or 8 kHz, FOV = 40×40 mm2, ST = 2 mm, matrix = 32×32, BW HS-AFP = 1.35 kHz, resolution = 0.4 μs, cutoff = 2 or 4%).
Simulations
SPGR, rapid acquisition with relaxation enhancement (RARE), and fid/echo steady state free precession (fid-SSFP, echo-SSFP) sequences were simulated in accordance with steady-state, closed-form, signal and signal-to-noise ratio (SNR) equation formulations, as described in the S1 Appendix. Parametric SNR maps were generated in MATLAB (Version 2010b, Mathworks, Natick, MA, USA) using typical 19F relaxation times and imaging parameters for TFA solutions and CPCs, as these were determined in this study. Estimated SNR values for SPGR and SSFP sequences were normalized to the maximum signal over the entire parametric space. SNR normalization in the case of RARE adhered to the recent analysis presented by Mastropietro et al. [25].
Pulse sequence comparison
Extensive prior in vivo preclinical work has favored RARE imaging [25], and more recently bSSFP [24]. The choice of the optimal pulse sequence for 19F MRI was based on experimental work using the homogeneous birdcage coil by comparing the SNR responses of SPGR, RARE, and SSFP sequences, based on 2D acquisitions of a 100 mM TFA solution. The imaging parameters were: (SPGR/SSFP: TR/TE = 6.7 and 6.8/3.4 ms, flip angle = 50/30°, NEX = 988/580, 32×32, FOV = 40×40 mm2, ST = 2 mm, BW = 6 kHz, and RARE: TR/TE = 1100/10.5 ms. NEX = 96, NEX = 1024, 32×32, ETL = 32, FOV = 40×40 mm2, ST = 2 mm, BW = 6 kHz, total acquisition time = 3.5 min).
SNR maps were generated by dividing the reconstructed images by the noise SD, estimated from background regions using standard methodologies. Given the relatively large SNR values (>10) of reconstructed images for cell and phantom studies, no bias corrections for the magnitude reconstruction were applied.
Relaxation measurements
To allow SNR optimization and direct image-based quantification, T1 and T2 measurements of aqueous TFA, NaF, NP solutions, and labeled (without and with the use of FuGENE) CPCs, were conducted with the birdcage/solenoid coils using conventional inversion recovery and Carr-Purcell-Meiboom-Gill (CPMG) pulse sequences, with the following imaging parameters: a) T1: TR = 5–8 s, 512 points, NEX = 2–16, BW = 4 kHz; b) T2: TR = 5 s, TE = 2 ms, 2048 points, NEX = 8, BW = 4 kHz.
For the relaxation measurements of CTs, cells were suspended in Eppendorf tubes and were maintained in IMDM media in ice-cold baths at 4°C and under normoxic conditions (normal oxygen tension). The tubes were then allowed to reach room temperature before measurements were conducted. The partial pressures of oxygen (pO2) of the tested solutions were not monitored during experiments.
In comparison to the T1 values, the T2 values of labeled CT cells were not quantified owing to the low 19F signal elicited from these labeled cells, often leading to exceedingly long acquisition times using nonlocalized spectroscopy that may lead to compromised oxygenation status and viability.
19F detection threshold in TFA and nanoparticle solutions
To assess the MRI detection threshold and the optimal labeling dose (SPGR vs. SSFP), a a) TFA, and b) a phantom with PFCE label solutions were imaged with the birdcage coil using conventional 1H SPGR (TR/TE = 4.9/2.47ms, flip angles = 20°, NEX = 1, BW = 50 kHz, FOV = 40×40 mm2, ST = 2 mm, matrix = 128×128) and 19F SPGR and SSFP sequences (TR/TE = 4.9–6.37/2.47–3.2ms, flip angles = 50–60°, NEX = 512, BW = 4–6 kHz, FOV = 40×40 or 60×60 mm2, ST = 2 mm, matrix = 32×32). Both phantoms were constructed with 0.7 and 1.7 ml Eppendorf tubes containing: a) one 100 mM TFA reference standard (middle vial) and five TFA solutions (1–100 mM), b) a 25 mM TFA reference standard (middle vial) and six PLGA-PFCE solutions at different NP concentrations (2.5–10 mg/ml, or equivalently, 0.32–3.2 mM).
Detection threshold in labeled cells in vitro
Multiple CT cells (1, 0.75, 0.5, 0.25 million cells) were seeded, and labeled with FuGENE with NPs at a concentration of 10 mg/ml/million cells. The cells were subsequently suspended, maintained in IMDM media, and placed in multiple Eppendorf tubes for imaging using SPGR with the butterfly coil (1H: TR/TE = 62.34 ms, flip angle = 50°, NEX = 2, BW = 20 kHz, FOV = 40×40 mm2, slice thickness = 2 mm, 8 slices, matrix = 128×128; 19F: TR/TE = 8.54/4.29ms, flip angle = 180°, NEX = 1024, BW = 6 kHz, FOV = 40×40 mm2, slice thickness = 20 mm, matrix = 32×32, pulse width (HS-AFP) = 3 ms, BW = 1.35 kHz, resolution = 4 μs, and 4% cutoff. For MRS, the solenoid acquisitions used the following parameters: TR = 500 ms, 512 points, NEX = 256, BW = 20 kHz.
MRS and image-based concentration quantification
Implementation was achieved in phantom TFA solutions, in labeled CPCs, and in the post-mortem mouse following intracardial administration of CPCs using a setup that could be used in vivo.
In vitro quantification: Direct MRS and SPGR-image-based quantification was implemented using the birdcage coil and the multivial TFA phantom described above, based on region-of-interest (ROI) estimation. A reference calibration curve was generated with four TFA solutions (25–100 mM), whereby signals were estimated in a blinded fashion and referenced against a standard of known concentration (middle vial) (19F: TR/TE = 6.21/1.75, flip angle = 20°, NEX = 256, BW = 8 kHz, FOV = 40×40 mm2, ST = 5 mm, axial, matrix = 32×32).
Labeled CPCs: This effort was extended in CPCs (cell densities of 0.5, 0.75, 1, and 2 million) and in macrophages (cell densities ~2–6 million) using the solenoid (both cell types) and butterfly (CPCs using adiabatic excitation) coils. Imaging and quantification was achieved with MRS/MRI (1H: TR/TE = 4.9/2.47 ms, flip angle = 20°, NEX = 1, BW = 50 kHz, FOV = 40×40 mm2, ST = 15 mm, coronal, matrix = 128×128; 19F (SPGR): TR/TE = 3.46 or 11.71/1.75 or 5.87 ms, flip angle = 50° or 227.5°, NEX = 2048, BW = 10 kHz, FOV = 40×40 mm2, ST = 15 mm, coronal, matrix = 32×32).
To facilitate extension of the methodology to the post-mortem mouse, a 3D SPGR acquisition protocol was successfully tested in vitro (flip angle = 200°) using the butterfly coil and reference (800k FuGENE-labeled CPCs) and test (~1.3 million CPCs before their injection—please see (c) below) Eppendorf vials. The two vials were positioned on the surface of the butterfly coil at comparable positions to those used in the post-mortem case. T1 and T2 correction was achieved based on fully relaxed MRS or the steady-state, closed form, signal equations of SPGR and echo-SSFP, in accordance to Eqs 8 and 16 (S1 Appendix). CPC cell numbers and 19F content from samples with unknown content were extrapolated based on known cell densities of reference standards (validated using Trypan Blue), against a 25 mM TFA phantom.
Post-mortem mouse: For these tests, the butterfly coil was used. The reference vial was positioned on the anterior thorax. The cell pellet position was approximately at the same level (along the inferior-superior direction of the animal, inclined ~-45°) as the injection site location. Estimation of the injected cell number was based on the ratio of the total 19F MRI signal of the injected cells (upper mid-ventricular myocardium) and the reference CPC signal, obtained using the 3D SPGR adiabatic excitation (optimized in vitro—see (b) above) using a 220° flip angle.
Post-mortem and in vivo animal models—Skeletal and cardiac muscle applications
To assess the feasibility and reproducibility of the imaging protocol, labeled cells (~1.5–2.5 million suspended in ~50–100 μl of IMDM media) were injected in the femoral areas of mouse hindlimbs (n = 3), and in the anterior left ventricular (LV) muscle post-mortem (n = 3) (mouse strains included PHD3f/f, PHD2flox/flox, and C57BL/6). The mice were then positioned on the butterfly coil and imaged. 1H images were acquired with the 3D SPGR sequence using TR = 2.49 ms, TE = 1.26 ms, flip angle = 20°, NEX = 3, BW = 50–150 kHz, FOV = 40×40 mm2, matrix = 192×192×192, in 4.36 min. For 19F MRI (3D-SPGR) the acquisition parameters were: TR/TE = 5.64–2.84ms, flip angles = 20°, NEX = 48, BW = 6 kHz, FOV = 40×40 mm2, matrix = 32×32×32, in a total of 4.37 min.
In vivo cardiac muscle applications
Labeled cells (~1.5 million suspended in ~50 μl of IMDM media) were injected in two C57BL/6J mice (male weight range = 20–30 g) that underwent thoracotomy, followed by recovery. Induction was achieved using 4% ISO, and the mice were maintained with 1.5–2.0% ISO, mixed in 100% oxygen.
The mice were subsequently imaged. All animals were placed on a specially constructed cradle, and were allowed to breathe freely throughout the study. A homeostatically controlled hot-air system was used to maintain mouse body temperature at approximately 37°C. Electrocardiographic (ECG) and breathing rates were monitored using a gating system. Heart rates were maintained at 300–500 beats/min.
Ungated/gated, non-localized 19F MRS of the mouse thorax were acquired (following shimming at the 1H nucleus) first confirming the presence and resonance of PLGA-NPs (TR = 800–1200 ms, flip angle = 90°, NEX = 32–128, 512 points, BW = 20 kHz). 3D cardiac 1H images were then acquired with ungated 3D SPGR (TR/TE = 1–2/2–2.5 ms, flip angle = 10–50°, NEX = 32–128, FOV = 40×40×40 mm3, matrix = 128×128×128, BW = 100 kHz) and standard, gated, 2D segmented k-space pulse sequences). Ungated 2D or 3D 19F MRI were subsequently acquired (that matched exactly the spatial orientation and FOV of the 1H MRI) of the mouse thorax (TR/TE = 9/1 or 16.5/8.3ms, flip angle = 50° or 20–30°, NEX = 768 or 12–32, BW = 2 kHz, FOV = 40×40 mm2or 40×40×40 mm3, matrix = 128×128 or 32×32×32). The receiver bandwidths of 19F MRI ranged between 1.5–2 kHz and allowed spectral selection of the NP vs. ISO peaks. Total 19F MRI signals and maximum SNR values were estimated from the focal areas of hyper-enhancement following labeled CPC cardiac injections in mice (two post-mortem and two in vivo).
Histology
Post-mortem histological evaluation was performed in mouse hearts on the same day following cellular implantation to confirm CPC localization. In brief, the excised hearts were dehydrated and fixed (in a 4% methanol-free formaldehyde solution), processed, embedded in paraffin, and stored. Serial transverse paraffin sections were subsequently cut, processed, and imaged on a Leica bright-field optical microscope,
Image processing
19F images were processed in MATLAB (Mathworks, Natick, MA, USA) or ImageJ (NIH, Bethesda, MD, USA), and MRS in CSX (Johns Hopkins, USA) and IDL (Harris Geospatial, USA). The overlay of the 19F and 1H images was achieved using up-interpolation of 19F MRI using bicubic splines followed by merging (at an opacity of 50%) in ImageJ. Flow cytometric data processing and exporting was achieved using FlowJo (FlowJo LLC, Version 10, Ashland, OR, USA).
The field uniformity of the birdcage coil was assessed following high-order shimming (B0 homogeneity linewidths of ~30–70 Hz), based on the signal variability (coefficient of variation [CV] = SD/mean) from multiple one-dimensional (1D) profiles spanning the central regions of multiple 2D images (along both image dimensions), acquired using high-concentration TFA phantoms. SNR was estimated as the mean intensity from selected ROIs divided by the background standard deviation.
Statistical analyses
All results are reported as mean±standard deviation (SD). Two-tailed Student’s t-tests, were also used (XLSTAT, Addinsoft, New York) to determine whether labeling led to significant changes in transverse relaxation times.
Results
Highest field uniformity was achieved by the birdcage coil that yielded a coefficient of B1 variation in the central 25×32×32 mm3 region of the coil of 2.5% (Fig 1). Additionally, the butterfly yielded the highest SNR, while increased field penetration was noted when adiabatic (HS-AFP) excitation was used. Successful implementation of the adiabatic pulses is also demonstrated with 1H and 19F MRI of the multivial sensitivity phantom using the butterfly coil, indicating the ability to image 19F concentrations of aqueous TFA solutions down to approximately 1 mM, at increased depths of penetration compared to nonadiabatic imaging (Fig 1C–1E).
Normalized parametric SNR maps of the flip angle and ETL versus the normalized TR/T1 values for SPGR, RARE, and SSFP imaging, are shown in Fig 2, depicting the zones from which acquisition parameters were selected for fast imaging.
To allow a direct, experimental comparison of SPGR, RARE, and SSFP sequences, solution phantoms (75–100 mM TFA) were used to perform a direct comparison using the homogeneous birdcage coil (Fig 3). Maximum SNR values (100 mM TFA) are elicited by the fid-SSFP sequence (139±10) that outperformed both the SPGR (106±7) and RARE (48±4) sequences. Based on TFA phantom imaging, SSFP achieved the highest mean SNR (1.3- and 2.9-fold higher than SPGR and RARE) (Fig 3).
T1/T2 relaxation values for TFA, NaF, NP solutions, and labeled/FuGENE labeled CT cells were 2.73±0.06/2.36±0.16 s, 1.54±0.07/1.11±0.22 s, 0.77±0.02/0.36±0.03 s, 1.32±0.9 s (T1), and 1.36 ± 0.09 / 0.53 ± 0.07 s, respectively (Table 1). T2 values in labeled cells could not be measured primarily owing to the low 19F signal of labeled cells, and the prohibitively long measurement times using CPMG (that would impose cellular viability risks). Labeling led to significant increases in longitudinal relaxation values in CPCs, compared to NPs in media solutions (p<0.00047 (labeled), p<0.001 (FuGENE-labeled), α = 5%).
Table 1. In vitro relaxation values of fluorinated compounds.
Compound | T1 (ms) | T2 (ms) | Field strength (T) | Comment | Reference |
---|---|---|---|---|---|
NP (PCE) | 1010 | 526 | 1.5 | PFCE—Gd in lipid monolayer surrounding NP | Neubauer et al. 2008 [32] |
NP (Gd) | 238 | 63 | |||
NP (PCE) | 990 | 238 | 4.7 | ||
NP-Gd | 500 | 24 | |||
NP (PCE) | 763.3 | 71 | 11.7 | ||
NP-Gd | 625 | 11 | |||
PFCE in dendritic cells | 950±30 | 50±8 | 7 | MRSI—birdcage/solenoid 15 mm in diameter | Bonetto et al. 2012 [33] |
Saline | 1400±30 | 440±25 | 9.4 | 18 mm surface adiabatic excitation | van Heeswijk 2012 [16] |
PCE in venous blood (in vitro) | 1350±40 | 25±2 | |||
Free PFPE | 280±20 | 153±4 | 11.7 | Surface coils 9–25 mm diameter | Boehm-Sturm et al. 2011 [13] |
Labeled human neural stem cells | 380±4 | 68±3 | |||
TFA | 2728±16 | 2365±18 | 9.4 | Birdcage/Solenoid | This study |
PLGA-PFCE (solution) | 773±16 | 360±34 | |||
NP labeled CT cells | 1324±89 | - | |||
NP FuGENE-labeled CT cells | 1360±95 | 533±70 |
Based on phantom imaging and MRS, the minimum detectable NP dose was 2.5 mg/ml (0.8mM) (birdcage) (Fig 4). As before, superior SNR performance is demonstrated by echo-SSFP (compared to SPGR) in 19F imaging of the NP solutions, whereas lack of substantially improved signals is also documented for NP loading beyond 7.5 mg/ml for the studied CPCs (Fig 4E).
Fig 5 presents flow cytometry results in justification of the successful labeling protocol. Confocal microscopy images of control and labeled CT cells are shown in (Fig 5A–5F), in justification of the lower sensitivity of detection of the labeled population by confocal microscopy (and correspondingly by MRI). The estimated percentage of viable labeled cells using confocal microscopy was 10% using simple labeling, and significantly increased upon use of FuGENE to 80% or higher [30]. A significant labeling heterogeneity was also noted for CPCs (Fig 5G–5I). Additionally, MRS/MRI allowed imaging and quantification of CPC NP concentrations (fully relaxed MRS for a label loading concentration of 7.5–10 mg/ml/million cells) (Fig 5J–5L) of the order of 0.3–0.5 mM (solenoid).
Image-based quantification (SPGR images, error of ≤ 3%, 10–100 mM) (Fig 6A–6C) confirmed linear signal-concentration dependence in TFA solutions (Actual concentration [mM] = 1.11×Estimated concentration [mM]-5.5, R2 = 0.997), with detection limits for the butterfly/birdcage equal to ~0.5 and 10 mM in imaging acquisitions that spanned ~3 min). Fig 6F–6H confirmed that the cellular load detection threshold for the butterfly coil was approximately 0.5 million labeled CPCs (without FuGENE) (Table 2) using the imaging protocol. More importantly, the ability to conduct in vitro, image-based cell quantification was confirmed using TFA phantoms (Fig 6A and 6B), labeled CPC (Fig 6E) and macrophage cells (Actual concentration [mM] = 0.17×Estimated concentration [mM]+0.03, R2 = 0.99), using fast imaging and MRS. The effort was extended successfully in the post-mortem mouse using the butterfly coil.
Table 2. Comparison of acquisition parameters and cell detection limits in this and prior studies.
19F label | Cell type | Field strength (T) | Voxe size (μl) | Total imaging (min) | Cell detection limit (per voxel) | 19F Atom detection limit/voxel (×1016) | Reference |
---|---|---|---|---|---|---|---|
PLGA-PFCE | Human dendriti cells | 7 | 2–3.5 | 27 | <30,000 | - | Srinivas et al. 2010 [7] |
PLGA-PFCE | Human dendritic cells | 7 | 13 | 60 | 5000 | 3.5 | Bonetto et al. 2012 [33] |
PFCE emulsion | Human SC/progenitor | 11.7 | 27 | 4 | 6100 | 3.8 | Partlow et al. 2007 [9] |
PFCE emulsion | Murine dendritic cells | 9.4 | 0.1 | 106 | 106 | 170 | Waiczies et al. 2011 [34] |
PFCE emulsion | Murine macrophages | 9.4 | 0.2–0.4 | 19 | 200 | 0.18 | Flogel et al. 2008 [15] |
PFCE emulsion | Murine neural SCs | 9.4 | 0.3 | 5 | 140 pmol PFCE/cell |
- | Ruiz-Cabello et al. 2008 [10] |
PCE emulsion | Dendrit c cells | 9.4 | 0.1 | 70 | 105 | 100 | Waiczies et al. 2013 [35] |
PFC emulsion | Monocytes Macrophages Dendritic cells Granulocytes |
9.4 | 0.4 | 32 | - | - | Van Heeswijk et al. 2013 [18] |
PLGA-PFCE (no FuGENE) |
CDC CT |
9.4 | 2 | 3–5 | 10400 | - | This study |
The applicability of the presented methodologies and optimization strategies for in vivo imaging are justified by the fast, post-mortem imaging of injected labeled CPCs in cardiac muscle (in vivo and post-mortem) and femoral areas (post-mortem) of the C57BL/6 mouse [37] (Fig 7). Reported findings were confirmed with histology, whereby cellular injection localization was identified using bright field histological imaging (Fig 7).
From the quantification viewpoint, the in vitro measurement using 19F MRI (based on the labeled CPC reference standard) estimated the number of labeled cells as 1.26 million based on the labeled CPC reference standard (compared to the actual number of 1.3 million estimated based on Trypan Blue). Only 1 million cells were quantified from the post-mortem images. The discrepancy is attributed to injection losses (resuspension fluid loss before injection, and other injection losses).
From the post-mortem and in vivo imaging viewpoints, the total 19F signal of injected CPCs in mice (post-mortem vs. in vivo) yielded a CV of 24%, whereas the maximum SNR values were comparable (10±1.8, range = 8–11.8). The reported variability is a result of the expected discrepancies in the injected number of cells and injection losses in the four studied cases.
Discussion
Despite the technical, biological, and imaging advances and development of 19F MRI [5, 6, 9, 10, 15, 33, 38] and cellular tracking during the last decade [39, 12, 13, 24, 35, 40], no prior imaging and quantification work focused on CPCs [12, 13, 36].
While Faber and Schmid [26] have recently reported theoretical and experimental comparisons of 19F pulse imaging strategies, the approach adopted herein is mathematically rigorous and analytic, and includes reference to the explicit mathematical formulations, allowing theoretical comparison of SSFP, SPGR, and RARE sequences. Additionally, the 2D/3D comparison is also formulated mathematically, thereby justifying the efficiency and expected SNR improvements in 3D (as tabulated in Faber and Schmid [26]). Additionally, there are important practical implications relevant to the translatability and implementation of the bSSFP sequence for cardiac 19F MRI that have not been addressed previously as they pertain to artifacts and SNR performance for exogenously administered SCs in the in vivo murine beating heart, while the explicit mathematical formulation for the signal of the SSFP sequence (fid vs. echo) is lacking (useful and relevant for quantification). Furthermore, there have been no prior direct, experimental comparisons of image SNR under controlled phantom conditions for SPGR, SSFP, and RARE. Prior publications on the performance of SPGR imaging in 19F MRI, including the recent work by Faber and Schmid [26], have shown merit only for Ernst angle imaging. Nevertheless, in this work, the inefficient aspects of (slow) Ernst angle and the beneficial aspects of fast SC imaging are emphasized using SPGR.
Technical development of cardiac 19F MRI using exogenously administered SCs has been lacking, while optimizations of MR imaging acquisitions in association with SCs have been limited [26].
More importantly, imaging times for adopted methodologies in most applications in vivo were excessively long (often > 60 min) [36] due to the low cellular label concentration and the choice of the imaging sequences, often becoming prohibitive for translational work. Interestingly, recent implementations of compressed sensing in 19F MRI [41, 42] have led to reduced imaging times, albeit limited by the low SNR of multinuclear in vivo studies.
We have presented the first in vivo cardiac 19F MRI data from labeled CPCs injected in the murine myocardium. We have demonstrated feasibility and reproducibility of fast (of the order of a few minutes), in vivo cardiac 19F-MRI of exogenously administered CPCs in the murine heart using SPGR/SSFP sequences.
The imaging protocol can be easily and readily adopted for any other labeling agent in preclinical work, and has potential for use in translational work. Additionally, the presented theoretical and experimental schemes are label-independent and can be readily applied to any other fluorinated compounds.
In contrast to imaging of nuclei other-than-protons (e.g., 23Na MRI), where the intrinsically abundant sodium nucleus exhibits fast longitudinal relaxation (and where the optimal SNR is elicited at the Ernst angle favoring TR≈T1), 19F MRI depends on exogenously administered NPs, emulsions, or labeled cells, with relatively long T1 values and small T2/T1 ratios. Despite the expected maximal SNR increases for SPGR sequences at long TRs, it is recommended that fast acquisitions are used in conjunction with averaging and with an appropriate choice of the flip angle, based on theoretical evaluations. This study has also assessed the cellular load detection thresholds (0.5 mM for the butterfly versus 10 mM for the birdcage coil) for 19F NP labels, based on fluorinated phantom solution comparisons and fast acquisitions of the order of a few minutes at voxel resolutions of ~2 μl or less.
Despite recent efforts to chemically modify the T1/T2 characteristics of labeled NPs [43] to speed up acquisitions, such approaches require complex chemical syntheses. The PLGA-PFC NPs used in this study have elicited T1 and T2 values that are in agreement with prior reports [36], and a T1-effect is demonstrated post-cellular loading. The inability to quantify T2 upon cellular loading is primarily attributed to the low labeling efficiency (and hence the low 19F signal), and to the intracellular endosomal/vesicular packaging of the NP label leading to a relatively short transverse relaxation times (of the order of a few ms) [39, 36]. Correspondingly, prolonged spectroscopic acquisition times will invariably impose additional issues in terms of the cellular oxygenation status (oxygen tension, leading to hypoxia [41]), and altered viability, compromising T2 estimates. Given the limited in vivo/post-mortem visibility of labeled cells (without FuGENE), and the reported dependence of the T2 values of PFCE-NPs on i) temperature [44, 45], ii) label concentration [46], and iii) cell type, and the iv) extremely low 19F signal of labeled cells, increased complexity and limited usefulness and consistency, are anticipated from the measurement of these values.
Our reported relaxation values in NP solutions (T1 and T2) are smaller in value than those recently reported by Colotti et al. [46] at 24°C at 3T. This difference, however, is expected in view of the B0-field relaxation dependence trends reported by de Vries et al. for 19F emulsions [44].
Our successful labeling protocol was confirmed using flow cytometric and confocal microscopy validations. Based on validation studies presented herein, no additional benefits are expected by increasing the label concentration beyond 7.5–10 mg/ml per million cells, given the constancy of elicited signal responses in NP solutions using both the SPGR and SSFP acquisitions.
The minimum cellular detectable load (no FuGENE) was determined to be approximately 500k cardiac stem cells (or equivalently~10k cells per voxel) in fast acquisitions (~3–5 min) using the butterfly coil. This finding can be justified by the inefficient process of cellular label uptake in these cardiac stem cells in association with their much smaller cellular size (~30 μm3 isotropic) compared to dendritic cells or macrophages. Evidently, FuGENE significantly decreases this detection limit, thereby achieving 19F cardiac MRI [30].
Regarding quantification, we have demonstrated direct spectroscopy, and direct image-based quantification of absolute 19F concentration in TFA solutions, in labeled CPCs and macrophage cells, and injected cells post-mortem, with responses that scale linearly with increased fluorinated label or fluorine concentrations ex vivo. The ability to conduct both 19F MRI/MRS provides an easy/efficient methodological pathway to study focal myocardial disease.
However, unlike prior quantification studies [6, 47, 48], a major limitation of the quantitative capacity of in vivo 19F MRI is the fact that it cannot be applied to CPCs given their low and heterogeneous label uptake at insufficient levels for their MRI detection. To overcome such a limitation, we have validated an in vitro quantification scheme (in combination with adiabatic excitation) that can be adopted to in vivo applications, where a secondary reference phantom containing a known number of labeled CPCs can be used.
The present study is associated with various limitations, including the focus on two particular CPC types that consist of a heterogeneous cell mixture, and their low label uptake that ultimately hinders T2 measurements. Given the stringent spatial requirements of the high-field bore system, temperature and oxygen tension effects cannot be easily monitored during relaxation measurements, or following cellular injection. Furthermore, our study has focused on PFCE with a single resonant peak, compared to other fluorinated compounds (e.g., PFOB) that exhibit complex, multi-peak responses.
Another major limitation is the lack of quantitative accuracy in vivo primarily owing to the stringent time limitations for additional data acquisitions for achieving adiabaticity and B1 and motion corrections. Even still, an additional confounding factor that may limit quantification is the diminished viability of injected cells as a result of the hypoxic environment in which they are injected [1]. Correspondingly, evoked 19F signal hyper-enhancements that may be attributed to viable cells, or released NPs from lysed cells (immediately following injection, or in tracking studies), ought to be interpreted with care. Such present the primary limitations in translating this work to a true experimental setting in humans.
Furthermore, the spatio-temporal effects of motion on 19F cardiac MRI may need to be addressed in more detail. However, preliminary tests with the implemented protocols indicate lack of spatial discrepancies between ungated and gated 1H MRI scans (voxel size = 0.2 μl) in normal mice. Correspondingly, possible motion effects between ungated and gated 19F MRI acquisitions are expected to be minimal, considering the large voxel (voxel volume>2 μl), and the spatio-temporal averaging over the acquisition time intervals. Although gated cardiac 19F scans are possible, they are prohibitively long (exceeding at least 30 min, thereby imposing beat-to-beat variability issues) and yield inadequate image SNR responses that would disallow NP detectability and localization.
Supporting information
Acknowledgments
The lead author is most grateful to Professor Craig Lygate for his support, advice, and guidance throughout the study. We are also thankful to Dr. J. Brown, Mr. A. Hale, Dr. J. Patel, Dr. A. Al Haj Zen, and A. Worth, for their support with the cell cultures, confocal microscopy, and flow cytometry studies. We are also most appreciative to Drs. A. Yavari and S. Ghaffari for providing us with FuGENE for the initial experimentation studies.
We also thank Dr. V. Clark, Mrs. H. Gray, Mrs. L. Stork, and Mrs. A. Vernet, and Dr. J. Beglov, for their help with the animal studies. Dr. D. McClymont is thanked for checking the mathematical computational formulations, and Mr. J. Prentice is acknowledged for his help with the design and construction of the frame for the birdcage coil. We are also most appreciative to Professor C. Hayes and to Dr. L. Cochlin for their advice and guidance with the construction of the birdcage coil.
Data Availability
All relevant data are within the paper and its Supporting Information files.
Funding Statement
The project has received funding from the European Union’s Horizon 2020 research and innovation programme under the Marie Sklodowska-Curie grant agreement No. 652986 to Christakis Constantinides, European Research Council grants ERC-2013-StG-336454 and ERC-PoC-713524, Radboud Universiteit SWT grant 14716 (MS), Wellcome Trust Core Award (grant 090532/Z/09/Z), and British Heart Foundation (BHF) grant FS/11/50/29038 (JS). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1.Feyen DA, Gaetani R, Doevendans PA, Sluijter JP. Stem cell-based therapy: Improving myocardial cell delivery. Advanced Drug Delivery Reviews 2016; 106(PtA):104–115. doi: 10.1016/j.addr.2016.04.023 [DOI] [PubMed] [Google Scholar]
- 2.Michalet X, Pinaud FF, Bentolila LA, Tsay JM, Doose S, Li JJ, et al. Quantum dots for live cells, in vivo imaging, and diagnostics. Science 2005; 307(5709):538–544. doi: 10.1126/science.1104274 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Chen IY, Wu JC. Cardiovascular molecular imaging: focus on clinical translation. Circulation. 2011;123(4):425–43. doi: 10.1161/CIRCULATIONAHA.109.916338 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Stuckey DJ, Carr CA, Martin-Rendon E, Tyler DJ, Willmott C, Cassidy PJ, et al. Iron particles for noninvasive monitoring of bone marrow stromal cell engraftment into, and isolation of viable engrafted donor cells from, the heart. Stem Cells 2006; 24(8):1968–75. doi: 10.1634/stemcells.2006-0074 [DOI] [PubMed] [Google Scholar]
- 5.Ahrens ET, Flores R, Xu H, Morel PA. In vivo imaging platform for tracking immunotherapeutic cells. Nature Biotechnology 2005; 23(8):983–987. doi: 10.1038/nbt1121 [DOI] [PubMed] [Google Scholar]
- 6.Srinivas M, Morel PA, Ernst LA, Laidlaw DH, Ahrens ET. Fluorine-19 MRI for visualization and quantification of cell migration in a diabetes model. Magnetic Resonance in Medicine 2007; 58:725–734. doi: 10.1002/mrm.21352 [DOI] [PubMed] [Google Scholar]
- 7.Srinivas M, Cruz LJ, Boneto F, Heerschap A, Figdor CG, de Vries IJM. Customizable, multi-functional fluorocarbon nanoparticles for quantitative in vivo imaging using 19F MRI and optical imaging. Biomaterials 2010; 31:7070–7077. doi: 10.1016/j.biomaterials.2010.05.069 [DOI] [PubMed] [Google Scholar]
- 8.Gomes RSM, das Neves RP, Cochlin L, Lima A, Carvalho R, Korpisalo P, et al. Efficient pro-survival/angiogenic miRNA delivery by an MRI-detectable nanomaterial. ACS Nano 2013; 7(4):3362–3372. doi: 10.1021/nn400171w [DOI] [PubMed] [Google Scholar]
- 9.Partlow KC, Chen J, Brant JA, Neubauer AM, Meyerrose TE, Creer MH, et al. 19F magnetic resonance imaging for stem/progenitor cell tracking with multiple unique perfluorocarbon nanobeacons. FASEB Journal 2007; 21(8):1647–54. doi: 10.1096/fj.06-6505com [DOI] [PubMed] [Google Scholar]
- 10.Ruiz-Cabello J, Walczak P, Kedziorek DA, Chacko VP, Schmieder AH, Wickline SA, et al. In vivo "hot spot" MR imaging of neural stem cells using fluorinated nanoparticles. Magnetic Resonance in Medicine 2008; 60(6):1506–11. doi: 10.1002/mrm.21783 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Jacoby C, Temme S, Mayenfels F, Benoit N, Krafft MP, Schubert R, et al. Probing different perfluorocarbons for in vivo inflammation imaging by 19F MRI: image reconstruction, biological half-lives and sensitivity. NMR in Biomedicine 2014; 27:261–271. doi: 10.1002/nbm.3059 [DOI] [PubMed] [Google Scholar]
- 12.Ahrens ET, Helfer BM, O’Hanlon CF, Schirda C. Clinical cell therapy imaging using a perfluorocarbon tracer and fluorine-19 MRI. Magnetic Resonance in Medicine 2014; 72:1696–1701. doi: 10.1002/mrm.25454 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Boehm-Sturm P, Mengler L, Wecker S, Hoehn M, Kallur T. In vivo tracking of human neural stem cells with 19F magnetic resonance imaging. PLoS One 2011; 6(12):e29040 doi: 10.1371/journal.pone.0029040 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Aday S, Paiva J, Sousa S, Gomes RSM, Pedreiro S, So PW, et al. Inflammatory modulation of stem cells by Magnetic Resonance Imaging (MRI)-detectable nanoparticles. RSC Advances 2014; 4:31706–31709. [Google Scholar]
- 15.Flogel U, Ding Z, Hardung H, Jander S, Reichmann G, Jacoby C, et al. In vivo monitoring of inflammation after cardiac and cerebral ischemia by fluorine magnetic resonance imaging. Circulation 2008; 118(2):140–8. doi: 10.1161/CIRCULATIONAHA.107.737890 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.van Heeswijk RB, Pilloud Y, Flögel U, Schwitter J, Stuber M. Fluorine-19 magnetic resonance angiography of the mouse. PLoS One 2012; 7(7): e432236. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Ye Y-X, Basse-Lusebrink TC, Arias-Loza P-A, Kocoski V, Kampf T, Gan Q, et al. Monitoring of monocyte recruitment in reperfused myocardial infarction with intramyocardial hemorrhage and microvascular obstruction by combined fluorine 19 and proton cardiac magnetic resonance imaging. Circulation 2013; 128(17):1878–1888. doi: 10.1161/CIRCULATIONAHA.113.000731 [DOI] [PubMed] [Google Scholar]
- 18.van Heeswijk RB, De Blois J, Kania G, Gonzalez C, Blyszczuk P, Stuber M, et al. Selective in vivo visualization of immune-cell infiltration in a mouse model of autoimmune myocarditis by fluorine-19 cardiac magnetic resonance. Circulation Cardiovascular Imaging 2013; 6(2):277–284. doi: 10.1161/CIRCIMAGING.112.000125 [DOI] [PubMed] [Google Scholar]
- 19.Jacoby C, Borg N, Heusch P, Sauter M, Bonner F, Kandolf R, et al. Visualization of immune cell inflitration in experimental viral myocarditis by 19F MRI in vivo. Magnetic Resonance Materials Physics 2014; 27(1):101–106. [DOI] [PubMed] [Google Scholar]
- 20.Chalmers KH, Kenwright AM, Parker D, Blamire AM. 19F-lanthanide complexes with increased sensitivity for 19F-MRI: optimization of the MR acquisition. Magnetic Resonance in Medicine 2011; 66(4):931–936. doi: 10.1002/mrm.22881 [DOI] [PubMed] [Google Scholar]
- 21.Schmid F, Holtke C, Parker D, Faber C. Boosting 19F MRI—SNR efficient detection of paramagnetic contrast agents using ultrafast sequences. Magnetic Resonance in Medicine 2013; 69(4):1056–1062. doi: 10.1002/mrm.24341 [DOI] [PubMed] [Google Scholar]
- 22.Goette MJ, Keupp J, Rahmer J, Lanza GM, Wickline SA, Caruthers SH. Balanced UTE-SSFP for 19F MR imaging of complex spectra. Magnetic Resonance in Medicine 2015; 74:537–543. doi: 10.1002/mrm.25437 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Giraudeau C, Flament J, Marty B, Boumezbeur F, Meriaux, Robic C, et al. A new paradigm for high-sensitivity 19F magnetic resonance imaging of perfluorooctylbromide. Magnetic Resonance in Medicine 2010; 63:1119–1124. doi: 10.1002/mrm.22269 [DOI] [PubMed] [Google Scholar]
- 24.Ribot EJ, Gaudet JM, Chen Y, Gilbert KM, Foster PJ. In vivo MR detection of fluorine-labeled human MSC using the bSSFP sequence. International Journal of Nanomedicine 2014; 9:1731–1739. doi: 10.2147/IJN.S59127 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Mastropietro A, De Bernadi E, Breschi GL, Zucca I, Cametti M, Soffientini CD, et al. Optimization of rapid acquisition with relaxation enhancement (RARE) pulse sequence parameters for 19F MRI studies. Journal of Magnetic Resonance Imaging 2014; 40:162–170. [DOI] [PubMed] [Google Scholar]
- 26.Faber C, Schmid F. Pulse sequence considerations and schemes In: Flogel U, Ahrens ET, editors. Fluorine Magnetic Resonance Imaging. Singapore: Pan Stanford Publishing: 2016; 1–28. [Google Scholar]
- 27.Chimenti I, Smith RR, Li TS, Gerstenblith G, Messina E, Giacomello A, et al. Relative roles of direct regeneration versus paracrine effects of human cardiosphere-derived cells transplanted into infarcted mice. Circulation Research 2010; 106(5):971–980. http://circres.ahajournals.org/content/106/5/971. doi: 10.1161/CIRCRESAHA.109.210682 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Tan SC, Gomes RS, Yeoh KK, Perbellini F, Malandraki-Miller S, Ambrose L, et al. Preconditioning of cardiosphere-derived cells with hypoxia or prolyl-4-hydroxylase inhibitors increases stemness and decreases reliance on oxidative metabolism. Cell Transplantation 2016; 25(1):35–53. doi: 10.3727/096368915X687697 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Malandraki-Miller S., Tyser R., Riley P P., Carr C.A.. Comparative characterization of cardiac atrial progenitor cell populations for use in cell therapy. Heart 2014; 100:A14. [Google Scholar]
- 30.Constantinides C, McNeill E, Benson M, Urruela RS, Padilla S, Malandraki-Miller S, et al. Improvements in the cellular uptake of perfluorocarbon nanoparticles and 19F MRS/MRI detectability using the transfection agent FuGENE. ISMRM Workshop on Molecular and Cellular MRI: Focus on Integration: Amsterdam, Netherlands, June 2016.
- 31.Silver MS, Joseph RI, Hoult DI. Highly selective π/2 and π pulse generation. Journal of Magnetic Resonance 1984; 59:347–351. [Google Scholar]
- 32.Neubauer AM, Myerson J, Caruthers SD, Hockett FD, Winter PM, Chen J, et al. Gadolinium-modulated 19F signals from perfluorocarbon nanoparticles as a new strategy for molecular imaging. Magnetic Resonance in Medicine 2008; 60:1066–1072. doi: 10.1002/mrm.21750 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Bonetto F, Srinivas M, Weigelin B, Cruz LJ, Heerschap A, Friedl P, et al. A large-scale 19F MRI-based cell migration assay to optimize cell therapy. NMR in Biomedicine 2012; 25:1095–1103. doi: 10.1002/nbm.2774 [DOI] [PubMed] [Google Scholar]
- 34.Waiczies H, Lepore S, Janitzek N, Hagen U, Seifert F, Ittermann B, et al. Perfluorocarbon particle size influences magnetic resonance signal and immunological properties of dendritic cells. PLoS One 2011; 6(7):e21981 doi: 10.1371/journal.pone.0021981 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Waiczies H, Guenther M, Skowoski J, Lepore S, Pohlmann A, Niendorf T, et al. Monitoring dendritic cell migration using 19F/1H magnetic resonance imaging. Journal of Visualized Experiments 2013; 20(73):e50251 doi: 10.3791/50251 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Srinivas M, Philipp Boehm-Sturm P, Figdor CG, de Vries IJ, Hoehn M. Labeling cells for in vivo tracking using 19F MRI. Biomaterials 2012; 33:8830–8840. doi: 10.1016/j.biomaterials.2012.08.048 [DOI] [PubMed] [Google Scholar]
- 37.Constantinides C, Maguire ML, Malandraki-Miller S, Swider E, Srinivas M, Carr CA, et al. Fast, Quantitative 19F MRI: Optimized Imaging Strategies. ESMRMB 2016, Vienna, Austria, October 2016, #122.
- 38.Hanicke W, Vogel HU. An analytical solution for the SSFP signal in MRI. Magnetic Resonance in Medicine 2003; 49:771–775. doi: 10.1002/mrm.10410 [DOI] [PubMed] [Google Scholar]
- 39.Hoehn M, Kustermann E, Blunk H, Wiedermannm D, Trapp T, Wecker S, et al. Monitoring of implanted stem cell migration in vivo: a highly resolved in vivo magnetic resonance imaging investigation of experimental stroke in rat. Proceedings of the National Academy of Sciences 2002; 99(25):16267–16272. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Gaudet JM, Ribot EJ, Chen Y, Gilbert KM, Foster PJ. Tracking the fate of stem cell implants with fluorine-19 MRI. PLoS One 2015; 10(3):e0118544 doi: 10.1371/journal.pone.0118544 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Kampf T, Fischer A, Basse-Lusebrink TC, Ladewig G, Breuer F, Stoll G, et al. Application of compressed sensing to in vivo 3D 19F CSI. Journal of Magnetic Resonance 2010; 207:262–273. doi: 10.1016/j.jmr.2010.09.006 [DOI] [PubMed] [Google Scholar]
- 42.Zhong J, Mills PH, Hitchens K, Ahrens ET. Accelerated fluorine-19 MRI cell tracking using compressed sensing. Magnetic Resonance in Medicine 2013; 69:1683–1690. doi: 10.1002/mrm.24414 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Kislukhin AA, Xu H, Adams SR, Narshinh KH, Tsien RY, Ahrens ET. Paramagnetic fluorinated nanoemulsions for sensitive cellular fluorine-19 magnetic resonance imaging. Nature Materials 2016; 15:662–669. doi: 10.1038/nmat4585 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.de Vries A, Moonen R, Yildirim M, Langereis S, Lamerichs R, Pikkemaat JA, et al. Relaxometric studies of gadolinium-functionalized perfluorocarbon nanoparticles for MR imaging. Contrast Media Molecular Imaging 2014; 9:83–91. doi: 10.1002/cmmi.1541 [DOI] [PubMed] [Google Scholar]
- 45.Kadayakkara DK, Damodaran K, Hitchens TK, Bulte JWM.19F spin-lattice relaxation of perfluoropolyethers: Dependence on temperature and magnetic field strength (7.0–14.1 T). Journal of Magnetic Resonance 2014; 242:18–22. doi: 10.1016/j.jmr.2014.01.014 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Colotti R, Bastiaansen JAM, Wilson A, Flogel U, Gonzalez C, Schwitter J, et al. Characterization of perfluorocarbon relaxation times and their influence on the optimization of fluorine-19 MRI at 3 Tesla. Magnetic Resonance in Medicine 2016. doi: 10.1002/mrm.26317 [DOI] [PubMed] [Google Scholar]
- 47.Srinivas M, Heerschap A, Ahrens ET, Figdor CG, de Vries IJM. 19F MRI for quantitative in vivo cell tracking. Trends in Biotechnology 2010; 28(7):363–370. doi: 10.1016/j.tibtech.2010.04.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Lee D, Marro K, Shankland E, Mathis M. Quantitative 19F imaging using inductively coupled reference signal injection. Magnetic Resonance in Medicine 2010; 63:570–573. doi: 10.1002/mrm.22298 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All relevant data are within the paper and its Supporting Information files.